Analysis of bond strength between a nylon reinforcement structure and dental resins

Leonardo-Jiro-Nomura Nakano 1, Guilherme-da Rocha-Scalzer Lopes 1,, Aline-Silva Firmino 2, Jefferson-David-Melo de Matos 1, Rubens-Nisie Tango 1, Tarcisio-José-de Arruda Paes-Junior 1
PMCID: PMC8106934  PMID: 33981399

Abstract

Background

Nylon is a polymer that its use to reinforce dental resins has shown positive results such as increased flexural strength. The aim of this study was to evaluate the bond strength between dental resins and a nylon reinforcement.

Material and Methods

Forty cylindrical nylon blocks with 13 x 23 mm with 0.5% by volume of silica and 40 without were made. Half of the samples of each nylon composition were sandblasted with aluminum oxide (50μm) for 3 s (2.8 bar pressure, distance: 20 mm, incidence angle: 90o). On the nylon blocks, cylinders of chemically activated acrylic resin and indirect composite resin were made, with a bonding area of 6,28 mm2. Eight different groups were obtained according to the material used and the surface treatment (n = 10): Acrylic Resin + Nylon; GAS: Acrylic Resin + Nylon with Silica; GAT: Acrylic Resin + Nylon (Al2O3); GAST: Acrylic Resin + Nylon with Silica (Al2O3); GC: Composite Resin + Nylon; GCS: Composite Resin + Nylon with Silica; GCT: Composite Resin + Nylon (Al2O3); GCST: Composite Resin + Nylon with Silica (Al2O3). The shear test was carried out. The Student’s and the Kruskal-Wallis test was adopted.

Results

There was no statistically difference in the bond strength for nylon with silica for the acrylic resin group. For the composite groups, nylon with silica did not present a statistically difference without surface treatment (p = 0.10) and with surface treatment the bond strength decreased (p = 0.000). The GCT showed a higher bond strength (0.89 MPa). The surface treatment improved the bond strength for the both groups.

Conclusions

The presence of silica in the nylon composition did not influence the bond strength between materials evaluated. However, the surface treatment with aluminum oxide proved to be favorable for this adhesion.

Key words:Nylons - Resins, Synthetic - Structures Strengthening - Dental Research.

Introduction

Since the introduction of acrylic resin, or polymethylmethacrylate (PMMA), in dental practice, there is a continuous search to change processing techniques and their composition to improve their physical and mechanical properties. Even so, this material is widely used in dentistry for its qualities such as biocompatibility, easy handling, reliability, stability in the oral environment, color stability, favorable aesthetics and low cost (1,2). However, despite these advantages, this material has some characteristics that can be improved for a better clinical performance, including flexural strength, maximum fracture load resistance and hardness (3-5).

The lowest flexural strength of acrylic resin denture bases for interocclusal records is considered the main factor of clinical failures (6,7). Several methods have been developed to increase the strength of the material such as plasticization, copolymerization with rubber, use of metal reinforcement and the use of fibers (8-10). In the last few years, there has been a considerable increase in the use of fiber-reinforced composites, especially glass, aramid, carbon, polyester, polyethylene and nylon fibers (11-23,24).

The use of the fibers has as main objective to overcome the mechanical limitations of the polymers. The fiber reinforcement behavior depends on the percentage of material added, the length and orientation of the fibers, and adhesion between fibers and resin. The fiber arrangement can be used in different directions, for example, unidirectional or bidirectional, changing the strength of the material (24,25).

Nylon is a thermoplastic polymer of the polyamide class, and it is produced by the condensation reaction between a diamine and a dibasic acid. It’s useful in the dental resins is due to their durability and strength properties (24). It can also be noted that the use of this polymer to reinforce composite resins has shown positive results such as increased flexural strength (26-28).

The nylon surface treatment is a factor that can contribute to the bond between nylon and dental resins, improving its mechanical behavior. In this sense, the presence of silica causes the mesh to chemically bond to polymeric materials (29). However, there is still no study on the contribution of silica to the nylon structure with regard to adhesion. Therefore, the aim of this study was to analyze the bond strength between dental resins and silica-nylon reinforcement.

Material and Methods

Forty cylindrical nylon blocks with 13 x 23 mm (Natmar Moldes e Plásticos Ltda company) were made. In addition, 40 samples were developed with the same material incorporating 0.5% by volume of silica (ICT / UNESP, São José dos Campos, Patent nº: BR1020120281198) (21,30).

