Off-axis digital lensless holographic microscopy based on spatially multiplexed interferometry

José Ángel Picazo-Bueno a,b,*, Steffi Ketelhut a, Jürgen Schnekenburger a, Vicente Micó b, Björn Kemper a,*
PMCID: PMC11331263  PMID: 39161785

Abstract.

Significance

Digital holographic microscopy (DHM) is a label-free microscopy technique that provides time-resolved quantitative phase imaging (QPI) by measuring the optical path delay of light induced by transparent biological samples. DHM has been utilized for various biomedical applications, such as cancer research and sperm cell assessment, as well as for in vitro drug or toxicity testing. Its lensless version, digital lensless holographic microscopy (DLHM), is an emerging technology that offers size-reduced, lightweight, and cost-effective imaging systems. These features make DLHM applicable, for example, in limited resource laboratories, remote areas, and point-of-care applications.

Aim

In addition to the abovementioned advantages, in-line arrangements for DLHM also include the limitation of the twin-image presence, which can restrict accurate QPI. We therefore propose a compact lensless common-path interferometric off-axis approach that is capable of quantitative imaging of fast-moving biological specimens, such as living cells in flow.

Approach

We suggest lensless spatially multiplexed interferometric microscopy (LESSMIM) as a lens-free variant of the previously reported spatially multiplexed interferometric microscopy (SMIM) concept. LESSMIM comprises a common-path interferometric architecture that is based on a single diffraction grating to achieve digital off-axis holography. From a series of single-shot off-axis holograms, twin-image free and time-resolved QPI is achieved by commonly used methods for Fourier filtering-based reconstruction, aberration compensation, and numerical propagation.

Results

Initially, the LESSMIM concept is experimentally demonstrated by results from a resolution test chart and investigations on temporal stability. Then, the accuracy of QPI and capabilities for imaging of living adherent cell cultures is characterized. Finally, utilizing a microfluidic channel, the cytometry of suspended cells in flow is evaluated.

Conclusions

LESSMIM overcomes several limitations of in-line DLHM and provides fast time-resolved QPI in a compact optical arrangement. In summary, LESSMIM represents a promising technique with potential biomedical applications for fast imaging such as in imaging flow cytometry or sperm cell analysis.

Keywords: quantitative phase imaging, off-axis lensless holography, digital lensless holographic microscopy, label-free imaging, digital holographic microscopy, phase retrieval, spatially multiplexed interferometric microscopy

1. Introduction

Digital holographic microscopy (DHM) is a label-free microscopy technique providing accurate and time-resolved quantitative phase information of transparent biological samples.13 In contrast to other label-free techniques such as bright field, Zernike phase contrast,4 and differential interference contrast (DIC) microscopy,5 DHM quantitatively recovers the optical path delay of the light passing through semi-transparent samples.6,7 This enables quantitative phase imaging (QPI) with up to a nanometer sensitivity and extended depth of field, employing a non-invasive, full-field, real-time capable, non-contact, and static (without mechanical parts) operational principle.811 In earlier studies, various biomedical applications of DHM3,12,13 were demonstrated; these include cancer research,1416 in vitro cytotoxicity testing,1719 cell manipulation monitoring,20 immune cell analysis,2123 sperm cell assessment,24,25 detection of viral infections,26 diabetes screening,27 and SARS-Cov-2 detection and classification.28

Digital lensless holographic microscopy (DLHM) is a variant of DHM and offers a simple and compact microscopy scheme with high resolution and a wide field of view (FOV)29 by implementing a digital version of Gabor’s initial holography approach,30 in which a point source illuminates the sample in transmission and a digital sensor records the diffraction pattern.31 Then, the sample image is recovered by numerical propagation from the recording/hologram plane to the image plane.32 However, the initially proposed in-line holographic configuration causes an overlap between the numerically focused image and the unfocused diffraction pattern of the conjugate complex twin image. This overlap not only affects the quality of the final image by reducing the signal-to-noise ratio (SNR) but also prevents an accurate acquisition of QPI images.29

Various approaches have been developed to address the twin-image problem in DLHM.3349 These approaches include phase-shifting techniques33,34 and multi-height phase-retrieval procedures,35,4249 which require recording multiple holograms to eliminate the twin image contribution. However, these earlier reported methods are limited or not suitable for fast dynamic events due to the need for multiple hologram recordings. Alternatively, single-shot techniques, which allow for twin-image-free image reconstruction from a single hologram, have been devised.3641 These methods typically rely on iterative phase-retrieval algorithms with object mask constraints36,37 or multi-wavelength illumination.3841 Although effective in mitigating the twin image, iterative algorithms can be computationally intensive and prone to convergence issues.50 Other approaches involved deep learning, which showed promising results in rapidly removing the twin image for specific applications, although they required training the algorithm with a substantial amount of suitable ground-truth data.5155

An alternative approach to overcoming the twin-image problem in DLHM involves adopting an off-axis DLHM configuration.5659 The off-axis approach allows for image reconstruction from single captured holograms without twin image presence and without a requirement for recording multiple images, iterative algorithms, object constraints, multi-illumination sources, or machine learning. For example, Lu et al.56 utilized an off-axis DLHM setup using two pinholes to achieve an interferometric configuration. Serabyn et al.57 employed a pair of small gradient-index (GRIN) lenses for off-axis holographic recording. Rostykus and Moser58 proposed the use of a prism with a photopolymer layer to record two-volume hologram gratings, enabling the implementation of an off-axis lensless configuration. In addition, Ebrahimi et al.59 presented a common-path off-axis DLHM scheme that utilizes a Fresnel biprism to generate two spherical waves for off-axis holographic recordings.

Here, we report on a common-path off-axis DLHM approach entitled LEnslesS spatially multiplexed interferometric microscopy (LESSMIM) to achieve both, single-shot QPI and twin-image elimination. LESSMIM is inspired by a previously reported DHM technique known as spatially multiplexed interferometric microscopy (SMIM),6063 which extended a bright field microscope to a holographic one by introducing coherent illumination, leaving a clear region at the sample plane, and incorporating a diffraction grating. SMIM has been applied for super-resolution imaging,64,65 noise-reduced QPI,61,66 opposed-view QPI,67 and multimodal imaging.68 In this study, the application of SMIM is expanded to the field of DLHM-based QPI. Hence, LESSMIM implements a common-path off-axis interferometric configuration by assembling an illumination unit including a one-dimensional (1D) diffraction grating and by spatially multiplexing the sample plane. From a recorded off-axis hologram, LESSMIM achieves QPI by Fourier filtering-based reconstruction, aberration compensation, and numerical propagation.

Section 2 describes the concept and layout of our approach. Section 3 presents the experimental validation of LESSMIM using technical test targets, microspheres, and living cells. Experimental validations with living cells comprise evaluations with adherent cells having different sizes as well as evaluations with cells in flow using a microfluidic system with hydrodynamic focusing. Finally, Sec. 4 presents the discussion and conclusions of the proposed technique.

