Abstract
Purpose.
To demonstrate accurate measurement of corneal elastic moduli in vivo with non-contact and non-invasive optical coherence elastography.
Methods.
Elastic properties (in-plane Young’s modulus E and both in-plane, μ, and out-of-plane, G, shear moduli) of rabbit cornea were quantified in vivo using non-contact dynamic Acoustic micro-Tapping Optical Coherence Elastography (AμT-OCE). The IOP-dependence of measured mechanical properties was explored in extracted whole globes following in vivo measurement. A nearly-incompressible transverse isotropic (NITI) model was used to reconstruct moduli from AμT-OCE data. Independently, cornea elastic moduli were also measured ex vivo with traditional, destructive mechanical tests (tensile extensometry and shear rheometry).
Results.
Our study demonstrates strong anisotropy of corneal elasticity in rabbits. The in-plane Young’s modulus, computed as E = 3μ, was in the range of 20–44 MPa, whereas the out-of-plane shear modulus was in the range of 34–261 kPa. Both pressure-dependent ex vivo OCE and destructive mechanical tests performed on the same samples within an hour of euthanasia strongly support the results of AμT-OCE measurements.
Conclusions.
Non-contact AμT-OCE can non-invasively quantify cornea anisotropic elastic properties in vivo.
Translational Relevance.
As OCT is broadly accepted in Ophthalmology, these results suggest the potential for rapid translation of AμT-OCE into clinical practice. In addition, AμT-OCE can likely improve diagnostic criteria of ectatic corneal diseases, leading to early diagnosis, reduced complications, customized surgical treatment, and personalized biomechanical models of the eye.
Introduction
Early detection and monitoring of corneal diseases require better understanding of biomechanics to predict and prevent future corneal deformation. However, there are no current clinical tools that can accurately quantify corneal stiffness parameters that, along with the intraocular pressure (IOP), control corneal shape and focusing power. Personalized biomechanical models of the cornea can potentially be used to study disease progression and may play an important role in developing patient specific treatment plans. Non-invasive and non-destructive quantitative measurement of corneal mechanical moduli is needed so that personalized biomechanical models can be developed.
The macro-structure of corneal lamellae produce a highly anisotropic biomechanical deformation response depending on the type of applied force. As such, measured corneal stiffness is very different when a shear stress is applied along the lamellar plane1–3 compared to a tensile stress along the same plane4–7. To account for the anisotropic deformation response, a nearly incompressible transverse isotropic (NITI) material model based on collagen fiber macro-structure was developed. It was shown that the cornea must be described by at least two shear moduli (in-plane tensile and out-of-plane shear moduli)8, a departure from commonly used single-modulus models. Additionally, it was shown that elastic waves in dynamic OCE studies of the cornea were accurately described by the NITI model. Non-contact acoustic micro-tapping optical coherence elastography (AμT-OCE) was used to measure anisotropic mechanical properties in ex vivo porcine cornea, where the cornea’s bounded structure forced guided wave propagation and enabled reconstruction of in-plane, μ (where E = 3μ), and out-of-plane, G, shear moduli by fitting experimentally obtained dispersion curves in the wavenumber-frequency domain to the theoretical (NITI) model.8
The NITI model was further explored using both AμT-OCE and traditional destructive mechanical testing.7 Specifically, the Young’s modulus, E, and out-of-plane shear modulus, G, of fresh porcine whole globes inflated to controlled intraocular pressures (IOP) were measured with AμT-OCE before being cut into corneal buttons and tested using a parallel-plate rheometer and then stripped and tested via in-plane extension loading. Results were consistent with an order of magnitude difference between in-plane, E, and out-of-plane, G, elastic moduli for all testing methods. While there were some differences in corneal condition during testing (such as curvature, boundary conditions, loading type, preconditioning, etc.), it was shown that moduli quantified from OCE data analyzed with the NITI model were accurate and provided a non-contact, non-destructive path to measure corneal anisotropic biomechanical properties.
Because the host-imaging method in AμT-OCE (Optical Coherence Tomography, OCT) can accurately (and without contact) image corneal structure and shape in addition to quantifying mechanical properties, AμT-OCE can potentially provide the in vivo measurements required for personalized biomechanical models of the cornea. However, robust and reliable measurement of anisotropic elastic moduli has not yet been demonstrated in living species. Although elastic properties of in vivo rabbit cornea were estimated previously with OCE, the assumed material models resulted in inaccurate values of elastic moduli because they did not account for cornea bounding and/or cornea mechanical anisotropy.9–11 Thus, to date, there are no studies demonstrating non-contact, non-invasive measurements of cornea shear elastic anisotropy in vivo.
The primary goal of this study was to test the AμT-OCE method using an in vivo rabbit model (New Zealand White Rabbits). It was chosen due to the similarities in geometrical and functional properties with human cornea. In the present study, rabbit cornea elastic moduli were measured with AμT-OCE in vivo and the intraocular pressure (IOP) was recorded. Following in vivo measurements, whole globe samples were extracted and inflated to controlled IOP for comparison between in vivo moduli and ex vivo, pressure-dependent values. Finally, samples were sectioned and tested using both a parallel-plate rheometer (for independent determination of out-of-plane shear modulus, G) and tensile extensometer (for independent determination of in-plane Young’s modulus, E). The results demonstrated that AμT-OCE can reliably generate and track elastic waves in rabbit cornea in vivo, enabling accurate reconstruction of corneal elasticity. In vivo results were compared with both controlled ex vivo and destructive mechanical tests that currently serve as the gold standard in biomechanical testing.
Methods & Materials
Sample preparation
In this study, five (5) adult New Zealand White Rabbits (3 Female, 2 Male, mean ~4 kg) were acquired and housed at the University of Washington Vivarium for a minimum of 5 days prior to experiments. All rabbits were treated in accordance with the Association for Research in Vision and Ophthalmology Statement for the Use of Animals in Ophthalmic and Vision Research. All procedures were approved by UW IACUC (PROTO202000139). While in vivo tests with rabbits do not exactly match the blinking, respiration, and saccades seen in humans, this animal study presents a good model for similar types of motion expected in human cornea for clinical scans.
Each rabbit was transported from the UW vivarium to a research facility for imaging with veterinarian assistance. A trained veterinarian staff anesthetized the rabbit using a one-time dose of 2 mg/kg Xylazine and 50 mg/kg Ketamine, followed by a secondary dose of 15 mg/kg ketamine at the mid-point of the experiment. All animals were kept on 1–5% isoflurane until in vivo testing was completed.
For in vivo OCE testing, rabbits were placed on their side on a sterile pad under the imaging arm of the OCE system (Figure 1a). Sterile cotton pads were used to stabilize the rabbit’s head and align the eye for imaging. The eyelid was held open using a pediatric eyelid speculum (Figure 1b) and the intraocular pressure of each eye was tested using a contact tonometer (Tono-Pen XL, Reichert, Inc., Buffalo, NY). Following tonometry, at least 10 non-contact OCE scans were performed on each eye.
Figure 1.
a) Schematic of AμT-OCE system b) in vivo imaging of rabbit cornea c) ex vivo whole globe with needle insertion to control IOP.
Following imaging, rabbits were euthanized using a 1.5 ml Euthasol injection and whole globe corneas were harvested no more than 30 minutes following initial anesthesia. All excess tissue was removed to expose the sclera, and each globe was rinsed with balanced saline solution (BSS).
Whole globes were placed in a mold containing a damp sterile cotton pad to stabilize samples and mimic in vivo boundary conditions (Figure 1c). A 20-gauge needle connected to a bath filled with BSS was inserted through the temporal wall of the sclera to apply a controlled internal hydrostatic pressure (intraocular pressure, IOP). The IOP was controlled by raising and lowering the bath to impart controlled pressures between 1 and 18 mmHg, at increments of 2 mmHg. Each sample was held at the corresponding pressure for 5 minutes before scanning, over which BSS was applied lightly to prevent corneal dehydration. Five (5) repeat OCE scans were acquired at each pressure. Each sample was scanned at room temperature and imaging took no longer than 1 hour per sample.
Following OCE, cornea-scleral rings were extracted for mechanical testing, where each ring was used to test the out-of-plane shear modulus, G, via rotational rheometry.
Following rheometry, each sample was cut into strips for tensile testing of the in-plane Young’s modulus, E, using an extensometer. All tests were performed within 6 hours of euthanasia and all samples were transported using cloths dampened with BSS.
Nearly incompressible transverse isotropic (NITI) model
The macro-structure of corneal lamellae creates a highly anisotropic deformation response to force. The primary lamellae orientation (arranged along in-plane layers12,13) contributes to a stress-strain relationship that can be approximated using four independent elastic constants λ, δ, G, and μ (assuming small-strain deformation).7,14 The elasticity matrix in the cornea (in Voigt notation) can thus be described as:
| (1) |
where σij denotes engineering stress, εij denotes engineering strain, τij denotes shear stress, γij = 2εij denotes shear strain, and the subscripts x, y, and z denote standard Cartesian coordinates. Because the cornea is nearly incompressible, the longitudinal modulus λ does not define deformation, and the corneal strain response to an applied stress can be fully defined by the out-of-plane and in-plane shear modulus (G and μ, respectively), and an additional term, δ.
As shown previously, the in-plane Young’s modulus depends on both μ and δ.7 However, corneal structure constrains the latter to the range −2μ < δ < 0, which restricts the Young’s modulus to the range:
| (2) |
Although δ cannot be determined from guided wave propagation, the limited range for ET suggests that μ and G can provide very close approximations of full cornea deformations assuming small deformation. While the exact relationship between the in-plane Young’s modulus (ET) and δ has not yet been determined, we assume tensile isotropy here so that E = ET ≅ 3μ. The approximation of δ = 0 assumes corneal tensile isotropy and likely leads to a slight overestimation of ET.
Acoustic micro-tapping OCE system
A home built AμT-OCE system launched mechanical waves in the cornea and tracked their propagation in space and time. The system in Figure 1a has been detailed previously.7 Briefly, elastic waves were generated using a cylindrically focused air-coupled piezoelectric transducer (AμT) driven with a 100 μs-long chirped (1 MHz-1.1 MHz) waveform providing a temporally localized and spatially focused acoustic ‘push’. The resulting elastic wave was measured using a stable Michelson-type fiber-optic interferometer where a broadband superluminescent diode (SLD1018P, Thorlabs, NJ) with central wavelength 1310 nm (45 nm full-width-half-maximum bandwidth) was coupled into polarization maintaining fibers and components for depth encoded (1.5 mm effective imaging range) OCT imaging and motion detection.
The generation and tracking of propagating elastic waves were achieved using an acoustic pulse aligned in time with the start of multiple (256 consecutive) OCT A-scans performed at the same spatial location (referred to as an M-scan). Sequential acoustic pulses were then generated at a fixed location while the OCT M-scan was performed at different locations (256 spatial locations) perpendicular to the AμT line source (generating what is referred to as an MB-scan15). Each MB-scan provided a 3D dataset with 1024 depth × 256 lateral × 256 temporal dimensions. The scan rate of the system (determined by the line-scan camera) was 90 kHz, corresponding with a total scan time of around 750 milliseconds. For ex vivo scans at low IOP, 512 temporal scans were taken to allow the elastic wave to propagate across the full 10 mm field of view.
The MB data-set was used to reconstruct elastic wave propagation based on local vertical particle vibration velocity,16 where the measured displacement sensitivity of the system was approximately 1 nm in water. The log-compressed real-part of the OCT signal was used to reconstruct corneal structure, and the surface of the cornea was detected using an automatic segmentation algorithm. The vibration velocity along the surface of the cornea was determined using a weighted-average (one half of a Gaussian window, HWHM = 90 μm, weight decreasing with depth) along the anterior 183 μm of the cornea, providing raw space-time (x-t) maps of the vertical displacement from propagating guided elastic waves detected along the air-tissue boundary.
For in vivo imaging, bulk motion due to pulsatile vibrations, breath, and rabbit movements can make it difficult to accurately reconstruct elastic wave propagation. In this study, broadband elastic waves generated by the acoustic excitation travel with energy concentrated in the multiple-kHz range. Because much of the bulk motion associated with in vivo imaging occurs in the low-frequency range, a temporal bandpass filter was applied to x-t data where vibration frequencies below 50 Hz and above 4 kHz were removed. Additionally, randomly propagating diffuse wavefields17 were limited by a directional filter, where a two-dimensional Fast Fourier Transform (2D FFT) was performed on the x-t plot and a mask applied to the 2nd and 4th quadrant of the k-f space, followed by an inverse 2D FFT.18 Finally, unwanted reflections and forward propagating waveforms from sidelobes in the acoustic excitation, or remaining diffuse propagations, were removed by applying a moving temporal window centered on the peak of the vertical velocity in the x-t plot. The moving temporal window utilized a super-Gaussian (SG) function (Eq. 3) that followed the maximum vibration velocity of the wavefield at each discrete position x:
| (3) |
with σt = 0.5 ms. The resulting surface x-t information was used to reconstruct elastic moduli assuming a NITI material.
Reconstruction of corneal elastic moduli
Since the cornea is a bounded anisotropic material, an appropriate model is required to describe complex wavefields and perform accurate analysis for moduli reconstruction. As shown previously and described above, a bounded nearly incompressible transversally isotropic (NITI) material model can be used for elastic wave propagation in the cornea.
Once surface waveforms were captured (Figure 2a), the space-time (x-t) plots (Figure 2b) were subject to a 2D FFT to display the waveform in wavenumber-frequency (k-f) space. An inversion method based on the solution to guided elastic waves in a NITI material quantified out-of-plane shear modulus, G, as well as in-plane Young’s modulus E (assuming EL=ET).7,8 As the resolution limit (determined in guided materials by the wavelength and corneal thickness19) was approximately 1 mm, the corneal surface area available for imaging (~10 mm) was sufficient for accurate reconstruction of elastic moduli.
Figure 2.
a) Measured corneal surface vibrations in a whole globe rabbit sample inflated to a pressure of 7 mmHg. b) Filtered and windowed x-t plot. c) Best fit of A0 mode plotted on top of k-f plot. d) Optimization function showing best-fit and uncertainty around G and e) μ. Uncertainty intervals (red lines) calculated for the representative example where standard deviation in gNITI was 0.3%. The best-fit solution provided G = 20.5 kPa, with uncertainty of 19.8 kPa – 21.3 kPa, and μ = 4.0 MPa, with uncertainty of 2.2 MPa – 8.7 MPa.
Because only the A0 -mode is expected in the cornea for the frequency ranges produced by AμT, the A0 -mode solution (detailed previously8) was solved in k-f space for a broad range of input shear moduli (G, and μ) with a fixed corneal thickness. The central cornea thickness was measured in each scan using the OCT image assuming a refractive index of 1.3831.
An iterative routine estimated in-plane tensile and out-of-plane shear moduli (μ and G, respectively), where the best-fit theoretical dispersion relation was performed by maximizing the following objective function using simplex optimization (fminsearch, MATLAB, MathWorks, Natick, MA):
| (4) |
The normalized 2D Fourier spectrum in k-f space is , the A0 mode solution for a NITI material is decribed by w(f, k; μ, G), and the nearly incompressible assumption is defined by β (set to 1 based on an L-curve analysis).2 The value of λ was updated at each iteration according to . The corneal density was assumed to be ρ = 1000 kg/m3 and corneal longitudinal wave speed cL = 1540 m/s. The cornea was assumed bounded from below by water with a density of 1000 kg/m3 and longitudinal wave speed of 1540 m/s. To avoid convergence to a local (as opposed to global) maximum in Eq. (4), five independent fits were performed, with quasi-random (within a reasonable range of expected values) initial values of G0 and μ0. The final output of μ and G were set to those corresponding to the highest value in Eq. 4. An example of the resulting best-fit A0 -mode superimposed on top of the k-f spectrum in a rabbit cornea is shown in Figure 2c.
While the optimization function provides the theoretical A0 mode that most closely matches experimental data, it does not provide information on the fit quality. Although the best-fit line generally follows the spectral maxima (Figure 2c), the degree to which the theoretical A0 mode follows the measured spectral maxima is not directly determined by the fitting regime. By normalizing the best-fit line to the unconstrained global maximum of the spectral peaks at each spectral bin in k-f space (Φmax), the quality of the fit can be estimated:
| (5) |
where gNITI describes the fraction of the maximum mode energy captured by the best-fit A0 dispersion curve. A value of gNITI = 1 would indicate that the experimental dispersion curve has the exact match with the NITI model (ΦNITI captures all the measured mode’s maximum energy). Because low values of gNITI correspond with a failure to converge to a solution that captures most of the mode energy, it can be used to exclude unreliable scans (due to misalignment, rabbit motion, model inaccuracies, etc.). The output values of G and μ were considered inaccurate for any fit that corresponded to gNITI below 0.87, and 0.92, respectively. Details on determining exclusion parameters can be found in Supplementary Methods.
Due to the maximization approach for weighted fitting, residual errors are not computed. To estimate uncertainty in the final fit value, G and μ were varied independently around the optimum values of ΦNITI. For each combination of G and μ, Eq. 4 was used to calculate Φ(μ, G), which was then used to determine:
| (6) |
Multiple fits were performed on independent scans of the same sample, and both model and system error were considered to determine uncertainty intervals around the best-fit values. The best-fit estimate (gNITI) was determined independently for at least 5 repeat scans (in both in vivo and ex vivo data sets). The standard deviation of gNITI was then used to determine the cut-off value in ψ for each set of scans.
Consider the representative example shown in Figure 2d, e, where the mean value of gNITI (gNITI = ψmax) was 0.98 with a standard deviation of 0.003 (or 0.3%) across 5 independent scans. The variation in gNITI determined the uncertainty interval using the corresponding G and μ values for ψ at 0.3% below the max (i.e. ψ = .977 for G and μ). This method produced uneven error bars due to the shape of ψ. Note that the absolute value of gNITI provides an estimate of model error. When gNITI is reduced, the shape of ψ widens for both moduli and, therefore, model uncertainty increases.
Because n = 5 repeat scans were taken, five independent measurements produced corresponding values for G and μ, and their respective uncertainty ranges. The mean value of G and μ for all repeat fits was considered for each cornea (at each IOP). Uncertainty ranges were calculated as mean-squared values of lower and upper uncertainty limits in independent scans. The process provided a value for the OCE-measured out-of-plane shear modulus, G, and in-plane Young’s modulus, E (assuming E = 3μ), as well as error bars associated with the uncertainty of reconstruction. Uncertainty intervals for each independent OCE measured value can be found in the Supplementary Material.
Mechanical testing of ex vivo samples
Following OCE testing of corneas, corneo-scleral rings were extracted and used to test the out-of-plane shear modulus, G. A rheometer (Anton Paar MCR 301 Physica) assessed the frequency-dependent shear behavior (storage, G′(ω), and loss, G″(ω)) of corneal buttons over a range of 0.16–16 Hz. A 5 N compressive preload was applied and the peak shear strain was ~.1%. The test was performed twice, and the mean of each run was used for the final value.
Following rheometry, corneal-sclera buttons were cut into strips and pneumatically clamped (2752–005 BioPuls submersible pneumatic grips, 250 N max load). A 50 mN pre-load was applied to each sample and stretched at 2 mm/min up to 10% strain. Two load-unload cycles were performed to precondition the tissue. Three rounds of force-elongation followed by relaxation were performed and converted to stress-strain according to sample geometry. A second order exponential was fit to three-sets of raw data to determine the stress-strain curve. The in-plane Young’s modulus, E, was defined as the tangential slope of the stress-strain curve. Extension testing provided a value for the strain-dependent Young’s modulus, E, up to 10% strain.
Results
In vivo pressure (Tono-Pen)
In vivo intraocular pressure was measured in each eye using a contact tonometer (Tono-Pen XL, Reichert, Depew, NY). Because contact tonometers have been reported to have significant inaccuracies (with errors increasing with greater IOP),20,21 a correction factor was applied to all in-vivo tonometry measurements to facilitate comparison between in vivo and ex vivo measurements. The correction factor in this study was determined using a direct comparison of Tono-Pen measurements made on 5 ex vivo whole globe samples inflated to known intraocular pressures.
As the inflation pressure was incrementally raised from 3 mmHg to 21 mmHg in a subset of ex vivo whole globes, Tono-Pen measurements were recorded at each pressure (Figure 3). The dependence of the Tono-Pen -measured IOP on the inflated pressure was averaged over 5 independent measures and then fit with a linear function:
| (7) |
Figure 3.
Tono-Pen measured IOP value (mmHg) as the internal pressure was raised using a lifted water bath. Colored dots and lines correspond with each sample, black squares and error bars denote mean and standard deviation, respectively, and the black solid line corresponds with the best-fit linear function (Eq. 7). The dotted black line is the one-to-one (slope=1) line for visualization.
As such, in vivo IOP values underwent a correction factor associated with Eq. (7), where IOPControl (corresponding to a coarse estimate of the ‘actual’ IOP) was found for each Tono-Pen determined value (IOPTonopen). The result of this test suggests that a Tono-Pen measured value of 10 mmHg, for example, corresponds to an actual value of 13 mmHg (with uncertainty associated with the standard deviation of the values at each pressure between ~6 mmHg and 21 mmHg). The mean difference between the Tono-Pen and actual values was ~6.5 mmHg across all pressures for the sample size used. As such, error bars corresponding to ± 6.5 mmHg are included in the adjusted in vivo pressure values displayed in Figure 4. All error bars were cut-off at 0 mmHg for display.
Figure 4.
Summary of elasticity measurements performed in vivo and ex vivo. The squares denote in vivo values, and the triangles are the mean of ex vivo measurements. Each black dot is the OCE measured ex vivo modulus for a single sample at the associated pressure (all results shown in Supplementary Materials).
Quantification of elastic moduli
Anisotropic elastic moduli were quantified for each sample in vivo, following extraction (ex vivo), and via destructive mechanical testing for each sample. A summary of the values measured via OCE is shown in Figure 4. In vivo measurements of G (blue squares) and E (red squares) are plotted at the corrected pressure (horizontal error bars are associated with the range of possible ‘actual’ pressure values). Ex vivo moduli (triangles) were measured and displayed at the actual IOP value. The vertical error bars in OCE measured values correspond with the standard deviation of the mean in each cluster, at each IOP. Note that of the 10 samples tested, one had a fit quality below the exclusion criteria and was omitted. Because the goodness of fit generally decreased with increasing IOP for both moduli, there are fewer data-points at pressures greater than 17 mmHg (Supplementary Methods).
A summary of the anisotropic elastic moduli measured in vivo, ex vivo, and with destructive mechanical testing is presented in Figure 5. OCE- measured values represent the mean and standard deviation of the entire group of 9 samples. For in vivo samples, the IOP range (after correction using Fig.3) 11 mmHg on average. The range of in vivo G was 34 kPa- 261 kPa, and for E was 20 MPa- 44 MPa. For both in vivo and ex vivo whole globes measured via OCE, the stiffness generally increased with increasing pressure. In ex vivo whole globes, G increased from 31 kPa (± 15 kPa) in the IOP range from 3 mmHg to 5 mmHg, to 98 kPa (±65 kPa) in the IOP range from 11 mmHg to 13 mmHg, and E increased for the same IOP ranges respectively from 27 MPa (± 9 MPa) to 47 MPa (± 13 MPa). The values are not reported at higher pressures due to decreased sample size with reliable quality of fit. The mean and standard deviation of rheometry measured values of G′(ω) (Figure 5a) at 16 Hz was 75 kPa (± 43 kPa). Note that shear rheometry was performed over a lower range of frequencies than OCE, which presumably should lead to a slightly lower estimate of G, as shown. The tensile modulus, E, increased with strain from 2.8 MPa (± 1.1 MPa) at 1% to 32 MPa (± 20 MPa) at 10%. The mean and standard deviation of the high strain value is displayed in Figure 5c. Corresponding frequency-dependent values of G′(ω) and G″(ω) (measured in all corneal buttons with shear rheometry), as well as the strain-dependent value of the Young’s Modulus, E, along with OCE measured values, can be found in Supplementary Methods for each sample included in the analysis. As can be seen in Figure 5, the range of values measured with all methods (for comparable IOP in OCE) were in close agreement.
Figure 5.
The mean and standard deviation for all corneas measured via OCE in vivo and ex vivo, as well as by destructive mechanical testing. a) mean shear, G, modulus and b) Young’s, E, modulus for: in vivo OCE at the fixed physiological IOP (measured with the tonopen and equal to 11 mmHg on average after the correction presented in Fig.3), ex vivo OCE for three different ranges of IOP indicated in the legend, the mean rheometry value of the storage modulus at 16 Hz, and tensile testing value of the tangential modulus at 10% strain. Standard deviation corresponds to the deviation of measured moduli across the population of 9 cornea samples.
Discussion
The non-contact nature and consistency of AμT-OCE make it a promising method to evaluate corneal elasticity, monitor keratoconus, and potentially serve as a tool to develop personalized biomechanical models that can predict corneal response to ophthalmic interventions. Unlike other imaging methods, the host imaging modality of OCT allows fast, non-contact, potentially high-resolution imaging due to scanning-point focusing of the laser light used to recreate structural images and track propagating elastic waves.
AμT-OCE potentially can change diagnostic criteria of ectatic corneal diseases, leading to early diagnosis, reduced complications, customized surgical treatment, and new opportunities to develop personalized biomechanical models of the eye.
Optimizing the imaging instrument for human use requires live animal testing because there are no equivalent phantoms that mimic the low-frequency motion associated with both cardiac and respiratory cycles. Because the animals used in this study were anesthetized and stable, bulk motion associated with skeletal muscular movement was minimal over the ~3 second scan time. Singh et al.32 have shown (in their case for anesthetized rodents) that the frequency of respiratory motion and the heartbeat do no exceed 5Hz and that the associated RMS motion amplitude of both cornea and retina does not exceed 1.5μm (sub-pixel in our case). A typical time for blood pressure-change-induced axial motion is about 1s. On the other hand, the time between each M-scan in our study is about 3ms (~256/90e3). As such, we can infer that recordings are not highly influenced by axial motion of the eye. In addition, the low frequency filtering induced by windowing the x-t plot removes the residual low frequency component. Finally, because the top-surface displacement is in-fact an averaged and weighted displacement from the first 200μm below the surface of the cornea, small motion effects can be disregarded. Bulk motion in awake animal and human models would likely require faster scan times or surface correction algorithms to account for micrometer and millimeter scale movements.
Although OCT can provide micron-scale spatial resolution, reconstructed quantitative mechanical properties for bounded materials such as the cornea have significantly degraded spatial resolution.19,22 Dense-scanning of propagating elastic waves creates high-resolution images of local group velocity,23 but these images do not map anisotropic elastic moduli at the same resolution. In reconstructing quantitative elastic moduli using guided waves, the resolution is determined by (i) the elastic wavelength of the generated mechanical wave22 and (ii) the corneal thickness.19 For AμT excitation, the theoretical elastic wavelengths generated can yield spatially resolved elastic moduli maps with mm-scale resolution. Demonstration of local anisotropic elastic moduli remains an area of future focus.
An appropriate mechanical model is required to reconstruct elastic moduli from experimental wave propagation data. Mechanical anisotropy has previously been confirmed in both ex vivo porcine and human cornea, where the in-plane tensile (Young’s) modulus, E, is on average multiple orders of magnitude larger than the out-of-plane shear modulus, G. The NITI model used in this work introduced two independent shear moduli, separating in-plane from out-of-plane moduli. It helps to explain the order(s) of magnitude difference in corneal stiffness estimates extracted from shear- and tensile-based mechanical measurements. The results of this study show that AμT- OCE can be used to quantify anisotropic mechanical properties in living corneal tissue. As reported above, anisotropic mechanical moduli in rabbit corneas measured via OCE under both in vivo and ex vivo conditions were in close agreement with the values measured via destructive ex vivo testing.
The propagation speed of a vertically polarized bulk or surface shear wave in the cornea plane of a NITI medium is determined by the modulus, G, only. As such, the shear modulus, μ, cannot be determined in a semi-infinite or bulk material. However, guided elastic wave propagation in the cornea plane also creates a weak dependence on μ through boundary conditions.8 As shown here and in our recent work,7,8a change in μ results in a small change in the low-kHz region of the A0-mode dispersion. Thus, a deviation of the NITI model from the actual cornea may further worsen the goodness of fit and result in a large confidence interval for reconstructed modulus μ. A recent study28, for example, suggests that the wave dispersion can also be affected by corneal prestress induced by IOP and its in-plane boundary conditions, which are not taken into account in our study. . Although the reconstruction of G seems to be quite stable to small inaccuracies in the model used here, this model can be further refined for better quantification of μ. Therefore, the reconstruction of μ from experimental data is less accurate compared to the reconstruction of G. Under these circumstances, it is important to evaluate whether the measured elastic waveform can be appropriately described using the NITI model. To determine whether μ can be accurately determined assuming a NITI model, a standardized method to determine data inclusion and model appropriateness is very important. In this work we developed a method to quantify data quality based on what we refer to as the goodness of fit (between measured and theoretical dispersion curves). Using a statistical analysis of experimental data, we have shown that the reconstruction of G and μ can be considered correct for any fit when gNITI is above 0.87, and 0.92, respectively. Details on the determination of exclusion parameters can be found in Supplementary Methods.
Another important note is that this work assumed tensile isotropy of elastic moduli in the cornea, i.e. ET = EL = 3μ. Most likely, this assumption requires further refinement of the NITI model, particularly in the definition of the proportionality coefficient between ET and μ. Since corneal shear anisotropy is extremely strong, we expect this relationship to be closer to ET = 2μ. Measuring the exact relationship on a large population of cornea samples will be a part of our future studies. Here, however, we prefer to keep the relationship E = ET = EL = 3μ until it is accurately measured.
The cornea is also unique in that the fibers are pre-stressed by the intraocular pressure. Both G and E increased with intraocular pressure for in vivo and ex vivo measurements. Additionally, the stress-strain curves in Supplementary Materials show that tensile Young’s Modulus E is strain dependent. Clearly, the pre-strain condition will play a role in measured corneal stiffness. As an aside, it has also been shown that increased stress plays a role in the propagation behavior of elastic waves, even for linear elastic materials.24,29 Different techniques have shown the possibility of disentangling the effect of IOP induced pre-stress from the change of stiffness for linearly isotropic solids.28–30 Decoupling non-linear mechanical responses from strain-induced changes in wave speed for anisotropic materials remains an area of future interest.
In this work we also assumed that both E and G are real moduli, i.e. the loss modulus related to tissue viscosity is assumed negligibly small. As can be seen in the rheometry plots shown in Supplementary Methods, G has a non-zero loss modulus and the frequency response is consistent with a measurable viscosity. Methods to estimate corneal viscosity (such as by tracking elastic wave energy) are currently under investigation. However, we notice here that cornea viscosity mostly affects the higher frequency range of the dispersion relationship for the A0 mode,25,26 and thus has very little effect on values estimated from the real part of moduli. In other words, the cornea loss modulus is not measured in this study, but introducing complex elastic moduli into the NITI model would produce small changes in the reconstructed elastic (real) part of moduli.
The in vivo measurement of intraocular pressure was performed using a contact Tono-Pen. Consistent with previously reported results,20,21 the mean tonometry-measured IOP generally underestimated the actual IOP at high pressures, with very large individual variability between samples. One limitation in the present study was that only 5 samples were used to create the correction factor. Due to large individual variations, a more accurate correction factor would require a much larger population sample size. Because contact tonometers (such as the Tono-Pen) use a simple model to estimate internal pressure based on displacement of the cornea, the final value it measures is inherently linked to not only IOP, but also to cornea biomechanics. Independent measurements of corneal elastic moduli may help to develop better mechanical models that can potentially lead to more accurate estimates of IOP.
In comparing measured moduli between in vivo and ex vivo studies and between OCE measurement and mechanical tests, we note that sample boundary conditions were quite different for the different methods. In vivo scans were performed with the cornea intact, where the whole globe was held in place via ocular muscles and under normal IOP. Ex vivo scans attempted to match in vivo conditions; however, whole globes were no longer vascularized and ocular muscles no longer attached. Even more significant, rheometry and tensile testing required that the cornea was excised from the sclera and physically altered. Such alterations change the macro-structure and can release tension within lamellae. Additionally, the loading direction was different in ex vivo samples. Thus, identical values of moduli extracted with the different methods used in this study was not expected. Still, the results were in relatively close agreement.
In addition, shear rheometry and OCE methods determine the out-of-plane shear modulus, G, at different frequencies. Whereas rheometry operates in the Hz-range, OCE operates in the multiple-kHz regime. As can be seen in the individual plots shown in Supplementary Methods, the apparent modulus, G, increases with increasing frequency, suggesting that the OCE results reported in Figure 5 would be higher than the expected shear modulus of the cornea under lower frequency shear strain, as well as under typical biological shear strain rates (due to eye rubbing, for example). However, the results obtained with all methods were still in relatively close agreement, which supports the NITI model for reconstruction of cornea moduli from OCE data.
We also note a decrease in fit quality with increasing IOP (Supplementary Methods). Because increasing IOP has been shown to introduce an additional degree of anisotropy,27 rabbit cornea may not be accurately described by the NITI model at high IOP (above 15 mmHg), and an orthotropic model (and, therefore, a larger number of elastic moduli) may be required.
Conclusions
Humans and rabbits share many common genetic features and by examining the physiology, anatomy and mechanical structures of the living rabbit, scientists can gain valuable insights into human function. In this study, the anisotropic shear moduli of rabbit corneas were measured in living animals and validated using multiple ex vivo measurement techniques on the same set of samples. The results of this work suggest that accurate in vivo measurements of human corneal moduli can be made using AμT-OCE, an important step toward supporting clinical adoption.
Supplementary Material
Acknowledgements
The authors wish to thank Nicholas Reyes at UW Animal Facility Operations and Gary Fye at UW Veterinary Services for veterinarian assistance.
Funding Information
This work was supported, in part, by NIH grants R01-EY026532, R01EY028753, R01-EB016034, R01-CA170734, and R01-AR077560, Life Sciences Discovery Fund 3292512, the Coulter Translational Research Partnership Program, an unrestricted grant from the Research to Prevent Blindness, Inc., New York, New York, and the Department of Bioengineering at the University of Washington.
Footnotes
Commercial Relationships Disclosure
Disclosure: M. A. Kirby, None; G. Regnault, None; I. Pelivanov, None; M. O’Donnell, None; R. K. Wang, None; T. T. Shen, None.
Data Availability
The authors declare that all data from this study are available within the Article and its Supplementary Information. Raw data for the individual measurements are available upon reasonable request.
References
- 1.Petsche S. J., Chernyak D., Martiz J., Levenston M. E. & Pinsky P. M. Depth-dependent transverse shear properties of the human corneal stroma. Investig. Ophthalmol. Vis. Sci. 53, 873–880 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Hatami-Marbini H. Viscoelastic shear properties of the corneal stroma. J. Biomech. 47, 723–728 (2014). [DOI] [PubMed] [Google Scholar]
- 3.Sloan S. R., Khalifa Y. M. & Buckley M. R. The location- and depth-dependent mechanical response of the human cornea under shear loading. Investig. Ophthalmol. Vis. Sci. 55, 7919–7924 (2014). [DOI] [PubMed] [Google Scholar]
- 4.Zeng Y., Yang J., Huang K., Lee Z. & Lee X. A comparison of biomechanical properties between human and porcine cornea. J. Biomech. 34, 533–537 (2001). [DOI] [PubMed] [Google Scholar]
- 5.Wollensak G., Spoerl E. & Seiler T. Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking. J. Cataract Refract. Surg. 29, 1780–1785 (2003). [DOI] [PubMed] [Google Scholar]
- 6.Boschetti F., Triacca V., Spinelli L. & Pandolfi A. Mechanical characterization of porcine corneas. J. Biomech. Eng. 134, (2012). [DOI] [PubMed] [Google Scholar]
- 7.Kirby M. A. et al. Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests. Ophthalmol. Sci. 100058 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Pitre J. J. et al. Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE. Sci. Rep. 10, 1–14 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.In Z. I. J. et al. In-vivo 3D corneal elasticity using air-coupled ultrasound optical coherence elastography. 10, 6272–6285 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Zhu J. et al. Simultaneously imaging and quantifying in vivo mechanical properties of crystalline lens and cornea using optical coherence elastography with acoustic radiation force excitation Simultaneously imaging and quantifying in vivo mechanical properties of cryst. 106102, (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Vietcovich F. E. Z. et al. In vivo assessment of corneal biomechanics under a localized cross-linking treatment using confocal air-coupled optical coherence elastography. 13, 2644–2654 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Meek K. M. & Knupp C. Corneal structure and transparency. Prog. Retin. Eye Res. 49, 1–16 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Quantock A. J. et al. From nano to macro: Studying the hierarchical structure of the corneal extracellular matrix. Exp. Eye Res. 133, 81–99 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.O’Donnell M. & Skovoroda A. R. Prospects for Elasticity Reconstruction in the Heart. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 51, 322–328 (2004). [PubMed] [Google Scholar]
- 15.Song S. et al. Shear modulus imaging by direct visualization of propagating shear waves with phase-sensitive optical coherence tomography. J. Biomed. Opt. 18, 121509 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Wang R. K., Kirkpatrick S. & Hinds M. Phase-sensitive optical coherence elastography for mapping tissue microstrains in real time. Appl. Phys. Lett. 90, (2007). [Google Scholar]
- 17.Nguyen T. et al. Diffuse shear wave imaging : toward passive elastography using low-frame rate spectral-domain optical coherence tomography. 21, (2017). [DOI] [PubMed] [Google Scholar]
- 18.Deffieux T., Gennisson J. & Bercoff J. On the effects of reflected waves in transient shear wave elastography. IEEE Transacctons Ultrason. Ferroelectr. Freq. Control 58, 2032–2035 (2011). [DOI] [PubMed] [Google Scholar]
- 19.Regnault G. et al. Spatial resolution in optical coherence elastography of bounded media. Biomed. Opt. Express 13, 4851–4869 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ma D. et al. Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen. Sci. Rep. 6, 1–7 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Lim K. S., Wickremasinghe S. S. & Cordeiro M. F. Accuracy of Intraocular Pressure Measurements in New Zealand White Rabbits. 2419–2423 (2005). [DOI] [PubMed]
- 22.Kirby M. A. et al. Spatial resolution in dynamic optical coherence elastography. J. Biomed. Opt. 24, 1 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Ambroziński Ł. et al. Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity. Sci. Rep. 6, 38967 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Li G. Y., Gower A. L., Destrade M. & Yun S. H. Non-destructive mapping of stress and strain in soft thin films through sound waves. Commun. Phys. 5, 1–7 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Ramier A., Tavakol B. & Yun S.-H. Measuring mechanical wave speed, dispersion, and viscoelastic modulus of the cornea using optical coherence elastography. Opt. Express 27, 16635 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Ramier A. et al. In vivo measurement of shear modulus of the human cornea using optical coherence elastography. Sci. Rep. 10, 1–10 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Li J. et al. Revealing anisotropic properties of cornea at different intraocular pressures using optical coherence elastography. Opt. Elastography Tissue Biomech. III 9710, 97100T (2016). [Google Scholar]
- 28.Sun MG, Son T, Crutison J, Guaiquil V, Lin S, Nammari L, Klatt D, Yao X, Rosenblatt MI, Royston TJ. Optical Coherence Elastography for Assessing the Influence of Intraocular Pressure on Elastic Wave Dispersion in the Cornea. J. Mech. Behav. Biomed. Mater. 128, 105100 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Dore M, Luna A, Royston TJ. Biaxial Tensile Prestress and Waveguide Effects on Estimates of the Complex Shear Modulus Using Optical-Based Dynamic Elastography in Plate-Like Soft Tissue Phantoms. ASME Journal of Engineering and Science in Medical Diagnostics and Therapy. In Press (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Crutison J, Sun MG, Royston TJ. The Combined Importance of Finite Dimensions, Anisotropy, and Pre-stress in Acoustoelastography. J. Acoust. Soc. Amer. 151(4), 2403–13 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Lin R. C., Shure M. A., Rollins A. M., Izatt J. A., Huang D. Group index of the human cornea at 1.3-μm wavelength obtained in vitro by optical coherence domain reflectometry. Optics letters 29(1), 83–85 (2004). [DOI] [PubMed] [Google Scholar]
- 32.Singh K., Dion C., Costantino S., Wajszilber M., Lesk M. R., Ozaki T. Development of a novel instrument to measure the pulsatile movement of ocular tissues. Experimental eye research 91(1), 63–68 (2010). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The authors declare that all data from this study are available within the Article and its Supplementary Information. Raw data for the individual measurements are available upon reasonable request.





