Diagnostic Reference Levels for nuclear medicine imaging in Austria: A nationwide survey of used dose levels for adult patients

David Wachabauer a,, Thomas Beyer b,c, Manfred Ditto d, Hans-Jürgen Gallowitsch c,e, Michael Hinterreiter f, Bettina Ibi g,h, Phillipp Malle i,j, Siroos Mirzaei c, Florian Smetana e,h, Anton Staudenherz k,l, Boris Warwitz h,m, Georg Zettinig n,o, Ivo Rausch b,l,⁎⁎
PMCID: PMC9948827  PMID: 35067426

Abstract

Purpose

To assess dose levels in routine nuclear medicine (NUC) procedures in Austria as a prior to a legislative update of the National Diagnostic Reference Levels (NDRL).

Method

As part of a nationwide survey of common NUC-examinations between June 2019 and November 2019, data sets were collected from 33 Austrian hospitals with NUC equipment. All hospitals were asked to report the NUC imaging devices in use (model, type, year of manufacture, detector material, collimators), the standard protocol parameters for selected examinations (standard activity, collimator, average acquisition time, reconstruction type, use of time-of-flight) and to report data from 10 representative examinations (e.g. injected activity, weight), incl. the most common NUC-examinations for planar imaging/SPECT and PET. Median/mean values for injected activity were calculated and compared to current Austrian and international NDRL. A Pearson correlation coefficient was computed comparing different variables.

Results

In total, all 33 hospitals (100% response rate) reported data for this study for 60 SPECT devices, 21 PET/CT devices and 23 scintigraphy devices. Fixed activity values for scintigraphy/SPECT and PET were employed by about 90% and 56% of the hospitals, respectively. The most widely performed examinations for scintigraphy/SPECT are bone imaging, thyroid imaging, renal imaging (with MAG3/EC) and lung perfusion imaging (in 88% of the hospitals) and F-18 FDG-PET studies for oncology indications (in 100% of the hospitals). Significant correlations were found for patient weight and injected activity (scintigraphy/SPECT), use of iterative reconstruction and injected activity (PET) as well as size of field-of-view and injected activity (PET).

Conclusions

The reported injected activity levels were comparable to those in other countries. However, for procedures for which NDRL exist, deviations in injected activities of >20% compared to the NDRL were found. These deviations are assumed to result mainly from advances in technology but also from deviations between NDRL and prescribed activities as given in the information leaflets of the radiopharmaceuticals.

Abbreviations: BGO, Bismuth Germanate; CT, computed tomography; CZT, cadmium-zinc-telluride; DRL, diagnostic reference levels; EU, European Union; FOV, field of view; HTML, hypertext markup language; ICRP, International Commission on Radiological Protection; IR, iterative reconstruction; LSO, Oxyorthosilicate; NaI, Natriumiodide; NDRL, national diagnostic reference levels; NUC, nuclear medicine imaging; PET, positron emission tomography; SD, standard deviation; SPECT, single photon emission computed tomography; TOF, time-of-flight

Keywords: Radiation Protection, Nuclear medicine, Diagnostic reference levels

Introduction

Over the past century, medical imaging has evolved to an indispensable part of diagnostic procedures. However, some of the most common imaging methods, such as X-ray, computed tomography (CT) as well as nuclear medicine imaging (NUC) procedures, such as scintigraphy, single photon emission computed tomography (SPECT) and positron emission tomography (PET) make use of ionizing radiation, and, thus, inherently imply radiation exposure to patients [1]. Even though the exposure from diagnostic procedures is small (in the range of 0.01 mSv to 20 mSv effective dose) and eventual negative effects of such small exposures are controversially discussed [2], [3], [4], in sum they substantially contribute to the general radiation exposure of the population [1], [5], [6].

To limit the general radiation exposure of the population and potential adverse effects to individual patients, diagnostic procedures including ionizing radiation need to follow the ALARA (As Low As Reasonable Achievable) principle. Thus, exposure from medical imaging should be limited to the lowest level necessary to reliably answer the respective diagnostic question. To achieve this, diagnostic procedures should be continuously optimized, among others considering technical advances in imaging technologies and changes in diagnostic practice.

To monitor and facilitate the dose optimization process, the International Commission for Radiologic Protection (ICRP) introduced the concept of diagnostic reference levels (DRL) [7], [8], [9]. DRL are standard levels of easily measurable quantities, such as the dose length product in CT or injected activity in NUC, for common procedures. DRL set a general guidance for the dose in clinical operation and help to identify if routine doses are unusually high or low. They do not apply directly to individual examinations and patients [7].

According to the recommendations of the ICRP [7] and European legislation [10], DRL are set and updated on national level based on the assessment of observed dose distributions. Therefore, national radiological and NUC procedures need to be regularly surveyed to account for changes in diagnostic practice and technological advances.

In the standard literature, two main approaches for setting national diagnostic reference levels (NDRL) in NUC are discussed: first, as outlined in Radiation Protection Radiation Protection N° 109 [11] of the European Commission, NDRL in NUC are administered doses necessary for a good image during a standard procedure, which follow the concept of “optimal doses”. The DRL should be approached as closely as possible and a substantial deviation of these optimal doses is only permitted for special occasions, such as for obese patients. Consequently, following this concept, NDRL for NUC are not set at the third quartile but rather at the median or mean of each examination type and these recommendations should be developed in close cooperation with relevant societies and experts. This concept is implemented in Austria [12] or Switzerland [13] and was recommended by the European Commission in 2014 [14]. Another approach would be to use the same definition and system for DRL as it is valid for CT or radiography. Consequently, NDRL are determined based on the third quartile of a dose distribution. If these NDRL are on average constantly exceeded, justification is necessary. The ICRP 103 [7] recommends using this approach and several publications were recently published [15], [16], [17]. Also a combination of these two outlined approaches is possible and was described by A Shahzad, S Bashir and A Anwar [18].

