Abstract
Background and Purpose
The timing of decision‐making for a surgical intervention in patients with mild degenerative cervical myelopathy (DCM) is challenging. Spinal cord motion phase contrast MRI (PC‐MRI) measurements can reveal the extent of dynamic mechanical strain on the spinal cord to potentially identify high‐risk patients. This study aims to determine the comparability of axial and sagittal PC‐MRI measurements of spinal cord motion with the prospect of improving the clinical workup.
Methods
Sixty‐four DCM patients underwent a PC‐MRI scan assessing spinal cord motion. The agreement of axial and sagittal measurements was determined by means of intraclass correlation coefficients (ICCs) and Bland‐Altman analyses.
Results
The comparability of axial and sagittal PC‐MRI measurements was good to excellent at all cervical levels (ICCs motion amplitude: .810‐.940; p < .001). Significant differences between axial and sagittal amplitude values could be found at segments C3 and C4, while its magnitude was low (C3: 0.07 ± 0.19 cm/second; C4: −0.12 ± 0.30 cm/second). Bland‐Altman analysis showed a good agreement between axial and sagittal PC‐MRI scans (coefficients of repeatability: minimum −0.23 cm/second at C2; maximum −0.58 cm/second at C4). Subgroup analysis regarding anatomic conditions (stenotic vs. nonstenotic segments) and different velocity encoding (2 vs. 3 cm/second) showed comparable results.
Conclusions
This study demonstrates good comparability between axial and sagittal spinal cord motion measurements in DCM patients. To this end, axial and sagittal PC‐MRI are both accurate and sensitive in detecting pathologic cord motion. Therefore, such measures could identify high‐risk patients and improve clinical decision‐making (ie, timing of decompression).
Keywords: cervical stenosis, degenerative cervical myelopathy, phase contrast MRI, spinal cord compression, spinal cord oscillation, spinal cord motion
INTRODUCTION
Degenerative changes of the cervical spine may result in cervical spinal stenosis with consecutive spinal cord compression leading to the clinical imprint of degenerative cervical myelopathy (DCM). DCM is the most common cause of nontraumatic incomplete spinal cord injury. 1 , 2 While standard of care is decompressive surgery in DCM patients with moderate and severe impairments, treatment decisions (operative vs. conservative) in mild DCM are still challenging. 3 The clinical assessments such as modified Japanese Orthopaedic Association (mJOA) score are good ad hoc measures but are rather insensitive to reveal deterioration at an early stage of DCM and to identify patients at risk. 4 The pathophysiology of DCM is attributed to immediate (ie, direct or static) cord compression, spinal malalignment leading to altered cord tension, impaired vascular supply, and repeated dynamic injury. 5 , 6 , 7 , 8 Dynamic spinal cord injury is often narrowed to segmental hypermobility; however, cardiac‐related periodic cord motion may play a by far underestimated role in this pathophysiological consideration. The cervical spinal cord is subject to a physiological craniocaudal motion supposedly due to cardiac pulse wave dynamics, which can be assessed by phase contrast MRI (PC‐MRI). 9 , 10 Previous studies, applying 2‐dimensional PC‐MRI, demonstrated increased spinal cord motion at the level of a cervical spinal stenosis. 11 , 12 , 13 , 14 , 15 Interestingly, increased motion was associated with sensorimotor deficits 12 , 14 and impaired electrophysiological readouts. 14 , 15 Thus, altered spinal cord motion may be a potential biomarker of spinal cord dysfunction, identifying patients at risk for disease progression. However, to date no standardized PC‐MRI protocol has been established (plethora of readouts of spinal cord motion 12 , 14 , 15 , 16 , 17 ; axial 9 , 10 , 11 , 15 , 18 and sagittal PC‐MRI 12 , 16 , 17 , 19 ). Recently, sagittal phase contrast measurements 16 , 17 and the axial imaging 9 , 11 were reported to assess craniocaudal spinal cord motion. However, the agreement between these two approaches remains unclear. While axial measurements are adapted to the alignment of the spinal cord (slice orientation perpendicular to the spinal cord; Figure 1A), sagittal measurements may be confounded by a misalignment of the spinal cord in the sagittal field of view, causing a systematic measurement error.
This study therefore aims to determine the agreement of axial and sagittal PC‐MRI measurements of spinal cord motion with the prospect of improving the clinical workup of DCM patients. We hypothesized that sagittal spinal cord motion measurements may be confounded by the misalignment of the spinal cord within the field of view and can therefore not be as accurate as axial measurements that are aligned perpendicular to the spinal cord.
METHODS
Study design and population
This prospective, cross‐sectional, single‐center study recruited 64 DCM patients from the outpatient clinic of the University Spine Center, University Hospital Balgrist, Zurich between March 2019 and August 2021. Inclusion criteria were as follows: a cervical spinal stenosis on T2‐weighted (T2w) MRI; clinical symptoms consistent with DCM (ie, pain, sensory, or motor deterioration in the upper or lower limbs, gait, or bladder dysfunction); no other neurological disorder (identified by medical history and neurological examination); age 18‐80 years; and no MRI contraindications. Symptom severity and functional impairment were assessed with the mJOA score 20 and complementary clinical examinations (ie, a thorough standardized neurological examination according to the International Standards for Neurological Classification of Spinal Cord Injury including segmental sensory assessments and motor scores 21 ; the Graded and Redefined Assessment of Strength and Prehension for hand function 22 ). DCM was diagnosed in patients with at least one clinical symptom and sign of cervical myelopathy 23 complemented by corresponding stenosis in cervical spine MRI. Some patients in our population were rated with an mJOA score of 18, while symptoms consistent with DCM were observed in the complementary assessments.
