Skip to main content
. 2023 Feb 19;15(2):571. doi: 10.3390/v15020571

Table 2.

Limit of detection for the HIV-1 target-specific sequencing using iSeq100.

Concentration (cp/mL) Number of Samples Tested Number of Correctly Identified Samples Percentage of Correctly Identified Samples
2000 13 13 100%
1000 10 10 100%
500 10 10 100%