A numerical human brain phantom for dynamic glucose-enhanced (DGE) MRI: On the influence of head motion at 3 T

PM Lehmann 1, A Seidemo 1, M Andersen 2,3, X Xu 4,5, X Li 5,6, N Yadav 5,6, R Wirestam 1, P Liebig 7, F Testud 8, PC Sundgren 3,9,10, PCM van Zijl 5,6, L Knutsson 1,5,6
PMCID: PMC9992166  NIHMSID: NIHMS1856961  PMID: 36579955

Abstract

Purpose:

Dynamic glucose-enhanced (DGE) MRI relates to a group of exchange-based MRI techniques where the uptake of glucose analogues is studied dynamically. However, motion artefacts can be mistaken for true DGE effects, while motion correction may alter true signal effects. The aim was to design a numerical human brain phantom to simulate a realistic DGE MRI protocol at 3 T that can be used to assess the influence of head movement on the signal before and after retrospective motion correction.

Methods:

MPRAGE data from a tumor patient were used to simulate dynamic Z-spectra under the influence of motion. The DGE responses for different tissue types were simulated, creating a ground truth. Rigid head movement patterns were applied as well as physiological dilatation and pulsation of the lateral ventricles and head-motion-induced B0-changes in presence of 1st order shimming. The effect of retrospective motion correction was evaluated.

Results:

Motion artefacts similar to those previously reported for in vivo DGE data could be reproduced. Head movement of 1 mm translation and 1.5 degrees rotation led to a pseudo-DGE effect on the order of 1 % signal change. B0 effects due to head motion altered DGE changes due to a shift in the water saturation spectrum. Pseudo DGE effects were partly reduced or enhanced by rigid motion correction depending on tissue location.

Conclusion:

DGE MRI studies can be corrupted by motion artefacts. Designing post-processing methods using retrospective motion correction including B0 correction will be crucial for clinical implementation. The proposed phantom should be useful for evaluation and optimization of such techniques.

Keywords: CEST, D-glucose, glucoCEST, motion correction, simulations, dynamic glucose enhanced

1. INTRODUCTION

Characterization of the microvasculature and quantification of microcirculatory parameters such as cerebral perfusion and capillary permeability by MRI is important in several clinical applications.1 One major application is brain cancer, a disease that may cause multiple disabilities and death. Imaging therefore plays a crucial role for diagnosis and therapeutic monitoring.2,3 By intravenously injecting a contrast agent bolus and tracking it with dynamic T2/T2*-or T1-weighted MRI, quantitative and qualitative information on cerebral perfusion and/or blood brain barrier permeability can be obtained.4,5 However, these methods often use a gadolinium (Gd) –chelate contrast agent and, while these are safe for most patient populations, it has been shown that Gd accumulates in bone and tissue.69 Some of these agents are also potentially toxic for patients with kidney disease, leading to nephrogenic systemic fibrosis.10,11 Therefore, it is of interest to investigate non-Gd contrast agents.

Dynamic glucose-enhanced (DGE) MRI employs chemical exchange saturation transfer (CEST) or exchange-enhanced relaxation to study tissue uptake of glucose analogues, which, for D-glucose, is determined by tissue perfusion, transport and metabolism.1221 DGE MRI explores the change in signal intensity of several sugar hydroxyl protons, located approximately between 0.6 ppm and 3.0 ppm relative to the proton resonance of free water in the water saturation spectrum (Z-spectrum). The inherently slow (minute time scale) infusion and uptake of D-glucose requires long scan times of around 10–20 minutes. The calculation of difference images as a function of time is therefore susceptible to image artefacts related to head motion. In addition, B0 changes due to motion and field drift are problematic if a single frequency selection is used for RF saturation. Voxel displacement due to motion can render the shim settings determined at the beginning of the scan invalid,22 and heating of the iron plates in the room temperature shim trays due to eddy currents may cause field drift.23,24 Other sources of signal change may be dilatation of the lateral ventricles caused by D-glucose uptake and ventricular pulsation related to, for example, the cardiac pulsation.20,2527

A previous study investigated the effect of motion using difference images of Z-spectra without glucose infusion to create synthetic data before and after simulated motion, also taking B0-fluctuations into consideration.28 Hypo- and hyperintense signal changes of the order of 1 % were observed in the vicinity of tissue boundaries, causing pseudo-DGE effects of the same order of magnitude as the DGE signal changes. Another study concluded that standard motion correction algorithms reduced pseudo-DGE effects.29 While such retrospective motion correction may be a necessary step in post-processing, interpolation and smoothing can erroneously alter the true DGE signal entangled with residual motion artefacts at tissue interfaces.28

The aim of this study was to implement a realistic customizable framework for DGE MRI to simulate dynamic Z-spectra under the influence of motion and dynamic B0-changes. As a first application of this numerical human brain phantom, we simulated a DGE experiment that used simulated dynamic contrast responses for different tissue types as ground truth. Rigid body movement patterns and physiological volumetric changes of the lateral ventricles were included in the simulations and their influence on the DGE effects before and after retrospective motion correction were studied.

2. METHODS

2.1. MRI data acquisition

Experimental data were acquired for deriving different realistic patterns of motion and to aid in modelling the anatomical basis for the numerical human head phantom. Six subjects (3 male, 3 females, ages 39–74 years), including five patients (4 gliomas, 1 meningioma) and one healthy volunteer, were scanned at a 3 T MAGNETOM Prisma with a 20-channel head coil system (Siemens Healthcare, Erlangen Germany). The project was approved by the local ethics committee (The Regional Ethical Review Board in Lund, EPN 2017/673) and written informed consent was obtained from all participants.

All participants fasted for 4–6 hours before examination to stabilize the baseline blood glucose and insulin levels. D-glucose solution (50 mL, 50 % dextrose) was administered intravenously in one arm for a duration of 4 min using a power injector. The glucose infusion was followed by a saline flush of 30 mL included in the total infusion duration.

Four of the patients (3 gliomas, 1 meningioma) and the healthy volunteer were scanned with a prototype CEST sequence with a 3D gradient echo (GRE) readout consisting of 12 partitions (TR / TE / FA = 10 ms / 3 ms / 12°).30 Saturation was accomplished using 10 Gaussian-shaped pulses (B1 = 1.6 μT) of 100 ms duration and 10 ms interpulse delay at a single saturation offset of 2.0 ppm from the water resonance frequency. One non-saturated (S0) image was acquired by saturating at 1500 ppm at the start of the CEST image acquisition. Morphologic images were acquired using a MPRAGE sequence with 1 mm 3 isotropic voxel size and examined by an experienced neuroradiologist (PCS).

One volunteer and one patient were examined using a temporal resolution of 5 s, a voxel size of 3.3×3.3×4.0 mm 3, an acquisition time of 599 s with a 180 s baseline. Two patients were scanned with the same temporal resolution and voxel size as previously mentioned, but with a total acquisition time of 798 s including a baseline of 120 s. One patient was examined using a temporal resolution of 9.4 s, a voxel size of 1.7×1.7×4.0 mm 3 and an acquisition time of 468 s including a 180 s baseline. Finally, one patient was scanned with a 3D turbo spin echo (TSE) CEST sequence, but these data were not included for the retrieval of motion patterns for the phantom. Instead, the MPRAGE images acquired from this patient were selected for anatomical modelling of the phantom.

2.2. Simulation pipeline

The phantom was based on an MPRAGE brain image segmentation, used to model dynamic Z-spectra with or without D-glucose infusion under the influence of B0-changes and head motion. An overview of the simulation pipeline is shown in Figure 1.

Figure 1.

Figure 1

Steps of the numerical head phantom framework illustrated in a chronologically ordered pipeline (A): Segmentation of high-resolution anatomical images results in different tissue components, i.e., air, bone, fat, and intracranial tissue, in which veins (V), arteries (A), grey matter (GM), white matter (WM), CSF and tumor (Gb) were later identified. Each voxel has the option to add dynamic contrast. Motion patterns representing a least severe pattern (LSP) and a most severe pattern (MSP) of movement in the subject group can be included followed by adding Rician noise, smoothing and downsampling to a voxel size of 2×2×3 mm 3. The arrows indicate the order of the simulation steps. (B): Simulations steps over time for the ground truth without motion (left) and for the cases with motion (right).

