Summary
Deep brain stimulation (DBS) and responsive neurostimulation (RNS) use high frequency stimulation (HFS) per the pivotal trials and manufacturer-recommended therapy protocols. However, not all patients respond to HFS. In this retrospective case series, 10 patients implanted with the RNS System were programmed with Low Frequency Stimulation (LFS) to treat their seizures; 9 of these patients were previously treated with HFS (100 Hz or greater). LFS was defined as frequency less than 10 Hz. Burst duration was increased to at least 1000 msec. With HFS patients had a median seizure reduction (MSR) of 13% (IQR −67 to 54) after a median of 19 months (IQR 7–49). In contrast, LFS was associated with a 67% MSR (IQR 13–95) when compared to HFS and 76% MSR (IQR 43–91) when compared to baseline prior to implantation. Charge delivered per hour and pulses per day were not significantly different between HFS and LFS, although time stimulated per day was longer for LFS (228 min) than for HFS (7 min). There were no LFS-specific adverse effects reported by any of the patients. LFS could represent an alternative, effective method for delivering stimulation in focal DRE patients treated with the RNS System.
Keywords: neurostimulation, neuromodulation, low frequency stimulation, responsive neurostimulation, alternative parameters
Introduction
Intracranial neurostimulation is a palliative approach for the treatment of adult patients with drug resistant epilepsy (DRE). It involves electrical stimulation with physician-defined current, pulse width and frequency over a determined period of time to alter neural activity at seizure foci and network nodes. There are two FDA-approved intracranial stimulation modalities for focal epilepsy: anterior thalamic nuclei deep brain stimulation (ANT-DBS) and responsive neurostimulation (RNS). Both modalities are typically programmed with high frequency stimulation (HFS) as used in the pivotal trials1, 2. For this study we consider stimulation of 100 Hz or greater as HFS, consistent with prior observations3. The RNS System is a closed loop system that detects and records intracranial epileptiform activity arising from seizure foci and/or network nodes and stimulates automatically according to physician-programed settings. One therapy is comprised of two programable bursts and can be repeated up to 5 times if the abnormal electrical activity continues, though only the first delivered therapy is counted by the RNS system. The recommended initial stimulation settings are frequency of 200 Hz (or pulses per second), pulse width of 160 μsec and burst duration of 100 ms. These parameters have been informed in part by the experience of long-term treatment trials that resulted in a 75% median seizure reduction at nine years4.
Low frequency stimulation (LFS) has been studied in animals as a potential antiepileptic strategy specially in rodent kindling models. In humans, LFS has been studied in DRE patients through stimulation of a wide variety of targets including hippocampus5, fornix6, thalamus7 and cortex8. Regarding the latter, chronic subthreshold stimulation (CSS) involves open-loop, continuous electrical stimulation of seizure foci in focal DRE patients through LFS and may be particularly useful when stimulating eloquent cortex9. However, LFS for intracranial stimulation is not often considered, in part due to concerns that it may even worsen seizures3. To our knowledge, a within patient comparison of HFS and LFS for implanted intracranial stimulation devices has not been published.
In this report, we aimed to evaluate the clinical response of patients treated the RNS System programed with LFS. We suggest that when patients do not respond to the HFS of standard RNS settings, a lower stimulation frequency in addition to longer burst duration (LFS) are reasonable stimulation parameters to consider.
