Abstract
Objective:
To demonstrate similar image quality with deep learning image reconstruction (DLIR) in reduced contrast medium (CM) and radiation dose (double-low-dose) head CT angiography (CTA), in comparison with standard-dose and adaptive statistical iterative reconstruction-Veo (ASIR-V).
Methods:
A prospective study was performed in 63 patients who under head CTA using 256-slice CT. Patients were randomized into either the standard-dose group (n = 38) with 40 ml of Iopromide (370 mgI ml−1 at 4.5 ml s−1); or a double-low-dose group (n = 25) with CM of 25 ml at 3.0 ml s−1. For image reconstruction, the double-low-dose group used DLIR-M and DLIR-H strength, and the standard-dose group used ASIR-V with 50% strength. The CT value and standard deviation, signal-to-noise ratio and contrast-to-noise ratio of posterior fossa, neck muscles, carotid, vertebral and middle cerebral arteries were measured. The image noise, vessel edge and structure blurring and overall image quality were assessed by using a 5-grade method.
The double-low-dose group reduced CM dose by 37.5% and CT dose index by 41% compared with the standard-dose group. DLIR further reduced the standard deviation value of the middle cerebral artery and posterior fossa and provided better overall subjective image quality (p < 0.05).
Conclusion:
DLIR significantly reduces image noise and provides higher overall image quality in the double-low-dose CTA.
Advances in knowledge
It is feasible to reduce CM dose by 37.5% and volume CT dose index by 41% with the combination of 80 kVp and DLIR in head CTA. Compared with ASIR-V, DLIR further reduces image noise and achieves better image quality with reduced contrast and radiation dose.
Introduction
The number of CT scans has been increasing with the increase of clinical requirements1,2 and radiation dose reduction has been the focus of many researchers in the last a few decades. However, reducing radiation dose will inevitably increase image noise which will adversely affect the image quality of CT,3 and the observation of subtle lesions.4,5
There are many ways to reduce image noise and improve image quality under low-dose CT scanning. One of the effective methods is to use iterative reconstruction (IR) algorithms such as the adaptive statistical iterative reconstruction-Veo (ASIR-V) algorithm. In fact, the IR algorithm was used for image reconstruction when the first CT came out, but its development was limited due to its long reconstruction time and the need for expensive computer hardware6 in the early days. Thus, filtered back projection (FBP) became the main method of CT image reconstruction due to its fast reconstruction speed for a long time. However, the image noise of FBP images is large in low-dose conditions.
As people demand higher spatial and contrast resolution, lower noise, faster scanning time and lower radiation dose at the same time, new reconstruction techniques have emerged. At present, a variety of reconstruction algorithms using artificial intelligence have been applied to CT images and have demonstrated good image noise reduction capabilities in early clinical applications.7,8 One of such algorithms is the deep learning image reconstruction (DLIR algorithm developed by GE Healthcare (TrueFidelityTM, Milwaukee). DLIR features a deep neural network (DNN), which was trained with high quality FBP data sets acquired at high radiation doses to learn how to differentiate noise from signals, and to intelligently suppress the noise without impacting anatomical and pathological structures.
The purpose of our research was to investigate the feasibility of using DLIR to significantly reduce both radiation dose and contrast medium dose while maintaining or improving image quality in head CTA compared with the scans using standard radiation and contrast medium (CM) doses and reconstructed with the state-of-the-art ASIR-V algorithm.
Methods and materials
This was a prospective study. Data were collected from patients who underwent CTA examinations in the First Affiliated Hospital of Xi’an Jiaotong University from July 2020 to September 2020. The procedure followed in this study was in line with the standards set by the hospital medical ethics committee, and the informed consent was signed by the participants. Inclusion criteria: patients who were clinically required to undergo cranial CTA examination. The exclusion criteria were: ① renal insufficiency [Glomerular filtration rate <30 ml/min]; ② inability to perform venipuncture; ③ known allergies; ④ heart failure which will affect the adequate contrast flow due to poor heart function; ⑤ external causes causing image quality too poor to be evaluated (One case with emergency cerebral hemorrhage had severe motion artifacts due to irritability was excluded).
Finally, this study collected 63 patients with an average age of 55.39 ± 13.21 years, range (19–83 years). The patients were divided into a standard dose group and a double low-dose group using a random table method.
