. 2022 Nov 8;3(1):73–81. doi: 10.1021/acsmeasuresciau.2c00055

Enhancing 19F Benchtop NMR Spectroscopy by Combining para-Hydrogen Hyperpolarization and Multiplet Refocusing

Ana I Silva Terra , Matheus Rossetto , Claire L Dickson , George Peat , Dušan Uhrín , Meghan E Halse †,*
PMCID: PMC9936801  PMID: 36817010

Abstract

graphic file with name tg2c00055_0008.jpg

Benchtop NMR spectrometers provide a promising alternative to high-field NMR for applications that are limited by instrument size and/or cost. 19F benchtop NMR is attractive due to the larger chemical shift range of 19F relative to 1H and the lack of background signal in most applications. However, practical applications of benchtop 19F NMR are limited by its low sensitivity due to the relatively weak field strengths of benchtop NMR spectrometers. Here we present a sensitivity-enhancement strategy that combines SABRE (Signal Amplification By Reversible Exchange) hyperpolarization with the multiplet refocusing method SHARPER (Sensitive, Homogeneous, And Resolved PEaks in Real time). When applied to a range of fluoropyridines, SABRE-SHARPER achieves overall signal enhancements of up to 5700-fold through the combined effects of hyperpolarization and line-narrowing. This approach can be generalized to the analysis of mixtures through the use of a selective variant of the SHARPER sequence, selSHARPER. The ability of SABRE-selSHARPER to simultaneously boost sensitivity and discriminate between two components of a mixture is demonstrated, where selectivity is achieved through a combination of selective excitation and the choice of polarization transfer field during the SABRE step.

Keywords: benchtop NMR spectroscopy, SABRE hyperpolarization, NMR sensitivity enhancement, SHARPER, 19F NMR

Introduction

In recent years, benchtop NMR spectrometers with moderate fields of 1–2 T and line widths of <0.5 Hz, have emerged as a promising alternative to traditional high-field NMR spectrometers for applications where instrument size and/or cost are a limiting factor and for use in nontraditional environments, such as within a fume cupboard in a synthetic lab.15 While benchtop NMR spectrometers benefit from increased portability and lower purchase and maintenance costs, they suffer from reduced sensitivity and chemical shift dispersion due to their lower magnetic field strengths.6 The resolution challenge of benchtop NMR is particularly acute for protons because of their narrow chemical shift range, which leads to signal overlap and strong coupling effects.

Fluorine NMR is an attractive alternative as it has a wider chemical shift range than 1H and is the most sensitive of all spin-1/2 nuclei other than 1H, due to its 100% natural abundance and high gyromagnetic ratio. Furthermore, 19F NMR is of particular analytical interest due, for example, to the high prevalence of fluorine in pharmaceuticals,79 the use of fluorinated tags in chemical biology,10 and reaction monitoring applications.1119F is a convenient target in benchtop NMR because fluorine measurements can usually be performed using the proton channel due to the proximity of the 19F and 1H Larmor frequencies at 1–2 T. Despite the sensitivity challenges, benchtop 19F NMR has been investigated for practical applications such as monitoring the degradation of persistent fluorinated organic pollutants.12

One strategy to overcome the sensitivity challenge of benchtop NMR is to use hyperpolarization, which increases the population difference between nuclear spin states and, as a result, the detected signal. para-Hydrogen-based hyperpolarization methods, such as the Signal Amplification By Reversible Exchange (SABRE)13 method, are particularly attractive for use with compact NMR devices because they do not require large or expensive instrumentation and the level of polarization generated is independent of the detection field strength.14 This has led to the combination of SABRE with benchtop NMR detection by many groups to deliver enhancements of several orders of magnitude for a range of nuclei, including 1H, 13C, 15N, and 19F.1527

In SABRE, the high spin order of the nuclear singlet isomer of hydrogen, para-hydrogen (p-H2), is harnessed to create highly polarized states in a target molecule through reversible binding to an iridium catalyst (Figure 1A).13 Reversible binding of both p-H2 and the target analyte (the substrate) to the complex establishes a temporary J coupling network through which spontaneous transfer of the polarization can occur. This transfer is facilitated by a weak magnetic field, referred to as the polarization transfer field (PTF). The optimal PTF varies according to the specific spin system and is typically in the range of 0 < PTF ≤ 10 mT.28,29 In the presence of the relevant PTF and under a continual supply of p-H2, substrate hyperpolarization is built up in free solution over a period of seconds due to the chemical exchange of both the substrate and p-H2.30 Enhanced NMR spectra are observed following introduction of the sample into the NMR spectrometer for detection (Figure 1B). The chemical identity of the substrate is not altered in this process and so it can be rehyperpolarised by adding fresh p-H2.13

Figure 1.

Figure 1

(A) Diagram showing the SABRE hyperpolarization process followed by measurement using benchtop NMR. para-Hydrogen is added to the headspace of an NMR tube containing the SABRE catalyst and target substrate. It is then shaken in the desired PTF to allow for a buildup of hyperpolarized substrate in free solution. Finally, the sample is introduced into a benchtop NMR spectrometer for detection. B) Comparison of a nonhyperpolarised (top) and a SABRE-enhanced (bottom) 1H NMR spectrum of 3,5-difluoropyridine (1) zoomed in between 6.5 and 9 ppm. SABRE hyperpolarization (PTF = 6 mT) provides a 1200-fold improvement in the single-scan signal-to-noise ratio (SNR).

Another approach to improving the sensitivity of 19F NMR is the multiplet refocusing method SHARPER (Sensitive, Homogeneous, And Resolved PEaks in Real time).31 SHARPER increases signal-to-noise ratios (SNR) by eliminating peak-splitting through the use of FID acquisition periods interleaved with a series of 180° refocusing pulses. Importantly, this is achieved by applying radio frequency (RF) pulses only at the frequency of the acquired nucleus. An additional benefit of the SHARPER approach is that it minimizes the effect of field inhomogeneity, producing narrow peaks that approach their natural line width. We recently introduced adapted versions of SHARPER and the chemically selective variant, selSHARPER, for use on benchtop NMR spectrometers that remove the need for pulsed field gradients (PFG), which are not available on many benchtop instruments.32 When combined with optimized data processing strategies, including a matched filter, 19F SNR improvements by more than an order of magnitude were achieved.

