Abstract
Objective. Pencil beam scanning (PBS) proton therapy target dose conformity can be improved with energy layer-specific collimation. One such collimator is the dynamic collimation system (DCS), which consists of four nickel trimmer blades that intercept the scanning beam as it approaches the lateral extent of the target. While the dosimetric benefits of the DCS have been demonstrated through computational treatment planning studies, there has yet to be experimental verification of these benefits for composite multi-energy layer fields. The objective of this work is to dosimetrically characterize and experimentally validate the delivery of dynamically collimated proton therapy with the DCS equipped to a clinical PBS system. Approach. Optimized single field, uniform dose treatment plans for 3 × 3 × 3 cm3 target volumes were generated using Monte Carlo dose calculations with depths ranging from 5 to 15 cm, trimmer-to-surface distances ranging from 5 to 18.15 cm, with and without a 4 cm thick polyethylene range shifter. Treatment plans were then delivered to a water phantom using a prototype DCS and an IBA dedicated nozzle system and measured with a Zebra multilayer ionization chamber, a MatriXX PT ionization chamber array, and Gafchromic™ EBT3 film. Main results. For measurements made within the SOBPs, average 2D gamma pass rates exceeded 98.5% for the MatriXX PT and 96.5% for film at the 2%/2 mm criterion across all measured uncollimated and collimated plans, respectively. For verification of the penumbra width reduction with collimation, film agreed with Monte Carlo with differences within 0.3 mm on average compared to 0.9 mm for the MatriXX PT. Significance. We have experimentally verified the delivery of DCS-collimated fields using a clinical PBS system and commonly available dosimeters and have also identified potential weaknesses for dosimeters subject to steep dose gradients.
Keywords: proton therapy, collimation, Monte Carlo, dosimetry
1. Introduction
Pencil beam scanning proton therapy (PBS-PT) is an increasingly common form of radiotherapy that utilizes a scanned beam of protons in successive energy layers to deliver a complex radiation dose distribution (Lomax 1999, Tian et al 2018). Target dose conformity in PBS-PT is dominated by the spot size, which is dictated by the standard deviation of the measured Gaussian fluence distribution in air at isocenter () (Moignier et al 2016a, Grewal et al 2021, Nelson et al 2021a). Clinical spot sizes can be as small as 3 mm and greater than 20 mm at isocenter. Spot sizes depend on many factors such as the nozzle design, the energy selection system upstream, the requested beam energy, the beam divergence, and air gap between the treatment nozzle and patient surface (Gillin et al 2010, Titt et al 2010, Kralik et al 2015, Langner et al 2017, Smith 2019, Rahman et al 2020, Nelson et al 2021a). For example, Wang et al found that a spot size of less than 4.3 mm for PBS-PT was required to produce a superior treatment to state-of-the-art photon modalities for the treatment of intracranial lesions (Wang et al 2014). Widesott et al found that values must be less than 3 mm, 4 mm and 6 mm, to maintain clinical OAR doses relative to advanced photon therapy techniques for prostate, head and neck, and malignant mesothelioma cases, respectively. Many vendors have employed external collimation devices to reduce the effective spot size at the patient surface. Collimators used in PBS-PT include fixed per-field apertures (Moteabbed et al 2016, Winterhalter et al 2018, Maes et al 2019), multileaf collimators adapted from x-ray modalities, (Bues et al 2005, Daartz et al 2009, Daartz and Maughan 2014) and other novel collimation devices (Hyer et al 2014a, Kang and Pang 2020, Vilches-Freixas et al 2020, Chiang et al 2021, Grewal et al 2021). These devices have greatest impact on the treatment of shallow tumors, which require lower beam energies with larger spot sizes and range shifters that introduce considerable beam-broadening scatter.
A dynamic collimation system (DCS) is currently under development to reduce the lateral penumbra in PBS-PT to enable dynamically collimated proton therapy (DC-PT) (Hyer et al 2014a, Geoghegan et al 2020, Smith et al 2020). The DCS consists of two pairs of nickel trimmer blades (four total), an optional 4 cm thick polyethylene range shifter (WET = 4.1 cm5) to treat shallow (<4 cm) targets, and is mounted to a telescoping accessory nozzle that can produce various trimmer-to-surface distances (TSDs) ranging from 5 to 18.15 cm to accompany a wide array of patient and setup geometries. Each trimmer has a range of motion that provides up to 7.5 cm between the central axis of the beam coordinate system and the medial edge of the trimmer, resulting in a maximum field of view of 15 × 15 cm2 at isocenter. Polyethylene was selected as the range shifter material due to its minimal scattering power and thus, preservation of the in-air spot size and resulting dose conformity (Shen et al 2015). Nickel was selected for the trimmers based off previous results of Smith (2019), who showed that the secondary neutron ambient dose equivalence for PBS with the DCS was less than reported for uniform scanning with static brass apertures (Islam et al 2017, Smith et al 2019a).
