Monoaxial Mechanical Tests on Porcino Knee Ligaments *

Rodrigo Ribeiro Pinho Rodarte 1,2,, João Antônio Matheus Guimarães 1, Brenno Tavares Duarte 2, Paulo Pedro Kenedi 2, William Ribeiro Pinho 3
PMCID: PMC10038718  PMID: 36969770

Abstract

The failure of ligament reconstruction has different risk factors, among which we can highlight the period before its incorporation, which is a mechanically vulnerable period. Loss of resistance over time is a characteristic of living tissues. Dissection with bone insertions of the cruciate ligaments of animal models is not described; however, it is essential for monoaxial assays to extract information from tests such as relaxation. The present work describes the dissection used for the generation of a test body for the performance of nondestructive tests to evaluate the mechanical behavior. We performed dissection of four porcino knee ligaments, proposing a dissection technique for the cruciate ligaments with bone inserts for comparison with collateral ligaments. The ligaments were submitted to relaxation tests and had strain gauges placed during the tests. The results showed viscoelastic behavior, validated by strain gauges and with a loss over time; with some ligaments presenting with losses of up to 20%, a factor to be considered in future studies. The present work dissected the four main ligaments of the knee demonstrating the posterior approach that allows maintaining their bone insertions and described the fixation for the monotonic uniaxial trials, besides being able to extract the viscoelastic behavior of the four ligaments of the knee, within the physiological limits of the knee.

Keywords: dissection; biomechanical phenomena; knee joint; tensile strength; models, animal

Introduction

The failure of ligament reconstruction has different risk factors, among which we can highlight the period before its incorporation, which is a mechanically vulnerable period. 1

Knee ligament injuries are high-frequency lesion due to sports practice, among which we highlight the injuries of the four main ligaments: the medial collateral ligament, (MCL), the lateral collateral ligament (LCL), the anterior cruciate ligament (ACL), and the posterior cruciate ligament (PCL).

Loss of resistance over time is a characteristic of living tissues; for example, what happens with the viscoelastic behavior of each structure. If there is no ligament reconstruction in this behavior, this may be the cause of failure, even with normal resonances, as described by Ekdahl et al., 2 resulting in an inadequate redistribution of efforts in knee structures.

Several studies performed limit tests of ligament resistance, 3 4 5 mimicking ligament or graft ruptures. However, there is no work in the literature that studied ligaments in isolation within the physiological range of deformation and extracted the information of this behavior specially in anterior and posterior cruciate ligaments.

Studies can be destructive or nondestructive, although limit tests of the destructive type are the most found in the literature. However, for the extraction of ligament behaviors and to simulate behavioral situations that occur within the normal physiology, some authors suggest performing the tests within the physiological limits of deformations (from 3 to 6%).

Dissection with the bone insertions of the cruciate ligaments of animal models is not described; however, it is essential for monoaxial essays to extract information from tests such as relaxation, as proposed by Duenwald et al. 6 The present work describes the dissection of the four ligaments used for the generation of a test body and the performance of nondestructive tests (relaxation) in monoaxial traction tests, with emphasis on the ACL of the porcine knee.

Material and Method

The study protocol was approved by the Ethics Committee on the Use of Animals as prescribed by the Regulation of Sanitary and Industrial Inspection of Animal Products (MAPA, in the Portuguese acronym).

The four pig knee ligaments (MCL, LCL, ACL, and PCL) were dissected with their bone insertions as described below.

Dissection was performed starting with the collateral ligaments, the first being the medial collateral due to its ease of approach, because it is very superficial. The bone inserts were maintained with a 30-mm length at each end to serve as fixation points. In each bone fragment (proximal and distal), a 6-mm hole was made to install the connecting pins with the INSTRON material testing machine (Instron 5966, Norwood, Massachusetts, USA). After removal of the collaterals, the cruciate ligaments are removed in sequence. Dissection should be performed through the posterior view ( Fig. 1 ).

Fig. 1.

Fig. 1

Posterior vision of the knee with osteotomy of the tibial insertion.

The cut is performed between the posterior and anterior cruciates at the femoral level, through a posterior approach ( Fig. 2 ).

Fig.

Fig.

2 Visualization of the osteotomy site after medialization of the bone plug of the posterior cruciate ligament.

After separation of the proximal extremity, the PCL is removed with the bone fragment in the proximal tibia. This removal allows the visualization of the tibial insertion of the ACL, allowing the removal of the tibia with the bone fragment ( Fig. 3 ).

Fig.

Fig.

3 Osteotomy of the femur between the cruciate ligaments.

We made a modification in the technique of removal of the cruciate ligaments proposed in the article by Skelley et al., 7 who removed the ligaments without the bone fragments.