Imaging technique and post-processing
All patients underwent head CTA on a 256-slice spiral CT scanner (Revolution CT, GE Healthcare, Milwaukee, USA). All patients were injected with CM (Iopromide, 370 mgI ml−1) using a 20G cannula via median elbow vein, and contrast-enhanced scans were manually triggered. The scan range was from the top of the skull to the second cervical vertebra. The scanning parameters and CM dosage were: for the standard-dose group (n = 38), CM volume, 40 ml, injection rate, 4.5 ml s−1, tube voltage, 100 kVp, noise index (NI), 11.5, resulted in a volume CT dose index (CTDIvol) of 9.67 ± 1.14 mGy Images were reconstructed using ASIR-V with 50% strength (ASIR-V50%); for the double low-dose group (n = 25), CM volume, 25 ml, injection rate, 3.0 ml s−1, tube voltage, 80 kVp, NI, 15, resulted in a CTDIvol of 5.69 ± 0.53 mGy. Images were reconstructed using DLIR with medium (DLIR-M) and high (DLIR-H) strength. Both groups used automatic tube current modulation to meet the NI setting for that group, with an additional 40 ml saline injection after CM at a flow rate of 4.5 ml s−1. Compared with the standard-dose group, the double-low-dose group had a 37.5% reduction in CM volume and a 41% reduction in CTDIvol. The image reconstruction layer thickness was 0.625 mm.
Objective evaluation
The reconstructed images were sent to a GE Advantage Workstation (AW4.7) for measurement and analysis by two general radiologists with 12 and 10 years of experience in CT imaging and diagnosis. Mean CT value and standard deviation (SD) of the posterior fossa, neck muscles, carotid artery, vertebral artery and middle cerebral artery of all subjects were measured to calculate the signal-to-noise ratio (SNR = ROIvessel/SDvessel) and contrast-to-noise ratio (CNR) for arteries using muscle as the background: CNR=[(ROIvessel – ROImuscle) / SDmuscle], where ROI vessel and ROI muscle represent the average attenuation of the vessel of interest and paravertebral muscle, respectively, and SD muscle represents image noise.9 The area of region of interest (ROI) was 50 mm2 for the posterior fossa and neck muscle and 2 mm2 for the arteries, and ROI was placed in the center of the blood vessel as much as possible during measurement avoiding calcification and plaque areas. Beam hardening, streaks, and/or partial volume artifacts are prone to appear in the posterior fossa due to the large bony structures in this area10,11 which will artificially increase the non-uniform distribution of CT number or standard deviation. Therefore, in addition to the inherent image noise associated with the scanner and patient-related factors, the SD values in the posterior fossa can also reflect the amount of change in CT numbers caused by beam hardening artifacts.12–14 However, since DLIR further reduces the overall image noise, to reflect the net change of CT number non-uniformity caused by beam hardening artifacts, we adapted the beam hardening artifact (BHA) index concept introduced by Lin et al15 to reduce the influence of background image noise of different reconstructions. Specifically, the BHA was defined as: BHA = , where SDp2 is the measured SD value for the posterior fossa and SDm2 is the SD value for the neck muscle which was served as background in our study. CT radiation dose, including CTDIvol, dose–length product (DLP), were recorded and the effective radiation dose (effective dose, ED) was calculated using the formula ED = DLP×k, where k = 0.0021 mSv/(mGy·cm) for the head. The noise reduction rate for DLIR was calculated as follows: noise reduction rate (DLIR) (%) = (SD(ASIR-V) – SD(DLIR)) / SD(ASIR-V) × 100, where DLIR indicates DLIR-M or DLIR-H).14
Subjective image assessment
The same two radiologists used a 5-level scoring method to assess the degree of image noise, blood vessel edge sharpness and structure clarity, and overall image quality. They were blinded to the reconstruction settings and results from the objective image quality analysis by changing the image orders between the objective measurement and subjective rating in two different image reviews. Two general radiologists read the films independently first, any disagreements in scoring were resolved after discussion to reach the final results. The five grades were as follows: image noise (Grade 0, slight; Grade 1, mild; Grade 2, moderate; Grade 3, high; Grade 4, severe) and sharpness and clarity (Grade 0, no blurring; Grade 1, slightly blurred; Level 2, moderately blurred; Level 3, highly blurred, Level 4, severely blurred).16 The overall image quality of each image was defined as the combination of image noise and blur levels.
