Biventricular assist devices and total artificial heart: Strategies and outcomes

Taiyo Kuroda 1, Chihiro Miyagi 1, Kiyotaka Fukamachi 1,2, Jamshid H Karimov 1,2,*
PMCID: PMC9853410  PMID: 36684573

Abstract

In contrast to the advanced development of the left ventricular assist device (LVAD) therapy for advanced heart failure, the mechanical circulatory support (MCS) with biventricular assist device (BVAD) and total artificial heart (TAH) options remain challenging. The treatment strategy of BVAD and TAH therapy largely depends on the support duration. For example, an extracorporeal centrifugal pump, typically referred to as a temporary surgical extracorporeal right ventricular assist device, is implanted for the short term with acute right ventricular failure following LVAD implantation. Meanwhile, off-label use of a durable implantable LVAD is a strategy for long-term right ventricular support. Hence, this review focuses on the current treatment strategies and clinical outcomes based on each ventricle support duration. In addition, the issue of heart failure post-heart transplantation (post-HT) is explored. We will discuss MCS therapy options for post-HT recipients.

Keywords: biventricular heart failure, left ventricular assist device, heart transplantation, Impella®, right heart failure, HeartMate 3®, Syncardia® TAH, ProtekDuo®

1. Introduction

The clinical data of mechanical circulatory support (MCS) for biventricular heart failure (BHF) is limited due to the low prevalence of treating severe right heart failure (RHF). The complexity of the BHF treatment strategy also makes the situation challenging. The reason for the complexity is that the BHF treatment strategy will be determined with the consideration of support duration for both ventricles. Furthermore, the clinical scenario will be important to determine the device; for example, the device selection may differ with the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) patient profiles and baseline patient characteristics, including the diagnosis (1, 2).

BHF support conditions have combinations of chronic/temporary support duration and left/right ventricle, which is as follows: (1) Chronic left ventricular (LV) support with chronic right ventricular (RV) support; (2) Chronic LV support with temporary RV support; (3) Temporary LV support with chronic RV support; and (4) Temporary LV support with temporary RV support. In this review, we will focus on 1, 2, and 4, because the temporary LV support with chronic RV support situation is quite rare due to RHF pathogenesis (3).

Lastly, this review article will discuss populations in need of BHF support among post-heart transplantation (post-HT) patients (4). Since the first successful human heart transplantation (HT) reported in 1967 in South Africa, over 120,000 patients have received HT therapy (5). HT volumes have been increasing steadily for more than a decade (6). Also, the survival rate is improving, as median survival after adult HT performed in 2002–2008 now exceeds 12 years; therefore, the number of post-HT patients are increasing. However, the leading causes of death among post-HT patients, remain the rate of graft failure and infection (6). We will discuss the possibility of indications for BHF support MCS in this situation.

2. Devices for chronic LV support with chronic RV support

There are two options for chronic LV support with chronic RV support: using a continuous-flow (CF) left ventricular assist device (LVAD) and using one more LVAD as a right ventricular assist device (RVAD); or a total artificial heart (TAH). Patients requiring RVAD after LVAD implantation, or lower INTERMACS patient profile patients with biventricular failure diagnosis, are the most frequent example for this clinical situation.

Regarding the first listed population (patients requiring RVAD after LVAD implantation; biventricular assist device, BVAD), according to the Twelfth INTERMACS Report, approximately 2,700–3,000 patients in the United States (US) receive an LVAD implant each year; and, in 2020, CF LVADs have accounted for the most (7). In the HeartMate 3 ® (HM3) (Abbott, Abbott Park, IL, USA) pivotal and post-pivotal trial study, 4.1 and 7.4% of patients, respectively required BVAD (8). Several reasons for RHF during the LVAD therapy have been hypothesized. For instance, the shift of the interventricular septum (IVS) to the left side by LVAD therapy reduces the septal contribution of the RV contraction and thus increased workload is a concern for worsening RV function (912). In addition, the shift of the IVS may distract the septal papillary muscle with systolic restriction of septal leaflet motion, which may intensify the tricuspid regurgitation (13). Furthermore, increased venous return created by increased cardiac output from the LVAD may worsen the potential RV dysfunction that LVAD patients already have to some degree (14).

