Diagnostic accuracy of magnetic resonance elastography and point-shear wave elastography for significant hepatic fibrosis screening: Systematic review and meta-analysis

João Paulo L Schambeck 1,2, Gabriele C Forte 2,3,*, Luana M Gonçalves 1,2, Guilherme Stuker 2, João Bruno F Kotlinski 3, Giacomo Tramontin 2, Stephan Altmayer 1, Guilherme Watte 4, Bruno Hochhegger 1,2,4,5
Editor: Wan-Long Chuang6
PMCID: PMC9894488  PMID: 36730265

Abstract

The hepatic diseases are extremely common in clinical practice. The correct classification of liver fibrosis is extremely important, as it influences therapy and predicts disease outcomes. The purpose of this study is to compare the diagnostic performance of point-shear wave elastography (pSWE) and magnetic resonance elastography (MRE) in the hepatic fibrosis diagnostic. A meta-analysis was carried out based on articles published until October 2020. The articles are available at following databases: MEDLINE, EMBASE, Cochrane Central Register of Controlled Trials, Scientific Electronic Library Online, LILACS, Scopus, and CINAHL. Diagnostic performances were analyzed per METAVIR F2, using 3.5kPa as target fibrosis. Assessment of the methodological quality of the incorporated papers by the QUADAS-2 tool for pSWE and MRE. A total 2,153 studies articles were evaluated and 44 studies, comprising 6,081 patients with individual data, were included in the meta-analysis: 28 studies for pSWE and 16 studies for MRE. The pooled sensitivity and specificity were 0.86 (95%CI 0.80–0.90) and 0.88 (95%CI 0.85–0.91), respectively, for pSWE, compared with 0.94 (95%CI 0.89–0.97) and 0.95 (95%CI 0.89–0.98) respectively, for MRE. The pooled SROC curve for pSWE shows in the area under the curve (AUC) of 0.93 (95%CI 0.90–0.95), whereas the AUC for MRE was 0.98 (95%CI 0.96–0.99). The diagnostic odds ratio for pSWE and MRE were 41 (95%CI 24–72) and 293 (95%CI 86–1000), respectively. There was statistically significant heterogeneity for pSWE sensitivity (I² = 85.26, P<0.001) and specificity (I² = 89.46, P<0.001). The heterogeneity for MRE also was significant for sensitivity (I² = 73.28, P<0.001) and specificity (I² = 87.24, P<0.001). Therefore, both pSWE and MRE are suitable modalities for assessing liver fibrosis. In addition, MRE is a more accurate imaging technique than pSWE and can be used as alternative to invasive biopsy.

Introduction

The hepatic diseases are extremely common in clinical practice [1]. Constant cell damage can lead to progressive fibrosis and, consequently, to the final stage, cirrhosis [2]. The right staging is extremely important given that the amount of fibrosis influences the therapy and predicts the diseases outcomes [3, 4]. Even in the final stage, the patient may remain “compensated” for months or years. However, after cirrhosis is established, it is estimated that the annual mortality rates can reach 57% [5].

For the impairment grading of liver parenchyma and diagnosis of fibrosis, liver biopsy is still considered the reference standard. However, it is an invasive technique that requires some considerations. Hospitalization for several hours is needed [5]. Although the fibrosis involvement tends to be diffuse, it does not have a uniform distribution in the hepatic parenchyma and we often see some areas more affected by fibrosis than others [69]. Besides, intra- and inter observer variability is another limitation which may lead to misdiagnosis and incorrect staging [10, 11]. In light of this, a liver biopsy may have uncertain accuracy, feasibility, and reliability [12]. Consequently, non-invasive techniques are tempting for avoid iatrogenic complications, being a safer approach for the follow-up monitoring [12].

Among the alternatives, we emphasize the elastography techniques, which are based on the measurement of mechanical properties of the interested tissues [4, 13, 14]. A decrease in elasticity may represent more advanced fibrosis staging. Point-shear wave elastography (pSWE) is an ultrasound-based evaluation with easy access, quick attainment, and low cost. It is able to measure shear wave velocity estimating the tissue stiffness, as well a simultaneous evaluation of the inner structures of the liver and surrounding [4, 15]. The required equipment is becoming progressively more compact, which allow inpatient and outpatient evaluation. However, this method has some limitations, such as being operator dependent, which may lead to inter and intra-observer variance, and the evaluation is considerably impaired in patients with ascites and obesity [16].

