Abstract
This short review has the aim of helping the radiologist to identify medical devices when interpreting a chest X-ray, as well as looking for their most commonly detectable complications. Nowadays, many different medical devices are used, often together, especially in critical patients. It is important for the radiologist to know what to look for and to remember the technical factors that need to be considered when checking each device’s positioning.
Keywords: CXR, malposition, CVC, pneumothorax, device, complications
1. Introduction
The chest radiograph, or chest X-ray (CXR), is a well-established imaging modality that is widely used to check for lung, pleural and mediastinal abnormalities [1]. In addition, it is also a first-line examination to monitor the correct positioning of medical devices.
A CXR is usually performed to check for immediate and late-onset complications, even before they can potentially become clinically significant (e.g., the early identification of a small pneumothorax can prevent it from degenerating into a tension pneumothorax).
Potential confounding factors are EKGs’ leads, external tubes, artifacts and overlapping of different devices. The need to have at least two orthogonal projections to correctly locate a device can be a limit in patients in which only an anteroposterior projection is feasible. Moreover, technical factors must be considered, such as the orientation of the X-ray tube and patient rotation (Figure 1) [2].
2. Classification of Medical Devices
In this article, medical devices are disclosed according to their anatomical location: vascular devices (central vein catheter, peripherally inserted central catheter, port, midline and Swan-Ganz catheter), airway devices (endotracheal tubes, tracheostomy tubes), cardiac devices (loop recorder, pacemaker, automatic implantable cardioverter-defibrillator, ventricular assistance devices, Impella, intra-aortic balloon pump and extracorporeal membrane oxygenation), gastro-intestinal devices (nasogastric tube and nasoenteric tubes), chest tubes and miscellaneous other devices that detectable upon CXR (Table 1).
Table 1.
Device Classification | |
---|---|
Intravascular devices | Central vein catheter |
Peripherally inserted central catheter | |
Port Swan-Ganz catheter | |
Airway devices | Endotracheal tubes |
Tracheostomy tubes | |
Cardiac devices | Loop recorder |
Pacemaker and automatic implantable cardioverter-defibrillator Ventricular assistance devices | |
Impella | |
Intra-aortic balloon pump Other cardiac devices | |
Gastro-intestinal devices | Nasogastric tubes |
Nasoenteric tubes | |
Chest tubes | |
Miscellaneous | |
All X-ray images have been anonymized.
3. Intravascular Devices
3.1. Central Vein Catheter (CVC)
CVC refers to a wide range of catheters that are used to obtain reliable central venous access in patients needing prolonged intravenous therapy or in critically ill patients. There are a variety of these catheters, varying in size and shape, but on a CXR, they all appear radiopaque and can be easily detected [3].
CVCs are usually inserted from the jugular or subclavian vein, both on the left or right side (they can also be introduced from femoral veins, but this review considers only the thoracic location) (Figure 2) [4].
The tip of the catheter should be positioned into the superior vena cava (SVC) at the superior cavoatrial junction or in the right atrium [5].
The most common CXR-detectable complications are pneumothorax, malposition (looping, inferior vena cava (IVC) positioning of the tip, wrong thoracic vein positioning), or, less often, vascular perforation, with the catheter located inside the thorax (Figure 3, Figure 4 and Figure 5, Table 2) [6,7].
Table 2.
Device | Most Frequent CXR-Detectable Anomalies and Complications |
---|---|
CVC | Malposition (looping, IVC tip positioning, wrong thoracic vein cannulation) |
Pneumothorax | |
Vascular perforation (less frequent) | |
PICC | Misposition |
Pneumothorax/hemothorax (infrequent) | |
PORT | Pneumothorax |
Midline | Misposition |
Swan-Ganz catheter | Malposition (looping, knotting) |
RA or pulmonary artery wall perforation (diagnosis mostly clinical- or echocardiography-based) |
If there are any doubts about vascular perforation, chest CT should be performed before removing the catheter to assess its exact location [8]. CXR is not able to diagnose catheter-related infections or catheter thrombosis. Arterial puncture is almost always diagnosed by the operator by seeing bright red and pulsating blood in the cannula.
As previously mentioned, technical factors such as the rotation of the patient should be considered, because they can alter the projection of the catheter, a potential confounding factor leading to the wrong diagnosis of a displaced device.
3.2. Peripherally Inserted Central Catheter (PICC)
PICC is a venous catheter that is used for medium- and long-term medical therapy administration. It is a central catheter, but a peripheral arm vein is used as an insertion point. It is usually seen in CXR arising from the left or right arm, with the ideal tip position at the level of the SVC or at the cavoatrial junction (Figure 6). Due to the peripheral insertion point, PICC is usually less associated with the iatrogenic pneumothorax. However, if forced against the vessel wall, the PICC can still cause pneumothorax or hemothorax (Table 2) [9].
3.3. PORT, or Totally Implantable Vascular Access Device
A PORT is an infusion device made of a catheter and a small chamber placed beneath the skin into which drugs can be injected through a silicone membrane. It allows for long-term intravenous therapy, and it is specially used in hematology and oncology patients. The PORT can be seen on a CXR against pectoral tissues with a catheter arising from a small radiopaque device, usually round-shaped, placed via the jugular or subclavian vein, with its tip ending inside the SVC or in the right atrium (Figure 7). The most common CXR-diagnosable complication of port implantation is pneumothorax (Figure 8, Table 2) [10].