One of the keys to a successful BVAD therapy may be estimating the support duration. In HM3 patients, approximately 40% of those upgrading to BVAD were performed within 0–2 days after HM3 implantation, and 23% of upgrades were performed within 3–14 days (8). The severe late RV failure among LVAD patients, which is defined by the requirement for an RVAD at 3–12 months from LVAD implant, is very rare (15). In addition, over 60% of successful RVAD weaning rates were reported with a median support duration of 13–17 days (1619). Therefore, paradoxically, if BVAD intervention after LVAD implantation took place in an earlier period, BVAD support duration is expected to be short (up to 17 days). An investigation using the United Network for Organ Sharing (UNOS) database showed that 1% of LVAD patients were transitioned to durable BVAD support, and 0.2% of LVAD patients were transitioned to TAH support (20). Appropriate timing for the intervention is also critical. Preoperative hemodynamic parameters are used to assess the RV function after LVAD implantation (Table 1) (21). For example, with preoperative central venous pressure (CVP) greater than 15 mm Hg and CVP/pulmonary capillary wedge pressure (PCWP) greater than 0.63, a significantly higher risk of RHF was reported (17). The preoperative RV stroke work index (RV SWI) was reported as another predictor for RVAD implant function, which suggests that RV SWI was lower in the RVAD group (22, 23). The preoperative pulmonary artery pulsatility index was reported as a predictor for early RV failure with a cutoff value of 2.0 (23, 24). Also, using echocardiography, the RV global longitudinal strain predicted an early acute and post-implant RV failure with a cutoff value of −9.7% (25). However, a single parameter still may not be sufficient in predicting the RHF after LVAD implantation (26).

TABLE 1.

Preoperative predictors for right ventricle failure following left ventricular assist device (LVAD) implantation.

Parameters Description
CVP > 15 mm Hg OR: 2.1 (IQR 1.2–3.6)
CVP/PCWP > 0.63 OR: 2.5 (IQR 1.4–4.6)
RV SWI RVAD [151 ± 75 (mm Hg × mL/m2)] > no RVAD [368 ± 245 (mm Hg × mL/m2)] (p = 0.01)
PAPi < 2.0 AUC, 0.77; sensitivity, 74%; specificity, 67%
RV GLS > −9.7% AUC, 0.86; sensitivity, 89%; specificity, 78%

CVP, central venous pressure; OR, odds ratio; IQR, interquartile range; PCWP, pulmonary capillary wedge pressure; RV SWI, right ventricle stroke work index; RVAD, right ventricle assist device; PAPi, pulmonary artery pulsatility index; AUC, area under curve; RV GLS, right ventricle longitudinal strain.

Using durable implantable CF LVAD as RVAD, such as HM3, currently is an off-label use. There is a lack of a proven, long-term MCS devices except for the dual Berlin Heart EXCOR ® system (Berlin Heart, GmbH, Berlin, Germany), which is mainly used in pediatric patients. Due to anatomical and physiological limitations (12), a modified technique was used to implant dual HM3 (27, 28). The clinical outcome for dual HM3 varied among two studies; the survival at 18 months was 54.6–91.7% (28, 29). Another retrospective study using INTERMACS database (n = 38, multi-center) described that survival outcomes among the BVAD patients (BVAD was defined as LVAD and RVAD implanted in the same operation) were 68% at 6 months and 62% at 12 months (30). This study found that 11 patients died with device in place, 9 patients survived to reach HT, and 18 patients were alive on support at the mean follow-up of 5.08 months.

Another option for chronic LV support with chronic RV support is TAH. SynCardia® TAH (Figure 1A) (SynCardia Systems. LLC, Tucson, Arizona, USA) is the most commonly used TAH. It is a pneumatically driven FDA-approved, volume displacement pump TAH with a size line-up of 50 mL and 70 mL. The clinical data reported that the TAH patient mortality while waiting for HT was 7.4%, while 87% of patients reached HT (20). An analysis from the INTERMACS database (n = 450, all patients received SynCardia 70 ml TAH) showed that survival at 1 year was 53.2% and 33.9% at 2 years (2). HT was performed in 266 patients, and 162 patients died on support. Among those, 80% of patients in this study were INTERMACS patient profiles 1 or 2, and 20% of patients were supported with extracorporeal membrane oxygenation (ECMO). Notably, survival was superior in the earlier era (2006–2011) than the most recent era (2012–2017), which suggests that the difference in survival was largely influenced by the experience of the facility (31); therefore, the TAH survival may improve with surgical and patient management experience including the patient selection and the timing of intervention (2). Furthermore, the baseline condition of the TAH cohort was sicker than LVADs patients, which may have affected the survival. Currently, the largest difference between BVAD and TAH is the strategy implemented. TAH implantation procedure resect both ventricles; therefore, a bridge to recover treatment concept is not feasible, and HT is the only way forward. However, destination therapy with SynCardia 70 mL TAH, in which a clinical trial is ongoing in 2022, may change the future strategy.

FIGURE 1.

