Sound for the human voice is produced via flow-induced vocal fold vibration. The vocal folds consist of several layers of tissue, each with differing material properties1. Normal voice production relies on healthy tissue and vocal folds, and occurs as a result of complex coupling between aerodynamic, structural dynamic, and acoustic physical phenomena. Voice disorders affect up to 7.5 million annually in the United States alone2 and often result in significant financial, social, and other quality-of-life difficulties. Understanding the physics of voice production has the potential to significantly benefit voice care, including clinical prevention, diagnosis, and treatment of voice disorders.
Existing methods for studying voice production include in vivo experimentation using human and animal subjects, in vitro experimentation using excised larynges and synthetic models, and computational modeling. Owing to hazardous and difficult instrument access, in vivo experiments are severely limited in scope. Excised larynx experiments have the benefit of anatomical and some physiological realism, but parametric studies involving geometric and material property variables are limited. Further, they are typically only able to be vibrated for relatively short periods of time (typically on the order of minutes).
Overcoming some of the limitations of excised larynx experiments, synthetic vocal fold models are emerging as a complementary tool for studying voice production. Synthetic models can be fabricated with systematic changes to geometry and material properties, allowing for the study of healthy and unhealthy human phonatory aerodynamics, structural dynamics, and acoustics. For example, they have been used to study left-right vocal fold asymmetry3,4, clinical instrument development5, laryngeal aerodynamics6-9, vocal fold contact pressure10, and subglottal acoustics11 (a more comprehensive list can be found in Kniesburges et al.12)
Existing synthetic vocal fold models, however, have either been homogenous (one-layer models) or have been fabricated using two materials of differing stiffness (two-layer models). This approach does not allow for representation of the actual multi-layer structure of the human vocal folds1 that plays a central role in governing vocal fold flow-induced vibratory response. Consequently, one- and two-layer synthetic vocal fold models have exhibited disadvantages3,6,8 such as higher onset pressures than what are typical for human phonation (onset pressure is the minimum lung pressure required to initiate vibration), unnaturally large inferior-superior motion, and lack of a “mucosal wave” (a vertically-traveling wave that is characteristic of healthy human vocal fold vibration).
In this paper, fabrication of a model with multiple layers of differing material properties is described. The model layers simulate the multi-layer structure of the human vocal folds, including epithelium, superficial lamina propria (SLP), intermediate and deep lamina propria (i.e., ligament; a fiber is included for anterior-posterior stiffness), and muscle (i.e., body) layers1. Results are included that show that the model exhibits improved vibratory characteristics over prior one- and two-layer synthetic models, including onset pressure closer to human onset pressure, reduced inferior-superior motion, and evidence of a mucosal wave.
Keywords: Vocal folds, larynx, voice, speech, artificial biomechanical models
Protocol Text
The fabrication sequence (see Fig. 1) consists of making molds for vocal fold model layers, sequentially casting silicone layers, and mounting the models for testing. The model has four distinct layers: body, ligament, superficial lamina propria, and epithelium, in addition to a single fiber. A backing layer is added to facilitate the accurate placement of individual layers to the vocal fold model. The model geometric parameter definitions are shown in Fig. 2, with parameter values for the current model given in Table 1. In the following sections, different silicone mixing ratios are specified for the different layers; these produce material properties that are similar to those reported for human vocal fold tissue in the small strain regime13 (see Table 2).
Table 1.
Parameter | Ligament Model | Description |
---|---|---|
θ 1b,c | 50° | Inferior glottal angle |
θ 2b,c | 5° | Intraglottal angle |
θ 3b,c | 90° | Superior glottal angle |
r1c | 6.0 mm | SLP entrance radius |
r2c | 0.987 mm | SLP exit radius |
r1b | 2.0 mm | Ligament entrance radius |
r2b | 0.513 mm | Ligament exit radius |
T | 0.1 mm | Vertical glottal thickness |
t | 1.6 mm | Inferior and superior cover layer thickness |
δ | 2.0 mm | Maximum medial cover layer thickness |
D | 8.4 mm | Lateral depth |
A | 1 mm | Ligament layer thickness |
Table 2.
Model Section | Mixing Ratio | Young's Modulus (kPa) |
---|---|---|
Body | EF 1:1:1 | 11.8 |
Ligament | EF 1:1:4 | 1.6 |
SLP | EF 1:1:8 | 0.2 |
Epithelium | DS 1:1:1 | 49.8 |
1) Mold fabrication and preparation
1.1) Create solid models of three vocal fold layers: superficial lamina propria, ligament, and body layers. This is typically done by creating 3D computer-aided design (CAD) models with the desired geometries, exporting the CAD models as stereolithography (STL) files, and sending the STL files to a custom machine shop for rapid prototyping.