In Austria, NDRL are regulated in the Austrian Medical Radiation Protection Ordinance [12], [19]. For NUC, this ordinance states “Diagnostic reference levels for nuclear medicine examinations are defined as activities to be administered to standard-sized adult persons.” [12] and a deviation from the NDRL by more than 20% must be justified. Consequently, in the course of this study, NDRL were developed based on the median (and mean) values in close cooperation with NUC experts and under involvement of all relevant societies. Additionally, third quartiles of the dose distribution are provided for comparison with other NDRL derived based on the third quartile.

The last evaluation of diagnostic procedures and respective doses of NUC examinations in Austria was performed in 2008 [20], [21]; the NDRL were updated in 2010 accordingly [19]. Since then, significant technological advances in imaging technologies have been made and new radiopharmaceuticals have come to use. Technical advances include the introduction of direct photon detection with cadmium-zinc-telluride (CZT) based detectors for scintigraphy and SPECT imaging [22]. For specific organ examinations, such as myocardial perfusion imaging, dedicated SPECT devices and optimized collimators are more commonly available now [23]. In PET imaging, a general trend to extended axial field-of-views (FOV), and, thus, improved sensitivity can be seen [24], [25]. Furthermore, the implementation of semiconductor-based PET detectors and improved detection electronics led to significant advances in time-of-flight (TOF) resolution, and, in turn, improvements in signal-to-noise ratios [26]. Finally, progress in image reconstruction has been made within the last decade for SPECT as well as PET imaging [27], [28].

In parallel to the technical development, new diagnostic agents and respective procedures were established. For example, in the last decade Tc-99m- and Ga-68 labelled PSMA was introduced for the assessment of prostate cancer [29]. F-18 labelled amyloid tracers for dementia imaging are now commonly used [30]. F-18 labelled choline has been established for parathyroid imaging [31] and cardiac SPECT examinations shifted from mainly based on thallium-chloride to the use of technetium labelled perfusion tracers [32].

The Federal Ministry of Health and the Austrian National Public Health Institute engage in updating the Austrian NDRL for medical imaging procedures [33], [34]. To assess these above-mentioned changes in technology and clinical practice and to update the enforced NDRL for NUC, a survey of NUC procedures was performed across Austria. This included the assessment of the used technology, examination protocols and administered activities. Here, we report on the methodology and the outcome of this survey which was the basis of the update of the NDRL for NUC in Austria.

Material and methods

The study design of this nationwide survey was conducted with reference to similar international studies and based on the recommendations by the ICRP 135 [7].

Examination types

To define relevant examination types for the survey, Austrian NDRL for NUC (which were enforced at that time) were used [19]. In addition, the last study on Austrian NDRL for NUC by A Stemberger, T Leitha and A Staudenherz [21], a report by the European Commission [14] on European NDRL, NDRL from Switzerland [13] and Germany [35] were included and discussed with a multidisciplinary expert group consisting of radiologists, medical physicists and radiographers as well as representatives from relevant societies. Frequencies of relevant NUC-examinations were taken from the Austrian-wide routine documentation data held and maintained by the Austrian Federal Ministry of Health [6], [36]. Table 1 and Table 2 depict the examination types and indications included into the survey besides obligatory and optional parameters, respectively.

Table 1.

Types of planar scintigraphy imaging/SPECT examinations as well as mandatory and optional survey parameters.

Procedure Radionuclide Imaging agent survey parameters
Mandatory optional
Brain imaging I-123 I-123 benzamide, beta-CIT a) Per device:
• Manufacturer
• Model
• Year of Manufacture
• Type [scintigraphy  device,  SPECT, SPECT/CT]
• Detector material  [NaI, CZT]
Salivary gland scan Tc-99m Tc-99m pertechnetate
Thyroid imaging Tc-99m Tc-99m pertechnetate
Thyroid whole body imaging I-131 I-131 sodium iodide
Parathyroid imaging Tc-99m Tc-99m isonitrile
Myocardial perfusion imaging (rest and stress) Tc-99m Tc-99m isonitrile;
1-day and 2-day protocol
Myocardial perfusion/vitality imaging Tl-201 Tl-201 chloride
Lung perfusion imaging Tc-99m Tc-99m macroaggregates b) Per examination protocol:
• Activity [MBq], fixed or per kg
• Collimator [LEHR, LEGP, MEGP, HEGP, other]
b) Per examination protocol:
• Acquisition time per patient [min]
Gastric imaging Tc-99m marked chyme Tc-99m marked chyme colloid
Renal imaging Tc-99m Tc-99m MAG3/EC
Renal imaging Tc-99m Tc-99m DMSA
Renal imaging Tc-99m Tc-99m DTPA
Adrenal imaging I-123 I-123 MIBG
Bleeding Tc-99m Tc-99m pertechnetate, erythrocytes
Inflammation imaging Tc-99m Tc-99m anti-granulocytes antibodies
SNL lymphoscintigraphy Tc-99m
Tc-99m colloid – 1-day and 2-day protocol
Bone imaging Tc-99m Tc-99m bisphosphonate
Tumour/inflammation/receptor imaging In-111 In-111 somatostatin receptor antagonist c) Per patient:
• injected activity [MBq]
• Weight [kg]
c) Per patient:
• Height
• Sex
• Year of birth
Tumour/inflammation/receptor imaging Tc-99m Tc-99m PSMA
Tumour/inflammation/receptor imaging Tc-99m Tc-99m somatostatin receptor antagonist

Table 2.

Types of PET examinations as well as mandatory and optional survey parameters.