Preliminary dosimetric and mechanical validation steps have been taken with the DCS mounted to the IBA dedicated nozzle (DN) system located at the Miami Cancer Institute (MCI), including the development of a Monte Carlo model and a central axis alignment device for trimmer positioning (Geoghegan et al 2022, Nelson et al 2021a). While the dosimetric benefits of DC-PT have been extensively evaluated through treatment planning studies, (Hyer et al 2014b, Smith et al 2016, Moignier et al 2016a, Moignier et al 2016b, 2016c) direct measurements of dynamically collimated proton dose distributions have yet to be performed. Therefore, the purpose of this work is to experimentally validate volumetric treatment plans with the DCS mounted to the IBA DN system using multiple detectors. This work will also provide an insight into detector performance for benchmarking the sharp dose gradients produced by the DCS.
2. Materials and methods
2.1. Treatment planning and dose optimization
The open-source Dynamic Collimation Monte Carlo package (Nelson et al 2021a), or DCMC, was used to commission TOPAS-based Monte Carlo models of the DCS and IBA DN nozzle to create and optimize three-dimensional uncollimated and collimated treatments for 3 × 3 × 3 cm3 cubic planning target volumes (PTVs). While this model has been previously validated against measurements of individual collimated beamlets, it has yet to be validated against composite treatment fields. Each PTV was placed at various depths to parametrize the penumbra sharpening as a function of depth in water. The impact of the trimmer-to-surface distance was also investigated for range shifted (RS) and non-range shifted (NRS) treatments. A summary of the target volumes and beam parameters investigated is in table 1.
Table 1.
Summary of relevant plan parameters for 3 × 3 × 3 cm3 target volumes used to investigate penumbra sharpening as a function of the target’s center depth. Plan names are listed using a CX-DY-Z nomenclature where X is the side length of the cube in cm, Y is the depth of the cube center in cm, and Z specifies if the plan utilized range shifted (RS) or non-range shifted (NRS) beamlets.
| Cube size (cm3) | Depth (cm) | Range shifter present? (Y/N) | Number of energy layers | Energy range (MeV) | Plan name |
|---|---|---|---|---|---|
| 5 | Y | 11 a | 96.8–120.9 | C3-D5-RS | |
| 3 × 3 × 3 | 6 | N | 12 | 71.4–100.6 | C3-D6-NRS |
| 10 | N | 12 | 104.3–127.0 | C3-D10-NRS | |
| 15 | N | 12 | 136.2–155.7 | C3-D15-NRS |
Proximal layer dropped due to minimum MU constraints.
For each PTV, candidate beam spots were placed within a 2.5 mm expansion of PTV in the lateral and depth dimensions to ensure a uniform dose within the PTV. A 3 mm lateral in-plane spot spacing was utilized with a ∼3 mm energy layer spacing that resulted in 12 planned energy layers for all PTVs. For the collimated plans, the medial face of each trimmer was positioned to collimate the outer-most beamlets and trimmers were moved in between each energy layer to match the divergent field edge to the cubic target volume.
TOPAS (Perl et al 2012) (Version 3.8.p1) was used to simulate each candidate beamlet incident on a 40 × 40 × 40 cm3 water phantom and the dose to water was scored in Gy/proton on a 3D voxelized dose grid with a 1 mm isotropic resolution. Each plan consisted of 2028 candidate beamlets that were simulated using the computing resources and assistance of the UW-Madison Center for High Throughput Computing. Each simulation consisted of 1e6 histories per beamlet and utilized the default TOPAS physics modules: g4em-standard_opt4, g4h-phy_QGSP_BIC_HP, g4decay, g4ion-binarycascade, g4h-elastic_HP, and g4stopping.
Subsequent beamlet weight optimizations were performed using an in-house MATLAB-based linear least squares optimizer developed by Flynn et al (2007), (Flynn 2007) where the number of protons was optimized to provide a uniform dose to the PTV. Following the optimization process, the number of protons was converted to MU using a previously-established Monte Carlo model conversion factor that describes the number of protons per MU required to reproduce the IAEA TRS-398 reference dose measurement (Nelson et al 2021a, IAEA. IAEA TRS 398 2000). A uniform dose of 5 Gy was prescribed to the PTV to (1) minimize the amount of beamlets dropped due to minimum MU constraints during experimental validation while, (2) ensuring that a majority of the dose was within the dose calibration window of the EBT3 film that was later used to benchmark these dose distributions. The proximal energy layer for the range shifted case was removed following weight optimization due to the minimum MU constraint of 0.05 MU at the Miami Cancer Institute. Optimizations were performed to achieve identical 90% target prescription dose coverage between the uncollimated and collimated plans.
2.2. Treatment delivery
Following the treatment planning and optimization, spot positions, energies, and the optimized weights were used to write PBS layer definition (PLD) files that allow for delivery of scanned fields from the accelerator console in a physics/service mode. The PLD contains details of the energy required for each layer and, for each layer, the number of MUs and position for each spot. Because the optimized beamlet weights are in units of number of initially simulated protons, these weights were converted to MUs using an energy-dependent TRS-398 derived Monte Carlo conversion factor
where and are the measured and simulated doses under TRS-398 reference conditions, respectively.
