Introduction
Percutaneous suture-mediated patent foramen ovale (PFO) closure with the “device-less” NobleStitch system (HeartStitch) has been employed in clinical practice as an effective and safe alternative to traditional device closure1,2,3. In previous studies, recurrence of a residual atrial right-to-left shunt (RLS) during follow-up was estimated at almost 20%1,2,3. In any case, specific mechanisms of failure have not been systematically described and they still remain unknown. The aim of the study was to describe the mechanisms of failure after NobleStitch from a prospective single-centre cohort.
Methods
Study population
Between June 2016 and September 2021, 122 patients (female 50.0%; mean age 48±12 years) underwent percutaneous suture-mediated PFO closure with NobleStitch at our institution. The overall population was divided into 2 groups on the basis of recurrence of atrial RLS ≤1 or RLS ≥2 grade at echocardiographic follow-up with transcranial doppler (TCD) or transthoracic echocardiography (TTE) bubble study, within 6 months following the procedure. Patients with RLS ≥2 grade were then studied with transoesophageal echocardiography (TOE), in order to define specific mechanisms of ineffective closure. Study design is illustrated in Supplementary Figure 1. Echocardiographic assessment is specified in Supplementary Appendix 1.
Data analyses
Statistical analyses are illustrated in Supplementary Appendix 2.
Results
The baseline characteristics of both groups, with and without significant residual RLS, are summarised in Table 1.
Table 1. Baseline demographic, clinical and echocardiographic characteristics of the study population.
Overall (n=122) |
RLS ≤1 (n=93) |
RLS ≥2 (n=29) |
p-value | |
---|---|---|---|---|
Baseline characteristics | ||||
Age, years (mean±SD) | 48±12 | 48±11 | 46±14 | 0.349 |
Female, n (%) | 61 (50.0) | 50 (53.8) | 11 (37.9) | 0.137 |
BMI, kg/m2 (IQR) | 21 (19-23) | 21 (19-24) | 21 (18-23) | 0.702 |
RoPE score, points (IQR) | 6 (5-7) | 6 (5-7) | 6 (5-7) | 0.478 |
RoPE ≥6, n (%) | 71 (61.2) | 56 (60.2) | 15 (65.2) | 0.659 |
Medical history | ||||
Hypertension, n (%) | 29 (24.0) | 23 (24.7) | 6 (21.4) | 0.720 |
Diabetes, n (%) | 3 (2.5) | 3 (3.2) | 0 (0.0) | 1.000 |
Smoking history, n (%) | 26 (21) | 20 (21.5) | 6 (20.6) | 0.993 |
Dyslipidaemia, n (%) | 31 (25.4) | 24 (25.8) | 14 (24.0) | 0.932 |
Family history of CVD, n (%) | 16 (13.1) | 10 (10.7) | 6 (20.7) | 0.144 |
Prior deep vein thrombosis, n (%) | 4 (3.3) | 2 (2.2) | 2 (7.1) | 0.229 |
Thrombotic diathesis, n (%) | 13 (10.7) | 8 (8.6) | 5 (17.9) | 0.166 |
Clinical presentation | ||||
Cryptogenic stroke, n (%) | 47 (38.8) | 32 (34.4) | 15 (53.6) | 0.068 |
TIA, n (%) | 39 (32.2) | 30 (32.3) | 9 (32.1) | 0.991 |
Intractable migraine, n (%) | 26 (21.5) | 19 (20.4) | 7 (25.0) | 0.606 |
Decompression sickness, n (%) | 11 (9.0) | 10 (10.8) | 1 (3.4) | 0.457 |
Neuroimaging ischaemic lesion, n (%) | 87 (71.9) | 65 (69.9) | 22 (78.6) | 0.669 |
Anatomical characteristics | ||||
Atrial septal aneurysm, n (%) | 19 (15.6) | 10 (10.8) | 9 (31.0) | 0.009 |
Atrial septal hypermobility, n (%) | 48 (39.3) | 30 (32.3) | 18 (62.1) | 0.004 |
Septum cribrosus, n (%) | 2 (1.7) | 1 (1.1) | 1 (3.4) | 0.423 |
Chiari network, n (%) | 84 (69.4) | 58 (63.0) | 26 (89.7) | 0.007 |
Fossa ovalis diameter, mm (IQR) | 20 (17-22) | 20 (17-22) | 19 (17-23) | 0.941 |
PFO tunnel length, mm (IQR) | 11 (9-13) | 11 (9-12) | 12 (9-15) | 0.058 |
PFO width, mm (IQR) | 2.6 (2-3.8) | 2.5 (2.0-3.0) | 3.0 (2.3-4.0) | 0.020 |
Functional characteristics | ||||
Interatrial shunting at rest, n (%) | 34 (28.1) | 20 (21.7) | 14 (48.3) | 0.006 |