1.2) Create a box-shaped mold form using thin pieces of acrylic material. Approximate dimensions (not critical) are 2.54 cm high × 5.72 cm wide × 6.35 cm deep. Make the bottom of the form by adhering it to a flat acrylic plate. Seal all interior edges with vacuum grease.
1.3) Place a small amount of vacuum grease on the lateral side of the solid model of desired geometry (i.e., the body, the ligament, or the superficial lamina propria). Press model into bottom of mold form cavity, vacuum grease side down, so that the vacuum grease holds the part in place. Liberally coat mold form and solid model with release agent. Using a paint brush, ensure release agent reaches into all corners of the mold form cavity.
1.4) Mix 10 parts A and one part B of Smooth-Sil 950 platinum silicone rubber (parts measured by weight) and pour into mold form cavity. To remove air bubbles place mold form with uncured silicone rubber in a vacuum chamber and reduce pressure (e.g., to around 23 inHg below atmospheric pressure) for approximately three minutes (or more or less as necessary). Remove from vacuum chamber and place on level surface. Allow to cure for 24 hours and remove mold from mold form.
1.5) Repeat steps 1.1 through 1.4 to create molds for each of the superficial lamina propria, ligament, and body layers.
1.6) Cut ligament layer mold at the center of the medial surface in the anterior-posterior direction with a straight razor to allow for fiber insertion.
2) Casting of each layer
2.1.1) Body Layer
Apply thin layer of release agent to body mold cavity with paint brush. Mix one part B and one part A of Ecoflex® 00-30 Supersoft Platinum Silicone (by weight). Add one part Silicone Thinner (by weight) to reduce the eventual cured stiffness of the material. Mix together for 30 seconds and place in vacuum chamber for one minute to remove entrapped air. Remove mixture from vacuum and pour into the body mold cavity, but do not fill to the top of the entire mold cavity. Place in oven at 250 °F for 30 minutes. Remove from oven and cool.
2.1.2) Backing
Mix one part B and one part A of Dragon Skin® and add one part Silicone Thinner (by weight). Mix vigorously for 30 seconds, place in vacuum for 1 minute, and pour into the body mold cavity until full. Place in oven at 250 °F for 30 minutes. Remove mold from oven and cool. Remove model from mold, allow to cool to room temperature, and remove any release agent on the surface of the body layer with paper towel.
2.1.3) Ligament Layer
Apply thin layer of release agent on ligament mold cavity surface with paint brush. Place a 30 cm thread in the mold by pushing it into the cut from the straight razor. Thoroughly mix one part B and one part A of Ecoflex® 00-30 and four parts of Silicone Thinner (by weight). Place in vacuum chamber to remove air bubbles and pour mixture into ligament mold cavity.
2.1.4) Ligament Layer (continued)
Press body-backing model (from Steps 2.1.1 and 2.1.2) into the ligament mold cavity. Begin insertion at one side and gently move to the other so that the model pushes the excess uncured silicone and air bubbles out of the mold cavity. If air bubbles are present, remove the model from the mold cavity, refill with uncured silicone, and repeat pressing model into the mold. Place mold in oven for 30 minutes, remove, and cool to room temperature. Remove model from mold. Remove excess release agent with paper towel.
2.1.5) Superficial Lamina Propria Layer
Apply thin layer of release agent on superifical lamina propria (SLP) mold cavity surface with paint brush. Mix one part B, one part A of Ecoflex® 00-30, and 8 parts Silicone Thinner by weight. Vacuum as done previously and pour into SLP mold cavity. Use the same process described in Step 2.1.4 to insert the ligament-body-backing model into the superficial lamina propria mold cavity. Place in an oven at 250 °F and cure for one hour. Remove from oven and allow to cool. Remove model slowly and with extreme care so that the superficial lamina propria remains intact.
2.1.6) Epithelium Layer
Place vocal fold model on a flat surface with the backing down. Remove support material with a straight razor. Suspend threads in air by attaching them to an object of greater height than the model. Mix one part B and one part A of Dragon Skin® with one part of Silicone Thinner, mix, vacuum, then pour over the model and allow to cure for one hour. Repeat the process to create a thicker layer. Remove excess material with a straight razor.
2.1.7) Optional
If each layer is desired to be a different color (for visible inspection of different layers), add dye to part B of either the Ecoflex® or Dragon Skin® during the mixing process.
