Summary
We report a noninvasive strategy for electrically stimulating neurons at depth. By delivering to the brain multiple electric fields at frequencies too high to recruit neural firing, but which differ by a frequency within the dynamic range of neural firing, we can electrically stimulate neurons throughout a region where interference between the multiple fields results in a prominent electric field envelope modulated at the difference frequency. We validated this temporal interference (TI) concept via modeling and physics experiments, and verified that neurons in the living mouse brain could follow the electric field envelope. We demonstrate the utility of TI stimulation by stimulating neurons in the hippocampus of living mice without recruiting neurons of the overlying cortex. Finally, we show that by altering the currents delivered to a set of immobile electrodes, we can steerably evoke different motor patterns in living mice.
Graphical Abstract
Introduction
Physical means of brain stimulation, such as the use of implanted electrodes for deep brain stimulation (DBS), have led to widespread excitement about the possibility of repairing neural dysfunction through direct control of brain circuit dynamics, including multiple FDA-approved therapies for previously intractable brain disorders (Greenberg et al. 2010; Kalia et al. 2013). Electrical stimulation via implanted electrodes sparsely activates distributed sets of neurons (Histed et al. 2009), in a fashion different from direct optogenetic stimulation of local cells (Gradinaru et al. 2009). The impact of electromagnetic stimulation on brain circuitry is an emergent function of the fields themselves, and the excitability properties of the neurons themselves, as well as the configuration of the neural network in which they are embedded (Merrill et al. 2005). As a result of this complexity, physical means of brain stimulation are often used in a phenomenological way, especially because the excitability properties of neurons vary across different cell types, and thus understanding how a given brain stimulation method impacts a given brain function may be complex.
However, some properties of neurons are universal – for example, the intrinsic low-pass filtering of electrical signals by the neural membrane (Hutcheon & Yarom 2000), which prevents neural electrical activity from following very high frequency oscillating (e.g., > 1 kHz) electric fields. We here explore whether the biophysics underpinning such a universal property might support novel strategies for electrical brain stimulation. In particular, if we apply high-frequency oscillating electric fields at multiple sites outside the brain, neurons in the brain will not be able to follow these high-frequency fields directly. However, if two such electric fields are applied at high frequencies that differ by a small amount, which corresponds to a low frequency that neurons can follow, neurons in the brain may be able to demodulate and follow the envelope modulation that results from the interference between these two applied fields, and which oscillates at the difference frequency. If the amplitude of the envelope modulation reaches a maximum at a site deep in the brain, it might be possible to drive deep-lying neurons without recruiting overlying ones. We here test this concept, which we call temporal interference (TI) stimulation, using computational modeling and phantom measurements, as well as electrophysiological measurements in vivo. We demonstrate the ability of TI stimulation to mediate activation of hippocampal neurons without recruiting overlying cortical neurons, and steerably probe motor cortex functionality without physically moving electrodes by altering the current magnitudes delivered to a fixed set of electrodes.
Results
Temporal interference stimulation: concept and validation of neural firing recruitment
We first set out to examine whether the TI concept could indeed result in well-defined low frequency envelope modulated electric fields. In the TI concept (Figure 1A), electric currents are applied at high frequencies f1 and f2 = f1 + Δf that fall outside the range of normal neural operation, but which differ by a small amount Δf that falls within the frequency range that neurons can respond to. The superposition of the two electric fields inside the brain results in an electric field at a frequency of (f1 + f2)/2 whose envelope is modulated at the frequency Δf (Figure 1B). The amplitude of the envelope modulation at a particular location depends on the vectorial sum of the two applied field vectors at that point, and as a result can have a maximum at a point distant from the electrodes, potentially even deep in the brain (Figure 1C). The location of this envelope maximum depends on the electrode configuration, as well as properties of the applied waveforms. For the trapezoidal configuration shown in Figure 1A, the low frequency envelope oscillates at a frequency of 40 Hz, with waveforms in Figure 1B plotted at the two specific points highlighted by Roman numerals in Figure 1A. For example, Figure 1Bi shows a large envelope modulation amplitude at a location where the two fields are large and aligned, whereas Figure 1Bii shows a small envelope modulation amplitude at a location where the two fields are less aligned.
To assess whether such low frequency field envelopes could effectively drive neural spiking activity, we applied TI stimulation transcranially to anesthetized living mice, and recorded the responses using whole-cell patch clamp neural recording. Currents were applied via two electrodes on the skull (with a ~0.5 mm gap between their edges), and recordings made in the somatosensory cortex. We found that interferential stimulation with two sinusoids at 2.01 kHz and 2 kHz, resulting in a Δf envelope frequency of 10 Hz, was able to recruit neurons to fire at 10 Hz (Figure 1D), as efficaciously as direct 10 Hz stimulation (Figure 1E) that would be expected to broadly affect neural activity (Miranda et al. 2013). High-frequency stimulation (with one sinusoid at 2 kHz, and no TI) did not result in activity (Figure 1F), beyond a brief transient associated with the beginning of stimulation in some cells (n = 4 out of 6 cells from 2 mice) when 0.25 s sinusoidal ramp-up times were used. When 0.5 s ramp-up times were used, no such transient activity was observed in any cells (n = 5 cells from 2 mice), suggesting that the transient spiking activity observed earlier was due to the speed of the 0.25 s duration ramp-up (Figure 1J). We validated TI stimulation on a population of cortical cells (Figure 1Ii), finding that interferential stimulation with a difference frequency of 10 Hz resulted in spike frequencies of 10.21 ± 0.83 Hz (mean ± standard deviation (st.d.)), for a 1 kHz carrier frequency (n = 6 cells from 2 mice) and 9.68 ± 0.85 Hz for a 2 kHz carrier frequency (n = 7 cells from 3 mice; see Table S1 for full statistics associated with Figure 1Ii).
To validate whether neuronal firing can be manipulated at different depths in tissue, we performed patch clamp recording in the mouse hippocampus. Currents were applied via two electrodes that were located on the skull with proximal edges 1.5–2 mm apart. We found that interferential stimulation (with two sinusoids at 2.01 kHz and 2 kHz, resulting in a Δf envelope frequency of 10 Hz) was able to recruit neural firing in synchronization with the envelope -- with either single spikes (n = 3 cells from 2 mice) or brief bursts of spikes (n = 5 cells from 3 mice; a burst was defined as a < 50 ms spiking event with inter-spike interval ≤ 15 ms; 1.3 ± 0.37 mean spikes per burst ± st.d; 9.07 ± 3.2 ms inter-spike interval) elicited by the TI stimulation (in detail: mouse 1 had 1 cell with a burst response; mouse 2 had 2 cells with single spike responses; mouse 3 had 2 cells with a burst response; mouse 4 had 2 cells with a burst response and 1 cell with a single spike response) (Figure 1G). Direct application of high-frequency stimulation (with two sinusoids on the two electrodes, both at 2 kHz) did not result in activity (Figure 1H). No spiking transient was observed because we used the slower, 0.5 s duration ramp-up that we had previously observed to eliminate this transient (Figure 1J; n = 5 cells from 3 mice). We found (Figure 1Iii) that interferential stimulation with a difference frequency of 10 Hz resulted in spike or burst occurrence frequencies of 10.23 ± 0.61 Hz for a 2 kHz carrier frequency (n = 8 cells from 4 mice; see Table S1 for full statistics associated with Figure 1Iii). The timing of the spikes or the first spikes of bursts, relative to the peak of the TI envelope, was −2.8 ± 4.8 ms, i.e., when the envelope amplitude was >97% of its peak amplitude, which was not different from the timing of spikes evoked by 10 Hz stimulation relative to the 10 Hz sinusoid peak (−1.3 ± 2.2 ms; pairwise t-test, p=0.47).
The membrane potential of neurons undergoing TI stimulation repolarized between single spikes, or between brief bursts of spikes, to the baseline membrane potential (cortex, −10.36 ± 27.84 mV, mean difference from baseline ± st.d.; p = 0.74, pairwise t-test; n = 13 cells from 5 mice; hippocampus, 5.5 ± 7.89 mV; p = 0.34; n = 8 cells from 4 mice). The spike frequency during the 20th bout of TI stimulation (tested in 3 cells in the somatosensory cortex from one mouse; 2 s stimulation followed by 2 s rest) was 9.93 ± 0.2 Hz (mean ± st.d), not different from the spike frequency during the 1st bout (p = 0.95; pairwise t-test), and the spike amplitude during the 20th bout of TI stimulation was not different from the spike amplitude during the 1st bout (5.3 ± 3.5 mV, mean amplitude difference ± st.d; p = 0.75, pairwise t-test); see Figure S1L–N for representative traces. The membrane potential of neurons undergoing high-frequency stimulation (with two sinusoids on the two electrodes, both at 2 kHz or at 1 kHz) in both the cortex and the hippocampus was not different from the baseline membrane potential before the stimulation (cortex, 1.67 ± 4.87 mV, mean difference from baseline ± st.d., measured 1 s after stimulation onset; p = 0.66, pairwise t-test; n = 11 cells from 5 mice; hippocampus, −1.7 ± 5.39 mV; p = 0.91; n = 5 cells from 3 mice).
