Abstract
Purpose
The purpose of this study was to demonstrate the potential utility of cone‐beam computed tomography (CBCT)‐guided online adaptive radiotherapy (ART) under end‐exhalation breath‐hold (EE‐BH) conditions for pancreatic cancer (PC).
Methods
Eleven PC patients who underwent 15‐fraction volumetric‐modulated arc therapy under EE‐BH conditions were included. Planning CT images and daily 165 CBCT images were imported into a dedicated treatment planning system. The prescription dose was set to 48 Gy in 15 fractions. The reference plan was automatically generated along with predefined clinical goals. After segmentation was completed on CBCT images, two different plans were generated: One was an adapted (ADP) plan in which re‐optimization was performed on the anatomy of the day, and the other was a scheduled (SCH) plan, which was the same as the reference plan. The dose distributions calculated using the synthetic CT created from both planning CT and CBCT were compared between the two plans. Independent calculation‐based quality assurance was also performed for the ADP plans, with a gamma passing rate of 3%/3 mm.
Results
All clinical goals were successfully achieved during the reference plan generation. Of the 165 sessions, gross tumor volume D 98% and clinical target volume D 98% were higher in 100 (60.1%) and 122 (74.0%) ADP fractions. In each fraction, the V 3 Gy < 1 cm3 of the stomach and duodenum was violated in 47 (28.5%) and 48 (29.1%), respectively, of the SCH fractions, whereas no violations were observed in the ADP fractions. There were statistically significant differences in the dose–volume indices between the SCH and ADP fractions (p < 0.05). The gamma passing rates were above 95% in all ADP fractions.
Conclusions
The CBCT‐guided online ART under EE‐BH conditions successfully reduced the dose to the stomach and duodenum while maintaining target coverage.
Keywords: CBCT‐guided adaptive radiotherapy, dosimetric analysis, independent calculation‐based QA, pancreatic cancer
1. INTRODUCTION
Pancreatic cancer (PC) is the fourth leading cause of death in the United States, and morbidity rates have been increasing in recent years. 1 Most cases are inoperable at the time of initial diagnosis, and multidrug chemotherapy has a prolonged prognosis 2 ; however, the 5‐year survival rate is still low at 10%. 1 Because of the poor prognosis, PC needs to be treated with multimodality therapies, such as surgery, radiation therapy, and chemotherapy. From this point of view, radiation therapy, which can be used in combination with chemotherapy, is increasingly important.
Escalating radiation dose may contribute to the improvement of treatment outcomes 3 , 4 , 5 ; however, the pancreas is located adjacent to the stomach or duodenum that are sensitive to radiation, and these structures move with respiration, 6 , 7 making it difficult to deliver high doses to the lesions even with volumetric‐modulated arc therapy (VMAT). Several studies have been published on the relationship between gastrointestinal (GI) adverse events and dose–volume constraints, 8 , 9 , 10 and the initial treatment plan is typically generated with reference to the published dose–volume constraints. However, the stomach and duodenum have respiratory motion and interfractional physiologic deformation, leading to large dose deviations in both organs and target from the planned dose. 11 , 12 From a recent phase I dose‐escalation trial with stereotactic body radiation therapy (SBRT) for PC, it was reported that radiation‐induced severe GI adverse events were observed at a nontrivial rate. 13 Although offline adaptive radiotherapy (ART) with respiratory motion management technique is one of the approaches to manage respiratory motion and organ deformation, interfractional dosimetric variations are not always predictable even after replanning.
In recent years, the use of online ART, which allows us to immediately generate a new plan reflecting the daily organ deformation, has allowed the delivery of a high dose to the target safely while sparing doses to organs at risk (OARs). There are three types of online ART techniques: magnetic resonance (MR)‐guided, 14 , 15 , 16 , 17 , 18 , 19 computed tomography (CT)‐on‐rails‐guided, 20 , 21 and cone‐beam CT (CBCT)‐guided approaches. 22 , 23 , 24 , 25 , 26 It is well known that MR images provide superior soft‐tissue contrast and delineation precision compared to CBCT. With the advantages of these characteristics, MR‐guided ART has been clinically applied to PC and has demonstrated its dosimetric advantages. 17 , 18 , 19 ART with CT‐on‐rails has the advantage of daily treatment planning with image quality comparable to that of planning CT; however, there is a positional error associated with couch rotation. 27 More recently, the image quality of CBCT has been greatly improved, even in the upper GI region owing to high‐speed gantry rotation, enhanced noise‐canceling grid, and advanced image reconstruction algorithms. 28 , 29 , 30 Several studies have demonstrated the clinical efficacy of CBCT‐guided ART in patients with prostate, bladder, cervical, and rectal cancer. 22 , 23 , 24 , 25 , 26 Li et al. used in‐house‐developed graphics processing unit‐based replanning system and found the dosimetric utility of daily adaptive replanning with SBRT for PC 31 ; however, they did not employ respiratory motion management techniques, and their re‐optimization protocol was based on equal priority, making it difficult to adjust doses, especially to the serial organs. In addition, they did not conduct patient‐specific quality assurance (QA).
In this study, we demonstrated the potential utility of CBCT‐guided online ART under end‐exhalation breath‐hold (EE‐BH) conditions for PC. We created a daily treatment plan on synthetic CT generated from the initial planning CT and daily CBCT in accordance with clinical intent and evaluated the differences in dose distributions to the target and OARs with and without daily adaptation for each session in patients with PC previously treated with 15‐fractionation VMAT under EE‐BH conditions at our institution. In addition to dosimetric comparison, independent calculation‐based QA was performed for daily adapted plans.
