Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2023 Feb 22.
Published in final edited form as: Nat Biomed Eng. 2022 Aug 15;6(9):1017–1030. doi: 10.1038/s41551-022-00913-2

Non-invasive monitoring of blood oxygenation in human placentas via concurrent diffuse optical spectroscopy and ultrasound imaging

Lin Wang 1,*, Jeffrey M Cochran 1, Tiffany Ko 1,3, Wesley B Baker 1,3, Kenneth Abramson 1, Lian He 1, David R Busch 1,4, Venki Kavuri 1, Rebecca L Linn 5, Samuel Parry 2, Arjun G Yodh 1, Nadav Schwartz 2
PMCID: PMC9944515  NIHMSID: NIHMS1862069  PMID: 35970929

Abstract

Direct assessment of blood oxygenation in the human placenta can provide information about placental function. However, the monitoring of placental oxygenation involves invasive sampling or imaging techniques that are poorly suited for bedside use. Here we show that placental oxygen haemodynamics can be non-invasively probed, in real time, up to 4.2 cm below the body surface via concurrent frequency-domain diffuse optical spectroscopy and ultrasound imaging. We developed the multimodal instrument to facilitate assessment of anterior placental properties by leveraging image-reconstruction algorithms that integrate ultrasound information about the morphology of tissue layers with optical information on haemodynamics. In a pilot investigation involving pregnancies in the third trimester, we studied women with normal (15) or some type of abnormal (9) placental function. We found no significant differences in baseline haemoglobin properties, but we found statistically significant differences in the placental haemodynamic responses to maternal hyperoxia. The results suggest that non-invasive monitoring of placental oxygenation may aid the early detection of placenta-related adverse pregnancy outcomes and maternal vascular malperfusion.

Editorial summary

Concurrent frequency-domain diffuse optical spectroscopy and ultrasound imaging can be used to non-invasively monitor placental oxygen haemodynamics in pregnant women in real time.


Abnormal placental development is widely accepted as the cause of common adverse pregnancy outcomes (APOs) such as hypertensive disorders, foetal growth restriction and stillbirth. Moreover, placental dysfunction has been associated with morbidities in offspring, including perinatal mortality and long-term neurodevelopmental and cardiovascular consequences13. To better understand human placental dysfunction associated with these APOs, non-invasive methods that measure placental oxygen dynamics are needed. Ultrasound imaging is the primary clinical modality used for assessing pregnancy. Although ultrasound imaging can provide indirect information about flow resistance in large blood vessels, it is used primarily to derive morphological information. In fact, clinical ultrasound guidelines do not include direct functional assessment of the placenta4.

Current knowledge about placental biology has been gleaned largely from ex vivo tissue and from animal research5,6. Yet these models have limitations for the understanding of dynamical changes in placental pathophysiology during pregnancy7,8. Similarly, the literature on placental oxygenation is derived largely from decades-old sheep studies and from scant human data using invasive sampling techniques that have yielded varying results7,9. And magnetic resonance imaging (MRI) approaches for assessing placental oxygenation10,11 are poorly suited for bedside monitoring, and generally rely on indirect signals12.

Here we report the development of an instrument and methodology based on diffuse optical spectroscopy (DOS) and ultrasound that facilitates the measurement of oxygen haemodynamics in complex organs such as the placenta, which is buried far below the tissue surface amidst intervening layered tissues. DOS measures oxy-haemoglobin and deoxy-haemoglobin concentrations, and it has been successfully employed for the assessment of tissue haemodynamics1321 in clinical problems such as breast cancer diagnosis and therapy monitoring19,22, brain function23 and injury monitoring17. In most of these applications, however, reflected light penetration was limited to less than 2 cm below the surface24,25. The new instrumentation and algorithms that we here report provide improvements in methodology needed to measure the oxygen haemodynamics of the anterior placenta amidst intervening heterogeneous tissue layers. Our human clinical measurements probed placental tissue located as deep as 4.2 cm below the surface, albeit more typically in the 2.3–3.3-cm range. These improvements enable continuous data collection for functional studies of the placenta at the bedside, and may create opportunities for the investigation of haemodynamics in other deep organs.

Continuous-wave diffuse optical spectroscopy (CW-DOS) has been explored for the non-invasive measurement of placental blood oxygenation2628, and some of this early research suggested that higher placental tissue oxygen at baseline can arise in patients with intrauterine growth restriction26,28. Although encouraging, these early measurements had limitations. The instruments used comparatively short source-detector separations (~4 cm) on the tissue surface, which limited the penetration of light to ~2 cm. Furthermore, the analysis of CW data requires major assumptions about tissue homogeneity and tissue scattering that prevent the quantification of absolute oxy-haemoglobin and deoxy-haemoglobin concentrations29 and that do not account for the layered morphology of the abdomen. This early work also pointed to clear avenues for improvement. Our work benefits from a more accurate implementation of light-transport models1315,30, and from sophisticated frequency-domain and time-domain (FD-DOS, TD-DOS) optical instrumentation that permits the relaxation of assumptions about tissue scattering and homogeneity13,14,31. Tissue-phantom experiments indicate signal-to-noise ratios (SNRs) sufficient to permit source-detector separations (SDSs) as large as 10 cm, which could enable light penetration of ~5 cm, thus improving on previous SDS records3234. Importantly, the optical instrument is integrated with ultrasound imaging in the same probe head22. This multimodal FD-DOS/US combination facilitates the integration of anatomic ultrasound information about tissue-layer morphology with functional haemodynamic information about deep tissues from FD-DOS. The anatomic information enables tissue-specific and layered image reconstruction that separates the haemodynamic properties and responses of deep tissues, such as the placenta, from those of overlying layers.

We validated the methodology in layered tissue phantoms, and show its feasibility and utility by direct in vivo assessment of human placental oxygenation in 24 participants. Specifically, we measured placental oxy-haemoglobin HbO2 and deoxy-haemoglobin Hb concentrations or, equivalently, total haemoglobin HbT concentration and oxygen saturation (StO2). We performed reproducibility and stability tests to characterize the technology. We also collected average tissue properties from each woman. We also show the detection of changes in dynamic placental oxygenation in maternal hyperoxia experiments by varying the maternal position. Notably, our pilot study shows that placental oxygen haemodynamics during maternal hyperoxia are significantly associated with placenta-related APOs and with placental maternal vascular malperfusion (MVM), a primary histopathologic pattern characteristic of placental dysfunction strongly associated with APO and with the risk of long-term disease35,36. Our findings suggest the possibility of the non-invasive detection of placental dysfunction for the generation of improved clinical understanding of placental pathophysiology in vivo.

Results

FD-DOS instrumentation

To measure placental oxygen haemodynamics in vivo, we built a low-noise heterodyne instrument for frequency-domain diffuse optical spectroscopy (FD-DOS). Tissue-phantom experiments showed that the instrument has a sufficient dynamic range and SNR to perform accurate FD-DOS measurements at SDSs of 10 cm. In the clinic, these capabilities facilitated quantitative determination of HbO2, Hb and HbT concentrations as well as StO2 of anterior placental tissue located as deep as 4.2 cm below the skin surface. Fig. 1a shows key features of the custom heterodyne FD-DOS instrument (details in Methods). It employs three lasers with wavelengths of 785, 808 and 830 nm. The output of each laser is radiofrequency (RF) amplitude-modulated at f1=100 MHz. A new technical feature of the instrument is its nearly 100% laser-modulation depth. To achieve this improvement, we divided the source driver signal into four sub-signals, amplified each sub-signal in multiple stages with low-noise linear amplifiers, and then recombined and impedance-matched the amplified sub-signals for input to the laser drivers. Each laser’s RF driver power was individually optimized to achieve >90% light-modulation depth, thereby increasing the modulated diffusive-wave amplitude and decreasing the (unmodulated) background diffuse light. As a result, measurement SNRs were significantly better than in previous work32,33 (specifically, by more than 20dB for an SDS of 8 cm), enabling longer distance SDS measurements with low laser powers (~40 mW).

Fig. 1 |. Integrated FD-DOS / US placenta instrumentation and three-layer modelling.

Fig. 1 |

a, Schematic of the hybrid FD-DOS/US instrumentation showing laser-amplitude-modulation electronics, tissue light transmission, PMT detection, heterodyne mixing for frequency down-conversion, lock-in detection of light wave amplitude/phase, simultaneous ultrasound-image segmentation, and optical-image reconstruction of placental haemoglobin properties. b, 3D rendering and bottom-view of the probe showing FD-DOS/US integration into a single-probe head. c, Ultrasound image showing anatomical ‘segmented’ regions and the corresponding three-layer model for optical-image reconstruction.

The optoelectronic components were fibre-coupled into a custom optical probe head, within which a commercial ultrasound probe was mounted. Optical source fibres in this probe offer 17 SDSs for measurements ranging between 1 and 9 cm in humans (Fig. 1b). During measurements in the participants, 10 source-fibre locations are chosen to optimize coverage over the anatomic regions of interest for each woman, and we scanned sequentially through them. At the end of each cycle, a dark count measurement is made to correct for systemic noise. A high-transmission liquid light guide (detector fibre) with 5 mm in core diameter collects and directs light to a photomultiplier tube (PMT) detector. The PMT electrical signal is mixed with another RF wave at f2=100.2 MHz to generate heterodyne down-converted signals Δf=f1f2=0.2 MHz. A high-sampling-rate lock-in amplifier captures amplitude and phase of the diffuse light waves.

Concurrent optical and ultrasound imaging

Custom integration of optics with a commercial ultrasound system (9L-D probe, Voluson E10, GE Healthcare) provides another substantial technical improvement over prior work37. The custom probe facilitates concurrent measurement of tissue layer morphology and tissue physiological properties. This concept has been employed in breast cancer research22, but much deeper light penetration is required for the placenta. The ultrasound transducer at the probe’s centre generates images (Fig. 1c) that we used to segment target tissue into distinct layers that constrain optical reconstruction algorithms.

Differently from prior placenta research2628, we used tissue layer morphology from ultrasound imaging to constrain the photon-diffusion tomographic inverse problem. In practice, we model the abdomen as three-layers: adipose, rectus/uterus, and placental tissue. We approximate each layer as homogeneous and laterally infinite, but with thickness and depth determined by ultrasound (Fig. 1c).

