Abstract
Purpose
Conventional computed tomography (CT) images are severely affected by metal artifacts in patients with intracranial coils. Monoenergetic images have been suggested to reduce metal artifacts.
The aim of this study was to assess metal artifacts in virtual monoenergetic images (VMIs) reconstructed from spectral brain CT.
Methods
Thirty-two consecutive patients with intracranial coils examined by spectral non contrast brain CT (NCCT) at our center between November 2017 and April 2019 were included. Attenuation and standard deviations were measured in regions of interest (ROIs) at predefined areas in artifact-free and artifact-affected areas. Measurements were performed in conventional polyenergetic images (CIs) and the corresponding data for VMIs were retrieved through spectral diagrams for the each ROI. Subjective analysis was performed by visual grading of CIs and specific VMIs by two neuroradiologists, independently.
Results
In artefact-affected image areas distal from the metal objects, the attenuation values decreased with higher energy level VMIs. The same effect was not seen for artefact-affected image areas close to the metal.
Subjective rating of the artefact severity was significantly better in VMIs at 50 keV for one of the two reviewers compared to the CIs. Overall image quality and tissue differentiation scores were significantly higher for both reviewers in VMIs at 60 and 70 keV compared to CIs.
Conclusion
Our quantitative and qualitative image analysis shown that there is a small significant reduction of intracranial coils artifacts severity by all monoenergetic reconstructions from 50 to 200 keV with preserved or increased overall subjective image quality compared to conventional images.
Abbreviations: CT, computed tomography; DECT, dual energy computed tomography; keV, kiloelectron volt; VMIs, virtual monoenergetic images; MRI, magnetic resonance imaging; DSA, digital subtraction angiography; CIs, conventional images; PACS, Picture Archiving and Communication System; DLP, dose length product; WM, white matter; ROI, region of interest; SD, standard deviation; HU, Hounsfield units; SNR, signal-to-noise ratio; IQR, interquartile range; CSF, cerebrospinal fluid
Keywords: Diagnostic imaging, X-ray computed tomography, Metal artifacts, Monoenergetic imaging, Aneurysm
1. Introduction
Artifacts caused by metal implants are known to impair evaluation of brain computed tomography (CT). The artifacts are, in part, caused by beam hardening i.e. dense materials or tissues that selectively filter out low energy x-ray beams, resulting in distorted attenuation in the surrounding tissues [1], [2], [3]. Metal implants are not uncommon and are used in a variety of treatments for intracranial pathology, such as aneurysm occlusion and symptomatic treatment with deep brain stimulation-electrodes in patients with Parkinson’s disease [4], [5].
Unlike conventional polyenergetic CT, dual energy CT (DECT) enables viewing of virtual images at different kiloelectron volts (keVs). Dual energy imaging can be acquired by using different tube voltages, tubes that rapidly switch voltage or through a dual layer detector. The technique is usually manufacturer-dependent, but an in common benefit is the possibility of reconstructing virtual monoenergetic images (VMIs) at a spectrum of energy levels. Reconstructing VMIs at a high keV could decrease the amount of metal artifacts since the beam hardening effect is reduced, as shown in previous studies [1], [6]. Other potential advantages of DECT over conventional CT include improved tissue differentiation at low keVs, and the possibility of reconstructing virtual non contrast (VNC) images [7], [8], [9], [10], [11].
Patients with intracranial metal implants are often followed up with CT imaging after receiving the implant but may also be referred for other reasons. Reduced metal artifacts on CT images is therefore a major clinical need.
The aim of this study was to assess metal artifacts in virtual monoenergetic images (VMIs) reconstructed from spectral brain CT.
2. Material and methods
2.1. Study population
We performed a retrospective search in our Picture Archiving and Communication System (PACS) for patients who had undergone non-contrast brain CT in the IQon Spectral CT (Philips Healthcare Inc., The Netherlands) at our institute between November 21, 2017 and April 4, 2019. All patients aged 18 or more with intracranial coils were included. If a patient had multiple exams that met our inclusion criteria, the one performed closest to the time of the endovascular intervention was chosen.
Examinations lacking spectral images in PACS were excluded.
The study was approved by the Swedish Ethical Review Authority (reference number 2019–02225) and individual informed consent was waived.
2.2. CT parameters
If a patient had been treated with coils in more than one location, the largest skein of coils was used for measurements, and the measurements were done in images with artifacts from that skein of coils only.