FIGURE 1

Mechanical circulatory support (MCS) devices for right ventricular/biventricular failure. (A) SynCardia® total artificial heart 70 cc (left) and 50 cc (right), re-use from Villa et al. (60), (B) Aeson® total artificial heart (Image from Carmat SA, used with permission), (C) Impella RP® (Image from Impella RP instruction manual, used with permission), and (D) LifeSparc® pump with ProtekDuo® cannula (Image from Livanova Investor Day 2021 presentation, used with permission).

Aeson® TAH (Figure 1B) (Carmat SA, Vélizy-Villacoublay, France), is a volume displacement, electro-hydraulically driven TAH which has recently been approved by the FDA to conduct an early feasibility study (32). Currently, the only clinical data available for Aeson® TAH is an implantation report for four patients, which describe a support period of 20–270 days, with two patients being discharged to return home (33). In addition, several TAHs are in development, such as Cleveland Clinic CF TAH (CFTAH) (Cleveland Clinic, Cleveland, OH, USA) (3436), BiVACOR® TAH (BiVACOR, Inc., Houston, TX, USA) (3739), and Realheart® TAH (Scandinavian Real Heart AB, Vesteras, Sweden) (4042).

3. Devices for chronic LV support with temporary RV support

For chronic LV support with short to intermediate RV support duration (0–30 days), an extracorporeal centrifugal pump has been widely used. Customarily, this configuration is called a surgical extracorporeal RVAD (sRVAD). In sRVAD, the inflow cannula is placed directly, or (via the femoral vein) into the right atrium (RA), and the outflow cannula is placed directly, or (via a sutured prosthesis graft) into the pulmonary artery (PA) (16, 43). Both cannula will pass through the skin, and the circuit will be connected to a centrifugal pump system, such as the CentriMag® blood pump (Abbott, Abbott Park, IL, USA). The advantage of this system is that it is easy to place, so if RV failure presents during the HM3 implantation procedure, the sRVAD can be placed through a sternotomy. Also, this system is able to provide much larger flows than a veno-arterial extracorporeal membrane oxygenation system. The disadvantage of this system is that it may require a re-sternotomy to explant the sRVAD, which increases the risk of infection.

The RVAD implantation following LVAD implantation occurs within 0–2 days (8), and a study reported that 86% of patients who received sRVAD after LVAD implantation were successfully weaned from support with a duration of 16.4 ± 9.6 days (44). Another analysis using the INTERMACS database reported that the 1-, 6-, and 12-month survival rates for the chronic BVAD patients were 89, 68, and 62%, respectively, and there was no significant difference between the patients with chronic LV support with temporary RV support (30).

Sternotomy for HM3 implantation allows easy access to both RA and PA for sRVAD implantation; however, an effort to implant HM3 with minimally invasive surgery, preferably by thoracotomy, is under review (45); therefore, access to the RA and PA would be limited. Since the most common support duration is temporary to intermediate (0–15 days), percutaneous RV assist device (p-RVAD), such as Impella RP® (Figure 1C) (Abiomed, Danvers, MA, USA) and LifeSPARC® Pump with ProtekDuo® dual-lumen cannula (Figure 1D) (LivaNova, Houston, TX, USA), may be a good solution. A prospective cohort study, which includes patients who received Impella RP implantation following LVAD implantation (n = 31), described that 77.4% of patients reached the primary end-point which was survival at 30 days, or discharge after device explant, or induction of anesthesia for a long-term therapy (46). Another retrospective study reported that 27 LVAD patients received the LifeSPARC Pump with ProtekDuo dual-lumen cannula system implantation, and device weaning occurred in 86% of patients, with 15% resulting in-hospital mortality (47). Those outcomes seems acceptable; thus, in cases of unplanned p-RVAD implantation, and if high-risk RV failure patients are having surgery, it may be better to perform LVAD implantation surgery in the hybrid operating room.

4. Devices for temporary LV support with temporary RV support

There are emergent, acute scenarios that require temporary BVAD support. The percutaneous LVAD (p-LVAD), such as Impella® (Abiomed, Danvers, MA, USA), has been increasingly used as temporary LV support, and sometimes it is used in combination with a p-RVAD. In a retrospective study among 5 U.S. hospitals (n = 20),with a combination of left side Impella and Impella RP, called BiPella (48), it was reported that in-hospital mortality was 50% (49). The advantage of this system is its ease to implant and explant the device; however, a concern of this therapy is adequate LV unloading and it should be monitored to ensure appropriate support is supplied. If not, the pulmonary vasculature is over-pressurized due to the high pulmonary resistance resulting from high left atrial pressure. The combination of left side Impella and LifeSPARC Pump with ProtekDuo dual-lumen cannula has been described in several reports (21, 5052).

