Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2024 Apr 1.
Published in final edited form as: Magn Reson Med. 2022 Nov 13;89(4):1429–1440. doi: 10.1002/mrm.29532

Ultrafast Z-spectroscopic Imaging in vivo at 3T using through-slice spectral encoding (TS-UFZ)

Chongxue Bie 1,2,3, Peter C M van Zijl 1,2, Deng Mao 1,4, Nirbhay N Yadav 1,2,*
PMCID: PMC9892239  NIHMSID: NIHMS1847055  PMID: 36373181

Abstract

Purpose:

Acquisition of high-resolution Z-spectra for CEST or MTC MRI requires excessive scan times. Ultrafast Z-spectroscopy (UFZ) has been proposed to address this, however, the quality of in vivo UFZ spectra has been insufficient. Here, we present a simple approach to improve this.

Theory and Methods:

UFZ imaging acquires full Z-spectra by encoding the spectral dimension spatially via a gradient applied concurrently with the RF saturation pulse. Different from previous implementations, both this saturation gradient and its readout were applied in the slice direction, resulting in a relatively uniform voxel composition. Phase-encoding was applied in both in-plane directions, allowing additional under-sampling and acceleration.

Results:

In phantoms, UFZ imaging with through-slice Z-spectral encoding (TS-UFZ) provided Z-spectra of salicylic acid and egg white in excellent agreement with conventional acquisition. In vivo brain Z-spectra were influenced by flow through the imaging slice which affected the Z-spectral baseline. Still, CEST signals could be quantified after baseline fitting and mapping the residual CEST signal. Amide proton transfer (APT) contrast intensities obtained by TS-UFZ were on the same order of magnitude as conventional CEST but with different contrast across slice which likely is a result of different tissue regions contributing.

Conclusion:

TS-UFZ approach improves signal stability and spectral uniformity over previous implementations and allows high spectral-resolution imaging of saturation transfer effects in the human brain at 3T. This implementation allows for further acceleration by reducing phase encoding steps and thus opens up the possibility of mapping dynamic CEST signals in vivo with a practical temporal resolution.

Keywords: CEST, MTC, imaging, through-slice ultrafast Z-spectroscopy, APT

1. Introduction

Chemical exchange saturation transfer (CEST) MRI has generated great interest for molecular imaging applications since it can detect low concentration solute molecules via the water proton signal.15 In CEST MRI, exchangeable solute protons are selectivity saturated using radiofrequency (RF) irradiation and after the exchange of these protons to the bulk water pool, a reduction of water signal is observed. Through continuous RF saturation and repeated chemical exchange, multiple saturation transfer events lead to an enhanced reduction of the bulk water signal thus CEST MRI is able to amplify signals from low concentration molecules by several orders of magnitude1,6.

However, in vivo, multiple signal sources such as semi-solid magnetization transfer contrast (MTC)7, direct water saturation (DS)8, multiple types of exchangeable protons9, and relayed Nuclear Overhauser effect (rNOEs) of mobile macromolecules1,1014 complicate the interpretation of CEST images. CEST images are therefore often acquired as a function of saturation frequency offset to create voxel-based water saturation spectra containing all of these effects.15 These so-called Z-spectra, which, similar to MR spectroscopy, are frequency-referenced (but here to the water solvent at 0 ppm) aid in the discrimination of signal components and allows their assessment using spectral quantification methods. A larger number of frequency offsets results in higher spectral resolution but the sequential acquisition of a proportionally larger number of images leads to excessive scan times.

To accelerate the acquisition of full Z-spectra for the purpose of measuring MTC effects, Swanson proposed applying a constant magnetic field gradient concurrently with the saturation RF pulse to encode the frequency offsets spatially.16 Xu et al17 and Boutin et al18 also showed the utility of this approach to measure CEST and hyperpolarized agents, respectively, named ultrafast Z-spectroscopy (UFZ). These initial implementations demonstrated the feasibility of this approach in single uniform samples in vitro. Subsequent studies showed that the incorporation of slice selection can allow multiple sample tubes to be measured simultaneously1921. Although these approaches have been shown useful to accelerate the screening of a large number of samples in vitro, they are not yet easily applicable to map dynamic processes in vivo. More recently, several groups have implemented UFZ approaches in vivo using different strategies.22 Liu et al.23 implemented the UFZ approach in a point-resolved spectroscopy (PRESS) sequence, making it available for in vivo single-voxel spectroscopy. Wilson et al.24 further extended the UFZ method by using the localized PRESS sequence with reversed gradient for the acquisition of in vivo Z-spectroscopy. However, these previous in vivo implementations of UFZ could only obtain Z-spectra of a single voxel and not CEST maps across a slice.

Here, we improve on previous UFZ implementations by applying the saturation gradient and its imaging readout in the slice direction, while phase-encoding is applied in-plane in two directions. Application of the saturation gradient and its readout across the relatively narrow slice thickness, produces a Z-spectrum over a relatively uniform tissue volume. In addition, this approach allows further acceleration of the phase-encoding applied in-plane. This UFZ imaging with through-slice Z-spectral encoding (TS-UFZ) implementation was validated on a 3T clinical scanner using samples of salicylic acid and egg white. Finally, in vivo feasibility was shown by mapping amide proton transfer (APT) contrast in the human brain.