2. METHODS
2.1. Patients
Institutional ethical approval was obtained before the study (approval number: R2762). Twelve consecutive patients with PC who underwent 15‐fraction VMAT under EE‐BH conditions with TrueBeam STx (Varian Medical Systems, Palo Alto, CA, USA) between November 2019 and July 2020 at our institution were enrolled. Of these, one patient (patient 7) was excluded from this study because daily CBCT, which was acquired with a lower mAs value for the patient's body shape, was not available at the segmentation stage (described below) due to extremely poor image quality. Consequently, 11 patients with PC were analyzed in this study (Table 1). All patients had histologically proven pancreatic adenocarcinoma and were diagnosed with borderline resectable or locally advanced unresectable PC on the tumor board at our institution. Oral intake was stopped except for drugs and water for at least 3 h before CT or CBCT acquisition.
TABLE 1.
Patients’ characteristics
| Pt. # | Age (year) | Sex | Primary site | Stage | TNM (UICC 8th) | GTV size (cm3) | CTV size (cm3) | PTV size (cm3) | Overlap ratio (%) |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 72 | M | Body | BR | T4N0M0 | 12.3 | 94.9 | 173.1 | 10.7 |
| 2 | 59 | F | Body | BR | T4N1M0 | 111.6 | 217.5 | 348.1 | 18.5 |
| 3 | 71 | M | Head | BR | T4N0M0 | 38.1 | 144.3 | 249.2 | 15.2 |
| 4 | 79 | M | Head | UR | T4N0M0 | 13.7 | 101.2 | 198.9 | 12.7 |
| 5 | 57 | M | Body | UR | T4N0M0 | 87.3 | 203.0 | 330.7 | 8.9 |
| 6 | 71 | F | Head | BR | T2N0M0 | 31.3 | 119.1 | 119.1 | 16.7 |
| 8 | 78 | M | Head | BR | T2N0M0 | 5.3 | 104.2 | 190.9 | 6.9 |
| 9 | 78 | M | Body | BR | T4N0M0 | 17.9 | 103.2 | 187.3 | 11.5 |
| 10 | 69 | M | Body | BR | T2N0M0 | 18.2 | 133.7 | 236.2 | 15.2 |
| 11 | 64 | F | Head | BR | T4N0M0 | 28.1 | 91.6 | 184.8 | 15.6 |
| 12 | 62 | F | Head | UR | T4N0M0 | 89.8 | 184.0 | 297.7 | 11.6 |
Note: Patient 7 was excluded from this study because daily CBCT, which was performed with a lower mAs value for the patient's body shape, was not available at the segmentation stage due to extremely poor image quality. Overlap refers to the area where the PTV and PRV overlap. The overlap ratio was calculated as the overlap volume to the PTV volume.
Abbreviations: BR, borderline resectable; CBCT, cone‐beam computed tomography; CTV, clinical target volume; GTV, gross tumor volume; PRV, planning organs at risk volume; PTV, planning target volume; UR, unresectable.
2.2. Initial planning CT acquisition, contouring, and daily CBCT acquisition
All patients were immobilized in a headfirst supine position with both upper limbs elevated using an individualized vacuum pillow (Body Fix; Elekta, Stockholm, Sweden). Non‐contrast and transvenous contrast CT were then acquired under EE‐BH conditions with a real‐time position management system (RPM; Varian Medical Systems). The operator instructed the patient to hold their breath while monitoring the patient's respiratory signal with the RPM system. Patients held their breath according to the operator's instruction only without visual feedback. This was done in accordance with a previous study where further details can be found. 32
The gross tumor volume (GTV) included the primary tumor and metastatic lymph nodes. The clinical target volume (CTV) was defined as the GTV plus a 5 mm margin and the potential para‐aortic lymph node and neuroplexus involvement between the celiac axis and the superior mesenteric artery. A planning target volume (PTV) margin of 5 mm was isotropically added to the CTV. The surrounding OARs, such as the stomach and duodenum, were also contoured. Planning organs at risk volume (PRV) margins of 3‐ and 5 mm were added to the duodenum and stomach, respectively, based on previous data. 33 These structures were defined by a board‐certified radiation oncologist.
On the day of treatment, CBCT was acquired just before beam delivery under EE‐BH conditions with the RPM system. One gantry rotation was divided into three to four segments to acquire CBCT images, depending on the patient's BH ability. The CBCT field‐of‐view (FOV) was 26.3 cm, and the slice thickness was 2 mm. The images were reconstructed using an iterative reconstruction algorithm with a medium noise suppression level. 34
2.3. Workflow of CBCT‐based online adaptive radiotherapy
The overall workflow of this study is shown in Figure 1, which consisted of initial planning, daily plan generation, and independent calculation‐based QA.
FIGURE 1.

