Abstract
The report describes the rehabilitation of a maxillary arch with limited bone volume in a 67‐year‐old female taking antiresorptives due to osteopenia. One 10‐mm and two extra‐short 4‐mm implants were inserted, and implant‐supported splinted crowns were fabricated. The 5‐year follow‐up showed stable bone levels, despite poor initial stability (ISQ: 14–51).
Keywords: alveolar bone loss, dental implants, maxilla, osteopenia, short dental implants, survival rate
1. INTRODUCTION
Low bone density and deficient bone dimensions present a challenge for implant insertion in the posterior maxilla. In cases where the volume of residual bone is reduced to less than 8 mm due to sinus pneumatisation and/or extensive postextraction bone resorption, several surgical options may be considered during the formulation of a treatment plan for implant‐supported prosthetic rehabilitation: a transcrestal (closed) sinus lift combined with the use of standard‐length implants; a simultaneous or delayed lateral (open) sinus lift; or the use of short (<8 mm) and ultrashort (<6 mm) implants. 1
Short and extra‐short implants have slowly become recognized by patients and clinicians in the past years. They require less aggressive surgical techniques and result in shorter intraoperative periods, reduced morbidity and lower treatment prices. 2 The predictability of mid‐term outcomes in single‐tooth and splinted short‐implant‐supported rehabilitations can be attributed to recent technological improvements and adjusted treatment protocols. Besides the lower incidence of comorbidities and complications in comparison to sinus lift procedures, 3 the rationale for using short implants is based on the observation that the maximum stress concentration occurs on the first 5–6 mm of the implant, irrespective of its length. 4 After loading, short implants may experience more failures than standard‐length implants because of their biomechanical shortcomings, particularly in the maxilla. 5
In general, osteoporosis or low‐dose oral antiresorptive drug (ARD) intake for its treatment does not compromise implant therapy; that is, patients taking oral ARDs do not lose more implants, experience inferior osseointegration or get more implant‐related complications/failures compared to patients without a history of such medication. 6 , 7 , 8 Nevertheless, if implant loss occurs, it is more likely to occur within a short time after loading (early loss 9 ) and in the posterior maxilla. 6 , 8 In addition, posterior dental implants are difficult to access and clean, making the development of peri‐implantitis more likely. 10 No information, however, exists regarding peri‐implant marginal bone levels in patients taking ARDs. 6 To the best of our knowledge, success and failure rates of short implants in patients with osteoporosis or patients taking ARDs have not yet been evaluated.
In our recent case series study, 11 , 12 we tested the possibility of rehabilitating posterior edentulous maxillary areas with 4‐mm extra‐short implants splinted to 10‐mm implants. We have already presented the treatment protocol 11 as well as the 1‐ 12 and 3‐year 13 results. The survival and success rates were found to be 100% in all 11 cases. The present case report describes the successful rehabilitation and 5‐year follow‐up period of a patient with poor maxillary bone quality, extremely low primary implant stability, osteopenia and a 9‐year history of preventive ARD therapy.
2. MATERIALS AND METHODS
This paper describes a case of a 67‐year‐old woman with a medical record of primary osteopenia. Her medical record revealed primary osteopenia diagnosed by dual x‐ray absorptiometry (DXA) with a T‐score of −1.9. 14 She had been undergoing preventive ARD treatment for the past 9 years: the first 7 years with alendronate, 70 mg p.o./week (Fosavance®, Merck Sharp & Dohme BV) and the last 2 years with denosumab 30 mg s.c./6 months (Prolia®, Amgen). She was also taking vitamin D and calcium supplements. Implant insertion was performed 5 months after the last denosumab injection and 6 months after extraction of teeth #2 and #5.
The clinical evaluation revealed a bilaterally shortened maxillary dental arch. Her natural dentition was preserved from tooth #6 to tooth #12. The vertical alveolar ridge dimension seemed satisfactory, yet the CBCT scan revealed an expanded maxillary sinus. The reduced vertical bone dimension measured 5.06 and 5.98 mm at the edentulous sites of former teeth no. #3 and #4, respectively. The vertical bone dimension at the edentulous site of former tooth #5 measured 12.35 mm (Figure 1). The alveolar ridge width in the right edentulous maxillary region measured 7 mm or more. In the mandible, the dental formula was from tooth #19 to tooth #31. Tooth #20 had been replaced with a titanium (Ti) implant 12 years ago and tooth #30 had been restored with a porcelain‐fused‐to‐metal crown. The patient's oral hygiene was good (full mouth plaque score: 18%), gingival inflammation was negligible and periodontitis‐associated bone loss was minimal. There was no sign of mucosal pathology.