2. Methods

2.1. TS-UFZ Pulse Sequence

The TS-UFZ imaging technique employs a saturation gradient Gsat applied concurrently with a pseudo-continuous wave (CW) saturation pulse as well as a turbo spin-echo (TSE) readout, both in the slice direction; phase-encoding is applied in-plane (Figure 1). When the center of the slice coincides with Gsat isocenter (Δs = 0, whereΔs is the distance of the slice center to the Gsat isocenter), the presence of Gsat along the slice shifts the resonance frequency (Δω) of each spin pool depending on the gradient strength and the location (d) from Gsat center:

Δω(d)=γGsatd [1]

where γ is the gyromagnetic ratio of the observed nuclei (γ ≈ 267.5 × 106 rad/T for protons) and Gsat is the saturation gradient amplitude. Thus, an RF saturation pulse applied at the water resonance frequency (F0) will directly saturate water at the Gsat isocenter (d = 0). In this case, the frequency of the applied saturation pulse (Δωsat) is equal to the center frequency of the Z-spectrum (Δωc), i.e., Δωsats = 0) = Δωc = 0 ppm. For other values of d, the resonance frequency for each pool will shift in line with Eq. [1] thus the saturation pulse applied at F0 will saturate a different proton pool depending on d. Therefore, an entire z-spectrum can be encoded along the spatial direction of the saturation gradient. This spectral information is recovered by applying a readout gradient in the same direction as Gsat. The CEST spectral bandwidth, BWsat, is thus determined by the saturation gradient strength over the slice thickness (Ls) in rad/s:

BWsat=γGsatLs [2]

with the frequency offset range between [γGsat(−Ls/2), γGsat(Ls/2)]. The number of saturation offsets is decided by the number of datapoints in the slice direction (N), so the resolution of Z-spectrum is given by:

δωsat=BWsat/N [3]

When applied through a slice, each offset corresponds to a sub-slice. Thus, for a 10 mm slice thickness with 100 offsets, signal for data points in the Z-spectrum is coming from 100 μm sub-slices at each offset. Practically, the slice is often off-centered from the Gsat isocenter (Δs ≠ 0), resulting the Z-spectrum will not be centered at slice. Thus, the frequency of the applied saturation pulse Δωsat needs to be adjusted according to the shift to ensure the Z-spectrum is at the slice middle as well as Δωc = 0 ppm:

Δωsat(Δs)=Δωc+ΔsLsBWsat [4]

Also, it is not necessary for the center of the Z-spectrum to be placed in the middle of the slice, and in fact, shifting the Z-spectra (i.e., shifting Δωc) may be useful for sampling select Z-spectral regions of interest whilst minimizing signal loss due to spectral dispersion.

Figure 1.

Figure 1.

Principle of TS-UFZ imaging pulse sequence. A series of short saturation gradient pulses of strength Gsat is applied in the slice direction concurrently with the saturation pulses. The saturation gradient Gsat and turbo spin-echo sequence are applied in the slice direction and phase-encoding is applied in-plane. To compensate for possible droop on the gradient amplifier during the prolonged CEST irradiation period, a series of shorter gradient pulses (a total of 10 in this paper) were applied for spatial frequency encoding during preparation instead of a single long gradient pulse.

Following the acquisition of the saturation image, a reference scan is acquired with the saturation pulse center applied far off-resonance (e.g., Δωc=200 ppm). This reference signal (S0) at an offset where there is no saturation difference expected over the slice thickness is required to correct the Z-spectrum for proton density differences between the sub-slices, and to eliminate effects arising from gradient non-linearity and spatial irregularities of the object. By dividing saturated signal (Ssat) at each offset with the corresponding S0 at each, the normalized Z-spectral information can be recovered.

2.2. MRI acquisition

TS-UFZ was validated in vitro and then tested in vivo on three healthy volunteers. All MRI scans were performed on a 3T Philips Elition RX system (Philips Healthcare), with a 32-channel phased array head coil used as signal receiver and a quadrature body coil used as radiofrequency transmitter (Philips Healthcare). Gradients of up to 95 mT/m and 220 T/m/s in all three directions were available.

The TS-UFZ pulse sequence, as shown in Figure 1, used a 3D spin-echo (SE) with multi-shot TSE readout and consisted of a saturation period of 2 s with a pseudo-continuous saturation pulse consisting of a train of 40 sinc-gaussian shape pulses, each 50 ms in duration and B1,sat = 0.5 μT. Spectral encoding was achieved using a train of 10 gradient pulses (each 200 ms in duration) applied concurrently with the RF saturation pulse. The saturation period was followed by the TSE sequence. To allow for the magnetization recovery, a delay of 3s was included before each saturation module.

For comparison, conventional Z-spectral images using a 2D SE with single-shot TSE readout were acquired with the same saturation duration and B1 as the TS-UFZ imaging method, but Z-spectral data was acquired by varying the frequency offsets of the saturation pulse. See below for detailed parameters.

2.2.1. In vitro experiments

Salicylic acid phantom

Salicylic acid contains exchangeable protons resonating at 9.3 ppm from water.25 A 50 mM salicylic acid solution (Sigma Aldrich, St Louis, MO) was prepared in phosphate buffered saline (PBS) solution (pH of 7.5).

TS-UFZ imaging was performed with the following parameters: field of view (FOV) = 144 × 144 × 20 mm3, the slice thickness along the Gsat was Ls = 20 mm, the typical number of acquired datapoints in the slice direction (N) was 200, leading to a voxel size for the sub-slices of 8 × 8 × 0.1 mm3; Gsat = 4, 6, 8, 10 mT/m, the corresponding BWsat values were 21,400, 32,100, 42,800, and 53,500 rad/s (i.e., 26.6, 39.9, 53.2, and 66.7 ppm, respectively). TSE factor = 40, TE/TR = 21/5,873 ms, excitation/refocusing pulse angle = 90°/120°. The sensitivity encoding (SENSE)26 factors were set to 1 in both in-plane directions. Encoding was done with Δωc= 0 ppm with adjusted Δωsat (Eq. [4]), resulting the Z-spectrum was centered at the slice; S0 was acquired with Δωc = 200 ppm. The total scanning time including both Ssat and S0 was 2 mins and 15 s.