An MPRAGE brain image from a patient with a glioblastoma was used to segment grey matter (GM), white matter (WM), cerebrospinal fluid (CSF), other soft tissue and bone using SPM12.31 In addition, regions of interest (ROIs) were manually drawn to delineate soft tissue in veins (V), arteries (A), and tumor tissue (Gb for glioblastoma) as well as CSF in lateral ventricles (LV).

2.2.1. Dynamic Z-spectra

Using a custom script written in Python Z-spectra as a function of B1 (0.6 μT, 1.2 μT and 1.6 μT) were derived using the Bloch–McConnell equations for 10 Gaussian-shaped saturation pulses (total saturation time of tSat = 1 s) and a field strength of 3 T. It was assumed that five protons exchange with bulk water (three protons at 1.2 ppm and one proton each at 2.2 ppm and 2.8 ppm, respectively) and that magnetization transfer contrast (MTC) originated from the semisolid macromolecular protons. Six types of tissue were used in the simulations, i.e. arterial blood, venous blood, GM, WM, CSF and tumor. The change in D-glucose concentration was assumed to be 10 mM in arterial and venous blood and ¼ of this concentration change was assumed in GM and WM tissue.32,33 These levels were added to the typical baseline concentration of 5 mM in vessels and ¼ of that in GM and WM. To obtain the Z-spectra in CSF we assumed that the D-glucose concentration in CSF was 60 % of that in arterial blood.34 For the Z-spectra in the tumor we used a vessel volume of 8 % since angiogenesis will result in a larger vessel volume than in normal tissue.35 Together with the extravascular extracellular space (EES), which can account for as much as 25 % of the tumor volume, one can assume that the D-glucose increase is 33 % of 10 mM in the tumor.36 In addition to the CEST effect, the presence of hydroxyl protons that are chemically shifted induces an exchange-related transverse relaxation component that can be described by the Swift-Connick equation, giving a transverse relaxivity r2ex = 0.012 s−1 mM−1.18 Tumors show an increased lactate production (i.e. the Warburg effect)37 leading to a lower pH in the EES and thereby a change in the exchange rates of the hydroxyl protons. For the tumor, we assumed an EES pH of 6.8 leading to a reduction in the exchange rates from k1.2 ppm = 4000 Hz to 3100 Hz, k2.2 ppm = 8000 Hz to 2500 Hz, k2.8 ppm = 10000 Hz to 6300 Hz approaching the intermediate exchange regime.38 Taking this pH into account, we also have the tumor D-glucose relaxivity increased to r2ex = 0.014 s−1 mM−1. Protons on the semi-solid macromolecules resulting in an MTC effect were assumed to have concentrations of 5500 mM and 15400 mM for GM and WM, respectively. For the tumor, a concentration of 2750 mM was assumed because of its location between GM and WM and having a higher water content.

Time-dependent glucoCEST uptake curves were simulated, based on the overall shape of DGE response curves previously observed at 3 T.13,20,32,39 A dynamic time series with an acquisition duration of 690 s and a temporal resolution of 5 s was simulated. The time series consisted of S0 (non-saturated signal) data acquisition (20 s), baseline (20–140 s), D-glucose infusion (140–380 s), signal increase and decay (380–565 s). The previously simulated Z-spectra as a function of D-glucose concentration were used to create dynamic Z-spectra under the influence of D-glucose infusion. The tissue specific dynamic Z-spectra were combined with the corresponding tissue segmentations, to obtain dynamic Z-spectral maps.

2.2.2. Motion parameters

Using Elastix,40,41 rigid body motion estimates were derived from all five participants in the glucoCEST DGE study using the first saturated image SSat as positional reference. Two patients were selected to represent two different motion patterns, i.e. a least severe pattern (LSP) of movement and a most severe pattern (MSP) of movement, within this group of subjects (Figs. 2AD). Three additional motion patterns were created by modifying the MSP: (1) 2×MSP where all rigid motions were multiplied by a factor of 2 (Figure S1.B), (2) 2×MSP where only rotations were included (Figure S1.C), (3) 2×MSP where only translations were included (Figure S1.D). An arbitrary motion pattern was also created to investigate the effect of non-continuous motion (Figure S1.E). Thus, a total of six different motion patterns were generated.

Figure 2.

Figure 2

Parameters used for rigid motion simulation (A-D): Rigid motion estimates chosen from an experimental DGE study including four patients and one volunteer. Least severe pattern (LSP: A, C) and most severe pattern (MSP: B, D) cases were selected to represent two extreme conditions with regards to magnitude. The first SSat image was assigned as positional reference. (A, B): Translation time series; (C, D) rotation time series. (E, F): Two examples of ventricular variation. Scale coefficient for affine transformation of LV volume oscillatory changes over time due to cardiac pulsation and LV volume increase after infusion (with and without infusion) related to glucose uptake.

Additional motions exists because LVs undergo an oscillating volume change of approximately 1 % due to cardiac pulsation, while D-glucose loading leads to ventricular dilatation by about 2.4 %.20,2527 Therefore, non-rigid LV movement was simulated by adding an affine transformation with a scaling coefficient changing over time to the LV segmentation. The scaling coefficient was varied randomly to mimic undersampled pulsation patterns due to, for example, cardiac pulsation. This is superimposed on a linear increase from the subject specific start of infusion, simulating an increasing D-glucose uptake (Figs. 2 EF). Increases in LV volume were offset by reductions in WM volume and vice versa.

2.2.3. Dynamic B0-changes

The previously described segmentations were used to create a model incorporating the volume magnetic susceptibility of air (χ ≈ 0.36 ppm)42 in the paranasal cavities and head surroundings, brain tissue and fluids (χ ≈ −9.05 ppm)43 surrounded by the skull (χ ≈ −11.56 ppm)44, and fat/skin (χ ≈ −8.50 ppm)45,46. Smaller inter-tissue susceptibility differences (ranging from about −0.05 to 0.25 ppm)47 were neglected under the assumption that a partial volume effect of a few percent will not affect the frequency shift substantially. The tissue contributions to the magnetic field were then numerically estimated by convolution with a magnetic dipole kernel48,49, after expanding the field of view (FOV) by a factor of two (zero-padding) in each dimension to increase computational accuracy.50 The effect of a static field in form of zero and first order static shims was obtained by using the first volume in the series of the simulated B0-field, and then adding these shim fields to retrieve a final B0 map. In this case of motion, the rigid body motion parameters were then applied to the susceptibility model with the Matlab (The MathWorks, Natick, MA, USA) function imwarp using a rigid body 3D transformation and cubic interpolation, creating a time series with reassigned fields in each voxel for each time point. At each time point, the original shims were added to produce the final motion-corrupted B0-field time series. Finally, to return to the initial frame of reference for Z-spectral analysis, the inverse of the previously applied rigid body motion was applied. Subtracting the map for the first time point scan (baseline) from the individual maps for each time point in the time series then provided the field change per time point, thus providing insight into possible B0-changes that shift the Z-spectrum due to motion.

2.2.4. Generation of DGE time series affected by motion

Figure 1 illustrates how the effects of the dynamic B0-changes were applied to the dynamic Z-spectra for each voxel leading to a shift along the Z-spectral frequency axis. An offset of 2 ppm and a B1 of 1.6 μT were selected to create a single offset time series of SSat(t). Afterwards, the rigid body motion parameters and non-rigid LV movement were applied with the Matlab function imwarp using cubic interpolation, and the DGE image time series affected by movement was down-sampled spatially to 2×2×3 mm3. Smoothing using a Gaussian kernel was applied to obtain an image quality similar to in vivo conditions. Rician noise with a standard deviation of 0.05% was added to the magnitude signal. In the normalized signal (S/S0), this corresponds to 0.10 % standard deviation. As the effect size in DGE is about 0.50% in terms of normalized signal, 0.10% noise corresponds to a noise level of about 20% in the DGE signal change.

2.3. Numerical phantom experiments and analysis

The ground truth was set to be an experiment with D-glucose infusion, without motion, and thus no dynamic B0-field changes, but a static B0-field combined with a first order shim, with a total duration of 565 s. Subsequently, experiments with and without infusion, and with six rigid head motion patterns (LSP, MSP, 2×MSP, 2×MSP only rotations, 2×MSP only translations and arbitrary motion, summarized in Fig. 2 and Supporting Information Figures S1.AE)) including ventricular pulsation, and with and without ventricular expansion, were simulated while keeping the zero and first order static shim constant, i.e. adding it to each time point after motion occurred (Fig. 1B).

2.3.1. Retrospective motion correction

Retrospective motion correction was applied to the simulations using Elastix.40,41 The details of the parameter settings are provided in the Supporting Information document. The first SSat image was assigned as positional reference.