Prior to irradiation, the DCS trimmers were aligned to isocenter using the on-board x-ray imager to ensure accurate trimmer positioning. The trimmers were then moved to the first set of intended positions and verified with subsequent x-ray images. Trimmers were then moved in between energy layers using the ‘pause after layer’ function during the spot scanning delivery. While this did increase the overall measurement time, pausing between layers was necessary to move the trimmers because the DCS control system and IBA scanning controller system were operated independently. In the future, deliveries that require more than one trimmer position in an energy layer will require integration of the DCS control system with the IBA scanning controller, allowing for trimmer positions to be set prior to delivery of a given spot.
2.3. Experimental dosimetry
2.3.1. Spread-out Bragg peak (SOBP) verification
Spread out Bragg peak (SOBP) profiles were measured for the NRS plans (6, 10 and 15 cm depths) using the Zebra multilayer ionization chamber (MLIC, IBA Dosimetry, Germany) to verify the SOBP range (depth of distal 90%), width, and uniformity of the evaluate plans agreed with the simulated SOBP distributions. The Zebra MLIC consists of 180 parallel plate ionization chambers (2.5 cm diameter) with an effective measurement resolution of 2 mm along the beam axis, and has been shown to be a quick and effective device for measuring proton range and SOBP width for both pristine beamlets and composite fields (Dhanesar et al 2013, Bäumer et al 2015). All collimated irradiations were performed using a 5 cm TSD.
2.3.2. Lateral profile measurements
Lateral profiles of the cube dose distributions were measured using GafchromicTM EBT3 film and a MatriXX PT ionization chamber array (referred to as ‘MatriXX’ onward) in the DigiPhant PT (IBA Dosimetry, Germany) scanning water tank. Planar 2D measurements were performed and the center of the SOBP for each of the NRS plans investigated while EBT3 film was used to measure the range shifted plan. Additional near-surface measurement was taken with the MatriXX and film at a depth of 4.1 cm for the 15 cm deep cube to investigate a near-surface dose enhancement with collimation. The surface dose enhancement with collimation was further assessed for the other plans through lateral profile analysis of the Monte Carlo dose distributions at a depth of 5 mm by comparing the dose horns between the uncollimated and collimated dose distributions. Surface dose enhancement, mainly found near the target edges, is a result of the increased relative weights on collimated beamlets that also have an increased entrance dose on a per-beamlet basis from low-energy scattered protons originating from the trimmers.
The MatriXX, shown along with the DigiPhant water tank in figure 1(a), is a well characterized ion chamber array that utilizes 1020 vented parallel plate chambers with a 7.6 mm center-to-center spacing, each with an active volume of 32 mm3 (Arjomandy et al 2008, 2010, Lin et al 2015, Chan et al 2017). The effective measurement resolution with the MatriXX was increased by merging two datasets: one acquired with the MatriXX in its nominal position and another where the MatriXX was shifted in the horizontal and vertical directions by 3.8 mm, or half the distance between the ionization chambers. Merged datasets were then interpolated to a 1.5 mm dose grid for further analysis. Despite its extensive use for patient-specific quality assurance for passive scattering and PBS delivery, the MatriXX has yet to be used to benchmark extremely sharp dose gradients expected to be produced by the DCS (up to 20%/mm mid-SOBP).
Figure 1.

(a) Photograph of the MatriXX array and the DigiPhant water tank used for lateral profile measurements. (b) Rendering of DigiPhant-compatible EBT3 film holder.
Because of these gradients, high-resolution EBT3 film was utilized to corroborate MatriXX results. A custom polyethylene based EBT3 film holder was developed to mount film within the DigiPhant PT. The film holder, shown in figure 1(b), was designed such that the effective points of measurement between the film and MatriXX were identical and reproducible. The film holder was machined to conform to the specifications of 8′ × 10′ sheets of EBT3, and ultimately presents a 17 × 20 cm2 cross-sectional area of film for measurement directly in water. Following irradiation, films were read out using an Epson 10000XL flatbed scanner following irradiation and converted to dose to water using a 13-point 6 MV calibration curve created using a Varian TrueBeam Linear Accelerator at the University of Wisconsin Medical Radiation Research Center. Because of the known under response of EBT3 film in high proton LET environments (Smith et al 2019b), the use of EBT3 film in this work was limited to relative dosimetry and not used to verify absolute doses. Following irradiation, both the MatriXX and film dose distributions underwent a 2D rigid optimization consisting of rotations and translations was performed to align the measured profiles with the simulated profiles.
3. Results
3.1. Monte Carlo penumbra sharpening
Following the beamlet weight optimization, penumbra widths (80%–20%) were tabulated as a function of target depth for the uncollimated and collimated plans. Figure 2 displays uncollimated and collimated lateral profiles for the C3-D6-NRS target plan (see table 2 for definition of plan name) and the Monte Carlo-simulated penumbra widths as a function of depth. For the NRS plans, penumbra reductions with the DCS ranged from 33% to 64%, or 2.5 to 5.5 mm, as the depth of the target decreases. For the RS plan at a depth of 5 cm, the penumbra was reduced by nearly 50%, or 4.2 mm.
