Abstract
We study the potential of the commercial mounting medium Slowfade diamond as a buffer for STORM microscopy. We show that although it does not work with the popular far-red dyes typically used for STORM imaging, such as Alexa Fluor 647, it performs really well with a wide variety of green-excited dyes such as Alexa Fluor 532, Alexa Fluor 555 or CF 568. Moreover, imaging can be performed several months after the samples are mounted in this environment and kept in the fridge, providing a convenient way to preserve samples for STORM imaging, as well as to keep calibration samples, for example for metrology or teaching in particular in imaging facilities.
1. Introduction
Single Molecule Localization Microscopy (SMLM) [1–5] is one of the most popular family of optical super-resolution methods since it requires a simple inverted microscope to achieve a gain in resolution of an order of magnitude and therefore enable the study of biological processes hidden behind the diffraction limit [6]. However, it requires samples to be prepared in a particular way, so as to enable stochastic switching between a fluorescent and a non-fluorescent state for the fluorophore imaged. In STORM microscopy this is achieved using standard dyes in a special chemical buffer that allows efficient transitions between a fluorescent an a nonfluorescent state [7–9]. Most STORM imaging is performed in a buffer fairly similar to the one used in the original demonstration [1,7], which combines a reducing thiol such as 2-Mercaptoethylamine/Cysteamine (MEA) or -Mercaptoethanol ( ME) and an enzymatic oxygen scavenging system based on the combination of glucose, glucose-oxydase and catalase. However, since this oxygen scavenging system acidifies with time [10,11], several alternatives have been proposed to alleviate this issue [12], including the use of other enzymatic systems such as PCA/PCD [13,14] and pyranose oxidase [14,15]. The blinking mechanism of dyes in this type of bufferis well studied [16–18] and several screens have been performed to identify the best fluorophores [19,20]. However, there have also been a few reports of different types of buffer aimed at simplifying the imaging protocols. In particular, Vectashield [21–23] and Prolong Diamond [24] are two commercial mounting media that have been found to work well for STORM imaging for different subsets of dye. Interestingly, while Vectashield worked best for the far-red dyes typically used in STORM such as Cy5 and Alexa Fluor 647, Prolong Diamond was reported to work best with Alexa Fluor 594 [24]. This ability to image red dyes motivated us in evaluating the properties of the related non-hardening mounting media Slowfade Diamond as a STORM buffer.
2. Methods
Optical setup and data acquisition
Dye Screening Dye screening was performed on an IX83 Inverted microscope (Olympus) using a 100x 1.3 NA objective (Olympus), and an Orca Fusion sCMOS camera (Hamamatsu). 3 free-space single mode lasers at 488 nm, 532 nm and 640 nm (Voltran, 100 mW, 40 mW and 100 mW) were used, as well as a 750 nm multimode laser (1.2 W Oxxius). The microscope was equipped with Chroma filters: 532 nm and 640 nm-excited fluorophores were imaged using a ZT532/640rpc 2-color dichroic mirror, and a ET605/70 or ET700-75 emission filter. Due to the double-deck design, where the top deck is equipped with only a dichroic mirror and the lower deck with full sets (Emission filter, dichroic mirror and excitation filter) the emitted light also goes through a T550lpxr dichroic (red fluorophores) or a T660lpxr dichroic (far-red fluorophores). 750 nm-excited fluorophores were imaged using a T760lpxr dichroic, and a ET810-90 emission filter. Similarly, the emitted light also goes through an additional T760lpxr dichroic. Green dyes excited at 488 nm were imaged using a GFP filter cube (49002 Chroma). For the dye screening 10,000-15,000 images were acquired with ms integration time using the slowest "ultra quiet scan" read-out speed of the camera and z-drift was corrected manually. The camera and microscope were controlled using micro-manager [25]. Laser power at 488 nm, 532 nm, 640 nm and 750 nm was typically between 1 and 4 kW. .