Validation of steerability using computational modeling and tissue phantom
To explore the effects of interferential stimulation at a physics level, we both modeled the interferential electric field envelope magnitude as in Figure 1A but for a variety of electrode configurations, and also experimentally assessed these fields using a tissue phantom comprising a plastic cylinder filled with saline. Current sources were isolated in the circuitry as described in Figure S3. We found that by altering the locations of the electrodes and by setting the currents appropriately, we could enable the interferential envelope modulation to be targeted to specific locations. For example, when the electrodes were arranged in a trapezoidal geometry as in Figure 1A, but with a narrow base, we obtained, both in simulation and in the phantom, a peak of envelope modulation near the surface of the cylinder, at a point in between the two electrodes (Figure 2A), when both electrode pairs were conducting equal currents (1 mA in this example). By moving the electrodes that comprise the narrow base of the trapezoid further and further apart from each other, holding the currents constant, we could steer the location of the peak envelope modulation deeper into the tissue (Figure 2B), approaching the center of the cylinder as the trapezoid converged to a rectangle (Figure 2C). The envelope locus (i.e., distance out to 1/e of the envelope maximum) in Figure 2C was approximately 2 times larger, and the peak envelope amplitude 10 times weaker, than in Figure 2A (see Table 1 for numerical values associated with these three panels). Thus, it is possible to steer the envelope peak to have its maximum at essentially any depth throughout a volume, albeit with a tradeoff between the locus depth and its width and strength. Varying the locations of electrodes causes large changes in steering, with electrode size variation playing a more minor role (Figure S2A–C). These analyses were conducted with a cylindrical phantom, but similar field distributions were obtained in simulations with a spherical phantom (Figure S2G–L).
Table 1. Summary of envelope amplitude attributes in cylindrical phantom.
Electrodes’ trapezoid | Envelope modulation amplitude |EAMx̂|max | Envelope modulation amplitude |EAMŷ|max | |||||||
---|---|---|---|---|---|---|---|---|---|
Figure 2 panel | Bottom base | Top base | Current ratio (gray to black) | Peak (V/m) (simulated, measured) |
Normalized peak location (x,y) | Normalized half width half maximum (HWHMx̂, HWHMŷ) |
Peak (V/m) (simulated, measured) |
Normalized peak location (x,y) | Normalized half width half maximum (HWHMx̂, HWHMŷ) |
A | 0.39 | 1.96 | 1:1 | (3.7,1) | (0,−1) | (0.18,0.1) | (2.5,1.1) | (0,−0.77) | (0.18,0.37) |
B | 1.11 | 1.96 | 1:1 | (0.9,0.75) | (0,−0.87) | (0.33,0.25) | (0.9,0.7) | (0,−0.37) | (0.4,0.52) |
C | 1.96 | 1.96 | 1:1 | (0.13,0.15) | (0,−0.59) | (0.37,0.42) | (0.21,0.23) | (0,0) | (0.42,0.49) |
D | 1.96 | 1.96 | 1:2.5 | (0.13,0.14) | (−0.2,−0.59) | (0.37,0.42) | (0.2,0.22) | (−0.2,0) | (0.38,0.4) |
E | 1.96 | 1.96 | 1:4 | (0.12,0.13) | (−0.35,0) | (0.36,0.42) | (0.19,0.2) | (−0.35,0) | (0.37,0.34) |
We next explored whether tuning the currents, while holding electrode locations constant, could be used to control the field envelope locus, and in particular to steer the envelope modulation peak away from the centroid of the electrode locations. We started by taking the electrode configuration of Figure 2C, with its rectangular geometry, and adjusting the ratio of currents across the gray vs. black electrodes from 1:1 to 1:2.5 (Figure 2D) and 1:4 (Figure 2E). We found that by changing the current ratio between the electrode pairs, the peak envelope modulation became increasingly close to the electrode pair with the lower current, with the peak moving 20% of the radius away from the center in the 1:2.5 case (Figure 2D) and 35% of the radius away from the center in the 1:4 case (Figure 2E). This suggests the possibility of “live steering” of activity from one deep site to another within the brain, without having to physically move the electrodes themselves. By having a larger number of electrodes on the scalp, and tuning the current frequencies and amplitudes appropriately, it may be possible to make the deep targeted stimulation volume smaller, as we computationally model in Figure S2M–N.
Stimulation of mouse hippocampus but not overlying cortex
We next aimed to stimulate a deep structure (i.e., mouse hippocampus) while not recruiting overlying structures (i.e., the cortex). We performed simulations like those we did before, but now for the mouse brain, and predicted that 10 Hz transcranial stimulation applied to sites at the surface of the skull (Figure S4A) would broadly recruit neural activity throughout both superficial and deep structures. In contrast performing TI stimulation with a 10 Hz difference frequency (Figure S4B) would, in our models, result in a peak of 10 Hz envelope modulation at a deep site, with lower envelope modulation amplitudes in more superficial structures. Of course, such models make assumptions about brain electrical parameters and geometry that may vary from brain to brain (Peratta & Peratta 2010), possess limited spatial resolution (our anatomical mouse model had x, y, and z resolutions of 42 μm, 42 μm, and 700 μm, respectively) and do not take into account differences in neural excitability across cell types and brain regions. Thus, to assess whether TI stimulation could indeed recruit activity in deep neural circuits without driving overlying ones, we compared 10 Hz vs. TI stimulation in anesthetized mice, using the immediate early gene c-fos as a marker of neural activity, as has been used previously to gauge the focality of brain stimulation, since c-fos functions in the mouse cortex and hippocampus as an indicator of activated neurons (Chen et al. 2015; Dragunow & Robertson 1987; Reijmers et al. 2007).
As expected, 10 Hz transcranial stimulation (10 s on then 10 s off, for 20 minutes) resulted in widespread c-fos expression (measured 90 minutes after stimulation) in both the cortex and in the hippocampus underlying the electrodes (Figure 3A–B), with 13.6% + 2.2% (mean + st.d. used throughout) of cortical cells (as indicated by DAPI-stained nuclei) and 63.9% + 5.7% of hippocampal cells c-fos-positive underneath the electrodes (Figure 3C). In contrast, there was essentially no c-fos activation on the contralateral side (see Table S2 for complete statistics for Figure 3). Driving the brain with a 2 kHz transcranial current (with the same current magnitude and durations as in the 10 Hz case) resulted in essentially no c-fos positive cells (Figure 3D–F), in either the cortex or the hippocampus, and on either the electrode-bearing or contralateral side. In contrast, when TI stimulation was applied with frequencies of 2 kHz and 2.01 kHz (with the same current magnitude and duration as in the 10 Hz case), the hippocampus was strongly activated, Figure 3G–H, with c-fos in 53.12% ± 14.5% of DAPI-labeled cells (Figure 3I), not significantly different from that recruited by the 10 Hz stimulation (Figure 3C). But, despite the strong hippocampal recruitment, there was essentially zero c-fos in cortical cells - both at a site between the stimulating electrodes where the cortical envelope modulation field would be anticipated to be at its highest value in the cortex, with c-fos in 0.48% ± 0.47% of DAPI-labeled cells, or directly underneath an electrode, with c-fos in 0.32% ± 0.29% of DAPI-labeled cells (Figure 3I). Thus, TI stimulation can recruit neural activation in a deep structure such as the hippocampus, without recruiting the overlying cortex. As a control experiment, we separated the electrodes by a larger distance, which would be expected to reduce the envelope modulation field amplitude (Figure S4C), and obtained no activation in the cortex or hippocampus, Figure S4D–F.
We did not observe seizures during or after any of these stimulation paradigms (i.e., 10 Hz, 2 kHz or TI stimulation). Furthermore, c-fos staining was always observed only on the ipsilateral side of stimulation, and not in other analyzed regions including below the hippocampus, or in the contralateral hippocampus, consistent with a local (as opposed to propagating) neural activity profile. The high c-fos expression observed in the dentate gyrus is consistent with c-fos expression patterns observed in rats after strong, unilateral electrical stimulation of the hippocampus via an implanted electrode in animals treated with carbamazepine, which prevents seizures and also prevents bilateral c-fos staining upon unilateral stimulation (Dragunow & Robertson 1987).
