Inter-vendor and inter-observer reliability of diffusion tensor imaging in the musculoskeletal system: a multiscanner MR study

Vito Chianca 1,2,, Domenico Albano 3, Stefania Rizzo 1, Mario Maas 4,5, Luca Maria Sconfienza 3,6, Filippo Del Grande 1
PMCID: PMC9911574  PMID: 36757529

Abstract

Background

To evaluate the inter-observer and inter-vendor reliability of diffusion tensor imaging parameters in the musculoskeletal system.

Methods

This prospective study included six healthy volunteers three men (mean age: 42; range: 31–52 years) and three women (mean age: 36; range: 30–44 years).

Each subject was scanned using different 3 Tesla magnetic resonance scanners from three different vendors at three different sites bilaterally. First, the intra-class correlation coefficient was used to determine between-observers agreement for overall measurements and clinical sites. Next, between-group comparisons were made through the nonparametric Friedman’s test. Finally, the Bland–Altman method was used to determine agreement among the three scanner measurements, comparing them two by two.

Results

A total of 792 measurement were calculated. ICC reported high levels of agreement between the two observers. ICC related to MD, FA, and RD measurements ranged from 0.88 (95% CI 0.85–0.90) to 0.95 (95% CI 0.94–0.96), from 0.85 (95% CI 0.81–0.88) to 0.95 (95% CI 0.93–0.96), and from 0.89 (0.85–0.90) to 0.92 (0.90–0.94).

No statistically significant inter-vendor differences were observed. The Bland–Altmann method confirmed a high correlation between parameter values.

Conclusion

An excellent inter-observer and inter-vendor reliability was found in our study.

Supplementary Information

The online version contains supplementary material available at 10.1186/s13244-023-01374-0.

Keywords: Muscle, Diffusion tensor imaging, Magnetic resonance, Reliability, Reproducibility

Key points

  1. Diffusion tensor imaging shows excellent inter-observer reliability.

  2. Diffusion tensor imaging shows excellent inter-vendor reliability.

  3. Quantitative results in musculoskeletal application of diffusion tensor imaging are reproducible in all the three magnetic resonance scanners.

  4. DTI, as a reproducible magnetic resonance sequence, can be used for a quantitative evaluation of muscle microstructures during daily practice.

Supplementary Information

The online version contains supplementary material available at 10.1186/s13244-023-01374-0.

Introduction

Since its first clinical application in the musculoskeletal (MSK) system, muscle diffusion tensor imaging (m-DTI) has become a valuable sequence for the evaluation of architectural changes of fibre microenvironments [14]. M-DTI provides parameters such as fractional anisotropy (FA), radial diffusivity (RD), and mean diffusivity (MD) that allow extracting quantitative information on integrity of muscle fibres [5, 6]. Inflammatory pathologies, traumatic injuries, neuromuscular disorders, or atrophic conditions are the principal areas of application of m-DTI in conjunction with conventional sequences in the assessments of early structural changes [711].

Furthermore, post-processing of DTI is able to generate fibre tractography and to assess 3D muscular structure from the origin to distal insertion by calculating architectural parameters such as fibre length, number and volume, and pennation angle [12].

Despite its potential to assess structural changes of the muscles, the application of m-DTI in clinical practice is still controversial because of several factors influencing DTI signals, such as field strength, gradient strength, b-values, and post-processing algorithms [13]. Several studies reported an acceptable agreement of DTI measurements on the brain, whereas m-DTI studies, which were mainly conducted on lower limb muscle group, reported relatively high variations with FA values ranging from 0.28 to 0.6 [6, 14]. Moreover, some studies on brain DTI reported conflicting results regarding inter-site, intra-site, and inter-vendor reliability [1517]. To the best of our knowledge, no studies assess m-DTI inter-vendor agreement. The aims of our study were to assess the inter-reader reliability and the inter-vendor reliability on 3 T magnetic resonance (MR) for m-DTI measurements.

Materials and methods

Study subjects

The local ethics committee approved our prospective study, and all participants signed an informed consent before starting the examination. The study was conducted in compliance with the Declaration of Helsinki.

