Abstract

Graphene-based materials are of interest in electrochemical biosensing due to their unique properties, such as high surface areas, unique electrochemical properties, and biocompatibility. However, the scalable production of graphene electrodes remains a challenge; it is typically slow, expensive, and inefficient. Herein, we reported a simple, fast, and maskless method for large-scale, low-cost reduced graphene oxide electrode fabrication; using direct writing (laser scribing and inkjet printing) coupled with a stamp-transferring method. In this process, graphene oxide is simultaneously reduced and patterned with a laser, before being press-stamped onto polyester sheets. The transferred electrodes were characterized by SEM, XPS, Raman, and electrochemical methods. The biosensing utility of the electrodes was demonstrated by developing an electrochemical test for Escherichia coli. These biosensors exhibited a wide dynamic range (917–2.1 × 107 CFU/mL) of low limits of detection (283 CFU/mL) using just 5 μL of sample. The test was also verified in spiked artificial urine, and the sensor was integrated into a portable wireless system driven and measured by a smartphone. This work demonstrates the potential to use these biosensors for real-world, point-of-care applications. Hypothetically, the devices are suitable for the detection of other pathogenic bacteria.
Keywords: Graphene electrodes, Fabrication, Biosensing, Nanomaterials, Bacteria detection
Introduction
Electrochemical biosensors have many inherent advantages over other sensing modalities, including quantitative output, low cost, and high sensitivity.1−3 Graphene-based electrodes have shown great potential for developing sensitive electrochemical sensors due to the tunable electrical conductivity, high surface area, versatile functionality, and biocompatibility, with numerous applications in the field of health care, food safety, and environmental monitoring.4−6 Low defect graphene is usually produced by physical exfoliation, in which expensive/toxic solvents (N-methyl-2-pyrrolidone, dimethyl sulfoxide, and N,N-dimethylformamide) are used,6 or chemical vapor deposition (CVD),7 which involves high temperatures and tedious transferring process and has low yields, limiting the real-world applicability of CVD graphene electrodes. To address these issues, “functionalized” graphene,8−10 such as graphene oxide (GO), has emerged as an alternative for graphene electrode fabrication. However, the poor conductivity of GO restricts its use in electronic devices.11 Reduced graphene oxide (rGO) displays good conductivity and a high electron transfer rate and preserves enough oxygen groups for further functionalization.12 Chemical,13 thermal,14 and electrochemical15 methods to reduce GO to rGO have all been reported. Each results in graphene electrodes with different surface compositions and defect densities.
The patterning of graphene electrodes is important for electrochemical sensor fabrication.4,16,17 Lithographic techniques are commonly used to pattern low defect graphene;17,18 however, the need for expensive equipment and clean room facilities makes these methods inaccessible to most researchers. Template based methods, such as vacuum filtration,19 chemical covalent bonding,20 and contact transfer printing,21 have also been used to pattern graphene, but they require physically or chemically patterned templates, which are laborious to prepare and can require harmful chemicals, limiting their scalability. Printing approaches have been developed to realize the scalable production of low-cost graphene electrodes.22 However, in order to make the graphene printable, additives and fillers are often needed to adjust the rheological properties of the inks; this can affect the electrochemical performance of the final printed electrodes. Likewise, the substrates may require chemical modification to have stable and uniform ink adhesion. Additionally, high temperature thermal annealing is often required after printing,23 which can affect the material properties of the fabricated devices and limit the choice of substrates to temperature resistant materials. Since printing is a solution-based fabrication process, restacking of graphene flakes is also possible during sintering.9 This restacking, driven by van der Waals forces and π–π stacking, reduces the graphene surface area.
One promising method of patterning graphene is with lasers, to create laser-induced graphene (LIG) electrodes.24−26 It is an easy, efficient, low-cost, chemical, and mask-free method of simultaneously reducing and patterning GO (or alternatively the in situ creation of graphene). Previous works have been reported by the Kaner group,27,28 in which LIG electrodes were fabricated using a standard LightScribe DVD driver; the total cost of this process, including purchasing the laser, was less than $20. Lin et al.29 were the first to produce porous graphene via the direct laser scribing of a commercial polyimide sheet with a CO2 laser in 2014. Ever since, polyimide substrates have dominated the field, although there has been much research into alternative materials, including other polymers,16 carbon, and natural materials (such as wood and leaves) produced by different laser types.16,30,31 There have been various works in the literature that focus on not only electrode fabrication but also biosensor development on electrodes formed by laser scribing polyimide films.16,24,26 However, at the time of writing, the authors are unaware of any publications showing biosensor development on electrodes created by the laser-induced reduction of GO. The main difficulties that prevent the development of biosensors by this method are as follows. (1) The reduction happens in situ, meaning that there will be GO and rGO in the same plane. Since GO is highly soluble in water, it will dissolve in any biological media.9,32 (2) LIG is very fragile, even being damaged by gentle gas flows.33 To overcome these issues, the Nam group developed a method of transferring LIG from polyethylene terephthalate (PET), but this method is only applicable to substrates that do not absorb the excitation laser.34 The Merkoçi group recently reported a novel, stamp transfer method, capable of fabricating rGO electrodes on a wide variety of substrates.35 In this method, a GO film is formed on a polyvinylidene fluoride filter membrane by vacuum filtration, before being simultaneously patterned and reduced by a LightScribe DVD driver. The LIG was then transferred to various substrates with a mechanical press. In this process, the rGO is transferred and the GO is not, while the LIG is compressed, improving its mechanical properties. This proof-of-concept work only verified that the transfer stamped LIG electrodes were capable of electrochemical sensing, not biosensing. In addition, the process required a vacuum pump, could only produce small batch sizes, required expensive filter membranes, and was relatively slow.
Herein we report advancements on our previous work on laser reduced graphene oxide (LRGO) electrode fabrication, to make the production simpler and more scalable, as well as showing their biosensing ability. The LRGO working electrodes were stamped onto polyester (PE) sheets ,and silver inks were inkjet-printed to create silver contacts and Ag/AgCl reference electrodes. A LRGO electrode based biosensor is demonstrated for the first time; as a proof-of-concept, it was developed for the detection of Escherichia coli (E. coli) by creating an electrochemical enzyme linked immunosorbent assay (ELISA). The sensor was validated in both phosphate buffered saline (PBS) and artificial urine (AU), with a commercial potentiostat and a portable, wireless system driven by a smartphone,36 displaying its capability for point-of-care applications.
Experimental Section
Materials
GO solution (GO water dispersion, 10 mg/mL) was purchased from Graphenea (San Sebastian, Spain). Polyester (PE) sheets (AUTOSTAT HT1802168201) were obtained from MacDermid Autotype Ltd. (Wantage, UK). Phosphate buffered saline tablets, Tween 20, and Tryptic Soy Agar were ordered from Merck Life Science S.L.U (Madrid, Spain). Phosphate buffered saline buffer (PBS, 10 mM, pH 7.4) was prepared by dissolving PBS tablets in ultrapure water. PBS with Tween 20 (PBST) buffer was prepared by adding Tween 20 (final concentration 0.05 wt %) in PBS buffer (10 mM). Polyclonal anti-E. coli antibody (PA1–7213) was obtained from Fisher Scientific (Madrid, Spain). HRP labeled antibodies (ab20425) and 3,3′,5,5′-tetramethylbenzidine (TMB) substrate (ab171523) were purchased from Abcam plc (Cambridge, UK). Escherichia coli (E. coli, CECT 4972) and Staphylococcus aureus subsp. Aureus (SA, CECT 5190) were purchased from CECT (Valencia, Spain), and Salmonella typhimurium (ST, ATCC 14028) from ATCC (Virginia, USA).