Statistical analysis
The SPSS 22 software was used for statistical analysis of the data, and the measurement data were expressed as mean ± standard deviation. Radiation dose-related values between the two groups were compared using the independent sample t-test. The objective measurements (CT value, SD, SNR and CNR) from the two scan groups with three different reconstructions (ASIR50% in the standard dose group and DLIR-M and DLIR-H in the double-low-dose group) were compared using the One-way ANOVA test while the subjective image quality scores were compared using the Kruskal–Wallis test. Cohen’s κ test was used to study the interobserver agreement between the two radiologists for the objective and subjective assessment. A κ value lower than 0.19 was considered poor; 0.20–0.39 was considered fair; 0.40–0.59 was considered moderate; 0.60–0.79 was considered substantial; and 0.80–1.00 was considered perfect. A p < 0.05 in the analysis would indicate that the difference is statistically significant.
Results
In this study, 63 patients who completed head CTA examination were enrolled, including 35 male patients and 28 female patients. They were grouped by random table method and divided into the standard-dose group (n = 38) and double-low-dose group (n = 25). There was no significant difference in the general information of subjects between the two groups (Table 1).
Table 1.
Normal information (objective evaluation)
| Standard-gose group | Double-low-dose group | p | |
|---|---|---|---|
| CTDIvol (mGy) | 9.67 ± 1.14 | 5.69 ± 0.53 | <0.001 |
| DLP (mGy·cm) | 280.78 ± 81.92 | 183.72 ± 60.96 | <0.001 |
| ED (mSv) | 0.54 ± 0.27 | 0.34 ± 0.16 | <0.001 |
| CM (ml) | 40 | 25 | - |
| Rate (ml/s) | 4.5 | 3 | - |
| Age (years) | 56.62 ± 13.66 | 53.97 ± 12.74 | 0.708 |
| Gender (%) | |||
| Male | 61 | 48 | |
| Female | 39 | 52 |
CM: contrast medium;CTDIvol: volumetric CT dose index; DLP: dose–length product; ED: effective dose.
Quantitative analysis of the image noise, SNR, CNR and beam hardening artifact index
The mean CT value and SD, SNR and CNR of posterior fossa, neck muscles, carotid artery, vertebral artery, and middle cerebral artery were measured, and BHA index of posterior fossa were calculated. The measurement results of the two radiologists had good agreement with ICC values greater than 0.79 for all measurements. CT values were similar among different reconstruction groups. The background image noise (neck muscle SD values) was reduced with DLIR and further decreased as DLIR level increased. The noise reduction rate further increases with the increase of DLIR level. The noise reduction rates of DLIR-M and DLIR-H in the muscle were 14.03 and 32.59%, respectively. For the SNR and CNR measurements, except for the SNR values of the middle cerebral artery (where DLIR-H and DLIR-M showed better values (p < 0.05)), there was no difference in quantitative measurements among the three groups (p > 0.05). (Table 2, Figure 1). DLIR significantly reduced the BHA in the posterior fossa area with much smaller SD values (Table 2) and BHA index values for DLIR-M and DLIR-H images (p < 0.001) (Figure 2).
Table 2.
Quantitative analysis (objective evaluation)
| Standard-dose group | Double-low-dose group | p | ||