The goal of this work is to optimize the sensitivity of 19F benchtop NMR spectroscopy through the combination of SABRE and SHARPER. To explore the efficacy of this approach, we first optimize the 19F SABRE enhancement of three fluoropyridines using benchtop NMR detection and a range of PTFs. The additional SNR gains that can be achieved by combining SHARPER with SABRE are demonstrated and compared to a standard 1H decoupling approach. Finally, a modified version of the selSHARPER experiment is applied to a mixture of two analytes to establish the ability of SABRE-SHARPER to deliver both SNR enhancement and resonance discrimination.

Experimental Section

Single component SABRE samples contained 100 mM of the substrate (1, 2, or 3 in Scheme 1) and 5 mM of precatalyst [IrCl(COD)(IMes)] (where COD: 1,5-cyclooctadiene and IMes: 1,3-bis(2,4,6-trimethyl-phenyl)-imidazolium) in 0.7 mL of HPLC grade methanol. The exact masses used in the preparation are provided in the SI (Table S1). All samples were sonicated to aid dissolution and then introduced into a 5 mm NMR tube with a J-Young valve and degassed with 3 cycles of freeze/pump/thaw using a Schlenk line and an acetone and dry ice bath. The degassing step is included to improve the dissolution of p-H2 and to remove dissolved oxygen. The headspace of the NMR tubes were then filled with 4 bar of p-H2, which was generated by cooling H2 gas to 28 K to produce >99% p-H2 using a generator described previously.14 For each SABRE acquisition, the sample was shaken for 10 s inside the desired polarization transfer field (PTF) and then was introduced into the spectrometer for data acquisition. Typical sample transfer times were on the order of 2 s. The desired PTF was achieved by shaking the sample (a) inside a hand-held magnetic array with an average field PTF = 6.2 mT,33 (b) in the ambient Earth’s magnetic field adjacent to the spectrometer (PTF ∼ 50 μT), or (c) inside a μ-metal shield that reduces the ambient field by a factor of ∼300-fold to achieve a PTF ∼ 0.2 μT. Between each SABRE measurement, the p-H2 in the sample was refreshed by evacuating the headspace of the NMR tube and refilling with fresh p-H2. For each fresh sample, the filling-shaking-acquiring process was repeated four to six times to activate the SABRE catalyst. The 1 and 2 SABRE mixture samples were prepared using the same procedure with 50 mM of each compound, respectively (Table S1). Each SABRE measurement was repeated three times. The SABRE SNR values were calculated as the average of the three repetitions and are reported with the corresponding standard error.

Scheme 1. 3,5-Difluoropyridine (1), 3-(Difluoromethyl)pyridine (2), and 3,4,5-Trifluoropyridine (3).

Scheme 1

Benchtop NMR spectra were acquired using a 43 MHz (1 T) NMR spectrometer (Spinsolve Carbon, Magritek, Aachen, Germany) equipped with 1H/19F and 13C channels. Shimming and frequency calibrations were performed on a reference sample containing a H2O:D2O mixture (10%:90%) before data was collected. Data was acquired using the Spinsolve Expert software (version 1.41) and processed either by Prospa (version 3.63), MestReNova (version 14.1.2.25024) or Matlab (version R2020a).

Unless otherwise stated, all of the SHARPER and selSHARPER spectra were acquired in a single scan using an acquisition chunk time of τ = 3.2 ms within each echo loop and a 180° pulse duration of 225 μs. Complete sets of pulse sequence parameters for all experiments are provided in the SI (Section S2).

Free induction decay data were apodized with a matched exponential decay filter, zero-filled and Fourier transformed to produce the spectra, which were manually phased and baseline corrected as needed. For non-SHARPER acquisitions, a filter was chosen that optimized SNR without significantly compromising spectral resolution. For full details see the SI (Section S2). For all SHARPER and selSHARPER spectra, the imaginary signal channel was zeroed prior to Fourier Transformation to improve SNR, as described previously.32

Signal-to-noise values were calculated as the ratio of the signal height of the tallest peak and the standard deviation of a region of spectral noise. When reference spectra were obtained by averaging more than one scan, the inverse of the square root of the number of scans was used as the correction factor to obtain a signal-to-noise ratio per one scan. The SNR enhancement factors (εSNR) were calculated as the ratio of the SNR of the hyperpolarised spectrum and the SNR per one scan of the reference spectrum.

Results and discussion

A range of fluoropyridines, including 3,5-difluoropyridine (1), have been shown previously to provide efficient 1H and 19F SABRE hyperpolarization when combined with high-field NMR detection.17,34,35 Therefore, the three fluorinated pyridines selected for this work, 3,5-difluoropyridine (1), 3-(difluoromethyl)pyridine (2), and 3,4,5-trifluoropyridine (3) (Scheme 1), are expected to deliver good levels of SABRE enhancement. In addition, the chosen substrates provide three different 19F coupling environments to test the efficacy of the SABRE-SHARPER approach. Specifically, 1 has a single 19F resonance with a medium strength heteronuclear coupling constant of 3JHF = 9 Hz, 2 has a single 19F resonance with a larger heteronulcear coupling constant of 2JHF = 55 Hz, and 3 contains two 19F resonances with both homonuclear and heteronuclear J couplings.

SABRE-enhanced spectra of 1, 2, and 3 were recorded using PTF = 6.2 mT and benchtop (1 T) NMR detection (Figure S1). 1H SNR enhancements of 1200-fold, 600-fold, and 1400-fold were observed for 1, 2, and 3, respectively, indicating efficient SABRE activity for all three species. To optimize polarization transfer to 19F, three different PTF regimes were tested for each compound: PTF = 6.2 mT, PTF = 50 μT (the ambient Earth’s magnetic field outside the spectrometer) and PTF ∼ 0.2 μT (field within a μ-metal shield). Optimized SABRE-enhanced 19F benchtop NMR spectra for 1, 2, and 3 are presented in Figure 2, with thermally polarized NMR spectra included for reference.

Figure 2.

Figure 2

Comparison of SABRE-enhanced and thermally polarized 19F benchtop NMR spectra of 100 mM of (A) 1, (B) 2, and (C) 3 and 5 mM precatalyst in methanol. All SABRE spectra were acquired with a single scan. The vertical scale of the thermal spectra was increased to aid visualization. SABRE SNR values are reported as the average over three repeat measurements. Fluorine SNR enhancement factors (εSNR) are calculated as the ratio of the average SNR of the SABRE spectra and the SNR per one scan of the reference spectrum.