The measurement geometry and modelling are shown in Fig. 1c. ρ is the SDS on the tissue surface. With measured layer thicknesses and optical and physiological properties for each layer as inputs, standard methods38,39 can be readily employed to generate predictions for the detected light-fluence rate on the tissue surface. Fig. 2a outlines our three-step reconstruction procedure (details in Methods). Briefly, each step of the three-step reconstruction finds ‘best’ tissue properties by minimizing the differences between the measured data and the predictions of diffuse optical tissue models of increasing complexity.

Fig. 2 |. Three-layer-model reconstruction algorithm and phantom-validation experiments.

Fig. 2 |

a, Three-layer-model initialization and reconstruction procedures. b, Tissue-simulating-phantom experiment for validating the accuracy and depth sensitivity of the optical properties. A translation stage moves the source fibre across the phantom for measurements at SDSs ranging from 6.2 cm to 10 cm. Nonlinear semi-infinite fitting is performed to reconstruct phantom absorption and scattering coefficients λ=830 nm. c, Schematic of a two-layer phantom experiment.

Step 1 assumes that the underlying tissue is semi-infinite and homogeneous, and employs the simplest analytical model for optical-property reconstruction. Step 2 used solutions from Step 1 as initial guesses in a two-layer diffuse optical tissue reconstruction. Step 3 uses solutions from Step 2 as initial guesses in a three-layer diffuse optical tissue reconstruction. In all steps, layer thicknesses are fixed based on ultrasound imaging, but other tissue properties are permitted to vary to determine the best physiological and optical property solutions for each layer. The 3-step reconstruction approach provides accurate determination of the optical haemoglobin properties for each layer while maintaining reasonable processing time. Moreover, the three-step algorithm helps to prevent the fit-search from becoming trapped in local solution minima.

Importantly, the image reconstructions rely on the simultaneous fitting of data from all SDSs and all wavelengths. This multispectral multi-SDS approach builds global constraints directly into the inverse problem and is critical for robust fitting40. To avoid reconstruction overfitting, Tikhonov regularization is employed to reduce ill-posedness (fitting and regularization procedures are detailed in Methods).

Validation and characterization with tissue-simulating phantoms

We first characterized the performance of the FD-DOS instrument and its SNR by using homogeneous liquid phantoms comprised of ink and 20% Intralipid (Baxter) (Fig. 2b); the phantom’s optical properties are known based on ink concentration (for absorption) and Intralipid concentration (for scattering)41. In the first study, fittings based on semi-infinite homogeneous solutions of the diffusion equation were employed to reconstruct the phantom’s optical properties (that is, the absorption coefficient, μa cm1) and the reduced scattering coefficient μs cm1)13. The data show good SNR at SDSs up to 10 cm (Fig. 2b), wherein the mean signal intensity was 28 times greater than the standard deviation of the measured intensity. The reconstructed optical properties at each wavelength had accuracies of 3%–9%.

In two-layer phantom experiments, a solid phantom with fixed optical properties was positioned inside a liquid phantom, and the optical probe (Fig. 1b) was placed on the liquid surface (Fig. 2c). The optical properties of the liquid phantom are known based on ink and Intralipid concentrations. The optical properties of the solid phantom were provided by the phantom manufacturer (INO). An absorption-titration experiment evaluated the instrument’s sensitivity to the absorption coefficient, holding the overlayer liquid phantom thickness (3 cm) constant while incrementally increasing the absorption coefficient of the top layer. A depth-changing experiment tested the sensitivity to superficial-layer thickness; here, the liquid phantom had fixed optical properties, and the superficial-layer thickness was increased from 1.5 cm to 3.0 cm. A deep two-layer phantom experiment verified the instrument’s ability to extract deep-layer optical properties for a superficial layer thickness of 4.3 cm. Each measurement was repeated three times; the resultant means and standard deviations are reported in Supplementary Table 1. These experimental results from tissue-simulating phantoms show that the instrument and algorithms extract deep-layer optical properties accurately with errors <10% in absorption and <15% in scattering. Semi-infinite fitting also produces an estimate of the optical properties, but as a weighted average of both layers and with a stronger weighting of superficial layers.

Validation with finite-element simulations

We generated simulated experimental data using a finite-element simulation tool (TOAST)42 that facilitated the creation of a three-layer model with segmented optical properties based on the layer morphology extracted from a participant’s ultrasound image (Fig. 1c). Since the participant’s layer interfaces are curved, we generated test data from curved layer interfaces (Fig. 1c, left). For the inverse problem, however, we assumed each layer interface to be flat (Fig. 1c, right). We carried out computer-simulation studies with simulated experimental data (Supplementary Table 2) based on the range of participant-derived haemoglobin optical properties observed in the clinical studies (that is, we employed median and upper (lower) 75th (25th) percentile properties from Supplementary Table 3). The three-layer reconstructions (Supplementary Table 2) produced good estimates for mean StO2, HbT, and HbO2 in each layer, with errors of 18% and more typically 10%. By contrast, the semi-infinite fitting generates a weighted average of the optical properties of all three layers and is more sensitive to superficial layers; its placental layer errors are comparatively large; 30–42%. The data presented in Supplementary Table 2 compares semi-infinite fitting to three-layer reconstruction for a moderate range of variation of placental HbO2, and a moderate range of variation of HbT and StO2 in the overlayers. The four datasets in the Supplementary Table 2 may be interpreted as four different participants with different haemoglobin properties. The three-layer reconstruction predicts placental properties with significantly increased accuracy compared to semi-infinite fitting for all participants. Specifically, by comparing the three-layer reconstructions, estimates of HbT and HbO2 based on the semi-infinite fitting are systematically worse, with ~3-fold greater percent-error. Percent-errors in semi-infinite fitting estimates of StO2 across these participants are similar or greater than the three-layer reconstruction estimates. Additionally, it is important to appreciate that semi-infinite fittings are not typically used; they benefit from data at very large SDSs, which is made possible by our instrumentation. The simulated datasets can also be paired to compare two different dynamic conditions within the same participant, such as would arise in response to hyperoxia versus baseline. For this comparison, the three-layer reconstruction produces good estimates of HbT and HbO2 and their changes, but the semi-infinite fittings of placental HbT and HbO2 dynamics are significantly underestimated.

Placental oxygen dynamics in vivo

We performed a pilot clinical study of human placental oxygen-related haemodynamic properties (the study design is detailed in Methods). The study enrolled women with singleton pregnancies in their third trimester and with anterior placentas. The central region of the placenta was targeted for monitoring. Adipose, rectus/uterus, and placenta layers were characterized by ultrasound and FD-DOS.

Four experiments were conducted: (1) a reproducibility experiment (n = 18); (2) a stability experiment (n=24); (3) a maternal left tilt experiment (n = 3); (4) a maternal hyperoxia experiment (n = 24). The women in the hyperoxia study had a median (interquartile range (IQR)) gestational age of 34.5 (32.9, 35.4) weeks and pre-gravid BMI (body mass index) of 27.7 (24.6, 30.2). The median (IQR) depth of the anterior placentas was 2.7 (2.3, 3.3) cm beneath the skin surface, and the nulliparity of the participants was 29.2%. Women in the maternal tilt study had gestational age of 37.0 (35.1, 37.3) weeks and pre-gravid BMI of 30.4 (29.2, 31.7); the median (IQR) depth of these anterior placentas was 2.3 (1.9, 2.4) cm beneath the skin surface.

Reproducibility was assessed by calculating the intra-class correlation coefficient (ICC) for 3 repeated measurements at the same placental location in 18 participants. Placental StO2, HbT, and HbO2 were highly repeatable within participants (ICC ≥ 0.9).

Stability was assessed from the standard deviation (S.D.) of data derived during a continuous 10-frame (3.5 minutes) baseline measurement at the same location on each participant. This single-participant S.D. was then averaged over all 24 participants to derive a mean measurement S.D., which served as our stability criteria. The resulting mean values of S.D. for placental StO2, HbT and HbO2 were small (2.2%, 0.9 μM, and 0.8 μM, respectively). Optically derived haemodynamic properties were stable at baseline (Fig. 6a), and ≥ 88% of the 24 stability measurements had S.D. for placental StO2, HbT and HbO2 of less than 3%, 2.0 μM, and 1.5 μM, respectively (Fig. 6b).

Fig. 6 |. Static (baseline) and dynamic (during maternal hyperoxia) placental haemoglobin properties for participants with NPP or MVM.

Fig. 6 |

a, Baseline placental StO2, HbT, HbO2 illustrate that they are not associated with placental pathology. The data show no significant differences between NPP and MVM. The p-values are 0.83, 0.69, and 0.69 respectively. b, Placental ΔStO2, ΔHbT and ΔHbO2 during maternal hyperoxia.ΔStO2, ΔHbT and ΔHbO2 exhibit clear separation, no difference, and clear separation, respectively, amongst participants with NPP versus MVM. The p-values are 0.05, 0.15, and 0.01, respectively. In these boxplots, the red centre line denotes the median value, and the blue box shows the IQR (25% to 75% of dataset). The blue whiskers mark the non-outlier minimum and non-outlier maximum; outliers are marked with symbol ‘+’. Values of the median and IQR are presented in Table 1. p-values were calculated via a two-sided Wilcoxon rank sum test.

The maternal left tilt experiment is sensitive to positional changes in cardiac output and utero-placental perfusion, which can lead to an increase in maternal cardiac output of up to 20% (ref. 43). We measured placental haemoglobin properties of 3 participants in the supine position, and then had them tilt to the left lateral position without removing the probe. The averaged relative increases in StO2, HbT and HbO2 were 2.4%, 8.4% and 10.9%, respectively. Mean values of StO2, HbT, and HbO2 for each participant before and after the maternal tilt are shown in Fig. 3c. Although the number of participants is small, we observed a trend; HbT and HbO2 increased in the lateral position (p = 0.05 and p = 0.02, respectively, two-sided paired sample t-test), suggestive of an accompanying increase in placental perfusion with oxygenated maternal blood.

Fig. 3 |. Placental haemoglobin properties during stability-test measurements and the maternal left-tilt experiment.

Fig. 3 |

a, Mean of placental StO2, HbT and HbO2 versus gestational stage (GA) derived from the stability test measurements (n = 24; error bars indicate the S.D.). b, Histograms showing the standard deviation of placental StO2, HbT and HbO2 obtained from the stability test (n = 24). c, Mean of placental StO2, HbT and HbO2 before and after maternal left tilt (n = 3; error bars indicate the S.D.).

Finally, we measured placental haemodynamic responses to maternal hyperoxia. The participants were given 100% FiO2 via a facemask for 20 frames (~7 min). We monitored haemoglobin concentrations continuously before, during and after maternal hyperoxia. Fig. 3a,b present a case example of variations in StO2 and HbO2. Overall (n = 24), the method easily resolved changes in placental blood oxygenation owing to maternal hyperoxia. StO2 and HbO2 were found to increase, by a median (IQR) of 7.1 (4.9, 9.3)% and 1.9 (1.1, 3.3) μM, respectively.

Optical biomarkers of placental dysfunction

Another goal of the maternal-hyperoxia study was to examine associations between placental oxygen dynamics and both APO and MVM. To this end, APO was defined as a composite of gestational hypertension (GHTN), preeclampsia, or intrauterine growth restriction (IUGR), and MVM was determined from the examination of delivered placentas by a single placental pathologist (R.L.L.). None of the participants had clinical evidence of GHTN, preeclampsia or IUGR at the time of their ultrasound/FD-DOS measurement, and the mean time interval from optical measurement to delivery was 3.1 (2.1, 5.4) weeks. APO was found in 9 of the 24 participants; MVM was found in 8 of the 24 participants. 2 of the 15 participants with normal pregnancy outcomes (NPOs) had MVM, and 3 of the 9 participants with APO had a normal placental pathology assessment (NPP).

We determined the absolute values of StO2, HbT and HbO2, as well as their variation relative to baseline during maternal hyperoxia; that is, rStO2, rHbT and rHbO2. Relative data were obtained by normalizing their time-series values to the mean of the last four frames of the baseline period.

The 24 participants were categorized into two groups, according to their pregnancy outcomes: NPO or APO. Both rStO2 and rHbO2 increased substantially in response to maternal hyperoxia for the NPO group (Fig. 3c). However, the same parameters in participants with APO showed a more blunted response (Fig. 4d).

Fig. 4 |. Continuous monitoring of placental haemoglobin properties during maternal hyperoxia.

Fig. 4 |

a,b, Exemple case of StO2 and HbO2 during maternal hyperoxia. c,d, Averaged placental rStO2 (purple), rHbT(blue), and rHbO2 (red) for NPO participants (n = 15) and for participants with APO (n = 9). e,f, Averaged placental rStO2 (purple), rHbT(blue) and rHbO2 (red) for NPP participants (n = 16) and for participants with MVM (n = 8). Shaded regions represent the standard error. The cohort-averaged rStO2 and rHbO2 exhibit a significant increase during maternal hyperoxia for the NPO and NPP groups, but a blunted response for the APO and MVM groups.

As for the participant’s placental histopathology, the same 24 participants were categorized into two groups: NPP or MVM. We observed significant, large and positive rStO2 and rHbO2 in the NPP group (Fig. 4e). These same parameters, however, showed a blunted response in the MVM group (Fig. 4f). rHbT was comparatively constant for all groups.

We next sought to quantitatively determine whether placental haemoglobin properties were significantly associated with APO. For this analysis, mean baseline StO2, HbT and HbO2 were calculated using the final 4 frames of the baseline period. Hyperoxia-induced changes from baseline (that is, ΔStO2, ΔHbT, and ΔHbO2) were defined using mean values in the 4-frame window wherein maximum StO2 occurred. Baseline StO2, HbT and HbO2 were not associated with APO (Fig. 4a, Table 1). On the other hand, ΔHbO2 and ΔStO2 were significantly reduced in cases with APO compared to NPO (Fig. 4b, Table 1).

Table 1 |. Correlations of haemodynamic variables with placental dysfunction.

For all correlations, the total number of participants is 24. The parameters are summarized as median (IQR) within each group. The p-values were obtained using a two-sided Wilcoxon rank.

StO2 % HbT μM HbO2 μM ΔStO2 % ΔHbT μM ΔHbO2 μM
Between placental static and dynamic haemoglobin properties and APO
NPO (n = 15) 69.5 (67.6, 73.4) 26.6 (22.5, 39.1) 19.9 (15.6, 28.6) 7.6 (5.6, 10.2) 0.0 (-0.8, 1.1) 2.6 (1.8, 4.4)
APO (n = 9) 75.9 (68.5, 77.3) 29.2 (15.4, 36.6) 23.2 (9.3 26.0) 4.9 (3.9, 7.1) -0.1 (-0.9, 0.5) 1.3 (0.8, 1.8)
p-value 0.37 0.51 0.63 0.03 0.63 0.02
Between placental static and dynamic haemoglobin properties and MVM
NPP (n = 16) 69.5 (67.9, 75.2) 27.9 (23.6, 35.0) 20.8 (15.8, 24.7) 7.6 (5.7, 10.1) 0.2 (-0.8, 1.7) 2.8 (1.5, 4.1)
MVM (n = 8) 73.5 (65.1, 76.5) 27.7 (14.8, 39.3) 19.3 (9.0, 30.1) 5.2 (3.7, 7.1) -0.1 (-1.7, 0.2) 1.1 (0.7, 1.8)
p-value 0.83 0.69 0.69 0.05 0.15 0.01

Similarly, we determined whether placental haemoglobin properties were significantly associated with MVM.

Baseline StO2, HbT and HbO2 were not associated with MVM (Fig. 5a, Table 1). ΔHbO2 and ΔStO2 were significantly reduced in cases with MVM, and ΔStO2 (Fig. 5b, Table 1).

Fig. 5 |. Static (baseline) and dynamic (during maternal hyperoxia) placental haemoglobin properties for participants with NPO or APO.

Fig. 5 |

a, Baseline placental StO2, HbT and HbO2 illustrate that they are not associated with pregnancy outcome. The data show no significant differences between NPO and APO. p-values are 0.37, 0.51, and 0.63 respectively. b, Placental ΔStO2, ΔHbT and ΔHbO2 during maternal hyperoxia.ΔStO2, ΔHbT and ΔHbO2 exhibit trending separation, no difference, and clear separation, respectively, amongst participants with NPO versus APO. The p-values are 0.03, 0.63 and 0.02 respectively. In these boxplots, the red centre line denotes the median value, and the blue box shows the IQR (25% to 75% of dataset). The blue whiskers mark the non-outlier minimum and non-outlier maximum; outliers are marked with symbol ‘+’. Values of the median and IQR are presented in Table 1. p-values are calculated via a two-sided Wilcoxon rank sum test.

As an additional check, we carried out the same univariate analysis with standard parameters such as maternal age, nulliparity, gestational age (GA) at study visit, maternal BMI, placental depth d, and uterine artery Doppler pulsatility index (UtA PI), which has been proposed as a surrogate indicator of trophoblastic invasion44. We did not find any significant association between APO and/or MVM and either maternal age, nulliparity, GA at study visit or UtA PI (Table 2). However, we did observe a slight trend (not significant) towards larger d and BMI in participants with APO (p = 0.1, p = 0.1, respectively, two-sided Wilcoxon rank-sum test). Although our sample size is small and although these variables are not necessarily uncorrelated, for completeness we ran binary logistic regressions with pairs of variables: d,ΔStO2; d, ΔHbO2; BMI, ΔStO2; BMI, ΔHbO2. The results confirmed that a trend towards significant association between optically derived haemodynamic properties and outcomes of interest remained (Supplementary Table 4).

Table 2 |. Correlations of demographic variables with placental dysfunction.

For all correlations, the total number of participants is 24. The parameters are summarized as median (IQR) within each group except for nulliparity, which is presented as a percentage. The p-values were obtained using a two-sided Wilcoxon rank sum test except for nulliparity, for which they were obtained using a two-sided Fisher’s exact test.

Age (years) Nulliparity (%) GA (weeks) BMI d cm UtA PI
Between other variables and APO
NPO (n = 15) 27.0 (25.3, 32.3) 20.0 34.4 (32.8, 35.1) 25.7 (24.2, 28.9) 2.5 (2.3, 2.9) 0.7 (0.6, 0.8)
APO (n = 9) 27.0 (22.8, 33.3) 44.4 34.6 (32.9, 36.4) 30.1 (24.9, 33.4) 3.2 (2.5, 3.6) 0.8 (0.6, 1.1)
p-value 0.93 0.36 0.53 0.10 0.10 0.21
Between other variables and MVM
NPP (n = 16) 26.5 (24.0, 31.5) 18.8 34.2 (32.8, 35.2) 26.5 (24.6, 29.2) 2.5 (2.3, 3.0) 0.7 (0.6, 0.8)
MVM (n = 8) 29.5 (24.0, 34.0) 50.0 34.7 (33.6, 35.9) 30.2 (23.4, 34.3) 3.2 (1.9, 3.5) 0.7 (0.6, 1.1)
p-value 0.52 0.17 0.58 0.34 0.56 0.65

Lastly, our methodology permitted the study of adipose and rectus/uterus layers at baseline and during maternal hyperoxia (Supplementary Tables 3 and 5). The resultant adipose and rectus/uterus layer haemodynamic properties typically differed from those of the placenta. We did not find any statistically significant association of any overlayer haemodynamic property with APO or MVM (Supplementary Table 5). Collectively, these data suggest that without multilayer modelling, the computed placenta responses would have been attenuated; that is, the deep-tissue signal would have then been a weighted average of the placenta, rectus/uterus and adipose layers.

Discussion

Direct and non-invasive clinical methods to assess placental function in vivo and at the bedside would be desirable. The depth of the placenta below the skin surface, and the variability in the properties of the overlying layers, presents significant challenges for optical diagnostics. In this study, we report the development and performance of an instrument and methodology that substantially expands the capabilities of DOS to enable bedside dynamic monitoring of the placenta and potentially of other organs buried far below the tissue surface.

Tissue-simulating phantom experiments showed that the instrument has sufficient dynamic range and SNR to perform measurements at long-distance SDSs (up to 10 cm), thereby suggesting that deep placental layers can be optically interrogated in the clinic. Moreover, by coupling FD-DOS instrumentation with ultrasound imaging, we directly mapped the morphology of overlying layers of the abdominal wall and uterus. This mapping permits multilayer modelling of tissue properties that effectively isolates the placenta’s optical and physiological properties. We validated the methodology using tissue phantoms and finite-element simulations, and carried out a pilot study of third-trimester pregnancies.