Half of the samples of each nylon composition were sandblasted with aluminum oxide (50μm) for 3 s (2.8 bar pressure, distance: 20 mm, incidence angle: 90o).

On the nylon blocks, cylinders of chemically activated acrylic resin (Vipi-Flash, VIPI Produtos Odontológicas) and indirect composite resin NanolabZ (WILCOS do Brasil Ind. E Com. Ltda) were made using a teflon matrix, with a bonding area of 6,28 mm2. The composite resin was deposited in increments and light-cured for 2 min under vacuum in the Visio Beta Vario oven (3M ESPE, Seefeld, Germany), following the manufacturer’s recommendations (Fig. 1).

Figure 1.

Figure 1

Specimen preparation. a) nylon base; b) nylon base positioned for surface treatment; c) teflon matrix; d) finished sample (nylon base + composite or acrylic resin).

Eight different groups were obtained according to the material used and the surface treatment (n = 10) (Table 1): GA: Acrylic Resin + Nylon; GAS: Acrylic Resin + Nylon with Silica; GAT: Acrylic Resin + Nylon (Al2O3); GAST: Acrylic Resin + Nylon with Silica (Al2O3); GC: Composite Resin + Nylon; GCS: Composite Resin + Nylon with Silica; GCT: Composite Resin + Nylon (Al2O3); GCST: Composite Resin + Nylon with Silica (Al2O3).

Table 1. Distribution of groups according to the type of resin, presence of silica in nylon base and surface treatment.

graphic file with name jced-13-e505-t001.jpg

After 24 hours of polymerization, the shear test was carried out in a universal testing machine (EMIC DL 1000, São José dos Pinhais, Brazil). For the test, the samples were fixed to a metallic device, where a knife chisel directed perpendicularly carried out the loading. The movement was automatically stopped when the specimens debonded (Fig. 2).

Figure 2.

Figure 2

Shear bond test. a) sample positioned for the shear bond test; b) sample after performing the test.

-Failure and statistical analysis

The fractured specimens were examined by stereomicroscope (Stereo Discovery V20, Zeiss, Göttingen, Germany), with a 30× magnification and the failures were classified as an adhesive (resin cement totally present in the composite resin), predominantly adhesive (60% or higher amount of cement in the composite resin) or cohesive. Only the predominantly adhesive and adhesive fractures were considered for statistical analysis. The bond strength (MPa) was calculated by dividing the failure load (N) by the adhesive area (mm2).

The present study evaluated two groups for each factor: material (PMMA x composite), silica (added or not) and surface treatment (blasted or not). A normality analysis was applied (assumption of several parametric tests) for the bond strength data with Shapiro-Wilk’s test and for a normal distribution data the Student’s t-test was adopted and for a non-normal distribution data the Kruskal-Wallis test was adopted.

Statistical tests were performed in R-project 3.2.0 statistical software (R Core Team, Vienna, Austria). The level of significance established for the tests was 5%, which established a 95% confidence interval for the presented results, and the power of a statistical test was 80%.

Results

Regarding the failure mode distribution, a predominantly adhesive and adhesive were observed for all specimens, as shown in Figures 3 and 4.

Figure 3.

Figure 3

Stereomicroscopic image of the representative predominantly adhesive failure of the GAT specimen under 7.5x magnification (a), and the area delimited in red under 67x magnification (b); acrylic resin (red arrow).

Figure 4.

Figure 4

Stereomicroscopic image of the representative predominantly adhesive failure of the GCT specimen under 7.5x magnification (a), and the area delimited in red under 67x magnification (b); composite resin (red arrow).

There was no statistically difference in the bond strength data for nylon with silica for the acrylic resin group without (p = 0.14) and with surface treatment (p = 0.83). For the composite groups, nylon with silica did not present a statistically difference without surface treatment (p = 0.10) and with surface treatment the bond strength decreased (p = 0.000).