Overview of the present study
The initial planning CT and structures used for the previous treatment were imported to the Ethos therapy solution emulator version 1.1 (Varian Medical Systems). Reference plans with 3‐arc VMAT were generated according to an institutional treatment protocol for locally advanced PC. Dose distributions were calculated using Acuros XB with a 2.5 × 2.5 × 2.5 mm3 grid resolution. The prescription dose was 48 Gy in 15 fractions with a dose covering 95% of the target (D 95%) to the PTV‐PRV, a volume that subtracted PRVs from the PTV. A PTV‐PRV of D 95% ≥ 48 Gy and PTV D 98% ≥ 36 Gy were preferable; however, if this was difficult to achieve within the constraints of the GI tract, the PTV‐PRV and PTV were reduced to D 95% ≥ 45.6 Gy and D 98% ≥ 34.2 Gy, respectively. The dose–volume constraints for OARs were also the same as in a previous institutional trial. 35 All clinical goals, the dose–volume constraints and the priority are shown in Table 2.
TABLE 2.
Clinical goals and priority used for the reference plan and adaptive plan
| Priority | Importance | Structure | Goals | Acceptable variation |
|---|---|---|---|---|
| 1 | Most important | PTV‐PRVa |
D 95% ≥ 48 Gy (D 95% ≥ 45.6 Gy)b |
|
| 2 | Very important | Stomach PRVa | V 36 Gy < 45 cm3 | |
| 2 | Very important | Overlap |
D 98% ≥ 36 Gy (D 98% ≥ 34.2 Gy)b |
|
| 2 | Very important | PTV‐PRV | D max < 55 Gy | D max < 56 Gy |
| 2 | Very important | Duodenum PRVa | V 36 Gy < 45 cm3 | |
| 2 | Very important | Spinal corda | D max < 36 Gy | |
| 2 | Very important | Spinal cord PRVa | < 39 Gy | |
| 2 | Very important | Stomach PRVa | V 39 Gy < 15 cm3 | V 39 Gy < 30 cm3 |
| 2 | Very important | Duodenum PRVa | V 39 Gy < 15 cm3 | V 39 Gy < 30 cm3 |
| 2 | Very important | Duodenuma | V 45 Gy < 1 cm3 | |
| 2 | Very important | Duodenuma | V 42 Gy < 5 cm3 | |
| 2 | Very important | Duodenuma | V 39 Gy < 25 cm3 | |
| 2 | Very important | Stomacha | V 45 Gy < 1 cm3 | |
| 2 | Very important | Stomacha | V 42 Gy < 5 cm3 | |
| 2 | Very important | Stomacha | V 39 Gy < 25 cm3 | |
| 3 | Important | Lt kidneya | V 20 Gy < 30% | |
| 3 | Important | Rt kidneya | V 20 Gy < 30% | |
| 3 | Important | Livera | D mean < 30 Gy | |
| 3 | Important | Overlap | D 5% ≤ 48 Gy | D 5% ≤ 49 Gy |
| 3 | Important | PTV‐PRV | D 50% ≥ 51 Gy | D 50% ≥ 50 Gy |
Abbreviations: D max, maximum dose; D mean, mean dose; Dxx %, dose covering xx%; Lt, left; PRV, planning organ at risk volume; PTV, planning target volume; Rt, right; Vxx Gy, the volume received by xx Gy.
aDose–volume constraints are the same in our facility for locally advanced pancreatic cancer.
b As one patient (patient 2) did not meet the clinical goals due to a large overlapping volume between PTV and PRVs, the case was allowed a PTV‐PRV of D 95% ≥ 45.6 Gy (95% of 48 Gy) and PTV of D 98% ≥ 34.2 Gy (95% of 36 Gy).
Daily CBCT images for each patient were also imported into the Ethos emulator in sequence. The following adaptive session was performed for each CBCT: First, the predefined OARs (duodenum, stomach, and liver) were automatically segmented using full image deep convolutional neural networks with proprietary architectures that share many similarities with U‐Net and DenseNet implemented on the Ethos system. 34 In this study, the duodenum, stomach, and liver were automatically segmented. The contours for the other OARs (kidneys and spinal cord) were then automatically propagated from the initial planning CT using a modified basis spline (B‐spline) deformable image registration (DIR) algorithm, followed by a simple post‐processing and smoothing of the results. 36 The small and large intestines were not contoured on the CBCT images in this study. In contrast, the targets were rigidly propagated from the initial planning CT via rigid registration. These structures should be reviewed and manually modified until clinically acceptable levels were reached by a board‐certified radiation oncologist in reference to the initial planning CT when needed.
Thereafter, the adapted (ADP) plan was generated under the same clinical goals as the reference plan, with daily approved contours. Optimization and dose distribution were conducted on the daily synthetic CT. The synthetic CT was generated by deforming the initial planning CT using the B‐spline DIR algorithm. 36 To compensate for the limited FOV as well as the motion artifacts on a CBCT, each CBCT voxel Hounsfield unit value was replaced with those on the planning CT. Furthermore, the planning CT voxel information was propagated via a rigid registration to the areas outside the FOV of the CBCT to compensate for the limited scan length. In order to increase registration quality for CT‐to‐CBCT DIR, only those voxels that lied inside the CBCT reconstruction radius were used in the cost function. This avoided registration artifacts in cases where the reconstruction radius truncates the body. In parallel with the ADP plan generation, the dose distribution was calculated on the daily synthetic CT using the initial plan information, which was termed the scheduled (SCH) plan. After calculating the dose distributions, the absence of inappropriate hot spots outside the target was confirmed visually.
2.4. Dosimetric analysis
All plans were exported to the Eclipse version 16.1 (Varian Medical Systems), and dose–volumetric indices (DVIs) in the stomach, duodenum, GTV, and CTV were examined for the ART and SCH fractions.