Magnetic resonance elastography (MRE) is another attractive approach as non-invasive assessment [4, 13]. Beyond the stiffness measurement using complex algorithms, it offers the possibility of morphological study of the entire liver and upper abdomen. MRE is becoming more assessable, although the cost is relatively higher than the pSWE study.

The aim of this meta-analysis was to compare the diagnostic performance of pSWE and MRE for the diagnostic of hepatic fibrosis.

Material and methods

This systematic review and meta-analysis were performed in accordance with the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) statement guidelines [17]. A protocol was designed a priori and registered at PROSPERO: International prospective register of systematic reviews (PROSPERO 2020 CRD42020162774). PIRO (P = adult patients; I = MRE and pSWE; R = Liver biopsy (METAVIR score); O = hepatic fibrosis).

Search strategy

MEDLINE (via PUBMED), EMBASE, Cochrane Central Register of Controlled Trials (CENTRAL, The Cochrane Library), Scientific Electronic Library Online (SciELO), LILACS, Scopus, and CINAHL database were searched through October 2020. Reference list of identified studies and reviews were also hand-searched. The search strategy included the descriptors (MeSH terms and other entry terms) related to pSWE, MRE, METAVIR, and hepatic fibrosis (S1 File).

Eligibility criteria

Full papers without language restrictions that evaluated pSWE or MRE in the diagnosis of liver fibrosis (stage 2), using liver biopsy as the reference standard and classified according to METAVIR score were included.

The following exclusion criteria were used: (a) duplicated publications or studies additional to those already included; (b) biopsy proven which uses other than METAVIR score; (c) study not published; (d) case reports, letters to the editor, reviews, abstracts and meta-analysis; (e) study not available; (f) study with other outcomes than hepatic fibrosis (stage 2 or higher); (g) study with insufficient data for 2x2 table; (h) studies that evaluated exclusively nonalcoholic fatty liver disease (NAFLD).

Study selection

Two investigators (G.S. and G.T.) independently reviewed the titles and abstracts of each article identified in the literature search. All articles that clearly did not meet the inclusion criteria were excluded. The selected articles were retrieved for full-text analysis and eligible articles were identified. In case of disagreement, the articles were reviewed aiming at a consensus position, and if no consensus could be achieved, a third investigator resolved discrepancies (G.C.F.).

Data extraction

Extraction of data from each study included in this review was also conducted independently by two investigators (J.B.F.K and L.M.G.), using a standardized instrument. The following data were extracted: country of study’s origin, year of publication, study design, patient number, patient age, sex and body mass index, technical failures in undertaking liver elastography, histological score used, true positive, true negative, false positive, and false negative pSWE and MRE results.

Methodological quality assessment

Two reviewers independently performed the quality assessment of the RCTs according to Quality Assessment of Diagnostic Accuracy Studies (QUADAS)-2 tool [18]. The patient selection, index test, reference standard, and flow and timing domains were evaluated. This tool classifies studies as low-risk (if most of the information is classified as having a low risk of bias), uncertain-risk (if reporting is insufficient to allow assessment), or high-risk (if the proportion of high-risk information is sufficient to affect interpretation of study results). A third reviewer (J.P.L.S.) resolved discrepancies between the two reviewers.

Statistical analysis

The pooled sensitivities, specificities, and 95% confidence intervals (CIs) were calculated by using random-effect analysis. The pooled positive likelihood ratio (PLR), negative likelihood ratio (NLR), and diagnostic odds ratios (DORs) were also obtained. Summary receiver operating characteristic curves were constructed, and the areas under the curve were obtained. To assume an approximate normal distribution, we used the distribution of logit-transformed sensitivity and specificity and the natural logarithm of DOR. Heterogeneity for pooled sensitivities and specificities was calculated in terms of I2. The threshold effect was quantified using Spearman’s correlation coefficient between logit sensitivity and logit specificity and a coefficient (ρ) ≥ -0.6 was considered significant. If the threshold effect was not significant, further subgroup analysis stratifying for study characteristics was planned to identify potential sources of heterogeneity for each imaging modality if at least three studies met the subgroup characteristics. A likelihood ratio test was used to compare the regression models of subgroups. A continuity correction of 1 was used when calculating the logit transformed sensitivity and specificity. The Deeks funnel plot was used to display possible publication bias. Interstudy heterogeneity was also evaluated by using Galbraith plots. All analyses were performed by using Stata, version 12.0 (Stata, College Station, Tex).

