Abstract
Objective
To determine the diagnostic agreement of non‐cycloplegic and cycloplegic refraction in children.
Method
The study methodology followed Preferred Reporting Items for Systematic Reviews and Meta‐Analyses (PRISMA) guidelines. Electronic databases were searched for comparative studies exploring refraction performed on children under non‐cycloplegic and cycloplegic conditions. There was no restriction on the year of publication; however, only publications in the English language were eligible. Inclusion criteria consisted of children aged ≤12 years, any degree or type of refractive error, either sex and no ocular or binocular co‐morbidities. The QUADAS‐2 tool was used to evaluate the risk of bias. Meta‐analysis was conducted to synthesise data from all included studies. Subgroup and sensitivity analyses were undertaken for those studies with a risk of bias.
Results
Ten studies consisting of 2724 participants were eligible and included in the meta‐analysis. The test for overall effect was not significant when comparing non‐cycloplegic Plusoptix and cycloplegic autorefractors (Z = 0.34, p = 0.74). The pooled mean difference (MD) was −0.08 D (95% CI −0.54 D, +0.38 D) with a prediction interval of −1.72 D to +1.56 D. At less than 0.25 D, this indicates marginal overestimation of myopia and underestimation of hyperopia under non‐cycloplegic conditions. When comparing non‐cycloplegic autorefraction with a Retinomax and Canon autorefractor to cycloplegic refraction, a significant difference was found (Z = 9.79, p < 0.001) and (Z = 4.61, p < 0.001), respectively.
Discussion
Non‐cycloplegic Plusoptix is the most useful autorefractor for estimating refractive error in young children with low to moderate levels of hyperopia. Results also suggest that cycloplegic refraction must remain the test of choice when measuring refractive error ≤12 years of age. There were insufficient data to explore possible reasons for heterogeneity. Further research is needed to investigate the agreement between non‐cycloplegic and cycloplegic refraction in relation to the type and level of refractive error at different ages.
Keywords: child, cycloplegia, eye examination, refraction
Key points.
Without cycloplegia, the Plusoptix is the most accurate machine to measure the spectacle prescription in children below 12 years of age.
However, the Plusoptix cannot be substituted for a measurement which includes the use of cycloplegic drops.
In future studies, the precision of cycloplegia needs to be examined across different spectacle prescriptions and children's ages.
INTRODUCTION
Cyclopentolate hydrochloride is a synthetic antimuscarinic cycloplegic agent. 1 It is the first choice for providing excellent short‐term paralysis of accommodation. 2 , 3 Cycloplegic refraction (CR) is an effective way of reducing fluctuations in accommodation or spasm of the ciliary muscle. 4 , 5 , 6 The temporary paralysis of accommodation is useful when refracting young children as their accommodative system is vigorous, leading to inaccuracies in non‐cycloplegic refraction (NCR) 7 with overestimation of myopia and underestimation of hyperopia. 8
Cyclopentolate hydrochloride has several ocular side effects, including irritation, lacrimation, allergic blepharoconjunctivitis, conjunctival hyperaemia and systemic side effects such as drowsiness, disorientation, incoherent speech and visual hallucinations. 9 From a clinical perspective, CR is considered the gold standard for measuring refractive errors due to high accuracy. 10 However, CR is an invasive procedure involving the use of eye drops which many patients find uncomfortable, potentially causing distress amongst younger children. 11 Many parents and children refuse cycloplegia due to the stinging sensation experienced on insertion of the drop, the resulting blurred vision and other side effects such as light sensitivity. 11 , 12 , 13 , 14 Therefore, the use of these diagnostic drops could deter parents and children from attending an eye examination. 15
The cycloplegic effects begin between 25 and 75 min after the administration of the drug with recovery up to 24 h later which may prove a deterrent for the patient 16 , 17 , 18 , 19 and has significant cost implications for the UK National Health Service (NHS) with finite resources. 20 The cost‐effectiveness of cycloplegic use has been explored in 2005, and in a cohort of 78 children, a median total cost of 2.08 Pounds Sterling per patient for this agent has been reported. 21
Despite the side effects, time and cost implications, the Royal College of Ophthalmologists (UK) recommends that children under the age of 12 years require CR. 22 This is particularly true when hyperopia is suspected or when a binocular vision anomaly is present and the full hyperopic prescription needs to be known. 22 This is also the case when prescribing the least minus prescription to a myopic child, as over prescribing can drive accommodation and axial length growth resulting in myopic progression. 23 In addition, the College of Optometrists (UK) advises optometrists to consider performing cycloplegic examination when refracting young children (without age specification) to obtain an accurate refraction and the best possible view of the fundus. 24
Research suggests that there might be differences in the diagnostic agreement of NCR and CR based on the type of refractive error, 8 the level of refractive error, 11 , 13 the patient's age 10 or the method of refraction. 25 , 26 Older patients are less likely to show significant differences between non‐cycloplegic (NC) and cycloplegic (C) refractive error measurements. 27 , 28 Previous work has explored NC and C retinoscopy and autorefraction and found more myopic/less hyperopic measurements during NC assessments, resulting in underestimation of hyperopia and overestimation of myopia in children. 11 , 13 , 26 Therefore, it is important to understand whether NCR and CR are comparable in some clinical scenarios. Evidence suggests that the NCR with the Plusoptix autorefractor (plusoptix.com) has high sensitivity in the detection of myopia and astigmatism 29 , 30 and shows good agreement with C retinoscopy, the latter being conducted by a paediatric ophthalmologist. 31
Uncorrected refractive error is the leading cause of moderate to severe visual impairment and the second most common cause of blindness. 32 In addition, a refractive error that remains uncorrected or not appropriately corrected can lead to manifest strabismus and anisometropia, which are common risk factors for amblyopia, and have the potential to result in the permanent loss of binocular function. 33 , 34 , 35 Accurate measurement of refractive error is essential to ensure that children achieve optimal visual acuity and binocular status. Despite ample research investigating NCR versus CR, the inclusion criteria, results and quality of research are variable. While comparisons have been made between cycloplegic and other refraction methods in children, there are no clear indications of the need for cycloplegia in children, based on age, type or level of refractive error. Therefore, more guidance is required on the need for CR in infants and children and whether NC options are appropriate in some scenarios.
