Abstract
Background
Four-dimensional flow cardiovascular magnetic resonance (4D flow CMR) is an emerging non-invasive imaging technology used to visualise and quantify intra-cardiac blood flow. The aim of this systematic review is to assess the literature on the current clinical applications of intra-cardiac 4D flow CMR.
Methods
A systematic review was conducted to evaluate the literature on the intra-cardiac clinical applications of 4D flow CMR. Structured searches were carried out on Medline, EMBASE and the Cochrane Library in October 2016. A modified Critical Skills Appraisal Programme (CASP) tool was used to objectively assess and score the included studies. Studies were categorised as ‘highly clinically applicable’ for scores of 67–100%, ‘potentially clinically applicable’ for 34–66% and ‘less clinically applicable’ for 0–33%.
Results
Of the 1608 articles screened, 44 studies met eligibility for systematic review. The included literature consisted of 22 (50%) mechanistic studies, 18 (40.9%) pilot studies and 4 (9.1%) diagnostic studies. Based on the modified CASP tool, 27 (62%) studies were ‘highly clinically applicable’, 9 (20%) were ‘potentially clinically applicable’ and 8 (18%) were ‘less clinically applicable’.
Conclusions
There are many proposed methods for using 4D flow CMR to quantify intra-cardiac flow. The evidence base is mainly mechanistic, featuring single-centred designs. Larger, multi-centre studies are required to validate the proposed techniques and investigate the clinical advantages that 4D flow CMR offers over standard practices.
PROSPERO = CRD42016051438.
Keywords: Intra-cardiac, Systematic review, Four-dimensional, 4D flow CMR, 4D flow MRI, Cardiovascular magnetic resonance
1. Introduction
Cardiovascular disease is the leading cause of mortality worldwide, with an estimated 17.7 million deaths in 2015 alone [1]. Blood flow is a vital parameter in the assessment of cardiovascular disease and hence, requires precise measurement.
Non-invasive imaging techniques, such as echocardiography and two-dimensional cine phase contrast magnetic resonance imaging (2D PC MRI), are standard components of a thorough cardiovascular investigation. Despite this, current quantification methods cannot fully assess the complex, three-dimensional and multi-directional nature of intra-cardiac blood flow.
Four-dimensional flow cardiovascular magnetic resonance (4D flow CMR) is a technique that enables comprehensive haemodynamic flow assessments. In recent years, 4D flow acquisition and post-processing techniques have greatly advanced, accompanied by the development of several new quantification methods. The latest 4D flow consensus document supplies a detailed overview of current 4D flow uses, both cardiac and non-cardiac [2]. This has been supplemented by several other review articles [3], [4], [5]. However, existing articles tend to encompass a large breadth of applications. The most recent 4D flow reviews addressing cardiac applications specifically, are from 2011, featuring non-systematic methodologies [6], [7]. With the rapid advancement of 4D flow, a comprehensive and archival update is warranted. This is especially relevant for the ageing population, given that adult cardiac disease, such as heart failure and valvular heart disease, presents a major and increasing health burden [8], [9]. Therefore, the aim of this systematic review is to methodically summarise new mechanistic intra-cardiac findings, along with information regarding the potential clinical applications of these 4D flow CMR techniques.
2. Methods
This systematic review was registered in the international database of prospectively registered systematic reviews (PROSPERO, registration number = CRD42016051438). The review protocol can be accessed online via the PROSPERO website [10]. The Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) checklist was adhered to when structuring this article [11].
2.1. Search strategy
A comprehensive search was undertaken in October 2016 on the following electronic databases: Medline, EMBASE and the Cochrane Library. The search was limited to ‘Humans’ and ‘English Language’ only. There was no limitation on time periods. To minimise bias, OpenGrey and Copac databases were also searched to identify any grey literature. In addition, clinicaltrials.gov was searched to find any ongoing studies. The supplemental file provides an in-depth description of the search strategy. Citation tracking and manual reference searching was carried out through the OvidSP databases.
2.2. Article screening
Once duplicates were removed, the titles and abstracts of the search results were assessed using a screening algorithm, shown in Fig. 1, Panel A. This was performed by two independent reviewers (S·C and P.G) and cross-checked against a third independent reviewer (M.E). Cases of disagreement were discussed between the reviewers to reach a suitable conclusion. The studies that adhered to this screening had their full texts evaluated.
Fig. 1.
Overview of study selection process. Panel A = Article screening algorithm. Two reviewers independently screened the titles and abstracts of 936 studies using this system. The number of studies excluded at each stage is shown. Panel B = Flow diagram used for identifying the included studies. The full-texts of the 73 studies that were identified from the screening process were assessed. Of these, 4 (5.5%) were excluded as being irrelevant to the systematic review. The remaining 69 studies underwent citation tracking through the OvidSP databases, as well as manual reference searching. This process identified a further 4 relevant studies for inclusion. Of the final 73 studies included, 44 (60.3%) were full studies, whereas 29 (39.7%) were abstracts only. Flow diagram adapted from Moher et al. [11]. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA Statement. PLOS Medicine 2009. 6 (7):e1000097. The PRISMA Statement is distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium.
2.3. Quality assessment
The quality of the included studies was assessed by P.G using a modified Critical Appraisal Skills Programme (CASP) tool, provided in the supplemental file. Answers of ‘yes’ scored 1 point, whereas answers of ‘no’ or ‘can't tell’ scored 0 points. Total scores were converted to percentages and studies were allocated to one of three categories; ‘highly clinical applicable’ for a score of 67–100%, ‘potentially clinically applicable’ for 34–66% and ‘less clinically applicable’ for 0–33%. Study quality was used as an estimate of clinical applicability for the purposes of this review.
2.4. Quantitative assessment
A meta-analysis was deemed inappropriate for this systematic review as much of the research is exploratory, with considerable heterogeneity in the included studies. As a result, a narrative review is provided.
3. Results
The search yielded a total of 1608 studies. These consisted of 539 (33.5%) and 1019 (63.4%) studies sourced from electronic databases Medline and EMBASE respectively, 18 (1.1%) from the Cochrane Library database, 20 (1.2%) from grey literature searches (Copac and OpenGrey databases) and 12 (0.7%) from clinicaltrials.gov.
