. 2022 Aug 16;2(6):568–575. doi: 10.1021/acsmeasuresciau.2c00036

A Solid-Contact Reference Electrode Based on Silver/Silver Organic Insoluble Salt for Potentiometric Ion Sensing

Shiyu Gan , Chunxian Liao , Rongfeng Liang , Sanyang Du , Lijie Zhong †,*, Yitian Tang , Tingting Han , Yu Bao , Zhonghui Sun , Yingming Ma , Li Niu †,*
PMCID: PMC9886000  PMID: 36785773

Abstract

graphic file with name tg2c00036_0007.jpg

Solid-contact ion-selective electrodes are a type of ion measurement devices that have been focused in wearable biotechnology based on the features of miniaturization and integration. However, the solid-contact reference electrodes (SC-REs) remain relatively less focused compared with numerous working (or indicator) electrodes. Most SC-REs in wearable sensors rely on Ag/AgCl reference electrodes with solid electrolytes, for example, the hydrophilic electrolyte salts in polymer matrix, but face the risk of electrolyte leakage. Herein, we report a type of SC-REs based on the silver/silver tetraphenylborate (Ag/AgTPB) organic insoluble electrode. The SC-RE consists of a Ag substrate, a solid contact (AgTPB), and a plasticized poly(vinyl chloride) (PVC) membrane containing the hydrophobic organic salt of tetrabutylammonium tetraphenylborate (TBATPB). The potentiometric measurements demonstrated that the SC-RE of Ag/AgTPB/PVC-TBATPB showed a reproducible standard potential in various electrolytes and disclosed high long-term stability. This SC-RE was further fabricated on a flexible substrate and integrated into all-solid-state wearable potentiometric ion sensor for sweat Cl monitoring.

Keywords: reference electrode, solid-contact, potentiometric sensing, ion-selective electrodes, wearable sensors

1. Introduction

Solid-contact ion-selective electrodes (SC-ISEs) are a type of miniaturized potentiometric analytical devices for ion and biochemical analysis and monitoring in complex environments.15 For example, environmental trace heavy metal analysis remains a significant issue hitherto.6 Recently focused wearable sensors have also accelerated the development of SC-ISEs due to the requirements of ion monitoring in human biological fluids,712 typically like sweat. Since the rudiment of SC-ISEs (coated wire electrodes)13 was proposed in 1971, intensive research has focused on improving the performances, particularly for the potential stability through development of advanced solid contacts.2,4,14 In addition to the working (or indicator) electrodes, the solid-contact reference electrodes (SC-REs) as an indispensable part of integrated SC-ISEs are of the same significance and even more challenging. A reliable SC-RE is a prerequisite for the development of highly reproducible15 and even calibration-free16 solid-contact potentiometric sensors. Traditional liquid-junction REs consist of a reference element and a salt bridge filled with concentrated aqueous salt solution (internal reference solution). The liquid-junction REs have a stable and fixed potential because of the inner solution with a constant concentration and ionic composition that determines the phase boundary potential, typically like the interface between the metal/metal insoluble salt and inner electrolyte solution.

Based on a similar strategy for traditional salt bridges, a few approaches of constructing SC-REs have been proposed.1719 The popular RE element is the Ag/AgCl electrode due to its simple fabrication and biocompatibility. The crucial point is the solid electrolyte compositions. The often-used method is mixing concentrated electrolytes into hydrophobic polymeric membranes, such as, hydrophilic and lipophilic salts.2024 For example, most SC-REs in wearable sensors are based on Ag/AgCl with polymer mixed electrolytes (KCl or NaCl).2529 However, there is a risk that the electrolytes could continuously and slowly leak into the sample solution. The other representative SC-REs are the ionic-liquid-based reference electrodes.3034 A stable potential could be formed between ionic liquid and aqueous phases by the distribution equilibrium. However, the ionic liquids still exhibit a certain solubility in an aqueous solution.35 Conducting polymers36 and carbon materials3739 have been also used as SC materials for the fabrication of SC-REs but face the same challenges of solid electrolyte leaking. Recently reported Ag/AgI SC-RE has identified a truly all-solid-state RE without a solid electrolyte layer.40 The electrolyte (I) was in situ produced by a current pulse (self-referencing pulstrode). This Ag/AgI SC-RE has been demonstrated for feasibility in sample analysis. However, the local I concentration should be enough to maintain the potential stability and avoid interferences from sulfide containing species. In addition, the current pulse might be a relatively complex method in the application of wearable sensors.

In this work, we proposed a type of SC-RE based on a silver/silver tetraphenylborate organic insoluble salt (Ag/AgTPB) electrode containing tetrabutylammonium tetraphenylborate (TBATPB) as a solid electrolyte (Figure 1a). The TBATPB is a hydrophobic electrolyte that has very low solubility in water (∼0.26 μM),41 which could overcome the issue of leaking. We investigated the potential stability of Ag/AgTPB SC-RE in various electrolyte solutions with examinations of concentration, light/gas, and redox interferences. This SC-RE even shows long-term stability in deionized water. Moreover, the Ag/AgTPB SC-RE was fabricated on a flexible substrate and integrated with a Cl-SC-ISE for on-body sweat analysis.