The present review and analysis aims to provide guidance on the use of CR in children ≤12 years old by synthesising data from relevant studies at low risk of bias, and to determine the diagnostic agreement of NCR and CR in children ≤12 years of age.
METHODS
Eligibility criteria
Comparative studies including participants who had undergone refraction with and without cyclopentolate hydrochloride as a cycloplegic agent were included; studies using all other cycloplegic agents were excluded. There was no restriction on the year of publication or location of study; however, only publications in the English language were eligible due to the reviewers' capabilities. Published and unpublished journals (abstracts and dissertations) were searched. The reference lists from relevant retrieved studies were also searched. Citation alerts were used to ensure that the more recently published studies were included. Studies were included if the study participants met the following inclusion criteria: children ≤12 years old of either sex, with any type or level of refractive error and without other ocular or binocular vision anomalies. To avoid duplication of results, studies were excluded if they reported re‐analysis or re‐publication from initial data.
Search methods
The Preferred Reporting Items for Systematic Reviews and Meta Analyses (PRISMA) guidelines were followed. 36 Electronic searches were conducted using both thesaurus controlled and text terms to increase sensitivity. The following resources were checked on 24 June 2020 to identify whether a systematic review or a similar research question had been proposed or if one had been published: PROSPERO, Cochrane Library, National Institute for Health and Care Excellence (NICE) evidence, TRIP, EBSCOhost and OVID online. No current review protocol exists under NCR and CR (https://www.crd.york.ac.uk/PROSPERO/).
Selection process
Two review authors (SW and MC) independently assessed the titles and abstracts of all investigations retrieved via the electronic searches to exclude irrelevant publications. Studies were marked as ‘definitely relevant’, ‘possibly relevant’ or ‘definitely not relevant’. Those investigations marked as ‘definitely not relevant’ by both review authors were excluded. Studies marked as ‘definitely relevant’ or ‘possibly relevant’ by both review authors were independently assessed against the inclusion criteria to determine whether they were relevant to this review. This was done by obtaining full copies of the relevant papers. Any disparities at this stage were initially solved through discussion to reach consensus.
Data collection process
The two review authors (SW and MC) independently extracted data from the included studies using a standardised data collection form. If disagreements arose during this process, they were resolved through discussion until consensus was achieved. Data were then entered into Review Manager 5 (RevMan 5). 37
Study risk of bias assessment
The two reviewing authors independently assessed each eligible study for risk of bias and assessed the quality of the body of evidence in this review using the QUADAS‐2 tool. 38 This assessment tool has undergone evaluation to ensure validity and usefulness. 38 , 39 Bias is a systematic error or a deviation from the truth in the results or inferences leading to over or underestimating test accuracy. 39 A checklist approach was used with categories ‘yes’, ‘no’ and ‘unclear’ to assess the quality of each domain (patient selection, index test, reference standard and flow and timing). 40 Any discrepancies were resolved through discussion until consensus had been obtained. The questions used to evaluate the risk of bias can be found in Table A1 in Appendix 1.
Effect measures
The main outcome measure was the agreement between NCR and CR. In addition, the mean difference (MD) with a corresponding 95% confidence interval (CI) was calculated. 41 The 95% limits of agreement (LoA) were not obtained due to limitations with the available data. Having reviewed previous research in this area, we used a value of ±0.85 D which falls outside the published intra‐examiner limits of repeatability for C retinoscopy in 4‐year‐olds, 42 , 43 and was rounded up to 1.00 D as refractive error is measured in 0.25 D steps. 44 The 95% prediction interval was calculated to quantify the impact of between‐study heterogeneity. The pooled estimate of MD and random error was calculated using the Der Simonian and Laird random‐effects method. 45 Moreover, the mean difference between NC and C conditions with the Plusoptix autorefractor could not be undertaken, as the Plusoptix has been designed to be used on undilated pupils. 46
Synthesis methods
Statistical analysis and data synthesis were conducted in line with methods described in Chapter 10 of the Cochrane Handbook for Systematic Reviews of Diagnostic Test Accuracy. 40
We included studies that either evaluated only one eye of each participant or both eyes. This was due to the nature of the data available which was averaged for each sample making extraction of individual data unfeasible. However, this is likely to overestimate the mean difference when the analysis is concluded. The eligible studies were tabulated according to the number and age of participants, range of refractive error explored and the type of refraction undertaken.
Heterogeneity was evaluated using the Cochrane Q test and I2 index. 47 To explore between‐study variance, tau‐squared (τ 2) was calculated. 48 The prediction interval, which is the index of dispersion, was also calculated to examine how widely the scores varied and the variance of the effect size.
Sensitivity analyses were undertaken to assess the impact of the risk of bias on test accuracy by repeating the analysis after removing the studies with a high risk of bias. Heterogeneity was explored using subgroup analysis based on the mode of refraction (retinoscopy or autorefractor) and the child's age.
RESULTS
Study selection
Of 150 reports, 131 full text manuscripts were obtained. Ten of these studies met the inclusion criteria (Figure 1). During the study selection process, 119 reports were excluded as they did not meet the review criteria. Those studies that were eligible for this review are shown in Table 1.
FIGURE 1.
Preferred reporting items for systematic reviews and meta‐analyses (PRISMA) flow diagram illustrating the screening process. 36 CET, Continuing education and training.
TABLE 1.
