One-person versus two-person mask ventilation in preterm infants at birth: a pilot randomised controlled trial

Dharmesh Shah 1,2,, Mark B Tracy 1,2, Murray Kenneth Hinder 1,2, Nadia Badawi 3,4
PMCID: PMC9906374  PMID: 36746525

Abstract

Background

Mask leak and airway obstruction are common with mask ventilation in newborn infants, leading to suboptimal ventilation. We aimed to perform a pilot study measuring respiratory mechanics during one-person and two-person mask ventilation in preterm infants at birth.

Methods

Infants less than 30 weeks’ gestation were eligible for the study. In the two-person method, one person holds the mask in place and the other provides positive pressure ventilation compared with the standard one-person mask hold. A respiratory function monitor was used in line with a T-piece resuscitator to measure mask leak and airway obstruction. Deferred consent was obtained.

Results

Twenty-five infants were recruited. The mean (SD) birth weight was 920.4 g (188.3), and mean (SD) gestational age was 27.3 weeks (3.0). Percentage mask leak was higher in the one-person mask method (26.4±18.5) compared with the two-person mask method (17.6±9.3) (p=0.018). The mean (SD) expired tidal volume (VTe, mL) in breaths with leak was 3.9 (1.57) in the one-person method compared with 3.05 (1.0) the two-person method (p=0.31). A significantly lower mean (SD) end-tidal carbon dioxide (EtCO2, mm Hg) was measured at 25.3 (9.9) in breaths with mask leak, compared with 30.8 (12.1) in breaths without leak. The breaths with airway obstruction had lower mean EtCO2 (25.9 vs 30.8, p=0.003) and lower mean VTe (1.71 vs 6.95, p<0.001).

Conclusion

Mask leak and airway obstruction are common in resuscitation of preterm infants at birth. The use of the two-person mask technique is effective and it could be a useful option if mask ventilation with the one-person method is not effective.

Trial registration number

ACTRN12614000245695.

Keywords: Neonatology, Resuscitation


WHAT IS ALREADY KNOWN ON THIS TOPIC

  • Mask leak and airway obstruction are common in preterm infants and they often not recognised by the resuscitator.

  • A respiratory function monitor has been identified to be feasible at resuscitation for preterm infants at birth.

  • A two-person method of mask ventilation leads to better tidal volume and less mask leak in a manikin model.

WHAT THIS STUDY ADDS

  • A two-person method of mask ventilation is feasible in resuscitation of preterm infants at birth.

  • Significant airway obstruction and mask leak leading to reduction in tidal volumes are common with mask ventilation.

  • A two-person method of mask ventilation could be an effective option for clinicians if mask ventilation is not effective with a one-person method.

HOW THIS STUDY MIGHT AFFECT RESEARCH, PRACTICE OR POLICY

  • Our results show that the use of two-person mask ventilation in resuscitation of preterm infants at birth is feasible.

  • The two-person method of mask ventilation needs to be evaluated in a larger group or cohort.

Introduction

Extreme preterm infants need ventilatory support to establish lung aeration at birth. However, escalation of ventilatory support could inadvertently induce barotrauma and/or volutrauma.1 Recognition of lung aeration prior to intubation is vital but reliable assessment of this at birth is difficult, even by an experienced hand.2

Some infants require positive pressure ventilation (PPV) to establish sufficient alveolar ventilation and functional residual capacity to correct hypoxia. Significant mask leak and airway obstruction are very common, reducing effective alveolar ventilation and thus effectiveness of PPV.3 4 Recognition of mask leak at resuscitation is clinically difficult.5 Studies have highlighted the finding of upper airway obstruction (hyperflexion/hyperextension of the neck) as a frequent event in extremely premature infants, which can coexist with a mask leak and potentially render the PPV ineffective.6