Figure 1.

Figure 1

Principle and performances for the SC-RE of Ag/AgTPB/PVC-TBATPB electrodes. (a) Basic structure and response principle. (b) Potentiometric measurements for the Ag/AgTPB/PVC-TBATPB electrodes (n = 5) in a LiCl electrolyte under the concentration range of 10–5 to 10–1 M as an example. The numbers on the curves represent the logarithms of the electrolyte concentrations. The inset shows the schematic of fabricated SC-RE. (c) The potential responses for the SC-RE in different electrolytes. The EMF is shown in an offset model along the y-axis. (d) Interferences tests of light, gas, and redox couples. The potential measurements for gas and light were carried out in 0.1 M NaCl.

2. Experimental Section

2.1. Reagents

Tetrabutylammonium chloride (TBACl), sodium tetraphenylborate (NaTPB), and acetonitrile were purchased from Innochem. High molecular weight poly(vinyl chloride) (PVC), 2-nitrophenyl octyl ether (NPOE), tetrahydrofuran (THF), and sodium nitrate were obtained from Sigma-Aldrich. Other inorganic salts were used at least in analytical grade. All the reagents were used by ultrapure water (>18.2 MΩ·cm) to prepare aqueous solutions without further treatment.

2.2. Preparation of Solid-Contact Reference Electrodes (SC-REs)

The silver disk electrodes (3 mm diameter) were polished with 1 and 0.3 μm alumina powder and then ultrasonically cleaned with water–ethanol–water successively. The silver electrodes were then placed in a 0.01 M NaTPB-0.067 M NaNO3 mixture solution. The electrodeposition was carried out through a constant current of 5 μA for 2800 s by a Gamry electrochemical workstation (reference 600 plus). A thin layer of silver tetraphenylborate (AgTPB) was deposited on the Ag electrode (Ag/AgTPB). The Ag/AgTPB was further covered by dropping 30 μL of PVC-based reference electrolyte membrane solution and dried at ambient temperature overnight. The reference electrolyte membrane solution was obtained by dissolving 200 mg of the membrane cocktail in 2 mL of THF: 28.5 wt % PVC and 56.9 wt % NPOE and 14.6 wt % tetrabutylammonium tetraphenylborate (TBATPB). The TBATPB was prepared by the precipitation reaction between NaTPB and TBACl in a molar ratio of 1:1 in aqueous solution and was recrystallized in acetone. The final SC-RE is described as Ag/AgTPB/PVC-TBATPB. Before the test, the SC-RE was conditioned in 10–5 M NaCl solution overnight.

2.3. Electrochemical Characterizations

All potentiometric measurements of the SC-REs of Ag/AgTPB/PVC-TBATPB electrodes were recorded by a multi-channel potentiometer EMF6 (Lawson Lab, Inc). The potential response tests were measured in various electrolyte solutions with the concentration range from 10–5to 10–1 M. The potentiometric tests were also performed through other two modes including different electrolytes under the same concentration and random electrolytes with random concentrations. The anti-interference experiments toward light, gas, and redox were performed. The water-layer tests for the Ag/AgTPB/PVC-TBATPB electrodes were carried out through varied solutions of 0.1 M NaCl/ 0.1 M LiCl/0.1 M NaCl. The long-term stability was examined in both 0.1 M NaCl and deionized (DI) water. A commercial Hg/Hg2Cl2/ saturated KCl/1 M LiOAc (SCE) was used as a reference electrode. The ion activities were calculated using the Debye–Hückel equation.

2.4. Preparation of Flexible Sensor

A polyethylene terephthalate (PET) substrate (7 × 7 cm) was cleaned by water–acetone–isopropanol successively followed by O2 plasma etching for 2 min. Then, a 300 nm silver layer was deposited on the PET using ultra-high vacuum sputtering technology (AJA Orion 5) through a silver target. The sputtering parameters include the sputtering speed of 2.5 A/s, the sputtering power of 60 W and the sputtering time of 1200 s. A polydimethylsiloxane (PDMS) insulating layer was further coated except for the regions of the working and reference electrodes (d = 5 mm). The electrode was then dried at 90 °C for 50 min. The flexible SC-RE was further obtained by an oxidizing sputtered silver electrode in acetonitrile solution containing 0.01 M NaTPB under a constant current of 5 μA for 30 min (Ag/AgTPB), and then, 20 μL of reference electrolyte membrane solution was dropped and dried at room temperature (Ag/AgTPB/PVC-TBATPB). The flexible SC-RE was finally conditioned in 10–5 M NaCl solution overnight before use. For the working electrode of Cl-SC-ISEs, the process is similar to a previous work.42 The sputtered Ag electrode was further oxidized in 0.05 M FeCl3 for 10 s to obtain the flexible Ag/AgCl electrode for sweat Cl detection. Three Cl-SC-ISEs were fabricated in parallel, which could simultaneously measure the concentration of sweat Cl.