Procedure Radionuclide Imaging agent survey parameters
Mandatory optional
Brain imaging F-18 F-18 FDG a) Per device:
• Manufacturer
• Model
• Year of Manufacture
axial FOV in cm (in Z-direction)
• Detector material [LSO, BGO]
Brain imaging F-18 F-18 tyrosine
Myocardial imaging F-18 F-18 FDG
Tumour imaging Cu-64 Cu-64 PSMA b) Per examination:
• Activity [MBq], fixed or per kg
• average acquisition time per bed position [min]
• iterative reconstruction [yes, no]
• Time of Flight [yes, no]
Tumour imaging F-18 F-18 choline
Tumour imaging F-18 F-18 FDG
Tumour imaging F-18 F-18 fluoride
Tumour imaging Ga-68 Ga-68 PSMA c) Per patient:
• injected activity [MBq]
• weight [kg]
c) Per patient:
• Height
• Sex
• Year of birth
Tumour imaging Ga-68 Ga-68 somatostatin receptor antagonist

Survey design and data entry form

In Austria, approximately 39% of the NUC examinations are performed in private imaging centres (= ambulatory care sector) and 61% in hospitals (only including examinations which are fully covered by the social insurance system) [30]. The predominant examination in the ambulatory care sector, approx. 83%, are thyroid imaging examinations. The inclusion of private imaging centres would have caused practical issues by means of availability of contact data, thus, the decision was made that this survey will only include hospitals. We expect that the data collected is representative also for private imaging centres as the majority of examinations are performed in hospitals and thyroid imaging is a well standardised examination.

All 33 hospitals with NUC equipment were invited via e-mail, followed up by e-mail reminders and, if necessary, telephone calls to non-responding hospitals to increase response rates. The data collection period lasted from June 2019 to October 2019. The survey was carried out using a HTML based online data entry form. Consequently, this offered the possibility to identify individual data entries, create standardized reports and retrospectively trace and analyse each data entry.

The survey consisted of three parts; (a) the assessment of the imaging devices used, (b) the definition the local standard protocols and (c) the collection of individual patient data.

(a) First, each hospital was asked to declare the NUC imaging devices in use. This included the manufacturer, model and year of manufacture of the NUC-imaging devices, detector material and the axial FOV (for PET) (Table 1 and Table 2).

(b) For each device, the hospitals were asked to enter the standard protocol parameters for the selected examinations. This included the standard activity (fixed or per kg), the collimator as well as (for PET) the average acquisition time per bed position, use of iterative reconstruction (IR) and TOF (Table 1 and Table 2).

(c) For each examination type, the hospitals were asked to report data from 10 representative examinations. The required information was the injected activity and the weight of the patient. On an optional basis, the form allowed to report the height, sex and year of birth of the patient.

Data quality

Measures were taken to increase response rates and improve data quality. To incentivise study participation, analyses comparing the median activity values of an individual participating hospital to the new NDRL recommendations were advertised and made available to the participating hospitals. During the data collection phase, a questions and answers hotline was installed to allow participating hospitals to directly call and talk to experts in order to assist with and answer questions regarding data entry. Automated pop-up warning messages were integrated into the data entry form to directly inform the participants on entries with highly implausible values. In such cases the software would prompt the user to check the dose entered. To further improve data quality, data entries were centrally monitored during the data collection period and hospitals entering implausible dose values were contacted in short order to clarify eventual faulty entries. After the data entry period ended, data were again checked for plausibility.

After this plausibility check, data cleaning was performed especially for the Tc-99m DMSA examination type as several patient datasets were entered including children. Consequently all datasets with patients younger than 18 years were excluded (for planar/SPECT 123 datasets and for PET 0 datasets). For datasets without the year of birth provided, patient weight was used as a proxy and all datasets with patient weight equal or lower 35 kg were excluded (for planar/SPECT 21 datasets and for PET 0 datasets). The number of patient datasets with weight between 35 kg and 70 kg was 1,403 (40% of all datasets) for planar/SPECT and 353 (34% of all datasets) for PET. Furthermore, for myocardial imaging, values for stress and rest examinations had to be reported. If only stress examination were reported for myocardial imaging and no rest value was available, these datasets were excluded (in total 5 protocol datasets and 62 patient datasets). 13 patient datasets for the examination type thyroid whole body imaging were excluded as they were performed for therapy purpose.

Statistical analysis

All analyses were conducted with the statistical software package R (Version 3.3.1; The R Foundation, Free Software Foundation) and Microsoft Excel (Office 365, Microsoft). Median, third quartiles, mean, standard deviation (SD), maximum and minimum were calculated for each examination type. Pearson's correlation coefficient was calculated to analyse the association between different variables such as injected activity and year of manufacture of the devices. Significance was set at a P value of <0.01.

Results

In total, all 33 hospitals (100% response rate) reported data for this study. 64% (N = 21) of the hospitals have a dedicated scintigraphy device (a scintigraphy device without a SPECT option, e.g. dedicated thyroid cameras) in operation, 94% (N = 31) of the responding hospitals are equipped with a SPECT or SPECT/CT device, and 55% (N = 17) of those hospitals with a SPECT or SPECT/CT device additionally operate a PET/CT. 58% (N = 18) of the hospitals with SPECT or SPECT/CT have more than one SPECT or SPECT/CT device in operation, whereas only 24% (N = 4) of the hospitals operating a PET/CT have more than one PET/CT device. Table 3 and 4 summarizes the number of hospitals per examination type.

Table 3.

Planar imaging/SPECT - Number (N) of hospitals/devices, data for examination protocols and collected patient data.