Safety characterization of temporal interference stimulation
To characterize the safety profile of TI stimulation, we immunohistochemically examined cellular and synaptic molecular profiles in the cortex and the hippocampus after unilateral TI stimulation (2 kHz and 2.01 kHz, 10 s on then 10 s off, for 20 minutes), as in Figure 3G–I but in awake, behaving mice. Mice were sacrificed and transcardially perfused following a 24 hr recovery period to allow for detection of persistent effects (e.g., caspase-3 activation) after a bout of stimulation. Brain sections were fluorescently stained with antibodies for the neuronal marker NeuN (Wolf et al. 1996), the apoptotic marker cleaved caspase-3 (D’Amelio et al. 2012), the DNA damage marker γH2AX (Mah et al. 2010), the microglial marker Iba1 (Ito et al. 1998), the astrocyte marker GFAP (Eng et al. 2000), and the synaptic protein synaptophysin (Syp) (Tarsa & Goda 2002). We compared fluorescence profiles in the brain regions that were stimulated (‘Stim+’), with fluorescence profiles in the contralateral, non-stimulated hemisphere (‘Stim-’), as well as with fluorescence profiles in mice that underwent the same procedure but with current amplitudes set to 0 μA (‘Sham’). We found that TI stimulation did not alter the neuronal density or affect the number of apoptotic cells (Figure 4A–B, Figure S5A–B, Figure S5J), or induce DNA damage (Figure 4C–D, Figure S5C–D, Figure S5K), at least as reflected by the stains above, relative to unstimulated or sham stimulated brains. In addition, TI stimulation did not alter the intensity and density of Iba1 positive cells (Figure 4E–F, Figure S5G–H, Figure S5M) or GFAP positive cells (Figure S5N–O), suggesting a lack of reactive microglia and astrocytes, respectively, in response to TI stimulation. Finally, TI stimulation did not alter synaptophysin intensity, suggesting no changes to synapse density (Figure 4G–H, Figure S5E–F, Figure S5L). See Table S3 for full statistics for these experiments, analyzed over cortical and hippocampal regions, respectively.
To assess whether high frequency electric fields heat the brain, we measured the brain temperature during stimulation with 2 kHz fields (60s with 0.5 ramp-up and ramp-down periods) that were applied via an electrode configuration as in Figure 4A–H. We measured temperature with a 1 mm-diameter thermocouple probe inserted into the cortex underneath the lateral electrode. We found that the maximal temperature increase at this cortical location during stimulation was 0.069 ± 0.05 C° (mean ± st.d. increase from baseline temperature; n = 6 mice). This change in brain temperature was not larger than the largest spontaneous deviations from baseline seen during the pre- and post-stimulation periods (Figure 4I; p = 0.81, one-way ANOVA; see Table S3 for full statistics of Figure 4I).
Steerable probing of motor functionality without electrode movement
We next explored the capability of TI stimulation to activate neurons so as to drive mouse movements. Using ketamine-xylazine anesthetized mice, we applied a current I1 via an electrode that was positioned on the skull above the motor cortex region associated with a movement of the contralateral forepaw and a current I2 via a second electrode that was positioned on the contralateral skull, above the motor cortex region associated with movement of the whiskers ipsilateral to the I1 electrode (Figure 5A) (Tennant et al. 2011).
We first established the motor threshold by systematically increasing the current sum I1 + I2 in steps of 50 μA while keeping the current ratio I1:I2 fixed at 1:4. We found that if I1 and I2 were applied with a carrier frequency of 2 kHz and a difference frequency of 10 Hz, the stimulation evoked a 10 Hz periodic movement of the contralateral forepaw with a motor threshold of 916 ± 129 μA (Figure 5B; mean ± st.d., n = 6 mice). If I1 and I2 were applied at the same frequency, no motor movement was observed (n = 9 mice; assessed up to a current sum value of 2 mA). Changing the difference frequency between 1 Hz and 15 Hz changed the motion frequency accordingly, but not the motor threshold (p = 0.88; one-way ANOVA; n = 6 mice; Figure 5B and Video S1; see Table S4 for full statistics associated with Figure 5B). Increasing the carrier frequency from 1 kHz to 4 kHz linearly increased the motor threshold with a slope of 250 μA/kHz (linear regression, R2 = 0.99; Figure 5C and Video S2; see Table S4 for full statistics associated with Figure 5C); a 5 kHz carrier resulted in no response at the maximum current sum value tested (2 mA).
We next sought to test whether steering the site of stimulation by changing the current ratio I1:I2, as in our physics experiments (Figure 2C–E), would shift the site of motor cortex activation. We changed the current ratio I1:I2, keeping I1 + I2 fixed, and measured movements evoked in the forepaws, whiskers and ears (see Video S3 for representative video). We found that when I1 < I2 (that is, stimulation was steered towards the I1 electrode), TI stimulation evoked a movement of the forepaw (Figure 5Di, p=0.00004, Fisher’s exact test, n = 9 mice) and the whiskers (Figure 5Ei, p = 0.002, n = 9 mice) contralateral to the I1 electrode, but no movements ipsilateral to the I1 electrode (Figure 5Di–Fi; see Table S4 for full statistics associated with Figure 5Di–Fi). In contrast, when I1 > I2 (stimulation was steered towards the I2 electrode), TI stimulation evoked a movement of the whiskers (Figure 5Ei, p = 0.002, n = 9 mice) and ear (Figure 5Fi, p = 0.03, n = 9 mice) ipsilateral to the I1 electrode but no movements contralateral to the I1 electrode (Figure 5Di–Fi).
The movement of the forepaw contralateral to the I1 electrode was maximal when the current ratio I1:I2 was 1:8, i.e., our condition in which stimulation was maximally steered towards the I1 electrode (1.24 ± 0.36 mm, mean movement ± st.d. used throughout; p = 0.000007, unpaired t-test vs. null hypothesis of no movement, thresholding at p < 0.0025 Bonferroni corrected for multiple comparisons; n = 9 mice) and gradually decreased as the current ratio increased (Figure 5Dii). (Perhaps because the electrodes were not placed symmetrically, the movement of the forepaw ipsilateral to the I1 electrode, though not statistically significant, showed an opposite pattern, with a maximal movement when the current ratio was 8:1, i.e. when stimulation was maximally steered towards the I2 electrode (Figure 5Dii, 0.2 ± 0.6 mm, p = 0.35, n = 9 mice).) The movement of the whiskers contralateral to the I1 electrode was maximal when the current ratio I1:I2 was 1:4, i.e., stimulation was partially steered towards the I1 electrode (0.4 ± 0.41 mm, p = 0.018, n = 9 mice) and gradually decreased as the current ratio varied from this maximum (Figure 5Eii; the 1:8 condition steers the stimulation more laterally than the 1:4 condition, which perhaps stimulated cortical regions more lateral to the whisker region (Tennant et al. 2011), and thus elicited a lower whisker movement amplitude than that elicited by the 1:4 condition). In contrast, perhaps due to the asymmetrical location of our electrodes, the movement of the whisker ipsilateral to the I1 electrode was maximal when the current ratio I1:I2 was 8:1, i.e., stimulation was maximally steered towards the I2 electrode (0.96 ± 0.75 mm, p = 0.0016, n = 9 mice) and gradually decreased as the current ratio decreased (Figure 5Eii). A similar trend was seen for the ears (Figure 5Fii; see Table S4 for full statistics associated with Figure 5Dii–Fii). Thus, TI stimulation can support steering of brain stimulation without physical electrode movement, resulting in tunable elicitation of movements.
Discussion
In this paper, we present TI stimulation, validating the concept using modeling as well as both physics and neurophysiology experiments, and demonstrate its utility by performing stimulation of a deep region (mouse hippocampus) without stimulating overlying neurons (cortex), as well as steerable brain stimulation of motor patterns without physical electrode movement. Future studies, perhaps using larger numbers of electrodes and multiple sets of interfering fields, may be able to pinpoint even smaller regions of the brain, or multiple regions of the brain. An open question is how small a focal volume may be achieved. At some point, inhomogeneities in the gray and white matter of the brain may cause difficulty in improving the resolution below that spatial scale, although MRI scans and data-driven sculpting of the electric fields may be able to compensate for this to some degree.
How generalizable might be the effects observed here? There have been reports that strong kHz-frequency electric fields can block the propagation of compound action potentials in peripheral nerves (Cuellar et al. 2013; Kilgore & Bhadra 2014). Such effects were localized to the immediate vicinity of the electrodes, leaving regions a few millimeters away, perhaps where the magnitude of the fields was lower, unaffected. The magnitude of the fields used in our study to transcranially recruit neural activity in the brain were perhaps 1–2 orders of magnitude weaker than in these earlier studies, so we anticipate that such effects may have been minimal in our study. This is consistent with our repeatable observation of a lack of physiological effect of 2 kHz electric fields on brain activity. However, future studies might seek to explore how stronger kHz-frequency electric fields affect the brain. Such data might also present an upper limit to the field strengths applicable for TI stimulation. There has been a report of kHz-frequency transcranial electric field stimulation (1–5 kHz) that resulted in neural plasticity similar to that resulting from anodal DC stimulation (Chaieb et al. 2011), but a later report found that 5 kHz transcranial electric fields, grouped in theta burst patterns, did not result in neural plasticity (Kunz et al. 2016). In our current dataset, we did not observe effects of kHz-frequency electric fields beyond a brief transient that was observed when short ramp-up times were used (but not with longer ramp-up times), suggesting that in studies using kHz-frequency fields, subtle parameters of the stimulation may help determine exactly what effects on neural activity result, and presenting an area for future exploration. We found that TI stimulation at amplitudes sufficient to recruit deep brain structures, such as the hippocampus, did not alter the neuronal and synaptic integrity of the underlying tissue 24 h after stimulation, at least as reflected by the stains we used. Additional time points other than 24 h post stimulation may provide in the future a more detailed picture of the safety of TI stimulation. Furthermore, the safety profile of TI stimulation associated with evoked behavior patterns, such as those explored here, should be explored in the future.
Given the remarkable therapeutic benefits of deep brain stimulation for patients with otherwise treatment-resistant movement and affective disorders (Kringelbach et al. 2007), the prospects for noninvasive deep brain stimulation using electricity are potentially exciting. Other methods for noninvasive deep brain stimulation have been proposed, e.g. using transcranial ultrasound (Legon et al. 2014) or using expression of heat-sensitive receptors and injection of thermomagnetic nanoparticles (Chen et al. 2015), but the poorly understood mechanism of action (Plaksin et al. 2014) and the need to genetically manipulate the brain, respectively, may limit their immediate use in humans. TI stimulation may thus represent a practical strategy for noninvasively stimulating neurons deep in the brain. It uses familiar and well-tested electric fields (Stavroulakis 2014; International Commission on Non-Ionizing Radiation Protection 2009) and does not require chemical or genetic manipulation of the brain. We anticipate that it might rapidly be deployable into human clinical trials, as well as studies of the human brain.
STAR Methods
Further information and requests for resources and reagents should be directed to and will be fulfilled by the Lead Contact, Ed Boyden (esb@media.mit.edu).
EXPERIMENTAL MODEL AND SUBJECT DETAILS
Mouse: C57BL/6
Sex
Male.
Age
8–12 weeks old.
Source
All animals were purchased from Taconic Biosciences, Inc.
Housing and husbandry
Mice were housed in standard cages in the MIT animal facility with ad libitum food and water in a controlled light-dark cycle environment, with standard monitoring by veterinary staff.
Allocation of animals to experimental groups
Randomly assigned.
Committee approval
All animal procedures were approved by the Massachusetts Institute of Technology (MIT) Committee on Animal Care (CAC, Protocol Number: 1115-111-18), and all experiments conformed to the relevant regulatory standards.
Mouse: CK-p25
Sex
Male.
Age
4 months old.
Source
The CK-p25 transgenic mouse was created in the Tsai lab (Cruz et al. 2003)
Housing and husbandry
Mice were housed in standard cages in the MIT animal facility with ad libitum food and water in a controlled light-dark cycle environment, with standard monitoring by veterinary staff.
Allocation of animals to experimental groups
Randomly assigned.
Committee approval
All animal procedures were approved by the Massachusetts Institute of Technology (MIT) Committee on Animal Care, and all experiments conformed to the relevant regulatory standards.
METHOD DETAILS
Design and implementation of the TI stimulator
The kHz currents were generated using a custom made device consisting of two electrically isolated current sources. To isolate the channels, each waveform was supplied via a balanced pair of current sources that were driven in precisely opposite phase with a ground electrode carrying any imbalance currents (<1%) from the paired current sources, preventing charging of the body relative to earth ground (Figure S3A). Each current source reliably drove 2 mA of current on 1 kΩ loads up to a frequency of 10 kHz with a resolution of 0.02Hz (Figure S3C–D). At load resistances higher than 10 kΩ the current source eventually saturated. The current output had a leakage level <0.1 μA root mean square (RMS) at 100 kHz bandwidth, measured on a 1 kΩ load resistor with a differential amplifier (7A22, Tektronix). The total harmonic distortion of the current source was <0.08% at 100 Hz and <0.4% at 10 kHz (measured with 9 harmonics on a 1 kΩ load resistor). The total harmonic distortion and frequency cross-talk were measured using an FFT spectrum analyzer (SR770, Stanford Research). When the two current sources were applied to a common conductive load, e.g., a resistive bridge (Figure S3B) or a saline bath (Figure S3E), the cross-talk at the terminals of each channel was <0.1%, allowing almost 100% of the interference to build up inside the load. In comparison, without the anti-phasic drive, approximately 30% cross-talk was measured at both the channel terminals and inside the conductive medium (Figure S3B, Figure S3F).
In-vivo rodent electrophysiology
Surgical procedures
On the day of the experiment, the mice were injected with Meloxicam (1mg/kg) and buprenorphine (0.1mg/kg) and anesthetized with 1–2% (vol/vol) isoflurane in oxygen. Ophthalmic ointment (Puralube Vet Ointment, Dechra) was applied to the eyes. The scalp and the ventral torso were shaved and sterilized with Betadine and 70% ethanol. Two electrodes made of saline-filled polyimide tubes (Vention Medical Inc) with 1.5 mm outer diameter and 1.4 mm inner diameter or two electrodes made of adhesive electrogel with 1.5 mm diameter (SignaGel, ParkerLabs Inc.) were affixed to the skull (polyimide tubes were affixed using dental acrylic). During cortical recording, the positions of the skull electrodes relative to bregma were anteroposterior (AP) −1 mm, mediolateral (ML) −1.5 mm, and AP −1 mm, ML −2 mm; during hippocampus recording, their positions relative to bregma were: AP −2 mm, ML 0.25 mm, and AP −2 mm, ML −2.75 mm.
In vivo transcranial stimulation
Transcranial stimulation was applied to anesthetized mice via the two skull electrodes, described above. Each skull electrode was paired with a cloth-base electrode (11 mm diameter conductive area; EL504, BioPac Inc) that was attached to the ventral torso with adhesive electrode gel (for the experiments of Figure 1D–F,Ii; SignaGel, Parker Laboratories Inc; 10–15 mm spacing between the edge of the conductive area of the torso electrodes) or with an adhesive electrode gel based electrode (SignaGel, ParkerLabs Inc.) on the cheeks (for the experiments of Figure 1G–H,Iii; approximately 11 mm diameter conductive area). Stimulation was applied for 1.5–2 s periods, with 0.25–0.5 s duration ramp-on periods and 0–0.5 s duration ramp-off periods, with at least a 2 s rest period between consecutive stimulations.
Whole cell patch clamp recording
In vivo whole cell patching in current clamp mode (i.e., 0 pA holding current) was conducted in the cortex (depth of ~500 μm below the dura) and CA1 layer of the hippocampus (depth of 1131 ±157 μm below the dura) of anesthetized mice with an autopatcher (Kodandaramaiah et al. 2012). Data were acquired using pClamp software (Molecular Devices) at a 400 kHz sampling rate. Patch electrodes were pulled from thin-walled borosilicate glass capillary tubing using a P-97 puller (Sutter Instruments). Tip electrode resistance was 4.6–7.4 MΩ in artificial cerebrospinal fluid (ACSF), containing 126 mM NaCl, 3 mM KCl, 1.25 mM NaH2PO4, 2 mM CaCl2, 2 mM MgSO4, 24 mM NaHCO3 and 10 mM glucose). The patch electrode solution consisted of (in mM) potassium gluconate 122.5, KCl 12.5, KOH-HEPES 10, KOH-EGTA 0.2, Mg-ATP 2, Na3-GTP 0.3, NaCl 8 (pH 7.35, mOsm 296), with 0.2–0.4 mg/ml biocytin added immediately before use. Capacitance, series resistance and input resistance were frequently measured throughout recording to monitor patch quality and cell health, using 10-pA hyperpolarization/depolarization square current pulses; a 300 pA ramp depolarization over 500 ms was used for AP generation.
Data analysis
Data were analyzed using Matlab (MathWorks). The recorded traces from each neuron were split into blocks corresponding to each trial within an experiment. Each block consisted of a single stimulation period of 1.5 −2 s duration with 1 s of baseline recorded before and after each stimulation period. To reduce stimulation artifacts for spike identification, traces were filtered using a 5th order Butterworth band-stop filter with cutoff frequencies of 1 kHz and 15 kHz and then with a 3rd order Butterworth high-pass filter with a cutoff frequency of 100 Hz (representative traces from the cortex are shown in the supplemental materials without filtering and after filtering with only the band-stop filter; representative traces from the hippocampus are shown after filtering with only the band-stop filter). Single spikes were identified using a running window average that picked out depolarizations of >40 mV above baseline, which were “peaky” (that is, exhibited amplitudes larger than the average amplitudes of the nearest 3 data points before and after, by >0.001 mV). Consecutive spikes with inter-spike interval ≤ 15 ms, which occurred during a period of 50 ms or less, were defined as a spiking burst. Mean spiking frequency during stimulation periods (not including the ramping periods) was computed for each stimulation block and then averaged across neurons for each stimulation condition. Mean spontaneous firing rate was computed by a similar averaging of the firing rates across neurons, but for the 1 s interval before stimulation began. In the case of control 1 kHz or 2 kHz stimulation with no TI, we analyzed data from all complete blocks, without a requirement for a minimal number of spikes. Mean membrane potential was computed for a 500 ms period before the onset of 2 kHz or 1 kHz stimulation and was compared with a similar 500 ms period 1 s after stimulation onset, by dividing each period to 10 equally sized epochs and averaging across epochs. Overall, 18 neurons from 8 mice were analyzed with a minimal and maximal number of neurons per mouse of 1 and 4, respectively.
In-vivo stimulation followed by c-fos staining
Surgical procedure
On the day of the experiment, to reduce background fos staining, mice were kept undisturbed in the dark, with full access to food and water, for at least an hour after being transferred from the animal facility to the experimental procedure room. Mice were injected with Meloxicam (1mg/kg) and buprenorphine (0.1mg/kg) and anesthetized with 1–2% (vol/vol) isoflurane in oxygen. The scalp and the ventral torso were shaved and sterilized with Betadine and 70% ethanol. Two cloth-base electrodes with 11 mm diameter conductive area (EL504, BioPac Inc) were attached to the ventral torso with saline electrode gel (SignaGel, Parker Laboratories Inc). The spacing between the edge of the two conductive areas was 10–15 mm. The mice were then placed in a custom stereotax, with ophthalmic ointment (Puralube Vet Ointment, Dechra) applied to the eyes, and with Betadine and 70% ethanol used to sterilize the surgical area. Two polyimide tubes (Vention Medical Inc) with 1.5 mm outer diameter and 1.4 mm inner diameter were affixed to the skull using dental acrylic (C&B Metabond, Parkell). One polyimide tube was located at stereotactic coordinates (relative to bregma) of anteroposterior −2 mm, mediolateral, −0.25 mm. The second polyimide tube was located to the left of the first tube, with a gap between the edges of the electrodes of between 1.5 and 4 mm. Once the dental acrylic set (~20 min), the mice were removed from the stereotactic apparatus and placed in a custom-built low profile holder. The polyimide tubes were filled with saline solution and a silver wire of 0.25 mm diameter (Sigma-Aldrich Co. LLC) was inserted into each tube for connection to the stimulator.
In vivo transcranial stimulation
The cranial tube electrodes and the cloth-base ventral torso electrodes were connected to the stimulator so that each cranial tube electrode was paired with one cloth-base ventral torso electrode. The complex impedance between each pair of electrodes was established by applying short currents of low amplitude (10 Hz and 2 kHz, 10 μA, 0.5 s) and measuring the applied current and resultant voltage waveforms. When the resistance between a pair of electrodes was higher than 1 MΩ, a dental drill was used to thin the skull area inside polyimide tubes (high impedance often resulted from a layer of dental acrylic that leaked into the polyimide tubes and hardened on the skull in that area). 10 s intervals of electrical stimulation, with 0.25 s ramp-on and ramp-off periods, were applied interspersed with 10 s rest intervals over a 20 min experimental timecourse. Mice were sacrificed after 90 min to allow for c-fos expression.
Histology, immunohistochemistry and imaging
Mice were deeply anesthetized with ketamine/xylazine and transcardially perfused with cold phosphate-buffered saline (PBS) followed by cold 4% paraformaldehyde (PFA) in 1 × PBS. The brains were dissected and post-fixed in 4% PFA in PBS at 4°C overnight. Free-floating vibratome coronal sections (35 μm) were cut and incubated in a blocking solution containing 10% normal donkey serum, 0.2% Triton-X 100, 3% bovine serum albumin, and 0.02% sodium azide in 1 × PBS for 2 hrs at room temperature (RT). Sections were labeled with primary anti-c-Fos antibody (1:400; sc-52; Santa Cruz Biotechnology, USA) in the blocking solution at 4°C overnight, followed by the Alexa488-conjugated (1:1000; Invitrogen, ThermoFisher Scientific, USA) secondary antibody for 1 hr at RT. Slices were incubated for 20min in 1 × DAPI dye (Invitrogen, ThermoFisher Scientific, USA) in PBS at RT to label cell nuclei. Samples were then washed 4 × 15 min in PBS with 0.1% Triton-X 100. Immunolabeled brain sections were mounted onto glass slides using ProLong Gold anti-fade reagent (Invitrogen, ThermoFisher Scientific, USA) and stored at −20°C. Images of the cortical and hippocampal areas from the stimulated and unstimulated sides of the brain were acquired using a high-resolution multi-channel (sequential) scanning confocal microscope (LSM 510, Zeiss, Germany), using a 10× air objective (NA 0.45).
c-fos quantification
Greyscale images were analyzed using Matlab (MathWorks). Each greyscale image was processed with contrast-limited adaptive histogram equalization (128 tiles per image) followed by Wiener adaptive noise-removal lowpass filtering (using 5×5 neighboring pixels to estimate the local image mean and standard deviation). DAPI and GFP greyscale images were converted to black and white (BW) masks with global image thresholds established using Otsu’s method. BW masks were smoothed using a morphological disc kernel with a radius of 2 pixels. Masks were visually inspected and Otsu’s threshold was adjusted when required. Cells were quantified from the masked DAPI images. A cell was defined as a region with more than 20 and less than 100 connected pixels. Regions with less than 20 connected pixels were ignored. The number of cells in regions with more than 100 connected pixels was estimated by dividing the number of connected pixels by 100 - the maximal number of connected pixels defined per cell. DAPI cells expressing GFP were quantified from the corresponding GFP BW mask. A cell expressing GFP was defined as a DAPI cell region with connected GFP pixels. The percentage of c-fos expressing cells (a DAPI cell region with connected GFP pixels) was computed in 512 μm × 512 μm regions of interest in the cortex underneath the lateral electrode, the cortex underneath the medial electrode, the cortex between the electrodes, and the dentate gyrus region of the hippocampus, as well as the corresponding contralateral cortical and hippocampal regions.
Brain slice montages
GFP greyscale images were converted to RGB images. Images from a single brain slice were visually rearranged and overlapped to form a montage of the imaged slice.
In-vivo stimulation in awake mice followed by staining for cell and synapse markers
Surgical procedure and animal habituation
On the day of the surgery, the mice were injected with Meloxicam (1mg/kg) and buprenorphine (0.1mg/kg) and were anesthetized with 1–2% (vol/vol) isoflurane in oxygen and placed in a stereotactic frame. The scalp was shaved, ophthalmic ointment (Puralube Vet Ointment, Dechra) was applied to the eyes, and Betadine and 70% ethanol were used to sterilize the surgical area. The scalp was opened and a custom stainless steel headplate was affixed to the skull using dental acrylic (C&B Metabond, Parkell), and the mice were then recovered. Headfixed awake mice were then habituated to restraint for three consecutive days for 15, 30 and 45 min respectively. During this time the animals were placed in a cylinder-shaped tube. Animals were rewarded with sweetened condensed milk (diluted 1:2 in water) every 5–10 min during habituation.
In vivo transcranial stimulation
On the day of stimulation, part of the dental cement was removed to enable cranial electrode placement. The stimulation procedure was as described in the ‘In-vivo stimulation followed by c-Fos staining’ section, but electrodes were made of conductive paste (Ten20, Weaver and Company Inc.) instead of polyimide tubes filled with saline. The centre of one cranial electrode was placed at a point at the midline and −1.5 mm anteroposterior from bregma, and the center of the second cranial electrode was located laterally at +2 mm mediolateral from bregma and at the same anteroposterior coordinate. Each cranial electrode was paired with a larger (approximately 8 mm diameter) electrode that was made of the same conductive paste and was located on the ipsilateral cheek. The complex impedance between each pair of electrodes was established by applying short currents of low amplitude (10 Hz and 2 kHz, 10 μA, 0.5 s) and measuring the applied current and observed voltage waveforms. The stimulation protocol comprised 10 s intervals of electrical stimulation with 0.25 s ramp-on and ramp-off periods, with 10 s rest intervals in between, over a 20 min period. The lateral electrode was driven at a frequency of 2 kHz and current amplitude of 125 μA and the medial electrode was driven at a frequency of 2.01 kHz and a current amplitude of 125 μA. In the case of sham stimulation, the amplitude of both currents was set to 0 μA; the rest of the procedure was identical.
Histology, immunohistochemistry and imaging
Mice were sacrificed and perfused (cold 4% paraformaldehyde in 1 × PBS) 24 h later to assess cellular and synaptic integrity by labelling for the neuronal marker NeuN (1:1000, Synaptic Systems, #266004), astrocyte marker GFAP (1:500, Cell Signalling Technology, #12389), microglia marker Iba1 (1:500, Wako Cehmicals, #019-19741), presynaptic marker synaptophysin (1:500, SVP-38, Sigma, #S5768), apoptosis marker cleaved caspase-3 (1:250, Cell Signalling Technology, #9664) and the DNA damage marker γH2AX (1:500. anti-phospho-histone H2A.X, Millipore, #05-636). For immunostaining, 40 μm sections were incubated with blocking buffer (5% normal donkey serum and 0.3% Triton X-100 in PBS) for 1 h. Primary antibodies were diluted in blocking buffer and incubated with the sections overnight at 4 °C. Primary antibodies were visualized using the appropriate secondary antibody conjugates (Alexa Fluor 488, Alexa Fluor 594 and Alexa Fluor 647, ThermoFisher Scientific). We used the CK-p25 transgenic mouse, an established mouse model of neurodegeneration, which exhibits neuronal atrophy, reduced synaptic density and pronounced DNA damage (Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008)(Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008)(Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008)(Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008)(Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008)(Cruz et al. 2003; Dobbin et al. 2013; Kim et al. 2008), as a positive staining control for yH2AX antibody staining. The samples were then washed, stained with DAPI (Sigma, #10236276001) and mounted onto glass slides. Images were acquired on a Zeiss LSM710 laser-scanning confocal microscope using 20× and 40× air objectives, and subsequently analysed in ImageJ.
In-vivo temperature measurement
Surgical procedure
Surgical procedures were as described in the ‘In-vivo stimulation followed by c-Fos staining’ section.
In vivo transcranial stimulation
Stimulation currents (I1, 2 kHz, 500 μA; I2, 2 kHz, 500 μA) were simultaneously applied with 0.5 ramp-up and ramp-down periods via cranial electrodes that were configured on the skull as described in the ‘In-vivo stimulation in awake mice followed by staining for cell and synapse markers’ section.
Intracranial temperature measurement
A 1 mm diameter thermocouple (type “K” dual 36 Gauge with Teflon insulation. 36 TT-K-36, OMEGA Engineering) was inserted to the brain underneath the lateral electrode via a 2 mm diameter craniotomy, and continuous temperature measurements were obtained with a temperature logger (NI USB TC01, National Instruments Inc.) during 60 s of stimulation as described above, as well as during 30 s periods before and after stimulation.
In-vivo stimulation with spatial probing of motor cortex functionality
Surgical procedure and in vivo transcranial stimulation
On the day of the experiment, mice were anesthetized with 100 mg/kg ketamine and 10 mg/kg xylazine. Ophthalmic ointment (Puralube Vet Ointment, Dechra) was applied to the eyes. The scalp was shaved and sterilized with Betadine and 70% ethanol and the mice were headfixed. Two electrically isolated currents I1 and I2 were applied transcranially via electrodes that were made of conductive paste (1 mm diameter; Ten20, Weaver and Company Inc.) and connected to the stimulator via thin silver wires. Current I1 was applied via the skull electrode that was located at coordinates relative to bregma AP −1.5 mm, ML +2 mm (n=5 mice) or ML −2 mm (n=4 mice). Current I2 was applied via the skull electrode that was located 2.5 mm laterally to the I1 electrode (distance between centers of electrodes). Both skull electrodes were paired with a 5–8 mm diameter electrode, made of adhesive electrode gel (SignaGel, ParkerLabs Inc.), that was attached to the ipsilateral cheek. Stimulation blocks comprised a 0.5 s ramp-up period, a 6 s stimulation period and a 0.5 s ramp-down period. There was approximately 10 s interval between consecutive stimulation blocks.
Recording of motor activity
Motor activities were recorded using a video camera (Fujinon, YV10×5B-2, 1:1 3/5–50mm 1/3″ CS), over a period lasting from 1.5 s pre-stimulation until 1.5 s post-stimulation. The period of stimulation was indicated to the camera with a green LED that was positioned behind a post to avoid a direct illumination of the animal eye.
Data analysis
Motor activities were analyzed offline using Matlab (MathWorks Inc.). Movements of the forepaws and whiskers were measured with the help of an in-frame ruler. In the case of whisker movement, we analyzed movements of the whisker that showed the maximal periodic movement amplitude during stimulation. Movements of the ears were scored by three independent researchers who were blind to the stimulation conditions, per the following scale: 0, no movement; 1, weak movement; 2, strong movement; 3, very strong movement. Prior to scoring movements, the researchers were shown one example video with a weak movement and one example video with a very strong movement, to help calibrate their numerical judgments (these training videos were not included in the analysis).
Phantom electric field measurements
Phantom construction
A phantom was constructed from a 50 mm diameter petri dish. We mounted sixteen 1mm diameter silver wire electrodes with equal spacing along the circumference of the phantom (i.e., an interelectrode spacing of 9.8mm). The electrodes were connected to a 24-channel adapter box that was connected to the TI stimulator. The phantom was filled with sodium chloride solution. The salt concentration was adjusted until an interelectrode impedance of 3 kOhm was measured between two opposite electrodes.
Electric field measurement
The electric field was measured using two orthogonal 3.6 mm-spaced dipole electrodes constructed from medical stainless steel needle electrodes. The location of the probe was adjusted across an 18 mm × 18 mm matrix with 6 mm steps using two large range motorized linear stages (Compumotor NEMA 23 types Model S57-51-MO, Parker Hannifin Corporation). The signal from each dipole electrode was fed into two separate custom built ultra-high input impedance differential electrometer type buffer amplifiers and then the outputs of these amplifiers were differentially fed into lock-in amplifiers (SR830, Stanford Research Systems) before readout with a digital multimeter (3457A, Hewlett Packard). The movements of the probe and the readouts of the measurements from the digital multimeter were controlled by a Labview script. The measurements at each location were averaged several times to reduce noise.
Post-processing
For 2D electric field maps, measurement points were linearly interpolated using Matlab’s interp2 function.
Electromagnetic FEM simulation
Electromagnetic field computation
Electromagnetic simulations were performed on the Sim4Life platform (Zurich MedTech AG) using the ‘stationary current’ solver - a real valued quasielectrostatic finite element method (FEM) solver for the ohmic current dominated regime. The simulation solved the equation ∇σ∇ϕ = 0, where σ is the local electrical conductivity and ϕ is the electric potential from which the electric field and the current density can be obtained as E⃗ = −∇ϕ and J⃗ = σ · E⃗ respectively. The solver is suitable for the frequencies used in this paper, as displacement currents can be neglected compared to ohmic ones. The solver discretizes the model using adaptive, rectilinear meshes. Tissue properties have been assigned according to the IT’IS Foundation database of tissue properties (Hasgall et al. 2015). The same conductivity values were used at all frequencies. The stimulating currents were normalized by integrating the normal current density over a surface surrounding an electrode. The simulations were performed with Dirichlet boundary conditions at active electrodes. The spatial distribution of the envelope modulation amplitude caused by temporal interference was computed from the fields of both electrode pairs using |E⃗AM(↛,⇉)| = ||(E⃗1(⇉) + E⃗2(⇉)) · ↛| − |(E⃗1(⇉) − E⃗2(⇉)) · ↛|| where E⃗1(⇉) and E⃗2(⇉) are the fields generated by the first and second electrode pair, respectively, at the location ⇉(x, y, z) and = ↛ is an unit vector along the direction of interest (e.g. normal to the surface). The maximal envelope modulation amplitude along any orientation which results from the vector fields E⃗1(⇉) and E⃗2(⇉) at location ⇉(x, y, z) was computed. Assuming without loss of generality (as the numbering of the channels can be swapped and the sign of E⃗2 can be inverted) that |E⃗1| > |E⃗2| and that the angle α (angle between E⃗1 and E⃗a) is smaller than , the maximal modulation amplitude is obtained using:
Phantom model
The homogenous phantom model consisted of a saline medium with a conductivity of σ = 0.333 S/m. The inhomogeneous (‘4-layer’) model consisted of scalp (d = 0.05R, σ = 0.333 S/m), skull (d = 0.085R, σ = 0.0083 S/m), cerebrospinal fluid (d = 0.023R, σ = 1.79 S/m) and brain (d = 0.83R, σ = 0.333 S/m) layers, where d is the layer thickness normalized to the overall sphere’s radius R.