Approval for the planned treatment as part of the case series study was obtained from the Republic of Slovenia's National Ethics Committee (No. 30/10/2015). Written informed consent was obtained from the patient prior to treatment and paper publication. The treatment plan was to provide her with 2 extra‐short 4‐mm implants in the edentulous sites of former teeth #3 and #4 and one 10‐mm implant at the edentulous site of former tooth #5. Restoration with splinted crowns supported by these three implants was planned to occur 6 months after implant insertion. An acrylic surgical guide for a pilot drill was fabricated to obtain optimal position and inclination of the implants. After the application of local anesthesia (Ultracain®, Hoechst), a midcrestal incision was made and a full‐thickness flap reflected. The standard implant bed preparation sequence was followed as recommended by the manufacturer (Institut Straumann AG). The profile drill was omitted for the 4‐mm implant beds. Finally, one 10‐mm × 4.1‐mm diameter and two 4‐mm × 4.1‐mm diameter (0.8 mm thread pitch and a 1.8 mm polished collar) tissue‐level (Standard Plus, titanium‐zirconium [Ti‐Zr] alloy, SLActive®) implants (Institut Straumann AG) were manually inserted. Implant stability was measured immediately after insertion using a hand ranchette and Ostell® equipment (Integration Diagnostics). Finally, closure screws (height: 2 mm) were inserted onto the implants and flaps were approximated and sutured using 5–0 non‐resorbable polypropylene sutures (Prolen®). In the same session, one 10‐mm × 4.1 mm bone‐level (BL) implant was inserted at the edentulous site of former tooth #13. A soft diet and chlorhexidine rinse 2×/day was recommended for 2 weeks. Diclofenac 75 mg p.o./8 h (Naklofen duo®, Krka) was prescribed for 2 days. Secondary herpetic lesions appeared in the operative region 4 days after implant insertion; valacyclovir 500 mg p.o./12 h (Valtrex®, GSK) was prescribed for 14 days. The sutures were removed 2 weeks after implant insertion and the implant position checked by radiograph.
After 6 months, the implants were deemed ready for prosthetic rehabilitation. Following an X‐ray of the three implants, their stability was tested using resonance frequency analysis (Osstell®, Integration Diagnostics). Then, an implant‐level definitive open tray impression was made with polyether impression material (Impregum®, 3M ESPE). The metal framework was designed and manufactured by a computer‐aided software using Coron®, a cobalt‐chromium alloy (Institut Straumann AG). Finally, the patient was restored with a metal‐ceramic fixed dental prosthesis cemented onto individually tailored original abutments, which were screwed to the three implants with a torque of 35 Ncm.
The patient was thereafter recalled every 4–6 months. At each visit, we reinforced her oral hygiene instructions, and meticulously cleaned any tooth deposits. Every 12 months, a detailed periodontal examination (evaluation of full mouth of plaque score, gingival bleeding index, pocket probing depth, recession, bleeding on probing) was performed in addition to a radiographic examination (without bite blocks or standardization methods) of the implants, whereby radiographic bone levels were accurately measured. The clinical follow‐up is presented in Figure 2.
3. RESULTS
Based on its radiographic appearance, bone quality was estimated as type III at the edentulous sites of former teeth #4 and #5 and type IV at the edentulous site of former tooth #3. Insertion torques of all implants were below 15 Ncm, and initial ISQ values measured 14, 49, and 52 for implants at edentulous sites #3, #4, and #5, respectively.
Four days following implant insertion, the patient complained of pain at the right maxillary surgical site. A detailed clinical examination revealed ruptured vesicles and fibrine exudate, characteristic for secondary herpetic mucosal lesions. Fourteen days later, the lesions were successfully cured, and sutures removed. After 6 months, the ISQ values for implants at edentulous sites #3, #4, and #5 increased to 56, 51, and 60, respectively. Slight marginal bone resorption was noticed, typical for tissue level implants.
The annual follow‐up clinical and radiographic examinations showed a stable condition of the examined implants and other parts of the dentition even 5 years after implant insertion (Figures 3 and 4). However, the 2.5‐year clinical examination revealed bleeding on probing at the buccal aspects of the examined implants and increased probing pocket depths around teeth #7, #8, #9, and #10. An additional nonsurgical intervention was needed to overcome this issue, comprised of supra‐ and subgingival plaque removal at teeth #7 –#10 using piezoelectric ultrasonic instruments (PiezoLED ultrasonic scaler with Piezo Scaler tip 203 [KaVo dental]). Due to COVID‐19, one scheduled maintenance session was skipped in 2020 and one in 2021.