To date, there is no description of removals of the cruciate ligaments with their bone insertions. The removal with the bone fragments of the ACL inserts prevents the slipping of the ligament specimens in the claws, providing an adequate reading of the viscoelastic behavior of the ligament within physiological limits.

On the day of dissection, a model was thawed at air-conditioned room temperature (22°C) for 5 hours. Then, it was dissected isolating the four main ligaments of the knee (ACL, PCL, MCL, and LCL), maintaining their bone insertions with 30 mm of bone fragment. A 6-mm hole was made, cross-sectional to each bone insert, for pin coupling for correct positioning in the material testing machine. The pig ligament specimens were then placed in plastics and soaked in gauze with 0.9% saline and kept until their testing, which was carried out on the same day.

The models were then placed in the INSTRON material testing machine of the LADES laboratory of CEFET/RJ with a load cell of 1 kN. At an initial moment, the ligament is subjected to an axial load of 10 N during a period of 5 minutes to reach its original length ( Fig. 4 ), as described by Troyer et al. 8

Fig.

Fig.

4 Ligament coupled with strain gauge on the Instron.

Regarding the relaxation tests, the three-step protocol proposed by Duenwald et al. was used 6 with maximum physiological deformations of 6% and with small 3% deformations being usually found, as proposed by the work of Gardiner et al. 9

The proposed protocol, known as Hill-Valley-Hill, served to simulate behavior in an in vivo situation. It is characterized by performing deformations within the expected physiological limit for adequate joint mobility, ranging between 3% (lower limit) and 6% (upper limit) over a period of 100 seconds and done in three stages: one performed as a relaxation test (6%), followed by recovery (3%) and by another relaxation test (6%). 6 Then, data from the behavioral results over time were extracted, with the objective of characterizing the viscoelastic behavior of the ligaments.

Results

The results of the tests performed on the four ligaments are shown in the force versus time graphs, in which force signals were obtained from the INSTRON load cell ( Fig. 5 ).

Fig.

Fig.

5 ( A ) Graph of the force reading x medial collateral ligament time; ( B ) Graph of the strength reading x time of the lateral collateral ligament; ( C ) Graph of the force reading x time of the posterior cruciate ligament; ( D ) Force reading graph x anterior cruciate ligament time.

When analyzing the behavior of the MCL, there was a loss of < 10% of its load in the relaxation test in the 1 st cycle, attenuating in the 2 nd cycle. The LCL behavior also decreased < 10%; however, the 2 nd cycle still showed a loss of load, but of lower decrease, stabilizing with 100 seconds.

The loss by the viscous component of the PCL was evident and the peak in deformation at 6% was of 160 N and showed a fall in the 2 nd cycle and it was of a lower value than that of the 1 st cycle, demonstrating its viscoelastic behavior.

The loss by the viscous component of the anterior cruciate ligament was evident and the 120 N peak in deformation at 6% showed a drop in the 2 nd cycle, being less distinct than that of the 1 st cycle, which reached values < 100 N, with an absolute difference of 20 N (which was greater than that of the 2 nd cycle).

The results showed viscoelastic behavior with a loss over time, which is a factor to be considered in future studies.

Final Comments

The present technical note dissected the four main ligaments of the knee demonstrating the step by posterior vision, extracting with the ligaments and described the fixation for the monotonic uniaxial essays and was able to extract the behavior of the viscoelastic pattern of the four knee ligaments, within the physiological limits of the knee, through adequate specimens obtained with their bone insertions. All ligaments tested showed loss of load over time. These findings point to the importance of new experimental and ligament trials and tendon options for reconstruction, including serving as a reference for graft choices with ligament-like behaviors and simulation models. 10

Funding Statement

Suporte Financeiro O presente estudo não recebeu nenhum suporte financeiro de fontes públicas, comerciais ou sem fins lucrativos

Financial Support The present study did not receive any financial support from public, commercial, or not-for-profit sources.

Conflito de Interesses Os autores declaram não haver conflito de interesses.

*

Trabalho desenvolvido no Centro Federal de Educação Tecnológica Celso Suckow da Fonseca (CEFET/RJ), Rio de Janeiro e Instituto Nacional de Traumatologia e Ortopedia (INTO) , Rio de Janeiro, RJ, Brasil.

*

Work developed in the Centro Federal de Educação Tecnológica Celso Suckow da Fonseca (CEFET/RJ), Rio de Janeiro e Instituto Nacional de Traumatologia e Ortopedia (INTO), Rio de Janeiro, RJ, Brazil.

Referências

  • 1.Ménétrey J, Duthon V B, Laumonier T, Fritschy D. “Biological failure” of the anterior cruciate ligament graft. Knee Surg Sports Traumatol Arthrosc. 2008;16(03):224–231. doi: 10.1007/s00167-007-0474-x. [DOI] [PubMed] [Google Scholar]
  • 2.Ekdahl M, Wang J H, Ronga M, Fu F H. Graft healing in anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc. 2008;16(10):935–947. doi: 10.1007/s00167-008-0584-0. [DOI] [PubMed] [Google Scholar]
  • 3.Górios C, Hernandez A J, Amatuzzi M M et al. Estudo da rigidez do ligamento cruzado anterior do joelho e dos enxertos para sua reconstrução com o ligamento patelar e com os tendões dos músculos semitendíneo e grácil. Acta Ortop Bras. 2001;9(02):26–40. [Google Scholar]
  • 4.Monaco E, Labianca L, Speranza A et al. Biomechanical evaluation of different anterior cruciate ligament fixation techniques for hamstring graft. J Orthop Sci. 2010;15(01):125–131. doi: 10.1007/s00776-009-1417-9. [DOI] [PubMed] [Google Scholar]
  • 5.Adam F, Pape D, Schiel K, Steimer O, Kohn D, Rupp S. Biomechanical properties of patellar and hamstring graft tibial fixation techniques in anterior cruciate ligament reconstruction: experimental study with roentgen stereometric analysis. Am J Sports Med. 2004;32(01):71–78. doi: 10.1177/0095399703258608. [DOI] [PubMed] [Google Scholar]
  • 6.Duenwald S E, Vanderby R, Jr, Lakes R S. Stress relaxation and recovery in tendon and ligament: experiment and modeling. Biorheology. 2010;47(01):1–14. doi: 10.3233/BIR-2010-0559. [DOI] [PubMed] [Google Scholar]
  • 7.Skelley N W, Castile R M, Cannon P C, Weber C I, Brophy R H, Lake S P. Regional Variation in the Mechanical and Microstructural Properties of the Human Anterior Cruciate Ligament. Am J Sports Med. 2016;44(11):2892–2899. doi: 10.1177/0363546516654480. [DOI] [PubMed] [Google Scholar]
  • 8.Troyer K L, Shetye S S, Puttlitz C M. Experimental characterization and finite element implementation of soft tissue nonlinear viscoelasticity. J Biomech Eng. 2012;134(11):114501. doi: 10.1115/1.4007630. [DOI] [PubMed] [Google Scholar]
  • 9.Gardiner J C, Weiss J A, Rosenberg T D. Strain in the human medial collateral ligament during valgus loading of the knee. Clin Orthop Relat Res. 2001;(391):266–274. doi: 10.1097/00003086-200110000-00031. [DOI] [PubMed] [Google Scholar]
  • 10.Completo A, Noronha J C, Oliveira C, Fonseca F. Análise biomecânica da reconstrução do ligamento cruzado anterior. Rev Bras Ortop. 2019;54(02):190–197. doi: 10.1016/j.rbo.2017.11.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
Rev Bras Ortop (Sao Paulo). 2022 Jun 6;58(1):173–178. [Article in Portuguese]

Ensaios mecânicos monoaxiais nos ligamentos do joelho porcino *

Resumo

A falha da reconstrução ligamentar tem diferentes fatores de risco, dentre os quais podemos destacar o período antes da sua incorporação, o qual configura um período mecânico vulnerável. A perda de resistência ao longo do tempo é uma característica dos tecidos vivos. A dissecção com as inserções ósseas dos ligamentos cruzados de modelos animais não é descrita; todavia, para os ensaios monoaxiais, é fundamental extrair as informações de ensaios como os de relaxação. O presente trabalho realiza a descrição da dissecção utilizada para a geração de corpo de prova para a realização de ensaios não destrutivos para avaliar o comportamento mecânico. Realizamos dissecção de quatro ligamentos de joelho porcino, propondo uma técnica de dissecção para os ligamentos cruzados com as inserções ósseas para comparação com os colaterais. Os ligamentos foram submetidos a testes de relaxação e foram colocadas strain gauges durante os testes. Os resultados mostraram comportamento viscoelástico, validado pelas strain gauges e com uma perda ao longo do tempo, sendo que, em alguns ligamentos, as perdas chegaram a até 20%, fator este a ser considerado em trabalhos futuros. O presente trabalho dissecou os quatro principais ligamentos do joelho, demonstrando a abordagem posterior que permite manter as suas inserções ósseas e descrevendo a fixação para os ensaios uniaxiais monotônicos, além de ter conseguido extrair o comportamento viscoelástico dos quatro ligamentos do joelho dentro dos limites fisiológicos do joelho.

Palavras-chave: dissecção, fenômenos biomecânicos, articulação do joelho, resistência a tração, modelo animal

Introdução

A falha da reconstrução ligamentar tem diferentes fatores de risco, dentre os quais podemos destacar o período antes de sua incorporação, o qual configura um período mecânico vulnerável. 1

As lesões ligamentares do joelho apresentam uma frequência elevada devido a prática esportiva, dentre os quais destacamos as lesões dos quatro principais ligamentos: ligamento colateral medial (LCM), ligamento colateral lateral (LCL), ligamento cruzado anterior (LCA) e ligamento cruzado posterior (LCP).

A perda de resistência ao longo do tempo é uma característica dos tecidos vivos, como, por exemplo, o que ocorre com o comportamento viscoelástico de cada estrutura. Caso não haja na reconstrução ligamentar a mimetização deste comportamento, esta pode ser a causa de falha, mesmo com ressonâncias normais, como descrito por Ekdahl et al., 2 causando, em consequência, a redistribuição inadequada dos esforços nas estruturas do joelho.

Inúmeros trabalhos realizaram ensaios limite da resistência dos ligamentos, 3 4 5 mimetizando a ruptura ligamentar ou de enxertos. Todavia, não existe na literatura nenhum trabalho que estude os ligamentos de forma isolada dentro da faixa fisiológica de deformação e que tenha extraído as informações deste comportamento principalmente dos ligamentos cruzado anterior e posterior.

Os estudos podem ser destrutivos ou não destrutivos, embora os ensaios limite do tipo destrutivo sejam os mais encontrados na literatura. Todavia, para a extração dos comportamentos dos ligamentos e simular situações de comportamento que ocorrem dentro da fisiologia normal, alguns autores sugerem realizar os testes dentro dos limites fisiológicos de deformações (de 3 a 6%).

A dissecção com as inserções ósseas dos ligamentos cruzados de modelos animais não é descrita; todavia, para os ensaios monoaxiais, é fundamental extrair as informações de ensaios como os de relaxação, como proposto por Duenwald et al. 6 O presente trabalho realiza a descrição da dissecção dos quatro ligamentos utilizada para a geração de corpo de prova e a realização de ensaios não destrutivos (relaxação) em ensaios de tração monoaxial, com ênfase no LCA do joelho do porco.

Material e Método

O protocolo do estudo foi aprovado pela Comissão de Ética na Utilização de Animais como prescreve o Regulamento de Inspeção Sanitária e Industrial de Produtos de Origem Animal (MAPA).

Foram dissecados os quatro ligamentos (LCM, LCL, LCA e LCL) de joelho de porco com as suas inserções ósseas conforme descrito a seguir.

A dissecção foi realizada iniciando pelos ligamentos colaterais, sendo o primeiro o colateral medial, pela sua facilidade de abordagem, que é bem superficial. As inserções ósseas foram mantidas com o comprimento de 30 mm em cada extremidade para servir de pontos de fixação. Em cada fragmento ósseo (proximal e distal) foi realizado um furo de 6 mm para instalar os pinos de ligação com a máquina de ensaio de materiais Instron (5966 Norwood, Massachusetts, USA). Após a retirada dos colaterais, os cruzados são retirados em sequência. A dissecção deve ser realizada através da visão posterior ( Fig. 1 ).

Fig. 1.

Fig. 1

Visão posterior do joelho com a realização da osteotomia da inserção tibial.

O corte é realizado entre o cruzado posterior e anterior ao nível do fêmur, através de abordagem posterior ( Fig. 2 ).

Fig. 2.

Fig. 2

Visualização do local de osteotomia após medialização do plugue ósseo do ligamento cruzado posterior.

Após a separação da extremidade proximal, é realizada a retirada do ligamento cruzado posterior com o fragmento ósseo na tíbia proximal. Esta retirada permite a visualização da inserção tibial do ligamento cruzado anterior, permitindo a retirada da tíbia com fragmento ósseo ( Fig. 3 ).

Fig. 3.

Fig. 3

Osteotomia do fêmur entre os ligamentos cruzados.

Realizamos uma modificação na técnica de retirada dos ligamentos cruzados proposta no artigo de Skelley et al., 7 que retirava os ligamentos sem os fragmentos ósseos.

Até o presente momento, não há descrição de retiradas dos ligamentos cruzados com as suas inserções ósseas. A retirada com os fragmentos ósseos das inserções do LCA evita o escorregamento dos corpos de prova de ligamentos nas garras, proporcionando uma leitura adequada do comportamento viscoelástico do ligamento dentro dos limites fisiológicos.

No dia da dissecção, um modelo foi descongelado em temperatura ambiente climatizada (22°C) durante 5 horas, sendo depois dissecados e isolados os quatro principais ligamentos do joelho (LCA, LCP, LCM e LCL). As inserções ósseas dos 4 ligamentos foram mantidas com 30 mm de fragmento ósseo e foi realizado um furo de 6mm, transversal a cada inserção óssea, para acoplamento de pino para o correto posicionamento na máquina de ensaios de materiais. Os corpos de prova de ligamento de porco foram então colocados em plásticos e embebidos em gazes com soro fisiológico 0,9% e mantidos até a sua testagem, realizada no mesmo dia.

Os modelos foram então colocados na máquina de ensaio de materiais INSTRON, existente no laboratório LADES do CEFET/RJ com uma célula de carga de 1 kN. Em um momento inicial, o ligamento é submetido a uma carga axial de 10 N durante um período de 5 minutos para atingir o seu comprimento original ( Fig. 4 ), conforme descrito por Troyer et al. 8

Fig. 4.

Fig. 4

Ligamento acoplado com strain gauge no Instron.

Quanto aos ensaios de relaxação, foram utilizados o protocolo de 3 etapas proposto por Duenwald et al. 6 com deformações máximas fisiológicas e 6% e com deformações pequenas normalmente encontradas de 3%, como proposto pelo trabalho de Gardiner et al. 9

O protocolo proposto, conhecido como Hill-Valley-Hill, serviu para simular o comportamento em uma situação in vivo. Ele é caracterizado por realizar deformações dentro dos limites fisiológicos esperados para a mobilidade articular adequada, dentro de 3% (limite inferior) e 6% (limite superior) durante um período de 100 segundos e feito em 3 etapas: uma realizada como teste de relaxação (6%), seguida por uma de recuperação (3%) e por outra de relaxação (6%). 6 Depois, foram extraídos os dados dos resultados do comportamento ao longo do tempo com o objetivo de caracterização do comportamento viscoelástico dos ligamentos.

Resultados

Os resultados encontrados dos ensaios realizados nos quatro ligamentos são mostrados nos gráficos de força versus tempo, cujos sinais de força foram obtidos da célula de carga da INSTRON ( Fig. 5 ).

Fig. 5.

Fig. 5

( A ) Gráfico da leitura de força x tempo do ligamento colateral medial; ( B ) Gráfico da leitura de força x tempo do ligamento colateral lateral; ( C ) Gráfico da leitura de força x tempo do ligamento cruzado posterior; ( D ) Gráfico da leitura de força x tempo do ligamento cruzado anterior.

Ao analisar o comportamento do ligamento colateral medial, houve uma perda de > 10% da sua carga no teste de relaxação no primeiro ciclo, atenuando-se no segundo ciclo. O comportamento do LCL também apresentou uma queda > 10%; todavia, o segundo ciclo ainda apresentou uma perda de carga, porém de menor decréscimo, se estabilizando com 100 segundos.

A perda pelo componente viscoso do LCP foi evidente; o pico na deformação a 6% 160 N apresentou uma queda no 2° ciclo, apresentando um valor menor do que o do 1° ciclo, demonstrando o seu comportamento viscoelástico.

A perda pelo componente viscoso do ligamento cruzado anterior foi evidente e o pico na deformação a 6% 120 N apresentou uma queda no 2° ciclo, apresentando um valor menor do que o do 1° ciclo, que atingiu valores < 100 N, com uma diferença absoluta de 20 N (maior do que a do 2° ciclo).

Os resultados mostraram comportamento viscoelástico com uma perda ao longo do tempo, fator este a ser considerado em trabalhos futuros.

Comentários Finais

A presente nota técnica dissecou os quatro principais ligamentos do joelho demonstrando a etapa pela visão posterior extraindo com os ligamentos e descreveu a fixação para os ensaios uniaxiais monotônicos e conseguiu extrair o comportamento do padrão viscoelástico dos quatro ligamentos do joelho, dentro dos limites fisiológicos do joelho, através de corpos de prova adequados obtidos com as suas inserções ósseas. Todos os ligamentos testados mostraram perda da carga ao longo do tempo. Os achados apontam a importância de novos ensaios experimentais tanto de ligamentos quanto das opções tendinosas para a reconstrução, inclusive servindo de referência para escolhas de enxertos com comportamentos semelhantes aos ligamentos e para modelos de simulação. 10


Articles from Revista Brasileira de Ortopedia are provided here courtesy of Sociedade Brasileira de Ortopedia e Traumatologia

RESOURCES