Of the 1608 studies identified, 672 (41.8%) were removed as duplicates. After screening the title and abstracts of the remaining studies, a further 863 (53.7% of original 1608) were removed as irrelevant. Assessing the full texts of the remaining 73 studies resulted in 4 further exclusions. Citation tracking resulted in no new relevant studies whereas manual reference searching identified 4 relevant records. From the final 73 studies, 29 (39.7%) were abstracts and are presented in the supplemental file. The remaining 44 (60.3%) records were full-text studies. The PRISMA algorithm for study inclusion is presented in Fig. 1, Panel B.
3.1. Included studies
In total, 44 studies were included in this systematic review. Of these, 22 (50%) were mechanistic studies, 18 (40.9%) were pilot studies and 4 (9.1%) were diagnostic studies. The modified CASP tool found that 27 (62%) studies were ‘highly clinically applicable’, 9 (20%) were ‘potentially clinically applicable’ and 8 (18%) were ‘less clinically applicable’. Graphs of the percentage of clinical applicability against various study design factors are depicted in Fig. 2, A to D. The studies were divided according to intra-cardiac structure, 34 of which were relevant to the left heart, 12 addressing the valves of the heart, and 8 for the right heart. Some studies were relevant to more than one category. All 44 included studies were single-centred. A summary of the included studies' characteristics for the left heart, heart valves and right heart can be found in Table 1. The breakdown of scoring for each study is provided in the supplemental file.
Fig. 2.
Graphical representations of the percentage of clinical applicability against various study design factors, Panel A = against study type, Panel B = against study year, Panel C = against the 4D flow methods used and Panel D = against intra-cardiac structure. KE = kinetic energy, TKE = turbulent kinetic energy, Haem forces = haemodynamic forces, RVT = retrospective valve tracking, JSLD = jet shear layer detection method, VT = volume tracking, RV = regurgitant volume, LA = left atrium, LV = left ventricle, MV = mitral valve, AV = aortic valve, TV = tricuspid valve, PV = pulmonary valve, RA = right atrium, RV = right ventricle.
Table 1.
Summary of included studies.
Author(s), year | Study type | Cohort | Methods | Applicability |
---|---|---|---|---|
Left atrium | ||||
Lee et al., 2016 [12] | Mechanistic | 30 HV, 40 AF | Velocity profile | Highly |
Markl et al., 2016 [13]* | Retrospective mechanistic | 8 HV, 62 AF | Velocity profile | Highly |
Markl et al., 2016 [14] | Mechanistic | 30 HV, 81 AF | Velocity profile and its maps | Highly |
Markl et al., 2016 [15] | Retrospective mechanistic | 15 HV, 60 AF | Velocity profile | Highly |
Steding-Ehrenborg et al., 2015 [16]* | Retrospective mechanistic | 14 HV, 14 athletes | KE | Highly |
Suwa et al., 2015 [17] | Prospective mechanistic | 9 HV, 15 No OHD, 17 OHD | Pathline visualisation of vortex | Highly |
Arvidsson et al., 2013 [18]* | Mechanistic | 15 HV | KE | Highly |
Fluckiger et al., 2013 [19] | Pilot | 19 HV, 10 AF | Flow velocity distribution | Less |
Foll et al., 2013 [20]* | Mechanistic | 24 HV | Vortex and flow visualisation | Potentially |
Dyverfeldt et al., 2011 [21] | Pilot | 2 HV, 5 MR | Particle tracing visualisation and turbulent KE | Potentially |
Fyrenius et al., 2001 [22] | Mechanistic | 11 HV | Particle tracing visualisation | Highly |
Left ventricle | ||||
Eriksson et al., 2016 [23] |
Mechanistic | 10 HV, 10 DCM | Global haemodynamic forces | Potentially |
Suwa et al., 2016 [24] | Mechanistic | 22 pEF, 14 rEF | Pathline visualisation of vortex | Highly |
Svalbring et al., 2016 [25] | Mechanistic | 10 HV, 26 IHD | KE | Potentially |
Van Ooij et al., 2016 [26] | Mechanistic | 10 HV, 35 HCM | Pressure gradients and viscous energy loss | Highly |
Wong et al., 2016 [27] | Prospective mechanistic | 35 HV, 10 rEF | KE | Highly |
Al-Wakeel et al., 2015 [28] | Pilot | 7 HV, 10 MR | KE | Potentially |
Eriksson et al., 2015 [29] | Mechanistic | 12 HV | Pressure gradients | Potentially |
Kanski et al., 2015 [30] | Prospective mechanistic | 12 HV, 29 HF | KE | Highly |
Kanski et al., 2015 [31] | Pilot | 8 HV, 20 HF | Particle tracing visualisation and kinetic energy | Highly |
Steding-Ehrenborg et al., 2015 [16]* | Retrospective mechanistic | 14 HV, 14 athletes | KE | Highly |
Zajac et al., 2014 [32] | Mechanistic | 11 HV, 9 DD | Turbulent KE | Less |
Elbaz et al., 2014 [33] | Pilot | 24 HV | Vortex visualisation | Highly |
Eriksson et al., 2013 [34] | Prospective mechanistic | 10 HV, 10 DCM | Pathline visualisation and KE | Highly |
Foll et al., 2013 [20]* | Mechanistic | 24 HV | Particle tracing and velocity vector visualisation | Potentially |
Carlsson et al., 2012 [35]* | Pilot | 9 HV | KE | Potentially |
Toger et al., 2012 [36] | Mechanistic | 9 HV, 4 DCM | Vortex visualisation | Less |
Brandts et al., 2011 [37] | Prospective diagnostic | 47 HF | Diastolic function using retrospective valve tracking | Highly |
Eriksson et al., 2011 [38] | Mechanistic | 12 HV, 1 DCM | Pathline visualisation and KE | Less |
Kumar et al., 2011 [39] | Prospective pilot | 15 DD | Flow visualisation | Less |
Toger et al., 2011 [40] | Pilot | 8 HV, 1 ALVA | Volume tracking | Less |
Eriksson et al., 2010 [41] | Pilot | 6 HV, 3 DCM | Pathline visualisation | Highly |
Bolger et al., 2007 [42] | Pilot | 17 HV, 1 DCM | Particle tracing and KE | Potentially |
Ebbers et al., 2002 [43] | Pilot | 1 HV | Pressure gradients | Less |
Kim et al., 1995 [44] | Mechanistic | 26 HV | Flow visualisation | Highly |
Mitral valve | ||||
Marsan et al., 2009 [45] | Diagnostic | 64 MR | Regurgitant volume, 4D flow as reference method | Highly |
Roes et al., 2009 [46]* | Pilot | 22 HV, 29 IC | Valve flow quantification using retrospective valve tracking | Highly |
Westenberg et al., 2008 [47]* | Retrospective pilot | 10 HV, 20 MR/TR | Valve flow quantification using retrospective valve tracking | Highly |
Westenberg et al., 2005 [48] | Pilot | 10 HV, 10 MR | Valve flow quantification using retrospective valve tracking | Highly |