Methods
This institutional review board-approved retrospective case series included all DRE patients implanted with the RNS System followed at our center with active LFS at last clinical follow up. All variables, including clinical seizure frequency, were obtained through the electronic health record. The RNS System implantation was performed as part of clinical care2. HFS has been defined as stimulation >45 Hz elsewhere3. We defined LFS as a stimulation frequency of less than 10 Hz and used a burst duration of 5 sec. We used 7 Hz theta frequency, which we have used previously given an association with the limbic system10–12. Pulse width was 160 μsec, except for patient 5 during HFS when it was 240 μsec. Specifically, stimulation was typically 5 sec trains of 35 biphasic pulses with phase width of 160 μsec. Stimulation amplitude was adjusted using charge density as the relevant metric, per typical clinical practice. ECoG events were not analyzed as detection parameters were changed during clinical care precluding direct comparisons over time. Responder rate was defined as clinical seizure frequency reduction 226550%. We determined total time of stimulation per day, reported in minutes per day (min/d); pulses delivered per day, reported as pulses in a 24-hour period; and calculated charge delivered per hour, reported as millicoulombs per hour (mC/h):
Since the RNS System has 2 programmable bursts per therapy. When the same lead was used for both burst 1 and burst 2, the charges were added. When burst settings were different, the burst providing the largest charge was used. The 1.27 factor represents the average number of therapy repetitions delivered until the sensed abnormal electrical activity was no longer detected.
All statistics and graphs were performed on GraphPad Prism version 9.3.1 for Windows (GraphPad Software, San Diego, CA). Continuous and categorical variables are described as median with interquartile ranges (IQR) and percentages, respectively. Chi-squared and Fisher’s exact tests were used for comparison of proportions and frequencies. Mann-Whitney U or Wilcoxon signed ranks tests were used for median comparison between groups as appropriate. Spearman Rho was used for correlation analysis. P values ≤0.05 were considered statistically significant. De-identified data are available upon request.
Results
A total of 39 patients implanted with the RNS System were followed at our center between August 2004 and April 2022, and 10 patients were included in our analysis. Two patients were programmed with LFS in the past but were not on LFS at last clinical follow-up: one was implanted at an outside center, was trialed on LFS in our clinic for less than 3 months, and then programmed back to HFS by their primary neurologist. The other patient was on a mixed HFS and LFS protocol at last follow up. Eleven patients were on LFS at last clinical follow up. One of these patients was excluded due to unreliable clinical seizure reporting. Of the remaining 10 patients, one patient was initially started on LFS and has never been on HFS; this patient received promising LFS via temporary trial stimulation during stereo EEG evaluation9, 13. Table 1 summarizes patient baseline characteristics and seizure frequencies.
Table 1 –
Baseline Characteristics
| Pt Num | Sex/Age at implant (y) | Seizure onset / Type of EEG | MRI findings | Electrode location, orientation and type | Number of ASD trialed before RNS/ASD at time of implant | ASD at last follow-up | Baseline Seizure Fq (sz/mo) | Seizure at last HFS Fq (sz/mo) | Seizure Fq at last LFS (sz/mo) | Follow-up time (mo) | LFS time (mo) | Comments |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | F/27 | Left temporal / Subdural electrodes | Post left ATL with residual MTS | Left temporal neocortex, subdurals | 5/FBM, LEV, LTG | FBM, LEV, LTG | 4.5 | 3 | 1 | 206 | 37 | |
| 2 | M/49 | Bilateral temporal / SEEG | Non-lesional MRI | Bilateral mesial temporal, longitudinal depths | 5/LTG | LTG | 6 | 10 | 7.5 | 65 | 6 | Implanted at different center. Orthogonal bitemporal leads replaced with longitudinal leads |