5. Devices for post-heart transplantation biventricular failure

HT is one of the de facto goals of current end-stage HF treatment, which has an overall median survival of 12.5 years, and a conditional survival curve of 14.8 years for those who survived the first year (53). Hospitalization due to late graft failure was observed in 33% of patients (54). Among those populations, some may progress to cardiogenic shock, and may need a MCS. Notably, a retrospective study (n = 26) reported that 42% of patients with late graft failure were treated with BVAD and TAH (4). Among the patients with BVAD implantation in this study, 60% received dual CentriMag configuration, 20% received dual implantable LVAD configuration (details unknown), 10% received dual HVAD, and 10% received HVAD with sRVAD. The outcomes were that 23% were weaned (including single VA-ECMO), 19% underwent Re-heart transplantation (Re-HT), and 15% were discharged with durable MCS. The mortality rate was 42%.

Re-HT may be an option for patients who develop refractory graft failure. The indication is rare, but the population receiving second and third Re-HT has increased (55). Although MCS is not commonly performed as a bridge to Re-HT, the TAH is theoretically an ideal option because the antigen would be removed from the body; however, the outcome for the TAH as a bridge to Re-HT is reported as very high risk, and the potential for improving survival remains uncertain (56). In addition, TAH as a destination therapy may change future treatment strategies.

6. Discussion

In this review, we mainly discussed biventricular failure with post-LV failure pathogenesis based on the classification of support duration. The treatment strategy differed in support duration, and there is an interesting, ongoing effort in both acute and chronic biventricular MCS. In addition, congenital heart disease is still another condition that may require biventricular support. The most widely used MCS among congenital heart disease population is the Berlin Heart EXCOR® system, which offers a variety of pump sizes. Treatment strategies for the pediatric population may differ due to patient growth.

Regarding chronic BVAD, dual HM3 is the most commonly used configuration. The adverse event most noted is RVAD pump thrombosis, which has been consistently reported with occurrence of 36–37% (12, 57, 58). However, it seems lesser than previous reports (28, 29), which can be explained by the low thrombosis risk of the HM3. HM3 is a centrifugal pump with a displacing volume of 80 mL, and it is slightly larger than HeartWare® Ventricular Assist Device (HVAD) (Medtronic Inc., Minneapolis, MN, USA) (50 mL); this size difference influences device location inside the patient’s chest. HM3 is likely to be implanted into the free wall of the RA. The pump is wrapped in Gore-Tex® (W.L. Gore, Flagstaff, AZ, USA), or a polytetrafluoroethylene sheet, and is placed as it protrudes into the right thoracic cavity through a slit in the pericardium (27, 29). Although the relationship with pump positioning was uncertain, the evacuation of a right hemothorax, effusion, and Aspergillus species infection in the RVAD cavity were reported. On the other hand, HVADs were implanted into either the RA (38%) or RV (63%) (59). RA-implanted HVAD was supported longer than RV HVAD (p = 0.02), and did not show significant differences in postoperative complications, such as pump thrombosis. Therefore, a small, durable pump implanted into the RA appears to be a viable option for RVAD used in chronic BVAD.

In conclusion, the percutaneous temporary RVAD may increase its prevalence in temporary biventricular support. As for long-term biventricular support, the development of a durable, specifically designed RVAD, with a wide operation range and suitable inflow cannulas, is expected. Furthermore, the development of a durable BVAD, including TAH, may provide a valid solution for the management of heart failure among post-HT patients.