Figure 2.

(a) Monte Carlo-simulated uncollimated (red, solid) and collimated (blue, dashed) lateral profiles at the center of the C3-D6-NRS target shown in black. (b) Simulated penumbra widths (uncollimated = red, solid; collimated = blue, dashed) are plotted as function of target depth on the left y-axis with percent change in penumbra width (black, solid) plotted on the right y-axis.
Table 2.
Summary of the simulated and measured penumbra widths (80%–20%) and penumbra width reductions with collimation for the plans and dosimeters investigated. The bottom row contains the average, absolute differences observed relative to the Monte Carlo values.
| Cube Depth (cm) | Uncollimated (mm) | Collimated (mm) | ||||
|---|---|---|---|---|---|---|
| MC | MatriXX | Film | MC | MatriXX | Film | |
| 6 | 8.4 | 8.6 | 7.7 | 3.0 | 4.4 | 2.7 |
| 10 | 7.6 | 8.0 | 7.2 | 3.9 | 5.0 | 3.8 |
| 15 | 7.5 | 7.9 | 7.2 | 5.0 | 5.9 | 5.1 |
| 15 (measured at 4.1 cm) | 5.7 | 6.2 | 5.4 | 1.9 | 3.8 | 1.4 |
| (mm) | — | 0.4 | 0.4 | — | 1.3 | 0.2 |
Figure 3 displays the simulated penumbra widths for three of the treatments investigated in this work under various trimmer-to-surface distances (TSDs). The penumbra for the range shifted plan, C3-D5-RS, illustrates a large dependency on the TSD, where an increase in the penumbra width of 3.6 mm was seen between the 5 and 18.15 cm TSDs. A very weak to no correlation between the collimated penumbra widths and the TSD was observed for the NRS plans.
Figure 3.

Bar chart of simulated penumbra widths for the C3-D5-RS, C3-D6-NRS, and C3-D15-NRS plans. Penumbra widths are shown for the uncollimated plans and the collimated plans under 5, 10, and 18.15 cm trimmer-to-surface distances (TSD).
3.2. Experimental validation
3.2.1. SOBPs
For the NRS plans shown in table 1, the depth of the distal 90% SOBP doses were verified to within 0.5 mm on average with a maximum difference of 0.7 mm and SOBP widths were within 1.5 mm on average with a max deviation of 1.7 mm.
3.2.2. MatriXX and film profiles
Figure 4 presents film and MatriXX lateral profile measurements for the uncollimated and collimated C3-D15-NRS plans at a depth of 4.1 cm. For the uncollimated case, excellent agreement is observed between simulation and measurements with 2D gamma (2%/2 mm/5% threshold) pass rates of 94.6% and 100% for comparisons with the MatriXX array and EBT3 film, respectively. When the MatriXX is used to measure the comparatively higher dose gradients for the collimated case, however, the pass rate is reduced to 83.7%, whereas film maintained a pass rate of 90.2%. MatriXX-measured penumbra widths (80%–20%) were 0.6 mm and 1.8 mm wider than Monte Carlo for the uncollimated and collimated dose distributions, respectively. For EBT3 film, these differences were reduced to 0.2 and 0.5 mm for the uncollimated and collimated plans, respectively.
Figure 4.

Monte Carlo, film, and MatriXX lateral profiles simulated and measured at a depth of 4.1 cm for the uncollimated (a) and collimated (b) 3 × 3 × 3 cm3 cube centered at a depth of 15 cm.
Figure 5 displays lateral dose profile results for measurements at the middle of the SOBPs for the uncollimated and collimated C3-D6-NRS, C3-D10-NRS, and C3-D15-NRS plans. Table 2 contains a comparison of the simulated and measured penumbra widths for the plans investigated. Uncollimated penumbra widths were determined to within 0.4 mm of Monte Carlo on average for both dosimeters while collimated penumbra widths were determined to within 1.3 and 0.2 mm for the MatriXX and EBT3 film, respectively. Penumbra reductions were measured to within 0.9 and 0.3 mm of Monte Carlo using the MatriXX and EBT3 film, respectively.
Figure 5.

Monte Carlo, film, and MatriXX lateral profiles simulated and measured at the middle of the SOBPs for the C3-D6-NRS (top row), C3-D10-NRS (middle row), and C3-D15-NRS (bottom row) for uncollimated (left column) and collimated (right column) plan deliveries.
Relative 2D gamma pass rates are shown for the NRS plans and dosimeters investigated in table 3. Overall good agreement is observed, with average pass rates exceeding 96% across all planes of measurements, regardless of collimation. For the measurements performed proximal to the C3-D15-NRS SOBP, slightly better agreement was observed with EBT3 film, a result that can corroborated by the improved penumbra width agreement shown in table 2.
Table 3.