2. LESSMIM Concept and Experimental Setup

Figure 1 illustrates the optical concept of LESSMIM, consisting of an illumination unit and a digital recording sensor. Collimated coherent illumination is achieved by a fiber-coupled laser diode applying a collimating lens with the focal length fCL. A sinusoidal diffraction grating diffracts the illumination wave into three fractions, which are focused by an additional lens with an image focal length fFL to generate three spatially separated point sources in the focal plane. These replicas are laterally separated a distance

d=NλfFL, (1)

where N is the spatial frequency of the grating, defined as the inverse of the spatial period of the grating periodic structures, and λ represents the wavelength of the illuminating laser light.69 The three resulting point sources, which correspond to the three diffraction orders of the grating, are located at a distance z1 from the sample plane. The sample plane is divided into three regions from which one is blocked [object (o), reference (r), and blocking (x)], similar to previously reported SMIM implementations.60,63,65 The blocking region prevents spurious interferences, whereas the object and reference regions are areas with and without the sample for simultaneous object and reference wave transmission, respectively. The arrangement of these regions is shown in the upper panel of Fig. 1. The blocking region can be achieved in different ways: either a customized chamber with an opaque region is fabricated, a limiting aperture is placed at the plane where the point sources are generated to block one of the point sources, or a border of the opaque frame of the diffraction grating is used to block only a portion (one third) of the beam to produce the blocking effect. The most general use case to provide the blocking region is using an opaque border of the grating, but the choice of the different blocking options will depend on the experimental requirements of the specific setup. A digital sensor is positioned at a distance z2 from the sample plane to record an off-axis digital hologram. The recorded hologram results from the coherent overlapping between the Fresnel diffraction pattern from the o region (0th diffraction order) and the non-perturbed light from the sample free r region (1st order). Further details about the system alignment are provided in Sec. S1 of the Supplemental Material. Due to the divergence of the sample illumination, the recorded diffraction pattern of the object region is geometrically magnified with magnification M according to the relation70

M=(z1+z2)/z1. (2)

Fig. 1.

Fig. 1

Scheme of the LESSMIM layout consisting of an illumination unit and a digital recording sensor. fCL is the focal length of the collimating lens for object illumination with laser light, fFL is the focal length of the focusing lens for the creation of three-point sources, z1 is the distance between the point light sources and sample, z2 is the distance between the sample and digital sensor, o is the object region, r is the reference region, and x is the blocking region.

The field of view (FOV) is determined by the size of the CMOS recording area and magnification as

FOV=(Lp/M)2, (3)

where L is the number of pixels in each transversal direction and p is the pixel pitch of the digital sensor.70 The lateral resolution ρ is diffraction-limited and depends on the distance between the sample and sensor as well as on the sensor size as follows:70

ρ=2λz2/Lp. (4)

The experimental validation of the LESSMIM concept was performed by utilization of a single-mode fiber-coupled laser diode (Thorlabs LP515-SF3, output power 3 mW, central wavelength 515 nm) as the coherent illumination source. An achromatic collimating lens (Linos, AC focal length fCL=100  mm) was used for collimation of the illuminating laser light, and a 20×/0.40 NA microscope lens (ZEISS LD Achroplan) was employed as the lens for beam focusing. Note that a specific microscope objective was used in the experimental validation of our concept, but more cost-effective optical lenses can be utilized. A Ronchi ruling (Edmund Optics, spatial frequency 120 line pairs/mm) served as a diffraction grating to generate the three-point sources. To generate the blocking region, we blocked one third of the beam with one opaque border of the Ronchi ruling frame. A monochrome CMOS sensor (TheImagingSource DMK23UP1300, 1280×1024  pixels, 4.8  μm pixel size, image acquisition rate 95 fps) was employed for recording the generated digital off-axis holograms. Holograms were transferred to a computer via a USB 3.0 interface for numerical reconstruction and image processing downstream using MATLAB 2021b. Despite the rectangular shape of the digital sensor, a squared region of interest (ROI) of 1024×1024  pixels was defined to achieve an equal lateral resolution in the x and y directions, according to Eq. (4). The sample was positioned at a distance of z1=1.0±0.1  mm from the point source and z2=10.0±0.1  mm from the digital sensor. The entire setup had a length of 200 mm from the optical fiber tip to the digital sensor. According to Eq. (2), a magnification of M=11× was set, resulting in an FOV of 447×447  μm2, considering Eq. (3). Considering the above described optical and geometrical system parameters and Eq. (4), the theoretical spatial resolution is ρ=2.16±0.02  μm.

3. Experimental Validation

3.1. Demonstration of the Operation Principle with a Resolution Target

In the first step, the operation principle of LESSMIM was evaluated. For that, off-axis holograms from a resolution test target (positive 1951 USAF target, Thorlabs Inc., New Jersey, United States) were recorded. Figure 2 illustrates the subsequently performed numerical reconstruction process. Figure 2(a) shows a recorded off-axis Fresnel digital hologram with an included magnified inset depicting the generated parallel off-axis carrier fringe pattern. Figure 2(b) displays the corresponding well-separated site bands of the diffraction orders in the spatial frequency domain after a fast Fourier transformation. Fourier filtering was applied to one of the cross-correlation terms, as indicated with a dotted red circle in Fig. 2(b), to retrieve the complex object wave, as well as the corresponding amplitude [Fig. 2(c)] and phase [Fig. 2(d)] distributions of the diffracted light. The reconstructed phase image in Fig. 2(d) exhibits a comatic aberration. This aberration was generated by the propagation of the tilted reference wave across the focusing lens through a region outside the optical axis. To compensate for the comatic effect, an additional blank reference hologram of a clear region of the USAF test target without sample information was recorded. Alternatively, it is also possible to acquire a reference hologram without a sample in the optical path. The retrieved amplitude and phase distributions are included in Figs. 2(e) and 2(f). Subsequently, the retrieved complex waves with [Figs. 2(c) and 2(d)] and without [Figs. 2(e) and 2(f)] sample information, including amplitude and phase information, were subtracted by complex division of the object wave and the wave from the blank reference hologram. As evident in Figs. 2(g) and 2(h), the subtraction process efficiently removed aberrations, spurious reflections, and dust, resulting in a homogeneous image background. The further numerical propagation of the subtracted complex object wave to the image plane was performed utilizing the angular spectrum method71 yielding focused amplitude [Fig. 2(i)] and phase [Fig. 2(j)] images of the test chart structures. To compare LESSMIM with conventional DLHM, amplitude and phase images obtained by DLHM were recovered as follows: the diffraction grating was simply removed to record a conventional DLHM hologram without a separate reference wave. The reconstruction of the conventional hologram was performed by numerical propagation as described in Ref. 71, and a reference hologram without a sample was recorded and employed for achieving a homogeneous image background in the amplitude and phases images70 [Figs. 2(k) and 2(l)]. In both DLHM images [Figs. 2(k) and 2(l)], disturbances by twin-image presence are evident, whereas the images provided by LESSMIM [Figs. 2(i) and 2(j)] exhibit a considerably higher contrast. The quality improvement of the amplitude and phase images achieved by LESSMIM is also indicated by the enlarged color-coded areas shown in Fig. 2(m) [see also blue and yellow rectangular frames in Figs. 2(i)2(l)], in which the smallest elements of the applied USAF resolution test target are resolved (element 6 - group 7, G7-E6, period 4.38  μm). These observations are also supported by the cross-section profiles through the amplitude images [Fig. 2(n)] along the blue and red lines included in Figs. 2(i) and 2(k). Due to the smallest available structures of the applied USAF test target, the confirmed spatial resolution is lower than the theoretical value ρ=2.16  μm calculated in Sec. 2. For a more precise verification of the maximum achievable spatial resolution, a calibrated high-resolution USAF target, including group 8, element 6, which corresponds to a period of 2.19  μm, is required.