Given the difficulty with recognising mask leak or airway obstruction during PPV without a complex respiratory mechanics’ monitor, a suboptimal response is likely to be attributed to ‘bad lungs’ rather than poor mask technique, resulting in peak inflation pressures being increased. Observation of chest rise is inadequate for estimating appropriate tidal volumes.7 Assuming the endotracheal tube is well placed, this situation may expose the premature infant to excessive tidal volumes and minute ventilation. These are both directly injurious to fragile immature lung tissue and cause indirect injury to the brain via the effect of hypocarbia, inducing a reduction in cerebral blood flow.8

Finer et al demonstrated that more than 70% of extremely premature infants needing mask ventilation had a median of 14 breaths with airway obstruction before adequate alveolar ventilation was established.9 Woods et al demonstrated leak at the face mask resulted in inadequate ventilation, with operators unable to reliably detect leak during PPV.10 A study by Tracy et al, using a two-person technique for mask ventilation in a manikin model, demonstrated that mask leak was halved compared with the standard one-hand technique.11 This was the first study to support the use of two-person mask ventilation in newborn infants. The UK Resuscitation Guidelines (2015) support the use of a two-person mask technique for neonatal resuscitation if ventilation is not effective.12 The two-person technique was reported in the adult literature as early as 1985, by Jesudian et al and Thomas et al who studied the technique on an adult manikin and reported higher tidal volumes with this method.13 14 The ‘EC-Clamp’ technique with a two-handed jaw-thrust technique has been described to deliver significantly higher minute ventilation in adults as compared with a one-hand technique.15

The aim of this study was to compare one-person versus two-person methods of mask ventilation in preterm infants and to measure respiratory mechanics, including mask leak and airway obstruction, using these different methods of mask hold.

Methods

The study was conducted from June 2015 to February 2016. The pilot study was prospectively registered with Australia New Zealand Trial Registry (ACTRN12614000245695) on 7 March 2014. Eligible infants were randomised to either a one-person mask-hold method (the current standard technique)—in which the resuscitator holds the mask in place with one hand using a two-point top hold and the other hand provides PPV—or the two-person method in which one person holds the mask on the upper rim at four points, applying downward force with a jaw-thrust chin lift, and the other person provides the PPV. If needed clinically, PPV was supplied using a T-piece resuscitator (Neopuff, Fisher & Paykel Healthcare, Auckland, New Zealand), with start settings of: 25 cm H2O for peak inflation pressure, 5 cm H2O for positive end expiratory pressure; and fresh blended gas flow of 10 L/min at 30% FiO2 to Neopuff. A respiratory function monitor (RFM) (Respironics NM3, Philips Healthcare, Best, The Netherlands) was used in line with the mask to measure respiratory mechanics. This includes an end-tidal carbon dioxide (EtCO2) monitor with an accuracy of 2 mm Hg at 0–40 mm Hg. Airway obstruction was defined as a 75% reduction in expired tidal volume compared with the baseline of 10 inflations prior to the obstructed inflation (figure 1A). Percentage mask leak was defined as: (tidal volume inspired (VTi)−tidal volume expired (VTe)/VTi×100); and a mask leak of >75% was considered significant (figure 1B).16 The operators were blinded to the RFM.

Figure 1.

Figure 1

(A) Airway obstruction during mask ventilation in infant at 28 weeks’ gestation. The solid arrow on the top graph depicts reduction in gas flow with simultaneous reduction in EtCO2 (broken arrows) and tidal volumes. (B) Mask leak in another infant at 28 weeks’ gestation with the solid line showing inspiratory flow towards the infant but very little coming out. The broken arrow shows the reduction in measured inspiratory pressure with simultaneous reduction in EtCO2 and tidal volumes.

Participants

All infants born prematurely at less than 30 weeks’ gestation were eligible for the study if they needed respiratory support in form of mask ventilation. The decision to provide mask ventilation included apnoea requiring PPV, persistent desaturation and increased work of breathing. A round mask (35 or 42 mm external diameter) was used and the resuscitators were senior neonatal fellows who were trained to perform different techniques of mask ventilation (figure 2).

