Validation of a simplified HPV genotyping assay designed for cervical screening in low-resource settings

Kanan T Desai 1,2, Kayode O Ajenifuja 3, Clement A Adepiti 3, Federica Inturrisi 1,, Casey Dagnall 1,4, Amanda C Hoffman 1,4, Didem Egemen 1, Julia C Gage 1, Nicolas Wentzensen 1, Silvia de Sanjose 1,5, Mark Schiffman 1
Editor: Elitza S Theel6
PMCID: PMC11837489  PMID: 39745462

ABSTRACT

Human papillomavirus (HPV) genotype predicts cervical cancer risk, and genotyping could help guide the management of HPV positives as part of cervical screening. An isothermal amplification HPV extended genotyping test (ScreenFire HPV RS assay) can assay up to 96 samples/controls in 1 hour plus preparation time. A novel format with pre-aliquoted reagents and an anti-contamination component (Zebra BioDome) could simplify the HPV testing process and reduce the chances of post-amplification contamination. We validated the Zebra BioDome formulation prior to its clinical use. Residual provider-collected cervical samples (n = 450) from a population-based study in rural Nigeria were retested with ScreenFire, once using the standard assay version (liquid reagents combined onsite) and twice with Zebra BioDome. HPV results with adequate DNA (N = 427) were analyzed channel-by-channel and using the cervical cancer risk-based hierarchy of HPV type channels (HPV16, else 18/45, else 31/33/35/52/58, else 39/51/56/59/68, else high-risk HPV negative) to evaluate Zebra BioDome repeatability and accuracy against the standard version. Zebra BioDome reduced the number of pipetting steps to run the ScreenFire HPV assay. Following amplification, the BioDome material formed a sealant layer above the reaction components. Zebra BioDome had excellent repeatability and agreement with the standard version, both at the channel-specific analysis (positive percent agreement between 88.4% [HPV39/51/56/59/68] and 100% [HPV16]; negative percent agreement between 97.8% [HPV31/33/35/52/58] and 100% [HPV39/51/56/59/68]) and hierarchical analysis (overall agreement 97.2%). The assay version utilizing Zebra BioDome performed similarly to the previously validated standard version of the ScreenFire HPV assay and is now undergoing field evaluation. This solution has the potential to reduce assay preparation time and risk of contamination, providing a simpler, low-cost, near-point-of-care HPV testing and extended genotyping solution for cervical screening in lower-resource settings. The potential application of Zebra BioDome technology to other PCR assays should be considered.

IMPORTANCE

This work validates a novel pre-packed formulation for the ScreenFire human papillomavirus (HPV) assay, which has the potential to simplify the HPV testing process and to reduce the chances of post-amplification contamination, providing a simpler, low-cost, near-point-of-care HPV testing, and extended genotyping solution for cervical screening in resource-limited settings as part of the ultimate public health goal to accelerate cervical cancer prevention. This technology can also have broad applications for other DNA amplification assays beyond HPV.

KEYWORDS: cervical cancer screening, HPV testing, HPV genotyping, point-of-care, LMICs

INTRODUCTION

Cervical cancer rates and deaths tend to be highest in regions with limited resources (1). The World Health Organization call to eliminate cervical cancer globally promotes human papillomavirus (HPV) vaccination, HPV-based screening during mid-adulthood, and treatment of both precancerous and cancerous conditions (2). Highly sensitive HPV testing of self-collected samples is being promoted worldwide (3), because of the crucial role of HPV in the development of cervical cancer (4), and the pressing need to screen many millions of women. It is impractical to treat every infected woman in high-prevalence areas because HPV infections are extremely common and often benign. Most HPV infections are controlled by the immune system within 1 to 2 years of detection, and only persistently detectable infections lead to precancer and cancer (4). Because HPV detection on population screening can be very common, and follow-up to distinguish persistence is often impractical, a triage test to identify biomarkers of persistent/progressive infections is necessary for prioritization of care by need. HPV typing is a promising means of triage because the likelihood of progression depends on the HPV type and varies significantly among the 13 high-risk HPV types (5, 6) classified by the International Agency for Research on Cancer as Group 1 (known carcinogens) or Group 2A (probable carcinogens) (7). Risk-based extended HPV typing is proposed to stratify HPV-positive women according to their risk of developing cervical cancer and to guide clinical management (6).