Fig. 2.

Fig. 2

Illustration of the LESSMIM working principle for the example of a USAF resolution test target: (a) recorded off-axis hologram; (b) Fourier transformation of panel (a) with the filtered spectral region (marked with a dotted red circle); (c) amplitude and (d) phase reconstructions after Fourier filtering application; (e) amplitude and (f) phase reconstructions of the reference hologram; (g) amplitude and (h) phase images after subtraction of the amplitude and phase distributions retrieved from the reference hologram; (i) amplitude and (j) phase images after numerical propagation to the focus plane; and (k) amplitude and (l) phase images recovered by conventional in-line DLHM without a separate reference wave for direct comparison; (m) color-coded magnified images of regions in panels (i)–(l) containing the smallest elements of the resolution target; and (n) intensity profiles along blue and red lines in panels (i) and (k), respectively. The scale bar in panel (a) corresponds to a length of 20  μm.

3.2. Temporal Stability

The temporal stability of LESSMIM was assessed by measuring phase fluctuations over time. Therefore, a blank microscope slide was placed at the sample location, and 300 off-axis holograms were sequentially recorded over a period of 5 min at a hologram acquisition rate of 1 Hz. After numerical QPI image reconstruction, within an ROI of 300×300  px, the variation Δφ(t)¯=φ(t)¯φ(t=0)¯ of the average phase of each frame φ(t) with respect to the average phase of the initially acquired QPI image φ(t=0)¯ was determined. The plot of the temporal development Δφ(t)¯ is presented in Fig. 3(a). Figure 3(b) depicts the spatial distribution corresponding of the standard deviation (STD) σt(m,n)=1300t=0300Δφ(m,n,t)¯2 that was calculated for each pixel of the entire stack of 300 QPI images. The corresponding histogram of the computed values σt is depicted in Fig. 3(c). The mean value σt¯=0.07  rad indicates a high temporal phase stability.

Fig. 3.

Fig. 3

Temporal phase stability of LESSMIM during a period of 300 s within an ROI of 300×300  pixels utilizing a blank microscope slide as the sample location (hologram acquisition rate of 1 Hz). (a) temporal (t) development of the averaged phase Δφ(t)¯; (b) spatial distribution of the standard deviation σt(m,n) of the phase differences per pixel determined from the entire stack of 300 QPI images; and (c) histogram of the σt from the data in panel (b). The mean value σt¯=0.07  rad quantifies a high temporal stability of the setup. The scale bar in panel (b) corresponds to 20  μm.

3.3. Evaluation of QPI Accuracy Using Microspheres

The performance of LESSMIM for QPI was validated experimentally by analyzing PMMA microspheres with a diameter of 9.8  μm (PolyAn GmbH, lot number PT1140130FS) that were immersed in a mixture of glycerin and water (90%/10%). Figure 4(a) depicts a recorded digital off-axis hologram of 3  μA. The enlarged area of the yellow-framed region visualizes the off-axis carrier fringe pattern. Following the recovery process described in Sec. 3.1, the numerically focused QPI image of the microsphere enclosed in the red-framed square of Fig. 4(a) is obtained [Fig. 4(b)]. Here, a blank reference hologram was achieved by imaging a region of the sample without microspheres. Considering the refractive indices of the microspheres nPMMA=1.494 (Ref. 72) and the immersion liquid nmedium=1.458 (measured with an Abbe refractometer at λ=515  nm), the thickness distribution d(x,y)=Δθ(x,y)·λ/2πΔn of the microsphere was computed from the background-corrected phase distribution Δθ(x,y) of the QPI image in Fig. 4(b) and the RI difference Δn=nPMMAnmedium between the microspheres and the surrounding glycerol-water mixture.73 A pseudo-three-dimensional (3D) representation of the microsphere thickness distribution is presented in Fig. 4(c). Figure 4(d) shows a thickness profile along the dotted blue line in Fig. 4(b). To obtain the maximum thickness of the microsphere, we evaluated 10 data points around the center of the microsphere from which the mean value as well as the standard deviation were calculated. The obtained value dmax=9.7±0.2  μm for the maximum thickness matches the microsphere diameter of 9.8  μm provided by the manufacturer and validates LESSMIM for accurate QPI.

Fig. 4.

Fig. 4

Validation of QPI image retrieval with LESSMIM by analysis of PMMA microspheres in a glycerol/water mixture (90%/10%). (a) Recorded off-axis hologram, (b) reconstructed numerically focused QPI image, (c) gray level coded pseudo-3D representation of thickness distribution computed from phase data in panel (b), and (d) thickness profile along the dotted blue line in panel (b). Yellow and white scale bars in panels (a) and (b) correspond to 100 and 20  μm, respectively.

3.4. QPI of Living Adherently Grown Pancreatic Tumor Cells

To assess the capability of LESSMIM for QPI of biological samples, investigations on living adherent pancreatic tumor cells (PaTu 8988T) were performed.74,75 Subconfluently grown cells were observed in Dulbecco modified Eagle’s medium (DMEM) between a glass object carrier slide and a cover slip (thickness of 175  μm). A blank reference hologram was achieved by imaging a sample region without cells. Figure 5 presents experimental results from adherent cells. Rows include different adherent cells exhibiting different morphologies and thicknesses. The first column (a1 to a4) in Fig. 5 displays the recorded off-axis holograms. In the second column (b1 to b4) of Fig. 5, the corresponding numerically focused QPI images are shown. Similar to the analyzed microspheres in Sec. 3.3, the cell thickness was determined by considering an average cellular refractive index of nPaTuT=1.365476 and a refractive index of nDMEM=1.339 of the culture medium (measured by an Abbe refractometer). Pseudo-3D visualizations of the resulting cell thickness distributions are plotted in the third column (c1 to c4) of Fig. 5. Finally, Figs. 5(d1)5(d4) include thickness profiles along the dotted blue lines in Figs. 5(b1)5(b4). Additional results from isolated PaTu 8988T single cells are shown in Fig. S1 in the Supplemental Material. The experimental results demonstrate that LESSMIM successfully provided QPI images of both adjacent and isolated living adherent cells having different morphologies and axial dimensions, even for the case of thin adherent cells (as shown in row 4 of Fig. 5).

Fig. 5.

Fig. 5

Evaluation of LESSMIM for QPI of living adherent pancreatic tumor cells (PaTu 8988T). Rows (1)–(4): ROIs containing cells with various morphologies and thicknesses. First column (a1–a4): recorded off-axis holograms; second column (b1–b4): reconstructed focused QPI images; third column (c1–c4): gray level coded pseudo-3D plots of the cell thickness distributions calculated from panels (b1–b4); and fourth column (d1–d4): thickness profile along blue dotted lines marked in panels (b1–b4). Scale bars in panels (a1–a4) correspond to 20  μm.