3.4. Midline
A midline is a mid- and long-term peripheral venous access catheter. It is usually inserted from the brachial vein, and the tip is usually placed in the axillary or subclavian vein. In a normal CXR, only the tip of the catheter can be evaluated.
It is important to know whether the patient has a midline or a PICC, because on a CXR, a midline can appear just like a mispositioned PICC catheter, with the CXR being able to study just the distal part of it.
Considering the location of the midline, the possibility of a pneumothorax as a complication is remote. If the tip of the catheter is too proximal or is misplaced in a brachial vein, the CXR is not able to provide a diagnosis, because the catheter would be out of the field of view (Table 2).
3.5. Swan-Ganz Catheter
The Swan-Ganz catheter is a balloon-tipped catheter that is used to measure pulmonary artery pressure. It is usually inserted from the right jugular vein, making its way to the right atrium, right ventricle and main pulmonary artery. Usually, the tip is wedged into a right divisional artery (Figure 9) [11].
Complications include perforation of the right atrium or of the pulmonary artery wall; these are life-threatening situations that usually require surgical treatment, and the diagnosis is mainly based on clinical features or bedside echocardiographic findings. The CXR has a role in diagnosing looping or knotting of the catheter (Table 2) [6,12].
4. Airway Devices
4.1. Endotracheal Tubes
Endotracheal tubes (ETTs) are essential devices in anesthesiological practice, allowing for the artificial ventilation of the lungs.
They can be single- (SLT) or double (DLT)-lumen tubes, the latter allowing for differential ventilation or anatomic separation of the lungs; they are usually inserted through the oral cavity, although nasal access can be used when the previous method is not viable.
SLTs usually consist of wide-bore plastic tubes (visible in a plain radiograph through a radiopaque strip within them) with an inflatable balloon at their tip, in order to fix them in place and prevent the aspiration of gastric contents, blood, secretions and other fluids.
In adults, a correctly positioned SLT has its deep extremity placed in the trachea, 5 cm (±2 cm) proximally to the carina (Figure 10); when assessing its position, eventual flexion/extension attitude of the neck should be taken into account (by looking at the mandibula, when included in the radiogram), adjusting the measurement by up to +2/−2 cm, respectively [3].
If the carina is not clearly recognizable, the aortic knob can be taken as a point of reference; the tip of the ETT should be located right above its side [3].
ETT malposition is a relatively common situation following an intubation procedure: it can be placed too high, risking a cord trauma, or too low, selectively cannulating a mainstem bronchus (usually the right one, for anatomic reasons) (Figure 11), and therefore causing atelectasis of the contralateral lung (accompanied by onset/worsening of hypoxemia) and overinflation of the ipsilateral lung (risk of barotrauma and subsequent iatrogenic pneumothorax) [13,14].
Oesophageal intubation should be suspected if an anomalous gaseous distention of the stomach is noted while lungs remain hypo-expanded; in addition to not improving the respiratory function, such a malposition of the ETT can dramatically lead to an oesophageal perforation. If not detected clinically, it is therefore crucial to promptly identify the perforation on plain radiography [15].
In this regard, it is important to note that such a misplaced ETT is not always clearly projected outside the tracheal lumen on the AP projection, making it challenging to provide a confident diagnosis. In these situations, an oblique projection can be helpful [16].
The last condition that should be checked in a CXR after ETT positioning is eventual overinflation of the balloon, which can cause throat pain, tissue ischemia and tracheal perforation/fistulae formation (Table 3) [17,18].
Table 3.
Device | CXR-Detectable Mispositions | Complications |
---|---|---|
ETT | Laryngeal tip positioning | Cord trauma |
Bronchus cannulation | Contralateral atelectasis, ipsilateral overinflation (risk of barotrauma, pneumothorax) | |
Oesophageal intubation | Non-improving respiratory function, esophageal perforation | |
Balloon overinflation | Throat pain, tissue ischemia, tracheal perforation, fistulae formation | |
Tracheostomy tube | Subcutaneous emphysema Hematoma Pneumomediastinum Tracheal stenosis (late complication) |
4.2. Tracheostomy Tubes
Tracheostomy is usually performed when long-term artificial ventilation is needed.
The tip of the tracheostomy tube should be located one-half to two-thirds of the distance from the stoma to the carina (Figure 12), and contrary to the ETT, it does not variate its position with flexion-extension movements of the neck [19].
Complications associated with tracheostomy positioning include subcutaneous emphysema, hematoma and pneumomediastinum (although a small amount of subcutaneous and/or mediastinal air can normally be seen); tracheal stenosis derived from granulation tissue formation and fibrosis may develop at the site of the stoma (Table 3) [3,19].
5. Cardiac Devices
A variety of devices used to monitor heart rhythm and for therapeutic aims can be found on a CXR when analyzing the mediastinum.