Increases in 19F SNR by factors of 210 ± 10, 240 ± 20, and 360 ± 40 were observed for 1, 2, and 3, respectively (Figure 2). Interestingly, while the optimal PTF for 1 and 3 was the Earth’s magnetic field (EF ∼ 50 μT), PTF = 6.2 mT was found to be most efficient for 2. The level anticrossing (LAC) theory of efficient SABRE polarization transfer suggests that the optimal PTF for 19F lies in the μT regime, while direct transfer to 1H is optimized in mT fields.28,29,36 This framework is consistent with the findings of Chukanov et al.,34 who studied the SABRE polarization transfer mechanisms for 3-fluoropyridine and observed more efficient transfer to 19F in microtesla fields, such as the Earth’s magnetic field, compared to transfer in mT fields. In the case of 2 we observe much higher enhancements in the mT regime than in the Earth’s field. This suggests that 19F is polarized indirectly, with polarization being transferred first to 1H nuclei on the substrate and then spin-relayed to 19F, with the transfer within the substrate being mediated by the large heteronuclear coupling constant, 2JHF = 55 Hz.

Inspection of the SABRE-enhanced 19F NMR spectra in Figure 2 suggests that significant further 19F SNR enhancements are possible by removing inhomogeneous peak broadening and collapsing the multiplets into a single resonance using the SHARPER approach. Figure 3a presents the SHARPER pulse sequence developed for use on benchtop NMR spectrometers that does not rely on the use of pulsed field gradients within the sampling loop.32 In the first step of the experiment, a broadband 90° pulse is applied and a first half chunk of the free induction decay (FID) is acquired. This is followed by a CPMG-style loop that includes additional short acquisition chunks interleaved with 180° refocusing pulses. The complete FID is assembled by combining the acquisition chunks acquired within each spin–echo period. In this way, for on-resonance spins, all heteronuclear J couplings and inhomogeneous broadening are refocused, leading to a single narrow peak.

Figure 3.

Figure 3

Pulse sequence diagrams for (A) SHARPER and (B) selSHARPER with nonselective pulses in the pulse train. The filled and empty rectangles represent 90° and 180° hard pulses, while the smoothed empty shape depicts a selective Gaussian 180° pulse. The chunk time is defined as τ = Nδt, where N is the number of points per chunk and δt is the dwell time. The total acquisition time is tacq = (n + 1/2)τ, where n is the total number of loops, (A) SHARPER: φ1 = x; φ2 = y; ψ = x. (B) selSHARPER φ1 = x; φ2 = y; φ3 = −y; ψ = x.

A consequence of the piece-wise collection of the FID is the appearance of sideband artifacts at multiples of the inverse of the chunk time, τ = Nδt, where N is the number of points per chunk and δt is the dwell time. These artifacts are minimized for short chunk times, which do not allow for significant chemical shift or J coupling evolution. However, very short sampling intervals limit the overall SNR gains that can be achieved because the time spent applying the refocusing pulses increases as a proportion of the total acquisition time. Therefore a balance is required to achieve clean spectra with high SNR.31,32

Usually, optimal performance with minimal artifacts is achieved in multiple scans through the use of phase cycling. However, single scan methods are preferable for combining with SABRE because of the transient nature of the hyperpolarization. While we exclusively use single scans for the SABRE-SHARPER experiments presented here, automated methods for moving the sample between the polarization transfer and detection fields, via flow or sample shuttling for example, have been shown to enable multistep experiments, including SABRE-enhanced 2D benchtop NMR spectroscopy.15,16,37,38 These multidimensional experiments could be combined with the SABRE-SHARPER approach presented here to provide a wider range of structural and dynamic information.

Figure 4 presents a comparison of 19F SABRE spectra (A–C) with corresponding SABRE-SHARPER spectra (D–F) for 1, 2, and 3, respectively. A relatively short acquisition period of τ = 3.2 ms was used in each loop of the SHARPER sequence. In all cases, the SHARPER sequence has succeeded in collapsing the spectra into a single resonance. While all SABRE-SHARPER spectra demonstrate SNR improvements, the extent of enhancement varies for each molecule: 8.9-fold for 1, 17-fold for 2, and 7.2-fold for 3. The differences in SHARPER efficacy are also evident in the line widths achieved in each case. We note that due to the use of a matched exponential filter to optimize SNR, the observed line width is approximately double the fundamental line-width achieved by the SHARPER sequence. The effective Inline graphic relaxation times and corresponding line-widths Inline graphic for all SHARPER experiments are provided in the Supporting Information (Section S3).

Figure 4.

Figure 4

Comparison of (A–C) SABRE and (D–F) SABRE-SHARPER 19F benchtop (1 T) NMR spectra of 100 mM 1, 2 or 3 and 5 mM SABRE precatalyst in methanol. All spectra were acquired in a single scan and were apodized by a matched exponential filter (see Table S3). The SNR values represent averages over three repeat experiments. Full width at half-maximum values are shown for the SABRE SHARPER spectra. We note that these line-widths are effectively double the fundamental SHARPER line width due to the application of the matched filter. Individual traces were magnified as required to aid visualization.

In the case of 1, the SABRE-SHARPER sequence achieves a line width of Inline graphic = 0.97 Hz prior to the application of the matched filter. When SHARPER is applied to the sample without SABRE hyperpolarization, a much narrower line width of Inline graphic = 0.18 Hz is achieved (Figure S2A). This indicates that the nature of the hyperpolarization generated by SABRE is interfering with the optimal performance of SHARPER in this case. It is well established that, since the origin of the hyperpolarization in SABRE is the singlet state of p-H2, higher-order spin states can be enhanced alongside single-spin magnetization.39 Therefore a possible explanation for the imperfect refocusing of the SHARPER sequence is interference from higher-order terms (e.g., two-spin order 19F–1H states) that behave differently under the train of refocusing pulses. Nevertheless, even with imperfect refocusing, an overall SNR advantage of 1900-fold is obtained by the combined effects of SABRE and SHARPER.

