Abstract
The application of cross sectional echocardiography and pulsed Doppler ultrasound to the management of symptomatic neonates with suspected congenital heart disease was studied in 96 consecutive cases. The ability of echocardiography to establish a complete and accurate diagnosis and a correct management plan was evaluated. Sequential segmental analysis of data from cardiac catheterisation and necropsy identified 536 cardiovascular anomalies. Of the 536 anomalies 512 were correctly diagnosed by echocardiography (sensitivity 95.5%). Seven false positive echocardiographic diagnoses were made (specificity 98.6%). Nearly all the missing diagnoses and all the false positive diagnoses made by echocardiography were extracardiac vascular anomalies. Normal cardiovascular anatomy was at all times correctly identified by echocardiography. In 12 babies (12.5%) a management plan could not be established because of inconclusive echocardiographic findings. Of the 84 proposed plans based on the echocardiographic findings, eight were found to be inappropriate after catheterisation. Thus, 76 babies (79.2%) could have been correctly managed without cardiac catheterisation. The combination of cross sectional echocardiography and pulsed Doppler ultrasound not only allows diagnosis of congenital cardiac anomalies in most neonates but can facilitate appropriate clinical management. Only a minority of neonates with suspected congenital heart disease require cardiac catheterisation and angiography.
Full text
PDF









Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Gutgesell H. P., Huhta J. C., Latson L. A., Huffines D., McNamara D. G. Accuracy of two-dimensional echocardiography in the diagnosis of congenital heart disease. Am J Cardiol. 1985 Feb 15;55(5):514–518. doi: 10.1016/0002-9149(85)90237-1. [DOI] [PubMed] [Google Scholar]
- Hagler D. J., Tajik A. J., Seward J. B., Mair D. D., Ritter D. G. Real-time wide-angle sector echocardiography: atrioventricular canal defects. Circulation. 1979 Jan;59(1):140–150. doi: 10.1161/01.cir.59.1.140. [DOI] [PubMed] [Google Scholar]
- Hatle L., Angelsen B. A., Tromsdal A. Non-invasive assessment of aortic stenosis by Doppler ultrasound. Br Heart J. 1980 Mar;43(3):284–292. doi: 10.1136/hrt.43.3.284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hatle L., Brubakk A., Tromsdal A., Angelsen B. Noninvasive assessment of pressure drop in mitral stenosis by Doppler ultrasound. Br Heart J. 1978 Feb;40(2):131–140. doi: 10.1136/hrt.40.2.131. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huhta J. C., Smallhorn J. F., Macartney F. J., Anderson R. H., de Leval M. Cross-sectional echocardiographic diagnosis of systemic venous return. Br Heart J. 1982 Oct;48(4):388–403. doi: 10.1136/hrt.48.4.388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huhta J. C., Smallhorn J. F., Macartney F. J. Two dimensional echocardiographic diagnosis of situs. Br Heart J. 1982 Aug;48(2):97–108. doi: 10.1136/hrt.48.2.97. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MACMAHON B., MCKEOWN T., RECORD R. G. The incidence and life expectation of children with congenital heart disease. Br Heart J. 1953 Apr;15(2):121–129. doi: 10.1136/hrt.15.2.121. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Patel A. K., Rowe G. G., Dhanani S. P., Kosolcharoen P., Lyle L. E., Thomsen J. H. Pulsed Doppler echocardiography in diagnosis of pulmonary regurgitation: its value and limitations. Am J Cardiol. 1982 May;49(7):1801–1805. doi: 10.1016/0002-9149(82)90262-4. [DOI] [PubMed] [Google Scholar]
- Quinones M. A., Young J. B., Waggoner A. D., Ostojic M. C., Ribeiro L. G., Miller R. R. Assessment of pulsed Doppler echocardiography in detection and quantification of aortic and mitral regurgitation. Br Heart J. 1980 Dec;44(6):612–620. doi: 10.1136/hrt.44.6.612. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rice M. J., Seward J. B., Hagler D. J., Mair D. D., Feldt R. H., Puga F. J., Danielson G. K., Edwards W. D., Tajik A. J. Impact of 2-Dimensional echocardiography on the management of distressed newborns in whom cardiac disease is suspected. Am J Cardiol. 1983 Jan 15;51(2):288–292. doi: 10.1016/s0002-9149(83)80052-6. [DOI] [PubMed] [Google Scholar]
- Rice M. J., Seward J. B., Hagler D. J., Mair D. D., Tajik A. J. Definitive diagnosis of truncus arteriosus by two-dimensional echocardiography. Mayo Clin Proc. 1982 Aug;57(8):476–481. [PubMed] [Google Scholar]
- Shinebourne E. A., Macartney F. J., Anderson R. H. Sequential chamber localization--logical approach to diagnosis in congenital heart disease. Br Heart J. 1976 Apr;38(4):327–340. doi: 10.1136/hrt.38.4.327. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smallhorn J. F., Huhta J. C., Anderson R. H., Macartney F. J. Suprasternal cross-sectional echocardiography in assessment of patient ducts arteriosus. Br Heart J. 1982 Oct;48(4):321–330. doi: 10.1136/hrt.48.4.321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smallhorn J. F., Sutherland G. R., Tommasini G., Hunter S., Anderson R. H., Macartney F. J. Assessment of total anomalous pulmonary venous connection by two-dimensional echocardiography. Br Heart J. 1981 Dec;46(6):613–623. doi: 10.1136/hrt.46.6.613. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stevenson J. G., Kawabori I., Dooley T., Guntheroth W. G. Diagnosis of ventricular septal defect by pulsed Doppler echocardiography. Sensitivity, specificity and limitations. Circulation. 1978 Aug;58(2):322–326. doi: 10.1161/01.cir.58.2.322. [DOI] [PubMed] [Google Scholar]
- Stevenson J. G., Kawabori I., Guntheroth W. G. Pulsed Doppler echocardiographic diagnosis of patent ductus arteriosus: sensitivity, specificity, limitations, and technical features. Cathet Cardiovasc Diagn. 1980;6(3):255–263. doi: 10.1002/ccd.1810060306. [DOI] [PubMed] [Google Scholar]
- Sutherland G. R., Godman M. J., Smallhorn J. F., Guiterras P., Anderson R. H., Hunter S. Ventricular septal defects. Two dimensional echocardiographic and morphological correlations. Br Heart J. 1982 Apr;47(4):316–328. doi: 10.1136/hrt.47.4.316. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tajik A. J., Seward J. B., Hagler D. J., Mair D. D., Lie J. T. Two-dimensional real-time ultrasonic imaging of the heart and great vessels. Technique, image orientation, structure identification, and validation. Mayo Clin Proc. 1978 May;53(5):271–303. [PubMed] [Google Scholar]
- Varghese P. J., Celermajer J., Izukawa T., Haller J. A., Jr, Rowe R. D. Cardiac catheterization in the newborn: experience with 100 cases. Pediatrics. 1969 Jul;44(1):24–29. [PubMed] [Google Scholar]