Summary table of the included studies in this review
Study | Country / Region | Number of participants | Age | Refractive error range (D = Dioptres) | Range of spherical equivalent refractive error of sample (D = Dioptres) | Refractive error definition given |
Type or refraction NC = non‐cycloplegic C = cycloplegic |
---|---|---|---|---|---|---|---|
Akil 49 | Turkey | 112 |
Range 2–12 years Mean 6.78 ± 2.61 years |
No restriction | The range was not disclosed. | No |
NC Canon RK‐F1 NC Retinomax K‐plus 3 C Canon RK‐F1 C Retinomax K‐plus 3 C Retinoscopy |
Demirci 50 | Turkey | 118 |
Range 1–12 years Mean 4.9 ± 2.6 years |
Spherical range −7.00 D to +5.00 D Cylindrical range −7.00 D to +5.00 D |
Plusoptix (−5.38 D to +2.75 D) Topcon (−5.38 D to +2.50 D) Retinoscopy (−3.88 D to +2.75 D) |
No |
NC Plusoptix S08 (PS08) C Topcon RM8800 C Retinoscopy |
Harvey 51 | North America | 36 |
Range 3.6–5.6 years Mean 4.7 years |
No restriction | Retinoscopy (−3.75 D to +3.50 D) | No |
NC Retinomax K+ C Retinomax K+ C Retinoscopy |
Iuorno 52 | North America | 91 |
Range 37–107 months (3.1–8.9 years) Mean 97 months (8.1 years) |
No restriction |
Welch Allyn Sure Sight (WASS) (−3.375 D to +4.375 D) Nidek AR‐820 (NAR) (−3.50 D to +6.75 D) Retinoscopy (−3.50 D to +6.875 D) |
No |
NC Welch Allyn SureSight (WASS) C Nidek AR‐820 (NAR) C Retinoscopy |
Ozdemir 53 | Turkey | 98 |
Range 12–60 months (1–5 years) Mean 28.8 ± 18.5 months (2.4 ± 1.5 years) |
Spherical range −7.00 D to +5.00 D Cylindrical range −7.00 D to +5.00 D |
The range was not disclosed. | No |
NC Plusoptix A09 C Plusoptix A09 C Refraction (Retinoscopy /WASS) |
Seymen 54 | Turkey | 194 |
Range 3–34 months (0.25–2.8 years) Mean 16.65 ± 10.04 months (1.4 ± 1.3 years) |
Spherical range −7.00 D to +5.00 D Cylindrical range −7.00 D to +5.00 D |
Plusoptix (−3.38 D to +5.88 D) Retinomax (−2.88 D to +6.50 D) HandyRef‐K (−3.00 D to +6.63 D) |
No |
NC Plusoptix A09 C Retinomax K‐Plus 3 C HandyRef‐K |
Vasudevan 55 | North America | 5 | Range 5–12 years |
Exclude: Anisometropia >2.00 D Spherical refraction more than ±4.00 D or cylindrical component more than 1.00 D |
The range was not disclosed. | No |
NC WAM‐5500 C Retinoscopy |
Won 56 | Korea | 40 | Range 2–10 years | Spherical range −7.00 D to +5.00 D | The range was not disclosed. | No |
NC Plusoptix S09 NC Canon RK‐F1 C Canon RK‐F1 |
Yilmaz 18 | Turkey | 200 |
Range 4–12 years Mean 6.2 ± 2.8 years |
Spherical range −7.00 D to +5.00 D Cylindrical range −7.00 D to +5.00 D |
The range was not disclosed. | No |
NC Plusoptix A09 C Retinomax K‐Plus3 C Retinoscopy |
Li 57 | China | 1830 |
Range 5.89–10.32 years Mean 6.38 ± 0.46 years |
No restriction |
Non‐cycloplegic (−8.63 D to +8.94 D) Cycloplegic (−8.50 D to +8.63 D) |
Yes |
NC KR‐800 Topcon C KR‐800 Topcon |
Study characteristics of included studies
Risk of bias in studies
Figures 2 and 3 summarise the risk of bias and applicability. The risk of bias in the included studies was due to a different criterion being used for selecting children that would have NCR, a refractive threshold introduced for NCR limiting the range of refractive error being explored, lack of masking or an insufficient interval between NCR and CR. Unclear concern about applicability was found for only one study due to participant selection and refraction criteria.
FIGURE 2.
Risk of bias and applicability concerns: Review authors' judgements about each domain are presented as percentages of all the included studies.
FIGURE 3.
Risk of bias and applicability concerns summary: Review authors' judgements on the included studies for each domain.
Results of individual studies
Comparison of NC and C autorefractors
The test for overall effect was not significant in the NC Plusoptix subgroup (MD −0.08 D) 95% CI (−0.54 D, +0.38 D) (Z = 0.34, p = 0.74) (Table 2). The prediction interval for the NC Plusoptix autorefractor is −1.72 D to +1.56 D, indicating that in 95% of all populations, the true effect size will fall in this range. However, the NC Retinomax and NC Canon subgroups illustrate a significant overall effect (MD −1.17 D) 95% CI (−1.41 D, −0.94 D) (Z = 9.79, p < 0.00001) and (MD −1.20 D) 95% CI (−1.71 D, −0.69 D) (Z = 4.61, p < 0.00001), respectively. A significant difference was found between the subgroups (Q = 17.93, df = 2, p = 0.00001).
TABLE 2.
Data from studies illustrated in the forest plot comparing non‐cycloplegic (NC) to cycloplegic (C) autorefraction using different autorefractors
Comparison of NC autorefractors and C retinoscopy
The forest plot in Table 3 compares NC autorefraction and C retinoscopy. The overall effect (MD) is −0.55 D with a 95% confidence interval of −1.13 D to +0.04 D (Z = 1.84, p = 0.07). Therefore, on average, refractive error estimation is more myopic under NC than C conditions by ≈0.50 D. The Q‐value is 84.50 with 5 degrees of freedom and p < 0.0001 indicating a significant difference between NCR and CR.