2.5. Sweat Analysis

A commercial sweat guide band was used to fix the prepared flexible electrode. Combining with a multi-channel mini-potentiometer (MEK1000, Guangdong Dynechem Electronic Technology Co., Ltd.) and wireless data transmission mode, the variation trend of chloride ion concentration in the sweat produced by the volunteer during exercising was analyzed. Three Cl sensors were worn by the volunteer and measured at the same time. Before starting the exercise, the volunteer’s forehead was wiped with medical alcohol and deionized water in turn. Then, the volunteer exercised for a period to produce sweat. The concentration of chloride ion in sweat was recorded by the flexible wearable sensor. The sweat sample was collected and diluted for ex situ analysis by ion chromatography (CIC-D120, Qingdao Shenghan, Co., Ltd). It should be noted that the volunteer agrees with this test of wearable sensor for sweat analysis.

3. Results and Discussion

3.1. Characterizations of the SC-RE of Ag/AgTPB/PVC-TBATPB

The SC-RE was prepared by the electrodeposition of AgTPB on polished clean Ag electrodes in a NaTPB-NaNO3 electrolyte solution (see the Experimental Section for details). After electrodeposition for 2800 s at a constant current of 5 μA, a thin layer of AgTPB was covered on the Ag electrode. Further, the solid polymeric electrolyte containing PVC, NPOE, and TBATPB electrolytes was drop-casted on the surface of AgTPB to form final SC-RE of Ag/AgTPB/PVC-TBATPB (Figure 1a). The PVC-TBATPB is a hydrophobic membrane (∼113o water contact angle, Figure S1), and the TBATPB electrolyte has very low solubility in water (∼0.26 μM).41 In addition, the AgTPB should be either insoluble in organic NPOE since it has been used as an organic interface electrode in the early studies of liquid/liquid interface electrochemistry.43,44 The basic principle is like the case of Ag/AgCl-based SC-REs. An electron-coupled ion transfer reaction mechanism occurs at Ag/AgTPB and AgTPB/PVC-TBATPB interfaces, which can be described by AgTPB (s) + e ↔ Ag(s) + TPB(PVC) (Figure 1a). A reliable SC-RE should not respond to a broad concentration range of any electrolyte solutions. To verify this point, the potential responses to a series of salt solutions were evaluated in a concentration range from 10–5 to 10–1 M against a commercial double-junction SCE. For example, the potential response curves in a LiCl electrolyte under a four magnitude concentration are shown in Figure 1b. The Ag/AgTPB/PVC-TBATPB electrodes show small potential variation even when exposed to different ionic activities of electrolyte solutions. The E° value was determined to be −42.2 ± 1.1 mV. For other electrolytes, the potential response curves are presented in Figure 1c and Figure S2. The potentials basically disclose small variations for the whole concentrations. Relatively apparent potential fluctuations are observed for NH4+ and Ca2+ at a high concentration of 0.1 M, which is possibly ascribed to cation interaction with the TPB anion. However, there is no apparent effect under concentration of less than 10 mM. At least, it should work for sweat analysis since their concentration levels in sweat are relatively low.7 The E° is between −38.5 and −44.9 mV, and the maximum standard derivation (STD) is ∼2.5 mV for these electrolytes (Figure S2). The reproducibility for E° in these six types of electrolytes is determined to −41.7 ± 2.3 mV (Figure S2f). This performance is comparable to state-of-the-art Ag/AgCl-based SC-REs with STD of ∼2 mV for E° (Table S1).

The anti-interference abilities of the SC-RE of Ag/AgTPB/PVC-TBATPB toward light, gas, and redox were further evaluated. As shown in Figure 1d, the SC-RE is almost unaffected by the light (< 0.5 mV). For the gas interference, the SC-RE shows a stable potential under N2 and O2, while a minor potential variation (∼ 5 mV) resulted for CO2. It should be noted that the concentration of CO2 in the solution herein is close to a saturated state by purging CO2 gas in the solution. However, for most of systems, the low concentration of CO2 could have less effect on the potential fluctuation. For the redox sensitivity, the SC-RE was measured in five different concentration ratios of K3[Fe(CN)6]/K4[Fe(CN)6] (log CFe3+/Fe2+ = −1 to 1). The SC-RE does not either show considerable sensitivity to redox species. The total potential variation is about 5 mV when the concentration is changed by two magnitudes.

The potential responses for the Ag/AgTPB/PVC-TBATPB electrodes were further examined in randomly selected types and concentrations of electrolytes according to analytical quality protocol for solid-state reference electrodes.45,46 Under different electrolytes with the same concentration of 1 mM (Figure 2a), the potential variation is determined to 4.6 ± 0.9 mV. For the other two concentrations of 10–5and 10–1 M (Figure 2b and Figure S3), the electrodes show the potential variations of 4.6 ± 0.8 and 10.6 ± 1.8 mV, respectively. It is reasonable that the high ionic strength has a relatively significant effect since the phase boundary potential between PVC-TBATPB and water interfaces is affected due to ion distribution in the two immiscible phases. Further harsh examinations were tested in randomly selected electrolytes (Figure 2c,d). The potential variation is around 11.1 ± 0.4 mV. This result is comparable and even better than representative Ag/AgCl-based SC-REs based on lipophilic salts23 (∼11.7 mV) and ionic liquids-based SC-REs30 (15–25 mV).