5.1. Loop Recorder
A loop recorder is a long-term hearth rhythm monitor that is usually seen in a CXR as a small radiopaque device with no catheters or electro stimulator wires against the mediastinum (Figure 13). It is located in the subcutaneous tissues and should not cause any CXR-diagnosable complications (the migration of the loop recorder along the chest wall is anecdotical) [20,21]
5.2. Pacemaker (PM) and Automatic Implantable Cardioverter-Defibrillator (AICD)
PMs and AICDs are cardiac conduction devices that are made to monitor and to intervene upon cardiac rhythm. They can be external or internal, usually placed under the skin just above the clavicle, and they have electrical leads running inside the veins towards the heart chambers.
In the case of an external PM, only the stimulator wires can be recognized in a CXR. They are usually inserted from a jugular or femoral vein, or in some cases, directly during heart surgery. Their tips can be recognized as projecting against the right atrium and ventricle.
AICD and PM can be distinguished on a CXR by the presence of ICD shock coils: a thick metal segment placed in the end of the lead. Combined ICD-PM devices also exist, known as “cardiac resynchronization therapy” (Figure 14).
CXR is useful after the positioning of these devices to check for the correct location of the electrical leads and to check for their integrity on follow-up examinations. Twiddler’s Syndrome, the consequence of the patient manipulating the device in its pouch, can result in misplacement of the wires that are visible on the CXR as twisted around the PM body [22].
5.3. Ventricular Assistance Devices (VADs)
VADs are mechanical circulatory support devices that can be implanted both in the right (RVAD) or left (LVAD) ventricle; in some cases, RVAD and LVAD can be placed together.
These devices are surgically implanted, with a pump communicating with the ventricle through an inflow cannula while an outflow cannula delivers blood to the aorta (in the case of LVAD) (Figure 15) or to the pulmonary artery (in the case of RVAD) [23].
A CXR can provide information regarding the correct positioning of the cannula. Bleeding from the insertion point can cause hemopericardium, leading to cardiac tamponade or hemothorax; other possible complications of VADs (thrombosis/venous thromboembolism, cardiac arrhythmias, solid organ dysfunction, driveline infections) are not detectable on a CXR [24,25,26,27].
Frequently, the pump of the VAD covers part of the thorax on a CXR, therefore making it difficult for the radiologist to evaluate lung parenchyma and pleural space in that particular site (Table 4).
Table 4.
Device | Most Frequent CXR-Detectable Anomalies and Complications |
---|---|
Loop recorder | Migration (anecdotical) |
PM and AICD | Leads misposition/breakage |
Twiddler’s syndrome | |
VADs | Hemopericardium, hemothorax |
Impella | Dislocation |
IABP | Misposition |
ECMO | Misposition |
5.4. Impella
An Impella is a left ventricular assistance device. It is placed inside the left ventricle through the aortic valve, and it pumps blood from the left ventricle into the ascending aorta (Figure 16). Principal complications are related to the arterial femoral access, and an Impella is usually placed using radioscopy, therefore reducing the risk of mispositioning [25,28]; nevertheless, cases of Impella dislocation are reported in literature [29] (Table 4).
5.5. Intra-Aortic Balloon Pump (IABP)
An intra-aortic balloon pump is an inflatable device that improves perfusion of the coronary arteries.
An IAPB catheter is radiolucent, except for two radiopaque marks at both endings to allow for its identification on CXR. The ideal positioning of the IABP should be at the level of the descending aorta, below the origin of the left subclavian artery and above the splanchnic vessels, to avoid complications such as the occlusion of these vessels or losing of device functionality. On a CXR, an IABP should be at the aortopulmonary level (Figure 17).
The main complications of IABPs include aortic dissection and malpositioning. The CXR has a role in diagnosing mainly the latter, the first complication being a clinical emergency that needs an emergency CT angiography (Table 4) [30].
5.6. Extracorporeal Membrane Oxygenation (ECMO)
ECMO includes a series of bypass techniques that allow for blood oxygenation in patients with severe respiratory or cardiocirculatory failure [31].
Traditional circuits include veno-venous (VV) and veno-arterial (VA) ECMO [31].
The VV-ECMO is used mainly in the setting of severe isolated respiratory failure, draining blood from a cannula usually placed in the SVC and pumping it back into a peripheral vein (usually the femoral vein); the patient’s own heart then pumps the blood throughout the body (Figure 14) [31,32,33].
The VA-ECMO, on the other hand, is the preferred modality in patients with isolated cardiac or combined cardiopulmonary failure. It drains deoxygenated blood from a large vein (usually the femoral vein or the IVC) and reintroduces it into a large artery (usually the subclavian or the femoral artery) after oxygenation, therefore bypassing the heart [31,33].
Even though a CXR is not indicated to monitor the correct functioning of the ECMO circuit, it can still be useful in recognizing the position of ECMO cannulas, considering their variable configuration according to the type of ECMO and the kind of vascular access performed [34,35].
A suspicion of misplacement or migration can be raised according to CXR and should be confirmed via CT angiography (Table 4) [36].
5.7. Other Cardiac Devices
The following section discusses a series of CXRs depicting other cardiac devices that are not included in the previously mentioned ones (Figure 18, Figure 19 and Figure 20).