|---|---|---|---|---|
| DLIR-M | DLIR-H | |||
| Radiologist 1 carotid artery |
||||
| HU | 387.35 ± 73.18 | 372.76 ± 95.33 | 375.61 ± 82.58 | 0.728 |
| SD | 11.89 ± 6.95 | 12.64 ± 8.86 | 10.03 ± 8.74 | 0.759 |
| SNR | 52.73 ± 31.47 | 58.93 ± 25.78 | 69.32 ± 28.07 | 0.728 |
| CNR | 60.85 ± 32.59 | 64.48 ± 41.37 | 81.07 ± 39.49 | 0.937 |
| Vertebral artery | ||||
| HU | 397.53 ± 83.77 | 385.75 ± 89.63 | 385.82 ± 90.03 | 0.634 |
| SD | 11.42 ± 9.25 | 11.96 ± 8.46 | 11.07 ± 6.31 | 0.903 |
| SNR | 73.64 ± 75.49 | 77.35.39 ± 49.27 | 79.50 ± 49.31 | 0.877 |
| CNR | 64.22 ± 32.65 | 67.61 ± 43.71 | 78.94 ± 48.30 | 0.961 |
| Middle cerebral artery | ||||
| HU | 384.91 ± 67.36 | 369.79 ± 82.77 | 369.63 ± 82.07 | 0.619 |
| SD | 15.66 ± 12.08 | 7.93 ± 3.93 | 5.44 ± 3.06 | <0.05 |
| SNR | 56.26 ± 43.83 | 106.78 ± 98.70 | 110.04 ± 134.12 | <0.05 |
| CNR | 58.79 ± 38.04 | 42.94 ± 24.33 | 48.47 ± 24.90 | 0.439 |
| Posterior fossa | ||||
| HU | 47.38 ± 5.73 | 49.84 ± 5.21 | 49.11 ± 6.07 | 0.720 |
| SD | 21.06 ± 3.56 | 13.95 ± 2.16 | 9.49 ± 2.32 | <0.001 |
| BHA | 18.81 ± 2.73 | 11.06 ± 2.05 | 7.02 ± 2.46 | <0.001 |
| Neck muscles | ||||
| HU | 52.32 ± 6.09 | 54.88 ± 7.2 9 | 56.35 ± 6.27 | 0.589 |
| SD | 9.48 ± 1.36 | 8.50 ± 2.36 | 6.39 ± 3.01 | 0.079 |
| Radiologist 2 carotid artery |
||||
| HU | 381.11 ± 86.54 | 377.26 ± 83.23 | 376.61 ± 86.32 | 0.754 |
| SD | 11.54 ± 6.15 | 12.37 ± 9.74 | 10.72 ± 8.65 | 0.728 |
| SNR | 55.87 ± 26.65 | 59.81 ± 26.61 | 73.28 ± 31.19 | 0.736 |
| CNR | 65.93 ± 33.85 | 64.94 ± 43.70 | 77.49 ± 47.73 | 0.918 |
| Vertebral artery | ||||
| HU | 395.82 ± 82.49 | 389.26 ± 87.98 | 389.42 ± 91.15 | 0.637 |
| SD | 12.62 ± 9.12 | 12.96 ± 8.37 | 11.59 ± 6.04 | 0.941 |
| SNR | 70.30 ± 79.18 | 74.04 ± 34.79 | 79.93 ± 48.48 | 0.893 |
| CNR | 63.38 ± 37.87 | 67.19 ± 43.89 | 81.62 ± 46.81 | 0.972 |
| Middle cerebral artery | ||||
| HU | 376.94 ± 65.38 | 369.15 ± 78.30 | 368.25 ± 82.97 | 0.663 |
| SD | 15.09 ± 13.15 | 6.26 ± 3.49 | 5.25 ± 3.07 | <0.05 |
| SNR | 55.14 ± 47.59 | 106.46 ± 103.97 | 113.47 ± 128.74 | <0.05 |
| CNR | 52.25 ± 36.59 | 46.63 ± 22.44 | 50.31 ± 22.61 | 0.401 |
| Posterior fossa | ||||
| HU | 45.48 ± 5.27 | 48.35 ± 5.23 | 49.17 ± 5.63 | 0.759 |
| SD | 22.16 ± 3.90 | 12.62 ± 2.63 | 9.21 ± 1.89 | <0.001 |
| BHA | 19.98 ± 1.97 | 9.63 ± 2.37 | 6.18 ± 2.84 | <0.001 |
| Neck muscles | ||||
| HU | 52.20 ± 6.13 | 54.28 ± 8.37 | 55.17 ± 6.33 | 0.594 |
| SD | 9.59 ± 1.34 | 8.16 ± 2.48 | 6.83 ± 3.26 | 0.084 |
Data are mean value ±standard deviation. DLIR-M, and DLIR-H = deep learning reconstruction with medium and high strength levels, respectively; HU = mean CT number, SD = image noise, SNR = ROIvessel/SDvessel, and CNR = [(ROIvessel – ROImuscle) / SDmuscle],
BHA (Beam hardening artifact) = , where SDp2 is the measured SD value for the posterior fossa and SDm2 is the SD value for the neck muscle.
BHA, beam hardening artifact; CNR, contrast-to-noise ratio; DLIR, deep learning image reconstruction; HU, Hounsfield unit; SD, standard deviation; SNR, signal-to-noise ratio.
Figure 1.

Axial images with 100 kVp ASIR-V50% (a, d, g), 80 kVp with DLIR-H (b, e, h) and DLIR-M (c, f, i) at the second cervical level, median level of orbit, and posterior fossa. DLIR, deep learning image reconstruction.