Figure 2.

Figure 2

Potential responses for the Ag/AgTPB/PVC-TBATPB electrodes (n = 5) in (a, b) different electrolytes with the same concentration and (c, d) in randomly selected types and concentrations of electrolytes. The EMF in (b) is shown in an offset model along the y-axis.

For the proposed SC-RE of Ag/AgTPB/PVC-TBATPB, there are in fact three phase boundary interfaces, including Ag/AgTPB, AgTPB/PVC-TBATPB, and PVC-TBATPB/solution interfaces. The phase boundary potentials at first two interfaces are determined by the Nernst equation of Ag and AgTPB redox couple. The phase boundary potential at the PVC-TBATPB/solution interface is dependent on the ion distribution between the hydrophobic membrane phase of PVC-TBATPB and the sample aqueous solution. According to the early studies of ion transfers at the liquid/liquid interface,47 the phase boundary potential is determined to by their standard Gibbs transfer energies of TBA+ and TPB ions (see the supplementary discussion in the Supporting Information). The phase boundary potential could keep a constant value. This might be the reason that the potential of the proposed SC-RE could be basically stable even though the type or concentration of ions changes.

3.2. Water-Layer Test and Potential Stability

Another significant basis for the SC-REs is the potential stability. We first performed the water-layer tests for the Ag /AgTPB/PVC-TBATPB electrode since it consists of a PVC-based solid electrolyte membrane. The water layer might exist at the interface between the AgTPB and PVC membrane. According to a general protocol,48 the electrode was tested in two types of solutions, first in 0.1 M NaCl, then in 0.1 M LiCl, and back to 0.1 M NaCl (Figure 3a). When the solution was changed from NaCl to LiCl, the potential variation was 1.2 ± 1.2 mV. Upon going back to NaCl from LiCl, the response reaches equilibrium rapidly with a potential variation of only 0.3 ± 0.2 mV. This result indicates no existence of a water layer. Subsequently, we examined the long-term stability (23 days) of the electrodes in 0.1 M NaCl aqueous solution. As shown in Figure 3b, the potential fluctuation (ΔE) is determined around ∼14.5 ± 4.4 μV h–1. It should be noted that the fluctuation was determined from the difference between the maximum and minimum potentials. The SC-RE is generally relatively stable in electrolyte solution, for example, the quasi-reference electrode, typically a bare silver wire, which could keep somewhat stable in an electrolyte solution at a constant concentration. Thus, a harsh examination was performed in deionized (DI) water. As shown in Figure 3c, the SC-RE maintains relatively stable (∼39.1 ± 6.1 μV h–1) near 18 days although there is a potential oscillation possibly due to the high resistance. We have compared the stability with representative Ag/AgCl-based reference electrodes (Table S1). The stability is comparable or even better than those electrodes.

Figure 3.

Figure 3

Water-layer and stability measurements. (a) Water-layer tests for the Ag/AgTPB/PVC-TBATPB electrodes (n = 6). (b, c) The long-term stability was examined in both 0.1 M NaCl and deionized (DI) water. The EMF is shown in an offset model along the y-axis. The EMF in the general non-offset model is shown in Figure S4.

3.3. Flexibility of Ag/AgTPB/PVC-TBATPB

The above measurements have confirmed the feasibility of the Ag/AgTPB SC-RE. It should be noted that all electrochemical characterizations are based on solid Ag column electrodes. In this section, we fabricate a flexible Ag/AgTPB SC-RE to examine the possibility in wearable sensing. The electrode was prepared by sputtering an Ag layer on a flexible PET substrate and followed by running the same procedure as the Ag column electrode. As shown in Figure 4a, the flexible Ag/AgTPB/PVC-TBATPB could undergo bending at 60 and 90° angles. Cross-section SEM images disclose that the thickness of the flexible Ag/AgTPB/PVC-TBATPB including the polyethylene terephthalate (PET) substrate was ∼200 μm (Figure 4b). The solid electrolyte layer does not separate or fall off under 60 and 90° bending (Figure 4c,d). Subsequently, potentiometric measurements for the flexible Ag/AgTPB/PVC-TBATPB were characterized. Under no bending state, the potentiometric responses in various electrolytes exhibit similar results like the Ag column electrode (Figure 4e and Figure S5). Under 60 and 90° bending, the electrodes still maintain relatively stable potentials (Figure 4f–g and Figures S6 and S7). However, it should be noted that the electrode-to-electrode reproducibility under 90° is less than that at 60°, possibly due to the decreased interfacial adhesion under high strength bending. We also examined the reproducibility of a single Ag/AgTPB/PVC-TBATPB electrode at 90° bending (Figure S8). It is found that the potential fluctuation for three tests is around 1–5 mV, which indicates a good reproducibility for a single electrode. Overall, the endurability under 60–90° bending should meet the requirements for the wearable sweat sensors.