Results

The initial search returned 2,153 studies, from which 468 were duplicate. We screened the remaining 1,685 titles and abstracts of which 1,460 were excluded. Of 225 articles full-text articles assessed for eligibility, we excluded 180 studies. Finally, 44 studies, comprising 6,081 patients with individual data, were included in the meta-analysis: 28 studies for pSWE and 16 studies for MRE (Fig 1).

Fig 1. Study selection for meta-analysis.

Fig 1

Point-shear wave elastography (pSWE). MRI = magnetic resonance imaging.

Table 1 contains the main features of the pSWE studies included in this systematic review and meta-analysis. Most of the studies were conducted in European countries, followed by Asian countries. There were a total of three in Italy [1921], four in Romania [2225], two in France [2628], one in Spain [29], one in Indonesia [30], and two German [31, 32], two in Brazil [33, 34], one in United State [35], seven in China [3642], three in Japan [4345], one in Egypt [46], and one in South Korea [47]. Many of the studies were prospective in design and performed in a single center. The mean age of the 4,465 patients was 52.8 years [SD 2.8], with a predominance of men (n = 2,331, 52.2%), and a mean body mass index was 24.9 kg/m² (SD 1.1). A total of 14 studies included patients with viral etiology only [20, 23, 25, 30, 3234, 3639, 41, 43, 46]. The other 14 studies were performed mostly in a mixed set of patients, including viral etiologies, autoimmune liver diseases (primary biliary cirrhosis, autoimmune hepatitis, among others), alcoholic cirrhosis, and small subset of patients with NALFD [19, 21, 22, 24, 2729, 31, 35, 40, 42, 44, 45, 47].

Table 1. General characteristics of the pSWE selected articles.

Author, year Country Study design Center Sample size Mean age (y) Male sex BMI (kg/m²)
Dhyani, 2018 USA Prospective Single 20 54 12 ND
Karlas, 2011 Germany Prospective Single 97 42.7 68 24.0
Nishikawa, 2014 Japan Prospective Single 108 59.5 56 22.5
Liu, 2015 China Prospective Single 108 40.8 81 21.9
Liu, 2017 China Retrospective Single 174 36.8 107 ND
Liu, 2016 China Prospective Single 187 34.9 111 ND
Lin, 2016 Taiwan Prospective Single 60 51.8 40 26.7
Colombo, 2012 Italy Prospective Single 54 55 38 25.8
Tomita, 2013 Japan Prospective Single 22 6.3 13 ND
Tai, 2015 Taiwan Prospective Single 204 52.9 48 ND
Gani, 2017 Indonesia Prospective Single 43 47.3 31 ND
Rust, 2009 Germany Prospective Single 86 48 46 26
Elhosary, 2016 Egypt Prospective Single 190 53.3 142 ND
Crespo, 2012 Spain Prospective Single 146 54 90 25.5
Chung, 2013 South Korea Prospective Single 74 47.3 35 ND
Chen, 2015 China/Taiwan Prospective Single 137 54 63 24.1
Chen, 2012 China/Taiwan Prospective Single 142 51.6 59 24.6
Cassinotto, 2014 France Prospective Multiple 349 54.8 188 27.4
Cassinotto, 2013 France Prospective Single 321 54.4 192 27
Takahashi, 2010 Japan Prospective Single 55 59.9 30 23.5
Sporea, 2010 Romania Prospective Single 114 46.9 53 ND
Sporea, 2011 Romania Prospective Multiple 197 50 78 ND
Sporea, 2012 Romania Retrospective Multiple 914 55.7 423 24.7
Sporea, 2011 Romania Prospective Single 233 48 90 ND
Silva, 2014 Brazil Prospective Single 51 53.8 18 25.1
Rizzo, 2011 Italy Prospective Single 139 55 83 26
Ragazzo, 2017 Brazil Prospective Single 107 49.1 53 24.9
Piscaglia, 2011 Italy Prospective Single 133 58 83 ND

BMI = body mass index; US = ultrasound; ND = not described.

The characteristics of the MRE studies were summarized in Table 2. The study centers were located in Netherlands (n = 1) [48], in Belgium (n = 1) [49], in United State (n = 4) [5052], in China (n = 2) [53, 54], in Taiwan (n = 2) [55, 56], in Singapore (n = 2) [57, 58], in South Korea (n = 2) [59, 60], and in Japan (n = 2) [61, 62]. Eight studies (50%) were prospective and fifteen were performed in single center. Twelve studies (75%) were performed with MRE 1.5 Tesla. Taken together, the studies reported data from 1,616 subjects. The mean age was 52.8 years (SD 7.6), with majority men (n = 1,000, 61.8%). The mean body mass index was 24.5 kg/m² (SD 1.5). There were 6 studies including only patients with chronic viral liver disease [48, 5355, 57, 58], while the other 10 studies had a more diverse patient population including several etiologies of chronic liver disease in the same study [13, 4952, 56, 5962].