We enrolled six healthy volunteers: Three were males (mean age: 42; range: 31–52 years) and three women (mean age: 36; range: 30–44 years).

Inclusion criteria were: 18 year or older, no neuromuscular diseases in their personal and/or family history, no present or past muscle strains in the muscular group under evaluation, and no participation in any sports activity three weeks before the examination.

Exclusion criteria for enrolment were: usual contraindications to MR imaging, positive pregnancy test, and objects in the body that could obscure the target muscle groups through artefacts. After having optimized the sequences in collaboration with the specialists of the different vendors, each volunteer was scanned once on three different anatomical sites bilaterally (middle third of the arm, middle third of the leg, and middle third of the thigh). All scans were acquired on the same day within 6 h to reduce any possible bias and were checked for image quality and artefacts.

MR examination

MR examinations were performed using three 3 T (T) MR of our institution: Signa Pioneer (GE Healthcare, Milwaukee, WI, USA), Achieva (Philips Healthcare, Best, Netherlands), and Skyra (Siemens Healthineers, Erlangen, Germany). The total MR examination time for Signa Pioneer, Achieva, and Skyra was 18.05, 17.56, and 17.52 min, respectively. MR protocol acquisition parameters including RF-coils are summarized in Tables 1 and 2.

Table 1.

DTI acquisition parameters

GE Philips Siemens
DTI parameters (Thigh)
 Coil 16 channels 32 channels 18 channels
 Directions 12 15 12
 b-value (s/mm2) 400 400 400
 TR (ms) 7350 7500 7500
 TE (ms) Minimum 72 83.0
 Matrix 128 × 128 128 × 128 128 × 128
 Parallel factor 2 (asset) 2 (sense) 2 (grappa)
 Voxel size (mm) 3.1 × 3.1 × 4.0 3.03 × 2.97 × 4.0 3.1 × 3.1 × 4.0
 Acquisition time (minutes:seconds) 3.18 4.07 3.32
DTI parameters (leg)
 Coil 16 channels 32 channels 18 channels
 Directions 12 12 12
 b-value (s/mm2) 400 400 400
 TR (ms) 7950 7500 7500
 TE (ms) Minimum 66 83
 Matrix 128 × 128 128 × 128 128 × 128
 Parallel factor 2 (asset) 2 (sense) 2 (grappa)
 Voxel size (mm) 2.3 × 2.3 × 4.0 2.3 × 2.3 × 4.0 2.3 × 2.3 × 4.0
 Acquisition time (minutes:seconds) 3.35 4.07 3.32
DTI parameters (arm)
 Coil 16 channels 8 channels 18 channels
 Directions 12 12 12
 b-value (s/mm2) 400.0 400 400
 TR (ms) 8600 7837 7900
 TE (ms) Minimum 79 83
 Matrix 122 × 122 122 × 122 122 × 122
 Parallel factor 2 (asset) 2 (sense) 2 (grappa)
 Voxel size (mm) 1.6 × 1.6 × 4.0 1.6 × 1.6 × 4.0 1.6 × 1.6 × 4.0
 Acquisition time (minutes: seconds) 3.52 4.18 3.43

Table 2.

T1 TSE acquisition parameters

GE Philips Siemens
Thigh
 TR (ms) 604 519 600
 TE (ms) Minimum 13 11
 Matrix 448 448 448
 Parallel factor 2 (asset) (2 sense) 2 (grappa)
 Voxel size (mm) 0.9 × 0.9 × 4.0 0.9 × 0.9 × 4 0.9 × 0.9 × 4.0
 Acquisition time (minutes: seconds) 2.01 1.36 1.52
Leg
 TR (ms) 618 519 600
 TE (ms) Minimum 13 11
 Matrix 384 × 384 384 × 384 384 × 384
 Parallel factor 2 (asset) 2 (sense) 2 (grappa)
 Voxel size (mm) 0.8 × 0.8 × 4.0 0.8 × 0.8 × 4.0 0.8 × 0.8 × 4.0
 Acquisition time (minutes: seconds) 1.49 1.33 1.38
Arm
 TR (ms) 635.0 519 605.0
 TE (ms) Minimum 13 11.0
 Matrix 256 × 256 256 × 256 256 × 256
 Parallel factor 2 (asset) 2 (sense) 2 (grappa)
 Voxel size (mm) 0.7 × 0.7 × 4.0 0.7 × 0.7 × 4.0 0.7 × 0.7 × 4.0
 Acquisition time (minutes: seconds) 1.40 1.25 1.29