Author contributions

TK: manuscript preparation. CM: critical manuscript review. KF: critical revision of article. JK: critical revision and approval of article. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1.Caraffa R, Bejko J, Carrozzini M, Bifulco O, Tarzia V, Lorenzoni G, et al. A device strategy-matched comparison analysis among different intermacs profiles: A single center experience. J Clin Med. (2022) 11:4901. 10.3390/jcm11164901 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Arabía F, Cantor R, Koehl D, Kasirajan V, Gregoric I, Moriguchi J, et al. Interagency registry for mechanically assisted circulatory support report on the total artificial heart. J Heart Lung Transplant. (2018) 37:1304–12. 10.1016/j.healun.2018.04.004 [DOI] [PubMed] [Google Scholar]
  • 3.Konstam M, Kiernan M, Bernstein D, Bozkurt B, Jacob M, Kapur N, et al. Evaluation and management of right-sided heart failure: A scientific statement from the american heart association. Circulation. (2018) 137:e578–622. 10.1161/CIR.0000000000000560 [DOI] [PubMed] [Google Scholar]
  • 4.D’Angelo A, Naka Y, Sanchez J, Kaku Y, Witer L, Fried J, et al. Outcomes of mechanical support for cardiogenic shock associated with late cardiac allograft failure. J Card Surg. (2020) 35:3381–6. 10.1111/jocs.15089 [DOI] [PubMed] [Google Scholar]
  • 5.Quader M, Toldo S, Chen Q, Hundley G, Kasirajan V. Heart transplantation from donation after circulatory death donors: Present and future. J Card Surg. (2020) 35:875–85. 10.1111/jocs.14468 [DOI] [PubMed] [Google Scholar]
  • 6.Khush K, Cherikh W, Chambers D, Goldfarb S, Hayes D, Jr., Kucheryavaya A, et al. The international thoracic organ transplant registry of the international society for heart and lung transplantation: Thirty-fifth adult heart transplantation report-2018; focus theme: Multiorgan transplantation. J Heart Lung Transplant. (2018) 37:1155–68. 10.1016/j.healun.2018.07.022 [DOI] [PubMed] [Google Scholar]
  • 7.Shah P, Yuzefpolskaya M, Hickey G, Breathett K, Wever-Pinzon O, Khue-Ton V, et al. Twelfth interagency registry for mechanically assisted circulatory support report: Readmissions after left ventricular assist device. Ann Thorac Surg. (2022) 113:722–37. 10.1016/j.athoracsur.2021.12.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Mehra M, Cleveland J, Jr., Uriel N, Cowger J, Hall S, Horstmanshof D, et al. Primary results of long-term outcomes in the momentum 3 pivotal trial and continued access protocol study phase: A study of 2200 heartmate 3 left ventricular assist device implants. Eur J Heart Fail. (2021) 23:1392–400. 10.1002/ejhf.2211 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Farrar D. Ventricular interactions during mechanical circulatory support. Semin Thorac Cardiovasc Surg. (1994) 6:163–8. [PubMed] [Google Scholar]
  • 10.Moon M, Bolger A, Deanda A, Komeda M, Daughters G, II, Nikolic S, et al. Septal function during left ventricular unloading. Circulation. (1997) 95:1320–7. 10.1161/01.CIR.95.5.1320 [DOI] [PubMed] [Google Scholar]
  • 11.Foschi M, Di Mauro M, Tancredi F, Capparuccia C, Petroni R, Leonzio L, et al. The dark side of the moon: The right ventricle. J Cardiovasc Dev Dis. (2017) 4:18. 10.3390/jcdd4040018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Karimov J, Sunagawa G, Horvath D, Fukamachi K, Starling R, Moazami N. Limitations to chronic right ventricular assist device support. Ann Thorac Surg. (2016) 102:651–8. 10.1016/j.athoracsur.2016.02.006 [DOI] [PubMed] [Google Scholar]
  • 13.Krishan K, Nair A, Pinney S, Adams D, Anyanwu A. Liberal use of tricuspid-valve annuloplasty during left-ventricular assist device implantation. Eur J Cardiothorac Surg. (2012) 41:213–7. 10.1016/j.ejcts.2011.05.047 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Lampert B, Teuteberg J. Right ventricular failure after left ventricular assist devices. J Heart Lung Transplant. (2015) 34:1123–30. 10.1016/j.healun.2015.06.015 [DOI] [PubMed] [Google Scholar]
  • 15.Rame J, Pagani F, Kiernan M, Oliveira G, Birati E, Atluri P, et al. Evolution of late right heart failure with left ventricular assist devices and association with outcomes. J Am Coll Cardiol. (2021) 78:2294–308. 10.1016/j.jacc.2021.09.1362 [DOI] [PubMed] [Google Scholar]