6. Gastro-Intestinal Devices
6.1. Nasogastric Tubes
Nasogastric intubation is a medical process involving the insertion of a radiopaque-marked plastic tube (nasogastric tube or NG tube) through the nose (sometimes through the oral cavity, referred to in this case as an orogastric tube) to reach down through the stomach.
The NG tube has a double function: it can be used for both the administration of enteral nutrition/oral drugs and for the drainage of gastric contents [37].
A properly positioned NG tube descends into the thorax and crosses the diaphragm on the midline, ending about 10 cm below the oesophago-gastric junction (Figure 21) [38]; note that often, an abdomen X-ray may be additionally required in order to define the abdominal position of the tube.
NG tube malposition includes a tip that is placed too high, mostly in the distal oesophagus (Figure 22) (note that in case of a pH probe, this is considered the adequate placement) or pushed too low (Figure 23), especially in the duodenum (where it could be harmful) (Figure 24); there could also be coiling in the upper ways (Figure 25) or kinking in the gastric lumen (Figure 26) [19,39,40].
Insertion into the respiratory tree (Figure 27) (tip in the trachea/mainstem bronchus/lobar bronchus) can lead to subsequent aspiration pneumonia [37].
Breaching of the gastric wall can be easily recognized if the tip lies in the abdominal cavity, and signs of perforation (i.e., subdiaphragmatic free gas) are noticeable [19].
Intracranial insertion of the NG tube is rare, mostly reported in cases of skull base trauma or surgery (Table 5) [41].
Table 5.
Device | Most Frequent CXR-Detectable Anomalies and Complications |
---|---|
Nasogastric tubes | Proximal (oesophagus)/distal (duodenum) * misposition Coiling/kinking Insertion into the respiratory tree → aspiration pneumonia Gastric wall perforation Intracranial insertion (rare) * |
Nasoenteric tubes | Comparable to NG tubes |
6.2. Nasoenteric Tubes
Nasoenteric tubes are small-bore plastic tubes that are used when prolonged artificial feeding is needed (especially Dobhoff tubes); they are not capable of suction.
Considering that the most correct location of their tip is in the second portion of the duodenum [19], one should be aware of their existence in order to avoid mistaking them for mispositioned nasogastric tubes (Table 5).
7. Chest Tubes
The insertion of tubes through the chest wall into the pleural space (tube thoracostomy) is a well-assessed practice in the hospital setting, allowing for both drainage of collected air/fluid (pneumothorax, pleural effusion, hemothorax, cardio-thoracic surgery, etc.) and drug administration (antibiotics, saline, chemical pleurodesis) [42].
Chest tubes include a wide variety of sizes (small- and large-bore chest tubes) and shapes (straight, angled, pig-tailed), all provided with a radiopaque strip presenting a small gap in correspondence of the first side-hole [42].
The proper position of the tube depends on its aim, an antero-apical orientation being recommended for drainage of the pneumothorax (Figure 28 and Figure 29); conversely, in the case of evacuation of a pleural effusion, the tip should lie in the inferior-posterior region (Figure 30 and Figure 31) [3].
Kinking and malpositioning of the tube are not infrequently seen and should always be suspected in the case of ineffective pleural drainage [19].
The most common sites for misplaced chest tubes include extrapleural soft tissues, intrafissural/intraparenchymal positioning and juxtaposition against the mediastinum. Meanwhile, though inadvertent advancement of the chest tube into the mediastinum appears to be uncommon, there are reports of iatrogenic placement of the chest tube through the diaphragm into the abdomen, causing laceration of the liver, spleen or stomach [19,43,44,45,46].
Intermittent obstruction of chest tubes is usually attributable to clotting blood, debris or pus; however, such abnormalities are not directly visible in a CXR.
Once a chest tube is removed, a residual thickened pleural or parenchymal line may be noticed along its previous tract on CXR. This is not a pathological finding and should not be misinterpreted as a pneumothorax (Table 6) [19].
Table 6.
Device | Most Frequent CXR-Detectable Anomalies and Complications |
---|---|
Chest tubes | Kinking Extrapleural/intrafissural/intraparenchymal/misposition Mediastinum juxtaposition Diaphragmatic trespassing Mediastinal invasion (uncommon) |
8. Miscellaneous
8.1. Cerebrospinal Fluid (CSF) Shunts
The main indication of CSF shunts is the treatment of hydrocephalus.
Whereas the proximal catheter is usually inserted intp one of the lateral ventricles, the distal catheter can theoretically be placed in any fluid-reabsorbing body cavity, peritoneum being the preferential location (ventriculoperitoneal or VP shunt) (Figure 32), as it is associated with fewer complications. The right atrium (ventriculoatrial or VA shunt) or pleural space (ventriculopleural shunt) are also possible [3].
Complications that are visible on plain radiographs include breakages (Figure 33), disconnections and migrations of the distal catheter, pneumothorax, subcutaneous emphysema and, in VA shunts, features of pulmonary hypertension (Table 7) [3,47].
Table 7.
Device | Most Frequent CXR-Detectable Anomalies and Complications |
---|---|
CSF shunts | Breakages Disconnection/migration of the distal catheter Pneumothorax Subcutaneous emphysema Pulmonary hypertension (VA shunts only) |
8.2. Vagal Nerve Stimulator (VNS)
VNS are used for long-term management of seizures in refractory patients.