Conventional CEST imaging with FOV = 144 × 144 mm2, slice thickness = 20 mm, voxel size = 8 × 8 mm2, TSE factor = 9, TE/TR = 4.9/6,000 ms, excitation/refocusing pulse angle = 90°/120°. SENSE factor was set to 2 in the phase-encoding direction. A total of 58 saturated images (Ssat) were acquired from −5 to 12 ppm, and followed by 5 unsaturated images (S0) where the saturation pulse was applied at 100 ppm. The acquisition time for one scan was 6 mins and 24 s.

Egg white phantom

The experiments on egg white phantom had the same FOV as the salicylic acid phantom but used slice thicknesses (Ls) = 10 mm. TS-UFZ imaging was conducted for Gsat = 4, 6, 8, 10 mT/m, the corresponding BWsat values were 10,700, 16,050, 21,400, and 26,750 rad/s (i.e., 13.3, 20.0, 26.6, and 33.3 ppm, respectively). The number of N was 100. The scan at each Gsat was repeated three times to assess signal reproducibility. SENSE factors were set to 2 and 1 in two in-plane directions, respectively. And the total scan time of acquiring both Ssat and S0 was 53 s. For conventional CEST, a total of 54 saturated images (Ssat) were acquired from −5 to 5 ppm with 0.2 ppm step increment, as well as 7, 10, and 15 ppm, and 5 unsaturated images (S0). The acquisition time was 6 mins and 6 s.

2.2.2. In vivo experiments

Human studies were approved by the Johns Hopkins Medicine Institutional Review Board and performed on three healthy subjects, after informed consent was obtained.

The in vivo TS-UFZ experiments were performed similar to egg white with modified FOV parameters in both low and high in-plane resolutions, FOV = 150 × 206 × 10 mm3, Ls = 10 mm. Low spatial resolution: voxel size = 8 × 8 × 0.104 mm3, N was 96, TE/TR = 21/5,873 ms, the total scan time including both Ssat and S0 was 1 min and 28 s. High spatial resolution: voxel size = 5 × 5 × 0.12 mm3, N was 84, TE/TR = 19/5,792 ms, the total scan time for both Ssat and S0 was 3 mins and 58 s. Gsat = 6 mT/m, and the corresponding BWsat was 16,050 rad/s (i.e., 20.0 ppm).

Conventional CEST imaging was also performed with both low-resolution and high-resolution image acquisitions with FOV = 150 × 205 mm2, slice thickness = 10 mm. Low-resolution: voxel size = 8 × 8 mm2, TSE factor = 13. High-resolution: voxel size = 5 × 5 mm2, TSE factor = 21. Both TE/TR = 4.9/6,000 ms. The saturation frequency list and total scan time were same as egg white phantom.

2.3. Data processing

Both TS-UFZ and conventional CEST images were downsampled to achieve the same in-plane matrix size (32×32 for phantom, 33×33 for in vivo low-resolution, and 60×60 for in vivo high-resolution). For TS-UFZ, the Z-spectrum was constructed by normalizing the Ssat spectrum with the S0 spectrum. In contrast to the conventional WASSR-based B0 correction methods which shift the saturation spectrum, B0 field inhomogeneities were corrected for the position of the direct water saturation peak, where the nominal saturation frequency list for each voxel (calculated from Eqs. [2, 3]) was shifted so the frequency offset of water peak was at 0 ppm. The approach was based on the fact that shifting the spectrum would also have resulted in shifts in the spatial position of the voxels. For conventional CEST, the images with saturation were normalized by the corresponding image without saturation, and B0 correction was performed voxel-by-voxel by shifting the direct water saturation to 0 ppm27. Then, the Z-spectrum was interpolated to 0.2 ppm step size and employed PCA-based denoising28 to improve signal-to-noise (SNR) for both techniques.

The APT signal was extract using a modified PLOF method29,30. In detail, a partial Z-spectrum (Zacq) ranging from 1 to 6 ppm was used for fitting. First, the background spectrum (Zback, excluding the APT signal between [2.4, 4.4] ppm, including DS, MTC, and other CEST effects, was fitted with a mixed Lorentzian and polynomial function. Second, by subtracting Zback from the Zacq, a residual spectrum (Zresd) was obtained and integrated over the 3 to 4 ppm range to generate an APT contrast map. In addition, APTw images were obtained using the averaged MTR asymmetry (MTRasym) between 3 and 4 ppm, MTRasym (Δω) = Ssat (-Δω)/S0 - Ssat (+Δω)/S0. Finally, the in vivo APT contrast and APTW maps were smoothed by 3 × 3 median filter.

All code was written in-house with Python.

3. Results

Salicylic acid phantom

To assess the implementation of the TS-UFZ pulse sequence, experiments were performed on a salicylic acid phantom (exchangeable proton resonance around 9.3 ppm from water). The profile of the Ssat and S0 TS-UFZ images across the slice direction are shown in Figure 2. On the Ssat image, the black band at the center of the slice shows that the frequency of the applied saturation pulse Δωsat was adjusted thus the DS was at the center of the slice. A faint dark band labeled by the red arrow indicates the CEST effect corresponding to 9.3 ppm from water. From the saturated image, the Z-spectrum was recovered by normalizing Ssat with S0. As shown in Figure 2C, the Z-spectra averaged over four acquisitions with varied Gsat shows a small standard deviation (SD) across the spectrum and clearly shows the salicylic acid CEST peak, illustrating that this implementation of the TS-UFZ method can provide clear CEST signals and has good reproducibility.

Figure 2.

Figure 2.

TS-UFZ for a salicylic acid phantom (50 mM, pH = 7.5). (A) The illustration of slice chosen for TS-UFZ imaging. The salicylic acid phantom is labelled by yellow circle. (B) The profile of UFZ images across the slice direction with saturation (Ssat) and without saturation (S0) for the salicylic acid phantom. The red arrow labels the CEST effect on the Ssat image. (C) Comparison of Z-spectra from TS-UFZ imaging (mean ± SD) and conventional CEST MRI. The standard deviation (SD) is over 4 acquisitions using Gsat = 4, 6, 8, 10 mT/m.

Egg white phantom

Experiments were performed on egg white to investigate the ability of the TS-UFZ technique to map the APT effect. As shown in Figure 3AD, TS-UFZ Ssat and S0 spectra of egg white displayed negligible deviations over three repeat scans for each Gsat. The resulting Z-spectra with small SDs (n = 3) were in good agreement with conventional CEST and successfully distinguished the APT effect under various Gsat (Figures 3EH). However, a slight broadening of the direct water saturation peak was observed with increasing Gsat. The APT contrast was quantified from the residual spectrum by removing background signals from the acquired Z-spectrum, as illustrated in Figures 3IL.

Figure 3.

Figure 3.

TS-UFZ Z-spectra from an ROI in an egg white phantom for different saturation gradient strengths (Gsat = 4, 6, 8, 10 mT/m). (A-D) Mean intensities of Ssat and S0 signals with standard deviation (SD) of TS-UFZ. The SD is over 3 repeat acquisitions for each Gsat. (E-H) The comparison between Z-spectra (mean ± SD) of TS-UFZ (blue line) and conventional CEST (red circles). (I-L) The extraction of APT signals for TS-UFZ (Z-spectrum from one scan for each Gsat). The acquired Z-spectrum (black open circles), background data points included in the fit (blue crosses inside the black circles), and fitted background (black line) are scaled on the left vertical axis. The APT contrast is quantified by integrating the residual spectrum (red dotted line, scale on the right vertical axis) between 3 to 4 ppm (gray shadow).

APT contrast maps from the egg white phantom obtained using TS-UFZ and conventional CEST MRI are shown in Figures 4A and 4B, respectively, demonstrating that the TS-UFZ contrast is comparable to conventional CEST. Interestingly, TS-UFZ water shift maps displayed a clear dependence on Gsat and differed from water shift maps determined using conventional CEST (Figures 4C, D). The water peak shift of conventional CEST is caused by B0 inhomogeneity, whereas the TS-UFZ shift is due to a combination of B0 inhomogeneity and an imprecise calculation of the off-center slice position (Δs). Consequently, Δωc was not completely set to the water resonance frequency thus the water peak shifted depending on Gsat. This shift was corrected by shifting the nominal saturation frequency list in each voxel so that the frequency offset of water peak was assigned to 0 ppm.

Figure 4.

Figure 4.

Comparison of APT contrast (A, B) and water peak shift (C, D) between TS-UFZ and conventional CEST MRI for egg white phantom. APT contrast maps for (A) TS-UFZ under Gsat = 4, 6, 8, 10 mT/m and (B) conventional CEST. (C, D) The water peak shift maps for both acquisition techniques.

Human brain

The feasibility of TS-UFZ imaging was further tested on healthy volunteers, by imaging the coronal orientation of the brain. The Ssat and S0 TS-UFZ profiles (high in-plane resolution scan) across the slice direction are shown in Figures 5A, B. DS was centered at the middle sub-slice, which is clearly visible as a broad black band on the Ssat profile. TS-UFZ Ssat and S0 for a selected region in white matter are shown in Figure 5C. After normalization, the TS-UFZ Z-spectral baseline shows slightly lower saturation when compared to conventional CEST (Figure 5D). We tentatively attribute this to the saturation gradient (Gsat) making the sequence sensitive to flow through the slice. To account for this baseline difference, the APT signal was extracted by removing the background from the acquired Z-spectra, as shown in Figures 5E, F.

Figure 5.

Figure 5.

In vivo TS-UFZ results for subject 1 from the high in-plane resolution scan. (A, B) Profiles of TS-UFZ images in the slice direction with saturation (A, Ssat) and without saturation (B, S0). (C) Intensities of Ssat and S0 signals for the white matter ROI shown on a TS-UFZ image. (D) The Z-spectra of TS-UFZ and conventional CEST. (E, F) Extraction of APT signals for TS-UFZ and conventional CEST. The acquired Z-spectral data are shown as black open circles (scale on the left vertical axis), with selected background data points (circles filled with blue crosses) and fitted background (black line) also indicated. The APT contrast is quantified by integrating the residual spectrum (red line, scale on the right vertical axis) over the 3 to 4 ppm (gray shadow).

APT contrast maps were then obtained by integrating the residual curve, as shown in Figure 6. The TS-UFZ results show signal intensities in the Z-spectra that are on the same order of magnitude as conventional CEST results, but APT contrast across the slice differed between methods. The difference is likely due to the TS-UFZ image (1 ppm integral) representing a 0.5 mm sub-slice while the conventional CEST signal is the average over a 10 mm slice. In Figures 6D, G, there are clear differences in the APTW maps generated from the two methods. The TS-UFZ maps showed unexpected heterogeneity, possibly due to Z-spectral baseline shifts and tissue heterogeneity between the images acquired at each frequency about the water peak.

Figure 6.

Figure 6.

Comparison of APT contrast maps for a coronal slice through the midbrain of subject 1. (A) T2w image. (B, C) TS-UFZ and (E, F) conventional CEST for both low (voxel size = 8 × 8 mm2) and high in-plane resolution (voxel size = 5 × 5 mm2). (D, G) APTW maps obtained by using MTRasym analysis for the two techniques (high-resolution image).