Figure 2.
Axial images with 100 kVp ASIR-V50% (a), 80 kVp with DLIR-M (b) and DLIR-H (c) at the posterior fossa. ROI was drawn in the interpetrous region (the area indicated by the yellow arrow) to analyze the artifact index. DLIR significantly reduced the beam hardening artifacts in the posterior fossa area. ASIR, adaptive statistical iterative reconstruction; DLIR, deep learning image reconstruction; ROI, region of interest.
Qualitative analysis
The subjective qualitative analysis results for both radiologists are shown in Table 3 and image quality comparison is shown in Figures 1 and 2. The subjective evaluation scores showed that in general as the DLIR intensity level increased, the score gradually decreased from high (worse) to low (better). The observation results of the two radiologists were in substantial agreement (the κ value was 0.658, p < 0.005). Compared with the standard-dose ASIR-V50% images, the overall image quality of the double-low-dose DLIR images were significantly improved between DLIR-H and ASIR-V50% (both radiologists, p < 0.001) (Radiologist 1: 0.15 ± 0.23 vs 2.72 ± 0.58, p < 0.001; Radiologist 2: 0.23 ± 0.35 vs 2.86 ± 0.43, p < 0.001); and between DLIR-M and ASIR-V50% (Radiologist 1: 0.93 ± 0.65 vs 2.72 ± 0.58, p < 0.001; Radiologist 2: 0.83 ± 0.65 vs 2.86 ± 0.43, p < 0.001).
Table 3.
Qualitative analysis
| General group Low-dose group | ||||
|---|---|---|---|---|
| (n = 46) | ASIRV50% | DLIR-M | DLIR-H | p |
| Radiologist 1 | ||||
| Mottle | 2.72 ± 0.58 | 0.93 ± 0.65 | 0.08 ± 0.26 | <0.001 |
| Blurring | 0 | 0 | 0.07 ± 0.26 | 0.019 |
| Overall | 2.72 ± 0.58 | 0.93 ± 0.65 | 0.15 ± 0.23 | <0.001 |
| Radiologist 2 | ||||
| Mottle | 2.86 ± 0.43 | 0.83 ± 0.65 | 0.12 ± 0.35 | <0.001 |
| Blurring |
0 | 0 | 0.11 ± 0.24 | 0.021 |
| Overall | 2.86 ± 0.43 | 0.83 ± 0.65 | 0.23 ± 0.35 | <0.001 |
Data are mean value ±standard deviation.
ASIR, adaptive statistical iterative reconstruction; DLIR, deep learning image reconstruction.
Discussion
The purpose of our study was to compare the image noise and sharpness of the state-of-the-art ASIR and DLIR algorithms in head CTA. We evaluated the DLIR images under reduced radiation dose and contrast medium dose. The results of our study showed that the double-low-dose group using DLIR-M and DLIR-H could further reduce image noise and BHA and achieve better image quality compared with the ASIR-V50% images in the standard-dose group and DLIR images with high intensity level showed the best overall subjective image quality score. The CM dose was reduced by 37.5% and the radiation dose was reduced by 37%.