Image analysis

Following data acquisition and after removing all patient identifying information, a radiologist with eight years of experience in MSK MR interpretation assessed image quality [18]. Then, m-DTI parameters on different muscle compartments were independently assessed by two radiologists (8 and 10 years of experience in the MSK field) using a commercially available software (Olea sphere 3.0). The muscle regions of interest (ROIs) were selected as described in Fig. 1. Post-processing was performed on the DTI images. Motion-related misalignments and adjacent image noise were corrected with automated image registration. Both readers manually drew the ROIs on the same slices at the middle third of the thigh, leg, and arm on axial T1w sequences as shown in Fig. 1. FA, RD, and MD values of the different muscle areas were calculated. Fibre tractography of the thigh, leg, and arm is shown in Fig. 2.

Fig. 1.

Fig. 1

Axial T1w images showing ROIs of the different anatomical compartments. a 1 rectus femoris, 2 vastus medialis, 3 vastus lateralis, 4 vastus intermedius, 5 sartorius, 6 gracilis, 7 biceps femoris, 8 semitendinosus, 9 semimembranosus. b 1 Medial gastrocnemius, 2 lateral gastrocnemius, 3 soleus, 4 anterior tibialis, 5 peroneal muscles, 6 posterior tibialis, 7 flexor digitorum longus, 8 flexor hallucis longus. c 1 Medial head of triceps brachii, 2 lateral head of triceps brachii, 3 long head of triceps brachii, 4 biceps brachii, and 5 coraco brachialis

Fig. 2.

Fig. 2

Axial color FA map with overlaid tractography of the thighs (a), legs (b), and arm (c)

Statistical analysis

Results are reported as medians and interquartile ranges (IQR). Data distributions were checked for normality using the Shapiro–Wilk test, which showed that all data were non-normally distributed (p < 0.05). Next, the intra-class correlation coefficient (ICC) was used to determine between-observers agreement for overall measurements and clinical sites, keeping in this last case, the distinction between left and right side, so to use the contralateral side as a double check of agreement. Next, between-group comparisons were made through the nonparametric Friedman’s test. We applied Friedman’s test for each observer’s measurements to prevent biased findings. Finally, the Bland–Altman method was used to determine agreement among the three scanner measurements, comparing them two by two. A p value less than 0.01 was considered to be statistically significant. Statistical analyses were executed by MedCalc Statistical Software version 19.2.6 (MedCalc Software bv, Ostend, Belgium; https://www.medcalc.org; 2020).

Results

Agreement between observers

The ICC reported high levels of agreement between the two observers as summarized in Table 3.

Table 3.

Intra-class correlation

GE Philips Siemens
MD 0.95 (0.94–0.96) 0.92 (0.90–0.94) 0.88 (0.85–0.90)
FA 0.91 (0.88–0.93) 0.95 (0.93–0.96) 0.85 (0.81–0.88)
RD 0.89 (0.86–0.91) 0.92 (0.90–0.94) 0.89 (0.85–0.90)

Good to excellent ICC values (higher than 0.69) were assessed between the two observers according to anatomical sites, except for FA measurement on Siemens MR (0.62, 95% CI 0.43–0.76). Detailed results are reported in Table 4.

Table 4.