In the case of 2, SABRE-SHARPER achieves an excellent line width of Inline graphic = 0.19 Hz prior to apodization, which is comparable to the performance of SHARPER without SABRE hyperpolarization (Figure S2B). Thus, for 2, the combination of SABRE and SHARPER results in an overall SNR advantage relative to the thermally polarized spectrum of 4100-fold.

In the case of 3, the SHARPER sequence efficiently refocuses the chemical shift difference between the two 19F resonances (1.9 ppm = 75 Hz) as well as the heteronuclear (JHF) and to some extent the homonuclear (JFF) couplings. As with 1, the line width achieved prior to apodization for 3, Inline graphic = 2.1 Hz, is much broader than the expected natural line width. However, here it is comparable to what is achieved for this sample using SHARPER without SABRE hyperpolarization (Figure S2C). Unlike 1 and 2, the spin system in 3 contains a significant homonuclear coupling of 3JFF = 18 Hz, which will continue to evolve during the τ = 3.2 ms period. In order to improve the efficiency of the line-narrowing, we repeated the SHARPER and SABRE-SHARPER experiments on 3 with a much shorter chunk time of τ = 0.8 ms. We note that the use of very short delays between RF pulses is more straightforward to achieve on a benchtop NMR spectrometer, when compared to high-field NMR, because the lower Larmor frequency requires lower RF power and therefore less stringent duty cycle constraints on long trains of RF pulses. Figure 5 presents a comparison of the SABRE-SHARPER FIDs acquired with τ = 3.2 ms and τ = 0.8 ms. We observe a significant increase in the apparent relaxation time from Inline graphic = 154 ms to Inline graphic = 547 ms. This corresponds to a decrease in line width from Inline graphic = 2.1 Hz to Inline graphic = 0.58 Hz, prior to apodization, and a corresponding increase in the SNR advantage of SABRE-SHARPER relative to SABRE from 7.2-fold to 16-fold. Overall, SABRE-SHARPER with τ = 0.8 ms provides a total SNR increase of 5700-fold relative to a standard 19F acquisition. We note that in the case of 1 and 2, no additional SNR improvements were observed when shorter chunk lengths were used (τ < 3.2 ms).

Figure 5.

Figure 5

Comparison of 19F SABRE-SHARPER FIDs (A,C) and spectra (B,D) of 100 mM of 3 with 5 mM SABRE precatalyst in methanol, acquired with chunk lengths of τ = 3.2 ms (A,B) and τ = 0.8 ms (C,D). Spectra were acquired in a single scan and were apodized by matched exponential filters (see Table S3). The SNR values represent averages over three repeat experiments. Full width at half-maximum values are shown for the SABRE-SHARPER spectra and include the effects of apodization.

An alternative strategy to boost SNR and simplify the 19F peaks is 1H decoupling. However, this is not straightforward to implement on the benchtop NMR instrument because 19F and 1H share the same coil and so heteronuclear 1H decoupling can only be implemented in an interleaved manner. In practice we find that, while SABRE-enhanced 19F {1H} benchtop NMR spectra are simplified, no net SNR gain is observed (see Figure S3). In addition, the decoupling conditions were challenging to optimize and led to significant peak and baseline distortions in many cases.

A limitation of the SHARPER approach for the analysis of mixtures is that it collapses all signals into a single peak and therefore does not distinguish between components. This can be addressed by using the selective variant, selSHARPER,31 in which the broadband excitation is replaced with a single pulsed-field gradient spin–echo (SPFGSE) element. On the Magritek Spinsolve benchtop NMR spectrometer used in this work, we are able to use the first-order shims to provide the pulsed-field gradient (PFG) pulses required for selective excitation. However, if no PFG is available, we have proposed alternative selection strategies previously.32

In the original version of selSHARPER, the rectangular pulses inside the acquisition loop are replaced by selective shaped pulses in order to maximize selectivity and refocus homonuclear couplings.31 However, we found that for the SABRE hyperpolarised samples explored here, this was not required to achieve efficient refocusing due to the short echo times that can be achieved on the benchtop spectrometer. Furthermore, the introduction of shaped pulses within the loop significantly reduced the observed SNR due to the time required to apply the refocusing pulses as a proportion of the total acquisition time. Therefore, we have implemented a variant of selSHARPER, shown in Figure 3B, where nonselective pulses are used in the loop. By applying this simplified variation of selSHARPER, it is possible to isolate the signal of interest in a mixture and boost its SNR in a single scan. If multiple components are of interest, the sequence can be repeated once for each target resonance. An additional benefit of this approach, is that selectivity can be improved without incurring an SNR penalty by increasing the duration of the initial selective RF pulse without changing the nonselective RF pulses inside the refocusing loop.

To evaluate the performance of SABRE combined with selSHARPER, a sample containing 50 mM each of 1 and 2 was analyzed. As these compounds have different optimal PTF values, we are able to achieve an element of selectivity from both SABRE and SHARPER. In addition, the resonances for these compounds are separated by 11.1 ppm (∼ 450 Hz at 1 T) and therefore provide a good test for the selectivity of the selSHARPER sequence. The results of combining SABRE and selSHARPER are shown in Figure 6, where SABRE hyperpolarization in Figure 6A and B was achieved using PTF = 6.2 mT, which is optimal for 2, while PTF ∼ 50 μT was used for Figure 6C and D to optimize the signal of 1.

Figure 6.

Figure 6

Comparison between SABRE-enhanced 19F benchtop NMR spectra with (A,C) standard and (B,D) selSHARPER acquisition for a mixture containing 1 and 2. SABRE spectra were acquired with (A,B) PTF = 6.2 mT and (C,D) PTF ∼ 50 μT. SABRE-selSHARPER spectra were acquired with selection of 2 in B and selection of 1 in D. All spectra were acquired in a single scan and were apodized by a matched exponential filter (see Table S3). The SNR values represent averages over three repeat experiments. Full width at half-maximum values are shown for the SABRE-selSHARPER spectra. Vertical scales were increased as required to aid visualization.

Considering first the effect of the PTF, we observe that for PTF = 6.2 mT (Figure 6A), the enhancement of 2 is maximized while 1 is significantly reduced relative to Figure 6C. In the case of PTF ∼ 50 μT, the SABRE enhancement of 1 is optimized; however, the enhancement of 2 is not suppressed. Instead, we observe strongly antiphase signals for 2, which are characteristic of SABRE enhancement of a two-spin-order term, rather than the desired in-phase single-spin magnetization.39 Nevertheless, in both cases, as observed in Figure 6B and D, the combination of SABRE with selSHARPER successfully isolates and narrows the target signal from the mixture, while increasing the SNR by factors of 3.6-fold and 15-fold, for 1 and 2, respectively.