2.1.8) Optional
If material property data will be collected, create tensile and rheological specimens simultaneously with fabrication of each model layer. Do this by pouring extra uncured material into release agent-treated molds of desired material property specimen shape and size.
2.1.9) Optional
If measurements of layer thickness are desired, cut a cross section of the model with a straight razor and inspect with microscope.
3) Final model preparation for testing
3.1) Mount each completed vocal fold model into an acrylic mounting plate by first applying a thin layer of silicone glue on the back (lateral) and side (anterior-posterior) model surfaces. Insert model into recessed cut of mounting plate. Align the model medial surface with the top of the acrylic plate. Wipe away excess glue. Allow glue to cure for one hour.
3.2) Apply talc powder to the model surface to reduce surface tackiness.
3.3) For medial surface tracking use a fine-point Sharpie pen to mark dots on the model. Best results occur if marking is done after application of talc powder.
3.4) Place long bolts through holes of the mounting plate with the threaded ends pointing toward the model to which the existing model will be paired. Lay threads over the bolts. Put closed cell foam over the bolts to close any air gaps.
3.5) Pair this prepared model with another vocal fold model that has been similarly mounted to an acrylic holder using Steps 3.1 and 3.2. Tighten screws to compress the foam and bring the medial surfaces together until the desire pre-vibratory gap is reached. Ensure both sets of threads are placed over the bolts and extend outward from the acrylic plates in the anterior-posterior direction.
3.6) Mount vocal fold pair on air supply tube.
3.7) Tie the anterior threads together to form a loop. Repeat for the posterior threads. Hang desired weight on the loops simultaneously.
3.8) Models are now ready for testing and data collection.
Representative Results
Vibratory response data from one model created using this fabrication process are as follows; these results are typical. With tension of approximately 31 g applied to the fibers, the onset pressure was 400 Pa. At a subglottal pressure of 10% above onset pressure (440 Pa), the model vibrated at 115 Hz with a glottal flow rate of 210 ml/s. These values are in good agreement with values reported for those of humans (Table 3). Using high-speed videokymography to analyze model motion showed evidence of a phase difference between the superior and inferior margins, i.e., the superior margin concealed the inferior margin during the open phase of the vibration period (Fig. 3). Trajectories extracted from stereo images of the dots applied to the medial and inferior surfaces of the vocal fold model showed that the model exhibited an alternating convergent-divergent profile that is typical of human phonation, a mucosal wave-like motion, and a lower inferior-superior motion than in previous models (Fig. 4).
Table 3.
Onset Pressure (kPa) | Frequency (Hz) | Flow Rate (ml/s) | |
---|---|---|---|
Synthetic Model | 0.40 | 115 | 210 |
Adult Human | 0.29–0.4915 | 98–24716 | 70 – 20015 |
Discussion
This method of fabricating synthetic vocal fold models yields models that exhibit vibratory behavior similar to that of human vocal folds. The multi-layer concept results in significant advantages over previous one- and two-layer model designs3,6,8,15, in terms of reduced onset and improved model motion (convergent-divergent profile during oscillation, mucosal wave-like motion, and reduced inferior superior displacement). The method presented here is demonstrated on a somewhat idealized model in terms of geometry, but it can be applied to models with different geometries. For example, a model based on human imaging geometric data (e.g., MRI17, CT) could be fabricated using this method. Additionally, this fabrication process concept may find application in other research areas in which flow-induced vibrations and/or multiple layers of soft materials are central elements, e.g., investigations of flow through blood vessels, sleep apnea, and animal locomotion (particularly swimming and flying).
The model described here has some limitations that could be subjects for future research and development. The materials have linear stress-strain response characteristics, and an anticipated future improvement includes the incorporation of nonlinear stress-strain materials. Use of biological rather than synthetic materials in this fabrication process is also possible. Because of the extreme flexibility of the lamina propria layer, the model is less robust under vibration than previous one- and two-layer models. However, keeping the subglottal pressure beneath approximately 1 kPa and occasionally applying talc powder to minimize surface adhesion should allow for the model to be used for durations on the order of days with minimal changes in model behavior, typically far exceeding those possible using excised larynges.