2.3.2. DGE maps and statistics

The DGE signal SDGE(t) was calculated using the following equation

SDGE(t)=SbaseSSat(t)S0, (1)

where Sbase is the average baseline signal (over 24 time points in this study), SSat(t) is the saturated signal at time t and S0 is the average non-saturated signal (over 4 time-points in this study). A positive SDGE value can be interpreted as an increased D-glucose concentration. ROIs were drawn in tumor, WM, GM and CSF to extract DGE signal curves before and after rigid motion correction. Maps of the mean area under curve AUCmean were calculated using:

AUCmean=0tSDGE(t)N, (2)

where N is the number of time points over a selected time interval. Dividing by N normalizes the signal magnitude over the number of dynamics used during a certain time interval. AUCmean maps were derived from the DGE maps for the baseline time interval (120 s, N = 24), the time interval during infusion (0–240 s, N = 48) and two post-infusion time intervals (240–305 s, N = 13 and 305–425 s, N = 24).

For WM and tumor tissue, the change in DGE AUCmean signal over the time frame 240–305 s compared to the time frame 0–240 s was evaluated using the two-sided Wilcoxon signed-rank test. This statistical test was performed for the motion patterns “ground truth”, “LSP” and “MSP”, with and without motion correction (for LSP and MSP), and with and without D-glucose infusion.

3. RESULTS

The experimentally measured dynamic time series for LSP (Figs. 2A, 2C) and MSP (Figs. 2B, 2D) show an overall continuously increasing displacement over time with random fluctuations. Larger motion is seen from the start of the glucose infusion initiating a diverging motion pattern with superimposed sudden jumps. Especially the negative pitch in MSP (Fig. 2D) shows that the head falls back into the head coil, with a large magnitude. For the LV, the scale coefficients in the two examples in Figs. 2E, F show a steady increase after the start of the glucose infusion, simulating a volume dilatation of up to approximately 2% in the LV. The curve is superimposed with a randomly varying constant interval simulating an oscillatory expansion and reduction of approximately 1 % in the volume of the LV due to, for example, cardiac pulsation, but not limited to this. Oscillatory deformation of LVs is complex and cannot be fully captured with the low temporal resolution (5 s) in this phantom study.27

The simulations illustrated in Figure 3 define the ground truth. Baseline Z-spectra for the tissues are shown at three different B1 values in the top row of Figure 3A, while the differences between spectra before and after infusion are shown in the bottom row. The estimated DGE signals at an offset frequency of 2 ppm and a B1 of 1.6 μT are SDGE,A = 1.11 %, SDGE,V = 0.51 %, SDGE,GM = 0.28 %, SDGE,WM = 0.06 %, SDGE,CSF = 0.90 % and SDGE,Gb = 0.38 %. The simulated normalized signal curves for the different tissues are shown in Figure 3B (left) and the resulting DGE signal in 3B (right). About one minute after the end of the infusion, the DGE signal reached a peak followed by a slow steady decrease. Anatomical reference images for an axial slice in the mid-section of the brain, including the MPRAGE image used for segmentation, are shown in Figure 3C. Gadolinium-enhanced MPRAGE, FLAIR, and segmented images are added for visual comparison. Applying the signal curves to each separate tissue segmentation created the DGE images in Figure 3D, visualized in form of AUCmean maps for 4 different time intervals. The AUCmean map series starts with the baseline, showing no signal change, followed by time intervals after D-glucose infusion showing the individual tissue responses. In particular, changes in CSF and tumor are clearly visible.

Figure 3.

Figure 3

(A) Top row: Simulated Z-spectra at three B1 values (0.6 μT, 1.2 μT and 1.6 μT, tSat = 1 s) before D-glucose infusion, using a baseline of 5 mM D-glucose. Bottom row: Spectral differences between before and after infusion. (B) Left: Simulated DGE signal curves for veins, arteries, tumor, grey matter, white matter and CSF, including baseline, infusion and post-infusion periods. Right: The corresponding resulting DGE curves. (C): From left to right: MPRAGE image, Gd-enhanced MPRAGE image, FLAIR image, segmented image. (D): AUCmean-maps defining the ground truth.

AUCmean maps for the ground truth and different cases without/with infusion and before/after motion correction are shown for the different motion patterns in Figures 4A, 5A, and Supporting Information Figures S4, S6, S8, S10. The ground truth shows a clear dynamic tissue response (Fig. 4A, Fig. 5A). For all motion patterns without infusion, added motion clearly affects the AUCmean-maps by introducing artefacts (pseudo-DGE contrast) seen as hyper- and hypo-intensities at tissue interfaces such as the tumor rim. Infusion tends to highlight especially CSF over other tissues, while a small signal increase is seen within the motion-induced signal pattern in the tumor rim. These changes are amplified with increasing motion (cf. Supporting Information Fig. S4), showing increased pseudo-DGE effects, again at tissue interfaces such as the tumor rim, and especially some hyper- and hypointense signal changes in the cortex and within and around the ventricles. The AUCmean-maps show that motion correction reduced the overall appearance of pseudo-DGE effects but could also introduce new effects, especially in regions with many tissue interfaces such as GM and the lateral ventricles (Figs. 4A, 5A, and Supporting Information Figures S4, S6, S8, S10). For all motion patterns, the tumor rim enhancement was reduced when motion correction was applied and the tumor itself became visible after infusion, both with and without motion correction.

Figure 4.

Figure 4

(A): Illustration of LSP AUCmean map evolution separated in time intervals of baseline, during infusion (0–240 s), early post-infusion (240–305 s) and late post-infusion (305–425 s), with and without infusion before motion correction (NoMoco) and after motion correction (Moco). Ground truth with infusion and without motion together with the AUCmean intensity scale is shown below. (B): Extracted dynamic response curves, based on manually drawn ROIs in white matter (yellow in C) and tumor tissue (purple in C) before motion correction and after motion correction using Elastix, in comparison to the ground truth with and without infusion.

Figure 5.

Figure 5

(A): Illustration of MSP AUCmean map evolution separated in time intervals of baseline, during infusion (0–240 s), early post-infusion (240–305 s) and late post-infusion (305–425 s), with and without infusion before motion correction (NoMoco) and after motion correction (Moco). Ground truth with infusion and without motion together with the AUCmean intensity scale is shown below. (B): Extracted dynamic response curves, based on manually drawn ROIs in white matter (yellow in C) and tumor tissue (purple in C) before motion correction and after motion correction using Elastix, in comparison to the ground truth with and without infusion.

The extracted dynamic response curves for the ROIs, both before and after motion correction, are displayed in Figures 4B, 5B, and Supporting Information Figures S5, S7, S9, S11. Without infusion, the DGE signal in the tumor showed a pattern (Supporting Information Figs. S9.E1E2) similar to the purely translational motion (cf. Supporting Information Fig. S1.D). For the motion pattern using only rotations (Supporting Information Figs. S7.E1E2), there were hardly any pseudo-DGE effects visible in the tumor. In particular, pseudo-DGE effects were not present in WM and the resulting DGE signal change in WM was within the noise fluctuations. However, GM and CSF showed clear pseudo-DGE effects for all motion scenarios, except in CSF when only rotations were applied (cf. Supporting Information Figs. S7.H1H4). Adding infusion led to increased DGE signal, and motion correction was able to decrease the DGE signal in tumor and WM to levels closer to the ground truth. This can be seen, for example, in the MSP case (Fig. 5B), where in tumor tissue, the situation without infusion shows an erroneous large positive signal. After motion correction, this signal is reduced but still higher than in the ground truth.

The effect of only B0-changes on the DGE signal for LSP and MSP is illustrated in Figure 6. The AUCmean signal is shown in the transverse view for LSP (6AC) and MSP (6DF) without infusion and with infusion in the late post-infusion time interval (305–425 s). The dynamic B0-changes (Figs. 6A,D) induced a gradual spatial intensity change on the AUCmean DGE signal (Figs. 6B,E), that gave decreases as large as 0.5% in the anterior and posterior/inferior brain for the MSP case, but an increase in the tumor region in the middle of the brain. Infusion caused a dominating intensity effect for CSF, but the B0-induced DGE signal variations contributed in the MSP case. The ground truth with infusion in the late post-infusion time interval (305–425 s) is shown in Fig. 6G.

Figure 6.