Summary of relative 2D gamma pass rates between measurement (MatriXX and film) and simulation at a 2%/2 mm criterion with a 5% dose threshold applied on measured data points.
| Cube Depth (cm) | Uncollimated | Collimated | ||
|---|---|---|---|---|
| MatriXX | Film | MatriXX | Film | |
| 6 | 100 | 97.0 | 99.0 | 100 |
| 10 | 100 | 100 | 99.3 | 100 |
| 15 | 100 | 93.3 | 98.0 | 91.0 |
| 15 (measured at 4.1 cm) | 94.6 | 100 | 83.7 | 90.2 |
3.3. Surface dose enhancement
For the shallow measurements of the C3-D15-NRS plan, dose enhancements of 2.9 ± 1.6% and 1.3 ± 2.8% were observed using the MatriXX and EBT3 film. For the same plan, the Monte Carlo-simulated dose enhancement was 3.9 ± 0.4%, which is within the standard deviation of the MatriXX-measured values. To illustrate this surface dose enhancement effect, figure 6 displays lateral profiles for the C3-D5-RS and C3-D15-NRS plans at a depth of 5 mm. Table 4 summarizes the simulated dose enhancements at a depth of 5 mm for all plans investigated. In general, an increase in the dose enhancement was observed with increasing treatment depth.
Figure 6.

Simulated uncollimated and collimated lateral profiles at a depth of 5 mm for the C3-D5-RS and C3-D15-NRS plans. The arrows indicate the location of the evaluated dose points that were used to quantify the collimation-induced dose enhancement.
Table 4.
Monte Carlo-simulated dose enhancement at a depth of 5 mm for the plans investigated. Dose values are reported relative to the prescription dose and are the average of the max dose horns in each lateral dimension, for a total of four evaluated dose points per plan. Standard deviations are reported across the evaluated dose points. For the dose enhancement, collimated and uncollimated standard deviations were summed in quadrature.
| Cube depth (cm) | Uncollimated dose (%) | Collimated dose (%) | Enhancement [Coll—uncoll] (%) |
|---|---|---|---|
| 5 (RS) | 68.6 ± 0.1 | 71.1 ± 0.1 | 2.5 ± 0.1 |
| 6 | 62.6 ± 0.1 | 64.6 ± 0.3 | 2.0 ± 0.3 |
| 10 | 59.7 ± 0.03 | 62.9 ± 0.5 | 3.2 ± 0.5 |
| 15 | 60.8 ± 0.2 | 67.6 ± 0.9 | 6.9 ± 0.9 |
4. Discussion
This work is the first to experimentally validate the penumbra reduction capabilities of the DCS equipped to a clinical proton therapy system using set of MU-optimized pencil beams. In the future, we aim to integrate the DCS control system with the beam scanning controller system to enable the delivery of treatment fields that require multiple trimmer positions in an energy layer rather than the single trimmer position per energy layer shown in this work. This work is a crucial step in the development of the DCS where the simulated penumbra sharpening capabilities of the DCS were compared to direct measurements of uncollimated and collimated treatment fields. This work has also served as an overall validation of the Monte Carlo planning and optimization techniques that were used to create the multi-spot treatment fields.
4.1. Monte Carlo simulations
For NRS plans, penumbra reductions ranged from 64% to 33% with increasing depth while the RS plan yielded penumbra reductions of around 50% at a depth of 5 cm in water. In general, uncollimated penumbral widths were below 9 mm and decreased with increasing depth, an effect that can be attributed to the reduced spot sizes and scatter cross sections for higher energy beams used to treat at deeper depths. For the collimated plans, an increase in the penumbra is evident with the inclusion of the range shifter (C3-D5-RS). This can be attributed to scattered protons originating from the range shifter that subsequently transmit through portions of the DCS trimmers. Once the range shifter is removed for the 6 cm depth and beyond, penumbra widths were all below 5.2 mm and found to linearly increase with depth at a rate of 0.23 mm per cm of water (R 2 = 1). The surface dose enhancement with the DCS was also observed to be no greater than 4.7% at a reference depth of 2 cm in water across all plans investigated.
Variation in TSD was found to impact the dose distribution only for fields that utilized the range shifter. This effect can be attributed to large-angle scattered protons originating from that range shifter that result in subsequent partial trimmer transmission, ultimately reducing penumbra widths on a per-beamlet basis, an effect that has been previously addressed in the context of analytical modeling of the DCS (Nelson et al 2022). These results also indicate a near-negligible influence of in-air scatter on the achievable penumbra width for DCS-collimated treatments over the range of TSDs evaluated.
4.2. Experimental validation
Experimental validation in this work consisted of measurements of the SOBP distributions and lateral profile measurements of the NRS plans. The Zebra MLIC was used to measure SOBPs, where the depths of the distal 90% dose point and SOBP widths were verified to within 0.5 and 1.5 mm, respectively. These results agree well with those previously reported by Dhanesar et al (2013), who benchmarked the performance of the MLIC against scanned ionization chamber measurements performed in water. Disagreements between measured and simulated SOBPs were observed in the plateau region and are a result of variation in the stopping power ratio between MLIC-equivalent material and water near the end of range (Nelson et al 2021b).