  • 16.Shimada S, Nawata K, Kinoshita O, Ono M. Mechanical circulatory support for the right ventricle in combination with a left ventricular assist device. Expert Rev Med Devices. (2019) 16:663–73. 10.1080/17434440.2019.1635006 [DOI] [PubMed] [Google Scholar]
  • 17.Kormos R, Teuteberg J, Pagani F, Russell S, John R, Miller L, et al. Right ventricular failure in patients with the heartmate ii continuous-flow left ventricular assist device: Incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg. (2010) 139:1316–24. 10.1016/j.jtcvs.2009.11.020 [DOI] [PubMed] [Google Scholar]
  • 18.Yoshioka D, Takayama H, Garan R, Topkara V, Han J, Kurlansky P, et al. Contemporary outcome of unplanned right ventricular assist device for severe right heart failure after continuous-flow left ventricular assist device insertion. Interact Cardiovasc Thorac Surg. (2017) 24:828–34. 10.1093/icvts/ivw409 [DOI] [PubMed] [Google Scholar]
  • 19.Khorsandi M, Schroder J, Daneshmand M, Bishawi M, Bouamra O, Winterton P, et al. Outcomes after extracorporeal right ventricular assist device combined with durable left ventricular assist device support. Ann Thorac Surg. (2019) 107:1768–74. 10.1016/j.athoracsur.2018.11.051 [DOI] [PubMed] [Google Scholar]
  • 20.Coyan G, Huckaby L, Diaz-Castrillon C, Miguelino A, Kilic A. Trends and outcomes following total artificial heart as bridge to transplant from the unos database. J Card Surg. (2022) 37:1215–21. 10.1111/jocs.16329 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kapur N, Esposito M, Bader Y, Morine K, Kiernan M, Pham D, et al. Mechanical circulatory support devices for acute right ventricular failure. Circulation. (2017) 136:314–26. 10.1161/CIRCULATIONAHA.116.025290 [DOI] [PubMed] [Google Scholar]
  • 22.Fukamachi K, McCarthy P, Smedira N, Vargo R, Starling R, Young J. Preoperative risk factors for right ventricular failure after implantable left ventricular assist device insertion. Ann Thorac Surg. (1999) 68:2181–4. 10.1016/S0003-4975(99)00753-5 [DOI] [PubMed] [Google Scholar]
  • 23.Morine K, Kiernan M, Pham D, Paruchuri V, Denofrio D, Kapur N. Pulmonary artery pulsatility index is associated with right ventricular failure after left ventricular assist device surgery. J Card Fail. (2016) 22:110–6. 10.1016/j.cardfail.2015.10.019 [DOI] [PubMed] [Google Scholar]
  • 24.Kang G, Ha R, Banerjee D. Pulmonary artery pulsatility index predicts right ventricular failure after left ventricular assist device implantation. J Heart Lung Transplant. (2016) 35:67–73. 10.1016/j.healun.2015.06.009 [DOI] [PubMed] [Google Scholar]
  • 25.Liang L, Jamil A, Mazurek J, Urgo K, Wald J, Birati E, et al. Right ventricular global longitudinal strain as a predictor of acute and early right heart failure post left ventricular assist device implantation. Asaio J. (2022) 68:333–9. 10.1097/MAT.0000000000001467 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Houston B, Shah K, Mehra M, Tedford R. A new “twist” on right heart failure with left ventricular assist systems. J Heart Lung Transplant. (2017) 36:701–7. 10.1016/j.healun.2017.03.014 [DOI] [PubMed] [Google Scholar]
  • 27.Lavee J, Mulzer J, Krabatsch T, Marasco S, McGiffin D, Garbade J, et al. An international multicenter experience of biventricular support with heartmate 3 ventricular assist systems. J Heart Lung Transplant. (2018) 37:1399–402. 10.1016/j.healun.2018.08.008 [DOI] [PubMed] [Google Scholar]
  • 28.Hetzer R, Krabatsch T, Stepanenko A, Hennig E, Potapov E. Long-term biventricular support with the heartware implantable continuous flow pump. J Heart Lung Transplant. (2010) 29:822–4. 10.1016/j.healun.2010.02.012 [DOI] [PubMed] [Google Scholar]
  • 29.McGiffin D, Kure C, McLean J, Marasco S, Bergin P, Hare J, et al. The results of a single-center experience with heartmate 3 in a biventricular configuration. J Heart Lung Transplant. (2021) 40:193–200. 10.1016/j.healun.2020.12.006 [DOI] [PubMed] [Google Scholar]
  • 30.Arabía F, Milano C, Mahr C, McGee E, Jr., Mokadam N, Rame J, et al. Biventricular support with intracorporeal, continuous flow, centrifugal ventricular assist devices. Ann Thorac Surg. (2018) 105:548–55. 10.1016/j.athoracsur.2017.08.019 [DOI] [PubMed] [Google Scholar]