Figure 6

(A, D): Average of dynamic B0-changes without tissue mixing under conditions of a static first order shim in the transversal slice orientation for LSP and MSP, in the late post-infusion interval (305–425 s). (B, E): AUCmean map derived from DGE maps without infusion illustrating the influence of isolated B0-changes. (C, F): AUCmean map derived from DGE maps with infusion und the influence of isolated dynamic B0-changes. (G): Ground truth with infusion in the same time interval.

Box-and-whisker plots of the AUCmean signal of all voxels in the previously defined ROIs in tumor tissue and WM are shown in Figures 7 and 8, respectively. Three cases are shown for LSP and MSP: i) only dynamic B0-changes without tissue mixing, ii) B0-changes and movement without motion correction and iii) B0-changes and movement followed by motion correction, with and without infusion. Tumor tissue without infusion, influenced only by B0-changes, showed slight hyperintense signal for MSP (cf. Fig. 7). Addition of the infusion led to a signal change higher than the ground truth. Motion led to hyperintensities, which were increased after motion correction in MSP. These effects were enhanced by infusion. Motion led to a wider quartile range and larger total range, with a larger variability for MSP. Motion correction led to a slightly wider quartile range, and the median signal after infusion was higher for MSP than for LSP and for ground truth. White matter influenced only by dynamic B0-changes showed hyperintensities for the MSP case without infusion, and adding infusion led to a larger signal intensity increase compared to the ground truth. Comparing the AUCmean DGE signal before and after motion correction showed no significant differences, but the signal trended to be larger than the ground truth due to B0-changes (Fig. 8, bottom, Fig. 5B).

Figure 7.

Figure 7

Box-and-whisker plots showing the AUCmean DGE signal for tumor tissue (purple ROI in Figure 4C) under the influence of motion patterns LSP and MSP, with and without infusion. Top row: Effect of only dynamic B0-changes without tissue mixing (in the presence of a static first order shim) on the DGE signal without motion correction (NoMoco) (cf. Figs. 6B, 6C, 6E, 6F). Middle row: Effect of dynamic B0-changes (in the presence of a static first order shim) with tissue mixing and without motion correction (NoMoco) (cf. Figs. 4B, 5B). Bottom row: Dynamic B0-changes (in the presence of a static first order shim) with tissue mixing and with motion correction (Moco) (cf. Figs. 4B, 5B). The ground truth with infusion for tumor is shown separately beneath. The upper and lower borders of the blue box indicate the upper and lower quartile, respectively, the red line shows the median value, and the whiskers represent the maximum and minimum AUCmean values. The ‘+’ denotes outliers, defined as values that are more than 1.5 times the interquartile range away from the bottom or top of the box.

Figure 8.

Figure 8

Box-and-whisker plots showing the AUCmean DGE signal for white matter (yellow ROI in Figure 4C) under the influence of motion patterns LSP and MSP, with and without infusion. Top row: Effect of only dynamic B0-changes without tissue mixing (in the presence of a static first order shim) on the DGE signal without motion correction (NoMoco) (cf. Figs. 6B, 6C, 6E, 6F). Middle row: Effect of dynamic B0-changes (in the presence of a static first order shim) with tissue mixing and without motion correction (NoMoco) (cf. Figs. 4B, 5B). Bottom row: Dynamic B0-changes (in the presence of a static first order shim) with tissue mixing and with motion correction (Moco) (cf. Figs. 4B, 5B). The ground truth with infusion for white matter is shown separately beneath. The upper and lower borders of the blue box indicate the upper and lower quartile, respectively, the red line shows the median value, and the whiskers represent the maximum and minimum AUCmean values. The ‘+’ denotes outliers, defined as values that are more than 1.5 times the interquartile range away from the bottom or top of the box.

The P-values from the statistical test for LSP and MSP are listed for tumor tissue (Table 1) and for WM (Table 2). Both tissues showed significant DGE AUCmean signal change when infusion was added in all cases (ground truth, with/without motion). In tumor tissue, DGE AUCmean signal changes were non-significant without infusion for LSP before and after motion correction. For white matter, DGE AUCmean signal changes were non-significant without infusion before and after correction. For both tissue types, DGE AUCmean changes for MSP were significant without infusion and before and after motion correction.

Table 1.

P-values for AUCmean DGE signal change between the time intervals 0–240 s and 240–305 s in tumor tissue (for ROI colored in purple in Figure 4C). NoMoco - before motion correction; Moco - motion correction; LSP - least severe pattern; MSP - most severe pattern

Ground Truth LSP MSP
Without
infusion
With
infusion
Without
infusion
With
infusion
Without
infusion
With
infusion
NoMoco 0.82 * 0.74 * * *
Moco N/C N/C 0.44 * * *

The asterisk (*) indicates P< 0.05. N/C = not considered

Table 2.

P-values for AUCmean DGE signal change between the time intervals 0–240 s and 240–305 s in white matter (for ROI colored in yellow in Figure 4C). NoMoco - before motion correction; Moco - motion correction; LSP - least severe pattern; MSP - most severe pattern

Ground Truth LSP MSP
Without
infusion
With
infusion
Without
infusion
With
infusion
Without
infusion
With
infusion
NoMoco 0.78 * 0.47 * * *
Moco N/C N/C 0.57 * * *

The asterisk (*) indicates P < 0.05. N/C = not considered

4. DISCUSSION

In order to understand the influence of motion on DGE data, we developed a numerical human brain phantom. Availability of such a ground truth that takes only the DGE effect into consideration, without any motion, opens up possibilities for testing and validation of different motion patterns and correction and analysis approaches to account for these.

The introduced DGE response curve shapes were based on experimental glucoCEST time series at 3 T and 7 T.13,20,39 Experimental DGE response curves reported in the literature usually show high variability in curve shape between different scans and tissues, due to low contrast-to-noise ratio and artefacts mainly related to motion. In this study, response curves were simulated with peak signal based on the results of Bloch-McConnell simulations at 3 T, giving a sufficient approximation of experimentally measured DGE signal changes due to the glucoCEST effect.32 Inclusion of a full description of the kinetics for D-glucose uptake for the human brain would be required for a more realistic approach in future studies.51 The availability of a numerical human brain phantom is likely to increase the understanding of the effect of motion during a DGE experiment, and an actual Bloch-simulator may enable addition of additional MR-related artefacts, for example, intra-volume motion-induced blurring.52,53 On the other hand, bulk head motion, which could be simulated in the present approach, can be considered the main source of intensity artefacts, as shown, for example, in Figures 4 and 5.

Even though the number of participants was small, the LSP and MSP motion patterns (Fig. 2) are likely to represent rigid motion occurring during a DGE experiment with a slow infusion duration of approximately 240 s. We interpret that the slowly varying characteristics originate from the involuntary relaxation of the subject’s neck muscles, leading to a translation in Z-direction and a pitch rotation similar to nodding.5355 Ventricular volume variations are caused by cardiac pulsation and loading of D-glucose20,25,26 These effects were simulated together in Figures 4 and 5 for cases with and without glucose infusion, leading to signal variations similar to those reported in vivo at 3 T (0.25 % - 1.5 % in tumors).20,28,29,32 Note that development of new pulse sequences56 or use of other sugars such as 3-ortho-methylglucose (3OMG)57,58 or D-glucosamine (D-GlcN)59 is expected to increase these effects. The example motion patterns showed different amounts and types of movement, and different motion scenarios lead to individual pseudo-DGE patterns. Larger motion changes led generally to stronger pseudo-DGE effects, as demonstrated when comparing LSP, MSP and 2xMSP in the Supporting Information. In addition, the resulting dynamic response curve depended on the pattern of the motion. This could be confirmed using the arbitrary motion pattern demonstrated in Supporting Information Figure S11. In this case, we did not observe the typical dynamic response obtained using the other motion patterns. This observation is similar to what has been described previously in the fMRI literature, i.e. motion may mimic activated voxels when the motion pattern is correlated with the stimulus.60,61

It has previously been reported that B0-changes due to motion can affect the detection of DGE signal changes.20,28 The reported variations are around 0.03 ppm for 1 mm translations and 1 degree rotations at 3 T. These effects could be reproduced, inducing global signal variations across tissue boundaries. The images in Figure 6, and Supporting Information Figures S2, S4, S6, S8, S10 demonstrate that the amount of motion and the brain location govern where erroneous signal occurs. Dominant positive field alterations could be found in the anterior and posterior parts, and negative field changes were observed in the mid-section of the brain, corresponding to the location of some air-filled cavities and the particular motion pattern. We previously recommended using an offset frequency at 2 ppm and a B1 of 1.6 μT at 3 T.32 Saturation closer to water (as commonly done at 1.2 ppm) has a higher glucoCEST effect, even for a B1 of 0.6 μT, as expected based on the merged water and D-glucose resonances in this intermediate to fast exchange regime. However, it will also introduce more pseudo-DGE signal changes, especially those caused by dynamic B0-changes, because the signal slope in the Z-spectrum is steeper when closer to the direct water saturation (Fig. 3 A). One approach to correct for B0 effects may be dynamic shimming62,63, which would be technically simple to add to the phantom. However, this approach is complex in practice in vivo, where dynamic shimming of each time point may not help because shimming is global. As a consequence, it is unclear if the field in the voxel after the new shimming will be the same as before the motion (i.e., shimming is just improved summed over the whole brain unless one would do slice-by-slice shimming) and the Z-spectrum can therefore still be shifted. In addition, motion correction would then be applied to the brain following the shim correction, which may again change the average field shift in many of the voxels. The easiest way for field correction would be acquisition of the Z-spectrum as a function of time, which would allow the field to be corrected by simply shifting the spectrum at each time point, which is our goal for DGE in the future.