Standard protocol approvals, registrations, and patient consents
This prospective study was approved by the local ethics committee (Kantonale Ethikkommission Zurich, KEK‐ZH 2012‐0343, BASEC Nr. PB_2016‐00623) and registered (www.clinicaltrials.gov; NCT 02170155). The work described has been carried out in accordance with The Code of Ethics of the World Medical Association (Declaration of Helsinki) for experiments involving humans. Informed consent was obtained from all participants prior to study enrolment. Study data were collected and managed using REDCap electronic data capture tools hosted at Balgrist University Hospital, Zurich, Switzerland. 24
Imaging
The MRI setup was supported by a radiologist (RS) and a physicist (MK). All subjects underwent a 3T MRI scan (MAGNETOM Prisma; Siemens Healthcare GmbH, Erlangen, Germany) including sagittal T2w and axial T2w MRI. Spinal cord motion was assessed with sagittal and axial PC‐MRI retrospectively cardiac gated using a finger clip (MRI parameters are listed in Table 1). The 2‐dimensional phase contrast sequence encoded craniocaudal spinal cord motion. The velocity encoding (venc) value was set to 2 or 3 cm/second (from April 2020) based on previous findings of cord motion. 9 , 12 , 13 , 14 , 15 , 18 Velocity signal was assessed within 20 time points during a cardiac cycle and 128 cardiac cycles were acquired per measurement. Slice orientation in axial PC‐MRI (measurement at the corresponding intervertebral disc level) was adjusted perpendicular to the spinal cord (Figure 1A); sagittal phase contrast measurements were placed midsagittal into the spinal cord (Figure 1D). While in sagittal PC‐MRI no adaption to the alignment of the spinal cord is possible for each individual cervical segment, it may be confounded by the misalignment of the cord to the vertical reference line within the field of view (Figure 2), which may cause a systematic measurement error. Total acquisition time was approximately 23 minutes.
TABLE 1.
Axial T2w | Sagittal T2w | Axial PC | Sagittal PC | |
---|---|---|---|---|
TE (ms) | 93 | 87 | 12.36 | 12.36 |
TR (ms) | 3600 | 3760 | 60.84 | 60.28 |
Slice thickness (mm) | 3 | 2.5 | 5 | 5 |
Flip angle (°) | 90/150 | 90/160 | 10 | 10 |
Field of view (mm) | 160 | 220 | 140 | 180 |
Bandwidth (Hz/px) | 284 | 260 | 355 | 355 |
Base resolution | 320 | 384 | 256 | 256 |
Phase resolution | 80% | 75% | 50% | 50% |
Spatial resolution (mm3) | 0.5 × 0.6 × 3.0 | 0.6 × 0.8 × 2.5 | 0.3 × 0.6 × 5.0 | 0.4 × 0.8 × 5.0 |
Parallel imaging technique | GRAPPA 2 | None | None | None |
Abbreviations: GRAPPA, GeneRalized Autocalibrating Partial Parallel Acquisition; TE, echo time; TR, repetition time; w, weighted.
Cervical segments were classified as “stenotic” or “nonstenotic.” A stenotic segment was defined as a loss of the cerebrospinal fluid (CSF) signal in axial T2w imaging ventral and dorsal to the spinal cord. Segments with visible CSF signal in axial T2w imaging ventral and/or dorsal to the spinal cord were defined as nonstenotic. In axial PC‐MRI in patients with a monosegmental stenosis, the C2 segment and the stenotic segment were measured. In patients with a multisegmental stenosis, the C2 segment and the most stenotic segment (defined as maximum spinal canal narrowing) were measured. The stenotic respectively the most stenotic segment was judged visually in T2w imaging by two investigators (NP, MH; consultant neurologists). In all patients, additional axial PC‐MRI measurements at as many as possible other stenotic (multisegmental stenosis) and nonstenotic (monosegmental and multisegmental stenosis) cervical segments were obtained. In sagittal PC‐MRI, all cervical segments were measured.
Imaging analysis
Image processing was performed using the Horos free DICOM viewer (www.horosproject.org) by two experienced investigators (NP, MH; consultant neurologists). Phase contrast images were visually controlled for artifacts prior to image processing. In axial phase contrast measurements, craniocaudal spinal cord motion was determined by a predefined ellipsoid‐shaped region of interest (30.56 mm2) mid‐centered into the spinal cord (Figure 1C). In sagittal measurements, a predefined round‐shaped region of interest (20.03 mm2) was centered on the spinal cord at the corresponding cervical intervertebral disc level (Figure 1E). Velocity calculation was conducted as reported previously. 9 , 11 PC‐MRI spinal cord motion readouts included the amplitude (maximum negative peak to maximum positive peak) of the velocity signal over the cardiac cycle. The misalignment angle α was measured between the spinal cord orientation at the segmental cervical disc level (perpendicular to the axial PC‐MRI slice; red dotted line; Figure 2) and the vertical reference line (orientation of the craniocaudal velocity measurement; green arrow; Figure 2) within the field of view.
Statistical analysis
Statistical analysis was conducted using SPSS (IBM SPSS Statistics for Windows, Version 26.0. Armonk, NY; IBM Corp). Metrics are reported as group mean ± standard deviation (SD). Statistical significance was set at <.05. Spinal cord motion values were compared in each cervical segment (C2 = segment C2/C3; C3 = segment C3/C4; C4 = segment C4/C5; C5 = segment C5/C6; C6 = segment C6/C7) separately, as previous work showed significant differences of spinal cord motion between the cervical segments. 9 , 16 At segment C7/T1, no comparison of axial and sagittal PC‐MRI was possible as no axial measurement was available.