These DVIs were analyzed statistically between the ADP and SCH fractions using the Wilcoxon rank sum test. The significance level was set at 5%. All statistical analyses were conducted using EZR (Saitama Medical Center, Jichi Medical University, Japan).
2.5. Independent calculation‐based QA
The DICOM dataset (CT images, RT structures, RT plan, and RT dose) was exported to an independent dose calculation software, MobiusAdapt (Mobius Medical Systems, LLC, Houston, TX, USA). Currently, MobiusAdapt is the only software to conduct independent calculation‐based QA for the Ethos system. A collapsed cone convolution/superposition algorithm was implemented as the dose calculation algorithm. MobiusAdapt calculated dose distribution based on the RT plan and then compared it with the dose distribution stored in RT dose. The imported RT dose was used as reference. In this study, the global gamma passing rates (3%/2 and 3%/3 mm) with 10% dose threshold criteria were evaluated. 22 , 26 , 37 , 38
To investigate the change in the gamma passing rate, Pearson's correlation between the gamma passing rate and monitor units (MUs) difference between the reference plan and ADP fractions or overlap ratio of PRVs to PTV was analyzed.
2.6. Time evaluation
The time required to complete segmentation, optimization, and dose calculation was measured with a stopwatch or extracted from the system log.
3. RESULTS
3.1. Initial plan generation
All clinical goals listed in Table 2 were successfully achieved at the stage of initial plan generation, except for one patient (patient 2) who had a large overlap volume between PTV and PRVs. The clinical goals shown in parentheses for the target volumes were applied in this case (Table 2).
3.2. Comparison of the ADP plan with the SCH plan
A total of 165 sessions (11 patients × 15 CBCT images) were analyzed. Figure 2 shows the scatter plots of the DVIs of the GTV, CTV, duodenum, and stomach for the ADP and SCH fractions. Of 165 sessions, GTV D 98% and CTV D 98% were higher in 100 (60.1%) and 122 (74.0%) ADP fractions, respectively. The median doses of GTV D 98% and CTV D 98% were 2.96 and 2.72 Gy in SCH fractions and 2.98 and 2.84 Gy in ADP fractions, respectively. Meanwhile, the V 3 Gy < 1 cm3 of the stomach and duodenum was violated in 47 (28.5%) and 48 (29.1%) SCH fractions, respectively, whereas no violations were observed in the ADP fractions (Figure 2c,f). The median doses of the stomach and duodenum were 2.75 and 2.88 Gy in SCH, and 2.54 and 2.67 Gy in ADP fractions, respectively. There were statistically significant differences in the DVIs between the SCH and ADP fractions (p < 0.05). Figure 3 shows a typical example of dose distribution and dose–volume histograms. In the SCH fractions, the 100% isodose line (a surrogate for 45 Gy in 15 fractions) largely covered the duodenum, and dose coverages to the GTV and CTV were insufficient; however, these two drawbacks were resolved in the ADP plan.
FIGURE 2.

Scatter plots of the dose–volume indices of (a) gross tumor volume (GTV) D 98%, (b) clinical target volume (CTV) D 98%, (c) duodenal , (d) stomach , (e) duodenal , and (f) stomach in the scheduled and adapted fractions. If the plots are below the dotted line, it means that the value in the scheduled plan is higher than the value in the adapted plan. The color‐filled circles mean that CTV D 98% was higher in the scheduled plan than in the adapted plan in more than half of the 15 fractions. The yellow dashed lines in (c)–(f) show the dose–volume constraints used in this study.
FIGURE 3.

Representative (upper) dose distribution and (lower) dose–volume histograms for which the adaptive plan was effective. In the scheduled plan, the isodose lines indicated by yellow arrows (greater than 100% dose) largely cover the duodenum, and clinical target volume (CTV) coverage was insufficient as shown by the blue arrows; however, these two drawbacks were resolved in the adapted plan.
The overlap ratio of the stomach or duodenum PRV to PTV at the initial planning had a potential predictor for the number of violations of V 42 Gy > 5 cm3 in the SCH fractions (Figure S1). The correlation coefficients were 0.85 and 0.69 for the stomach V 42 Gy and duodenum V 42 Gy, respectively. These results indicated that the violations did not occur randomly but tended to occur in specific patients.
3.3. Independent calculation‐based QA
The mean ± standard deviation of the gamma passing rate was 93.5% ± 3.7% (range 82.8%–99.3%) and 98.9% ± 0.9% (range 95.7%–100.0%) with gamma criteria of 3%/2 and 3%/3 mm, respectively. Figure 4 shows the relationship between the difference in the MU from the reference plan or the overlap ratio and gamma passing rate, with the gamma criteria of 3%/2 and 3%/3 mm. The gamma passing rate decreased with a larger MU, and the correlation coefficient was −0.75 and −0.70, with the gamma criteria of 3%/2 and 3%/3 mm, respectively (Figure 4a,c). In addition, the correlation coefficient between the overlap ratio and gamma passing rate was −0.66 and −0.54, with the gamma criteria of 3%/2 and 3%/3 mm, respectively (Figure 4b,d). Of 165 ADP fractions, 29 (17.6%) had a gamma passing rate below 90% with the gamma criterion of 3%/2 mm. Of these 29 ADP fractions, 19 (65.5%) violated the dose–volume constraints for OARs if the SCH fraction was selected.
FIGURE 4.