Examination type N hospitals/devices Prescribed activities in the examination protocols
mean/median
Reported injected activities from the collected patient data
for protocols with a defined total activity [MBq] for protocols with a weight based prescription [MBq/kg] Number of reported datasets Weight mean [kg] Injected activity mean & SD [MBq] Injected activity median [MBq] Injected activity 3rd quartile [MBq] Injected activity Min-max [MBq]
Brain imaging: I-123 benzamide beta-CIT 21/28 177/185 2.4/2.4 222 78 181 ± 19 185 187 129-230
Salivary gland scan: Tc-99m pertechnetate 14/20 137/111 2.6/2.6 135 71 162 ± 75 157 186 71-434
Thyroid imaging: Tc-99m pertechnetate 29/38 73/74 1.0/1.0 321 75 76 ± 16 76 85 37-122
Thyroid whole-body imaging: I-131 sodium iodide 15/19 194/185 - 132 76 178 ± 89 185 186 74-371
Parathyroid imaging: Tc-99m isonitrile 22/34 470/400 4.9/4.9 249 76 465 ± 135 400 578 302-768
Myocardial perfusion imaging: Tc-99m isonitrile 1-day protocol (rest and stress) 20/23 total: 1090/1050 total: 12.75/12.75 180 81 1108 ± 165 1084 1141 731-1884
Myocardial perfusion imaging: Tc-99m isonitrile 2-day protocol (rest and stress) 4/5 595/550 per day 4/4 30 80 total: 1085 ± 376 total: 1000 1558 509-1585
Myocardial perfusion/vitality imaging: TI-201 chloride 7/9 103/100 3.0/3.0 80 79 129 ± 88 100 111 64-376
Lung perfusion imaging: Tc-99m macroaggregates 29/42 129/125 1/1 345 79 128 ± 39 126 150 21-387
Gastric imaging: Tc-99m marked chyme 15/21 49/40 0.5/0.5 165 71 51 ± 29 41 55 10-165
Renal imaging: Tc-99m MAG3/EC 29/43 103/100 1.2/1 331 75 101 ± 30 98 110 57-205
Renal imaging: Tc-99m DMSA 17/23 110/100 1.4/1.4 110 73 109 ± 37 98 109 59-185
Renal imaging: Tc-99m DTPA 2/4 170/170 - 20 78 222 ± 15 223 232 176-247
Adrenal imaging: I-123 MIBG 7/9 302/300 - 42 81 288 ± 106 260 370 185-450
Bleeding: Tc-99m pertechnetate, erythrocytes 9/16 625/600 8.1/8.1 116 71 565 ± 73 556 597 400-808
Inflammation imaging: Tc-99m anti-granulocytes antibodies 13/15 661/700 - 104 81 688 ± 121 700 757 408-925
SNL lymphoscintigraphy: Tc-99m colloid – 1-day protocol 23/37 89/94 - 284 74 123 ± 138 98 138 21-781
SNL lymphoscintigraphy: Tc-99m colloid – 2-day protocol 5/8 94/110 - 62 74 119 ± 53 136 160 22-219
Bone imaging: Tc-99m bisphosphonate 29/50 627/650 8.9/9.1 424 77 652 ± 102 663 727 402-985
Tumour/inflammation/receptor imaging: In-111 somatostatin receptor antagonist 3/6 173/185 2.2/2.2 60 73 172 ± 32 160 199 118-257
Tumour/inflammation/receptor imaging: Tc-99m somatostatin receptor antagonist 8/10 731/740 - 79 77 751 ± 61 740 759 525-911
Tumour/inflammation/receptor imaging: Tc-99m PSMA 3/3 618/600 - 14 87 635 ± 57 607 661 587-786

Table 4.

PET - Number (N) of hospitals/devices, data for the examination protocols and collected patient data.

Examination type N hospitals/devices Prescribed activities in the examination protocols
mean/median
Reported injected activities from the collected patient data
for protocols with a defined total activity [MBq] for protocols with a weight based prescription [MBq/kg] Number of reported datasets Weight mean [kg] Injected activity mean & SD [MBq] Injected activity median [MBq] Injected activity 3rd quartile [MBq] Injected activity Min-max [MBq]
Brain imaging: F-18 FDG 13/16 179/185 2.6/2.6 150 75 199 ± 65 197 230 70-397
Brain imaging: F-18 tyrosine 7/9 252/250 2.6/2.6 78 76 271 ± 51 260 295 172-393
Myocardial imaging: F-18 FDG 9/11 275/275 3.9/4.0 93 84 297 ± 61 300 348 154-452
Tumour imaging: F-18 FDG 17/21 271/260 3.8/3.6 207 76 279 ± 58 272 317 99-476
Tumour imaging: Ga-68 somatostatin receptor antagonist 12/15 150/150 2.2/2.0 146 78 159 ± 36 162 182 53-259
Tumour imaging: Cu-64 PSMA 3/3 250/250 4/4 30 79 288 ± 48 274 297 183-395
Tumour imaging: F-18 choline 11/14 250/250 3.7/3.5 123 78 278 ± 55 272 316 152-388
Tumour imaging: F-18 fluoride 4/6 217/200 - 50 78 246 ± 98 248 337 104-425
Tumour imaging: Ga-68 PSMA 13/17 156/150 2.1/2.0 168 83 175 ± 37 173 189 111-330

a) Devices: In total, data was reported for 60 SPECT or SPECT/CT devices and for 21 PET/CT devices in operation. In addition, data were submitted for 23 scintigraphy devices. On average, scintigraphy devices were installed in 2009 (min: 1999; max: 2017), for SPECT or SPECT/CT devices 2010 (min: 2000; max: 2019), and for PET/CT devices 2013 (min: 2008; max: 2018). 100% (N = 23) of the scintigraphy devices as well as 97% (N = 58) of the SPECT or SPECT/CT devices use Natriumiodide (NaI) as a scintillator material whereas only 3% (N = 2) of the SPECT or SPECT/CT devices use CZT. 95% (N = 20) of the PET/CT devices use Lutetium Oxyorthosilicate (LSO) as a scintillation detector material whereas only 5% (N = 1) use Bismuth Germanate (BGO). In Table 3 and Table 4 the number of devices per examination type is illustrated.

b) Examination protocols: In total, data for 463 examination protocols for planar imaging/SPECT and 112 examination protocols for PET were collected. The administered activity was based on a fixed activity value for 87% (N = 401) of the planar/SPECT-examination protocols, whereas for 13% (N = 64) of the protocols administered activity was based on activity per body weight. For PET fixed administered activities were used in 56% (N = 63) of the protocols, while for 44% (N = 49) an activity per body weight approach was used. Table 3 and Table 4 summarize the examination protocols for planar imaging, SPECT and PET/CT examinations together with the mean and median values of the fixed activities and activity per body weight as defined in the protocols for the two activity determination methods, respectively.

c) Patients: In total, data for 3,505 individual patients were submitted for planar imaging/SPECT and for 1,045 individual patients for PET. The number (N) of reported values together with the mean weight in kg, mean and median injected activity (in MBq), standard deviation as well as minimum and maximum activity values are reported in Table 3 and 4. Cave: In contrast to the PET examinations where weight was reported for all examinations, for planar imaging/SPECT patient weight was reasonably reported in 96% (N = 3,364) of the patient datasets. In Table 5 the mean and median activity values for patient data from this study (planar/SPECT) are compared to the old NDRL for Austria, NDRL for Switzerland, new NDRL from Germany [37], Croatia and EU. In Table 6 the same is illustrated for PET.