When the original selSHARPER sequence,32 with 5 ms selective pulses inside the loop, is applied to the mixture sample under SABRE conditions, there was a significant sensitivity penalty that led to SNR enhancements of only 2.7-fold and 7.3-fold for 1 and 2, respectively (Figure S4). This reduction in SNR performance relative to the results in Figure 6 is due to the different way the couplings are removed in the original selective SHARPER experiments, where longer selective pulses within the acquisition loop accelerate the apparent T2 relaxation, causing signal broadening and hence lower SNR. Therefore, selSHARPER with nonselective pulses in the loop and a short chunk time, τ (Figure 3b), is the preferred method for optimizing SNR with the combined SABRE-selSHARPER approach.

Another popular strategy for removing peak splittings is to use pure shift techniques to simplify spectra via broadband homonuclear decoupling.40 A benefit of pure shift approaches is that chemical shift information is retained; however, this generally comes at the cost of signal loses, which, depending on the method, can be as much as 75–90%.40 Furthermore, the widely used PSYCHE (pure shift yielded by chirp excitation) methods,41 which retain ca. 25% of the 1H NMR signal, use a pseudo 2D acquisition strategy that requires multiple acquisitions to produce a 1D spectrum and therefore cannot easily be combined with single-shot SABRE experiments. An exception is the HOBS (homodecoupled band selective) method,42 which provides additional sensitivity and can, in principle, be acquired in a single scan. A limitation of this approach, particularly for benchtop NMR applications where chemical shift dispersion is limited by the lower magnetic field strength, is the requirement for chemical shift separation between the coupled peaks. Additionally, the standard pure shift methods do not compensate for magnetic field inhomogeneity. Two pure shift methods that are designed to generate high resolution spectra in the presence of field inhomogeneity have been introduced.43,44 However, these are pseudo 2D methods that cannot be acquired in a single scan. Therefore, if the goal is to maximize the SNR in a single shot SABRE experiment, SHARPER is the superior approach, as it achieves homo- and heteronuclear decoupling at no expense to sensitivity in a single scan while compensating for magnetic field inhomogeneity. Additionally, multiresonance SHARPER acquisition can be achieved and has been applied to reaction monitoring followed by simultaneous acquisition of one signal from each of the reactant and product.45

All of the SABRE-SHARPER spectra presented here were acquired in a single scan immediately following SABRE hyperpolarization. However, if there are significant background nonhyperpolarized signals present, these can give rise to off-resonance artifacts in the SHARPER spectra. Such signals can be efficiently removed by acquiring a reference nonhyperpolarised second scan (Figure S5). However, this second scan will necessarily decrease the overall SNR as it adds in extra noise but no further hyperpolarised signal. Therefore, it should only be used if the artifacts introduced by any background signals overlap or interfere with the target signal in the SABRE-SHARPER spectrum.

Conclusions

We have presented an optimized approach to 19F benchtop NMR spectroscopy that exploits the combined SNR enhancements of SABRE hyperpolarization and the multiplet-refocusing method, SHARPER. The SABRE-SHARPER method was tested on three fluorinated pyridines, 3,5-difluoropyridine, 3-(difluoromethyl)pyridine, and 3,4,5-trifluoropyridine, which are characterized by a range of homo- and heteronuclear J coupling constants and chemical shift differences. Significant 19F SNR enhancements of up to 360-fold were obtained by SABRE hyperpolarization in a single scan, following optimization of the polarization transfer field (PTF). The combination of SABRE with SHARPER achieved a further boost in SNR of up to 17-fold by removing inhomogeneous broadening and refocusing homo- and/or heteronuclear J couplings and chemical shift differences on the order of 75 Hz. Taken together, the SABRE-SHARPER approach achieved SNR enhancements of up to 5700-fold relative to a standard 19F NMR acquisition. SNR increases were not observed when using 1H decoupling to remove heteronuclear couplings due to the limitations of the shared 1H/19F channel. A selective variant of SHARPER was combined with SABRE to enhance the sensitivity of the two components in a mixture of 3,5-difluoropyridine and 3-(difluoromethyl)pyridine, which are separated by only 450 Hz at 1 T. While a separate experiment is required to enhance each component, excellent selectivity and overall SNR enhancements of 630-fold and 4300-fold were achieved for 3,5-difluoropyridine and 3-difluoromethylpyridine, respectively. For analytical applications, the quantification of hyperpolarised NMR signals is a challenge because the observed signal intensity depends on the hyperpolarization efficiency, which is analyte specific and can vary with experimental parameters, including analyte concentration. A route to overcome this has been proposed by Tessari and co-workers.46 They have demonstrated a linear dependence of the observed 1H SABRE hyperpolarization as a function of analyte concentration for a range of N-heterocycles, where one or more cosubstrates are used to stabilize the active SABRE catalyst.46,47 A feature of this method is the requirement that the target analytes are present in substoichiometric concentrations relative to the catalyst. In this way, the limit of detection for SABRE-enhanced analytes can be reduced to submillimolar concentrations. Indeed, Eshuis et al. have reported the detection of SABRE-enhanced 1H NMR signals for analyte concentrations on the order of 10 μM using NMR detection at 11.7 T (600 MHz). Combining this strategy with the SABRE-SHARPER approach provides a promising route to achieving submillimolar limits of detection for benchtop 19F NMR spectroscopy. In addition, we note that the SHARPER acquisition itself is quantitative, as has been demonstrated previously for its use in reaction monitoring applications.31 While we have focused here on 19F benchtop NMR, we anticipate that equally significant SNR gains can be achieved by applying the SABRE-SHARPER approach to other low-sensitivity nuclei, such as 13C. Equally, we observe comparable SNR gains for 1H benchtop NMR spectra with SABRE-SHARPER; however, the ability to selectively enhance a target species in this case is limited by the comparatively narrow chemical shift range for 1H.

Acknowledgments

This research was supported by funding from EPSRC (EP/M020983/1, EP/R028745/1, EP/S016139/1, EP/N509644/1, and EP/R030065/1) and a Department of Chemistry Wild Fund Scholarship (AIST). We thank Dr. Victoria Annis for synthesis of the SABRE catalyst and Dr Craig Eccles from Magritek Ltd for technical assistance.