The gamma passing rate as a function of (a and c) the difference in monitor units (MUs) from the reference plan and (b and d) overlap ratio of the planning organs at risk volumes (PRVs) to planning target volume (PTV). (a and b) A gamma passing rate of 3%/2 mm, and (c and d) a gamma passing rate of 3%/3 mm
3.4. Time evaluation
The means ± standard deviations of the time to complete segmentation, optimization, and dose calculation were 537.6 ± 107.1 s (range 378.0–723.0 s), 305.4 ± 20.6 s (range 239.6–359.7 s), and 28.8 ± 1.0 s (range 26.9–31.4 s), respectively.
4. DISCUSSIONS
This study aimed to demonstrate the clinical feasibility of CBCT‐based online ART for PC. We showed that 28.5% and 29.1% of SCH fractions violated the dose–volume constraints for the stomach and duodenum, respectively, whereas no violations for these organs were observed without reducing the target coverage of ADP fractions, and 60.1% and 74.0% of the ADP fractions had higher GTV D 98% and CTV D 98% than for the SCH fractions. The absence of a violation ensures that the Ethos system successfully optimized the ADP plan regardless of patient anatomy.
It is difficult to estimate the daily dose to the stomach and duodenum during initial planning because interfractional variations are unpredictable. Niedzielski et al. recently reported large dosimetric uncertainties resulting from interfractional variations in patients with PC; in four out of 11 patients, duodenal received a higher dose of 1 Gy per fraction than the planned dose. 12 The impact is especially pronounced in SBRT. Courtney et al. observed grades 4–5 late toxicities in a phase I dose‐escalation trial of SBRT for PC without online ART. 13 However, recent stereotactic MR‐guided online ART studies reported that an incident rate of grade 3 GI toxicity was acceptable after high‐dose SBRT (50 Gy in five fractions). 17 , 18 From these findings, the observation of dose–volume constraints of every single day would be essential, which can be realized with CBCT‐guided online ART.
Although there were differences in prescription doses, PRV margin sizes, and chemotherapy, previous studies have shown that the frequency of GI toxicity increases when certain dose–volume constraints are not achieved in the initial plan 8 , 9 , 10 From the present study, it is possible to understand the daily dose distribution. Although the visualization of the dose distribution for each session is a major progress, the accumulated dose of OAR should be a better indicator of adverse events. The DIR is one of the approaches to obtain the accumulated dose; however, its accuracy is low, especially in abdominal and pelvic regions. 39 , 40 Ziegler et al. accumulated the delivered dose for pancreatic patients based on CBCTs using DIR; however, they pointed out that there were still large residual registration errors, such as huge movement of air pocket. 39 Thus, a few CBCT datasets with severe artifacts were unavailable for dose accumulation in their study. Swamidas et al. pointed out that the current DIR algorithms are not yet robust. 40 Developments of reliable accumulated dose and clinical results are expected in the future.
To reduce the intrafractional error with online adaptive therapy, respiratory motion management should obviously be used, as the pancreas and OARs have been reported to move more than 10 mm along with respiration. 6 , 7 , 41 In the current clinical practice, RPM (Varian Medical Systems) cannot be used with the current Ethos version. If venders implement their own respiratory motion management devices in the near future or facilities use third‐party ones appropriately, such as AlignRT (Vision RT, London, UK) 42 and SpiroDynr'X system (SDX) (DYN'R, Aix‐en‐Provence, France), 43 our clinical workflow can be achieved.
In this study, we did not assess the dose to the bowel, considering their clinical impact and the reliability of the contouring. In our clinical intensity‐modulated radiotherapy (IMRT), the PRV for bowels was not generated because it was not always in the same position due to peristalsis and daily change in the amount of bowel content, and no critical adverse events have been observed so far. 33 , 44 Although McGinn et al. reported transverse colon stenosis in a dose‐escalation trial of 15 fractions, 45 they did not use IMRT. When employing hypofractionation, such as 5 fractions, it is more important to evaluate the bowel dose even if it takes more time to adaptation. In the Ethos platform, the high dose areas are automatically optimized not to extend beyond the target, and the radiation oncologist and medical physicist confirmed visually that the 45 Gy out of the PTV did not cover the bowel.
Patient‐specific QA is required to verify treatment delivery and dose calculation by the treatment planning system. Although patient‐specific QA is typically based on measurements, calculation‐based QA is the standard for online ART because the patient is on the treatment couch during daily plan generation. 38 In this study, independent calculation‐based QA was conducted for all ADP fractions with the gamma criteria of 3%/2 and 3%/3 mm. With the gamma criterion of 3%/2 mm recommended by AAPM TG218, 37 29 (17.6%) and 5 (3.0%) fractions had gamma passing rates below 90% and 85%, respectively. If these criteria are employed in clinical practice, some ADP fractions will not be acceptable. In contrast, the gamma passing rates were above 95% with the gamma criteria of 3%/3 mm, which were used by several investigators, 22 , 26 in all ADP fractions. Even with the gamma criteria of 3%/2 and 3%/3 mm, the gamma passing rate exhibited negative correlation with the MU difference between the reference plan and ADP fractions or the overlap ratio, which were supported by the results of Sibolt et al. 46 As stated in another study by Sibolt et al., 22 higher modulation factor (MU/Gy) might contribute to the negative correlation. AAPM TG219 recommends that users develop and use confidence limits from clinical data points. From our findings and the recommendation from AAPM TG219, careful determinations of the tolerance and action level are required before clinical application at each facility.
