Test-retest analysis of cerebral oxygen extraction estimates in healthy volunteers: comparison of methods based on quantitative susceptibility mapping and dynamic susceptibility contrast magnetic resonance imaging

Ronnie Wirestam a,, Anna Lundberg a, Arthur Chakwizira a, Danielle van Westen b,c, Linda Knutsson a,d,e, Emelie Lind a
PMCID: PMC9801129  PMID: 36590544

Abstract

Background

Estimation of the oxygen extraction fraction (OEF) by quantitative susceptibility mapping (QSM) magnetic resonance imaging (MRI) is promising but requires systematic evaluation. Extraction of OEF-related information from the tissue residue function in dynamic susceptibility contrast MRI (DSC-MRI) has also been proposed. In this study, whole-brain OEF repeatability was investigated, as well as the relationships between QSM-based OEF and DSC-MRI-based parameters, i.e., mean transit time (MTT) and an oxygen extraction index, referred to as apparent OEF (AOEF).

Method

Test-retest data were obtained from 20 healthy volunteers at 3 T. QSM maps were reconstructed from 3D gradient-echo MRI phase data, using morphology-enabled dipole inversion. DSC-MRI was accomplished using gradient-echo MRI at a temporal resolution of 1.24 s.

Results

The whole-brain QSM-based OEF was (40.4±4.8) % and, in combination with a previously published cerebral blood flow (CBF) estimate, this corresponds to a cerebral metabolic rate of oxygen level of CMRO2 = 3.36 ml O2/min/100 g. The intra-class correlation coefficient [ICC(2,1)] for OEF test-retest data was 0.73. The MTT-versus-OEF and AOEF-versus-OEF relationships showed correlation coefficients of 0.61 (p = 0.004) and 0.52 (p = 0.019), respectively.

Discussion

QSM-based OEF showed a convincing absolute level and good test-retest results in terms of the ICC. Moderate to good correlations between QSM-based OEF and DSC-MRI-based parameters were observed. The present results constitute an indicator of the level of robustness that can be achieved without applying extraordinary resources in terms of MRI equipment, imaging protocol, QSM reconstruction, and OEF analysis.

Keywords: Magnetic resonance imaging, Contrast media, Cerebrovascular circulation, Hemodynamics, Oxygen consumption, Magnetometry


Magnetic resonance imaging; Contrast media; Cerebrovascular circulation; Hemodynamics; Oxygen consumption; Magnetometry.

1. Introduction

Magnetic resonance imaging (MRI) methods for estimation of cerebral blood flow (CBF) as well as the venous oxygen saturation level, closely related to the oxygen extraction fraction (OEF), are well established, offering the opportunity to assess also the cerebral metabolic rate of oxygen (CMRO2). MRI susceptometry, based on phase information, is a common approach for quantification of the venous oxygen saturation levels [1, 2], and a more recent expansion of this technique is based on quantitative susceptibility mapping (QSM) [3, 4, 5]. For example, global or whole brain OEF estimation using QSM has recently received substantial attention in the characterization of multiple sclerosis (MS) lesions [6], and promising attempts to achieve regional evaluation or visualization have also been introduced [7, 8]. The use of QSM for OEF estimation is clearly well documented, but QSM reconstruction is a delicate mathematical procedure known to be sensitive to noise, choice of algorithm, settings of regularization parameters, choice of brain mask, etc., and systematic evaluation of acquired OEF estimates is thus warranted.

With regard to MRI-based CBF assessment, the most common method in clinical environments is dynamic susceptibility contrast MRI (DSC-MRI) [9]. DSC-MRI is generally a robust and time efficient technique, but a number of methodological impediments that hamper CBF quantification in absolute terms need to be considered [10]. DSC-MRI data processing includes retrieval of the tissue residue function, by deconvolution of the tissue contrast agent (CA) concentration curve with the arterial input function (AIF). This implies that DSC-MRI is, in principle, able to provide information related to the blood transit time distribution and, hypothetically, also oxygen extraction occurring during the passage from the arterial inlet to the venous outlet of the local capillary network. An intriguing theoretical framework for assessment of an OEF-related metric (referred to as oxygen extraction capacity), based on DSC-MRI capillary transit time distributions, has been proposed by Jespersen & Østergaard and Mouridsen et al. [11, 12].

However, a crucial prerequisite for accurate estimation of the shape of the tissue residue function is a reliable deconvolution algorithm. The true residue function is, by definition, characterized by non-negative values and strict decrease, while the most commonly used deconvolution algorithms, based on singular value decomposition [9, 13], tend to return residue function estimates that suffer from severe oscillations [10]. Hence, improved deconvolution algorithms, with the ability to return physiologically plausible residue functions in DSC-MRI, have attracted recent attention [e.g. 14, 15, 16]. The use of a control point interpolation (CPI) method, as introduced by Mehndiratta et al. [15], is an attractive approach to impose the desired constraints on the output, while at the same time avoiding the need for a parametric or model-dependent algorithm. However, in CPI, the residue function must pass through all control points. Hence, as a further development, deconvolution based on Bézier curves was recently introduced [16]. The main advantage with such an algorithm is that a Bézier curve needs only be bounded by the convex hull of its control points, which significantly reduces the number of optimization parameters.

The tissue residue function or transit time distribution, obtained by DSC-MRI, is not expected to carry all pieces of information that are required to quantify true OEF, but the shape of the tissue residue function may still provide complementary understanding of tissue oxygenation and viability. Hence, attempts to compare such an OEF-related metric with an established OEF method are indeed of interest. In the present study, both conventional mean transit time (MTT) and an OEF-related metric, below referred to as apparent OEF (AOEF), retrieved from DSC-MRI data, were directly compared with the corresponding QSM-based whole-brain OEF estimates in a group of healthy volunteers. Additionally, test-retest analyses of the extracted parameters were carried out.