On a CXR, they resemble a pacemaker or ICD, differentiating themselves from these by the fact that the lead is positioned in the neck to stimulate the left vagal nerve in the carotid sheath [3].
8.3. Other Devices
The following section presents a series of CXRs depicting various devices that are not included in the previous descriptions, which can also be found on a plain chest radiograph (Figure 34, Figure 35, Figure 36, Figure 37, Figure 38, Figure 39 and Figure 40).
Note that in some cases, abdominal devices can be seen on a CXR (Figure 37, Figure 38, Figure 39 and Figure 40), especially if the radiogram is extended to the upper abdomen.
9. Deep-Learning-Based Algorithms for Chest Devices’ Detection on CXR: A Glimpse into the Future
The use of artificial intelligence (AI) software has already demonstrated improved sensibility, specificity and accuracy if used as a decision support for the radiologist when evaluating CXR for thoracic pathologies; other algorithms, on the other hand, have proven to help the radiologist when evaluating follow-up X-rays of the same patient [48,49].
The accurate identification of chest devices by AI in the context of a hypothetical CXR-automated reporting software presents two main purposes, the first being a better evaluation of eventual concomitant chest pathologies, obtainable with the reduction of false positives represented by devices opacities [50]; on the other hand, when the assessment of the location of chest devices via CXR is necessary for many patients at a time (as is often the case in the clinical care context of a hospital setting), a high-sensitivity deep-learning-based algorithm could potentially reduce the time required for the diagnosis of any misplacement, allowing for a more immediate solution and helping to decrease possible misinterpretations [9].
AI algorithms can also be used to monitor potential device-related thoracic complications, such as pneumothorax or pleural effusion [51,52].
Even though a few examples of deep-learning systems for the automatic detection of chest devices have already been proposed [9,50,53,54,55,56], further studies and efforts are necessary in order to provide a comprehensive software tool that can be concretely and reliably applied in the variegated clinical scenario of the hospital reality.
10. Conclusions
Multiple chest- and non-chest-related devices can be seen on a CXR, the use of some of which has only recently been introduced in the hospital setting. It is important for the radiologist to know the normal appearances of such devices as well as their proper location, in order to promptly recognize any mispositioning and correctly address clinical management before any major complications develop.
The automated detection of chest devices through deep-learning-based algorithms will hopefully not be only a helpful tool in this scope for the radiologist, but will also improve AI sensitivity and specificity in the assessment of chest pathologies through CXR examination.
Author Contributions
Conceptualization, A.D.C., J.-D.G., N.S. and M.G.; methodology, G.C. and J.Z.B.; software, G.B.; validation, P.N., E.Q. and F.C.; resources, M.G. and N.S.; writing—original draft preparation, M.G. and N.S.; writing—review and editing, A.D.C., J.-D.G.; supervision, P.N., E.Q. and F.C. All authors have read and agreed to the published version of the manuscript.
Conflicts of Interest
The authors declare no conflict of interest.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Quaia E., Baratella E., Poillucci G., Kus S., Cioffi V., Cova M.A. Digital Tomosynthesis as a Problem-Solving Imaging Technique to Confirm or Exclude Potential Thoracic Lesions Based on Chest X-ray Radiography. Acad. Radiol. 2013;20:546–553. doi: 10.1016/j.acra.2012.12.009. [DOI] [PubMed] [Google Scholar]
- 2.Hunter T.B., Taljanovic M.S., Tsau P.H., Berger W.G., Standen J.R. Medical Devices of the Chest. Radiographics. 2004;24:1725–1746. doi: 10.1148/rg.246045031. [DOI] [PubMed] [Google Scholar]
- 3.Mathew R.P., Alexander T., Patel V., Low G. Chest radiographs of cardiac devices (Part 1): Lines, tubes, non-cardiac medical devices and materials. S. Afr. J. Radiol. 2019;23:1729. doi: 10.4102/sajr.v23i1.1729. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Weissbach A., Gendler Y., Lakovsky Y., Kadmon G., Nahum E., Kaplan E. Routine chest X-ray following ultrasound-guided internal jugular vein catheterization in critically ill children: A prospective observational Study. Pediatr. Anesthesia. 2020;30:1378–1383. doi: 10.1111/pan.14030. [DOI] [PubMed] [Google Scholar]