Table 2. General characteristics of the MR elastography selected articles.

Author, year Country Study design Center Sample size Mean age (y) Male sex BMI (kg/m²) Magnetic field (T)
Kim, 2011 South Korea Prospective Single 55 58.3 46 22.3 1.5
Huwart, 2007 Belgium Prospective Single 88 54 37 25 1.5
Ye, 2012 South Korea Retrospective Single 173 57.2 129 22.7 1.5
Hennedige, 2017 Singapore Retrospective Single 63 50.1 44 24.9 1.5
Ichikawa, 2015 Japan Retrospective Single 182 66.4 127 ND 3.0
Shi, 2014 China Prospective Single 113 42 48 21.7 3.0
Toguchi, 2017 Japan Retrospective Single 51 59.9 ND ND 1.5
Venkatesh, 2013 Singapore Prospective Multiple 63 50 44 24.8 1.5
Venkatesh, 2014 USA Retrospective Single 62 54.6 31 ND 1.5
Wu, 2017 Taiwan Retrospective Single 104 60.6 87 24.5 1.5
Bohte, 2014 Netherlands Prospective Single 85 45 55 25.5 3.0
Besa, 2018 USA Retrospective Single 83 58.4 59 25.7 1.5
Batheja, 2015 USA Prospective Single 54 38.5 0 30 1.5
Wu, 2015 Taiwan Retrospective Single 185 53.2 135 24 1.5
Wang, 2011 USA Prospective Single 76 55 50 ND 1.5
Shi, 2016 China Prospective Single 179 42.9 108 23 3.0

BMI = body mass index; MR = magnetic resonance; T = Tesla; ND = not described.

Quality appraisal

Assessment of the methodological quality of the incorporated papers by the QUADAS-2 tool for pSWE and MR elastography is depicted in Fig 2. In the “patient selection” domain, 31 studies were at relatively low risk of bias and 13 unclear. In “index test” domain, all studies were at low risk of bias. In “reference standard”, 42 studies were regarded as low risk and two were unclear. In terms of “flow and timing, 24 studies were scored with low risk of bias, seven, high risk, and 13 unclear.

Fig 2.

Fig 2

(A) Proportion pSWE and (B) MR elastography studies with low, high, and uncertain risk of bias according to the domains of the QUADAS-2 quality tool.

Diagnostic accuracy of hepatic fibrosis

Diagnostic performances were analyzed per fibrosis (METAVIR F2) using 3,5 kPa as target fibrosis in all studies included. Forest plots for the sensitivities and specificities with theirs corresponding 95% confidence intervals (CI) of pSWE and MRE are shown in Figs 3 and 4, respectively. The pooled sensitivity and specificity were 0.86 (95%CI 0.79–0.90) and 0.87 (95%CI 0.83–0.91), respectively, for pSWE, compared with 0.94 (95%CI 0.89–0.97) and 0.95 (95%CI 0.89–0.98) respectively, for MRE. The pooled SROC curve for pSWE (Fig 5A) shows in the area under the curve (AUC) of 0.93 (95%CI 0.90–0.95), whereas the AUC for MRE was 0.98 (95%CI 0.96–0.99) (Fig 5B).

Fig 3. Forest plot of estimates of sensitivity and specificity of pSWE for diagnosis of hepatic fibrosis.

Fig 3

The 95% confidence intervals (CI) are shown around point estimates and the pooled result. Plots show (A) sensitivity and (B) specificity of pSWE.

Fig 4. Forest plot of estimates of sensitivity and specificity of MR elastography for diagnosis of hepatic fibrosis.

Fig 4

The 95% confidence intervals (CI) are shown around point estimates and the pooled result. Plots show (A) sensitivity and (B) specificity of MR elastography.

Fig 5.

Fig 5

(A) Summarized receiver operating characteristic (SROC) curves for pSWE and (B) MRE for the diagnosis of hepatic fibrosis.

The diagnostic odds ratio for pSWE and MRE were 41 (95%CI 24–72) and 293 (95%CI 86–1000), respectively. The Deeks’ funnel plot regression revealed no statistical evidence of asymmetry for pSWE (p = 0.40) and MRE (p = 0.90) (Fig 6), which suggests no asymmetry and major publication bias.

Fig 6.

Fig 6

(A) Funnel plot for assessment of potential publication bias in the pSWE studies. (B) Funnel plot for the MRE studies.

Heterogeneity analysis and subgroup analysis

There was statistically significant heterogeneity for pSWE sensitivity (I² = 0.86, P<0.001) and specificity (I² = 0.88, P<0.001). The heterogeneity for MRE also was significant for sensitivity (I² = 0.73, P<0.001) and specificity (I² = 0.87, P<0.001). The threshold effect was neither significant for pSWE (ρ = -0.14), nor for MRE (ρ = 0.34). Therefore, subgroup analyses were conducted for both pSWE and MRE to investigate potential factors contributing to the heterogeneity.

The subgroup analysis for the pSWE studies (Table 3) revealed that the number of centers (multicentric vs. single) was a significant contributor to heterogeneity, with multicentric studies presenting lower sensitivity and specificity (p = 0.03). However, etiology of cirrhosis (p = 0.09) and the country where the study was performed (p = 0.07) also showed a trend towards different diagnostic performance between groups. Analysis by design (prospective vs. retrospective) was attempted, but only two studies were included in the retrospective group therefore the comparison was not conducted.

Table 3. Subgroup analyses of the diagnostic performance of pSWE for the evaluation of liver fibrosis (28 Studies).

Characteristics No. of studies Sensitivity (95% CI) Specificity (95% CI) p-value
Year publication 0.25
 Before 2015 17 0.84 (0.77–0.89) 0.83 (0.77–0.88)
 ≥ 2015 11 0.88 (0.73–0.95) 0.91 (0.83–0.96)
Region of study 0.07
 Non-Asia 15 0.79 (0.68–0.87) 0.89 (0.80–0.94)
 Asia 13 0.90 (0.85–0.93) 0.86 (0.79–0.91)
Number of centers 0.03
 Single center 25 0.86 (0.79–0.91) 0.88 (0.83–0.92)
 Multicenter 3 0.83 (0.72–0.91) 0.83 (0.76–0.89)
Sample size 0.57
 ≥100 19 0.84 (0.77–0.90) 0.86 (0.79–0.91)
 <100 9 0.87 (0.77–0.93) 0.89 (0.81–0.94)
Etiology of cirrhosis 0.09
 Viral 14 0.86 (0.79–0.91) 0.89 (0.83–0.93)
 Mixed 14 0.83 (0.72–0.91) 0.83 (0.76–0.89)

95%CI = 95% confidence interval; pSWE = point-shear wave elastography.

The subgroup analysis for MRE revealed that the study design significantly contributed to heterogeneity, as the studies with prospective design presented significantly higher sensitivity and specificity than the retrospective studies included (Table 4). Analysis regarding number of centers was not possible due to limited number of multicentric studies found in the literature.

Table 4. Subgroup analyses of the diagnostic performance of MRE for the evaluation of liver fibrosis (16 studies).

Characteristics No. of studies Sensitivity (95% CI) Specificity (95% CI) p-value
Year publication 0.09
 Before 2015 8 0.96 (0.90–0.98) 0.98 (0.89–0.99)
 ≥ 2015 8 0.91 (0.84–0.94) 0.88 (0.83–0.91)
Design 0.001
 Prospective 8 0.95 (0.88–0.98) 0.96 (0.90–0.99)
 Retrospective 8 0.92 (0.85–0.95) 0.92 (0.81–0.97)
Region of study 0.57
 Non-Asia 6 0.90 (0.82–0.95) 0.92 (0.84–0.96)
 Asia 10 0.94 (0.90–0.97) 0.95 (0.87–0.98)
Sample size 0.70
 ≥100 6 0.95 (0.85–0.98) 0.95 (0.79–0.99)
 <100 10 0.92 (0.86–0.96) 0.94 (0.89–0.97)
Etiology of cirrhosis 0.42
 Viral 6 0.93 (0.87–0.97) 0.98 (0.85–0.99)
 Mixed 10 0.93 (0.86–0.97) 0.92 (0.84–0.96)

95%CI = 95% confidence interval.

Discussion

In the present meta-analysis, it was evaluated the diagnostic performance of ultrasound elastography, evaluated by pSWE and magnetic resonance elastography in the staging 2 of liver fibrosis, as reported in 44 studies (28 for pSWE and 16 for MRE). The use of METAVIR F2 (set by 3,5 kPa) as a cut-off value for pathologic findings dues to its importance in clinical practice: the begin of clinical treatment to reduce the progression of liver fibrosis. Both MRE and pSWE proved to be an important tool for early diagnosis of liver fibroses, reducing the role of biopsies by encompassing a greater part of liver parenchyma besides being a non-invasive diagnosis method, especially MRE according to our findings in this meta-analysis.

Our results showed that pSWE and MRE could be used to diagnose liver fibrosis. Both imaging methods provide excellent diagnostic accuracy for staging 2 liver fibrosis, with AUROC of 0.93 and 0.98 for pSWE and MRE, respectively. However, the sensitivity and specificity of MRE shows superior results compared to pSWE for the diagnosis of stage two of liver fibrosis. pSWE and MRE showed probability of 86% and 94%, respectively, correctly to diagnose liver fibrosis following a “positive” measurement.

Previous meta-analysis demonstrated inferior sensitivity and specificity compared to the present study, for both point-shear wave elastography and resonance elastography. Tsochatzis et al. [2] demonstrated accuracy of transient elastography for diagnose the severity of fibrosis in chronic liver disease. The summary sensitivity and specificity detected in stage F2 (31 studies) was 0.79 and 0.78, respectively. Su et al. [63] when assessing the accuracy of MRE for stage F2 liver fibrosis, showed results of sensitivity and specificity, respectively, 0.87 and 0.92. Guo et al. [6] show sensitivity 0.76 for pSWE and 0.87 for MRE, and significance was found in AUROC between pSWE (0.85) and MRE (0.97) for the diagnosis of stage 2 liver fibrosis.

Although in the study by Guo et al. [6] considerable heterogeneities were not observed in the MRE and pSWE studies, our study revealed significant heterogeneity in both imaging modalities for the evaluation of significant liver fibrosis. Tsochatzis et al. [2] showed results similar to the present meta-analysis finding statistically significant heterogeneity for stage 2 (I² = 67%, p<0.001), but not for the others. In our study, heterogeneity was not fully explained by threshold effect and further sub analysis was conducted. Three factors were shown to be related to heterogeneity in pSWE studies (number of centers, etiology, country of origin), although only the number of centers was statistically significant with multicentric presenting lower sensitivity and specificity than single center studies. For MRE, only the design of the studies were found to be in part contributing to the heterogeneity, with prospective studies demonstrating higher sensitivity and specificity compared to the retrospective group. Nonetheless, the summarized diagnostic performances of both modalities should be interpreted with caution due to high heterogeneity.

Although liver biopsy yet is the reference standard for evaluating and classifying stage of liver fibrosis, it has several limitations. It is invasive method and can cause minor complications including temporary pain until major complications, such as bleeding, hemothorax and even death [64, 65]. Accurate staging of liver fibrosis is very important, since hepatic fibrosis has a potential for reversal when in initial stages [66]. Therefore, the presence of significant fibrosis (F2) is already considered an important finding of progressive disease and needs special attention [67].

We adopted a systematic search and analysis strategy to assess the accuracy of pSWE and MRE for diagnose of significant liver fibrosis. However, there are still limitations in our meta-analysis. First, we have only included full-text analysis with histopathological score METAVIR. Second, we have not included patients with NAFLD to control for some of the hepatic inflammation which could have contributed to the heterogeneity of the studies, but it may limit some of the representativeness of our results. Third, there was significant heterogeneity in the meta-analysis of both modalities that were not fully accounted for the threshold effect. Our analysis was limited because there is not studies assessing joint pSWE and MRE in the same population. There is a single study that evaluated MRE and ultrasound by elastography, but it used the transient elastography instead of pSWE. Despite the heterogeneity and limitations found in this study, the meta-analysis results reported non-invasive clinical practice for the diagnosis of liver fibrosis. Furthermore, our study included 44 studies with a large sample size and most prospective design studies.

In conclusion, our meta-analysis shows pSWE and MRE provide excellent diagnostic accuracy for significant liver fibrosis. These methods, especially the MRE, can be used as an alternative to invasive biopsy. We suggest further studies with an adequate design and sample size comparing different elastography techniques.

Supporting information

S1 File. Search strategy.

(DOCX)

S2 File. PRISMA checklist.

(PDF)

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

The author(s) received no specific funding for this work.

References

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

S1 File. Search strategy.

(DOCX)

S2 File. PRISMA checklist.

(PDF)

Data Availability Statement

All relevant data are within the paper and its Supporting Information files.