ICC for clinical sites

GE Philips Siemens
MD
 Right arm 0.97 (0.95–0.99) 0.89 (0.78–0.95) 0.76 (0.56–0.88)
 Left arm 0.96 (0.91–0.98) 0.96 (0.93–0.98) 0.69 (0.44–0.84)
 Right thigh 0.92 (0.87–0.95) 0.94 (0.90–0.96) 0.94 (0.89–0.96)
 Left thigh 0.94 (0.89–0.96) 0.86 (0.77–0.91) 0.97 (0.94–0.98)
 Right leg 0.94 (0.91–0.97) 0.96 (0.93–0.98) 0.91 (0.84–0.95)
 Left leg 0.96 (0.94–0.98) 0.88 (0.79–0.93) 0.92 (0.86–0.95)
FA
 Right arm 0.96 (0.92–0.98) 0.86 (0.73–0.93) 0.88 (0.76–0.94)
 Left arm 0.71 (0.49–0.85) 0.92 (0.83–0.96) 0.94 (0.87–0.97)
 Right thigh 0.93 (0.88–0.96) 0.96 (0.93–0.98) 0.62 (0.43–0.76)
 Left thigh 0.97 (0.94–0.98) 0.94 (0.90–0.96) 0.96 (0.94–0.98)
 Right leg 0.86 (0.76–0.92) 0.93 (0.89–0.96) 0.93 (0.88–0.96)
 Left leg 0.93 (0.88–0.96) 0.95 (0.92–0.97) 0.87 (0.77–0.92)
RD
 Right arm 0.98 (0.95–0.99) 0.89 (0.78–0.94) 0.84 (0.69–0.92)
 Left arm 0.96 (0.93–0.98) 0.96 (0.92–0.98) 0.99 (0.98–0.99)
 Right thigh 0.92 (0.86–0.95) 0.93 (0.89–0.96) 0.92 (0.87–0.95)
 Left thigh 0.93 (0.89–0.96) 0.87 (0.79–0.92) 0.97 (094–0.98)
 Right leg 0.95 (0.90–0.97) 0.94 (0.90–0.97) 0.90 (0.83–0.94)
 Left leg 0.96 (0.93–0.98) 0.87 (0.78–0.93) 0.89 (0.81–0.94)

Inter-vendor reliability

No statistically significant inter-vendor differences were observed for both readers and for all the parameters. (Table 5). MD measurements for reader two were close to significance (p = 0.0573).

Table 5.

Inter-vendor differences in MD, FA, and RD

GE Philips Siemens p value*
Median (IQR) Median (IQR) Median (IQR)
1° Observer
 MD 1.563 (0.295) 1.533 (0.253) 1.571 (0.346) 0.1152
 FA 0.303 (0.059) 0.299 (0.071) 0.300 (0.072) 0.3772
 RD 1.319 (0.257) 1.301 (0.252) 1.340 (0.318) 0.1643
 CV of MD 0.17 0.14 0.23
 CV of FA 0.15 0.16 0.17
 CV of RD 0.18 0.15 0.23
2° Observer
 MD 1.559 (0.299) 1.541 (0.264) 1.592 (0.356) 0.0573
 FA 0.301 (0.059) 0.300 (0.069) 0.300 (0.067) 0.7046
 RD 1.320 (0.283) 1.312 (0.258) 1.343 (0.344) 0.3258
 CV of MD 0.18 0.14 0.23
 CV of FA 0.15 0.16 0.19
 CV of RD 0.18 0.14 0.25

*p value was determined through Friedman’s Test

MD mean diffusivity, RA radial anisotropy, FA fractional anisotropy, RD radial diffusivity, CV coefficient of variation

Bland–Altman plots comparing MD (Fig. 3), FA (Fig. 4), and RD (Fig. 5), were drawn. The Bland–Altmann method confirmed a high correlation between parameter values due to the slight deviation obtained in the mean values and the difference between them: All values were drawn in the limits of agreement (LoA ± 1.96 standard deviation).

Fig. 3.

Fig. 3

MD-A: Philips versus Siemens; B: GE versus Siemens; C: GE versus Philips

Fig. 4.

Fig. 4

FA-A: Philips versus Siemens; B: GE versus Siemens; C: GE versus Philips

Fig. 5.