3.5. Quantitative Phase Imaging of Cells in Flow Within a Microfluidic Device

The capability of LESSMIM for QPI of fast events and imaging flow cytometry (IFC) was evaluated by observation of living suspended PaTu 8988 T cells in a microfluidic chip within a rectangular cross-section (channel width of 1 mm and channel height of 20  μm) and hydrodynamic focusing capabilities as sketched in Fig. 6(a). The microfluidic device employed a syringe pump (Nemesys, CETONI GmbH, Korbußen, Germany) to direct the cells toward the microfluidic channel. The microfluidic channel was realized by standard soft lithography in polydimethylsiloxane (PDMS, Dow Corning, Midland, Michigan, United States), with a substrate that was 1 mm thick, and the PDMS was bonded on a glass coverslip (thickness of 175  μm) by an air-plasma activation as described in Refs. 77 and 78. To achieve lateral hydrodynamic focusing, the chip circuit was designed in co-flow architecture with an included inlet for splitting the sheath fluid into two separate fractions at the lateral channel borders and another inlet for the sample fluid in the center. The sheath fluid was introduced with two channels at an angle of 30 deg with respect to the channel for the sample fluid. The enlarged green-framed region in the right panel of Fig. 6(a) indicates the sample plane imaged by the LESSMIM setup and illustrates the need for hydrodynamic focusing to prevent the presence of cells within the reference area.

Fig. 6.

Fig. 6

Evaluation of LESSMIM for IFC by investigations on living Patu 8988T cells in a microfluidic chip with hydrodynamic focusing capabilities. (a) Scheme of the utilized microfluidic chip with the enlarged region (green-framed rectangular) considered for lensless holographic imaging by spatial multiplexing (x is the blocking region, o is the object region, and r is the reference region). (b)–(d) Representative images of a recorded movie of 5 s that was recorded at a hologram acquisition rate of 40 fps with an exposure time of 2 ms (Video 1). (b) Representative off-axis hologram (the yellow arrow indicates the direction of the sample fluid stream); (c) focused QPI image reconstructed from panel (b); and (d) pseudo-3D thickness distribution computed from panel (c), with green outlines being the area retrieved by segmentation and indicating the cell boundaries. (e)–(f) Simultaneous visualization of all detected cells in the recorded QPI image stack achieved by maximum intensity projection (MIP); (e) MIP QPI image; and (f) corresponding pseudo-3D thickness representation with outlined cell boundaries. (Video 1, MOV, 7.67 MB [URL: https://doi.org/10.1117/1.JBO.29.S2.S22715.s1]).

For the experiments, Patu 8988 T cells were suspended in DMEM at concentrations of 1.5·106  cells/ml and observed with LESSMIM in a laminar flow at flow rates of 10  μl/min and 2  μl/min for the sheath fluid and the sample fluid, respectively. Off-axis holograms were recorded for 5 s at an acquisition rate of 40 fps and an exposure time of 2 ms. A blank reference hologram was recorded prior to the suspended cells passing through the microfluidic channel. The resulting holograms and QPI images are presented in Fig. 6 and Video 1. Figure 6(b) shows an exemplary off-axis hologram of a recorded image stack. The yellow arrow indicates the direction of the sample fluid stream. A representative-focused QPI image recovered from the hologram in Fig. 6(b) is depicted in Fig. 6(c). Figure 6(d) presents a pseudo-3D plot of the thickness distribution calculated from Fig. 6(c) by considering the same average RI value for PaTu 8988T cells as for the experiments in Sec. 3.4. Green outlines in Fig. 6(d) that were generated by the Otsu thresholding algorithm68 illustrate a reliable segmentation of cells in the QPI images and indicate the cellular boundaries. Figures 6(e) and 6(f) show images generated by maximum intensity projection (MIP) using Fiji,79 in which all cells that were detected in the recorded QPI image stack are visible. Figure 6(e) displays the combined image of the recovered focused QPI images, and Fig. 6(f) presents the corresponding image of the pseudo-3D thickness distributions with outlined cellular boundaries.

To further evaluate the suitability of LESSMIM for IFC, biophysical features including the projected radius, integral refractive index, and cellular dry mass of the investigated cells in flow were determined from the recorded QPI image stacks. A total of 300 single cells (included in Video 1) were segmented and analyzed, whereas clustered cells in the respective images were discarded. Therefore, in the first step, segmentation was performed in each QPI image of a recorded stack, as illustrated in Fig. 6(c), to determine the projected area S of every analyzed single cell. Subsequently, assuming a spherical shape, as typical for suspended PaTu 8988 T cells,75 the cell radius R was calculated from the parameter S as R=S/π. Moreover, as described in Refs. 75 and 76, the cellular dry mass DM=10λΔφ¯S/2πβ was calculated from the average phase value Δφ¯ induced by the analyzed cell and the RI increment β of the intracellular content, which was estimated to be 0.002  m3/kg for the investigated pancreatic tumor cell line.80 Finally, the integral RI n=nDMEM+Δφ¯·λ/2πh¯ of the cells was computed, where h¯ represents the mean cell thickness determined from QPI images of spherical cells as described in Refs. 75 and 76. Figure 7(a) shows the plot of the retrieved integral cellular RI n versus the corresponding projected cell radius R for the 300 analyzed cells, and Fig. 7(b) displays the relative frequency histogram of the corresponding dry mass DM values. The mean values R¯=8.64±0.06  μm, n¯=1.366±0.001, and DM¯=376±7  pg are in good agreement with experimental data from previous investigations on PaTu 8988T cells.76

Fig. 7.

Fig. 7

Determination of biophysical parameters from 300 individually analyzed suspended pancreatic tumor cells (Patu 8988T) in flow as illustrated in Fig. 6. (a) Scatter plot of the integral cellular RI n versus the projected cell radius R and (b) the relative frequency histogram of the dry mass DM.

4. Discussion and Conclusions

In this study, we propose lensless spatially multiplexed interferometric microscopy (LESSMIM) as a single-shot DLHM technique to provide accurate QPI reconstruction and twin-image removal. LESSMIM is a simplified variant of the previously reported SMIM concept,60,63 in which imaging lenses are no longer required. LESSMIM comprises a common-path interferometric architecture that is based on a single diffraction grating to achieve digital off-axis holography. QPI of biomedical samples is achieved from single off-axis holograms using Fourier filtering and numerical propagation procedures. First, we characterized the technique concerning the lateral resolution (Fig. 2), temporal stability (Fig. 3), and QPI accuracy (Fig. 4) by utilizing a USAF resolution test target, a blank object carrier slide, and PMMA microspheres, respectively. Moreover, we demonstrated, for the example of pancreatic tumor cells, the capabilities of LESSMIM for QPI of living adherent cells (Fig. 5, Fig. S1 in the Supplemental Material) with different morphologies and thicknesses and for the retrieval of sets of biophysical parameters from fast-moving cellular specimens within a microfluidic channel with hydrodynamic focusing capabilities (Figs. 6 and 7).