Data Availability Statement

The NMR data underlying this study are openly available through the York Research Database at https://doi.org/10.15124/252bc49d-8f1f-4c0d-b1bb-9aeadcdde0c8.

Supporting Information Available

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsmeasuresciau.2c00055.

  • SABRE sample preparation details; experimental and processing parameters; effective relaxation times, T2S, for SHARPER experiments; supplementary figures; implementation of Gaussian pulse with RF nonlinearity correction (PDF)

Author Present Address

Oxford Instruments Magnetic Resonance, High Wycombe, HP12 3SE, UK

The authors declare no competing financial interest.

Supplementary Material

tg2c00055_si_001.pdf (20.6MB, pdf)

References

  1. Soyler A.; Cikrikci S.; Cavdaroglu C.; Bouillaud D.; Farjon J.; Giraudeau P.; Oztop M. H. Multi-scale benchtop 1H NMR spectroscopy for milk analysis. LWT 2021, 139, 110557. 10.1016/j.lwt.2020.110557. [DOI] [Google Scholar]
  2. Edgar M.; Percival B. C.; Gibson M.; Jafari F.; Grootveld M. Low-field benchtop NMR spectroscopy as a potential non-stationary tool for point-of-care urinary metabolite tracking in diabetic conditions. Diabetes Res. Clin. Pract. 2021, 171, 108554. 10.1016/j.diabres.2020.108554. [DOI] [PubMed] [Google Scholar]
  3. Gilbert N.; Mewis R. E.; Sutcliffe O. B. Fast & fluorinated – Development and validation of a rapid benchtop NMR approach and other routine screening methods for the detection and quantification of synthesized fluorofentanyl derivatives. Forensic Chem. 2021, 23, 100321. 10.1016/j.forc.2021.100321. [DOI] [Google Scholar]
  4. Shilliday E. R.; Barrow B.; Langford D.; Ling N. N.; Robinson N.; Johns M. L. Quantitative measurement of Mono-Ethylene Glycol (MEG) content using low-field Nuclear Magnetic Resonance (NMR). J. Nat. Gas Sci. Eng. 2022, 101, 104520. 10.1016/j.jngse.2022.104520. [DOI] [Google Scholar]
  5. Finch N.; Percival B.; Hunter E.; Blagg R. J.; Blackwell E.; Sagar J.; Ahmad Z.; Chang M.-W.; Hunt J. A.; Mather M. L.; Tasker S.; De Risio L.; Wilson P. B. Preliminary demonstration of benchtop NMR metabolic profiling of feline urine: chronic kidney disease as a case study. BMC Res. Notes 2021, 14, 469. 10.1186/s13104-021-05888-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Grootveld M.; Percival B.; Gibson M.; Osman Y.; Edgar M.; Molinari M.; Mather M. L.; Casanova F.; Wilson P. B. Progress in low-field benchtop NMR spectroscopy in chemical and biochemical analysis. Anal. Chim. Acta 2019, 1067, 11–30. 10.1016/j.aca.2019.02.026. [DOI] [PubMed] [Google Scholar]
  7. Inoue M.; Sumii Y.; Shibata N. Contribution of Organofluorine Compounds to Pharmaceuticals. ACS Omega 2020, 5, 10633–10640. 10.1021/acsomega.0c00830. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Howe P. W. Recent developments in the use of fluorine NMR in synthesis and characterisation. Prog. Nucl. Magn. Reson. Spectrosc. 2020, 118–119, 1–9. 10.1016/j.pnmrs.2020.02.002. [DOI] [PubMed] [Google Scholar]
  9. Okaru A. O.; Brunner T. S.; Ackermann S. M.; Kuballa T.; Walch S. G.; Kohl-Himmelseher M.; Lachenmeier D. W. Application of 19F NMR Spectroscopy for Content Determination of Fluorinated Pharmaceuticals. J. Anal. Methods Chem. 2017, 2017, 1–7. 10.1155/2017/9206297. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Gimenez D.; Phelan A.; Murphy C. D.; Cobb S. L. 19F NMR as a tool in chemical biology. Beilstein J. Org. Chem. 2021, 17, 293–318. 10.3762/bjoc.17.28. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Wei R.; Hall A. M. R.; Behrens R.; Pritchard M. S.; King E. J.; Lloyd-Jones G. C. Stopped-Flow 19F NMR Spectroscopic Analysis of a Protodeboronation Proceeding at the Sub-Second Time-Scale. Eur. J. Org. Chem. 2021, 2021, 2331–2342. 10.1002/ejoc.202100290. [DOI] [Google Scholar]
  12. Heerah K.; Waclawek S.; Konzuk J.; Longstaffe J. G. Benchtop 19F NMR spectroscopy as a practical tool for testing of remedial technologies for the degradation of perfluorooctanoic acid, a persistent organic pollutant. Magn. Reson. Chem. 2020, 58, 1160–1167. 10.1002/mrc.5005. [DOI] [PubMed] [Google Scholar]
  13. Adams R. W.; Aguilar J. A.; Atkinson K. D.; Cowley M. J.; Elliott P. I. P.; Duckett S. B.; Green G. G. R.; Khazal I. G.; López-Serrano J.; Williamson D. C. Reversible Interactions with para-Hydrogen Enhance NMR Sensitivity by Polarization Transfer. Science 2009, 323, 1708–1711. 10.1126/science.1168877. [DOI] [PubMed] [Google Scholar]
  14. Richardson P. M.; John R. O.; Parrott A. J.; Rayner P. J.; Iali W.; Nordon A.; Halse M. E.; Duckett S. B. Quantification of hyperpolarisation efficiency in SABRE and SABRE-Relay enhanced NMR spectroscopy. Phys. Chem. Chem. Phys. 2018, 20, 26362–26371. 10.1039/C8CP05473H. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Robinson A. D.; Richardson P. M.; Halse M. E. Hyperpolarised 1H–13C Benchtop NMR Spectroscopy. Appl. Sci. 2019, 9, 1173. 10.3390/app9061173. [DOI] [Google Scholar]
