Rapid, Multianalyte Detection of Opioid Metabolites in Wastewater

Narendra Kumar 1,#, Muhit Rana 2,#, Michael Geiwitz 3, Niazul Islam Khan 4, Matthew Catalano 5, Juan C Ortiz-Marquez 6, Hikari Kitadai 7, Andrew Weber 8, Badawi Dweik 9, Xi Ling 10, Tim van Opijnen 11, Avni A Argun 12, Kenneth S Burch 13
PMCID: PMC9949512  NIHMSID: NIHMS1867956  PMID: 35201755

Abstract

By monitoring opioid metabolites, wastewater-based epidemiology (WBE) could be an excellent tool for real-time information on the consumption of illicit drugs. A key limitation of WBE is the reliance on costly laboratory-based techniques that require substantial infrastructure and trained personnel, resulting in long turnaround times. Here, we present an aptamer-based graphene field effect transistor (AptG-FET) platform for simultaneous detection of three different opioid metabolites. This platform provides a reliable, rapid, and inexpensive method for quantitative analysis of opioid metabolites in wastewater. The platform delivers a limit of detection 2–3 orders of magnitude lower than previous reports, but in line with the concentration range (pg/mL to ng/mL) of these opioid metabolites present in real samples. To enable multianalyte detection, we developed a facile, reproducible, and high-yield fabrication process producing 20 G-FETs with integrated side gate platinum (Pt) electrodes on a single chip. Our devices achieved the selective multianalyte detection of three different metabolites: noroxycodone (NX), 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), and norfentanyl (NF) in wastewater diluted 20× in buffer.

Keywords: G-FET, opioid metabolites, wastewater, multianalyte, aptamers, LOD

Graphical Abstract

graphic file with name nihms-1867956-f0001.jpg

INTRODUCTION

Effective responses to the opioid epidemic require real-time and local information on the type and usage frequency of illicit drugs.1,2 A recent approach has emerged that avoids many of the difficulties associated with individual testing, namely, wastewater-based epidemiology (WBE). This strategy enables community tracing of drug metabolites and tracking the spread of infectious diseases.37 Wastewater monitoring can provide near real-time feedback on the introduction and continued usage of psychoactive substances without stigmatizing communities, households, or individuals.8,9 However, successful WBE requires a highly sensitive and specific detection technique as the concentrations of metabolites in wastewater are very low (pg/mL to ng/mL) due to excessive dilution.10,11 The current gold standard method to detect these drug metabolites is high-pressure liquid chromatography tandem mass spectrometry (HPLC-MS), which requires advanced equipment, sample analyses, and skilled personnel and cannot be performed on site.1214 As such, the implementation of HPLC-MS has been limited by a long turnaround time and cost. Therefore, to make WBE more meaningful and reliable, a rapid, highly sensitive, cost-effective, and easy to use detection method is required for on-site analysis of drug metabolites.

Attempts to achieve these aims have employed different techniques such as colorimetric, fluorescence, surface-enhanced Raman spectroscopy (SERS), lateral flow immunoassay (LFIA), and electrochemical detection.10,1521 In these approaches, optical detection using nanomaterial-based aptasensors has been extensively investigated for rapid analysis of illicit drugs.15 Nanomaterials are utilized to achieve high sensitivity and lower limit of detection (LOD) values, while aptamer probes possess excellent affinity, stability at room temperature, and smaller size and can be chemically synthesized on a large scale and at low cost.16,22 For example, a gold nanoparticle conjugated assay was reported with a LOD of 0.5 nM (0.15 ng/mL) and 3.3 nM (1 ng/mL) for methamphetamine and cocaine, respectively.23 The optical assay-based techniques using nanomaterials are limited by high LOD, miniaturization, complex equipment, and cost. On the other hand, the LFIA and electrochemical sensors have the capability to solve several challenges, but have yet to achieve high sensitivity and stability in real wastewater samples.17,18 For instance, a Nafion-coated carbon nanotube electrode can specifically detect oxycodone with a LOD of 85 nM (27 ng/mL), which is quite high considering the very low amount (~pg/mL) of several drug metabolites present in wastewater samples.10,13,19,24 Similarly, an LFIA-based sensor showed sensitivity (LOD) values of 5–50 ng/mL for detecting fentanyl (norfentanyl as metabolite) but were only tested with urine, phosphate-buffered saline (PBS), and saliva samples and not in wastewater.18 The LFIA still suffers from low sensitivity and quantification, while the electrochemical sensors require complex fabrication due to dependence on nanomaterial modification to achieve the desired detection limit.25,26 Moreover, most experimental approaches such as enzyme-linked immunosorbent assay (ELISA) and LFIA mostly rely on antibodies that might be incompatible with water while suffering from inconsistencies between vendors and product lots.20,21 To the best of our knowledge, graphene field effect transistors (G-FETs) with aptamer probes have not yet been implemented into a field-deployable wastewater sensor. Herein we report the development of a miniaturized G-FET platform utilizing highly specific aptamers (AptG-FET) (Figure 1) for rapid, sensitive, and simultaneous detection of drug metabolites in wastewater.

Figure 1.

Figure 1.

Schematic illustration of an on-site chip-based rapid detection platform for near real-time monitoring of opioid metabolites in wastewater using AptG-FET sensor technology. Step 1, wastewater collection, filtration, and dilution (as needed); step 2, on-chip sample (10 μL) incubation; step 3, characterization of the sensor to estimate the concentration of targeted drug metabolites.

In the past decade, G-FET-based biosensors have emerged as sensors with a large potential due to their high sensitivity, biocompatibility, noncovalent functionalization, and scalable fabrication on various substrates.2732 The electrical resistance of graphene is highly sensitive to the target bioanalytes (or the conformal changes of the probe), enabling direct and rapid readout.27,29,3336 In our recent work, we demonstrated a highly sensitive G-FET for the detection of biomarkers such as CA-I (oral diseases biomarker) in saliva37 and antibiotic-resistant bacteria, both at clinically relevant concentrations.38 However, the G-FET design in our prior work was limited to detection of a single target, with each chip functionalized with a single probe, provided minimal passivation, and required a platinum (Pt) wire as a separate reference electrode.

Thus, our previous G-FETs were not appropriate for WBE, where one requires multianalyte detection and robust devices on a self-contained chip. To achieve this, we developed a facile and reproducible fabrication process for multiple, isolated sets of G-FETs with a Pt reference electrode on a single chip (1.2 cm × 1.2 cm). We segregated the chip in four different sets of devices, and polydimethylsiloxane (PDMS) wells were mounted to functionalize the chip with four different probes for multianalyte detection of opioid metabolites (Figure 1). To further ensure the accuracy and robustness of the sensor, we kept five G-FET devices in each well to average the output signal and calculate the standard deviation. Furthermore, we eliminated the external Pt reference electrode by fabricating on-chip side gate electrodes.3941 To ensure the robustness of the G-FETs in wastewater, we passivated the devices with an aluminum oxide (AlOx) layer, exposing only the active area of graphene for selected attachment of probes. To keep the graphene surface pristine, a reliable device design and fabrication protocols were developed that ensure excellent stability and reproducibility (see Methods section). The devices were made in our cleanroom in a glovebox, significantly reducing the cost of production and fabrication time as well as limiting the exposure of the graphene to atmospheric contaminants.42

The design and fabrication enhancements enable our AptG-FET platform to simultaneously detect four different drug metabolites from a single sample of wastewater. To demonstrate this, we first validated the aptamer’s binding affinity with the respective drug metabolites in standard buffer and wastewater using plasmonic and electrochemical detection techniques. Then, one of the aptamers was functionalized over the G-FET sensor to confirm the sensitivity, selectivity, and detection limit. Finally, the multianalyte detection of all three targeted metabolites was performed on the same chip, and their sensitivity, affinity, and selectivity were tested.