LESSMIM offers several advantages over other DHM techniques. It utilizes a common-path interferometric architecture that leads to higher temporal stability and lower demands on the coherence properties of the utilized light source than double-path schemes.64,8183 Although in our study, the geometry and the quality of optics in the illumination unit were selected to achieve a flexible experimental arrangement, LESSMIM permits a simpler, more compact, and more cost-effective design than typical Mach-Zehnder interferometer-based DHM arrangements.8,14,15,84 In particular, when LESSMIM is compared with common-path DHM systems such as diffraction phase microscopy,82 LESSMIM avoids the use and precise alignment of spatial filters, but at the cost of requiring a blank reference region near the specimen. Moreover, due to the spherical divergent wavefront employed for sample illumination, it enables variable magnification by simply changing the axial position of the point sources while maintaining spatial resolution.70 This characteristic allows for flexibility in the FOV size, which can be extended up to the size of the digital sensor for a magnification equal to 1 [see Eq. (3)], as in wide-field on-chip microscopy.48,85 With respect to previously reported in-line DLHM approaches, LESSMIM overcomes the limitation of the twin image presence and enables an accurate reconstruction of QPI images. Moreover, in comparison with earlier reported off-axis DLHM concepts that are based on two-pinholes, Fresnel biprism, a pair of GRIN lenses, or a prism with two-volume hologram gratings, the diffraction grating-based off-axis principle of LESSMIM offers a simplified arrangement, alignment, and handling; improved light transmission efficiency; and increased cost effectiveness.5659 In addition, LESSMIM provides similar or even higher spatial resolution than previously reported off-axis DLHM approaches5659 and an FOV comparable to or slightly higher than that described in previous works.56,59 Regarding accuracy and temporal stability, only a few previously reported related approaches58,59 analyzed these features. In this study, we demonstrated that LESSMIM provides accurate QPI images and a temporal stability at an observation period of 5 min that is comparable to previously reported approaches.58,59

However, LESSMIM also has limitations. In LESSMIM, an object-free region near the sample is required, which limits the analysis of extended samples, as happens in lens-based SMIM systems.60,63 However, this issue can be addressed by customized micropatterned sample carriers.86 When LESSMIM is combined with a microfluidic system for IFC, this drawback can be overcome by either designing a microfluidic channel with a small width or, as in our study, employing microfluidic systems with hydrodynamic focusing. The spatially multiplexed approach also reduces the object region to one third of the illuminated FOV. However, in practice, this limitation is not significant as the illuminated sample region typically is larger than the recorded FOV. Another limitation of LESSMIM is the recording of out-of-focus holograms because it requires numerical propagation algorithms to achieve sharply focused QPI images, which increases the computational amounts and prevents correlative imaging with other modalities, such as fluorescence or bright field microscopy. Moreover, the system introduces coma aberration, which can be compensated for by either recording an additional reference hologram (as shown in Fig. 2), using various established state-of-the-art computational methods87,88 such as the utilization of Zernike polynomials,89 or more recent approaches such as principal component analysis90 or deep learning,91 etc.92,93 It is worth noting that this issue is not critical for the application of LESSMIM to IFC because a reference hologram can be recorded at the beginning of the experiment when no sample fluid is present inside the microfluidic channel. When compared with state-of-the-art DLHM systems, the LESSMIM setup presents relatively large axial dimensions (around 200 mm), but it can be further reduced by optimization of the axial distances (lenses with lower focal lengths and/or a smaller separation between lenses).

In summary, LESSMIM has been demonstrated to be a compact and cost-effective method for accurate QPI that is also capable of imaging fast biomedical events. Future developments include the further miniaturization of the setup, investigations on optical components for further cost reduction, and methods for digital correction of phase aberrations. In particular, we will investigate changing the microscope objective, utilized in our study for illumination focusing, to a more cost-effective lens with a high numerical aperture for which additional aberrations can be assumed and that can be addressed numerically by adapted aberration correction algorithms. In addition, investigations to further reduce optical elements and space within the illumination unit will be performed to minimize the dimensions of the LESSMIM setup and to increase the cost effectiveness. LESSMIM will be useful for applications in various biomedical fields, particularly those for which analysis of fast-moving samples is required such as imaging flow cytometry94 or sperm assessment,25 as it is a DHM technique particularly valuable for point-of-care diagnosis and laboratories with limited budgets.

Supplementary Material

JBO_029_S22715_SD001.pdf (395.2KB, pdf)
Download video file (7.7MB, mov)

Acknowledgments

This work has been partially funded by MCIN/AEI/10.13039/501100011033 (Grant No. PID2020-120056GB-C21). J.A.P.B. is supported by the Spanish Grant “Margarita Salas” (Grant No. MS21-100) of the Ministry of Universities of the Government of Spain (Grant No. UP2021-044) funded by the European Union, NextGenerationEU. We further acknowledge support from the Open Access Publication Fund of the University of Muenster. Moreover, the authors would like to thank the Nonlinear Photonics Group from the Institute of Applied Physics at the University of Muenster for providing the microfluidics system with special thanks go to Dr. Riccardo Zamboni for his assistance and technical support with the microfluidic devices.

Biographies

José Ángel Picazo-Bueno is a postdoctoral fellow at the University of Valencia, Spain, specializing in physics. He received his BS and MS degrees and his PhD in physics from the University of Valencia in 2013, 2014, and 2020, respectively. With over 25 published journal papers and contributions to over 20 international conferences, his research focuses on digital holographic microscopy, quantitative phase imaging, lensless holography, and super-resolution imaging. Currently, he is conducting research at the Biomedical Technology Center of the University of Münster in Münster, Germany. He is an active member of SPIE.

Biographies of the other authors are not available.

Contributor Information

José Ángel Picazo-Bueno, Email: J.Angel.Picazo@uv.es.

Steffi Ketelhut, Email: ketelhut@uni-muenster.de.

Jürgen Schnekenburger, Email: schneke@uni-muenster.de.

Vicente Micó, Email: vicente.mico@uv.es.

Björn Kemper, Email: bkemper@uni-muenster.de.

Disclosures

The authors declare that there are no conflicts of interest, financial or otherwise, related to this article.

Code and Data Availability

Data underlying the results presented in this article are available by a reasonable request to the corresponding authors.