Statistical analyses were conducted in the whole group (all collected measurements); additional subgroup analysis was performed with regard to the velocity encoding (2 vs. 3 cm/second) and in stenotic and nonstenotic segments separately. All data were tested for normal distribution by means of histograms and Shapiro‐Wilk test. If differences of axial minus sagittal spinal cord motion values were normally distributed (all measurements: amplitudes C4, C5, and C6; subgroup venc 2 cm/second: amplitudes C4, C5, and C6; subgroup venc 3 cm/second: amplitudes C2, C3, C5, and C6; subgroup nonstenotic segments: amplitudes C3 and C5; subgroup stenotic segments: amplitudes C4, C5, and C6), a one‐sample t‐test was used to test if the mean differences of the amplitude values between the two methods (ie, axial minus sagittal PC‐MRI value) were significantly different from zero. In not normally distributed data, a one‐sample Wilcoxon test was used. If the mean or median difference of a parameter was different from zero, there was no agreement between axial and sagittal PC‐MRI measurements. Comparability of axial and sagittal PC‐MRI was examined by intraclass correlation coefficients (ICCs; single measures, two‐way mixed effects model, absolute agreement) and Bland‐Altman analysis 25 to estimate limits of agreement between the two (ie, axial vs. sagittal PC‐MRI) methods. The ICC values were characterized as follows: “poor” for <.50, “moderate” for .50‐.74, “good” for .75‐.90, and “excellent” for >.90. 26 Between subgroup differences were evaluated with the Mann‐Whitney U test (amplitude differences; amplitude values; age; misalignment angle of the spinal cord) and Fisher´s exact test (number of stenotic segments; proportion of measurements in stenotic segments; site of the maximum spinal cord compression; sex). Correlations between the amplitude differences and the misalignment angle α and between the amplitude difference and the corresponding amplitude mean value were tested by calculation of Spearman‐Rho coefficients.
Data availability
The MRI data acquired and processed in the present study are available from the corresponding author upon request with the need for a formal data sharing agreement.
RESULTS
Patient demographics
Sixty‐four patients (40 (62.5%) male; age: 56.6 ± 14.7 years) were enrolled (Table 1). Twenty‐three patients (35.9%) presented with a single stenosis, and 41 (64.1%) had multiple stenotic segments (number of stenotic segments: 2.05 ± 1.02). mJOA score ranged from 11 to 18 (15.70 ± 2.14). Twenty‐five (39.1%) patients showed hyperintensity on T2w MRI consistent with myelopathy. In 64 patients with available sagittal phase contrast imaging (320 segmental measurements in segments C2‐C6), a total of 270 segmental axial PC‐MRI measurements were performed (Table 2). Five axial (1.9%) and 18 sagittal (5.6%) segmental PC‐MRI measurements were excluded from analysis due to artifacts (Table 2). Artifacts in PC‐MRI could be attributed to aliasing due to low velocity encoding in axial and sagittal measurements, to infolding artifacts of the shoulders in caudal cervical segments in axial measurements, and to unproper slice placement and motion artifacts (2 patients = 12 segmental datasets) in sagittal measurements. In total, 62 values of axial and sagittal measurements could be compared at segment C2, 43 at C3, 51 at C4, 54 at C5, and 41 at C6.
TABLE 2.
Basic demographics (N = 64) | |
---|---|
Sex (male) | 40 (62.5%) |
Age (years) | 56.6 ± 14.7 |
Monosegmental stenosis | 23 (35.9%) |
Multisegmental stenosis | 41 (64.1%) |
Number of stenotic segments | 2.05 ± 1.02 |
mJOA | 15.70 ± 2.14 |
T2w myelopathy | 25 (39.1%) |
Number of phase contrast measurements | |||
---|---|---|---|
Axial segment | N | Stenotic segment | Nonstenotic segment |
C2 | 64 | 0 | 64 |
C3 | 44 | 20 | 24 |
C4 | 57 | 32 (1) | 25 (1) |
C5 | 60 | 50 (2) | 10 |
C6 | 45 | 21 | 24 (1) |
Sagittal segment | N | Stenotic segment | Nonstenotic segment |
---|---|---|---|
C2 | 64 | 0 | 64 (2) |
C3 | 64 | 20 (1) | 44 (1) |
C4 | 64 | 31 (3) | 33 (2) |
C5 | 64 | 51 (2) | 13 (3) |
C6 | 64 | 23 (1) | 41 (3) |
Number of phase contrast measurements with regard to venc | ||||
---|---|---|---|---|
venc 2 | venc 3 | |||
Axial segment | Stenotic segment | Nonstenotic segment | Stenotic segment | Nonstenotic segment |
C2 | 0 | 29 | 0 | 35 |
C3 | 10 | 0 | 10 | 24 |
C4 | 14 (1) | 8 (1) | 18 | 17 |
C5 | 23 (2) | 2 | 27 | 8 |
C6 | 8 | 2 | 13 | 22 (1) |
venc 2 | venc 3 | |||
---|---|---|---|---|
Sagittal segment | Stenotic segment | Nonstenotic segment | Stenotic segment | Nonstenotic segment |
C2 | 0 | 29 (1) | 0 | 35 (1) |
C3 | 10 | 19 (1) | 10 (1) | 25 |
C4 | 15 (1) | 14 (1) | 18 (2) | 17 (1) |
C5 | 24 (2) | 5 | 27 | 8 (3) |
C6 | 10 | 19 (1) | 13 (1) | 22 (2) |
Note: Inside parenthesis is the number of measurements with artifacts.
Abbreviations: mJOA, modified Japanese Orthopaedic Association; N, number; venc, velocity encoding; w, weighted.
Spinal cord misalignment in sagittal PC‐MRI
Misalignment of the spinal cord to the vertical reference within the field of view in sagittal MRI (Figure 2), potentially causing a systematic measurement error, was most prominent at C6 (13.24 ± 8.27°), but less at segments C2 (6.43 ± 4.53°), C3 (7.32 ± 4.47°), C4 (8.26 ± 6.07°), and C5 (8.53 ± 7.10°). No correlations could be found between the amplitude differences and the misalignment angles in any segment (p > .169).
Differences of spinal cord motion values between axial and sagittal measurements
Segmental axial and sagittal amplitude values are illustrated in Table 3 (group mean) and Figure 3 (all measurements; individual values). Over all measurements (venc 2 and 3 cm/second; stenotic and nonstenotic segments), the test of the median or mean differences of the amplitude values (axial minus sagittal PC‐MRI values) showed significant differences to zero at the cervical segments C3 (0.07 ± 0.19 cm/second; p = .001) and C4 (−0.12 ± 0.30 cm/second; p = .008; Table 3).