  • 31.Pagani F. Clinical implications of the total artificial heart: Adversity and progress. J Heart Lung Transplant. (2018) 37:1298–300. 10.1016/j.healun.2018.07.019 [DOI] [PubMed] [Google Scholar]
  • 32.Han J. Aeson—the carmat total artificial heart is approved for enrollment in the united states. Artif Organs. (2021) 45:445–6. 10.1111/aor.13959 [DOI] [PubMed] [Google Scholar]
  • 33.Latrémouille C, Carpentier A, Leprince P, Roussel J, Cholley B, Boissier E, et al. A bioprosthetic total artificial heart for end-stage heart failure: Results from a pilot study. J Heart Lung Transplant. (2018) 37:33–7. 10.1016/j.healun.2017.09.002 [DOI] [PubMed] [Google Scholar]
  • 34.Fukamachi K, Horvath D, Massiello A, Fumoto H, Horai T, Rao S, et al. An innovative, sensorless, pulsatile, continuous-flow total artificial heart: Device design and initial in vitro study. J Heart Lung Transplant. (2010) 29:13–20. 10.1016/j.healun.2009.05.034 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Karimov J, Moazami N, Kobayashi M, Sale S, Such K, Byram N, et al. First report of 90-day support of 2 calves with a continuous-flow total artificial heart. J Thorac Cardiovasc Surg. (2015) 150:687–93. 10.1016/j.jtcvs.2015.06.023 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Kuroda T, Miyamoto T, Miyagi C, Polakowski A, Flick C, Kuban B, et al. Pulsatility hemodynamics during speed modulation of continuous-flow total artificial heart in a chronic in vivo model. Artif Organs. (2022) 46:1555–63. 10.1111/aor.14237 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Greatrex N, Kleinheyer M, Nestler F, Timms D. The maglev heart. IEEE Spectrum. (2019) 56:22–9. 10.1109/MSPEC.2019.8818588 [DOI] [Google Scholar]
  • 38.Timms D, Fraser J, Hayne M, Dunning J, McNeil K, Pearcy M. The bivacor rotary biventricular assist device: Concept and in vitro investigation. Artif Organs. (2008) 32:816–9. 10.1111/j.1525-1594.2008.00633.x [DOI] [PubMed] [Google Scholar]
  • 39.Emmanuel S, Jansz P, McGiffin D, Kure C, Watson A, Connellan M, et al. Anatomical human fitting of the bivacor total artificial heart. Artif Organs. (2022) 46:50–6. 10.1111/aor.14077 [DOI] [PubMed] [Google Scholar]
  • 40.Fresiello L, Najar A, Brynedal Ignell N, Zieliński K, Rocchi M, Meyns B, et al. Hemodynamic characterization of the realheart® total artificial heart with a hybrid cardiovascular simulator. Artif Organs. (2022) 46:1585–96. 10.1111/aor.14223 [DOI] [PubMed] [Google Scholar]
  • 41.Kelly N, McCree D, Fresiello L, Brynedal Ignell N, Cookson A, Najar A, et al. Video-based valve motion combined with computational fluid dynamics gives stable and accurate simulations of blood flow in the realheart total artificial heart. Artif Organs. (2022) 46:57–70. 10.1111/aor.14056 [DOI] [PubMed] [Google Scholar]
  • 42.Pieper I, Sonntag S, Meyns B, Hadi H, Najar A. Evaluation of the novel total artificial heart realheart in a pilot human fitting study. Artif Organs. (2020) 44:174–7. 10.1111/aor.13542 [DOI] [PubMed] [Google Scholar]
  • 43.Bhatia M, Jia S, Smeltz A, Kumar P. Right heart failure management: Focus on mechanical support options. J Cardiothorac Vasc Anesth. (2022) 36(8 Pt B):3278–88. 10.1053/j.jvca.2022.02.022 [DOI] [PubMed] [Google Scholar]
  • 44.Borisenko O, Wylie G, Payne J, Bjessmo S, Smith J, Yonan N, et al. Thoratec centrimag for temporary treatment of refractory cardiogenic shock or severe cardiopulmonary insufficiency: A systematic literature review and meta-analysis of observational studies. Asaio J. (2014) 60:487–97. 10.1097/MAT.0000000000000117 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Ahmad U, Khattab M, Schaelte G, Goetzenich A, Foldenauer A, Moza A, et al. Combining minimally invasive surgery with ultra-fast-track anesthesia in heartmate 3 patients: A pilot study. Circ Heart Fail. (2022) 15:e008358. 10.1161/CIRCHEARTFAILURE.121.008358 [DOI] [PubMed] [Google Scholar]
  • 46.Anderson M, Morris D, Tang D, Batsides G, Kirtane A, Hanson I, et al. Outcomes of patients with right ventricular failure requiring short-term hemodynamic support with the impella rp device. J Heart Lung Transplant. (2018) 37:1448–58. 10.1016/j.healun.2018.08.001 [DOI] [PubMed] [Google Scholar]