References

  • 1.Kim M. K., Digital holographic microscopy. Principles, techniques, and applications, Springer US; (2011). [Google Scholar]
  • 2.Park Y. K., Depeursinge C., Popescu G., “Quantitative phase imaging in biomedicine,” Nat. Photonics 12(10), 578–589 (2018). 10.1038/s41566-018-0253-x [DOI] [Google Scholar]
  • 3.Kemper B., et al. , “Label-free quantitative in vitro live cell imaging with digital holographic microscopy,” in Bioanalytical Reviews, Wegener J., Ed., Vol. 2, Springer, Cham: (2019). [Google Scholar]
  • 4.Zernike F., “Phase contrast, a new method for the microscopic observation of transparent objects,” Physica 9(7), 686–698 (1942). 10.1016/S0031-8914(42)80035-X [DOI] [Google Scholar]
  • 5.Allen R., David G., Nomarski G., “The Zeiss-Nomarski differential interference equipment for transmitted-light microscopy,” Zeitschrift fur wissenschaftliche Mikroskopie 69(4), 193–221 (1969). [PubMed] [Google Scholar]
  • 6.Marquet P., Depeursinge C., Magistretti P. J., “Review of quantitative phase-digital holographic microscopy: promising novel imaging technique to resolve neuronal network activity and identify cellular biomarkers of psychiatric disorders,” Neurophotonics 1(2), 020901 (2014). 10.1117/1.NPh.1.2.020901 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Majeed H., et al. , “Quantitative phase imaging for medical diagnosis,” J. Biophotonics 10(2), 177–205 (2017). 10.1002/jbio.201600113 [DOI] [PubMed] [Google Scholar]
  • 8.Marquet P., et al. , “Digital holographic microscopy: a noninvasive contrast imaging technique allowing quantitative visualization of living cells with subwavelength axial accuracy,” Opt. Lett. 30(5), 468–470 (2005). 10.1364/OL.30.000468 [DOI] [PubMed] [Google Scholar]
  • 9.Ferraro P., et al. , “Extended focused image in microscopy by digital holography,” Opt. Express 13(18), 6738–6749 (2005). 10.1364/OPEX.13.006738 [DOI] [PubMed] [Google Scholar]
  • 10.Kemper B., von Bally G., “Digital holographic microscopy for live cell applications and technical inspection,” Appl. Opt. 47(4), A52–A61 (2008). 10.1364/AO.47.000A52 [DOI] [PubMed] [Google Scholar]
  • 11.Kim M. K., “Principles and techniques of digital holographic microscopy,” SPIE Rev. 1(1), 018005 (2010). 10.1117/6.0000006 [DOI] [Google Scholar]
  • 12.Popescu G., Quantitative phase imaging of cells and tissues, McGraw-Hill; (2011). [Google Scholar]
  • 13.Picart P., New techniques in digital holography, iSTE Press and John Wiley & Sons, London: (2015). [Google Scholar]
  • 14.Dubois F., et al. , “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt. 11(5), 054032 (2006). 10.1117/1.2357174 [DOI] [PubMed] [Google Scholar]
  • 15.Kemper B., et al. , “Investigation of living pancreas tumor cells by digital holographic microscopy,” J. Biomed. Opt. 11(3), 034005 (2006). 10.1117/1.2204609 [DOI] [PubMed] [Google Scholar]
  • 16.Ugele M., et al. , “Label-free high-throughput leukemia detection by holographic microscopy,” Adv. Sci. 5(12), 1800761 (2018). 10.1002/advs.201800761 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Kühn J., et al. , “Label-free cytotoxicity screening assay by digital holographic microscopy,” Assay Drug Dev. Technol. 11(2), 101– 107 (2013). 10.1089/adt.2012.476 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Eder K. M., et al. , “Label-free digital holographic microscopy for in vitro cytotoxic effect quantification of organic nanoparticles,” Cells 11(4), 644 (2022). 10.3390/cells11040644 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Marzi A., et al. , “Interlaboratory evaluation of a digital holographic microscopy–based assay for label-free in vitro cytotoxicity testing of polymeric nanocarriers,” Drug Deliv. Transl. Res. 12(9), 2207–2224 (2022). 10.1007/s13346-022-01207-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Barroso Á., et al. , “Three-dimensional exploration and mechano-biophysical analysis of the inner structure of living cells,” Small 9(6), 885–893 (2013). 10.1002/smll.201201851 [DOI] [PubMed] [Google Scholar]
  • 21.Calin V. L., et al. , “Digital holographic microscopy evaluation of dynamic cell response to electroporation,” Biomed. Opt. Express 12(4), 2519–2530 (2021). 10.1364/BOE.421959 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Steike D. R., et al. , “Digital holographic microscopy for label-free detection of leukocyte alternations associated with perioperative inflammation after cardiac surgery,” Cells 11(4), 755 (2022). 10.3390/cells11040755 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.vom Werth K. L., et al. , “Application of digital holographic microscopy to analyze changes in t-cell morphology in response to bacterial challenge,” Cells 12(5), 762 (2023). 10.3390/cells12050762 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Di Caprio G., et al. , “Holographic imaging of unlabelled sperm cells for semen analysis: a review,” J. Biophotonics 8(10), 779–789 (2015). 10.1002/jbio.201400093 [DOI] [PubMed] [Google Scholar]
  • 25.Soler C., et al. , “Effect of counting chamber depth on the accuracy of lensless microscopy for the assessment of boar sperm motility,” Reprod. Fertil. Dev. 30(6), 924–934 (2018). 10.1071/RD17467 [DOI] [PubMed] [Google Scholar]
  • 26.Anand A., et al. , “Automatic identification of malaria-infected RBC with digital holographic microscopy using correlation algorithms,” IEEE Photonics J. 4(5), 1456–1464 (2012). 10.1109/JPHOT.2012.2210199 [DOI] [Google Scholar]
  • 27.Doblas A., et al. , “Diabetes screening by telecentric digital holographic microscopy,” J. Microsc. 261(3), 285–290 (2016). 10.1111/jmi.12331 [DOI] [PubMed] [Google Scholar]
  • 28.Goswami N., et al. , “Label-free SARS-CoV-2 detection and classification using phase imaging with computational specificity,” Light Sci. Appl. 10, 176 (2021). 10.1038/s41377-021-00620-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Wu Y., Ozcan A., “Lensless digital holographic microscopy and its applications in biomedicine and environmental monitoring,” Methods 136, 4–16 (2018). 10.1016/j.ymeth.2017.08.013 [DOI] [PubMed] [Google Scholar]
  • 30.Gabor D., “A new microscopic principle,” Nature 161(4098), 777–778 (1948). 10.1038/161777a0 [DOI] [PubMed] [Google Scholar]
  • 31.Garcia-sucerquia J., et al. , “Digital in-line holographic microscopy,” Appl. Opt. 45(5), 836–850 (2006). 10.1364/AO.45.000836 [DOI] [PubMed] [Google Scholar]
  • 32.Xu W., et al. , “Digital in-line holography for biological applications,” Proc. Natl. Acad. Sci. 98(20), 11301–11305 (2001). 10.1073/pnas.191361398 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Micó V., et al. , “Phase-shifting Gabor holography,” Opt. Lett. 34(10), 1492–1494 (2009). 10.1364/OL.34.001492 [DOI] [PubMed] [Google Scholar]
  • 34.Micó V., García J., “Common-path phase-shifting lensless holographic microscopy,” Opt. Lett. 35(23), 3919–3921 (2010). 10.1364/OL.35.003919 [DOI] [PubMed] [Google Scholar]
  • 35.Rivenson Y., et al. , “Sparsity-based multi-height phase recovery in holographic microscopy,” Sci. Rep. 6, 37862 (2016). 10.1038/srep37862 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Mudanyali O., et al. , “Compact, light-weight and cost-effective microscope based on lensless incoherent holography for telemedicine applications,” Lab Chip 10, 1417–1428 (2010). 10.1039/c000453g [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Biener G., et al. , “Combined reflection and transmission microscope for telemedicine applications in field settings,” Lab Chip 11(16), 2738–2743 (2011). 10.1039/c1lc20169g [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Noom D. W. E., et al. , “High-speed multi-wavelength Fresnel diffraction imaging,” Opt. Express 22(25), 30504–30511 (2014). 10.1364/OE.22.030504 [DOI] [PubMed] [Google Scholar]
  • 39.Sanz M., et al. , “Improved quantitative phase imaging in lensless microscopy by single-shot multi-wavelength illumination using a fast convergence algorithm,” Opt. Express 23(16), 21352 (2015). 10.1364/OE.23.021352 [DOI] [PubMed] [Google Scholar]
  • 40.Sanz M., et al. , “Compact, cost-effective and field-portable microscope prototype based on MISHELF microscopy,” Sci. Rep. 7(1), 1–12 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Picazo-Bueno J. Á., et al. , “Multi-illumination single-holographic-exposure lensless Fresnel (MISHELF) microscopy: principles and biomedical applications,” Sensors 23(3), 1472 (2023). 10.3390/s23031472 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Pedrini G., et al. , “Whole optical wave field reconstruction from double or multi in-line holograms by phase retrieval algorithm,” Opt. Express 11(24), 3234–3241 (2003). 10.1364/OE.11.003234 [DOI] [PubMed] [Google Scholar]
  • 43.Pedrini G., Osten W., Zhang Y., “Wave-front reconstruction from a sequence of interferograms recorded at different planes,” Opt. Lett. 30(8), 833–835 (2005). 10.1364/OL.30.000833 [DOI] [PubMed] [Google Scholar]
  • 44.Bao P., et al. , “Phase retrieval using multiple illumination wavelengths,” Opt. Lett. 33(4), 309–311 (2008). 10.1364/OL.33.000309 [DOI] [PubMed] [Google Scholar]
  • 45.Camacho L., et al. , “Quantitative phase microscopy using defocusing by means of a spatial light modulator,” Opt. Express 18(7), 6755–6766 (2010). 10.1364/OE.18.006755 [DOI] [PubMed] [Google Scholar]
  • 46.Bao P., et al. , “Lensless phase microscopy using phase retrieval with multiple illumination wavelengths,” Appl. Opt. 51(22), 5486–5494 (2012). 10.1364/AO.51.005486 [DOI] [PubMed] [Google Scholar]
  • 47.Greenbaum A., Ozcan A., “Maskless imaging of dense samples using pixel super-resolution based multi-height lensfree on-chip microscopy,” Opt. Express 20(3), 3129 (2012). 10.1364/OE.20.003129 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Greenbaum A., et al. , “Imaging without lenses: achievements and remaining challenges of wide-field on-chip microscopy,” Nat. Methods 9, 889–895 (2012). 10.1038/nmeth.2114 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Greenbaum A., et al. , “Wide-field computational imaging of pathology slides using lens-free on-chip microscopy,” Sci. Transl. Med. 6(267), 175 (2014). 10.1126/scitranslmed.3009850 [DOI] [PubMed] [Google Scholar]
  • 50.Fienup J. R., “Phase retrieval algorithms: a comparison,” Appl. Opt. 21(15), 2758–2769 (1982). 10.1364/AO.21.002758 [DOI] [PubMed] [Google Scholar]
  • 51.Rivenson Y., et al. , “Phase recovery and holographic image reconstruction using deep learning in neural networks,” Light Sci. Appl. 7(2), 17141–17141 (2017). 10.1038/lsa.2017.141 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Wu Y. C., et al. , “Air quality monitoring using mobile microscopy and machine learning,” Light Sci. Appl. 6(9), e17046 (2017). 10.1038/lsa.2017.46 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Göröcs Z., et al. , “A deep learning-enabled portable imaging flow cytometer for cost-effective, high-throughput, and label-free analysis of natural water samples,” Light Sci. Appl. 7, 66 (2018). 10.1038/s41377-018-0067-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Rivenson Y., Wu Y., Ozcan A., “Deep learning in holography and coherent imaging,” Light Sci. Appl. 8, 85 (2019). 10.1038/s41377-019-0196-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Liu T., et al. , “Deep learning-based holographic polarization microscopy,” ACS Photonics 7(11), 3023–3034 (2020). 10.1021/acsphotonics.0c01051 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Lu Y., Liu Y., Lau T. K., “Simple, portable, and low-cost microscope based on off-axis digital holography using two spherical waves,” Opt. Lett. 39(15), 4549–4552 (2014). 10.1364/OL.39.004549 [DOI] [PubMed] [Google Scholar]
  • 57.Serabyn E., et al. , “Compact, lensless digital holographic microscope for remote microbiology,” Opt. Express 24(25), 28540–28548 (2016). 10.1364/OE.24.028540 [DOI] [PubMed] [Google Scholar]
  • 58.Rostykus M., Moser C., “Compact lensless off-axis transmission digital holographic microscope,” Opt. Express 25(14), 16652–16659 (2017). 10.1364/OE.25.016652 [DOI] [PubMed] [Google Scholar]
  • 59.Ebrahimi S., et al. , “Common-path lensless digital holographic microscope employing a Fresnel biprism,” Opt. Lasers Eng. 128, 106014 (2020). 10.1016/j.optlaseng.2020.106014 [DOI] [Google Scholar]
  • 60.Mico V., et al. , “Spatially-multiplexed interferometric microscopy (SMIM): converting a standard microscope into a holographic one,” Opt. Express 22(12), 14929 (2014). 10.1364/OE.22.014929 [DOI] [PubMed] [Google Scholar]
  • 61.Picazo-Bueno J. Á., et al. , “Spatially multiplexed interferometric microscopy with partially coherent illumination,” J. Biomed. Opt. 21(10), 106007 (2016). 10.1117/1.JBO.21.10.106007 [DOI] [PubMed] [Google Scholar]
  • 62.Trusiak M., et al. , “Single-shot two-frame π-shifted spatially multiplexed interference phase microscopy,” J. Biomed. Opt. 24(9), 096004 (2019). 10.1117/1.JBO.24.9.096004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Picazo-Bueno J. Á., et al. , “Spatially multiplexed interferometric microscopy: principles and applications to biomedical imaging,” J. Phys. Photonics 3(3), 034005 (2021). 10.1088/2515-7647/abfc59 [DOI] [Google Scholar]
  • 64.Anand A., Chhaniwal V., Javidi B., “Tutorial: common path self-referencing digital holographic microscopy,” APL Photonics 3(7), 071101 (2018). 10.1063/1.5027081 [DOI] [Google Scholar]