Further imaging Further imaging of the best dyes identified was performed on a Nikon Ti:E microscope with an Apo TIRF 100X 1.49 Oil objective (Nikon) on an EMCCD camera (Quant EM:512SC, Photometrics), using the microscope’s 1.5x extra magnification and resulting in a pixel size of 107 nm. The illumination came from a fiber-coupled laser combiner (L6CC Oxxius) equipped with a 640 nm 300 mW laser (LPX-640L-300-CSB-OE-900, Oxxius) and a 532 nm 500 mW laser (LPX-532L-500-CSB-OE-900, Oxxius). The filters used were a 4 band dichroic (Di03-R405/488/532/635-t1-25x36, Semrock) and the associated 4-band emission filter (FF01-446/510/581/703-25). This microscope was controlled using NIS, with camera exposure times of 30 ms and an EM gain of 1000. For most datasets, 20,000 frames were recorded.
Sample preparation and immunofluorescence staining
Dye Screening African green monkey kidney cells (COS-7) were cultured in DMEM-Glutamax (Gibco 10566016) supplemented with 10% FBS in a cell culture incubator ( C and 5% ). Cells were plated at low confluency on ethanol-cleaned 25 mm #1.5 thickness round coverglass (VWR) for imaging. Prior to fixation, all solutions were pre-warmed to 37 C. 24h after plating, cells were pre-extracted for 30 s in 0.25% Triton X-100 (Sigma-Aldrich) in PHEM (60 mM PIPES, 25 mM HEPES, 10 mM EGTA, 4 mM MgSO4) washed in PHEM, fixed for 8 min in −20 C Methanol (Sigma-Aldrich), then washed 3 times with PBS. The samples were then blocked for 1 hour in 5% BSA, before being incubated for 1.5 hour at room temperature with 1:500 Mouse anti alpha-tubulin (Sigma T6199) in 1% BSA diluted in PBS-0.2% Triton (BSA-PBST), followed by 3 washes in PBST, and then incubated for 1 hour in BSA-PBST with a secondary antibodies (see list in Table S1) and finally washed 3 times in PBST.
The sample were imaged in an Attofluor imaging chamber (Invitrogen, A7816), with of Slowfade Diamond (Invitrogen S36963) and another 18 mm round coverglass on top to limit air exchanges.
Further imaging Human hCMEC/D3, hDBEC, and HDFa cells were cultured as in [26,27]. Briefly, Human hCMEC/D3 cells were grown in EBM-2 medium (Lonza) supplemented with 5% fetal calf serum 1.4 M hydrocortisone (Lonza), 5 g/mL ascorbic acid (Lonza), and 1 ng/mL b-FGF (Lonza) and grown on rat tail collagen-1 (BD Biosciences) coated flasks while hDBEC cells grown on rat tail collagen-1 (BD Biosciences) coated flasks in Endothelial Cell Growth Medium MV2 (Promocell) supplemented with Endothelial Cell Growth Medium Supplement Mix (Promocell). Cells were plated at low confluency on ethanol-cleaned 22x22mm #1.5 thickness square coverglass also coated with rat tail collagen-1. 24 hours after plating, the cells were rinsed in PHEM, then fixed in 4% PFA + Glutaraldehyde 0,2% in PHEM with 0,5% Triton for 12 min, then rinsed once in PBS 0.1% Triton and 3 times in PBS. The samples were then blocked for 1 hour in 3% BSA (Sigma-Aldrich A4503) at room temperature under gentle agitation, before being incubated for 1 hour at C with 1:500 mouse anti-alpha-tubulin antibodies (Sigma T9026) in 3% BSA, followed by 3 washes with 3% BSA, and then incubated at C for 1 hour in 3% BSA with secondary antibodies (see list in Table S1) followed by 3 washes in PBS. The samples were then post-fixed using 4% Formaldehyde (Sigma 8.18708) during 5 minutes, rinsed 3 times in PBS, quenched in NH4Cl 50 mM for 5 minutes and finally washed 3 times in PBS.
The samples were then mounted on a microscope slide with a drop of Slowfade Diamond ( ) and sealed with twinsil (Picodent). After imaging, the slide was kept in the fridge in the dark.
Data processing
Raw STORM image stacks were processed using using a FIJI macro that runs an analysis with Detection of Molecules (DoM) [28] and Thunderstorm [29], including drift-correction and grouping of consecutive localizations [30]. Further quantification of the blinking quality (Localization precision and number of detections per frame)were performed with UNLOC [31] (see Figures S1-S8). FRC resolution was calculated using a pixel size of 10 nm using the BIOP FIJI plugin [32] with DoM localizations exported in Thunderstorm format to create two images consisting of the odd and even localizations modified from [33,34].