The composite showed a higher bond strength (0.89 MPa) with a statistically difference of acrylic resin (0.24 MPa) (p = 0.000). The surface treatment improved the bond strength for the both groups. For the acrylic resin group without (p = 0.005) and with silica (p = 0.000); and for the composite group without (p = 0.000) and with silica (p = 0.001). Thus, the statistically tests presented the difference between the groups. In Table 2, it is possible to verify the results by correlating the factors (material, presence of silica and surface treatment) and the bond strength means of homogeneous groups

Table 2. Bond strength means of homogeneous groups.

graphic file with name jced-13-e505-t002.jpg

Discussion

In the present study, the composite resin group that does not have silica and has undergone surface treatment obtained greater bond strength values, and the groups in which they were subjected to surface treatment, regardless of whether they have silica or not, showed the best results. This result corroborates with what was obtained in the acrylic resin groups and reinforces the effectiveness of the surface treatment in relation to the increase of the bond strength between the studied materials.

There are no studies that performed the nylon surface treatment, therefore, the sandblasting methodology was adopted (16). The methodology of this study used a shear test to evaluate the bond strength, already used in several studies (11,14,15). Only samples that showed real adhesive values (31), being adhesive or predominantly adhesive, were included in the statistical computation.

In previous studies, the added of silica in the nylon mesh presented an increase in fracture load, flexural strength (21,23), and bond strength (29). According the adhesion properties, the presence of silica can improve the chemical bonding with the polymeric radical of the resins. Despite this, it is assumed that one of the reasons for this difference in the results obtained was the influence of the shape and architecture of the specimens, in which, depending on the structure used, there may be a difference in the amount of silica in contact with polymers, such as in previous studies where a nylon grid system was used as reinforcement material and obtained positive results with the presence of silica in nylon (21,23).

The surface treatment was evaluated, where aluminum oxide sandblasting was carried out to verify its influence on nylon adhesion with dental polymers. Corroborating the study by other, the surface treatment increases the roughness and the surface area, allowing greater micromechanical adhesion through the imbrication between the polymer and the nylon surface (32).

The results obtained (Table 2) revealed that the groups where the nylon surface treatments were performed achieved the highest bond strength values. It is assumed that the sandblasting may have activated the silane present in the nylon structure, which positively influenced the result, in addition to providing an increase in the surface area, which improve the adhesion between the materials (21,23). Silane is essential for the charge particles of the composite resin to remain adhered to the resinous matrix, and this allows the polymeric matrix to transfer tensions to the charge particles, which are more rigid (1).

In other study (21), the fracture strength of temporary resins was evaluated, using a nylon mesh with and without silica as reinforcement, and the group in which the nylon contained silica showed the higher strength values. Within the limitations of the study, the presence or not of silica did not improve the bond strength, but the indication of maintaining silica to improve this adhesion is not discarded. Other studies evaluating the architecture of the nylon mesh and the silica distribution are necessary to investigate possible influences on bond strength.

In this study, although the main proposal was to verify the influence of silica on the bond strength, it was found that the surface treatment was a major factor than others. Thus, it can be suggested the surface treatment with aluminum oxide in nylon reinforcement systems containing silica, regardless of the dental polymer, to improve the bond strength.

Conclusions

It can be concluded that:

The presence of silica in the nylon composition did not improve the bond strength between the nylon and the evaluated resins. However, the surface treatment with aluminum oxide proved to be favorable for this adhesion.

Acknowledgments

Ackcnowledgements None declared.

${Availability of data and materials} None declared.

Authors contributions None declared.

Ethics None declared.

Conflicts of interest The authors declare that there are no conflicts of interest.

Funding None declared.