The MatriXX ionization chamber array and EBT3 film were used to measure lateral profiles at the middle of the SOBP for all plans and proximal to the SOBP of the C3-D15-NRS plan. Proximal measurements of this higher energy plan were performed to assess detector performance under the extremely steep dose gradients present at this depth, where the simulated penumbra width of the collimated plans was 1.9 mm, yielding a dose gradient of over 30% per mm. For lateral profile measurements at the middle of the SOBPs, overall excellent agreement was observed between Monte Carlo, MatriXX, and EBT3 film using 2D gamma analyses. For measurements performed proximal to the C3-D15 SOBP, EBT3 film agreed best with Monte Carlo in terms of 2D gamma analysis and penumbra width, whereas the MatriXX exhibited a significant amount of volume averaging as shown in figure 4(b) resulting in a lower gamma pass rate and wider penumbra width of greater than 2 mm. Considering clinical patient specific dosimetry is often performed within the SOBP using ionization chamber arrays, these findings support the use of the MatriXX for verification of collimated penumbrae within the SOBP for dose gradients up to 20% per mm, as seen in the C3-D6-NRS plan. For relative dose verification proximal to SOBPs, or dose gradients that exceed 20% per mm, the use of a high-resolution 2D dosimeter such as EBT3 film is recommended.
The surface dose enhancement with the DCS was also found to be small (<7%) at a depth of 5 mm for all plans and located at the lateral edges of the targets downstream from the trimmers. The surface dose enhancement with collimation arises from an increased surface dose on a per-beamlet basis that results from low-energy scattered protons penetrating to shallow depths in the phantom. This effect is also more pronounced due to the increased weighting placed on these beamlets that is required to achieve identical target coverage for the uncollimated case. Although the impact of scattered protons is relatively small for the cases investigated in this work, it is necessary to model in the context of a treatment planning system, where treatments that utilize larger, more complex sets of spot and trimmer positions will be calculated.
5. Conclusions
Predictions of collimated penumbra sharpening, previously limited to theoretical models and Monte Carlo simulation, have been confirmed by experimental measurements for a set of treatment plans that consider a range of scenarios relevant to expected clinical use of the DCS. The use of dynamic collimation can reduce penumbra by up to 60% as compared to deliveries with uncollimated spots, and the effect of the air gap on the achievable spot collimation is negligible if a range shifter is absent. Ion chamber array measurements are sufficient to characterize collimated gradient penumbrae in the SOBP for the set of plans measured, however higher resolution detectors such as film are required in proximal regions where gradients were more pronounced. Additional work is underway to determine the appropriateness of these detectors for more complicated treatment plans and spot patterns.
Acknowledgments
Research reported in this publication was supported by the National Cancer Institute of the National Institutes of Health under Award Number R37CA226518. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. This research was performed using the compute resources and assistance of the UW-Madison Center For High Throughput Computing (CHTC) in the Department of Computer Sciences. The CHTC is supported by UW-Madison, the Advanced Computing Initiative, the Wisconsin Alumni Research Foundation, the Wisconsin Institutes for Discovery, and the National Science Foundation, and is an active member of the OSG Consortium (Pordes et al 2007, Sfiligoi et al 2009), which is supported by the National Science Foundation and the U.S. Department of Energy’s Office of Science.
Conflict of interest.
DEH, RTF, and PMH are co-inventors on a patent that has been licensed to IBA.
References
- Arjomandy B, Sahoo N, Ciangaru G, Zhu R, Song X, Gillin M. Verification of patient-specific dose distributions in proton therapy using a commercial two-dimensional ion chamber array. Med. Phys. 2010;37:5831–7. doi: 10.1118/1.3505011. [DOI] [PubMed] [Google Scholar]
- Arjomandy B, Sahoo N, Ding X, Gillin M. Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance. Med. Phys. 2008;35:3889–94. doi: 10.1118/1.2963990. [DOI] [PubMed] [Google Scholar]
- Bäumer C, Koska B, Lambert J, Timmermann B, Mertens T, Talla P T. Evaluation of detectors for acquisition of pristine depth-dose curves in pencil beam scanning. J. Appl. Clin. Med. Phys. 2015;16:151–63. doi: 10.1120/jacmp.v16i6.5577. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bues M, Newhauser W D, Titt U, Smith A R. Therapeutic step and shoot proton beam spot-scanning with a multi-leaf collimator: A Monte Carlo study. Radiat. Prot. Dosimetry. 2005;115:164–9. doi: 10.1093/rpd/nci259. [DOI] [PubMed] [Google Scholar]