DGE contrast was slightly improved (closer to the ground truth) after motion correction. However, residual motion artefacts, especially at the grey matter/CSF tissue interfaces, were still prominent both with and without glucose infusion. The correction algorithm seemed to introduce both hypo- and hyperintensities in the LVs due to their non-rigid motion characteristics. The signal in the tumor tissue was small and strongly affected by motion, but the ring enhancement at the tissue interface (tumor border) could be reduced by motion correction. Fig. 7 (top row) shows that a small additional hyperintensity was introduced by the B0 shift for the MSP case. Without motion correction (Fig. 7 middle row), tissue mixing leads to additional increases or decreases in signal, spreading the range of pseudo-DGE effects. The different motion patterns also confirmed that pseudo-DGE effects can originate both from motion (tissue mixing) and by applying motion correction. These effects originate from regions where tissue interfaces exist, and a large contrast is seen. This is similar to what has been observed in fMRI experiments, where motion correction can introduce false activiations.60,61 However, motion correction will have no effect on the signal changes introduced by B0 shifts. In our study, the simulated tumor was homogenous, while, in reality, a tumor is heterogeneous, which likely would lead to more pseudo-CEST effect due to the presence of more tissue interfaces.

Extracted response curves in tumor tissue, corrupted by motion, generally showed strong signal variations, and the pseudo-DGE effects using LSP and the different MSP movement patterns varied in the range up to 1 % signal change, in agreement with a previous study.28 The signal variation depended on how heterogeneous the structure was, and WM contrast was thus less affected. Furthermore, gradual small local field changes caused by B0-changes due to motion cannot be corrected when a single frequency point is acquired. In addition, static B0 inhomogeneities initially reduced by shimming will affect the overall contrast before and after motion. The negative pitch movement used in the different MSP movements (MSP, 2×MSP, 2×MSP only rotations), which can be interpreted as leaning backwards in the head coil, is often seen in practice due to relaxation of the neck muscles over long scanning durations. This situation leads to movement of the air-filled nasal sinuses towards the transversal slice of interest. Ultimately, fast acquisition of full Z-spectra is required to improve this situation. Contaminated dynamic contrast after retrospective motion correction is known from other fields of varying contrast MRI. We attribute these effects to smoothing and interpolation errors, and this problem has been reported in, for example, quantitative T1 mapping in a previous study.64

Aimed to be used for studying different scenarios of motion, the proposed phantom can be applied to study different regions of the brain individually. The modular simulation design enables the use of Z-spectra and curve shapes that reflect different tissues and effect sizes (e.g., due to different concentrations of infusions). DGE MRI using D-glucose is only one of several dynamic contrast modalities that can be simulated using this simulation framework. Modalities such as dynamic susceptibility contrast MRI4,65, dynamic contrast enhanced MRI5,66 and dynamic CEST imaging using other sugars, such as 3OMG57,58 , D-GlcN59, and dextran67,68 that have different uptake curves, can also be assessed using this phantom.

5. CONCLUSIONS

A numerical human brain phantom simulating a DGE experiment in combination with rigid-head movement and dynamic LV dilatation was developed. Motion-induced pseudo-DGE effects were reduced in some areas and increased in other areas by rigid motion correction. This was especially observed at tissue interfaces, where the induced pseudo-CEST effects were tangled with true CEST contrast complicating image interpretation. B0-shifts due to dynamic field changes as well as added shim settings for different positions after motion affected the spectral intensities. These effects cannot be corrected by rigid motion correction in DGE without acquiring large parts of the Z-spectra including the water proton frequency. Ultimately, advanced post-processing methods using retrospective motion correction, and perhaps dynamic B0-measurements and motion corrections using dual navigators,69 are needed when implementing the DGE technique into the clinical practice. The phantom can be used for testing these approaches for dynamic CEST experiments, as well as for other types of dynamic scans affected by motion, for example, functional MRI.

Supplementary Material

supinfo

Supporting Information Figure S1 Patterns used for rigid head motion simulation in the cases listed in Supporting Information Tables S1S5. (A): MSP. (B): 2×MSP; all motions in (A) are multiplied by a factor of 2. (C): 2×MSP, but only rotations. (D): 2×MSP, but only translations. (E): Arbitrary motion pattern. (F): Scale coefficient to simulate oscillatory ventricular volume changes over time due to cardiac pulsation and LV volume swelling after infusion (with and without infusion) related to glucose uptake.

Supporting Information Figure S2 AUCmean maps over the early post-infusion (240–305 s) interval for the first motion scenario (Supporting Information Figure S1.A: MSP) in Supporting Information Table S1. The different cases causing tissue mixing are (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction. Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth (infusion only). Case-IDs for each scenario (cf. Supporting Information Table S1) are placed in the top left corner of each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S3 Column-wise organized analyses for the first motion scenario (Supporting Information Figure S1.A: MSP and Supporting Information Table S1) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S4 AUCmean maps over the early post-infusion (240–305 s) interval for the second motion scenario (Supporting Information Figure S1.B: 2×MSP) in Supporting Information Table S2. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S2) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S5 Column-wise organized analyses for the second motion scenario (Supporting Information Figure S1.B: 2×MSP: all rigid head motions multiplied by a factor of 2) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S6 AUCmean maps over the early post-infusion (240–305 s) interval for the third motion scenario (Supporting Information Figure S1.C: 2 × MSP with only rotations) in Supporting Information Table S3. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S3) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S7 Column-wise organized analyses for the third motion scenario (Supporting Information Figure S1.C: 2×MSP with only rotations) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; ;(H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S8 AUCmean maps over the early post-infusion (240–305 s) interval for the fourth motion scenario (Supporting Information Figure S1.D: 2 × MSP with only translations) in Supporting Information Table S4. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S4) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S9 Column-wise organized analyses for the fourth motion scenario (Supporting Information Figure S1.D: 2×MSP with only translations) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S10 AUCmean maps over the early post-infusion (240–305 s) interval for the fifth motion scenario (Supporting Information Figure S1.E: arbitrary rotational and translational motions) in Supporting Information Table S5. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively; (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S5) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S11 Column-wise organized analyses for the fifth motion scenario (Supporting Information Figure S1.E: arbitrary rotational and translational motions) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Table S1 Cases evaluated for the first scenario (MSP) with respect to with/without infusion, with/without B0-changes as a consequence of the rigid head motion pattern, and with/without ventricular and rigid head motion separately and combined.

The case-ID has five digits with the following coding: (1–5: number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with rigid head motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S2 Scenarios evaluated for 2×MSP with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with rigid head motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S3 Scenarios evaluated for 2×MSP with only rotations with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S4 Scenarios evaluated for 2 × MSP with only translations with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S5 Scenarios evaluated for arbitrary rotational and translational head motions with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

ACKNOWLEDGMENTS

We acknowledge Dr. Akansha Sehgal (Johns Hopkins University and Kennedy Krieger Institute) for additional support and discussions. This project was supported by Swedish Research Council grants 2017-00995, 2019-01162 and 2019-03637. Swedish Cancer Society grants CAN 2018/468, CAN 2018/550, and 21 1652 Pj. Swedish Brain Foundation grant FO2017-0236, FO2021-0328, Regional Research Funding F2018/1490, and National Institutes of Health grants RO1 EB019934 and R00 EB026312.