Westenberg et al., 2004 [49] | Pilot | 10 HV | Valve flow quantification using retrospective valve tracking | Highly |
Aortic valve | ||||
Chelu et al., 2016 [50] | Prospective diagnostic | 54 AR | Visual grading of AR | Highly |
Garcia et al., 2014 [51] | Retrospective pilot | 10 HV, 40 AS | Jet shear layer detection method | Highly |
Ewe et al., 2013 [52] | Retrospective diagnostic | 32 AR | Regurgitant Volume with 4D flow as reference method | Highly |
Roes et al., 2009 [46]* | Pilot | 22 HV, 29 IC | Valve flow quantification using retrospective valve tracking | Highly |
Tricuspid valve | ||||
Roes et al., 2009 [46]* | Pilot | 22 HV, 29 IC | Valve flow quantification using retrospective valve tracking | Highly |
Westenberg et al., 2008 [47]* | Retrospective pilot | 10 HV, 20 MR/TR | Valve flow quantification using retrospective valve tracking | Highly |
Pulmonary valve | ||||
Roes et al., 2009 [46]* | Pilot | 22 HV, 29 IC | Valve flow quantification using retrospective valve tracking | Highly |
Right atrium | ||||
Callaghan et al., 2016 [53] | Prospective pilot | 12 HV | Particle tracing and kinetic energy | Highly |
Markl et al., 2016 [13]* | Retrospective mechanistic | 8 HV, 62 AF | Velocity profile | Highly |
Steding-Ehrenborg et al., 2015 [16]* | Retrospective mechanistic | 14 HV, 14 athletes | KE | Highly |
Arvidsson et al., 2013 [18]* | Mechanistic | 15 HV | KE | Highly |
Right ventricle | ||||
Han et al., 2015 [54] | Pilot | 9 HV, 10 PAH | KE | Less |
Steding-Ehrenborg et al., 2015 [16]* | Retrospective mechanistic | 14 HV, 14 athletes | KE | Highly |
Carlsson et al., 2012 [35]* | Pilot | 9 HV | KE | Potentially |
Fredriksson et al., 2011 [55] | Mechanistic | 10 HV | Pathline visualisation and KE | Potentially |
Summary of the 44 included studies. Of these, 35 address the left heart, 12 for the heart valves and 8 for the right heart. Clinical applicability has been divided into the following groups based on the score achieved using the modified CASP tool: ‘highly’ for 67–100%, ‘potentially’ for 34–67% and ‘less’ for 0–33%. Studies that are relevant to more than one intra-cardiac structure are denoted by an asterisk (*). HV = healthy volunteers, AF = atrial fibrillation, OHD = organic heart disease, MR = mitral regurgitation, DCM = dilated cardiomyopathy, pEF = preserved ejection fraction, rEF = reduced ejection fraction, IHD = ischaemic heart disease, HCM = hypertrophic cardiomyopathy, HF = heart failure, DD = diastolic dysfunction, ALVA = apical left ventricular aneurysm. IC = ischaemic cardiomyopathy, TR = tricuspid regurgitation, AR = aortic regurgitation, AS = aortic stenosis. PAH = pulmonary arterial hypertension.
4. Discussion
To the authors' knowledge, this is the first systematic review of intra-cardiac 4D flow CMR. This systematic review incorporated 44 (2.7%) full papers and 29 (1.8%) abstracts from the 1608 records identified.
4.1. Left heart
4.1.1. Left atrium
A 2001 study was one of the first to define normal left atrial (LA) flow patterns using particle trace visualisation [22]. These findings were supported by a 2015 study from Suwa et al. [17], who proposed that the presence of vortices in the LA may have a role in minimising blood stasis and thus, preventing thrombus formation. This concept has since been investigated widely by Markl et al. [13], [14], [15]. In a study investigating velocity profiles, patients with atrial fibrillation (AF) had 11–19% higher blood stasis compared to controls [13]. Taken further, velocity maps are reliable tools for detecting both regional and global flow patterns in the LA of AF patients [14]. Additionally, they found stasis was significantly higher (P < 0.001) at the wall of the LA, compared to the LA centre, a finding which complements preliminary work by Fyrenius et al. [22]. Recommendations by a third study [15] as well as Lee et al. [12], suggest the need for longitudinal studies to assess whether intra-atrial flow dynamics, as derived by 4D flow CMR, offer better predictions for thromboembolic events than the current, epidemiologically based, CHA2DS2-VASc score [19].
Dyverfeldt et al. assessed turbulent kinetic energy (TKE), in mitral regurgitation (MR) patients [21]. TKE encompasses the energy eventually lost as heat in turbulent flow. Mean LA TKE correlated with the severity of MR (r2 = 0.983, P < 0.001), however this study was limited to only 5 patients.
In health, younger individuals showed higher LA velocity inside the vortex than older individuals (P = 0.012) [20], suggesting that changes in intra-LA haemodynamics are part of normal ageing physiology. Arvidsson et al. found a weak correlation between early diastolic KE of the LA and left ventricular (LV) mass (r2 = 0.28, P < 0.05) [18]. They proposed that the elastic recoil of the LV causes a rise in LA KE in early diastole and so, diastolic suction is likely to be responsible for LV filling.
4.1.2. Left ventricle
The LV is the most extensively studied structure within the intra-cardiac 4D flow literature. This began with Kim et al. identifying the presence of LV vortices as well as its close relationship with anterior mitral leaflet motion [44]. This validated the speculations produced by in vitro flow visualisation models at the time.
In recent years, various studies have used 4D flow CMR to further understand normal LV blood flow [20], [29], [33], [35], [43]. Foll et al. described the effects of multiple demographic variables on LV haemodynamics [20]. It was found that increasing age has an inverse relationship with the number of both LV vortices [(correlation coefficient (CC) = − 0.51, P = 0.01] and diastolic vortices (CC = − 0.49, P = 0.03). In addition, older subjects had reduced size and velocity within their basal LV vortices, whilst women had smaller basal LV vortices than men. Supplementing this, Wong et al. found older individuals have a lower peak diastolic KE compared with both children (P = 0.0001) and young adults (P = 0.025), suggesting a progressive decline with advancing age [27]. These studies highlight the wide variety of normal LV blood patterns.