Figure 4.

Figure 4

Potentiometric measurements of flexibility Ag/AgTPB SC-RE. (a) Photo image of Ag/AgTPB/PVC-TBATPB under no bending and 60 and 90° bending. (b–d) Scanning electronic microscopy (SEM) images of Ag/AgTPB/PVC-TBATPB under different bendings. (e–g) Potentiometric measurements in different electrolytes under different bendings (n = 3). The EMF is shown in an offset model along the y-axis.

3.4. Wearable Cl Sensor for Sweat Analysis

Sweat not only contains a variety of electrolytes and metabolites that can provide healthy information but also has non-invasive characteristic of sampling and analysis.7 As the main anion in extracellular fluid, the chloride ion (Cl) plays an important role in maintaining acid–base balance, osmotic pressure distribution, and muscular activity.49 Herein, we integrate a working electrode of Cl-SC-ISE and a SC-RE of Ag/AgTPB/PVC-TBATPB on a flexible PET substrate for on-body sweat Cl analysis (Figure 5a). The Cl-SC-ISE was based on the bare Ag/AgCl electrode. The detailed preparation is presented in the Experimental Section. The integrated wearable Cl sensor was worn on the volunteer’s forehead in combination with a mini-potentiometer and a mobile battery. While the volunteer exercised on a spinning bike, the sensor performs a real-time analysis of sweat Cl. At the same time, the potential signal is displayed wirelessly transmitted in a customized mobile application program that matches the above mini-potentiometer.

Figure 5.

Figure 5

Wearable potentiometric Cl sensing. (a) Photo image of the potentiometric Cl sensor chip integrated with a Cl indicator electrode of a bare Ag/AgCl electrode and SC-RE of Ag/AgTPB/PVC-TBATPB. (b) On-body sweat Cl measurements (n = 3). (c) Ex situ analysis for sweat Cl by ion chromatography. The sweat was collected after cooling down. (d) Comparison of testing results between the wearable sensor and ion chromatograph. (e) Potentiometric calibration curves for the Cl sensor before and after sweat measurements.

As shown in Figure 5b, after ∼13 min exercise, the perspiration starts, and the potential response shows a relatively stable result. With the amount of sweat further increasing, the concentration of Cl increases gradually to a stable value of ∼57.7 ± 8.5 mM. After cooling down, a microtube was used to collect the sweat dripping. After dilution, the sweat sample was analyzed by the ion chromatograph for comparison (Figure 5c). The Cl concentration was determined 54 mM by ex situ analysis. The test results demonstrate that the wearable sensor has a high accuracy with a relative error of 7.0% compared with the ion chromatograph (Figure 5d). Subsequently, the sensor was taken off and measured again for the calibration curve (Figure 5e). The sensor maintains a Nernst response toward Cl after sweat tests. The calibration curves show the potential difference ∼10–20 mV for the low concentrations (10–5 and 10–4 M) but nearly overlap over 1 mM (Figure 5e). Since the Cl in sweat is around 50 mM, which demonstrates the accuracy of test result. Overall, the flexible wearable sensor integrated with Ag/AgTPB/PVC-TBATPB has shown relatively high reliability for on-line sweat analysis.

4. Conclusions

SC-REs remain a daunting challenge in the solid potentiometric ion sensing. In this work, we have developed an Ag/AgTPB/PVC-TBATPB organic insoluble salt electrode as an SC-RE. Compared with the state-of-the-art Ag/AgCl-based SC-REs, the Ag/AgTPB/PVC-TBATPB has demonstrated high potential stability, particularly under the rigorous examination in deionized water. Additionally, the SC-RE exhibited flexibility and could maintain potential stability under a certain mechanical bending strength. A proof-of-concept application of sweat Cl monitoring finally demonstrated that this SC-RE could be used for wearable sensors.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (21974032, 21974031, and U2006208), Science and Technology Research Project of Guangzhou (202102020622), and the Department of Science and Technology of Guangdong Province (2019B010933001).

Supporting Information Available

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsmeasuresciau.2c00036.

  • Supplementary figures for the hydrophobicity of membrane (Figure S1) and potentiometric responses of SC-REs (Figures S2–S8) and a comparison of analytical performances (Table S1) (PDF)

The authors declare no competing financial interest.