TABLE 3.
Axial | Sagittal | Difference Axial – Sagittal | Coefficient of repeatability | Intraclass correlation coefficients | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Segment | N | Mean | SD | Mean | SD | Mean | SD | 1.96 × SD | p | ICC | CImin | CImax | p |
C2 | 62 | 0.51 | 0.28 | 0.48 | 0.25 | 0.02 | 0.12 | 0.23 | .186 | .902 | .842 | .940 | <.001 |
C3 | 43 | 0.83 | 0.56 | 0.76 | 0.55 | 0.07 | 0.19 | 0.37 | .001 | .934 | .871 | .966 | <.001 |
C4 | 51 | 0.83 | 0.48 | 0.94 | 0.53 | –0.12 | 0.30 | 0.58 | .008 | .810 | .669 | .892 | <.001 |
C5 | 54 | 1.05 | 0.54 | 1.04 | 0.53 | 0.01 | 0.19 | 0.36 | .676 | .940 | .899 | .965 | <.001 |
C6 | 41 | 0.89 | 0.46 | 0.88 | 0.41 | 0.02 | 0.22 | 0.43 | .600 | .876 | .779 | .932 | <.001 |
Note: The motion amplitude values (all measurements) are reported per cervical segment for axial and sagittal measurements. Additionally, the difference of axial minus sagittal values for each cervical segment is shown. The amplitude difference between axial and sagittal measurements was significantly different from 0 at segment C3 and C4. Intraclass correlation coefficients (ICCs; two‐way mixed model, single measures, absolute agreement) with its 95% confidence interval (CI) are presented for each cervical segment.
Abbreviations: N, number of measurements; SD, standard deviation.
While results in the whole patient group may be confounded by different measurement techniques (ie, different venc used for measurements) and anatomic conditions (ie, measurements in stenotic vs. nonstenotic segments), subgroup analyses were conducted.
In a subgroup analysis with regard to the different venc used for the measurements, no significant amplitude differences could be found in measurements with venc 2 cm/second (p > .165; Table 4). In venc 3 cm/second measurements, amplitude differences were significantly different from 0 in segments C3 (0.07 ± 0.11 cm/second; p = .001) and C4 (−0.24 ± 0.25 cm/second; p < .001; Table 4).
TABLE 4.
Velocity encoding 2 cm/second | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Axial | Sagittal | Bland‐Altman coefficients | Intraclass correlation coefficients | ||||||||||
Segment | N | Mean | SD | Mean | SD | Mean | SD | 1.96 × SD | p | ICC | CImin | CImax | p |
C2 | 28 | 0.54 | 0.33 | 0.50 | 0.27 | 0.04 | 0.14 | 0.28 | .362 | .882 | .762 | .943 | <.001 |
C3 | 10 | 0.97 | 0.64 | 0.88 | 0.62 | 0.10 | 0.36 | 0.70 | .285 | .843 | .509 | .958 | .001 |
C4 | 19 | 1.05 | 0.59 | 0.96 | 0.58 | 0.09 | 0.26 | 0.52 | .165 | .894 | .749 | .958 | <.001 |
C5 | 22 | 1.05 | 0.53 | 1.03 | 0.49 | 0.02 | 0.21 | 0.42 | .638 | .915 | .808 | .964 | <.001 |
C6 | 10 | 1.00 | 0.25 | 0.96 | 0.24 | 0.04 | 0.19 | 0.36 | .559 | .732 | .235 | .926 | .006 |
Velocity encoding 3 cm/second | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Axial | Sagittal | Bland‐Altman coefficients | Intraclass correlation coefficients | ||||||||||
Segment | N | Mean | SD | Mean | SD | Mean | SD | 1.96 × SD | p | ICC | CImin | CImax | p |
C2 | 34 | 0.47 | 0.22 | 0.46 | 0.24 | 0.01 | 0.09 | 0.17 | .358 | .931 | .867 | .965 | <.001 |
C3 | 33 | 0.79 | 0.53 | 0.72 | 0.53 | 0.07 | 0.11 | 0.21 | .001 | .973 | .914 | .989 | <.001 |
C4 | 32 | 0.70 | 0.34 | 0.93 | 0.51 | –0.24 | 0.25 | 0.48 | <.001 | .734 | .167 | .899 | <.001 |
C5 | 32 | 1.05 | 0.55 | 1.04 | 0.56 | 0.00 | 0.17 | 0.33 | .921 | .956 | .912 | .978 | <.001 |
C6 | 31 | 0.86 | 0.51 | 0.85 | 0.45 | 0.01 | 0.23 | 0.45 | .767 | .887 | .780 | .944 | <.001 |
Note: The motion amplitude values for the subgroups with venc 2 cm/second (upper table) and venc 3 cm/second (lower table) measurements are reported per cervical segment for axial and sagittal measurements. The amplitude difference of axial minus sagittal measurements was significantly different from 0 at segment C3 and C4 in the venc 3 cm/second subgroup. Intraclass correlation coefficients (ICCs; two‐way mixed model, single measures, absolute agreement) with its 95% confidence interval (CI) are presented for each cervical segment.
Abbreviations: N, number of measurements; SD, standard deviation.
