How to analyze postoperative radiographs after total hip replacement

Martyna Barbara Budzińska 1, Bartosz Michał Maciąg 1,, Krystian Żarnovsky 1, Tomasz Kordyaczny 1, Inga Magdalena Kowalczyk 1, Olga Adamska 1, Artur Stolarczyk 1
PMCID: PMC9813239  PMID: 36063354

Abstract

Total hip replacement is one of the most widely performed surgeries. It is stated as the most efficient method of treating end-stage osteoarthritis of the hip joint. What is more, it significantly improves the quality of patients’ lives, relieves them from pain and restores decreased range of motion, provided that is conducted properly. Aim of this article is to indicate which constituents of prosthetic placement can be easily measured on postoperative radiographs and point out how to interpret obtained results. Multiple mechanical factors, such as center of rotation, femoral offset, acetabular offset, acetabular inclination, acetabular anteversion and leg length discrepancy can be measured on postoperative radiographs. To provide a successful surgery and to acquire both radiological and clinical satisfying results, proper prosthetic placement is crucial. Malpositioning of each element, in varying degrees may lead to dislocation or reoperation.

Keywords: Total hip replacement, Radiological evaluation, Radiographs

Introduction

Total hip replacement (THR) is one of the most widely performed surgeries. There were 98.649 hip replacement procedures performed in 2019 in the United Kingdom [1]. Not only is it cost-effective treatment, but what is more important it is a very successful one, which relieves patients from pain, improves their quality of life and restores decreased range of motion. Unfortunately, there still exists a group of patients, which does not derive any advantage from THR. It is most often caused by the impact of mechanical aspects of THR on clinical outcome.

In everyday clinical practice, radiographs are the most widely used tool for imaging. It is the most widespread and commonplace method. Classical radiography is associated with a lower dose of radiation in comparison to computed tomography. Its advantage over other imaging methods results also from the lowest price. Computed tomography can be used for postoperative assessment as well. It is certainly more accurate than radiographs, but it has its limitations, such as radiation dose and limited availability [2]. There is also one method of imaging—magnetic resonance imaging, which is far less accessible and more expensive than previously mentioned imaging modalities. However, there are situations in which magnetic resonance imaging is helpful to obtain a diagnosis. They include tendinopathy, implants loosening, persistent postoperative pain or other conditions that are hard or even impossible to detect using computed tomography and classical radiography.

In connection with this, in the majority of cases, radiographs are perfectly adequate. What is more, the application of this imaging technique is not associated with the artifacts caused by metal implants, in contrast to the use computed tomography or magnetic resonance.

Aim of this article is to indicate which constituents can be measured on postoperative radiographs and point out how to interpret obtained results.

Multiple mechanical and biomechanical factors, such as center of rotation (COR), femoral offset, acetabular offset, acetabular inclination (AI), acetabular anteversion and leg length discrepancy can be measured on radiographs. Each one of them, in varying degrees, can affect the outcome of the surgery.

Materials and methods

The literature was reviewed. Articles related to the subject, published in the years 1978–2021 were researched. Therefore publications including keywords such as total hip replacement, center of rotation, femoral offset, acetabular offset, cup anteversion, cup inclination and leg length discrepancy were searched in the PubMed database. Research was focused on English language papers, available abstracts, studies performed on people and articles. Inclusion of the articles was determined on the basis of titles, then abstracts, eventually entire articles. As terms of exclusion non-English language articles, papers ahead of print and only titles or abstract available were chosen. All studies presenting procedures conducted on animals were also excluded. If there occurred any signs of unreliability or relation to the topic were insignificant, the articles were eliminated during further evaluation. Afterwards, every selected article was verified another time. Any duplicates or obsolete information were removed. Research and error risk assessments were performed by one author. All information was selected and verified individually. Analysis and synthesis of studies were prepared independently.

All methods of measurements apply to radiographs taken in the supine position at a source-to-film distance of 100–115 cm with the X-ray beam directed to the midpoint of the pubic symphysis and perpendicular to the patient.

The described radiographic technique is a non-weight-bearing view. It is connected with the lesser radiation dose, allows to obtain a better quality radiographs, but it does not take into account the functional anatomy in contrast to weight-bearing view.