- Chan M F, et al. Patient-specific QA of spot-scanning proton beams using radiochromic film. Int. J. Med. Phys. Clin. Eng. Radiat. Oncol. 2017;06:111–23. doi: 10.4236/ijmpcero.2017.62011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chiang B H, Bunker A, Jin H, Ahmad S, Chen Y. Developing a monte carlo model for MEVION S250i with HYPERSCAN and adaptive APERTURETM pencil beam scanning proton therapy system. J. Radiother. Pract. 2021;20:279–86. doi: 10.1017/S1460396920000266. [DOI] [Google Scholar]
- Daartz J, Bangert M, Bussìre M R, Engelsman M, Kooy H M. Characterization of a mini-multileaf collimator in a proton beamline. Med. Phys. 2009;36:1886–94. doi: 10.1118/1.3116382. [DOI] [PubMed] [Google Scholar]
- Daartz J, Maughan R L. The disadvantages of a multileaf collimator for proton radiotherapy outweigh its advantages. Med. Phys. 2014;41:1–3. doi: 10.1118/1.4824437. [DOI] [PubMed] [Google Scholar]
- Dhanesar S, et al. Quality assurance of proton beams using a multilayer ionization chamber system. Med. Phys. 2013;40:1–9. doi: 10.1118/1.4817481. [DOI] [PubMed] [Google Scholar]
- Flynn R. ProQuest Diss Theses. Published online. 2007. Modulated proton therapy for the delivery of non-uniform dose. [Google Scholar]
- Geoghegan T, et al. Characterization of an isocenter alignment device for the proton PBS dynamic collimation system. Med. Phys. 2022;49:e178. [Google Scholar]
- Geoghegan T J, Nelson N P, Flynn R T, Hill P M, Rana S, Hyer D E. Design of a focused collimator for proton therapy spot scanning using Monte Carlo methods. Med. Phys. 2020;47:2725–34. doi: 10.1002/mp.14139. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gillin M T, et al. Commissioning of the discrete spot scanning proton beam delivery system at the university of texas m.d. anderson cancer center, proton therapy center, houston. Med. Phys. 2010;37:154–63. doi: 10.1118/1.3259742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Grewal H S, Ahmad S, Jin H. Characterization of penumbra sharpening and scattering by adaptive aperture for a compact pencil beam scanning proton therapy system. Med. Phys. 2021;48:1508–19. doi: 10.1002/mp.14771. [DOI] [PubMed] [Google Scholar]
- Hyer D E, Hill P M, Wang D, Smith B R, Flynn R T. A dynamic collimation system for penumbra reduction in spot-scanning proton therapy: proof of concept. Med. Phys. 2014a;41:1–9. doi: 10.1118/1.4837155. [DOI] [PubMed] [Google Scholar]
- Hyer D E, Hill P M, Wang D, Smith B R, Flynn R T. Effects of spot size and spot spacing on lateral penumbra reduction when using a dynamic collimation system for spot scanning proton therapy. Phys. Med. Biol. 2014b;59:N187–96. doi: 10.1088/0031-9155/59/22/N187. [DOI] [PubMed] [Google Scholar]
- Andreo P. IAEA TRS-398 Absorbed dose determination in external beam radiotherapy: an International code of practice for dosimetry based on standards of absorbed dose to water. International Atomic Energy Agency. 2000;18:35–6. [Google Scholar]
- Islam M R, Zheng Y, Collums T L, Monson J M, Ahmad S, Benton E R. Measurement and simulation of secondary neutrons from uniform scanning proton beams in proton radiotherapy. Radiat. Meas. 2017;96:8–18. doi: 10.1016/j.radmeas.2016.11.008. [DOI] [Google Scholar]
- Kang M, Pang D. Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system. Med. Phys. 2020;47:3496–510. doi: 10.1002/mp.13972. [DOI] [PubMed] [Google Scholar]
- Kralik J C, Xi L, Solberg T D, Simone C B, Lin L. Comparing proton treatment plans of pediatric brain tumors in two pencil beam scanning nozzles with different spot sizes. J. Appl. Clin. Med. Phys. 2015;16:41–50. doi: 10.1120/jacmp.v16i6.5389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Langner U W, Eley J G, Dong L, Langen K. Comparison of multi-institutional Varian ProBeam pencil beam scanning proton beam commissioning data. J. Appl. Clin. Med. Phys. 2017;18:96–107. doi: 10.1002/acm2.12078. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lin L, et al. Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance. J. Appl. Clin. Med. Phys. 2015;16:270–6. doi: 10.1120/jacmp.v16i3.5323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lomax A. Intensity modulation methods for proton radiotherapy. Phys. Med. Biol. 1999;44:185–205. doi: 10.1088/0031-9155/44/1/014. [DOI] [PubMed] [Google Scholar]
- Maes D, et al. Parametric characterization of penumbra reduction for aperture-collimated pencil beam scanning (PBS) proton therapy. Biomed. Phys. Eng. Express. 2019;5:35002. doi: 10.1088/2057-1976/ab0953. [DOI] [Google Scholar]
- Moignier A, et al. Toward improved target conformity for two spot scanning proton therapy delivery systems using dynamic collimation. Med. Phys. 2016a;43:1421–7. doi: 10.1118/1.4942375. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Moignier A, et al. Theoretical benefits of dynamic collimation in pencil beam scanning proton therapy for brain tumors: dosimetric and radiobiological metrics. Int. J. Radiat. Oncol. Biol. Phys. 2016b;95:171–80. doi: 10.1016/j.ijrobp.2015.08.030. [DOI] [PubMed] [Google Scholar]