LRGO Electrode Production
PE substrates were washed with isopropanol (IPA) and dried with nitrogen or compressed air. Afterward, 5 mL of GO solution was drop-casted on an 8 × 8 cm2 square defined on the PE sheet with paper tape (to prevent the solution from spreading), shaken gently by hand, and then dried at 60 °C for 2 h. A CO2 laser engraver (Rayjet 50 Laser Engraver, Trotec Laser GmbH, Marchtrenk, Austria) was used to scribe the GO film to produce laser reduced graphene oxide (LRGO). The laser settings were as follows: power = 2.07 W and speed = 1.5 m/s. The LRGO was placed face down on the desired substrate, fixed tightly in place with scotch tape, sandwiched between four sheets of A4 paper (2 layers above and 2 layers below the sample), and placed in a hydraulic press (SPECAC manual hydraulic press 15 ton, UK). A 6-ton force was applied for 1 min to transfer the LRGO onto the hosting substrate, forming both the working electrode (WE) and counter electrode (CE). Silver contacts were inkjet printed with a Dimatix 2831 (FUJIFILM Dimatix, USA. Print temperature 30 °C, plate temperature 60 °C, drop space 25 μm, 3 layers), using a silver ink (Sicrys I40DM-106, PV Nano Cell Ltd., Migdal Ha’Emek, Israel) and then oven-cured at 160 °C for 2 h. Polydimethylsiloxane (PDMS, SYLGARD 184 silicon elastomer kit, Dow corning, MI, USA) was used as an insulating layer; a schematic summary of the insulation process has been included in Figure S1. Ag/AgCl pseudoreference electrodes (RE) were obtained by chlorinating the printed Ag trace with 0.4% NaClO for 3 min.37
Characterizations of LRGO
Cyclic voltammetry (CV) was done with a commercial potentiostat (PalmSens4, PalmSens BV, Houten, Netherlands), in an aqueous solution of 0.5 mM potassium ferrocyanide, 0.5 mM potassium ferricyanide, and 0.1 M KCl at scan rate of 50 mV/s. For defining the electrochemical active surface area (ECSA), CVs were performed in 0.1 M NaClO4 (in ethanol) by using a 3 mm diameter LRGO disk as the WE, four LRGO squares (Each 1 cm2) as CEs, and a commercial Ag/AgCl electrode (CHI111, CH Instruments, Inc.) as the RE. The height of the films was measured by an optical profilometer (Profilm 3D, Filmetrics Inc.) equipped with a 10× objective and using a stage movement speed of 0.12 mm/s. Scanning electron microscopy (SEM) was performed with a Quanta 650 (UK) at a working distance of 10 mm and voltage 10 kV; the samples were mounted on aluminum stubs and sputtered with Au prior to imaging. The sheet resistance was obtained by measuring square (1 × 1 cm2) samples with a Keithley 2400 using a four-point probe system (van der Pauw method). Raman spectra were obtained with a Raman microscope (Alpha300 R-Raman Imaging Microscope, Oxford Instruments) using a 488 nm laser with a spot size of approximately 3 μm. The parameters used were as follows: laser power = 1 mW, grating = 600 gr/nm, objective = 50×, exposure time = 10 s, accumulations = 2, waveform range = 500–3500/cm. X-ray photoelectron spectroscopy (XPS) measurements were performed at room temperature with a SPECS PHOIBOS 150 hemispherical analyzer (SPECS GmbH, Berlin, Germany) at a base pressure of 5 × 10–10 mbar using monochromatic Al Kα radiation (1486.74 eV) as the excitation source, operated at 300 W. The energy resolution as measured by the full width at half-maximum (fwhm) of the Ag 3d 5/2 peak for a sputtered silver foil, and was found to be 0.62 eV. The fitting of the obtained spectra was done with XPSPEAK 41 with a Shirley background correction.
LRGO Electrode Functionalization
The LRGO working electrode was activated with 1-pyrenebutanoic acid succinimidyl ester solution (PBASE, 1 μL of 2 mg/mL in dimethylformamide) at room temperature for 4 h. The electrode was then cleaned twice with IPA and four times with ultrapure water. The capture antibody (cAb, 15 μg/mL in 10 mM PBS) was spontaneously immobilized by incubating 5 μL of the solution on the WE (4 °C, overnight). Ethylamine (4 μL, 0.1 M in PBS) was put on the WE for 30 min to block any unreacted PBASE, and then the electrode was further blocked with bovine serum albumin (BSA, 5 μL, 5% in PBS) at 37 °C for 1 h to prevent nonspecific fouling in complex media. Between each step, the electrode was washed twice with PBST and twice with PBS; all the steps above were performed in a humidity chamber.
Bacteria Preparation
Bacteria were cultured overnight in Tryptic Soy Agar at 37 °C. Afterward, colonies were picked and suspended in PBS. Then the bacterial solutions were diluted to obtain an OD600 value as close to 109 CFU/mL as possible.38 Finally, the bacteria were heat-killed at 65 °C for 20 min.39
Electrochemical Bacterial Detection
To perform the bacteria detection, 5 μL of the bacteria-containing solution was pipetted onto the WE, left for 30 min, and then washed twice with PBST and twice with PBS. Directly after cleaning, 4 μL of the detection antibody solution (dAb, ab20425, 2 μg/mL) was pipetted onto the WE and left for 30 min, and then washed twice with PBST and 4 times with PBS. Subsequently, 50 μL of the TMB ELISA solution was added, and the chronoamperometric sensing started. Chronoamperometry (CA) was performed at +0.125 V, vs the on-board Ag/AgCl RE, with both the PalmSens4 and a wireless smartphone driven potentiostat developed in house and fabricated according to a previous publication.36
ELISA Bacterial Detection
The ELISA test was performed according to a previously established protocol with minor modifications.40 First, 100 μL of capture antibody (PA1–7213, 2 μg/mL in Carbonate-Bicarbonate Buffer, pH 9.6) was aliquoted into a 96-well ELISA plate (44–2404–21, MaxiSorp, Fisher Scientific, Spain) and incubated at 4 °C overnight. Second, 200 μL of BSA (3% in PBS) was added into the wells and incubated at 37 °C for 1 h, to prevent nonspecific fouling of the wells. Third, 100 μL of bacterial samples was added and kept at room temperature for 1h. The wells were washed three times with PBST between each step, and then incubated with 100 μL of HRP-labeled detection antibodies (HRP-dAb, 0.33 μg/mL ab20425, in PBS) for 1 h at room temperature. Finally, the ELISA wells were washed 5 times with PBST. Afterward, 100 μL of the TMB substrate solution was added into each well and left for 10 min at room temperature. Finally, 50 μL of 2 M H2SO4 was added into the wells and then the absorbance of each well was detected at 450 nm using a SpectraMax iD3Multi-Mode Microplate Reader (Molecular Devices, USA).
Results and Discussion
In this work, all the digital designs were created with AutoCAD 2018 (Autodesk, USA) and then sent to the laser engraver/inkjet printer to produce the patterns, this direct writing strategy along with the stamp-transferring method enables the simple, fast, maskless and large-scale production of LRGO electrodes (Figure 1). SEM image of the GO film surface (Figure 2a), shows that the surface has the bumps and ravines typical of graphene-based materials.9,32,41 After laser treatment, the film becomes a foam-like, porous structure, composed of flakes 10s of microns long (Figure 2a and b). Additionally, the GO changed from a brown color to black (Figure 1, i, ii) upon reduction, in agreement with the literature.35 Although the LRGO surface is smoother and less porous after the stamp-transferring process, it still exhibits a highly porous 3D framework (Figure 2c and d). The data from the optical profilometer measurements, Figure S2, show that the GO film is 3–4.5 μm thick; after laser irradiation, the surface becomes much more heterogeneous but noticeably increases to a maximum height of 18.2 μm in some areas. The laser scribing of GO is a photothermal process; although the exposure time is only microseconds, the instantaneous temperature at the surface is over 1000 °C,16 which causes the surface to rapidly expanded and blister, this is proposed as the root cause for the changes in surface morphology in these experiments. A large pressure is applied to the film so it is pressed flat during transferring, hence the decrease in thickness to 0.4–1.3 μm.
Figure 1.
(a) Schematic of the laser reduced graphene oxide (LRGO) electrode fabrication process. (i) A graphene oxide (GO) film was formed on a polyester (PE) sheet by drop casting GO solution in a defined area and drying at 60 °C; (ii) laser scribing to reduce and pattern the GO film, creating LRGO; (iii) LRGO was placed face down on the desired substrate; (iv) pressure was applied to stamp the LRGO; (v) after separation, a mirror image of the LRGO had been transferred; (vi) inkjet printing of silver connections; (vii) polydimethylsiloxane (PDMS) was cast as a dielectric layer; (vii) NaClO was used to convert Ag to Ag/AgCl as reference electrode. (b) Photographs corresponding to each of the steps shown in (a); (ix) magnification of the final electrode, consisting of LRGO working electrode (black dot), LRGO counter electrode (black arc) and Ag/AgCl pseudoreference electrode (dark brown arc) and PDMS insulation layer inside the dashed square.