At present, a large number of studies have shown that thin-slice CT helps to improve the ability to detect lesions.17 Because the thinner-layer reconstruction in cranial CT angiography will cause an increase in image noise, the image quality will be reduced under the same radiation dose. However, increasing the radiation dose to improve the quality of thin-layer images is against the principle of ALARA (as low as reasonably achievable); at the same time, at the same time, increased use of contrast media also increases the risk of CIN (contrast media nephropathy). Therefore, how to balance radiation dose reduction and image quality is a big challenge. Currently, IR algorithm technology is usually used to reduce radiation dose while ensuring image quality.18–20
Although the use of IR algorithms can significantly reduce image noise, under very low dose conditions, the image texture and spatial resolution may be altered, especially when using high-intensity IR algorithms, causing “excessive smoothing”, or simply “unnatural looking”.21,22 Deep-learning techniques have been used in recent years to try to alleviate this problem. Akagi et al7 evaluated the use of a deep learning reconstruction in abdominal ultra-high-resolution CT and indicated that the deep-learning based reconstruction algorithm could reduce image noise and improve image quality compared with other IR algorithms under the same scan conditions. Shen et al8 proposed a modularized DNN for low-dose CT and demonstrated the competitive performance of this algorithm in reducing noise and maintaining structural fidelity compared with other commercial IR algorithms. DLIR is another deep-learning based reconstruction algorithm that uses artificial intelligence to improve the image quality of CT through deep convolutional neural networks (DNN). These networks optimize the images by minimizing the difference between their output and the ideal training sample images. DLIR is developed to overcome the drawbacks of the conventional IR algorithms. By using the high-dose FBP reconstructions as the gold-standard for training, DLIR can significantly reduce noise amplitude while maintain image sharpness, texture, and structural fidelity. DLIR has been applied clinically23 and the technical details describing the DLIR algorithm (TrueFidelityTM) can be found in the manufacturer’s white paper.24 There have been several publications related to the evaluation of DLIR. In our study, we extended the application of DLIR to head where BHA may introduce additional factor to negatively affect image quality and studied the performance of DLIR under the “double low” dose conditions (lower radiation dose and contrast dose). Although previous studies focused on CT images of different organs, our results were still consistent with their results in general. Cao et al23 pointed out that in contrast-enhanced abdominal CT with extremely low radiation dose, DLIR-H yielded a significantly lower image noise, higher CNR and higher overall image quality than the ASIR-V50% under low signal conditions.23 Jensen et al25 reported that in abdomen CTs, TruefidelityTM images showed higher attenuation, lower image noise, and higher CNR values than ASIR-V images with a blending factor of 30%. Park et al26 compared image noise and sharpness of vessels and muscle in lower extremity CT angiography between ASIR-V and DLIR algorithms and found that DLIR-H was the most balanced image in terms of image noise and sharpness among the examined image combinations. Compared with the latest generation IR algorithm (ASIR-V) provided by the same manufacturer, our results showed that the new DLIR algorithm significantly reduced image noise and BHA around the posterior fossa area in head CTA, similar to the results of recently published paper by Kim et al in the routine non-contrast brain CT scans.14 The low-dose images generated by DLIR had similar spatial resolution and noise texture properties to those of standard-dose ASIR images. On the other hand, DLIR algorithms further reduced the magnitude of image noise, resulted in better overall image quality. Using DLIR to process images greatly benefited the display of low-dose head CT images.
Our research has some limitations. This study had a relatively small number of patients, so the results of the study need further validation. Our research requires further studies with more patients to confirm our findings. In this study, only two radiologists performed subjective image quality assessment. The subjective results of our study need to be evaluated by more neuroradiologists.25 Our research only evaluated the similarity between low-dose scanning using DLIR reconstruction and standard-dose IR; it did not evaluate the disease. The potential diagnostic accuracy improvement for small lesions with low-dose scanning and DLIR needs to be investigated in future research. Our research performed subjective evaluation on spatial resolution but did not explicitly measure it. In the future, modular transfer function and noise power spectrum measurement from phantom and patient studies may be added to have more objective evaluation of how DLIR with different strengths influence it. In addition, whether reducing image layer thickness together with DLIR can improve lesion detection may also be added to the research lists.27
In summary, the application of DLIR technology in cranial CT angiography significantly reduces image noise and further improves overall image quality under lower radiation and contrast doses. DLIR provides clinically acceptable image quality in head CTA with 37% radiation dose reduction and 37.5% contrast medium dose reduction compared with the conventional ASIR-V algorithm using standard radiation and contrast doses.
Footnotes
Acknowledgments: This work was supported by the Key R&D Program of Shaanxi Province Universities and Colleges [NO. 2020GXLH-Y-026]. 3D Printing Medical Research Funding Project of the First Affiliated Hospital of Xi'an Jiaotong University [NO. XJTU1AF-3D-2018-003] and Science Development Foundation of the First Affiliated Hospital of Xi’an iaotong University (2018MS-27).
Declaration of Interest statement: One of the authors (J.L.) is an employee of GE Healthcare, the manufacturer of the CT system used in this study. The other authors, who are not GE Healthcare employees, had control of the images and information that might have represented a conflict of interest for (J.L.).
The authors Xin Huang and Wenzhe Zhao contributed equally to the work.
Contributor Information
Xin Huang, Email: 459030854@qq.com.
Wenzhe Zhao, Email: zhaowz@xjtufh.edu.cn.
Geliang Wang, Email: wgeliang@163.com.
Yiming Wang, Email: 386468534@qq.com.
Jianying Li, Email: jianying.li@med.ge.com.
