Abstract
Study Design
Cadaveric biomechanical study.
Objectives
Quantify the effects of vertebral body augmentation on biomechanics under axial compression by a total disc replacement (TDR) implant.
Summary of Background Data
TDR is a surgical alternative to lumbar spinal fusion to treat degenerative disc disease. Osteoporosis in the adjacent vertebrae to the interposed TDR may lead to implant subsidence or vertebral body fracture. Vertebral augmentation is used to treat osteoporotic compression fracture. The study sought to evaluate whether vertebral augmentation improves biomechanics under TDR axial loading.
Methods
Forty-five L1-L5 lumbar vertebral body segments with intact posterior elements were used. Peripheral quantitative computed tomography scans were performed to determine bone density, block randomizing specimens by bone density into augmentation and control groups. A semi-constrained keeled lumbar disc replacement device was implanted providing 50% endplate coverage. Vertebral augmentation of 17.6 ± 0.9% vertebral volume fill with Cortoss was performed on augmentation group. All segments underwent axial compression at a rate of 0.2 mm/s to 6mm.
Results
The load-displacement response for all specimens was non-linear. Subfailure mechanical properties with augmentation were significantly different from control; in all cases the augmented group was 2× higher than control. At failure, the maximum load and stiffness with augmentation was not significantly different from control. The maximum apparent stress and modulus with augmentation were 2× and 1.3× greater than control, respectively. The subfailure stress and apparent modulus with augmentation was moderately correlated with bone density while the control subfailure properties were not. The augmented maximum stress was not correlated with bone density, while the control was weakly correlated. The maximum apparent modulus was moderately correlated with bone density for both the augmented and control groups.
Conclusion
Augmentation improved the mechanical properties of the lumbar vertebral body for compression by a TDR implant.
Introduction
Total disc replacement (TDR) is a viable surgical alternative to lumbar spinal fusion to treat painful degenerative disc disease (DDD). The procedure is FDA approved for use in the United States (2006) with long-term prospective studies1 in addition to having a broad application worldwide with over ten years of follow-up reported 2-4. A relevant concern among the current population of TDR patients is the future development of osteoporosis. Osteoporosis is an age-related progressive skeletal disease characterized by low bone mass, micro-architectural deterioration, a decrease in bone strength, and an increased susceptibility to fracture5. While osteoporosis is currently a contraindication for TDR 1,6,7, many of the aging population of current TDR patients will predictably become osteoporotic. Osteoporosis in the adjacent vertebrae to the interposed TDR presents a potential clinical problem of implant subsidence or vertebral body fracture 7,8. Subsidence, defined as a 2 mm deformation into the vertebral body, is dependent upon, among other factors, the stiffness and strength of the underlying trabecular bone and the endplate-implant interface.
Osteoporosis is responsible for 700,000 compressive vertebral body fractures annually in the US and osteoporotic vertebral fractures occur in approximately 20% of individuals over the age of 70 9,10. Vertebroplasty, percutaneous injection of cement into a collapsed vertebral body, was introduced in 1984 for treating vertebral haemangiomas and osteolytic neoplasms 5. Clinical usage of bone cement in the spine via vertebroplasty and kyphoplasty procedures has since become an accepted medical practice for treating osteoporotic compression fracture11,12,5,13,14. Injecting a material into the trabecular interior of the vertebral body provides strength and support. The mechanical stabilization by augmentation may reduce fracture-site motion and mitigate pain 15. Several materials have been used for augmentation, including polymethylmethacrylate (PMMA) bone cement and a synthetic bone filler Cortoss (Orthovita, Inc ™, Malvern, PA), a bioactive glass-ceramic composite material with a matrix of bis-phenol glycidyl dimethacrylate, bis-phenol ethoxy dimethacrylate, and triethyleneglycol dimethacrylate 15-17. Cortoss is biocompatible, exhibits a lower setting exotherm than PMMA, is bone bonding, and has material properties more similar to bone 15-17. Cortoss is both a non-biologic mechanical reinforcing agent and has properties of bonding to bone (like a biologic agent). There is no evidence in the literature or clinically that metabolic bone disease is associated with the use of Cortoss. Further to the reality of progressive osteoporosis, the use of a reinforcing agent (of either type) will minimize the risk of the natural disease progression creating a non-indicated (osteoporotic) condition for TDRs. Cortoss has completed IDE investigation in a multi-center prospective randomized study, and been approved by the FDA for use in vertebral augmentation in the US, and is marketed for use in Europe and Australia.