Additionally, amplitude differences at the C4 level in the subgroup with venc 3 cm/second measurements were higher compared to the subgroup with venc 2 cm/second (p < .001), while no significant disagreement could be observed between subgroups in other segments (C2: p = .832; C3: p = .600; C5: p = .481; C6: p = .754). There was no significant difference at the C4 level between subgroups with venc 2 cm/second and venc 3 cm/second measurements in terms of age (52.2 ± 17.8 vs. 57.1 ± 12.4 years; p = .354), sex (36.8% vs. 40.6% female; p = 1.000), the proportion of measurements in stenotic segments (68.4% vs. 50.0%; p = .250), the number of stenotic segments within the patient (2.3 ± 1.0 vs. 2.0 ± 1.0; p = .830), the misalignment angle of the spinal cord (10.0 ± 6.7° vs. 7.2 ± 5.5°; p = .071), and the site of the maximum spinal cord compression (venc 2 cm/second: C3, 21.1%; C4, 31.6%; C5, 47.4%; C6, 0.0% vs. venc 3 cm/second: C3, 15.6%; C4, 18.8%; C5, 50.0%; C6, 15.6%; p = .292), but for amplitude values in axial measurements (Table 4; p = .031).
In nonstenotic segments, a significant difference from 0 was found for the amplitude differences at segments C3 (0.06 ± 0.10 cm/second; p = .010) and C4 (−0.14 ± 0.27 cm/second; p ≤ .010), while this could be observed in stenotic segments for the amplitude at C3 only (0.10 ± 0.27 cm/second; p = .049; Table 5). Additionally, amplitude differences between measurements in stenotic and nonstenotic segments did not show a significant disagreement at any cervical segment (C2: no stenotic segments; C3: p = .732; C4: p = .287; C5: p = .389; C6: p = .130).
TABLE 5.
Nonstenotic segments | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Axial | Sagittal | Bland‐Altman coefficients | Intraclass correlation coefficients | ||||||||||
Segment | N | Mean | SD | Mean | SD | Mean | SD | 1.96 × SD | p | ICC | CImin | CImax | p |
C2 | 62 | 0.51 | 0.28 | 0.48 | 0.25 | 0.02 | 0.12 | 0.23 | .186 | .902 | .842 | .940 | <.001 |
C3 | 24 | 0.63 | 0.24 | 0.58 | 0.23 | 0.06 | 0.10 | 0.19 | .010 | .896 | .719 | .958 | <.001 |
C4 | 22 | 0.69 | 0.34 | 0.83 | 0.40 | –0.14 | 0.27 | 0.53 | .004 | .687 | .353 | .861 | <.001 |
C5 | 7 | 0.80 | 0.48 | 0.82 | 0.43 | –0.02 | 0.07 | 0.13 | .378 | .990 | .948 | .998 | <.001 |
C6 | 21 | 0.74 | 0.41 | 0.76 | 0.40 | –0.02 | 0.20 | 0.39 | .903 | .883 | .734 | .951 | <.001 |
Stenotic segments | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Axial | Sagittal | Bland‐Altman coefficients | Intraclass correlation coefficients | ||||||||||
Segment | N | Mean | SD | Mean | SD | Mean | SD | 1.96 × SD | p | ICC | CImin | CImax | p |
C2 | 0 | na | Na | na | na | na | na | na | na | na | na | na | na |
C3 | 19 | 1.09 | 0.72 | 0.99 | 0.73 | 0.10 | 0.27 | 0.52 | .049 | .928 | .822 | .972 | <.001 |
C4 | 29 | 0.94 | 0.54 | 1.03 | 0.61 | –0.09 | 0.31 | 0.62 | .115 | .842 | .691 | .923 | <.001 |
C5 | 47 | 1.08 | 0.54 | 1.07 | 0.54 | 0.02 | 0.20 | 0.39 | .587 | .934 | .884 | .962 | <.001 |
C6 | 20 | 1.06 | 0.47 | 1.00 | 0.39 | 0.06 | 0.23 | 0.46 | .277 | .849 | .663 | .937 | <.001 |
Note: The motion amplitude values in subgroups with measurements in nonstenotic (upper table) and stenotic (lower table) segments are reported per cervical segment for axial and sagittal measurements. The amplitude difference between axial minus sagittal measurements was significantly different from 0 at segment C3 and C4 in nonstenotic segments and at segment C3 in stenotic segments. Intraclass correlation coefficients (ICCs; two‐way mixed model, single measures, absolute agreement) with its 95% confidence interval (CI) are presented for each cervical segment.
Abbreviations: N, number of measurements; SD = standard deviation.
Agreement of axial and sagittal measurements
Over all measurements (venc 2 and 3 cm/second; stenotic and nonstenotic segments), ICCs of axial and sagittal amplitude values showed good to excellent values in all segments (.810‐.940; p < .001; Table 3). The Bland‐Altmann readouts, that is, the measurement bias and limits of agreement, of the amplitude values in each cervical segment can be found in Table 3. The coefficients of repeatability (±1.96 SD of the mean difference between tests) ranged from 0.23 cm/second at segment C2 to 0.58 cm/second at C4. A summary of comparability analyses, that is, ICCs and Bland–Altman analysis, can be found in Table 3 and Figure 4.
In a subgroup analysis with regard to the different venc used in measurements, the ICCs ranged from .732 to .915 in measurements with venc 2 cm/second and from .734 to .973 in measurements with venc 3 cm/second (Table 4). The coefficients of repeatability in venc 2 cm/second measurements ranged from 0.28 cm/second at segment C2 to 0.70 cm/second at C3 (Table 4; Figure 5). In venc 3 cm/second measurements, it ranged from 0.17 cm/second at segment C2 to 0.48 cm/second at C4 (Table 4; Figure 5).
In measurements in nonstenotic segments, ICCs ranged from .687 to 0.990, and in stenotic segments from .842 to .934 (Table 5). The coefficients of repeatability in measurements in nonstenotic segments ranged from 0.13 cm/second at segment C5 to 0.53 cm/second at C4 (Table 5; Figure 6). In stenotic segments, it ranged from 0.39 cm/second at segment C5 to 0.62 cm/second at C4 (Table 5; Figure 6).
Amplitude differences with regard to the amplitude mean
A correlation between the segmental amplitude mean values and the differences of the amplitude values was found at C2 (r = .256; p = .045) and a trend for a correlation at segment C6 (r = .284; p = .072).