Atrial RLS after Valsalva, n (%) | 122 (100.0) | 93 (100.0) | 29 (100.0) | - |
Left-to-right shunt, n (%) | 16 (13.2) | 9 (9.8) | 7 (24.1) | 0.047 |
Spontaneous colour Doppler shunt, n (%) | 10 (8.3) | 8 (8.7) | 2 (7.1) | 1.000 |
Grade of RLS after Valsalva at TTE/TOE | ||||
Mild, n (%) | 10 (8.2) | 10 (10.8) | 0 (0.0) | 0.115 |
Moderate, n (%) | 68 (55.7) | 53 (57.0) | 15 (51.7) | 0.618 |
Severe, n (%) | 44 (36.1) | 30 (32.3) | 14 (48.3) | 0.117 |
Grade of RLS after Valsalva at TCD, n (%) | ||||
Mild, n (%) | 9 (9.9) | 9 (13.4) | 0 (0.0) | 0.105 |
Moderate, n (%) | 40 (44.0) | 32 (47.8) | 8 (33.3) | 0.222 |
Severe, n (%) | 42 (46.2) | 26 (38.8) | 16 (66.7) | 0.019 |
Categorical variables are expressed as n (%). Continuous variables are expressed as median and interquartile range (IQR), as appropriate. BMI: body mass index; CVD: cardiovascular disease; PFO: patent foramen ovale; RLS: right-to-left shunt, RoPE score: Risk of Paradoxical Embolism score; TCD: transcranial Doppler; TIA: transient ischaemic attack; TOE: transoesophageal echocardiography; TTE: transthoracic echocardiography |
The NobleStitch procedure was successfully performed in 116 (95.1%) patients (in 6 patients traditional devices were then used, due to RLS ≥2 at the end of the NobleStitch procedure). Follow-up was complete in 100.0%. At a median follow-up of 335 days, atrial RLS ≥2 occurred in 19.8% (23/116) of patients. Three different mechanisms of late failure (Figure 1, Supplementary Figure 2) were identified by TOE: partial stitch detachment, atrial septal tear and KwiKnot (Heartstitch) embolisation.
Discussion
The NobleStitch system is a good alternative for PFO closure when compared to classic device closure, although there is still a lack of information about the long-term follow-up of this procedure and its real effectiveness. In our analysis, the occurrence of effective PFO closure after the NobleStitch procedure - with atrial RLS ≤1 - was 80.2%, which was in line with the rate in previous reports1,2,3, although lower than the results obtained with traditional devices4. Although the initial learning curve may have had an impact, specific selection criteria are mandatory in order to improve the outcomes of this procedure.
In this prospective study, we introduced a systematic approach during the follow-up with a bubble test at TTE and TCD on all patients and a further investigation with TOE only in patients with RLS ≥2 to investigate failure mechanisms. After excluding the 6 patients with early technical failure who were managed with a classic device for closure, there remained 116 patients who were successfully treated, of whom 23 (19.8%) presented RLS ≥2. Among them, specific mechanisms of failure were detected in 17 (73.9%). Six patients, all asymptomatic, presented a large residual atrial RLS and refused further TOE. A partial stitch detachment was found as the main mechanism in 52.2% of cases. In these cases, the KwiKnot was attached only on the septum primum or septum secundum, with the clear identification of 2 separated septal sheets, leading to an opening tunnel with consequent atrial RLS. A second mechanism found was an atrial septal tear (13%) occurring really close to the stitch, with consequent atrial RLS. Finally, the third mechanism identified was a complete embolisation of the stitch and the knot into the pulmonary circulation (8.7%). To the best of our knowledge, this is the first study in which partial stitch detachment and KwiKnot embolisation have been identified as possible mechanisms of RLS ≥2, whereas atrial septal tear was previously described in an anecdotal case report5.