In the last year, the patient was diagnosed with two additional medical issues: hypertension and hypothyroidism. As a result, she started taking perindopril 5 mg/day (Bioprexanil, Servier) and levothyroxine 25 mcg/day (Eutirox, Merck). Five years after implant insertion, her bone density (measured with DXA) decreased to a T‐score of −2.1, the diagnosis remaining osteopenia.
4. DISCUSSION
Several recent studies have investigated the possibility of rehabilitating shortened maxillary arches with extra‐short implants, finding them to exhibit lower costs, shorter recovery times, minimal invasiveness, and promising med‐term results in cases of single crowns 15 , 16 , 17 as well as splinted crowns in partially edentulous maxillae with gaps between the premolar and distal molar. 18 , 19 Similarly, our recent case series yielded excellent 3‐year survival and success rates of splinted crowns supported by one or two extra‐short 4‐mm dental implants and one standard 10‐mm dental implant in shortened maxillary arches, despite the fact that more than half of the treated patients presented with concomitant systemic illnesses, which could potentially compromise implant survival. 13 The present case report with a 5‐year follow‐up period describes a patient whose bone regeneration capacity was disrupted due to osteopenia and ARD therapy. In addition, the patient refused to undergo any surgical procedure aimed at vertical bone augmentation due to her systemic health and wanted a fixed, minimally invasive option in the shortest possible time.
Despite poor bone quality, extremely low primary implant stability, and the reactivation of herpetic lesions in the early post‐operative period, the stability of the implants improved after 6 months. Although the stability of both 4‐mm implants measured below 60 ISQ–a threshold value that is considered to guarantee safe rehabilitation in splinted restorations, 20 the prosthetic rehabilitation remained favorable even after 5 years and provided the patient with excellent function. In addition, the 5‐year survival rate of the non‐bicortically stabilized extra‐short implant placed at edentulous site #4 was not compromised even though it has been found that bi‐cortical stabilization is key for short implant stability. Ng et al. showed that only 51% of 6‐mm single‐tooth maxillary implants without bi‐cortical stabilization survived a 5‐year follow‐up period in contrast to 100% of bi‐cortically stabilized 6‐mm single‐tooth implants. 21
However, short implants should be used cautiously since they are associated with a higher risk of failure compared to standard implants. Only patients with sufficient horizontal bone width, that is, those with a buccal and palatal plate measuring at least 1–1.5 mm, should be considered as candidates for prosthetic rehabilitation with extra‐short implants to avoid critical remodeling of marginal bone. 1 Splinting is also highly encouraged since it serves as a measure of protection against excessive masticatory forces; in addition, only edentulous regions with no or minor vertical bone resorption seem to be suitable because of the unfavorable crown‐to‐implant ratios at short implant sites. 12 In the present case study, the crown‐to‐implant ratio was more than 1:2 at the sites of the two extra‐short implants, yet no technical complications were found even at the 5‐year follow‐up visit, most probably because we were very careful when adjusting the occlusal contacts, ensuring disclusion during lateropulsion. Even though the number of case series showing favorable results of 4‐mm implants is on the rise, 15 , 16 , 17 , 18 , 19 no RCT has been performed to directly compare the effectiveness and success rates of sinus lifts versus extra‐short implants.