Mouse model
A computational mouse model (IT’IS Foundation, Male OF1 Mouse) was developed by segmentation of a male OF1 Mouse, weighing 35.5g, according to the method described in (Kainz et al. 2006). The resolution of the model in the x, y, z directions was 42μm, 42μm and 700μm, respectively. The model did not include CSF. The model was fitted with two small cranial electrodes of radius 0.17 mm filled with saline (σ = 0.333 S/m) and two large surface electrodes of radius 2.2 mm with conductive gel (σ = 1.79 S/m) on the ventral torso. The grid resolution in the electrode vicinity was high (0.04 mm) to properly resolve field gradients, while the coarsest grid step in the exposed area was in the order of 0.12 mm.
QUANTIFICATION AND STATISTICAL ANALYSIS
The reasoning behind our sample sizes is not based upon a power analysis, since the goal was to create a new technology. As noted in (Dell et al. 2002), “In experiments based on the success or failure of a desired goal, the number of animals required is difficult to estimate...” As noted in the aforementioned paper, “The number of animals required is usually estimated by experience instead of by any formal statistical calculation, although the procedures will be terminated [when the goal is achieved].” These numbers reflect our past experience in developing neurotechnologies.
In-vivo rodent electrophysiology (related to Figure 1, Table S1)
Definition of center and dispersion
Spike firing rate (Figure 1Ii), spike firing or bursting rate (Figure 1Iii): shown values are mean ± standard deviation (st.d.).
Definition of n
Number of cells.
Statistical test and definition of significance
Significance (p < 0.05) was characterized by one-way ANOVA followed by post-hoc test with Bonferroni correction for multiple comparisons.
Randomization strategy
The order of stimulation conditions (i.e., 10 Hz stimulation, 2 kHz stimulation, and TI stimulation) was randomized.
Inclusion/exclusion of data
We included all cells that responded to a control 10 Hz stimulation apart from one neuron that was excluded from the hippocampus analysis due to an unphysiologically high rest potential of ~−37 mV.
Statistical details can be found in the legends of Figure 1Ii and Figure 1Iii, and results section ‘Temporal interference stimulation: concept and validation of neural firing recruitment’, and in Table S1.
In-vivo stimulation followed by c-fos staining (related to Figure 3, Figure S4, and Table S2)
Definition of center and dispersion
Percentage of c-fos expressing cells (a DAPI cell region with connected GFP pixels) was computed in 512 μm × 512 μm regions of interest: shown values are mean values ± st.d.
Definition of n
Number of animals.
Statistical test and definition of significance
Significance (p < 0.05) was characterized by one-way ANOVA followed by post-hoc test with Bonferroni correction for multiple comparisons.
Randomization strategy
Each animal was exposed to a single stimulation condition (i.e., 10 Hz stimulation, 2 kHz stimulation, TI stimulation, or TI stimulation with a large inter-electrode distance).
Inclusion/exclusion of data
We included all animals that underwent the stimulation procedure. Statistical details can be found in the legends of Figure 3 and Figure S4, results section ‘Stimulation of mouse hippocampus but not overlying cortex’, and in Table S2.
In-vivo stimulation in awake mice followed by staining for cell and synapse markers (related to Figure 4, Figure S5, Table S3)
Definition of center and dispersion
Immunohistochemical: shown values are mean ± standard error of the mean (s.e.m.).
Definition of n
Number of brain sections.
Statistical test and definition of significance
Significance (p < 0.05) characterized by one-way ANOVA.
Randomization strategy
Each animal was exposed to a single stimulation condition (i.e., TI stimulation or sham).
Inclusion/exclusion of data
We included all animals that underwent the stimulation procedure. Statistical details can be found in the legends of Figure 4, Figure S5, results section ‘Safety characterization of temporal interference stimulation’, and in Table S3. In-vivo temperature measurement (related to Figure 4I)
Definition of center and dispersion
Instantaneous change in brain temperature from baseline: shown values are mean ± st.d.
Definition of n
Number of animals.
Statistical test and definition of significance
Significance (p < 0.05) characterized by one-way ANOVA.
Randomization strategy
N/A.
Inclusion/exclusion of data
We included all animals that underwent the stimulation procedure. Statistical details can be found in the legend of Figure 4I.
In-vivo stimulation with spatial probing of motor cortex functionality (related to Figure 5, Table S4)
Definition of center and dispersion
Motor threshold (Figure 5B–C) shown values are mean ± st.d.; number of responsive animals (Figure 5 panels Di, Ei and Fi); evoked movements (Figure 5 panels Dii, Eii and Fii) shown values are mean ± s.e.m.
Definition of n
Number of animals.
Statistical test and definition of significance
Significance (p < 0.05) was characterized by one-way ANOVA followed by post-hoc test with Bonferroni correction for multiple comparisons.
Randomization strategy
The order of stimulation conditions (i.e., TI stimulation at different difference frequencies, Figure 5B; TI stimulation at different carrier frequencies, Figure 5C; TI stimulation at different current ratios, Figure 5 panels Dii, Eii and Fii) was randomized.
Inclusion/exclusion of data
We included all animals that underwent the stimulation procedure. Statistical details can be found in the legend of Figure 5, results section ‘Steerable probing of motor functionality without electrode movement’, and in Table S4.
All the statistical analyses in this manuscript were performed using Matlab statistics toolbox (Mathworks).
Supplementary Material
Highlights.
Noninvasive TI stimulation electrically stimulates neurons at depth selectively
Neurons are stimulated by interference between multiple electric fields
Neurons in mouse hippocampus can be stimulated without affecting the overlying cortex
Acknowledgments
E.S.B. was funded by NIH Director’s Pioneer Award 1DP1NS087724 and NIH Director’s Transformative Research Award 1R01MH103910-01, the New York Stem Cell Foundation-Robertson Investigator Award, the MIT Center for Brains, Minds, and Machines NSF CCF-1231216, Jeremy and Joyce Wertheimer, Google, NSF CAREER Award CBET 1053233, the MIT Synthetic Intelligence Project, the MIT Media Lab, the MIT McGovern Institute, and the MIT Neurotechnology Fund. N.G. was funded by a Wellcome Trust MIT fellowship. L.H.-T. was funded by the NIH grants 1RF1AG047661 and NS051874, the Belfer Neurodegeneration Consortium, the Glenn Foundation for Aging Research, the Alana Foundation, Cure Alzheimer’s Fund, and the JBP Foundation. A.R. was funded by a NARSAD grant. A.P.L. was partly supported by the Sidney R. Baer Jr. Foundation, the NIH (R01MH100186, R01HD069776, R01NS073601, R21 NS082870, R21 MH099196, R21 NS085491, R21 HD07616), the Football Players Health Study at Harvard University, and Harvard Catalyst | The Harvard Clinical and Translational Science Center (NCRR and the NCATS NIH, UL1 RR025758). We acknowledge, for helpful discussions, A.C. Singer. The content is solely the responsibility of the authors and does not necessarily represent the official views of Harvard Catalyst, Harvard University, and its affiliated academic health care centers, MIT, the National Institutes of Health, or the Sidney R. Baer Jr. Foundation.
