Abstract
The temporal intensity fluctuations contain important information about the light source and light-medium interaction and are typically characterized by the intensity autocorrelation function, . The measurement of is a central topic in many optical sensing applications, ranging from stellar intensity interferometer in astrophysics, to fluorescence correlation spectroscopy in biomedical sciences and blood flow measurement with dynamic light scattering. Currently, at a single point is readily accessible through high-frequency sampling of the intensity signal. However, two-dimensional wide-field measurement of is still limited by camera frame rates. We propose and demonstrate a 2-pulse within-exposure modulation approach to break through the camera frame rate limit and obtain the quasi map in wide field with cameras of only ordinary frame rates.
Keywords: intensity fluctuation, autocorrelation function, camera, within-exposure modulation
1. Introduction
The intensity correlation function is widely used for quantifying optical fluctuations and its measurement has great physical and physiological significance in many optical sensing applications. It was first introduced in 1956 as intensity interferometry to measure the apparent angular diameter of stars1 and recently used to study stellar emission processes and calibrate star distances in astrophysics2,3. It plays an essential role in fluorescence correlation spectroscopy (FCS) in determining the diffusion coefficient of molecules and investigating biomolecular interaction processes4–8. It is used to achieve super-resolution optical fluctuation imaging (SOFI )9–11. It is used for particle sizing in dynamic light scattering12–14. In addition, tissue blood flow and perfusion information can be extracted from it, on which the diffuse correlation spectroscopy15–17 (DCS) and laser speckle contrast imaging18–22 (LSCI) are developed.
Generally, the intensity autocorrelation function measures the similarity of the intensity signal with itself between now and a moment later. Technically, it is defined as where is called the time lag, is the intensity signal of interest and denotes averaging. To resolve the intensity autocorrelation function, high temporal resolution detectors, such as avalanche photo-diodes (APD) or single photon avalanche diodes (SPAD), are required to record the intensity at sufficiently high sampling rates. However, these one-dimensional detectors are only capable of single-point measurements.
Light sheet or total internal reflection microscopy enables 2-dimensional (2D) FCS measurements at thousands of locations simultaneously by camera-based fluorescence intensity recording23,24. However, the electron multiplying charge coupled devices (EM-CCD), which are currently considered as the most suitable option in comprehensive comparison with other types of 2D detectors25,26, are still limited in frame rates (~ 1000 frames per second) and suffer from high instrumentation cost.
In DCS and LSCI, some high-speed cameras have been used to record the raw laser speckle signal and measure in a 2D field of view (FOV)27,28. In addition, SPAD arrays are utilized by speckle contrast optical spectroscopy to create synthetic multiple-exposure speckle contrast data29–31. However, both methods suffer from limited field of view and high instrumentation cost to resolve the signal at sufficient frame rates. Recently rolling shutters have been demonstrated helpful in alleviating the frame rate limit of cameras32. But the method trades the spatial resolution for temporal resolution and cannot measure of slowly varying dynamics due to the limited length of elongated speckle patterns created by the elliptical aperture.
Overall, is still measured based on the fully time-resolved signal, from which, however, high-frequency signal sampling is inevitable. As such, current methods must sacrifice either field of view or spatial resolution to accelerate the signal sampling. Here we propose a method to measure without resolving the fast temporal dynamics of the signal, thereby enabling characterization of rapid intensity fluctuations even at low camera frame rates.
Our method borrows the idea of speckle contrast from LSCI. The relationship between speckle contrast and has been well established in LSCI, where the pixel intensity is defined as
(1) |
The speckle contrast is then defined as
(2) |
Speckle contrast can be calculated either spatially or temporally. Spatially, a sliding window is typically used across the image to generate the speckle contrast of the center pixel by computing the standard deviation and mean of all pixel intensities within the window under the assumption of ergodicity21. Temporally, a series of images with the same camera exposure time can be acquired to calculate the speckle contrast at a certain pixel by computing the standard deviation over the mean of the pixel’s intensity in those images.
Note that speckle contrast can be measured with a much lower frame rate than since it is based on statistical properties of the integrated signal, within the camera exposure time . The integrated signal’s speckle contrast within different can be measured by multiple exposures of different exposure times33–36. The camera exposures do not have to be consecutive or acquired with a fast frame rate as long as the statistical property of the signal remains unchanged over the multiple exposures.