References

  • 1.Frazer RQ, Byron RT, Osborne PB, West KP. PMMA: an essential material in medicine and dentistry. J Long Term Eff Med Implants. 2005;15:629–39. doi: 10.1615/jlongtermeffmedimplants.v15.i6.60. [DOI] [PubMed] [Google Scholar]
  • 2.Ladha K, Shah D. An in-vitro evaluation of the flexural strength of heat-polymerized poly (methyl methacrylate) denture resin reinforced with fibers. J Indian Prosthodont Soc. 2011;11:215–20. doi: 10.1007/s13191-011-0086-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Jagger DC, Jagger RG, Allen SM, Harrison A. An investigation into the transverse and impact strength of "high strength" denture base acrylic resins. J Oral Rehabil. 2002;29:263–7. doi: 10.1046/j.1365-2842.2002.00830.x. [DOI] [PubMed] [Google Scholar]
  • 4.Ali IL, Yunus N, Abu-Hassan MI. Hardness, flexural strength, and flexural modulus comparisons of three differently cured denture base systems. J Prosthodont. 2008;17:545–9. doi: 10.1111/j.1532-849X.2008.00357.x. [DOI] [PubMed] [Google Scholar]
  • 5.Yadav NS, Elkawash H. Flexural strength of denture base resin reinforced with aluminum oxide and processed by different processing techniques. Journal of Advanced Oral Research. 2011;2:33–36. [Google Scholar]
  • 6.Johnston EP, Nicholls JI, Smith DE. Flexure fatigue of 10 commonly used denture base resins. J Prosthet Dent. 1981;46:478–83. doi: 10.1016/0022-3913(81)90232-8. [DOI] [PubMed] [Google Scholar]
  • 7.Chitchumnong P, Brooks SC, Stafford GD. Comparison of three- and four-point flexural strength testing of denture-base polymers. Dent Mater. 1989;5:2–5. doi: 10.1016/0109-5641(89)90082-1. [DOI] [PubMed] [Google Scholar]
  • 8.Ruffino AR. Effect of steel strengtheners on fracture resistance of the acrylic resin complete denture base. J Prosthet Dent. 1985;54:75–8. doi: 10.1016/s0022-3913(85)80074-3. [DOI] [PubMed] [Google Scholar]
  • 9.Rodford RA. Further development and evaluation of high impact strength denture base materials. J Dent. 1990;18:151–7. doi: 10.1016/0300-5712(90)90056-k. [DOI] [PubMed] [Google Scholar]
  • 10.Jagger DC, Harrison A, Jandt KD. The reinforcement of dentures. J Oral Rehabil. 1999;26:185–94. doi: 10.1046/j.1365-2842.1999.00375.x. [DOI] [PubMed] [Google Scholar]
  • 11.Kelly E. Fatigue failure in denture base polymers. J Prosthet Dent. 1969;21:257–66. doi: 10.1016/0022-3913(69)90289-3. [DOI] [PubMed] [Google Scholar]
  • 12.DeBoer J, Vermilyea SG, Brady RE. The effect of carbon fiber orientation on the fatigue resistance and bending properties of two denture resins. J Prosthet Dent. 1984;51:119–21. doi: 10.1016/s0022-3913(84)80117-1. [DOI] [PubMed] [Google Scholar]
  • 13.Solnit GS. The effect of methyl methacrylate reinforcement with silane-treated and untreated glass fibers. J Prosthet Dent. 1991;66:310–4. doi: 10.1016/0022-3913(91)90255-u. [DOI] [PubMed] [Google Scholar]
  • 14.Dixon DL, Breeding LC. The transverse strengths of three denture base resins reinforced with polyethylene fibers. J Prosthet Dent. 1992;67:417–9. doi: 10.1016/0022-3913(92)90261-8. [DOI] [PubMed] [Google Scholar]
  • 15.Ramos V Jr, Runyan DA, Christensen LC. The effect of plasma-treated polyethylene fiber on the fracture strength of polymethyl methacrylate. J Prosthet Dent. 1996;76:94–6. doi: 10.1016/s0022-3913(96)90348-0. [DOI] [PubMed] [Google Scholar]
  • 16.Vallittu PK. The effect of glass fiber reinforcement on the fracture resistance of a provisional fixed partial denture. J Prosthet Dent. 1998;79:125–30. doi: 10.1016/s0022-3913(98)70204-5. [DOI] [PubMed] [Google Scholar]
  • 17.Uzun G, Hersek N, Tinçer T. Effect of five woven fiber reinforcements on the impact and transverse strength of a denture base resin. J Prosthet Dent. 1999;81:616–20. doi: 10.1016/s0022-3913(99)70218-0. [DOI] [PubMed] [Google Scholar]