Online ART workflow itself is time‐sensitive; thus, time evaluation is essential. As expected, the time to complete segmentation was the longest, followed by optimization and dose calculation. As the accuracy of the segmentation determines the quality of the dose distribution, careful confirmation of the contouring is necessary. There is room for reconsideration regarding the time required for optimization. Additional internal review revealed that the time required for optimization was reduced by one third regardless of the number of ports when fixed gantry IMRT is selected. If the dose distribution is equivalent between IMRT and VMAT, IMRT would be the more practical choice.
This study has several limitations. First, the number of OAR violations during the course of treatment has not been shown to lead to severe adverse clinical events. Accurate determination of the clinical threshold is expected, as actual dose and clinical results are accumulated in the future. The second is contouring error. Even though the CBCT image quality has improved, artifacts caused by air pockets are still present, which were observed around stomach, duodenum, and bowels. As stated in the Methods section, OAR contours were visually checked and manually corrected if necessary. Niedzielski et al. concluded that the effect of contouring variability on the OAR was negligible by expanding or shrinking structures contoured on CT‐on‐rail images. 12 In this study, dose–volume constraints for not only OAR but were fulfilled, which would ensure the accurate delineation of OAR contours. The shape of the GTV is less deformable than the peristaltic intestine; therefore, the same shape at the initial planning CT can be used for daily plan generation. Taking oral contrast agents prior to treatment may be useful for improving image quality. 47 The third is the intrafraction variation. Intrafraction variations occurred immediately after the CBCT acquisition. One study reported that intramuscular butylscopolamine improved the MR image quality of the pancreas by suppressing GI peristalsis 48 ; thus, premedication may reduce intrafractional error.
5. CONCLUSIONS
The clinical feasibility of CBCT‐guided online ART for PC was investigated. Dose distributions were compared between ADP and SCH fractions using 15‐fraction VMAT. We have demonstrated that the ADP fractions met the dose–volume constraints in each session, although 28.5% and 29.1% of SCH fractions violated the stomach and duodenum doses, respectively. From the results of the independent calculation‐based QA, more MUs from the initial plan or a larger overlap ratio of PRVs to PTV would cause a decrease in the gamma passing rate with the criteria of 3%/2 mm; however, the decline was mitigated with the criteria of 3%/3 mm. In conclusion, CBCT‐guided online ART successfully reduced the dose to the stomach and duodenum while maintaining target coverage.
AUTHOR CONTRIBUTIONS
Ayaka Ogawa acquired, processed, and analyzed the data and wrote the manuscript. He also performed the statistical analysis. Mitsuhiro Nakamura and Hiraku Iramina helped to acquire the data, provided input for the study, and helped to write the manuscript. Mitsuhiro Nakamura, Michio Yoshimura, and Takashi Mizowaki conceptualized the project and oversaw all aspects of the work. All the authors reviewed and provided feedback on the manuscript.
CONFLICT OF INTEREST
Prof. Mizowaki and Dr. Nakamura have a collaborative research agreement with Varian Medical Systems. All other authors have no conflict of interest.
Supporting information
Supporting Information
ACKNOWLEDGMENTS
The authors thank Carol Huang, Masashi Kimura, and Takafumi Okoshi at Varian Medical Systems for their technical support.
Ogawa A, Nakamura M, Iramina H, Yoshimura M, Mizowaki T. Potential utility of cone‐beam CT‐guided adaptive radiotherapy under end‐exhalation breath‐hold conditions for pancreatic cancer. J Appl Clin Med Phys. 2023;24:e13827. 10.1002/acm2.13827
DATA AVAILABILITY STATEMENT
Research data are stored in an institutional repository and will be shared upon request to the corresponding author.
REFERENCES
- 1. Siegel RL, Miller KD, Fuchs HE, et al. Cancer Statistics, 2021. CA Cancer J Clin. 2021;71:7‐33. [DOI] [PubMed] [Google Scholar]
- 2. Conroy T, Desseigne F, Ychou M, et al. Folfirinox versus gemcitabine for metastatic pancreatic cancer. N Engl J Med. 2011;364:1817‐1825. [DOI] [PubMed] [Google Scholar]
- 3. Rudra S, Jiang N, Rosenberg SA, et al. Using adaptive magnetic resonance image‐guided radiation therapy for treatment of inoperable pancreatic cancer. Cancer Med. 2019;8:2123‐2132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Krishnan S, Chadha AS, Suh Y, et al. Focal radiation therapy dose escalation improves overall survival in locally advanced pancreatic cancer patients receiving induction chemotherapy and consolidative chemoradiation. Int J Radiat Oncol Biol Phys. 2016;94:755‐765. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Reyngold M, O'Reilly EM, Varghese AM, et al. Association of ablative radiation therapy with survival among patients with inoperable pancreatic cancer. JAMA Oncol. 2021;7:735‐738. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Uchinami Y, Suzuki R, Katoh N, et al. Impact of organ motion on volumetric and dosimetric parameters in stomach lymphomas treated with intensity‐modulated radiotherapy. J Appl Clin Med Phys. 2019;20:78‐86. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Umezawa R, Wakita A, Katsuta Y, et al. A pilot study of synchronization of respiration‐induced motions in the duodenum and stomach for the primary tumor in radiation therapy for pancreatic cancer using 4‐dimensional computed tomography. Adv Radiat Oncol. 2021;6:100730. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Nakamura A, Shibuya K, Matsuo Y, et al. Analysis of dosimetric parameters associated with acute gastrointestinal toxicity and upper gastrointestinal bleeding in locally advanced pancreatic cancer patients treated with gemcitabine‐based concurrent chemoradiotherapy. Int J Radiat Oncol Biol Phys. 2012;84:369‐375. [DOI] [PubMed] [Google Scholar]