  • 65.Picazo-Bueno J. Á., et al. , “Superresolved spatially multiplexed interferometric microscopy,” Opt. Lett. 42(5), 927 (2017). 10.1364/OL.42.000927 [DOI] [PubMed] [Google Scholar]
  • 66.Picazo-Bueno J. Á., et al. , “Hilbert–Huang single-shot spatially multiplexed interferometric microscopy,” Opt. Lett. 43(5), 1007 (2018). 10.1364/OL.43.001007 [DOI] [PubMed] [Google Scholar]
  • 67.Picazo-Bueno J. A., Micó V., “Opposed-view spatially multiplexed interferometric microscopy,” J. Opt. 21(3), 35701 (2019). 10.1088/2040-8986/ab000e [DOI] [Google Scholar]
  • 68.Picazo-Bueno J. Á., et al. , “Single capture bright field and off-axis digital holographic microscopy,” Opt. Lett. 48(4), 876–879 (2023). 10.1364/OL.478674 [DOI] [PubMed] [Google Scholar]
  • 69.Goodman J. W., Introduction to Fourier Optics, Roberts; (2005). [Google Scholar]
  • 70.Picazo-Bueno J. A., et al. , “Design, calibration, and application of a robust, cost-effective, and high-resolution lensless holographic microscope,” Sensors 22(2), 553 (2022). 10.3390/s22020553 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Shen F., Wang A., “Fast-Fourier-transform based numerical integration method for the Rayleigh-Sommerfeld diffraction formula,” Appl. Opt. 45(6), 1102–1110 (2006). 10.1364/AO.45.001102 [DOI] [PubMed] [Google Scholar]
  • 72.Polyanskiy M. N., “Refractiveindex.info database of optical constants,” Sci. Data 11, 94 (2024). 10.1038/s41597-023-02898-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Born M., Wolf E., “Principles of optics 7th edition,” in principles of optics electromagnetic theory of propagation interference and diffraction of light, Farley-Born M., Wolf E., Eds., Elsevier; (1999). [Google Scholar]
  • 74.Elsässer H. P., et al. , “Establishment and characterisation of two cell lines with different grade of differentiation derived from one primary human pancreatic adenocarcinoma,” Virchows Arch. B 61(1), 295–306 (1992). 10.1007/BF02890431 [DOI] [PubMed] [Google Scholar]
  • 75.Kastl L., et al. , “Quantitative phase imaging for cell culture quality control,” Cytom. Part A 91(5), 470–481 (2017). 10.1002/cyto.a.23082 [DOI] [PubMed] [Google Scholar]
  • 76.Min J., et al. , “Quantitative phase imaging of cells in a flow cytometry arrangement utilizing Michelson interferometer-based off-axis digital holographic microscopy,” J. Biophotonics 12(9), e201900085 (2019). 10.1002/jbio.201900085 [DOI] [PubMed] [Google Scholar]
  • 77.Zanini L., et al. , “Opto-microfluidic integration of the Bradford protein assay in lithium niobate lab-on-a-chip,” Sensors 22(3), 1144 (2022). 10.3390/s22031144 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.Zamboni R., et al. , “Opto-microfluidic coupling between optical waveguides and tilted microchannels in lithium niobate,” Opt. Express 31(17), 28423–28436 (2023). 10.1364/OE.495406 [DOI] [PubMed] [Google Scholar]
  • 79.Schindelin J., et al. , “Fiji: an open-source platform for biological-image analysis,” Nat. Methods 9(7), 676–682 (2012). 10.1038/nmeth.2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Barer R., “Interference microscopy and mass determination,” Nature 169(4296), 366–367 (1952). 10.1038/169366b0 [DOI] [PubMed] [Google Scholar]
  • 81.Kim T., et al. , “White-light diffraction tomography of unlabelled live cells,” Nat. Photonics 8(3), 256–263 (2014). 10.1038/nphoton.2013.350 [DOI] [Google Scholar]
  • 82.Bhaduri B., et al. , “Diffraction phase microscopy: principles and applications in materials and life sciences,” Adv. Opt. Photonics 6(1), 57–119 (2014). 10.1364/AOP.6.000057 [DOI] [Google Scholar]
  • 83.Zhang J., et al. , “A review of common-path off-axis digital holography: towards high stable optical instrument manufacturing,” Light Adv. Manuf. 2(3), 333–349 (2021). 10.37188/lam.2021.023 [DOI] [Google Scholar]
  • 84.Ferraro P., et al. , “Compensation of the inherent wave front curvature in digital holographic coherent microscopy for quantitative phase-contrast imaging,” Appl. Opt. 42(11), 1938 (2003). 10.1364/AO.42.001938 [DOI] [PubMed] [Google Scholar]
  • 85.Bishara W., et al. , “Lensfree on-chip microscopy over a wide field-of-view using pixel super-resolution,” Opt. Express 18(11), 11181 (2010). 10.1364/OE.18.011181 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Mues S., et al. , “Micro patterned surfaces: an effective tool for long term digital holographic microscopy cell imaging,” Proc. SPIE 10074, 1007410 (2017). 10.1117/12.2253280 [DOI] [Google Scholar]
  • 87.Colomb T., et al. , “Total aberrations compensation in digital holographic microscopy with a reference conjugated hologram,” Opt. Express 14(10), 4300 (2006). 10.1364/OE.14.004300 [DOI] [PubMed] [Google Scholar]
  • 88.Colomb T., et al. , “Automatic procedure for aberration compensation in digital holographic microscopy and applications to specimen shape compensation,” Appl. Opt. 45(5), 851–863 (2006). 10.1364/AO.45.000851 [DOI] [PubMed] [Google Scholar]
  • 89.Colomb T., et al. , “Numerical parametric lens for shifting, magnification, and complete aberration compensation in digital holographic microscopy,” J. Opt. Soc. Am. A 23(12), 3177 (2006). 10.1364/JOSAA.23.003177 [DOI] [PubMed] [Google Scholar]
  • 90.Zuo C., et al. , “Phase aberration compensation in digital holographic microscopy based on principal component analysis,” Opt. Lett. 38(10), 1724–1726 (2013). 10.1364/OL.38.001724 [DOI] [PubMed] [Google Scholar]
  • 91.Ma S., et al. , “Phase-aberration compensation via deep learning in digital holographic microscopy,” Meas. Sci. Technol. 32(10), 105203 (2021). 10.1088/1361-6501/ac0216 [DOI] [Google Scholar]
  • 92.Di J., et al. , “Phase aberration compensation of digital holographic microscopy based on least squares surface fitting,” Opt. Commun. 282(19), 3873–3877 (2009). 10.1016/j.optcom.2009.06.049 [DOI] [Google Scholar]
  • 93.Liu S., et al. , “Automatic and robust phase aberration compensation for digital holographic microscopy based on minimizing total standard deviation,” Opt. Lasers Eng. 134, 106276 (2020). 10.1016/j.optlaseng.2020.106276 [DOI] [Google Scholar]
  • 94.Turko N. A., Shaked N. T., “Erythrocyte volumetric measurements in imaging flow cytometry using simultaneous three-wavelength digital holographic microscopy,” Biomed. Opt. Express 11(11), 6649–6658 (2020). 10.1364/BOE.404368 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

JBO_029_S22715_SD001.pdf (395.2KB, pdf)
Download video file (7.7MB, mov)

Data Availability Statement

Data underlying the results presented in this article are available by a reasonable request to the corresponding authors.


Articles from Journal of Biomedical Optics are provided here courtesy of Society of Photo-Optical Instrumentation Engineers

RESOURCES