Laboratory studies using cadaveric human segments show that vertebral body augmentation increases strength under axial compression 13,15,18. Importantly, Higgins et al. augmented aged intact (non-fractured) osteoporotic segments with PMMA, increasing the failure strength over unfilled controls 18. Tan et al. injected the pedicle screw tracts with PMMA for augmenting below an interbody fusion device 13. Augmentation significantly increased failure load and strength under compression by the interbody device compared with un-augmented pedicle screws 13. These results suggested that cement augmentation of pedicle screws may reduce interbody device subsidence in osteoporotic patients 13. Cement augmentation into vertebral bodies and pedicle screw tracts have gained clinical application to reduce interbody fusion device subsidence 13,14,19.
Building on early clinical experience with two cases of subsidence, one of the current authors (Dr. Bertagnoli) routinely performs prophylactic vertebral body augmentation with bone cement in selected patients following TDR implant placement 6. While early clinical evidence suggests that subsidence or compressive fracture can be avoided by vertebral augmentation in TDR patients with osteopenia and osteoporosis 6, the biomechanical basis to support TDR vertebral augmentation has not been established. The objective of this study was to quantify the effects of vertebral augmentation with Cortoss on vertebral body mechanics under axial compression by a TDR implant. We hypothesize that augmentation will increase the compressive stiffness and strength of the endplate-implant interface in a bone density-dependent manner.
Materials and Methods
Nine fresh-frozen lumbar spines were obtained (6 female and 3 male donors, average age 79, range 69 – 88 years), providing a total of 45 L1-L5 lumbar levels. Antero-posterior and lateral radiographs were performed on intact spines to exclude segments based on structural deformities such as fractures and endplate irregularities (n=3 excluded), leaving 42 specimens. Peripheral quantitative computed tomography (pQCT, XCT2000, Stratec Medisintechnik, Pforzheim, Germany) scans were performed on the intact lumbar spines to determine the vertebral body bone density. Three images were acquired below the bony endplate based and the trabecular bone density within a 10 mm diameter region of interest were averaged (Figure 1). Specimens were block randomized according to bone density and divided into two groups, augmented (n=22) and control (n=20).
Figure 1.
Representative peripheral quantitative computed tomography (pQCT) scan of lumbar vertebral specimen. Trabecular bone density was evaluated in a 10 mm diameter region of interest (white circle).
Specimen Preparation
Single level vertebral body segments were prepared by excising the intervertebral disc tissue. The posterior elements were kept intact to maintain vertebral body integrity. The antero-posterior width and medial-lateral width of the superior endplate and the height of the vertebral body were measured three times with digital calipers and averaged. The endplate vertebral body was modeled as an elliptical cylinder to calculate the endplate area and vertebral body volume 18,20. A curette was used to remove the remaining cartilaginous endplate, taking care to keep the bony endplate intact. Endplate preparation has little effect on the yield or ultimate compressive strength of vertebral bodies 21. The inferior endplate was prepared in the same fashion and the sample was potted superior surface up in a cylinder with PMMA to cover 2-3 mm of the vertebral body. A level was used to ensure a flat testing surface. Each specimen was then stored in vacuum sealed bag and refrozen at -20°C.
Prior to augmentation and testing each specimen was thawed at 3°C for 24 hours. Medial-lateral fluoroscopic views were taken pre-augmentation, post-augmentation, and post-testing. A semi-constrained keeled lumbar disc replacement device sized to provided approximately 50% endplate coverage was implanted using techniques similar to clinical procedures 1. The goal of TDR insertion, similar to that of interbody fusion cages, is to place the implant with maximal apophseal rim coverage to prevent subsidence, 50% endplate coverage was chosen as a best estimate of what is feasible to place clinically. Tan et al. found that endplate coverage of 40% was biomechanically superior to 20%22. Anterior placement of an interbody implant has been shown to lead to early failure 23; while a posterior placement increased the implant-vertebral body unit strength up to 20%24. Ensuring 50% endplate coverage along with posterior placement our study minimizes the variability that is introduced by different endplate coverage options, and allows us to isolate the effect of the augmentation.
