Abstract
A very sensitive micronephelometer has been used to measure the light-scattering of dilute mixtures of antigen solutions and precipitating antibodies. At equivalent ratios of antigen and antibody, the degree of light scattering is proportional to the protein content of the complex thus allowing estimates of antibody protein to be made. In the region of antibody excess, light scattering is proportional to the concentration of antigen thus allowing a calibration curve to be drawn which may be used in the estimation of antigen. The technique is quick and simple and the results are not markedly affected by moderate changes in pH, salt concentration, temperature or period of incubation.
Full text
PDF








Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Bukantz S. C., Cooper A., Bullowa J. G. The Photoelectric Titration of Pneumococcal Capsular Polysaccharide. J Bacteriol. 1941 Jul;42(1):15–27. doi: 10.1128/jb.42.1.15-27.1941. [DOI] [PMC free article] [PubMed] [Google Scholar]
- LOWRY O. H., ROSEBROUGH N. J., FARR A. L., RANDALL R. J. Protein measurement with the Folin phenol reagent. J Biol Chem. 1951 Nov;193(1):265–275. [PubMed] [Google Scholar]
- Vincent W. F., Harris E. W., Yaverbaum S. The estimation of precipitating antibody using a turbidimetric technique. Immunology. 1970 Feb;18(2):143–147. [PMC free article] [PubMed] [Google Scholar]