Dose length product (DLP) was noted and the effective dose was calculated as DLP[mGycm]*0.0024[mSv/mGycm] [12].
Measurements to calculate the coil volume were obtained from 3D DSA images and calculated as 4/3, using the largest diameter to calculate the radius [13].
All images were assessed in the axial plane of 4 mm slice thickness images, in line with local routine.
2.3. Quantitative image analysis
Quantitative image analysis was based on measurements in predefined regions of interest (ROIs) manually drawn at specific locations in the CT images with conventional reconstruction (CI);.
-
1)
artifact-free white matter (WM) between vertex and the lateral ventricle in either hemisphere;
-
2)
artifact-free cerebrospinal fluid (CSF) in the left lateral ventricle;
-
3)
artifact-affected distal white matter (≥2 cm from the coils);
-
4)
artifact-affected proximal white matter (<2 cm from the coils).
All ROIs were of circular shape and drawn to a diameter of 10 ± 1 mm. ROI size was only adjusted when necessary to avoid adjacent tissue, for example in slim lateral ventricles. All measurements were performed in ISP (IntelliSpace Portal, version 10.1.4.21403, Philips healthcare Inc., The Netherlands).
The mean attenuation and the standard deviation (SD) measured in Hounsfield units (HU) were noted for each ROI in the CIs and, for the same ROIs, retrieved through spectral diagrams for monoenergetic reconstructions ranging from 40 to 200 keV with 10 keV increments.
2.4. Qualitative analysis
Qualitative analysis was performed by visual grading of the conventional image series and monoenergetic reconstructions at 40, 50, 60, 70, 80, 120, 140 and 200 keV, by two independent reviewers, both senior interventional neuroradiologists with over 10 years’ experience (BR and JW). Chosen keV levels that were included in this part of the analysis were based on the attenuation values from the quantitative data that indicated large changes mainly between the lower keV reconstructions, and from previous work [6], [7], [9]. Prior to the visual grading, example cases were used in order to establish the grading scale to the reviewers.
All images used for the qualitative analysis were de-identified and presented to the reviewers in random order without any patient- or reconstruction information using the Viewdex software version 2.272 (Viewer for Digital Evaluation of X-ray images) [14], [15]. The images were displayed on an image workstation with dedicated monitors for diagnostic radiology (Coronis® Fusion MDCC-6430 6MP, Barco, Kortrijk, Belgium). The contrast/brightness levels could be adjusted by the reviewer. All images were assessed on a 5-point Likert scale (1 =Non diagnostic/massive artifacts, 2 = Poor/severe artifacts 3 =Fair/moderate artifacts, 4 = Good/minor artifacts and 5 = Excellent/no artifacts) in coherence with previous work [7], [16], [17] regarding: 1) overall image quality, 2) presence of metal artifacts and 3) grey-white matter differentiation in artifact-free tissue.
Statistical analysis All analyses were performed using IBM SPSS Statistics version 25 (IBM Corp, Armonk, New York, USA). Descriptive data are presented as means and/or medians. A Wilcoxon signed-rank test was used for pairwise comparison of attenuation between specific VMIs. The level of statistical significance were predefined to p ≤ 0.05.
Interobserver agreement was calculated as a weighted Cohens kappa coefficient using MedCalc Statistical Software version 20.009 (MedCalc Software, Ostend, Belgium) and Wilcoxon signed-rank test was used for comparison of scores between different reconstructions (Fig. 1).
3. Results
3.1. Patient population and coil parameters
The final study population was 32 patients, 19 women and 13 men. The mean age was 58 years. The most common coiling location was anterior communicating artery aneurysms, in 16 of the patients. The remaining coiling locations were the internal carotid artery, basilar artery, posterior inferior cerebellar artery, pericallosal artery, and vertebral artery. Baseline and coil parameters are shown in Table 1.
Table 1.