Speckle contrast is related to in a way that is an integral of weighted by a right triangle function if the illumination is held constant within the camera exposure37
(3) |
Recently, Siket et al. have generalized the relationship to cases where the illumination can be modulated by an arbitrary waveform38 (Supplemental section S1), namely
(4) |
where is the modulated signal intensity and where is the modulation waveform within the camera exposure and ranges from 0 to 1. is the autocorrelation of the modulation waveform defined as
One important observation from Eq. 4 is that if we could find a modulation waveform such that , then we can measure by measuring at a much lower frame rate. We achieve this by 2 -pulse modulated multiple-exposure imaging with two illumination pulses placed inside the exposure and their temporal separation (denoted as ) varied across exposures (Fig. 1.a). The laser illumination pulses are created by externally modulating the laser with an acoustic optical modulator (AOM). The idea is that when the pulse duration is approaching 0, weighted by would become the sum of two delta functions (Fig. 1b) and Eq. 4 would be reduced to
(5) |
where represents the square of the 2 -pulse modulated speckle contrast and is a constant that is independent of (Methods 4.1). Since forms a linear relationship with , we call it quasi .
Cameras of ordinary frame rates as low as 1 Hz are sufficient for our measurement approach as long as the signal’s statistical property remains invariant within the measurement. The camera-based characterization of at various time lags is first validated against the curve obtained by the traditional single-point photodiode measurement with 1 MHz sampling rate (instrumentation shown in Fig. 1c). Furthermore, wide-field quasi measurement and correlation time mapping are demonstrated in case of in vivo blood flow imaging (workflow summarized in Fig. 1d).
2. Results
2.1. A 10μs pulse duration is short enough such that
In this section, we first verify the equivalency between normalized and (denoted as and , respectively) with numerical simulation using a pulse duration. As highlighted by the green dashed line in Fig. 2a, the 2-pulse modulated decreases to the flat level earlier than the unmodulated but at about the same time as . The shape of curve is also more similar to that of compared with . Further normalization reveals the consistency between the normalized and curves (Fig. 2b). Such equivalency holds for curves over a wide range of decreasing speeds. The correlation time is varied from 5 μs to 0.1s, which covers the whole spectrum of reported by Postnov et al27. The discrepancy between and drastically increases when is reduced to 5 μs (Fig. 2c, d). However, maintains a good consistency with throughout the range (Fig. 2c, d). The maximum relative percentage discrepancy is below 10−5%.
The consistency between normalized and with a 10 μs pulse duration holds experimentally as well. For in vitro microfluidics experiments, raw images of different exposure times under 2-pulse modulation are shown in Fig. 3a. The average pixel intensity is approximately the same across different camera exposures, which is expected since the effective exposure time is kept the same in those exposures, i.e. all 20 μs. But the corresponding speckle contrast shows significant decrease when the temporal separation between the two illumination pulses, , increases from 50 to 100 μs (Fig. 2b). Such trend is further illustrated in Fig. 3c where the in the microfluidic channel area decreases as increases. In addition, of higher flow rates begins to decrease earlier than those of lower rates. Such relationship between flow rate and start time of decreasing is also reflected in curves which are derived from APD measurements of 1 MHz sampling rate (Fig. 3d). Most importantly, When and curves are normalized, they overlap on each other (Fig. 3e).
The downtick in the tail of curves (Fig. 3c), which is not present in corresponding curves (Fig. 3d), arises from the incomplete gating of light by AOM between the two illumination pulses. When the distance between two illumination pulses, , becomes too large relative to the pulse duration, the effects of non-zero residual illumination accumulated in between are no longer negligible and can result in a lowered value (Supplemental Material section S3). Removing the downtick will be the topic of a future publication.
Apart from comparing the values of normalized and , we also compared the inverse correlation time (ICT), i.e. , extracted from and curves. First, the electric field autocorrelation model identification capability of 2-pulse modulated multiple-exposure speckle imaging (2PM-MESI) is validated against APD-based direct measurements. When the flow is zero, for the best model, . When the flow is non-zero among the tested flow rates, (Supplemental Fig. S2). Second, as expected, the larger the flow rate, the larger the ICT (Fig. 3f). Third, ICT extracted from curves consisting of only 15 values of is consistent with that from evaluated at a much denser set of , Fig. 3f). This suggests that there is redundancy in curves and that the correlation time of can be estimated from only a few key data points.
