Skip to main content
JAMA Network logoLink to JAMA Network
. 2023 Mar 9;149(5):466–467. doi: 10.1001/jamaoto.2023.0035

Audiometric Validation of a Smart Watch Decibel Meter

Ethan Gregory Muhonen 1, Mehdi Abouzari 1, Ye Yang 2, Fan-Gang Zeng 1,2, Hamid R Djalilian 1,3,
PMCID: PMC9999267  PMID: 36892831

Abstract

This diagnostic study assesses the performance of the Apple Watch Noise application in comparison with a class 1 sound level meter.


The Noise application on the Apple Watch measures the intensity of ambient sounds and is available on the Apple Watch SE and Apple Watch Series 4 and later.1 One prior study using the Noise application in a critical-care setting recommended further studies to explore the role of wearable devices in measuring noise levels.2 In addition, the promise of consumer wearables to help inform clinical decision-making has been emphasized as an important direction of future research in otolaryngology.3 To our knowledge, no studies have formally assessed the performance of the Noise application. Therefore, the aim of this diagnostic study was to assess the performance of the Apple Watch Noise application in comparison with a class 1 sound level meter.4

Methods

Both an Apple Watch Series 6 (Apple) and a Type 2250 Light handheld analyzer with the ZC-0032 microphone (Brüel & Kjær) were placed 50 cm from a speaker inside of an audiometric booth, and pure tones were played in half octaves from 125 to 8000 Hz in an amplitude range from 0.0001 to 1. A total of 21 amplitude measurements were performed within each half-octave frequency to allow sufficient statistical power in device comparison. The BZ-5503 Measurement Partner Suite software (Brüel & Kjær) was used with a D-75A amplifier (Crown) to produce the pure tones. Amplitude was provided as an input variable within this software. The measurement data for each amplitude across different frequencies were averaged for each device and compared with paired sample t test. SPSS software, version 17.0 (IBM) was used for statistical analysis, with P = .002 considered statistically significant after adjustment. This study was exempt from institutional review board approval because no human participants were involved.

Results

Device measurements demonstrated a range of sound from a minimum of 22.8 dB sound pressure level (A-weighted) to a maximum of 117.3 dB sound pressure level (A-weighted). Measurements between the 2 devices only displayed statistically significant differences between the 2 device outputs at 4 and 8 kHz (Table). In addition, correlation of dB measurements between the 2 devices at all frequencies was 0.961 or greater.

Table. Comparison of Measured Data by the 2 Devices for 21 Amplitude Measurements Across Different Frequencies.

Frequency, Hz Sample size Average measurement, mean (SD), dB P valuea 95% CI Effect size Relative size
Watch Meter
125 21 64.05 (21.10) 66.52 (22.70) .04 −4.86 to −0.08 0.47 Small
250 21 71.62 (22.20) 72.97 (22.99) .26 −3.80 to 1.09 0.25 Small
500 21 78.81 (23.01) 79.60 (23.47) .45 −2.97 to 1.38 0.17 Negligible
750 21 84.90 (23.22) 85.54 (23.64) .51 −2.60 to 1.33 0.15 Negligible
1000 21 86.57 (23.67) 86.99 (24.01) .67 −2.45 to 1.61 0.09 Negligible
1500 21 85.76 (23.53) 86.45 (23.43) .47 −2.62 to 1.25 0.16 Negligible
2000 21 81.52 (24.16) 82.99 (24.19) .14 −3.45 to 0.52 0.36 Small
3000 21 80.90 (22.50) 82.54 (23.55) .12 −3.76 to 0.48 0.35 Small
4000 21 85.90 (22.49) 80.06 (22.66) <.001b 3.87 to 7.81 −1.35 Large
6000 21 75.90 (23.11) 75.62 (24.84) .77 −1.70 to 2.27 −0.07 Negligible
8000 21 59.76 (20.09) 64.30 (21.65) .002b −7.27 to −1.81 0.76 Moderate
a

P values were calculated by paired sample t test.

b

P value was below the threshold for statistical significance (P < .002).

Discussion

Very little is known about the noise measurement capabilities of the Noise application given the proprietary nature of the algorithms and modeling assumptions used to create it.2 This diagnostic study is, to our knowledge, the first to formally quantify these capacities. Based on the present data, the smart watch application was as accurate as a class 1 sound level meter in the frequency range of 125 to 3000 Hz and at 6000 Hz, with minor (<5 dB) differences at 4000 and 8000 Hz. Prior work by Crossley et al in 2021 considered a correlation of 0.9 to represent acceptable accuracy for commercially available iPhone applications.5 By this metric, the Noise application can be considered accurate across all frequencies tested in this analysis, given correlation of at least 0.961.

The World Health Organization has estimated that 1.1 billion teenagers and young adults are at risk for unsafe sound exposure.6 Given the present results, this smart watch application may be useful in identifying harmful noise exposure environments where a class 1 sound level meter is unavailable.

This study did not assess performance outside of an audiometric booth and only studied pure-tone sound exposures. Specific variables at work in real-life settings that may limit sound measurement include interference with microphones by wind and blockage of sound by clothing. In addition, sound exposure at the wrist is not a perfect representation of sound exposure at an individual’s ear. Finally, harmful sound exposure in work environments is rarely composed of pure tones, and the accuracy of the smart watch application at measuring exposures comprising more complex sounds remains a future opportunity for research.

Supplement.

Data Sharing Statement

References

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplement.

Data Sharing Statement


Articles from JAMA Otolaryngology-- Head & Neck Surgery are provided here courtesy of American Medical Association

RESOURCES