Our patient's medical history revealed that she had been consecutively treated with two ARDs prior to implant placement—first with alendronate, a second‐generation bisphosphonate, for 7 years, followed by denosumab, a human monoclonal antibody, for 2 years. It is known that bisphosphonates remain circulating in the body for up to 10 years even after their discontinuation, resulting in impaired angiogenesis and prolonged inhibition of osteoclast action. 22 This presented a risk factor for potential bone grafting aimed at augmenting the patient's low maxillary floor since such procedures rely on a functioning vascular recipient site, intact osteoclastic resorption and unimpaired osteoblastic bone formation. 23 Considering the recommendation that bone augmentations such as sinus lifts should be, if possible, avoided in such cases due to very limited data concerning their safety, success, and risk for osteonecrosis, 6 prosthetic rehabilitation with extra‐short implants was deemed a better alternative treatment option for our patient. Nevertheless, it has been suggested that dose and cumulative duration of use (>3–4 years) rather than the route of ARD administration is associated with a greater risk for medication‐related osteonecrosis of the jaws (MRONJ). 6 , 9 , 23 Accordingly, Woo et al. 24 found that approximately 94% of all patients with bisphosphonate‐related osteonecrosis of the jaws received bisphosphonates intravenously (prescribed for the treatment of cancer/metastases) and only 6% orally (prescribed for the treatment of osteoporosis). It is also known that MRONJ is more likely to appear in the mandible. 22 Therefore, patients with high‐dose ARD intake, patients on oral ARDs over a longer period of time, and patients with comorbidities should be considered as high‐risk candidates for MRONJ. 6 These patients should be placed on regular long‐term recall schedules. 25 The American Association Of Oral And Maxillofacial Surgeons, however, still has not reached a consensus regarding the need for a “drug holiday” prior to dentoalveolar surgery in patients receiving oral or intravenous ARD therapy—if, however, the choice were made in favor of treatment suspension, it is recommended that implant insertion be performed 4 months following the last dose of ARD administration. 25
In cases of osteopenia‐associated poor bone quality, titanium‐zirconium (Ti‐Zr) implants with hydrophilic surfaces may enhance osteointegration compared to pure Ti implants with hydrophobic surfaces. 26 Ti alloys containing 13%–15% Zr may aid osteointegration in osteoporotic bone, since they have been shown to require a higher removal torque in comparison to pure Ti implants in a rabbit osteoporosis model. 27 Four‐millimeter implants have narrower thread pitches (0.8 mm) than 10‐mm implants (1.25 mm), which increases their implant‐to‐bone contact ratios.
5. CONCLUSION
After 5 years of strict maintenance, two 4‐mm extra‐short tissue level Ti‐Zr implants, splinted to a standard‐size implant, enabled the successful rehabilitation of our patient's shortened maxillary dental arch despite poor bone quality, low primary implant stability, osteopenia, and a history of ARD therapy. Within the limitation of the study, extra‐short 4‐mm implants splinted to a standard‐size implant may be used as a less invasive, faster, and more economical alternative compared to sinus lifts. However, additional trials with a high number of participants and extended observation periods are needed to provide guidelines for such treatment concepts.
AUTHOR CONTRIBUTIONS
Katja Povšič: Data curation; project administration; writing – original draft; writing – review and editing. Čedomir Oblak: Data curation; investigation; methodology; writing – original draft; writing – review and editing. Michel Dard: Project administration; writing – original draft; writing – review and editing. Rok Gašperšič: Conceptualization; data curation; formal analysis; investigation; methodology; project administration; supervision; writing – original draft; writing – review and editing.
CONFLICT OF INTEREST STATEMENT
No conflict of interest relevant to this article is reported by the authors.
CONSENT
Written informed consent was obtained from the patient to publish this report in accordance with the journal's patient consent policy.
ACKNOWLEDGMENTS
The study was supported by the Ministry of Science and Technology of the Republic of Slovenia, grant no. J7‐4639. Implants were donated by Institut Straumann AG, Basel, Switzerland, and suprastructures by Dentalia, Ljubljana, Slovenia (a local Straumann distributor). Doctor Michel Dard is an employee of Institut Straumann AG. The authors thank Mr. Jure Garbajs, dental technician, for his skillful technical work.
Povšič K, Oblak Č, Dard M, Gašperšič R. Implant rehabilitation of a posterior maxilla with 4‐mm long implants splinted to a 10‐mm long implant in a patient with osteopenia taking antiresorptive drugs: A 5‐year follow‐up case report. Clin Case Rep. 2023;11:e07291. doi: 10.1002/ccr3.7291
DATA AVAILABILITY STATEMENT
Data sharing not applicable to this article as no datasets were generated or analysed during the current study.