Each augmented specimen was held at 37°C physiologic temperature in phosphate buffered saline (PBS) for 20 minutes prior to injection to ensure flow and dispersion. Fill volume was determined based upon dividing the amount of material injected by the anterior 2/3 vertebral body volume with a factorization for trabecular bone. Vertebral augmentation (17.6 ± 0.9% vertebral volume fill) with Cortoss was performed via two posteriomedially-directed 16G needles placed under the lateral borders of the implant (Figure 2A). The fill volume was selected based on experimental and modeling studies recommending 15-20% fill for vertebral augmentation to restore vertebral body stiffness and strength following compression fracture or in the case of osteoporosis 18,25. Fluroscopic guidance was utilized to prevent material extravasation during augmentation, this is commonly utilized clinically during vertebroplasty to achieve optimal device placement. The TDR implant was inserted and the sample wrapped in 1% PBS soaked gauze then kept at 37°C for 24 hours allowing for the Cortoss to harden. The control group underwent the same procedures except for no needle insertion or augmentation.
Figure 2.
Fluoroscopic images A) pre-augmentation (Arrow indicates point of Cortoss injection), B) post-augmentation, C) post-testing
Mechanical Testing and Data Analysis
Axial compression at a rate of 0.2 mm/s to 6 mm was performed in an Instron 8874 (Instron Corporation, Canton, MA) and load and displacement were recorded 22. The 6 mm amount of total displacement corresponds to the clinical radiographic diagnosis of a compression fracture, which is a 25% collapse 26. Since load-displacement is nonlinear, mechanical properties were calculated in 2 regions: subfailure and maximum. The subfailure load was determined at 2 mm displacement, which corresponds to the clinical definition of implant subsidence27,28. Subfailure stiffness was calculated by linear regression as the slope of the load-displacement response between 200 N and 400 N. Maximum load was taken as the peak in the load-displacement curve. In some cases the load-displacement response exhibited an early drop in load; as long as the load continued to rise over the subsequent 0.5 mm, these loads were not recorded as the maximum load. Similar definitions permitting a small drop in load that does not represent failure are established in the literature 29,30. Stiffness at max was calculated as the slope of the load-displacement response over a 200 N range 20% from the max load. Apparent material properties at subfailure and maximum were calculated by normalizing for geometry. Apparent stress was calculated as load divided by cross-sectional area. Apparent modulus was calculated as stiffness multiplied by initial specimen height divided by cross-sectional area.
Statistical Analysis
Statistical analysis included a t-test for an effect of augmentation (treatment) compared to intact control for each mechanical property. A Pearson's correlation was performed between bone density with apparent stress and modulus, separately for control and augmented groups. Significance was set at p<0.05 and a trend was defined as 0.05≤p<0.1.
Results
Fluoroscopic images confirmed fill and repeatability of the augmentation technique. The only difference observed between pre-augmentation (Figure 2A) and post-augmentation (Figure 2B) is the removal of the remaining intervertebral disc. Extravasation of material occurred through the superior endplate and vertebral foramen during a few of the augmentation procedures, however based on previous studies, extravasation does not alter compression mechanics 18. Cortoss maintained its structural integrity between post-augmentation (Figure 2B) and post-testing (Figure 2C)
The load-displacement response for all specimens was non-linear (Figure 3). Subfailure mechanical properties with augmentation were significantly different from control (Figure 4); in all cases the augmented group was significantly higher than control. The augmented subfailure load was 1795 N, 2X greater than the intact control (Figure 4A). The augmented stiffness was 698 N/m, 2X greater than control (Figure 4B). Upon normalization for geometry, the augmented subfailure stress was 1 MPa, 2X control (Figure 4C) and modulus were 10 MPa, 2X control (Figure 4D).
Figure 3.
Representative load-displacement response from three specimens demonstrating variability in maximum load. Most samples exhibited the “typical” response of uniform load to failure. Some samples exhibited a small break in the curve (square) or an early drop followed by a continued increase to failure. The maximum load determined for each sample is indicated by #.
Figure 4.
Average (standard deviation) sub-failure and maximum properties for control (white) and augmented (black) specimens. A) load, B) stiffness, C) apparent stress, and D) apparent modulus, *p≤0.05.