  • 18.John J, Gangadhar SA, Shah I. Flexural strength of heat-polymerized polymethyl methacrylate denture resin reinforced with glass, aramid, or nylon fibers. J Prosthet Dent. 2001;86:424–427. doi: 10.1067/mpr.2001.118564. [DOI] [PubMed] [Google Scholar]
  • 19.Bae JM, Kim KN, Hattori M, Hasegawa K, Yoshinari M, Kawada E. Fatigue strengths of particulate filler composites reinforced with fibers. Dent Mater J. 2004;23:166–74. doi: 10.4012/dmj.23.166. [DOI] [PubMed] [Google Scholar]
  • 20.Segerström S, Ruyter IE. Mechanical and physical properties of carbon-graphite fiber-reinforced polymers intended for implant suprastructures. Dent Mater. 2007;23:1150–6. doi: 10.1016/j.dental.2006.06.050. [DOI] [PubMed] [Google Scholar]
  • 21.Almeida CS, Amaral M, de Cássia Papaiz Gonçalves F, de Arruda Paes-Junior TJ. Effect of an experimental silica-nylon reinforcement on the fracture load and flexural strength of bisacrylic interim partial fixed dental prostheses. J Prosthet Dent. 2016;115:301–5. doi: 10.1016/j.prosdent.2015.08.009. [DOI] [PubMed] [Google Scholar]
  • 22.Nagakura M, Tanimoto Y, Nishiyama N. Effect of fiber content on flexural properties of glass fiber-reinforced polyamide-6 prepared by injection molding. Dent Mater J. 2017;36:415–421. doi: 10.4012/dmj.2016-252. [DOI] [PubMed] [Google Scholar]
  • 23.Gonçalves FCP, Amaral M, Borges ALS, Gonçalves LFM, Paes-Junior TJA. Fracture load of complete-arch implant-supported prostheses reinforced with nylon-silica mesh: An in vitro study. J Prosthet Dent. 2018;119:606–610. doi: 10.1016/j.prosdent.2017.05.018. [DOI] [PubMed] [Google Scholar]
  • 24.Vallittu PK. Comparison of two different silane compounds used for improving adhesion between fibres and acrylic denture base material. J Oral Rehabil. 1993;20:533–9. doi: 10.1111/j.1365-2842.1993.tb01640.x. [DOI] [PubMed] [Google Scholar]
  • 25.Vallittu PK. Glass fiber reinforcement in repaired acrylic resin removable dentures: preliminary results of a clinical study. Quintessence Int. 1997;28:39–44. [PubMed] [Google Scholar]
  • 26.Fong H. Electrospun nylon 6 nanofiber reinforced BIS-GMA/TEGDMA dental restorative composite resins. Polymer. 2004;45:2427–32. [Google Scholar]
  • 27.Tian M, Gao Y, Liu Y, Liao Y, Xu R, Hedin NE. Bis-GMA/TEGDMA Dental Composites Reinforced with Electrospun Nylon 6 Nanocomposite Nanofibers Containing Highly Aligned Fibrillar Silicate Single Crystals. Polymer (Guildf) 2007;48:2720–28. doi: 10.1016/j.polymer.2007.03.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Borges ALS, Münchow EA, de Oliveira Souza AC, Yoshida T, Vallittu PK, Bottino MC. Effect of random/aligned nylon-6/MWCNT fibers on dental resin composite reinforcement. J Mech Behav Biomed Mater. 2015;48:134–144. doi: 10.1016/j.jmbbm.2015.03.019. [DOI] [PubMed] [Google Scholar]
  • 29.Hamanaka I, Shimizu H, Takahashi Y. Bond strength of a chairside autopolymerizing reline resin to injection-molded thermoplastic denture base resins. J Prosthodont Res. 2017;61:67–72. doi: 10.1016/j.jpor.2016.04.006. [DOI] [PubMed] [Google Scholar]
  • 30.Paes-Junior TJDA, De Castro HL, Borges ALS, Della Bona A, Gonçalves FDCP. A novel silica-nylon mesh reinforcement for dental prostheses. Advances in Materials Science and Engineering. 2017;3709171:1–7. [Google Scholar]
  • 31.Diniz V, Monteiro JB, Rodrigues JVM, Prado PHCO, Melo RMD. Impact of Acid Concentration and Firing on the Long-term Bond Strength of a Zirconia-Lithium Silicate Ceramic Following Adhesive Cementation. The journal of adhesive dentistry. 2019;21:355–363. doi: 10.3290/j.jad.a42998. [DOI] [PubMed] [Google Scholar]
  • 32.Şanlı S, Çömlekoğlu MD, Çömlekoğlu E, Sonugelen M, Pamir T, Darvell BW. Influence of surface treatment on the resin-bonding of zirconia. Dent Mater. 2015;31:657–68. doi: 10.1016/j.dental.2015.03.004. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical and Experimental Dentistry are provided here courtesy of Medicina Oral S.L

RESOURCES