- 9. Huang J, Robertson JM, Ye H, et al. Dose‐volume analysis of predictors for gastrointestinal toxicity after concurrent full‐dose gemcitabine and radiotherapy for locally advanced pancreatic adenocarcinoma. Int J Radiat Oncol Biol Phys. 2012;83:1120‐1125. [DOI] [PubMed] [Google Scholar]
- 10. Liu X, Ren G, Li L, et al. Predictive dosimetric parameters for gastrointestinal toxicity with hypofractioned radiotherapy in pancreatic adenocarcinoma. Onco Targets Ther. 2016;9:2489‐2494. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Liu F, Erickson B, Peng C, et al. Characterization and management of interfractional anatomic changes for pancreatic cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2012;83:e423‐e429. [DOI] [PubMed] [Google Scholar]
- 12. Niedzielski JS, Liu Y, Ng SSW, et al. Dosimetric uncertainties resulting from interfractional anatomic variations for patients receiving pancreas stereotactic body radiation therapy and cone beam computed tomography image guidance. Int J Radiat Oncol Biol Phys. 2021;111:1298‐1309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Courtney PT, Paravati AJ, Atwood TF, et al. Phase I trial of stereotactic body radiation therapy dose escalation in pancreatic cancer. Int J Radiat Oncol Biol Phys. 2021;110:1003‐1012. [DOI] [PubMed] [Google Scholar]
- 14. Henke L, Kashani R, Robinson C, et al. Phase I trial of stereotactic MR‐guided online adaptive radiation therapy (SMART) for the treatment of oligometastatic or unresectable primary malignancies of the abdomen. Radiother Oncol. 2018;126:519‐526. [DOI] [PubMed] [Google Scholar]
- 15. Paulson ES, Ahunbay E, Chen X, et al. 4D‐MRI driven MR‐guided online adaptive radiotherapy for abdominal stereotactic body radiation therapy on a high field MR‐linac: implementation and initial clinical experience. Clin Transl Radiat Oncol. 2020;23:72‐79. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Hyer DE, Cai B, Rong Y. Future mainstream platform for online adaptive radiotherapy will be using on‐board MR rather than on‐board (CB) CT images. J Appl Clin Med Phys. 2021;22:4‐9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Hassanzadeh C, Rudra S, Bommireddy A, et al. Ablative five‐fraction stereotactic body radiation therapy for inoperable pancreatic cancer using online MR‐guided adaptation. Adv Radiat Oncol. 2021;6:100506. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Chuong MD, Bryant J, Mittauer KE, et al. Ablative 5‐fraction stereotactic magnetic resonance‐guided radiation therapy with on‐table adaptive replanning and elective nodal irradiation for inoperable pancreas cancer. Pract Radiat Oncol. 2021;11:134‐147. [DOI] [PubMed] [Google Scholar]
- 19. Bohoudi O, Bruynzeel AME, Meijerink MR, et al. Identification of patients with locally advanced pancreatic cancer benefitting from plan adaptation in MR‐guided radiation therapy. Radiother Oncol. 2019;132:16‐22. [DOI] [PubMed] [Google Scholar]
- 20. Papalazarou C, Klop GJ, Milder MTW, et al. CyberKnife with integrated CT‐on‐rails: system description and first clinical application for pancreas SBRT. Med Phys. 2017;44:4816‐4827. [DOI] [PubMed] [Google Scholar]
- 21. Li X, Quan EM, Li Y, et al. A fully automated method for CT‐on‐rails‐guided online adaptive planning for prostate cancer intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2013;86:835‐841. [DOI] [PubMed] [Google Scholar]
- 22. Sibolt P, Andersson LM, Calmels L, et al. Clinical implementation of artificial intelligence‐driven cone‐beam computed tomography‐guided online adaptive radiotherapy in the pelvic region. Phys Imaging Radiat Oncol. 2021;17:1‐7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. de Jong R, Visser J, van Wieringen N, et al. Feasibility of conebeam CT‐based online adaptive radiotherapy for neoadjuvant treatment of rectal cancer. Radiat Oncol. 2021;16:136. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Moazzezi M, Rose B, Kisling K, et al. Prospects for daily online adaptive radiotherapy via ethos for prostate cancer patients without nodal involvement using unedited CBCT auto‐segmentation. J Appl Clin Med Phys. 2021;22:82‐93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Yock AD, Ahmed M, Ayala‐Peacock D, et al. Initial analysis of the dosimetric benefit and clinical resource cost of CBCT‐based online adaptive radiotherapy for patients with cancers of the cervix or rectum. J Appl Clin Med Phys. 2021;22:210‐221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Åström L, Behrens C, Calmels L, et al. Online adaptive radiotherapy of urinary bladder cancer with full re‐optimization to the anatomy of the day: initial experience and dosimetric benefits. Radiother Oncol. 2022;171:37‐42. [DOI] [PubMed] [Google Scholar]
- 27. Lim‐Reinders S, Keller BM, Al‐Ward S, et al. Online adaptive radiation therapy. Int J Radiat Oncol Biol Phys. 2017;99:994‐1003. [DOI] [PubMed] [Google Scholar]
- 28. Cai B, Laugeman E, Mazur TR, et al. Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a ring shape radiation therapy unit. Med Phys. 2019;46:1355‐1370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Henke LE, Green OL, Price A, et al. Feasibility of AI‐assisted CBCT‐guided stereotactic online adaptive radiotherapy (CT‐STAR) for upper abdominal SBRT: results of a prospective in silico clinical trial. Int J Radiat Oncol Biol Phys. 2020;108:S129‐S129. [Google Scholar]
- 30. Gardner SJ, Mao W, Liu C, et al. Improvements in CBCT image quality using a novel iterative reconstruction algorithm: a clinical evaluation. Adv Radiat Oncol. 2019;4:390‐400. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Li Y, Hoisak JD, Li N, et al. Dosimetric benefit of adaptive re‐planning in pancreatic cancer stereotactic body radiotherapy. Med Dosim. 2015;40:318‐324. [DOI] [PubMed] [Google Scholar]
- 32. Sasaki M, Nakamura M, Ono T, et al. Positional repeatability and variation in internal and external markers during volumetric‐modulated arc therapy under end‐exhalation breath‐hold conditions for pancreatic cancer patients. J Radiat Res. 2020;61:755‐765. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Nakamura M, Nakao M, Mukumoto N, et al. Statistical shape model‐based planning organ‐at‐risk volume: application to pancreatic cancer patients. Phys Med Biol. 2021;66:014001. [DOI] [PubMed] [Google Scholar]
- 34. Ethos Algorithms Reference Guide Publication ID: P1035867‐330‐C. 2019.