Results

Center of rotation (COR)

One of the goals of THR is to reconstruct the COR. As the study shows, restoring COR is an extremely important factor affecting operation result. It should be restored within 5 mm medial and 3 mm superior to the normal side. If optimal reconstruction is unattainable, the ability to control hip offset meticulously is limited. Shifting COR over 1 cm superiorly or medially causes early radiological signs of loosening. Malposition of COR may lead to abnormal gait, abductors insufficiency, increased risk of impingement and dislocation [3]. What is more, well reconstructed COR reduces the number of failures and revision surgeries [4]. Proper position of COR diminishes the risk of leg length discrepancy.

Ranawat’s method is an old but still applicable method used for the definition of the COR [5]. To determine the center of rotation on anteroposterior pelvis radiograph, two horizontal lines must be drawn. One at the level of iliac crests and the second one, at the level of ischial tuberosities. These lines must be connected by a perpendicular line passing through a point, which is located 5 mm lateral to the intersection of Kohler’s and Shenton’s lines. Point B and point C are situated along the horizontal line at the level of the subchondral roof of the cup. Point B is at an equal distance from both A and C points. COR is located half the length of the AC line [3, 5] (Fig. 1).

Fig. 1.

Fig. 1

Determination of the COR. A. point located 5 mm lateral to the intersection of the Kohler’s and Shenton’s lines; B. point situated along the horizontal line at the level of the subchondral roof of the cup; at equal distance from both A and C points; C. point situated along the horizontal line at the level of the subchondral roof of the cup; D. point located half the length of the AC line

Femoral offset

Femoral offset is stated as a distance from the COR of the femoral head to a line bisecting the long axis of the femur [3] (Fig. 2). Restored femoral offset enhances biomechanics, such as abduction strength and range of motion (ROM) by improving flexion and internal rotation of the hip [9, 10].

Fig. 2.

Fig. 2

Femoral offset measurement. A. line passing through the COR; B. line bisecting the long axis of the femur

In addition, femoral offset restored within 5 mm to the normal side, decreases both linear and volumetric polyethylene wear [4]. Failing to restore offset < 5 mm may cause pain aggravation and deterioration of the joint. Reduced offset decreases soft-tissue tightness and predisposes to dislocation. According to the studies, well-positioned femoral offset may be the critical mechanical factor preventing dislocation after THR. However, it is still not stated which values of femoral offset are indications for revision [610].

Acetabular offset

Hip offset is not only femoral offset. There is another component part-acetabular offset, which should not be omitted because its value varies from person to person. Acetabular offset is the distance from COR to the medial wall of the quadrilateral plate [11, 12]. It is measured as a vertical distance between the center of rotation and teardrop on the same side (Fig. 3). The unreconstituted acetabular offset reduces the lever arm of body weight. As a result, gluteus medius and minimus muscles get a more vertical line of action [11].

Fig. 3.

Fig. 3

Acetabular offset measurement. A. line passing through the medial wall of the quadrilateral plate; B. line passing through the COR

Acetabular inclination (AI)

Acetabular inclination is also known as an abduction angle. Radiographic AI is measured on anteroposterior radiographs. It is an angle between transischial line and a line conducted through the cup margins [10] (Fig. 4). AI affects range motion and wear of the acetabular component. When the abduction angle is lesser than 45°, flexion and abduction decreases. On the other hand, AI over 45° reduces adduction and rotation. There is also higher wear of acetabular polyethylene when the AI is over 45° [13, 14]. It cannot be univocally stated, without additional studies, what is a safe range for AI [15, 16].

Fig. 4.

Fig. 4

Acetabular inclination measurement. α—an angle between transischial line and a line conducted through the cup margin

Acetabular anteversion

Computed tomography facilitates accurate determination of acetabular cup position, especially anteversion, but it is not widely used in clinical practice because of its high cost, limited availability and additional radiation exposure [17, 18]. Because of that, plain radiographs are in common use. Plenty of methods can be used to measure anteversion, and there is no validated and most efficient one [14]. This is due to the fact that pelvic tilt has a greater impact on measuring anteversion on anteroposterior radiographs rather than determining abduction angle on anteroposterior radiographs. Nevertheless, as studies show, there is a method that enables measurements that do not differ from measurements carried out using computed tomography. It is a method created by Liaw [19, 20].