RESULTS AND DISCUSSION

Selection and Validation of Aptamers.

Three different opioid metabolites were chosen in this study, namely, noroxycodone (NX), 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), and norfentanyl (NF). NX is commonly used for analgesic effects and pain management, EDDP is prescribed to overcome opioid addiction, and fentanyl is a more potent, synthetic opioid used in intravenous anesthetics.20,43,44 To ensure the selected aptamers have strong binding affinities, we performed a standard nanoplasmonic assay for characterization of aptamers binding to their specific targets. We focused on confirming the binding time, sensitivity, specificity, and dynamic detection range.

Briefly, citrate-reduced gold nanoparticles (AuNPs) possess negative charge, and the electrostatic repulsive forces account for their characteristic red color when dispersed in a medium.15,23 In the presence of a negatively charged aptamer and ~30 mM NaCl, a negative charge cloud protects AuNPs from any aggregation (Figure 2a). When an aptamer binds to the target, it leaves the particle surface, reducing the interparticle distance. Next salt-induced aggregation of the AuNPs takes place, resulting in a red-to-purple color transition in less than a minute.4547 This simple mechanism allows us to obtain quantitative binding information by monitoring the optical density (OD) at each wavelength and comparing the ratio of (OD520/OD700) (Figure 2a inset), demonstrating aptamer functionality in buffer and challenging wastewater matrices. A strong peak in OD520 is observed in the absence of the target, which upon target binding is reduced by ~40%. We have successfully used this strategy to validate aptamers for NX, NF, and EDDP in buffer. This assay is also used to determine the LOD (≥6.05 nM) and selectivity of each aptamer in the relevant wastewater media for individual target drug metabolites.

Figure 2.

Figure 2.

Aptamer validation procedure that utilizes the relative optical density change resulting from aggregation of AuNPs in the presence of targets. (a) Schematic showing the change in color of AuNPs with and without the presence of the target analyte (inset: change in OD). (b) Characterization of AuNPs functionalized with an opioid aptamer (such as NX) to obtain the sensitivity by measuring change in OD values. (c) Specificity analysis of NX-aptamer in buffer (1× PBS), very little variation observed with 100 nM EDDP and NF, while a strong decrease in OD with 100 nM NX target. (d) NX detection in filtered and 20× diluted wastewater, a minimum of 300 nM NX target required against 100 nM aptamer to achieve a significant shift in signal (inset: change in OD). All experiments were performed in triplicate (n = 3).

Using the colorimetric assay, we performed sensitivity and specificity analysis of the aptamers for NX, EDDP, and NF in 1× PBS, pH 7.4. Figure 2b demonstrates a dose-dependent linear correlation between the absorbance reading (OD520/OD700) and various NX (one of the representative opioids in our study) levels with a LOD of 6.05 nM (1.8 ng/mL). As expected, a more drastic color change was observed when a higher dose of NX was introduced in the buffer. We have observed similar findings for other opioid aptamers (EDDP and NF) with their target drug analytes (Figure S3).

To determine the specificity of these aptamers, we tested them with noncomplementary target analytes. As shown in Figure 2c, in the presence of NX-aptamer, when a noncomplementary target analyte (NF or EDDP) was introduced, the absorbance reading showed nearly no difference compared to the control (only aptamer) result. The salt-induced aggregation only takes place with the correct aptamer and NX target, and the ratio (OD520/OD700) equivalent to the aggregation state is measured. These data confirm the absence of any false positive or false negative detection and thus verified the specificity of the chosen aptamer (high rates of false positives and false negatives are major drawbacks for existing immunoassays that rely on antibody binding).20,21

Having established the sensitivity and specificity of the aptamers in buffer solutions, we turned to confirm the aptamers would work with wastewater samples. The influent (raw and untreated) wastewater samples were collected from The Massachusetts Alternative Septic System Test Center (MASSTC), located in Sandwich, MA. To avoid interference in detection due to the presence of bigger objects and species, filtering as-received wastewater samples was performed using a 0.22 μm filter.14 To achieve the successful binding of aptamers, the sample was further diluted to 1:20 in binding buffer solution (1× PBS + 2 mM MgCl2 + 1% methanol) and spiked with different concentrations of opioid metabolites. As seen in Figure 2d, a distinguishable detection of opioids in wastewater samples via the AuNPs requires at least 300 nM (90 ng/mL) NX target compared to 100 nM NX target when measured in buffer. So, the working condition of the aptamer:target is 1:1 (Figure 2c) in buffer and 1:3 (Figure 2d) in wastewater samples. The increase of LOD in wastewater samples is attributed to the interference caused by the presence of several other analytes and species. Overall, these data indicate the functionality and selectivity of all aptamers for detection of the three opioid metabolites (NX, NF, and EDDP) both in buffer and in wastewater samples (NF and EDDP shown in Figure S3).

Detection of NX with G-FET in 1× PBS.

Having confirmed the aptamer’s binding and specificity utilizing nanoplasmonic and electrochemical detection, we turned to our G-FET sensing platform. There are two different ways to operate a G-FET to perform biosensing; one is back gate and another through an ionic liquid. Traditional back-gated G-FETs offer reference electrode free devices but require substantial voltages (>60 V) with special electronics.29 Liquid-gated G-FET sensors significantly lower the required voltage (below 2 V) as well as keep the probes and analytes in their original size and conformation. Our prior work, as well as others,37,38 shows liquid gating is a reliable approach with less complex electronics required for a back-gated FET. In this work, we utilized our upgraded G-FET platform having on-chip coplanar Pt side gate electrodes,39,40 which provides a miniaturized G-FET platform and allows upscaling of the number of devices on the same chip while measuring them simultaneously (see Figure 1). A detailed description of the multiplexed G-FET design is provided in the following section, with a schematic shown in Figure 3a along with a microscopic image of the device in Figure 3b. Furthermore, to achieve selective functionalization in the channel region, a graphene sensing window of 10 μm × 40 μm was defined by depositing and selectively etching 50 nm of AlOx. This thickness of AlOx around the contact pads further improves the stability of the sensor by passivating the source/drain electrodes and minimizing leakage current. For the specific detection of the NX target, the G-FETs were first functionalized with 10 mM 1-pyrenebutyric acid N-hydroxysuccinimide ester (PBASE) linker dissolved in dimethylformamide (DMF) for 1 h and rinsed with DMF, isopropyl alcohol (IPA), and DI water. Then, NX-aptamer with 10 μM concentration was incubated for 1 h and rinsed with PBS and DI water. Raman spectroscopy confirmed the attachment of aptamers to graphene (see Figure S4). To obtain a resistance vs liquid gate voltage plot, the measurements were performed in 0.01× PBS to minimize the Debye screening effect.37 Figure 3c shows the shift in Dirac voltage (VD) after successive functionalization with PBASE linker and NX-aptamers. A further shift in VD was observed upon incubation of NX target at concentrations of 1 and 10 nM, while saturating when measured with 100 nM. Several devices have been tested to obtain the concentration-dependent shift in VD as shown in Figure 3d. No significant shift was observed at 1 pM, while an incremental shift was recorded at 10 pM and higher values of NX concentration (Figure S5).

