Abstract
Background
To assess the accuracy of fully automated deep learning (DL) based coronary artery calcium scoring (CACS) from non-contrast computed tomography (CT) as acquired for attenuation correction (AC) of cardiac single-photon-emission computed tomography myocardial perfusion imaging (SPECT-MPI).
Methods and Results
Patients were enrolled in this study as part of a larger prospective study (NCT03637231). In this study, 56 Patients who underwent cardiac SPECT-MPI due to suspected coronary artery disease (CAD) were prospectively enrolled. All patients underwent non-contrast CT for AC of SPECT-MPI twice. CACS was manually assessed (serving as standard of reference) on both CT datasets (n = 112) and by a cloud-based DL tool. The agreement in CAC scores and CAC score risk categories was quantified. For the 112 scans included in the analysis, interscore agreement between the CAC scores of the standard of reference and the DL tool was 0.986. The agreement in risk categories was 0.977 with a reclassification rate of 3.6%. Heart rate, image noise, body mass index (BMI), and scan did not significantly impact (p=0.09 - p=0.76) absolute percentage difference in CAC scores.
Conclusion
A DL tool enables a fully automated and accurate estimation of CAC scores in patients undergoing non-contrast CT for AC of SPECT-MPI.
Supplementary Information
The online version contains supplementary material available at 10.1007/s12350-022-02940-7.
Keywords: CAD, Atherosclerosis, CT, SPECT, Diagnostic and prognostic application
Introduction
Single-photon-emission computed tomography myocardial perfusion imaging (SPECT-MPI) is a well-established cardiac imaging modality that provides diagnostic and prognostic information in patients with known or suspected coronary artery disease (CAD).1–5 In SPECT-MPI, stress-induced perfusion abnormalities are visualized and quantified. Thus, the extent and severity of ischemia, scar burden and left ventricular systolic function and volume can be assessed.6
As SPECT-MPI is susceptible to soft-tissue attenuation, potentially compromising its diagnostic value,1 international guidelines recommend the use of an attenuation correction (AC) method. Non-contrast computed tomography is commonly used for AC as it provides further clinically relevant image information. Moreover, non-contrast CT enables the visualization of calcifications which can be leveraged to quantify coronary artery calcium (CAC).6 The latter is considered an important biomarker and independent predictor of cardiovascular mortality and all-cause mortality in patients with CAD. Importantly, patients’ risk is stratified according to categories, with CAC scores > 400, for example, implying substantial risk of cardiovascular disease events and mortality.7–10
However, the quantitative assessment of CAC (so-called calcium scoring, CACS) is a time-consuming and tedious manual task.11–13 Nevertheless, performing CACS in patients undergoing SPECT-MPI is clinically valuable, and as previous studies have shown, CACS provides additional complementary and prognostically relevant information to SPECT-MPI findings.6,14,15
With recent advances in the field of artificial intelligence for medical imaging,11–13,16,17 deep learning (DL) based CACS has become feasible. Such tools enable the automatic quantification of CAC, thus alleviating the need to perform CACS manually.11–13,17 Thus, a DL-based CACS approach may enable the extraction of potentially valuable data from non-contrast CT scans as performed for AC of SPECT-MPI without having to allocate time and effort to perform CACS manually.
Accordingly, we sought to test the feasibility of performing automated DL-based CACS on non-contrast CT scans for AC of myocardial SPECT-MPI. We hypothesized that a DL tool would provide accurate CACS results in a fully automated manner relative to manual measurements as the standard of reference.
Materials and Methods
Study Subjects
In this single-center, institutional review board-approved intra-individual comparative study, we identified and enrolled 56 patients (46 male (82%), age: 63 ± 9 years, body mass index (BMI): 28 ± 5 kg/m2) who underwent clinically indicated SPECT-MPI on a CZT technology-based camera and non-contrast CT based AC for the exclusion of CAD (Table 1). Some of the patients in the current investigation have been included in previous studies.1,8 All patients were part of a prospective study cohort in which two non-contrast CT scans for AC were acquired per patient (NCT03637231): The first scan was acquired prior to pharmacological heart rate control. The second scan was acquired after pharmacological heart rate control to achieve a target heart rate of below 65 beats per minute (BPM). For each patient, both scans were analyzed to quantify scan-rescan reproducibility and variability of CAC scores.