The 2-pulse modulated curve is also compared with in vivo. Fig. 3g shows the location of three points (P1–3) where single-point direct measurements are performed with an APD. Note that their vessel radii are different, i.e., P1 the largest, P2 the smallest and P3 in the middle. As expected, their curves are also separated, i.e. of P1 starts decreasing first while that of P2 does last (Fig. 3h). The normalized in vivo curve is not as consistent with that of as it is in vitro. It could be due to the stronger flow disturbance in vivo as evident with the large error bars in the ICT plot (Fig. 3i). Note that APD and 2PM-MESI measurements are not performed simultaneously since 2PM-MESI requires modulating the illumination while the other not. The better consistency between normalized and in vitro could arise from the better flow control in vitro.
The model identification capability is also degraded in vivo. ICT extracted from 2PM-MESI curves is consistent with that from curves measured with APD when the model is fixed to for both APD and 2PM-MESI measurements (Fig. 3i, ). Unfixing the model and let the algorithms choose the optimal based on the fitting performance results in a degraded consistency of ICT between APD and 2PM-MESI measurements (Supplemental Fig. S3, ). It indicates that for complex flow dynamics in vivo, there is still room for the current settings of of 2PM-MESI, e.g. the number of exposures and values of , to be further optimized. In addition, a single model might be insufficient in vivo and a mixed model might be warranted27.
2.2. Widefield Quasi Measurement and Correlation Time Mapping
The correlation time is an important indicator of blood flow speed in LSCI. To demonstrate the 2D quasi measurement and correlation time mapping capability of our method, we present in this section widefield illumination results. APD results are not shown because they are dominated by noise in this illumination regime. Fig. 4a–d show the 2-pulse modulated quasi images in wide-field illumination at various . Small vessels gradually appear as increases, indicating slower intensity fluctuations. Note that the image size is as large as 1000×750 pixels (the corresponding FOV under 2× magnification: ~ 2.9 × 2.2 mm2). Fig. 4e–g show the inverse correlation time (ICT) maps extracted with the three different models. It can be seen that the ICT map of optimal (Fig. 4h) preserves the high ICT values in vascular regions in and ICT maps (Fig. 4e, f) as well as the low ICT values in parenchyma regions in ICT map (Fig. 4g). In addition, the distribution of optimal across the field of view (Fig. 4i) is consistent with what is reported by Postnov and Liu et al measuring with high-speed cameras27,39. The fitting results of curves at three representative points are shown in Fig. 4j–l ( and 0.5, respectively, position shown in Fig. 4i). The fitting results indicate that 2PM-MESI is capable of identifying one proper model according to the measured quasi curve.
3. Discussion
The measurement of intensity autocorrelation function, is a fundamental tool in many optical sensing applications to quantify intensity fluctuations and thus investigate the light source and light-medium interaction. However, measurements at short time lags are limited by the camera frame rate. To properly sample the rapid dynamics of intensity fluctuations, traditional two-dimensional measurement methods must sacrifice either field of view or spatial resolution to increase the temporal sampling rate. We propose the 2-pulse within-exposure modulation approach to break through the camera frame limit and change the problem from fast acquisition of raw images to the fast modulation of laser illumination. We showed that the normalized can be well approximated by the normalized , the 2-pulse modulated speckle contrast. With our method, can be measured at short time lags independent of camera frame rate.
The smallest time lag at which can be characterized by the 2-pulse modulated multiple-exposure imaging depends on the smallest value of that can be achieved. Since must be greater than or equal to the pulse duration , the question becomes how short the illumination pulse could be made while achieving a sufficient signal-to-noise ratio. We demonstrated that with a 10 μs pulse duration with ~100 mW laser power input into the AOM, the quasi can be measured with a decent signal-to-noise ratio even in widefield illumination, which is already beyond the capability of most cameras to measure with the traditional method. With a 10 μs pulse duration, we are able to evaluate quasi at the smallest time lag of , which would otherwise require a camera frame rate of 100 kHz with traditional methods.