REFERENCES
- 1. Stacchi C, Spinato S, Lombardi T, et al. Minimally invasive management of implant‐supported rehabilitation in the posterior maxilla, part II. Surgical techniques and decision tree. Int J Periodontics Restorative Dent. 2020;40(3):e95‐e102. doi: 10.11607/prd.4498 [DOI] [PubMed] [Google Scholar]
- 2. Lozano‐Carrascal N, Anglada‐Bosqued A, Salomó‐Coll O, Hernández‐Alfaro F, Wang HL, Gargallo‐Albiol J. Short implants (<8 mm) versus longer implants (≥ 8mm) with lateral sinus floor augmentation in posterior atrophic maxilla: a meta‐analysis of RCT's in humans. Med Oral Patol Oral Cir Bucal. 2020;25(2):e168‐e179. doi: 10.4317/MEDORAL.23248 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Molina A, Sanz‐Sánchez I, Sanz‐Martín I, Ortiz‐Vigón A, Sanz M. Complications in sinus lifting procedures: classification and management. Periodontol 2000. 2022;88(1):103‐115. doi: 10.1111/PRD.12414 [DOI] [PubMed] [Google Scholar]
- 4. Himmlová L, Dostálová T, Kácovský A, Konvičková S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent. 2004;91(1):20‐25. doi: 10.1016/j.prosdent.2003.08.008 [DOI] [PubMed] [Google Scholar]
- 5. Amato F, Polara G, Spedicato G. Immediate loading of fixed partial dental prostheses on extra‐short and short implants in patients with severe atrophy of the posterior maxilla or mandible: an up‐to‐4‐year clinical study. Int J Oral Maxillofac Implants. 2020;35(3):607‐615. doi: 10.11607/JOMI.7943 [DOI] [PubMed] [Google Scholar]
- 6. Stavropoulos A, Bertl K, Pietschmann P, Pandis N, Schiødt M, Klinge B. The effect of antiresorptive drugs on implant therapy: systematic review and meta‐analysis. Clin Oral Implants Res. 2018;29(Suppl 18):54‐92. doi: 10.1111/CLR.13282 [DOI] [PubMed] [Google Scholar]
- 7. Jung RE, Al‐Nawas B, Araujo M, et al. Group 1 ITI consensus report: the influence of implant length and design and medications on clinical and patient‐reported outcomes. Clin Oral Implants Res. 2018;29:69‐77. doi: 10.1111/CLR.13342 [DOI] [PubMed] [Google Scholar]
- 8. Sher J, Kirkham‐Ali K, Luo JD, Miller C, Sharma D. Dental implant placement in patients with a history of medications related to osteonecrosis of the jaws: a systematic review. J Oral Implantol. 2021;47(3):249‐268. doi: 10.1563/AAID-JOI-D-19-00351 [DOI] [PubMed] [Google Scholar]
- 9. Aljohani S, Fliefel R, Ihbe J, Kühnisch J, Ehrenfeld M, Otto S. What is the effect of anti‐resorptive drugs (ARDs) on the development of medication‐related osteonecrosis of the jaw (MRONJ) in osteoporosis patients: a systematic review. J Craniomaxillofac Surg. 2017;45(9):1493‐1502. doi: 10.1016/J.JCMS.2017.05.028 [DOI] [PubMed] [Google Scholar]
- 10. Jacobsen C, Metzler P, Rössle M, Obwegeser J, Zemann W, Grätz KW. Osteopathology induced by bisphosphonates and dental implants: clinical observations. Clin Oral Investig. 2013;17(1):167‐175. doi: 10.1007/S00784-012-0708-2 [DOI] [PubMed] [Google Scholar]
- 11. Gašperšič R, Dard M, Linder S, Oblak Č. The use of 4‐mm implants splinted to 10‐mm implants for replacement of multiple missing teeth in the posterior maxilla region with expanded maxillary sinus. An observational cases series: patient characteristics and preliminary results. Int J Periodontics Restorative Dent. 2021;41(2):261‐268. doi: 10.11607/PRD.4389 [DOI] [PubMed] [Google Scholar]
- 12. Gašperšič R, Dard M, Linder S, Oblak Č. One‐year results assessing the performance of prosthetic rehabilitations in the posterior maxilla supported by 4‐mm extrashort implants splinted to 10‐mm implants: a prospective case series. Int J Oral Maxillofac Implants. 2021;36(2):371‐378. doi: 10.11607/jomi.8645 [DOI] [PubMed] [Google Scholar]
- 13. Gašperšič R, Povšič K, Dard M, Linder S, Žarković Gjurin S, Oblak Č. No TitlExtra‐short 4‐mm implants splinted to 10‐mm implants in the posterior maxilla: 3‐year results. Int J Oral Maxillofac Implant. 2022. [DOI] [PubMed] [Google Scholar]