Table 1.
Demographics of study patients (n = 56)
Number of patients | 56 |
Male patients, n (%) | 46 (82%) |
Age, years | 63 ± 9 |
Height, cm | 174.6 ± 9.7 |
BMI, kg/m2 | 28 ± 5 |
Cardiovascular risk factors, n (%) | |
Smoking | 17 (30%) |
Diabetes mellitus | 6 (11%) |
Hypertension | 32 (57%) |
Dyslipidemia | 29 (52%) |
Cardiac history, n (%) | |
Previous MI | 0 (0%) |
Previous ICA | 1 (0%) |
Previous CABG | 0 (0%) |
Symptoms, n (%) | |
Asymptomatic | 12 (21%) |
Typical angina pectoris | 6 (11%) |
Atypical chest pain | 21 (37%) |
Dyspnea | 14 (25%) |
Values given are mean±standard deviation or absolute numbers and percentages in brackets
BMI Body mass index; CABG coronary artery bypass grafting; ICA invasive coronary angiography; MI myocardial infarction
Cardiac SPECT-MPI and CACS Imaging
All patients underwent a 1-day 99mTc-tetrofosmin stress-rest myocardial perfusion imaging (MPI) protocol in accordance with current guidelines18. Exercise bicycle stress test was performed according to a modified Bruce protocol, and pharmacological stress was induced by intravenous regadenoson, adenosine, or dobutamine infusion. At peak stress (i.e., after 3 minutes of induced stress by adenosine or after reaching 85% of the patient’s predicted maximum heart rate), a weight-adjusted dose of 99mTc-tetrofosmin was injected. Electrocardiogram (ECG)-gated stress images were acquired on a gamma camera with CZT-based detectors (Discovery 530 NMc, GE Healthcare, Milwaukee, WI, USA) 60-90 minutes after the isotope injection. Thereafter, a tracer dose three times the stress dose was administered at rest, followed by image acquisition with the same protocol as for stress. Patients with a BMI < 35 kg·m−2 received approximately 130 megabecquerel (MBq) for stress and 390 MBq for rest acquisition, whereas patients with a BMI ≥ 35 kg·m−2 received 180 MBq for stress and 540 MBq for rest acquisition.
Additionally, all patients underwent a non-contrast CT on a newest-generation 256-slice CT scanner (Revolution CT, GE Healthcare) for the creation of attenuation maps of the chest. All scans were performed in cranio-caudal direction during inspiratory breath-hold with prospective ECG-triggering as reported previously.19 In each patient, two separate scans (tube current 440.5 ± 96 milliampers and tube voltage 120 kilovolts) were acquired: The first one prior to pharmacological heart rate control and the second one after the application of metoprolol to decrease heart rate below 65 BPM. For each acquisition, the following scan parameters were chosen: collimation of 256 × 0.625 mm, z-coverage of 12-16 cm, field of view of 25 cm, and gantry rotation time of 280 milliseconds. As recommended elsewhere, images were reconstructed using filtered back projection (FBP).20,21 Effective radiation dose as quantified by the dose length product (DLP) multiplied with a conversion factor (0.014 mSv × mGy−1 × cm−1)22 was 0.6 ± 0.3 mSv.
Manual Calcium Scoring
Images were transferred to a dedicated workstation (Advantage AW 4.4, GE Healthcare) with a dedicated CACS software (SmartScore 4.0, GE Healthcare). All pixels with an attenuation equal or above the lowest threshold (e.g., ≥ 130 HU for 120-kV scans) having an area ≥ 1 mm2 are automatically color marked, and lesions are manually selected by creating a region of interest around all lesions found in a coronary artery. The software then calculates the CAC score, as previously described.20 In brief, a score for each region of interest is calculated by multiplying the density score (i.e., the thresholds) and the area of calcifications. A total CAC score is then determined by adding up the scores for each CT slice. Importantly, the software computes an overall CAC score and vessel-wise CAC scores. Of note, the thresholds for CACS are only applied to pixels with a density equal or larger than the lowest threshold and an area of ≥ 1 mm2. This eliminates single pixels with a density above the thresholds due to noise. All datasets were analyzed by two experienced readers in random order, and measurements from both readers were averaged.8 CAC risk categories were defined according to the following CAC score boundaries: 0, 1-100, 101-400, > 400.
Deep Learning Calcium Scoring
CAC scoring was performed by a fully automated DL-based CAC scoring tool (AVIEW CAC, Coreline Soft, access via https://cloud.corelinesoft.eu/login). In brief, the tool was trained on a 3-dimensional U-net architecture using non-enhanced cardiac CT scans acquired from multiple vendors and scanners as input data. No specific training data were included in this current study. A detailed description of this DL tool can be found elsewhere.11,23 In short, the network makes predictions with patches, and the segmentation mask is reconstructed with output patches. After reconstruction, only the coronary artery areas (i.e., left main coronary artery [LM], left anterior descending coronary artery [LAD], left circumflex coronary artery [LCX], and right coronary artery [RCA]) near the heart are filtered using the predicted Ventricle area. A multi-branch network is used consisting of the same features for coronary arteries and other coronary structures. It is based on U-Net (see also https://arxiv.org/abs/1505.04597) with residual units (see also https://arxiv.org/abs/1603.05027) in the feature extractors, and the input layers have shape 64 × 64 × 64 × 1 (3D patch with single channel). The output layers of two networks have shape 64 × 64 × 64 × channels. One branch has the last layer with 5 channels (with background) for coronary arteries, the other has the last layer with 8 channels (with background) for other coronary structures.
Image Analysis
To quantify image noise, one reader (A.G.G., with 7 years of experience in cardio-thoracic imaging) placed region of interests (ROI) on three consecutive CT image slices in the left ventricle. The standard deviation of values in each ROI was considered as image noise. The values from the three measurements were averaged and were then considered representative for further analyses.
Statistical Analysis
Descriptive statistics including mean ± standard deviation, counts, and percentages were used to present the results. To quantify the agreement of CAC scores, linear regression models were fitted, Bland-Altman analysis was performed, and two-way intraclass correlation coefficients (ICC) were computed. To quantify the agreement of CAC risk categories, weighted Kappa analysis was performed. The following scale was used to classify correlation coefficients: values of less than 0.20 were indicative of poor agreement; 0.21-0.40, fair agreement; 0.41-0.60, moderate agreement; 0.61-0.80, good agreement; and 0.81-1.00, excellent agreement. To compare values stratified by groups (such as CAC scores or heart rate), paired t-tests or wilcoxon signed-rank tests were used. Potential differences in reclassification rates of CAC score categories, paired z-tests of proportions were used. Lastly, a generalized linear model (GLM) using iteratively reweighted least squares to find the maximum likelihood estimates was fitted. The absolute percentage difference in CAC scores between the standard of reference and the DL tool was implemented as dependent variable and heart rate, image noise, BMI, and scan were considered as predictors. Two-tailed p-values < 0.05 were considered significant. All statistical analyses were performed in the R programming language (version 4.0.2; R Foundation for Statistical Computing, Vienna, Austria, https://www.R-project.org).
Results
Overall Accuracy of DL-CACS
A visual representation of the study results is provided in Figure 1. An overview of the data is provided in Table 2. For the overall analysis, each scan was considered for statistical analysis. Thus, 112 scans from 56 patients were considered. The DL tool successfully managed to automatically perform CAC scoring in all cases within 63 ± 48 s. The overall CAC score was 352.6 ± 491 for the standard of reference and 352.1 ± 461 for the DL tool. The regression model using CAC scores of the standard of reference as dependent variable and the CAC scores of the DL tool as predictor (R2 = 0.98, p < 0.001), exhibited a slope of 1.1 and an intercept of -18. Bland-Altman analysis (Fig. 1) revealed a bias of 0.44, a lower limit of agreement of − 156.8, and an upper limit of agreement of 157.7. Interscore agreement (ICC) between the CAC scores of the standard of reference and the DL tool was 0.986 (95%CI 0.979, 0.99).
Figure 1.
Visual representation of the study data including the data of all scans (n = 112): Linear regression model (A), Bland-Altman analysis (B), as well as CAC scores stratified by risk categories (C) of the manual coronary artery calcium scoring (i.e., standard of reference) and fully automated DL-CACS
Table 2.
Confusion matrices of risk categories between DL-CACS and manual CACS. Weighted kappa value was 0.977
Manual CACS | DL-CACS | Totala | Underestimationb | Overestimationb | Concordanceb | |||
---|---|---|---|---|---|---|---|---|
0 | 1-100 | 101-400 | > 400 | |||||
0 | 0 | 1 | 0 | 0 | 1 | 0 (0%) | 1 (100%) | 0 (0%) |
1-100 | 0 | 49 | 0 | 0 | 49 | 0 (0%) | 0 (0%) | 49 (100%) |
101-400 | 0 | 1 | 23 | 2 | 26 | 1 (3.8%) | 2 (7.7%) | 23 (88.5%) |
>400 | 0 | 0 | 0 | 36 | 36 | 0 (0%) | 0 (0%) | 36 (100%) |
CACS coronary artery calcium scoring; DL deep learning
an = 112 scans from 56 patients
b i.e., manual coronary artery calcium scoring as standard of reference
In terms of CAC risk categories, a reclassification rate of 3.6% (4 of 112 cases) was found between the standard of reference and the DL tool. In all 4 cases, reclassification was only by one category. Specifically, the DL tool falsely increased the class by one category in 3 cases and falsely decreased the class by one category in 1 case. Interclass agreement was 0.977 as determined by weighted Kappa analysis. Reasons for CAC score reclassification (i.e., under- or overestimation) of the DL tool in the 4 cases are provided in Table 3.
Table 3.
Cases (4/112) of patients where DL-CACS led to reclassification of coronary risk
N | Age | Manual-CACSa | DL-CACS | DL-Classification | Reason for wrong classification of DL tool |
---|---|---|---|---|---|
1 | 75 | 174 | 90 | Underestimated | Missed calcification in LCX |
2 | 66 | 315 | 551 | Overestimated | Aortic valve calcification falsely rated as CAC |
3 | 66 | 315 | 575 | Overestimated | Aortic valve calcification falsely rated as CAC |
4 | 55 | 0 | 3 | Overestimated | Image noise falsely rated as CAC |
CACS coronary artery calcium scoring; DL deep learning; LCX left circumflex artery
i.e., manual coronary artery calcium scoring as standard of reference
The generalized model with heart rate, image noise, BMI, and scan (i.e., scan 1 or 2) as predictors revealed no significant impact of all variables (p = 0.09 – p = 0.76) on absolute percentage difference in CAC scores between the DL tool and the standard of reference.
Per Vessel Analysis of DL-CACS
On a per-vessel basis, interscore agreement (ICC) between the standard of reference and the DL tool was 0.635 (95%CI 0.511, 0.734), 0.95 (95%CI 0.928, 0.965), 0.928 (95%CI 0.897, 0.95), and 0.991 (95%CI 0.988, 0.994) for LM, LAD, LCX, and RCA, respectively. Specifically, the CAC scores (standard of reference / DL tool) were 13.9 ± 32.1 / 29.9 ± 50.2 for LM, 174 ± 260.7 / 152.6 ± 212.2 for LAD, 54.5 ± 101.3 / 53.2 ± 106.2 for LCX, and finally 110.1 ± 194.6 / 116.4 ± 198.6 for RCA.
Scan-Rescan Reproducibility and Variability
Heart rate was 65.7 ± 13.4 BPM for the first scan and 58.9 ± 6.4 BPM for the second scan (p < 0.001). An average CAC score variability of 6.7 ± 64.7 and 13.6 ± 57.6 between both scans was observed for the standard of reference and the DL tool, respectively, without significant differences between both methods (p = 0.45). Interscore agreement between both scans as quantified by ICC was 0.991 (95%CI 0.985, 0.995) and 0.992 (95%CI 0.987, 0.995) for the standard of reference and the DL tool, respectively.
In terms of quantification accuracy of the DL tool relative to manual measurements as the standard of reference, ICC of CAC scores was 0.984 (95%CI 0.973, 0.991) for the first scan and 0.988 (95%CI 0.979, 0.993) for the second scan. Reclassification rate for CAC risk categories was 1.8% for the first scan and 5.4% for the second scan (p < 0.001). Interclass agreement of CAC risk categories was 0.988 and 0.966 for the first and second scan, respectively, as determined by weighted Kappa analysis.
A representative case of the DL-CACS tool correctly identifying the coronary calcium burden in a patient is presented in Fig. 2.
Figure 2.
Representative CT images of a 64-year-old woman with a body mass index of 22.9 kg/m2 with severe coronary artery calcifications. Images from gated non-contrast CT scanner for the creation of attenuation maps is presented in the upper row and fully automated deep learning coronary artery calcium scoring (DL-CACS) is presented in lower row. Coronary calcifications in the left main (LM), left anterior descending (LAD), left circumflex artery (LCX), and right coronary artery (RCA), were correctly marked by the DL-CACS tool resulting in a total score of 894. The score from manual readout of CAC scan was 871
Discussion
In this single-center, prospective, intra-individual comparative study, we compared the quantitative accuracy of CACS as performed fully automatically on non-contrast CT acquired for AC of SPECT-MPI by a DL tool relative to manual measurements as the standard of reference.
Our data indicate that the accuracy of the DL tool is very high, with interscore agreement of CAC scores and interclass agreement of CAC risk categories being excellent as well. Furthermore, the scan-rescan evaluation confirmed quantification stability and robustness of the DL tool on par with that of manual measurements. Further, quantitative accuracy was not significantly influenced by heart rate, image noise or BMI, thus confirming the overall validity and value of the DL tool. Given these results, our study suggests that a DL tool may derive accurate CAC scores from non-contrast CT scans as acquired for AC of SPECT-MPI in a fully automated manner.
The use of AI-based solutions for automated CACS has only gained traction recently. Notably, Van Velzen et al. developed a DL-backed CACS approach that achieved ICC scores of as high as 0.99 and risk category reclassification rates of as low as 3% when using calcium scans from CCTA as input data.12 Vonder et al. showed that a DL-CACS tool achieved an ICC score of 0.96 and a risk category reclassification rate of 2% relative to manual measurements in a cohort of 997 patients who underwent calcium scans of CCTA.11 To the best of our knowledge, this is the first study systematically assessing the validity of a DL approach for CACS on CT data as acquired for SPECT-MPI. Here, we achieved a similarly high performance as reported in previous studies with an ICC score as high as 0.986 and risk category reclassification rates of 3.6% between the DL tool and manual measurements.
Notably, while the overall agreement of CAC scores was very high (ICC: 0.986), we observed differences between the various coronary arteries. Specifically, for the LM artery, the interscore agreement of CAC scores was lower (ICC: 0.635) than that of the other coronary arteries (ICC ranging from 0.928 to 0.991). On average, the LM artery exhibited less CAC than the other vessels. Even small deviations in CAC scores between the reference standard and the DL tool may thus have caused considerable reductions in the the ICC. Furthermore, it should ne noted that the DL tools allocation of CAC to the various vessels may have differed slightly from that of the expert readers. Thus, in some cases the DL tool may have accurately recognized CAC but may have falsely allocated it to a different vessel which would then have resulted in lower ICC values for a given vessel.
On another note, it should be mentioned that our reclassification rates were slightly higher than those of previous studies. The reason is, firstly, that at least compared to Vonder et al. we had more and narrower risk categories.11 Secondly, our data correspond to a real validation set. That is, the DL tool was not retrained and optimized with data from our study cohort and institution. In contrast, Van Velzen et al. implemented additional protocol-specific training to further optimize the performance of their DL tool.12 Furthermore, it should be noted that the study cohort may have differed between ours and previous studies. In our study, nearly all patients presented with CAC as our patients all had a moderate to high pre-test probability of CAD, as can be expected form patients referred for SPECT-MPI. In contrast, in patients undergoing CCTA, the pre-test probability of CAD is lower. Notably, Vonder et al. used patient data from a population-based cardiovascular screening trial.11 In such a cohort, the proportion of patients with very little to no calcium is probably much higher than in our cohort.
Lastly, we would like to highlight the value of CACS in patients undergoing SPECT-MPI. Chang et al. followed up a cohort of 1126 generally asymptomatic patients without previous CAD who had undergone SPECT-MPI and CACS within a close period of time. The authors concluded that at a median follow-up of 6.9 years, CACS and SPECT-MPI results provided both independent and complementary prognostic information for total cardiac events and all-cause death/myocardial infarction.14 Furthermore, Engbers et al. showed that in a cohort of 4897 symptomatic patients, CACS and SPECT-MPI findings were independent predictors of major adverse cardiac events (MACE). The authors thus recommended routinely performing CACS in adjunct to SPECT-MPI in patients suspected of CAD in an effort to further improve risk prediction during follow-up.15
In addition, it should be mentioned that besides the prognostic and diagnostic value of CACS, it is also otherwise useful to extract the maximum information from the data acquired as part of an examination. Especially when considering that a SPECT-MPI examination involves radiation exposure, one should strive to maximize the information content and diagnostic value of the examination.24 Due to the fact that DL-based CACS does not require any further time investment, the use of such a tool is particularly useful and desirable in this context. Importantly, it should be considered that the results of this study carry implications for a wide range of diagnostic imaging modalities in nuclear cardiology. Specifically, the DL tool presented in this study may also be deployed for other imaging modalities such as hybrid cardiac PET/CT examinations.
Our study has the following limitations: First, this was a single-center study with a limited number of subjects. Secondly, all images were acquired on a single scanner with a single imaging protocol. We acknowledge that varying scan parameters may influence the results. Thirdly, we acknowledge that quantification accuracy may depend on the study cohort examined. Future studies should assess the quantification accuracy of the DL tool in patients with no or very high CAC.
New Knowledge Gained
A deep learning powered approach can fully automatically and accurately perform coronary artery calcium scoring from non-contrast computed tomography images as acquired for attenuation correction of single-photon-emission computed tomography myocardial perfusion imaging. This approach allows for the opportunistic and effortless extraction of coronary artery calcium scores in patients undergoing single-photon-emission computed tomography myocardial perfusion imaging.
Conclusion
Our study shows that a DL tool enables a fully automated and accurate estimation of CAC scores in patients undergoing SPECT-MPI. Thus, DL-based CACS may facilitate the further implementation of CAC scores as a routine imaging marker determined during the workup of SPECT-MPI examinations.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
Dr. Antonio G. Gennari and Dr. Michael Messerli are supported by a grant from the CRPP AI Oncological Imaging Network of the University of Zurich. Dr. Stephan Skawran is supported by a research grant from the Palatin Foundation, Switzerland. The authors would like to thank Vanja Azabagic-Thursar, Viviane Sasu and Markus Gumbert from Coreline Europe for their excellent technical support.
Abbreviations
- SPECT-MPI
Single-photon-emission computed tomography myocardial perfusion imaging
- CAD
Coronary artery disease
- AC
Attenuation correction
- CAC
Coronary artery calcium
- CACS
Coronary artery calcium scoring
- DL
Deep learning
- BMI
Body mass index
- BPM
Beats per minute
- ROI
Region of interest
- ICC
Intraclass correlation coefficient
- GLM
Generalized linear model
- LM
Left main coronary artery
- LAD
Left anterior descending coronary artery
- LCX
Left circumflex coronary artery
- RCA
Right coronary artery
- CCTA
Coronary computed tomography angiography
- MACE
Major adverse cardiac events
Funding
Open access funding provided by University of Zurich.
Footnotes
The authors of this article have provided a PowerPoint file, available for download at SpringerLink, which summarises the contents of the paper and is free for re-use at meetings and presentations. Search for the article DOI on SpringerLink.com.
All editorial decisions for this article, including selection of reviewers and the final decision, were made by guest editor Jeroen J Bax, MD.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Grossmann M, et al. Ultra-low-dose computed tomography for attenuation correction of cadmium-zinc-telluride single photon emission computed tomography myocardial perfusion imaging. J Nucl Cardiol. 2020;27:228–237. doi: 10.1007/s12350-018-1303-y. [DOI] [PubMed] [Google Scholar]
- 2.Clerc OF, et al. Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: Impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction. Eur Heart J Cardiovasc Imaging. 2017;18:31–38. doi: 10.1093/ehjci/jew031. [DOI] [PubMed] [Google Scholar]
- 3.Calabretta R, et al. Prognostic value of divergent pattern detection by 99mTc-sestamibi gated SPECT in patients with anterior acute myocardial infarction. J Nucl Cardiol. 2021 doi: 10.1007/s12350-021-02874-6. [DOI] [PubMed] [Google Scholar]
- 4.Megna R, et al. External validation of the CRAX2MACE model in an Italian cohort of patients with suspected coronary artery disease undergoing stress myocardial perfusion imaging. J Nucl Cardiol. 2021 doi: 10.1007/s12350-021-02855-9. [DOI] [PubMed] [Google Scholar]
- 5.Liu H, et al. Post-reconstruction attenuation correction for SPECT myocardium perfusion imaging facilitated by deep learning-based attenuation map generation. J Nucl Cardiol. 2021 doi: 10.1007/s12350-021-02817-1. [DOI] [PubMed] [Google Scholar]
- 6.Trpkov C, et al. Visually estimated coronary artery calcium score improves SPECT-MPI risk stratification. IJC Heart Vasc. 2021;35:100827. doi: 10.1016/j.ijcha.2021.100827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Peng AW, et al. Very high coronary artery calcium (≥1000) and association with cardiovascular disease events, non-cardiovascular disease outcomes, and mortality: Results from MESA. Circulation. 2021;143:1571–1583. doi: 10.1161/CIRCULATIONAHA.120.050545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Gräni C, et al. Ultra-low-dose coronary artery calcium scoring using novel scoring thresholds for low tube voltage protocols—a pilot study. Eur Heart J Cardiovasc Imaging. 2018;19:1362–1371. doi: 10.1093/ehjci/jey019. [DOI] [PubMed] [Google Scholar]
- 9.Horiguchi J, et al. Effect of heart rate and body mass index on the interscan and interobserver variability of coronary artery calcium scoring at prospective ECG-triggered 64-slice CT. Korean J Radiol. 2009;10:340. doi: 10.3348/kjr.2009.10.4.340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Clerc OF, et al. Non-invasive screening for coronary artery disease in asymptomatic diabetic patients: A systematic review and meta-analysis of randomised controlled trials. Eur Heart J Cardiovasc Imaging. 2018;19:838–846. doi: 10.1093/ehjci/jey014. [DOI] [PubMed] [Google Scholar]
- 11.Vonder M, et al. Deep learning for automatic calcium scoring in population-based cardiovascular screening. JACC: Cardiovasc Imaging. 2021;15:366–367. doi: 10.1016/j.jcmg.2021.07.012. [DOI] [PubMed] [Google Scholar]
- 12.van Velzen SGM, et al. Deep learning for automatic calcium scoring in CT: Validation using multiple cardiac CT and chest CT protocols. Radiology. 2020;295:66–79. doi: 10.1148/radiol.2020191621. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Eng D, et al. Automated coronary calcium scoring using deep learning with multicenter external validation. Digit Med. 2021;4:88. doi: 10.1038/s41746-021-00460-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Chang SM, et al. The coronary artery calcium score and stress myocardial perfusion imaging provide independent and complementary prediction of cardiac risk. J Am Coll Cardiol. 2009;54:1872–1882. doi: 10.1016/j.jacc.2009.05.071. [DOI] [PubMed] [Google Scholar]
- 15.Engbers EM, et al. Prognostic value of coronary artery calcium scoring in addition to single-photon emission computed tomographic myocardial perfusion imaging in symptomatic patients. Circ Cardiovasc Imaging. 2016;9:e003966. doi: 10.1161/CIRCIMAGING.115.003966. [DOI] [PubMed] [Google Scholar]
- 16.Schwyzer M, et al. Artificial intelligence for detecting small FDG-positive lung nodules in digital PET/CT: Impact of image reconstructions on diagnostic performance. Eur Radiol. 2020;30:2031–2040. doi: 10.1007/s00330-019-06498-w. [DOI] [PubMed] [Google Scholar]
- 17.Winkel DJ, et al. Deep learning for vessel-specific coronary artery calcium scoring: validation on a multi-centre dataset. Eur Heart J Cardiovasc Imaging. 2021 doi: 10.1093/ehjci/jeab119. [DOI] [PubMed] [Google Scholar]
- 18.Henzlova MJ, Duvall WL, Einstein AJ, Travin MI, Verberne HJ. ASNC imaging guidelines for SPECT nuclear cardiology procedures: Stress, protocols, and tracers. J Nucl Cardiol. 2016;23:606–639. doi: 10.1007/s12350-015-0387-x. [DOI] [PubMed] [Google Scholar]
- 19.Schepis T, et al. Use of coronary calcium score scans from stand-alone multislice computed tomography for attenuation correction of myocardial perfusion SPECT. Eur J Nucl Med Mol Imaging. 2007;34:11–19. doi: 10.1007/s00259-006-0173-8. [DOI] [PubMed] [Google Scholar]
- 20.Gebhard C, et al. Coronary artery calcium scoring: Influence of adaptive statistical iterative reconstruction using 64-MDCT. Int J Cardiol. 2013;167:2932–2937. doi: 10.1016/j.ijcard.2012.08.003. [DOI] [PubMed] [Google Scholar]
- 21.Messerli M, et al. Impact of advanced modeled iterative reconstruction on coronary artery calcium quantification. Acad Radiol. 2016;23:1506–1512. doi: 10.1016/j.acra.2016.08.008. [DOI] [PubMed] [Google Scholar]
- 22.Hausleiter J. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301:500. doi: 10.1001/jama.2009.54. [DOI] [PubMed] [Google Scholar]
- 23.Lee J-G, et al. Fully automatic coronary calcium score software empowered by artificial intelligence technology: Validation study using three CT cohorts. Korean J Radiol. 2021;22:1764. doi: 10.3348/kjr.2021.0148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Gennari AG, et al. Low-dose CT from myocardial perfusion SPECT/CT allows the detection of anemia in preoperative patients. J Nucl Cardiol. 2022 doi: 10.1007/s12350-021-02899-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.