Theoretically, the measurement of can be made at almost arbitrary time lags except those smaller than the pulse width . The sampling of the curve is determined by the number and values of the time lags. Finer sampling of the curve requires more images to be acquired, but is still independent of camera frame rate. The advantage of 2-pulse modulated multiple-exposure imaging over the traditional measurement method is that it enables cameras of even ordinary frame rates to measure at user-specified time lags. Even though it requires the use of an AOM or similar gating hardware to modulate the illumination within the camera exposure, the overall instrumentation cost is still substantially lower than that of high-speed cameras. In addition, the use of pulsed illumination reduces the average power incident upon the sample compared with continuous illumination, which can reduce tissue damage or photo-bleaching of fluorophores.
Note that we perform the modulation of the illumination, but our method is not limited to this case, especially in applications where modulation in the signal detection end is more convenient, for example intensity interferometry. In this case, modulation happens after the light interacts with the medium. Modulation could be also applied in the signal post-processing phase instead of the imaging phase, which we will revisit soon in the following discussion.
Even though the 2-pulse modulation strategy borrows the idea of speckle contrast from LSCI, it has the potential of being generalized to other optical applications than LSCI. This is because speckle contrast is, in definition, identical to the variation coefficient of a general signal. The relationship between speckle contrast and holds without special properties that would distinguish speckle from other types of intensity signal (Supplemental sections S1 and S2). Therefore, the idea of approximating with speckle contrast is not limited to speckle intensity signals. We verify the hypothesis with fluorescent intensity signal. We confirm that the 2-pulse modulation strategy works in FCS as well in principle by applying 2-pulse modulation to a realistic fluorescence intensity signal in the signal post-processing phase (Fig. 5).
We have demonstrated the relative equivalency between and 2-pulse modulated speckle contrast, i.e. . This is enough if we only care about the correlation time of since correlation time is invariant to linear transformations of . However, sometimes the absolute value of matters. Does our method still work in this case? We look into this question through applying 2-pulse modulation in the signal post-processing phase again. According to Eq. 5, the absolute value of can be estimated from 2-pulse modulated speckle contrast , i.e. . As shown in Fig. 6, the curves of realistic APD signals and the curves calculated from the same APD signal match with each other absolutely even though not perfectly. It suggests that given the same signal collected by exactly the same instrumentation, the absolute value of of the signal can be estimated from its 2-pulse modulated speckle contrast. However, to recover the absolute values of accurately, the requirement on the pulse duration is higher than to just recover the relative values (Supplemental section S4).
In summary, we proposed the 2-pulse modulation waveform to address the question of measuring without resolving the fast temporal dynamics of the intensity signal of interest and demonstrated wide-field intensity fluctuation imaging. Under the 2-pulse modulation, the problem is essentially converted from how fast the raw intensity images can be acquired to how fast the laser illumination or the detected signal can be modulated within the camera exposure. With a short pulse duration, the normalized can be approximated by the normalized . The multiple exposures to acquire at different do not need to be consecutive or acquired with a fast frame rate. It allows cameras of even ordinary frame rates to characterize the decay of intensity autocorrelation function. The method is expected to enable the 2-dimensional measurement of quasi and facilitate extracting the correlation time in wide field with a substantially lower instrumentation cost.
4. Method
4.1. Theory
In this section, we explain why is true when the pulse duration is approaching 0 in 2-pulse modulation. For a 2-pulse modulation waveform (Fig. 1b), can be written as
(6) |
where is the duration of a single illumination pulse and is the period of the modulation waveform. The duty cycle is . In this case, Eq. 4 can be simplified to
(7) |
where the subscript denotes the speckle contrast under 2-pulse modulation. The corresponding autocorrelation function of the modulation waveform, , is a pulse train consisting of two triangle pulses and (Fig. 1b),
(8) |
(9) |
and is valid for either or .
Note that the first right triangle pulse is solely dependent on and independent of the temporal separation of the two illumination pulses . In addition, the shape of , specifically the width and height, is independent of too. The horizontal position of , however depends on . Therefore, when is varied while holding constant, the second triangle pulse will move horizontally. Since is the integral of the product of and , the second triangle pulse, , sweeps the curve as changes, through which selective sampling of the curve is achieved.