Placental StO2, HbT and HbO2 were noninvasively measured at the bedside, differentiating placenta from tissue overlayers. We also showed the repeatability and stability of the optical metrics. The baseline placenta properties between the normal and abnormal pregnancies were statistically similar. This finding is in agreement with a prior CW study27 but is different from observations of higher placental oxygenation level in IUGR participants from two early CW studies26,28 and from observations of lower placental oxygen level in women with pregnancy complications from a recent CW study45. Regarding study comparisons, it should be noted that, besides important differences in technical approach (such as layer averaging, light penetration, participant BMI and other assumptions), the adverse outcomes in our study were generally not as severe as in some of the others26,28. Moreover, we made assessments weeks prior to clinical disease and delivery. Hence, although studies with larger sample size are clearly warranted, the similar baseline placenta-tissue oxygen levels that we observed are not at odds with physiological expectations, even in the presence of underlying pathology.

Since gold-standard references for placental tissue haemoglobin values do not exist, we used functional perturbations to show sensitivity to expected physiologic changes. Maternal left lateral tilt positioning relieves caval compression and augments venous return to the heart, improving cardiac output. In pregnancy, ~20% of maternal cardiac output is directed to uteroplacental perfusion. Thus, this manoeuvre generally increases blood flow to the placenta. In fact, this position is used clinically as a foetal resuscitation manoeuvre during labour to maximize placental perfusion46. Despite the limited sample size (n = 3), the FD-DOS/US device detected an increase for both HbT and HbO2, consistent with the expected increased maternal blood flow into the intervillous vascular spaces of the placenta.

More importantly, the maternal-hyperoxia experiments showed that the method can detect real time changes in placental HbO2 and StO2 owing to inflow of the oxygen-enriched maternal blood into the placenta, whereas HbT remained comparably stable. These perturbation experiments validated our clinical expectations, and confirmed the sensitivity of the optical measures to underlying utero-placental haemodynamics. They are also consistent with MRI findings; for example, this effect was demonstrated in normal foetuses and in those with congenital heart disease11, and a similar delayed response in abnormal versus normal placentas in twin pregnancies was observed with hyperoxic BOLD MRI47, again consistent with our observations.

We also studied how optical metrics obtained during maternal-hyperoxia correlates with pregnancy outcome and placental pathology. APOs are significantly associated with both short-term and long-term morbidity and mortality13. Similarly, MVM are associated with both APOs and long-term adverse outcomes35, with prevalence estimates as high as 32% in term births and over 50% in preterm births48. Our pilot study indicates that the non-invasive optical monitoring of placental responses to maternal hyperoxia is feasible and that it holds potential as a methodology to detect signs of poor placental perfusion weeks before delivery.

The adipose and rectus/uterus layer haemoglobin properties of healthy participants during baseline and during maternal hyperoxia were not significantly different from participants with APO or MVM (Supplementary Table 5). These null results for the overlayers are not necessarily surprising from a physiological perspective. For example, since MVM presents a recognizable pattern of placental injury related to altered uterine and intervillous blood flow, we would not expect a significant association between adipose haemoglobin properties and MVM. However, we noticed that a few overlayer parameters exhibit trending associations (p-values as small as 0.06 for the rectus/uterus layer ΔStO2 and APO, two-sided Wilcoxon rank-sum test); this observation should encourage further study in a larger sample size; that is, one might hypothesize that subclinical vascular dysfunction may predispose to poor placentation or be an indicator of placental dysfunction.

The need for new tools to assess placental function is well known to clinicians. Critical knowledge gaps exist in obstetrical care as a result. For this reason, some investigators have turned to MRI to derive functional parameters related to placental oxygenation. MRI does not have the same issues of depth penetration that limit optics, and specialized MRI techniques such as T2*, BOLD, and magnetic-resonance susceptibility are correlated with oxygen content1012. However, these correlations with oxygen have limitations, and MRI is not suitable for bedside measurements. DOS, by contrast, can directly measure deoxy-haemoglobin and oxy-haemoglobin concentrations at the bedside and non-invasively during clinical care. Here we have described a unique instrument that permits quantitative dynamic monitoring of the human placenta at the bedside, and the associations that we have found with APO and MVM support its continued development. However, our work thus far has some limitations: to date we have probed only anterior placentas within ~4 cm of the surface, and the placenta optical signals are due to a combination of maternal and foetal blood. Although the methodology holds clinical promise, a clinical study with larger sample sizes is desirable to corroborate the findings, and to enable more sophisticated statistical analyses that explore possible confounding variables and that generate composite metrics with improved specificity and sensitivity.

The technology will need to be refined. With improved spatial information from ultrasound, such as 3D imaging, we may be able to improve on the uniform slab-layer tissue model and derive optical properties with greater fidelity. With improved time resolution, the temporal responses to functional activation could be explored as a test variable. Additionally, an improved time resolution would increase the sensitivity to changing conditions and would facilitate the efficient evaluation of multiple placental sites; the latter is especially important given the potential spatial heterogeneity of pathology within the placenta. Exploration of placental oxygenation at earlier gestational ages, although challenging, could lead to the identification of early signs of placental insufficiency, and comparisons of uterine and placental responses during hyperoxia could improve our understanding of maternal and foetal oxygen consumption. Real-time optical monitoring feedback could alert clinicians to suspicious data owing to contraction or movement and thus improve data-acquisition quality. Currently, it takes ~90 minutes to determine the regularization hyperparameters; three-layer reconstruction of each frame data takes ~90 seconds on our 3.2GHz quad-core computer. These times can be reduced to ~12 minutes and ~12 seconds, respectively, with a 3.5GHz 32-core workstation. Looking forward, with machine learning and information from large datasets, we believe that the regularization hyperparameter could be determined in ~1 minute at the beginning of the measurement and thereby facilitate real-time layer-model reconstructions. Broadly, we anticipate that optical metrics of placental haemodynamics will enable scientists to better understand placental pathophysiology. Moreover, the instrumentation and methodology that we have reported here is potentially suited for in vivo studies of oxygen function in other internal organs buried deep below the tissue surface, such as the uterus and the kidney.

Methods

FD-DOS instrumentation

The details of the construction and operation of the custom heterodyne FD-DOS instrument are schematically shown in Extended Data Fig. 1. Briefly, three sinusoidal electromagnetic waves at radio frequency (RF) f1=100 MHz, i.e., one for each laser diode, and one sinusoidal wave at frequency f2=100.2 MHz, were generated from four low-noise, fractional-N phase-locked-loop signal generators (HMC833, Hittite Microwave Corporation). The waves were synchronized by an ultra-low-jitter programmable reference clock (f0=50 MHz, LMK61E2, Texas Instruments). Each f1 wave from the signal generator was amplified (ZX60-P103LN+ by Mini-Circuits), filtered (DC to 98 MHz, SLP-100+, Mini-Circuits), and divided (2-way splitter, Z99SC-62-S+, Mini-Circuits) into two f1 waves; one was prepared for the reference signal (‘reference f1’) and the other was prepared for driving amplitude modulation for one laser (‘signal f1’). Simultaneously, the f1 wave from the signal generator was also amplified, filtered, and divided (4-way splitter, ZB4PD-52–20W-S+, Mini-Circuits) into four f2 waves; three of these were prepared for frequency-mixing with the three detected signals (‘signal f2’), and the other was prepared for frequency-mixing with the reference signal (‘reference f2’).

The three of ‘signal f1’ were further amplified and input into laser controllers (CLD1011LP, Thorlabs), which drive the light amplitude modulation of the three near-infrared lasers with wavelengths of 785 nm (LP785-SF1000, Thorlabs), 808 nm (LDPC-T3–808-62.5/125-M-35–3S-3–0.5–70-150LD, OZ Optics), and 830 nm (LDPC-T3–830-62.5/125-M-35–3S-3–0.5–70-150LD, OZ Optics). The laser controllers also maintain thermal stability using digital proportional-integral-derivative control. To enhance signal-to-noise ratio (SNR), the achievement of a modulation depth or amplitude modulation index, i.e., the ratio of the modulation excursions of the RF signal to the level of unmodulated carrier, of greater than 90% for each laser is critically important. Achievement of these large modulation depths required individually optimized RF amplification according to each laser’s characteristics. Each laser has a specific threshold current, maximum current, and optical power efficiency (mW/mA). To minimize unmodulated background light and maximize SNR, each laser has a unique RF driver that drives its amplitude with modulation depth >90% (93%, 95% and 98% respectively). Specifically, each ‘signal f1wave was divided into four sub-signals via a 4-way power splitter. Each sub-signal was then amplified in one or two amplification stages (in each stage, the signal was amplified by ~8.6dB); the 830nm channel had 2 amplification stages and the 785nm and 805nm channels had 1 amplification stage. The sub-signals were then combined via a 4-way power splitter. A custom-built circuit containing a low noise amplifier (ZX60-P103LN+, Mini-Circuits) and low pass filter (DC to 98 MHz, SLP-100+, Mini-Circuits) was used in each amplification stage.

The three amplitude modulated laser diodes were fibre-coupled to an optical switch (MEMS 91545C4, Dicon), which was in turn connected to the 10 source fibres (400 µm core, 0.5 NA, FP400URT-Custom, Thorlabs) on the probe head (see main text Fig. 1 (b)). The optical switch sequentially cycled each laser diode through each source position and also a “dark count position” (i.e., a cycle of 3×11=33 sequential measurements; 21 seconds per cycle). Of note, for the dark count measurement, no fibre was connected to the 11th position on the switch (i.e., no light was delivered to the tissue).

Multiply scattered light emerging from the tissue at the detector position was collected by a high-transmission liquid light guide (5 mm core, 0.59 NA, LLG5–8H, Thorlabs) that was coupled to a high-sensitivity photomultiplier (PMT) detector (R12829, Hamamatsu). The PMT converts the diffuse light wave to a proportional electric voltage signal, which is then amplified and filtered by a high-speed current amplifier (DHPCA-100, FEMTO), and finally frequency-mixed (ZP-3-S+, Mini-Circuits) with one ‘signal f2’. Mixing produces a heterodyne down-converted signal, related to the diffusive light wave (i.e., its amplitude and phase are proportional to those of the diffusive light wave), at frequency Δf=f1f2=0.2 MHz. This lower-frequency signal, which can be very accurately quantified, is the heterodyne detected signal31. Simultaneously, a reference signal with the fixed frequency Δf, amplitude Ar, and phase φr is generated by mixing the ‘reference f1’ and ‘reference f2’.