Figure 2.

Figure 2

A two-person technique of mask ventilation.

Exclusion criteria

The exclusion criteria included neonates with lethal congenital anomalies, upper airway anomalies and unavailability of the research team.

Outcome measures

The primary outcomes included mask leak and airway obstruction. The secondary outcome was need for airway intubation in the delivery suite with EtCO2.

Randomisation

The infants were block randomised (1:1) using a sealed opaque envelope with computer-generated random numbers. This was a pilot study and not powered directly to detect significant changes in the primary outcomes. Parental consent was sought, if possible (from women in the antenatal ward), but retrospective consent was sought for infants born in emergency situations.

Parent and public involvement

There was no involvement of parents or the public, in the concept, design, reporting or dissemination plans of the research.

Statistical analysis

Statistical Analysis was performed using Stata V.MP18 (Stata Corp. College Station, Texas, USA). The medians and SD of the number of breaths with mask leak and airway obstruction during ventilation were computed for the one-person and two-person method groups. A number of breaths were compared rather than proportion of leak due to small sample size. Between groups, differences in the unadjusted mean number of breaths with mask leak and airway obstruction were investigated using two-sample t-tests. These differences were further investigated via the use of negative binomial regression models including offset terms, to account for each participant having been observed for a different number of breaths. These models were used to compute the adjusted mean differences between the groups. Mean differences between the one-person and two-person method groups for physiological parameters, adjusting for mask leak and airway obstruction, were calculated using linear-mixed models. Paired t-tests were used to compare physiological parameters in breaths with and without mask leak and airway obstruction. A significance level of p<0.05 was used for all analyses.

Results

Of the 60 eligible infants, 30 infants were recruited and 30 were excluded due to unavailability of the research personnel. The data was unavailable for five infants. Thirteen infants were randomised to the two-person method. The mean birth weight was 920.4 g (±188.3) and mean gestational age was 27.3 weeks (±3.0). Fifteen infants were male infants and 60% received a complete course of antenatal steroids.

Comparison of one-person versus two-person mask ventilation

There were no significant differences in birth weight, gestational age, mode of delivery, antenatal steroids, 1 and 5 min Apgar scores between the groups. Seven infants were intubated at birth in one-person versus eight in two-person method group (p=0.87). A mean (SD) total of 161.3 (96.55) inflations were analysed (146.6 one-person method and 174.8 two-person method). The median (IQR) VTe was 6.4 mL (5.2–7.4) in the one-person method group compared with 4.7 (3.9–7.4) in the two-person group (table 1). The median (IQR) VTe with leak was 3.9 mL (2.9–4.7) in the one-person method group compared with 3.1 (2.3–3.6) in the two-person method (p=0.05), whereas the median VTe in breaths with obstruction was 1.8 (1.6–1.8) and 1.8 (1.5–1.9) between groups (p=0.57).

Table 1.

Comparison of one-person and two-person method of mask ventilation

One-person method mean (SD) n=12 Two-person method mean (SD) n=13 P value
Birth weight 948.5 (194.7) 894.5 (186.3) 0.48
Gestational age 27.3 (1.37) 27.3 (1.65) 0.97
Mode of delivery
 LSCS 9 (75%) 10 (76.9%) 0.99
Apgar (median, IQR)
 1 min 6 (3–7) 5 (4–7) 0.82
 5 min 8 (7–8) 7 (6–9) 0.81
Antenatal steroids (complete) 7 (58.3%) 8 (61.5%) 0.99
Intubation in DS n (%) 7 (58.3) 8 (61.5) 0.87
Median (IQR) VTe (mL) (all breaths) 6.4 (5.2–7.4) 4.7 (3.9–7.4) 0.13
Median (IQR) VTe (mL) with leak 3.9 (2.9–4.7) 3.1 (2.3–3.6) 0.05
Median (IQR) VTe (mL) with obstruction 1.8 (1.6–1.8) 1.8 (1.5–1.9) 0.57

Mann-Whitney non-parametric test was used to compare the significance among the medians

DS, delivery suite; LSCS, Lower Segment Caesarean Section; VTe, tidal volume expired.