Table 5.

Planar Imaging/SPECT - Comparison of the median values from this study with published data (NDRL in MBq).

Examination type Patient data this study mean/median NDRL Austria old (2010) [19] NDRL Switzerland (2018) [13] NDRL Germany (2021) [37] Proposed NDRL Croatia (2020) [15]
(Median)
EU
(2014) [14]
Brain imaging: I-123 benzamide, beta-CIT 181/185 185 180
Salivary gland scan: Tc-99m pertechnetate 162/157 110 400 (369)
Thyroid imaging: Tc-99m pertechnetate 76/76 110 75 70 200 (185) 80
Thyroid whole body imaging: I-131 sodium iodide 178/185 185 (185)
Parathyroid imaging: Tc-99m isonitrile 465/400 740 550 550 710 (555)
Myocardial perfusion imaging: Tc-99m isonitrile 1-day protocol (rest and stress) 1108/1084 1200 1200 total: 1000 1390 (1184) 1200
Myocardial perfusion imaging: Tc-99m isonitrile 2-day protocol (rest and stress) total: 1085/1000 740 per day 600 per day, 1200 total 400 per application 730 (634)
Myocardial perfusion/vitality imaging: TI-201 chloride 129/100 110 100 110
Lung perfusion imaging: Tc-99m macroaggregates 128/126 150 180 160 180 (140) 150
Gastric imaging: Tc-99m marked chyme 51/41 110
Renal imaging: Tc-99m MAG3/EC 101/98 110 100 100 150 (137) 100
Renal imaging: Tc-99m DMSA 109/98 110 120 130 (122)
Renal imaging: Tc-99m DTPA 222/223 185 150 (130)
Adrenal imaging: I-123 MIBG 288/260 200
Bleeding: Tc-99m pertechnetate, erythrocytes 565/556 740 750
Inflammation imaging: Tc-99m anti-granulocytes antibodies 688/700 740 800 780 (582)
SNL lymphoscintigraphy: Tc-99m colloid – 1-day protocol 123/98 Total: 80 40 60 (19)
SNL lymphoscintigraphy: Tc-99m colloid – 2-day protocol 119/136 150
Bone imaging: Tc-99m bisphosphonate 652/663 740 700 8 MBq/kg 740 (663) 600
Tumour/inflammation/receptor imaging: In-111 somatostatin receptor antagonist 172/160 180 150
Tumour/inflammation/receptor imaging: Tc-99m somatostatin receptor antagonist 751/740 700 750 740 (740)
Tumour/inflammation/receptor imaging: Tc-99m PSMA 635/607

Table 6.

PET - Comparison of the mean and median values from this study with published data (NDRL in MBq).

Examination type Patient data this study NDRL Austria old (2010) [19] NDRL Switzerland (2018) [13] NDRL Germany (2021) [35]
Brain imaging: F-18 FDG 199/197 3 MBq/kg
Brain imaging: F-18 tyrosine 271/260 210
Myocardial imaging: F-18 FDG 297/300
Tumour imaging: F-18 FDG 279/272 400 250 3 MBq/kg
Tumour imaging: Ga-68 somatostatin receptor antagonist 159/162 2.0 MBq/kg
Tumour imaging: Cu-64 PSMA 288/274
Tumour imaging: F-18 choline 278/272 210
Tumour imaging: F-18 Fluoride 246/248
Tumour imaging: Ga-68 PSMA 175/173 2.5 MBq/kg

The most widely executed examination types for planar imaging/SPECT are bone, thyroid, renal imaging with MAG3/EC and lung perfusion imaging (Figure 1). Data for these examinations was submitted from 88% (N = 29) of all hospitals. The least widely used examination is renal imaging with DTPA receiving data from only 6% (N = 2) of the hospitals. For PET, tumour imaging with FDG is by far the most extensively performed examination (100% of all hospitals with PET/CT; N = 17) whereas tumour imaging with Cu-64 PSMA was documented in only 18% (N = 3) of all hospitals.

Figure 1.

Figure 1

Percentages of hospitals in which examinations are conducted.

ICRP 135 [7] recommends using a weight interval for average sized patients from 60 to 80 kg. To investigate the effects of light and heavy patients on the calculation of the median activities, a new analysis was performed excluding those datasets related to patients weighing less than 60 kg or more than 80 kg (indicated here as “weight-restricted dataset”). Percentage differences between the analysis with all patients and the weight-restricted dataset are shown in Figure 2. With imposing weight restrictions, the overall mean weight decreased from 77 kg to 70 kg for planar imaging/SPECT and from 78 to 71 kg for PET. For planar imaging/SPECT the most impact on mean weight (-16%) was found for Tc-99m PSMA examinations, inflammation imaging (-14%) and myocardial perfusion imaging (-13%). Weight restriction did not impact the findings for “bleeding”. Regarding the differences between median activity, highest difference was found for adrenal imaging with -29%. However, for most of the examination types only minor or no differences were present in the median injected activities after restriction of body weight. Similar could be observed for PET imaging with highest weight differences for myocardial imaging and lowest for brain imaging, however, with associated median activity differences of only 1% to 3%.

Figure 2.

Figure 2

Percentage differences between data without and with (60 kg to 80 kg) weight restriction and impact on median activity per examination type.