| 3 | F/23 | Bilateral temporal / SEEG | Right hippocampal atrophy Left MTS |
Bilateral mesial temporal, longitudinal depths | 8/CBZ, CLB, CZP | CLB, CNB | 6 | 10 | 3 | 61 | 14 | |
| 4 | F/31 | Bilateral temporal / Scalp EEG | Non-lesional MRI | Bilateral mesial temporal, longitudinal depths | 2/PRP, LTG | CLB, LEV, OXC | 6 | 1.5 | 0 | 51 | 6 | Implanted at different center. Switched to LFS at our center |
| 5 | F/37 | Left temporal / Scalp EEG | Bilateral MTS | Bilateral mesial temporal, longitudinal depths | 2/ECBZ, LCM | BRV, CLB, CNB, LEV | 12 | 30 | 3 | 47 | 35 | Infection of initial device leading to explant, subsequently reimplanted |
| 6 | F/37 | Bilateral temporal / Scalp EEG | Bilateral MTS | Bilateral mesial temporal, longitudinal depths | 9/GBP, LCM, LEV | GBP, LCM, LEV | 4 | 3.5 | 2 | 29 | 10 | Patient underwent left temporal LITT since 95% of ECoG detections were L sided |
| 7 | F/21 | Left temporal / SEEG | Non-lesional MRI | Left hippocampal and parahippocampal areas, longitudinal depths | 4/LEV | LEV | 6 | 4.6 | 4.6 | 22 | 2 | Patient committed suicide, known MDD |
| 8 | M/49 | Bilateral temporal / Scalp EEG | Right MTS | Bilateral mesial temporal, longitudinal depths | 4/CBZ, LCM, LEV, LZP | CBZ, LCM, LEV, LTG, LEV | 3.5 | 0.25 | 0.5 | 22 | 14 | |
| 9 | F/53 | Left frontal / SEEG | Left frontal FCD | Left frontal FCD, orthogonal depths | 3/LCM, OXC, TPM | LCM, OXC, TPM | 16 | NA | 2 | 26 | 26 | Patient on LFS since implant |
| 10 | M/18 | Bilateral temporal / Scalp EEG | Left temporal encephalocele s/p resection | Bilateral mesial temporal, longitudinal depths | 6/LCM, LTG | BRV, LCM | 3 | 4 | 0 | 9 | 4 |
ASD, antiseizure drug; ATL, anterior temporal lobectomy; CBZ, carbamazepine; CLB, clobazam; CNB, cenobamate, CZP, clonazepam; ECBZ, eslicarbazepine; ECoG, electrocorticography; FBM, felbamate; Fq, frequency; GBP, gabapentin; FCD, focal cortical dysplasia; LCM, lacosamide; LEV, levetiracetam; LITT, laser interstitial thermal therapy; LTG, lamotrigine; LZP, lorazepam; MDD, major depressive disorder; mo, moths; MTS, mesial temporal sclerosis; Num, number; OXC, oxcarbazepine; pt, patient; SEEG, stereoelectroencephalography; sz/mo, seizures per month; TPM, topiramate
Median seizure frequency at baseline prior to RNS System implantation was 6 seizures per month (sz/mo) (IQR 4–8). After a median of 19 mo (IQR 8–49) on HFS, patients (n=9) had a median seizure reduction (MSR) of 13% (IQR −67 to 54). Seizure frequency after HFS was not significantly different compared to baseline (4 sz/mo, IQR 2–10 vs. 6 sz/mo, IQR 4–8; p=0.88). Patients were on LFS for a median of 12 mo (IQR 5–28), and patients with LFS had significantly fewer seizures (2 sz/mo, IQR 0.4–3) compared to HFS (4 sz/mo, IQR 2–10; p=0.02) and baseline (6 sz/mo, IQR 4–8; p=0.006). The MSR associated with LFS when compared to baseline and HFS was 76% (IQR 43–91) and 67% (IQR 13–95), respectively. LFS had an 80% responder rate compared to baseline, and a 56% responder rate compared to HFS (Figure 1). When compared to baseline, LFS had a significantly higher proportion of responders than HFS (80% vs. 22%, p=0.02).
Figure 1 –

Upper left panel; Individual seizure frequency reduction after LFS compared to HFS (n = 10) from worse (upper left) to best (upper right). Mid left panel; percentage of responders after HFS from baseline, after LFS from HFS, and after LFS from baseline, respectively. Upper right panel; comparison of median charge delivered per hour (p=0.197). Mid left panel; comparison of time stimulated per day (p=0.004). Bottom panel; visualization of an ECoG long episode with LFS after adjusting the ECoG record length from 90secs to 180secs and increasing the long episode length to exceed the maximum time of therapy. In this example above, Burst 1 and Burst 2 are set to 5000ms (5secs), ie, up to 50 seconds of stimulation total can be delivered per therapy.