Supplementary Material

tg2c00036_si_001.pdf (1.1MB, pdf)

References

  1. Bobacka J.; Ivaska A.; Lewenstam A. Potentiometric Ion Sensors. Chem. Rev. 2008, 108, 329–351. 10.1021/cr068100w. [DOI] [PubMed] [Google Scholar]
  2. Hu J.; Stein A.; Bühlmann P. Rational design of all-solid-state ion-selective electrodes and reference electrodes. TrAC, Trends Anal. Chem. 2016, 76, 102–114. 10.1016/j.trac.2015.11.004. [DOI] [Google Scholar]
  3. Zdrachek E.; Bakker E. Potentiometric Sensing. Anal. Chem. 2019, 91, 2–26. 10.1021/acs.analchem.8b04681. [DOI] [PubMed] [Google Scholar]
  4. Zdrachek E.; Bakker E. Potentiometric Sensing. Anal. Chem. 2021, 93, 72–102. [DOI] [PubMed] [Google Scholar]
  5. Ding J.; Qin W. Recent advances in potentiometric biosensors. TrAC, Trends Anal. Chem. 2020, 124, 115803 10.1016/j.trac.2019.115803. [DOI] [Google Scholar]
  6. Lisak G. Reliable environmental trace heavy metal analysis with potentiometric ion sensors-reality or a distant dream. Environ. Pollut. 2021, 289, 117882 10.1016/j.envpol.2021.117882. [DOI] [PubMed] [Google Scholar]
  7. Bariya M.; Nyein H. Y. Y.; Javey A. Wearable sweat sensors. Nat. Electron. 2018, 1, 160–171. 10.1038/s41928-018-0043-y. [DOI] [Google Scholar]
  8. Bandodkar A. J.; Jeang W. J.; Ghaffari R.; Rogers J. A. Wearable Sensors for Biochemical Sweat Analysis. Annu. Rev. Anal. Chem. 2019, 12, 1–22. 10.1146/annurev-anchem-061318-114910. [DOI] [PubMed] [Google Scholar]
  9. Gao W.; Ota H.; Kiriya D.; Takei K.; Javey A. Flexible Electronics toward Wearable Sensing. Acc. Chem. Res. 2019, 52, 523–533. 10.1021/acs.accounts.8b00500. [DOI] [PubMed] [Google Scholar]
  10. Parrilla M.; Cuartero M.; Crespo G. A. Wearable potentiometric ion sensors. TrAC, Trends Anal. Chem. 2019, 110, 303–320. 10.1016/j.trac.2018.11.024. [DOI] [Google Scholar]
  11. Lyu Y.; Gan S.; Bao Y.; Zhong L.; Xu J.; Wang W.; Liu Z.; Ma Y.; Yang G.; Niu L. Solid-Contact Ion-Selective Electrodes: Response Mechanisms, Transducer Materials and Wearable Sensors. Membranes 2020, 10, 128. 10.3390/membranes10060128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Mohan A. M. V.; Rajendran V.; Mishra R. K.; Jayaraman M. Recent advances and perspectives in sweat based wearable electrochemical sensors. TrAC, Trends Anal. Chem. 2020, 131, 116024 10.1016/j.trac.2020.116024. [DOI] [Google Scholar]
  13. Cattrall R. W.; Freiser H. Coated wire ion-selective electrodes. Anal. Chem. 1971, 43, 1905–1906. 10.1021/ac60307a032. [DOI] [PubMed] [Google Scholar]
  14. Shao Y.; Ying Y.; Ping J. Recent advances in solid-contact ion-selective electrodes: functional materials, transduction mechanisms, and development trends. Chem. Soc. Rev. 2020, 49, 4405–4465. 10.1039/C9CS00587K. [DOI] [PubMed] [Google Scholar]
  15. Cheong Y. H.; Ge L.; Lisak G. Highly reproducible solid contact ion selective electrodes: Emerging opportunities for potentiometry - A review. Anal. Chim. Acta 2021, 1162, 338304 10.1016/j.aca.2021.338304. [DOI] [PubMed] [Google Scholar]
  16. Rousseau C. R.; Buhlmann P. Calibration-free potentiometric sensing with solid-contact ion-selective electrodes. TrAC, Trends Anal. Chem. 2021, 140, 116277 10.1016/j.trac.2021.116277. [DOI] [Google Scholar]
  17. Guth U.; Gerlach F.; Decker M.; Oelssner W.; Vonau W. Solid-state reference electrodes for potentiometric sensors. J. Solid State Electrochem. 2009, 13, 27–39. 10.1007/s10008-008-0574-7. [DOI] [Google Scholar]
  18. Vonau W.; Oelssner W.; Guth U.; Henze J. An all-solid-state reference electrode. Sens. Actuators, B 2010, 144, 368–373. 10.1016/j.snb.2008.12.001. [DOI] [Google Scholar]
  19. Michalska A. All-Solid-State Ion Selective and All-Solid-State Reference Electrodes. Electroanalysis 2012, 24, 1253–1265. 10.1002/elan.201200059. [DOI] [Google Scholar]