Figure 3.

Figure 3.

(a) Schematic of fabricated G-FET. (b) Microscopic image of a single G-FET with source/drain and side gate electrode and graphene sensing window with AlOx passivation. (c) G-FET characteristics upon functionalization with aptamer probes and detection of noroxycodone. (d) Concentration dependence of Dirac voltage shift along with the shift from a high concentration of the negative control (EDDP). Error bars were calculated with the data from 3 to 5 devices at each concentration.

To understand the binding kinetics of the aptamers with NX, we fit the calibration curve to Hill’s equation: VD=VDmaxCnKDn+Cn. Here VD is the measured Dirac voltage shift at different concentrations of NX, VDmax is the Dirac voltage shift when all the binding sites are saturated, C is the concentration of NX, KD is the dissociation constant, and n is the Hill coefficient. As shown in Figure 3d, the resulting fit provides an excellent description of the concentration dependence. Interestingly, we find a KD of 94 pM, which is much better than the conventional fluorescent and HPLC techniques yet comparable to G-FET-based biosensors used for other types of biomarkers.19,23,25,48 LOD was calculated using the 3-sigma rule (3 σ/slope), where σ is the standard deviation of controls, while the slope is obtained by linearly fitting the calibration curve.47,5153 The obtained LOD from the calibration curve was approximately 10 pM, which is nearly 2 orders of magnitude higher than that obtained with the fluorescent assay and mass spectrometry19,49 and is comparable to that obtained for oxycodone utilizing electrodes modified with different complex nanostructures and immobilization processes.25,26 To confirm the specificity, we tested the NX-aptamers-functionalized G-FET with 100 nM EDDP, and only a small shift of 8 mV was seen, which is much lower than that (~60 mV) obtained with a 10 pM NX concentration (see Figure 3d). This confirmed that aptamers used for NX detection are also highly specific.

Multianalyte Detection of Three Different Targets in Wastewater.

After confirming the sensitivity and specificity of the NX-aptamer in 1× PBS, we moved to test the three different opioid metabolites (NX, EDDP, and NF) together in real wastewater samples. To test them all together, we utilized our upgraded G-FET detection platform. We note the entire platform is 5 cm × 8 cm including the G-FET chip with four wells, enabling ease of use and portability (Figure 4a). Since we just need a digital multimeter for measuring resistance (0–5 kΩ) and a voltage supply (0–2 V), the electronics can be easily miniaturized for future point-of-need diagnostics. Each PDMS well can hold 10 μL of solution. The devices in all the wells are measured simultaneously before and after functionalization. Three different wells were first functionalized with PBASE linker followed by three different aptamers, i.e., NX-Apt, EDDP-Apt, and NF-Apt, while the fourth well is used as a control. In this way, each well has an aptamer for one respective target. The as-fabricated G-FETs were tested with raw wastewater, which resulted in minimal variation in characteristics (i.e., VD, mobility, resistance), confirming the stability of the devices (see Figure S6). However, no shift in VD was observed until 1 μM of NX target because of the interference caused by several other analytes and species. So, we utilized the filtered and diluted (20× in binding buffer) wastewater spiked with different concentrations of targets. Then, the simultaneous detection of all three targets was performed. Representative plots of resistance vs voltage at each stage of functionalization and detection of different concentrations of targets are shown in Figure S7. To avoid nonspecific binding at the graphene surface and minimize the interference of different analytes likely present in wastewater, both end amine-terminated polyethylene glycol (PEG 1 kDa) was mixed in a 1:1 ratio with the aptamers during functionalization.28 There are two different ways to attach the blocking agents. One is after the incubation of aptamers, and another is together with aptamers. We chose the second method because it minimizes the number of functionalization steps, which also minimizes chances of the devices being damaged during the incubation and measurement process, as well as reducing the eventual cost of the device fabrication. Another advantage is that the PEG can be uniformly distributed between the aptamers as reported earlier by Gao et al. by using mixtures of PEG/ethanolamine and PEG/DNA aptamers for specific and sensitive detection of Prostate Specific Antigen in high-strength buffers.28 The utilization of PEG mixed with aptamers resulted in more stable behavior with minimized drift and standard deviation in calculated error bars (see data without PEG in Figure S8). The other two targets were utilized as negative controls while detecting the third target. For example, high concentrations (100 nM) of EDDP and NF were used as negative controls for the well functionalized with NX-Apt; then successive detection of different NX concentrations was performed (Figure 4b). No significant shift in VD was observed with EDDP and NF, while an incremental voltage shift was observed with increasing NX concentration. The shift in VD saturated around 100 nM, suggesting most available binding sites of aptamer probes are occupied with added integration of the target analyte. Similar trends of concentration-dependent voltage shifts were observed with the other two targets (EDDP and NF), as shown in Figure 4c and d. Juxtaposed to NX and EDDP, the devices functionalized with NF-Apt showed a higher shift (~50 mV) with NF target concentration at 10 pM, which could be attributed to the higher binding affinity of this aptamer (Figure 4d). However, the devices functionalized with NF-Apt also showed a voltage shift of ~40 mV when tested in the presence of NX and EDDP targets of concentrations (100 nM). This confirms some level of binding of other components of wastewater (Figure 4d).

Figure 4.

Figure 4.

(a) G-FET chip (1.2 cm × 1.2 cm) with 4 PDMS wells. (b) Calibration curve for NX in wastewater diluted 20× in PBS with negative controls of EDDP and NF, KD value of 490 pM, and LOD of 126 pM. (c) Calibration curve for EDDP with negative controls of NX and NF, KD value of 115 pM, and LOD of 96 pM. (d) Calibration curve for NF with negative controls of NX and EDDP, KD value of 60 pM, and LOD of 183 pM. Error bars were calculated with the data from 5 devices at each concentration.