The adjusted mean (SD) inflations with leak were significantly higher in the one-person method group (26.5±18.6) versus the two-person method (17.7±9.3, p=0.018). The mean (SD) breaths with obstruction were similar between the groups (22.9 vs 27.4, p=0.315). A number of consecutive breaths with mask leak and airway obstruction were also similar between groups (table 2). The SD for number of breaths with airway obstruction was large due to two outliers with 83 out of 182 and 86 out of 319 breaths with airway obstruction.

Table 2.

Difference between one-person and two-person methods of mask ventilation for breath-to-breath occurrence of mask leak and airway obstruction

Overall mean (SD) One-person method mean (SD) Two-person method mean (SD) P value
Total number of breaths analysed 161.3 (96.5) 146.7 (102.9) 174.9 (92.4) 0.477
Number of breaths with mask leak
 Raw mean (SD) 18.0 (12.7) 22.0 (16.6) 14.3 (6.2) 0.156
 Adjusted mean (SD) 26.5 (18.6) 17.7 (9.3) 0.018
Consecutive breaths with mask leak
 Raw mean (SD) 6.6 (3.7) 7.0 (4.5) 6.2 (3.1) 0.618
 Adjusted mean (SD) 10.0 (7.0) 8.6 (4.5) 0.315
Number of breaths with airway obstruction
 Raw mean (SD) 23.2 (21.3) 20.8 (21.9) 25.5 (21.5) 0.593
 Adjusted mean (SD) 22.9 (16.1) 27.5 (14.5) 0.983
Consecutive breaths with airway obstruction
 Raw mean (SD) 7.7 (5.2) 5.7 (4.0) 9.6 (5.6) 0.061
 Adjusted mean (SD) 6.9 (4.9) 13.2 (7.0) 0.145

The p values for the raw mean were calculated from a two-sample test, and the p values for the adjusted means are from a negative binomial regression model.

Analysis of differences between two-person versus one-person methods on VTe and EtCO2, adjusted for mask leak and airway obstruction showed no statistical significance. The mean difference (95% CI) in VTe was −1.55 (−3.84 to 0.75) adjusted for mask leak and −0.86 (−1.96 to 0.33) adjusted for airway obstruction. The analysis of EtCO2 showed a mean difference (95% CI) of 4.83 (−3.65 to 13.32) adjusted for mask leak and 4.26 (−4.04 to 12.57) adjusted for airway obstruction.

Comparison of breaths with leak and without leak demonstrated significant difference in mean VTe, mean oxygen saturation, EtCO2 and heart rate (table 3). The mean VTe was 6.76 (3.42) in breaths without leak compared with 3.47 (1.39) in breaths with leak. Similarly, the mean VTe was 6.95 (2.97) in breaths without obstruction compared with 1.72 (0.19) in breaths with obstruction. The mean EtCO2 was higher in breaths without mask leak (30.89 vs 25.39, p=0.002), and in breaths without obstruction (30.81 vs 25.98, p=0.003).

Table 3.

Comparison of physiological parameters in breaths with mask leak and airway obstruction

Without mask leak, mean (SD) With mask leak, mean (SD) P value
VTe (mL) 6.8 (3.4) 3.5 (1.4) <0.001
Oxygen saturation (%) 71.6 (16.1) 63.2 (13.3) 0.010
EtCO2 (mm Hg) 30.9 (12.2) 25.39 (10.0) 0.002
Heart rate (breaths per minute) 123.9 (30.2) 112.4 (26.6) <0.001
Without airway obstruction With airway obstruction
VTe (mL) 7.0 (3.0) 1.7 (0.2) <0.001
Oxygen saturation (%) 70.0 (16.3) 69.1 (10.2) 0.693
EtCO2 (mm Hg) 30.8 (11.1) 26.0 (10.4) 0.003
Heart rate (breaths per minute) 121.3 (29.1) 118.7 (28.8) 0.269

The p values were calculated from a paired t-test.

EtCO2, end-tidal carbon dioxide; VTe, tidal volume expired.

Table 4 compares the respiratory parameters in infants who were intubated and non-intubated at birth. There was no difference in mask leak and airway obstruction. There were significant numbers of infants with mean VTe outside the preferred target of tidal volume (4–8 mL/kg) as shown in figure 3.

Table 4.

Mask leak and airway obstruction in infants intubated versus non-intubated at birth

Intubated mean (SD) Non-intubated mean (SD) P value
n 15 10
Mean VTe (all breaths) 6.4 (3.0) 5.5 (1.5) 0.40
Number of breaths with leak 14.2 (9.1) 23.6 (15.5) 0.07
Mean VTe (mL) with leak 3.3 (1.5) 3.7 (1.1) 0.46
Number of breaths with obstruction 23.0 (26.0) 23.6 (12.8) 0.95
Mean VTe (mL) with obstruction 1.7 (0.20) 1.7 (0.19) 0.63

VTe, tidal volume expired.

Figure 3.

Figure 3

Graph displaying individual patient mean tidal volume expired (VTe) mL/kg using one-person and two-person methods, with the area between the horizontal dotted lines depicting the preferred VTe (4–8 mL/kg) for individual infants.

Discussion

We demonstrated that there is significant airway obstruction and mask leak in extreme preterm infants resuscitated at birth in the delivery suite. Our study was a pilot study and hence not powered to show significant differences in respiratory parameters and short-term ventilator outcomes, although it does demonstrate significant differences in VTe in breaths with airway obstruction and mask leak. While mask leak and airway obstruction are reported in the literature, our study prospectively demonstrated and quantified the frequency and severity of these events in extreme preterm infants at birth. As transition at birth is a complex process in these preterm infants, provision of optimal airway support is extremely important.

Mask leak is not routinely recognised without specialised volume measurements in resuscitation and potentially reduces alveolar ventilation and thus the effectiveness of the mask PPV at resuscitation. Foglia et al reviewed the respiratory function data in infants at 23–27 weeks’ gestation as post-hoc analysis of the sustained inflation trial.6 They reported significant mask leak (15% of breaths) and airway obstruction (17%) in infants receiving sustained inflation and 3% leak with 17% obstructed breaths in infants receiving PPV.

Our data reflects significant airway obstruction and mask leak in infants receiving mask ventilation. The tidal volumes vary a lot and are quite often outside the target range. In a study of infants under 29 weeks’ gestation, Murthy et al reported lower expired tidal volumes (mL/kg) in infants receiving mask ventilation (1.6, range 0.2–13.3) compared with infants resuscitated via endotracheal tube (2.8, range 0.2–9.9).17 The lower EtCO2 (kPa) was recorded (0.36, range 0–1.8) with mask ventilation compared with endotracheal tube (0.2, range 0–0.9), suggestive of mask leak and/or airway obstruction. This was comparative to our study, with lower EtCO2 in infants with significant mask leak and airway obstruction.

Yang et al prospectively reviewed RFM data in infants under 32 weeks’ gestation receiving PPV.18 They reported mask leak in 32% and airway obstruction in 12% of inflations. They used the mask adjustment, reposition airway, suction mouth and nose, open mouth, pressure increase and alternative airway as intervention, although it was not enough to provide adequate ventilation, with worsening of airway obstruction in four out of five instances. They also reported inflations with low tidal volumes, and only small numbers were associated with bradycardia. Although our study had a small sample size, we demonstrated a significant number of breaths with leak associated with lower heart rate.

Various methods to overcome mask leak and airway obstruction are reported in the literature. A report by Kamlin et al of infants <34 weeks’ gestation confirmed the frequency of airway obstruction with PPV and use of oropharyngeal airways.19 A higher incidence of airway obstruction was noted with use of the oropharyngeal airway. Davidovic et al compared the mean tidal volumes per weight and peak pressures generated by the one-person and two-person techniques of bag-valve-mask ventilation in a paediatric model.20 Two-person bag-valve-mask ventilation provided greater mean tidal volumes and peak pressures in the infant and child manikin models compared with the one-person technique. The two-person technique generated higher mean peak pressures than the one-person technique (infant: 27.4 vs 22.0; child: 27.2 vs 21.8). Wilson et al studied a two-handed mask-hold technique with a two-point top and spider hold in a term newborn manikin. They reported no significant difference in mask leak among the groups, with a median leak of 9% (2%–51%) in the two-handed technique.21 In our study, the median VTe was lower in the two-person mask method (6.4 in one-person vs 4.7 in two-person mask method, p=0.13), and the effect of jaw-thrust and chin lift on airway obstruction needs to be further evaluated.

In a study of preterm infants at resuscitation using a preterm manikin and RFM, there was a wide variation of leak at the mask–face interphase during each resuscitation and between resuscitators.22 They highlighted the importance of correct size mask and RFM to accurately assess the face mask leak and VTe. The RFM has been demonstrated to be useful in recording flow and pressure measurements in premature infants and was associated with significantly less mask leak.23 A study by van Zanten et al randomised infants under 28 weeks’ gestation to a visible or non-visible RFM at resuscitation. Percentage leak (>60%) was 17.4 (7.2–33.3) in the visible RFM versus 13.6 (3.7–32.1) in the non-visible group.24 The per cent of inflations within the target range was 30%, which was comparable to our study. Our study strengthens the importance of RFM in recognition of mask leak, airway obstruction and monitoring EtCO2 at resuscitation of extreme preterm infants.

Hooper et al showed that EtCO2 values may indicate aeration of distal airways.2 Their data on extreme preterm infants at birth demonstrated time to achieve EtCO2 more than 10 mm Hg from birth was 92 s (46–150) and 51 s (27–91) from start of mask ventilation. They reported a significant relationship of tidal volume and EtCO2 (EtCO2 increasing 0.15% with each 1 mL/kg increase in tidal volume). Our study demonstrated the use of EtCO2 in the delivery suite and its value in guiding ventilation. There was association of lower EtCO2 in breaths associated with significant airway obstruction and mask leak.

In adult patients, effectiveness of two resuscitators in terms of greater mask pressures and tidal volumes has been reported and well established.13 Resuscitation in newborn infants seeks to correct bradycardia/asystole and hypoxia by providing adequate alveolar ventilation. Breath-by-breath variation in mask leak may result in underinflation to over inflation with the potential for barotrauma and/or volutrauma, particularly in preterm infants. The optimal mask technique to provide adequate ventilation in resuscitation of extreme preterm infants needs to be further evaluated and the incidence of airway obstruction and mask leak further studied with use of a two-person mask method.

Strengths

The strengths of our study are that it is a prospectively randomised trial with documented RFM data from the first breath provided with PPV. The study validates the feasibility of a two-person mask technique in extreme preterm infants’ resuscitation at birth. We were prospectively able to demonstrate significant airway obstruction and mask leak in the delivery suite.

Limitations

This is a pilot study and hence the numbers are small within a single centre. The definition of mask leak and airway obstruction is not validated, although we used a strict definition of >75% mask leak/obstruction.

Conclusion

Mask leak and airway obstruction are common in resuscitation of preterm infants at birth. The use of two-person mask ventilation could be an effective option for clinicians if mask ventilation is not effective with a one-person method. Further studies are needed to further evaluate efficacy and safety of the two-person mask ventilation method.

Supplementary Material

Reviewer comments
Author's manuscript

Acknowledgments

We would like to acknowledge Ms Annabel Webb for her statistical analysis and Ms Merran Laver for her professional editing of the manuscript.

Footnotes

Contributors: DS and MBT conceptualised the study. DS performed the study. MKH transformed the digital data to.csv file and performed the analysis. DS wrote the manuscript. MBT, MBH and NB critically appraised the manuscript. DS is responsible for the overall content as the guarantor.

Funding: The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

Competing interests: None.

Patient and public involvement: Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.

Provenance and peer review: Not commissioned; externally peer reviewed.

Data availability statement

All data relevant to the study are included in the article or uploaded as an online supplemental information.

Ethics statements

Patient consent for publication

Not applicable.

Ethics approval

This study involves human participants and was approved by HREC2013/10/4.4(3831). Participants gave informed consent to participate in the study before taking part.

References

  • 1.Schmölzer GM, Te Pas AB, Davis PG, et al. Reducing lung injury during neonatal resuscitation of preterm infants. J Pediatr 2008;153:741–5. 10.1016/j.jpeds.2008.08.016 [DOI] [PubMed] [Google Scholar]
  • 2.Hooper SB, Fouras A, Siew ML, et al. Expired CO2 levels indicate degree of lung aeration at birth. PLOS ONE 2013;8:e70895. 10.1371/journal.pone.0070895 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Schilleman K, Witlox RS, Lopriore E, et al. Leak and obstruction with mask ventilation during simulated neonatal resuscitation. Arch Dis Child Fetal Neonatal Ed 2010;95:F398–402. 10.1136/adc.2009.182162 [DOI] [PubMed] [Google Scholar]
  • 4.Murthy V, Dattani N, Peacock JL, et al. The first five inflations during resuscitation of prematurely born infants. Arch Dis Child Fetal Neonatal Ed 2012;97:F249–53. 10.1136/archdischild-2011-300117 [DOI] [PubMed] [Google Scholar]
  • 5.Schmölzer GM, Kamlin OCOF, O’Donnell CPF, et al. Assessment of tidal volume and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed 2010;95:F393–7. 10.1136/adc.2009.174003 [DOI] [PubMed] [Google Scholar]
  • 6.Foglia EE, Kirpalani H, Ratcliffe SJ, et al. Sustained inflation versus intermittent positive pressure ventilation for preterm infants at birth: respiratory function and vital sign measurements. J Pediatr 2021;239:150–4. 10.1016/j.jpeds.2021.08.038 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Poulton DA, Schmölzer GM, Morley CJ, et al. Assessment of chest rise during mask ventilation of preterm infants in the delivery room. Resuscitation 2011;82:175–9. 10.1016/j.resuscitation.2010.10.012 [DOI] [PubMed] [Google Scholar]
  • 8.Rosenberg AA. Response of the cerebral circulation to profound hypocarbia in neonatal lambs. Stroke 1988;19:1365–70. 10.1161/01.str.19.11.1365 [DOI] [PubMed] [Google Scholar]
  • 9.Finer NN, Rich W, Wang C, et al. Airway obstruction during mask ventilation of very low birth weight infants during neonatal resuscitation. Pediatrics 2009;123:865–9. 10.1542/peds.2008-0560 [DOI] [PubMed] [Google Scholar]
  • 10.Wood FE, Morley CJ, Dawson JA, et al. Assessing the effectiveness of two round neonatal resuscitation masks: study 1. Arch Dis Child Fetal Neonatal Ed 2008;93:F235–7. 10.1136/adc.2007.117713 [DOI] [PubMed] [Google Scholar]
  • 11.Tracy MB, Klimek J, Coughtrey H, et al. Mask leak in one-person mask ventilation compared to two-person in newborn infant manikin study. Arch Dis Child Fetal Neonatal Ed 2011;96:F195–200. 10.1136/adc.2009.169847 [DOI] [PubMed] [Google Scholar]
  • 12.Fawke J, Wyllie J, Madar J. Resuscitation council of UK. newborn resuscitation and support of transition of infants at birth guidelines; 2021. Available: https://www.resus.org.uk/library/2021-resuscitation-guidelines [Accessed 05 May 2022]. [DOI] [PubMed]
  • 13.Jesudian MC, Harrison RR, Keenan RL, et al. Bag-valve-mask ventilation; two rescuers are better than one: preliminary report. Crit Care Med 1985;13:122–3. 10.1097/00003246-198502000-00015 [DOI] [PubMed] [Google Scholar]
  • 14.Thomas AN, Dang PT, Hyatt J, et al. A new technique for two-hand bag valve mask ventilation. Br J Anaesth 1992;69:397–8. 10.1093/bja/69.4.397 [DOI] [PubMed] [Google Scholar]
  • 15.Joffe AM, Hetzel S, Liew EC. A two-handed jaw-thrust technique is superior to the one-handed “ EC-clamp ” technique for mask ventilation in the apneic unconscious person. Anesthesiology 2010;113:873–9. 10.1097/ALN.0b013e3181ec6414 [DOI] [PubMed] [Google Scholar]
  • 16.Schmölzer GM, Dawson JA, Kamlin COF, et al. Airway obstruction and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed 2011;96:F254–7. 10.1136/adc.2010.191171 [DOI] [PubMed] [Google Scholar]
  • 17.Murthy V, Shah R, Fox G, et al. Comparison of the efficacy of face mask to endotracheal tube resuscitation in prematurely born infants. Arch Dis Child 2012;97:A133. 10.1136/archdischild-2012-301885.318 [DOI] [Google Scholar]
  • 18.Yang KC, Te Pas AB, Weinberg DD, et al. Corrective steps to enhance ventilation in the delivery room. Arch Dis Child Fetal Neonatal Ed 2020;105:605–8. 10.1136/archdischild-2019-318579 [DOI] [PubMed] [Google Scholar]
  • 19.Kamlin COF, Schmölzer GM, Dawson JA, et al. A randomized trial of oropharyngeal airways to assist stabilization of preterm infants in the delivery room. Resuscitation 2019;144:106–14. 10.1016/j.resuscitation.2019.08.035 [DOI] [PubMed] [Google Scholar]
  • 20.Davidovic L, LaCovey D, Pitetti RD. Comparison of 1- versus 2-person bag-valve-mask techniques for manikin ventilation of infants and children. Ann Emerg Med 2005;46:37–42. 10.1016/j.annemergmed.2005.02.005 [DOI] [PubMed] [Google Scholar]
  • 21.Wilson EV, O’Shea JE, Thio M, et al. A comparison of different mask holds for positive pressure ventilation in a neonatal manikin. Arch Dis Child Fetal Neonatal Ed 2014;99:F169–71. 10.1136/archdischild-2013-304582 [DOI] [PubMed] [Google Scholar]
  • 22.Kelm M, Dold SK, Hartung J, et al. Manual neonatal ventilation training: a respiratory function monitor helps to reduce peak inspiratory pressures and tidal volumes during resuscitation. J Perinat Med 2012;40:583–6. 10.1515/jpm-2012-0023 [DOI] [PubMed] [Google Scholar]
  • 23.Schmölzer GM, Morley CJ, Wong C, et al. Respiratory function monitor guidance of mask ventilation in the delivery room: a feasibility study. J Pediatr 2012;160:377–381. 10.1016/j.jpeds.2011.09.017 [DOI] [PubMed] [Google Scholar]
  • 24.van Zanten HA, Kuypers K, van Zwet EW, et al. A multi-centre randomised controlled trial of respiratory function monitoring during stabilisation of very preterm infants at birth. Resuscitation 2021;167:317–25. 10.1016/j.resuscitation.2021.07.012 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Reviewer comments
Author's manuscript

Data Availability Statement

All data relevant to the study are included in the article or uploaded as an online supplemental information.


Articles from BMJ Paediatrics Open are provided here courtesy of BMJ Publishing Group