An isothermal DNA amplification technology-based ScreenFire HPV risk stratification (RS) assay (Atila Biosystems Inc., Sunnyvale, CA, USA) can provide extended genotyping for 13 high-risk HPV types in the four established risk channels: 16, 18/45, 31/33/35/52/58, and 39/51/56/59/68 (8, 9). It was designed to support clinical management according to the risk of cervical cancer associated with the grouped types, hence the emphasis on risk stratification, in which the highest-risk positive channel is reported when multiple concurrent infections are present. It is relatively rapid and low-cost, making it suitable for extended HPV genotyping beyond high-resource settings. Its clinical accuracy has already been shown (811) and the assay is currently undergoing field validation as part of the HPV-Automated Visual Evaluation (PAVE) protocol in nine low- and middle-income countries (LMICs) worldwide (12). Two aspects of performing the ScreenFire HPV RS assay emerged during the validation to date and are addressed here: the need to further decrease the time and manual handling needed to run the assay, particularly during the assay preparation phase, and the risk of laboratory contamination. Despite being a simple assay, the ScreenFire HPV RS test in its current format still requires that a trained laboratory technician carry out a number of preparatory steps: thawing reagents, preparing the master mix (by combining reaction mix and primer mix), dispensing the master mix into reaction tubes, and then adding the sample (Fig. 1A). These steps involve pipetting small volumes of reagents and samples, demanding careful attention to detail and increasing the risk of human error, cross-contamination, and variability in results. Moreover, these steps add to the overall time required for HPV testing, making single-day screening strategies more difficult.

Fig 1.

Composite figure with three panels describing sample processing, reaction assembly, and HPV DNA amplification for two protocols to run the ScreenFire HPV RS assay using the standard and the pre-packed Zebra BioDome assay versions.

Schematics of the steps to run the ScreenFire HPV RS assay. (A) Assay in its standard liquid version. (B) Assay in the novel Zebra BioDome version. (C) Details of the Zebra BioDome version before (i) and after (ii) amplification. The blue boxes on (A) and (B) highlight the steps impacted by the introduction of Zebra BioDome.

Besides, like any other DNA amplification technique (13), ScreenFire HPV RS assay is susceptible to contamination of the laboratory environment by amplified products. The problem arises because the post-amplification concentration of amplicons in a positive sample/control can reach billions of copies per microliter. If the tube content is not strictly contained after amplification, the high concentration of amplicons can contaminate laboratory surfaces, equipment, pipettors, and personal protective gear like gloves and laboratory coats. Identifying and cleaning up this contamination is time-consuming and resource-intensive, both in terms of labor and money, and it precludes HPV testing activities until full decontamination has been verified.

To reduce the number of pipetting steps and mitigate the risk of contamination, the assay’s manufacturer Atila BioSystems developed an assay format with pre-aliquoted reagents and an anti-contamination component, known as “Zebra BioDome”. This format removes the need for preparation and dispensing of the master mix, only requiring the addition of the lysate to the Zebra BioDome tubes (Fig. 1B) as the two components of the master mix are suspended between gel matrix layers. The gel matrix melts during amplification enabling the mixing of sample and reagents and solidifies again at the top of the reaction volume during cooling, thus creating a physical barrier for the amplified products (Fig. 1C).

This article aims to independently validate the Zebra BioDome version of the ScreenFire HPV RS assay by comparing it with the established standard liquid version of the assay and assessing its repeatability.

MATERIALS AND METHODS

Study population and design

The present study drew from 1,339 residual cervical samples stored frozen in PreservCyt solution (Hologic, Marlborough, MA, USA), collected by providers from a population-based screening project in rural Nigeria (Project Itoju) involving women aged 16 to 88 years (mean [standard deviation] = 44.0 [15.8]). The methodology of Project Itoju has been detailed elsewhere (14). In brief, between 2009 and 2010, approximately 1,420 eligible women—who were not pregnant, were sexually active, had not undergone a hysterectomy, were at least 15 years old, and had been residing in their household for more than 3 months—were invited from homes in the village of Irun, Nigeria. Participants in the research effort provided written informed consent. During the screening, locally trained nurses conducted cervical examinations and collected samples using a broom-type device and an endocervical brush, which were then placed in PreservCyt for liquid-based cytology (LBC) and HPV testing.

One milliliter aliquots of the frozen residual cytology samples were tested in the United States (Albert Einstein Cancer Center) for HPV using an AmpliTaq Gold MY09‐MY11 polymerase chain reaction (PCR)‐based test that included additional type‐specific primers, and primers to amplify a cellular beta‐globin fragment as an internal control for amplification as previously described (15). PCR products were genotyped with dot‐blot hybridization using type‐specific probes for 13 high-risk and >20 other HPV types, using a semi-quantitative measure of signal strength (1–5 from lowest to highest) as a proxy of viral load (16). Of note, the PCR signal strength variable (1–5) is not strictly linear, in that a signal strength of 5 has no upper bound and is, on average, particularly strong. Samples were also tested for HPV16 and HPV18 using a research-use-only HPV DNA test, the Luminex assay.