2. Material and methods

2.1. Theory

2.1.1. QSM-based OEF estimation

The underlying susceptibility distribution χ(r) of the tissue is related to a corresponding phase shift Δ(r) according to Eq. (1) [17]:

Δr=γ·B0·TE·dχr, (1)

where r is the voxel position given in spherical coordinates, γ is the gyromagnetic ratio, B0 is the magnetic flux density of the main magnetic field, TE is the echo time of a gradient-echo (GRE) sequence, d=3cos2ϑ14πr3 is the z-component of the unit dipole kernel (ϑ is the observation angle relative the main magnetic field) and the symbol “⊗” denotes convolution. The mathematical inversion required to obtain the spatial susceptibility distribution from measured phase is referred to as QSM [18, 19]. Assuming that the arterial oxygen saturation is 100%, OEF can be expressed as OEF=1Y, where Y is the fractional venous oxygen saturation level. Eq. (2) shows the relationship between OEF and the susceptibility difference Δχ between the surrounding tissue (here assumed to represent arterial blood) and the venous vessel [1]:

OEF=ΔχΔχdo·Hct, (2)

where Δχdo=0.193·4π ppm (in SI units) is the susceptibility difference per unit hematocrit between fully deoxygenated and fully oxygenated blood [20]. Hct is the fractional hematocrit, in this study assumed to be 0.42 for males and 0.40 for females [21].

The cerebral metabolic rate of oxygen is given by:

CMRO2=OEF·CBF·Ca=OEF·CBF·MCHC·Hct·c (3)

Ca = [Hb]⋅c = MCHC⋅Hct⋅c is the oxygen concentration of blood, where [Hb] is the hemoglobin concentration in the blood, c is the Hb-carrying capacity of oxygen and MCHC is the mean corpuscular Hb concentration in red blood cells. In the calculations presented here, MCHC = 34 g/dl [21] and c = 1.368 ml/g [22] were used.

2.1.2. Dynamic susceptibility contrast MRI

DSC-MRI [9, 10, 23] is based on monitoring of the CA bolus passage in tissue and artery, after intravenous CA injection in a peripheral vein. Measured contrast agent concentration is given by Cm(t) = -ln[S(t)/S0]/(TE r2∗), where r2∗ is the T2∗ relaxivity, t is time and S0 is the baseline signal. Gadolinium-based CAs are plasma tracers, so a factor kH is applied, accounting for brain density and hematocrit in large and small blood vessels, i.e., tissue concentration is Ct(t) = kH⋅Cm(t). The general tracer-kinetic relationship in Eq. (4) is employed for quantification.

Ct(t)=CBF[R(t)Cart(t)], (4)

where Ct(t) is the tissue CA concentration and Cart(t) is the arterial input function (AIF). CBF⋅R(t), obtained by deconvolution, is the tissue impulse response function and R(t) is the tissue residue function. The transit time distribution is given by h(t) = -dR(t)/dt and the MTT was obtained according to Eq. (5):

MTT=0R(t)dt (5)

An OEF-related DSC-MRI metric can be obtained according to Jespersen & ​Østergaard [11] and Mouridsen et al. [12], here referred to as apparent OEF (AOEF):

AOEF=0hτQτdτ, (6)

where h(τ) is the capillary transit time distribution and Q(τ) is oxygen extraction along a single capillary with transit time τ. The AOEF is a sum of the oxygen extraction in a single capillary over all transit times, weighted by the distribution of transit times over all capillaries. Determining the dependence of the single-capillary oxygen extraction Q on transit time τ requires a model, and here we follow the approach presented by Jespersen & ​Østergaard [11], where the capillary is treated as a three-compartment system comprising hemoglobin, oxygen in plasma and oxygen in tissue. The transport of oxygen across the capillary membrane is modelled as an exchange process described by a fixed rate constant k. Variation of oxygen concentration (C) along the capillary with a transit time τ is described by the differential equation in Eq. (7):

dCdx=kταHP50CBC1/hCtissue (7)

where x is the normalized distance along the capillary, αH=3.1105 (mm Hg)−1 (Henry's constant), P50=26 mm Hg (oxygen pressure required for half saturation), B=0.1943 ml/ml (maximum amount of oxygen bound to hemoglobin), h=2.8 (Hill coefficient) and Ctissue=25 mm Hg (tissue oxygen concentration) [11, 12]. From the solution of Eq. (7) above, Q is computed according to Q = 1−C(1)/C(0) as a function of , enabling estimation of AOEF.

2.1.3. Subjects and measurements

All measurements and associated postprocessing were approved by the local ethics committee (The Regional Ethical Review Board in Lund), and written informed consent was obtained from each subject. In light of ethically motivated restraints related to repeated injections of a gadolinium contrast agent in healthy volunteers, the current study relies on an independent re-analysis of image data acquired according to a previously reported data collection protocol [24, 25]. The current study is based on a separate scientific hypothesis and employs post-processing approaches and analyses that were not available at the time of the previous publications. Twenty healthy volunteers (10 females and 10 males in two age groups of equal size, 25–34 years and 51–84 years) were examined according to a test-retest protocol, with a time interval of 7–20 days between the two investigations (below referred to as visit 1 and visit 2). The volunteers underwent a neurological physical examination, including basic tests to verify their cognitive ability. The medical history of the volunteers was reviewed to exclude the risk that any medication could have altered their blood flow. Morphological images acquired before and after CA administration were evaluated to rule out the presence of any pathology.

Measurements were performed using a 3 T MRI unit (Philips Achieva, Philips Healthcare, Best, The Netherlands), and the Smart-Exam feature [26] was used for the planning of slice orientation and field of view (FOV). Gadolinium contrast agent (0.1 mmol/kg body weight, 5 ml/s, Dotarem, Guerbet, Paris, France) was injected, followed by a 20 ml saline flush at the same injection rate. DSC-MRI was carried out using single-shot GRE echo-planar imaging (EPI) with temporal resolution 1.24 s, FOV 220 × 220 mm2, image matrix 128 × 128, slice thickness 5 mm, 20 slices, flip angle 60°, TE 29 ms, and SENSE factor 2.5. The DSC-MRI experiment was preceded by a pre-bolus administration of a low dose of contrast agent, not further considered in the present study. The phase maps used for QSM-based OEF estimation were acquired using 3D GRE imaging with flow compensation. Magnitude and phase images were collected for 50 axial slices orthogonal to the external magnetic field, with spatial resolution 0.98 × 0.98 × 1.15 mm3, field of view (FOV) = 220 × 220 mm2, repetition time (TR) = 45 ms, TE = 20 ms, flip angle (FA) = 20°, bandwidth = 218 Hz/pixel.