To investigate the relationship between different variables a Pearson correlation coefficient was computed. No significant correlation was found between the following variables:

  • the year of manufacture of NUC-devices and injected activity (planar imaging/SPECT): Pearson correlation coefficient of r = −0.02, p = 0.14, n = 3649 and PET: Pearson correlation coefficient of r = 0.01, p = 0.87, n = 1045),

  • the use of TOF and injected activity (PET: Pearson correlation coefficient of r = −0.03, p = 0.36, n = 845)

There is a significant correlation between:

  • Patient weight and injected activity (planar/SPECT): Pearson correlation coefficient of r = 0.17, p < 0.01, n = 3508 and PET: Pearson correlation coefficient of r = 0.13, p < 0.01, n = 1045) indicating higher activity values for patients with higher weight (positive correlation, Figure 3).

  • Use of IR and the injected activity (PET): Pearson correlation coefficient of r = −0.13, p < 0.01, n = 1045) indicating lower activity values when using iterative reconstruction (negative correlation).

  • FOV size and the injected activity (PET): Pearson correlation coefficient of r = −0.09, p < 0.01, n = 1045) indicating lower activity values with larger FOV (negative correlation).

Figure 3.

Figure 3

Scatter plots for planar imaging/SPECT and PET comparing patient weight and injected activity.

Table 7 shows the NDRL for Austria developed based on this study.

Table 7.

Austrian NDRL for nuclear medicine as enforced in 2020.

Procedure Radionuclide and radiopharmaceutical NDRL [MBq]
Planar imaging/SPECT
 Brain imaging I-123 ioflupane 185
 Salivary gland scan Tc-99m pertechnetate 110
 Thyroid imaging Tc-99m pertechnetate 75
 Thyroid whole body imaging I-131 sodium iodide 185*
 Parathyroid imaging Tc-99m isonitrile 460
 Myocardial perfusion imaging Tc-99m isonitrile 1-day protocol total: 1100
 Myocardial perfusion/vitality imaging TI-201 chloride 100
 Lung perfusion imaging Tc-99m macroaggregates 130**
 Renal imaging Tc-99m MAG3/EC and Tc-99m DMSA 110
 Adrenal imaging I-123 MIBG 200
 Bleeding Tc-99m pertechnetate, erythrocytes 740
 Bone imaging Tc-99m bisphosphonate 650



PET
 Brain imaging F-18 FDG 180
 Brain imaging F-18 tyrosine 250
 Myocardial imaging F-18 FDG 300
 Tumour imaging F-18 FDG 280
 Tumour imaging F-18 choline 250
 Tumour imaging Ga-68 somatostatin receptor antagonist 150
 Tumour imaging Ga-68 PSMA 150

Comments:

*

370 MBq in patients with thyroid stimulation.

**

for previous inhalation.

Discussion

This study summarizes the outcome of a recent survey of the employed activity levels in nuclear medicine procedures in Austria with the aim to provide a well-founded basis for the update of the NDRL. The study design was tailored to achieve a maximum response rate to enable a valid assessment of the used dosages given the comparatively low absolute number of nuclear medicine departments in Austria. This was successfully achieved for in hospital nuclear medicine departments with a response rate of 100%.

The collected dose data are in general comparable to those found in other recent studies (Table 5 and Table 6) [13], [14], [15], [35], [38]. This is reasonable as today several international guidelines and recommendation for most of the standard examinations exist, which were developed in multilateral consensus between the national professional societies (e.g. [39], [40], [41]). However, compared to the, at the time of the survey valid NDRL in Austria, deviations of >20% in average activity were found for nearly half (8/17) of the surveyed procedures (Table 5 and 6). For most of the examinations with deviation of >20%, injected activities were found to be lower than the NDRL. For example, the average injected dose for tumour imaging with F-18-FDG was found to be 30% below the NDRL. This is expected to result from technological improvements, such as 3D PET data acquisition, IR and increased sensitivities.

During the data analysis also different reasons for deviating injected activities where found. For some of the radiopharmaceuticals the NDRL valid at the time of the survey where not in agreement with the dose recommendation for the radiopharmaceuticals. For example, Thyroid imaging using Tc-99m pertechnetate was done on (median) injecting 76 MBq and parathyroid imaging using Tc-99m isonitrile using (median) 400 MBq (Table 5). Both median activities were below the NDRL of 110 MBq and 740 MBq, respectively. However, the injected activities were in agreement with the recommended dosage as defined in the pharmaceutical information leaflet of 20-80 MBq and 200-700 MBq, respectively. Such inconsistencies in prescribed doses and NDRL should be avoided. Thus, prescribed doses need to be reviewed and taken into account during updates of NDRL.

In general, the results nicely show the necessity to update DRL on a regular basis to account for changes in clinical practice and technological advances. As seen in the correlation analysis, especially IR and increased sensitivity due to extended FOVs in PET devices seemed to contribute to a general dose reduction. While being reasonable as such, it is also in line with findings for the use of IR in CT [34]. Another interesting finding pertains to the correlation of administered activities with weight. Although, a general significant correlation of patient weight with dose was found, a re-evaluation of the data in a subset of standard patients with weights of 70 ±10 kg revealed no practically considerable changes in injected activity levels for most of the protocols (Figure 2). The reason therefore is expected to be the relatively high number of protocols with fixed activity levels (87%). This facilitates the derivation of DRL as deviations of data reported for individuals from a standard patient do not influence the general average doses. However, rigor data inspections and weight restrictions might be necessary to derive reasonable DRL in cases where patient weight-based dosage is used in the majority of sites.

Limitations

To keep the workload for participating hospitals reasonable and to achieve a high participation rate, we limited the minimum number of patient datasets per examination type to 10. This approach was also used in previous studies [33], [34] and is regarded as appropriate, if the participation rate is high. As NUC examinations are more standardised than radiological examinations and all hospitals with NUC equipment participated, we believe the data quality in this work represents a good compromise between theoretical optimum and practicability of the data collection.

Conclusion

This work presents data on injected activities used in clinical practice for diagnostic nuclear medicine procedures in Austria to provide a base for the update of the NDRL. The injected activities found are comparable with those reported in other countries. However, for procedures for which NDRL exist, deviations in injected activities of >20% compared to the NDRL were found. Reasons therefore are assumed to result mainly from advances in technology but also from inconsistencies between NDRL and prescribed dosages as given in the information leaflets of the radiopharmaceuticals.

Funding

This study has received funding by the Austrian Federal Ministry of Health.

Conflict of interest

Michael Hinterreiter is an employee of GE Healthcare Handels GmbH (Technologiestraße 10, 1120 Vienna, AUT). The other authors declare that they have no conflict of interest.

Acknowledgements

The support of all hospitals that provided data for this study and the input of the expert group are greatly acknowledged.

Contributor Information

David Wachabauer, Email: David.wachabauer@goeg.at.

Ivo Rausch, Email: ivo.rausch@meduniwien.ac.at.

References

  • 1.United Nations Scientific Committee on the Effects of Atomic Radiation . 2008. Sources and effects of ionizing radiation. [Google Scholar]
  • 2.Duncan J.R., Lieber M.R., Adachi N., Wahl R.L. Radiation dose does matter: mechanistic insights into DNA damage and repair support the linear no-threshold model of low-dose radiation health risks. J Nucl Med. 2018;59(7):1014–1016. doi: 10.2967/jnumed.118.210252. [DOI] [PubMed] [Google Scholar]
  • 3.Doss M. Are we approaching the end of the linear no-threshold era? J Nucl Med. 2018;59(12):1786–1793. doi: 10.2967/jnumed.118.217182. [DOI] [PubMed] [Google Scholar]
  • 4.Siegel J.A., Sacks B. Eliminating use of the linear no-threshold assumption in medical imaging. J Nucl Med. 2017;58(6):1014–1015. doi: 10.2967/jnumed.117.189928. [DOI] [PubMed] [Google Scholar]
  • 5.European Commission . Unit D3 — Radiation Protection; Luxembourg: 2014. Medical radiation exposure of the European population, Radiation Protection No. 180. [Google Scholar]
  • 6.Wachabauer D., Mathis-Edenhofer S., Moshammer H. Medical radiation exposure from radiological and interventional procedures in Austria. Wiener Klin Wochensc. 2020;132(19):563–571. doi: 10.1007/s00508-019-01557-0. [DOI] [PubMed] [Google Scholar]
  • 7.International Commission on Radiological Protection 135 Diagnostic reference levels in medical imaging. ICRP Publication 135. Ann ICRP. 2017;46(1) doi: 10.1177/0146645317717209. Oxford. [DOI] [PubMed] [Google Scholar]
  • 8.International Commission on Radiological Protection 60 1990 recommendations of the International Commission on Radiological Protection. ICRP publication no. 60. Ann ICRP. 1991;21 Pergamon, Oxford. [PubMed] [Google Scholar]
  • 9.International Commission on Radiological Protection 73 Radiological protection and safety in medicine. ICRP publication no 73. Ann ICRP. 1996;26 Oxford. [PubMed] [Google Scholar]
  • 10.The Council of the European Union . Brussels; 2013. Council Directive 2013/59/Euratom. [Google Scholar]
  • 11.Radiation Protection No. 109 . Directorate-General Environment NSaCP; 1999. Guidance on diagnostic reference levels (DRLs) for medical exposures. [Google Scholar]
  • 12.Medizinische Strahlenschutzverordnung . 2020. Verordnung der Bundesministerin für Gesundheit und Frauen über Maßnahmen zum Schutz von Personen vor Schäden durch Anwendung ionisierender Strahlung im Bereich der Medizin, BGBl. II Nr. 197/2020. [Google Scholar]
  • 13.Bundesamt für Gesundheit . Bundesamt für Gesundheit, Schweiz; Abteilung Strahlenschutz: 2018. Diagnostische Referenzwerte (DRW) für nuklearmedizinische Untersuchungen. [Google Scholar]
  • 14.European Commission Diagnostic reference levels in thirty-six European countries. Rad Prot. 2014;180 [Google Scholar]
  • 15.Dundara Debeljuh D., Jurković S., Pribanić I., Poljak F., Kralik I., Krstonošić B., et al. National survey to set diagnostic reference levels in nuclear medicine single photon emission imaging in Croatia. Phys Med. 2020;78:109–116. doi: 10.1016/j.ejmp.2020.09.005. [DOI] [PubMed] [Google Scholar]
  • 16.Abe K., Hosono M., Igarashi T., Iimori T., Ishiguro M., Ito T., et al. The 2020 national diagnostic reference levels for nuclear medicine in Japan. Ann Nucl Med. 2020;34(11):799–806. doi: 10.1007/s12149-020-01512-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Song H.-C., Na M.H., Kim J., Cho S.-G., Park J.K., Kang K.-W. Korean Society of Nuclear Medicine Diagnostic Reference Level Task F: Diagnostic reference levels for adult nuclear medicine imaging established from the National Survey in Korea. Nucl Med Mol Imag. 2019;53(1):64–70. doi: 10.1007/s13139-019-00585-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Shahzad A., Bashir S., Anwar A. Establishment of age-specific reference levels and achievable doses for children and adults undergoing nuclear medicine exams. Radioprotection. 2019;54(3):187–194. [Google Scholar]
  • 19.Medizinische Strahlenschutzverordnung . 2010. Verordnung der Bundesministerin für Gesundheit und Frauen über Maßnahmen zum Schutz von Personen vor Schäden durch Anwendung ionisierender Strahlung im Bereich der Medizin, BGBl. II Nr. 197/2010. [Google Scholar]
  • 20.Stemberger A., Staudenherz A. Nuklearmedizinische Untersuchungen in Österreich - Häufigkeiten und Strahlendosen. Im Auftrag Bundesminister Gesundheit. 2008 [Google Scholar]