Fig. 5

RD-A: Philips versus Siemens; B GE versus Siemens; C GE versus Philips

Discussion

Our study shows almost perfect inter-reader reliability for MD, FA, and RD on three different MR scanner and overall no statically significance differences among the three different vendors. A slightly decrease of inter-reader agreement was detected on Skyra MR for FA measurement in the right thigh. We suppose that this is due to the inclusion of tiny fatty areas of the subcutaneous tissue within the ROI [19].

Similar to previous studies on the nervous system, in our study the DTI values showed slightly statistical differences among different muscles as reported in Additional file (1) [13, 15, 1820]. We assessed highest FA (0.348; IQR: 0.097) on gracilis muscle and lowest FA on vastus intermedius (second observer GE FA = 0.229; IQR: 0.038). Nevertheless, we believe that these differences, albeit slightly statically different, are not clinically significant because there is no overlapping between our FA values and those reported in patients with spinal muscular atrophy or muscular dystrophies ranging from 0.7 to 0.41 [9, 2325]. Our results support the findings found by Fourè and colleagues who reported some differences among the FA values of the muscles of the lower limb [26]. However, values of the different muscles reported in this study are slightly different compared to ours. We believe that value discrepancies between the two studies may be due to different magnetic fields strength that determines higher SNR provided by acquisition on 7 T which improve fibre tracking compared to 3 T [27, 28].

Another study conducted on ten volunteers showed good reproducibility on both 3 and 7 T MR (Siemens Healthcare GmbH, Erlangen, Germany) with an SNR increase in the 7 T MR of up to 111% [12]. However, on the 7 T MR, the authors found higher FA and lower MD values in the soleus muscle, while the results of the remaining muscle compartments did not show significant statistical difference of the quantitative values between the MR. The authors justified the heterogeneity of the muscular values as the results of the same effect described by Polders et al., who found a higher uncertainty in peripheral areas of the brain on 7 T [29].

Interestingly, one study conducted on 18 healthy subjects on a single 3 T MR (Philips Medical Systems, Best, Netherlands) on the entire lower leg [30] and reported a statistically significant intra-muscle difference in FA between the origin of the muscle and the muscle belly. This is probably due to different chemical–physical properties of the actin-myosin components and the different amount and organization of collagen fibres.

M-DTI may be influenced by contraction and activity. Muscle contraction induces muscle fibre shortening and increases cross-sectional area (CSA), producing higher MD and lower FA values. Mazzoli et al. [31] performed MRI examinations of the lower leg on five volunteers during muscle contraction and found lower FA values of anterior tibialis in dorsiflexion compared with plantarflexion contraction and no contraction. The assessment of DTI values during muscle contraction is complex due to the length of current MRI sequences, which do not allow for constant, homogeneous fibre contraction. For this reason, we preferred to perform our MR examinations during muscle rest state. However, it is possible that in the future, with the increasingly widespread use of ultra-fast MRI sequences, the evaluation of muscle contraction will soon become available [32].

M-DTI may be influenced by activity, as well [33]. Hooijmans and colleagues described higher DTI values in upper legs muscles of 12 marathon runners in the post-marathon acquisition. The increase of MD and the decrease of FA are related to interstitial oedema and the alteration of diffusivity cellular barriers caused by muscle micro-trauma. These higher values returned to baseline (i.e. those values observed in the pre-marathon phase at the follow-up) after 3 weeks. MD and RD values are the first to return to the resting phase values, while FA values show a more prolonged alteration. This is the reason why we acquired MR examinations on the same days (within 6 h), without any sport activities performed the 4 weeks before the MR examination [34].

The first limitation of our study is the small sample size of volunteers, anyhow we assessed a large amount of data. Second, MR protocol parameters are not perfectly identical among the three MR scanners, because vendor-specific characteristics prevented us from applying exactly the same parameters for all the MR. However, other authors have used coils with different numbers of channels and DTI sequences with slightly different parameters to evaluate inter-observer and intra-observer agreement on the brain, obtaining promising results [13]. Moreover, the good results obtained with some parametric differences indirectly allow to obtain an even more significant inter-vendor agreement for clinical applications.

We believe that these reasons would make this sequence even more usable in clinical practice.