DISCUSSION
This study demonstrates a good comparability and agreement between axial and sagittal spinal cord motion PC‐MRI in DCM patients. A possible measurement error in sagittal PC‐MRI through the spinal cord misalignment to the vertical reference line within the field of view appears to be negligible. However, a potential future clinical application would greatly benefit from a standardized MRI protocol and postprocessing pipeline, thus improving comparability of findings across different clinical centers.
Different techniques and numerous readouts measuring spinal cord motion reported in previous studies challenge the comparability between sites. 9 , 10 , 11 , 12 , 13 , 14 , 15 , 16 , 17 , 18 , 19 , 27 , 28 , 29 , 30 In our study, we decided to analyze the agreement of the motion amplitude between axial and sagittal measurements as an easy‐to‐collect readout with no need for a phase drift correction. Varying phase drift was reported across different scanners 31 and also over time in the same scanner, 32 potentially limiting the use of readouts with need for a phase drift correction like the area under the curve of the motion signal. To our knowledge, reproducibility of phase contrast measurements across different scanners was rarely analyzed, 33 where a substantial agreement but also a significant bias could be observed across different readouts of pulmonary blood flow measurements. Hence, the agreement for different spinal cord motion readouts has to be further evaluated across scanners. Importantly, the PC‐MRI postprocessing for velocity calculations should be automated as far as possible, but also easy to be transferred and introduced across different clinical centers. Wolf et al. reported an evaluation technique using a 3‐dimensional hierarchical deep convolutional neural network (CNN) for anatomical segmentation and spinal cord motion measurements. 16 , 17 This approach appears promising, as motion evaluation could be conducted automatically after appropriate training of the CNN. However, it may be challenging to widely transfer this technique to other centers in order to foster clinical application. Our group reported a manual method, using free available DICOM tools and manual calculation of velocity values. 9 , 11 This approach could be easily introduced at other centers, but remains more time consuming and lacks further automatization. Reliability between different evaluation techniques (ie, automated vs. manual calculation) is not known. To further promote the implementation of spinal cord motion PC‐MRI measurements within the clinical setting, comparison of different postprocessing techniques (ie, CNN vs. manual evaluation) between sites is mandatory. Furthermore, a proper venc in PC‐MRI is important for spinal cord motion measurements, adapted as far as possible to the expected velocities. 34 , 35 While improper low venc causes aliasing artifacts, noise in velocity images increases with higher venc, probably leading to measurement errors. 36 In subgroup analyses with regard to different venc used in our measurements, a significant difference between axial and sagittal amplitude values could only be found at C3 and most obvious at C4 in venc 3 cm/second measurements and amplitude differences were higher in venc 3 cm/second compared to venc 2 cm/second measurements at C4. An increasing measurements error with higher velocity encoding was previously reported, 36 while this was relevant with an extensive higher venc only (ie, a deviation of more than 10% if venc increases by more than three times the velocity in the vessel). However, the magnitude of differences was low and aliasing artifacts could be observed in 5.6% of patients using a velocity encoding of 2 cm/second. In principle, aliasing artifacts within PC‐MRI measurements could be corrected, 37 but this leads to more complicated and time‐consuming postprocessing of the measurements. The question now turns to, whether it is advisable to choose a venc of 3 cm/second or a venc of 2 cm/second for spinal cord motion measurements. Considering the pros and cons, we recommend a velocity encoding of 3 cm/second for future measurements, where no further aliasing was noticed and the magnitude of differences observed appears to be acceptable for clinical use. However, venc may be further adjusted to an optimal value (eg, 2.5 cm/second) within establishment of a “gold standard” protocol for clinical implementation.
Previous studies reported excellent interrater and test‐retest reliability of spinal cord motion measurements with axial and sagittal PC‐MRI. 9 , 11 , 15 , 17 In addition, in our study we demonstrate a good to excellent 26 comparability and agreement of these two different PC‐MRI measurement techniques (ie, axial and sagittal phase contrast imaging) of spinal cord motion in DCM patients. We hypothesized that a difference of velocity values in sagittal measurements compared to axial values might be mostly attributed to a systematic measurement error due to the misalignment of the spinal cord within the field of view. Significant differences between axial and sagittal amplitude values could only be found at C3 and C4, but not in segments C5 and C6 with higher spinal cord misalignment. Additionally, no correlation between the misalignment angle and the amplitude differences could be found, underlining the weak effect of this potential confounder. Thus, effect size of a systematic measurement error due to spinal cord misalignment within the field of view is supposed to be weak as values of sagittal measurements showed a good agreement to axial data without the need for a correction. The reason for less agreement between axial and sagittal amplitude values in segments C3 and C4 compared to other segments remains unexplained. With regard to the segmental anatomic conditions (ie, measurements in stenotic vs. nonstenotic segments), subgroup analysis showed a good to excellent comparability between axial and sagittal values in both subgroups. In previous reports, reliability of spinal cord motion measurements was not reported for stenotic and nonstenotic segments separately. Interestingly, measures of reliability in our population showed comparable to even better values of reliability (ie, higher ICC values) in stenotic segments and no disagreement of differences could be found between the subgroups with measurements in stenotic and nonstenotic segments. While the spinal cord is subject not only to a craniocaudal oscillation, but also to anterior‐posterior and right‐left motion, 38 measurements in stenotic segments may be more reliable due to narrowed anatomic conditions. Restricted anterior‐posterior motion may limit potential partial volume effects between the spinal cord and CSF within the cardiac cycle reducing measurements errors. A weak correlation between the amplitude difference and the magnitude of the amplitude mean could be found at segment C2 only. While overall the magnitude of differences between axial and sagittal motion values appears acceptable for future clinical application, in a few measurements relevant differences could be observed, underlining the importance of a standardized measurement protocol to further improve comparability. In summary, phase contrast measurements of spinal cord motion have potential to be applied in the clinical settings, but pros and cons of different evaluation techniques have to be considered.