A high-sampling-rate lock-in amplifier (MFLI 500 kHz, Zurich Instruments) compares reference and detected signals to derive the amplitude A and phase φ of the diffuse light wave (i.e., the lock-in output in-phase I=Acosφ and quadrature Q=Asinφ signals, from which A and φ are calculated). Note, a 3-to-1 RF switch (G4J-520120, Charter Engineering Inc.) was employed to pair the correct reference signal with the corresponding detected wavelength. Note also, prior to computing A and φ, the Q and I for each wavelength at every source position were subtracted by the corresponding Qnoise and Inoise obtained from the dark count position in the same cycle. In summary, we collect diffuse light waves from 10 source-detector pairs with source-detector separations (SDSs) ranging from ~1 to ~9 cm in the human probe (and up to 10 cm in the tissue-phantom experiment); these data enable the depth-dependent optical determination of tissue properties.

Three-layer photon diffusion model and Green’s function

The human abdomen is multi-layered. We model it as a three-layer medium, wherein each layer is assumed homogeneous and laterally infinite in extension. The experimental geometry is described using cylindrical coordinates in the main text Fig. 1(c); the depth is denoted by z cm, and the source-detector separation (SDS) is denoted by ρ cm. Both source and detector are positioned on the surface z=0. In the diffusive medium, sources on the tissue boundary are well-modelled as an isotropic point source in the medium at depth of z0=ltr,1 cm, which depends on tissue optical properties and is defined below. The diffusion equations (in the frequency-domain) for the spatially-dependent amplitude of the diffusive waves in each of the layers of the three-layer medium are38,39,49:

2U1ρ,zμa,1D1+i2πf1v1D1U1ρ,z=MS0D1δρρ0,zz0;0z<d1; (1–1)
2U2(ρ,zμa,2D2+i2πf1v2D2U2ρ,z=0 ;d1z<d1+d2; (1–2)
2U3(ρ,zμa,3D3+i2πf1v3D3U3ρ,z=0;d1+d2z. (1–3)

Here, the diffusive wave, ΦAC,kρ,z,t=Ukρ,zei2πf1t Wcm2; it is a complex representation of photon fluence rate within layer k k=1,2,3. S0 W is the time averaged power emitted by the light source, f1 Hz and M dimensionless are the frequency f1=100 MHz and modulation depth of the source, respectively. The source is point-like and located at ρ0,z0.dkcm, μa,k cm1, and μs,k cm1 are the layer k thickness, layer k light absorption coefficient, and layer k reduced scattering coefficient, respectively. ltr,k1/μa,k+μs,kcm and Dk=1/3ltr,kcm are the photon transport mean-free path and the diffusion coefficient of layer k, respectively. vk=c/nkcm/s is the light velocity in layer k, where c cm/s is the speed of light in vacuum and nk is the refractive index of layer k. The boundary conditions for the photon fluence rate and its normal derivative across the interfaces are well-known and are used to derive solutions50.

Assuming large SDS, the equations can be solved using a Fourier transform approach and extrapolated-zero boundary conditions. The analytical Green’s function for the three-layer diffusion equation in the z=0 plane is38,49:

G3Lρ,z=0,ρs=ρ0, zs=z0=12π0G˜3LssJ0sρds; (2–1)
G˜3Ls=Z3LsY3Ls/α1D1; (2–2)
Z3Ls=N12M23expα1z0α1d1+α2d2+α1zb,1N12M23expα1z0α1d1+α2d2α1zb,1+N12N23expα1z0+α1d1α2d2+α1zb,1N12N23expα1z0+α1d1α2d2α1zb,1+M12N23expα1z0α1d1α2d2+α1zb,1M12N23expα1z0α1d1α2d2α1zb,1+M12M23expα1z0+α1d1+α2d2+α1zb,1M12M23expα1z0+α1d1+α2d2α1zb,1; (2–3)
Y3Ls=M12M23expα1d1+α2d2+α1zb,1+N12M23expα1d1+α2d2α1zb,1;+ N12N23expα1d1α2d2+α1zb,1+M12N23expα1d1α2d2α1zb,1; (2–4)
M12=α1D1+n22n12α2D2;M23=α2D2+n32n22α3D3;N12=α1D1n22n12α2D2;N23=α2D2n32n22α3D3; (2–5)
αk=Dks2+vkμa,k+i2πf1Dk ;k=1,2,3. (2–6)

Here, J0 is the Bessel function of the first kind and zero-order. zb,12/3ltr,11+Reff/1Reffcm is the extrapolation length, where Reff is related to the indices of refraction of the media50. In practice, we solve the integral in equation (2-1) numerically by applying Gauss-Laguerre quadrature of 5000 points. Note, in order to minimize numerical errors, the hyperbolic functions are expanded and simplified.

Note, the analytical Green’s function for the two-layer diffusion equation in the z=0 plane can be derived from a special case of the three-layer model, where μa,1=μa,2, μs,1=μs,2, n1=n2, d1+d2=d, and D1=D2, α1=α2.

G2Lρ,z=0,ρs=ρ0, zs=z0=12π0G˜2LssJ0sρds; (3–1)
G˜2Ls=Z2LsY2Ls/α1D1; (3–2)
Z2Ls=expα1z0expα1z02α1zb,1 +βexpα1z02α1dβexpα1z02α1d2α1zb,1; (3–3)
Y2Ls=1+βexp2α1d2α1zb,1; (3–4)
β=α1D1n32/n12α3D3α1D1+n32/n12α3D3 . (3–5)

The photon fluence rate measured on the tissue surface is essentially the Green’s function multiplied by the constant amplitude of the source MS0. The detected signal intensity is directly proportional this photon fluence rate:

U1ρ,z=0,ρs=ρ0, zs=z0=MS0Gρ,z=0,ρs=ρ0, zs=z0; (4–1)
Imρ=KU1ρ,z=0,ρs=ρ0, zs=z0=CGρ,z=0,ρs=ρ0, zs=z0; (4–2)
Amρ=Imρ; θmρ=argImρ . (4–3)

Here, C=CaexpiCp is the complex light coupling coefficient (a proportionality constant); Ca and Cp, are determined using calibration phantoms with known optical properties before/after the measurement. With equations (2) and (4), the signal at the tissue surface boundary can be calculated, given input optical properties for each layer and layer thicknesses. Such a calculation is called a solution of the forward problem. The reconstruction algorithms underlying the FD-DOS analysis solves the inverse problem by finding the tissue properties that minimize the difference between measured signal and the forward problem (theoretical) solution (with specific input properties).

Global optimization with multiple spectral/SDS channels

In practice, the two-layer or three-layer optical properties based on the photon diffusion model were reconstructed by solving a global optimization problem. Specifically, we carried out the data inversion using all source-detector pairs and wavelengths simultaneously. This approach builds-in global constraints about chromophore absorption and layer geometry into the inverse problem and is critical for robust fitting. Note, the accuracy of our determination of Hb and HbO2 concentration (and thus StO2, HbT) is superior to that based on independent determination of absorption and scattering coefficients as a function of wavelength. This is because the use of the known chromophore extinction coefficients, the Mie power-law model for scattering, and the full collection of multi-spectral and multi-SDS data to fit to all wavelengths and SDSs simultaneously, effectively constrains the reconstruction problem40.

The multispectral fitting assumed that μa,kλj, the absorption coefficient at layer k at wavelength λj j=1,2,3, λj=785, 808, 830 nm, is due to the absorption of HbO2k, Hbk, H2Ok, and lipidk.

μa,kλj=εHbO2λjHbO2k+εHbλjHbk+εH2OλjH2Ok+εlipidλjlipidk. (5)

In equation (5), εHbO2λj, εHbλj, εH2O, and εlipidλj are wavelength-dependent extinction coefficients for each chromophore, which are known51. H2Ok and lipidk are the concentration of water and lipid in layer k, which are assumed51. HbO2k and Hbk are the concentration of oxy- and deoxy-haemoglobin in layer k; they are determined by solving the inverse problem.

Notice, the total haemoglobin concentration HbTk and the tissue blood oxygen saturation StO2,k in layer k can also be readily obtained from:

HbTk=HbO2k+Hbk;StO2,k=HbO2k/HbTk;k=1,2,3. (6)

For multi-spectral fitting, we also assumed a Mie scattering model (equation just below) for the tissue scatterers15, wherein the scattering coefficient in layer k is a power law function with scattering amplitude γk and scattering power bk. Here, λ0=700 nm is a reference wavelength chosen based on the range of the three wavelengths.

μs,kλj=γkλj/λ0bk. (7)

With equations (5), (6), and (7), StO2,k and HbTk can be directly determined by global optimization using all data:

argminj=13l=1Lεl,jX¯2; (8–1)
ΨX¯=εl,jX¯2=j=13l=1LlnAc,l,j/Ac,l0,jlnAm,l,j/Am,l0,j2+θc,l,jθc,l0,jθm,l,jθm,l0,j2; (8–2)
X¯=StO2,1  HbT1Υ1 b1  StO2,2  HbT2  Υ2  b2  StO2,3  HbT3  Υ3  b3 . (8–3)

In this global optimization, X¯ is an array of all fitting variables, including haemoglobin concentrations and variants thereof, and Mie scattering model parameters for scattering. The objective function ΨX¯ is a “residual” function of X¯; it is essentially a Chi-squared function that compares calculated to measured data. Represents the L2 norm of the vector. Ac,l,j and θc,l,j are the calculated (predicted) amplitude and phase using the forward solver with estimated optical properties. Am,l,j and θm,l,j are the measured amplitude and phase. Subscripts l and j represent the lth SDS and jth wavelength, respectively. Notice, the normalization amplitude and phase factors are also incorporated into the objective function; these terms are denoted with the subscript, l0.

Regularization and initialization of the optimization problem

To avoid overfitting in the reconstruction, regularization is employed to reduce the ill-posedness of the inverse problem. A Tikhonov regularization term is added to the original objective function to provide additional constraints. Thus, the new objective function ΨX¯ is the sum of the residual function in (8-2) and a weighted regularization term52:

ΨX¯=j=13l=1Lεl,j(X¯)2+ζRR(X¯);R(X¯)=X¯X¯02. (9)

RX¯ is the Tikhonov regularization term, and ζR is the regularization hyperparameter, which calibrates the relative weight of the residual function and regularization term. Tikhonov regularization, as used here, seeks to minimize the difference between the initial estimated value X¯0 and the reconstructed value X¯. The value for the regularization hyperparameter is determined using an L-curve method, which is a convenient graphical tool to find the optimized regularization parameter which balances the trade-off between fluctuation size and fluctuation smoothing. Once the ζR is determined, the optimization problem to minimize ΨX¯ can be solved iteratively. In practice, the iterative algorithm is performed by MATLAB fmincon function with a parallel search function- multi-start for global minimum53.

The first and very important step in the iterative search is initialization, wherein a reasonable estimated value X¯0 is set. It is important that the initial guess is chosen to be reasonably close to the true value; otherwise, the iterative search may not converge to a meaningful solution. Main text Fig. 2 (a) schematically outlines our three-step reconstruction procedure for initialization and determination of placental haemodynamic properties.

In essence, each step of the three-step reconstruction finds “best” tissue properties by minimizing the difference between measured data and the predictions of diffuse optical tissue models of increasing complexity. STEP 1 assumes the underlying tissue is semi-infinite and homogeneous; it utilizes short SDSs (based on the superficial layer thickness, longest SDS ≥ 2*Depth+2) to derive an initial estimate for properties of the near-surface region and the full set of SDSs to derive an initial property estimate for the whole region. For predictions, a standard semi-infinite homogeneous medium analytic solution is used.

STEP 2 utilizes estimates from step one as initial guesses in an analytic two-layer diffuse optical tissue model. The two-layer (and three-layer) models fix layer thicknesses based on ultrasound data, but other tissue properties within each layer such as scattering, StO2, and HbT, are permitted to vary to minimize the difference between measurement and tissue model predictions. Short SDSs and a two-layer analytical light transport model are utilized to derive best estimates for properties of adipose and rectus/uterus layers; all SDSs, and this same two-layer analytic light transport model, are utilized to derive best estimates for the over-layer (adipose plus rectus/uterus) and placental regions.

STEP 3 utilizes estimates for each layer from step two as initial guesses in the final three-layer diffuse optical tissue model. Again, all layer thicknesses are fixed by ultrasound-image segmentation, but the other tissue properties within each layer are permitted to vary to minimize the difference between the measurement and the predictions of the three-layer analytic light transport model.

Validation and characterization with tissue-simulating phantoms

We characterized FD-DOS instrument performance using tissue-simulating phantoms with known optical properties (μa=0.10 cm1, μs=8.9 cm1 for λ1=830 nm). The simplest tissue phantoms were comprised of water, ink for absorption, and 20% Intralipid (Baxter) for scattering. During this experiment, we used a translational stage and a motion control motor to accurately change the source-detector distance with minimum change of coupling coefficient. Briefly, the detector fibre was fixed at the liquid surface and a source fibre was physically translated in the same surface plane with SDSs ranging from 6.2 cm to 10 cm using a translation stage (see main text Fig. 2 (b)). To good approximation, the coupling coefficients in equation (4) were the same at every SDS, and a nonlinear fitting based on semi-infinite homogeneous solutions of the diffusion equation was employed to reconstruct tissue phantom optical properties13. The measurement SNR is defined as the mean intensity of 10 distinct continuous measurement trains divided by the standard deviation of the same 10 measurements for a specified SDS; accuracies are defined as the difference of measured and expected values, divided by the expected value.

In the two-layer phantom experiments, a solid phantom (μa,b=0.10 cm1, μs,b=5.0 cm1 for λ1=785 nm) was positioned under the liquid phantom, and the full Ultrasound/FD-DOS optical probe was set on the liquid surface (see main text Fig. 2 (c)). An absorption-titration experiment tested sensitivity to absorption coefficient (chromophore concentration). The over-layer liquid phantom thickness was held constant at 3 cm, the over-layer scattering coefficient was held constant (μs,t=9.10 cm1 for λ1=785 nm), and the absorption coefficient was incrementally increased in the top layer from μa,t=0.08 to 0.13 cm1. Results are given in Supplementary Table 1a. A depth-changing experiment tested sensitivity to superficial layer thickness (see main text Fig. 2 (c)). Here, the liquid had fixed optical properties (μa,t=0.13 cm1, μs,t=9.10 cm1 for λ1=785 nm), and the superficial layer thickness was increased from 1.5 to 3.0 cm; note this range encompasses the majority of our clinical placenta depths. Results are given in Supplementary Table 1b. The results are summarized as mean (S.D.) of 3 continuous measurements. The experiment accuracy is defined as the difference of measured and expected values, divided by the expected value. In Supplementary Table 1c, we show results from a two-layer phantom experiment with superficial layer thickness of 4.3 cm. Finally, we successfully carried out a deep layer absorption-titration phantom experiments to test our sensitivity to deep layer absorption coefficient. In these studies, the over-layer liquid phantom thickness was held constant at 3.25 cm, the top-layer optical properties were fixed, and the bottom-layer absorption was titrated. We reconstructed bottom-layer absorption variation of μa=0.10 cm1, μa=0.18 cm1, and μa=0.26 cm1 (all at λ=785 nm) with error of 8.4%, -3.6%, and -14.8%, respectively. These experiments (and results) are consistent with other work in the literature37. While our phantom work used larger source-detector separations to probe deep-layer optical properties, but the signal-to-noise and depth-to-SDS-ratio in our measurements and those of reference37 was comparable; importantly, our inverse problem was more robust than in reference37 because the overlayer thickness was known.

Ultrasound-image segmentation

The ultrasound transducer at the centre of optical probe was critical because it enabled us to derive tissue layer morphology and geometry information needed for the optical reconstruction algorithms. At the beginning of each ‘optical frame’, we captured an ultrasound image. Based on this ultrasound image, the clinician determined the depth below the tissue surface of the adipose, rectus, uterus, and placenta tissue layers in the left, middle and right sections of the ultrasound image. The difference between the left, middle and right depths was thus determined to ensure effective reconstruction of tissue optical properties. Note, if this difference was less than 0.5 cm, then we took data; on the rare occasion when the difference was larger than 0.5 cm, we repositioned the probe position/angle to decrease the difference and then took data.

Pregnancy outcome

After delivery, medical records were reviewed and relevant data on outcomes were extracted by a reviewer blinded to the optical data. Gestational hypertension (GHTN) and preeclampsia (PE) were defined per American College of Obstetricians and Gynecologists criteria54; intrauterine growth restriction (IUGR) was defined as a birth weight below the 5th percentile for gestational age55.

Placental histopathology

The delivered placentas were evaluated using a standard procedure36,56 by a single placental pathologist (R.L.L.) who was blinded to the optical properties. Maternal vascular malperfusion (MVM) was defined as a pattern of injury including placental hypoplasia (small for gestational age), villous infarcts, retroplacental haemorrhage (abruptio placentae), distal villous hypoplasia, villous agglutination, accelerated villous maturation and decidual arteriopathy. The minimum findings required for MVM diagnosis included decidual arteriopathy or at least 2 other features including accelerated villous maturation.

In vivo monitoring of placental oxygen dynamics

We designed a pilot clinical study of human placental oxygen-related haemodynamic properties. The study enrolled women with singleton pregnancies in the third trimester, anterior placentas, and pre-gravid BMI<40. For each experiment, participants were placed in supine, semi-recumbent position, and the central region of placenta was monitored. Prior to proceeding with the study, approval by the Institutional Review Board (IRB) at the University of Pennsylvania was obtained. Each participant signed the resultant informed consent forms prior to participating in the study.

Four experiments were completed: (1) a reproducibility experiment, wherein a 2-frame measurement was made both before and after lifting and placing the probe at approximately the same location for three times (n=18); (2) a stability experiment, wherein continuous data was collected for 10 frames (~3.5 minutes) with participant breathing room air (n=24); (3) a maternal left tilt experiment wherein 4 frames of data were collected, first in supine and then in left-lateral decubitus position, to characterize haemodynamic changes related to increased maternal cardiac output and uterine perfusion (n=3); (4) a maternal hyperoxia experiment (n=24), wherein the placenta was monitored continuously for 10 frames (~3.5min) of baseline at room air, 20 frames (~7 min) of maternal hyperoxia (100%FiO2), and 10 frames (~3.5min) of recovery at room air again. Note, a single “optical frame” corresponds to a measurement cycle through 11 light source-detector pairs and 3 wavelengths in ~21 seconds.

In total, n=24 participants participated in this study. Detailed information about the participants is provided in Supplementary Table 6. Note, data from two other participants were excluded because their signals were either too small (tissue optical absorption coefficient was very large, μa>0.2 cm1) or too unstable (due to large fluctuations during baseline period).

Measurement reproducibility was evaluated using the Intra-class Correlation Coefficient (ICC). We measured the haemoglobin properties multiple times at the same placental location in 18 participants.

The stability test (n=24) results were represented by the standard deviation (S.D.) during the continuous 10 frames measurements (Fig. 6a). Note, occasionally during data acquisition, substantial movement artifacts can occur causing a single frame to exhibit >10% fluctuations in StO2, HbT, or HbO2 compared to the values of nearby frames. We identified these motion artifacts and filtered them out from the data.

To further validate the US/FD-DOS instrumentation and methodology in a physiologic context, we performed a left tilt experiment (n=3). In this experiment, the impact of increased maternal cardiac output on placental oxygen haemoglobin properties were determined. Briefly, in each participant, we measured StO2, HbT, and HbO2 for 4 frames both before and after the maternal tilt. Fig. 3c presents the mean values of StO2, HbT, and HbO2 for each participant, before and after the maternal tilt. To calculate the mean (S.D.) of relative increases, we first determined the “before/after” difference in mean value of each parameter for each participant. Then we normalized this difference by the mean of the “before” value. Finally, we averaged these fractional changes across all 3 participants (Fig. 3c). To calculate the P-values, the relative haemoglobin properties were obtained by normalizing the “after” mean values to the “before” mean values, and paired t test analysis was applied to the relative “before/after” values.

The placental haemodynamic response to maternal hyperoxia was examined by monitoring the placental haemoglobin properties (Hb, HbO2, HbT, StO2) before, during, and after maternal hyperoxia. After an initial baseline period, participants were given 100% FiO2 via facemask for ~ 7 minutes (20 frames). Concurrent ultrasound and FD-DOS data were acquired throughout the process. Overall (n=24), the experimental methodology easily resolved changes in placental blood oxygenation due to maternal hyperoxia, making use of all source-detector pairs. Nevertheless, in processing we identified and excluded specific SDSs from our reconstructions. For example, occasionally an SDS would saturate or very unstable (owing to human movement), or, occasionally, the longest SDS was very noisy (owing to large absorption); such SDSs would be excluded from further processing. Similarly, during the hyperoxia measurements, movement artifacts occasionally occurred causing a single frame to exhibit unphysiologically large (negative) fluctuations compared to the nearby frames (and baseline); such a frame would be excluded.