However, other studies with a larger sample of healthy volunteers are needed to confirm this claim.

Conclusions

Our results highlight the inter-vendor and inter-reader reproducibility of m-DTI values, and we strongly believe that the use of this sequence should be more included in the MRI protocols during daily clinical practice for the evaluation of MSK pathologies.

Supplementary Information

13244_2023_1374_MOESM1_ESM.pdf (192.2KB, pdf)

Additional file 1: Table S1. DTI values of each single muscle ROI of the arms of both observers. Table S2. DTI values of each single muscle ROI of the legs of both observers. Table S3. DTI values of each single muscle ROI of the thighs of both observers.

Abbreviations

CSA

Cross-sectional area

FA

Fractional anisotropy

ICC

Intra-class correlation coefficient.

IQR

Interquartile ranges

MD

Mean diffusivity

M-DTI

Muscle diffusion tensor imaging

MR

Magnetic resonance

MSK

Musculoskeletal

RD

Radial diffusivity

ROIs

Muscle regions of interest

T

Tesla

Author contributions

VC helped in conceptualization, data curation, measurements, and writing—original draft. DA contributed to methodology and measurements. SR helped in software and formal analysis. LMS was involved in review and data curation. MM reviewed the manuscript. FDG performed supervision and methodology. All authors read and approved the final manuscript.

Funding

This work has not received any funding.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

It was obtained from Swissethics, with waiver of informed consent.

Consent for publication

The authors of this manuscript received consent for publication.