- 14. Cheung AM, Papaioannou A, Morin S. Osteoporosis Canada Scientific Advisory Council. Postmenopausal osteoporosis. N Engl J Med. 2016;374(21):2096. doi: 10.1056/NEJMc1602599 [DOI] [PubMed] [Google Scholar]
- 15. Rossi F, Lang NP, Ricci E, Ferraioli L, Baldi N, Botticelli D. Long‐term follow‐up of single crowns supported by short, moderately rough implants—a prospective 10‐year cohort study. Clin Oral Implants Res. 2018;29(12):1212‐1219. doi: 10.1111/CLR.13386 [DOI] [PubMed] [Google Scholar]
- 16. Bolle C, Felice P, Barausse C, Pistilli V, Trullenque‐Eriksson A, Esposito M. 4 mm long vs longer implants in augmented bone in posterior atrophic jaws: 1‐year post‐loading results from a multicentre randomised controlled trial. Eur J Oral Implant. 2018;11(1):31‐47. [PubMed] [Google Scholar]
- 17. Lombardo G, Signoriello A, Pardo A, et al. Short and ultra‐short (<6‐mm) locking‐taper implants supporting single crowns in posterior areas (part II): a 5‐year retrospective study on periodontally healthy patients and patients with a history of periodontitis. Clin Implant Dent Relat Res. 2022;24(4):455‐467. doi: 10.1111/CID.13103 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Torassa D, Naldini P, Calvo‐Guirado JL, Fernández‐Bodereau E. Prospective, clinical pilot study with eleven 4‐mm extra‐short implants splinted to longer implants for posterior maxilla rehabilitation. J Clin Med. 2020;9(2):357‐373. doi: 10.3390/JCM9020357 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Estévez‐Pérez D, Bustamante‐Hernández N, Labaig‐Rueda C, et al. Comparative analysis of peri‐implant bone loss in extra‐short, short, and conventional implants. A 3‐year retrospective study. Int J Environ Res Public Health. 2020;17(24):1‐14. doi: 10.3390/IJERPH17249278 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Lages FS, Douglas‐de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: a systematic review. Clin Implant Dent Relat Res. 2018;20(1):26‐33. doi: 10.1111/CID.12565 [DOI] [PubMed] [Google Scholar]
- 21. Ng P, Hu X, Wan S, Mo H, Deng F. Clinical outcomes of bicortical engagement implants in atrophic posterior maxillae: a retrospective study with 1 to 5 years follow‐up. Int J Periodontics Restorative Dent. 2018;38(5):e96‐e104. doi: 10.11607/PRD.3241 [DOI] [PubMed] [Google Scholar]
- 22. Ruocco‐Vetucci V, de Souza Faloni AP, Faeda RS. Follow‐up of an implant‐supported rehabilitation after long‐term use of alendronate. J Craniofac Surg. 2019;30(8):e793‐e796. doi: 10.1097/SCS.0000000000005782 [DOI] [PubMed] [Google Scholar]
- 23. Granate‐Marques A, Polis‐Yanes C, Seminario‐Amez M, Jané‐Salas E, López‐López J. Medication‐related osteonecrosis of the jaw associated with implant and regenerative treatments: systematic review. Med Oral Patol Oral Cir Bucal. 2019;24(2):e195‐e203. doi: 10.4317/MEDORAL.22691 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Bin WS, Hellstein JW, Kalmar JR. Narrative [corrected] review: bisphosphonates and osteonecrosis of the jaws. Ann Intern Med. 2006;144(10):753‐761. doi: 10.7326/0003-4819-144-10-200605160-00009 [DOI] [PubMed] [Google Scholar]
- 25. Ruggiero SL, Dodson TB, Aghaloo T, Carlson ER, Ward BB, Kademani D. American Association of Oral and Maxillofacial Surgeons' position paper on medication‐related osteonecrosis of the jaws—2022 update. J Oral Maxillofac Surg. 2022;80(5):920‐943. doi: 10.1016/J.JOMS.2022.02.008 [DOI] [PubMed] [Google Scholar]
- 26. Mardas N, Schwarz F, Petrie A, Hakimi AR, Donos N. The effect of SLActive surface in guided bone formation in osteoporotic‐like conditions. Clin Oral Implants Res. 2011;22(4):406‐415. doi: 10.1111/J.1600-0501.2010.02094.X [DOI] [PubMed] [Google Scholar]
- 27. Wen B, Zhu F, Li Z, Zhang P, Lin X, Dard M. The osseointegration behavior of titanium‐zirconium implants in ovariectomized rabbits. Clin Oral Implants Res. 2014;25(7):819‐825. doi: 10.1111/CLR.12141 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data sharing not applicable to this article as no datasets were generated or analysed during the current study.