Limitations
Different limitations should be considered in the interpretation of our results. This was not a controlled study, so several confounding factors could have influenced our results; secondly, a small sample size from a single centre, with limited follow-up, was included, so our results need to be confirmed in multicentre analyses with longer follow-up. Reflecting real-world clinical practice, an independent core laboratory did not review the echocardiographic data. However, the analyses were conducted by dedicated, highly experienced echocardiographists, utilising validated methods.
Conclusion
Percutaneous closure of PFO by suturing with the NobleStitch system is an effective and safe technique that has obvious advantages over traditional device closure, although recurrence of RLS after PFO closure is not uncommon. Partial stitch detachment, atrial septal tear and KwiKnot embolisation were the main mechanisms encountered during follow-up.
TOE is an essential tool for assessing the anatomical and functional characteristics of PFO, in order to improve patient selection for the NobleStitch procedure, guide the operator in sutureorientation and correct placement, and identify possible mechanisms of failure during follow-up. Defining mechanisms and predictors of failure may help to improve long-term results.
Supplementary data
Acknowledgments
Conflict of interest statement
The authors have no conflicts of interest to declare.
Contributor Information
Jessica Zannoni, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Antonio Popolo Rubbio, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Maurizio B. Tusa, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Anca I. Corciu, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Matteo Casenghi, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Gaspare Cannone, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Marta Barletta, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Elisa Stefanini, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Valentina Mantovani, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Riccardo Gorla, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Nedy Brambilla, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Luca Testa, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Francesco Bedogni, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
Federico De Marco, Clinical and Interventional Cardiology Department, I.R.C.C.S. Policlinico San Donato, San Donato Milanese, Milan, Italy.
References
- Gaspardone A, De Marco, Sgueglia GA, De Santis, Iamele M, D’Ascoli E, Tusa M, Corciu AI, Mullen M, Nobles A, Carminati M, Bedogni F. Novel percutaneous suture-mediated patent foramen ovale closure technique: early results of the NobleStitch EL Italian Registry. EuroIntervention. 2018;14:e272–9. doi: 10.4244/EIJ-D-18-00023. [DOI] [PubMed] [Google Scholar]
- Gaspardone A, Sgueglia GA, De Santis, D’Ascoli E, Iamele M, Piccioni F, Giannico B, D’Errico F, Gioffrè G, Summaria F, Gaspardone C, Versaci F. Predictors of Residual Right-to-Left Shunt After Percutaneous Suture-Mediated Patent Fossa Ovalis Closure. JACC Cardiovasc Interv. 2020;13:2112–20. doi: 10.1016/j.jcin.2020.06.004. [DOI] [PubMed] [Google Scholar]
- Beneduce A, Ancona MB, Moroni F, Ancona F, Ingallina G, Melillo F, Russo F, Ferri LA, Bellini B, Chieffo CA, Agricola E, Montorfano M. A systematic transoesophageal echocardiography study of suture-mediated percutaneous patent foramen ovale closure. EuroIntervention. 2022;18:63–7. doi: 10.4244/EIJ-D-21-00242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Patent Foramen Ovale Closure or Antiplatelet Therapy for Cryptogenic Stroke. N Engl J Med. 2020;382:978. doi: 10.1056/NEJMx200001. [DOI] [PubMed] [Google Scholar]
- Baldetti L, Ferri LA, Ancona M, Bellini B, Visco E, Melillo F, Beneduce A, Chieffo A, Ancona F, Agricola E, Montorfano M. Interatrial Septal Tear After Patent Foramen Ovale Closure With the NobleStitch Device. JACC Cardiovasc Interv. 2019;12:e139–40. doi: 10.1016/j.jcin.2019.05.040. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.