3. Results
3.1. Alexa Fluor 555
We first tested Alexa Fluor 555 (Fig. 1(a)), and observed nice prolonged blinking with individual camera frames showing isolated fluorophores when continuously excited with a 532-nm laser with intensities higher than 1 (Fig. 1(b), the raw data is available on zenodo [35]). We imaged the sample continuously to record 20.000 frames , and could reconstruct a high-resolution STORM image (Fig. 1(c-d)). We noticed low photobleaching, and a good number of molecules per frame was still observed after 20,000 frames (See Figure S1). In order to have a quantitative measure of our resolution, we performed a FRC [36] calculation on a 20x20 region which yielded a value of 20.5 nm (Fig. 1(e)). Indeed, a profile on the two neighboring microtubules shows that the hollowness of the structure can be resolved (Fig. 1(f)) consistently with the FRC resolution.
3.2. Dye screening
Since Slowfade Diamond is a great buffer for Alexa Fluor 555, we were excited about its potential for multicolor imaging. We first tested a wide array of far-red dyes, as they are the most widely used in STORM, but achieved very poor results with both 640 nm excited dyes such as Alexa Fluor 647 and 750 nm excited dyes such as CF 750. Indeed, they all behaved so poorly (see Table 1) that it was impossible to generate a recognizable image of the structure of interest with any of them.
Table 1. Image quality scale: (-) cannot reconstruct an image. (+) can reconstruct an image at least as good as a widefield image (++) can reconstruct a good quality STORM image [see Figure S2-8 for further quantification] (+++) Best performing dye in Slowfade Diamond. (*) Image quality/FRC after background subtractions.
532 nm-excited dyes: | ||
---|---|---|
Fluorophore | image quality | FRC (nm) |
Alexa Fluor 532 | 29 | |
CF 532 | 28 | |
Alexa Fluor 546 | - | |
Dylight 549 | 21 | |
Atto 550 | 25 | |
Alexa Fluor 555 | 20 | |
CF 555 | - | |
Cy3 | - | |
CF 568 | 20 | |
Alexa Fluor 568 | - | |
Alexa Fluor 594 | 24 | |
Dylight 594 | 25 |
Other dyes: | ||
---|---|---|
Fluorophore | image quality | FRC (nm) |
Alexa Fluor 488 | - | |
CF 488 | - | |
Dylight 488 | (*) | 35(*) |
Alexa Fluor 514 | - | |
Alexa Fluor 647 | - | |
CF 647 | - | |
Dylight 649 | - | |
CF 660C | - | |
CF 680 | - | |
Alexa Fluor 700 | - | |
CF 750 | - | |
Dylight 755 | - |
Undeterred, we moved on to test dyes excited at 488 nm such as Alexa Fluor 488. All 4 dyes we tested (see Table 1) allowed us to at least reconstruct an image of the structure of interest, with Dylight-488 behaving best in that wavelength range. However, the presence of a non-negligible background (partially due to an un-optimized microscope configuration , but also due to the presence of a background signal made particularly significant by our use of epi-illumination that increases with time in our samples, see raw data [35]) required slightly more advanced data processing: using a temporal median filter [37,38] to remove the slowly varying background we could reconstruct a fairly decent image as can be seen in Fig. 2.
We then considered another option: since Slowfade Diamond works so well for Alexa Fluor 555 maybe it also works with other dyes in that wavelength range that could be used to do multicolor imaging using spectral unmixing [39–41] as is typically done in the usual STORM buffer with Alexa Fluor 647 and CF-680 [20]. We therefore tested 12 different dyes that could be excited at 532 nm, and identified 8 that worked well (see Table 1 and Figure S2-8). Out of these 8 dyes, we chose 3 that have the most distinct emission spectra: Alexa Fluor 532, CF-568 and Alexa Fluor 594 (see Fig. 3(a), spectra taken from FPbase [42]). Images taken with these dyes are shown in Fig. 3(b-d). Although dimmer than Alexa Fluor 555, they all provide high quality STORM images (the raw data is available on zenodo [35]). With the recent advent of publicly available spectral-unmixing software [41,43] we expect this combination to be the most promising for multicolor imaging with Slowfade Diamond.
3.3. Slowfade Diamond as a long-term STORM buffer for Alexa Fluor 555
Since Slowfade Diamond appears to be a suitable buffer for STORM imaging of Alexa Fluor 555, and can be used to preserve mounted samples, we wondered how long samples mounted in Slowfade Diamond could be imaged. We therefore kept a mounted sample in the fridge after imaging, and tried to image it again after 150 and 220 days. As can be seen in Fig. 4, not only could we still image the sample after several months, but the image quality is fairly stable as a function of time. Indeed, performing FRC calculations on the images of the old sample yielded values of 23 nm and 22 nm, compared to 20.5 nm on the fresh sample. We did notice that after 220 days it took longer to reach single molecule regime, and that fairly high laser powers were necessary to obtain a good image, so we expect that 1 year might be a maximum for sample preservation in the fridge. We checked that this was also true after 60 days for the other 532 nm-excited fluorophores rated (++) in Table 1.
4. Conclusion and perspectives
We tested 24 different fluorophores in Slowfade Diamond and found that a surprisingly large number of 532 nm excited dyes blinked very well, as well as to a lesser degree all the 488 nm excited dyes we tried. Interestingly, samples mounted in Slowfade and kept in the fridges can be imaged for several month which should be very useful for microscope calibration and comparisons, as well as for teaching (The bright signal of the single red fluorophores can be readily observed by eye on the eyepiece). The composition of Slowfade Diamond being proprietary information, it is hard to draw conclusion about the photophysics involved [17,18], but it is interesting to notice the high correlation between dyes that work and the excitation wavelength used, and the fact that the wavelength range that works well is very different to that of the usual buffer [19,20] or of Vectashield [21]. The index of refraction of Slowfade Diamond is 1.42, so TIRF and grazing incidence imaging are both possible (though they require the use of high NA objective), as well as the use of an active z-stabilization but the higher index compared to that of most water-based buffer limits depth-induced aberrations, which is especially relevant for 3D imaging. As was the case with Vectashield, one potential advantage of using a common mounting medium as Slowfade Diamond would be to combine SIM [44,45] and STORM for multicolor imaging, for example with 405-nm and 488-nm excited fluorophores used for SIM and one or two 532 nm excited dye used for STORM. Slowfade Diamond has also been used for STED microscopy [46,47] (see Figure S9) but since it is a scanning technique, straightforward combinations seem difficult. However, comparisons could be performed on the same sample between SIM, STED and STORM, for example for metrology or to determine which method is better adapted to a particular sample. Finally, since Vectashield works for far-red fluorophores, an intriguing possibility would be to mix the two anti-fading agents to achieve multi-color imaging.
Acknowledgments
We thank Sandrine Lévêque-Fort for the Cos-7 cells, Maxime Mauviel for his help with Cos-7 cell culture, and Mathieu Coureuil for his help with hCMEC/D3 cell culture. We also thank Vectorlab for the gift of the Dylight-488, Dylight-549, Dylight-594 and Dylight-649 antibodies, Biotium for the gift of the CF 532, CF 660C, and CF 680 secondary antibodies. We thank Debora Keller-Olivier for critical reading, and all the members of the "nano" group at LOB for scientific and technical discussions. N.O. thanks Christa Walther and the imaging facility at the University of Sheffield for preliminary experiments.
Funding
Institut de physique10.13039/501100017212 (Tremplin@INP 2019, Tremplin@INP 2021).
Disclosures
The authors declare no conflicts of interest
Data availability
Data underlying the results presented in this paper are available on Zenodo [35].
Supplemental document
See Supplement 1 (12.1MB, pdf) for supporting content.
References
- 1.Rust M. J., Bates M., Zhuang X., “Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (storm),” Nat. Methods 3(10), 793–796 (2006). 10.1038/nmeth929 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Betzig E., Patterson G. H., Sougrat R., Lindwasser O. W., Olenych S., Bonifacino J. S., Davidson M. W., Lippincott-Schwartz J., Hess H. F., “Imaging intracellular fluorescent proteins at nanometer resolution,” Science 313(5793), 1642–1645 (2006). 10.1126/science.1127344 [DOI] [PubMed] [Google Scholar]
- 3.Hess S. T., Girirajan T. P., Mason M. D., “Ultra-high resolution imaging by fluorescence photoactivation localization microscopy,” Biophys. J. 91(11), 4258–4272 (2006). 10.1529/biophysj.106.091116 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sharonov A., Hochstrasser R. M., “Wide-field subdiffraction imaging by accumulated binding of diffusing probes,” Proc. Natl. Acad. Sci. 103(50), 18911–18916 (2006). 10.1073/pnas.0609643104 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Jungmann R., Steinhauer C., Scheible M., Kuzyk A., Tinnefeld P., Simmel F. C., “Single-molecule kinetics and super-resolution microscopy by fluorescence imaging of transient binding on dna origami,” Nano Lett. 10(11), 4756–4761 (2010). 10.1021/nl103427w [DOI] [PubMed] [Google Scholar]
- 6.Lelek M., Gyparaki M. T., Beliu G., Schueder F., Griffié J., Manley S., Jungmann R., Sauer M., Lakadamyali M., Zimmer C., “Single-molecule localization microscopy,” Nat. Rev. Methods Primers 1(1), 39 (2021). 10.1038/s43586-021-00038-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bates M., Blosser T. R., Zhuang X., “Short-range spectroscopic ruler based on a single-molecule optical switch,” Phys. Rev. Lett. 94(10), 108101 (2005). 10.1103/PhysRevLett.94.108101 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Heilemann M., Margeat E., Kasper R., Sauer M., Tinnefeld P., “Carbocyanine dyes as efficient reversible single-molecule optical switch,” J. Am. Chem. Soc. 127(11), 3801–3806 (2005). 10.1021/ja044686x [DOI] [PubMed] [Google Scholar]
- 9.Steinhauer C., Forthmann C., Vogelsang J., Tinnefeld P., “Superresolution microscopy on the basis of engineered dark states,” J. Am. Chem. Soc. 130(50), 16840–16841 (2008). 10.1021/ja806590m [DOI] [PubMed] [Google Scholar]
- 10.Benesch R. E., Benesch R., “Enzymatic removal of oxygen for polarography and related methods,” Science 118(3068), 447–448 (1953). 10.1126/science.118.3068.447 [DOI] [PubMed] [Google Scholar]
- 11.Shi X., Lim J., Ha T., “Acidification of the oxygen scavenging system in single-molecule fluorescence studies: in situ sensing with a ratiometric dual-emission probe,” Anal. Chem. 82(14), 6132–6138 (2010). 10.1021/ac1008749 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Hartwich T. M., Chung K. K. H., Schroeder L., Bewersdorf J., Soeller C., Baddeley D., “A stable, high refractive index, switching buffer for super-resolution imaging,” bioRxiv, (2018). 10.1101/465492 [DOI]
- 13.Aitken C. E., Marshall R. A., Puglisi J. D., “An oxygen scavenging system for improvement of dye stability in single-molecule fluorescence experiments,” Biophys. J. 94(5), 1826–1835 (2008). 10.1529/biophysj.107.117689 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Olivier N., Keller D., Gönczy P., Manley S., “Resolution doubling in 3D-storm imaging through improved buffers,” PLoS One 8(7), e69004 (2013). 10.1371/journal.pone.0069004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Swoboda M., Henig J., Cheng H.-M., Brugger D., Haltrich D., Plumeré N., Schlierf M., “Enzymatic oxygen scavenging for photostability without pH drop in single-molecule experiments,” ACS Nano 6(7), 6364–6369 (2012). 10.1021/nn301895c [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Dempsey G. T., Bates M., Kowtoniuk W. E., Liu D. R., Tsien R. Y., Zhuang X., “Photoswitching mechanism of cyanine dyes,” J. Am. Chem. Soc. 131(51), 18192–18193 (2009). 10.1021/ja904588g [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Levitus M., “Photophysics of single-molecule probes,” in Spectroscopy and Dynamics of Single Molecules , (Elsevier, 2019), pp. 15–69. [Google Scholar]
- 18.Ha T., Tinnefeld P., “Photophysics of fluorescent probes for single-molecule biophysics and super-resolution imaging,” Annu. Rev. Phys. Chem. 63(1), 595–617 (2012). 10.1146/annurev-physchem-032210-103340 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Dempsey G. T., Vaughan J. C., Chen K. H., Bates M., Zhuang X., “Evaluation of fluorophores for optimal performance in localization-based super-resolution imaging,” Nat. Methods 8(12), 1027–1036 (2011). 10.1038/nmeth.1768 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Lehmann M., Lichtner G., Klenz H., Schmoranzer J., “Novel organic dyes for multicolor localization-based super-resolution microscopy,” J. Biophotonics 9(1-2), 161–170 (2016). 10.1002/jbio.201500119 [DOI] [PubMed] [Google Scholar]
- 21.Olivier N., Keller D., Rajan V. S., Gönczy P., Manley S., “Simple buffers for 3D storm microscopy,” Biomed. Opt. Express 4(6), 885–899 (2013). 10.1364/BOE.4.000885 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Kwakwa K., Savell A., Davies T., Munro I., Parrinello S., Purbhoo M. A., Dunsby C., Neil M. A., French P. M., “Easystorm: a robust, lower-cost approach to localisation and tirf microscopy,” J. Biophotonics 9(9), 948–957 (2016). 10.1002/jbio.201500324 [DOI] [PubMed] [Google Scholar]
- 23.Arsić A., Stajković N., Spiegel R., Nikić-Spiegel I., “Effect of vectashield-induced fluorescence quenching on conventional and super-resolution microscopy,” Sci. Rep. 10(1), 6441 (2020). 10.1038/s41598-020-63418-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Prakash K., “Laser-free super-resolution microscopy,” Phil. Trans. R. Soc. A 379(2199), 20200144 (2021). 10.1098/rsta.2020.0144 [DOI] [PubMed] [Google Scholar]
- 25.Edelstein A. D., Tsuchida M. A., Amodaj N., Pinkard H., Vale R. D., Stuurman N., “Advanced methods of microscope control using μmanager software,” J. Biol. Methods 1(2), e10 (2014). 10.14440/jbm.2014.36 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Weksler B., Subileau E., Perriere N., Charneau P., Holloway K., Leveque M., Tricoire-Leignel H., Nicotra A., Bourdoulous S., Turowski P., Male D. K., Roux F., Greenwood J., Romero I. A., Couraud P. O., “Blood-brain barrier-specific properties of a human adult brain endothelial cell line,” FASEB J. 19(13), 1872–1874 (2005). 10.1096/fj.04-3458fje [DOI] [PubMed] [Google Scholar]
- 27.Barnier J.-P., Meyer J., Kolappan S., Bouzinba-Ségard H., Gesbert G., Jamet A., Frapy E., Schönherr-Hellec S., Capel E., Virion Z., Dupuis M., Bille E., Morand P., Schmitt T., Bourdoulous S., Nassif X., Craig L., Coureuil M., “The minor pilin pilv provides a conserved adhesion site throughout the antigenically variable meningococcal type iv pilus,” Proc. Natl. Acad. Sci. 118(45), e2109364118 (2021). 10.1073/pnas.2109364118 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Chazeau A., Katrukha E. A., Hoogenraad C. C., Kapitein L. C., “Studying neuronal microtubule organization and microtubule-associated proteins using single molecule localization microscopy,” in Methods in Cell Biology , vol. 131 (Elsevier, 2016), pp. 127–149. [DOI] [PubMed] [Google Scholar]
- 29.Martens K. J., Bader A. N., Baas S., Rieger B., Hohlbein J., “Phasor based single-molecule localization microscopy in 3d (PSMLM-3d): An algorithm for mhz localization rates using standard cpus,” J. Chem. Phys. 148(12), 123311 (2018). 10.1063/1.5005899 [DOI] [PubMed] [Google Scholar]
- 30.“Fiji macros for storm analysis,” GitHub, https://github.com/LaboratoryOpticsBiosciences/STORM_Analysis.
- 31.Mailfert S., Touvier J., Benyoussef L., Fabre R., Rabaoui A., Blache M.-C., Hamon Y., Brustlein S., Monneret S., Marguet D., Bertaux N., “A theoretical high-density nanoscopy study leads to the design of unloc, a parameter-free algorithm,” Biophys. J. 115(3), 565–576 (2018). 10.1016/j.bpj.2018.06.024 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.B. (EPFL) , “Biop FRC imagej macro,” GitHub, 2022, https://github.com/BIOP/ijp-frc.