TABLE 3.
Data from studies illustrated in the forest plot comparing non‐cycloplegic autorefraction and cycloplegic retinoscopy
The true effect size varies between studies. The I2 statistic tells us that the proportion of the observed variance reflects differences in the true effect size rather than sampling error (I2 = 95%). The prediction interval is −2.65 D to +1.55 D, indicating that the true effect size is within this range. Subgroup analysis could not be performed due to the small number of studies.
Comparison of NCR and CR in children under and over 5 years of age
In comparison between NCR and CR in the different age subgroups (Table 4), the test for overall effect is significant in children over 5 years of age (MD −0.73 D) 95% CI (−1.39 D, −0.06 D) (Z = 2.14, p = 0.03) but not at younger ages (MD −0.50 D) 95% CI (−1.16 D, +0.15 D) (Z = 1.51, p = 0.13). For under and over 5 years of age, the prediction interval is −3.60 D to +2.60 D and −3.25 D to +1.79 D respectively, indicating a wide range of true effect. The model suggests no differences between the age subgroups (Q = 0.22, df = 1, p = 0.64). The I2 value is high in both subgroups, indicating there is most likely another covariant causing heterogeneity.
TABLE 4.
Data from studies illustrated in the forest plot comparing the difference in non‐cycloplegic and cycloplegic autorefraction in relation to age
Sensitivity analysis
Following the removal of studies with a high risk of bias (Tables 5, 6, 7), there was no change in terms of significance when assessing the type of autorefractor under NC and C conditions (Q = 13.14, df = 2, p = 0.001). Neither was there a change when comparing between age subgroups (Q = 0.01, df = 1, p = 0.93).
TABLE 5.
Data from studies illustrated in the forest plot comparing non‐cycloplegic and cycloplegic autorefraction, excluding studies with high risk of bias
TABLE 6.
Data from studies illustrated in the forest plot comparing non‐cycloplegic autorefraction and cycloplegic retinoscopy with the exclusion of studies with high risk of bias
TABLE 7.
Data from studies illustrated in the forest plot comparing the difference in non‐cycloplegic and cycloplegic autorefraction in age, excluding studies with a high risk of bias
Comparisons between NC and C autorefractors, NC autorefractors and C retinoscopy and NCR and CR in children either under or over 5 years of age are shown in Tables 5, 6 and 7, respectively.
DISCUSSION
This systematic review and meta‐analysis aimed to determine the diagnostic agreement between NCR and CR in children. We found the mean effect size for the Righton Retinomax (righton‐oph.com) to be −1.17 D. A negative mean effect size indicates that this method provides a higher estimate of myopia under NC conditions. As the 95% confidence intervals for the NC Retinomax fall outside the intra‐examiner limits of repeatability for C retinoscopy (>1.00 D), these results suggest that the Retinomax autorefractor is an inaccurate method for measuring refractive error under NC conditions. Similar findings were found with the NC Canon (RK‐F1) autorefractor (canon.ca). These findings are in line with previous reports that under NC conditions autorefractors over‐ and under‐estimate myopia and hyperopia, respectively. 5 , 42 , 58 The confidence intervals suggest that under NC conditions, both autorefractors are inaccurate methods for measuring refractive error in children under and over 5 years of age and up to 12 years.
Meta‐analysis reveals that NC Plusoptix autorefraction showed reasonable agreement with C retinoscopy and the least variability in the refractive error measurements with a MD close to zero. Since the 95% confidence intervals fall within the intra‐examiner limits of repeatability for C retinoscopy (<1.00 D), the results suggest that the NC Plusoptix is an accurate method for examining refractive error. Binocular open‐field autorefractors were developed to avoid accommodation that a monocular closed‐field autorefractor would generate without cycloplegia. 59 Research has shown less myopic findings with a binocular open‐field instrument than a monocular closed‐field 60 autorefractor, which may explain the agreement between the NC Plusoptix and C retinoscopy. However, these findings should be interpreted with caution because they specify the accuracy of the mean effect for all of the five studies included, but not the variance of the effect size within the population as a whole. Therefore, results cannot be extrapolated to patients with clinical characteristics beyond the five studies included.
Estimating refractive error with eccentric infrared photorefraction depends on calibration of the luminance slopes in the pupil (conversion of the distribution of light reflected across the pupil to refractive error), and research has shown that better agreement can be found with low levels of refractive error and significant errors can arise as the refractive error increases. 61 These findings and the prediction interval suggest that the NC Plusoptix should not be used exclusively due to the potential variability being significantly larger than 1.00 D. The level of variability exceeded the limits of maximum acceptable difference, meaning that NC Plusoptix cannot be substituted for a CR. In addition, the total sample size for studies with the NC Plusoptix was small (n = 650), which could have potentially led to an underestimation of the difference between NCR and CR.
Our findings suggest that when compared with C retinoscopy, NC autorefraction is an inaccurate method of measuring refractive error, as it often results in an overestimate of myopia and a wide range of effect sizes. Comparison between NC retinoscopy and C autorefraction or C retinoscopy was not possible as the relevant data were not available from the included papers. Future studies should examine this relationship as practitioners may find it easier to obtain a measurement using a retinoscope because younger children like to move around. In addition, a practitioner has some control over accommodation during NC retinoscopy, unlike an autorefractor with its inherently inflexible design. Moreover, it has been reported that 80% of intra‐ and inter‐examiner repeatability for C retinoscopy falls within ±0.50 D. 42
Accommodation is known to affect the precision of refractive error assessment as active accommodation produces a negative shift in the measurements. Rosenfield and Benzoni found a considerable decline in the amplitude of accommodation (3.80 D) between the age of 5–10 years, suggesting that inaccuracy of NCR is particularly likely in younger children. 62 Most of the evidence on refraction in children <12 years old indicates that myopia is over estimated in NC compared with C autorefraction 5 , 41 , 48 with greater differences in younger children. 63 These findings suggest that younger children on average should have larger effect sizes with increased variability. Results from the current study indicate a smaller effect size with less variability in children <5 years of age than for those >5 years old. One possible explanation for this unusual finding is that the type of autorefractor also differed between the studies and therefore confounded the results. This possibility is supported by the high I2 value suggesting that there is significant heterogeneity between the groups (e.g., a range of different autorefractors). The prediction intervals were also extremely large in both groups indicating that the effect size is hard to predict due to the large amount of heterogeneity. Unfortunately, due to the low number of studies included, we were unable to carry out a meta‐regression to verify this.
As indicated above, an important finding of this review was the level of heterogeneity between the results of individual studies. The meta‐analysis quantified the degree to which the findings differed between studies, and showed inconsistencies and variation between them. Several factors in the individual study findings limit the extent to which we can accurately represent the evidence and explore reasons for heterogeneity. Unfortunately, many investigations averaged all types and levels of refractive error measurements into one overall figure. As a direct result, we were unable to investigate how types and levels of refractive error influenced the differences between NCR and CR. In addition, several studies included both eyes per child, which resulted in the clustering of intra‐individual data 64 making it impossible to extract data for just one eye. Finally, accommodation can influence refractive error measurements and potentially affect the agreement between NCR and CR. These could not be formally investigated because of the limitations of the reported data. Recent findings have shown that a clinically significant change in spherical equivalent (≥0.50 D) between NC and C autorefraction is more likely to occur in children who have a lag of accommodation <1.15 D. 12 Additional work is therefore required to create guidance on the application of cyclopentolate, and when it can be avoided.
In future studies, bias should be minimised by ensuring that all subjects, irrespective of their level and type of refractive error, are randomised into one arm of the trial. Studies should specify their cycloplegia regimen, including the interval between administration of drops and refraction. In addition, future studies should implement the assessment of accommodative function and pupil size both pre‐ and post‐instillation of cyclopentolate.
The strengths of this review include the methodological rigour used to conduct the review process which is consistent with best practice. 65 A comprehensive search strategy was used to identify as many potential studies for inclusion as possible, with no clinical setting, study design or publication year restriction. Two review authors independently screened all titles and abstracts. The data extraction and quality assessment for each study using the QUADAS‐2 Tool was conducted independently.
There are some limitations that are worth highlighting. Many studies had a high or unclear risk of bias in at least one domain with substantial heterogeneity observed between studies. This should be taken into consideration when interpreting the review findings. In addition, we were unable to conduct the planned sensitivity analysis. A lack of data from various degrees and types of refractive error might affect the overall estimate in this review. The range of refractive error was not always disclosed in the included publications, and therefore, the level of agreement between NCR and CR was unknown. Accordingly, the findings should be interpreted with caution until more data are available. The assumptions made from this review are restricted to the autorefractors included in this study. In addition, only the analysis of low to moderate hyperopia can be established. Agreement for hyperopia >5.00 D is yet to be explored. Moreover, analysis of the NC Plusoptix and C Plusoptix (plusoptix.com) could not be undertaken to establish a mean difference between these autorefractors with and without cycloplegia.
The present systematic review and meta‐analysis highlight substantial gaps in our knowledge of the accuracy of refracting young children without cycloplegia. Unfortunately, as most studies averaged different types and levels of refractive error, we were unable to determine whether cycloplegia is needed for all children or whether it can be safely administered only to children with specific types or levels of refractive error. Further quality research is needed to allow this analysis to be conducted. This could either be addressed by a large primary study or potentially via a meta‐analysis of individual patient data obtained from study authors. In conclusion, many different forms of autorefractors can be used to help evaluate refractive error objectively. However, CR is still recommended to ensure diagnostic accuracy in children younger than 12 years of age.
AUTHOR CONTRIBUTIONS
Salma Wilson: Conceptualization (equal); data curation (equal); formal analysis (equal); investigation (equal); methodology (equal); validation (equal); visualization (equal); writing – original draft (equal); writing – review and editing (equal). Irene Ctori: Conceptualization (equal); methodology (equal); supervision (equal); visualization (equal); writing – original draft (equal); writing – review and editing (equal). Rakhee Shah: Conceptualization (equal); methodology (equal); supervision (equal); visualization (equal); writing – original draft (equal); writing – review and editing (equal). Catherine Suttle: Conceptualization (equal); methodology (equal); supervision (equal); visualization (equal); writing – original draft (equal); writing – review and editing (equal). Miriam L. Conway: Conceptualization (equal); formal analysis (equal); investigation (equal); methodology (equal); supervision (equal); validation (equal); visualization (equal); writing – original draft (equal); writing – review and editing (equal).
CONFLICT OF INTEREST
The authors have no proprietary or commercial interest in any materials discussed in this article.
APPENDIX 1.
TABLE A1.
QUADAS‐2 signalling questions
Study: | ||
---|---|---|
Domain | QUADAS‐2 questions | Author's judgement |
Patient selection |
Did the study avoid inappropriate exclusions? (yes/no/unclear) |
|
Did the study avoid inappropriate inclusion? (yes/no/unclear) |
||
Were the same patient selection criteria used for those assigned to the non‐cycloplegic test? (yes/no/unclear) |
||
If the children received the non‐cycloplegic test, was the decision made before the children were recruited? (yes/no/unclear) |
||
Did the study avoid using prior tests as inclusion criteria that were correlated with the non‐cycloplegic refraction test? (yes/no/unclear) |
||
Risk of bias Could the selection of children have introduced bias? (yes/no/unclear) = (High risk/ low risk/ unclear risk) |
||
Concerns regarding applicability Are the concerns that the included children in the study do not match the review questions? (yes/no/unclear) = (High concern/low concern /unclear concern) |
||
Index test |
Were the non‐cycloplegic test results conducted and interpreted without knowledge of the results of the cycloplegic refraction? (yes/no/unclear) |
|
Did the study avoid using index test thresholds that are likely to advantage some of the index tests? (yes/no/unclear) |
||
Risk of bias Could the conduct or interpretation of the non‐cycloplegic refraction test have introduced bias? (yes/no/unclear) = (High risk/ low risk/ unclear risk) |
||
Concerns regarding applicability Are the concerns with the non‐cycloplegic test, its conduct, or interpretation differing from the review question? (yes/no/unclear) = (High concern/low concern /unclear concern) |
||
Reference standard |
Is the cycloplegic refraction likely to correctly classify the target condition? (yes/no/unclear) |
|
Were the cycloplegic refraction results conducted and interpreted without knowledge of the results of the non‐cycloplegic refraction? (yes/no/unclear) |
||
Risk of bias Could the cycloplegic refraction, its conduct, or its interpretation have introduced bias? (yes/no/unclear) = (High risk/ low risk/ unclear risk) |
||
Concerns regarding applicability Are there concerns that the target condition, as defined by cycloplegic refraction, does not match the review question? (yes/no/unclear) = (High concern/low concern /unclear concern) |
||
Flow and timing |
Was there an appropriate interval between the non‐cycloplegic refraction and cycloplegic refraction? (yes/no/unclear) |
|
Did all children receive cycloplegic refraction? (yes/no/unclear) |
||
Did all children receive the same cycloplegic refraction method? (yes/no/unclear) |
||
Were all children included in the analysis? (yes/no/unclear) |
||
Risk of bias Could the patient flow have introduced bias? (yes/no/unclear) = (High risk/ low risk/ unclear risk) |
Wilson S, Ctori I, Shah R, Suttle C, Conway ML. Systematic review and meta‐analysis on the agreement of non‐cycloplegic and cycloplegic refraction in children. Ophthalmic Physiol Opt. 2022;42:1276–1288. 10.1111/opo.13022
The protocol is registered on PROSPERO (https://www.crd.york.ac.uk/prospero/; registration number: CRD42020208201).
REFERENCES
- 1. National Institute for Health and Care Excellence . Cyclopentolate hydrochloride. Available from: https://bnfc.nice.org.uk/drug/cyclopentolate‐hydrochloride.html Accessed 25 May, 2020.
- 2. Patel T. Diagnostic drugs: part 2 – cycloplegics. Optician. 2015;250:12–8. [Google Scholar]
- 3. Farhood Q. Cycloplegic refraction in children with cyclopentolate versus atropine. J Clin Exp Ophthalmol. 2012;3:239. 10.4172/2155-9570.1000239 [DOI] [Google Scholar]
- 4. Hopkins S, Sampson G, Hendicott P, Lacherez P, Wood J. Refraction in children. Optom Vis Sci. 2012;89:1734–9. [DOI] [PubMed] [Google Scholar]
- 5. Fotedar R, Rochtchina E, Morgan I, Wang JJ, Mitchell P, Rose KA. Necessity of cycloplegia for assessing refractive error in 12‐year‐old children: a population‐based study. Am J Ophthalmol. 2007;144:307–9. [DOI] [PubMed] [Google Scholar]
- 6. Zhao J, Mao J, Luo R, Li F, Pokharel G, Ellwein L. Accuracy of noncycloplegic autorefraction in school‐age children in China. Optom Vis Sci. 2004;81:49–55. [DOI] [PubMed] [Google Scholar]
- 7. Major E, Dutson T, Moshirfar M. Cycloplegia in children: an optometrist's perspective. Clin Optom. 2020;12:129–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Sankaridurg P, He X, Naduvilath T, Lv M, Ho A, Smith E 3rd, et al. Comparison of noncycloplegic and cycloplegic autorefraction in categorizing refractive error data in children. Acta Ophthalmol. 2017;95:e633–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Rengstorff RH, Doughty CB. Mydriatic and cycloplegic drugs: a review of ocular and systemic complications. Am J Optom Physiol Opt. 1982;59:162–77. [PubMed] [Google Scholar]
- 10. Morgan I, Iribarren R, Fotouhi A, Grzybowski A. Cycloplegic refraction is the gold standard for epidemiological studies. Acta Ophthalmol. 2015;93:581–5. [DOI] [PubMed] [Google Scholar]
- 11. Zhu D, Wang Y, Yang X, Yang D, Guo K, Guo Y, et al. Pre‐ and postcycloplegic refractions in children and adolescents. PLoS One. 2016;11:e0167628. 10.1371/journal.pone.0167628 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Jin CC, Pei RX, Du P, Liu GH, Jin N, Liu L. Lag of accommodation predicts clinically significant change of spherical equivalents after cycloplegia. Int J Ophthalmol. 2021;14:1052–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Ruttum M, Smith R. Pediatric distress from cyclopentolate: can a better drop be formulated? Am Orthopt J. 1994;44:112–5. [Google Scholar]
- 14. Hu Y, Wu J, Lu T, Wu H, Sun W, Wang X, et al. Effect of cycloplegia on the refractive status of children: the Shandong children eye study. PLoS One. 2015;10:e0117482. 10.1371/journal.pone.0117482 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Sutherland MS, Young JD. Does instilling proxymetacaine before cyclopentolate significantly reduce stinging? The implications of paediatric cycloplegia. Br J Ophthalmol. 2001;85:244–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Sani RY, Hassan S, Habib SG, Ifeanyichukwu EP. Cycloplegic effect of atropine compared with cyclopentolate‐tropicamide combination in children with hypermetropia. Niger Med J. 2016;57:173–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Bartlett J. Administration of and adverse reactions to cycloplegic agents. Optom Vis Sci. 1978;55:227–33. [DOI] [PubMed] [Google Scholar]
- 18. Yilmaz I, Ozkaya A, Alkin Z, Ozbengi S, Yazici A, Demirok A. Comparison of the Plusoptix A09 and Retinomax K‐plus 3 with retinoscopy in children. J Pediatr Ophthalmol Strabismus. 2015;52:37–42. [DOI] [PubMed] [Google Scholar]
- 19. Llewellyn S, Khandelwal P, Glaze S, Thomas P, Dahlmann‐Noor A. Time to drop the phenylephrine from the paediatric cycloplegia protocol: informing practice through audit. Eye. 2018;33:337–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Hingorani M. The invisible ophthalmologist. Eye. 2019;33:1193–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Ebri A, Kuper H, Wedner S. Cost‐effectiveness of cycloplegic agents: results of a randomized controlled trial in Nigerian children. Invest Ophthalmol Vis Sci. 2007;48:1025–31. [DOI] [PubMed] [Google Scholar]
- 22. Royal College of Ophthalmologists . Guidelines for the management of strabismus in childhood. Available from: https://www.rcophth.ac.uk/standards‐publications‐research/clinical‐guidelines/ Accessed 4 March, 2020.