2.2. Image processing and data analysis

2.2.1. QSM-based OEF

QSM maps were reconstructed using the morphology-enabled dipole inversion (MEDI) toolbox (Cornell University, https://pre.weill.cornell.edu/mri/pages/qsm.html) [17, 27, 28, 29]. Spatial phase unwrapping was performed using a region growing algorithm [30] followed by background field removal using a projection onto dipole field (PDF) method [31]. The regularization parameter λ was set to 800. Voxels representing venous blood in the superior sagittal sinus (SSS) were selected using a previously published procedure [32]. In brief, a slab of interest (SOI) consisting of five consecutive slices was first identified manually in the QSM volume. In each SOI, smaller volumes of interest (VOIs), including the vessel of interest, were selected. The SSS was then outlined by segmentation of the VOI venous voxels using a venous mask, defining SSS values as all susceptibility values larger than the mean + 2 standard deviations (mean + 2SDs) of the corresponding SOI values. Finally, the values of SSS voxels assumed not to be affected by partial volume effects (PVEs) were obtained by averaging of the 20% highest values within the SSS. In the Δχ calculation, according to Eq. (2), the venous blood susceptibility was represented by this mean PVE-free SSS value, while the susceptibility value of the surrounding tissue was given by the mean value of the SOI, after exclusion of SSS voxels.

2.2.2. MTT and AOEF obtained from DSC-MRI data

Tissue residue functions R(t) were retrieved according to Eq. (4) by pixelwise deconvolution of the tissue CA concentration curves with an AIF. Subject-specific global AIFs were measured in middle cerebral artery branches in the Sylvian fissure region. The employed deconvolution algorithm was based on Bézier curves including a compensation for time delays between the AIF and the tissue concentration curve [16]. The Bézier deconvolution approach was designed to return strictly decreasing residue curves, i.e., showing non-negative values without oscillations. MTT and AOEF were calculated according to Eq. 5 and Eq. 6, respectively. With regard to the rate constant k, one DSC-MRI dataset, originating from the volunteer that showed median QSM-based OEF level, was selected as a reference, and, similarly to previous approaches [12], k was set to return an oxygen extraction capacity value of 30% in a local white matter region in this dataset. This procedure yielded k = 25 s−1, and, in order to be able to compare AOEF estimates among subjects, this value of k was used in the analysis of all DSC-MRI datasets in this study. In our implementation, no attempts were made to interpret the AOEF values as absolute levels of oxygen extraction, i.e., the AOEF parameter should be understood as an OEF-related index in arbitrary units. In order to match the whole-brain OEF estimates obtained from QSM data, MTT and AOEF values were calculated as the mean value of all non-zero pixel values in the brain volume.

2.2.3. Statistics

Repeatability was assessed by comparing results from visit 1 and visit 2 in Bland-Altman plots [33] and, for QSM-based OEF, by calculating intra-class correlation coefficients ICC(2,1) [two-way, absolute agreement, same rater, single measurement] and ICC(3,1) [two-way, consistency, same rater, single measurement] (MedCalc Software Ltd, Ostend, Belgium, version 20.109). The variability between subjects was evaluated by the coefficient of variation (CoV). The association between parameters was assessed by linear regression and correlation analysis (significance level α = 0.05). The assumptions of linear regression were assessed by testing for a linear relationship, testing for normal distribution of residuals and by visual inspection of residuals with respect to independence and homoscedasticity (see Supplementary Material 1).

3. Results

3.1. Assumptions of linear regression

The results of testing the assumptions of linear regression, for all datasets that were analyzed, are provided in Supplementary Material 1.

3.2. Absolute values

The OEF, MTT and AOEF results are summarized in Table 1, including population mean values, standard deviations (SDs) and the corresponding coefficients of variation (CoVs) between subjects. From the previous study by Knutsson et al. [24], a whole-brain CBF estimate is available for the investigated group of volunteers, with a group mean value of 43.6 ml/min/100 g. Applying this CBF value, in combination with the overall population mean QSM-based OEF value of 40.4 % from the current results, implies a population mean CMRO2 of 3.36 ml O2/min/100 g.

Table 1.

Summary of OEF, MTT and AOEF results.

OEF [%]
Visit 1
OEF [%]
Visit 2
MTT [s]
Visit 1
MTT [s]
Visit 2
AOEF [a.u.]
Visit 1
AOEF [a.u.]
Visit 2
Mean 39.3 41.4 3.20 3.25 0.207 0.203
SD 4.72 5.44 0.31 0.58 0.016 0.027
CoV 0.120 0.131 0.096 0.178 0.079 0.131

SD – standard deviation.

CoV – coefficient of variation between subjects.

3.3. Repeatability

For OEF, a test-retest scatter plot (i.e., OEF visit 2 versus OEF visit 1) is displayed in Figure 1a and the corresponding Bland-Altman plot is shown in Figure 1b. The ICCs for the OEF test-retest data were ICC(2,1) = 0.73 and ICC(3,1) = 0.79. The repeatability features of MTT and AOEF are illustrated in the Bland-Altman plots in Figures 2a and 2b, respectively. Neither OEF, MTT nor AOEF test-retest data exhibited any outliers, according to the Bland-Altman analyses.

Figure 1.

Figure 1

Test-retest results for QSM-based OEF estimates in healthy volunteers. (a) Scatter plot showing OEF from visit 2 versus OEF from visit 1. The solid line is the result of a linear regression analysis, and the dotted line is the line of identity. (b) Bland-Altman plot showing the OEF difference between visit 2 and visit 1 as a function of the mean OEF estimate from visit 1 and visit 2.

Figure 2.

Figure 2

Bland-Altman plots of DSC-MRI-based (a) MTT and (b) AOEF estimates from visit 1 and visit 2, showing the difference between visit 2 and visit 1 as a function of the mean value from visit 1 and visit 2.

3.4. OEF and AOEF age dependences

The healthy volunteers were recruited to represent a large age interval, and Figures 3a and 3b show OEF and AOEF (mean values of visit 1 and visit 2), respectively, as a function of age. In Figures 3a and 3b, trendlines have been added for completeness, although the hypothesis that there is no linear relationship could not be rejected (see Supplementary Material 1). For OEF, the linear regression analysis implied an increase of approximately 0.066 percentage points/year.

Figure 3.

Figure 3

(a) QSM-based OEF (mean value of visit 1 and visit 2) as a function of age. The solid line is a trendline resulting from a linear regression analysis. (b) AOEF (mean value of visit 1 and visit 2) as a function of age. The solid line is a trendline resulting from a linear regression analysis.