Al-Wakeel et al. used KE for the assessment of surgical outcomes in MR patients [28], finding significant decreases in mean KE, systolic and early-diastolic KE peaks after mitral valve (MV) repair (P = 0.01, 0.02 and 0.01 respectively). A larger study found that heart failure patients had lower average normalised systolic KE when compared with healthy volunteers (6.3 ± 2.2 mJ/ml vs 8.0 ± 2.1 mJ/ml, P = 0.025) [30]. Carlsson et al. quantified the KE for the left and right ventricle, noting higher early diastolic KE in the LV (6.0 ± 0.6 mJ vs 3.6 ± 0.4 mJ, P = 0.004) [35]. This indicates that LV filling is more dependent on suction mechanisms. Unlike previous studies, these values weren't normalised for LV volumes.
KE has been proposed as a subclinical marker of LV dysfunction in dilated cardiomyopathy (DCM) patients, following observations that alterations in diastolic haemodynamics can be detected despite clinical compensation [34]. Moreover, ischaemic heart disease patients display reductions in ‘Direct flow’ and KE at end-diastole as LV volumes increase, compared to controls [25]. Changes in ‘these parameters may provide more sensitive clinical classifications of LV dysfunction over standard volumetric assessments.
Many studies have compared normal ventricular function with that of DCM patients [23], [34], [36], [38], [41], [42]. One study described the reduced efficiency of a DCM heart, preserving only 5% of the mitral inflow KE compared with 16% for healthy volunteers [42]. However, these are anecdotal findings, given this study featured only one patient. Eriksson et al. demonstrated that LV haemodynamic filling forces of DCM hearts are non-uniform [23] and the volume of LV inflow that is directly ejected, is decreased [34], [38]. The group also mapped out normal LV pressure differences [29] as well as validating semi-automatic quantification analysis as an accurate and reproducible method of assessing LV inflow and outflow volumes [41].
Colour vector analysis was proposed as a novel method of assessing LV diastolic flow in a study of 15 patients with diastolic dysfunction, by observing the termination of organised high velocity flow at the mid-LV [39]. Vortex size and vortex core locations were analysed between heart failure patients, in both preserved and reduced ejection fraction [24]. This study suggests diastolic vortex formation may be crucial for the LV ejection, as well as filling. Characteristic haemodynamic changes need to be established before this can be introduced clinically.
Volume tracking has been offered as an additional way to quantify blood flow [40]. It was tested against particle tracing, a common 4D flow CMR visualisation method. Volume tracking showed an average of 90.5% agreement with particle tracing in mid-diastole as well as strong inter-observer agreement (κ = 0.91). Despite this, the clinical applicability of this method is uncertain given its potential to accrue velocity errors and the unknown influence of factors such as blood viscosity.
Zajac et al. assessed TKE [32] as a measure of flow inefficiency. They found peak TKE at late diastolic filling was higher in DCM patients compared to healthy volunteers (3.0 ± 1.8 mJ vs 1.5 ± 0.8 mJ, P = 0.02). This supports earlier findings by Bolger et al. demonstrating the reduced efficiency of DCM hearts [42]. There is conflicting evidence regarding vortex characteristics in LV dysfunction [24], [36]. Toger et al. found a smaller proportion of blood volume is incorporated within vortices of these patients compared to healthy volunteers [36]. However, this is inconsistent with findings of increased sphericity, greater transverse lengths and larger area for LV vortices in patients with impaired LV function [24]. These variances could be due to chance given the small samples or the differences in how the patients' clinical features were defined.
In a study investigating flow changes in hypertrophic cardiomyopathy, patients with increased extracellular volume (ECV), from fibrosis, had greater energy losses (P < 0.001) and elevated pressure gradients (P < 0.001) at the LV outflow tract [26].
4.2. Heart valves
4.2.1. Mitral valve
Westenberg et al. validated the use of retrospective valve tracking for valvular flow quantification through comparisons with conventional CMR [47]. Retrospective valve tracking was used to accurately and reliably quantify flow, showing strong correlations with aortic systolic stroke volume reference standard (r = 0.96, P < 0.01 for MV flow). Roes et al. investigated the intra −/inter-observer variability of net flow volumes using retrospective valve tracking [46]. For the MV of healthy volunteers, retrospective valve tracking showed excellent agreement (intraclass correlation coefficient [ICC] = 0.97, P < 0.001 and ICC = 0.91, P < 0.001 for intra-/inter-observer variability respectively). Similar levels of agreement were demonstrated for patients with valvular regurgitation.
4.2.2. Aortic valve
Retrospective valve tracking is also reliable for net flow quantification across the aortic valve (AV), with an ICC of 0.93, P < 0.001 for intra-observer results, and an ICC of 0.98, P < 0.001 for inter-observer findings [46]. In a larger study, 4D flow CMR had a sensitivity and specificity of 100% and 98% respectively for identifying aortic regurgitation (AR) [50]. In a study grading AR severity, 2D echocardiography showed limited correlation with 4D flow (κ = 0.53) [52]. The authors speculate that these differences are the result of the assumptions 2D echocardiography requires for its calculation of AR, such as eccentric jets and noncircular orifices.
The only study evaluating aortic stenosis (AS) proposed using the jet shear layer detection (JSLD) method to calculate the AV's effective orifice area (EOA) [51]. Not only did 4D flow CMR show a strong agreement with current 2D EOA methods (r = 0.91, P < 0.001), but it allows for improvements in identifying the true position of the vena contracta, reducing sources of measurement error.
4.2.3. Tricuspid valve
Two studies have demonstrated the robustness of quantifying tricuspid valve (TV) flow using retrospective valve tracking, against the aortic systolic stroke volume reference standard (r = 0.88, P < 0.01) [47] and with strong intra-/inter-observer agreement (ICC = 0.93, P < 0.001 and ICC = 0.94, P < 0.01 respectively) [46].
4.2.4. Pulmonary valve
One study investigated pulmonary valve (PV) flow patterns using retrospective valve tracking [46]. This technique is highly repeatable for PV flow quantification in those with and without valvular regurgitation (intra-/inter-observer variability, ICC = 0.99 and 0.95, P < 0.001 respectively). The lack of studies investigating this structure could be due to the reduced incidence of PV pathology with respect to other valvular diseases [56].
4.3. Right heart
4.3.1. Right atrium
Arvidsson et al. suggested that the conversion of rotational flow into helical flow in the right atrium (RA) may be a method of conserving atrial KE during right ventricular (RV) filling [18]. This is supplemented by findings that 79% of RA stroke volume is comprised of a single vortex [53], with further evidence of helical flow existing between the RA and the RV [16].
4.3.2. Right ventricle
Several studies have shown RV haemodynamics are different to that of the LV [16], [35], [55]. One study found that RV ‘Direct flow’ was larger and made up a greater proportion of the blood volume at end-diastole, compared to LV ‘Direct flow’ (P < 0.01) [55]. It is thought that these conditions, specific to the RV, allow for maximal systolic ejection efficiency of the ‘Direct flow’ volume. This is consistent with findings of higher systolic RV KE compared to the LV (P = 0.004) [35].
Additional KE research found the percentage of viscous dissipation was significantly larger in pulmonary arterial hypertension (PAH) patients compared to healthy volunteers (21.1 ± 6.4% vs 2.2 ± 3.1%, P = 0.0001, ICC = 0.995) [54]. PAH patients also displayed a greater RV KE work density (94.7 ± 33.7 mJ/ml vs 61.7 ± 14.8 mJ/ml, P = 0.007, ICC = 0.990).
5. Clinical perspective
Current technological advances in acceleration methods have enabled 4D flow to be acquired in < 10 min, with a reasonable spatial and temporal resolution. Given these optimisations, it is feasible to incorporate 4D flow acquisition within current clinical CMR protocols. The results from this systematic review suggest that retrospective valve tracking for valvular flow quantification, velocity profiling/KE for various chambers of the heart and several visualisation techniques have the most clinical applicability.
Firstly, retrospective valve tracking is a highly reproducible and accurate 4D flow method (Fig. 3). Moreover, it now features within commercial software packages, for example CAAS by PIE Medical Imaging, Maastricht, The Netherlands. This facilitates its incorporation into clinical studies and increases its availability for routine use. Retrospective valve tracking circumvents issues seen with 2D imaging, chiefly through-plane motion errors when quantifying mitral inflow velocities. Despite the enhanced reliability of these measures, the added value that this may bring to diastolic assessments is yet to be validated clinically. Velocity measurements derived by 4D flow CMR have shown similar correlation to velocities calculated using transesophageal echocardiography (TOE) [15]. Compared to TOE, 4D flow is non-invasive and so, is a more tolerable investigation. It can also be used in patients with contraindications to TOE and does not carry the inherent risks associated with anaesthesia and intubation. Furthermore, velocity/KE profiles have characterised unique flow differences between health and disease. Given the complex classification and management of many cardiac pathologies, it is likely that these 4D flow methods can provide new parameters in which to risk stratify these patients, prioritising targeted preventative interventions. Finally, visualisation is a key strength of 4D flow. It is a versatile technique, including vector graphs, streamlines and pathline/particle tracing. Visualisation provides detailed flow maps which can aid in making distinct assessments of haemodynamic disturbances, including alterations to vortex architecture, to supplement velocity and KE findings. However, it is critical that the presentation of this visualisation data is standardised.
Fig. 3.
4D flow streamline visualisation and retrospective valve tracking quantification. Panels A and B = Four-dimensional mitral inflow in a patient with mitral regurgitation. Panel A shows the mitral regurgitation (yellow arrow) as well as tricuspid regurgitation (red arrow). Panel B = Mitral valve inflow quantification using retrospective valve-tracking. Panels C and D = Four-dimensional aortic flow in a patient with aortic root dilatation. Panel C shows pathological vortex formation in the ascending aorta (red arrow) as well as aortic regurgitation (yellow arrow). Panel D = Aortic valve flow quantification using retrospective valve-tracking.
6. Limitations
Systematic reviews can only synthesise existing literature, therefore any biases or limitations in the included studies will reduce the reliability of the results included for review. The studies reviewed showed considerable methodological heterogeneity, which may have affected inter-study comparisons. For this reason, meta-analyses were not performed. The aspects of subjectivity within the modified CASP tool were minimised through the use of various prompts for each scoring component. Assessing authors have 3 years of 4D flow experience, which may have influenced the results. Nevertheless, the culmination of over 10 years' experience in cardiovascular medicine, enables a qualified assessment of both study quality and the clinical feasibility of developing methods.
7. Conclusion
Current literature in 4D flow CMR is mainly single-centred and mechanistic. The available data offer novel insights into intra-cardiac flow patterns, their possible clinical relevance and demonstrate that 4D flow CMR is a precise and reliable tool for flow quantification. The developed methods with the most clinical applicability appear to be retrospective valve tracking, velocity profiling/KE and visualisation techniques. Prospective, randomised, multi-centred studies are required to investigate the incremental benefits 4D flow CMR may offer over standard practices.
A list of abbreviations and definitions is provided in the supplemental file.
Competing interests
None.
Funding
British Heart Foundation (FS/10/62/28409)
Acknowledgements
We would like to thank Lizzie Caperon from the Skills@Library team, University of Leeds for providing search strategy advice. Karen Lee, Audrey Zannese and Agi Karasinska assisted with the organisation of this work.
Footnotes
Supplementary data to this article can be found online at http://dx.doi.org/10.1016/j.ijcard.2017.07.023.
Appendix A. Supplementary data
Supplementary file
References
- 1.World Health Organisation Cardiovascular disease. (Fact sheet no. 317) 2017. http://www.who.int/mediacentre/factsheets/fs317/en/ Retrieved from.