Competing interests

LMS is a member of the Insights into Imaging Advisory Editorial Board. He has not taken part in the review or selection process of this article. All remaining authors declare that they have no conflict of interest.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Chianca V, Albano D, Messina C, et al. Diffusion tensor imaging in the musculoskeletal and peripheral nerve systems: from experimental to clinical applications. Eur Radiol Exp. 2017;1:12. doi: 10.1186/s41747-017-0018-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Zaraiskaya T, Kumbhare D, Noseworthy MD. Diffusion tensor imaging in evaluation of human skeletal muscle injury. J Magn Reson Imaging. 2006;24:402–408. doi: 10.1002/jmri.20651. [DOI] [PubMed] [Google Scholar]
  • 3.Bruno F, Arrigoni F, Mariani S, et al. Advanced magnetic resonance imaging (MRI) of soft tissue tumors: techniques and applications. Radiol Med. 2019;124:243–252. doi: 10.1007/s11547-019-01035-7. [DOI] [PubMed] [Google Scholar]
  • 4.Chianca V, Albano D, Messina C, et al. An update in musculoskeletal tumors: from quantitative imaging to radiomics. Radiol Med. 2021;126:1095–1105. doi: 10.1007/s11547-021-01368-2. [DOI] [PubMed] [Google Scholar]
  • 5.Wheeler-Kingshott CAM, Cercignani M. About “axial” and “radial” diffusivities. Magn Reson Med. 2009;61:1255–1260. doi: 10.1002/mrm.21965. [DOI] [PubMed] [Google Scholar]
  • 6.HeeMSKerk AM, Sinha TK, Wilson KJ, Ding Z, Damon BM. Repeatability of DTI-based skeletal muscle fiber tracking. NMR Biomed. 2010;23:294–303. doi: 10.1002/nbm.1463. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Budzik J-F, Balbi V, Verclytte S, Pansini V, Le TV, Cotten A. Diffusion tensor imaging in musculoskeletal disorders. Radiographics. 2014;34:E56–E72. doi: 10.1148/rg.343125062. [DOI] [PubMed] [Google Scholar]
  • 8.Williams SE, HeeMSKerk AM, Welch EB, Li K, Damon BM, Park JH. Quantitative effects of inclusion of fat on muscle diffusion tensor MRI measurements. J Magn Reson Imaging. 2013;38:1292–1297. doi: 10.1002/jmri.24045. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Li GD, Liang YY, Xu P, Ling J, Chen YM. Diffusion-tensor imaging of thigh muscles in Duchenne muscular dystrophy: correlation of apparent diffusion coefficient and fractional anisotropy values with fatty infiltration. AJR Am J Roentgenol. 2016;206:867–870. doi: 10.2214/AJR.15.15028. [DOI] [PubMed] [Google Scholar]
  • 10.Van Dyck P, Froeling M, De Smet E, et al. Diffusion tensor imaging of the anterior cruciate ligament graft. J Magn Reson Imaging. 2017;46:1423–1432. doi: 10.1002/jmri.25666. [DOI] [PubMed] [Google Scholar]
  • 11.Van Dyck P, Billiet T, Desbuquoit D, et al. Diffusion tensor imaging of the anterior cruciate ligament graft following reconstruction: a longitudinal study. Eur Radiol. 2020;30:6673–6684. doi: 10.1007/s00330-020-07051-w. [DOI] [PubMed] [Google Scholar]
  • 12.Giraudo C, Motyka S, Weber M, Feiweier T, Trattnig S, Bogner W. Diffusion tensor imaging of healthy skeletal muscles: a comparison between 7 T and 3 T. Invest Radiol. 2019;54:48–54. doi: 10.1097/RLI.0000000000000508. [DOI] [PubMed] [Google Scholar]
  • 13.Min J, Park M, Choi JW, Jahng GH, Moon WJ. Inter-vendor and inter-session reliability of diffusion tensor imaging: implications for multicenter clinical imaging studies. Korean J Radiol. 2018;19:777–782. doi: 10.3348/kjr.2018.19.4.777. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Galbán CJ, Maderwald S, Uffmann K, Ladd ME. A diffusion tensor imaging analysis of gender differences in water diffusivity within human skeletal muscle. NMR Biomed. 2005;18:489–498. doi: 10.1002/nbm.975. [DOI] [PubMed] [Google Scholar]
  • 15.Jovicich J, Marizzoni M, Bosch B, et al. Multisite longitudinal reliability of tract-based spatial statistics in diffusion tensor imaging of healthy elderly subjects. Neuroimage. 2014;101:390–403. doi: 10.1016/j.neuroimage.2014.06.075. [DOI] [PubMed] [Google Scholar]
  • 16.Teipel SJ, Reuter S, Stieltjes B, et al. Multicenter stability of diffusion tensor imaging measures: a European clinical and physical phantom study. Psychiatry Res Neuroimaging. 2011;194:363–371. doi: 10.1016/j.pscychresns.2011.05.012. [DOI] [PubMed] [Google Scholar]
  • 17.Vollmar C, O’Muircheartaigh J, Barker GJ, et al. Identical, but not the same: intra-site and inter-site reproducibility of fractional anisotropy measures on two 3.0T scanners. Neuroimage. 2010;51:1384–1394. doi: 10.1016/j.neuroimage.2010.03.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Kermarrec E, Budzik JF, Khalil C, Le Thuc V, Hancart-Destee C, Cotten A. In vivo diffusion tensor imaging and tractography of human thigh muscles in healthy subjects. AJR Am J Roentgenol. 2010;195:W352–366. doi: 10.2214/AJR.09.3368. [DOI] [PubMed] [Google Scholar]
  • 19.Damon BM. Effects of image noise in muscle diffusion tensor (DT)-MRI assessed using numerical simulations. Magn Reson Med. 2008;60:934–944. doi: 10.1002/mrm.21707. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Grech-Sollars M, Hales PW, Miyazaki K, et al. Multi-centre reproducibility of diffusion MRI parameters for clinical sequences in the brain. NMR Biomed. 2015;28:468–485. doi: 10.1002/nbm.3269. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Magnotta VA, Matsui JT, Liu D, et al. MultiCenter reliability of diffusion tensor imaging. Brain Connect. 2012;2:345–355. doi: 10.1089/brain.2012.0112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Fox RJ, Sakaie K, Lee JC, et al. A validation study of multicenter diffusion tensor imaging: reliability of fractional anisotropy and diffusivity values. AJNR Am J Neuroradiol. 2012;33:695–700. doi: 10.3174/ajnr.A2844. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Barp A, Carraro E, Albamonte E, et al. Muscle MRI in two SMA patients on nusinersen treatment: a two years follow-up. J Neurol Sci. 2020;417:117067. doi: 10.1016/j.jns.2020.117067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Ponrartana S, Ramos-Platt L, Wren TAL, et al. Effectiveness of diffusion tensor imaging in assessing disease severity in Duchenne muscular dystrophy: preliminary study. Pediatr Radiol. 2015;45:582–589. doi: 10.1007/s00247-014-3187-6. [DOI] [PubMed] [Google Scholar]
  • 25.Otto LAM, van der Pol WL, Schlaffke L, et al. Quantitative MRI of skeletal muscle in a cross-sectional cohort of patients with spinal muscular atrophy types 2 and 3. NMR Biomed. 2020;33:e4357. doi: 10.1002/nbm.4357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Fouré A, Ogier AC, Le Troter A, et al. Diffusion properties and 3D architecture of human lower leg muscles assessed with ultra-high-field-strength diffusion-tensor MR imaging and tractography: reproducibility and sensitivity to sex difference and intramuscular variability. Radiology. 2018;287:592–607. doi: 10.1148/radiol.2017171330. [DOI] [PubMed] [Google Scholar]
  • 27.Yao X, Yu T, Liang B, Xia T, Huang Q, Zhuang S. Effect of increasing diffusion gradient direction number on diffusion tensor imaging fiber tracking in the human Brain. Korean J Radiol. 2015;16:410–418. doi: 10.3348/kjr.2015.16.2.410. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Froeling M, Nederveen AJ, Nicolay K, Strijkers GJ. DTI of human skeletal muscle: the effects of diffusion encoding parameters, signal-to-noise ratio and T2 on tensor indices and fiber tracts. NMR Biomed. 2013;26:1339–1352. doi: 10.1002/nbm.2959. [DOI] [PubMed] [Google Scholar]
  • 29.Polders DL, Leemans A, Hendrikse J, Donahue MJ, Luijten PR, Hoogduin JM. Signal to noise ratio and uncertainty in diffusion tensor imaging at 1.5, 3.0, and 7.0 Tesla. J Magn Reson Imaging. 2011;33:1456–1463. doi: 10.1002/jmri.22554. [DOI] [PubMed] [Google Scholar]
  • 30.Schlaffke L, Rehmann R, Froeling M, et al. Diffusion Tensor Imaging of the Human Calf: variation of Inter- and Intramuscle-Specific Diffusion. Parameters. 2017;46:1137–1148. doi: 10.1002/jmri.25650. [DOI] [PubMed] [Google Scholar]
  • 31.Mazzoli V, Moulin K, Kogan F, Hargreaves BA, Gold GE. Diffusion tensor imaging of skeletal muscle contraction using oscillating gradient spin Echo. Front Neurol. 2021;15(12):608549. doi: 10.3389/fneur.2021.608549. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Del Grande F, Rashidi A, Luna R, et al. Five-minute five-sequence knee MRI using combined simultaneous multislice and parallel imaging acceleration: comparison with 10-minute parallel imaging knee MRI. Radiology. 2021;299:635–646. doi: 10.1148/radiol.2021203655. [DOI] [PubMed] [Google Scholar]
  • 33.Hooijmans MT, Monte JRC, Froeling M, et al. Quantitative MRI reveals microstructural changes in the upper leg muscles after running a marathon. J Magn Reson Imaging. 2020;52:407–417. doi: 10.1002/jmri.27106. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Froeling M, Oudeman J, Strijkers GJ. Muscle changes detected with diffusion-tensor imaging after long-distance running. Radiology. 2015;274:548–562. doi: 10.1148/radiol.14140702. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

13244_2023_1374_MOESM1_ESM.pdf (192.2KB, pdf)

Additional file 1: Table S1. DTI values of each single muscle ROI of the arms of both observers. Table S2. DTI values of each single muscle ROI of the legs of both observers. Table S3. DTI values of each single muscle ROI of the thighs of both observers.

Data Availability Statement

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.


Articles from Insights into Imaging are provided here courtesy of Springer