- 5.Wang L., Liu Z.-S., Wang C.-A. Malposition of Central Venous Catheter. Chin. Med. J. 2016;129:227–234. doi: 10.4103/0366-6999.173525. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Baratella E., Marrocchio C., Bozzato A.M., Roman-Pognuz E., Cova M.A. Chest X-ray in intensive care unit patients: What there is to know about thoracic devices. Diagn. Interv. Radiol. 2021;27:633–638. doi: 10.5152/dir.2021.20497. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Roldan C.J., Paniagua L. Central Venous Catheter Intravascular Malpositioning: Causes, Prevention, Diagnosis, and Correction. West. J. Emerg. 2015;16:658–664. doi: 10.5811/westjem.2015.7.26248. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Gayer G., Rozenman J., Hoffmann C., Apter S., Simansky D.A., Yellin A., Itzchak Y. CT diagnosis of malpositioned chest tubes. Br. J. Radiol. 2000;73:786–790. doi: 10.1259/bjr.73.871.11089474. [DOI] [PubMed] [Google Scholar]
- 9.Yu D., Zhang K., Huang L., Zhao B., Zhang X., Guo X., Li M., Gu Z., Fu G., Hu M., et al. Detection of peripherally inserted central catheter (PICC) in chest X-ray images: A multi-task deep learning model. Comput. Methods Programs Biomed. 2020;197:105674. doi: 10.1016/j.cmpb.2020.105674. [DOI] [PubMed] [Google Scholar]
- 10.Fu J.-Y., Wu C.-F., Ko P.-J., Wu C.-Y., Kao T.-C., Yu S.-Y., Liu Y.-H., Hsieh H.-C. Analysis of chest X-ray plain film images of intravenous ports inserted via the superior vena cava. Surg. Today. 2014;44:1513–1521. doi: 10.1007/s00595-014-0893-5. [DOI] [PubMed] [Google Scholar]
- 11.Rajaram S.S., Desai N.K., Kalra A., Gajera M., Cavanaugh S.K., Brampton W., Young D., Harvey S., Rowan K. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst. Rev. 2013;2018:CD003408. doi: 10.1002/14651858.CD003408.pub3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Rudziński P.N., Henzel J., Dzielińska Z., Lubiszewska B., Michałowska I., Szymański P., Pracoń R., Hryniewiecki T., Demkow M. Pulmonary artery rupture as a complication of Swan-Ganz catheter application. Diagnosis and endovascular treatment: A single centre’s experience. Adv. Interv. Cardiol. 2016;12:135–139. doi: 10.5114/aic.2016.59364. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Han J.-H., Lee S.-H., Kang Y.-J., Kang J.-M. Plain endotracheal tube insertion carries greater risk for malpositioning than does reinforced endotracheal tube insertion in females. Korean J. Anesthesiol. 2013;65:S23–S24. doi: 10.4097/kjae.2013.65.6S.S23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ioannidis G., Lazaridis G., Baka S., Mpoukovinas I., Karavasilis V., Lampaki S., Kioumis I., Pitsiou G., Papaiwannou A., Karavergou A., et al. Barotrauma and pneumothorax. J. Thorac. Dis. 2015;7:S38–S43. doi: 10.3978/j.issn.2072-1439.2015.01.31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Jougon J., Cantini O., Delcambre F., Minniti A., Velly J. Esophageal perforation: Life threatening complication of endotracheal intubation. Eur. J. Cardio-Thoracic Surg. 2001;20:7–11. doi: 10.1016/S1010-7940(01)00766-7. [DOI] [PubMed] [Google Scholar]
- 16.Smith G.M., Reed J.C., Choplin R.H. Radiographic detection of esophageal malpositioning of endotracheal tubes. Am. J. Roentgenol. 1990;154:23–26. doi: 10.2214/ajr.154.1.2104719. [DOI] [PubMed] [Google Scholar]
- 17.Kumar S.S., Young P. Over-inflation of the tracheal tube cuff: A case for routine monitoring. Crit. Care. 2002;6:P37. doi: 10.1186/cc1736. [DOI] [Google Scholar]
- 18.Striebel H.W., Pinkwart L.U., Karavias T. Trachealruptur durch zu stark geblockte Tubusmanschette. Der Anaesthesist. 1995;44:186–188. doi: 10.1007/s001010050146. [DOI] [PubMed] [Google Scholar]
- 19.Godoy M.C.B., Leitman B.S., de Groot P.M., Vlahos I., Naidich D.P. Chest Radiography in the ICU: Part 1, Evaluation of Airway, Enteric, and Pleural Tubes. Am. J. Roentgenol. 2012;198:563–571. doi: 10.2214/AJR.10.7226. [DOI] [PubMed] [Google Scholar]
- 20.Silveira I., Sousa M.J., Antunes N., Silva V., Roque C., Pinheiro-Vieira A., Lagarto V., Hipólito-Reis A., Luz A., Torres S. Efficacy and Safety Of Implantable Loop Recorder: Experience Of A Center. J. Atr. Fibrillation. 2016;9:1425. doi: 10.4022/jafib.1425. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Hasnie A.A., Hasnie A.A., Assaly R.A. The Case of the Migrating Loop Recorder. JACC Case Rep. 2019;1:156–160. doi: 10.1016/j.jaccas.2019.06.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Aguilera A.L., Volokhina Y.V., Fisher K.L. Radiography of Cardiac Conduction Devices: A Comprehensive Review. Radiographics. 2011;31:1669–1682. doi: 10.1148/rg.316115529. [DOI] [PubMed] [Google Scholar]