3.5. Comparison between OEF and DSC-MRI parameters (MTT and AOEF)

In the comparisons between OEF and MTT and between OEF and AOEF, the mean value of visit 1 and visit 2 is used to represent the respective parameter for each individual. The MTT-versus-OEF and AOEF-versus-OEF relationships are shown in Figures 4a and 4b, respectively, together with the results of a linear regression and correlation analysis. The association between MTT and OEF showed a correlation coefficient of r = 0.61 (p = 0.004) while the AOEF-versus-OEF relationship corresponded to r = 0.52 (p = 0.019).

Figure 4.

Figure 4

DSC-MRI-based estimates as a function of QSM-based OEF estimates (mean values of visit 1 and visit 2). Solid lines show the results of linear regression analyses. (a) MTT as a function of OEF. (b) AOEF as a function of OEF.

3.6. Numerical data

Numerical data are available in Supplementary Material 2.

4. Discussion

The present study concerns evaluation of QSM-based OEF data as well as initial investigation of the potential usefulness of a DSC-MRI-based OEF index based on the tissue residue function retrieved by deconvolution. Data were acquired using equipment and imaging protocols that should be readily available in a normal clinical environment.

QSM reconstruction is a delicate mathematical procedure, and quantitative results are influenced by a number of factors, such as quality of phase data, phase data preprocessing, type of algorithm for inversion, regulariza& ​Østergaardtion settings, noise, brain mask extraction, etc. Hence, evaluation of QSM-based quantitative parameters are indeed warranted, and the general idea was to conduct the QSM reconstruction procedure and OEF analysis in a fashion that would be feasible to apply in a routine workflow.

Regarding the absolute level, QSM-based OEF values were in good agreement with literature data. The mean value of 40.4% and the full range 33–50% are quite consistent with reference levels from textbooks and other scientific studies. For example, 15O positron emission tomography (PET) OEF measurements have resulted in 39 ± 6 % [34] and 43 ± 6 % [35]. The full range (min-max) of OEF values reported by Hattori et al. was 30–51 % [34]. In this context, is should be noted that there has been some debate with regard to the correct value of Δχdo. Previously published values include Δχdo = 0.18⋅4π ppm [1], Δχdo = 0.2⋅4π ppm [36, 37] and Δχdo = 0.27⋅4π ppm [38, 39]. In the present study, the result of the most recent investigation, focusing particularly on in vivo conditions, was employed, i.e., Δχdo = 0.193⋅4π ppm [20]. Furthermore, the hematocrit level can be influenced by several factors, and the use of a constant value for all subjects is not optimal. In a crude attempt to take the gender differences into account, the values Hct = 0.42 for males and Hct = 0.40 for females were applied in this study, as suggested by Peng et al. [40].

The OEF repeatability, described by the ICC(2,1) = 0.73, was good and similar to other hemodynamic parameters evaluated in the same group of volunteers. For example, for DSC-MRI-based whole-brain cerebral blood volume (CBV) and CBF estimates, ICC values of 0.64–0.85 were reported [24]. The reported increase in OEF with age corresponding to 0.066 percentage points/year (Figure 3a) is clearly not statistically significant, and should not be overinterpreted. However, the current observation is, at least, not in contradiction to previously observed patterns of age dependence. For comparison, the comprehensive PET study by Leenders et al. reported an OEF increase of 0.13 percentage points/year in healthy volunteers [41].

The previously reported CBF estimate [24] for the investigated group of volunteers enabled an assessment of the cerebral metabolic rate of oxygen as well, serving as an additional indication of the methodological quality. The observed population mean CMRO2 of 3.36 ml O2/100 g/min is in excellent agreement with textbook values of the normal CMRO2 level, provided by Clarke & Sokoloff (approximately 3.5 ml O2/min/100 g in young adults) [42] and by Seubert & Mahla (range 3.0–3.5 ml O2/min/100 g) [43].

The MTT value of 3.2 s was, in absolute terms, somewhat short, but still comparable to other DSC-MRI studies [44]. Also, in some cases, previous DSC-MRI investigations may have arrived at overestimated MTT (=CBV/CBF), considering that conventional deconvolution algorithms have tended to underestimate CBF [16]. Bézier deconvolution is also available with delay compensation in combination with a vascular transfer function (VTF) for modelling of dispersion [16]. For the present data, the version including VTF resulted in comparable (even slightly better) correlation with QSM-based OEF (r = 0.67, p = 0.001 for MTT and r = 0.52, p = 0.018 for AOEF), but unrealistically short MTT values in absolute terms (population mean MTT <1.5 s) (data not shown). This MTT underestimation was predicted by simulation of cases where a VTF was applied to situations with low or absent arterial dispersion (as can be expected in healthy subjects) [16]. With regard to the relationships between DSC-MRI parameters and OEF, it should be remembered that correlation coefficients must, in general, be interpreted with caution. For example, the OEF variability between individuals in a healthy population is limited, and the correlation coefficient tends, in general terms, to become lower when the inter-subject variability is small. A positive correlation between OEF and MTT has been predicted by cerebral microvasculature modelling [45], and experimentally observed, for example, in a DSC-MRI and PET study of patients with chronic occlusive cerebrovascular disease [46] and in a PET study with acetazolamide test [47]. In the present study, it is particularly interesting that we, using Bézier deconvolution, were able to see a moderate-to-good significant correlation also within the limited range of healthy volunteers.

The AOEF parameter is considerably less explored and the exact nature of the information that it provides remains to be established. Hence, it is probably fair to state that the observed degree of association between OEF and AOEF was not below expectations. In this context, it should be pointed out that previous studies have used a large range of different rate constant (k) values [11, 12, 48]. This indicates that the exact choice of k, on group level, is not crucial for obtaining reasonable AOEF estimates in relative terms. However, it also suggests that the numerical setting of the k value is not entirely straightforward, and it cannot be completely ruled out that a further optimized rate constant k in this study might have resulted in a slightly altered correlation with OEF.

In pathophysiological situations, the residue function shape can become substantially altered compared with the normal case [49, 50]; one might speculate that the AOEF metric would provide better sensitivity, compared with MTT, under such conditions considering the fact that two residue curves with markedly different shapes can, in principle, show the same MTT. The AOEF-vs-OEF coefficient of correlation was moderate, but still significant (p < 0.02), which is reasonably encouraging under the circumstances. Considering that assessment of oxygen metabolism requires combined information about CBF and OEF (cf. Eq. 3), it would certainly be advantageous to be able to extract data related to both these physiological parameters from the same image dataset. An alternative approach, likely to be more robust at present, would be to obtain OEF by, for example, MRI susceptometry or QSM measurements and CBF from pseudo-continuous arterial spin labelling (ASL) [51]. A general drawback with standard ASL implementations is that they do not provide estimates of conventional CBV or MTT. However, the use of multiple post-label delays allows for the assessment of arterial cerebral blood volume (aCBV) and arterial arrival time (AAT) which may provide comparable information [52, 53].

As implied above, the phase data for QSM reconstruction were acquired with a slightly dated protocol, and this should be considered in the interpretation of our results. However, a recent 7 T study with an up-to-date protocol resulted in a very similar group mean OEF value of 42 % in healthy volunteers (when applying Δχdo = 0.193⋅4π ppm and Hct = 0.40 for women and Hct = 0.42 for men), based on SSS QSM data and the same procedure of analysis as in the present study [32]. The DSC-MRI data acquisition in the present study was not regarded to be suboptimal in any way, and the preceding pre-bolus injection of a small dose of CA is not expected to negatively influence the main DSC-MRI experiment. For example, Jin et al. demonstrated, in a control experiment, that CBV estimates in normal (non-leaking) tissue were consistent between two sequential GRE-EPI DSC-MRI experiments (i.e. with repeated injections of Gd CA) [54], and, generally, DSC-MRI studies of brain tumours that are likely to show blood-brain barrier disruption frequently include administration of a small pre-bolus dose of CA to minimize T1 signal enhancement, caused by CA in the extravascular extracellular space [55]. Furthermore, the employed DSC-MRI deconvolution algorithm is novel and recently introduced by our group [16]. For ethical reasons, it would have been very difficult, or almost impossible, to undertake a similar test-retest contrast-agent injection scheme in healthy subjects today, and this was the main reason for employing a re-analysis of available datasets in this study.

Brain perfusion and related physiological parameters show variability or fluctuations over time and can be modified by a number of factors [56], and test-retest results may be influenced by such effects. However, systematic changes in resting-state OEF with age are not more than approximately 0.1 percentage points/year [41], so any such ageing effects should be negligible over approximately 1–3 weeks. With regard to short-term variability resulting from local brain activation, regional OEF is known to vary much less than CBF. Importantly, we investigated whole-brain estimates only, and neither whole-brain CBF nor whole-brain OEF is likely to vary significantly during local brain activation [57]. Furthermore, the aim of the present study was to assess parameters at rest, and the efficient autoregulation of the brain ensures that CBF at rest is fairly stable over time scales of days or weeks, as confirmed by previously reported ASL studies [58, 59].

In conclusion, QSM-based OEF showed a convincing absolute level and good test-retest results in terms of the ICC. Moderate-to-good correlations between DSC-MRI-based parameters (MTT and AOEF) and QSM-based OEF were observed. The results of the present study should be viewed upon as an indicator of the degree of robustness that can be achieved without applying extraordinary resources in terms of MRI equipment, imaging protocol, QSM preprocessing and reconstruction, and OEF analysis.

Declarations

Author contribution statement

Ronnie Wirestam: Conceived and designed the experiments; Performed the experiments; Analyzed and interpreted the data; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Anna Lundberg; Arthur Chakwizira: Analyzed and interpreted the data; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Danielle van Westen: Performed the experiments; Contributed reagents, materials, analysis tools or data.

Linda Knutsson: Performed the experiments; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Emelie Lind: Performed the experiments; Analyzed and interpreted the data; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Funding statement

Professor Ronnie Wirestam was supported by Vetenskapsrådet [MH 2017-00995].

Emelie Lind was supported by Hjärnfonden [PS2021-0045].

Data availability statement

Data will be made available on request.

Declaration of interest’s statement

The authors declare the following conflict of interests: Under a license agreement between Philips and the Johns Hopkins University, Dr. Knutsson's spouse, Dr. van Zijl and the University are entitled to fees related to an imaging device used in the study discussed in this publication. Dr. van Zijl also is a paid lecturer for Philips. This arrangement has been reviewed and approved by the Johns Hopkins University in accordance with its conflict of interest policies.

Additional information

Supplementary content related to this article has been published online at https://doi.org/10.1016/j.heliyon.2022.e12364.

Appendix B. Supplementary data

The following are the supplementary data related to this article:

Supplementary Material 1
mmc1.pdf (198.6KB, pdf)
Supplementary Material 2.xlsx
mmc2.xlsx (16.7KB, xlsx)

References

  • 1.Weisskoff R.M., Kiihne S. MRI susceptometry: image-based measurement of absolute susceptibility of MR contrast agents and human blood. Magn. Reson. Med. 1992;24:375–383. doi: 10.1002/mrm.1910240219. [DOI] [PubMed] [Google Scholar]
  • 2.Haacke E.M., Lai S., Reichenbach J.R., Kuppusamy K., Hoogenraad F.G., Takeichi H., Lin W. In vivo measurement of blood oxygen saturation using magnetic resonance imaging: a direct validation of the blood oxygen level-dependent concept in functional brain imaging. Hum. Brain Mapp. 1997;5:341–346. doi: 10.1002/(SICI)1097-0193(1997)5:5<341::AID-HBM2>3.0.CO;2-3. [DOI] [PubMed] [Google Scholar]
  • 3.Haacke E.M., Tang J., Neelavalli J., Cheng Y.C. Susceptibility mapping as a means to visualize veins and quantify oxygen saturation. J. Magn. Reson. Imaging. 2010;32:663–676. doi: 10.1002/jmri.22276. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Fan A.P., Bilgic B., Gagnon L., Witzel T., Bhat H., Rosen B.R., Adalsteinsson E. Quantitative oxygenation venography from MRI phase. Magn. Reson. Med. 2014;72:149–159. doi: 10.1002/mrm.24918. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Xu B., Liu T., Spincemaille P., Prince M., Wang Y. Flow compensated quantitative susceptibility mapping for venous oxygenation imaging. Magn. Reson. Med. 2014;72:438–445. doi: 10.1002/mrm.24937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Cho J., Nguyen T.D., Huang W., Sweeney E.M., Luo X., Kovanlikaya I., Zhang S., Gillen K.M., Spincemaille P., Gupta A., Gauthier S.A., Wang Y. Brain oxygen extraction fraction mapping in patients with multiple sclerosis. J. Cereb. Blood Flow Metab. 2022;42:338–348. doi: 10.1177/0271678X211048031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.McFadden J.J., Matthews J.C., Scott L.A., Parker G.J.M., Lohézic M., Parkes L.M. Optimization of quantitative susceptibility mapping for regional estimation of oxygen extraction fraction in the brain. Magn. Reson. Med. 2021;86:1314–1329. doi: 10.1002/mrm.28789. [DOI] [PubMed] [Google Scholar]
  • 8.Uchida Y., Kan H., Inoue H., Oomura M., Shibata H., Kano Y., Kuno T., Usami T., Takada K., Yamada K., Kudo K., Matsukawa N. Penumbra detection with oxygen extraction fraction using magnetic susceptibility in patients with acute ischemic stroke. Front. Neurol. 2022;13 doi: 10.3389/fneur.2022.752450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Østergaard L., Weisskoff R.M., Chesler D.A., Gyldensted C., Rosen B.R. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: mathematical approach and statistical analysis. Magn. Reson. Med. 1996;36:715–725. doi: 10.1002/mrm.1910360510. [DOI] [PubMed] [Google Scholar]
  • 10.Knutsson L., Ståhlberg F., Wirestam R. Absolute quantification of perfusion using dynamic susceptibility contrast MRI: pitfalls and possibilities. MAGMA. 2010;23:1–21. doi: 10.1007/s10334-009-0190-2. [DOI] [PubMed] [Google Scholar]
  • 11.Jespersen S.N., Østergaard L. The roles of cerebral blood flow, capillary transit time heterogeneity, and oxygen tension in brain oxygenation and metabolism. J. Cereb. Blood Flow Metab. 2012;32:264–277. doi: 10.1038/jcbfm.2011.153. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Mouridsen K., Hansen M.B., Østergaard L., Jespersen S.N. Reliable estimation of capillary transit time distributions using DSC-MRI. J. Cereb. Blood Flow Metab. 2014;34:1511–1521. doi: 10.1038/jcbfm.2014.111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wu O., Østergaard L., Weisskoff R.M., Benner T., Rosen B.R., Sorensen A.G. Tracer arrival timing-insensitive technique for estimating flow in MR perfusion-weighted imaging using singular value decomposition with a block-circulant deconvolution matrix. Magn. Reson. Med. 2003;50:164–174. doi: 10.1002/mrm.10522. [DOI] [PubMed] [Google Scholar]
  • 14.Mouridsen K., Friston K., Hjort N., Gyldensted L., Østergaard L., Kiebel S. Bayesian estimation of cerebral perfusion using a physiological model of microvasculature. NeuroImage. 2006;33:570–579. doi: 10.1016/j.neuroimage.2006.06.015. [DOI] [PubMed] [Google Scholar]
  • 15.Mehndiratta A., MacIntosh B.J., Crane D.E., Payne S.J., Chappell M.A. A control point interpolation method for the non-parametric quantification of cerebral haemodynamics from dynamic susceptibility contrast MRI. NeuroImage. 2013;64:560–570. doi: 10.1016/j.neuroimage.2012.08.083. [DOI] [PubMed] [Google Scholar]
  • 16.Chakwizira A., Ahlgren A., Knutsson L., Wirestam R. Non-parametric deconvolution using Bézier curves for quantification of cerebral perfusion in dynamic susceptibility contrast MRI. MAGMA. 2022;35:791–804. doi: 10.1007/s10334-021-00995-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Liu J., Liu T., de Rochefort L., Ledoux J., Khalidov I., Chen W., Tsiouris A.J., Wisnieff C., Spincemaille P., Prince M.R., Wang Y. Morphology enabled dipole inversion for quantitative susceptibility mapping using structural consistency between the magnitude image and the susceptibility map. Neuroimage. 2012;59:2560–2568. doi: 10.1016/j.neuroimage.2011.08.082. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Salomir R., de Senneville B.D., Moonen C.T.W. A fast calculation method for magnetic field inhomogeneity due to an arbitrary distribution of bulk susceptibility. Concepts Magn. Reson. Part B Magn. Reson. Eng. 2003;19B:26–34. [Google Scholar]
  • 19.Marques J.P., Bowtell R. Application of a Fourier-based method for rapid calculation of field inhomogeneity due to spatial variation of magnetic susceptibility. Concepts Magn. Reson. Part B Magn. Reson. Eng. 2005;25B:65–78. [Google Scholar]
  • 20.Eldeniz C., Binkley M.M., Fields M., Guilliams K., Ragan D.K., Chen Y., Lee J.M., Ford A.L., An H. Bulk volume susceptibility difference between deoxyhemoglobin and oxyhemoglobin for HbA and HbS: a comparative study. Magn. Reson. Med. 2021;85:3383–3393. doi: 10.1002/mrm.28668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Hall J.E., Hall M.E. Guyton and Hall Textbook of Medical Physiology. fourteenth ed. Elsevier; Philadelphia: 2021. Red blood cells, anemia, and polycythemia; pp. 439–447. [Google Scholar]