Figure 2.

(a) SEM image of the boundary between the LRGO and GO film, showing the morphological and topographical changes in the graphene upon reduction. (b) Magnified SEM image of the LRGO showing the flake-like structure. (c) SEM image of the LRGO after transferring onto a polyester (PE) sheet, dubbed T-LRGO. (d) Magnified image of the compressed flake-like structure of T-LRGO on PE; inset is a high magnification of the same region to evidence the flake structures.
GO, LRGO and T-LRGO were characterized via XPS, in all these samples the C and O peaks dominated (Figure 3a). The increase in the C/O ratio when the GO (2.7) is reduced to LRGO (22.0) is expected and confirms that the laser reduction was successful. During the stamp-transferring step, the surface of the LRGO is placed in contact with the new substrate and the surface of the T-LRGO is, what was once, the bulk of the LRGO. Therefore, the similar C/O ratio between T-LRGO (23.4) and LRGO (22.0) indicates that the laser-induced reduction is not a surface confined process but penetrates into the bulk of the GO. The high-resolution C 1s spectra of GO (Figure 3b) display peaks at 284.4, 286.4, and 287.7 eV, which can be assigned to sp2 hybridized carbon in the graphene lattice (C–C), sp3 carbons in the form of epoxide and hydroxyl carbons (C–O) and sp3 carbons in the form of carbonyl groups (C=O), respectively.28,42,43 Further analysis shows the relative peak areas are 39.0% (284.4 eV), 46.8% (286.4 eV), and 13.9% (287.7 eV), Table S1; the high proportion of oxygen containing functional groups shows the degree of oxidation of the GO. In the C 1s spectra of LRGO (Figure 3c), the presence of a disordered carbon peak at 285.0 eV (25.6%)28,43 with a small C–O peak (8.8%) and the disappearance of the C=O peak further proves the effective reduction of the GO by the laser scribing. The increase in sp2 hybridized carbon content (C–C, 56.4%) and significant increase in the π–π* (at 291.2 eV, 9.2%; and 0.3% in the case of GO) satellite peak shows that the graphene lattice has been restored upon reduction of the GO. The data from the LRGO and T-LRGO samples are the same within error, showing that the reduction is uniform to the transferred graphene depth and that the transferring has no effect on the chemistry of the graphene. The degree of reduction GO reduction is clearly very high for this process, given all the above, but was further proven by the improved conductivity of the materials, the GO had a sheet resistance of >20 MΩ/sq,27 which decreased to 90.6 ± 3.9 Ω/sq for LRGO (the thickness was assumed to be 1 μm).
Figure 3.

(a) XPS survey and high-resolution C 1s spectra of (b) GO, (c) LRGO, and (d) T-LRGO. The survey spectra show a significant decrease in O content after laser radiation; C 1s spectra of GO reveals an abundance of oxidized carbon C–O (286.4 eV) and C=O (287.7 eV); the dominance of the C–C (284.4 eV) bond and presence of π–π* (291.2 eV) in LRGO and T-LRGO proves GO has been reduced by the laser.
The Raman spectroscopy data (Figure 4a) shows that the characteristic D (∼1345 cm–1) and G (∼1580 cm–1) bands of graphene are present in all three samples, with intensity ratios (ID/IG) of 1.00 ± 0.02 for GO, 0.41 ± 0.07 for LRGO and 0.99 ± 0.01 for T-LRGO (n = 6). The 2D band, at ∼2700 cm–1, was only present in the LRGO samples. Previous publications41,44,45 have assigned the D peak to the primary in plane vibrations of graphene, so it can be used to estimate the degree of defects and disorder within the hexagonal lattice. The 2D peak is attributed to second-order in-plane vibrations of graphene. The significant decrease of ID/IG after laser exposure is therefore indicative of rGO formation, which is in accord with the XPS and sheet resistance findings. The normalized average 2D peaks from 5 LRGO spectra recorded at random locations on a single working electrode surface, Figure 4b, show small standard deviations; this indicates a homogeneous reduction of GO across the electrode surface. The full width at half maximum (FWHM) of the 2D bands were extracted using the Lorentzian function (Figure 4b), for the LRGO the FWHM was found to be 74.97 ± 2.93 cm–1, indicating that the LRGO film consists of only a few graphene layers.28,46,47 The Raman spectra of the T-LRGO are the same within error as that of the GO. It is proposed that there are residual amounts of unreduced GO on the surface of the transferred electrodes. Raman spectroscopy is a surface measurement so the GO lying on the surface of the electrode dominates the signal. This was not observed in the XPS since it has much larger sampling area (spot size 100s μm, and 3 μm in the case of Raman) and probes the chemistry up to a certain depth (depending on the substrate and experimental conditions), reporting the average chemical composition of a volume at the surface. We can therefore surmise that there is a thin layer of GO on the surface of the T-LRGO.
Figure 4.

(a) Representative Raman spectra of GO, LRGO, and T-LRGO on PE, showing the typical D (∼1345 cm–1) and G (∼1580 cm–1) bands of graphene; the 2D band and decreased ID/IG ratio in LRGO are indicative of GO reduction by laser scribing. (b) Average 2D band from 5 measurements of LRGO (shaded area refers to the standard deviation of the average), showing that the LRGO contains only a few graphene layers, i.e., is not multilayered graphene. (c) Cyclic voltammograms obtained in a solution of 0.1 M NaClO4 in ethanol vs a LRGO CE (4 cm2) and a Ag wire pseudoreference electrode, at a series of scan rates shown in the inset. (d) Current density from (c) plotted as a function of the scan rate, to determine the electrochemical active surface area (ECSA) of the electrodes. The black dots represent raw data and red line is a linear fit to the data.
The surface area, roughness, and porosity of electrodes can directly influence the performance of electrochemical sensors.1,25 To ascertain the electrochemically active surface area (ECSA) of the LRGO electrodes, non-Faradaic CV was performed in 0.1 M NaClO4 in ethanol. The CVs of the electrodes at scan rates from 20 to 200 mV/s are provided in Figure 4c. Plotting current density (j) against the scan rate (υ) (Figure 4d), a linear relationship was obtained, in which the slope of the linear plot is equal to the area specific double layer capacitance (CS), and the CS of the electrode is directly related to its ECSA through eqs 1–3:48
| 1 |
| 2 |
| 3 |
where S is the geometric surface area of the electrode; υ is scan rate; I is the current; j is current density; Cref is the area specific double-layer capacitance of a flat reference surface.
Equationss 1–3 and the data in Figures 4c,d and S3 were used to calculate the ECSA of the electrodes based on the Cref of graphite (∼20 μF/cm2);49,50 the average electrode area from three repeats was found to be 0.781 ± 0.022 cm2. The ECSA is more than 10 times larger than the geometric area (0.07 cm2), which is expected given the highly rough and porous nature of the electrodes. The standard deviation of the three repeats speaks to the reproducibility of the LRGO working electrodes.
Inkjet-printed silver was used to create electrode contacts and to fabricate the reference electrode.37 Open circuit potential measurements of the printed quasi-reference electrodes (N = 4) vs a commercial Ag/AgCl electrode (CHI111, CH Instruments, Inc.) exhibit potential fluctuations of less than 4 mV over 200 s, and a potential difference of less than 14 mV (Figure S4a), verifying the functionality of the prepared reference electrodes.
Figure 1b, ix shows a photograph of the final electrodes, consisting of LRGO working and counter electrodes, and Ag/AgCl quasi-reference electrode. The electrodes are made in batches; it is therefore prudent to assess the reproducibility of the electrochemical signal obtained from multiple batches. To do this, 3 batches of electrodes were fabricated and 6 electrodes from each batch selected at random for testing (a total of 18 electrodes were used). These were first modified with PBASE to make them hydrophilic, and CVs were recorded in 0.5 mM [Fe(CN)6]4–/[Fe(CN)6]3– with 0.1 M KCl. The peak potentials were identical for all the electrodes tested, which further shows the reproducibility of the quasi-reference electrode fabrication (Figure S4b). The relative standard deviation (RSD) of the anodic peak current shows the intrabatch variation is 4.2%, and the interbatch variation is 5.6%.