  • 21.Stemberger A., Leitha T., Staudenherz A. [Diagnostic reference value. Critical evaluation of the term with the example of nuclear medicine studies in Austria] Nuklearmedizin Nucl Med. 2011;50(2):68–73. doi: 10.3413/Nukmed-0379-11-01. [DOI] [PubMed] [Google Scholar]
  • 22.Ljungberg M., Pretorius P.H. SPECT/CT: an update on technological developments and clinical applications. Br J Rad. 2018;91(1081):20160402. doi: 10.1259/bjr.20160402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Slomka P.J., Patton J.A., Berman D.S., Germano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol: Off Publ Am Soc Nucl Cardiol. 2009;16(2):255–276. doi: 10.1007/s12350-009-9052-6. [DOI] [PubMed] [Google Scholar]
  • 24.Vandenberghe S., Moskal P., Karp J.S. State of the art in total body PET. EJNMMI Phys. 2020;7(1):35. doi: 10.1186/s40658-020-00290-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Spencer B.A., Berg E., Schmall J.P., Omidvari N., Leung E.K., Abdelhafez Y.G., et al. Performance evaluation of the uEXPLORER Total-body PET/CT scanner based on NEMA NU 2-2018 with additional tests to characterize long axial field-of-view PET scanners. J Nucl Med. 2020 doi: 10.2967/jnumed.120.250597. jnumed.120.250597. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Surti S., Karp J.S. Update on latest advances in time-of-flight PET. Phys Med: Eur J Med Phys. 2020;80:251–258. doi: 10.1016/j.ejmp.2020.10.031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Cabello J., Ziegler S.I. Advances in PET/MR instrumentation and image reconstruction. Br J Radiol. 2018;91(1081):20160363. doi: 10.1259/bjr.20160363. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Beyer T., Bidaut L., Dickson J., Kachelriess M., Kiessling F., Leitgeb R., et al. What scans we will read: imaging instrumentation trends in clinical oncology. Cancer Imaging. 2020;20(1):38. doi: 10.1186/s40644-020-00312-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Virgolini I., Decristoforo C., Haug A., Fanti S., Uprimny C. Current status of theranostics in prostate cancer. Eur J Nucl Med Mol Imag. 2018;45(3):471–495. doi: 10.1007/s00259-017-3882-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Chandra A., Valkimadi P.-E., Pagano G., Cousins O., Dervenoulas G., Politis M. Initiative ftAsDN: applications of amyloid, tau, and neuroinflammation PET imaging to Alzheimer's disease and mild cognitive impairment. Hum Brain Map. 2019;40(18):5424–5442. doi: 10.1002/hbm.24782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Evangelista L., Ravelli I., Magnani F., Iacobone M., Giraudo C., Camozzi V., et al. 18F-choline PET/CT and PET/MRI in primary and recurrent hyperparathyroidism: a systematic review of the literature. Ann Nucl Med. 2020;34(9):601–619. doi: 10.1007/s12149-020-01507-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Manabe O., Kikuchi T., Scholte A.J.H.A., El Mahdiui M., Nishii R., Zhang M.-R., et al. Radiopharmaceutical tracers for cardiac imaging. J Nucl Cardiol. 2018;25(4):1204–1236. doi: 10.1007/s12350-017-1131-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Wachabauer D., Röthlin F., Moshammer H.M., Homolka P. Diagnostic Reference Levels for conventional radiography and fluoroscopy in Austria: results and updated National Diagnostic Reference Levels derived from a nationwide survey. Eur J Radiol. 2019;113:135–139. doi: 10.1016/j.ejrad.2019.02.015. [DOI] [PubMed] [Google Scholar]
  • 34.Wachabauer D., Röthlin F., Moshammer H.M., Homolka P. Diagnostic Reference Levels for computed tomography in Austria: a 2018 nationwide survey on adult patients. Eur J Radiol. 2020:108863. doi: 10.1016/j.ejrad.2020.108863. [DOI] [PubMed] [Google Scholar]
  • 35.Bundesamt für Strahlenschutz . Bundesamt für Strahlenschutz; Deutschland: 2012. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für nuklearmedizinische Untersuchungen. Salzgitter. [Google Scholar]
  • 36.Wachabauer D., Stoppacher A., Mathis-Edenhofer S. Gesundheit Österreich, Vienna; 2017. Häufigkeiten medizinischer Anwendungen ionisierender Strahlung in Österreich. Analysen und Empfehlungen auf Basis des Datenjahres 2015. [Google Scholar]
  • 37.Bundesamt für Strahlenschutz . Salzgitter Deutschland; 2021. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für nuklearmedizinische Untersuchungen. BAnz AT 06.07.2021 B. [Google Scholar]
  • 38.Rausch I., Füchsel F.G., Kuderer C., Hentschel M., Beyer T. Radiation exposure levels of routine SPECT/CT imaging protocols. Eur J Rad. 2016;85(9):1627–1636. doi: 10.1016/j.ejrad.2016.06.022. [DOI] [PubMed] [Google Scholar]
  • 39.Fendler W.P., Eiber M., Beheshti M., Bomanji J., Ceci F., Cho S., et al. 68Ga-PSMA PET/CT: Joint EANM and SNMMI procedure guideline for prostate cancer imaging: version 1.0. Eur J Nucl Med Mol Imag. 2017;44(6):1014–1024. doi: 10.1007/s00259-017-3670-z. [DOI] [PubMed] [Google Scholar]
  • 40.Boellaard R., Delgado-Bolton R., Oyen W.J.G., Giammarile F., Tatsch K., Eschner W., et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imag. 2015;42(2):328–354. doi: 10.1007/s00259-014-2961-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Dam H.Q., Brandon D.C., Grantham V.V., Hilson A.J., Howarth D.M., Maurer A.H., et al. The SNMMI procedure standard/EANM practice guideline for gastrointestinal bleeding scintigraphy 2.0. J Nucl Med Technol. 2014;42(4):308–317. doi: 10.2967/jnmt.114.147959. [DOI] [PubMed] [Google Scholar]