We investigated potential associations between placental oxygen dynamics during maternal hyperoxia, (i.e., ΔStO2, ΔHbT, and ΔHbO2) and the APO / MVM outcomes. For these analyses, the mean baseline StO2, HbT, and HbO2 were calculated using the final 4 frames of the baseline period.ΔStO2, ΔHbT, and ΔHbO2 were defined as the difference between these mean baseline values and the “peak” values of the 4-frame window during maternal hyperoxia wherein maximum StO2 occurred. Each participant (N=24) was then categorized into two groups based on pregnancy outcome: NPO or APO. We observed significantly larger ΔStO2 and ΔHbO2 in response to maternal hyperoxia in the NPO group, compared to (the more blunted response in) the APO group (see main text Fig. 4 (b)). Similarly, when analysing placental histopathology as the outcome of interest, we observed significant (large) ΔStO2 and ΔHbO2 in the NPP group compared to a blunted response in the MVM group (see main text Fig. 5 (b)). Wilcoxon rank sum tests were performed to calculate the p-values for comparison of different variables between NPO vs APO groups and NPP vs MVM groups.

Haemodynamic properties of adipose and rectus/uterus layers

Using the three-layer model, the optical and haemodynamic properties of adipose and rectus/uterus layers can also be reconstructed. Generally, we expect the accuracy of these overlayer results to be less than that of the placenta, because the probe SDSs were optimized for the deeper placental tissue. Note also, due to very thin adipose or rectus/uterus layer thickness, 4 of the 24 participants were processed with two-layer rather than three-layer model reconstruction; therefore, we excluded these 4 participants in the statistical analysis of adipose or rectus/uterus layer. The StO2 and HbT of adipose and rectus/uterus layers are reported in Supplementary Table 3; these properties are within ~10% of values reported for somewhat similar tissues, e.g., breast57 (for comparison to adipose) and muscle58 (for comparison to rectus/uterus). For the placenta layer, we found good agreement between our reduced scattering coefficient and an ex-vivo placenta study59; the placenta layer absolute haemoglobin properties have not been previously reported (to our knowledge), and the placenta layer StO2 are roughly consistent with those observed by CW-NIRS44.

The adipose and rectus/uterus baseline haemoglobin properties of healthy participants were not significantly different from participants with APO or MVM (Supplementary Table 5). This finding parallels our baseline placenta results. We also analysed the adipose and rectus/uterus layer haemodynamic response to maternal hyperoxia, and we compared their differences across participant groups (Supplementary Table 5). We did not find significant association of the overlayer haemodynamic properties in normal participants versus participants with APO or MVM. This finding is different from the placenta results during maternal hyperoxia. In total, these data provide in vivo evidence underscoring the importance of the multi-layer modelling to separate layer responses. Without the multi-layer model and associated instrumentation, quantitative estimates of placenta response are contaminated by signals from the other layers.

The null overlayer results are not necessarily surprising from physiological perspective. For example, since MVM presents a recognizable pattern of placental injury related to altered uterine and intervillous blood flow, we would not expect significant association between adipose haemoglobin properties and MVM. However, we noticed that the rectus/uterine layer blunted response to maternal hyperoxia exhibited a trending but statistically insignificant association to APO (P=0.06). Although it is possible that impaired uterine perfusion may be involved in the pathophysiology of placental dysfunction, in our view, further study in a larger sample size will be important to confirm and further elucidate these relationships.

Uterine-artery Doppler-pulsatility index

Each uterine artery was identified using a transabdominal C1–5 ultrasound probe (GE Healthcare) via power Doppler mapping. Pulsed wave Doppler was then used to obtain three similar consecutive waveforms. PI was defined as the difference between peak systolic and end diastolic velocities divided by the mean velocity. The mean PI of the two uterine arteries was used for analysis. No association was found between UtA PI and APO or MVM (Table 2). Furthermore, when including UtA PI as a covariate, the associations between ΔStO2 and ΔHbO2 and our outcomes (i.e. APO and MVM) remained apparent (Supplementary Table 4).

Statistical analysis

Statistical analyses were performed using MATLAB 2019a. Accuracy in phantom experiments and simulations is defined as the difference between measured and expected values, divided by the expected value. Depending on the data-type, results from the clinical study are presented as mean (S.D.) or median (IQR). ICC in the reproducibility experiment was calculated by dividing the random effect variance by the total variance. Our study employed a relatively small sample-size (<30 participants), and the data in the two groups were not normally distributed. In this case, a non-parametric test is appropriate (i.e., a rank sum test with continuous data). Thus, P-values for studying correlations between different variables and placental dysfunction were obtained using the two-sided Wilcoxon rank sum test, a nonparametric test for two populations when samples are independent. P values for studying correlations between nulliparity and placental dysfunction were obtained by two-sided Fisher’s exact test. P-values for studying the before/after difference in the maternal left tilt experiment were calculated by two-sided paired sample t test analysis. Binary logistic regressions were also performed to study the correlation between APO/MVM and ΔStO2 or ΔHbO2 but with control of other variables: UtA PI, placental depth d, and pre-gravid BMI. We carried out this analysis for completeness with caveats that the sample size is small and that different pairs of variables might be partially correlated (and if so, that future inclusion of interactions in the statistical models is desirable). Supplementary Table 4 shows the resultant p-values of ΔHbO2 and ΔStO2 for prediction of APO or MVM from the binary logistic regression models. The results confirm that there remained a trend towards significant association between optically-derived haemodynamic properties and our outcomes of interest. For the future, a larger sample size will permit more sophisticated statistical analyses that explore the effects of possible confounding variables and that generate composite metrics with improved specificity and sensitivity.

Extended Data

Extended Data Figure 1.

Extended Data Figure 1

Detailed schematic of custom heterodyne FD-DOS instrument.

Supplementary Material

Supplementary Table 3

Baseline hemoglobin-related properties of adipose, rectus/uterus, and placenta layers (n=20*).

Supplementary Table 4

Binary logistic regression of placental properties with two independent variables (n=24).

Supplementary Table 5

Correlation of overlayer hemodynamic with placental dysfunction (n=20).

Supplementary Table 6
Supplementary Table 1

Two-layer phantom experiment results (λ =785 nm)

Supplementary Table 2

Reconstructed optical and hemoglobin-related properties of three-layer finite element numerical simulations with realistic layer morphology.

Acknowledgements

The work was supported by the NIH U01HD087180. J.M.C. was partially supported by NIH P41EB015893. T.K. was partially supported by NIH F31HD085731 and NIH T32HL007915. W.B. was partially supported by R01NS113945. A.G.Y. acknowledges partial support from NIH R01NS060653 and NIH P41EB015893. We gratefully acknowledge useful discussions with Daniel Licht, Brian White, Jerome Strauss, Yi Hong Ong for their advice and support as well as the clinic research coordinators of the Maternal and Child Health Research Center at Perelman School of Medicine, University of Pennsylvania.

Footnotes

Code availability

The custom code employed for processing the optical data and for performing the statistical analysis are available from figshare with the identifiers https://doi.org/10.6084/m9.figshare.19451882, https://doi.org/10.6084/m9.figshare.19451879 and https://doi.org/10.6084/m9.figshare.19451876. The LabVIEW code and simulation code are also available from the corresponding author on request.

Competing interests

The authors declare no competing interests.

Additional information

Supplementary information The online version contains supplementary material available at https://doi.org/10.1038/s41551-022-00913-2.

Peer review information Nature Biomedical Engineering thanks Carolyn Bayer, Christopher Contag and the other, anonymous, reviewer(s) for their contribution to the peer review of this work.

Data availability

The main data supporting the results in this study are available within the paper and its supplementary information. All optical data generated in this study, including source data and the data used to make the figures, are available from figshare with identifiers https://doi.org/10.6084/m9.figshare.19451882, https://doi.org/10.6084/m9.figshare.19451879 and https://doi.org/10.6084/m9.figshare.19451876. The raw clinical and ultrasound data are available from the corresponding author, subject to approval from the Institutional Review Board of the University of Pennsylvania.