To confirm the binding kinetics of these aptamers and targets in 20× diluted wastewater, the calibration curves obtained with all three AptG-FETs were fitted to Hill’s equation. The obtained “n” values for all three different opioid metabolites are 0.61, 0.71, and 0.53 for NX, EDDP, and NF, respectively, and agree with aptamers designed to have single binding sites.54 They all showed strong binding affinity, as evident from the resulting KD values of 490, 115, and 60 pM for NX, EDDP, and NF, respectively. The higher binding affinity (KD) of aptamers and sensitivity of G-FETs resulted in significantly lower statistical LOD values of 126 pM (38 pg/mL), 96 pM (27 pg/mL), and 183 pM (42 pg/mL) for NX, EDDP, and NF, respectively. Visually the LOD values look lower than the statically calculated values, specifically for norfentanyl, as shown in Figure 4d. This is attributed to fluctuation of the control measurements, which resulted in a higher standard deviation. These obtained values of KD and LOD are 2 to 3 orders of magnitude lower than those reported in the literature for all three drug metabolites measured by portable techniques and comparable to HPLC approaches.1,12,15,19,23 Considering the 1:20 dilutions of used wastewater, these obtained LOD values are still on the order of 100 pg/mL with the detection range of pg/mL to ng/mL, which is in line with the concentration range of metabolites present in wastewater.10,11,13,24 A detailed comparison of our results to the reported literature for real samples is shown in Table 1.

Table 1.

Comparison of the Detection of Opioid Metabolites in Real Samples (Additional Metabolites Can Be Found in References 10,11)

sensing platform opioid metabolite LOD sample type ref
LFIA norfentanyl 8 ng/ mL urine 18
LFIA norfentanyl 100 ng/mL saliva 18
ELISA fentanyl 1 ng/ mL urine 20
GC-MS EDDP 10 ng/ mL urine 44
HPLC-MS noroxycodone 0.06 ng/mL plasma 46
HPLC-MS noroxycodone 0.15 ng/mL urine 46
calorometric methamphetamine 0.15 ng/mL wastewater 23
calorometric cocaine 1 ng/ mL wastewater 23
electrochemical oxycodone 27 ng/ mL urine 55
G-FET noroxycodone 38 pg/ mL wastewater diluted 20× in PBS this work
G-FET EDDP 27 pg/ mL wastewater diluted 20× in PBS this work
G-FET norfentanyl 42 pg/ mL wastewater diluted 20× in PBS this work

Selective Detection in 20x Diluted Wastewater.

After confirming the sensitivity and specificity of the opioid metabolites in 20× diluted wastewater, their selectivity was tested. Here, the one specific target was mixed with the other two nonspecific targets, and detection was performed with G-FETs to see the hindrance in signal in comparison with that detected individually. Figure 5a shows the signal obtained from two different concentrations of NX target mixed with similar amounts of EDDP and NF; obtained values are slightly lower than that tested with NX alone as a target. A similar test was also performed for EDDP, which shows the voltage shift much closer to that obtained with a single target present (Figure 5b). However, the voltage shift with NF when mixed with NX and EDDP was significantly lower as compared to the other two targets (Figure 5c). This lower shift could be attributed to the interference of other targets, as we have already observed some level of unspecific binding with NF-aptamers. This confirms that these AptG-FETs possess a promising selectivity level along with their high sensitivity and affinity in 20× diluted wastewater.

Figure 5.

Figure 5.

(a) Selective detection of NX from mixed samples with EDDP and NF; (b) selective detection of EDDP from mixed samples with NX and NF; (c) selective detection of NF from mixed samples with NX and EDDP.

CONCLUSION AND FUTURE DIRECTIONS

In summary, we demonstrated the capabilities of aptamer probe-based G-FET sensors for rapid, selective, and simultaneous detection of three different drug metabolites in wastewater. Our AptG-FET platform provides multianalyte detection on a single chip (1.2 cm × 1.2 cm), which consists of four different PDMS wells each having five devices, on-chip coplanar side gate electrodes, and a passivation layer of AlOx. Our presented AptG-FET platform showed high specificity, sensitivity, and selectivity for all three opioid metabolites used in this work. The achieved LOD values of 38, 27, and 42 pg/mL for NX, EDDP, and NF, respectively, are well in line with the desired limits in wastewater and are 2–3 orders of magnitude better that what has been achieved with other portable techniques and are comparable to HPLC.10,11,18,19,23 All AptG-FETs have shown high binding affinity with KD values of 490, 115, and 60 pM for NX, EDDP, and NF, respectively. Thus, the presented AptG-FET platform is capable of being utilized for real-time monitoring of illicit drugs in wastewater and can provide a boost to the WBE. Our current device design is straightforward to scale to a larger number of wells for detecting an array of analytes. Although a lot of effort is involved in the G-FET-based technology, the presented platform can be easily scaled up to a 6 in. wafer, which can result in 100 chips per wafer with almost the same amount of process time. Also, in the future the linker can be preattached to the whole wafer before dicing, which will further eliminate the use of any chemicals, while aptamers can just be used in aqueous solutions. In the future, the same platform, with different probes, could be utilized for wastewater-based monitoring of a variety of analyte types including pathogens and other disease biomarkers in local health monitoring and epidemiology studies. Furthermore, our device’s design, size, rapid response, multianalyte capabilities, scalability, and ease of operation enable an upcoming era of wastewater epidemiology at the local level.

MATERIALS AND METHODS

Materials.

PBASE linker and DMF were obtained from Sigma-Aldrich. All aptamers (5′-amine-aptamer-3′: norfentanyl: CFA0071-GP5–25 AKA-H6AAZ; NX: CFA0079-GP5–25, AKA-H4LFD; and EDDP: CFA0661-GP5–25) and their resuspension buffer were purchased from Base Pair Biotechnologies, Inc., Pearland, TX, USA, which has developed aptamers that are capable of binding noroxycodone, EDDP, and norfentanyl and has readily available aptamers for morphine. Target noroxycodone hydrochloride, EDDP, and norfentanyl oxalate were purchased as ampules of 1 mL with a concentration of 1 mg/mL in methanol (as free base) from Sigma-Aldrich, St. Louis, MO, USA. Ambion DEPC-treated nuclease-free water (0.2 μm filtered and autoclaved) was purchased from Invitrogen, Thermo Fisher Scientific (Waltham, MA, USA) and utilized in all studies. To avoid any DNase contamination, DNA Away (DNA surface decontaminant) was purchased from Thermo Scientific and used before performing any experiments. All other reagents and buffers were purchased from Sigma-Aldrich. The influent (raw and untreated) wastewater samples were collected from MASSTC, located in Sandwich, MA. To avoid interference in detection due to the presence of bigger objects and species, the as-received wastewater sample was treated by the following process: initial filtering by a 0.22 μm filter,14 followed by further dilution to 1:20 in binding buffer solution (1× PBS + 2 mM MgCl2 + 1% methanol) and spiking with different concentrations of opioid metabolites. The dilution step was employed to ensure proper binding with the aptamers. In all reported LODs, the dilution has already been accounted for, such that the levels are those that would be present in the original sample.

Nanoplasmonic Assays: Functionalization and Characterization.