Sagittal PC‐MRI can assess the entire cervical spine in one PC‐MRI measurement. A potential systematic measurement error in sagittal PC‐MRI through the spinal cord misalignment within the field of view (Figure 1) could be shown to be negligible. In contrast, axial measurements might be more precise, accounting for the spinal cord misalignment by adapting the measurement slice perpendicular to the spinal cord (Figure 1A), but are more time consuming, as each cervical segment has to be analyzed separately (ie, causing longer MRI measurement time, difficult to implement in clinical protocols). Sagittal phase contrast measurements appear as a good compromise between accuracy (good agreement to axial values; good interrater and test‐retest reliability 17 ; no need for a correction for the spinal cord misalignment) and feasibility (ie, shorter measurement time compared to separate axial measurements in each cervical segment) within the clinical setting.
This study has few limitations. As no analysis was possible at segment C7, where the maximum spinal cord misalignment to the vertical reference line was observed, a relevant systematic measurement error in segments with higher grade of spinal cord misalignment could not be excluded. However, spinal canal stenosis in these segments is rare and measurements in segment C6 (also considerable spinal cord misalignment) showed a good agreement. Another limitation of our study is based on the use of only one MRI scanner. Scanners from different MRI vendors may influence comparability across measurements. Additionally, further insights of reliability between different PC‐MRI protocols and postprocessing techniques across centers are required to define a “gold standard” and to foster its clinical application in the diagnostic workup of DCM.
In conclusion, the confident assessment of spinal cord damage in DCM patients remains an unmet clinical need. There is evidence to support the use of spinal cord motion surrogate parameters in order to determine the degree of dynamic mechanical stress precipitating spinal cord damage. To this end, axial and sagittal PC‐MRI are both accurate and sensitive in detecting pathologic cord motion. Therefore, such measures could identify high‐risk patients and improve clinical decision‐making (eg, timing of decompressive surgery).
ACKNOWLEDGEMENTS AND DISCLOSURES
We would like to thank all patients for participation in this study. We thank Regula Schuepbach and Nathalie Kuehne for the excellent organization of the examinations and the management of the RedCap database. We also would like to thank the MRI team, that is, Natalie Hinterholzer, Alexandra Conte, and Zoe Volkart for conduction of the MRI measurements. Imaging was partly performed with support of the Swiss Center for Musculoskeletal Imaging, SCMI, Balgrist Campus AG, Zurich. Open access funding provided by Universitat Zurich.
All authors do not disclose commercial considerations in the past 2 years, such as equity interests, patent rights, or corporate affiliations including director roles, stock ownership, consultancy, speaking fees, and research support, for any product or process relevant to the submission. Balgrist University Hospital and Balgrist Campus have an academic research agreement with Siemens Healthineers.
Pfender N, Rosner J, Zipser CM, Friedl S, Vallotton K, Sutter R, et al. Comparison of axial and sagittal spinal cord motion measurements in degenerative cervical myelopathy. J Neuroimaging. 2022;32:1121–1133. 10.1111/jon.13035
REFERENCES
- 1. Zipser CM, Margetis K, Pedro KM, et al. Increasing awareness of degenerative cervical myelopathy: a preventative cause of non‐traumatic spinal cord injury. Spinal Cord. 2021;59:1216‐8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Kalsi‐Ryan S, Karadimas SK, Fehlings MG. Cervical spondylotic myelopathy: the clinical phenomenon and the current pathobiology of an increasingly prevalent and devastating disorder. Neuroscientist. 2013;19:409‐21. [DOI] [PubMed] [Google Scholar]
- 3. Fehlings MG, Kwon BK, Tetreault LA. Guidelines for the management of degenerative cervical myelopathy and spinal cord injury: an introduction to a focus issue. Global Spine J. 2017;7:6S‐7S. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Tetreault L, Nouri A, Kopjar B, et al. The minimum clinically important difference of the modified Japanese Orthopaedic Association scale in patients with degenerative cervical myelopathy. Spine. 2015;40:1653‐9. [DOI] [PubMed] [Google Scholar]
- 5. Karadimas SK, Moon ES, Yu WR, et al. A novel experimental model of cervical spondylotic myelopathy (CSM) to facilitate translational research. Neurobiol Dis. 2013;54:43‐58. [DOI] [PubMed] [Google Scholar]
- 6. Karadimas SK, Gatzounis G, Fehlings MG. Pathobiology of cervical spondylotic myelopathy. Eur Spine J. 2015;24:132‐8. [DOI] [PubMed] [Google Scholar]
- 7. Karadimas SK, Klironomos G, Papachristou DJ, et al. Immunohistochemical profile of NF‐kappaB/p50, NF‐kappaB/p65, MMP‐9, MMP‐2, and u‐PA in experimental cervical spondylotic myelopathy. Spine. 2013;38:4‐10. [DOI] [PubMed] [Google Scholar]
- 8. Beattie MS, Manley GT. Tight squeeze, slow burn: inflammation and the aetiology of cervical myelopathy. Brain. 2011;134:1259‐61. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Hupp M, Vallotton K, Brockmann C, et al. Segmental differences of cervical spinal cord motion: advancing from confounders to a diagnostic tool. Sci Rep. 2019;9:7415. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Mikulis DJ, Wood ML, Zerdoner OA, et al. Oscillatory motion of the normal cervical spinal cord. Radiology. 1994;192:117‐21. [DOI] [PubMed] [Google Scholar]
- 11. Hupp M, Pfender N, Vallotton K, et al. The restless spinal cord in degenerative cervical myelopathy. AJNR Am J Neuroradiol. 2021;42:597‐609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Chang HS, Nejo T, Yoshida S, et al. Increased flow signal in compressed segments of the spinal cord in patients with cervical spondylotic myelopathy. Spine. 2014;39:2136‐42. [DOI] [PubMed] [Google Scholar]