References

  • 1.Burton GJ, Fowden AL & Thornburg KL Placental Origins of Chronic Disease. Physiol. Rev 96, 1509–65 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Thornburg KL, O’Tierney PF & Louey S Review: The placenta is a programming agent for cardiovascular disease. Placenta 31 Suppl, S54–9 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hodyl NA et al. Child neurodevelopmental outcomes following preterm and term birth: What can the placenta tell us? Placenta 57, 79–86 (2017). [DOI] [PubMed] [Google Scholar]
  • 4.AIUM-ACR-ACOG-SMFM-SRU Practice Parameter for the Performance of Standard Diagnostic Obstetric Ultrasound Examinations. J. Ultrasound Med 37, E13–E24 (2018). [DOI] [PubMed] [Google Scholar]
  • 5.Turco MY et al. Trophoblast organoids as a model for maternal–fetal interactions during human placentation. Nature 564, 263–281 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hemberger M, Hanna CW & Dean W Mechanisms of early placental development in mouse and humans. Nature Reviews Genetics 21, 27–43 (2020). [DOI] [PubMed] [Google Scholar]
  • 7.Carter AM Animal Models of Human Placentation - A Review. Placenta 28, S41–S47 (2007). [DOI] [PubMed] [Google Scholar]
  • 8.Schmidt A, Morales-Prieto DM, Pastuschek J, Fröhlich K & Markert UR Only humans have human placentas: Molecular differences between mice and humans. J. Reprod. Immunol 108, 65–71 (2015). [DOI] [PubMed] [Google Scholar]
  • 9.Nye GA et al. Human placental oxygenation in late gestation: experimental and theoretical approaches. J. Physiol 596, 5523–5534 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Horsman MR, Mortensen LS, Petersen JB, Busk M & Overgaard J Imaging hypoxia to improve radiotherapy outcome. Nature Reviews Clinical Oncology 9, 674–687 (2012). [DOI] [PubMed] [Google Scholar]
  • 11.You W et al. Hemodynamic Responses of the Placenta and Brain to Maternal Hyperoxia in Fetuses with Congenital Heart Disease by Using Blood Oxygen–Level Dependent MRI. Radiology 294, 141–148 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Logothetis NK What we can do and what we cannot do with fMRI. Nature 453, 869–878 (2008). [DOI] [PubMed] [Google Scholar]
  • 13.Durduran T, Choe R, Yodh AG & Baker WB Diffuse optics for tissue monitoring and tomography. Reports Prog. Phys 73, (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Yodh AG & Boas DA Functional Imaging with Diffusing Light. in Biomedical Photonics Handbook: Biomedical Diagnostics 21 (2014). [Google Scholar]
  • 15.Boas DA, O’Leary MA, Chance B & Yodh AG Scattering of diffuse photon density waves by spherical inhomogeneities within turbid media: Analytic solution and applications. Proc. Natl. Acad. Sci. U. S. A 91, 4887–4891 (1994). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Waterhouse DJ, Fitzpatrick CRM, Pogue BW, O’Connor JPB & Bohndiek SE A roadmap for the clinical implementation of optical-imaging biomarkers. Nature Biomedical Engineering 3, 339–353 (2019). [DOI] [PubMed] [Google Scholar]
  • 17.Baker WB et al. Continuous non-invasive optical monitoring of cerebral blood flow and oxidative metabolism after acute brain injury. J. Cereb. Blood Flow Metab 39, 1469–1485 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Choe R et al. Transabdominal near infrared oximetry of hypoxic stress in fetal sheep brain in utero. Proc. Natl. Acad. Sci 100, 12950–12954 (2003). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Tromberg BJ et al. Predicting responses to neoadjuvant chemotherapy in breast cancer: ACRIN 6691 trial of diffuse optical spectroscopic imaging. Cancer Res 76, 5933–5944 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Boas DA, Elwell CE, Ferrari M & Taga G Twenty years of functional near-infrared spectroscopy: introduction for the special issue. Neuroimage 85, 1–5 (2014). [DOI] [PubMed] [Google Scholar]
  • 21.Yun SH & Kwok SJJ Light in diagnosis, therapy and surgery. Nat. Biomed. Eng 1, 8 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Zhu Q et al. Breast Cancer: Assessing Response to Neoadjuvant Chemotherapy by Using US-guided Near-Infrared Tomography. Radiology 266, 433–442 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Eggebrecht AT et al. Mapping distributed brain function and networks with diffuse optical tomography. Nat. Photonics 8, 448–454 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Konecky SD et al. Imaging complex structures with diffuse light. Opt. Express 16, 5048–5060 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Ntziachristos V Going deeper than microscopy: The optical imaging frontier in biology. Nature Methods 7, 603–614 (2010). [DOI] [PubMed] [Google Scholar]
  • 26.Kakogawa J, Sumimoto K, Kawamura T, Minoura S & Kanayama N Noninvasive monitoring of placental oxygenation by near-infrared spectroscopy. Am. J. Perinatol 27, 463–468 (2010). [DOI] [PubMed] [Google Scholar]
  • 27.Hasegawa J et al. Evaluation of placental function using near infrared spectroscopy during fetal growth restriction. J. Perinat. Med 38, 29–32 (2010). [DOI] [PubMed] [Google Scholar]
  • 28.Kawamura T et al. Measurement of Placental Oxygenation by Transabdominal Near-Infrared Spectroscopy. Am. J. Perinatol 24, 161–166 (2007). [DOI] [PubMed] [Google Scholar]
  • 29.Scholkmann F et al. A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology. NeuroImage 85, 6–27 (2014). [DOI] [PubMed] [Google Scholar]
  • 30.Boas DA, Pitris C & Ramanujam N Handbook of biomedical optics. Handbook of Biomedical Optics (2016). doi: 10.1201/b10951 [DOI] [Google Scholar]
  • 31.Fantini S & Sassaroli A Frequency-Domain Techniques for Tissue Spectroscopy and Imaging. in Handbook of Optical Biomedical Diagnostics, Second Edition, Volume 1: Light-Tissue Interaction 477–533 (SPIE PRESS, 2016). doi: 10.1117/3.2219603.ch7 [DOI] [Google Scholar]
  • 32.Choi J et al. Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach. J. Biomed. Opt 9, 221–229 (2004). [DOI] [PubMed] [Google Scholar]
  • 33.Liebert A et al. Assessment of inflow and washout of indocyanine green in the adult human brain by monitoring of diffuse reflectance at large source-detector separation. J. Biomed. Opt 16, 046011 (2011). [DOI] [PubMed] [Google Scholar]
  • 34.Pifferi A et al. New frontiers in time-domain diffuse optics, a review. J. Biomed. Opt 21, 091310 (2016). [DOI] [PubMed] [Google Scholar]
  • 35.Wright E et al. Maternal Vascular Malperfusion and Adverse Perinatal Outcomes in Low-Risk Nulliparous Women. Obstet. Gynecol 130, 1112–1120 (2017). [DOI] [PubMed] [Google Scholar]
  • 36.Ernst LM Maternal vascular malperfusion of the placental bed. APMIS 126, 551–560 (2018). [DOI] [PubMed] [Google Scholar]
  • 37.B H, A S & S F Optical characterization of two-layered turbid media for non-invasive, absolute oximetry in cerebral and extracerebral tissue. PLoS One 8, E64095 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Liemert A & Kienle A Light diffusion in N-layered turbid media: frequency and time domains. J. Biomed. Opt 15, 025002 (2010). [DOI] [PubMed] [Google Scholar]
  • 39.Ripoll J et al. Recovery of optical parameters in multiple-layered diffusive media: theory and experiments. J. Opt. Soc. Am. A 18, 821–830 (2001). [DOI] [PubMed] [Google Scholar]
  • 40.Corlu A et al. Diffuse optical tomography with spectral constraints and wavelength optimization. Appl. Opt 44, 2082–2093 (2005). [DOI] [PubMed] [Google Scholar]
  • 41.Pogue BW & Patterson MS Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry. J. Biomed. Opt 11, 041102 (2006). [DOI] [PubMed] [Google Scholar]
  • 42.Schweiger M & Arridge S The Toast++ software suite for forward and inverse modeling in optical tomography. J. Biomed. Opt 19, 040801 (2014). [DOI] [PubMed] [Google Scholar]
  • 43.Lee SWY, Khaw KS, Kee WDN, Leung TY & Critchley LAH Haemodynamic effects from aortocaval compression at different angles of lateral tilt in non-labouring term pregnant women. Br. J. Anaesth 109, 950–956 (2012). [DOI] [PubMed] [Google Scholar]
  • 44.O’Gorman N, Tampakoudis G, Wright A, Wright D & Nicolaides KH Uterine artery pulsatility index at 12, 22, 32 and 36 weeks’ gestation in screening for pre-eclampsia. Ultrasound Obstet. Gynecol 47, 565–572 (2016). [DOI] [PubMed] [Google Scholar]
  • 45.Nguyen T et al. Non-invasive transabdominal measurement of placental oxygenation: a step toward continuous monitoring. Biomed. Opt. Express 12, 4119–4130 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.American College of Obstetricians and Gynecologists. Management of intrapartum fetal heart rate tracings. Obstetrics and Gynecology 116, 1232–1240 (2010). [DOI] [PubMed] [Google Scholar]
  • 47.Luo J et al. In Vivo Quantification of Placental Insufficiency by BOLD MRI: A Human Study. Sci. Rep 7, 3713 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Catov JM et al. Neonatal outcomes following preterm birth classified according to placental features. Am. J. Obstet. Gynecol 216, 411.e1-411.e14 (2017). [DOI] [PubMed] [Google Scholar]
  • 49.Liemert A Light diffusion in N-layered turbid media: steady-state domain. J. Biomed. Opt 15, 025003 (2010). [DOI] [PubMed] [Google Scholar]
  • 50.Haskell RC et al. Boundary conditions for the diffusion equation in radiative transfer. J. Opt. Soc. Am. A. Opt. Image Sci. Vis 11, 2727–2741 (1994). [DOI] [PubMed] [Google Scholar]
  • 51.Jacques SL Optical properties of biological tissues: A review. Phys. Med. Biol 58, R37–61 (2013). [DOI] [PubMed] [Google Scholar]
  • 52.Hansen PC The L-curve and its use in the numerical treatment of inverse problems (1999). [Google Scholar]
  • 53.Ugray Z et al. Scatter search and local NLP solvers: A multistart framework for global optimization. INFORMS J. Comput 19, 328–340 (2007). [Google Scholar]
  • 54.Gestational Hypertension and Preeclampsia: ACOG Practice Bulletin, Number 222. Obstet. Gynecol 135, e237–e260 (2020). [DOI] [PubMed] [Google Scholar]
  • 55.Fenton TR & Kim JH A systematic review and meta-analysis to revise the Fenton growth chart for preterm infants. BMC Pediatr 13, 59 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Khong TY et al. Sampling and definitions of placental lesions Amsterdam placental workshop group consensus statement. in Archives of Pathology and Laboratory Medicine 140, 698–713 (College of American Pathologists, 2016). [DOI] [PubMed] [Google Scholar]
  • 57.Taroni P et al. Breast tissue composition and its dependence on demographic risk factors for breast cancer: Non-invasive assessment by Time Domain diffuse optical spectroscopy. PLoS One 10, (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Henry B et al. Hybrid diffuse optical techniques for continuous hemodynamic measurement in gastrocnemius during plantar flexion exercise. J. Biomed. Opt 20, 125006 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Khare SM et al. Evaluation of the human placenta optical scattering properties using continuous wave and frequency-domain diffuse reflectance spectroscopy. J. Biomed. Opt 25, 116001 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Wang L et al. Placental FDDOS data analysis with layered model reconstruction. Figshare, 10.6084/m9.figshare.19451882 (2022). [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Table 3

Baseline hemoglobin-related properties of adipose, rectus/uterus, and placenta layers (n=20*).

Supplementary Table 4

Binary logistic regression of placental properties with two independent variables (n=24).

Supplementary Table 5

Correlation of overlayer hemodynamic with placental dysfunction (n=20).

Supplementary Table 6
Supplementary Table 1

Two-layer phantom experiment results (λ =785 nm)

Supplementary Table 2

Reconstructed optical and hemoglobin-related properties of three-layer finite element numerical simulations with realistic layer morphology.

Data Availability Statement

The main data supporting the results in this study are available within the paper and its supplementary information. All optical data generated in this study, including source data and the data used to make the figures, are available from figshare with identifiers https://doi.org/10.6084/m9.figshare.19451882, https://doi.org/10.6084/m9.figshare.19451879 and https://doi.org/10.6084/m9.figshare.19451876. The raw clinical and ultrasound data are available from the corresponding author, subject to approval from the Institutional Review Board of the University of Pennsylvania.

RESOURCES