AuNPs were synthesized using the standard citrate reduction method.50 This nanoplasmonic test was designed according to published articles.47,5153 Briefly, 2 mL of 50 mM HAuCl4 was added into 98 mL of boiling DI water in an Erlenmeyer flask. Then 10 mL of 38.8 mM sodium citrate was added, and the mixture was stirred until the color turned wine-red. The synthesized homogeneous gold nanoparticles were characterized using UV–vis spectroscopy and stored at 4 °C. All aptamers were reconstituted in 1× PBS and 2 mM MgCl2, pH 7.4, and targets were resuspended in 1× PBS. All opioid aptamers were preheated at 95 °C for 5 min to remove any dimerization before utilizing in any experiment. For aptamer validation, 1 μL of 10 μM aptamer is added to 98 μL of 11 nM AuNPs (~13 nm size) to a final volume of 99 μL and incubated at room temperature (RT) for 5 min. After 5 min, 1 μL of 10 μM target (NX, EDDP, or NF) is added to the 99 μL of preincubated aptamer/AuNP solution for a final volume of 100 μL, resulting in final aptamer and target concentrations of 100 nM. After an additional 15–20 min of incubation at RT, 3 μL of 1 M NaCl is added to 100 μL of the nanoparticle solution to a final concentration of ~30 mM Na+. After the addition of NaCl, the color transition was observed within 1 min or less and recorded with a photograph. In the presence of salt addition, when the aptamer binds to the specific target, it desorbs from the gold nanoparticle surface, leaving gold nanoparticles unprotected and easily neutralized by Na+ and showing a color changed from red to purple. Similarly, if the aptamer does not bind to its target, the gold nanoparticle’s color will be unchanged upon salt (Na+) addition. The resulting nanoparticle assembly’s change in optical density at 520/700 nm (abs 520/700) was used to plot the aggregation rate and degree. UV–vis spectra of each sample were measured in 96-well plates using a BioTek microplate reader. Control experiments were performed in the absence of target (only aptamer + AuNP + adjusted reaction buffer, 1× PBS, 2 mM MgCl2, pH 7.4). We utilized a similar procedure for detecting and validating NX, EDDP, and NF in buffer and wastewater. In previous studies, the change in OD ratio at 520/700 nm of the resulting nanoprobe complex assembly was used to determine the limit of detection (3σ/slope) where σ is the standard deviation of controls, while slope is obtained by linearly fitting the calibration curve.47,5153

For sensitivity measurement, studies were also performed using various amounts (0, 25, 50, 100, 150 nM) of target NX in 100 μL of solution, and color changes were recorded within 1 min after incubation with ~30 mM NaCl. Control experiments were performed in the absence of target NX. We performed a similar procedure to detect opioids in wastewater where we spiked various amounts of target in the presence of fixed aptamer concentration. The OD value at 520/700 nm and the pictures of the nanoparticle suspensions were recorded. All experiments were performed in triplicate (n = 3) using 96-well plates. For specificity measurements, individual opioid aptamers with their target and/or nontarget analyte with a ratio of aptamer:target = 1:1 (for buffer) or 1:3 for wastewater were evaluated to verify false positive and false negative binding performance. The change in OD value at 520/700 nm was measured and plotted. Similar experimental procedures were employed to other opioid targets (NF and EDDP) to perform the sensitivity and specificity measurements.

G-FET Fabrication, Characterization, and Functionalization.

G-FET Fabrication and Characterization.

G-FETs were fabricated with chemical vapor deposition (CVD) of monolayer graphene transferred over SiO2/Si substrates. Monolayer graphene was grown on copper via low-pressure chemical vapor deposition. The copper foil (Alfa Aesar) was pretreated in Ni etchant (Transene) to remove any coatings or oxide layers from its surface. The tube furnace was evacuated to a read pressure of 200 mTorr with a constant flow of H2 (10 sccm). Prior to growth, the foil was annealed at 1010 °C (ramp rate 25 °C/min) for 35 min. Growth was done at 1010 °C with 68 sccm of H2 and 3.5 sccm of CH4 for 15 min. After growth, a poly(methyl methacrylate) (PMMA) layer was spin coated on one side of the copper foil and baked for 60 s at 60 °C. To facilitate smooth and fast etching of the copper foil, the backside graphene was etched using oxygen plasma with 60 W power for 60 s. The exposed copper was etched away in nickel etchant for 2 h at 60 °C. The remaining PMMA/graphene structure was washed in three DI water baths, the first and second water baths for 60 s each and the third for 30 min, to rinse away leftover etchant. To fabricate the presented G-FET platform, the source/drain along with coplanar gate electrodes were patterned on SiO2/Si chips of size 1.2 cm × 1.2 cm using a bilayer photoresist (LOR1A/S1805) and laser mask writer (Heidelberg Instruments) followed by Pt/Ti (20 nm/5 nm) deposition with an e-beam (Angstrom Engineering) and lift off using remover PG (MicroChem). To remove photoresist residues and improve the adhesion of electrodes, a 10 h baking was performed at 400 °C under vacuum, which resulted in clean and smoother electrodes. The PMMA/graphene was then transferred onto these prepatterned Pt/Ti electrodes. Any leftover water was slowly dried with argon gas, and finally the PMMA was dissolved in acetone vapors; IPA (Fisher) was used for a final wash. The chips were baked at 300 °C for 8 h under vacuum to ensure graphene adhesion and further clean photoresist residue. This was followed by deposition of 3 nm AlOx at room temperature by e-beam deposition to protect the graphene. Substrates were baked at 175 °C for 10 min before the lithography process. After that the graphene patterning was done with lithography using the same bilayer resists and then etched with oxygen plasma for 30 s at 75 W followed by 3 min of argon plasma at 100 W to remove any oxide formed over Pt gate electrodes. Devices were cleaned with remover PG, rinsed with IPA and DI water, and dried with argon followed by removal of the 3 nm AlOx layer by dipping in MF-321 developer for 30 s. Then, for electrode passivation to protect the electrodes and edges of the graphene for liquid gating, 50 nm AlOx was deposited using e-beam and AlOx crystals (Lesker) at an oxygen pressure of 7.5 × 105 mbar. Photolithography was done using S1805 to expose the sensing area (10 × 40 μm), gate electrodes, and contact pads while leaving remaining chips covered. The chips were postbaked at 120 °C for 5 min followed by AlOx etching in Transetch (Transene) for 7 h and 30 min at an 80 °C hot plate temperature. To hold the solution for experimental measurements and for functionalization, PDMS wells of size 1.5 × 1.2 mm were fabricated and placed over the chip segregating the four sets of devices with five devices in each well.

G-FET Functionalization and Measurements.

Fabricated G-FETs were functionalized with respective aptamers for specific and selective detection of opioids. G-FET chips were incubated for an optimized time of 1 h with high-concentration (10 mM) PBASE linker dissolved in DMF. Next, the G-FET was rinsed with DMF to remove adsorbed linker molecules followed by rinsing with IPA and DI water to clean the surface of solvents. The pyrene group in the PBASE linker stacks over the graphene surface through π–π interactions, while the N-hydroxysuccinimide (NHS) ester reacts with amine terminated at the 5′ end of the aptamers.23 Chips functionalized with linker were incubated for 1 h with an optimized aptamer concentration of 10 μM in PBS solution with 2 mM MgCl2 as reported in our prior work.37,38 This concentration provided the maximized surface coverage of the exposed graphene surface of the sensor, which helps to achieve high specificity and a lower LOD. Finally, the chips were rinsed with PBS to remove excess aptamers, followed by DI water to clean the salts from the graphene surface. For target detection, different incubation times were tested to maximize the binding, and 40 min was optimum to obtain the consistent signal at 10 pM concentration; the same is used for all other concentrations.