- 23. Chen AM, Erzurum SA, Chandler DL, Hercinovic A, Melia BM, Bhatt AR, et al. Overminus lens therapy for children 3 to 10 years of age with intermittent exotropia: a randomized clinical trial. JAMA Ophthalmol. 2021;139:464–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. College of Optometrists . Examining younger children. Available from: https://guidance.college‐optometrists.org/guidance‐contents/knowledge‐skills‐and‐performance‐domain/examining‐younger‐children/#open:74 Accessed 13 July, 2021.
- 25. Ilechie AA, Addo NA, Abraham CH, Owusu‐Ansah A, Annan‐Prah A. Accuracy of noncycloplegic refraction for detecting refractive errors in school‐aged African children. Optom Vis Sci. 2021;98:920–8. [DOI] [PubMed] [Google Scholar]
- 26. Kirschen D, Isenberg S. The effectiveness of an autorefractor with eye‐tracking capability in pediatric patients. J AAPOS. 2014;18:217–21. [DOI] [PubMed] [Google Scholar]
- 27. Sanfilippo P, Chu B, Bigault O, Kearns L, Boon M, Young T, et al. What is the appropriate age cut‐off for cycloplegia in refraction? Acta Ophthalmol. 2014;92:e458–62. [DOI] [PubMed] [Google Scholar]
- 28. Kearns LS, Hewitt AW, Ruddle JB, Bigault O, Staffieri SE, Sanfilliop P, et al. Up to what age is a cyclopleged refraction required? Results from the twins eye study Tasmania (TEST). Invest Ophthalmol Vis Sci. 2010;51:ARVO E‐Abstract 1716. [Google Scholar]
- 29. Payerols A, Eliaou C, Trezeguet V, Villain M, Daien V. Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia. BMC Ophthalmol. 2016;16:72. 10.1186/s12886-016-0247-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Paff T, Oudesluys‐Murphy A, Wolterbeek R, Swart‐van den Berg M, de Nie J, Tijssen E, et al. Screening for refractive errors in children: the plusoptiX S08 and the Retinomax K‐plus2 performed by a lay screener compared to cycloplegic retinoscopy. J AAPOS. 2010;14:478–83. [DOI] [PubMed] [Google Scholar]
- 31. Peterseim M, Papa C, Wilson M, Cheeseman E, Wolf B, Davidson J, et al. Photoscreeners in the pediatric eye office: compared testability and refractions on high‐risk children. Am J Ophthalmol. 2014;158:932–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Flaxman SR, Bourne R, Resnikoff S, Ackland P, Braithwaite T, Cicinelli MV, et al. Global causes of blindness and distance vision impairment 1990‐2020: a systematic review and meta‐analysis. Lancet Glob Health. 2017;5:e1221–34. [DOI] [PubMed] [Google Scholar]
- 33. Al‐Haddad C, Ismail K, Kheir W, Keaik M. Clinical profile and treatment outcomes of amblyopia across age groups. Middle East Afr J Ophthalmol. 2019;26:71–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Birch E, Holmes J. The clinical profile of amblyopia in children younger than 3 years of age. J AAPOS. 2010;14:494–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Von Noorden GK, Campos EC. Binocular vision and ocular motility. Theory and management of strabismus. St. Louis: Mosby; 2002. [Google Scholar]
- 36. Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA statement: an updated guideline for reporting systematic reviews. Br Med J. 2021;372:n71. 10.1136/bmj.n71 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Nordic Cochrane Centre, The Cochrane Collaboration . Review Manager 5 (Version 5) [computer program]. Available from: https://training.cochrane.org/online‐learning/core‐software‐cochrane‐reviews/revman/revman‐5‐download Accessed 25 May, 2020.
- 38. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS‐2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155:529–36. [DOI] [PubMed] [Google Scholar]
- 39. Whiting P, Weswood M, Rutjes A, Reitsma J, Bossuyt P, Kleijnen J. Evaluation of QUADAS, a tool for the quality assessment of diagnostic accuracy studies. BMC Med Res Methodol. 2006;6:9. 10.1186/1471-2288-6-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Macaskill P, Gatsonis C, Deeks JJ, Harbord RM, Takwoingi Y. Analysing and presenting results. In: Deeks JJ, Bossuyt PM, Gatsonis C, editors. Cochrane handbook for systematic reviews of diagnostic test accuracy version 1.0. Chichester: The Cochrane Collaboration; 2010. Available from: http://srdta.cochrane.org/ Accessed 26 May, 2020. [Google Scholar]
- 41. McAlinden C, Khadka J, Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt. 2011;31:330–8. [DOI] [PubMed] [Google Scholar]
- 42. Tipton E, Shuster J. A framework for the meta‐analysis of bland–Altman studies based on a limits of agreement approach. Stat Med. 2017;36:3621–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. McCullough S, Doyle L, Saunders K. Intra‐ and inter‐ examiner repeatability of cycloplegic retinoscopy among young children. Ophthalmic Physiol Opt. 2017;37:16–23. [DOI] [PubMed] [Google Scholar]
- 44. Doherty SE, Doyle LA, McCullough SJ, Saunders KJ. Comparison of retinoscopy results with and without 1% cyclopentolate in school‐aged children. Ophthalmic Physiol Opt. 2019;39:272–81. [DOI] [PubMed] [Google Scholar]
- 45. DerSimonian R, Laird N. Meta‐analysis in clinical trials. Control Clin Trials. 1986;7:177–88. [DOI] [PubMed] [Google Scholar]
- 46. PlusOptix . Products. Available from: https://www.plusoptix.com/products/vision‐screener/application Accessed 3 May, 2022.