Yanshou Li, Email: 321173057@qq.com.
Qiang Zeng, Email: 10534509506@qq.com.
Jianxin Guo, Email: gjx1665@xjtufh.edu.cn.
REFERENCES
- 1. Arndt C, Güttler F, Heinrich A, Bürckenmeyer F, Diamantis I, Teichgräber U. Deep learning CT image reconstruction in clinical practice. Rofo 2021; 193: 252–61. doi: 10.1055/a-1248-2556 [DOI] [PubMed] [Google Scholar]
- 2. Brenner DJ, Hall EJ. Computed tomography -- an increasing source of radiation exposure. N Engl J Med 2007; 357: 2277–84. doi: 10.1056/NEJMra072149 [DOI] [PubMed] [Google Scholar]
- 3. Ren Z, Zhang X, Hu Z, Li D, Liu Z, Wei D, et al. Reducing radiation dose and improving image quality in CT portal venography using 80 kv and adaptive statistical iterative reconstruction-V in slender patients. Acad Radiol 2020; 27: 233–43. doi: 10.1016/j.acra.2019.02.022 [DOI] [PubMed] [Google Scholar]
- 4. Chi J, Xu D, Yin S, Li M, Shen L, Ding N, et al. Reducing the radiation dose of pediatric paranasal sinus CT using an ultralow tube voltage (70 kvp) combined with iterative reconstruction: feasibility and image quality. Medicine (Baltimore) 2020; 99(34): e21886. doi: 10.1097/MD.0000000000021886 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Hill MD, Demchuk AM, Tomsick TA, Palesch YY, Broderick JP. Using the baseline CT scan to select acute stroke patients for IV-IA therapy. AJNR Am J Neuroradiol 2006; 27: 1612–16. [PMC free article] [PubMed] [Google Scholar]
- 6. Arcadu F, Nilchian M, Studer A, Stampanoni M, Marone F. A forward regridding method with minimal oversampling for accurate and efficient iterative tomographic algorithms. IEEE Trans Image Process 2016; 25: 1207–18. doi: 10.1109/TIP.2016.2516945 [DOI] [PubMed] [Google Scholar]
- 7. Akagi M, Nakamura Y, Higaki T, Narita K, Honda Y, Zhou J, et al. Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol 2019; 29: 6163–71. doi: 10.1007/s00330-019-06170-3 [DOI] [PubMed] [Google Scholar]
- 8. Shan H, Padole A, Homayounieh F, Kruger U, Khera RD, Nitiwarangkul C, et al. Competitive performance of a modularized deep neural network compared to commercial algorithms for low-dose CT image reconstruction. Nat Mach Intell 2019; 1: 269–76. doi: 10.1038/s42256-019-0057-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Marin D, Nelson RC, Schindera ST, Richard S, Youngblood RS, Yoshizumi TT, et al. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm -- initial clinical experience. Radiology 2010; 254: 145–53. doi: 10.1148/radiol.09090094 [DOI] [PubMed] [Google Scholar]
- 10. Pomerantz SR, Kamalian S, Zhang D, Gupta R, Rapalino O, Sahani DV, et al. Virtual monochromatic reconstruction of dual-energy unenhanced head CT at 65-75 keV maximizes image quality compared with conventional polychromatic CT. Radiology 2013; 266: 318–25. doi: 10.1148/radiol.12111604 [DOI] [PubMed] [Google Scholar]
- 11. Riordan AJ, Bennink E, Viergever MA, Velthuis BK, Dankbaar JW, de Jong HWAM. Ct brain perfusion protocol to eliminate the need for selecting a venous output function. AJNR Am J Neuroradiol 2013; 34: 1353–58. doi: 10.3174/ajnr.A3397 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Li X, Li Z, Li J, Song J, Yu Y, Liu B. Optimize non-contrast head CT imaging tasks using multiple virtual monochromatic image sets in dual-energy spectral CT. J Xray Sci Technol 2020; 28: 345–56. doi: 10.3233/XST-190617 [DOI] [PubMed] [Google Scholar]
- 13. Dodig D, Kovačić S, Matana Kaštelan Z, Žuža I, Benić F, Slaven J, et al. Comparing image quality of single- and dual-energy computed tomography of the brain. Neuroradiol J 2020; 33: 259–66. doi: 10.1177/1971400920920785 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Kim I, Kang H, Yoon HJ, Chung BM, Shin N-Y. Deep learning-based image reconstruction for brain CT: improved image quality compared with adaptive statistical iterative reconstruction-veo (ASIR-V). Neuroradiology 2021; 63: 905–12. doi: 10.1007/s00234-020-02574-x [DOI] [PubMed] [Google Scholar]