Out of the 1,339 residual samples, valid (i.e., adequate DNA) HPV results at the AmpliTaq Gold MY09‐MY11 PCR were obtained for 1,305 samples. A small subset (n = 8) was unblinded and sent to Atila BioSystems for updating the AmpFire assay chemistry to the ScreenFire RS HPV assay with its emphasis on sensitive detection of the most high-risk HPV types (8). Of the remaining 1,297 masked samples with valid results, we selected all 356 samples that tested positive for any HPV types (including low-risk types) by either PCR or HPV16/18 Luminex. Additionally, we randomly chose 100 of 941 specimens that tested negative for all HPV types (including low-risk) by both PCR and Luminex. These 456 samples were previously used to redesign and validate the ScreenFire HPV RS assay in its current standard format (8). Out of the 456 samples, six were found to have insufficient residual material (i.e., no more volume), leaving 450 usable samples.

These same 450 samples were retested with ScreenFire HPV RS assay using both the standard (9) and Zebra BioDome versions of the assay. Testing using the standard version was conducted at the US National Cancer Institute’s (NCI) Cancer Genomics Research (CGR) laboratory with visiting Atila BioSystems laboratory scientists. Testing using the Zebra BioDome version was conducted at the Atila BioSystems Laboratory by Atila BioSystems scientists. The Atila scientists were blinded to prior testing results and had no role in the analysis or reporting of the experiments.

ScreenFire HPV testing

As per the manufacturer’s protocol, ScreenFire HPV RS assay can be used with dry swab samples, cervical cell suspension, FFPE slice samples, and purified DNA samples. For the work presented here, the protocol for cervical cell suspension was used, plus pre-processing steps to prepare the frozen residual specimens. Initially, 1 mL of cervical cell suspension in PreservCyt was transferred to a 1.5 mL tube and centrifuged at maximum speed for 15 minutes. The supernatant was discarded, and 100 µL of 1× lysis buffer was added to the pellet. The mixture was vortexed thoroughly to resuspend the cell pellets. The resuspended cell pellets were then transferred to 96-well PCR plates, incubated at 95°C for 15 minutes, allowed to cool to room temperature, and stored at −20°C for future use.

The plates were thawed at room temperature prior to testing with ScreenFire HPV RS assay. This was done twice on one Powergene machine, once using the standard liquid version and once using the Zebra BioDome version, and once on a second Powergene machine using the Zebra BioDome version.

For testing with the standard liquid version, 5 µL of the prepared specimen (lysate) was combined with 20 µL of freshly prepared master mix (containing reaction mix and primer mix) in a 96-well PCR plate using a pipette. For testing with the Zebra BioDome version, 10 µL of the lysate was added to each well using the technique described below. Positive and negative template controls were included in each run. The plates or tube strips were sealed or capped, gently vortexed (only for the standard version), and centrifuged to ensure that all the material settled at the bottom of the wells.

The plates or Zebra BioDome tube strips were then placed into the Powergene PCR machines, set to the isothermal program mode (60°C for 60 minutes). Fluorescence signals were detected using CY5 for the HPV16 channel, ROX for the HPV18/45 channel, CY5.5 for the HPV31/33/35/52/58 channel, FAM for the HPV39/51/56/59/68 channel, and HEX for the human beta-globin gene as the internal control channel. A sample was deemed positive for an HPV type (group) if a signal was detected in the corresponding HPV channel within 60 minutes, regardless of the HEX channel signal. If no signal was detected in any of the HPV channels within 60 minutes, a signal in the HEX channel was required for a certain sample to be considered a valid HPV negative. If no signal was detected in any of the channels including the HEX channel, a sample was considered invalid suggesting inadequate DNA in the reaction.

All ScreenFire HPV RS testing was done using the Powergene 9600 Plus Real-Time PCR System. Instruments had the same design, software, and other characteristics and were thus considered interchangeable.

Zebra BioDome technology

Zebra BioDome is a proprietary format for isothermal amplification assays by Atila BioSystems. It features layers of reaction mix and primer mix separated by the BioDome anti-contamination gel matrix, with an additional BioDome layer on top for stability and containment. This assay format comes as pre-packed reaction tube strips covered by an adhesive film to prevent evaporation (Fig. 1C-i).

The version of Zebra BioDome described here is penetrable, meaning that at room temperature, the gel matrix is semi-solid—solid enough to prevent loss of the reaction mix but soft enough to be pierced with a standard pipette tip. Users only need to remove the adhesive film and pipette the lysate into the reaction tube. A standard volumetric pipette with a disposable tip (avoiding very narrow tips to prevent gel matrix blockage) is used to insert the sample/control into the bottom of the Zebra BioDome tube. Once the pipette tip touches the bottom, the pipettor is slightly withdrawn to allow dispensing of the lysate within the primer mix layer. The pipette is then removed and visually inspected to ensure no clogging of residual liquid in the tip. If clogging occurs, the volume transfer of 10 µL lysate is repeated into the same Zebra BioDome tube.

The gel matrix melts during amplification at 60°C enabling the mixing of the reaction components and re-solidifies at the top of the reaction during cooling, thus creating a physical barrier for the amplified products (Fig. 1C-ii) and resulting in a sealed reaction chamber.