Charge density and charge per hour were not significantly different between HFS and LFS (2.0 μC/cm2, range 0.5–4.6 vs. 3 μC/cm2, range 1.0–5.5; p=0.07), (0.5 mC/h, range 0.004–3.4 vs. 4.4 mC/h, range 0.02–24.5; p=0.2), respectively. Time stimulated per day was significantly longer with LFS (228 min, range 2–717) compared to HFS (7 min, range 0.4–27), p=0.006. Pulses per day were not significantly different between HFS and LFS (76,302 pulses, range 2,489–161,442 vs. 97,451 pulses, range 1,037–306,578; p=0.8). Therapies delivered per day (tpd) were not significantly different between HFS and LFS (1502 tpd, IQR 298–2519 vs. 1254 tpd, IQR 151–1978; p=0.7). Figure 1 (panel e) shows a long episode ECoG recording from a patient treated with LFS. Typically, 30 sec of detected abnormal activity are required to trigger the storage of a long episode (Long Episode Length). Here, the Long Episode Length was increased from 30 sec to 60 sec since LFS can provide up to 50 sec of stimulation per therapy (if two bursts of 5 sec are repeated 5 times). Additionally, the length of the Capture Window was increased from 90 sec to 180 sec. The Capture Window is divided such that two-thirds is reserved for pre-trigger activity (including the abnormal activity required to trigger the long episode storage), and one-third for post-trigger activity14. Thus, with a 60-sec Long Episode Length, increasing the Capture Window to 180 seconds allows for 60 sec to be recorded before the 60 sec long episode and 60 seconds of post-trigger activity14. Table 2 shows the HFS and LFS parameters used for each patient. There were no LFS-specific adverse effects reported in any of the patients.
Table 2 –
Stimulation parameters
| Patient | Frequency, Hz | Charge density, μC/cm2 | Burst duration, ms | Therapies delivered per day | Charge per hour, mC/h | Time stimulated per day, min/d | Pulses per day | Duration of stimulation, mo | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HFS | LFS | HFS | LFS | HFS | LFS | HFS | LFS | HFS | LFS | HFS | LFS | HFS | LFS | HFS | LFS | |
| 1 | 333 | 7 | 2.5 | 4.5 | 200 | 5000 | 100 | 191 | 0.3 | 0.5 | 0.8 | 40.4 | 16,916 | 17,223 | 169 | 37 |
| 2 | 100 | 7 | 2 | 2.5 | 200 | 5000 | 1502 | 1316 | 0.5 | 7.9 | 12.7 | 278.6 | 76,302 | 118,664 | 59 | 6 |
| 3 | 150 | 7 | 3.5 | 3 | 100 | 5000 | 3869 | 3400 | 3.4 | 24.5 | 16.4 | 719.7 | 147,409 | 306,578 | 46 | 14 |
| 4 | 100 | 7 | 4.6 | 4.6 | 100 | 5000 | 1266 | 604 | 0.5 | 0.8 | 5.4 | 65.2 | 32,156 | 27,776 | 45 | 6 |
| 5 | 200 | 7 | 1 | 5.5 | 100 | 5000 | 1723 | 2087 | 0.9 | 6.8 | 7.3 | 441.7 | 87,528 | 188,185 | 11 | 35 |
| 6 | 100 | 7 | 2 | 3 | 200 | 5000 | 495 | 23 | 0.2 | 0.02 | 4.2 | 2.4 | 25,146 | 1,037 | 18 | 10 |
| 7 | 100 | 7 | 2.5 | 2.5 | 200 | 5000 | 1860 | 1191 | 3.1 | 7.2 | 15.7 | 252.1 | 94,488 | 10,739 | 19 | 2 |
| 8 | 100 | 7 | 1.5 | 3 | 200 | 5000 | 3178 | 1618 | 1.6 | 11.7 | 26.9 | 342.5 | 161,442 | 145,895 | 8 | 14 |
| 9 | NA | 7 | NA | 3.5 | NA | 5000 | NA | 1941 | NA | 2.0 | NA | 205.4 | NA | 87,510 | NA | 26 |
| 10 | 100 | 7 | 0.5 | 1 | 200 | 5000 | 49 | 32 | 0.004 | 0.04 | 0.4 | 6.8 | 2,489 | 2,885 | 5 | 4 |
Pulse width was 160 μsec for all patients when on HFS except for patient 5 who was on 240 μsec
Pulse width was 160 μsec for all patients when on LFS
HFS values represent parameters for burst 1 which are the same for burst 2 except for patients 3 and 4
LFS values represent parameters for burst 1 which are the same for burst 2 except for patients 4,6, and 9
HFS, high frequency stimulation; Hz, Hertz; LFS, low frequency stimulation; mA, milliampere, μC/cm2, microcoulomb per centimeter squared; mC/h, millicoulomb per hour; mo, months; μsec, microsecond; ms, millisecond.