- 15. Lin XZ, Miao F, Li JY, Dong HP, Shen Y, Chen KM. High-Definition CT gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr 2011; 35: 294–97. doi: 10.1097/RCT.0b013e3182058d5c [DOI] [PubMed] [Google Scholar]
- 16. Park C, Choo KS, Kim JH, Nam KJ, Lee JW, Kim JY. Image quality and radiation dose in CT venography using model-based iterative reconstruction at 80 kvp versus adaptive statistical iterative reconstruction-V at 70 kvp. Korean J Radiol 2019; 20: 1167–75. doi: 10.3348/kjr.2018.0897 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Zhan Y, Peng X, Shan F, Feng M, Shi Y, Liu L, et al. Attenuation and morphologic characteristics distinguishing a ground-glass nodule measuring 5-10 mm in diameter as invasive lung adenocarcinoma on thin-slice CT. AJR Am J Roentgenol 2019; 213: W162–70. doi: 10.2214/AJR.18.21008 [DOI] [PubMed] [Google Scholar]
- 18. Goodenberger MH, Wagner-Bartak NA, Gupta S, Liu X, Yap RQ, Sun J, et al. Computed tomography image quality evaluation of a new iterative reconstruction algorithm in the abdomen (adaptive statistical iterative reconstruction-V) a comparison with model-based iterative reconstruction, adaptive statistical iterative reconstruction, and filtered back projection reconstructions. J Comput Assist Tomogr 2018; 42: 184–90. doi: 10.1097/RCT.0000000000000666 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Lee NK, Kim S, Hong SB, Kim TU, Ryu H, Lee JW, et al. Low-Dose CT with the adaptive statistical iterative reconstruction V technique in abdominal organ injury: comparison with routine-dose CT with filtered back projection. AJR Am J Roentgenol 2019; 213: 659–66. doi: 10.2214/AJR.18.20827 [DOI] [PubMed] [Google Scholar]
- 20. Xu Y, He W, Chen H, Hu Z, Li J, Zhang T. Impact of the adaptive statistical iterative reconstruction technique on image quality in ultra-low-dose CT. Clin Radiol 2013; 68: 902–8. doi: 10.1016/j.crad.2013.03.024 [DOI] [PubMed] [Google Scholar]
- 21. Verdun FR, Racine D, Ott JG, Tapiovaara MJ, Toroi P, Bochud FO, et al. Image quality in CT: from physical measurements to model observers. Phys Med 2015; 31: 823–43. doi: 10.1016/j.ejmp.2015.08.007 [DOI] [PubMed] [Google Scholar]
- 22. Samei E, Richard S. Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 2015; 42: 314–23. doi: 10.1118/1.4903899 [DOI] [PubMed] [Google Scholar]
- 23. Cao L, Liu X, Li J, Qu T, Chen L, Cheng Y, et al. A study of using A deep learning image reconstruction to improve the image quality of extremely low-dose contrast-enhanced abdominal CT for patients with hepatic lesions. Br J Radiol 2021; 94(1118): 20201086. doi: 10.1259/bjr.20201086 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Healthcare GE. A new era of image reconstruction: TrueFidelity; 2019. [Google Scholar]
- 25. Jensen CT, Liu X, Tamm EP, Chandler AG, Sun J, Morani AC, et al. Image quality assessment of abdominal CT by use of new deep learning image reconstruction: initial experience. AJR Am J Roentgenol 2020; 215: 50–57. doi: 10.2214/AJR.19.22332 [DOI] [PubMed] [Google Scholar]
- 26. Park C, Choo KS, Jung Y, Jeong HS, Hwang J-Y, Yun MS. Ct iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol 2021; 31: 3156–64. doi: 10.1007/s00330-020-07358-8 [DOI] [PubMed] [Google Scholar]
- 27. Higaki T, Nakamura Y, Tatsugami F, Nakaura T, Awai K. Improvement of image quality at CT and MRI using deep learning. Jpn J Radiol 2019; 37: 73–80. doi: 10.1007/s11604-018-0796-2 [DOI] [PubMed] [Google Scholar]