- 2.Dyverfeldt P., Bissell M., Barker A.J., Bolger A.F., Carlhall C.J., Ebbers T., Francios C.J., Frydrychowicz A., Geiger J., Giese D., Hope M.D., Kilner P.J., Kozerke S., Myerson S., Neubauer S., Wieben O., Markl M. 4D flow cardiovascular magnetic resonance consensus statement. J. Cardiovasc. Magn. Reson. 2015;17:72. doi: 10.1186/s12968-015-0174-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Bax J.J., Delgado V. Advanced imaging in valvular heart disease. Nat. Rev. Cardiol. 2017;14:209–223. doi: 10.1038/nrcardio.2017.1. [DOI] [PubMed] [Google Scholar]
- 4.Markl M., Frydrychowicz A., Kozerke S., Hope M., Wieben O. 4D flow MRI. J. Magn. Reson. Imaging. 2012;36:1015–1036. doi: 10.1002/jmri.23632. [DOI] [PubMed] [Google Scholar]
- 5.Stankovic Z., Allen B.D., Garcia J., Jarvis K.B., Markl M. 4D flow imaging with MRI. Cardiovasc. Diagn. Ther. 2014;4:173–192. doi: 10.3978/j.issn.2223-3652.2014.01.02. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ebbers T. Flow imaging: cardiac applications of 3D cine phase-contrast MRI. Curr. Cardiovasc. Imaging Rep. 2011;4:127–133. [Google Scholar]
- 7.Markl M., Kilner P.J., Ebbers T. Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J. Cardiovasc. Magn. Reson. 2011;13:7. doi: 10.1186/1532-429X-13-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Coffey S., Cairns B.J., Iung B. The modern epidemiology of heart valve disease. Heart. 2015 doi: 10.1136/heartjnl-2014-307020. [DOI] [PubMed] [Google Scholar]
- 9.Roger V.L. Epidemiology of heart failure. Circ. Res. 2013;113:646–659. doi: 10.1161/CIRCRESAHA.113.300268. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Crandon S., Garg P., Elbaz M.S., Plein S. 2016. Intra-cardiac clinical applications of 4D flow MRI (four-dimensional magnetic resonance imaging) review protocol. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Moher D., Liberati A., Tetzlaff J., Altman D. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6 [PMC free article] [PubMed] [Google Scholar]
- 12.Lee D.C., Markl M., Ng J., Carr M., Benefield B., Carr J.C., Goldberger J.J. Three-dimensional left atrial blood flow characteristics in patients with atrial fibrillation assessed by 4D flow CMR. Eur. Heart J. Cardiovasc. Imaging. 2016;17:1259–1268. doi: 10.1093/ehjci/jev304. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Markl M., Carr M., Ng J., Lee D.C., Jarvis K., Carr J., Goldberger J.J. Assessment of left and right atrial 3D hemodynamics in patients with atrial fibrillation: a 4D flow MRI study. Int. J. Cardiovasc. Imaging. 2016;32:807–815. doi: 10.1007/s10554-015-0830-8. [DOI] [PubMed] [Google Scholar]
- 14.Markl M., Lee D.C., Ng J., Carr M., Carr J., Goldberger J.J. Left atrial 4-dimensional flow magnetic resonance imaging stasis and velocity mapping in patients with atrial fibrillation. Investig. Radiol. 2016;51:147–154. doi: 10.1097/RLI.0000000000000219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Markl M., Lee D.C., Furiasse N., Carr M., Foucar C., Ng J., Carr J., Goldberger J.J. Left atrial and left atrial appendage 4D blood flow dynamics in atrial fibrillation. Circ. Cardiovasc. Imaging. 2016;9(9):e004984. doi: 10.1161/CIRCIMAGING.116.004984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Steding-Ehrenborg K., Arvidsson P.M., Toger J., Rydberg M., Heiberg E., Carlsson M., Arheden H. Determinants of kinetic energy of blood flow in the four-chambered heart in athletes and sedentary controls. Am. J. Physiol. Heart Circ. Physiol. 2016;310:H113–22. doi: 10.1152/ajpheart.00544.2015. [DOI] [PubMed] [Google Scholar]
- 17.Suwa K., Saitoh T., Takehara Y., Sano M., Nobuhara M., Saotome M., Urushida T., Katoh H., Satoh H., Sugiyama M., Wakayama T., Alley M., Sakahara H., Hayashi H. Characteristics of intra-left atrial flow dynamics and factors affecting formation of the vortex flow - analysis with phase-resolved 3-dimensional cine phase contrast magnetic resonance imaging. Circ. J. 2015;79:144–152. doi: 10.1253/circj.CJ-14-0562. [DOI] [PubMed] [Google Scholar]
- 18.Arvidsson P.M., Toger J., Heiberg E., Carlsson M., Arheden H. Quantification of left and right atrial kinetic energy using four-dimensional intracardiac magnetic resonance imaging flow measurements. J. Appl. Physiol. 2013;114:1472–1481. doi: 10.1152/japplphysiol.00932.2012. [DOI] [PubMed] [Google Scholar]
- 19.Fluckiger J.U., Goldberger J.J., Lee D.C., Ng J., Lee R., Goyal A., Markl M. Left atrial flow velocity distribution and flow coherence using four-dimensional FLOW MRI: a pilot study investigating the impact of age and pre- and postintervention atrial fibrillation on atrial hemodynamics. J. Magn. Reson. Imaging. 2013;38:580–587. doi: 10.1002/jmri.23994. [DOI] [PubMed] [Google Scholar]
- 20.Foll D., Taeger S., Bode C., Jung B., Markl M. Age, gender, blood pressure, and ventricular geometry influence normal 3D blood flow characteristics in the left heart. Eur. Heart J. Cardiovasc. Imaging. 2013;14:366–373. doi: 10.1093/ehjci/jes196. [DOI] [PubMed] [Google Scholar]
- 21.Dyverfeldt P., Kvitting J.P., Carlhall C.J., Boano G., Sigfridsson A., Hermansson U., Bolger A.F., Engvall J., Ebbers T. Hemodynamic aspects of mitral regurgitation assessed by generalized phase-contrast MRI. J. Magn. Reson. Imaging. 2011;33:582–588. doi: 10.1002/jmri.22407. [DOI] [PubMed] [Google Scholar]