Discussion
In this study of 10 DRE patients treated with the RNS System for predominantly bitemporal mesial epilepsy, LFS was an effective approach to improve seizure control after using standard HFS settings. The time of stimulation per day was significantly longer with LFS than HFS. No adverse events related to LFS were reported. These results suggest that LFS coupled with longer stimulation times may be effective for cortical intracranial stimulation.
In our study we coupled low frequency stimulation with longer burst durations, effectively increasing the time of stimulation per day while maintaining the total amount of charge delivered. In other words, stimulation was delivered over a longer period of time but there was not a significant difference in the charge delivered per hour; thus, we do not expect that LFS will have any significant negative impact of battery longevity (although this has not been verified). The number of therapies delivered per day was not significantly different between HFS and LFS, suggesting that a change in delivered pulses did not lead to the benefit associated with LFS. Similarly, pulses per day were also comparable between HFS and LFS. The benefit of LFS may be from increased stimulation time, lower stimulation frequency, or a combination of the two. Previous reports suggest that for some anatomical structures low frequency may provide a greater6 or lesser15 benefit than high frequency stimulation, which suggests the ideal stimulation frequency may depend on stimulation location. Another possibility is that some patients benefit from lower stimulation frequencies due to characteristics of their epileptic networks5, 16.
A concern regarding the use of LFS is the possibility of seizure exacerbation. However, there is evidence suggesting potential benefit from LFS in mesial temporal epilepsy17, similar to most patients in this study. LFS during invasive epilepsy monitoring has been used for seizure induction to facilitate epileptogenic zone identification, although higher frequencies (e.g. 50 Hz) have been noted to induce seizures more readily18. Chronic LFS of cortical structures has been safe and effective in epilepsy patients with predominantly eloquent seizures onset zones9. Seizure induction through transcranial magnetic stimulation (TMS) has been a safety concern, typically for frequencies of 10 Hz or greater19, although mechanisms underlying TMS differ significantly from invasive neurostimulation making comparisons difficult. For subcortical structures, the seizure exacerbation potential from invasive neurostimulation previously reported by Velasco et al. is restricted to bilateral, high voltage centromedian nucleus thalamic deep brain stimulation (6 Hz, 30 V) in generalized epilepsy with absence seizures20. Other seizure types, including generalized onset, have been treated with thalamic LFS without reported adverse effects11. Prior work suggests that anterior thalamic nucleus (ANT) stimulation at 15–45 Hz may increase synchronization between hippocampus and ANT3. We excluded one patient stimulated with 40 Hz, who noted an 82% seizure reduction compared to baseline and 53% seizure reduction compared to HFS (100 Hz) without side effects from stimulation.
One concern of LFS is of a more technical nature: increasing the burst duration increases the blanking duration of the amplifier to reduce artifact, making the electrographic activity during the seizure more difficult to visualize in the recorded ECoGs. This can be ameliorated by increasing the Long Episode Length such that it is greater than the maximum therapy time (e.g., at least 50 seconds if two bursts of 5 sec are each delivered 5 times) to record only ECoGs of interest. In addition, the Capture Window can be increased, e.g. from 90 sec to 180 sec to store more useful ECoGs (see Figure 1, panel e). Our study is limited by its retrospective nature that carries risks of inconsistencies related to data documentation in the electronic health record, lack of randomized control data and matched cohorts, and selection biases.
In conclusion, although HFS is generally effective in DRE patients treated with the RNS System, LFS offers a viable alternative approach and may be a beneficial RNS programming approach for patients who have not responded to standard high frequency settings. Other studies demonstrate that cortical9 and thalamic21 LFS can be effective, thus LFS as well as HFS may be effective for reducing seizure frequency.
Acknowledgements:
We thank Melinda Marthaler and Emily Mirro for their advice and help in the preparation of this manuscript.
Disclosures:
GW and BNL are named inventors for intellectual property licensed to Cadence Neuroscience Inc, which is co-owned by Mayo Clinic. BNL waived contractual rights to royalties.
GW, NMG, and BNL are investigators for the Medtronic EPAS trial and Medtronic-supported NIH grants (UH3-NS95495 and UH3-NS112826).
BNL is an investigator for Neuropace RNS System Responsive Stimulation for Adolescents with Epilepsy (RESPONSE) Study and Neuroelectrics tDCS for Patients with Epilepsy Study. Mayo Clinic has received consulting fees on behalf of BNL from Epiminder, Medtronic, Neuropace and Philips Neuro.
Mayo Clinic has received research support and consulting fees on behalf of GW from UNEEG, NeuroOne Inc., and Medtronic. GW has licensed intellectual property developed at Mayo Clinic to NeuroOne, Inc.
Funding:
The research was supported by the National Institutes of Health National Institute of Neurological Disorders and Stroke: K23NS112339 (B.N.L.)
References
- 1.Fisher R, Salanova V, Witt T, Worth R, Henry T, Gross R, et al. Electrical stimulation of the anterior nucleus of thalamus for treatment of refractory epilepsy Epilepsia. 2010. May;51:899–908. [DOI] [PubMed] [Google Scholar]
- 2.Morrell MJ, Group RNSSiES. Responsive cortical stimulation for the treatment of medically intractable partial epilepsy Neurology. 2011. Sep 27;77:1295–1304. [DOI] [PubMed] [Google Scholar]
- 3.Yu T, Wang X, Li Y, Zhang G, Worrell G, Chauvel P, et al. High-frequency stimulation of anterior nucleus of thalamus desynchronizes epileptic network in humans Brain. 2018. Sep 1;141:2631–2643. [DOI] [PubMed] [Google Scholar]
- 4.Nair DR, Laxer KD, Weber PB, Murro AM, Park YD, Barkley GL, et al. Nine-year prospective efficacy and safety of brain-responsive neurostimulation for focal epilepsy Neurology. 2020. Sep 1;95:e1244–e1256. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Lim SN, Lee CY, Lee ST, Tu PH, Chang BL, Lee CH, et al. Low and High Frequency Hippocampal Stimulation for Drug-Resistant Mesial Temporal Lobe Epilepsy Neuromodulation. 2016. Jun;19:365–372. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Chaitanya G, Toth E, Pizarro D, Iasemidis L, Murray TA, Riley K, et al. Acute modulation of the limbic network with low and high-frequency stimulation of the human fornix Epilepsy & behavior reports. 2020;14:100363. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stavropoulos I, Selway R, Hasegawa H, Hughes E, Rittey C, Jimenez-Jimenez D, et al. Low frequency centromedian thalamic nuclei deep brain stimulation for the treatment of super refractory status epilepticus: A case report and a review of the literature Brain stimulation. 2021. Mar-Apr;14:226–229. [DOI] [PubMed] [Google Scholar]