Statistical analysis

Valid ScreenFire HPV results were obtained for 427 out of 450 samples for both the standard and Zebra BioDome versions, which were included in the analysis. The remaining 23 were excluded from the analytical sample because of invalid results on either or both assay versions (7 invalids by Zebra BioDome, 7 invalids by standard, and 9 invalids by both).

Analyses were done to assess the repeatability of Zebra BioDome when testing on two different Powergene instruments and to compare the HPV results of Zebra BioDome to those of the standard assay version.

Two types of analyses were conducted: first, channel-by-channel in a non-hierarchical fashion, acknowledging that a sample could test positive for multiple HPV channels, then hierarchically using the risk-based grouping of HPV genotype channels intrinsic to ScreenFire’s design. In the hierarchical analysis, HPV results are categorized as HPV16 positive, otherwise positive for HPV18/45 (if HPV16 is not detected), otherwise positive for HPV31/33/35/52/58 (if HPV16/18/45 are not detected), otherwise positive for HPV39/51/56/59/68, and otherwise negative for the 13 high-risk types.

Positive and negative agreement statistics were calculated, with 95% confidence intervals (CI) estimated under normal approximation. Agreement statistics for the repeatability analysis were calculated based on either replicate positivity or negativity. Agreement statistics for the comparison of Zebra BioDome against the standard version were calculated using the standard assay version as the reference. Additionally, the extension of McNemar’s test for asymmetry was computed for the channel-specific analysis. The analyses were adjusted back to the original sample by weighting HPV negatives (i.e., 100 negatives were weighted by 9.41 for 941 negatives in the original sample using the formula 100 × 9.41 = 941) while retaining channel-specific estimates.

Discrepancies between the Zebra BioDome version and the standard version were investigated by examining PCR and Luminex results, as well as histopathology and LBC results when available. Data on valid histopathology and/or LBC were available for 409 samples and both missing for 18 samples.

Sensitivity analyses were carried out by restricting to HPV results of the study population of women within the age range of 25 to 49 years.

All analyses were carried out using Microsoft Excel, the Statistical Package for the Social Sciences (SPSS) (version 1.0.0.1089), and R Statistical Software (version 4.3.0).

RESULTS

Repeatability

There was excellent agreement between the two replicates of ScreenFire HPV testing with the Zebra BioDome version performed on Powergene instrument 1 vs instrument 2 across all channels in the non-hierarchical analysis (Table 1). There were no statistically significant differences between the two replicates. Notably, there were no discordants for the HPV16 and HPV18/45 channels, in agreement with the emphasis on the primer design of the ScreenFire HPV assay. Differences in the HPV31/33/35/52/58 and HPV39/51/56/59/68 channels did exist but were non-significant. Repeatability was excellent also in the hierarchical analysis (not shown).

TABLE 1.

Channel-specific repeatability of Zebra BioDome version of ScreenFire HPV RS assay, run on Powergene1 and Powergene2, amongst 426a samples with valid results

Zebra BioDome ScreenFire HPV results: Powergene instrument 1/instrument 2 +/+ (N) −/+ (N) +/− (N) −/− (N) Overall positive agreement (95% CI) Overall negative agreement (95% CI) McNemar P-value
HPV16 27 0 0 399 100% 100% NAb
HPV18/45 26 0 0 400 100% 100% NAb
HPV31/33/35/52/58 103 3c 7d 313 91.2%
(85.9%–96.4%)
96.9%
(95.0%–98.8%)
0.34
HPV39/51/56/59/68 35 4e 3f 384 83.3%
(72.1%–94.6%)
98.2%
(96.9%–99.5%)
1.00
Agreement on overall HPV (high-risk 13), irrespective of the channel 175 5 6 240 94.1%
(90.7%–97.5%)
95.6%g
(93.1%–98.2%)
1.00
a

Excluding invalids on either replicate.

b

McNemar test statistic cannot be derived because the disagreement is zero.

c

Three positives on PCR: one for HPV58 (signal strength = 2), one for HPV35 (signal strength = 2), one for HPV52 (signal strength = 2).

d

Three positives on PCR: one for HPV35 (signal strength = 3), one for HPV31 (signal strength = 1), one for HPV58 (signal strength = 2).

e

Two positives for HPV68 (signal strength = 4 and 5) on PCR.

f

One positive for HPV68 (signal strength = 3) and one positive for HPV51 (sign strength = 1) on PCR.

g

Overall negative agreement was 95.6% (93.1%–98.2%) without sampling weights (calculated as 240/[240 + 5 + 6] = 0.956 = 95.6%) and was estimated to be 99.0% (98.4%–99.6%) after applying sampling weights for negatives (calculated as [140 + 100 × 9.41]/[(140 + 100 × 9.41) + 5 + 6] = 0.990 = 99.0%).