- 22.Fyrenius A., Wigstrom L., Ebbers T., Karlsson M., Engvall J., Bolger A. Three dimensional flow in the human left atrium. Heart. 2001;86:448–455. doi: 10.1136/heart.86.4.448. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Eriksson J., Bolger A.F., Ebbers T., Carlhall C.J. Assessment of left ventricular hemodynamic forces in healthy subjects and patients with dilated cardiomyopathy using 4D flow MRI. Physiol. Rep. 2016;4(3) doi: 10.14814/phy2.12685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Suwa K., Saitoh T., Takehara Y., Sano M., Saotome M., Urushida T., Katoh H., Satoh H., Sugiyama M., Wakayama T., Alley M., Sakahara H., Hayashi H. Intra-left ventricular flow dynamics in patients with preserved and impaired left ventricular function: analysis with 3D cine phase contrast MRI (4D-flow) J. Magn. Reson. Imaging. 2016;44(6):1493–1503. doi: 10.1002/jmri.25315. [DOI] [PubMed] [Google Scholar]
- 25.Svalbring E., Fredriksson A., Eriksson J., Dyverfeldt P., Ebbers T., Bolger A.F., Engvall J., Carlhäll C.J. Altered diastolic flow patterns and kinetic energy in subtle left ventricular remodeling and dysfunction detected by 4D flow MRI. PLoS One. 2016;11(8):e0161391. doi: 10.1371/journal.pone.0161391. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Van Ooij P., Allen B.D., Contaldi C., Garcia J., Collins J., Carr J., Choudhury L., Bonow R.O., Barker A.J., Markl M. 4D flow MRI and T1-mapping: assessment of altered cardiac hemodynamics and extracellular volume fraction in hypertrophic cardiomyopathy. J. Magn. Reson. Imaging. 2016;43:107–114. doi: 10.1002/jmri.24962. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Wong J., Chabiniok R., deVecchi A., Dedieu N., Sammut E., Schaeffter T., Razavi R. Age-related changes in intraventricular kinetic energy: a physiological or pathological adaptation? Am. J. Physiol. Heart Circ. Physiol. 2016;310:H747–55. doi: 10.1152/ajpheart.00075.2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Al-Wakeel N., Fernandes J.F., Amiri A., Siniawski H., Goubergrits L., Berger F., Kuehne T. Hemodynamic and energetic aspects of the left ventricle in patients with mitral regurgitation before and after mitral valve surgery. J. Magn. Reson. Imaging. 2015;42:1705–1712. doi: 10.1002/jmri.24926. [DOI] [PubMed] [Google Scholar]
- 29.Eriksson J., Bolger A.F., Carlhall C.J., Ebbers T. Spatial heterogeneity of four-dimensional relative pressure fields in the human left ventricle. Magn. Reson. Med. 2015;74:1716–1725. doi: 10.1002/mrm.25539. [DOI] [PubMed] [Google Scholar]
- 30.Kanski M., Arvidsson P.M., Toger J., Borgquist R., Heiberg E., Carlsson M., Arheden H. Left ventricular fluid kinetic energy time curves in heart failure from cardiovascular magnetic resonance 4D flow data. J. Cardiovasc. Magn. Reson. 2015;17:111. doi: 10.1186/s12968-015-0211-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Kanski M., Toger J., Steding-Ehrenborg K., Xanthis C., Bloch K.M., Heiberg E., Carlsson M., Arheden H. Whole-heart four-dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use: a head-to-head comparison. BMC Med. Imaging. 2015;15:20. doi: 10.1186/s12880-015-0061-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Zajac J., Eriksson J., Dyverfeldt P., Bolger A.F., Ebbers T., Carlhall C.J. Turbulent kinetic energy in normal and myopathic left ventricles. J. Magn. Reson. Imaging. 2015;41:1021–1029. doi: 10.1002/jmri.24633. [DOI] [PubMed] [Google Scholar]
- 33.Elbaz M.S., Calkoen E.E., Westenberg J.J., Lelieveldt B.P., Roest A.A., van der Geest R.J. Vortex flow during early and late left ventricular filling in normal subjects: quantitative characterization using retrospectively-gated 4D flow cardiovascular magnetic resonance and three-dimensional vortex core analysis. J. Cardiovasc. Magn. Reson. 2014;16:78. doi: 10.1186/s12968-014-0078-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Eriksson J., Bolger A.F., Ebbers T., Carlhall C.J. Four-dimensional blood flow-specific markers of LV dysfunction in dilated cardiomyopathy. Eur. Heart J. Cardiovasc. Imaging. 2013;14:417–424. doi: 10.1093/ehjci/jes159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Carlsson M., Heiberg E., Toger J., Arheden H. Quantification of left and right ventricular kinetic energy using four-dimensional intracardiac magnetic resonance imaging flow measurements. Am. J. Physiol. Heart Circ. Physiol. 2012;302:H893–900. doi: 10.1152/ajpheart.00942.2011. [DOI] [PubMed] [Google Scholar]
- 36.Toger J., Kanski M., Carlsson M., Kovacs S.J., Soderlind G., Arheden H., Heiberg E. Vortex ring formation in the left ventricle of the heart: analysis by 4D flow MRI and Lagrangian coherent structures. Ann. Biomed. Eng. 2012;40:2652–2662. doi: 10.1007/s10439-012-0615-3. [DOI] [PubMed] [Google Scholar]
- 37.Brandts A., Bertini M., van Dijk E.J., Delgado V., Marsan N.A., van der Geest R.J., Siebelink H.M., de Roos A., Bax J.J., Westenberg J.J. Left ventricular diastolic function assessment from three-dimensional three-directional velocity-encoded MRI with retrospective valve tracking. J. Magn. Reson. Imaging. 2011;33:312–319. doi: 10.1002/jmri.22424. [DOI] [PubMed] [Google Scholar]
- 38.Eriksson J., Dyverfeldt P., Engvall J., Bolger A.F., Ebbers T., Carlhall C.J. Quantification of presystolic blood flow organization and energetics in the human left ventricle. Am. J. Physiol. Heart Circ. Physiol. 2011;300:H2135–H2141. doi: 10.1152/ajpheart.00993.2010. [DOI] [PubMed] [Google Scholar]
- 39.Kumar R., Charonko J., Hundley W.G., Hamilton C.A., Stewart K.C., McNeal G.R., Vlachos P.P., Little W.C. Assessment of left ventricular diastolic function using 4-dimensional phase-contrast cardiac magnetic resonance. J. Comput. Assist. Tomogr. 2011;35:108–112. doi: 10.1097/RCT.0b013e3181ffdbaf. [DOI] [PubMed] [Google Scholar]