- 23.Kadakia S., Moore R., Ambur V., Toyoda Y. Current status of the implantable LVAD. Gen. Thorac. Cardiovasc. Surg. 2016;64:501–508. doi: 10.1007/s11748-016-0671-y. [DOI] [PubMed] [Google Scholar]
- 24.Cohen D.G., Thomas J.D., Freed B.H., Rich J.D., Sauer A.J. Echocardiography and Continuous-Flow Left Ventricular Assist Devices. JACC Heart Fail. 2015;3:554–564. doi: 10.1016/j.jchf.2015.03.003. [DOI] [PubMed] [Google Scholar]
- 25.Sigakis C.J.G., Mathai S.K., Suby-Long T.D., Restauri N.L., Ocazionez D., Bang T.J., Restrepo C.S., Sachs P.B., Vargas D. Radiographic Review of Current Therapeutic and Monitoring Devices in the Chest. Radiographics. 2018;38:1027–1045. doi: 10.1148/rg.2018170096. [DOI] [PubMed] [Google Scholar]
- 26.George A.N., Hsia T.-Y., Schievano S., Bozkurt S. Complications in children with ventricular assist devices: Systematic review and meta-analyses. Heart Fail. Rev. 2021;27:903–913. doi: 10.1007/s10741-021-10093-x. [DOI] [PubMed] [Google Scholar]
- 27.Allen S.J., Sidebotham D. Postoperative care and complications after ventricular assist device implantation. Best Pr. Res. Clin. Anaesthesiol. 2012;26:231–246. doi: 10.1016/j.bpa.2012.03.005. [DOI] [PubMed] [Google Scholar]
- 28.Glazier J.J., Kaki A. The Impella Device: Historical Background, Clinical Applications and Future Directions. Int. J. Angiol. 2018;28:118–123. doi: 10.1055/s-0038-1676369. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Sugimura Y., Bauer S., Immohr M.B., Mehdiani A., Rellecke P., Westenfeld R., Aubin H., Boeken U., Lichtenberg A., Akhyari P. Outcome of Patients Supported by Large Impella Systems After Re-implantation Due to Continued or Recurrent Need of Temporary Mechanical Circulatory Support. Front. Cardiovasc. Med. 2022;9 doi: 10.3389/fcvm.2022.926389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Ginat D., Massey H.T., Bhatt S., Dogra V. Imaging of Mechanical Cardiac Assist Devices. J. Clin. Imaging Sci. 2011;1:21. doi: 10.4103/2156-7514.80373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Shah A., Dave S., Goerlich C.E., Kaczorowski D.J. Hybrid and parallel extracorporeal membrane oxygenation circuits. JTCVS Tech. 2021;8:77–85. doi: 10.1016/j.xjtc.2021.02.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Gross G.W., Cullen J., Kornhauser M.S., Wolfson P.J. Thoracic complications of extracorporeal membrane oxygenation: Findings on chest radiographs and sonograms. Am. J. Roentgenol. 1992;158:353–358. doi: 10.2214/ajr.158.2.1729797. [DOI] [PubMed] [Google Scholar]
- 33.Types of ECMO | Extracorporeal Membrane Oxygenation|ECLS. [(accessed on 31 January 2023)]. Available online: https://www.elso.org/ecmo-resources/types-of-ecmo.aspx.
- 34.Beck L., Burg M.C., Heindel W., Schülke C. Extrakorporale Membranoxygenierung bei Erwachsenen–Varianten, Komplikationen unter Therapie und die Rolle der radiologischen Diagnostik. Rofo. 2016;189:119–127. doi: 10.1055/s-0042-118885. [DOI] [PubMed] [Google Scholar]
- 35.Barnacle A.M., Smith L.C., Hiorns M.P. The Role of Imaging during Extracorporeal Membrane Oxygenation in Pediatric Respiratory Failure. Am. J. Roentgenol. 2006;186:58–66. doi: 10.2214/AJR.04.1672. [DOI] [PubMed] [Google Scholar]
- 36.Lidegran M.K., Ringertz H.G., Frenckner B., Lindén V.B. Chest and abdominal CT during extracorporeal membrane oxygenation: Clinical benefits in diagnosis and treatment1. Acad. Radiol. 2005;12:276–285. doi: 10.1016/j.acra.2004.11.027. [DOI] [PubMed] [Google Scholar]
- 37.Guthrie D.B., Pezzollo J.P., Lam D.K., Epstein R.H. Tracheopulmonary Complications of a Malpositioned Nasogastric Tube. Anesthesia Prog. 2020;67:151–157. doi: 10.2344/anpr-67-01-02. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Taylor S., Manara A.R. X-ray checks of NG tube position: A case for guided tube placement. Br. J. Radiol. 2021;94:20210432. doi: 10.1259/bjr.20210432. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Verpoorte R. Alkaloids. In: Worsfold P., Townshend A., Poole C., editors. Encyclopedia of Analytical Science. 2nd ed. Elsevier; Oxford, UK: 2005. pp. 56–61. [Google Scholar]
- 40.Misra S., Behera B.K., Sahoo A.K. Internal jaw thrust for nasogastric tube insertion. Indian J. Anaesth. 2018;62:911–913. doi: 10.4103/ija.IJA_400_18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Obiorah S.C.S., Moldovan K., Doberstein C. Intracranial insertion of a nasogastric tube following septoplasty: Case report and literature review. Interdiscip. Neurosurg. 2020;22:100879. doi: 10.1016/j.inat.2020.100879. [DOI] [Google Scholar]
- 42.Porcel J.M. Chest Tube Drainage of the Pleural Space: A Concise Review for Pulmonologists. Tuberc. Respir. Dis. 2018;81:106–115. doi: 10.4046/trd.2017.0107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Al Mosa A.F.H., Ishaq M., Ahmed M.H.M. Unusual Malposition of a Chest Tube, Intrathoracic but Extrapleural. Case Rep. Radiol. 2018;2018:1–4. doi: 10.1155/2018/8129341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Ohtaka K., Hase R., Chiba R., Miyasaka M., Sato S., Shoji Y., Ichimura T., Senmaru N., Kaga K., Matsui Y. Noninvasive management for iatrogenic splenic injury caused by chest tube insertion: A case report. Clin. Case Rep. 2016;4:1157–1160. doi: 10.1002/ccr3.726. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Serji B., MirAli H., Chablou M., Kamaoui I., El Harroudi T. Liver injury secondary to chest tube placement: A case report of conservative management and review of literature. Clin. Case Rep. 2017;6:45–48. doi: 10.1002/ccr3.1280. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Kesieme E.B., Dongo A., Ezemba N., Irekpita E., Jebbin N., Kesieme C. Tube Thoracostomy: Complications and Its Management. Pulm. Med. 2011;2012:1–10. doi: 10.1155/2012/256878. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Kluge S., Baumann H.J., Regelsberger J., Kehler U., Gliemroth J., Koziej B., Klose H., Meyer A. Pulmonary hypertension after ventriculoatrial shunt implantation. J. Neurosurg. 2010;113:1279–1283. doi: 10.3171/2010.6.JNS091541. [DOI] [PubMed] [Google Scholar]
- 48.Hwang E.J., Goo J.M., Yoon S.H., Beck K.S., Seo J.B., Choi B.W., Chung M.J., Park C.M., Jin K.N., Lee S.M. Use of Artificial Intelligence-Based Software as Medical Devices for Chest Radiography: A Position Paper from the Korean Society of Thoracic Radiology. Korean J. Radiol. 2021;22:1743–1748. doi: 10.3348/kjr.2021.0544. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Li D., Pehrson L.M., Lauridsen C.A., Tøttrup L., Fraccaro M., Elliott D., Zając H.D., Darkner S., Carlsen J.F., Nielsen M.B. The Added Effect of Artificial Intelligence on Physicians’ Performance in Detecting Thoracic Pathologies on CT and Chest X-ray: A Systematic Review. Diagnostics. 2021;11:2206. doi: 10.3390/diagnostics11122206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Singh R., Kalra M.K., Nitiwarangkul C., Patti J.A., Homayounieh F., Padole A., Rao P., Putha P., Muse V.V., Sharma A., et al. Deep learning in chest radiography: Detection of findings and presence of change. PLoS ONE. 2018;13:e0204155. doi: 10.1371/journal.pone.0204155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Irmici G., Cè M., Caloro E., Khenkina N., Della Pepa G., Ascenti V., Martinenghi C., Papa S., Oliva G., Cellina M. Chest X-ray in Emergency Radiology: What Artificial Intelligence Applications Are Available? Diagnostics. 2023;13:216. doi: 10.3390/diagnostics13020216. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Lee H., Mansouri M., Tajmir S., Lev M.H., Do S. A Deep-Learning System for Fully-Automated Peripherally Inserted Central Catheter (PICC) Tip Detection. J. Digit. Imaging. 2017;31:393–402. doi: 10.1007/s10278-017-0025-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Adams S.J., Henderson R.D.E., Yi X., Babyn P. Artificial Intelligence Solutions for Analysis of X-ray Images. Can. Assoc. Radiol. J. 2020;72:60–72. doi: 10.1177/0846537120941671. [DOI] [PubMed] [Google Scholar]
- 54.Subramanian V., Wang H., Wu J.T., Wong K.C.L., Sharma A., Syeda-Mahmood T. Automated Detection and Type Classifica-tion of Central Venous Catheters in Chest X-Rays. In: Shen D., Liu T., Peters T.M., Staib L.H., Essert C., Zhou S., Yap P.-T., Khan A., editors. Proceedings of the Medical Image Computing and Computer Assisted Intervention—MICCAI 2019. Springer International Publishing; Cham, Switzerland: 2019. pp. 522–530. [Google Scholar]
- 55.Frid-Adar M., Amer R., Greenspan H. Endotracheal Tube Detection and Segmentation in Chest Radiographs Using Synthetic Data. In: Shen D., Liu T., Peters T.M., Staib L.H., Essert C., Zhou S., Yap P.-T., Khan A., editors. Proceedings of the Medical Image Computing and Computer Assisted Intervention—MICCAI 2019. Springer International Publishing; Cham, Switzerland: 2019. pp. 784–792. [Google Scholar]
- 56.Yi X., Adams S.J., Henderson R.D.E., Babyn P. Computer-aided Assessment of Catheters and Tubes on Radiographs: How Good Is Artificial Intelligence for Assessment? Radiol. Artif. Intell. 2020;2:e190082. doi: 10.1148/ryai.2020190082. [DOI] [PMC free article] [PubMed] [Google Scholar]