. 2023 Feb 14;8(8):7615–7625. doi: 10.1021/acsomega.2c07175

Sequential Injection Analysis for Rapid Determination of Mercury in Skincare Products Based on Fluorescence Quenching of Eco-Friendly Synthesized Carbon Dots

Kanokwan Sakunrungrit , Cheewita Suwanchawalit , Kanokwan Charoenkitamorn , Apisake Hongwitayakorn , Kamil Strzelak §, Sumonmarn Chaneam †,∥,*
PMCID: PMC9979229  PMID: 36872964

Abstract

graphic file with name ao2c07175_0011.jpg

This work reports the analysis of mercury using a spectrofluorometric method combined with a sequential injection analysis (SIA) system. This method is based on the measurement of fluorescence intensity of carbon dots (CDs), which is quenched proportionally after adding mercury ions. Herein, the CDs underwent environmentally friendly synthesis using a microwave-assisted approach that provides intensive and efficient energy and shortens reaction time. After irradiation at 750 W for 5 min in a microwave oven, a dark brown CD solution with a concentration of 2.7 mg mL–1 was obtained. The properties of the CDs were characterized by transmission electron microscopy, X-ray diffractometry, X-ray photoelectron spectroscopy, Fourier-transform infrared spectroscopy, and UV–vis spectrometry. We presented for the first time the use of CDs as a specific reagent for the determination of mercury in skincare products with the SIA system to achieve rapid analysis and full automatic control. The as-prepared CD stock solution was diluted 10 times and used as a reagent in the SIA system. Excitation and emission wavelengths at 360 and 452 nm, respectively, were used to construct a calibration curve. Physical parameters affecting the SIA performance were optimized. In addition, the effect of pH and other ions was investigated. Under the optimum conditions, our method showed a linear range from 0.3 to 600 mg L–1 with an R2 of 0.99. The limit of detection was 0.1 mg L–1. Relative standard deviation was 1.53% (n = 12) with a high sample throughput of 20 samples per hour. Finally, the accuracy of our method was validated by comparison using inductively coupled plasma mass spectrometry. Acceptable recoveries were also presented without a significant matrix effect. This method was also the first time that uses the untreated CDs for the determination of mercury(II) in skincare products. Therefore, this method could be an alternative for mercuric toxic control in other sample applications.

1. Introduction

At present, there has been increasing concern about the safety of cosmetic and pharmaceutical products used in everyday life. Mercury is a toxic metal that presents intentionally or accidentally in some treatments of skin conditions, particularly whitening products. Mercury can be presented in these products, although it is not listed on the label or on the product information. However, it may accumulate in the skin and exert harmful side effects on the skin and internal organs when used in large quantities for a long period of time. For example, in the case of a woman who uses mercury-containing cream, mercury levels are high in the hair, blood, and urine. The long-term use of mercury-containing lighteners often produces a gray skin color.13 Therefore, many countries, including Thailand, have prohibited the use of mercury in these products not higher than 1 mg L–1.4 In Thailand, face whitening cream holds a 60% share of the national market for facial lotion, with approximately 2100 million Baht ($70 million).5 Therefore, increasing public awareness about mercury contamination in skincare products is necessary.

Traditional techniques have been utilized for detecting mercury ions (Hg2+) in skincare and cosmetic products such as titration,6 neutron activation analysis,7 and cold vapor atomic absorption spectrometry.8 Inductively coupled plasma mass spectrometry (ICP-MS) has become a powerful analytical technique for Hg2+ analysis.9,10 However, the instrument is expensive, and it requires highly skilled operators. An alternative for optical sensing systems for the detection of Hg2+ can also be based on the fluorescence assay because of its high sensitivity and fast analysis. Many fluorescent probes, including organic molecules, have been developed for fluorescent Hg2+ detection in various samples based on the ionophore/fluorophore reagent.1114 However, some reagents have difficult and multi-step synthesis, a considerable amount of products, and chemical toxicity. In addition, most reagents are insoluble in an aqueous medium, which is an unfavorable condition of the sample preparation medium and of the flow-based system including the sequential injection analysis system (SIA). There are previous works reported using the SIA for determination of Hg2+, for example, SIA with cold vapor atomic absorption spectrometry,15 SIA with anodic stripping voltammetry,16 and SIA with spectrophotometry.17

Nanoparticles have been used to address the abovementioned problems. Recent advances in nanotechnology have provided new methodologies for Hg2+ sensing contaminants.18,19 For example, colorimetric methods based on surface plasmon resonance of gold nanoparticles have been used for testing Hg2+.20 However, these gold nanoparticles are unstable, and they require time-consuming modification processes. Among these nanoparticles, carbon-based nanomaterials such as carbon dots (CDs) are increasingly used in various fields, including food, agricultural, industrial, health, and medical purposes, because of their unique physical and chemical properties. CDs have easy synthesis, biocompatibility, and low toxicity.21 Elemental analysis showed that CDs are composed of C, H, O, and N, depending on the carbon precursors. In general, CDs should be further purified by using centrifugation, dialysis, electrophoresis, or another separation technique to control the size and improve the photoluminescence properties. Some studies use CDs as a specific sensor for heavy metals.2227 Most of the CD probes for Hg2+ were applied to analyze Hg2+ in water samples2832 and breast milk.33 In addition, CDs were investigated as the Hg2+ sensor for bioimaging.34,35

As described previously, CDs have been applied for Hg2+ analysis. However, CDs have never been applied for the assessment of Hg2+ in skincare products. Herein, we presented an easy preparation of CDs through a one-step and short-time method under a microwave-assisted approach by using common chemical reagents such as citric acid and urea as initial substances. The resultant CDs exhibited strong blue fluorescence emission in aqueous solution. After adding Hg2+ to the untreated CD solution, precipitation and aggregation of Hg-CDs occurred. Simultaneously, fluorescence emission of the solution turned off. Based on this phenomenon, a fluorometric detection was explored and used for quantitative analysis of mercury.

In this work, the feasibility of CDs for Hg2+ analysis in a real sample was also demonstrated by employing the SIA system for microliter handling to reduce the sample and reagent as well as achieve the automatic operation. The developed procedure was applied to monitor the Hg2+ content in skincare products, particularly whitening cream.

2. Experimental Section

2.1. Materials and Instruments

All the experiments were performed in aqueous solution. Citric acid anhydrous 99.5% (Loba Chemie Pvt., Ltd., India) and urea 99.5% (Loba Chemie Pvt., Ltd., India) were used for CD synthesis. The standard solution of 1000 mg/L Hg2+ was prepared by dissolving mercury(II) acetate (Sigma-Aldrich, USA) in deionized water and used as a stock solution. The working standard solution of 0.3–600 mg L–1 Hg2+ was prepared by diluting the stock solution. Acetate buffer (0.10 M) at pH 6 and 7 was prepared from acetic acid (Carlo Erba, Italy) and sodium acetate (Sigma-Aldrich, USA). Phosphate buffer (0.1 M) at pH 6 was prepared from sodium phosphate (Sigma-Aldrich, USA). All of the solutions were prepared in deionized water.

In this work, we aimed to quantitatively analyze Hg2+ in skincare products. Different products for skin conditions, particularly whitening cream, were purchased from the online market, local cosmetic shops, and supermarkets in Nakhon Pathom, Thailand. For analysis, the exact weight of 0.10 g of the sample was dissolved in 0.50 mL of 5% (v/v) HNO3 (Carlo Erba, Italy), and the volume was made to 25 mL in a volumetric flask by 0.1 M acetated buffer at pH 7.0, which was then injected to the SIA system. All samples were analyzed using the proposed flow system and the ICP-MS reference method.

2.2. Microwave-Assisted Synthesis of CDs

The CDs were synthesized by a microwave-assisted method of the previously reported procedure with a slight modification.36,37 The mixed solution composed of 1.00 g of citric acid and 1.00 g of urea to 10.00 mL of deionized water was irradiated at 750 W for 5 min in a microwave oven (Samsung model MW71B) to give the CD suspension. Finally, the as-synthesized CDs were collected and stored at 4 °C for further characterization and application of Hg2+ detection.

Fluorescence quantum yield (QY) was determined by a comparative method.3840 Quinine sulfate in 0.1 M H2SO4 was used as a standard solution to calculate the QY of synthesized CD samples, which were dissolved in deionized water. All the absorbance values of the solutions at the excitation wavelength were measured with a UV–vis spectrophotometer. Photoluminescence (PL) emission spectra of all the sample solutions were recorded by a PerkinElmer LS-50B luminescence spectrometer at an excitation wavelength of 360 nm. The fluorescence quantum yield of the as-prepared CDs was calculated by comparing the UV–vis absorbance values and the integrated fluorescence intensities (λex = 360 nm) of the CDs with those of quinine sulfate solution. The absorbance values of all solutions were maintained under 0.05 to minimize self-absorption. The fluorescence quantum yield of obtained CDs was calculated using the following equation:

2.2.

where φ is the quantum yield, A stands for the optical density, F represents the integrated emission intensity, and η is the refractive index of the solvent. “s” and “u” correspond to the standard and the sample, respectively.

2.3. Microwave Digestion and ICP-MS for Method Validation

For sample preparation, microwave digestion was used. The exact weight of 0.20 g of each sample was transferred into a microwave vessel. Then, 9.00 mL of nitric acid (conc. 65%, w/w) and hydrogen peroxide (conc. 30% w/w) was added into the microwave vessel as digestion reagents. Finally, under microwave digestion around 50 min, the sample solution was transferred into a volumetric flask, and the volume was made up to 100 mL with deionized water. Afterward, the prepared sample solutions were analyzed by ICP-MS. ICP-MS (7900 ICP-MS, Agilent, USA), with a detection limit of mercury of 0.001 mg L–1, was used as the standard method to validate mercury analysis.

2.4. Characterizations

The properties of the as-prepared CDs before and after reacting with Hg2+ were characterized extensively. The structures and morphologies were examined using X-ray diffractometry (XRD; Aris, PANalytical, UK), X-ray photoelectron spectroscopy (XPS) (Kratos Axis Ultra spectrometer, Manchester, UK, with a monochromic Al Kα source at 1486.7 eV), and transmission electron microscopy (TEM) (FE-TEM/STEM-EDS, Thermo Scientific Talos F200X STEM, USA). Fourier-transform infrared spectra of standard mercury(II) acetate, dried CDs, and dried Hg-CDs were recorded on Frontier, PerkinElmer, USA. UV–visible absorption spectra were recorded on Cary 60, Agilent, USA, and fluorescence spectra were recorded using a spectrofluorometer (LS 55, PerkinElmer, USA).

2.5. Sequential Injection Procedure

The SIA with a syringe pump (Cavro XLP 6000, Switzerland) and a selection valve (Cavro Smart Valve, Switzerland) was operated (Figure 1). A 5 mL zero-dead-volume syringe was fitted with a holding coil. The programmable pump and valve were automatically controlled, via an RS-232 communication port, using the program developed in C# language running under an MS-Windows environment. The graphical user interface of the program practically guided users to input parameters and steps of the procedure as necessary. The amount of Hg2+ in skincare products was determined using an SIA coupled spectrofluorometer. As shown in Table 1, the common sequential injection procedure has five steps for one cycle. First, a 3000 μL carrier was aspirated into the system at a flow rate of 10 mL min–1. Next, two 100 μL segments of standard or sample (port 2) partition with a 300 μL segment of the diluted CD solution (port 3) were sequentially aspirated into a holding coil, namely, sandwich pattern, at a flow rate of 10 mL min–1. Then, zone stacking was sent to a reaction coil (PTFE, 0.75 mm, 100 cm) and continuously propelled to the detection cell through port 4 at a flow rate of 2.5 mL min–1 with no need of any cleaning step. The fluorescence intensity was recorded at an excitation wavelength and emission wavelength of 360 and 452 nm, respectively. A decreased signal (compared with the blank sample) was observed because the fluorescence of the CDs was quenched by Hg2+ in the sample.

Figure 1.

Figure 1

Schematic diagram of the SIA system for the determination of mercury.

Table 1. SIA Procedure Operated in This Work.

step flow rate (mL min–1) volume (μL) flow direction event
1 10 3000 reverse carrier aspirated
2 10 100 reverse standard/sample zone segment 1 aspirated
3 10 300 reverse reagent zone aspirated
4 10 100 reverse standard/sample zone segment 2 aspirated
5 2.5 3500 forward zones sent to the spectrofluorometer

3. Results and Discussion

3.1. Concentration and Size of CDs

The CDs were synthesized by a microwave-assisted method as described above. A hundred microliters of the as-prepared CD solution was pipetted onto a glass slide and was evaporated at 80 °C for 24 h. Finally, the concentration of the synthesized CDs was found to be 2.7 mg mL–1. The fluorescence quantum yield of the CDs was calculated to be about 0.16%. This result might be due to the effect of low energy transfer of large untreated CDs. However, this QY value is sufficient for our application. To use the CDs as a reagent in the SIA system, the optimal concentration of the CD solution was investigated by dilution to 0.54, 0.27, and 0.14 mg mL–1 and the concentration of 0.27 mg L–1 showed the best sensitivity. Therefore, the as-prepared stock CD solution was diluted 10 times in deionized water. The clear diluted CD solution was used as a reagent for Hg2+ analysis with the SIA. The stability of the CDs was also investigated by measuring the fluorescence intensity and sensitivity of the flow system for Hg2+ determination. The results revealed that the sensitivity of our system was stable up to 3 months after preparation. Precision of synthesis was considered by measuring the emission intensity at 452 nm of the 10 times dilution of the as-prepared CDs obtained from each batch. Therefore, inter-batch precision of the synthesis of CDs was reported as 2.9%RSD (n = 3).

3.2. Characterization of CDs

In this paper, the CDs were synthesized using a microwave-assisted method. TEM was used to characterize the morphology of the untreated CDs. The TEM image (Figure 2A) shows spherical nanoparticles with consistent dispersion. In this work, about 71% of the size distribution is in the range of 10–20 nm (Figure 2B). After analyzing random particles, the CD mean size was 14.7 ± 4.8 nm. A small number of particles are relatively of bigger size because of the agglomeration of smaller particles. Figure 2C shows the XRD patterns of the CDs obtained in the range from 10 to 70°. The XRD pattern shows that the CDs have a broad diffraction peak at 27.2°, which indexes to the (002) plane of graphitic carbon.41,42 This peak reveals the predominately amorphous structure of the as-prepared CDs.

Figure 2.

Figure 2

Typical TEM image of as-prepared CDs (A) and CDs’ size distribution (B). XRD pattern of the as-prepared CDs (C). FT-IR spectra of the synthesized CDs and Hg-CDs (D). The UV–vis spectrum of the CD solution shows absorption at approximately 220–340 nm (E). Photoluminescence emission spectra of the CDs in aqueous solutions at different excitation wavelengths (300–400 nm) (F).

The surface functional groups of CDs were characterized by FT-IR (Figure 2D). A broad peak of approximately 3420 cm–1 was indicative of OH stretching. Absorption bands at 3177 cm–1 were attributed to NH2. These functional groups typically existed on the surface of CDs.38 The characteristic peaks of the C=O stretching vibration, which are a typical observation for CDs, are shown at 1701 and 1661 cm–1. The 1576 and 1350 cm–1 peaks were attributed to the stretching vibrations of C=C and C–N, respectively.43,44 The FT-IR spectrum of dried Hg-CDs was also evaluated. Two peaks at approximately 400–600 cm–1 were found dominant. Based on our reviews in previous works, the characteristic peaks at 576 and 468 cm–1 were defined as Hg–O in vibrational mode, which confirms the formation of Hg–O on the surface of Hg-CDs.45 Notably, the peaks of C=O and broad peaks of −OH and −NH2 were shifted, and their intensities were reduced. The UV–vis spectrum of the CD solution (Figure 2E) shows main absorption bands at approximately 220, 260, and 340 nm, which correspond to the carbon–carbon double bond’s π → π* and aromatic ring’s π → π* transitions. The absorption peak at 340 nm in the UV–vis absorption spectra of the synthesized CDs confirmed that the presence of the C=C functional group of the graphitic structure occurs during the carbonization process.46,47 As shown in Figure 2F, when the CDs were excited at 360 nm, strong fluorescence emission at 452 nm was observed. The maximum fluorescence peak shifted from 420 to 530 nm with a change of excitation wavelength from 300 to 400 nm, respectively.

Furthermore, elemental compositions of CDs and Hg-CDs were also investigated by the XPS technique.37,47,48Figure 3 presents the high-resolution XPS spectra of C (1s), N (1s), O (1s), and Hg (4f) of CDs and Hg-CDs. The XPS survey scan spectrum of the CDs (Figure 3A) revealed three apparent binding peaks corresponding to the O (1s), N (1s), and C (1s). The deconvolution spectrum of C (1s) showed four peaks at 284.64, 285.59, 287.05, and 288.44 eV, which can be attributed to C=C (sp2), C–N, C–OH/C–O–C, and C=O groups, respectively.49 The high-resolution N (1s) spectrum (Figure 3A) was deconvoluted into two peaks at 399.07 and 400.22 eV, corresponding to C–NH2 and O=N—C functional groups, respectively. The O (1s) spectrum (Figure 3A) was deconvoluted to three peaks at 530.84, 531.91, and 533.14 eV, indicating the presence of −C–O/–N–O, —C=O, and C–O–C functional groups, respectively.50,51 The XPS results confirmed that there are many functional groups (−NH2, −COOH, and −OH) on the surface of CDs corresponding with FT-IR results. The XPS spectra of the Hg-CDs (Figure 3B) are composed of O, N, C, and Hg elements. The C (1s) spectrum could be deconvoluted into four peaks at 284.94, 286.03, 286.96, and 288.52 eV, corresponding to the C=C (sp2), C–N, O—C=O, and C=O groups, respectively.49 The N (1s) spectrum reveals the presence of the amine group (C–NH2 (398.56 eV) and O=N—C (399.74 eV)). The O (1s) spectrum was deconvoluted to four peaks at 530.09, 531.47, 532.28, and 533.47 eV, suggesting the presence of −C–O/–N–O, —C=O, C–OH, and C–O–C groups, respectively.48,50,51 For Hg, the (4f) spectrum found two peaks of 4f5/2 and 4f7/2 at 105.18 and 101.13 eV, respectively.52 The results showed that mercury was adsorbed on CDs via oxygen-containing functional groups and amine group in the adsorption of mercury ions.48,52

Figure 3.

Figure 3

XPS spectra of the CD (A) and Hg-CD products (B) obtained.

3.3. Detection of Mercury Ions Using the Batch Method

Fluorescence quenching of CDs was achieved by hand mixing the aqueous solutions of CDs and Hg2+. At first, Figure 4 presents images of the untreated CDs and products after reacting with Hg2+ under visible (Figure 4A) and UV light (Figure 4B). The bright blue photoluminescence is strong and easily seen with the naked eyes. In addition, 100 μL of the diluted CDs was added to 3.00 mL of deionized water in a quartz cuvette, and the fluorescence intensity of CDs was recorded as blank. Then, the standard solution of 0–100 μL of 1000 mg L–1 Hg2+ was added into 10 μL of solution increment each time. Afterward, the solution was mixed by shaking the cuvette. After Hg2+ was introduced into the CD solution, Hg-CD aggregation was induced; thus, fluorescence intensity of the CDs disappears soon. The aggregation of CDs is enforced upon the addition of the mercury ion. Consequently, the color of solution turns from pale yellow to colorless. However, UV–visible absorption of CDs and CDs after adding Hg2+ solution is less sensitive, and it could not be applied to real samples. Therefore, photoluminescence was used by measuring the maximum emission peak at 452 nm with an excitation at 365 nm. As shown in Figure 5A,B, the fluorescence intensity decreased gradually with the increase in Hg2+ concentration. The formation of non-luminescent Hg-CD aggregation might lead to quenching of fluorescence intensity. A linear calibration curve (Figure 5C) was obtained, and it shows that the untreated CDs could be utilized for quantitative analysis of Hg2+. Based on the above results, high selectivity toward Hg2+ in aqueous solution could be probably a coordinative interaction with Hg2 + .and the carboxyl and hydroxyl groups on the surface of CDs. The photoluminescence intensity decreases with the increasing concentrations of Hg2+. The photoluminescence quenching of CDs was caused by Hg2+chelation between surface functional groups of CDs and Hg2 + via charge transfer, leading to the fluorescence quenching of the CDs.5356

Figure 4.

Figure 4

Photographs under visible (A; daylight lamp) and UV light (B; λex = 365 nm). (The concentration of CDs is 0.27 mg mL–1, and the concentration of Hg2+ is 60 mmol L–1). All photos were taken by S.C.

Figure 5.

Figure 5

Corresponding fluorescence images under a 365 nm UV lamp (A) taken by S.C. and fluorescence emission spectra of a mixture of CDs with the addition of Hg2+ (B). Calibration curve plot of fluorescence intensity at 452 nm at various concentrations of Hg2+ (C). Effect of pH on the fluorescence intensity (D).

A dependence of the CD fluorescence intensity on the pH value was reported. Similar to previous works,21 we found that the intensity decreased when the pH value of the CD solution was lower than 5. The fluorescence intensity of Hg-CDs is also pH-dependent (Figure 5D). At the acidic medium, low quenching efficiency was observed, which results from the dissociation of the Hg-CD compound caused by the protonation of surface-binding carboxyls. With the increase in pH, the deprotonation of the carboxylic groups in the CDs occurs. This phenomenon may strengthen the covalent bond between Hg2+ and CDs, which leads to higher quenching efficiency and lower fluorescence intensity.37 At the high basic medium, the precipitate of mercury hydroxide might occur and reduce the Hg-CD compound. From the results, pH 7.0 was used. Various types of medium solutions at pH 7 were further studied, which are DI water, acetate buffer, and phosphate buffer, and the results show that acetate buffer at pH 7 gives the best sensitivity.

3.4. Optimization of Physical Parameters Affecting the SIA Performance

The experiment is shown in Figure 1. The working standard solution of 0.3–600 mg L–1 Hg2+ was used to optimize the physical parameters. First, the volumes of the sample and reagent were optimized to reduce their amount (Figure 6A,B). Notably, the solution volume needed for the proposed SIA method can be reduced to the microliter level. We found that 200 μL of the sample (divided in two aliquots of 100 μL intercalated with the reagent aliquot) was sufficient to cover the analysis range with a satisfactory signal, and the CD reagent solution at 300 μL was selected for all further experiments. Next, the flow rate sent to the spectrofluorometer, which has a significant effect on the measured signal, was studied. In this experiment, the flow rate varied from 2.5, 5, 10, and 15 mL min–1. As shown in Figure 6C, the sensitivity decreased when the flow rate was increased. In accordance with the sensitivity, we selected 2.5 mL min–1 as the optimal flow rate. In addition, we observed that even if a flow rate higher than 10 mL min–1 was used, sample throughput was not significantly improved, but a large noise was observed. At a 2.5 mL min–1 flow rate, we obtained a satisfactorily high throughput of 20 sample h–1.

Figure 6.

Figure 6

Sensitivity of calibration curves obtained from physical parameter studies: sample volume (A), reagent volume (B), and flow rate to detector (C). The optimized conditions are as follows: a flow rate of 2.5 mL min–1, volumes of the sample and reagent of 0.2 and 0.3 mL, respectively, and no waiting time.

3.5. Effect of Other Metal Ions

Figure 7 shows the results of the effect of coexisting metal ions. First, batch experiments were performed by evaluating the fluorescence intensity under UV light. A hundred microliters of diluted CDs was added to a medium of 3.00 mL of DI water in a vial, and then, 80 μL of 100 mg L–1 metal ion solution was added (Figure 7A). The fluorescence intensity was recorded with an emission wavelength of 452 nm and excitation wavelength of 360 nm (Figure 7B). The effect of metal ions on Hg2+ detection was also investigated by injecting 200 μL of 100 mg L–1 for each metal ion solution into the SIA system, and the signal profile and bar graph are shown in Figure 7C and Figure 7D, respectively, which were consistent with the data obtained from the batch experiment (Figure 7A,B). Notably, interferences were not observed from Cd2+, Ba2+, Ni2+, Ca2+, Na+, K+, and Zn2+. Pb2+ and Mn2+ had a slight effect on the fluorescence intensity, which could be negligible. This result may be due to the stability constants between the Hg2+ and carboxylic group, which are higher than other metal ions leading to the formation of a Hg–O non-fluorescent metal adduct.52 In the case of Fe2+ and Cu2+, less fluorescence intensity might result from metal hydroxide precipitation. Quenching of CDs by Fe2+ and Cu2+ was also found in previous works.33 In addition, 2 and 20 mg L–1 Fe2+ and Cu2+ were tested by injecting to the SIA system. The results showed that our method can tolerate the presence of Fe2+ and Cu2+ at the tested levels. However, these ions are rarely found in the tested skincare products.

Figure 7.

Figure 7

Photograph of fluorescence change under UV light (A) taken by S.C. and emission spectra (B) of diluted CDs in acetate buffer pH 7.0 solution upon the addition of various metal ions, including Hg2+, Pb2+, Cd2+, Fe2+, Ba2+, Cu2+, Mn2+, Ni2+, Ca2+, Na+, K+, and Zn2+. SIA experiments with the same set of foreign metal ions used acetate buffer pH 7.0 solution as the carrier (C, D).

3.6. Analytical Features

Figure 8A illustrates the signal profile obtained from the SIA system with various concentrations of Hg2+ and calibration curve obtained from the signal profile for the determination of Hg2+ in samples (Figure 8B). The horizontal axis is the concentration of Hg2+ in mg L–1, and the vertical axis is the difference between fluorescence intensity of blank (I0) and fluorescence intensity of the sample (Is). The linear equation of 0.3–10 mg L–1 is y = (0.846 ± 0.040)x + (4.243 ± 0.192) with R2 = 0.991 and of 10–600 mg L–1 is y = (0.129 ± 0.006)x + (12.123 ± 2.119) with R2 = 0.991. The results showed that good precision with standard deviation and relative standard deviation for Hg2+ are 2.30 and 1.53%, respectively (n = 12). This method has short-time analysis with a sample throughput of 20 samples per hour. The limit of detection, that is, the ratio of 3 times the standard deviation of the background and the slope of the linear function (3 s/S), was calculated to be as low as 0.1 mg L–1, which is lower than the permitted limit of mercury in skincare products in Thailand.4 In addition, the U.S. FDA regulations also say about prohibited and restricted ingredients in cosmetics, which are limited to 65 mg Hg L–1 for eye area products and 1 mg Hg L–1 for all other cosmetics. Its presence is unavoidable under conditions of good manufacturing practice (21 CFR 700.13).57 It should be noted that the analytical features of our proposed method are sufficient for determination of mercury in skincare products. Table 2 shows the figures of merit for some of the previously reported methods and our method. This method has easy synthesis, and it uses untreated CDs without the need of any modifications. In addition, the measuring procedure is controlled automatically by a computer.

Figure 8.

Figure 8

Signal profile (A) and calibration curve (B) obtained from the proposed SIA operated under the optimum conditions. Error bars are derived from three injections.

Table 2. Figures of Merit for Some of the Previous Reports Utilizing CDs as a Reagent for Mercury Analysis.

3.6.

3.7. Sample Analysis and Method Validation

The applicability of the proposed method was evaluated by analyzing skincare products (sampled from Nakhon Pathom, Thailand) as real samples. First, the original samples were analyzed using the proposed SIA method and the reference ICP-MS. Notably, no Hg2+ was observed in all samples. Although Hg2+ was not found in the tested skincare samples, the result agreed well with that obtained from ICP-MS (with a detection limit of Hg2+ as low as 0.001 mg L–1). The matrix effect of skincare was investigated by using spiked samples of 1 mg L–1 Hg2+. For SIA, the sample solutions were directly injected into the system. For ICP-MS, a dilution factor of 100 was applied and then calculated backwards. Recoveries of each method are calculated and reported in Table 3. By comparison of %recoveries obtained from the SIA and ICP-MS, although %relative errors of some samples were carried out at ±5%,58 their %recoveries are acceptable according to the AOAC method performance requirements for heavy metal analysis, which are in the range of 80–115.59 The accuracy of our method was further evaluated by triplicate measurements of reference solution Hg2+ (Agilent part number 8500-6940-HG), certified as 1.0 mg L–1.60 The analysis of this sample using the proposed system showed that the fluorescence intensity measurement Hg2+ was 1.01 ± 0.02 mg L–1. This result indicates the potential application of this procedure in monitoring hazardous Hg2+.

Table 3. Analysis Results of the Detection of Hg2+ in Real Samples from Triplicate Analysis Comparison with ICP-MS and Recovery Results of the Actual Samples (ND = Not Detectable).

  Hg2+ (mg L–1)
 
standard method (ICP-MS) proposed method
 
samplea original %recoveryb,c original %recoveryb,c %relative errord
1 ND 101.10 ND 87.1 ± 2.4 –13.8
2 ND 100.83 ND 82.4 ± 3.9 –18.3
3 ND 90.77 ND 86.6 ± 1.0 –4.6
4 ND 98.66 ND 81.8 ± 6.6 –17.1
5 ND 109.56 ND 114.0 ± 5.7 4.05
6 ND 105.42 ND 104.2 ± 3.8 –1.16
7 ND 93.94 ND 85.2 ± 2.7 –9.30
8 ND 100.60 ND 88.7 ± 2.9 –11.8
9 ND 105.94 ND 92.6 ± 3.4 –12.6
a

Sample nos. 1–5 were the whitening face serum, no. 6 was the soothing gel, and nos. 7–9 were serum lotion.

b

%recovery = [(concentration of spiked standard mercury solution – concentration of the sample)/concentration of standard mercury solution] × 100.

c

The concentration of spiked standard mercury solution for ICP-MS and the proposed method is 1 mg L–1.

d

%relative error = [(%recoveryproposed method – %recoverystd method)/%recoverystd method] × 100.

4. Conclusions

Photoluminescent CDs have shown great application in potential health and medical fields. Here, we have utilized the microwave-assisted synthesis of untreated CDs as a specific reagent in the SIA system for the detection of heavy metal ions, which is mercury. In this work, although our CDs have a size of 14.7 ± 4.8 nm in average, the CDs exhibited good optical properties investigated corresponding to the “dots”.61 The CDs showed potential as mercury ion sensors with a detection limit of 0.1 mg L–1. There are several advantages, such as a wider linear range and easy synthesis without the need for later purification and modification steps. No significant difference was observed between the results from our method and the results from ICP-MS with an acceptable percentage recovery of 81.8–114. This method provides less recovery; however, the feature of cost-effectiveness and convenience is better compared to the standard ICP-MS. The SIA application based on CDs was developed for the first time, providing a simple, automatic, rapid, and low-cost analysis platform for the detection of mercury ions in skincare products and heavy metal ion contamination. Our method is environmentally friendly because the detection does not rely on any toxic chemical reaction. Therefore, this system has received considerable attention from scientists working with quality control or working for agencies of safety and inspection service. Furthermore, the SIA system could be further developed as a portable device for on-site analysis.

Acknowledgments

Financial support from the Reinventing University System Program by the Ministry of Higher Education, Science, Research, and Innovation is gratefully acknowledged (fiscal year 2021). The authors wish to express their gratitude to the Faculty of Science, Silpakorn University, for the support of this work.

The authors declare no competing financial interest.

References

  1. Bourgeois M.; Dooms-Goossens A.; Knockaert D.; Sprengers D.; Van Boven M.; Van Tittelboom T. Mercury intoxication after topical appilcation of a metallic mercury ointment. Dermatologica 2004, 172, 48–51. 10.1159/000249292. [DOI] [PubMed] [Google Scholar]
  2. Dyall-Smith D. J.; Scurry J. P. Mercury pigmentation and high mercury levels from the use of a cosmetic cream. Med. J. Aust. 1990, 153, 409–415. 10.5694/j.1326-5377.1990.tb125501.x. [DOI] [PubMed] [Google Scholar]
  3. Chan T. Y. K. Inorganic mercury poisoning associated with skin-lightening cosmetic products. Clin. Toxicol. 2011, 49, 886–891. 10.3109/15563650.2011.626425. [DOI] [PubMed] [Google Scholar]
  4. Ministry of Public Health , Ed., List of Prohibited in Cosmetic Production.; Ministry of Public Health: Bangkok: 2559. [Google Scholar]
  5. EARTH, E. A. a. R. T . Study of mercury contamination in face whitening products in Thailand; EARTH, E. A. a. R. T.: Thailand, 2021. DOI: https://ipen.org/sites/default/files/documents/EARTH%20Hg%20in%20Whitening%20-%20Report.pdf. [Google Scholar]
  6. Gaál F. F.; Abramović B. F. Determination of mercury content of some pharmaceutical products by catalytic titration. Microchim. Acta 1972, 1, 465–472. 10.1007/BF01197126. [DOI] [Google Scholar]
  7. Margosis M.; Tanner J. T. Determination of mercury in pharmaceutical products by neutron activation analysis. J. Pharm. Sci. 1972, 61, 936–938. 10.1002/jps.2600610625. [DOI] [PubMed] [Google Scholar]
  8. Mallkuci I.; Lazo P. MERCURY DETERMINATION IN DRUG AND COSMETIC PRODUCTS. Int. J. of Curr. Res. 2014, 6, 8077–8082. [Google Scholar]
  9. Passariello B.; Barbaro M.; Quaresima S.; Casciello A.; Marabini A. Determination of mercury by Inductively Coupled Plasma-Mass Spectrometry. Microchem. J. 1996, 54, 348–354. 10.1006/mchj.1996.0110. [DOI] [PubMed] [Google Scholar]
  10. Allibone J.; Fatemian E.; Walker P. J. Determination of mercury in potable water by ICP-MS using gold as a stabilising agent. J. Anal. At. Spectrom. 1999, 14, 235–239. 10.1039/A806193I. [DOI] [Google Scholar]
  11. Kraithong S.; Sangsuwan R.; Worawannotai N.; Sirirak J.; Charoenpanich A.; Thamyongkit P.; Wanichachewa N. Triple detection modes for Hg2+ sensing based on a NBD-fluorescent and colorimetric sensor and its potential in cell imaging. New J. Chem. 2018, 42, 12412–12420. 10.1039/c8nj01915k. [DOI] [Google Scholar]
  12. Petdum A.; Panchan W.; Sirirak J.; Promarak V.; Sooksimuang T.; Wanichacheva N. Colorimetric and fluorescent sensing of a new FRET system via [5]helicene and rhodamine 6G for Hg2+ detection. New J. Chem. 2018, 42, 1396–1402. 10.1039/c7nj04129b. [DOI] [Google Scholar]
  13. Rasheed T.; Nabeel F.; Li C.; Bilal M. Rhodamine-assisted fluorescent strategy for the sensitive and selective in-field mapping of environmental pollutant Hg(II) with potential bioimaging. J. Lumin. 2019, 208, 519–526. 10.1016/j.jlumin.2019.01.032. [DOI] [Google Scholar]
  14. Kaewnok N.; Sirirak J.; Jungsuttiwong S.; Wongnongwa Y.; Kamkaew A.; Petdum A.; Panchan W.; Sahasithiwat S.; Sooksimuang T.; Charoenpanich A.; Wanichacheva N. Detection of hazardous mercury ion using [5]helicene-based fluorescence probe with ″Turn ON″ sensing response for practical applications. J. Hazard. Mater. 2021, 418, 126242. 10.1016/j.jhazmat.2021.126242. [DOI] [PubMed] [Google Scholar]
  15. Bauzá De Mirabó F. M.; Thomas A. C.; Rubí E.; Forteza E.; Cerdá V. Sequential injection analysis system for determination of mercury by cold-vapor atomic absorption spectroscopy. Anal. Chim. Acta 1997, 355, 203–210. 10.1016/S0003-2670(97)00397-8. [DOI] [Google Scholar]
  16. Punrat E.; Chuanuwatanakul S.; Kaneta T.; Motomizu S.; Chailapakul O. Method development for the determination of mercury(II) by sequential injection/anodic stripping voltammetry using an in situ gold-film screen-printed carbon electrode. J. Electroanal. Chem. 2014, 727, 78–83. 10.1016/j.jelechem.2014.05.026. [DOI] [Google Scholar]
  17. Rattanakit P.; Prasertboonyai K.; Liawruangrath S. Development of sequential injection spectrophotometric method for determination of mercury (II) using pyrogallol red. Int. J. Environ. Anal. Chem. 2016, 96, 1415–1429. 10.1080/03067319.2016.1265955. [DOI] [Google Scholar]
  18. Du J.; Zhu B.; Peng X.; Chen X. Optical reading of contaminants in aqueous media based on gold nanoparticles. Small 2014, 10, 3461–3479. 10.1002/smll.201303256. [DOI] [PubMed] [Google Scholar]
  19. Zangeneh Kamali K.; Pandikumar A.; Jayabal S.; Ramaraj R.; Lim H. N.; Ong B. H.; Bien C. S. D.; Kee Y. Y.; Huang N. M. Amalgamation based optical and colorimetric sensing of mercury(II) ions with silver@graphene oxide nanocomposite materials. Microchim. Acta 2016, 183, 369–377. 10.1007/s00604-015-1658-6. [DOI] [Google Scholar]
  20. Lee J.-S.; Han M. S.; Mirkin C. A. Colorimetric Detection of Mercuric Ion (Hg2+) in Aqueous Media using DNA-Functionalized Gold Nanoparticles. Am. Ethnol. 2007, 119, 4171–4174. 10.1002/ange.200700269. [DOI] [PubMed] [Google Scholar]
  21. Baker S. N.; Baker G. A. Luminescent carbon nanodots: emergent nanolights. Angew. Chem., Int. Ed. 2010, 49, 6726–6744. 10.1002/anie.200906623. [DOI] [PubMed] [Google Scholar]
  22. Berlina A. N.; Zherdev A. V.; Dzantiev B. B. Progress in rapid optical assays for heavy metal ions based on the use of nanoparticles and receptor molecules. Microchim. Acta 2019, 186, 172. 10.1007/s00604-018-3168-9. [DOI] [PubMed] [Google Scholar]
  23. Yoo D.; Park Y.; Cheon B.; Park M. H. Carbon Dots as an Effective Fluorescent Sensing Platform for Metal Ion Detection. Nanoscale Res. Lett. 2019, 14, 272. 10.1186/s11671-019-3088-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Li P.; Li S. F. Y. Recent advances in fluorescence probes based on carbon dots for sensing and speciation of heavy metals. NANO 2021, 10, 877–908. 10.1515/nanoph-2020-0507. [DOI] [Google Scholar]
  25. El-Shafey A. M. Carbon dots: Discovery, structure, fluorescent properties, and applications. Green Process. Synth. 2021, 10, 134–156. 10.1515/gps-2021-0006. [DOI] [Google Scholar]
  26. Sekar A.; Yadav R.; Basavaraj N. Fluorescence quenching mechanism and the application of green carbon nanodots in the detection of heavy metal ions: a review. New J. Chem. 2021, 45, 2326–2360. 10.1039/d0nj04878j. [DOI] [Google Scholar]
  27. Nazri N. A. A.; Azeman N. H.; Luo Y.; Bakar A. A. A. Carbon quantum dots for optical sensor applications: A review. Opt. Laser Technol. 2021, 139, 106928. 10.1016/j.optlastec.2021.106928. [DOI] [Google Scholar]
  28. Ghanem A.; Al-Qassar Bani Al-Marjeh R.; Atassi Y. Novel nitrogen-doped carbon dots prepared under microwave-irradiation for highly sensitive detection of mercury ions. Heliyon 2020, 6, e03750 10.1016/j.heliyon.2020.e03750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Ye Z.; Zhang Y.; Li G.; Li B. Fluorescent Determination of Mercury(II) by Green Carbon Quantum Dots Synthesized from Eggshell Membrane. Anal. Lett. 2020, 53, 2841–2853. 10.1080/00032719.2020.1759618. [DOI] [Google Scholar]
  30. Omer K. M.; Hama Aziz K. H.; Mohammed S. J. Improvement of selectivity via the surface modification of carbon nanodots towards the quantitative detection of mercury ions. New J. Chem. 2019, 43, 12979–12986. 10.1039/c9nj03057c. [DOI] [Google Scholar]
  31. Yahyazadeh E.; Shemirani F. Easily synthesized carbon dots for determination of mercury(II) in water samples. Heliyon 2019, 5, e01596 10.1016/j.heliyon.2019.e01596. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Mohammadpour Z.; Safavi A.; Shamsipur M. A new label free colorimetric chemosensor for detection of mercury ion with tunable dynamic range using carbon nanodots as enzyme mimics. Chem. Eng. J. 2014, 255, 1–7. 10.1016/j.cej.2014.06.012. [DOI] [Google Scholar]
  33. Pajewska-Szmyt M.; Buszewski B.; Gadzala-Kopciuch R. Carbon dots as rapid assays for detection of mercury(II) ions based on turn-off mode and breast milk. Spectrochim. Acta, Part A 2020, 236, 118320. 10.1016/j.saa.2020.118320. [DOI] [PubMed] [Google Scholar]
  34. Tadesse A.; Hagos M.; RamaDevi D.; Basavaiah K.; Belachew N. Fluorescent-Nitrogen-Doped Carbon Quantum Dots Derived from Citrus Lemon Juice: Green Synthesis, Mercury(II) Ion Sensing, and Live Cell Imaging. ACS Omega 2020, 5, 3889–3898. 10.1021/acsomega.9b03175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Zhang Q.; Zhang X.; Bao L.; Wu Y.; Jiang L.; Zheng Y.; Wang Y.; Chen Y. The Application of Green-Synthesis-Derived Carbon Quantum Dots to Bioimaging and the Analysis of Mercury(II). J. Anal. Methods Chem. 2019, 2019, 8183134. 10.1155/2019/8183134. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Qu S.; Wang X.; Lu Q.; Liu X.; Wang L. A biocompatible fluorescent ink based on water-soluble luminescent carbon nanodots. Angew. Chem., Int. Ed. 2012, 51, 12215–12218. 10.1002/anie.201206791. [DOI] [PubMed] [Google Scholar]
  37. Qin X.; Lu W.; Asiri A. M.; Al-Youbi A. O.; Sun X. Microwave-assisted rapid green synthesis of photoluminescent carbon nanodots from flour and their applications for sensitive and selective detection of mercury(II) ions. Sens. Actuators, B 2013, 184, 156–162. 10.1016/j.snb.2013.04.079. [DOI] [Google Scholar]
  38. Sk M. P.; Chattopadhyay A. Induction coil heater prepared highly fluorescent carbon dots as invisible ink and explosive sensor. RSC Adv. 2014, 4, 31994–31999. 10.1039/c4ra04264f. [DOI] [Google Scholar]
  39. Han Z.; He L.; Pan S.; Liu H.; Hu X. Hydrothermal synthesis of carbon dots and their application for detection of chlorogenic acid. Luminescence 2020, 35, 989–997. 10.1002/bio.3803. [DOI] [PubMed] [Google Scholar]
  40. Bajpai S. K.; D’Souza A.; Suhail B. Blue light-emitting carbon dots (CDs) from a milk protein and their interaction with Spinacia oleracea leaf cells. Int. Nano Lett. 2019, 9, 203–212. 10.1007/s40089-019-0271-9. [DOI] [Google Scholar]
  41. Fu Y.; Zhao S.; Wu S.; Huang L.; Xu T.; Xing X.; Lan M.; Song X. A carbon dots-based fluorescent probe for turn-on sensing of ampicillin. Dyes Pigm. 2020, 172, 107846. 10.1016/j.dyepig.2019.107846. [DOI] [Google Scholar]
  42. Li X.; Zhang S.; Kulinich S. A.; Liu Y.; Zeng H. Engineering surface states of carbon dots to achieve controllable luminescence for solid-luminescent composites and sensitive Be2+ detection. Sci. Rep. 2014, 4, 4976. 10.1038/srep04976. [DOI] [Google Scholar]
  43. Hoan B. T.; Van Huan P.; Van H. N.; Nguyen D. H.; Tam P. D.; Nguyen K. T.; Pham V. H. Luminescence of lemon-derived carbon quantum dot and its potential application in luminescent probe for detection of Mo(6+) ions. Luminescence 2018, 33, 545–551. 10.1002/bio.3444. [DOI] [PubMed] [Google Scholar]
  44. Wang C.; Hu T.; Wen Z.; Zhou J.; Wang X.; Wu Q.; Wang C. Concentration-dependent color tunability of nitrogen-doped carbon dots and their application for iron(III) detection and multicolor bioimaging. J. Colloid Interface Sci. 2018, 521, 33–41. 10.1016/j.jcis.2018.03.021. [DOI] [PubMed] [Google Scholar]
  45. Liu W.; Diao H.; Chang H.; Wang H.; Li T.; Wei W. Green synthesis of carbon dots from rose-heart radish and application for Fe3+ detection and cell imaging. Sens. Actuators, B 2017, 241, 190–198. 10.1016/j.snb.2016.10.068. [DOI] [Google Scholar]
  46. Askarinejad A.; Morsali A. Synthesis and characterization of mercury oxide unusual nanostructures by ultrasonic method. Chem. Eng. J. 2009, 153, 183–186. 10.1016/j.cej.2009.05.031. [DOI] [Google Scholar]
  47. Tang Q.; Zhu W.; He B.; Yang P. Rapid Conversion from Carbohydrates to Large-Scale Carbon Quantum Dots for All-Weather Solar Cells. ACS Nano 2017, 11, 1540–1547. 10.1021/acsnano.6b06867. [DOI] [PubMed] [Google Scholar]
  48. Wei X.-M.; Xu Y.; Li Y.-H.; Yin X.-B.; He X.-W. Ultrafast synthesis of nitrogen-doped carbon dots via neutralization heat for bioimaging and sensing applications. RSC Adv. 2014, 4, 44504–44508. 10.1039/c4ra08523j. [DOI] [Google Scholar]
  49. Zhao C.; Li X.; Cheng C.; Yang Y. Green and microwave-assisted synthesis of carbon dots and application for visual detection of cobalt(II) ions and pH sensing. Microchem. J. 2019, 147, 183–190. 10.1016/j.microc.2019.03.029. [DOI] [Google Scholar]
  50. Edison T. N.; Atchudan R.; Sethuraman M. G.; Shim J. J.; Lee Y. R. Microwave assisted green synthesis of fluorescent N-doped carbon dots: Cytotoxicity and bio-imaging applications. J. Photochem. Photobiol., B 2016, 161, 154–161. 10.1016/j.jphotobiol.2016.05.017. [DOI] [PubMed] [Google Scholar]
  51. Barati A.; Shamsipur M.; Arkan E.; Hosseinzadeh L.; Abdollahi H. Synthesis of biocompatible and highly photoluminescent nitrogen doped carbon dots from lime: analytical applications and optimization using response surface methodology. Mater. Sci. Eng., C 2015, 47, 325–332. 10.1016/j.msec.2014.11.035. [DOI] [PubMed] [Google Scholar]
  52. Kuzmenko V.; Naboka O.; Staaf H.; Haque M.; Göransson G.; Lundgren P.; Gatenholm P.; Enoksson P. Capacitive effects of nitrogen doping on cellulose-derived carbon nanofibers. Mater. Chem. Phys. 2015, 160, 59–65. 10.1016/j.matchemphys.2015.04.006. [DOI] [Google Scholar]
  53. Shah H.; Xin Q.; Jia X.; Gong J. R. Single precursor-based luminescent nitrogen-doped carbon dots and their application for iron (III) sensing. Arabian J. Chem. 2019, 12, 1083–1091. 10.1016/j.arabjc.2019.06.004. [DOI] [Google Scholar]
  54. He J. H.; Cheng Y. Y.; Yang T.; Zou H. Y.; Huang C. Z. Functional preserving carbon dots-based fluorescent probe for mercury (II) ions sensing in herbal medicines via coordination and electron transfer. Anal. Chim. Acta 2018, 1035, 203–210. 10.1016/j.aca.2018.06.053. [DOI] [PubMed] [Google Scholar]
  55. Yan F.; Kong D.; Luo Y.; Ye Q.; He J.; Guo X.; Chen L. Carbon dots serve as an effective probe for the quantitative determination and for intracellular imaging of mercury(II). Microchim. Acta 2016, 183, 1611–1618. 10.1007/s00604-016-1788-5. [DOI] [Google Scholar]
  56. Li L.; Yu B.; You T. Nitrogen and sulfur co-doped carbon dots for highly selective and sensitive detection of Hg (II) ions. Biosens. Bioelectron. 2015, 74, 263–269. 10.1016/j.bios.2015.06.050. [DOI] [PubMed] [Google Scholar]
  57. https://www.fda.gov/cosmetics/cosmetics-laws-regulations/prohibited-restricted-ingredients-cosmetics
  58. Codex Alimentarius Commission . CAC/GL 40–1993 Guidelines on Good Laboratory Practice in Pesticide Residue Analysis; Codex Alimentarius Commission: Rome, Italy: 2010. [Google Scholar]
  59. AOAC International . AOAC SMPR 2012.007 Standard Method Performance Requirements for Heavy Metals in a Variety of Foods and Beverages; AOAC International: Rockville, USA: 2021. [DOI] [PubMed] [Google Scholar]
  60. https://www.agilent.com/store/en_US/Prod-8500-6940/8500-6940
  61. Liu J.; Li R.; Yang B. Carbon Dots: A New Type of Carbon-Based Nanomaterial with Wide Applications. ACS Cent. Sci. 2020, 6, 2179–2195. 10.1021/acscentsci.0c01306. [DOI] [PMC free article] [PubMed] [Google Scholar]