- 40.Toger J., Carlsson M., Soderlind G., Arheden H., Heiberg E. Volume tracking: a new method for quantitative assessment and visualization of intracardiac blood flow from three-dimensional, time-resolved, three-component magnetic resonance velocity mapping. BMC Med. Imaging. 2011;11:10. doi: 10.1186/1471-2342-11-10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Eriksson J., Carlhall C.J., Dyverfeldt P., Engvall J., Bolger A.F., Ebbers T. Semi-automatic quantification of 4D left ventricular blood flow. J. Cardiovasc. Magn. Reson. 2010;12:9. doi: 10.1186/1532-429X-12-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Bolger A.F., Heiberg E., Karlsson M., Wigstrom L., Engvall J., Sigfridsson A., Ebbers T., Kvitting J.P., Carlhall C.J., Wranne B. Transit of blood flow through the human left ventricle mapped by cardiovascular magnetic resonance. J. Cardiovasc. Magn. Reson. 2007;9:741–747. doi: 10.1080/10976640701544530. [DOI] [PubMed] [Google Scholar]
- 43.Ebbers T., Wigstrom L., Bolger A.F., Wranne B., Karlsson M. Noninvasive measurement of time-varying three-dimensional relative pressure fields within the human heart. J. Biomech. Eng. 2002;124:288–293. doi: 10.1115/1.1468866. [DOI] [PubMed] [Google Scholar]
- 44.Kim W.Y., Walker P.G., Pedersen E.M., Poulsen J.K., Oyre S., Houlind K., Yoganathan A.P. Left ventricular blood flow patterns in normal subjects: a quantitative analysis by three-dimensional magnetic resonance velocity mapping. J. Am. Coll. Cardiol. 1995;26:224–238. doi: 10.1016/0735-1097(95)00141-l. [DOI] [PubMed] [Google Scholar]
- 45.Marsan N.A., Westenberg J.J., Ypenburg C., Delgado V., van Bommel R.J., Roes S.D., Nucifora G., van der Geest R.J., de Roos A., Reiber J.C., Schalij M.J., Bax J.J. Quantification of functional mitral regurgitation by real-time 3D echocardiography: comparison with 3D velocity-encoded cardiac magnetic resonance. J. Am. Coll. Cardiol. Img. 2009;2:1245–1252. doi: 10.1016/j.jcmg.2009.07.006. [DOI] [PubMed] [Google Scholar]
- 46.Roes S.D., Hammer S., van der Geest R.J., Marsan N.A., Bax J.J., Lamb H.J., Reiber J.H., de Roos A., Westenberg J.J. Flow assessment through four heart valves simultaneously using 3-dimensional 3-directional velocity-encoded magnetic resonance imaging with retrospective valve tracking in healthy volunteers and patients with valvular regurgitation. Investig. Radiol. 2009;44:669–675. doi: 10.1097/RLI.0b013e3181ae99b5. [DOI] [PubMed] [Google Scholar]
- 47.Westenberg J.J., Roes S.D., Ajmone Marsan N., Binnendijk N.M., Doornbos J., Bax J.J., Reiber J.H., de Roos A., van der Geest R.J. Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR imaging with retrospective valve tracking. Radiology. 2008;249:792–800. doi: 10.1148/radiol.2492080146. [DOI] [PubMed] [Google Scholar]
- 48.Westenberg J.J., Doornbos J., Versteegh M.I., Bax J.J., van der Geest R.J., de Roos A., Dion R.A., Reiber J.H. Accurate quantitation of regurgitant volume with MRI in patients selected for mitral valve repair. Eur. J. Cardiothorac. Surg. 2005;27:462–466. doi: 10.1016/j.ejcts.2004.11.015. (discussion 467) [DOI] [PubMed] [Google Scholar]
- 49.Westenberg J.J., Danilouchkine M.G., Doornbos J., Bax J.J., van der Geest R.J., Labadie G., Lamb H.J., Versteegh M.I., de Roos A., Reiber J.H. Accurate and reproducible mitral valvular blood flow measurement with three-directional velocity-encoded magnetic resonance imaging. J. Cardiovasc. Magn. Reson. 2004;6:767–776. doi: 10.1081/jcmr-200036108. [DOI] [PubMed] [Google Scholar]
- 50.Chelu R.G., van den Bosch A.E., van Kranenburg M., Hsiao A., van den Hoven A.T., Ouhlous M., Budde R.P.J., Beniest K.M., Swart L.E., Coenen A., Lubbers M.M., Wielopolski P.A., Vasanawala S.S., Roos-Hesselink J.W., Nieman K. Qualitative grading of aortic regurgitation: a pilot study comparing CMR 4D flow and echocardiography. Int. J. Cardiovasc. Imaging. 2016;32:301–307. doi: 10.1007/s10554-015-0779-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Garcia J., Markl M., Schnell S., Allen B., Entezari P., Mahadevia R., Chris Malaisrie S., Pibarot P., Carr J., Barker A.J. Evaluation of aortic stenosis severity using 4D flow jet shear layer detection for the measurement of valve effective orifice area. Magn. Reson. Imaging. 2014;32:891–898. doi: 10.1016/j.mri.2014.04.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Ewe S.H., Delgado V., van der Geest R., Westenberg J.J., Haeck M.L., Witkowski T.G., Auger D., Marsan N.A., Holman E.R., de Roos A., Schalij M.J., Bax J.J., Sieders A., Siebelink H.M. Accuracy of three-dimensional versus two-dimensional echocardiography for quantification of aortic regurgitation and validation by three-dimensional three-directional velocity-encoded magnetic resonance imaging. Am. J. Cardiol. 2013;112:560–566. doi: 10.1016/j.amjcard.2013.04.025. [DOI] [PubMed] [Google Scholar]
- 53.Callaghan F.M., Arnott C., Figtree G.A., Kutty S., Celermajer D.S., Grieve S.M. Quantifying right atrial filling and emptying: a 4D-flow MRI study. J. Magn. Reson. Imaging. 2016 doi: 10.1002/jmri.25457. [DOI] [PubMed] [Google Scholar]
- 54.Han Q.J., Witschey W.R., Fang-Yen C.M., Arkles J.S., Barker A.J., Forfia P.R., Han Y. Altered right ventricular kinetic energy work density and viscous energy dissipation in patients with pulmonary arterial hypertension: a pilot study using 4D flow MRI. PLoS One. 2015;10 doi: 10.1371/journal.pone.0138365. ([Electronic Resource]) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Fredriksson A.G., Zajac J., Eriksson J., Dyverfeldt P., Bolger A.F., Ebbers T., Carlhall C.J. 4-D blood flow in the human right ventricle. Am. J. Physiol. Heart Circ. Physiol. 2011;301:H2344–50. doi: 10.1152/ajpheart.00622.2011. [DOI] [PubMed] [Google Scholar]
- 56.Iung B., Baron G., Tornos P., Gohlke-Bärwolf C., Butchart E.G., Vahanian A. Valvular heart disease in the community: a European experience. Curr. Probl. Cardiol. 2007;32:609–661. doi: 10.1016/j.cpcardiol.2007.07.002. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary file