The intra-aortic balloon pump (IABP) has been the most used mechanical circulatory support (MCS) device in the last 40 years. Its utility has been investigated in several clinical settings: cardiogenic shock, myocardial infarction, high-risk percutaneous coronary interventions, left ventricular unloading, and as a bridging strategy to high-risk cardiac surgery. However, no data on the feasibility of temporary MCS with an IABP during transfemoral transcatheter aortic valve implantation (TAVI) are available.
From November 2021 to November 2022, eleven patients with severe aortic stenosis (AS) and a left ventricular ejection fraction (LVEF) <30% (low-flow, low-gradient: n=7) underwent elective transfemoral TAVI with prophylactic transfemoral IABP support. Patients had a median age of 82 years (interquartile range [IQR]: 71.8-85.3), two of the eleven (18%) were female, the median LVEF was 20% (IQR: 18%-25%), and four patients (36%) had known coronary artery disease. Ten balloon-expandable valves (SAPIEN 3 Ultra; Edwards Lifesciences) and one self-expanding valve (Evolut PRO+; Medtronic) were implanted. No patients underwent balloon predilation. Procedural success with no haemodynamic instability or perioperative complications (i.e., cerebrovascular or access-site related complications) was achieved in all patients; the IABP was removed on the same day (n=7) or on day 1 post-procedure (n=4).
Prophylactic IABP counterpulsation was initiated during the index procedure preceding conventional TAVI. The IABP was inserted through the contralateral femoral artery access to the artery reserved for subsequent transcatheter heart valve (THV) delivery. Radial artery access was also obtained for pigtail catheter placement. After confirmation of accurate IABP positioning in the descending aorta under fluoroscopic imaging, standard TAVI protocol was followed. Because of the highly compliant properties of the IABP, neither the THV tracking in the descending aorta nor the proper functioning (i.e., balloon inflation and haemodynamic support) of the IABP were compromised (Figure 1, Moving image 1). Immediate IABP use was deferred in patients with greater than mild aortic insufficiency because of the risk of worsening valve insufficiency and acute left ventricular volume overload. In this subset of patients, IABP MCS was initiated after THV deployment. Otherwise, the IABP was kept functioning throughout the procedure, with the sole exception being during the balloon-expandable valve deployment under rapid pacing (Central illustration, Moving image 1).
Figure 1. The intra-aortic balloon pump can be kept functioning during the following procedural steps: loading of transcatheter heart valve (THV) in descending aorta (balloon-expandable THV), tracking of THV in descending aorta (balloon-expandable and self-expanding THV), and THV delivery (self-expanding THV).
Central Illustration. Prophylactic use of an intra-aortic balloon pump in transfemoral transcatheter aortic valve implantation with balloon-expandable (orange) and self-expanding (purple) valves.
IABP-assisted arterial waveforms before and after TAVI show an increase in aortic pressure. Valve deployment was performed with and without concurrent IABP counterpulsation according to pacing frequency in self-expanding (120 bpm) and balloon-expandable (180 bpm) valves, respectively. IABP use is deferred if more than mild AR is present or balloon predilation is planned because of the risk of AR worsening and acute left ventricular volume overload. *If post-dilatation is needed, IABP counterpulsation should be paused during rapid pacing (180 bpm). AR: aortic regurgitation; BEV: balloon-expandable valve; IABP: intra-aortic balloon pump; SBP: systolic blood pressure; SEV: self-expanding valve; TAVI: transcatheter aortic valve implantation; THV: transcatheter heart valve
TAVI is a consolidated treatment providing survival benefit to patients with severe AS. Yet, those with a severely reduced LVEF, especially in the case of low-flow, low-gradient AS, represent a group with increased mortality1. IABP-induced diastolic pressure augmentation might be particularly beneficial in patients with left ventricular dysfunction, low-flow state, and coronary artery disease. This pilot experience demonstrates the feasibility of temporary MCS with an IABP during transfemoral TAVI, with no impact on valve delivery through the descending aorta2. Although these findings are promising, prospective studies are warranted to validate the safety and efficacy of TAVI with prophylactic IABP support.
Supplementary data
Transcatheter aortic valve implantation with a prophylactic intra-aortic balloon pump.
Acknowledgments
Conflict of interest statement
A. Latib has served on advisory boards or as a consultant for Medtronic, Boston Scientific, Philips, Edwards Lifesciences, and Abbott. S. Ludwig has received travel compensation from Edwards Lifesciences and consulting fees from Bayer. The other authors have no conflicts of interest to declare related to the content of this paper.
Contributor Information
Andrea Scotti, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA; Cardiovascular Research Foundation, New York, NY, USA.
Pier Pasquale Leone, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Matteo Sturla, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Jonathan Curio, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Alexander M. Spring, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Gaia Ressa, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Sebastian Ludwig, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA; Cardiovascular Research Foundation, New York, NY, USA.
Tadahisa Sugiura, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Manaf Assafin, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Juan F. Granada, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA; Cardiovascular Research Foundation, New York, NY, USA.
Mei Chau, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
James Doolittle, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
Azeem Latib, Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center/Albert Einstein College of Medicine, New York, NY, USA.
References
- Ribeiro HB, Lerakis S, Gilard M, Cavalcante JL, Makkar R, Herrmann HC, Windecker S, Enriquez-Sarano M, Cheema AN, Nombela-Franco L, Amat-Santos I, Muñoz-García AJ, Garcia Del, Zajarias A, Lisko JC, Hayek S, Babaliaros V, Le Ven, Gleason TG, Chakravarty T, Szeto WY, Clavel MA, de Agustin, Serra V, Schindler JT, Dahou A, Puri R, Pelletier-Beaumont E, Côté M, Pibarot P, Rodés-Cabau J. Transcatheter Aortic Valve Replacement in Patients With Low-Flow, Low-Gradient Aortic Stenosis: The TOPAS-TAVI Registry. J Am Coll Cardiol. 2018;71:1297–308. doi: 10.1016/j.jacc.2018.01.054. [DOI] [PubMed] [Google Scholar]
- Shreenivas SS, Lilly SM, Szeto WY, Desai N, Anwaruddin S, Bavaria JE, Hudock KM, Thourani VH, Makkar R, Pichard A, Webb J, Dewey T, Kapadia S, Suri RM, Xu K, Leon MB, Herrmann HC. Cardiopulmonary bypass and intra-aortic balloon pump use is associated with higher short- and long-term mortality after transcatheter aortic valve replacement: a PARTNER trial substudy. Catheter Cardiovasc Interv. 2015;86:316–22. doi: 10.1002/ccd.25776. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Transcatheter aortic valve implantation with a prophylactic intra-aortic balloon pump.