Funding information

This project was supported by Swedish Research Council grants 2017-00995, 2019-01162 and 2019-03637. Swedish Cancer Society grants CAN 2018/468, CAN 2018/550, and 21 1652 Pj. Swedish Brain Foundation grant FO2017-0236, FO2021-0328, Regional Research Funding F2018/1490, and National Institutes of Health grants RO1 EB019934 and R00 EB026312.

CONFLICT OF INTEREST

Under a license agreement between Philips and the Johns Hopkins University, Dr. Knutsson’s spouse, Dr. van Zijl, and the University are entitled to fees related to an imaging device used in the study discussed in this publication. Dr. van Zijl also is a paid lecturer for Philips. This arrangement has been reviewed and approved by the Johns Hopkins University in accordance with its conflict of interest policies. Mads Andersen is employed by Philips Denmark A/S. Frederik Testud is employed by Siemens Healthcare AB, Sweden. Patrick Liebig is employed by Siemens Healthcare GmbH, Germany.

Abbreviations

AUC

area under curve

CEST

chemical exchange saturation transfer

CSF

cerebrospinal fluid

DGE

dynamic glucose-enhanced

DSC

dynamic susceptibility contrast

MTR

magnetization transfer ratio

PVC

peripheral venous catheter

WM

white matter

Footnotes

Supporting Information

Additional supporting information may be found in the Supporting Information section.