Patient characteristics | |
---|---|
Number of patients | 32 |
Men/women | 13/19 |
Mean age (range) [years] | 58 (25–84) |
Subarachnoid hemorrhage | 28 (88%) |
CT scanning parameters | |
Tube voltage [kV] | 120 |
Collimation [mm] | 64 × 0.625 |
Rotation time [s] | 0.33 |
Pitch | 0.359 |
Median CTDIvol (IQR) [mGy] | 40.3 (37–43) |
Median DLP (IQR) [mGycm] | 833 (744–891) |
Median effective dose (IQR) [mSv] | 2.0 (1.8–2.1) |
Coil parameters | |
Anterior communicating artery | 16 (50%) |
Internal carotid artery | 9 (28%) |
Basilar artery | 2 (6%) |
Other locations | 5 (16%) |
Median coil volume (IQR) [mm3] | 93 (34–296) |
n = number, mSv = milliSievert, IQR = Inter Quartile Range |
3.2. Quantitative image analysis
The attenuation values generally decreased with higher keV in ROIs 1–3. Box plots of HU values of artifact-free white matter, CSF and artifact-affected tissue distal and proximal to the metal implant are shown in Fig. 2.
Pairwise comparison of HU values was performed between CI and VMIs at 50, 60 and 70 keV, based on the results of the qualitative analysis. The tests showed significantly higher attenuation values in VMIs at 50 and 60 keV and non-significantly higher in VMIs at 70 keV reconstructions compared to CI for ROI 1 (artifact-free white matter).
For ROI 2 (CSF) the attenuation was significantly higher in 50 and 60 keV VMIs and significantly lower in 70 keV reconstructions compared to CIs. For ROI 3 (distal artifact) there were significant differences between CI and VMIs at 50 and 60 keV, with higher values in the VMIs, but no significant difference between CIs and 70 keV was found. No significant differences between CIs or the VMIs listed above were found for ROI 4 (proximal artifact).
3.3. Qualitative image analysis
The results of overall image quality, metal artifact severity and grey-white matter differentiation scores for both reviewers are shown in Fig. 3. Based on visual assessment of the graphs, significance was tested between CIs and VMIs at 50, 60 and 70 keV.
Typical image examples for CIs and VMIs at 50, 60, 70 and 120 keV are shown in Fig. 4.
3.4. Overall image quality
The median image quality score peaked at 4.0 for both reviewers (in 50 and 60 keV reconstructions for Reviewer 1, and in 60 keV for Reviewer 2). The median score for CIs was 3.0 for both reviewers. The scores for overall image quality were significantly higher for VMIs at 50, 60 and 70 keV for Reviewer 1, and for VMIs at 60 and 70 keV for Reviewer 2 compared to CIs. The inter-observer agreement was fair (Cohens weighted kappa coefficient = 0.3).
3.5. Artifact Severity
The median artifact severity score was 3.0 for all VMIs except for the 40 keV reconstructions (scores of 2.5, and 2 for Reviewers 1 and 2, respectively). The median score for CIs was 2.5 for Reviewer 1 and 3.0 for Reviewer 2. The scores were significantly higher in VMIs at 50 keV than for CIs for Reviewer 1, but not for any of the VMIs compared to CI for Reviewer 2. The interobserver agreement was moderate (Cohens weighted kappa coefficient = 0.6).
3.6. Grey-white matter differentiation
The median grey-white matter differentiation score peaked at 4.5 in VMIs at 50 keV for Reviewer 1, and at 4.0 in VMIs at 50, 60 and 70 keV for Reviewer 2. The scores for VMIs at 50, 60 and 70 keV were all significantly higher than for the CIs for both reviewers.
The interobserver agreement was moderate (Cohens weighted kappa coefficient = 0.5).
4. Discussion
The aim of this study was to assess metal artifacts in virtual monoenergetic images (VMIs) reconstructed from spectral brain CT.
In our quantitative image analysis, we found that in artifact-affected image areas ≥ 2 cm distal from the coils, the attenuation values decreased with higher energy level VMIs. This is consistent with the quantitative image analysis that show improved overall image quality, improved grey-white matter differentiation and reduced metal artifact severity in VMIs at lower energy levels compared to CIs.
The same benefit of VMIs is not seen in artifact-affected image areas within 2 cm of the coils. A likely explanation is that photon starvation is the main cause of the proximal artifacts, whereas the beam hardening effect, that the dual energy CT technique mainly reduces, is more important further distal to the coils.
In a phantom study, Winklhofer et al. [6] evaluated artifacts produced by clips and coils of various volumes and densities in VMIs at 120 keV in combination with the iMAR metal artifact reduction algorithm and found that VMIs in combination with the iterative algorithm resulted in the least severe artifact scores, but this improvement was small for large and densely packed coils which could possibly be due to the predominance of photon starvation adjacent to dense metal. Große Hokamp et al. [18] showed that the metal artifacts from deep brain electrodes was reduced by using VMIs in combination with the O-MAR iterative metal artifact reduction algorithm. The results of these studies regarding VMIs are in line with ours.
The effect of metal artifact reduction achieved by dedicated post-processing algorithms [6], [18] seem to be more prominent than the effect of VMIs, but this may differ for metal artifacts generated by many small objects, such as the Tantalum in ONYX® or Squid®-embolization used for dural arteriovenous fistulas (DAVF) or chronic subdural hematomas (CSDH) instead of a single large metal object. Therefore, the use of VMIs as a complimentary measure to reduce metal artifacts may be of greater importance in such cases.
Other studies have shown significant metal artifact reduction in VMIs for various types of orthopedic implants [19], [20], [21]. Although not directly comparable, the effect on VMIs on intracranial metal artifacts seem to be less pronounced for intracranial coils than for orthopedic implants, this may be due to the surrounding cranium that possibly absorbs the higher energy beams to a great extent, or by the different physical properties of metal implants. For example, platinum used in neurovascular coils has an approximately five-fold density compared to titanium used in orthopedic implants [22], [23], which could affect the severity of metal artifacts.
The artifacts produced by coils are often heterogenic with alternating hypo- and hyperdense streaks, due to irregularities in aneurysm shape and coil packing density within the aneurysm [24]. In quantitative measurements, the hypo- and hyperattenuating streaks could possibly cancel each other out, and thus it is important to combine quantitative and qualitative evaluation of artifact severity. In our study, the quantitative and qualitative image evaluations are consistent and in support of each other.
A benefit of metal artifact reduction by VMIs is that it does not increase the patient’s radiation dose, which would typically be necessary for adjustments of scanning parameters to decrease metal artifacts, such as increasing the x-ray tube voltage. In addition, VMI reconstructions is rapid and easily accessible in the clinical day-to-day routine post-processing work, for examples if the presence of coils was not known beforehand.
Since we find that the reduction of metal artifact severity is relatively constant across all VMIs compared to CI we may also conclude that any other benefits of VMIs, such as increased diagnostic ability for early ischemia [25] may be utilized without aggravating any metal artifacts.
4.1. Limitations
This study has several limitations, the first being a relatively small study population, however many of the results was still shown to be statistically significant. The second limitation is that all data collection was done on the same CT scanner and all image processing was done on only one vendor-specific software. In future studies it would be valuable to include multiple scanners capable of generating a spectrum of VMIs from 40 to 200 keV. The third limitation is that we only included cases where the metal artifacts were caused by neurovascular coils. In future studies it would be valuable to evaluate artifacts from other intracranial implants such as clips and electrodes as well as artifacts from multiple locations such as Tantalum in CSDH embolization.
5. Conclusions
Our quantitative and qualitative image analysis shown that there is a small significant reduction of intracranial coils artifacts severity by monoenergetic reconstructions with preserved or increased overall subjective image quality compared to conventional images.
None of the monoenergetic reconstructions from 50 to 200 keV increased the severity of intracranial coils artifacts compared to conventional images.
Further studies, should also address the diagnostic ability with VMIs with and without iterative metal artifact reduction algorithms in assessing brain CT with metal artifacts, such as the ability to assess subarachnoid hemorrhage or ischemia.
Ethics approval details
This study was approved and individual informed consent was waived by the Swedish Ethical Review Authority (reference number 2019–02225).
Fundings
This study was funded by regional ALF grants to HM, TU and JW, the Crafoord Foundation to JW and by SUS Stiftelser & Fonder to JW. None of the funders had any involvement in the planning of methodology, data retrieval or analysis in the study.
CRediT authorship contribution statement
HM, VF, KY, TU, JW and BR conceptualized and planned the study. HM and VF collected data. HM, VF, JL and TU processed and analyzed data. HM wrote the first manuscript draft. All authors revised the manuscript. All authors have read and approved the final version of the manuscript.
Conflicts of interest
None of the authors have any conflicts of interests related to the study. JW is a founder and shareholder of Uman Sense AB and has received speaker honoraria from Siemens Healthineers, BALT group and Medtronic Inc.
Acknowledgements
The authors wish to thank statistician Dr Sara Jesperson at Lund University for professional statistical services, Dr Lee Nolan is for professional language editing and all colleagues and patients at the department of medical imaging and physiology at Skåne University Hospital/Lund.
Data Availability
An anonymized dataset supporting the conclusions of this article may be upon reasonable request including the appropriate ethics approval.
References
- 1.Laukamp K.R., Gupta A., Große Hokamp N., Obmann V.C., Graner F.P., Ho V., et al. Role of spectral-detector CT in reduction of artifacts from contrast media in axillary and subclavian veins: single institution study in 50 patients. Acta Radiol. 2019 doi: 10.1177/0284185119868904. [DOI] [PubMed] [Google Scholar]
- 2.Boas F., Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012:4. doi: 10.2217/iim.12.13. [DOI] [Google Scholar]
- 3.Barrett J.F., Keat N. Artifacts in CT: recognition and avoidance. RadioGraphics. 2004;24(6):1679–1691. doi: 10.1148/rg.246045065. [DOI] [PubMed] [Google Scholar]
- 4.Ellis J.A., Nossek E., Kronenburg A., Langer D.J., Ortiz R.A. Intracranial aneurysm: diagnostic monitoring, current interventional practices, and advances. Curr. Treat. Options Cardiovasc. Med. 2018;20(12):94. doi: 10.1007/s11936-018-0695-y. [DOI] [PubMed] [Google Scholar]
- 5.Lee D.J., Dallapiazza R.F., De Vloo P., Lozano A.M. Current surgical treatments for Parkinson's disease and potential therapeutic targets. Neural Regen. Res. 2018;13(8):1342–1345. doi: 10.4103/1673-5374.235220. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Winklhofer S., Hinzpeter R., Stocker D., Baltsavias G., Michels L., Burkhardt J.-K., et al. Combining monoenergetic extrapolations from dual-energy CT with iterative reconstructions: reduction of coil and clip artifacts from intracranial aneurysm therapy. Neuroradiology. 2018;60(3):281–291. doi: 10.1007/s00234-018-1981-9. [DOI] [PubMed] [Google Scholar]
- 7.Neuhaus V., Abdullayev N., Große Hokamp N., Pahn G., Kabbasch C., Mpotsaris A., et al. Improvement of image quality in unenhanced dual-layer CT of the head using virtual monoenergetic images compared with polyenergetic single-energy CT. Investig. Radiol. 2017;52(8):470–476. doi: 10.1097/rli.0000000000000367. [DOI] [PubMed] [Google Scholar]
- 8.Pan Y.-N., Chen G., Li A.-J., Chen Z.-Q., Gao X., Huang Y., et al. Reduction of metallic artifacts of the post-treatment intracranial aneurysms: effects of single energy metal artifact reduction algorithm. Clin. Neuroradiol. 2019;29(2):277–284. doi: 10.1007/s00062-017-0644-2. [DOI] [PubMed] [Google Scholar]
- 9.Zhao X.-M., Wang M., Wu R.-Z., Dharaiya E., Feng F., Li M.-L., et al. Dual-layer spectral detector CT monoenergetic reconstruction improves image quality of non-contrast cerebral CT as compared with conventional single energy CT. Eur. J. Radiol. 2018;103:131–138. doi: 10.1016/j.ejrad.2018.04.015. [DOI] [PubMed] [Google Scholar]
- 10.Sauter A.P., Muenzel D., Dangelmaier J., Braren R., Pfeiffer F., Rummeny E.J., et al. Dual-layer spectral computed tomography: virtual non-contrast in comparison to true non-contrast images. Eur. J. Radiol. 2018;104:108–114. doi: 10.1016/j.ejrad.2018.05.007. [DOI] [PubMed] [Google Scholar]
- 11.Riederer I., Fingerle A.A., Baum T., Kirschke J.S., Rummeny E.J., Noël P.B., et al. Acute infarction after mechanical thrombectomy is better delineable in virtual non-contrast compared to conventional images using a dual-layer spectral CT. Sci. Rep. 2018;8(1):9329. doi: 10.1038/s41598-018-27437-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Huda W., Magill D., He W. CT effective dose per dose length product using ICRP 103 weighting factors. Med. Phys. 2011;38(3):1261–1265. doi: 10.1118/1.3544350. [DOI] [PubMed] [Google Scholar]
- 13.Goddard J.K., Moran C.J., Cross D.T., 3rd, Derdeyn C.P. Absent relationship between the coil-embolization ratio in small aneurysms treated with a single detachable coil and outcomes. AJNR Am. J. Neuroradiol. 2005;26(8):1916–1920. [PMC free article] [PubMed] [Google Scholar]
- 14.Börjesson S., Håkansson M., Båth M., Kheddache S., Svensson S., Tingberg A., et al. A software tool for increased efficiency in observer performance studies in radiology. Radiat. Prot. Dosim. 2005;114(1–3):45–52. doi: 10.1093/rpd/nch550. [DOI] [PubMed] [Google Scholar]
- 15.Håkansson M., Svensson S., Zachrisson S., Svalkvist A., Båth M., Månsson L.G. ViewDEX: an efficient and easy-to-use software for observer performance studies. Radiat. Prot. Dosim. 2010;139(1–3):42–51. doi: 10.1093/rpd/ncq057. [DOI] [PubMed] [Google Scholar]
- 16.Guggenberger R., Winklhofer S., Osterhoff G., Wanner G.A., Fortunati M., Andreisek G., et al. Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur. Radiol. 2012;22(11):2357–2364. doi: 10.1007/s00330-012-2501-7. [DOI] [PubMed] [Google Scholar]
- 17.Bamberg F., Dierks A., Nikolaou K., Reiser M.F., Becker C.R., Johnson T.R.C. Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur. Radiol. 2011;21(7):1424–1429. doi: 10.1007/s00330-011-2062-1. [DOI] [PubMed] [Google Scholar]
- 18.Große Hokamp N., Hellerbach A., Gierich A., Jordan D.W., Visser-Vandewalle V., Maintz D., et al. Reduction of artifacts caused by deep brain stimulating electrodes in cranial computed tomography imaging by means of virtual monoenergetic images, metal artifact reduction algorithms, and their combination. Invest. Radiol. 2018;53(7):424–431. doi: 10.1097/rli.0000000000000460. [DOI] [PubMed] [Google Scholar]
- 19.Wang Y., Qian B., Li B., Qin G., Zhou Z., Qiu Y., et al. Metal artifacts reduction using monochromatic images from spectral CT: Evaluation of pedicle screws in patients with scoliosis. Eur. J. Radiol. 2013;82(8):e360–e366. doi: 10.1016/j.ejrad.2013.02.024. [DOI] [PubMed] [Google Scholar]
- 20.Horat L., Hamie M.Q., Huber F.A., Guggenberger R. Optimization of monoenergetic extrapolations in dual-energy ct for metal artifact reduction in different body regions and orthopedic implants. Acad. Radiol. 2019;26(5):e67–e74. doi: 10.1016/j.acra.2018.06.008. [DOI] [PubMed] [Google Scholar]
- 21.Lee K.Y.G., Cheng H.M.J., Chu C.Y., Tam C.W.A., Kan W.K. Metal artifact reduction by monoenergetic extrapolation of dual-energy CT in patients with metallic implants. J. Orthop. Surg. 2019;27(2) doi: 10.1177/2309499019851176. [DOI] [PubMed] [Google Scholar]
- 22.National Center for Biotechnology Information. PubChem Element Summary for AtomicNumber 78, Platinum https://pubchem.ncbi.nlm.nih.gov/element/Platinum: PubChem [Internet]; n.d. Accessed on 4 October 2021 [Available from: https://pubchem.ncbi.nlm.nih.gov/element/Platinum.
- 23.National Center for Biotechnology Information. PubChem Element Summary for AtomicNumber 22, Titanium https://pubchem.ncbi.nlm.nih.gov/element/22: PubChem [Internet]; n.d. Accessed on 4 October 2021. Available from: https://pubchem.ncbi.nlm.nih.gov/element/22.
- 24.Sluzewski M., van Rooij W.J., Slob M.J., Bescós J.O., Slump C.H., Wijnalda D. Relation between aneurysm volume, packing, and compaction in 145 cerebral aneurysms treated with coils. Radiology. 2004;231(3):653–658. doi: 10.1148/radiol.2313030460. [DOI] [PubMed] [Google Scholar]
- 25.Mellander H., Bengtsson P., Fransson V., Ramgren B., Undrén P., Drake M., Ydström K., Lätt J., Hilal A., Wassélius J., Ullberg T. Virtual monoenergetic images by spectral detector computed tomography may improve image quality and diagnostic ability for ischemic lesions in acute ischemic stroke. Acta Radio. 2022 doi: 10.1177/02841851221130612. 2841851221130612. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
An anonymized dataset supporting the conclusions of this article may be upon reasonable request including the appropriate ethics approval.