Liaw’s anteversion is stated as:

Anteversion=arcsictanα

α is an angle between the principal axis of the ellipse and the vector connecting the endpoints of the main and the minor axes. In plan, ellipse is located on the margin of a cup (Fig. 5).

Fig. 5.

Fig. 5

Liaw’s acetabular anteversion measurement. α—an angle between the principal axis of the ellipse and the vector connecting the endpoints of the main and the minor axes

The positioning of the cup influences the risk of dislocation. There is no consensus among researchers on what really is a safe zone for acetabular anteversion. According to Lewinnek safe zone theory, cup anteversion should be oriented between 5° and 25° to minimize dislocation after THR. But in fact, recent studies show that there is no real safe range, whether for AI or acetabular anteversion [14, 15, 21].

Anteversion values can be exploited to monitor acetabular migration. Change of anteversion on postoperative radiographs over 1.59° is an early sign of cup loosening, which can manifest itself in hip pain [18].

Leg length discrepancy

Leg length discrepancy is one of the most frequent complications after THR. Lengthening occurs more often than shortening of the limb and is more noticeable in patients. According to the study, leg length discrepancy is perceived by patients when the operated limb is lengthened over 6 mm or shortened below 10 mm [22]. However, the greatest problem are inequalities above 10 mm due to their impact on everyday functioning. They can cause abnormal gait, instability, sciatica and back pain. Moreover, lengthening greater than 10 mm occurs with other complications: limping, pelvic obliquity and feeling disenchanted [15, 23].

The leg length discrepancy on anteroposterior radiographs is given as the difference in perpendicular distance between a line passing through the lower edge of the teardrop points to the corresponding tip of the lesser trochanter [10] (Fig. 6).

Fig. 6.

Fig. 6

Leg length discrepancy measurement. A and A’ are located at the level of the tips of the lesser trochanters. B and B’ are situated at the level of the lower edge of the teardrop points. AB and A’B’ lines are distances between a line passing through the lower edge of the teardrop points to the corresponding tip of the lesser trochanter. Difference between AB and A’B’ lengths is stated as leg length discrepancy

Conclusions

To provide a successful surgery and to acquire both radiological and clinical satisfying results, all of the mechanics and biomechanics must be restored, if it is not, surgery can aggravate the patient’s complaints instead of removing them and the results of performed THR can be unsatisfactory. This may also result in dislocation or the need for reoperation.

Author contributions

Conceptualization: MBB, BMM, AS; Literature search and data analysis: KŻ, TK, IMK, OA; Writing—original draft preparation: MBB, BMM, IMK, KŻ; Writing—review and editing: TK, AS, OA; Supervision: BMM, AS; All authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

Funding

This research received no external funding.

Declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.NJR Annual Reports. http://www.njrreports.org.uk/ (Date Last Accessed 31 Dec 2020).
  • 2.Reito A, Puolakka T, Paakkala A. Assessment of inter- and intra-observer reliability in the determination of radiographic version and inclination of the cup in metal-on-metal hip resurfacing. Int Orthop. 2012;36:519–525. doi: 10.1007/s00264-011-1328-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Lum ZC, Dorr LD. Restoration of center of rotation and balance of THR. J Orthop. 2018;15(4):992–996. doi: 10.1016/j.jor.2018.08.040. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lecerf G, Fessy MH, Philippot R, et al. Femoral offset: anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty. Orthop Traumatol Surg Res. 2009;95:210–219. doi: 10.1016/j.otsr.2009.03.010. [DOI] [PubMed] [Google Scholar]
  • 5.Ranawat CS, Dorr LD, Inglis AE. Total hip arthroplasty in protrusio acetabuli of rheumatoid arthritis. J Bone Jt Surg Am. 1980;62(7):1059–1065. doi: 10.2106/00004623-198062070-00002. [DOI] [PubMed] [Google Scholar]
  • 6.Little NJ, Busch CA, Gallagher JA. Acetabular polyethylene wear and acetabular inclination and femoral offset. Clin Orthop Relat Res. 2009;467(11):2895–2900. doi: 10.1007/s11999-009-0845-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.McGrory BJ, Morrey BF, Cahalan TD. Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Jt Surg (Br) 1995;77:865–889. doi: 10.1302/0301-620X.77B6.7593096. [DOI] [PubMed] [Google Scholar]
  • 8.Fordea B, Engelnb K, Bedairc H. Restoring femoral offset is the most important technical factor in preventing total hip arthroplasty dislocation. J Orthop. 2018;15(1):131–133. doi: 10.1016/j.jor.2018.01.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Xu B, Yang D, Aili R, Cao L. Effect of femoral offset change on pain and function after total hip arthroplasty. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013;27(7):843–846. [PubMed] [Google Scholar]
  • 10.Mahmood SS, Amiry BA, Mukka SS. Validity, reliability and reproducibility of plain radiographic measurements after total hip arthroplasty. Skeletal Radiol. 2015;44(3):345–351. doi: 10.1007/s00256-014-2055-7. [DOI] [PubMed] [Google Scholar]
  • 11.Matsushita A, Nakashima Y, Jingushi S, Yamamoto T, Kuraoka A, Iwamoto Y. Effects of the femoral offset and the head size on the safe range of motion in total hip arthroplasty. J Arthroplasty. 2009;24(4):646–651. doi: 10.1016/j.arth.2008.02.008. [DOI] [PubMed] [Google Scholar]
  • 12.Bonnin MP, Pooler HA, Basiglini L. Do we medialise the hip centre of rotation in total hip arthroplasty? Influence of acetabular offset and surgical technique. Hip Int. 2012;22(04):371–378. doi: 10.5301/HIP.2012.9350. [DOI] [PubMed] [Google Scholar]
  • 13.D’Lima DD, Uruhart AG, Buehler KO, Walker RH, Colwell CW., Jr The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Jt Surg Am. 2000;82(3):315–321. doi: 10.2106/00004623-200003000-00003. [DOI] [PubMed] [Google Scholar]
  • 14.Lee GC, Lee SH, Kang SW, Park HS, Jo S. Accuracy of planar anteversion measurements using anteroposterior radiographs. BMC Musculoskelet Disord. 2019;20(1):586. doi: 10.1186/s12891-019-2979-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Biedermann R, Tonin A, Krismer M. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Jt Surg Br. 2005;87(6):762–769. doi: 10.1302/0301-620X.87B6.14745. [DOI] [PubMed] [Google Scholar]
  • 16.Abdel MP, von Roth P. What safe zone? The vast majority of dislocated THAs are within the Lewinnek safe zone for acetabular component position. Clin Orthop Relat Res. 2016;474:386–391. doi: 10.1007/s11999-015-4432-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Loughead JM, Chesney D, Holland JP, McCaskie AW. Comparison of offset in Birmingham hip resurfacing and hybrid total hip arthroplasty. J Bone Jt Surg Br. 2005;87:163–166. doi: 10.1302/0301-620X.87B2.15151. [DOI] [PubMed] [Google Scholar]
  • 18.Yeh KL, Wu TY, Ma HH. Ellipse method for measuring Liaw’s anteversion of the acetabular component after total hip arthroplasty. BMC Musculoskelet Disord. 2020;21:667. doi: 10.1186/s12891-020-03669-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Park YS, Shin WC, Lee SM, Kwak SH, Bae JY, Suh KT. The best method for evaluating anteversion of the acetabular component after total hip arthroplasty on plain radiographs. J Orthop Surg Res. 2018;13(1):66. doi: 10.1186/s13018-018-0767-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Alzohiry MA, Abdelnasser MK, Moustafa M. Accuracy of plain antero-posterior radiographic-based methods for measurement of acetabular cup version. Int Orthop. 2018;42:2777–2785. doi: 10.1007/s00264-018-3984-x. [DOI] [PubMed] [Google Scholar]
  • 21.Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Jt Surg Am. 1978;60(2):217–220. doi: 10.2106/00004623-197860020-00014. [DOI] [PubMed] [Google Scholar]
  • 22.Konyves A, Bannister GC. The importance of leg length discrepancy after total hip arthroplasty. J Bone Jt Surg Br. 2005;87:155–157. doi: 10.1302/0301-620X.87B2.14878. [DOI] [PubMed] [Google Scholar]
  • 23.Flecher X, Ollivier M, Argenson JN. Lower limb length and offset in total hip arthroplasty. Orthop Traumatol Surg Res. 2016;102(1 Suppl):S9–20. doi: 10.1016/j.otsr.2015.11.001. [DOI] [PubMed] [Google Scholar]

Articles from Japanese Journal of Radiology are provided here courtesy of Springer

RESOURCES