Supplementary Material

Supporting Information (Final)

ACKNOWLEDGMENTS

N.K. is grateful for the support of the Office of Naval Research under Award No. N00014-20-1-2308. The work of M.G., M.C., and K.S.B. was supported by the National Science Foundation via award DMR-2003343. M.R., N.I.K., A.W., B.D., and A.A. acknowledge the funding support from National Institutes of Health National Institute on Drug Abuse (NIDA) through SBIR Phase I grant (Award No. 1R43DA051105-01). J.O.-M. and T.v.O. were supported by the National Institutes of Health, through Grant U01-AI124302. X.L. and H.K. acknowledge the support of the National Science Foundation (NSF) under Grant No. 1945364. X.L. acknowledges the membership of the Boston University Photonics Center. We would like to thank Mr. George Heufelder, Director of MASSTC, for providing us wastewater influent samples for this study and Prof. Kurunthachalam Kannan at the Department of Environmental Medicine, NYU, for helpful discussions and guidance.

Footnotes

The authors declare no competing financial interest.

ASSOCIATED CONTENT

Supporting Information

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsnano.1c07094.

Raman spectra of graphene before and after fabrication (Figure S1); G-FET characteristics obtained from a chip having all 20 devices with four different PDMS wells (Figure S2); NF and EDDP detection results in wastewater using nanoplasmonic assay (Figure S3); Raman spectra to confirm the functionalization (Figure S4); G-FET characteristics for detection of NX in 1× PBS (Figure S5); effect of raw wastewater on asfabricated G-FET (Figure S6); NX, NF, and EDDP detection results in wastewater with G-FETs (Figure S7); calibration curve for NX detection in wastewater without using PEG (Figure S8) (PDF)

Contributor Information

Narendra Kumar, Department of Physics, Boston College, Chestnut Hill, Massachusetts 02467, United States.

Muhit Rana, Giner Inc., Newton, Massachusetts 02466, United States.

Michael Geiwitz, Department of Physics, Boston College, Chestnut Hill, Massachusetts 02467, United States.

Niazul Islam Khan, Giner Inc., Newton, Massachusetts 02466, United States.

Matthew Catalano, Department of Physics, Boston College, Chestnut Hill, Massachusetts 02467, United States.

Juan C. Ortiz-Marquez, Department of Biology, Boston College, Chestnut Hill, Massachusetts 02467, United States

Hikari Kitadai, Department of Chemistry, Boston University, Boston, Massachusetts 02215, United States.

Andrew Weber, Giner Inc., Newton, Massachusetts 02466, United States.

Badawi Dweik, Giner Inc., Newton, Massachusetts 02466, United States.

Xi Ling, Department of Chemistry, Boston University, Boston, Massachusetts 02215, United States; Division of Materials Science and Engineering and The Photonics Center, Boston University, Boston, Massachusetts 02214, United States.

Tim van Opijnen, Department of Biology, Boston College, Chestnut Hill, Massachusetts 02467, United States.

Avni A. Argun, Giner Inc., Newton, Massachusetts 02466, United States

Kenneth S. Burch, Department of Physics, Boston College, Chestnut Hill, Massachusetts 02467, United States