- 8.Kinoshita M, Ikeda A, Matsumoto R, Begum T, Usui K, Yamamoto J, et al. Electric stimulation on human cortex suppresses fast cortical activity and epileptic spikes Epilepsia. 2004. Jul;45:787–791. [DOI] [PubMed] [Google Scholar]
- 9.Lundstrom BN, Gompel JV, Khadjevand F, Worrell G, Stead M. Chronic subthreshold cortical stimulation and stimulation-related EEG biomarkers for focal epilepsy Brain communications. 2019;1:fcz010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Child ND, Benarroch EE. Anterior nucleus of the thalamus: functional organization and clinical implications Neurology. 2013. Nov 19;81:1869–1876. [DOI] [PubMed] [Google Scholar]
- 11.Alcala-Zermeno JL, Gregg NM, Wirrell EC, Stead M, Worrell GA, Van Gompel JJ, et al. Centromedian thalamic nucleus with or without anterior thalamic nucleus deep brain stimulation for epilepsy in children and adults: A retrospective case series Seizure. 2021. Jan;84:101–107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Van Gompel JJ, Klassen BT, Worrell GA, Lee KH, Shin C, Zhao CZ, et al. Anterior nuclear deep brain stimulation guided by concordant hippocampal recording Neurosurgical focus. 2015. Jun;38:E9. [DOI] [PubMed] [Google Scholar]
- 13.Lundstrom BN, Worrell GA, Stead M, Van Gompel JJ. Chronic subthreshold cortical stimulation: a therapeutic and potentially restorative therapy for focal epilepsy Expert review of neurotherapeutics. 2017. Jul;17:661–666. [DOI] [PubMed] [Google Scholar]
- 14.NeuroPace RNS System Programming Manual. 2017.
- 15.Boex C, Vulliemoz S, Spinelli L, Pollo C, Seeck M. High and low frequency electrical stimulation in non-lesional temporal lobe epilepsy Seizure. 2007. Dec;16:664–669. [DOI] [PubMed] [Google Scholar]
- 16.Gregg NM, Sladky V, Nejedly P, Mivalt F, Kim I, Balzekas I, et al. Thalamic deep brain stimulation modulates cycles of seizure risk in epilepsy Sci Rep. 2021. Dec 20;11:24250. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Koubeissi MZ, Joshi S, Eid A, Emami M, Jaafar N, Syed T, et al. Low-frequency stimulation of a fiber tract in bilateral temporal lobe epilepsy Epilepsy & behavior : E&B. 2022. Mar 25;130:108667. [DOI] [PubMed] [Google Scholar]
- 18.Trebuchon A, Racila R, Cardinale F, Lagarde S, McGonigal A, Lo Russo G, et al. Electrical stimulation for seizure induction during SEEG exploration: a useful predictor of postoperative seizure recurrence? Journal of neurology, neurosurgery, and psychiatry. 2020. Oct 28. [DOI] [PubMed] [Google Scholar]
- 19.Rossi S, Hallett M, Rossini PM, Pascual-Leone A, Safety of TMSCG. Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology. 2009. Dec;120:2008–2039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Velasco F, Velasco M, Marquez I, Velasco G. Role of the centromedian thalamic nucleus in the genesis, propagation and arrest of epileptic activity. An electrophysiological study in man Acta Neurochir Suppl (Wien). 1993;58:201–204. [DOI] [PubMed] [Google Scholar]
- 21.Alcala-Zermeno JL, Gregg NM, Van Gompel JJ, Stead M, Worrell GA, Lundstrom BN. Cortical and thalamic electrode implant followed by temporary continuous subthreshold stimulation yields long-term seizure freedom: A case report Epilepsy & behavior reports. 2020;14:100390. [DOI] [PMC free article] [PubMed] [Google Scholar]