At failure, with augmentation the maximum load (2311 N, Figure 4A) and stiffness (840 N/mm, Figure 4B) was not significantly different from control (p=0.1 and 0.2). With augmentation the maximum apparent stress (1.4 MPa, Figure 4C) and apparent modulus (13.6 MPa, Figure 4D) were 2X (p=0.08) and 1.3X (p=0.07), respectively, greater than control.
The subfailure stress with augmentation was moderately correlated with bone density (R=0.61, p<0.05, Figure 5A), as was the augmented subfailure apparent modulus (R=0.44, p<0.05, Figure 5B). The control subfailure properties were not significantly correlated with bone density. In contrast, the augmented maximum stress was not correlated with bone density, while the control was weakly correlated (R=0.22, p<0.05, Figure 5C). The maximum apparent modulus was moderately correlated with bone density for both the augmented and control groups (R=0.71 and 0.66, respectively, p<0.05, Figure 5D).
Figure 5.
Correlation with bone density for apparent material properties, normalized by geometry. A) sub-failure stress, (B) sub-failure modulus, (C) maximum stress, (D) maximum modulus, *p≤0.05.
Discussion
The interface between the TDR implant and the bony endplate needs to provide axial compressive strength sufficient to resist implant subsidence or vertebral fracture, as was previously established for spinal interbody fusion implants 21,22. Augmentation improved the mechanical properties of the lumbar vertebral body for compression by a TDR implant. This finding is consistent with prior studies that have shown vertebral augmentation to improve compressive mechanics and restore mechanical properties towards pre-fracture values31 Adjacent level risks are a factor32 however internal stresses of the intervertebral disc are unaffected after vertebroplasty. The use of CORTOSS provides better material dispersion within a vertebral body compared to PMMA, resulting in fewer intervertebral stress concentrations, mitigating the incidence of secondary fractures 33.
While the age of the subjects were outside the range of those who would typically receive a total disc replacement, this was intentional, as the lower bone density represented the aging TDR population. Values for maximum stiffness and load were consistent with previous studies (stiffness 400-800 N/mm, load 300-1500 N) 13,18,22,29,34. Osteoporosis is a looming clinical problem for an aging population with risk of subsidence or vertebral fracture. By improving the overall strength of the vertebral body and creating a larger surface area for the implant will allow for improved bone-implant strength at the time of initial implantation.
Subfailure properties were significantly increased with augmentation, a finding which has not been previously reported in vertebral body augmentation studies. Consistent with previous studies, augmenting the vertebral body improved the maximum (failure) load and stress but did not improve the maximum stiffness and modulus 13,18; while not significant the average stiffness and modulus increased in the present and previous studies. The mechanics of the vertebral body are highly variable, due to age and bone density variation in the sample population, which may explain some of the limited significance in cadaver studies.
Effect of augmentation on subfailure mechanics
The bony endplate is a concave and thin shelf less than ½ mm thick. While it inherently provides little strength, it serves to distribute axially-applied loads more evenly over the underlying trabecular bone 30. Due to its concavity, under the initial compression loading, only the periphery of the implant is in contact with the bony endplate. As loading continues, it compresses the endplate and subsides to some degree before it achieves more contact with the underlying bone. Since in the sub-failure regime (at 2 mm displacement) the control stress and modulus did not correlate with trabecular bone density, these subfailure properties are likely dependent on the endplate and posterior rim, not the underlying trabecular bone. With augmentation, the Cortoss likely provided a buttressing support to the endplate. At this point, the implant itself acts as the endplate, distributing load over the trabeculae 30. This notion is supported by previous imaging studies during interbody device compression, where it was observed that the trabecular bone undergoes densification below the point of compression 22. Augmentation below the endplate as performed in the present study would further enhance the mechanics in the early displacement regime.
Effect of augmentation on failure mechanics
It is well known that trabecular bone strength is correlated with bone density35,36,18,21,37,38, as was observed in the present study. When augmented, the maximum stress was no longer correlated with density, as previously observed following augmentation of osteoportic segments 18. These findings confirm that augmentation increases the functional support provided by trabecular bone in the low density specimens. The performance of interbody fusion devices subsidence are also compromised by low bone mineral density 13,38-40, where in compression, trabecular bone failure occurs in a semi-elliptical zone underlying the interbody implant 22. The observed failure below the implant suggests that local cement augmentation could increase the strength at the implant-vertebra interface 13. Similarly, cement injection into pedicle screw tracts provides structural reinforcement to surrounding trabeculae. Together, the stiffer cement-bone composite located directly below the superior EP redistributes load onto large volumes of supporting trabecular bone and increases failure strength 13.