REFERENCES

  • (1).Ahmed SR; Chand R; Kumar S; Mittal N; Srinivasan S; Rajabzadeh AR Recent Biosensing Advances in the Rapid Detection of Illicit Drugs. TrAC - Trends in Analytical Chemistry 2020, 131, 116006. [Google Scholar]
  • (2).Nutt D; King LA; Saulsbury W; Blakemore C Development of a Rational Scale to Assess the Harm of Drugs of Potential Misuse. Lancet 2007, 369 (9566), P972. [DOI] [PubMed] [Google Scholar]
  • (3).Zarei S; Salimi Y; Repo E; Daglioglu N; Safaei Z; Güzel E; Asadi A A Global Systematic Review and Meta-Analysis on Illicit Drug Consumption Rate through Wastewater-Based Epidemiology. Environ. Sci. Pollut. Res 2020, 27, 36037–36051. [DOI] [PubMed] [Google Scholar]
  • (4).Lorenzo M; Picó Y Wastewater-Based Epidemiology: Current Status and Future Prospects. Current Opinion in Environmental Science and Health 2019, 9, 77–84. [Google Scholar]
  • (5).Gunnar T; Kankaanpää A The Practical Implications of Wastewater-Based Illicit Drug Epidemiology. Current Opinion in Environmental Science & Health 2019, 9, 49–57. [Google Scholar]
  • (6).Feng L; Zhang W; Li X Monitoring of Regional Drug Abuse through Wastewater-Based Epidemiology—A Critical Review. Science China Earth Sciences; Science in China Press, 2018; Vol. 61, pp 239–255 DOI: 10.1007/s11430-017-9129-x. [DOI] [Google Scholar]
  • (7).Choi PM; Tscharke BJ; Donner E; O’Brien JW; Grant SC; Kaserzon SL; Mackie R; O’Malley E; Crosbie ND; Thomas K. v.; Mueller JF Wastewater-Based Epidemiology Biomarkers: Past, Present and Future. TrAC - Trends in Analytical Chemistry 2018, 105, 453–469. [Google Scholar]
  • (8).European Monitoring Centre for Drugs and Drug Addiction. Assessing Illicit Drugs in Wastewater: Advances in Wastewater-Based Drug Epidemiology; Castiglioni S, Ed.; Insights 22, Publications Office of the European Union: Luxembourg, 2016; DOI: 10.2810/017397. [DOI] [Google Scholar]
  • (9).Banta-Green C; Field J City-Wide Drug Testing Using Municipal Wastewater. Significance 2011, 8 (2), 70–74. [Google Scholar]
  • (10).Subedi B; Kannan K Mass Loading and Removal of Select Illicit Drugs in Two Wastewater Treatment Plants in New York State and Estimation of Illicit Drug Usage in Communities through Wastewater Analysis. Environ. Sci. Technol. 2014, 48 (12), 6661–6670. [DOI] [PubMed] [Google Scholar]
  • (11).Sulej-Suchomska AM; Klupczynska A; Dereziński P; Matysiak J; Przybyłowski P; Kokot ZJ Urban Wastewater Analysis as an Effective Tool for Monitoring Illegal Drugs, Including New Psychoactive Substances, in the Eastern European Region. Sci. Rep. 2020, 10 (1), 4885. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (12).Castiglioni S; Zuccato E; Crisci E; Chiabrando C; Fanelli R; Bagnati R Identification and Measurement of Illicit Drugs and Their Metabolites in Urban Wastewater by Liquid Chromatography-Tandem Mass Spectrometry. Anal. Chem. 2006, 78 (24), 8421–8429. [DOI] [PubMed] [Google Scholar]
  • (13).Gul W; Stamper B; Godfrey M; Gul SW; Elsohly MA LC-MS-MS Method for Analysis of Opiates in Wastewater during Football Games II. Journal of Analytical Toxicology 2016, 40 (5), 330–337. [DOI] [PubMed] [Google Scholar]
  • (14).Foppe KS; Kujawinski EB; Duvallet C; Endo N; Erickson TB; Chai PR; Matus M Analysis of 39 Drugs and Metabolites, Including 8 Glucuronide Conjugates, in an Upstream Wastewater Network via HPLC-MS/MS. Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences 2021, 1176, 122747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (15).Mao K; Zhang H; Pan Y; Zhang K; Cao H; Li X; Yang Z Nanomaterial-Based Aptamer Sensors for Analysis of Illicit Drugs and Evaluation of Drugs Consumption for Wastewater-Based Epidemiology. TrAC - Trends in Analytical Chemistry 2020, 130, 115975. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (16).Urmann K; Modrejewski J; Scheper T; Walter JG Aptamer-Modified Nanomaterials: Principles and Applications. BioNanoMaterials; Walter de Gruyter GmbH, 2017; DOI: 10.1515/bnm-2016-0012. [DOI] [Google Scholar]
  • (17).de Rycke E; Stove C; Dubruel P; de Saeger S; Beloglazova N Recent Developments in Electrochemical Detection of Illicit Drugs in Diverse Matrices. Biosensors and Bioelectronics; Elsevier Ltd, 2020; DOI: 10.1016/j.bios.2020.112579. [DOI] [PubMed] [Google Scholar]
  • (18).Angelini DJ; Biggs TD; Maughan MN; Feasel MG; Sisco E; Sekowski JW Evaluation of a Lateral Flow Immunoassay for the Detection of the Synthetic Opioid Fentanyl. Forensic Science International 2019, 300, 75–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (19).Mynttinen E; Wester N; Lilius T; Kalso E; Mikladal B; Varjos I; Sainio S; Jiang H; Kauppinen EI; Koskinen J; Laurila T Electrochemical Detection of Oxycodone and Its Main Metabolites with Nafion-Coated Single-Walled Carbon Nanotube Electrodes. Anal. Chem. 2020, 92 (12), 8218–8227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (20).Mao CL; Zientek KD; Colahan PT; Kuo MY; Liu CH; Lee KM; Chou CC Development of an Enzyme-Linked Immunosorbent Assay for Fentanyl and Applications of Fentanyl Antibody-Coated Nanoparticles for Sample Preparation. J. Pharm. Biomed. Anal. 2006, 41 (4), 1332–1341. [DOI] [PubMed] [Google Scholar]
  • (21).Browne RW; Kantarci A; LaMonte MJ; Andrews CA; Hovey KM; Falkner KL; Cekici A; Stephens D; Genco RJ; Scannapieco FA; van Dyke TE; Wactawski-Wende J Performance of Multiplex Cytokine Assays in Serum and Saliva among Community-Dwelling Postmenopausal Women. PLoS One 2013, 8 (4), 59498. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (22).Cai S; Yan J; Xiong H; Liu Y; Peng D; Liu Z Investigations on the Interface of Nucleic Acid Aptamers and Binding Targets. Analyst; Royal Society of Chemistry, 2018; pp 5317–5338, DOI: 10.1039/c8an01467a. [DOI] [PubMed] [Google Scholar]
  • (23).Mao K; Ma J; Li X; Yang Z Rapid Duplexed Detection of Illicit Drugs in Wastewater Using Gold Nanoparticle Conjugated Aptamer Sensors. Sci. Total Environ. 2019, 688, 771–779. [DOI] [PubMed] [Google Scholar]
  • (24).Jacox A; Wetzel J; Cheng S-Y; Concheiro M Quantitative Analysis of Opioids and Cannabinoids in Wastewater Samples. Forensic Sciences Research 2017, 2 (1), 18–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (25).Hashemi P; Bagheri H; Afkhami A; Ardakani YH; Madrakian T Fabrication of a Novel Aptasensor Based on Three-Dimensional Reduced Graphene Oxide/Polyaniline/Gold Nanoparticle Composite as a Novel Platform for High Sensitive and Specific Cocaine Detection. Anal. Chim. Acta 2017, 996, 10–19. [DOI] [PubMed] [Google Scholar]
  • (26).Niu X; Huang L; Zhao J; Yin M; Luo D; Yang Y An Ultrasensitive Aptamer Biosensor for the Detection of Codeine Based on a Au Nanoparticle/Polyamidoamine Dendrimer-Modified Screen-Printed Carbon Electrode. Analytical Methods 2016, 8 (5), 1091–1095. [Google Scholar]
  • (27).Hwang MT; Landon PB; Lee J; Choi D; Mo AH; Glinsky G; Lal R Highly Specific SNP Detection Using 2D Graphene Electronics and DNA Strand Displacement. Proc. Natl. Acad. Sci. U.S.A. 2016, 113 (26), 7088–7093. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (28).Gao N; Gao T; Yang X; Dai X; Zhou W; Zhang A; Lieber CM Specific Detection of Biomolecules in Physiological Solutions Using Graphene Transistor Biosensors. Proc. Natl. Acad. Sci. U.S.A. 2016, 113 (51), 14633–14638. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (29).Ping J; Vishnubhotla R; Vrudhula A; Johnson ATC Scalable Production of High-Sensitivity, Label-Free DNA Biosensors Based on Back-Gated Graphene Field Effect Transistors. ACS Nano 2016, 10 (9), 8700–8704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (30).Nakatsuka N; Yang K-A; Abendroth JM; Cheung KM; Xu X; Yang H; Zhao C; Zhu B; Rim YS; Yang Y; Weiss PS; Stojanović MN; Andrews AM Aptamer-Field-Effect Transistors Overcome Debye Length Limitations for Small-Molecule Sensing. Science 2018, 362 (6412), 319–324. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (31).Xu S; Zhan J; Man B; Jiang S; Yue W; Gao S; Guo C; Liu H; Li Z; Wang J; Zhou Y Real-Time Reliable Determination of Binding Kinetics of DNA Hybridization Using a Multi-Channel Graphene Biosensor. Nat. Commun. 2017, 8, DOI: 10.1038/ncomms14902. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (32).Wang X; Hao Z; Olsen TR; Zhang W; Lin Q Measurements of Aptamer-Protein Binding Kinetics Using Graphene Field-Effect Transistors. Nanoscale 2019, 11 (26), 12573–12581. [DOI] [PubMed] [Google Scholar]
  • (33).Castro Neto AH; Guinea F; Peres NMR; Novoselov KS; Geim AK The Electronic Properties of Graphene. Rev. Mod. Phys. 2009, 81 (1), 109–162. [Google Scholar]
  • (34).Rahimi S; Tao L; Chowdhury SF; Park S; Jouvray A; Buttress S; Rupesinghe N; Teo K; Akinwande D Toward 300 Mm Wafer-Scalable High-Performance Polycrystalline Chemical Vapor Deposited Graphene Transistors. ACS Nano 2014, 8 (10), 10471–10479. [DOI] [PubMed] [Google Scholar]
  • (35).Seo G; Lee G; Kim MJ; Baek SH; Choi M; Ku KB; Lee CS; Jun S; Park D; Kim HG; Kim SJ; Lee JO; Kim BT; Park EC; Kim S. il. Rapid Detection of COVID-19 Causative Virus (SARS-CoV-2) in Human Nasopharyngeal Swab Specimens Using Field-Effect Transistor-Based Biosensor. ACS Nano 2020, 14 (4), 5135–5142. [DOI] [PubMed] [Google Scholar]
  • (36).Li J; Wu D; Yu Y; Li T; Li K; Xiao MM; Li Y; Zhang ZY; Zhang GJ Rapid and Unamplified Identification of COVID-19 with Morpholino-Modified Graphene Field-Effect Transistor Nanosensor. Biosens. Bioelectron. 2021, 183, DOI: 10.1016/j.bios.2021.113206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (37).Kumar N; Gray M; Ortiz-Marquez JC; Weber A; Desmond CR; Argun A; Opijnen T; Burch KS Detection of a Multi-Disease Biomarker in Saliva with Graphene Field Effect Transistors. MEDICAL DEVICES & SENSORS 2020, 3 (6), e10121. [Google Scholar]
  • (38).Kumar N; Wang W; Ortiz-Marquez JC; Catalano M; Gray M; Biglari N; Hikari K; Ling X; Gao J; van Opijnen T; Burch KS Dielectrophoresis Assisted Rapid, Selective and Single Cell Detection of Antibiotic Resistant Bacteria with G-FETs. Biosens. Bioelectron. 2020, 156, 112123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (39).Vieira NCS; Borme J; Machado G Jr; Cerqueira F; Freitas PP; Zucolotto V; Peres NMR; Alpuim P Graphene field-effect transistor array with integrated electrolytic gates scaled to 200 mm. J. Phys.: Condens. Matter 2016, 28, 085302. [DOI] [PubMed] [Google Scholar]
  • (40).Lerner MB; Pan D; Gao Y; Locascio LE; Lee K-Y; Nokes J; Afsahi S; Lerner JD; Walker A; Collins PG; Oegema K; Barron F; Goldsmith BR Large scale commercial fabrication of high quality graphene-based assays for biomolecule detection. Sens. Actuators, B 2017, 239, 1261–1267. [Google Scholar]
  • (41).Khan NI; Song E Detection of an IL-6 Biomarker Using a GFET Platform Developed with a Facile Organic Solvent-Free Aptamer Immobilization Approach. Sensors 2021, 21, 1335. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (42).Gray MJ; Kumar N; O’Connor R; Hoek M; Sheridan E; Doyle MC; Romanelli ML; Osterhoudt GB; Wang Y; Plisson V; Lei S; Zhong R; Rachmilowitz B; Zhao H; Kitadai H; Shepard S; Schoop LM; Gu GD; Zeljkovic I; Ling X; Burch KS A Cleanroom in a Glovebox. Rev. Sci. Instrum. 2020, 91 (7), 073909. [DOI] [PubMed] [Google Scholar]
  • (43).Thai PK; Lai FY; Bruno R; van Dyken E; Hall W; O’Brien J; Prichard J; Mueller JF Refining the Excretion Factors of Methadone and Codeine for Wastewater Analysis - Combining Data from Pharmacokinetic and Wastewater Studies. Environ. Int. 2016, 94, 307–314. [DOI] [PubMed] [Google Scholar]
  • (44).Larson MEM; Richards TM Quantification of a Methadone Metabolite (EDDP) in Urine: Assessment of Compliance. Clinical Medicine and Research 2009, 7 (4), 134–141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (45).Pfeiffer F; Mayer G Selection and Biosensor Application of Aptamers for Small Molecules. Frontiers in Chemistry 2016, 4 (25), 1–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (46).Mondal B; Ramlal S; Lavu PS; Bhavanashri N; Kingston J Highly Sensitive Colorimetric Biosensor for Staphylococcal Enterotoxin B by a Label-Free Aptamer and Gold Nanoparticles. Frontiers in Microbiology 2018, 9 (179), 1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (47).Rana M; Balcioglu M; Robertson NM; Hizir MS; Yumak S; Yigit M. v. Low Picomolar, Instrument-Free Visual Detection of Mercury and Silver Ions Using Low-Cost Programmable Nanoprobes. Chemical Science 2017, 8 (2), 1200–1208. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (48).Lerner MB; Matsunaga F; Han GH; Hong SJ; Xi J; Crook A; Perez-Aguilar JM; Park YW; Saven JG; Liu R; Johnson ATC Scalable Production of Highly Sensitive Nano-sensors Based on Graphene Functionalized with a Designed G Protein-Coupled Receptor. Nano Lett. 2014, 14 (5), 2709–2714. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (49).Pantano F; Brauneis S; Forneris A; Pacifici R; Marinelli E; Kyriakou C; Pichini S; Busardo FP̀. Determination of Oxycodone and Its Major Metabolites Noroxycodone and Oxymorphone by Ultra-High-Performance Liquid Chromatography Tandem Mass Spectrometry in Plasma and Urine: Application to Real Cases. Clinical Chemistry and Laboratory Medicine 2017, 55 (9), 1324–1331. [DOI] [PubMed] [Google Scholar]
  • (50).Liu J; Lu Y Preparation of Aptamer-Linked Gold Nanoparticle Purple Aggregates for Colorimetric Sensing of Analytes. Nat. Protoc. 2006, 1 (1), 246–252. [DOI] [PubMed] [Google Scholar]
  • (51).Rana M; Balcioglu M; Kovach M; Hizir MS; Robertson NM; Khan I; Yigit M v. Reprogrammable Multiplexed Detection of Circulating OncomiRs Using Hybridization Chain Reaction. Chem. Commun. 2016, 52 (17), 3524–3527. [DOI] [PubMed] [Google Scholar]
  • (52).Rana M; Augspurger EE; Hizir MS; Alp E; Yigit M v. Molecular Logic Gate Operations Using One-Dimensional DNA Nanotechnology. Journal of Materials Chemistry C 2018, 6 (3), 452–455. [Google Scholar]
  • (53).Balcioglu M; Rana M; Hizir MS; Robertson NM; Haque K; Yigit MV Rapid Visual Screening and Programmable Subtype Classification of Ebola Virus Biomarkers. Adv. Healthcare Mater. 2017, 6 (2), 1600739. [DOI] [PubMed] [Google Scholar]
  • (54).Zhang Z; Oni O; Liu J New insights into a classic aptamer: binding sites, cooperativity and more sensitive adenosine detection. Nucleic Acids Res. 2017, 45 (13), 7593–7601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (55).Mynttinen E; Wester N; Lilius T; Kalso E; Mikladal B.ør.; Varjos I; Sainio S; Jiang H; Kauppinen EI; Koskinen J; Laurila T Electrochemical Detection of Oxycodone and Its Main Metabolites with Nafion-Coated Single-Walled Carbon Nanotube Electrodes. Anal. Chem. 2020, 92 (12), 8218–8227. [DOI] [PMC free article] [PubMed] [Google Scholar]