  16. Richardson P. M.; Parrott A. J.; Semenova O.; Nordon A.; Duckett S. B.; Halse M. E. SABRE hyperpolarization enables high-sensitivity 1H and 13C benchtop NMR spectroscopy. Analyst 2018, 143, 3442–3450. 10.1039/C8AN00596F. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Olaru A. M.; Robertson T. B. R.; Lewis J. S.; Antony A.; Iali W.; Mewis R. E.; Duckett S. B. Extending the Scope of 19F Hyperpolarization through Signal Amplification by Reversible Exchange in MRI and NMR Spectroscopy. ChemistryOpen 2018, 7, 97–105. 10.1002/open.201700166. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Colell J. F. P.; Logan A. W. J.; Zhou Z.; Shchepin R. V.; Barskiy D. A.; Ortiz G. X.; Wang Q.; Malcolmson S. J.; Chekmenev E. Y.; Warren W. S.; Theis T. Generalizing, Extending, and Maximizing Nitrogen-15 Hyperpolarization Induced by Parahydrogen in Reversible Exchange. J. Phys. Chem. C 2017, 121, 6626–6634. 10.1021/acs.jpcc.6b12097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Birchall J. R.; Kabir M. S. H.; Salnikov O. G.; Chukanov N. V.; Svyatova A.; Kovtunov K. V.; Koptyug I. V.; Gelovani J. G.; Goodson B. M.; Pham W.; Chekmenev E. Y. Quantifying the effects of quadrupolar sinks via 15N relaxation dynamics in metronidazoles hyperpolarized via SABRE-SHEATH. Chem. Commun. 2020, 56, 9098–9101. 10.1039/D0CC03994B. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Salnikov O. G.; Chukanov N. V.; Svyatova A.; Trofimov I. A.; Kabir M. S. H.; Gelovani J. G.; Kovtunov K. V.; Koptyug I. V.; Chekmenev E. Y. 15 N NMR Hyperpolarization of Radiosensitizing Antibiotic Nimorazole by Reversible Parahydrogen Exchange in Microtesla Magnetic Fields. Angew. Chem., Int. Ed. Engl. 2021, 60, 2406–2413. 10.1002/anie.202011698. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Adelabu I.; Ettedgui J.; Joshi S. M.; Nantogma S.; Chowdhury M. R. H.; McBride S.; Theis T.; Sabbasani V. R.; Chandrasekhar M.; Sail D.; Yamamoto K.; Swenson R. E.; Krishna M. C.; Goodson B. M.; Chekmenev E. Y. Rapid 13C Hyperpolarization of the TCA Cycle Intermediate -Ketoglutarate via SABRE-SHEATH. Anal. Chem. 2022, 94, 13422–13431. 10.1021/acs.analchem.2c02160. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Adelabu I.; TomHon P.; Kabir M. S. H.; Nantogma S.; Abdulmojeed M.; Mandzhieva I.; Ettedgui J.; Swenson R. E.; Krishna M. C.; Theis T.; Goodson B. M.; Chekmenev E. Y. Order-Unity 13C Nuclear Polarization of [1-13C]Pyruvate in Seconds and the Interplay of Water and SABRE Enhancement. ChemPhysChem 2022, 23, e202100839 10.1002/cphc.202100839. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Shchepin R. V.; Barskiy D. A.; Coffey A. M.; Feldman M. A.; Kovtunova L. M.; Bukhtiyarov V. I.; Kovtunov K. V.; Goodson B. M.; Koptyug I. V.; Chekmenev E. Y. Robust Imidazole-15N2 Synthesis for High-Resolution Low-Field (0.05 T) 15N Hyperpolarized NMR Spectroscopy. ChemistrySelect 2017, 2, 4478–4483. 10.1002/slct.201700718. [DOI] [Google Scholar]
  24. Tennant T.; Hulme M.; Robertson T.; Sutcliffe O.; Mewis R. Benchtop NMR analysis of piperazine-based drugs hyperpolarised by SABRE. Magn. Reson. Chem. 2020, 58, 1151–1159. 10.1002/mrc.4999. [DOI] [PubMed] [Google Scholar]
  25. Semenova O.; Richardson P. M.; Parrott A. J.; Nordon A.; Halse M. E.; Duckett S. B. Reaction Monitoring Using SABRE-Hyperpolarized Benchtop (1 T) NMR Spectroscopy. Anal. Chem. 2019, 91, 6695–6701. 10.1021/acs.analchem.9b00729. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Chae H.; Min S.; Jeong H. J.; Namgoong S. K.; Oh S.; Kim K.; Jeong K. Organic Reaction Monitoring of a Glycine Derivative Using Signal Amplification by Reversible Exchange-Hyperpolarized Benchtop Nuclear Magnetic Resonance Spectroscopy. Anal. Chem. 2020, 92, 10902–10907. 10.1021/acs.analchem.0c01270. [DOI] [PubMed] [Google Scholar]
  27. Robertson T. B. R.; Antonides L. H.; Gilbert N.; Benjamin S. L.; Langley S. K.; Munro L. J.; Sutcliffe O. B.; Mewis R. E. Hyperpolarization of Pyridyl Fentalogues by Signal Amplification By Reversible Exchange (SABRE). ChemistryOpen 2019, 8, 1375–1382. 10.1002/open.201900273. [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Ivanov K. L.; Pravdivtsev A. N.; Yurkovskaya A. V.; Vieth H.-M.; Kaptein R. The role of level anti-crossings in nuclear spin hyperpolarization. Prog. Nucl. Magn. Reson. Spectrosc. 2014, 81, 1–36. 10.1016/j.pnmrs.2014.06.001. [DOI] [PubMed] [Google Scholar]
  29. Theis T.; Truong M. L.; Coffey A. M.; Shchepin R. V.; Waddell K. W.; Shi F.; Goodson B. M.; Warren W. S.; Chekmenev E. Y. Microtesla SABRE Enables 10% Nitrogen-15 Nuclear Spin Polarization. J. Am. Chem. Soc. 2015, 137, 1404–1407. 10.1021/ja512242d. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Cowley M. J.; Adams R. W.; Atkinson K. D.; Cockett M. C. R.; Duckett S. B.; Green G. G. R.; Lohman J. A. B.; Kerssebaum R.; Kilgour D.; Mewis R. E. Iridium N-heterocyclic carbene complexes as efficient catalysts for magnetization transfer from para-hydrogen. J. Am. Chem. Soc. 2011, 133, 6134–6137. 10.1021/ja200299u. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Jones A. B.; Lloyd-Jones G. C.; Uhrín D. SHARPER Reaction Monitoring: Generation of a Narrow Linewidth NMR Singlet, without X-Pulses, in an Inhomogeneous Magnetic Field. Anal. Chem. 2017, 89, 10013–10021. 10.1021/acs.analchem.7b02437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Dickson C. L.; Peat G.; Rossetto M.; Halse M. E.; Uhrín D. SHARPER-enhanced benchtop NMR: improving SNR by removing couplings and approaching natural linewidths. Chem. Commun. 2022, 58, 5534–5537. 10.1039/D2CC01325H. [DOI] [PubMed] [Google Scholar]