- 33.Letterier C., “Christorm fiji macros,” GitHub, 2022, https://github.com/cleterrier/ChriSTORM.
- 34.Leterrier C., Potier J., Caillol G., Debarnot C., Boroni F. R., Dargent B., “Nanoscale architecture of the axon initial segment reveals an organized and robust scaffold,” Cell Rep. 13(12), 2781–2793 (2015). 10.1016/j.celrep.2015.11.051 [DOI] [PubMed] [Google Scholar]
- 35.Boukhatem H., Durel B, Raimbault M., et al. , “Raw storm data,” Zenodo 2022, 10.5281/zenodo.7054330. [DOI]
- 36.Nieuwenhuizen R. P., Lidke K. A., Bates M., Puig D. L., Grünwald D., Stallinga S., Rieger B., “Measuring image resolution in optical nanoscopy,” Nat. Methods 10(6), 557–562 (2013). 10.1038/nmeth.2448 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Hoogendoorn E., Crosby K. C., Leyton-Puig D., Breedijk R. M., Jalink K., Gadella T. W., Postma M., “The fidelity of stochastic single-molecule super-resolution reconstructions critically depends upon robust background estimation,” Sci. Rep. 4(1), 3854 (2015). 10.1038/srep03854 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Lab H., “Fast temporal median filter fiji plugin,” Github, 2022, https://github.com/HohlbeinLab/FTM2.
- 39.Platonova E., Winterflood C. M., Ewers H., “A simple method for gfp-and rfp-based dual color single-molecule localization microscopy,” ACS Chem. Biol. 10(6), 1411–1416 (2015). 10.1021/acschembio.5b00046 [DOI] [PubMed] [Google Scholar]
- 40.Winterflood C. M., Platonova E., Albrecht D., Ewers H., “Dual-color 3d superresolution microscopy by combined spectral-demixing and biplane imaging,” Biophys. J. 109(1), 3–6 (2015). 10.1016/j.bpj.2015.05.026 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Andronov L., Genthial R., Hentsch D., Klaholz B. P., “A spectral demixing method for high-precision multi-color localization microscopy,” bioRxiv, (2021). 10.1101/2021.12.23.473862 [DOI] [PMC free article] [PubMed]
- 42.Lambert T. J., “Fpbase: a community-editable fluorescent protein database,” Nat. Methods 16(4), 277–278 (2019). 10.1038/s41592-019-0352-8 [DOI] [PubMed] [Google Scholar]
- 43.Ries J., “Smap: a modular super-resolution microscopy analysis platform for smlm data,” Nat. Methods 17(9), 870–872 (2020). 10.1038/s41592-020-0938-1 [DOI] [PubMed] [Google Scholar]
- 44.Gustafsson M. G., “Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy,” J. Microsc. 198(2), 82–87 (2000). 10.1046/j.1365-2818.2000.00710.x [DOI] [PubMed] [Google Scholar]
- 45.Schermelleh L., Carlton P. M., Haase S., Shao L., Winoto L., Kner P., Burke B., Cardoso M. C., Agard D. A., Gustafsson M. G., Leonhardt H., Sedat J. W., “Subdiffraction multicolor imaging of the nuclear periphery with 3d structured illumination microscopy,” Science 320(5881), 1332–1336 (2008). 10.1126/science.1156947 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Hell S. W., Wichmann J., “Breaking the diffraction resolution limit by stimulated emission: stimulated-emission-depletion fluorescence microscopy,” Opt. Lett. 19(11), 780–782 (1994). 10.1364/OL.19.000780 [DOI] [PubMed] [Google Scholar]
- 47.Görlitz F., Guldbrand S., Runcorn T. H., Murray R. T., Jaso-Tamame A. L., Sinclair H. G., Martinez-Perez E., Taylor J. R., Neil M. A., Dunsby C., French P. M., “Easyslm-sted: Stimulated emission depletion microscopy with aberration correction, extended field of view and multiple beam scanning,” J. Biophotonics 11(11), e201800087 (2018). 10.1002/jbio.201800087 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Citations
- Boukhatem H., Durel B, Raimbault M., et al. , “Raw storm data,” Zenodo 2022, 10.5281/zenodo.7054330. [DOI]
Data Availability Statement
Data underlying the results presented in this paper are available on Zenodo [35].