Data availability statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1.van Dijken BRJ, van Laar PJ, Smits M, Dankbaar JW, Enting RH, van der Hoorn A. Perfusion MRI in treatment evaluation of glioblastomas: Clinical relevance of current and future techniques. Journal of Magnetic Resonance Imaging. 2019;49:11–22. doi: 10.1002/jmri.26306 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Campbell BCV, Khatri P. Stroke. The Lancet. 2020;396(10244):129–142. doi: 10.1016/S0140-6736(20)31179-X [DOI] [PubMed] [Google Scholar]
  • 3.Leigh R, Knutsson L, Zhou J, van Zijl PCM. Imaging the physiological evolution of the ischemic penumbra in acute ischemic stroke. Journal of Cerebral Blood Flow and Metabolism. 2018;38(9):1500–1516. doi: 10.1177/0271678X17700913 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Knutsson L, Lindgren E, Ahlgren A, et al. Dynamic susceptibility contrast MRI with a prebolus contrast agent administration design for improved absolute quantification of perfusion. Magn Reson Med. 2014;72(4):996–1006. doi: 10.1002/mrm.25006 [DOI] [PubMed] [Google Scholar]
  • 5.Wirestam R. Using contrast agents to obtain maps of regional perfusion and capillary wall permeability. Imaging Med. 2012;4(4):423–442. doi: 10.2217/iim.12.24 [DOI] [Google Scholar]
  • 6.Errante Y, Cirimele V, Mallio CA, di Lazzaro V, Zobel BB, Quattrocchi CC. Progressive increase of T1 signal intensity of the dentate nucleus on unenhanced magnetic resonance images is associated with cumulative doses of intravenously administered gadodiamide in patients with normal renal function, suggesting dechelation. Invest Radiol. 2014;49(10):685–690. doi: 10.1097/RLI.0000000000000072 [DOI] [PubMed] [Google Scholar]
  • 7.Radbruch A, Weberling LD, Kieslich PJ, et al. Gadolinium retention in the dentate nucleus and globus pallidus is dependent on the class of contrast agent. Radiology. 2015;275(3):783–791. doi: 10.1148/radiol.2015150337 [DOI] [PubMed] [Google Scholar]
  • 8.Kanda T, Fukusato T, Matsuda M, et al. Gadolinium-based contrast agent accumulates in the brain even in subjects without severe renal dysfunction: Evaluation of autopsy brain specimens with inductively coupled plasma mass spectroscopy. Radiology. 2015;276(1):228–232. doi: 10.1148/radiol.2015142690 [DOI] [PubMed] [Google Scholar]
  • 9.Robert P, Lehericy S, Grand S, et al. T1-Weighted Hypersignal in the Deep Cerebellar Nuclei after Repeated Administrations of Gadolinium-Based Contrast Agents in Healthy Rats: Difference between Linear and Macrocyclic Agents. Invest Radiol. 2015;50(8):473–480. doi: 10.1097/RLI.0000000000000181 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Marckmann P, Skov L, Rossen K, et al. Nephrogenic systemic fibrosis: Suspected causative role of gadodiamide used for contrast-enhanced magnetic resonance imaging. Journal of the American Society of Nephrology. 2006;17(9):2359–2362. doi: 10.1681/ASN.2006060601 [DOI] [PubMed] [Google Scholar]
  • 11.Grobner T, Prischl FC. Gadolinium and nephrogenic systemic fibrosis. Kidney Int. 2007;72(3):260–264. doi: 10.1038/sj.ki.5002338 [DOI] [PubMed] [Google Scholar]
  • 12.Chan KWY, McMahon MT, Kato Y, et al. Natural D -glucose as a biodegradable MRI contrast agent for detecting cancer. Magn Reson Med. 2012;68:1764–1773. doi: 10.1002/mrm.24520 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Xu X, Yadav NN, Knutsson L, et al. Dynamic Glucose-Enhanced (DGE) MRI: Translation to Human Scanning and First Results in Glioma Patients. Tomography. 2015;1(2):105–114. doi: 10.18383/j.tom.2015.00175 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Xu X, Chan KWY, Knutsson L, et al. Dynamic glucose enhanced (DGE) MRI for combined imaging of blood-brain barrier break down and increased blood volume in brain cancer. Magn Reson Med. 2015;74(4):1556–1563. doi: 10.1002/mrm.25995 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Paech D, Schuenke P, Windschuh J, et al. T1ρ-weighted Dynamic Glucose-enhanced MR Imaging in the Human Brain. Radiology. 2017. Dec;285(3):914–922. doi: 10.1148/radiol.2017162351 [DOI] [PubMed] [Google Scholar]
  • 16.Schuenke P, Paech D, Koehler C, et al. Fast and Quantitative T1p-weighted Dynamic Glucose Enhanced MRI. Sci Rep. 2017;7:1–10. doi: 10.1038/srep42093 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Huang J, van Zijl PCM, Han X, et al. Altered d-glucose in brain parenchyma and cerebrospinal fluid of early Alzheimer’s disease detected by dynamic glucose-enhanced MRI. Sci Adv 2020;6:1–10. doi: 10.1126/sciadv.aba3884 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Yadav NN, Xu J, Bar-Shir A, et al. Natural D-glucose as a biodegradable MRI relaxation agent. Magn Reson Med. 2014;72:823–828. doi: 10.1002/mrm.25329 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Walker-Samuel S, Ramasawmy R, Torrealdea F, et al. In vivo imaging of glucose uptake and metabolism in tumors. Nat Med. 2013;19(8):1067–1073. doi: 10.1038/nm.3252 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Seidemo A, Lehmann PM, Rydhög A, et al. Towards robust glucose chemical exchange saturation transfer imaging in humans at 3 T : Arterial input function measurements and the effects of infusion time. NMR Biomed. 2021;e4624:1–13. doi: 10.1002/nbm.4624 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Knutsson L, Xu X, Zijl PCM, Chan KWY. Imaging of sugar-based contrast agents using their hydroxyl proton exchange properties. NMR Biomed. 2022;(May):1–25. doi: 10.1002/nbm.4784 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ward HA, Riederer SJ, Jack CR. Real-time autoshimming for echo planar timecourse imaging. Magn Reson Med. 2002;48:771–780. doi: 10.1002/mrm.10259 [DOI] [PubMed] [Google Scholar]
  • 23.Foerster BU, Tomasi D, Caparelli EC. Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging. Magn Reson Med. 2005;54(5):1261–1267. doi: 10.1002/mrm.20695 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Benner T, van der Kouwe AJW, Kirsch JE, Sorensen AG. Real-time RF pulse adjustment for Bo drift correction. Magn Reson Med. 2006;56(1):204–209. doi: 10.1002/mrm.20936 [DOI] [PubMed] [Google Scholar]
  • 25.Puri BK, Lewis HJ, Saeed N, Davey NJ. Volumetric change of the lateral ventricles in the human brain following glucose loading. Exp Physiol. 1999;84(1):223–226. doi: 10.1111/j.1469-445X.1999.tb00085.x [DOI] [PubMed] [Google Scholar]
  • 26.Zhu DC, Xenos M, Linninger AA, Penn RD. Dynamics of lateral ventricle and cerebrospinal fluid in normal and hydrocephalic brains. Journal of Magnetic Resonance Imaging. 2006;24:756–770. doi: 10.1002/jmri.20679 [DOI] [PubMed] [Google Scholar]
  • 27.Kiviniemi V, Wang X, Korhonen V, et al. Ultra-fast magnetic resonance encephalography of physiological brain activity - Glymphatic pulsation mechanisms? Journal of Cerebral Blood Flow and Metabolism. 2016;36:1033–1045. doi: 10.1177/0271678X15622047 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Zaiss M, Herz K, Deshmane A, et al. Possible artifacts in dynamic CEST MRI due to motion and field alterations. Journal of Magnetic Resonance. 2019;298:16–22. doi: 10.1016/j.jmr.2018.11.002 [DOI] [PubMed] [Google Scholar]
  • 29.Herz K, Lindig T, Deshmane A, et al. T1ρ-based dynamic glucose-enhanced (DGEρ) MRI at 3 T: method development and early clinical experience in the human brain. Magn Reson Med. 2019;82(5):1832–1847. doi: 10.1002/mrm.27857 [DOI] [PubMed] [Google Scholar]
  • 30.Zaiss M, Ehses P, Scheffler K. Snapshot-CEST: Optimizing spiral-centric-reordered gradient echo acquisition for fast and robust 3D CEST MRI at 9.4 T. NMR Biomed. 2018;31(4):1–14. doi: 10.1002/nbm.3879 [DOI] [PubMed] [Google Scholar]
  • 31.Penny W, Friston K, Ashburner J, Kiebel S, Nichols T. Statistical Parametric Mapping: The Analysis of Functional Brain Images. Amsterdam; Boston, MA: Academic. 2006;1(eBook ISBN: 9780080466507). doi: 10.1111/j.1365-294X.2006.02813.x [DOI] [Google Scholar]
  • 32.Xu X, Sehgal AA, Yadav NN, et al. d-glucose weighted chemical exchange saturation transfer (glucoCEST)-based dynamic glucose enhanced (DGE) MRI at 3T: early experience in healthy volunteers and brain tumor patients. Magn Reson Med. 2020;84:247–262. doi: 10.1002/mrm.28124 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Gruetter R, Ugurbil K, Seaquist ER. Steady-state cerebral glucose concentrations and transport in the human brain. J Neurochem. 1998;70(1):397–408. doi: 10.1046/j.1471-4159.1998.70010397.x [DOI] [PubMed] [Google Scholar]
  • 34.Ross K Principles of neurologic infectious diseases. In: 1, ed. McGraw-Hill Education; - Europe; 2005:p 4. [Google Scholar]
  • 35.Lüdemann L, Grieger W, Wurm R, Wust P, Zimmer C. Glioma assessment using quantitative blood volume maps generated by T1-weighted dynamic contrast-enhanced magnetic resonance imaging: A receiver operating characteristic study. Acta radiol. 2006;47(3):303–310. doi: 10.1080/02841850500539033 [DOI] [PubMed] [Google Scholar]
  • 36.Mills SJ, Soh C, Rose CJ, et al. Candidate Biomarkers of Extravascular Extracellular Space: A Direct Comparison of Apparent Diffusion Coefficient and Dynamic Contrast-Enhanced MR Imaging-Derived Measurement of the Volume of the Extravascular Extracellular Space in Glioblastoma Multiforme. American Journal of Neuroradiology. 2010;31:549–53. doi: 10.3174/ajnr.A1844 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Warburg O. On the Origin of Cancer Cells. Science (1979). 1956;123(3191):309–314. [DOI] [PubMed] [Google Scholar]
  • 38.Zaiss M, Anemone A, Goerke S, et al. Quantification of hydroxyl exchange of D-Glucose at physiological conditions for optimization of glucoCEST MRI at 3, 7 and 9.4 Tesla. NMR Biomed. 2019;32(9):e4113. doi: 10.1002/nbm.4113 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Knutsson L, Seidemo A, Rydhög A, et al. Arterial Input Functions and Tissue Response Curves in Dynamic Glucose-Enhanced (DGE) Imaging: Comparison Between glucoCEST and Blood Glucose Sampling in Humans. Tomography. 2018;4(4):164–171. doi: 10.18383/j.tom.2018.00025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW. Elastix: A toolbox for intensity-based medical image registration. IEEE Trans Med Imaging. 2010;29(1):196–205. doi: 10.1109/TMI.2009.2035616 [DOI] [PubMed] [Google Scholar]
  • 41.Shamonin DP, Bron EE, Lelieveldt BPF, Smits M, Klein S, Staring M. Fast parallel image registration on CPU and GPU for diagnostic classification of Alzheimer’s disease. Front Neuroinform. 2014;7(50):1–15. doi: 10.3389/fninf.2013.00050 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys. 1996;23:815–850. doi: 10.1118/1.597854 [DOI] [PubMed] [Google Scholar]
  • 43.Collins CM, Yang B, Yang QX, Smith MB. Numerical calculations of the static magnetic field in three-dimensional multi-tissue models of the human head. Magn Reson Imaging. 2002;20:413–424. doi: 10.1016/S0730-725X(02)00507-6 [DOI] [PubMed] [Google Scholar]
  • 44.Hopkins JA, Wehrli FW. Magnetic susceptibility measurement of insoluble solids by NMR: Magnetic susceptibility of bone. Magn Reson Med. 1997;37(4):494–500. doi: 10.1002/mrm.1910370404 [DOI] [PubMed] [Google Scholar]
  • 45.Dimov AV, Liu T, Spincemaille P, et al. Joint estimation of chemical shift and quantitative susceptibility mapping (chemical QSM). Magn Reson Med. 2015;73:2100–2110. doi: 10.1002/mrm.25328 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Wei H, Gibbs E, Zhao P, et al. Susceptibility tensor imaging and tractography of collagen fibrils in the articular cartilage. Magn Reson Med. 2017;78:1683–1690. doi: 10.1002/mrm.26882 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Lim IAL, FA V, Li X, et al. Human brain atlas for automated region of interest selection in quantitative susceptibility mapping: application to determine iron content in deep gray matter structures. Neuroimage. 2013;82:449–496. doi: 10.1016/j.neuroimage.2013.05.127 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Marques JP, Bowtell R. Application of a fourier-based method for rapid calculation of field inhomogeneity due to spatial variation of magnetic susceptibility. Concepts Magn Reson Part B Magn Reson Eng. 2005;25B(1):65–78. doi: 10.1002/cmr.b.20034 [DOI] [Google Scholar]
  • 49.Salomir R, Senneville BDDE, Moonen CTW. A Fast Calculation Method for Magnetic Field Inhomogeneity due to an Arbitrary Distribution of Bulk Susceptibility. Concepts Magn Reson Part B Magn Reson Eng. 2003;19B(1):26–34. doi: 10.1002/cmr.b.10083 [DOI] [Google Scholar]
  • 50.Koch KM, Papademetris X, Rothman DL, de Graaf RA. Rapid calculations of susceptibility-induced magnetostatic field perturbations for in vivo magnetic resonance. Phys Med Biol. 2006;51:6381–6402. doi: 10.1088/0031-9155/51/24/007 [DOI] [PubMed] [Google Scholar]
  • 51.van Zijl PCM, Davis D, Eleff SM, Moonen CTW, Parker RJ, Strong JM. Determination of cerebral glucose transport and metabolic kinetics by dynamic MR spectroscopy. Am J Physiol Endocrinol Metab. 1997;273:E1216–E1227. doi: 10.1152/ajpendo.1997.273.6.e1216 [DOI] [PubMed] [Google Scholar]
  • 52.Godenschweger F, Kägebein U, Stucht D, et al. Motion correction in MRI of the brain. Phys Med Biol. 2016;61(5):32–56. doi: 10.1088/0031-9155/61/5/R32 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI: a complex problem with many partial solutions. Journal of Magnetic Resonance Imaging. 2015;42:887–901. doi: 10.1002/jmri.24850 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Hess AT, Alfaro-Almagro F, Andersson JL, Smith SM. Head movement in UK biobank, analysis of 42,874 fMRI motion logs. In Proceedings of the ISMRM Workshop on Motion Detection & Correction, Oxford, UK, 2022. [Google Scholar]
  • 55.Eichhorn H, Vascan AV, Nørgaard M, et al. Characterisation of Children’s Head Motion for Magnetic Resonance Imaging With and Without General Anaesthesia. Frontiers in Radiology. 2021;1(December):1–10. doi: 10.3389/fradi.2021.789632 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Xu X, Xu J, Chan KWY, et al. GlucoCEST imaging with on-resonance variable delay multiple pulse (onVDMP) MRI. Magn Reson Med. 2019;81(1):47–56. doi: 10.1002/mrm.27364 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Sehgal AA, Li Y, Lal B, et al. CEST MRI of 3-O-methyl-D-glucose uptake and accumulation in brain tumors. Magn Reson Med. 2019;81(3):1993–2000. doi: 10.1002/mrm.27489 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Rivlin M, Tsarfaty I, Navon G. Functional molecular imaging of tumors by chemical exchange saturation transfer MRI of 3-O-Methyl-DGlucose. Magn Reson Med. 2014;72(5):1375–1380. doi: 10.1002/mrm.25467 [DOI] [PubMed] [Google Scholar]
  • 59.Rivlin M, Navon G. Glucosamine and N-acetyl glucosamine as new CEST MRI agents for molecular imaging of tumors. Sci Rep. 2016;6(August):4–10. doi: 10.1038/srep32648 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Freire L, Mangin JF. Motion correction algorithms may create spurious brain activations in the absence of subject motion. Neuroimage. 2001;14(3):709–722. doi: 10.1006/nimg.2001.0869 [DOI] [PubMed] [Google Scholar]
  • 61.Yakupov R, Lei J, Hoffmann MB, Speck O. False fMRI activation after motion correction. Hum Brain Mapp. 2017;38:4497–4510. doi: 10.1002/hbm.23677 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Gruetter R. Automatic, localized in vivo adjustment of all first-and second-order shim coils. Magn Reson Med. 1993;29:804–811 doi: 10.1002/mrm.1910290613 [DOI] [PubMed] [Google Scholar]
  • 63.Blamire AM, Rothman DL, Nixon T. Dynamic shim updating: A new approach towards optimized whole brain shimming. Magn Reson Med. 1996;36(1):159–165. doi: 10.1002/mrm.1910360125 [DOI] [PubMed] [Google Scholar]
  • 64.Nachmani A, Schurr R, Joskowicz L, Mezer AA. The effect of motion correction interpolation on quantitative T1 mapping with MRI. Med Image Anal. 2019;52:119–127. doi: 10.1016/j.media.2018.11.012 [DOI] [PubMed] [Google Scholar]
  • 65.Rempp K, Brix G, Wenz F, Becker C, Gückel F, Lorenz W. Quantification of regional cerebral blood flow and volume with dynamic susceptibility contrast-enhanced MR imaging. Radiology. 193(2):637–641. doi: 10.1148/radiology.193.3.7972800 [DOI] [PubMed] [Google Scholar]
  • 66.Tofts P, Kermode A. Measurement of the blood-brain barrier permeability and leakage space using dynamic MR imaging. 1. Fundamental concepts. Fundamental concepts Magnetic Resonance in Medicine. 1991;17(2):357–367. doi: 10.1002/mrm.1910170208 [DOI] [PubMed] [Google Scholar]
  • 67.Liu G, Banerjee S, Yang X, et al. A dextran-based probe for the targeted magnetic resonance imaging of tumours expressing prostate-specific membrane antigen. Nat Biomed Eng. 2017;1:977–982. doi: 10.1038/S41551-017-0168-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Han Z, Chen C, Xu X, et al. Dynamic contrast-enhanced CEST MRI using a low molecular weight dextran. NMR Biomed. 2022;35(3):1–11. doi: 10.1002/nbm.4649 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Simegn GL, van der Kouwe AJW, Robertson FC, Meintjes EM, Alhamud A. Real-time simultaneous shim and motion measurement and correction in glycoCEST MRI using double volumetric navigators (DvNavs). Magn Reson Med. 2019;81(4):2600–2613. doi: 10.1002/mrm.27597 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supinfo

Supporting Information Figure S1 Patterns used for rigid head motion simulation in the cases listed in Supporting Information Tables S1S5. (A): MSP. (B): 2×MSP; all motions in (A) are multiplied by a factor of 2. (C): 2×MSP, but only rotations. (D): 2×MSP, but only translations. (E): Arbitrary motion pattern. (F): Scale coefficient to simulate oscillatory ventricular volume changes over time due to cardiac pulsation and LV volume swelling after infusion (with and without infusion) related to glucose uptake.

Supporting Information Figure S2 AUCmean maps over the early post-infusion (240–305 s) interval for the first motion scenario (Supporting Information Figure S1.A: MSP) in Supporting Information Table S1. The different cases causing tissue mixing are (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction. Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth (infusion only). Case-IDs for each scenario (cf. Supporting Information Table S1) are placed in the top left corner of each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S3 Column-wise organized analyses for the first motion scenario (Supporting Information Figure S1.A: MSP and Supporting Information Table S1) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S4 AUCmean maps over the early post-infusion (240–305 s) interval for the second motion scenario (Supporting Information Figure S1.B: 2×MSP) in Supporting Information Table S2. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S2) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S5 Column-wise organized analyses for the second motion scenario (Supporting Information Figure S1.B: 2×MSP: all rigid head motions multiplied by a factor of 2) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S6 AUCmean maps over the early post-infusion (240–305 s) interval for the third motion scenario (Supporting Information Figure S1.C: 2 × MSP with only rotations) in Supporting Information Table S3. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S3) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S7 Column-wise organized analyses for the third motion scenario (Supporting Information Figure S1.C: 2×MSP with only rotations) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; ;(H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S8 AUCmean maps over the early post-infusion (240–305 s) interval for the fourth motion scenario (Supporting Information Figure S1.D: 2 × MSP with only translations) in Supporting Information Table S4. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively. (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S4) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S9 Column-wise organized analyses for the fourth motion scenario (Supporting Information Figure S1.D: 2×MSP with only translations) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S10 AUCmean maps over the early post-infusion (240–305 s) interval for the fifth motion scenario (Supporting Information Figure S1.E: arbitrary rotational and translational motions) in Supporting Information Table S5. Different cases causing tissue mixing (A1–A4): only ventricular motion, (B1–B4): only rigid head motion and (C1–C4): ventricular and rigid head motion combined. All cases are shown with and without motion correction, (D1–D2): Cases without tissue mixing with B0-changes, are shown in D1 and D2, without and with infusion, respectively; (D3): the ground truth for infusion. Corresponding case-IDs for each scenario (cf. Supporting Information Table S5) are in the top left corner in each map. With and without motion is indicated as with and without tissue mixing, respectively. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Figure S11 Column-wise organized analyses for the fifth motion scenario (Supporting Information Figure S1.E: arbitrary rotational and translational motions) in form of extracted dynamic response curves and histograms for different tissues based on manually drawn ROIs (right) in (E1–E4): tumor tissue; (F1–F4): WM; (G1–G4): GM; (H1–H4): CSF. Histograms correspond to the timepoint of the DGE signal time peak (305 s after infusion start). The ground truth is set as either with or without infusion. NoMoco: no motion correction; Moco: with motion correction.

Supporting Information Table S1 Cases evaluated for the first scenario (MSP) with respect to with/without infusion, with/without B0-changes as a consequence of the rigid head motion pattern, and with/without ventricular and rigid head motion separately and combined.

The case-ID has five digits with the following coding: (1–5: number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with rigid head motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S2 Scenarios evaluated for 2×MSP with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with rigid head motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S3 Scenarios evaluated for 2×MSP with only rotations with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S4 Scenarios evaluated for 2 × MSP with only translations with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Supporting Information Table S5 Scenarios evaluated for arbitrary rotational and translational head motions with respect to with and without infusion, B0-changes as a consequence of the rigid head motion pattern, and ventricular and rigid head motion leading both to tissue mixing.

The case-ID has five digits with the following coding: (1–5: case number of motion scenario, i/I: without infusion/with infusion, r/R: tissue mixing without/with motion pattern, b/B: without/with B0-changes due to rigid head motion pattern, v/V: tissue mixing without/with ventricular motion).

I) This indicates if tissue mixing due to the rigid head motion pattern under the listed case has been included or not in the simulations. II) This indicates whether the simulations included B0-changes as a consequence of the rigid head motion pattern. III) This indicates if tissue mixing due to ventricular motion under the listed case has been included or not in the simulations.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.