  33. Richardson P. M.; Jackson S.; Parrott A. J.; Nordon A.; Duckett S. B.; Halse M. E. A simple hand-held magnet array for efficient and reproducible SABRE hyperpolarisation using manual sample shaking. Magn. Reson. Chem. 2018, 56, 641–650. 10.1002/mrc.4687. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Chukanov N. V.; Salnikov O. G.; Shchepin R. V.; Svyatova A.; Kovtunov K. V.; Koptyug I. V.; Chekmenev E. Y. 19F Hyperpolarization of 15N-3-19F-Pyridine Via Signal Amplification by Reversible Exchange. J. Phys. Chem. C 2018, 122, 23002–23010. 10.1021/acs.jpcc.8b06654. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Shchepin R. V.; Goodson B. M.; Theis T.; Warren W. S.; Chekmenev E. Y. Toward Hyperpolarized 19F Molecular Imaging via Reversible Exchange with Parahydrogen. ChemPhysChem 2017, 18, 1961–1965. 10.1002/cphc.201700594. [DOI] [PubMed] [Google Scholar]
  36. Shchepin R. V.; Truong M. L.; Theis T.; Coffey A. M.; Shi F.; Waddell K. W.; Warren W. S.; Goodson B. M.; Chekmenev E. Y. Hyperpolarization of “Neat” Liquids by NMR Signal Amplification by Reversible Exchange. J. Phys. Chem. Lett. 2015, 6, 1961–1967. 10.1021/acs.jpclett.5b00782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. TomHon P.; Akeroyd E.; Lehmkuhl S.; Chekmenev E. Y.; Theis T. Automated Pneumatic Shuttle for Magnetic Field Cycling and Parahydrogen Hyperpolarized Multidimensional NMR. J. Magn. Reson. 2020, 312, 106700. 10.1016/j.jmr.2020.106700. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Kiryutin A. S.; Yurkovskaya A. V.; Zimmermann H.; Vieth H.-M.; Ivanov K. L. Complete magnetic field dependence of SABRE-derived polarization. Magn. Reson. Chem. 2018, 56, 651–662. 10.1002/mrc.4694. [DOI] [PubMed] [Google Scholar]
  39. Mewis R. E.; Atkinson K. D.; Cowley M. J.; Duckett S. B.; Green G. G. R.; Green R. A.; Highton L. A. R.; Kilgour D.; Lloyd L. S.; Lohman J. A. B.; Williamson D. C. Probing signal amplification by reversible exchange using an NMR flow system: Signal amplification by reversible exchange. Magn. Reson. Chem. 2014, 52, 358–369. 10.1002/mrc.4073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  40. Castañar L.; Parella T. Broadband 1H homodecoupled NMR experiments: recent developments, methods and applications. Magn. Reson. Chem. 2015, 53, 399–426. 10.1002/mrc.4238. [DOI] [PubMed] [Google Scholar]
  41. Foroozandeh M.; Adams R. W.; Meharry N. J.; Jeannerat D.; Nilsson M.; Morris G. A. Ultrahigh-Resolution NMR Spectroscopy. Angew. Chem., Int. Ed. 2014, 53, 6990–6992. 10.1002/anie.201404111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Castañar L.; Nolis P.; Virgili A.; Parella T. Full Sensitivity and Enhanced Resolution in Homodecoupled Band-Selective NMR Experiments. Chem.—Eur. J. 2013, 19, 17283–17286. 10.1002/chem.201303235. [DOI] [PubMed] [Google Scholar]
  43. Huang Y.; Cao S.; Yang Y.; Cai S.; Zhan H.; Tan C.; Lin L.; Zhang Z.; Chen Z. Ultrahigh-Resolution NMR Spectroscopy for Rapid Chemical and Biological Applications in Inhomogeneous Magnetic Fields. Anal. Chem. 2017, 89, 7115–7122. 10.1021/acs.analchem.7b01036. [DOI] [PubMed] [Google Scholar]
  44. Morris G. A.; Gibbs A. High-sensitivity, high-resolution NMR in inhomogeneous magnetic fields. J. Magn. Reson. 1988, 78, 594–596. 10.1016/0022-2364(88)90145-X. [DOI] [Google Scholar]
  45. Davy M.; Dickson C. L.; Wei R.; Uhrín D.; Butts C. P. Monitoring off-resonance signals with SHARPER NMR – the MR-SHARPER experiment. Analyst 2022, 147, 1702–1708. 10.1039/D2AN00134A. [DOI] [PubMed] [Google Scholar]
  46. Eshuis N.; Hermkens N.; van Weerdenburg B. J. A.; Feiters M. C.; Rutjes F. P. J. T.; Wijmenga S. S.; Tessari M. Toward nanomolar detection by NMR through SABRE hyperpolarization. J. Am. Chem. Soc. 2014, 136, 2695–2698. 10.1021/ja412994k. [DOI] [PubMed] [Google Scholar]
  47. Eshuis N.; vanWeerdenburg B. J. A.; Feiters M. C.; Rutjes F. P. J. T.; Wijmenga S. S.; Tessari M. Quantitative Trace Analysis of Complex Mixtures Using SABRE Hyperpolarization. Angew. Chem., Int. Ed. 2015, 54, 1481–1484. 10.1002/anie.201409795. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

tg2c00055_si_001.pdf (20.6MB, pdf)

Data Availability Statement

The NMR data underlying this study are openly available through the York Research Database at https://doi.org/10.15124/252bc49d-8f1f-4c0d-b1bb-9aeadcdde0c8.


Articles from ACS Measurement Science Au are provided here courtesy of American Chemical Society