Name of the reagent | Company | Catalogue number | Comments (optional) |
---|---|---|---|
High Vacuum Grease | Dow Corning | 01018817 | |
Pol-Ease 2300 | Polytek | Pol-Ease2300-1 | Release agent |
Smooth-Sil 950 | Smooth-On | Smooth-Sil 950 | Mold making material |
Vacuum Pump | Edwards | E2M2 | |
Vacuum Chamber | Kartell | 230 | |
Pressure Gage | Marsh Bellofram | 11308252A | |
Straight Razor | Husky | 008-045-HKY | |
Ecoflex 00-30 | Smooth-On | Ecoflex 00-30 | |
Silicone Thinner | Smooth-On | Silicone Thinner | |
Dragon Skin | Smooth-On | Dragon Skin 10 FAST | |
Thread | Omega | OmegaCrys | Use only clear fibers |
Silicone Dye | Smooth-On | Silc Pig Black | |
Silicone Glue | Smooth-On | Sil-Poxy | |
Talc Powder | Western Family |
Acknowledgments
The authors gratefully acknowledge Grants R03DC8200, R01DC9616, and R01DC5788 from the National Institute on Deafness and Other Communication Disorders for support of synthetic model development.
Footnotes
Disclosures:
The authors have nothing to disclose.
References
- 1.Hirano M, Kakita Y. Cover-body theory of vocal fold vibration. Speech Science: Recent Advances. 1985:1–46. [Google Scholar]
- 2.National Institute on Deafness and Other Communication Disorders 2010 http://www.nidcd.nih.gov/health/statistics/vsl.asp.
- 3.Pickup BA, Thomson SL. Influence of asymmetric stiffness on the structural and aerodynamic response of synthetic vocal fold models. Journal of Biomechanics. 2009;42(14):2219–2225. doi: 10.1016/j.jbiomech.2009.06.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zhang Z. Vibration in a self-oscillating vocal fold model with left-right asymmetry in body-layer stiffness. Journal of the Acoustical Society of America. 2010;128(5):EL279–EL285. doi: 10.1121/1.3492798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Popolo PS, Titze IR. Qualification of a Quantitative Laryngeal Imaging System Using Videostroboscopy and Videokymography. Annals of Otology, Rhinology & Laryngology. 2008;117(6):4014–412. doi: 10.1177/000348940811700602. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Thomson SL, Mongeau L, Frankel SH. Aerodynamic transfer of energy to the vocal folds. Journal of the Acoustical Society of America. 2005;118(3):1689–1700. doi: 10.1121/1.2000787. [DOI] [PubMed] [Google Scholar]
- 7.Neubauer J, Zhang Z, Miraghaio R, Berry DA. Coherent structures of the near field flow in a self-oscillating physical model of the vocal folds. Journal of the Acoustical Society of America. 2007;121(2):1102–1118. doi: 10.1121/1.2409488. [DOI] [PubMed] [Google Scholar]
- 8.Drechsel JS, Thomson SL. Influence of supraglottal structures on the glottal jet exiting a two-layer synthetic, self-oscillating vocal fold model. Journal of the Acoustical Society of America. 2008;123(6):4434–4445. doi: 10.1121/1.2897040. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Becker S, et al. Flow-structure-acoustic interaction in a human voice model. Journal of the Acoustical Society of America. 2009;125(3):1351–1361. doi: 10.1121/1.3068444. [DOI] [PubMed] [Google Scholar]
- 10.Spencer M, Siegmund T, Mongeau L. Experimental study of the self-oscillation of a model larynx by digital image correlation. Journal of the Acoustical Society of America. 2007;123(2):1089–1103. doi: 10.1121/1.2821412. [DOI] [PubMed] [Google Scholar]
- 11.Zhang Z, Neubauer J, Berry D. The influence of subglottal acoustics on laboratory models of phonation. Journal of the Acoustical Society of America. 2006;120(3):1558–1569. doi: 10.1121/1.2225682. [DOI] [PubMed] [Google Scholar]
- 12.Kniesburges S, et al. In vitro experimental investigation of voice production. Current Bioinformatics. doi: 10.2174/157489311796904637. In Press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Titze IR. The Myoelastic Aerodynamic Theory of Phonation. National Center for Voice and Speech; 2006. pp. 82–101. [DOI] [PubMed] [Google Scholar]
- 14.Murray PR. Master's Thesis. Brigham Young University; 2011. Flow-Induced Responses of Normal, Bowed, and Augmented Synthetic Vocal Fold Models. [Google Scholar]
- 15.Baken RJ, Orlikoff RF. Clinical Measurement of Speech and Voice. 2nd edn Singular Publishing; 2000. [Google Scholar]
- 16.Titze IR. Principles of Voice Production. 2nd printing National Center for Voice and Speech; 2000. [Google Scholar]
- 17.Pickup BA, Thomson SL. Flow-induced vibratory response of idealized vs. magnetic resonance imaging-based synthetic vocal fold models. Journal of the Acoustical Society of America. 2010;128(3):EL124–EL129. doi: 10.1121/1.3455876. [DOI] [PMC free article] [PubMed] [Google Scholar]