To prove that the electrodes have the potential to be used for biosensor development, an immunoassay for the detection of pathogenic bacteria was developed on them. E. coli was used as a model bacteria in these proof-of-concept tests, since their detection is important for medical, food safety, and environmental monitoring.39,51 First a conventional ELISA was run to select the best capture and detection antibody pairs and act as a reference with which the performance of the T-LRGO electrodes could be compared. The control ELISA was capable of detecting E. coli between 9.35 × 104–9.06 × 106 CFU/mL and had a limit of detection (LOD) of 1.30 × 104 CFU/mL (Figure S5a).
The same antibodies used in the conventional ELISA were then used to functionalize the T-LRGO and optimized to create a point of care biosensing system, that could be wirelessly driven by a smartphone, Figure 5a (see electric scheme in Scheme S1).36Figure 5b shows a schematic representation of the electrochemical sensing mechanism. In brief, HRP catalyzes the generation of OH radicals from H2O2, and these radicals oxidize TMB. The oxidized TMB (oxTMB) can be detected optically or electrochemically. When the oxTMB is reduced, the reductive current generated is proportional to the amount of oxTMB present. Given that the amount of oxTMB is determined by the amount of dAb, which is proportional to the amount of E. coli bound to the electrode surface, the reductive current gives a quantitative evaluation of the amount of E. coli in the sample. In other words, the amount of oxTMB is a directly proportional of the amount of E. coli bound to the electrode.
Figure 5.

(a) Smartphone controlled portable wireless system for E. coli detection (not to scale). (b) Illustration showing the working mechanism of the electrochemical enzyme linked immunosorbent assay (ELISA). (c) Cyclic voltammograms of TMB on the functionalized T-LRGO electrode. (d) Representative current vs time transients of various E. coli concentrations, recorded at +0.125 V vs Ag/AgCl.
Figure 5c displays a CV of TMB on the functionalized LRGO electrode. Oxidation peaks at +0.231 and +0.411 V and reduction peaks at +0.166 and +0.367 V can all be easily distinguished (all potentials vs the on-chip Ag/AgCl electrodes). The current time transients obtained after incubating the functionalized electrodes with E. coli concentrations between 0 and 109 CFU/mL are provided in Figure 5d. As expected, the normalized dose–response curve in Figure 6a shows a sigmoidal response. With the 4-Parameter Logistic model fitting52,53 the LOD was calculated to be 283 CFU/mL (blank + 3SD) with a detection range between 917–2.1 × 107 CFU/mL (lowest and highest concentrations of the analytes based on 10% and 90% of the maximum signal respond on the calibration curve). The improved sensing performance of the LRGO electrode based platform compared with the standard optical ELISA(Table S2), is in accord with other reported electrochemical ELISA platforms.52 The performance of the sensor is suitable for the detection of E. coli in urine at clinically relevant concentrations (103 – 105 CFU/mL)54,55 and it is advantageous that all the results were obtained from 5 μL samples. Ultrasensitive detection, even down to single cell, has been demonstrated,56−58 but tedious material synthesis and complicated sensor construction are typically required (Table 1). Our platform showed comparable performance with other electrochemical sensors,59−61 and could be improved with further optimization of the antibody concentrations (this work is mainly focused on the fabrication and characterization of the electrodes, with the sensor construction being merely a proof-of-concept).
Figure 6.

(a) Calibration curve of E. coli concentration vs normalized current response from the chronoamperometry at +0.125 V; the black stars were recorded with a commercial potentiostat in PBS, the green diamonds were recorded with the smartphone driven potentiostat developed in house, and the blue circles were obtained with a commercial potentiostat in artificial urine (AU). (b) Selectivity studies showing the current response without bacteria (orange), with Escherichia coli (green), Staphylococcus aureus (purple), Salmonella typhimurium (yellow), Escherichia coli and Staphylococcus aureus (blue), and Escherichia coli and Salmonella typhimurium (pink). In all cases, each bacterial concentration was 104 CFU/mL, with a 50:50 mix of each in the last two conditions. (c) Control experiments run either without the cAb, E. coli or without both. (d) Data from the stability studies of the sensors stored in different conditions for 1 month, measuring the current response to 105 CFU/mL of E. coli.
Table 1. Representative Nanomaterial-Based Biosensors for E. coli Detection Reported in Literaturea.
| platform | technique | detection range (CFU/mL) | LOD (CFU/mL) | response time (min) | sample volume (μL) | ref |
|---|---|---|---|---|---|---|
| metal NPs-LIG | EIS | 1 × 102–1 × 108 | 102 | 30 | (59) | |
| silica NPs | CV | 8 × 104–8 × 106 | 2 × 103 | 30 | 10 | (61) |
| Au NPs | LFA | 104–106 | 104 | 10 | 150 | (39) |
| Fe3O4@SiO2/ polymer/fluorescein | fluorescent | 4–4 × 108 | 3 | 75 | 1000 | (56) |
| MWCNTs/chitosan/thionine | CV | 102–109 | 102 | 180 | 200 | (60) |
| N-GQDs/MIP | ECL | 10–107 | 5 | 140 | (57) | |
| rGO/Al2O3/Au NPs | FET | 1–100 | single cell | 50 s | 1 | (58) |
| MoS2/Au/optical fiber | SPR | 103–8 × 109 | 94 | ∼15 | (62) | |
| LRGO | CA | 917–2.1 × 107 | 283 | 60 | 5 | this work |
NPs = nanoparticles; N-GQDs = nitrogen dopped graphene quantum dots; MWCNTs = multiwall carbon nanotubes.
To test the selectivity of the sensors, Salmonella enterica serovar Typhimurium (ST) and Staphylococcus aureus subsp. Aureus (SA) were run in place of and alongside E. coli. The results of these tests are given in Figure 6b, where it can be seen that neither of these bacteria gave a signal significantly above that of the blank. Likewise, when run alongside E. coli, the responses are the same within error as those of E. coli alone. The same trends were seen in the ELISA tests, Figure S5b; proving that the sensor is specific and that specificity is attributable to the antibodies. Other control experiments, Figure 6c, showed that the signal response is negligible (−0.083 ± 0.034 μA, N = 4) in the absence of E. coli and cAb; it increases to −0.349 ± 0.010 μA when only E. coli is excluded (blank control, N = 4), indicating that there is some cross-reaction between the antibodies, which is common when working with polyclonal antibodies.63,64 When only the cAb was omitted, the current (−0.391 ± 0.108 μA, N = 4) was the same within error as the blank control, indicating insignificant nonspecific binding of E. coli to the electrodes.
The stability of the sensors was tested in different storage conditions. After functionalization, batches of the sensors were stored in humidity chambers placed at room temperature (taken as 20 °C), in the fridge (4 °C) and in a freezer (−20 °C), all under N2. The E. coli detection was performed at several time points over 30 days, Figure 6d (N = 4). The signal variability at 20 and 4 °C was found to be similar (RSD ≤ 7.6%) within the first 10 days, comparable to that of the variation in the electrodes calculated earlier (5.6%); even stored for 30 days, the sensors showed similar performance in all cases (RSD of the current response, 12.8% for 4 °C, 13.0% for 20 °C, 14.0% for −20 °C); these data show that the sensors are relatively stable for 30 days and the storage temperature makes no significant difference to the sensor stability. The slightly higher RSD for the sensors stored at −20 °C may come from the freeze–thawing cycles when taking them in and out of the freezer, which could damage the antibodies.65
The efficacy of the sensors was further evaluated in spiked artificial urine. The results of these tests are shown in Figure 6a (the blue circles) and correlate very well with those obtained in PBS. The recovery rates obtained are 119% for 500 CFU/mL, 88% for 105 CFU/mL, and 90% for 106 CFU/mL of E. coli, showing that the system has the potential to be used in real world matrices.
In order to create a point-of-care (POC) test, there is a need for sensitive, inexpensive, wireless, portable detection platforms.6,66−68 A previously established smartphone-based wireless system36 was used to make this test more suitable for POC applications. Three concentrations of E. coli in PBS were tested with said wireless system and compared to a commercial potentiostat. Movie S1 shows the real-time response of the system. The current responses obtained for three E. coli concentrations are the same within error as those recorded with a commercial potentiostat (Figure 6a, green diamonds).
The scalability of the electrode fabrication has been proven by the production of ∼500 electrodes in 1 day, although the whole fabrication process for a single batch of 12 electrodes takes ∼4.5 h. In fact, more than 2000 electrodes could be made under ideal conditions, as calculated from Table S3. It is also worth mentioning that the cost of GO is estimated at € 0.06 per electrode, 4 times cheaper than polyimide. The total materials cost per electrode is € 0.09, and the whole sensor costs less than € 0.12 (Table S4). While the CO2 lasers used herein are not common laboratory equipment, there are other low-cost laser sources with wavelengths ranging from 248 nm to 10.6 μm that can be used for LRGO formation.16,25,27,35,69 The printing of Ag connections and the reference electrodes can also be realized by much cheaper methods, such as consumer inkjet printers, screen/stencil printing, or simply painting silver inks. These features mean that T-LRGO based sensing platforms can be produced in most research laboratories around the world.
Conclusions and Future Work
In this work, we reported a cheap and scalable method of fabricating LRGO-based electrodes on polyester substrates. The biosensing potential of the electrodes was proven by functionalizing them for the detection of E. coli in PBS and artificial urine. This proof-of-concept sensing platform could detect E. coli between 917 and 2.1 × 107 CFU/mL, with a LOD of 283 CFU/mL, within the clinically relevant range for E. coli in human urine. We believe that the platform can be utilized for the detection of other bacteria or biomarkers, simply by changing the antibodies.
Being able to transfer these graphene-based sensors onto different substrates would be of interest to other sensing applications, such as wearable sensing.6,36,44 An initial study showed that these LRGO films could be transferred onto a variety of other substrates, Figure S6, demonstrating the potential applicability of these electrodes. It is also of note that the maximum resolution of the laser will define the resolution with which electrodes can be patterned. With our equipment this was down to the micron scale, making microelectrode development with this technology possible. Future work will include the development of microelectrode arrays by this method and the optimization of printing and stamping onto other substrates for different applications.
Acknowledgments
ICN2 is funded by CERCA programme, Generalitat de Catalunya. Grant SEV-2017-0706 funded by MCIN/AEI/10.13039/501100011033. Lei Zhao, Liming Hu and Qiuyue Yang acknowledge the financial support from China Scholarship Council (CSC). F.D.P. acknowledge the Ministry of Education, University and Research (MIUR) and European Social Fund (ESF) for the PON R&I 2014-2020 program, action 1.2 ’AIM: Attraction and International Mobility’ (AIM1894039-3). Cecilia C. C. Silva acknowledges funding through CAPES – PRINT (Programa Institucional de Internacionalização; grant #88887.310281/2018-00 and 88887.467442/2019-00) and Mackpesquisa-UPM. This project has received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No 825694. Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union. The European Union can not be held responsible for them. We acknowledge Miguel Angel Aroca and Johann F. Osma (both from Department of Electrical and Electronic Engineering, Universidad de los Andes, Bogotá, Colombia) for the wireless smartphone-controlled system used in this work. The authors wish to express thanks to Marcos Rosado Iglesias, Antonio Pablo Pérez, Elena Del Corro García, Guillaume Sauthier and Raúl Pérez (all Catalan Institute of Nanoscience and Nanotechnology, CSIC and The Barcelona Institute of Science and Technology, Barcelona, Spain) for their expert assistance with SEM, Raman spectroscopy, XPS and surface profile, respectively.
Glossary
Abbreviations
- GO
graphene oxide
- rGO
reduced graphene oxide
- LIG
laser-induced graphene
- E. coli
Escherichia coli
- LRGO
laser reduced graphene oxide
- T-LRGO
transferred laser reduced graphene oxide
- PDMS
Poly dimethylsiloxane
- WE
working electrode
- CE
counter electrode
- RE
reference electrode
- PBASE
1-pyrenebutanoic acid succinimidyl ester solution
- cAb
capture antibody
- dAb
detection antibody
- LOD
limit of detection
- CV
cyclic voltammetry
- CA
chronoamperometry
- SA
Staphylococcus aureus subsp. Aureus
- ST
Salmonella typhimurium
- XPS
X-ray photoelectron spectroscopy
- SEM
scanning electron microscope
- AU
artificial urine
Supporting Information Available
The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsami.2c20859.
Illustration of PDMS insulation process; one-dimensional surface profile; XPS C 1s elemental composition; current density at different scan rates obtained with 0.1 M NaClO4 in ethanol of two T-LRGO electrodes; open circuit potential measurements of the as-prepared pseudoreference electrodes vs a commercial Ag/AgCl electrode, CV measurements of PBASE functionalized electrodes; ELISA calibration curve of E. coli and selectivity test by ELISA; 4-Parameter Logistic model fitting parameters for the stander optical ELISA and as-prepared electrochemical ELISA; the timetable of electrode fabrication in detail; detailed material costs for the device fabrication; electric scheme of the wireless electrochemical measurement unit; T-LRGO patterns on different substrates (PDF)
Real-time response of the portable wireless system (MP4)
Author Contributions
All authors dedicated in the revision of the manuscript. All authors have given approval to the final version of the manuscript.
The authors declare no competing financial interest.
This paper originally published ASAP on February 14, 2023. Figure 4 was missing information, and a new version reposted on February 22, 2023.
Supplementary Material
References
- Lahcen A. A.; Rauf S.; Beduk T.; Durmus C.; Aljedaibi A.; Timur S.; Alshareef H. N.; Amine A.; Wolfbeis O. S.; Salama K. N. Electrochemical Sensors and Biosensors Using Laser-Derived Graphene: A Comprehensive Review. Biosens. Bioelectron. 2020, 168, 112565. 10.1016/j.bios.2020.112565. [DOI] [PubMed] [Google Scholar]
- Tortorich R.; Shamkhalichenar H.; Choi J.-W. Inkjet-Printed and Paper-Based Electrochemical Sensors. Applied Sciences 2018, 8 (2), 288. 10.3390/app8020288. [DOI] [Google Scholar]
- Hammond J. L.; Formisano N.; Estrela P.; Carrara S.; Tkac J. Electrochemical Biosensors and Nanobiosensors. Essays in Biochemistry 2016, 60 (1), 69. 10.1042/EBC20150008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krishnan S. K.; Singh E.; Singh P.; Meyyappan M.; Nalwa H. S. A Review on Graphene-Based Nanocomposites for Electrochemical and Fluorescent Biosensors. RSC Adv. 2019, 9 (16), 8778. 10.1039/C8RA09577A. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Singh E.; Meyyappan M.; Nalwa H. S. Flexible Graphene-Based Wearable Gas and Chemical Sensors. ACS Appl. Mater. Interfaces 2017, 9 (40), 34544. 10.1021/acsami.7b07063. [DOI] [PubMed] [Google Scholar]
- Htwe Y. Z. N.; Mariatti M. Printed Graphene and Hybrid Conductive Inks for Flexible, Stretchable, and Wearable Electronics: Progress, Opportunities, and Challenges. Journal of Science: Advanced Materials and Devices 2022, 7 (2), 100435. 10.1016/j.jsamd.2022.100435. [DOI] [Google Scholar]
- Novoselov K. S.; Fal′ko V. I.; Colombo L.; Gellert P. R.; Schwab M. G.; Kim K. A Roadmap for Graphene. Nature 2012, 490 (7419), 192. 10.1038/nature11458. [DOI] [PubMed] [Google Scholar]
- Zhang H.; Zhang H.; Aldalbahi A.; Zuo X.; Fan C.; Mi X. Fluorescent Biosensors Enabled by Graphene and Graphene Oxide. Biosens. Bioelectron. 2017, 89, 96. 10.1016/j.bios.2016.07.030. [DOI] [PubMed] [Google Scholar]
- Georgakilas V.; Tiwari J. N.; Kemp K. C.; Perman J. A.; Bourlinos A. B.; Kim K. S.; Zboril R. Noncovalent Functionalization of Graphene and Graphene Oxide for Energy Materials, Biosensing, Catalytic, and Biomedical Applications. Chem. Rev. 2016, 116 (9), 5464. 10.1021/acs.chemrev.5b00620. [DOI] [PubMed] [Google Scholar]
- Zhu Y.; Murali S.; Cai W.; Li X.; Suk J. W.; Potts J. R.; Ruoff R. S. Graphene and Graphene Oxide: Synthesis, Properties, and Applications. Adv. Mater. 2010, 22 (35), 3906. 10.1002/adma.201001068. [DOI] [PubMed] [Google Scholar]
- Mousavi S. M.; Hashemi S. A.; Ghasemi Y.; Amani A. M.; Babapoor A.; Arjmand O. Applications of Graphene Oxide in Case of Nanomedicines and Nanocarriers for Biomolecules: Review Study. Drug Metabolism Reviews 2019, 51 (1), 12. 10.1080/03602532.2018.1522328. [DOI] [PubMed] [Google Scholar]
- Ren X.; Ma H.; Zhang T.; Zhang Y.; Yan T.; Du B.; Wei Q. Sulfur-Doped Graphene-Based Immunological Biosensing Platform for Multianalysis of Cancer Biomarkers. ACS Appl. Mater. Interfaces 2017, 9 (43), 37637. 10.1021/acsami.7b13416. [DOI] [PubMed] [Google Scholar]
- Titelman G. I.; Gelman V.; Bron S.; Khalfin R. L.; Cohen Y.; Bianco-Peled H. Characteristics and Microstructure of Aqueous Colloidal Dispersions of Graphite Oxide. Carbon 2005, 43 (3), 641. 10.1016/j.carbon.2004.10.035. [DOI] [Google Scholar]
- Wu Z.-S.; Ren W.; Gao L.; Liu B.; Jiang C.; Cheng H.-M. Synthesis of High-Quality Graphene with a Pre-Determined Number of Layers. Carbon 2009, 47 (2), 493. 10.1016/j.carbon.2008.10.031. [DOI] [Google Scholar]
- Kudr J.; Zhao L.; Nguyen E. P.; Arola H.; Nevanen T. K.; Adam V.; Zitka O.; Merkoçi A. Inkjet-Printed Electrochemically Reduced Graphene Oxide Microelectrode as a Platform for Ht-2 Mycotoxin Immunoenzymatic Biosensing. Biosens. Bioelectron. 2020, 156, 112109. 10.1016/j.bios.2020.112109. [DOI] [PubMed] [Google Scholar]
- Kurra N.; Jiang Q.; Nayak P.; Alshareef H. N. Laser-Derived Graphene: A Three-Dimensional Printed Graphene Electrode and Its Emerging Applications. Nano Today 2019, 24, 81. 10.1016/j.nantod.2018.12.003. [DOI] [Google Scholar]
- Wei T.; Bao L.; Hauke F.; Hirsch A. Recent Advances in Graphene Patterning. ChemPlusChem. 2020, 85 (8), 1655. 10.1002/cplu.202000419. [DOI] [PubMed] [Google Scholar]
- Backes C.; Abdelkader A. M.; Alonso C.; Andrieux-Ledier A.; Arenal R.; Azpeitia J.; Balakrishnan N.; Banszerus L.; Barjon J.; Bartali R.; Bellani S.; Berger C.; Berger R.; Ortega M. M. B.; Bernard C.; Beton P. H.; Beyer A.; Bianco A.; Bo̷ggild P.; Bonaccorso F.; Barin G. B.; Botas C.; Bueno R. A.; Carriazo D.; Castellanos-Gomez A.; Christian M.; Ciesielski A.; Ciuk T.; Cole M. T.; Coleman J.; Coletti C.; Crema L.; Cun H.; Dasler D.; De Fazio D.; Díez N.; Drieschner S.; Duesberg G. S.; Fasel R.; Feng X.; Fina A.; Forti S.; Galiotis C.; Garberoglio G.; García J. M.; Garrido J. A.; Gibertini M.; Gölzhäuser A.; Gómez J.; Greber T.; Hauke F.; Hemmi A.; Hernandez-Rodriguez I.; Hirsch A.; Hodge S. A.; Huttel Y.; Jepsen P. U.; Jimenez I.; Kaiser U.; Kaplas T.; Kim H.; Kis A.; Papagelis K.; Kostarelos K.; Krajewska A.; Lee K.; Li C.; Lipsanen H.; Liscio A.; Lohe M. R.; Loiseau A.; Lombardi L.; Francisca López M.; Martin O.; Martín C.; Martínez L.; Martin-Gago J. A.; Ignacio Martínez J.; Marzari N.; Mayoral Á.; McManus J.; Melucci M.; Méndez J.; Merino C.; Merino P.; Meyer A. P.; Miniussi E.; Miseikis V.; Mishra N.; Morandi V.; Munuera C.; Muñoz R.; Nolan H.; Ortolani L.; Ott A. K.; Palacio I.; Palermo V.; Parthenios J.; Pasternak I.; Patane A.; Prato M.; Prevost H.; Prudkovskiy V.; Pugno N.; Rojo T.; Rossi A.; Ruffieux P.; Samorì P.; Schué L.; Setijadi E.; Seyller T.; Speranza G.; Stampfer C.; Stenger I.; Strupinski W.; Svirko Y.; Taioli S.; Teo K. B. K.; Testi M.; Tomarchio F.; Tortello M.; Treossi E.; Turchanin A.; Vazquez E.; Villaro E.; Whelan P. R.; Xia Z.; Yakimova R.; Yang S.; Yazdi G. R.; Yim C.; Yoon D.; Zhang X.; Zhuang X.; Colombo L.; Ferrari A. C.; Garcia-Hernandez M. Production and Processing of Graphene and Related Materials. 2D Materials 2020, 7 (2), 022001. 10.1088/2053-1583/ab1e0a. [DOI] [Google Scholar]
- Nagar B.; Balsells M.; de la Escosura-Muñiz A.; Gomez-Romero P.; Merkoçi A. Fully Printed One-Step Biosensing Device Using Graphene/Aunps Composite. Biosens. Bioelectron. 2019, 129, 238. 10.1016/j.bios.2018.09.073. [DOI] [PubMed] [Google Scholar]
- Li Z.; Li K.; Wang S.; Teng C. Covalent Patterning of Graphene for Controllable Functionalization from Microscale to Nanoscale: A Mini-Review. Frontiers in Chemistry 2022, 10, 829614. 10.3389/fchem.2022.829614. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruan X. J.; Luo J. J.; Wang R.; Yao Y. B.; Guan J. J.; Liu T. Microcontact Printing with Laser Direct Writing Carbonization for Facile Fabrication of Carbon-Based Ultrathin Disk Arrays and Ordered Holey Films. Small 2019, 15 (44), 1902819. 10.1002/smll.201902819. [DOI] [PubMed] [Google Scholar]
- Ambaye A. D.; Kefeni K. K.; Mishra S. B.; Nxumalo E. N.; Ntsendwana B. Recent Developments in Nanotechnology-Based Printing Electrode Systems for Electrochemical Sensors. Talanta 2021, 225, 121951. 10.1016/j.talanta.2020.121951. [DOI] [PubMed] [Google Scholar]
- Kathirvelan J. Recent Developments of Inkjet-Printed Flexible Sensing Electronics for Wearable Device Applications: A Review. Sensor Review 2021, 41 (1), 46. 10.1108/SR-08-2020-0190. [DOI] [Google Scholar]
- Mishra N. K.; Patil N.; Anas M.; Zhao X.; Wilhite B. A.; Green M. J. Highly Selective Laser-Induced Graphene (Lig)/Polysulfone Composite Membrane for Hydrogen Purification. Applied Materials Today 2021, 22, 100971. 10.1016/j.apmt.2021.100971. [DOI] [Google Scholar]
- Chyan Y.; Ye R.; Li Y.; Singh S. P.; Arnusch C. J.; Tour J. M. Laser-Induced Graphene by Multiple Lasing: Toward Electronics on Cloth, Paper, and Food. ACS Nano 2018, 12 (3), 2176. 10.1021/acsnano.7b08539. [DOI] [PubMed] [Google Scholar]
- Jung Y.; Min J.; Choi J.; Bang J.; Jeong S.; Pyun K. R.; Ahn J.; Cho Y.; Hong S.; Hong S.; Lee J.; Ko S. H. Smart Paper Electronics by Laser-Induced Graphene for Biodegradable Real-Time Food Spoilage Monitoring. Applied Materials Today 2022, 29, 101589. 10.1016/j.apmt.2022.101589. [DOI] [Google Scholar]
- El-Kady M. F.; Strong V.; Dubin S.; Kaner R. B. Laser Scribing of High-Performance and Flexible Graphene-Based Electrochemical Capacitors. Science 2012, 335 (6074), 1326. 10.1126/science.1216744. [DOI] [PubMed] [Google Scholar]
- Strong V.; Dubin S.; El-Kady M. F.; Lech A.; Wang Y.; Weiller B. H.; Kaner R. B. Patterning and Electronic Tuning of Laser Scribed Graphene for Flexible All-Carbon Devices. ACS Nano 2012, 6 (2), 1395. 10.1021/nn204200w. [DOI] [PubMed] [Google Scholar]
- Lin J.; Peng Z.; Liu Y.; Ruiz-Zepeda F.; Ye R.; Samuel E. L. G.; Yacaman M. J.; Yakobson B. I.; Tour J. M. Laser-Induced Porous Graphene Films from Commercial Polymers. Nat. Commun. 2014, 5 (1), 5714. 10.1038/ncomms6714. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kumar R.; Singh R. K.; Singh D. P.; Joanni E.; Yadav R. M.; Moshkalev S. A. Laser-Assisted Synthesis, Reduction and Micro-Patterning of Graphene: Recent Progress and Applications. Coord. Chem. Rev. 2017, 342, 34. 10.1016/j.ccr.2017.03.021. [DOI] [Google Scholar]
- Wan Z.; Nguyen N.-T.; Gao Y.; Li Q. Laser Induced Graphene for Biosensors. Sustainable Materials and Technologies 2020, 25, e00205. 10.1016/j.susmat.2020.e00205. [DOI] [Google Scholar]
- Jiang H.; Zhao B.; Liu Y.; Li S.; Liu J.; Song Y.; Wang D.; Xin W.; Ren L. Review of Photoreduction and Synchronous Patterning of Graphene Oxide toward Advanced Applications. J. Mater. Sci. 2020, 55 (2), 480. 10.1007/s10853-019-03981-z. [DOI] [Google Scholar]
- Guan Y. C.; Fang Y. W.; Lim G. C.; Zheng H. Y.; Hong M. H. Fabrication of Laser-Reduced Graphene Oxide in Liquid Nitrogen Environment. Sci. Rep. 2016, 6, 28913. 10.1038/srep28913. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oh J.-S.; Kim S.-H.; Hwang T.; Kwon H.-Y.; Lee T. H.; Bae A.-H.; Choi H. R.; Nam J.-D. Laser-Assisted Simultaneous Patterning and Transferring of Graphene. J. Phys. Chem. C 2013, 117 (1), 663. 10.1021/jp309382w. [DOI] [Google Scholar]
- Giacomelli C.; Álvarez-Diduk R.; Testolin A.; Merkoçi A. Selective Stamping of Laser Scribed Rgo Nanofilms: From Sensing to Multiple Applications. 2D Materials 2020, 7 (2), 024006. 10.1088/2053-1583/ab68a7. [DOI] [Google Scholar]
- Yang Q.; Rosati G.; Abarintos V.; Aroca M. A.; Osma J. F.; Merkoçi A. Wearable and Fully Printed Microfluidic Nanosensor for Sweat Rate, Conductivity, and Copper Detection with Healthcare Applications. Biosens. Bioelectron. 2022, 202, 114005. 10.1016/j.bios.2022.114005. [DOI] [PubMed] [Google Scholar]
- da Silva E. T.; Miserere S.; Kubota L. T.; Merkoci A. Simple on-Plastic/Paper Inkjet-Printed Solid-State Ag/Agcl Pseudoreference Electrode. Anal. Chem. 2014, 86 (21), 10531. 10.1021/ac503029q. [DOI] [PubMed] [Google Scholar]
- Bergua J. F.; Álvarez-Diduk R.; Hu L.; Hassan A. H. A.; Merkoçi A. Improved Aliivibrio Fischeri Based-Toxicity Assay: Graphene-Oxide as a Sensitivity Booster with a Mobile-Phone Application. Journal of Hazard Materials 2021, 406, 124434. 10.1016/j.jhazmat.2020.124434. [DOI] [PubMed] [Google Scholar]
- Bergua J. F.; Hu L.; Fuentes-Chust C.; Álvarez-Diduk R.; Hassan A. H. A.; Parolo C.; Merkoçi A. Lateral Flow Device for Water Fecal Pollution Assessment: From Troubleshooting of Its Microfluidics Using Bioluminescence to Colorimetric Monitoring of Generic Escherichia coli. Lab Chip 2021, 21 (12), 2417. 10.1039/D1LC00090J. [DOI] [PubMed] [Google Scholar]
- Bergua J. F.; Álvarez-Diduk R.; Idili A.; Parolo C.; Maymó M.; Hu L.; Merkoçi A. Low-Cost, User-Friendly, All-Integrated Smartphone-Based Microplate Reader for Optical-Based Biological and Chemical Analyses. Anal. Chem. 2022, 94 (2), 1271. 10.1021/acs.analchem.1c04491. [DOI] [PubMed] [Google Scholar]
- Mbayachi V. B.; Ndayiragije E.; Sammani T.; Taj S.; Mbuta E. R.; khan A.u. Graphene Synthesis, Characterization and Its Applications: A Review. Results in Chemistry 2021, 3, 100163. 10.1016/j.rechem.2021.100163. [DOI] [Google Scholar]
- Mukherjee R.; Thomas A. V.; Krishnamurthy A.; Koratkar N. Photothermally Reduced Graphene as High-Power Anodes for Lithium-Ion Batteries. ACS Nano 2012, 6 (9), 7867. 10.1021/nn303145j. [DOI] [PubMed] [Google Scholar]
- Sokolov D. A.; Rouleau C. M.; Geohegan D. B.; Orlando T. M. Excimer Laser Reduction and Patterning of Graphite Oxide. Carbon 2013, 53, 81. 10.1016/j.carbon.2012.10.034. [DOI] [Google Scholar]
- Chen X. D.; Liu Z. B.; Zheng C. Y.; Xing F.; Yan X. Q.; Chen Y.; Tian J. G. High-Quality and Efficient Transfer of Large-Area Graphene Films onto Different Substrates. Carbon 2013, 56, 271. 10.1016/j.carbon.2013.01.011. [DOI] [Google Scholar]
- Rocha J. F.; Hostert L.; Bejarano M. L. M.; Cardoso R. M.; Santos M. D.; Maroneze C. M.; Gongora-Rubio M. R.; Silva C.d.C.C. Graphene Oxide Fibers by Microfluidics Assembly: A Strategy for Structural and Dimensional Control. Nanoscale 2021, 13 (14), 6752. 10.1039/D0NR08380A. [DOI] [PubMed] [Google Scholar]
- Wang L.; Zhang W.; Samavat S.; Deganello D.; Teng K. S. Vertically Aligned Graphene Prepared by Photonic Annealing for Ultrasensitive Biosensors. ACS Appl. Mater. Interfaces 2020, 12 (31), 35328. 10.1021/acsami.0c08036. [DOI] [PubMed] [Google Scholar]
- Ferrari A. C.; Meyer J. C.; Scardaci V.; Casiraghi C.; Lazzeri M.; Mauri F.; Piscanec S.; Jiang D.; Novoselov K. S.; Roth S.; et al. Raman Spectrum of Graphene and Graphene Layers. Phys. Rev. Lett. 2006, 97 (18), 187401. 10.1103/PhysRevLett.97.187401. [DOI] [PubMed] [Google Scholar]
- Voiry D.; Chhowalla M.; Gogotsi Y.; Kotov N. A.; Li Y.; Penner R. M.; Schaak R. E.; Weiss P. S. Best Practices for Reporting Electrocatalytic Performance of Nanomaterials. ACS Nano 2018, 12 (10), 9635. 10.1021/acsnano.8b07700. [DOI] [PubMed] [Google Scholar]
- Xia J.; Chen F.; Li J.; Tao N. Measurement of the Quantum Capacitance of Graphene. Nat. Nanotechnol. 2009, 4 (8), 505. 10.1038/nnano.2009.177. [DOI] [PubMed] [Google Scholar]
- Wang Y.; Shi Z.; Huang Y.; Ma Y.; Wang C.; Chen M.; Chen Y. Supercapacitor Devices Based on Graphene Materials. J. Phys. Chem. C 2009, 113 (30), 13103. 10.1021/jp902214f. [DOI] [Google Scholar]
- Lazcka O.; Campo F. J. D.; Munoz F. X. Pathogen Detection: A Perspective of Traditional Methods and Biosensors. Biosens. Bioelectron. 2007, 22 (7), 1205. 10.1016/j.bios.2006.06.036. [DOI] [PubMed] [Google Scholar]
- Farrow B.; Hong S. A.; Romero E. C.; Lai B.; Coppock M. B.; Deyle K. M.; Finch A. S.; Stratis-Cullum D. N.; Agnew H. D.; Yang S.; Heath J. R. A Chemically Synthesized Capture Agent Enables the Selective, Sensitive, and Robust Electrochemical Detection of Anthrax Protective Antigen. ACS Nano 2013, 7 (10), 9452. 10.1021/nn404296k. [DOI] [PubMed] [Google Scholar]
- Bhimji A.; Zaragoza A. A.; Live L. S.; Kelley S. O. An Electrochemical Elisa Featuring Proximal Reagent Generation: Detection of Hiv Antibodies in Clinical Samples. Anal. Chem. 2013, 85 (14), 6813. 10.1021/ac4009429. [DOI] [PubMed] [Google Scholar]
- Orenstein R.; Wong E. S. Urinary Tract Infections in Adults. Am. Fam Physician 1999, 59 (5), 1225. [PubMed] [Google Scholar]
- Huang T.; Yang J.; Zhou W.; Liu X.; Pan Y.; Song Y. Rapid Identification of Urinary Tract Infections Based on Ultrasensitive Bacteria Detection Using Volumetric Bar-Chart Chip. Sensors and Actuators, B: Chemical 2019, 298, 126885. 10.1016/j.snb.2019.126885. [DOI] [Google Scholar]
- Hu R.-R.; Yin Z.-Z.; Zeng Y.-B.; Zhang J.; Liu H.-Q.; Shao Y.; Ren S.-B.; Li L. A Novel Biosensor for Escherichia Coli O157:H7 Based on Fluorescein-Releasable Biolabels. Biosens. Bioelectron. 2016, 78, 31. 10.1016/j.bios.2015.11.018. [DOI] [PubMed] [Google Scholar]
- Chen S.; Chen X.; Zhang L.; Gao J.; Ma Q. Electrochemiluminescence Detection of Escherichia Coli O157:H7 Based on a Novel Polydopamine Surface Imprinted Polymer Biosensor. ACS Appl. Mater. Interfaces 2017, 9 (6), 5430. 10.1021/acsami.6b12455. [DOI] [PubMed] [Google Scholar]
- Thakur B.; Zhou G.; Chang J.; Pu H.; Jin B.; Sui X.; Yuan X.; Yang C.-H.; Magruder M.; Chen J. Rapid Detection of Single E. Coli Bacteria Using a Graphene-Based Field-Effect Transistor Device. Biosens. Bioelectron. 2018, 110, 16. 10.1016/j.bios.2018.03.014. [DOI] [PubMed] [Google Scholar]
- You Z.; Qiu Q.; Chen H.; Feng Y.; Wang X.; Wang Y.; Ying Y. Laser-Induced Noble Metal Nanoparticle-Graphene Composites Enabled Flexible Biosensor for Pathogen Detection. Biosens. Bioelectron. 2020, 150, 111896. 10.1016/j.bios.2019.111896. [DOI] [PubMed] [Google Scholar]
- Gayathri C. H.; Mayuri P.; Sankaran K.; Kumar A. S. An Electrochemical Immunosensor for Efficient Detection of Uropathogenic E. Coli Based on Thionine Dye Immobilized Chitosan/Functionalized-Mwcnt Modified Electrode. Biosens. Bioelectron. 2016, 82, 71. 10.1016/j.bios.2016.03.062. [DOI] [PubMed] [Google Scholar]
- Mathelié-Guinlet M.; Cohen-Bouhacina T.; Gammoudi I.; Martin A.; Béven L.; Delville M.-H.; Grauby-Heywang C. Silica Nanoparticles-Assisted Electrochemical Biosensor for the Rapid, Sensitive and Specific Detection of Escherichia Coli. Sens. Actuators, B 2019, 292, 314. 10.1016/j.snb.2019.03.144. [DOI] [Google Scholar]
- Kaushik S.; Tiwari U. K.; Pal S. S.; Sinha R. K. Rapid Detection of Escherichia Coli Using Fiber Optic Surface Plasmon Resonance Immunosensor Based on Biofunctionalized Molybdenum Disulfide (MoS2) Nanosheets. Biosens. Bioelectron. 2019, 126, 501. 10.1016/j.bios.2018.11.006. [DOI] [PubMed] [Google Scholar]
- Stathopoulou E. A.; Routsias J. G.; Stea E. A.; Moutsopoulos H. M.; Tzioufas A. G. Cross-Reaction between Antibodies to the Major Epitope of Ro60 Kd Autoantigen and a Homologous Peptide of Coxsackie Virus 2b Protein. Clin. Exp. Immunol. 2005, 141 (1), 148. 10.1111/j.1365-2249.2005.02812.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vaarala O.; Aho K.; Palosuo T.; Alfthan G.; Jauhiainen M.; Leirisalo-Repo M. Crossreaction between Antibodies to Oxidised Low-Density Lipoprotein and to Cardiolipin in Systemic Lupus Erythematosus. Lancet 1993, 341 (8850), 923. 10.1016/0140-6736(93)91213-6. [DOI] [PubMed] [Google Scholar]
- Bhatnagar B. S.; Bogner R. H.; Pikal M. J. Protein Stability During Freezing: Separation of Stresses and Mechanisms of Protein Stabilization. Pharm. Dev. Technol. 2007, 12 (5), 505. 10.1080/10837450701481157. [DOI] [PubMed] [Google Scholar]
- Vaghasiya J. V.; Křípalová K.; Hermanová S.; Mayorga-Martinez C. C.; Pumera M. Real-Time Biomonitoring Device Based on 2d Black Phosphorus and Polyaniline Nanocomposite Flexible Supercapacitors. Small 2021, 17 (38), 2102337. 10.1002/smll.202102337. [DOI] [PubMed] [Google Scholar]
- Vaghasiya J. V.; Mayorga-Martinez C. C.; Vyskočil J.; Sofer Z.; Pumera M. Integrated Biomonitoring Sensing with Wearable Asymmetric Supercapacitors Based on Ti3c2Mxene and 1t-Phase Ws2 Nanosheets. Adv. Funct. Mater. 2020, 30 (39), 2003673. 10.1002/adfm.202003673. [DOI] [Google Scholar]
- Vaghasiya J. V.; Mayorga-Martinez C. C.; Vyskočil J.; Pumera M. Flexible Wearable Mxene Ti3C2-Based Power Patch Running on Sweat. Biosens. Bioelectron. 2022, 205, 114092. 10.1016/j.bios.2022.114092. [DOI] [PubMed] [Google Scholar]
- Lu Z.; Wu L.; Dai X.; Wang Y.; Sun M.; Zhou C.; Du H.; Rao H. Novel Flexible Bifunctional Amperometric Biosensor Based on Laser Engraved Porous Graphene Array Electrodes: Highly Sensitive Electrochemical Determination of Hydrogen Peroxide and Glucose. Jounal of Hazard Material 2021, 402, 123774. 10.1016/j.jhazmat.2020.123774. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.