Footnotes
Author Contributions
N.G. designed TI stimulator device, simulations and experiments and wrote the paper. D.B. designed and implemented the TI stimulator. N.G., S.B. K. and H.J.S. designed and conducted in-vivo rodent stimulation and electrophysiology experiments. N.G., S.B. K. and A.R. designed and conducted in-vivo rodent stimulation experiments followed by c-fos staining. N.G. and N.D. designed and conducted in-vivo rodent stimulation experiments with evoke motor activities. H.J.S. and N.D. designed and conducted awake in-vivo rodent stimulation experiments, N.D. followed the experiments with staining and imaging for cell and synapses markers. A.C. and E.N designed, developed and conducted electromagnetic FEM simulations in phantom and mouse model. N.K., L.H.T. and A.P.L. oversaw and designed experiments. E.S.B designed devices, experiments, and wrote the paper, and oversaw experiments.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- Chaieb L, Antal A, Paulus W. Transcranial alternating current stimulation in the low kHz range increases motor cortex excitability. Restorative Neurology and Neuroscience. 2011;29(3):167–175. doi: 10.3233/RNN-2011-0589. [DOI] [PubMed] [Google Scholar]
- Chen R, et al. Wireless magnetothermal deep brain stimulation. Science (New York, NY) 2015;347(6229):1477–80. doi: 10.1126/science.1261821. Available at: http://www.ncbi.nlm.nih.gov/pubmed/25765068. [DOI] [PubMed] [Google Scholar]
- Cruz JC, et al. Aberrant Cdk5 activation by p25 triggers pathological events leading to neurodegeneration and neurofibrillary tangles. Neuron. 2003;40(3):471–483. doi: 10.1016/s0896-6273(03)00627-5. [DOI] [PubMed] [Google Scholar]
- Cuellar JM, et al. Effect of high-frequency alternating current on spinal afferent nociceptive transmission. Neuromodulation. 2013;16(4):318–327. doi: 10.1111/ner.12015. [DOI] [PubMed] [Google Scholar]
- D’Amelio M, Sheng M, Cecconi F. Caspase-3 in the central nervous system: Beyond apoptosis. Trends in Neurosciences. 2012;35(11):700–709. doi: 10.1016/j.tins.2012.06.004. [DOI] [PubMed] [Google Scholar]
- Dell RB, Holleran S, Ramakrishnan R. Sample size determination. ILAR journal / National Research Council, Institute of Laboratory Animal Resources. 2002;43(4):207–13. doi: 10.1093/ilar.43.4.207. Available at: http://www.ncbi.nlm.nih.gov/pubmed/12391396%5Cnhttp://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=PMC3275906. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dobbin MM, et al. SIRT1 collaborates with ATM and HDAC1 to maintain genomic stability in neurons. Nat Neurosci. 2013;16(8):1008–1015. doi: 10.1038/nn.3460. Available at: http://www.ncbi.nlm.nih.gov/pubmed/23852118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dragunow M, Robertson Ha. Kindling stimulation induces c-fos protein(s) in granule cells of the rat dentate gyrus. Nature. 1987;329(6138):441–442. doi: 10.1038/329441a0. [DOI] [PubMed] [Google Scholar]
- Eng LF, Ghirnikar RS, Lee YL. Glial Fibrillary Acidic Protein : GFAP-Thirty-One Years (1969–2000) Neurochemical Research. 2000;25(9–10):1439–1451. doi: 10.1023/a:1007677003387. [DOI] [PubMed] [Google Scholar]
- Gradinaru V, et al. Optical deconstruction of parkinsonian neural circuitry. Science (New York, NY) 2009;324(5925):354–9. doi: 10.1126/science.1167093. Available at: http://www.ncbi.nlm.nih.gov/pubmed/19299587. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Greenberg BD, et al. Deep brain stimulation of the ventral internal capsule/ventral striatum for obsessive-compulsive disorder: worldwide experience. Molecular psychiatry. 2010;15(1):64–79. doi: 10.1038/mp.2008.55. Available at: http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=3790898&tool=pmcentrez&rendertype=abstract. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hasgall P, et al. IT’IS Database for thermal and electromagnetic parameters of biological tissues. Version 3.0. 2015 Available at: www.itis.ethz.ch/database.
- Histed MH, Bonin V, Reid RC. Direct Activation of Sparse, Distributed Populations of Cortical Neurons by Electrical Microstimulation. Neuron. 2009;63(4):508–522. doi: 10.1016/j.neuron.2009.07.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hutcheon B, Yarom Y. Resonance, oscillation and the intrinsic frequency preferences of neurons. Trends in Neurosciences. 2000;23(5):216–222. doi: 10.1016/s0166-2236(00)01547-2. [DOI] [PubMed] [Google Scholar]
- International Commission on Non-Ionizing Radiation Protection. Guidelines for limiting exposure to time-varying electric and magnetic fields (1 Hz to 100 kHz) Health physics. 2009;99(6):818–36. doi: 10.1097/HP.0b013e3181f06c86. Available at: http://www.ncbi.nlm.nih.gov/pubmed/21068601%5Cnhttps://drive.google.com/open?id=0B-_EfabIdYlFTkU4NmlYMVBsVjA&authuser=0%5Cnhttps://drive.google.com/open?id=0B-_EfabIdYlFZDFlamwzZTM5Ym8&authuser=0. [DOI] [PubMed] [Google Scholar]
- Ito D, et al. Microglia-specific localisation of a novel calcium binding protein, Iba1. Molecular Brain Research. 1998;57(1):1–9. doi: 10.1016/s0169-328x(98)00040-0. [DOI] [PubMed] [Google Scholar]
- Kainz W, et al. Development of novel whole-body exposure setups for rats providing high efficiency, National Toxicology Program (NTP) compatibility and well-characterized exposure. Physics in Medicine and Biology. 2006;51(20):5211–5229. doi: 10.1088/0031-9155/51/20/009. Available at: http://stacks.iop.org/0031-9155/51/i=20/a=009?key=crossref.7ca12003586f2936a11716e77ecdc2ac. [DOI] [PubMed] [Google Scholar]
- Kalia SK, Sankar T, Lozano AM. Deep brain stimulation for Parkinson’s disease and other movement disorders. Current opinion in neurology. 2013;26(4):374–80. doi: 10.1097/WCO.0b013e3283632d08. Available at: http://www.ncbi.nlm.nih.gov/pubmed/23817213. [DOI] [PubMed] [Google Scholar]
- Kilgore KL, Bhadra N. Reversible nerve conduction block using kilohertz frequency alternating current. Neuromodulation. 2014;17(3):242–254. doi: 10.1111/ner.12100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim D, et al. Deregulation of HDAC1 by p25/Cdk5 in Neurotoxicity. Neuron. 2008;60(5):803–817. doi: 10.1016/j.neuron.2008.10.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kodandaramaiah SB, et al. Automated whole-cell patch-clamp electrophysiology of neurons in vivo. Nature methods. 2012;9(6):585–7. doi: 10.1038/nmeth.1993. Available at: http://dx.doi.org/10.1038/nmeth.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kringelbach ML, et al. Translational principles of deep brain stimulation. Nature reviews. Neuroscience. 2007;8(8):623–635. doi: 10.1038/nrn2196. [DOI] [PubMed] [Google Scholar]
- Kunz P, et al. 5 kHz transcranial alternating current stimulation: lack of cortical excitability changes when grouped in a theta burst pattern. Front Hum Neurosci. 2016;10(683) doi: 10.3389/fnhum.2016.00683. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Legon W, et al. Transcranial focused ultrasound modulates the activity of primary somatosensory cortex in humans. Nature neuroscience. 2014;17(2):322–9. doi: 10.1038/nn.3620. Available at: http://www.ncbi.nlm.nih.gov/pubmed/24413698. [DOI] [PubMed] [Google Scholar]
- Mah L-J, El-Osta a, Karagiannis TC. gammaH2AX: a sensitive molecular marker of DNA damage and repair. Leukemia : official journal of the Leukemia Society of America, Leukemia Research Fund, U.K. 2010;24(4):679–686. doi: 10.1038/leu.2010.6. Available at: http://dx.doi.org/10.1038/leu.2010.6. [DOI] [PubMed] [Google Scholar]
- Merrill DR, Bikson M, Jefferys JGR. Electrical stimulation of excitable tissue: Design of efficacious and safe protocols. Journal of Neuroscience Methods. 2005;141(2):171–198. doi: 10.1016/j.jneumeth.2004.10.020. [DOI] [PubMed] [Google Scholar]
- Miranda PC, et al. The electric field in the cortex during transcranial current stimulation. NeuroImage. 2013;70:48–58. doi: 10.1016/j.neuroimage.2012.12.034. [DOI] [PubMed] [Google Scholar]
- Peratta C, Peratta A. Modelling the human body exposure to elf electric fields. WIT Press; 2010. [DOI] [PubMed] [Google Scholar]
- Plaksin M, Shoham S, Kimmel E. Intramembrane cavitation as a predictive bio-piezoelectric mechanism for ultrasonic brain stimulation. Physical Review X. 2014;4(1) [Google Scholar]
- Reijmers LG, et al. Localization of a stable neural correlate of associative memory. Science (New York, NY) 2007;317(5842):1230–1233. doi: 10.1126/science.1143839. [DOI] [PubMed] [Google Scholar]
- Stavroulakis P, editor. Biological Effects of Electromagnetic Fields Mechanisms, Modeling, Biological Effects, Therapeutic Effects, International Standards, Exposure Criteria. Springer; Berlin Heidelberg: 2014. [Google Scholar]
- Tarsa L, Goda Y. Synaptophysin regulates activity-dependent synapse formation in cultured hippocampal neurons. Proceedings of the National Academy of Sciences of the United States of America. 2002;99(2):1012–6. doi: 10.1073/pnas.022575999. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11792847%5Cnhttp://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=PMC117422. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tennant KA, et al. The Organization of the Forelimb Representation of the C57BL/6 Mouse Motor Cortex as Defined by Intracortical Microstimulation and Cytoarchitecture. Cerebral Cortex. 2011;(21):865–876. doi: 10.1093/cercor/bhq159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wolf HK, et al. NeuN: a useful neuronal marker for diagnostic histopathology. The journal of histochemistry and cytochemistry : official journal of the Histochemistry Society. 1996;44(10):1167–1171. doi: 10.1177/44.10.8813082. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.