- 35. Goto Y, Nakamura A, Ashida R, et al. Clinical evaluation of intensity‐modulated radiotherapy for locally advanced pancreatic cancer. Radiat Oncol. 2018;13:118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Lawson J, Schreibmann E, Jani A, et al. Quantitative evaluation of a cone‐beam computed tomography–planning computed tomography deformable image registration method for adaptive radiation therapy. J Appl Clin Med Phys. 2007;8:96‐113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Miften M, Olch A, Mihailidis D, et al. Tolerance limits and methodologies for IMRT measurement‐based verification QA: recommendations of AAPM Task Group No. 218. Med Phys. 2018;45:e53‐e83. [DOI] [PubMed] [Google Scholar]
- 38. Zhu TC, Stathakis S, Clark JR, et al. Report of AAPM Task Group 219 on independent calculation‐based dose/MU verification for IMRT. Med Phys. 2021;48:e808‐e829. [DOI] [PubMed] [Google Scholar]
- 39. Ziegler M, Nakamura M, Hirashima H, et al. Accumulation of the delivered treatment dose in volumetric modulated arc therapy with breath‐hold for pancreatic cancer patients based on daily cone beam computed tomography images with limited field‐of‐view. Med Phys. 2019;46:2969‐2977. [DOI] [PubMed] [Google Scholar]
- 40. Swamidas J, Kirisits C, De Brabandere M, et al. Image registration, contour propagation and dose accumulation of external beam and brachytherapy in gynecological radiotherapy. Radiother Oncol. 2020;143:1‐11. [DOI] [PubMed] [Google Scholar]
- 41. Akimoto M, Nakamura M, Nakamura A, et al. Inter‐ and intrafractional variation in the 3‐dimensional positions of pancreatic tumors due to respiration under real‐time monitoring. Int J Radiat Oncol Biol Phys. 2017;98:1204‐1211. [DOI] [PubMed] [Google Scholar]
- 42. Nguyen D, Farah J, Barbet N, et al. Commissioning and performance testing of the first prototype of AlignRT InBore™ a Halcyon™ and Ethos™‐dedicated surface guided radiation therapy platform. Phys Med. 2020;80:159‐166. [DOI] [PubMed] [Google Scholar]
- 43. Lens E, van der Horst A, Versteijne E, et al. Considerable pancreatic tumor motion during breath‐holding. Acta Oncol. 2016;55:1360‐1368. [DOI] [PubMed] [Google Scholar]
- 44. Iwai T, Yoshimura M, Ashida R, et al. Hypofractionated intensity‐modulated radiotherapy with concurrent chemotherapy for elderly patients with locally advanced pancreatic carcinoma. Radiat Oncol. 2020;15:264. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. McGinn CJ, Zalupski MM, Shureiqi I, et al. Phase I trial of radiation dose escalation with concurrent weekly full‐dose gemcitabine in patients with advanced pancreatic cancer. J Clin Oncol. 2001;19:4202‐4208. [DOI] [PubMed] [Google Scholar]
- 46. Sibolt P, Booth J, Aland T, et al. Evaluation of calculation‐based patient specific QA for online adaptive radiotherapy. Varian White Papers (Last accessed: 18 Apr 2022). https://medicalaffairs.varian.com/Clinical-Notes-Technical-Notes-and-White-Papers [Google Scholar]
- 47. Chu KY, Eccles CL, Baker T, et al. Oral contrast improves soft tissue matching in image guided radiation therapy for gastrointestinal (GI) tumors. Int J Radiat Oncol Biol Phys. 2016;96:E170‐E171. [Google Scholar]
- 48. Wagner M, Klessen C, Rief M, et al. High‐resolution T2‐weighted abdominal magnetic resonance imaging using respiratory triggering: impact of butylscopolamine on image quality. Acta Radiol. 2008;49:376‐382. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information
Data Availability Statement
Research data are stored in an institutional repository and will be shared upon request to the corresponding author.