Channel-specific agreement

The number of invalid results was similar when using Zebra BioDome and the standard assay version of the ScreenFire assay: seven invalids by Zebra BioDome, seven invalids by standard, and nine invalids by both. Among the 427 valid results by both, there was very good agreement between the Zebra BioDome version and the standard version across all channels in the non-hierarchical analyses (Table 2). There were no statistically significant differences between the two. Differences were minor for the HPV16 and HPV18/45 channels, with all three additional positives by Zebra BioDome (two in the HPV16 channel and one in the HPV18/45 channel) positive for HPV16/18 by either PCR or Luminex. Differences for the two lower-risk channels approached but did not reach significance, with Zebra BioDome showing a slight increase in positivity for the HPV31/33/35/52/58 channel and the standard version exhibiting a slight increase in positivity for the HPV39/51/56/59/68 channel. Additional positives detected by either version of the assay did not show specific cross-reactivity patterns with other high-risk or low-risk HPV types identified by PCR except for the possible chance finding that three of the seven additional positives by Zebra BioDome for HPV31/33/35/52/58 were positive for HPV53. A similar agreement in the channel-specific analysis was found when repeating the analysis only on the age range of 25 to 49 years and when using the results of the Zebra BioDome replicate on Powergene instrument 2 (not shown).

TABLE 2.

Channel-specific agreement (non-hierarchical) between Zebra BioDome and the standard version of ScreenFire HPV RS assay amongst 427 samples with valid results by both versions

ScreenFire HPV results: Zebra BioDomea/Standard +/+ (N) −/+ (N) +/− (N) −/− (N) Positive agreement with standard as reference (95% CI) Negative agreement with standard as reference (95% CI) McNemar P-value
HPV16 25 0 2b 400 100% 99.5% (98.8%–100%) 0.48
HPV18/45 25 1c 1c 400 96.2% (88.8%–100%) 99.8% (99.3%–100%) 1.00
HPV31/33/35/52/58 104 1d 7e 315 99.0% (97.2%–100%) 97.8% (96.2%–99.4%) 0.08
HPV39/51/56/59/68 38 5f 0 384 88.4% (78.8%–98.0%) 100% 0.07
Agreement on overall HPV (high-risk 13), irrespective of the channel 174 2 8 243 98.9% (97.3%–100%) 96.8%g (94.6%–99.0%) 0.11
a

Zebra BioDome run on Powergene instrument one was used for this comparison.

b

One of the two was positive for HPV16 on Luminex; the other was positive for HPV16 on PCR (signal strength = 2).

c

Additional positives on Zebra BioDome or standard version were positive for HPV18 on PCR (signal strength = 2–3).

d

One positive for HPV31 on PCR (signal strength = 2).

e

Four positives on PCR: one for HPV52 (signal strength = 1), one for HPV35 (signal strength = 2), two for HPV58 (signal strength = 2); three additional were positive for HPV53.

f

Three positives on PCR: two for HPV68 (signal strength = 4–5), one for HPV59 (signal strength = 3).

g

Overall negative agreement was 96.8% (94.6%–99.0%) without sampling weights (calculated as 243/[243 + 8] = 0.968 = 96.8%) and was estimated to be 99.3% (98.8%–99.8%) after applying sampling weights for negatives (calculated as [143 + 100 × 9.41]/[(143 + 100 × 9.41) + 8] = 0.993 = 99.3%).

Hierarchical agreement

There was excellent agreement between the Zebra BioDome version and the standard version of the ScreenFire assay also in the hierarchical analyses (Table 3), which is relevant for risk-based clinical management. Following the hierarchical distribution of HPV channels by the standard version, Zebra BioDome detected 25 out of 25 (100%) HPV16 positives; 24 out of 25 (96.0%) HPV18/45 positives in the absence of HPV16; 88 out of 90 (97.8%) HPV31/33/35/52/58 positives in the absence of HPV16, or 18/45; 35 out of 36 (97.2%) otherwise HPV39/51/56/59/68 positives. The overall agreement between the Zebra BioDome version vs the standard version was 97.2% (95% CI: 95.6%–98.8%) without sampling weights and 99.1% (95% CI: 98.6%–99.6%) with sampling weights (Table 3). A similar agreement in the hierarchical analysis was found when repeating the analysis using the results of the Zebra BioDome replicate performed on Powergene instrument 2 (not shown).

TABLE 3.