4. Discussion

In this work, we proposed an UFZ imaging approach in which a saturation gradient and the readout gradient were applied in the slice direction and phase-encoding in both in-plane directions, named TS-UFZ. Previous in vivo UFZ implementations, such as ultrafast localized CEST-spectroscopy with PRESS (UCEPR) and gradient-reversed ultrafast Z-spectroscopy (GRUFZS)23,24, improved Z-spectral SNR by increasing the number of signal averages (NSA). The acquisition time was still significantly reduced compared to conventional CEST when the Z-spectroscopy of a single voxel was studied. However, increasing NSA is not practical for imaging a whole slice due to excessive scan times. Here, using TS-UFZ imaging, sufficient SNR could be achieved with NSA=1, the obtained APT contrast map resembled a conventional CEST approach for phantom experiments and on the same order of magnitude as conventional CEST for human brain.

The TS-UFZ Z-spectral resolution is determined by the resolution across the slice (Ls/N) at specific values of Gsat, which in this study (Ls = 10 mm, N = 100) resulted in a very high Z-spectral frequency resolution of 0.13 ppm when using Gsat = 4 mT/m, without requiring additional scan time. This frequency resolution ensures minimal chance of missing important Z-spectral features during CEST experiments without additional scan time as would be the case with conventional CEST where separate images need to be acquired per frequency leading to reduced spectral resolution for the same scan time. However, conventional CEST acquisition results in averaging over the whole slice thickness, while TS-UFZ acquisition presents signals from a sub-slice for each frequency offset and the thickness of each sub-slice is determined by N. The significant volume reduction for each frequency offset in TS-UFZ will result in the loss of in-plane SNR. Even though an exact comparison of TS-UFZ and conventional CEST is difficult due to different readout parameters, we compared the SNR of the in-plane S0 images using two repeat acquisitions, where SNR=(S0,1+S0,2)/2(S0,1S0,2).31 Table S1 shows the SNR ratios between conventional CEST and TS-UFZ, illustrating that conventional CEST does provide much higher in-plane SNR than TS-UFZ for both phantom and in vivo experiments.

To recover Z-spectral information, normalization with reference scan, S0, was required to remove proton density contrast and to compensate for possible gradient non-linearity. As was discussed by Xu et al20, UFZ imaging objects can have an irregular shape or even under severe B0 inhomogeneity but these irregularities needs to be accounted for when determining the center frequency of the Z-spectrum, Δωc using Eq. [4] above. In this work, Eq. [4] was used in addition to second-order shimming to minimize errors in placing the direct water saturation at the gradient isocenter. However, small B0 shifts were still observed and became severe at high Gsat values (Figure 4C), which might be due to the imperfect correction of Δωsat shifted Δωc. Along the slice direction, each sub-slice was approximately 0.1 mm and corresponded to a single frequency offset. Δωsat was corrected according to the object isocenter, and Δωc set at 0 ppm ideally. However, the mathematical calculation for object isocenter might not be perfectly accurate due to field inhomogeneities. For example, when the object isocenter was calculated with an error of 0.1 mm, Δωc would shift about 0.13 ppm at Gsat = 4 mT/m. Additionally, the in vivo situation is further complicated by subject motion where 0.1 mm movements can affect the Δωsat calculation.

We also noticed that the DS signal width for TS-UFZ imaging was increased for larger Gsat (Figures 3EH). The Eq. [2] illustrates that higher Gsat will expand BWsat for specific values of Ls. Also, the frequency step size becomes correspondingly larger when the total number of frequency offsets (N) is fixed. One possible reason of the DS deviation in Figure 3 might be that the spectral dispersion at higher Gsat results in less points around 0 ppm in the Z-spectrum and this in-turn reduces in the Z-spectrum intensity around 0 ppm. Additionally, while the saturation gradient was segmented to account for droop from the gradient amplifier, there might still be a small droop for Gsat. For example, for nominal Gsat = 10 mT/m and Ls = 10 mm, if the effective Gsat is 2% less than nominal Gsat, the effective BWsat will be 535 rad/s (0.67 ppm) less than nominal BWsat, which will be reflect as stretching of the frequency axis and become apparent around DS. Such BWsat decrease would show a dependence with increasing Gsat. In terms of possible droop and spectral dispersion, some signal variation around DS was observed when using a Gsat = 8, 10 mT/m in the egg white phantom and therefore a lower Gsat (6 mT/m) was selected for in vivo experiments. The gradient strength used here was higher than previous in vivo studies which used Gsat = 3 mT/m23,24. This was to ensure that the saturation gradient was uniform across the whole slice for the Z-spectral region of interest (we noticed non-uniform edges of the slice profile when using 3 mT/m).

Another important consideration for TS-UFZ in vivo is tissue heterogeneity across the saturation gradient direction. Each saturation offset in the Z-spectrum corresponds to a different spatial location and thus different Z-spectral regions may contain signals from different tissue types. Consequently, MTRasym-based quantification approach may provide erroneous results since the reference frequency offset does not correspond to the same tissue type (Figures 6D, S2D, S4D). Therefore, in this work, the APT signal was extracted using a fitting approach, and APT image contrast was quantified by integrating the signal curve over a narrow saturation frequency range of 1 ppm (~ 0.5 mm), over which voxels are assumed to have reasonably similar tissue properties. Another potential source for the in vivo error is that, compared to conventional CEST methods which saturate the whole brain, the TS-UFZ imaging saturation is very position dependent and bulk motion (e.g., flow) through the slice can lead to Z-spectral signal fluctuations. Unsaturated CSF and blood flowing into the slice region will affect signal intensities for Ssat and S0, which we think caused the difference in baseline with the conventional approach in Figure 5. The coronal orientation of brain was studied in this work rather than transverse orientation, as it had less flow interference. To eliminate the baseline difference, APT signal was extracted by removing the fitted background from the acquired Z-spectra, using a modified PLOF method29,30. The in vivo TS-UFZ approach was tested on three healthy human volunteers and provided APT signal intensities on the same order of magnitude in the Z-spectra as conventional CEST but with different contrast across the slice which likely is a result of different tissue regions contributing to the signal in each method (Figures 5, 6, and S1S4).

Additionally, the sub-slice thickness for each frequency offset is constant for a fixed N and a given Ls at different Gsat. However, higher Gsat increases BWsat (Eq. [2]). So the measured sub-slice thickness over the same frequency range will decrease with increasing Gsat and the sub-slice positions will change accordingly. For example, in the egg white phantom experiments, the 10 mm slice thickness at Gsat = 4, 6, 8, 10 mT/m corresponded to BWsat values of 13.3, 20.0, 26.6, and 33.3 ppm, respectively. In turn, the Z-spectral signal from 3 to 4 ppm corresponds to the measured slice thicknesses of 0.75, 0.5, 0.38 and 0.3 mm, respectively. Hence, different saturation gradient strengths can change the resulting contrast since the measured slice thickness and sub-slice positions contributing to the signal over a specific frequency range will be different. The APT contrasts from the egg white phantom under different Gsat resembled each other well (Figures 3, 4), because the sample was homogeneous. The in vivo situation is more complex due to the tissue heterogeneity across the saturation gradient direction and contrast will change with different Gsat.

Theoretically, the TS-UFZ method should speed up acquisition because saturation pulses at only two frequency offsets (Ssat and S0) are needed for the entire Z-spectral image series. However, TS-UFZ imaging using a traditional 3D SE imaging technique resulted in an acquisition time dependent on the acquired in-plane voxel size. Compared to conventional CEST, in vivo TS-UFZ imaging could save about three-quarters of the scan time for low image resolution (voxel size of 8 × 8 mm2). However, savings were not significant for high-resolution TS-UFZ imaging (voxel size of 5 × 5 mm2). Still, several acceleration methods can be applied to the phase-encoding steps to significantly speed up acquisition time, for instance compressed sensing and parallel imaging3235. This would allow for high spatiotemporal resolution dynamic studies. In this work, we first used SENSE factors of 1 (no acceleration in k-space) for both in-plane directions to guarantee good image quality for the salicylic acid phantom. However, there was still a temporal acceleration in the acquisition of TS-UFZ. Then the egg white phantom which with same in-plane FOV and voxel size as the salicylic acid phantom used an increased in-plane SENSE factor (2 and 1 in Figures 3 and 4; 2 and 2 in Figures S5 and S6), allowing a scan time saving by a factor of two and four with respect to the salicylic acid phantom, respectively (2 mins 15 s for 1 and 1, 53 s for 2 and 1, and 30 s for 2 and 2). Further, this method has the potential to be extended to whole brain coverage by incorporating multi-slice readout methods.

5. Conclusion

In this work, we propose a TS-UFZ imaging method which is an extension of gradient-encoded ultrafast Z-spectroscopy but has spectroscopic imaging capabilities. It was demonstrated on a 3T clinical scanner for both phantom and in vivo applications. TS-UFZ imaging was able to distinguish CEST effects as a function of frequency in both cases. In phantoms, APT contrast maps were in excellent agreement with those for conventional CEST MRI. In vivo, APT contrast intensities in Z-spectra were of the same order of magnitude as for conventional CEST, but maps showed differences that are likely due to the large difference in slice thickness for the integrated APT region.

Supplementary Material

SUPINFO

Figure S1. In vivo TS-UFZ results for subject 2 from the high in-plane resolution scan. (A, B) Profiles of TS-UFZ images in the slice direction with saturation (A, Ssat) and without saturation (B, S0). (C) Intensities of Ssat and S0 signals for the white matter ROI shown on a TS-UFZ image. (D) The Z-spectra of TS-UFZ and conventional CEST. (E, F) Extraction of APT signals for TS-UFZ and conventional CEST.

Figure S2. Comparison of APT contrast maps for a coronal slice through the midbrain of subject 2. (A) T2w image. (B, C) TS-UFZ and (E, F) conventional CEST for both low (voxel size = 8 × 8 mm2) and high in-plane resolution (voxel size = 5 × 5 mm2). (D, G) APTW maps obtained by using MTRasym analysis for the two techniques (high-resolution image).

Figure S3. In vivo TS-UFZ results for subject 3 from the high in-plane resolution scan. (A, B) Profiles of TS-UFZ images in the slice direction with saturation (A, Ssat) and without saturation (B, S0). (C) Intensities of Ssat and S0 signals for the white matter ROI shown on a TS-UFZ image. (D) The Z-spectra of TS-UFZ and conventional CEST. (E, F) Extraction of APT signals for TS-UFZ and conventional CEST.

Figure S4. Comparison of APT contrast maps for a coronal slice through the midbrain of subject 3. (A) T2w image. (B, C) TS-UFZ and (E, F) conventional CEST for both low (voxel size = 8 × 8 mm2) and high in-plane resolution (voxel size = 5 × 5 mm2). (D, G) APTW maps obtained by using MTRasym analysis for the two techniques (high-resolution image).

Figure S5. TS-UFZ Z-spectra from an ROI in an egg white phantom for different saturation gradient strengths (Gsat = 4, 6, 8, 10 mT/m) with SENSE factor of 2 and 2 in both in-plane (the total scan time of acquiring both Ssat and S0 was 45 s). (A-D) Mean intensities of Ssat and S0 signals with standard deviation (SD) of TS-UFZ. The SD is over 3 repeat acquisitions for each Gsat. (E-H) The comparison between Z-spectra (mean ± SD) of TS-UFZ (blue line) and conventional CEST (red circles). (I-L) The extraction of APT signals for TS-UFZ (Z-spectrum from one scan for each Gsat).

Figure S6. APT contrast of TS-UFZ for egg white phantom under Gsat = 4, 6, 8, 10 mT/m and SENSE factor of 2 and 2 in both in-plane (the total scan time of acquiring both Ssat and S0 was 45 s).

Table S1. SNR ratios between TS-UFZ and conventional CEST calculated for in-plane S0 images using two repeat acquisitions, SNR=(S0,1+S0,2)/2(S0,1S0,2).

Acknowledgements

The authors would like to thank Dr. Xiang Xu for valuable discussions. This research was supported by NIH grants P41031771 and R01 EB015032. C.B. thanks China Scholarship Council (201906970024) for financial support. Under a license agreement between Philips and the Johns Hopkins University, Dr. van Zijl and Johns Hopkins University are entitled to fees related to an imaging device used in the study discussed in this publication. Dr. van Zijl also is a paid lecturer for Philips and receives research support from Philips. This arrangement has been reviewed and approved by the Johns Hopkins University in accordance with its conflict of interest policies.

References

  • 1.van Zijl PCM, Yadav NN. Chemical exchange saturation transfer (CEST): what is in a name and what isn’t? Magn Reson Med. 2011;65:927–948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ward KM, Aletras AH, Balaban RS. A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST). J Magn Reson. 2000;143:79–87. [DOI] [PubMed] [Google Scholar]
  • 3.van Zijl PCM, Sehgal AA. Proton chemical exchange saturation transfer (CEST) MRS and MRI. eMagRes. 2016:1307–1332. [Google Scholar]
  • 4.van Zijl PCM, Lam WW, Xu J, Knutsson L, Stanisz GJ. Magnetization transfer contrast and chemical exchange saturation transfer MRI. features and analysis of the field-dependent saturation spectrum. NeuroImage. 2018;168:222–241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Jones KM, Pollard AC, Pagel MD. Clinical applications of chemical exchange saturation transfer (CEST) MRI. J Magn Reson Imaging. 2018;47:11–27. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Zhou J, Payen J-F, Wilson DA, Traystman RJ, van Zijl PCM. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nat Med. 2003;9:1085–1090. [DOI] [PubMed] [Google Scholar]
  • 7.Zu Z, Janve VA, Li K, Does MD, Gore JC, Gochberg DF. Multi-angle ratiometric approach to measure chemical exchange in amide proton transfer imaging. Magn Reson Med. 2012;68:711–719. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Dula AN, Smith SA, Gore JC. Application of chemical exchange saturation transfer (CEST) MRI for endogenous contrast at 7 Tesla. J Neuroimaging. 2013;23:526–532. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Zhang X-Y, Wang F, Xu J, Gochberg DF, Gore JC, Zu Z. Increased CEST specificity for amide and fast-exchanging amine protons using exchange-dependent relaxation rate. NMR Biomed. 2018;31: 10.1002/nbm.3863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Jones CK, Huang A, Xu J, et al. Nuclear Overhauser enhancement (NOE) imaging in the human brain at 7T. NeuroImage. 2013;77:114–124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Ling W, Regatte RR, Navon G, Jerschow A. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc Natl Acad Sci U S A. 2008;105:2266–2270. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.van Zijl PCM, Zhou J, Mori N, Payen J-F, Wilson D, Mori S. Mechanism of magnetization transfer during on-resonance water saturation. A new approach to detect mobile proteins, peptides, and lipids. Magn Reson Med. 2003;49:440–449. [DOI] [PubMed] [Google Scholar]
  • 13.Xu J, Zaiss M, Zu Z, et al. On the origins of chemical exchange saturation transfer (CEST) contrast in tumors at 9.4 T. NMR Biomed. 2014;27:406–416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zhou Y, Bie C, van Zijl PCM, Yadav NN. The relayed nuclear Overhauser effect in magnetization transfer and chemical exchange saturation transfer MRI. NMR Biomed. 2022:e4778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Grad J, Bryant RG. Nuclear magnetic cross-relaxation spectroscopy. J Magn Reson. 1990;90:1–8. [DOI] [PubMed] [Google Scholar]
  • 16.Swanson SD. Broadband excitation and detection of cross-relaxation NMR spectra. J Magn Reson. 1991;95:615–618. [Google Scholar]
  • 17.Xu X, Lee J-S, Jerschow A. Ultrafast Scanning of Exchangeable Sites by NMR Spectroscopy. Angew Chem Int Ed. 2013;52:8281–8284. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Boutin C, Léonce E, Brotin T, Jerschow A, Berthault P. Ultrafast Z-Spectroscopy for 129Xe NMR-Based Sensors. J Phys Chem Lett. 2013;4:4172–4176. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Döpfert J, Witte C, Schröder L. Slice-selective gradient-encoded CEST spectroscopy for monitoring dynamic parameters and high-throughput sample characterization. J Magn Reson. 2013;237:34–39. [DOI] [PubMed] [Google Scholar]
  • 20.Xu X, Yadav NN, Song X, et al. Screening CEST contrast agents using ultrafast CEST imaging. J Magn Reson. 2016;265:224–229. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Döpfert J, Zaiss M, Witte C, Schröder L. Ultrafast CEST imaging. J Magn Reson. 2014;243:47–53. [DOI] [PubMed] [Google Scholar]
  • 22.Zhang Y, Zu T, Liu R, Zhou J. Acquisition sequences and reconstruction methods for fast chemical exchange saturation transfer imaging. NMR in Biomedicine. 2022;n/a:e4699. [DOI] [PubMed] [Google Scholar]
  • 23.Liu Z, Dimitrov IE, Lenkinski RE, Hajibeigi A, Vinogradov E. UCEPR: Ultrafast localized CEST-spectroscopy with PRESS in phantoms and in vivo. Magn Reson Med. 2016;75:1875–1885. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Wilson NE, D’Aquilla K, Debrosse C, Hariharan H, Reddy R. Localized, gradient-reversed ultrafast z-spectroscopy in vivo at 7T. Magn Reson Med. 2016;76:1039–1046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Yang X, Yadav NN, Song X, et al. Tuning Phenols with Intra-Molecular Bond Shifted HYdrogens (IM-SHY) as diaCEST MRI Contrast Agents. Chem Eur J. 2014;20:15824–15832. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. [PubMed] [Google Scholar]
  • 27.Kim M, Gillen J, Landman BA, Zhou J, van Zijl PCM. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med. 2009;61:1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Breitling J, Deshmane A, Goerke S, et al. Adaptive denoising for chemical exchange saturation transfer MR imaging. NMR Biomed. 2019;32:e4133. [DOI] [PubMed] [Google Scholar]
  • 29.Chen L, Zeng H, Xu X, et al. Investigation of the contribution of total creatine to the CEST Z-spectrum of brain using a knockout mouse model. NMR Biomed. 2017;30:e3834. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Sui R, Chen L, Li Y, et al. Whole-brain amide CEST imaging at 3T with a steady-state radial MRI acquisition. Magn Reson Med. 2021;86:893–906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters. Journal of Magnetic Resonance Imaging. 2007;26:375–385. [DOI] [PubMed] [Google Scholar]
  • 32.Liang D, Liu B, Wang J, Ying L. Accelerating SENSE using compressed sensing. Magn Reson Med. 2009;62:1574–1584. [DOI] [PubMed] [Google Scholar]
  • 33.Feng L, Grimm R, Block KT, et al. Golden-angle radial sparse parallel MRI: Combination of compressed sensing, parallel imaging, and golden-angle radial sampling for fast and flexible dynamic volumetric MRI. Magn Reson Med. 2014;72:707–717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Zhu H, Jones CK, van Zijl PCM, Barker PB, Zhou J. Fast 3D chemical exchange saturation transfer (CEST) imaging of the human brain. Magn Reson Med. 2010;64:638–644. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Zhao X, Wen Z, Zhang G, et al. Three-dimensional turbo-spin-echo amide proton transfer MR imaging at 3-Tesla and its application to high-grade human brain tumors. Mol Imaging Biol. 2013;15:114–122. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

SUPINFO

Figure S1. In vivo TS-UFZ results for subject 2 from the high in-plane resolution scan. (A, B) Profiles of TS-UFZ images in the slice direction with saturation (A, Ssat) and without saturation (B, S0). (C) Intensities of Ssat and S0 signals for the white matter ROI shown on a TS-UFZ image. (D) The Z-spectra of TS-UFZ and conventional CEST. (E, F) Extraction of APT signals for TS-UFZ and conventional CEST.

Figure S2. Comparison of APT contrast maps for a coronal slice through the midbrain of subject 2. (A) T2w image. (B, C) TS-UFZ and (E, F) conventional CEST for both low (voxel size = 8 × 8 mm2) and high in-plane resolution (voxel size = 5 × 5 mm2). (D, G) APTW maps obtained by using MTRasym analysis for the two techniques (high-resolution image).

Figure S3. In vivo TS-UFZ results for subject 3 from the high in-plane resolution scan. (A, B) Profiles of TS-UFZ images in the slice direction with saturation (A, Ssat) and without saturation (B, S0). (C) Intensities of Ssat and S0 signals for the white matter ROI shown on a TS-UFZ image. (D) The Z-spectra of TS-UFZ and conventional CEST. (E, F) Extraction of APT signals for TS-UFZ and conventional CEST.

Figure S4. Comparison of APT contrast maps for a coronal slice through the midbrain of subject 3. (A) T2w image. (B, C) TS-UFZ and (E, F) conventional CEST for both low (voxel size = 8 × 8 mm2) and high in-plane resolution (voxel size = 5 × 5 mm2). (D, G) APTW maps obtained by using MTRasym analysis for the two techniques (high-resolution image).

Figure S5. TS-UFZ Z-spectra from an ROI in an egg white phantom for different saturation gradient strengths (Gsat = 4, 6, 8, 10 mT/m) with SENSE factor of 2 and 2 in both in-plane (the total scan time of acquiring both Ssat and S0 was 45 s). (A-D) Mean intensities of Ssat and S0 signals with standard deviation (SD) of TS-UFZ. The SD is over 3 repeat acquisitions for each Gsat. (E-H) The comparison between Z-spectra (mean ± SD) of TS-UFZ (blue line) and conventional CEST (red circles). (I-L) The extraction of APT signals for TS-UFZ (Z-spectrum from one scan for each Gsat).

Figure S6. APT contrast of TS-UFZ for egg white phantom under Gsat = 4, 6, 8, 10 mT/m and SENSE factor of 2 and 2 in both in-plane (the total scan time of acquiring both Ssat and S0 was 45 s).

Table S1. SNR ratios between TS-UFZ and conventional CEST calculated for in-plane S0 images using two repeat acquisitions, SNR=(S0,1+S0,2)/2(S0,1S0,2).

RESOURCES