- 47. Huedo‐Medina T, Sánchez‐Meca J, Marín‐Martínez F, Botella J. Assessing heterogeneity in meta‐analysis: Q statistic or I2 index? Psychol Methods. 2006;11:193–206. [DOI] [PubMed] [Google Scholar]
- 48. Smith G. Refraction and visual acuity measurements: what are their measurement uncertainties? Clin Exp Optom. 2006;89:66–72. [DOI] [PubMed] [Google Scholar]
- 49. Akil H, Keskin S, Çavdarli C. Comparison of the refractive measurements with hand‐held autorefractometer, table‐mounted autorefractometer and cycloplegic retinoscopy in children. Korean J Ophthalmol. 2015;29:178–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50. Demirci G, Arslan B, Özsütçü M, Eliaçık M, Gulkilik G. Comparison of photorefraction, autorefractometry and retinoscopy in children. Int Ophthalmol. 2014;34:739–46. [DOI] [PubMed] [Google Scholar]
- 51. Harvey E, Miller J, Dobson V, Tyszko R, Davis A. Measurement of refractive error in native American preschoolers: validity and reproducibility of autorefraction. Optom Vis Sci. 2000;77:140–9. [DOI] [PubMed] [Google Scholar]
- 52. Iuorno J, Grant W, Noël L. Clinical comparison of the Welch Allyn SureSight handheld autorefractor versus cycloplegic autorefraction and retinoscopic refraction. J AAPOS. 2004;8:123–7. [DOI] [PubMed] [Google Scholar]
- 53. Ozdemir O, Tunay ZO, Petricli IS, Acar DE, Erol MK. Comparison of non‐cycloplegic photorefraction, cycloplegic photorefraction and cycloplegic retinoscopy in children. Int J Ophthalmol. 2015;8:128–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Seymen Z, Vural E, Eris E, Vural A, Ogreden T, Aslan O, et al. Comparison of the latest handheld Autorefractokeratometer in patients younger than three years of age with Retinomax and PlusOptix. Beyoglu Eye J. 2019;4:32–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Vasudevan B, Ciuffreda K, Meehan K, Grk D, Cox M. Comparison of objective refraction in darkness to cycloplegic refraction: a pilot study. Clin Exp Optom. 2016;99:168–72. [DOI] [PubMed] [Google Scholar]
- 56. Won J, Shin H, Kim S, Lee Y. A comparison of the Plusoptix S09 with an autorefractometer of noncycloplegics and cycloplegics in children. Medicine. 2016;95:e4596. 10.1097/MD.0000000000004596 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Li L, Fu J, Chen W, Meng Z, Sun Y, Su H, et al. Difference of refractive status before and after cycloplegic refraction: the Lhasa childhood eye study. Jpn J Ophthalmol. 2021;65:526–36. [DOI] [PubMed] [Google Scholar]
- 58. Mallen EAH, Gilmartin B, Wolffsohn JS, Tsujimura SI. Clinical evaluation of the Shin‐Nippon SRW‐5000 autorefractor in adults: an update. Ophthalmic Physiol Opt. 2015;35:622–7. [DOI] [PubMed] [Google Scholar]
- 59. Carracedo G, Carpena‐Torres C, Batres L, Serramito M, Gonzalez‐Bergaz A. Comparison of two wavefront autorefractors: binocular open‐field versus monocular closed‐field. J Ophthalmol. 2020;2020:8580471. 10.1155/2020/8580471 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Bharadwaj SR, Sravani NG, Little JA, Narasaiah A, Wong V, Woodburn R, et al. Empirical variability in the calibration of slope‐based eccentric photorefraction. J Opt Soc Am. 2013;30:923–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Chen J, Xie A, Hou L, Su Y, Lu F, Thorn F. Cycloplegic and noncycloplegic refractions of Chinese neonatal infants. Invest Ophthalmol Vis Sci. 2011;14:2456–61. [DOI] [PubMed] [Google Scholar]
- 62. Rosenfield M, Benzoni JA. Amplitude of accommodation in children between 5 and 10 years of age. Invest Ophthalmol Vis Sci. 2007;48:ARVO E‐Abstract 975. [Google Scholar]
- 63. Fogel‐Levin M, Doron R, Wygnanski‐Jaffe T, Ancri O, Ben ZI. A comparison of plusoptiX A12 measurements with cycloplegic refraction. J AAPOS. 2016;20:310–4. 10.1016/j.jaapos.2016.04.006 [DOI] [PubMed] [Google Scholar]
- 64. Murdoch IE, Morris SS, Cousens SN. People and eyes: statistical approaches in ophthalmology. Br J Ophthalmol. 1998;82:971–3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65. Pollock M, Fernandes RM, Becker LA, Pieper D, Hartling L. Overviews of reviews. In: Higgins JPT, Thomas J, Chandler J, Cumpston M, Li T, Page MJ, et al., editors. Cochrane handbook for systematic reviews of interventions version 6.3 (updated February 2022). 2nd ed. Chichester: John Wiley & Sons; 2019. Available from: www.training.cochrane.org/handbook Accessed 28 May, 2022. [Google Scholar]