Scaling by as in Eq. 7, we find that the height/width ratio of its two triangle pulses increases as the illumination pulse duration decreases. In the special case where approaches 0, weighted by becomes the sum of two delta functions
(10) |
Therefore, Eq. 7 simplifies to Eq. 5 where is a constant representing the contribution of which is independent of . Specifically, when the impact of on is negligible. A rigorous proof of Eq. 5 in this case from the perspective of statistics is provided in Supplemental section S2. Note that this supplemental proof is valid without assuming Eq. 3, 4 or 7. When must be accounted for, can also be estimated from the curve since if decreases to 1 when is infinitely large. Removing the constant and the scaling factor in Eq. 5 through normalization, we have
(11) |
where represents the normalization of , i.e. .
4.2. Numerical simulation
For any given curve, the 2-pulse modulated can be simulated according to Eq. 7. For the same curve, the corresponding traditional without modulation can be simulated according to Eq. 3.
Particularly, when assumes the following form,
(12) |
the analytical solution of the corresponding speckle contrast in the 2-pulse modulation without any approximation would be
(13) |
where and . Similarly, plugging the model into Eq. 3, the speckle contrast without modulation would be
(14) |
where .
In simulation results presented in Fig. 2 and S5, assumes the form of Eq. 12, the parameters in which are , , or ranges from 10 μs to 0.1 s with a resolution of 1 μs. The correlation time is varied in simulation.
4.3. Instrumentation setup
A volume holographic grating (VHG) wavelength stabilized laser diode (785 nm; LD785-SEV300, Thorlabs) is used to provide the light source. An optical isolator (Electro-Optics Technology, Inc.) based on Faraday rotation effect is placed immediately after the laser output to prevent potential inadvertent back reflections from disrupting the laser source. The light passing through the isolator is then coupled into a single-mode fiber (P3–780A-FC-2, Thorlabs, Inc.) to reshape the beam profile into a circular Gaussian one. The output beam is sent into an acoustic optical modulator (AOM) (AOMO 3100–125, Gooch & Housego, Inc.) through which the power of the 1st order diffraction can be manipulated. The 0th order diffraction is filtered out by an aperture. Apart from the widefield illumination in which the output beam from AOM is deflected down and incident upon the imaging object directly, another focused illumination beam path is constructed with two convex lenses (focal length, L3: 100 mm, L4:50 mm) whose relative distance can be adjusted to focus the beam onto the imaging spot. For detection path, the light is collected by a Nikon 24mm camera lens (AF NIKKOR 24 mm f/2.8D, Nikon, Inc.) and split into two by a 50:50 plate beamsplitter (BSW17, Thorlabs, Inc.). The transmission split is focused on a camera (acA1920–155 um, Basler, Inc.) via a Nikon 50mm lens (AF NIKKOR 50 mm f/1.8D, Nikon, Inc.). The same model of lens is used to focus the reflection split onto a fiber coupler to which a single-mode fiber (P3–780A-FC-2, Thorlabs, Inc.) is attached. The output light of the fiber is collimated via a collimator (focal length: 11 mm; F220APC-780, Thorlabs, Inc.) and focused on the photosensitive area of the APD (APD410A, Thorlabs, Inc.) by a spherical lens (focal length: 40 mm; LBF254–040-B, Thorlabs, Inc.). The intensity signal measured by APD is filtered by a low-pass filter (500 kHz; EF506, Thorlabs) and sampled by the data acquisition board at 1 MHz (USB-6363, National Instrument, Inc.).
4.4. LSCI experimental validation in vitro and in vivo
A single-channel microfluidics device is used to test the method in vitro. Its bulk is manufactured with polydimethylsiloxane (Dow Corning Sylgard 184 PDMS) in a 10 to 1 base-to-curing agent mixture by weight. Titanium dioxide (CAS 1317–80-2, Sigma, USA) is added into the mixture (1.8% w/w) to create optical properties mimicking the tissue40. The scattering solution flowing through the channel is made by diluting the Latex microsphere suspensions (5100A, 10% w/w, Thermo Fisher Scientific, USA) in a 4.8% v/v ratio with distilled water to mimic the optical properties of blood.
The mouse cranial window preparation procedures were detailed by Kazmi et al.41. All animal procedures are approved by the Institutional Animal Care and Use Committee (IACUC) of University of Texas at Austin.
In 2-pulse modulated multiple-exposure imaging, 15 camera exposure times were used for demonstration of characterizing at multiple time lags. and ranges from 10 μs to 5 ms. Specifically, the 15 T are 10 μs, 12 μs, 15 μs, 20 μs, 30 μs, 40 μs, 50 μs, 75 μs, 100 μs, 250 μs, 500 μs, 750 μs, 1 ms, 2.5 ms and 5 ms. The raw image size is 1000×750. Speckle contrast is computed spatially from raw images according to Eq. 2 with a 7 × 7 sliding window. Focused illumination is employed for both APD and camera measurements. For APD measurement, the laser power is 100 mW. In camera measurements, the laser power is attenuated by AOM to avoid pixel saturation. For in vitro experiements, 150 raw speckle images are collected for each camera exposure time and the raw intensity signal is recorded by APD for 10 s. The measurement is repeated 5 times for each flow rate. The flow rate increases from 0 to 100 μL/min with a step size of 10 μL/min in each repetition. The maximum and minimum ICT values in those five repetitions are discarded and the rest three are used for the ICT comparison between camera and APD measurements. For in vivo measurements, 30 raw camera images are collected for each exposure time and 2 s APD signal is recorded. The measurement is repeated 5 times at each point. Data collection is performed at 28 points in cranial windows of 4 mice (C57BL/6, Charles River Laboratories Inc.).
The curve is calculated from APD recordings in software according to with equally spaced. The correlation time is extracted from curve by fitting to the following model
(15) |
where is the instrumentation factor ranging from 0 to 1, denotes the fraction of dynamic component in the detected light ranging from 0 to 1, and denotes the noise. determines the type of model to use. can be fixed to 1 or chosen from 2,1 or 0.5 based on . For equally spaced , is concentrated in the tail when is plotted in the logarithmic scale. To counteract the skewing effects of denser sampling towards larger in the logarithmic axis, weighted fitting is deployed with as the weighting function. The weighting function equalizes the integral weight of data points within different ranges of the same length in the logarithmic scale, i.e. for . Weighted fitting by improves the fitting performance in the head of curve (Supplemental Fig. S4). To match the range in 2PM-MESI, the curve is truncated in the head such that only data of is used for correlation time extraction.
For both focused and widefield illumination experiments, the correlation time is extracted from measured curves according to the following model
(16) |
where represents a constant term independent of , and have the same meaning as those in Eq. 15. is the period of the 2-pulse modulation waveform.
4.5. Applying 2-pulse modulation to FCS
The FCS data comes from a public FCS dataset (https://github.com/FCSIib/FCSlib/blob/master/Sample%20Data/Cy5.tif, the curve of this sample data is provided in the Figure 5.3 of its user guide). The sampling rate is 500 kHz. The curve is calculated according to with the binned photon count number as the is calculated from binned photon count with the pulse width . Two binned photon count numbers at different time points are added together and the sum’s variance over its mean squared is calculated as (Eq. 2). The temporal gap between the two time points elongates so that at different is obtained.
Supplementary Material
Highlights.
The quasi intensity autocorrelation function can be measured at almost arbitrary time lags
Cameras of ordinary frame rates are sufficient for the measurement
Illumination within the camera exposure is modulated by an acoustic optical modulator using two short pulses
Acknowledgements
We acknowledge the support of National Institutes of Health (NIH) (Grant NS108484, EB011556) and UT Austin Portugal Program. We thank Dr. Jeanne Stachowiak, Dr. Carl C. Hayden and Dr. Feng Yuan for the help and support in the FCS project.
Footnotes
Conflicts of Interest
The authors declared no conflicts of interest.
Data Availability
The sample FCS data and MATLAB processing scripts relevant to Fig. 5 can be accessed through Github (the 2PM-FCS project). URL: https://github.com/2010511951/2PM-FCS. Other experimental data and resources will be made available upon reasonable request.
References
- [1].Brown Robert Hanbury and Twiss Richard Q. “A test of a new type of stellar interferometer on Sirius”. In: Nature 178 (1956), pp. 1046–1048. [Google Scholar]
- [2].Guerin W et al. “Temporal intensity interferometry: photon bunching in three bright stars”. In: Monthly Notices of the Royal Astronomical Society 472.4 (2017), pp. 4126–4132. [Google Scholar]
- [3].Rivet JP et al. “Intensity interferometry of P Cygni in the H emission line: towards distance calibration of LBV supergiant stars”. In: Monthly Notices of the Royal Astronomical Society 494.1 (2020), pp. 218–227. [Google Scholar]
- [4].Elson Elliot L and Magde Douglas. “Fluorescence correlation spectroscopy. I. Conceptual basis and theory”. In: Biopolymers: Original Research on Biomolecules 13.1 (1974), pp. 1–27. [DOI] [PubMed] [Google Scholar]
- [5].Magde Douglas, Elson Elliot L, and Webb Watt W. “Fluorescence correlation spectroscopy. II. An experimental realization”. In: Biopolymers: Original Research on Biomolecules 13.1 (1974), pp. 29–61. [DOI] [PubMed] [Google Scholar]
- [6].Krichevsky Oleg and Bonnet Grégoire. “Fluorescence correlation spectroscopy: the technique and its applications”. In: Reports on Progress in Physics 65.2 (2002), p. 251. [Google Scholar]
- [7].Ries Jonas and Schwille Petra. “Fluorescence correlation spectroscopy”. In: BioEssays 34.5 (2012), pp. 361–368. [DOI] [PubMed] [Google Scholar]
- [8].Yu Lan et al. “A comprehensive review of fluorescence correlation spectroscopy”. In: Frontiers in physics 9 (2021), p. 644450. [Google Scholar]
- [9].Alva Alma et al. “Fluorescence fluctuation-based super-resolution microscopy: Basic concepts for an easy start”. In: Journal of Microscopy 288.3 (2022), pp. 218–241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Dertinger Thomas et al. “Fast, background-free, 3D super-resolution optical fluctuation imaging (SOFI)”. In: Proceedings of the National Academy of Sciences 106.52 (2009), pp. 22287–22292. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Dertinger Thomas et al. “Achieving increased resolution and more pixels with Superresolution Optical Fluctuation Imaging (SOFI)”. In: Optics express 18.18 (2010), pp. 18875–18885. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Berne Bruce J and Pecora Robert. Dynamic light scattering: with applications to chemistry, biology, and physics. Courier Corporation, 2000. [Google Scholar]
- [13].Stetefeld Jörg, McKenna Sean A, and Patel Trushar R. “Dynamic light scattering: a practical guide and applications in biomedical sciences”. In: Biophysical reviews 8 (2016), pp. 409–427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Kaszuba Michael et al. “Measuring sub nanometre sizes using dynamic light scattering”. In: Journal of nanoparticle research 10 (2008), pp. 823–829. [Google Scholar]
- [15].Boas David A and Yodh Arjun G. “Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation”. In: JOSA A 14.1 (1997), pp. 192–215. [Google Scholar]
- [16].Durduran Turgut and Yodh Arjun G. “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement”. In: Neuroimage 85 (2014), pp. 51–63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [17].Boas David A et al. “Establishing the diffuse correlation spectroscopy signal relationship with blood flow”. In: Neurophotonics 3.3 (2016), p. 031412. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Briers JD and Fercher AF. “Retinal blood-flow visualization by means of laser speckle photography”. In: Investigative ophthalmology visual science 22.2 (1982), pp. 255–259. [PubMed] [Google Scholar]
- [19].Briers J David. “Laser speckle contrast imaging for measuring blood flow”. In: (2007).
- [20].Briers David et al. “Laser speckle contrast imaging: theoretical and practical limitations”. In: Journal of biomedical optics 18.6 (2013), p. 066018. [DOI] [PubMed] [Google Scholar]
- [21].Boas David A and Dunn Andrew K. “Laser speckle contrast imaging in biomedical optics”. In: Journal of biomedical optics 15.1 (2010), p. 011109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Dunn Andrew K. “Laser speckle contrast imaging of cerebral blood flow”. In: Annals of biomedical engineering 40.2 (2012), pp. 367–377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Wohland Thorsten et al. “Single plane illumination fluorescence correlation spectroscopy (SPIM-FCS) probes inhomogeneous three-dimensional environments”. In: Optics express 18.10 (2010), pp. 10627–10641. [DOI] [PubMed] [Google Scholar]
- [24].Krieger Jan W et al. “Imaging fluorescence (cross-) correlation spectroscopy in live cells and organisms”. In: Nature protocols 10.12 (2015), pp. 1948–1974. [DOI] [PubMed] [Google Scholar]
- [25].Yu Lan et al. “A comprehensive review of fluorescence correlation spectroscopy”. In: Frontiers in physics 9 (2021), p. 644450. [Google Scholar]
- [26].Singh Anand Pratap et al. “The performance of 2D array detectors for light sheet based fluorescence correlation spectroscopy”. In: Optics express 21.7 (2013), pp. 8652–8668. [DOI] [PubMed] [Google Scholar]
- [27].Postnov Dmitry D et al. “Dynamic light scattering imaging”. In: Science advances 6.45 (2020), eabc4628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28].Yuan Yuan et al. “High-dynamic-range blood flow rate measurement in a large-diameter vessel”. In: Applied Optics 60.23 (2021), pp. 6837–6842. [DOI] [PubMed] [Google Scholar]
- [29].Valdes Claudia P et al. “Speckle contrast optical spectroscopy, a noninvasive, diffuse optical method for measuring microvascular blood flow in tissue”. In: Biomedical optics express 5.8 (2014), pp. 2769–2784. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Dragojević Tanja et al. “High-speed multi-exposure laser speckle contrast imaging with a single-photon counting camera”. In: Biomedical optics express 6.8 (2015), pp. 2865–2876. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].Dragojević Tanja et al. “Compact, multi-exposure speckle contrast optical spectroscopy (SCOS) device for measuring deep tissue blood flow”. In: Biomedical optics express 9.1 (2018), pp. 322–334. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [32].Yi Changyoon et al. “Single-shot temporal speckle correlation imaging using rolling shutter image sensors”. In: Optica 9.11 (2022), pp. 1227–1237. [Google Scholar]
- [33].Parthasarathy Ashwin B et al. “Robust flow measurement with multiexposure speckle imaging”. In: Optics express 16.3 (2008), pp. 1975–1989. [DOI] [PubMed] [Google Scholar]
- [34].Kazmi Syed Mohammad Shams et al. “Chronic imaging of cortical blood flow using Multi-Exposure Speckle Imaging”. In: Journal of Cerebral Blood Flow Metabolism 33.6 (2013), pp. 798–808. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Kazmi SM Shams, Balial Satyajit, and Dunn Andrew K. “Optimization of camera exposure durations for multi-exposure speckle imaging of the microcirculation”. In: Biomedical optics express 5.7 (2014), pp. 2157–2171. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Smith Christopher et al. “All fiber-based illumination system for multi-exposure speckle imaging”. In: Biomedical Optics Express 14.2 (2023), pp. 771–782. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Bandyopadhyay Ranjini et al. “Speckle-visibility spectroscopy: A tool to study time-varying dynamics”. In: Review of scientific instruments 76.9 (2005), p. 093110. [Google Scholar]
- [38].Siket Máté et al. “Time varied illumination laser speckle contrast imaging”. In: Optics Letters 46.4 (2021), pp. 713–716. [DOI] [PubMed] [Google Scholar]
- [39].Liu Chang et al. “Choosing a model for laser speckle contrast imaging”. In: Biomedical Optics Express 12.6 (2021), pp. 3571–3583. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [40].Arkin Jeremy Sandon. “Investigation into optimizing laser speckle contrast imaging illumination”. PhD thesis. 2017. [Google Scholar]
- [41].Kazmi SM Shams et al. “Flux or speed? Examining speckle contrast imaging of vascular flows”. In: Biomedical optics express 6.7 (2015), pp. 2588–2608. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The sample FCS data and MATLAB processing scripts relevant to Fig. 5 can be accessed through Github (the 2PM-FCS project). URL: https://github.com/2010511951/2PM-FCS. Other experimental data and resources will be made available upon reasonable request.