  • 22.Dijkhuizen P., Buursma A., Fongers T.M.E., Gerding A.M., Oesburg B., Zijlstra W.G. The oxygen binding capacity of human haemoglobin. Pflugers Arch. 1977;369:223–231. doi: 10.1007/BF00582188. [DOI] [PubMed] [Google Scholar]
  • 23.Wirestam R. Using contrast agents to obtain maps of regional perfusion and capillary wall permeability. Imag. Med. 2012;4:423–442. [Google Scholar]
  • 24.Knutsson L., Lindgren E., Ahlgren A., Osch MJP van, Bloch K.M., Surova Y., Ståhlberg F., Westen D van, Wirestam R. Dynamic susceptibility contrast MRI with a prebolus contrast agent administration design for improved absolute quantification of perfusion. Magn. Reson. Med. 2014;72:996–1006. doi: 10.1002/mrm.25006. [DOI] [PubMed] [Google Scholar]
  • 25.Kämpe R., Lind E., Ståhlberg F., van Westen D., Knutsson L., Wirestam R. Quantification of normal cerebral oxygen extraction and oxygen metabolism by phase-based MRI susceptometry: evaluation of repeatability using two different imaging protocols. Clin. Physiol. Funct. Imaging. 2017;37:211–220. doi: 10.1111/cpf.12288. [DOI] [PubMed] [Google Scholar]
  • 26.Young S., Bystrov D., Netsch T., Bergmans R., van Muiswinkel A., Visser F., Springorum R., Hoogenraad F., Gieseke J. Proceedings of the SPIE 6144 on Medical Imaging (Image Processing) 2006. Automated planning of MRI neuro scans. San Diego, California, USA. [Google Scholar]
  • 27.de Rochefort L., Liu T., Kressler B., Liu J., Spincemaille P., Lebon V., Wu J., Wang Y. Quantitative susceptibility map reconstruction from MR phase data using bayesian regularization: validation and application to brain imaging. Magn. Reson. Med. 2010;63:194–206. doi: 10.1002/mrm.22187. [DOI] [PubMed] [Google Scholar]
  • 28.Liu T., Liu J., de Rochefort L., Spincemaille P., Khalidov I., Ledoux J.R., Wang Y. Morphology enabled dipole inversion (MEDI) from a single-angle acquisition: comparison with COSMOS in human brain imaging. Magn. Reson. Med. 2011;66:777–783. doi: 10.1002/mrm.22816. [DOI] [PubMed] [Google Scholar]
  • 29.Liu T., Wisnieff C., Lou M., Chen W., Spincemaille P., Wang Y. Nonlinear formulation of the magnetic field to source relationship for robust quantitative susceptibility mapping. Magn. Reson. Med. 2013;69:467–476. doi: 10.1002/mrm.24272. [DOI] [PubMed] [Google Scholar]
  • 30.Cusack R., Papadakis N. New robust 3-D phase unwrapping algorithms: application to magnetic field mapping and undistorting echoplanar images. Neuroimage. 2002;16:754–764. doi: 10.1006/nimg.2002.1092. [DOI] [PubMed] [Google Scholar]
  • 31.Liu T., Khalidov I., de Rochefort L., Spincemaille P., Liu J., Tsiouris A.J., Wang Y. A novel background field removal method for MRI using projection onto dipole fields (PDF) NMR Biomed. 2011;24:1129–1136. doi: 10.1002/nbm.1670. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Lundberg A., Lind E., Olsson H., Helms G., Knutsson L., Wirestam R. Comparison of MRI methods for measuring whole-brain oxygen extraction fraction under different geometric conditions at 7T. J Neuroimaging. 2022;32:442–458. doi: 10.1111/jon.12975. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Bland J.M., Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–310. [PubMed] [Google Scholar]
  • 34.Hattori N., Bergsneider M., Wu H.M., et al. Accuracy of a method using short inhalation of 15O-O2 for measuring cerebral oxygen extraction fraction with PET in healthy humans. J. Nucl. Med. 2004;45:765–770. [PubMed] [Google Scholar]
  • 35.Bremmer J.P., van Berckel B.N., Persoon S., et al. Day-to-day test-retest variability of CBF, CMRO2, and OEF measurements using dynamic 15O PET studies. Mol. Imaging Biol. 2011;13:759–768. doi: 10.1007/s11307-010-0382-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Thulborn K.R., Waterton J.C., Matthews P.M., Radda G.K. Oxygenation dependence of the transverse relaxation time of water protons in whole blood at high field. Biochim. Biophys. Acta. 1982;714:265–270. doi: 10.1016/0304-4165(82)90333-6. [DOI] [PubMed] [Google Scholar]
  • 37.Plyavin Y.A., Blum E. Magnetic parameters of blood cells and high-gradient paramagnetic and diamagnetic phoresis. Magn Gidrodin. 1982;19:3–15. [Google Scholar]
  • 38.Spees W.M., Yablonskiy D.A., Oswood M.C., Ackerman J.J. Water proton MR properties of human blood at 1.5 Tesla: magnetic susceptibility, T1, T2, T2∗, and non-Lorentzian signal behavior. Magn. Reson. Med. 2001;45:533–542. doi: 10.1002/mrm.1072. [DOI] [PubMed] [Google Scholar]
  • 39.Jain V., Abdulmalik O., Propert K.J., Wehrli F.W. Investigating the magnetic susceptibility properties of fresh human blood for noninvasive oxygen saturation quantification. Magn. Reson. Med. 2012;68:863–867. doi: 10.1002/mrm.23282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Peng S.L., Dumas J.A., Park D.C., Liu P., Filbey F.M., McAdams C.J., Pinkham A.E., Adinoff B., Zhang R., Lu H. Age-related increase of resting metabolic rate in the human brain. Neuroimage. 2014;98:176–183. doi: 10.1016/j.neuroimage.2014.04.078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Leenders K.L., Perani D., Lammertsma A.A., Heather J.D., Buckingham P., Healy M.J., Gibbs J.M., Wise R.J., Hatazawa J., Herold S., et al. Cerebral blood flow, blood volume and oxygen utilization. Normal values and effect of age. Brain. 1990;113(Pt 1):27–47. doi: 10.1093/brain/113.1.27. [DOI] [PubMed] [Google Scholar]
  • 42.Clarke D.D., Sokoloff L. In: Basic Neurochemistry: Molecular, Cellular and Medical Aspects. sixth ed. Siegel G.J., editor. Lippincott-Raven; Philadelphia: 1999. Regulation of cerebral metabolic rate. Agranoff BW, Albers RW, Fisher SK, Uhler MD. [Google Scholar]
  • 43.Seubert C.N., Mahla M.E. In: Miller's Anesthesia., Miller R.D., editors. Vol. 1. Elsevier; Amsterdam: 2014. pp. 1487–1524. (Neurologic Monitoring). [Google Scholar]