  • 47.Salna M, Garan A, Kirtane A, Karmpaliotis D, Green P, Takayama H, et al. Novel percutaneous dual-lumen cannula-based right ventricular assist device provides effective support for refractory right ventricular failure after left ventricular assist device implantation. Interact CardioVasc Thorac Surg. (2020) 30:499–506. 10.1093/icvts/ivz322 [DOI] [PubMed] [Google Scholar]
  • 48.Tschöpe C, Van Linthout S, Klein O, Mairinger T, Krackhardt F, Potapov E, et al. Mechanical unloading by fulminant myocarditis: Lv-impella, ecmella, bi-pella, and propella concepts. J Cardiovasc Transl Res. (2019) 12:116–23. 10.1007/s12265-018-9820-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Kuchibhotla S, Esposito M, Breton C, Pedicini R, Mullin A, O’Kelly R, et al. Acute biventricular mechanical circulatory support for cardiogenic shock. J Am Heart Assoc. (2017) 6:e006670. 10.1161/JAHA.117.006670 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Nagy C, Jumean M, Pham D, Kiernan M, Denofrio D, Kapur N. Percutaneous circulatory support for biventricular failure. Circ Cardiovasc Interv. (2013) 6:e12–4. 10.1161/CIRCINTERVENTIONS.112.000018 [DOI] [PubMed] [Google Scholar]
  • 51.Kapur N, Jumean M, Ghuloom A, Aghili N, Vassallo C, Kiernan M, et al. First successful use of 2 axial flow catheters for percutaneous biventricular circulatory support as a bridge to a durable left ventricular assist device. Circ Heart Fail. (2015) 8:1006–8. 10.1161/CIRCHEARTFAILURE.115.002374 [DOI] [PubMed] [Google Scholar]
  • 52.Aghili N, Bader Y, Vest A, Kiernan M, Kimmelstiel C, DeNofrio D, et al. Biventricular circulatory support using 2 axial flow catheters for cardiogenic shock without the need for surgical vascular access. Circ Cardiovasc Interv. (2016) 9:e003636. 10.1161/CIRCINTERVENTIONS.116.003636 [DOI] [PubMed] [Google Scholar]
  • 53.Crespo-Leiro M, Costanzo M, Gustafsson F, Khush K, Macdonald P, Potena L, et al. Heart transplantation: Focus on donor recovery strategies, left ventricular assist devices, and novel therapies. Eur Heart J. (2022) 43:2237–46. 10.1093/eurheartj/ehac204 [DOI] [PubMed] [Google Scholar]
  • 54.López-Sainz Á, Barge-Caballero E, Barge-Caballero G, Couto-Mallón D, Paniagua-Martin M, Seoane-Quiroga L, et al. Late graft failure in heart transplant recipients: Incidence, risk factors and clinical outcomes. Eur J Heart Fail. (2018) 20:385–94. 10.1002/ejhf.886 [DOI] [PubMed] [Google Scholar]
  • 55.Bhalla S, Dubey G, Basu S, Kanangat S, Dobrescu C, Nath D. Description and analysis of patients and outcomes following third-time heart transplantation: An analysis of the united network for organ sharing database from 1985 to 2017. Clin Transplant. (2019) 33:e13482. 10.1111/ctr.13482 [DOI] [PubMed] [Google Scholar]
  • 56.Zadikany R, Cole R, Chang D, Patel J, Geft D, Czer L, et al. Total artificial heart as bridge to cardiac retransplantation. Asaio J. (2021) 67:e77–9. 10.1097/MAT.0000000000001217 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Shah P, Ha R, Singh R, Cotts W, Adler E, Kiernan M, et al. Multicenter experience with durable biventricular assist devices. J Heart Lung Transplant. (2018) 37:1093–101. 10.1016/j.healun.2018.05.001 [DOI] [PubMed] [Google Scholar]
  • 58.Tran H, Pollema T, Silva Enciso J, Greenberg B, Barnard D, Adler E, et al. Durable biventricular support using right atrial placement of the heartware hvad. Asaio J. (2018) 64:323–7. 10.1097/MAT.0000000000000645 [DOI] [PubMed] [Google Scholar]
  • 59.Maynes E, O’Malley T, Patel P, Prochno K, Wood C, Choi J, et al. Right atrial versus right ventricular heartware hvad position in patients on biventricular heartware hvad support: A systematic review. Artif Organs. (2020) 44:926–34. 10.1111/aor.13675 [DOI] [PubMed] [Google Scholar]
  • 60.Villa C, Morales D. The total artificial heart in end-stage congenital heart disease. Front Physiol. (2017) 8:131. 10.3389/fphys.2017.00131 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Frontiers in Cardiovascular Medicine are provided here courtesy of Frontiers Media SA

RESOURCES