  20. Diamond D.; McEnroe E.; McCarrick M.; Lewenstam A. Evaluation of a new solid-state reference electrode junction material for ion-selective electrodes. Electroanalysis 1994, 6, 962–971. 10.1002/elan.1140061108. [DOI] [Google Scholar]
  21. Eine K.; Kjelstrup S.; Nagy K.; Syverud K. Towards a solid state reference electrode. Sens. Actuators, B 1997, 44, 381–388. 10.1016/S0925-4005(97)00146-9. [DOI] [Google Scholar]
  22. Kisiel A.; Michalska A.; Maksymiuk K.; Hall E. A. H. All-Solid-State Reference Electrodes with Poly(n-butyl acrylate) Based Membranes. Electroanalysis 2008, 20, 318–323. 10.1002/elan.200704065. [DOI] [Google Scholar]
  23. Mattinen U.; Bobacka J.; Lewenstam A. Solid-Contact Reference Electrodes Based on Lipophilic Salts. Electroanalysis 2009, 21, 1955–1960. 10.1002/elan.200904615. [DOI] [Google Scholar]
  24. Ding R.; Fiedoruk-Pogrebniak M.; Pokrzywnicka M.; Koncki R.; Bobacka J.; Lisak G. Solid reference electrode integrated with paper-based microfluidics for potentiometric ion sensing. Sens. Actuators, B 2020, 323, 128680 10.1016/j.snb.2020.128680. [DOI] [Google Scholar]
  25. Gao W.; Emaminejad S.; Nyein H. Y. Y.; Challa S.; Chen K.; Peck A.; Fahad H. M.; Ota H.; Shiraki H.; Kiriya D.; Lien D.-H.; Brooks G. A.; Davis R. W.; Javey A. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 2016, 529, 509–514. 10.1038/nature16521. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Nyein H. Y. Y.; Gao W.; Shahpar Z.; Emaminejad S.; Challa S.; Chen K.; Fahad H. M.; Tai L.-C.; Ota H.; Davis R. W.; Javey A. A Wearable Electrochemical Platform for Noninvasive Simultaneous Monitoring of Ca2+ and pH. ACS Nano 2016, 10, 7216–7224. 10.1021/acsnano.6b04005. [DOI] [PubMed] [Google Scholar]
  27. Nyein H. Y. Y.; Tai L.-C.; Quynh Phuong N.; Chao M.; Zhang G. B.; Gao W.; Bariya M.; Bullock J.; Kim H.; Fahad H. M.; Javey A. A Wearable Microfluidic Sensing Patch for Dynamic Sweat Secretion Analysis. ACS Sens. 2018, 3, 944–952. 10.1021/acssensors.7b00961. [DOI] [PubMed] [Google Scholar]
  28. An Q.; Gan S.; Xu J.; Bao Y.; Wu T.; Kong H.; Zhong L.; Ma Y.; Song Z.; Niu L. A multichannel electrochemical all-solid-state wearable potentiometric sensor for real-time sweat ion monitoring. Electrochem. Commun. 2019, 107, 106553 10.1016/j.elecom.2019.106553. [DOI] [Google Scholar]
  29. Xu J.; Zhang Z.; Gan S.; Gao H.; Kong H.; Song Z.; Ge X.; Bao Y.; Niu L. Highly Stretchable Fiber-Based Potentiometric Ion Sensors for Multichannel Real-Time Analysis of Human Sweat. ACS Sens. 2020, 5, 2834–2842. 10.1021/acssensors.0c00960. [DOI] [PubMed] [Google Scholar]
  30. Cicmil D.; Anastasova S.; Kavanagh A.; Diamond D.; Mattinen U.; Bobacka J.; Lewenstam A.; Radu A. Ionic Liquid-Based, Liquid-Junction-Free Reference Electrode. Electroanalysis 2011, 23, 1881–1890. 10.1002/elan.201100137. [DOI] [Google Scholar]
  31. Zhang T.; Lai C.-Z.; Fierke M. A.; Stein A.; Bühlmann P. Advantages and Limitations of Reference Electrodes with an Ionic Liquid Junction and Three-Dimensionally Ordered Macroporous Carbon as Solid Contact. Anal. Chem. 2012, 84, 7771–7778. 10.1021/ac3011507. [DOI] [PubMed] [Google Scholar]
  32. Zou X. U.; Chen L. D.; Lai C.-Z.; Bühlmann P. Ionic Liquid Reference Electrodes With a Well-Controlled Co(II)/Co(III) Redox Buffer as Solid Contact. Electroanalysis 2015, 27, 602–608. 10.1002/elan.201400274. [DOI] [Google Scholar]
  33. Lindner E.; Guzinski M.; Khan T. A.; Pendley B. D. Reference Electrodes with Ionic Liquid Salt Bridge: When Will These Innovative Novel Reference Electrodes Gain Broad Acceptance?. ACS Sens. 2019, 4, 549–561. 10.1021/acssensors.8b01651. [DOI] [PubMed] [Google Scholar]
  34. Anderson E. L.; Chopade S. A.; Spindler B.; Stein A.; Lodge T. P.; Hillmyer M. A.; Bühlmann P. Solid-Contact Ion-Selective and Reference Electrodes Covalently Attached to Functionalized Poly(ethylene terephthalate). Anal. Chem. 2020, 92, 7621–7629. 10.1021/acs.analchem.0c00296. [DOI] [PubMed] [Google Scholar]
  35. Kakiuchi T. Electrochemical Aspects of Ionic-Liquid|Water Two-Phase Systems. Anal. Chem. 2007, 79, 6442–6449. 10.1021/ac0719554. [DOI] [Google Scholar]
  36. Kisiel A.; Marcisz H.; Michalska A.; Maksymiuk K. All-solid-state reference electrodes based on conducting polymers. Analyst 2005, 130, 1655–1662. 10.1039/b510868c. [DOI] [PubMed] [Google Scholar]
  37. Rius-Ruiz F. X.; Kisiel A.; Michalska A.; Maksymiuk K.; Riu J.; Rius F. X. Solid-state reference electrodes based on carbon nanotubes and polyacrylate membranes. Anal. Bioanal. Chem. 2011, 399, 3613–3622. 10.1007/s00216-011-4689-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Xavier Rius-Ruiz F.; Bejarano-Nosas D.; Blondeau P.; Riu J.; Xavier Rius F. Disposable Planar Reference Electrode Based on Carbon Nanotubes and Polyacrylate Membrane. Anal. Chem. 2011, 83, 5783–5788. 10.1021/ac200627h. [DOI] [PubMed] [Google Scholar]
  39. Hu J.; Ho K. T.; Zou X. U.; Smyrl W. H.; Stein A.; Bühlmann P. All-Solid-State Reference Electrodes Based on Colloid-Imprinted Mesoporous Carbon and Their Application in Disposable Paper-based Potentiometric Sensing Devices. Anal. Chem. 2015, 87, 2981–2987. 10.1021/ac504556s. [DOI] [PubMed] [Google Scholar]
  40. Gao W.; Zdrachek E.; Xie X.; Bakker E. A Solid-State Reference Electrode Based on a Self-Referencing Pulstrode. Angew. Chem., Int. Ed. 2020, 59, 2294–2298. 10.1002/anie.201912651. [DOI] [PubMed] [Google Scholar]
  41. Silver B. R.; Holub K.; Mareček V. Spontaneous emulsification at surfactantless liquid/liquid interfaces. J. Electroanal. Chem. 2017, 805, 91–97. 10.1016/j.jelechem.2017.10.027. [DOI] [Google Scholar]
  42. Liao C.; Zhong L.; Tang Y.; Sun Z.; Lin K.; Xu L.; Lyu Y.; He D.; He Y.; Ma Y.; Bao Y.; Gan S.; Niu L. Solid-Contact Potentiometric Anion Sensing Based on Classic Silver/Silver Insoluble Salts Electrodes without Ion-Selective Membrane. Membranes 2021, 11, 959. 10.3390/membranes11120959. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Clarke D. J.; Schiffrin D. J.; Wiles M. C. A tetraphenylborate internal reference electrode for immiscible electrolyte solutions and ion selective electrodes. Electrochim. Acta 1989, 34, 767–769. 10.1016/0013-4686(89)87108-7. [DOI] [Google Scholar]
  44. Shao Y.; Weber S. G. Direct Observation of Chloride Transfer across the Water/Organic Interface and the Transfer of Long-Chain Dicarboxylates. J. Phys. Chem. 1996, 100, 14714–14720. 10.1021/jp9536864. [DOI] [Google Scholar]
  45. Mousavi Z.; Granholm K.; Sokalski T.; Lewenstam A. An analytical quality solid-state composite reference electrode. Analyst 2013, 138, 5216–5220. 10.1039/c3an00852e. [DOI] [PubMed] [Google Scholar]
  46. Granholm K.; Mousavi Z.; Sokalski T.; Lewenstam A. Analytical quality solid-state composite reference electrode manufactured by injection moulding. J. Solid State Electrochem. 2014, 18, 607–612. 10.1007/s10008-013-2294-x. [DOI] [Google Scholar]
  47. Samec Z. Electrochemistry at the interface between two immiscible electrolyte solutions (IUPAC Technical Report). Pure Appl. Chem. 2004, 76, 2147–2180. 10.1351/pac200476122147. [DOI] [Google Scholar]
  48. Fibbioli M.; Morf W. E.; Badertscher M.; de Rooij N. F.; Pretsch E. Potential Drifts of Solid-Contacted Ion-Selective Electrodes Due to Zero-Current Ion Fluxes Through the Sensor Membrane. Electroanalysis 2000, 12, 1286–1292. . [DOI] [Google Scholar]
  49. Choi D.-H.; Kitchen G. B.; Jennings M. T.; Cutting G. R.; Searson P. C. Out-of-clinic measurement of sweat chloride using a wearable sensor during low-intensity exercise. npj Digit. Med. 2020, 3, 49. 10.1038/s41746-020-0257-z. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

tg2c00036_si_001.pdf (1.1MB, pdf)

Articles from ACS Measurement Science Au are provided here courtesy of American Chemical Society

RESOURCES