  • 44.Ibaraki M., Ito H., Shimosegawa E., Toyoshima H., Ishigame K., Takahashi K., Kanno I., Miura S. Cerebral vascular mean transit time in healthy humans: a comparative study with PET and dynamic susceptibility contrast-enhanced MRI. J. Cereb. Blood Flow Metab. 2007;27:404–413. doi: 10.1038/sj.jcbfm.9600337. [DOI] [PubMed] [Google Scholar]
  • 45.Park C.S., Payne S.J. Modelling the effects of cerebral microvasculature morphology on oxygen transport. Med. Eng. Phys. 2016;38:41–47. doi: 10.1016/j.medengphy.2015.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Mihara F., Kuwabara Y., Tanaka A., Yoshiura T., Sasaki M., Yoshida T., Masuda K., Matsushima T. Reliability of mean transit time obtained using perfusion-weighted MR imaging; comparison with positron emission tomography. Magn. Reson. Imaging. 2003;21:33–39. doi: 10.1016/s0730-725x(02)00629-x. [DOI] [PubMed] [Google Scholar]
  • 47.Hokari M., Kuroda S., Shiga T., Nakayama N., Tamaki N., Iwasaki Y. Combination of a mean transit time measurement with an acetazolamide test increases predictive power to identify elevated oxygen extraction fraction in occlusive carotid artery diseases. J. Nucl. Med. 2008;49:1922–1927. doi: 10.2967/jnumed.108.054379. [DOI] [PubMed] [Google Scholar]
  • 48.Eskildsen S.F., Gyldensted L., Nagenthiraja K., Nielsen R.B., Hansen M.B., Dalby R.B., Frandsen J., Rodell A., Gyldensted C., Jespersen S.N., Lund T.E., Mouridsen K., Brændgaard H., Østergaard L. Increased cortical capillary transit time heterogeneity in Alzheimer's disease: a DSC-MRI perfusion study. Neurobiol. Aging. 2017;50:107–118. doi: 10.1016/j.neurobiolaging.2016.11.004. [DOI] [PubMed] [Google Scholar]
  • 49.Park C.S., Payne S.J. A generalized mathematical framework for estimating the residue function for arbitrary vascular networks. Interface Focus. 2013;3 doi: 10.1098/rsfs.2012.0078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Østergaard L., Sorensen A.G., Chesler D.A., Weisskoff R.M., Koroshetz W.J., Wu O., Gyldensted C., Rosen B.R. Combined diffusion-weighted and perfusion-weighted flow heterogeneity magnetic resonance imaging in acute stroke. Stroke. 2000;31:1097–1103. doi: 10.1161/01.str.31.5.1097. [DOI] [PubMed] [Google Scholar]
  • 51.Alsop D.C., Detre J.A., Golay X., Günther M., Hendrikse J., Hernandez-Garcia L., Lu H., MacIntosh B.J., Parkes L.M., Smits M., van Osch M.J., Wang D.J., Wong E.C., Zaharchuk G. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: a consensus of the ISMRM perfusion study group and the European consortium for ASL in dementia. Magn. Reson. Med. 2015;73:102–116. doi: 10.1002/mrm.25197. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Bright M.G., Croal P.L., Blockley N.P., Bulte D.P. Multiparametric measurement of cerebral physiology using calibrated fMRI. Neuroimage. 2019;187:128–144. doi: 10.1016/j.neuroimage.2017.12.049. [DOI] [PubMed] [Google Scholar]
  • 53.Wang D.J., Alger J.R., Qiao J.X., Gunther M., Pope W.B., Saver J.L., Salamon N., Liebeskind D.S. UCLA Stroke Investigators. Multi-delay multi-parametric arterial spin-labeled perfusion MRI in acute ischemic stroke - comparison with dynamic susceptibility contrast enhanced perfusion imaging. Neuroimage Clin. 2013;3:1–7. doi: 10.1016/j.nicl.2013.06.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Jin S., Kang M., Cho H. Cerebral blood perfusion deficits using dynamic susceptibility contrast MRI with gadolinium chelates in rats with post-ischemic reperfusion without significant dynamic contrast-enhanced MRI-derived vessel permeabilities: a cautionary note. PLoS One. 2018;13(7) doi: 10.1371/journal.pone.0201076. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Boxerman J.L., Quarles C.C., Hu L.S., Erickson B.J., Gerstner E.R., Smits M., Kaufmann T.J., Barboriak D.P., Huang R.H., Wick W., Weller M., Galanis E., Kalpathy-Cramer J., Shankar L., Jacobs P., Chung C., van den Bent M.J., Chang S., Al Yung W.K., Cloughesy T.F., Wen P.Y., Gilbert M.R., Rosen B.R., Ellingson B.M., Schmainda K.M. Jumpstarting brain tumor drug development coalition imaging standardization steering committee. Consensus recommendations for a dynamic susceptibility contrast MRI protocol for use in high-grade gliomas. Neuro Oncol. 2020;22(9):1262–1275. doi: 10.1093/neuonc/noaa141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Clement P., Mutsaerts H.J., Václavů L., Ghariq E., Pizzini F.B., Smits M., Acou M., Jovicich J., Vanninen R., Kononen M., Wiest R., Rostrup E., Bastos-Leite A.J., Larsson E.M., Achten E. Variability of physiological brain perfusion in healthy subjects - a systematic review of modifiers. Considerations for multi-center ASL studies. J Cereb. Blood Flow Metab. 2018;38:1418–1437. doi: 10.1177/0271678X17702156. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Mintun M.A., Vlassenko A.G., Shulman G.L., Snyder A.Z. Time-related increase of oxygen utilization in continuously activated human visual cortex. Neuroimage. 2002;16:531–537. doi: 10.1006/nimg.2002.1114. [DOI] [PubMed] [Google Scholar]
  • 58.Parkes L.M., Rashid W., Chard D.T., Tofts P.S. Normal cerebral perfusion measurements using arterial spin labeling: reproducibility, stability, and age and gender effects. Magn. Reson. Med. 2004;51:736–743. doi: 10.1002/mrm.20023. [DOI] [PubMed] [Google Scholar]
  • 59.Kilroy E., Apostolova L., Liu C., Yan L., Ringman J., Wang D.J. Reliability of two-dimensional and three-dimensional pseudo-continuous arterial spin labeling perfusion MRI in elderly populations: comparison with 15O-water positron emission tomography. J. Magn. Reson. Imaging. 2014;39:931–939. doi: 10.1002/jmri.24246. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material 1
mmc1.pdf (198.6KB, pdf)
Supplementary Material 2.xlsx
mmc2.xlsx (16.7KB, xlsx)

Data Availability Statement

Data will be made available on request.


Articles from Heliyon are provided here courtesy of Elsevier

RESOURCES