Determination of failure load
Mechanical testing yielded three types of responses, a smooth rise to failure which was the typical response, a break in the data curve and an early drop in the data prior to the max load being achieved. This variability in shape is consistent with previous vertebral body compression studies 41,29,30,22. Similar variability has been seen during interbody device testing 21,34. The variability in load-displacement response (Figure 3) is likely due to implant settling and/or breaking through the posterior rim; this was observed during fluoroscopic video taken during mechanical testing. Determination of the failure load from the load – displacement curve can be difficult and several different definitions have been used. Some studies define failure from the first peak or plateau in the curve, 21,22,34. However, when considering the response in Figure 3, it isn't convincing that the early drop defines the specimen failure. Other studies use the overall peak load. However, with significant sub-catastrophic failure occurring over large displacements, this may overestimate the effective load carrying capacity. When the peak load was difficult to identify, the inflection point at the elastic limit has been used to identify the beginning of the failure mode 18, however, this point is difficult to identify and its selection can be subjective. The definition of maximum failure load utilized in the present study, the value to which load fails to rise over the subsequent 0.5 mm, allows for equal comparisons to other vertebral body compression studies and is consistent with previous studies where failure was defined as the maximum load before a load decrease of greater than 5% 30.
The measurement of bone mineral density
The compressive strength of bone can be accurately predicted by bone mineral density measurements made using dual energy X-ray asorptiometry (DXA), quantitative computed tomorgraphy (QCT) or peripheral QCT (pQCT) 42-45. The two-dimensional nature of DXA is influenced by surrounding non-vertebral tissue such as ribs, fat, aortic calcification, vertebral osteophytes, and posterior elements. The absence of body tissues such as fat limits the correspondence between vertebral DXA bone density measured in vitro and clinical in vivo density by over 40%46,13,42,47. Bone density was measured in the present study using pQCT, because with pQCT the region of interest can be placed in the three-dimensional space containing only trabecular bone and exclude irrelevant tissues 42,47.
In summary, augmentation improved the mechanical properties of the lumbar vertebral body for compression by a TDR implant. Subfailure properties correlated with bone density when augmented, maximum stress correlated with bone density in control only. These laboratory findings provide mechanical support for the early clinical evidence that prophylactic vertebral body augmentation with bone cement at the time of TDR implant placement may avoid future subsidence or compressive fracture in TDR patients 6. Furthermore, biomaterials developed to augment trabecular bone strength may serve as treatment to prevent further implant subsidence, and potentially future device failure, when detected on radiographs in the course of routine follow-up examinations. While detailed discussion of revision lumbar TDR strategies is beyond the scope of the current study, others have discussed in detail the treatment options in this challenging clinical situation48,49. Options include revision anterior spinal surgery with replacement of a new device, revision to an anterior spinal fusion using femoral ring allograft, coupled with posterior pedicle screw-based fixation, or posterior-only fixation/stabilization with pedicle screws. Longer-term follow-up from clinical studies will be required to determine the optimal treatment and to generate evidence-based guidelines. It remains to be determined whether this technique may be valuable to treat TDR patients that later develop osteoporosis or may be a technique to overcome current exclusion criteria for TDR in osteoporotic patients, which could extend disc replacement surgery to a more broad patient population in the future.
Acknowledgements
Supported by funds from Orthovita and by the Penn Center for Musculoskeletal Disorders
References
- 1.Zigler JE. Lumbar spine arthroplasty using the ProDisc II. Spine J. 2004;4:260S–7S. doi: 10.1016/j.spinee.2004.07.018. [DOI] [PubMed] [Google Scholar]
- 2.David T. Long-term results of one-level lumbar arthroplasty: minimum 10-year follow-up of the CHARITE artificial disc in 106 patients. Spine. 2007;32:661–6. doi: 10.1097/01.brs.0000257554.67505.45. [DOI] [PubMed] [Google Scholar]
- 3.Lemaire JP, Carrier H, Sariali el H, et al. Clinical and radiological outcomes with the Charite artificial disc: a 10-year minimum follow-up. J Spinal Disord Tech. 2005;18:353–9. doi: 10.1097/01.bsd.0000172361.07479.6b. [DOI] [PubMed] [Google Scholar]
- 4.Tropiano P, Huang RC, Girardi FP, et al. Lumbar total disc replacement. Seven to eleven-year follow-up. J Bone Joint Surg Am. 2005;87:490–6. doi: 10.2106/JBJS.C.01345. [DOI] [PubMed] [Google Scholar]
- 5.Ploeg WT, Veldhuizen AG, The B, et al. Percutaneous vertebroplasty as a treatment for osteoporotic vertebral compression fractures: a systematic review. Eur Spine J. 2006;15:1749–58. doi: 10.1007/s00586-006-0159-z. [DOI] [PubMed] [Google Scholar]
- 6.Bertagnoli R, Yue JJ, Nanieva R, et al. Lumbar total disc arthroplasty in patients older than 60 years of age: a prospective study of the ProDisc prosthesis with 2-year minimum follow-up period. J Neurosurg Spine. 2006;4:85–90. doi: 10.3171/spi.2006.4.2.85. [DOI] [PubMed] [Google Scholar]
- 7.Thierry D. Long-term Results of One-Level Lumbar Arthroplasty: Minimum 10-Year Follow-up of the CHARITE Artificial Disc in 106 Patients. Spine. 2007;32:661–6. doi: 10.1097/01.brs.0000257554.67505.45. [DOI] [PubMed] [Google Scholar]
- 8.van Ooij A, Oner FC, Verbout AJ. Complications of artificial disc replacement: a report of 27 patients with the SB Charite disc. J Spinal Disord Tech. 2003;16:369–83. doi: 10.1097/00024720-200308000-00009. [DOI] [PubMed] [Google Scholar]
- 9.Cohen LD. Fractures of the osteoporotic spine. Orthopedic Clinics of North America. 1990;21:143–50. [PubMed] [Google Scholar]
- 10.Riggs BL, Melton LJ., 3rd The worldwide problem of osteoporosis: insights afforded by epidemiology. Bone. 1995;17:505S–11S. doi: 10.1016/8756-3282(95)00258-4. [DOI] [PubMed] [Google Scholar]
- 11.Barr JD, Barr MS, Lemley TJ, et al. Percutaneous vertebroplasty for pain relief and spinal stabilization. Spine. 2000;25:923–8. doi: 10.1097/00007632-200004150-00005. [DOI] [PubMed] [Google Scholar]
- 12.Phillips FM, An H, Kang JD, et al. Biologic treatment for intervertebral disc degeneration: summary statement. Spine. 2003;28:S99. doi: 10.1097/01.BRS.0000076906.82028.03. [DOI] [PubMed] [Google Scholar]
- 13.Tan JS, Bailey CS, Dvorak MF, et al. Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body. Spine. 2007;32:334–41. doi: 10.1097/01.brs.0000253645.24141.21. [DOI] [PubMed] [Google Scholar]
- 14.Heini PF. The current treatment--a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon's perspective. Osteoporos Int. 2005;16(Suppl 2):S85–92. doi: 10.1007/s00198-004-1723-1. [DOI] [PubMed] [Google Scholar]
- 15.Belkoff SM, Mathis JM, Erbe EM, et al. Biomechanical evaluation of a new bone cement for use in vertebroplasty. Spine. 2000;25:1061–4. doi: 10.1097/00007632-200005010-00004. [DOI] [PubMed] [Google Scholar]
- 16.Erbe EM, Clineff TD, Gualtieri G. Comparison of a new bisphenol-a-glycidyl dimethacrylate-based cortical bone void filler with polymethyl methacrylate. Eur Spine J. 2001;10(Suppl 2):S147–52. doi: 10.1007/s005860100288. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Erbe EM, Pomrink GJ, Murphy JP. A comparisaon of mechanical properties of a new synthetic cortical bone void filler (Cortoss/Orthovita) and those of polymethyl methacrylate. Eur Spine J. 2000;9:288. [Google Scholar]
- 18.Higgins KB, Harten RD, Langrana NA, et al. Biomechanical effects of unipedicular vertebroplasty on intact vertebrae. Spine. 2003;28:1540–7. discussion 8. [PubMed] [Google Scholar]
- 19.Wuisman PI, Van Dijk M, Staal H, et al. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. Eur Spine J. 2000;9:528–33. doi: 10.1007/s005860000169. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Panjabi MM, Goel V, Oxland T, et al. Human lumbar vertebrae. Quantitative three-dimensional anatomy. Spine. 1992;17:299–306. doi: 10.1097/00007632-199203000-00010. [DOI] [PubMed] [Google Scholar]
- 21.Steffen T, Tsantrizos A, Aebi M. Effect of implant design and endplate preparation on the compressive strength of interbody fusion constructs. Spine. 2000;25:1077–84. doi: 10.1097/00007632-200005010-00007. [DOI] [PubMed] [Google Scholar]
- 22.Tan JS, Bailey CS, Dvorak MF, et al. Interbody device shape and size are important to strengthen the vertebra-implant interface. Spine. 2005;30:638–44. doi: 10.1097/01.brs.0000155419.24198.35. [DOI] [PubMed] [Google Scholar]
- 23.Polikeit A, Ferguson SJ, Nolte LP, et al. Factors influencing stresses in the lumbar spine after the insertion of intervertebral cages: finite element analysis. Eur Spine J. 2003;12:413–20. doi: 10.1007/s00586-002-0505-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Labrom RD, Tan JS, Reilly CW, et al. The effect of interbody cage positioning on lumbosacral vertebral endplate failure in compression. Spine. 2005;30:E556–61. doi: 10.1097/01.brs.0000181053.38677.c2. [DOI] [PubMed] [Google Scholar]
- 25.Liebschner MA, Rosenberg WS, Keaveny TM. Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine. 2001;26:1547–54. doi: 10.1097/00007632-200107150-00009. [DOI] [PubMed] [Google Scholar]
- 26.Schildhauer TA, Bennett AP, Wright TM, et al. Intravertebral body reconstruction with an injectable in situ-setting carbonated apatite: biomechanical evaluation of a minimally invasive technique. J Orthop Res. 1999;17:67–72. doi: 10.1002/jor.1100170111. [DOI] [PubMed] [Google Scholar]
- 27.Beutler WJ, Peppelman WC., Jr Anterior lumbar fusion with paired BAK standard and paired BAK Proximity cages: subsidence incidence, subsidence factors, and clinical outcome. Spine J. 2003;3:289–93. doi: 10.1016/s1529-9430(03)00061-5. [DOI] [PubMed] [Google Scholar]
- 28.Eck KR, Bridwell KH, Ungacta FF, et al. Analysis of titanium mesh cages in adults with minimum two-year follow-up. Spine. 2000;25:2407–15. doi: 10.1097/00007632-200009150-00023. [DOI] [PubMed] [Google Scholar]
- 29.Hasegawa K, Abe M, Washio T, et al. An experimental study on the interface strength between titanium mesh cage and vertebra in reference to vertebral bone mineral density. Spine. 2001;26:957–63. doi: 10.1097/00007632-200104150-00022. [DOI] [PubMed] [Google Scholar]
- 30.Oxland TR, Grant JP, Dvorak MF, et al. Effects of endplate removal on the structural properties of the lower lumbar vertebral bodies. Spine. 2003;28:771–7. [PubMed] [Google Scholar]
- 31.Luo J, Skrzypiec DM, Pollintine P, et al. Mechanical efficacy of vertebroplasty: influence of cement type, BMD, fracture severity, and disc degeneration. Bone. 2007;40:1110–9. doi: 10.1016/j.bone.2006.11.021. [DOI] [PubMed] [Google Scholar]
- 32.Berlemann U, Ferguson SJ, Nolte LP, et al. Adjacent vertebral failure after vertebroplasty. A biomechanical investigation. J Bone Joint Surg Br. 2002;84:748–52. doi: 10.1302/0301-620x.84b5.11841. [DOI] [PubMed] [Google Scholar]
- 33.Sun K, Tawackoli W, Fukshanksky M, Rhimes L, Mendel E, Burton A, Liebschner M. Disperse Cement Filling is More Important than Cement Material Properties in Biomechanics of Vertebroplasty. NASS 21st Annual Meeting: The Spine Journal. 2006:147S–8S. [Google Scholar]
- 34.Hollowell JP, Vollmer DG, Wilson CR, et al. Biomechanical analysis of thoracolumbar interbody constructs. How important is the endplate? Spine. 1996;21:1032–6. doi: 10.1097/00007632-199605010-00007. [DOI] [PubMed] [Google Scholar]
- 35.Brantigan JW, Cunningham BW, Warden K, et al. Compression strength of donor bone for posterior lumbar interbody fusion. Spine. 1993;18:1213–21. doi: 10.1097/00007632-199307000-00015. [DOI] [PubMed] [Google Scholar]
- 36.Hoshijima K, Nightingale RW, Yu JR, et al. Strength and stability of posterior lumbar interbody fusion. Comparison of titanium fiber mesh implant and tricortical bone graft. Spine. 1997;22:1181–8. doi: 10.1097/00007632-199706010-00002. [DOI] [PubMed] [Google Scholar]
- 37.Closkey RF, Parsons JR, Lee CK, et al. Mechanics of interbody spinal fusion. Analysis of critical bone graft area. Spine. 1993;18:1011–5. doi: 10.1097/00007632-199306150-00010. [DOI] [PubMed] [Google Scholar]
- 38.Jost B, Cripton PA, Lund T, et al. Compressive strength of interbody cages in the lumbar spine: the effect of cage shape, posterior instrumentation and bone density. Eur Spine J. 1998;7:132–41. doi: 10.1007/s005860050043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Oxland TR, Lund T, Jost B, et al. The relative importance of vertebral bone density and disc degeneration in spinal flexibility and interbody implant performance. An in vitro study. Spine. 1996;21:2558–69. doi: 10.1097/00007632-199611150-00005. [DOI] [PubMed] [Google Scholar]
- 40.Lund T, Oxland TR, Jost B, et al. Interbody cage stabilisation in the lumbar spine: biomechanical evaluation of cage design, posterior instrumentation and bone density. J Bone Joint Surg Br. 1998;80:351–9. doi: 10.1302/0301-620x.80b2.7693. [DOI] [PubMed] [Google Scholar]
- 41.Lindahl O. Mechanical properties of dried defatted spongy bone. Acta Orthop Scand. 1976;47:11–9. doi: 10.3109/17453677608998966. [DOI] [PubMed] [Google Scholar]
- 42.Ebbesen EN, Thomsen JS, Beck-Nielsen H, et al. Lumbar vertebral body compressive strength evaluated by dual-energy X-ray absorptiometry, quantitative computed tomography, and ashing. Bone. 1999;25:713–24. doi: 10.1016/s8756-3282(99)00216-1. [DOI] [PubMed] [Google Scholar]
- 43.Cheng XG, Nicholson PH, Boonen S, et al. Prediction of vertebral strength in vitro by spinal bone densitometry and calcaneal ultrasound. J Bone Miner Res. 1997;12:1721–8. doi: 10.1359/jbmr.1997.12.10.1721. [DOI] [PubMed] [Google Scholar]
- 44.Hansson T, Roos B, Nachemson A. The bone mineral content and ultimate compressive strength of lumbar vertebrae. Spine. 1980;5:46–55. doi: 10.1097/00007632-198001000-00009. [DOI] [PubMed] [Google Scholar]
- 45.McBroom RJ, Hayes WC, Edwards WT, et al. Prediction of vertebral body compressive fracture using quantitative computed tomography. J Bone Joint Surg Am. 1985;67:1206–14. [PubMed] [Google Scholar]
- 46.Svendsen OL, Hassager C, Skodt V, et al. Impact of soft tissue on in vivo accuracy of bone mineral measurements in the spine, hip, and forearm: a human cadaver study. J Bone Miner Res. 1995;10:868–73. doi: 10.1002/jbmr.5650100607. [DOI] [PubMed] [Google Scholar]
- 47.Hangartner TN, Johnston CC. Influence of fat on bone measurements with dual-energy absorptiometry. Bone Miner. 1990;9:71–81. doi: 10.1016/0169-6009(90)90101-k. [DOI] [PubMed] [Google Scholar]
- 48.Cunningham BW, Hu N, Beatson HJ, et al. Revision strategies for single- and two-level total disc arthroplasty procedures: a biomechanical perspective. Spine J. 2009 doi: 10.1016/j.spinee.2009.03.011. [DOI] [PubMed] [Google Scholar]
- 49.Brau SA, Delamarter RB, Kropf MA, et al. Access strategies for revision in anterior lumbar surgery. Spine (Phila Pa 1976) 2008;33:1662–7. doi: 10.1097/BRS.0b013e31817bb970. [DOI] [PubMed] [Google Scholar]