- Moignier A, et al. Improving head and neck cancer treatments using dynamic collimation in spot scanning proton therapy. Int. J. Part Ther. 2016c;2:544–54. doi: 10.14338/IJPT-15-00026.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Moteabbed M, Yock T I, Depauw N, Madden T M, Kooy H M, Paganetti H. Impact of spot size and beam-shaping devices on the treatment plan quality for pencil beam scanning proton therapy. Int. J. Radiat. Oncol. Biol. Phys. 2016;95:190–8. doi: 10.1016/j.ijrobp.2015.12.368. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nelson N P, et al. Development and validation of the dynamic collimation monte carlo simulation package for pencil beam scanning proton therapy. Med. Phys. 2021a;48:3172–85. doi: 10.1002/mp.14846. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nelson N P, Culberson W S, Hyer D E, Smith B R, Flynn R T, Hill P M. Investigating aperture-based approximations to model a focused dynamic collimation system for pencil beam scanning proton therapy. Biomed. Phys. Eng. Express. 2022;8:1–13. doi: 10.1088/2057-1976/ac525f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nelson N P, Culberson W S, Smith B R, Flynn R T, Hyer D E, Hill P M. Characterizing the entrance dose induced by a dynamic collimation system for pencil beam scanning proton therapy using a commercial multilayer ionization chamber. Med. Phys. 2021b;48:e542. [Google Scholar]
- Perl J, Shin J, Schümann J, Faddegon B, Paganetti H. An innovative proton Monte Carlo platform for research and clinical applications. Med. Phys. 2012;39:6818–37. doi: 10.1118/1.4758060. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pordes R, et al. The open science grid. J. Phys. Conf. Ser. 2007;78:1–15. doi: 10.1088/1742-6596/78/1/012057. [DOI] [Google Scholar]
- Rahman M, Bruza P, Lin Y, Gladstone D J, Pogue B W, Zhang R. Producing a beam model of the varian probeam proton therapy system using TOPAS monte carlo toolkit. Med. Phys. 2020;47:6500–8. doi: 10.1002/mp.14532. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sfiligoi I, Bradley D C, Holzman B, Mhashilkar P, Padhi S, Würthwein F. The pilot way to Grid resources using glideinWMS. 2009. WRI World Congr Comput Sci Inf Eng CSIE 2009. 2009;2:428–32. doi: 10.1109/CSIE.2009.950. [DOI] [Google Scholar]
- Shen J, et al. Impact of range shifter material on proton pencil beam spot characteristics. Med. Phys. 2015;42:1335–40. doi: 10.1118/1.4908208. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith B, et al. Technical note: a treatment plan comparison between dynamic collimation and a fixed aperture during spot scanning proton therapy for brain treatment. Med. Phys. 2016;43:4693–9. doi: 10.1118/1.4955117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith B R. ProQuest Diss Theses. Published online. 2019. Towards dynamic collimation in pencil beam scanning proton radiotherapy. [Google Scholar]
- Smith B R, Hyer D E, Hill P M, Culberson W S. Secondary neutron dose from a dynamic collimation system during intracranial pencil beam scanning proton therapy: a monte carlo investigation. Int. J. Radiat. Oncol. Biol. Phys. 2019a;103:241–50. doi: 10.1016/j.ijrobp.2018.08.012. [DOI] [PubMed] [Google Scholar]
- Smith B R, Pankuch M, Hammer C G, DeWerd L A, Culberson W S. LET response variability of GafchromicTM EBT3 film from a 60Co calibration in clinical proton beam qualities. Med. Phys. 2019b;46:2716–28. doi: 10.1002/mp.13442. [DOI] [PubMed] [Google Scholar]
- Smith B R, Pankuch M, Hyer D E, Culberson W S. Experimental and Monte Carlo characterization of a dynamic collimation system prototype for pencil beam scanning proton therapy. Med. Phys. 2020;47:5343–56. doi: 10.1002/mp.14453. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tian X, Liu K U N, Hou Y, Cheng J, Zhang J. The evolution of proton beam therapy : current and future status (review) Molecular and Clinical Oncology. 2018;8:15–21. doi: 10.3892/mco.2017.1499. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Titt U, et al. Adjustment of the lateral and longitudinal size of scanned proton beam spots using a pre-absorber to optimize penumbrae and delivery efficiency. Phys. Med. Biol. 2010;55:7097–106. doi: 10.1088/0031-9155/55/23/S10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vilches-Freixas G, et al. Beam commissioning of the first compact proton therapy system with spot scanning and dynamic field collimation. Br. J. Radiol. 2020;93:1–9. doi: 10.1259/bjr.20190598. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang D, et al. Impact of spot size on plan quality of spot scanning proton radiosurgery for peripheral brain lesions. Med. Phys. 2014;41:1–10. doi: 10.1118/1.4901260. [DOI] [PubMed] [Google Scholar]
- Winterhalter C, Lomax A, Oxley D, Weber D C, Safai S. A study of lateral fall-off (penumbra) optimisation for pencil beam scanning (PBS) proton therapy. Phys. Med. Biol. 2018;63:aaa2ad. doi: 10.1088/1361-6560/aaa2ad. [DOI] [PubMed] [Google Scholar]