- 13. Tanaka H, Sakurai K, Iwasaki M, et al. Craniocaudal motion velocity in the cervical spinal cord in degenerative disease as shown by MR imaging. Acta Radiol. 1997;38:803‐9. [DOI] [PubMed] [Google Scholar]
- 14. Vavasour IM, Meyers SM, Macmillan EL, et al. Increased spinal cord movements in cervical spondylotic myelopathy. Spine J. 2014;14:2344‐54. [DOI] [PubMed] [Google Scholar]
- 15. Wolf K, Hupp M, Friedl S, et al. In cervical spondylotic myelopathy spinal cord motion is focally increased at the level of stenosis: a controlled cross‐sectional study. Spinal Cord. 2018;56:769‐76. [DOI] [PubMed] [Google Scholar]
- 16. Wolf K, Reisert M, Beltran SF, et al. Spinal cord motion in degenerative cervical myelopathy: the level of the stenotic segment and gender cause altered pathodynamics. J Clin Med. 2021;10:3788. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Wolf K, Reisert M, Beltran SF, et al. Focal cervical spinal stenosis causes mechanical strain on the entire cervical spinal cord tissue ‐ a prospective controlled, matched‐pair analysis based on phase‐contrast MRI. Neuroimage Clin. 2021;30:102580. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Tanaka H, Sakurai K, Kashiwagi N, et al. Transition of the craniocaudal velocity of the spinal cord: from cervical segment to lumbar enlargement. Invest Radiol. 1998;33:141‐5. [DOI] [PubMed] [Google Scholar]
- 19. Levy LM, Di Chiro G, Mccullough DC, et al. Fixed spinal cord: diagnosis with MR imaging. Radiology. 1988;169:773‐8. [DOI] [PubMed] [Google Scholar]
- 20. Kato S, Oshima Y, Oka H, et al. Comparison of the Japanese Orthopaedic association (JOA) score and modified JOA (mJOA) score for the assessment of cervical myelopathy: a multicenter observational study. PLoS ONE. 2015;10:e0123022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Rupp R, Biering‐Sorensen F, Burns SP, et al. International standards for neurological classification of spinal cord injury: revised 2019. Top Spinal Cord Inj Rehabil. 2021;27:1‐22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Velstra I‐M, Fellinghauer C, Abel R, et al. The graded and redefined assessment of strength, sensibility, and prehension version 2 provides interval measure properties. J Neurotrauma. 2018;35:854‐63. [DOI] [PubMed] [Google Scholar]
- 23. Fehlings MG, Badhiwala JH, Ahn H, et al. Safety and efficacy of riluzole in patients undergoing decompressive surgery for degenerative cervical myelopathy (CSM‐Protect): a multicentre, double‐blind, placebo‐controlled, randomised, phase 3 trial. Lancet Neurol. 2021;20:98‐106. [DOI] [PubMed] [Google Scholar]
- 24. Harris PA, Taylor R, Thielke R, et al. Research electronic data capture (REDCap)–a metadata‐driven methodology and workflow process for providing translational research informatics support. J Biomed Inform. 2009;42:377‐81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Martin Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307‐10. [PubMed] [Google Scholar]
- 26. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155‐63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Winklhofer S, Schoth F, Stolzmann P, et al. Spinal cord motion: influence of respiration and cardiac cycle. Rofo. 2014;186:1016‐21. [DOI] [PubMed] [Google Scholar]
- 28. Figley CR, Stroman PW. Investigation of human cervical and upper thoracic spinal cord motion: implications for imaging spinal cord structure and function. Magn Reson Med. 2007;58:185‐9. [DOI] [PubMed] [Google Scholar]
- 29. Enzmann DR, Pelc NJ. Brain motion: measurement with phase‐contrast MR imaging. Radiology. 1992;185:653‐60. [DOI] [PubMed] [Google Scholar]
- 30. Matsuzaki H, Wakabayashi K, Ishihara K, et al. The origin and significance of spinal cord pulsation. Spinal Cord. 1996;34:422‐6. [DOI] [PubMed] [Google Scholar]
- 31. Minderhoud SCS, Van Der Velde N, Wentzel JJ, et al. The clinical impact of phase offset errors and different correction methods in cardiovascular magnetic resonance phase contrast imaging: a multi‐scanner study. J Cardiovasc Magn Reson. 2020;22:68. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Gatehouse PD, Rolf MP, Bloch KM, et al. A multi‐center inter‐manufacturer study of the temporal stability of phase‐contrast velocity mapping background offset errors. J Cardiovasc Magn Reson. 2012;14:72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Iraha R, Tsuchiya N, Yamashiro T, et al. Reproducibility of pulmonary blood flow measurements by phase‐contrast MRI using different 1.5 T MR scanners at two institutions. Acta Radiol Open. 2017;6:2058460116684370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Korbecki A, Zimny A, Podgórski PA, et al. Imaging of cerebrospinal fluid flow: fundamentals, techniques, and clinical applications of phase‐contrast magnetic resonance imaging. Pol J Radiol. 2019;84:240‐50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Villa G, Ringgaard S, Hermann I, et al. Phase‐contrast magnetic resonance imaging to assess renal perfusion: a systematic review and statement paper. MAGMA 2020;33:3‐21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Lotz J, Meier C, Leppert A, et al. Cardiovascular flow measurement with phase‐contrast MR imaging: basic facts and implementation. Radiographics. 2002;22:651‐71. [DOI] [PubMed] [Google Scholar]
- 37. Yang GZ, Burger P, Kilner PJ, et al. Dynamic range extension of cine velocity measurements using motion‐registered spatiotemporal phase unwrapping. J Magn Reson Imaging. 1996;6:495‐502. [DOI] [PubMed] [Google Scholar]
- 38. Oztek MA, Mayr NA, Mossa‐Basha M, et al. The dancing cord: inherent spinal cord motion and its effect on cord dose in spine stereotactic body radiation therapy. Neurosurgery. 2020;87:1157‐66. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The MRI data acquired and processed in the present study are available from the corresponding author upon request with the need for a formal data sharing agreement.