Hierarchical agreement between Zebra BioDome and the standard version of ScreenFire HPV RS assay according to HPV risk groups amongst 427 samples with valid results by both versionsc

ScreenFire HPV results: Zebra BioDome (Powergene instrument 1)/Standard Standard
HPV 16 else HPV 18/45 else HPV 31/33/35/52/58 else HPV 39/51/56/59/68 else High-risk 13 negative Total for Zebra BioDome
Zebra BioDome HPV16 25 0 1 0 1 27
Column% 100 0 1.1 0 0.4 6.3
else HPV18/45 0 24 0 0 1 25
Column% 0 96.0 0 0 0.4 5.9
else HPV31/33/35/52/58 0 1 88 0 6 95
Column% 0 4.0 97.8 0 2.4 22.2
else HPV39/51/56/59/68 0 0 0 35 0 35
Column% 0 0 0 97.2 0 8.2
else high-risk 13 negative 0 0 1 1 243 245
Column% 0 0 1.1 2.8 96.8a 57.4
Total for standard 25 25 90 36 251 427b
a

Overall negative agreement was 96.8% without sampling weights (calculated as 243/251 = 0.968 = 96.8%) and 99.3% after applying sampling weights for negatives (calculated as [143 + 100 × 9.41]/1,092 = 0.993 = 99.3%).

b

Overall agreement was 97.2% without sampling weights (calculated as [25 + 24 + 88 + 35 + 243]/427 = 0.972 = 97.2%) and 99.1% after applying sampling weights for negatives (calculated as [25 + 24 + 88 + 35 + (143 + 100 × 9.41)]/1,268 = 0.991 = 99.1%).

c

Highlighted in gray are concordant HPV risk group results.

DISCUSSION

Our results showed that the Zebra BioDome version of the ScreenFire HPV RS assay had excellent repeatability and agreement with the standard version of the assay, for each of the four type groups defined by the assay individually and hierarchically. Small differences in channel positivity and invalid results were seen but none were statistically significant.

The validation presented here aimed at comparing the novel Zebra BioDome version of ScreenFire HPV RS assay with the established standard version of the assay and assessing its repeatability on stored clinician-collected specimens under ideal laboratory conditions. A real-life trial of the clinical performance of Zebra BioDome post-shipping under realistic simpler laboratory conditions is beyond the scope of this work and is being addressed in ongoing field evaluation.

Following this positive evaluation, Zebra BioDome has been introduced at PAVE study sites (12) using ScreenFire HPV RS assay (Nigeria, Tanzania, Malawi, Eswatini, Dominican Republic, El Salvador, Honduras, Brazil, and Cambodia). This field evaluation will address the assay’s clinical accuracy and will evaluate its real-world use across a wide range of laboratory settings. Because the PAVE study uses dry vaginal self-samples for HPV testing (FLOQSwab, Copan, Italy), the field evaluation will also address the pre-analytics of ScreenFire HPV RS assay Zebra BioDome format on dry swabs, acknowledging the fact that sample processing, together with assay preparation, contribute to the overall time required for HPV testing.

The Zebra BioDome version of the ScreenFire HPV RS assay was designed to manage the reality of HPV testing in LMICs, which is the need for accurate, rapid tests that allow single-day screening strategies for cervical cancer. As compared to the standard version of the ScreenFire HPV assay (8, 9), Zebra BioDome seems to offer several advantages given its formulation with pre-aliquoted reagents and an anti-contamination component.

First, it eliminates the need for master mix preparation and aliquoting, hence reducing the number of manual pipetting steps. This translates into reduced hands-on time for assay preparation and reduced variability due to potentially unequal manually pipetted volumes (i.e., unequal volumes of primer mix and reaction mix combined during preparation of master mix, and/or unequal volumes of master mix dispensed in the reaction wells). The only manual pipetting step is the addition of 10 µL of lysate to the Zebra BioDome tube.

Second, the Zebra BioDome version reduces the risk of post-amplification contamination by sealing the products with the BioDome matrix material. Contamination of amplified PCR products in a molecular biology laboratory space is a critical concern, common to all PCR-based techniques (13). The assay’s high sensitivity means that even small amounts of DNA can be amplified, leading to a high number of copies post-amplification with a risk of leakage. Contamination is very damaging because it compromises the trustworthiness of results due to false positives leading to incorrect interpretations and because it potentially invalidates the entire laboratory due to the need to decontaminate and repeat experiments, wasting time, and resources. During the initial phases of PAVE fieldwork (12), several HPV laboratories across different settings experienced significant disruptions due to post-amplification contamination. For instance, at the Nigeria study site, the decontamination process took over two months to complete, halting routine laboratory activities and resulting in substantial resource demands, a growing backlog of samples, and delays in both HPV testing and the communication of results. This example highlights the risk mitigation opportunity using anti-contamination measures such as the novel Zebra BioDome to ensure the continuity and efficiency of laboratory operations. Separately from the work presented here, a few Zebra BioDome tubes were opened post-reaction, swabbed from the inside, and re-run with ScreenFire HPV assay. They were found to have minimal to no amplification, thus demonstrating effective reduction of contamination with amplified products. By reducing the risk of contamination, the BioDome matrix increases the reliability of HPV results for clinical management. Besides the anti-contamination role, this matrix also acts as a stabilizer of the Zebra BioDome formulation because it separates the layers of pre-aliquoted reagents (i.e., reaction mix and primer mix) avoiding their mixing and creates a barrier from the air in the tube.

The formulation of Zebra BioDome may present a versatile and cost-effective solution for a wide range of DNA amplification techniques besides HPV, particularly at the table-top level. The pre-aliquoted reagents and anti-contamination components are applicable to other DNA amplification assays such as endpoint PCR or qPCR tests, potentially providing a broader application in research and diagnostic settings. This innovation could prove instrumental in addressing generic problems of molecular biology testing, such as assay preparation time and risk of post-amplification contamination, and in advancing the accessibility of molecular testing in all settings where cost and contamination control are critical.

The additional advantages of Zebra BioDome do not compromise those that are characteristic of the ScreenFire HPV RS assay in its original formulation (8, 9). Most importantly, the risk-based hierarchy of HPV type channels because of the great difference in the risk of cervical cancer associated with HPV types (9). Hence, in case of positivity for multiple HPV channels, clinical management is better directed following the hierarchical reading of the type groups: HPV16 positive, otherwise positive for HPV18 or HPV45 (if HPV16 was not detected), otherwise positive for HPV31/33/35/52/58 (if HPV16/18/45 were not detected), otherwise positive for HPV39/51/56/59/68, and otherwise negative. Like the standard version, the Zebra BioDome version allows for the detection of HPV and human DNA control directly from clinical specimens without the need for DNA extraction and is compatible with dry self-collected samples without requiring collection media. It also allows for flexibility to run up to 96 samples/controls without product wastage (i.e., the pre-aliquoted strips of 8 tubes can be used in multiples of 8 or cut as needed). These optimizations do not translate into additional costs for reagents compared to the standard version (approx. $6 per test for scale-up). For transport and storage, the space requirement is approximately unchanged while the temperature should be kept cold (recommended 2°C–8°C). A note of caution goes for the transportation conditions because high temperatures may cause disruption of the layers and invalidate the use of the product due to premature mixing of reagents, hence the importance of maintaining a cold chain and implementing temperature checks for this penetrable version of Zebra BioDome. In the future, a re-formulation may be considered to allow greater stability at warm temperatures and to remove the need for a cold chain.

Conclusions

In conclusion, the novel Zebra BioDome version of the ScreenFire HPV assay evaluated here has demonstrated accurate risk-based genotype grouping while simplifying the HPV testing process and reducing the risk of laboratory contamination. If validated in real-world settings, the assay will be an example of a rapid HPV extended genotyping solution for resource-limited settings as part of the ultimate public health goal to accelerate cervical cancer prevention. The formulation may also have potential application for other DNA amplification assays.

ACKNOWLEDGMENTS

We acknowledge the participants and providers involved in Project Itoju.

Contributor Information

Federica Inturrisi, Email: f.inturrisi@outlook.com.

Elitza S. Theel, Mayo Clinic Minnesota, Rochester, Minnesota, USA

FUNDING

This study was funded by the NCI Cancer Moonshot and NIH Intramural Research Program. CGR was provided with equipment and reagents by Atila Biosystems for the sole purpose of this validation study. The authors declare no intellectual property or conflict of interest with regard to the manufacturer, whose scientists had no input in the design, analysis, interpretation, or drafting of this article.

DATA AVAILABILITY

The data that support the findings of our study are available from the corresponding author upon reasonable request.

ETHICS APPROVAL

The specimens used in this study were residual samples of Project Itoju, a population-based study led by the US National Cancer Institute (NCI) in Nigeria. Both Nigerian and NCI institutional review boards (IRBs) (NCT 00804466) approved the original study protocol that included consent for specimen storage for future research.

REFERENCES

  • 1. International agency for research on cancer. World Health Organization . 2021. GCO Cancer Today. Available from: https://gco.iarc.fr/today/
  • 2. World Health Organization . 2020. Global strategy to accelerate the elimination of cervical cancer as a public health problem. Available from: https://www.who.int/publications/i/item/9789240014107
  • 3. World Health Organization . 2021. WHO Guideline for Screening and Treatment of Cervical Pre-Cancer Lesions for Cervical Cancer Prevention, Second Edition. Geneva. World Health Organization. Available from: https://www.who.int/publications/i/item/9789240030824 [PubMed] [Google Scholar]
  • 4. Schiffman M, Castle PE, Jeronimo J, Rodriguez AC, Wacholder S. 2007. Human papillomavirus and cervical cancer. Lancet 370:890–907. doi: 10.1016/S0140-6736(07)61416-0 [DOI] [PubMed] [Google Scholar]
  • 5. de Sanjosé S, Serrano B, Tous S, Alejo M, Lloveras B, Quirós B, Clavero O, Vidal A, Ferrándiz-Pulido C, Pavón MÁ, Holzinger D, Halec G, Tommasino M, Quint W, Pawlita M, Muñoz N, Bosch FX, Alemany L, RIS HPV TT, VVAP and Head and Neck study groups . 2018. Burden of human papillomavirus (HPV)-related cancers attributable to HPVs 6/11/16/18/31/33/45/52 and 58. JNCI Cancer Spectr 2:pky045. doi: 10.1093/jncics/pky045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Demarco M, Hyun N, Carter-Pokras O, Raine-Bennett TR, Cheung L, Chen X, Hammer A, Campos N, Kinney W, Gage JC, Befano B, Perkins RB, He X, Dallal C, Chen J, Poitras N, Mayrand M-H, Coutlee F, Burk RD, Lorey T, Castle PE, Wentzensen N, Schiffman M. 2020. A study of type-specific HPV natural history and implications for contemporary cervical cancer screening programs. EClinMed 22:100293. doi: 10.1016/j.eclinm.2020.100293 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. International Agency for Research on Cancer . 2012. IARC Monographs on the Evaluation of Carcinogenic Risks to Humans, No.100B. Lyon, France. World Health Organization. Available from: https://www.ncbi.nlm.nih.gov/books/NBK304348/ [Google Scholar]
  • 8. Desai KT, Adepiti CA, Schiffman M, Egemen D, Gage JC, Wentzensen N, de Sanjose S, Burk RD, Ajenifuja KO. 2022. Redesign of a rapid, low-cost HPV typing assay to support risk-based cervical screening and management. Int J Cancer 151:1142–1149. doi: 10.1002/ijc.34151 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Inturrisi F, de Sanjosé S, Desai KT, Dagnall C, Egemen D, Befano B, Rodriguez AC, Jeronimo JA, Zuna RE, Hoffman A, Farhat Nozzari S, Walker JL, Perkins RB, Wentzensen N, Palefsky JM, Schiffman M. 2024. A rapid HPV typing assay to support global cervical cancer screening and risk-based management: a cross-sectional study. Int J Cancer 154:241–250. doi: 10.1002/ijc.34698 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Hou J, Belinson JL, Du H, Li C, Zhang W, Zhang L, Zhang Y, Qu X, Wu R. 2024. AmpFire HPV and ScreenFire RS HPV validation trial. Am J Clin Pathol 161:535–542. doi: 10.1093/ajcp/aqad181 [DOI] [PubMed] [Google Scholar]
  • 11. Mungo C, Guliam A, Chinula L, Inturrisi F, Msowoya L, Mkochi T, Jawadu S, de Sanjosé S, Schiffman M, Tang JH, Smith JS. 2024. Comparison of the ScreenFire and Xpert HPV assays for the detection of human papillomavirus and cervical precancer among women living with HIV in Malawi. Infect Agent Cancer 19:24. doi: 10.1186/s13027-024-00585-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. de Sanjosé S, Perkins RB, Campos N, Inturrisi F, Egemen D, Befano B, Rodriguez AC, Jerónimo J, Cheung LC, Desai K, Han P, Novetsky AP, Ukwuani A, Marcus J, Ahmed SR, Wentzensen N, Kalpathy-Cramer J, Schiffman M, PAVE Study Group . 2024. Design of the HPV-automated visual evaluation (PAVE) study: validating a novel cervical screening strategy. Elife 12:RP91469. doi: 10.7554/eLife.91469 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Borst A, Box ATA, Fluit AC. 2004. False-positive results and contamination in nucleic acid amplification assays: suggestions for a prevent and destroy strategy. Eur J Clin Microbiol Infect Dis 23:289–299. doi: 10.1007/s10096-004-1100-1 [DOI] [PubMed] [Google Scholar]
  • 14. Gage JC, Ajenifuja KO, Wentzensen NA, Adepiti AC, Eklund C, Reilly M, Hutchinson M, Wacholder S, Harford J, Soliman AS, Burk RD, Schiffman M. 2012. The age-specific prevalence of human papillomavirus and risk of cytologic abnormalities in rural Nigeria: implications for screen-and-treat strategies. Int J Cancer 130:2111–2117. doi: 10.1002/ijc.26211 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Castle PE, Schiffman M, Gravitt PE, Kendall H, Fishman S, Dong H, Hildesheim A, Herrero R, Bratti MC, Sherman ME, Lorincz A, Schussler JE, Burk RD. 2002. Comparisons of HPV DNA detection by MY09/11 PCR methods. J Med Virol 68:417–423. doi: 10.1002/jmv.10220 [DOI] [PubMed] [Google Scholar]
  • 16. Gravitt PE, Kovacic MB, Herrero R, Schiffman M, Bratti C, Hildesheim A, Morales J, Alfaro M, Sherman ME, Wacholder S, Rodriguez A-C, Burk RD. 2007. High load for most high risk human papillomavirus genotypes is associated with prevalent cervical cancer precursors but only HPV16 load predicts the development of incident disease. Int J Cancer 121:2787–2793. doi: 10.1002/ijc.23012 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of our study are available from the corresponding author upon reasonable request.


Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES