Dosimetric uncertainties in very small (≤1.5 × 1.5 cm2) photon fields are remarkably higher, which undermines the validity of the virtual cone (VC) technique with a diminutive and variable MLC fields. We evaluate the accuracy and reproducibility of the VC method with a very small, fixed MLC field setting, called a fixed virtual cone (fVC), for small target radiosurgery such as trigeminal neuralgia (TGN). The fVC is characterized by 0.5 cm x 0.5 cm high‐definition (HD) MLC field of 10MV FFF beam defined at 100 cm SAD, while backup jaws are positioned at 1.5 cm x 1.5 cm. A spherical dose distribution equivalent to 5 mm (diameter) physical cone was generated using 10–14 non‐coplanar, partial arcs. Dosimetric accuracy was validated using SRS diode (PTW 60018), SRS MapCHECK (SNC) measurements. As a quality assurance measure, 10 treatment plans (SRS) for TGN, consisting of various arc ranges at different collimator angles were analyzed using 6 MV FFF and 10 MV FFF beams, including a field‐by‐field study (n = 130 fields). Dose outputs were compared between the Eclipse TPS and measurements (SRS MapCHECK). Moreover, dosimetric changes in the field defining fVC, prompted by a minute (± 0.5–1.0 mm) leaf shift, was examined among TPS, diode measurements, and Monte Carlo (MC) simulations. The beam model for fVC was validated (≤3% difference) using SRS MapCHECK based absolute dose measurements. The equivalent diameters of the 50% isodose distribution were found comparable to that of a 5 mm cone. Additionally, the comparison of field output factors, dose per MU between the TPS and SRS diode measurements using the fVC field, including ± 1 mm leaf shift, yielded average discrepancies within 5.5% and 3.5% for 6 MV FFF and 10 MV FFF beams, respectively. Overall, the fVC method is a credible alternative to the physical cone (5 mm) that can be applied in routine radiosurgical treatment of TGN.
Keywords: dosimetric uncertainties, fixed virtual cone, Monte Carlo simulations, stereotactic radiosurgery, trigeminal neuralgia
1. INTRODUCTION
Dosimetry in small (≤ 3 cm x 3 cm) photon field is complex due to the presence of small‐field related conditions1: (1) Lack of lateral charged particle equilibrium (LCPE), (2) Partial occlusion of the primary radiation source by collimating devices, and (3) The field size is comparable or smaller than the detector size. The International Codes of Practice (CoPs) such as the IAEA TRS‐483 2 , 3 and AAPM TG‐155 4 that provide the comprehensive guidelines for accurate and consistent small‐field dosimetry, are now available for reference and relative dose determination. In modern radiation therapy, the use of the small (≤ 3 cm x 3 cm) 1 and very small (≤ 1.5 cm x 1.5 cm), 5 non‐standard fields (static or dynamic) are very common for advanced treatments such as Intensity Modulated Radiation Therapy (IMRT), 6 Volumetric Modulated Arc Therapy (VMAT), 7 Stereotactic Radiosurgery (SRS), 8 and Stereotactic Body Radiotherapy (SBRT) 9 to treat relatively small lesions.
Currently, the cone‐based SRS using the specialized machines such as GammaKnife, CyberKnife, and other LINACs remains the standard practice for the treatment of small intracranial lesions such as trigeminal neuralgia (TGN). 10 , 11 , 12 , 13 However, it poses several challenges; need of (1) a cone‐specific dose calculation, (2) ensuring geometric precision for each cone collimating system, (3) rigorous patient‐specific quality assurance (QA), and (4) avoidance of collision for cone‐based arcs. It should be noted that a well‐defined spherical or ellipsoidal dose distribution along with a high mechanical precision in dose delivery is required for such treatment. It is also crucial to have the advanced LINAC that can deliver radiation beams with a sub‐millimeter accuracy in addition to the correctly defined outputs and accurate modeling of very small field/beam during the treatment planning. With the introduction of advanced MLC (leaf width 2.5−3.0 mm) technology, there is ongoing research on the MLC‐based SRS for the treatment small targets to achieve similar dosimetric goals in more efficient manner. To our best knowledge, the first MLC‐based radiosurgery for trigeminal neuralgia was reported and validated on the Varian TrueBeam STx LINAC 11 , 14 in which the 6 MV FFF beam was delivered with 18−21 static conformal fields (field size range, 0.6−0.7 cm). Next, a virtual cone technique using different standardized MLC fields, was introduced to replicate the dose distributions generated by physical cones. In this technique, very small non‐square fields defined by two central leaves of 10 MV FFF photon beams were used in concert with 7−10 non‐coplanar arcs to produce the spherical dose distribution comparable to 4−5 mm (diameter) physical cones. 15 , 16 However, the use of such diminutive and non‐square MLC fields could add further dosimetric uncertainties and complexities to the treatment.
It is evident that the effect of gravity on the MLC or imprecision of MLC leaves travel can cause notable dosimetric variation in small‐field (SRS) arc therapy in addition to gantry, collimator, couch rotations, and individual characteristics of leaves (shape, design). 9 , 17 , 18 An infinitesimal leaf shift of ± (0.5−1.0 mm) is normally expected (depending on how good the QA is performed) during the gantry rotation and it is noteworthy when the gantry is at 90° or 270° angle and the collimator is at 0° angle. 19 , 20 As a result, meaningful dosimetric uncertainties can be observed due to the significant geometrical variation of MLC fields during the small‐field arc treatments. 21 , 22 , 23 To mitigate these uncertainties, we propose the virtual cone technique with a fixed field geometry (i.e., 0.5 cm x 0.5 cm symmetric HDMLC field defined by two pairs of central leaves while backup jaws are positioned at 1.5 cm x 1.5 cm, defined at 100 cm SAD) in conjunction with multiple non‐coplanar arcs that is suitable for the small target radiosurgery of TGN. This approach intends to not only reduce undesirable geometrical (i.e., MLC) uncertainties as compared to non‐fixed field techniques but also make the treatment more convenient, reliable, and robust than previous approaches. Moreover, we perform the Monte Carlo (MC) study for small‐field virtual cone that can provide further authentication and evaluate dosimetric uncertainties associated with potential geometrical variation of MLC fields during the treatment (gantry, collimator rotation). An integration of proper uncertainty correction during the treatment planning could notably improve the accuracy and robustness small target radiosurgery. 22 , 24
The purpose of the research is to (1) study the viability of the fixed virtual cone (fVC) approach for the radiosurgical treatments of small targets such as trigeminal neuralgia, and (2) evaluate its dosimetric robustness considering potential MLC leaf shift (± 0.5−1.0 mm) during the treatment from measurements and MC calculations.
2. MATERIALS AND METHODS
2.1. Experimental
Dosimetric measurements were performed on the Varian Edge radiosurgery system (Varian Medical Systems, Palo Alto, California, USA) at South Florida Proton Therapy Institute (Delray Beach, Florida, USA). The LINAC is equipped with a HDMLC that is capable of delivering 6 MV FFF and 10 MV FFF photon beams with a sub‐millimeter accuracy. 25 It was defined at 100 cm source to surface distance (SSD) with 1 cGy/MU at depth of maximum dose and calibrated at 95 cm SSD, 5 cm depth (100 cm SAD). 26 The dose outputs measurement geometry consisted of an isocentric setting of 100 cm SAD with: (1) 95 cm SSD, 5 cm depth, and (2) 90 cm SSD, 10 cm depth in the water phantom. 3 , 26 The dose calculation was performed in the Eclipse treatment planning system (TPS) using the Acuros XB algorithm (version 16.1) with a grid size of 1 mm x 1 mm x 1 mm. 26 , 27
A separate machine model (fVC) for TGN treatments was created in the Eclipse TPS and commissioned as described by the AAPM TG‐101, 9 TG‐119, 28 and Varian Medical Systems guidelines. 26 Previously measured dosimetric leaf gap (DLG) values were modified to improve the agreement between measurements and calculations for such a very small‐field IMRT treatments. The fVC model was not designated as an equivalent model in the Eclipse; thus, a machine override was required at the treatment console prior to beam delivery. The beam modeling (for both energy beams) was performed on the Eclipse TPS with an Acuros XB algorithm and validated by the corresponding absolute dose measurements using the SRS MapCHECK (Sun Nuclear Corporation, FL) The SRS MapCHECK consists of 1013 n‐type solid‐state diodes (SunPoint 2 diode with 0.48 × 0.48 mm2 cross‐section area, 0.007 mm3 active volume) arranged on a 77 mm x 77 mm array with inter‐detector spacing of 2.47 mm. 29 It is designed to be compatible with an end‐to‐end phantom, StereoPHAN (Sun Nuclear Corporation, Florida, USA), which provides a freedom of interchanging the SRS MapCHECK and a film holder inside of it enabling comparison between both measurements without introducing positional uncertainty. The mechanical as well as dosimetric accuracy (i.e., end‐to‐end test) of the final treatment plan were also verified with SRS MapCHECK measurement using StereoPHAN. The SRS MapCHECK was considered as a standalone QA device for all measurements. A unique CT and MR protocol was applied during image acquisitions to meet the requirement of high precision dose delivery, thereby optimizing the image quality of tumor site.
2.1.1. Fixed virtual cone technique
We employed a virtual cone technique with a fixed field geometry, called a “fixed virtual cone (fVC) technique” to generate a very small, spherical dose distribution equivalent to 5 mm physical cone. The fVC was characterized by 0.5 cm x 0.5 cm HDMLC field at 100 cm SAD, defined by 2 pairs of central leaves while backup jaws were positioned at 1.5 cm x 1.5 cm (Figure 1). The spherical dose distribution was generated using several (10−14) partial arcs with different collimator angles and couch positions (Figure 2). For instance, at each couch position, different partial arcs with a clockwise and counter‐clockwise gantry rotation (gantry angle, 0°−360°, width, 30°−90°) were used with different collimator angles (0°−90°). The remaining leaves were positioned at 3 cm outside of the secondary collimation to avoid any geometrical uncertainties (photons attenuation). The 10 MV FFF photon beam was chosen clinically because of its high dose rate (up to 2400 MU/min) allowing for steep dose gradients, low collimator scatter, energy spectrum, and ability for SRS treatments to be delivered with a single fraction. When performing a high precision SRS for trigeminal neuralgia, the dose was delivered using multiple partial arcs at different collimator angles to optimize the dose coverage, thereby minimizing the dose to critical organs (brainstem). The dose per degree of the arcs was non‐uniform and proportional to the sine of the gantry angle. The weighting of the arcs was chosen in such a way that in the limit of large N and 2⊓ irradiation, it results in uniform fluence per unit area that is ideal for generating a spherical dose distribution.
FIGURE 1.

A beam eye view of the collimator settings for the fixed virtual cone (fVC) technique characterized by 0.5 cm x 0.5 cm HDMLC field, defined by two pairs of central leaves and fixed openings of jaws (1.5 cm x 1.5 cm) at 100 cm SAD.
FIGURE 2.

Three‐dimensional (3D) view of a treatment plan with multiple arc fields for the treatment of a very small tumor in left trigeminal nerve using the fixed (0.5 cm x 0.5 cm HDMLC field) virtual cone (fVC) technique.
2.1.2. Patient‐specific quality assurance
The treatment plans (10 MV FFF) of 10 clinical cases (left, right trigeminal neuralgia with a prescribed dose 7000−8000 cGy per fraction) were selected for the verification process or quality assurance and re‐planned with the 6 MV FFF (dose rate 1400 MU/min) photon beams. The verification process consisted of the gamma evaluation of planar dose profiles through isocenter using global criteria of 3%/1 mm (field‐by‐field), 2%/1 mm (cumulative dose), and the comparison of dose outputs or dose/MU between the TPS and the SRS MapCHECK (Sun Nuclear Corporation, Florida, USA) measurements at 100 cm SAD using a spatial resolution of 0.10 cm for both energy beams. The dose parameters were compared on the field‐by‐field basis as well as whole plan using 10% threshold. The field‐by‐field study included the comparison of dose outputs (at central axis) for 130 fixed virtual cone fields with different arcs (width 30°−90°), collimator angles (0°−90°), and couch positions. It is noted that SSD was changing throughout the measurements (arc therapy) due to the elliptical cranial shape thereby keeping the same SAD of 100 cm. The collimator was rotated for differing fields prior to irradiation of the arcs to deliver uniform dose and reduce MLC transmission. Due to the leaf ends being rounded shape, the rounded leaf edges were moved outside the jaw settings to Bank A, where the jaw collimators shielded radiation for the MLC.
2.1.3. TGN cases study
Fifteen clinical cases treated for TGN (prescribed dose, 7000−8000 cGy) using the fVC method (10 MV FFF beam) were collected retrospectively and important clinical parameters such as number of arcs, mean and maximum doses to the target, and maximum dose at critical organ (brainstem) were recorded from treatment plans. Similarly, the irradiated area/volume encompassed by 90%, 50%, and 25% isodose curves (Figure 3) together with gamma indices (passing rate, criteria) were manually acquired from the SRS MapCHECK QA using the SNC patient software tool. Dimensions of isodose curve (ellipsoidal) was defined by major and minor diameters (i.e., d 1 and d 2) and corresponding diameter (d) of equivalent circle or sphere in a given volume was calculated as d = √ (d 1 * d 2). The equivalent diameters were compared between the TPS and measurements for further evaluation.
FIGURE 3.

A representation of a treatment plan for trigeminal neuralgia using the fixed virtual cone technique along with multiple isodose lines. The isodose lines correspond to 90% (blue color), 50% (red‐orange color), and 25% (orange color) of the maximum dose are also displayed. A critical organ (i.e., brainstem) is contoured with a light blue color (left to the target).
2.1.4. Output factor constancy
Output factors were determined by the intermediate field method (IFM) 2 , 3 using the micro‐chamber (PTW 31021) 30 and small field SRS diode (PTW 60018). 31 The output factor constancy was investigated using the SRS diode (spatial resolution of 0.1 cm). The QA plans (with 200 MU) for the single field (at 0° gantry angle and 90° collimator angle) defining the fVC, including the fields with diminutive (± 0.5−1.0 mm) leaf shift, were generated for each energy beam and clinical setting. The corresponding dosimetric parameters (i.e., field output factors and dose/MU) were measured and dosimetric uncertainties caused by such infinitesimal leaf shift were examined. For instance, for each energy beam and clinical setting, dosimetric parameters of four different combination of fields (with a leaf shift ± 0.5 and ± 1.0 mm) were analyzed. The difference of dosimetric parameters between the TPS calculations and diode measurements were determined. 32 Moreover, the dosimetric variation linked with potential fluctuations in MLC fields during the gantry rotation or arc treatment was quantified by considering a minute leaf shift and the dosimetric robustness of the fVC technique was also evaluated. Furthermore, the field output factors in small fields are specified as a function of the field size expressed by the equivalent square small‐field size (Sclin ), which is determined by explicitly measuring the FWHM of the profiles both in‐plane (X) and cross‐plane (Y) directions at the depth of dose measurements. 2 For any rectangular small field, the equivalent square small‐field is given b:y2 , 4
| (1) |
2.2. Monte Carlo simulations
The EGSnrc Monte Carlo (MC) simulation package was used to simulate the Varian EDGE (Varian Medical Systems) linac. 33 The MC simulations were performed in two steps: MC BEAM modelling of linac head with/without MLC using the BEAMnrc and 3D voxelized dose calculations in the phantom using the DOSXYZnrc. The MC beam model was designed including jaws (Y, X) and HD120 MLC by utilizing the phase‐space files (for 6 and 10 MV FFF beams) scored above the Y jaw those were provided by the vendor (Varian Medical Systems, Palo Alto, California, USA). Details of MC beam modeling provided in another article. The dose outputs (.3ddose files) from the DOSXYZnrc program were extracted and analyzed using the MATLAB 2020 (MathWorks.com) software. For the MC dose model validation, two‐dimensional (2D) dose profiles of 10 cm x 10 and 5 cm x 5 cm reference fields for both energy beams were compared between the MC simulations and Eclipse TPS. 34 , 35 , 36 The comparison was made by recording the FWHMs of profiles scored in water phantom at 100 cm SSD. The corresponding depth dose curves were generated, and depths of maximum doses were recorded. Second, the dosimetric parameters such as TMR, OF were compared between the MC simulations and the TPS for static MLC fields of (10 × 10−0.5 × 0.5) cm2 at 95 cm SSD, 5 cm depth (100 cm SAD) in water phantom. Additionally, the OFs were compared with corresponding diode (PTW 60018) measurements for further validation. Ultimately, the potential dosimetric variation in very small (< 1 cm x 1 cm) MLC fields using fixed virtual cone method were analyzed amongst the MC, TPS and measurements (SRS diode) for the complete assessment. An extensive evaluation process of MC method will be described in a separate article.
3. RESULTS
3.1. Experimental
Results of all dosimetric measurements including the case study of 15 clinical TGN cases treated using the fVC method are presented in the following sections.
3.1.1. Fixed virtual cone technique
The virtual cone beam model using the smallest (0.5 cm x 0.5 cm) but fixed HDMLC field is commissioned within 3% dose difference against measurements using the SRS MapCHECK and effectively implemented for the small‐target stereotactic radiosurgery of TGN patients (15 cases) at the clinic.
3.1.2. Patient‐specific quality assurance
Table 1 shows results of the field‐by‐field study of 10 trigeminal cases (n = 130 fields) treated with the fixed virtual cone technique. The absolute value of average dose/MU difference between the SRS MapCHECK measurements and TPS for the 10 MV FFF beam was found (2.10 ± 1.30) % compared to (4.25 ± 1.80) % for the 6 MV FFF beam. 37 , 38 However, the difference in dose outputs (at central axis) was observed up to 12% for certain fields (arcs) depending on the gantry and collimator angles, energy beams, spatial resolution. The discrepancy was observed higher for the arc field with the gantry at 90°/270° and collimator at 0° angles. In addition, the comparison of whole treatment plans (i.e., cumulative dose comparison) resulted relatively higher passing rate (≥ 97%) for the 10 MV FFF beam compared to the 6 MV FFF beam (≥ 95%) with 2%/1 mm gamma criterion. The beam delivery time for 10 MV FFF plan was ∼ 5 min (∼ 30% decrease) shorter than for the 6 MV FFF plan. Furthermore, the measured dose profiles defined by Sclin are slightly wider than the planned ones and the apparent widening of field size are clearly visible (Figure 4c). For instance, the clinical field sizes (Sclin) given by Equation (1) are reported as 0.57−0.60 and 0.60−0.64 cm for 0.5 cm fixed virtual cones of 6 and 10 MV FFF beams, respectively.
TABLE 1.
Comparison of dose (cGy) per monitor unit (MU) between the TPS (Acuros XB) and measurements (SRS MapCHECK) during the field‐by‐field study of 10 treatment plans (average values of 130 fields) of trigeminal cases treated with the fixed virtual cone technique.
| Dose (cGy)/ MU | |||||
|---|---|---|---|---|---|
| 6 MV FFF | 10 MV FFF | ||||
| Calculated | Measured | Difference (%) | Calculated | Measured | Difference (%) |
| 0.45 ± 0.02 | 0.47 ± 0.02 | 4.25 ± 1.80 | 0.46 ± 0.02 | 0.47 ± 0.02 | 2.10 ± 1.30 |
FIGURE 4.

A representative case for gamma evaluation of treatment plans (10 MV FFF beam) for trigeminal neuralgia using the fixed virtual cone technique. Sections (a) and (b) represent planner dosemaps obtained from the TPS (Acuros XB) and SRS MapCHECK measurement, respectively. The sections (c) and (d) show the corresponding dose profiles in x and y directions, respectively. Solid line represents the planned dose profile, whereas yellow circle represents the measured one.
3.1.3. TGN cases study
Figure 4 represents the gamma evaluation for a trigeminal treatment plan (10 MV FFF beam) using the SRS MapCHECK, in which the cumulative absolute doses between the TPS and measurement are compared with 2%/1 mm gamma criterion. The isodose lines/curves are appeared ellipsoidal in shape due to the usage of various non‐coplanar arcs and variable MUs (i.e., it produces a constant dose per degree of the arc) during the treatment (Figure 4a and b). The average major diameters of the 90%, 50%, and 25% isodose curves are (2.1 ± 0.3) mm, (7.0 ± 0.8) mm, and (11.5 ± 1.2) mm, respectively. Whereas the respective minor diameters are (1.82 ± 0.25) mm, (5.7 ± 0.7) mm, and (7.5 ± 0.8) mm. The equivalent diameters (circle) of the 90%, 50%, and 25% isodose curves (coronal plane) for calculated and measured dose distributions are provided in Table 2.
TABLE 2.
Comparison of calculated and measured equivalent diameters of 90%, 50%, and 25% isodose lines in coronal plane for 15 trigeminal cases treated with fixed virtual cone method.
| Equivalent diameter a (mm) | |||||
|---|---|---|---|---|---|
| 90% isodose curve | 50% isodose curve | 25% isodose curve | |||
| Calculated | Measured | Calculated | Measured | Calculated | Measured |
| 1.95 ± 0.30 | 1.97 ± 0.20 | 6.10 ± 0.70 | 6.40 ± 0.50 | 8.70 ± 0.95 | 9.20 ± 0.80 |
The equivalent diameter is the diameter of a circle or a sphere of the given volume.
The calculated diameters are slightly smaller than the measured ones. The equivalent diameters of 50% isodose line are comparable to that generated using the 5 mm (diameter) physical cone. In Figure 4, the x‐dose profile (c) is wider than the y‐profile (d) due to a notable photon attenuation through MLC leaf edges and this effect is more dominant for the 10 MV FFF beam. The average clinical field size, given by Equation (1), for the SRS MapCHECK measurements was found as 6.4 mm using the fVC with 10 MV FFF beam. Additionally, the maximum dose delivered to the critical organ such as brainstem was reported ∼ 18 Gy (< 20 Gy) on average. The average volume of brainstem receiving maximum dose was 0.03 cm3.
3.1.4. Output factor constancy
The output factor constancy was investigated with the SRS diode (PTW 60018) measurements using the intermediate field method and considering a static fVC field. Table 3 shows the comparison of dosimetric parameters (field output factors and dose/MU) between the calculations (TPS) and diode measurements for 0.5 cm x 0.5 cm fVC field of the 6 MV FFF and 10 MV FFF photon beams at (1) 95 cm SSD, 5 cm depth, and (2) 90 cm SSD, 10 cm depth. 37 , 38 The measured values are consistently higher compared to the calculated ones as expected. For 10 MV FFF beam, the absolute differences of field output factors between the calculated (TPS) and measured (SRS diode) were −1.26% and −1.00% at settings (1) and (2), respectively. Whereas such differences for the 6 MV FFF beam were −2.45% and −1.45% at respective settings. Similarly, difference in dose/MU for the 10 MV FFF beam were −1.27% and −1.20% at settings (1) and (2), respectively. Whereas the corresponding values were −3.35% and −2.56%, respectively, for the 6 MV FFF beam.
TABLE 3.
Comparison of field output factors and Dose/MU between the Eclipse TPS (Acuros XB) and diode measurements (PTW 60018) for the 0.5 cm x 0.5 cm fixed virtual cone field (gantry at 0° and collimator at 90°) of the 6 MV FFF and 10 MV FFF photon beams at (1) 95 cm SSD, 5 cm depth, and (2) 90 cm SSD, 10 cm depth settings.
| Beam + SSD/Depth | Field output factors | Dose (cGy)/MU | ||||
|---|---|---|---|---|---|---|
| Calculated | Measured | Diff. (%) | Calculated | Measured | Diff. (%) | |
| 6 MV FFF 95/5 cm | 0.605 | 0.620 | 2.45 | 0.558 | 0.577 | 3.35 |
| 10 MV FFF 95/5 cm | 0.550 | 0.557 | 1.30 | 0.547 | 0.554 | 1.10 |
| 6 MV FFF 90/10 cm | 0.560 | 0.568 | 1.42 | 0.424 | 0.435 | 2.56 |
| 10 MV FFF 90/10 cm | 0.510 | 0.515 | 0.98 | 0.432 | 0.440 | 1.85 |
Furthermore, Figure 5 shows the comparison of (a) field output factors and (b) dose (cGy)/MU between the Eclipse TPS (Acuros XB) and SRS diode measurements (PTW 60018) for the field defining fVC, including ± 1.0 mm leaf shift, for the 6 MV FFF and 10 MV FFF photon beams at clinical settings (1) and (2). Solid lines represent calculated parameters in the fVC fields (including ± 1.0 mm leaf shift), whereas dotted lines display corresponding measured parameters. The calculated dosimetric parameters are found consistently lower (i.e., 5−10% for the 6 MV FFF beam and 3−5% for the 10 MV FFF beam, respectively) as compared to the corresponding measured parameters. In particular, the trend of these parameters seemed unusual (i.e., sudden fall off in calculated values whereas abrupt rise in measured ones) in fVC field with – 1.0 mm leaf shift exhibiting very high dosimetric uncertainties. Additionally, the calculated results exhibited notably higher (σ ≥ 4.0%) dosimetric variation (i.e., represented by error bars) in those non‐square or asymmetric MLC fields compared to measured ones (σ ≤ 2%). In other words, the margin of error or uncertainty (± 2σ) in the calculated field OF, dose per MU is significantly higher (∼ 4.5%) as compared to measured ones (∼ 2%). It is also observed that the clinical setting (2) has exhibited more accurate and consistent results than the setting (1) regardless of beam energy. Overall, the 10 MV FFF beam has delivered more stable dose outputs than the 6 MV FFF beam and the beam at setting (2) yielded better results.
FIGURE 5.

Comparison of (a) field output factors and (b) dose (cGy)/MU between the Eclipse TPS (Acuros XB) and SRS diode (PTW 60018) measurements for the field defining fVC (i.e., 0.5 cm x 0.5 cm HDMLC field defined with a gantry at 0° and collimator at 0°), including ± 1.0 mm leaf shift, for the 6 MV FFF and 10 MV FFF photon beams at different clinical settings. Solid and dotted lines represent calculated and measured parameters, respectively.
3.2. Monte Carlo modeling
MC simulations were conducted for static fields with or without MLC and validated against the TPS. The optimal MC results (≤ 3% dose uncertainty and ≤ 3% dose difference to TPS) were obtained after simulating 30−50 millions of incident particles for 10 × 10 and 5 × 5 cm2 fields of both 6 MV FFF and 10 MV FFF beams at 100 cm SSD in water phantom. 39 However, the dose uncertainty was slightly increased for the smaller fields but did not exceed 5% threshold. Detailed results of the model validation will be presented in a separate research paper.
The MC study of fVC included simulations of the 0.50 cm x 0.50 cm HDMLC field and evaluation of dosimetric variation due to a minute leaf shift (± 0.5−1.0 mm) in the field. The equivalent square field size (Sclin ), given by Equation (1), of the MC simulated dose profiles were recorded slightly higher than that obtained from the TPS. Particularly, for nominal square field sizes of 0.50 cm, Scli n were 0.60 cm for MC as compared to 0.57 cm from TPS using 6 MV FFF beam, whereas 0.62 cm for MC versus 0.59 cm of TPS using 10 MV FFF beam. Additionally, the comparison field output factors has shown a reasonable (≤ 4% difference) agreement against measurements thereby the 10 MV FFF beam yielding comparatively better results than the 6 MV FFF beam. Furthermore, results of MC study for dosimetric (uncertainty) evaluation in the fVC fields including ± 1.0 mm leaf shift have shown similar (pattern) dosimetric variation (σ ≤ 2.5%) as the measured ones (σ ≤ 2.0%). 38 However, the model requires further investigation for comprehensive results.
4. DISCUSSION
4.1. Experimental
During the PSQA, the dose (at central axis) difference was observed notably higher (up to 12%) for certain fields (arcs) depending on the gantry and collimator angles, energy beams, which is on par with previous studies. 18 , 19 , 22 , 40 Note that the highest dose deviation was associated with the arc field when the gantry at 90°/270° and collimator at 0° angles that can be avoided or corrected during treatment planning. 9 , 19 , 40 , 41 Our approach minimizes these inconsistencies by utilizing a symmetric MLC field with a fixed jaws setting and using various non‐coplanar arcs with minimal error so that it is capable of delivering more accurate and precise dose to the target. Regarding dosimetric uncertainty analysis in fVC field, the 6 MV FFF beam exhibited relatively higher difference in outputs between measurements and TPS (Table 3) along with a larger variation (Figure 5) than 10 MV FFF beam, which in on par with previous studies. 3 , 27 , 31 , 42 Since dosimetric uncertainty increases (up to 5%) with decrease in field size upto 0.5 cm x 0.5 cm and discrepancy between calculated and measured parameters rises notably due to the small field, Acuros XB commissioning, and diode measurements related uncertainties. 2 , 42 , 43 When the field size < 1 cm x 1 cm, Dosimetric uncertainties tend to increase rapidly and become more challenging to quantify. 2 , 4 It is obvious that the higher deviation (Figure 5) in calculated values in the fVC fields (± 1.0 mm shift) and larger difference compared to measurements are fairly reasonable results. In addition, comparatively lower values for the 10 MV FFF beam than the 6 MV FFF beam indicate that more particles escape the field due to a larger lateral range of the secondary charged particles and the higher transmission through the penumbra for higher energy beams. 2 , 4 An overall dosimetric inconsistency in the fields with ± 0.5 mm leaf shift are found to be insignificant (difference ≤ 3%). Whereas it is noteworthy (up to 12% difference) with ± 1.0 mm shift. These results are consistent with the field‐by‐field study during the quality assurance process specifically for arc fields with the gantry at 90° or 270° and the collimator at 0° angles. Early estimation of dosimetric impact due to a tiny fluctuation in MLC field/leaf during the gantry or collimator rotation could give a flexibility of applying corrections in treatment planning if necessary.
Additionally, our method allows freedom to manipulate the dose delivery (i.e., with selective arcs, arc span, and arc weight, variable MUs) such that the tumor receives maximum dose and critical organs receive minimal dose. As the dose per degree of the arcs is usually non‐uniform and proportional to the sine of the gantry angle. 15 , 22 The weighting is allocated in such a way that it results uniform fluence over an area that is ideal for generating a spherical dose distribution. Due to the constancy of dose per degree of arcs, the isodose lines/curves for trigeminal treatment plans are appeared to be ellipsoidal with significant dose spill in anterior‐posterior direction as illustrated in Figure 4a, b. A critical advantage of generating such dose distributions is to maximize the dose to the target while minimizing that to the critical organ (brainstem). 15 , 16 , 22 Furthermore, the 10 MV FFF beam was chosen because it can deliver higher dose rate (up to 2400 MU/min) thereby minimizing the treatment time (∼ 5 min) and it also provides more dosimetric stability and reliable dose delivery. 15 Automated table motion during the beam delivery will further reduce treatment time thereby making our technique even more efficient. The dimensions (i.e., major and minor diameters) of 90%, 50%, and 25% isodose lines (ellipsoidal) are manually obtained from the treatment plans and measurements. The equivalent diameters of the corresponding circles or spheres of given volume are computed to approximate the prescribed, irradiated volumes, dose coverage, proximity of critical organ (brainstem) as reported in Table 2. The dose distribution (50% isodose line) of fVC is equivalent to 5 mm physical cone. Additionally, average treatment time is ∼ 45 min, including initial set up, cone beam CT (initial, mid treatment), beam delivery (∼ 20 min), which is comparable to previous studies. 28 , 29 , 30 The maximum dose delivered to the brainstem was restricted to < 25 Gy (average ∼ 18 Gy).
4.2. MC modeling
Although we discuss details of the MC results in a separate paper. It is worth mentioning that the MC calculated parameters (i.e., Scli n, field output factors) for the 0.5 cm x 0.50 cm fVC field are found relatively larger compared to the TPS, but smaller than measured for smaller fields, which is on a par with previous studies. 2 , 3 For instance, MC simulated Sclin was 0.60 cm for 10 MV FFF beam and 0.58 cm for 6 MV FFF beam, respectively. Whereas respective field OFs were 0.562 and 0.511, calculated at 90 SSD, 10 cm depth. It is due to the fact that the apparent widening effect becomes pronounced in very small fields 5 and even more for higher energy beams. 4 It means more particles can escape the field due to a larger lateral range of the secondary charged particles and the higher transmission through the penumbra for higher energy beams. The potential sources of the larger dosimetric uncertainties as well as discrepancies reported in very small fields could be due t:o 2 , 4 , 34 , 35 , 44 (1) the presence of prominent small‐field conditions, (2) an inaccurate modeling of geometry (MLC), (3) the adoption of different dose calculations modalities (i.e., a volume‐based dose calculation in MC versus a point‐based dose calculation in TPS), and (4) the difference in source type (i.e., energy spectrum or fine tuning versus monoenergetic beam).
Furthermore, we extended the MC study to evaluate dosimetric uncertainties in fixed virtual cones with ± (0.5−1.0) mm leaf shift. Early MC results (water phantom) demonstrate the dosimetric variation (pattern) are more aligned to measurements (SRS diode) rather than TPS ones, which indicates the MC fairly estimates (σ ∼ 2%) the uncertainties in these asymmetric fields, even better than the TPS (σ ∼ 4%) when TPS only uses symmetric fields. 37 , 38 It could result a significant dosimetic inconsistencies during small‐field SRS arc treatments using dynamic MLC fields unless these are properly addressed in advance. These results justify the importance of MC study specifically for smaller fields besides the TPS and measurements. However, the MC model needs further improvements that can be done with rigorous simulations by inserting the accurate geometrical (i.e., MLC, detector, and phantom) details for the complete assessment. Lately, the MC dose uncertainty has been further reduced (≤ 3%) after applying various variance reduction techniques and extensive simulations 44 , 45 and results are being evaluated quantitatively, which will be discussed elsewhere.
Overall, the 10 MV FFF beam has shown better performance than the 6 MV FFF beam in both clinical settings and our test setting of 90 cm SSD, 10 cm depth with more favorable results. These results substantiate that our fixed virtual cone with 10 MV FFF beam is a straightforward and reliable method, which can be effectively implemented in routine small target radiosurgery of TGN.
5. CONCLUSION
The fixed virtual cone technique using the 0.5 cm x 0.5 cm HDMLC field for small‐targeted stereotactic radiosurgery such as TGN is successfully validated within 3% dose difference against measurements using the SRS MapCHECK. As the quality assurance measure, the comparison of dosimetric parameters (field output factors, dose/MU) between the TPS and SRS MapCHECK measurements has shown relatively better agreement for the 10 MV FFF beam compared to the 6 MV FFF beam. The output factors constancy test using the TPS and SRS diode measurements for the fVC fields including a minute leaf shift (± 0.5−1.0 mm) has shown a notable (3−10%) difference along with 1.5−4.5% uncertainty, depending on beam energy and clinical settings. Furthermore, early results of independent MC study for the fVC fields have shown similar (trend) dosimetric variation as measured but relatively better than TPS results. However, the model can be improved further by simulating the accurate geometrical (i.e., MLC, detector, and phantom) details for the final assessment. Collectively, the fVC with 10 MV FFF beam has yielded more accurate and consistent results.
The feasibility of the fixed virtual cone technique for a small‐targeted, high‐precision stereotactic radiosurgery and its dosimetric robustness due to a potential leaf drift during the gantry rotation are evaluated using measurements and MC simulations. This is a convenient and reliable alternative to the physical cone (5 mm), which can be routinely applied for the SRS of trigeminal neuralgia. This research can be extended further by a machine learning approach based on MLC log files as well as radiomics analysis from QA images.
AUTHOR CONTRIBUTIONS
Taindra Neupane has substantial contributions to the conception or design of the work, or acquisition, analysis, interpretation of data for the work. He has drafted/revised it critically and makes a final decision for the approval of final version. The author agrees hold the accountability for all aspects of work in ensuring that questions related to the accuracy or integrity of the work.
Charles Shang has major contributions to the conception or design of the work and acquisition/analysis/interpretation of data for the work. He has revised it critically and makes a final decision for the approval of final version. The author agrees hold the accountability for all aspects of work in ensuring that questions related to the accuracy or integrity of the work.
Maxwell Kassel has considerable contributions to the acquisition/interpretation of data for the work. He has revised the draft critically as a native speaker and suggests in the decision making for the approval of final version. The author agrees hold the accountability for all aspects of work in ensuring that questions related to the accuracy or integrity of the work.
Wazir Muhammad has useful contributions to the interpretation of data for the work. He has contributed notably towards the scientific revision of the draft and supports in the decision making for the approval of final version. The author agrees hold the accountability for all aspects of work in ensuring that questions related to the accuracy or integrity of the work.
Theodora Leventouri has valuable contributions to the design of the work. She has contributed significantly towards the scientific revision of the draft as a native speaker and supports in the decision making for the approval of final version. The author agrees hold the accountability for all aspects of work in ensuring that questions related to the accuracy or integrity of the work.
Timothy R. Williams has valuable contributions to the inception, concept and design of this treatment technique and has a critical role in its implementation into the clinic. He has contributed notably towards the sceintific revision of the draft and gives the final decision for approval. The author agrees to hold accountability for all aspects of work in ensuring that questions related to the validity and integrity of the work.
CONFLICT OF INTEREST STATEMENT
The authors declare no conflicts of interest.
ACKNOWLEDGMENTS
The authors would like to thank the Varian Medical Systems, Palo Alto, California, USA, for providing the phase‐space files of True Beam (EDGE) linac and the physicists at the facility for their constructive support during this research.
Neupane T, Shang C, Kassel M, Muhammad W, Leventouri T, Williams TR. Viability of the virtual cone technique using a fixed small multi‐leaf collimator field for stereotactic radiosurgery of trigeminal neuralgia. J Appl Clin Med Phys. 2023;24:e14148. 10.1002/acm2.14148
Taindra Neupane and Charles Shang equally contributed for this work and manuscript.
REFERENCES
- 1. Das IJ, Ding GX, Ahnesjo A. Small fields: nonequilibrium radiation dosimetry. Med Phys. 2008;35(1):206‐215. doi: 10.1118/1.2815356 [DOI] [PubMed] [Google Scholar]
- 2. Palmans H, Andreo P, Huq MS, Seuntjens J, Christaki KE, Meghzifene A. Dosimetry of small static fields used in external photon beam radiotherapy: summary of TRS‐483, the IAEA‐AAPM international Code of Practice for reference and relative dose determination. Med Phys. 2018;45(11):e1123‐e1145. doi: 10.1002/mp.13208 [DOI] [PubMed] [Google Scholar]
- 3. Huq MS, Hwang MS, Teo TP, Jang SY, Heron DE, Lalonde RJ. A dosimetric evaluation of the IAEA‐AAPM TRS483 code of practice for dosimetry of small static fields used in conventional linac beams and comparison with IAEA TRS‐398, AAPM TG51, and TG51 Addendum protocols. Med Phys. 2018. doi: 10.1002/mp.13092 [DOI] [PubMed] [Google Scholar]
- 4. Das IJ, Francescon P, Moran JM, et al. Report of AAPM Task Group 155: megavoltage photon beam dosimetry in small fields and non‐equilibrium conditions. Med Phys. 2021;48. [DOI] [PubMed] [Google Scholar]
- 5. Charles PH, Cranmer‐Sargison G, Thwaites DI, et al. A practical and theoretical definition of very small field size for radiotherapy output factor measurements. Med Phys. 2014;41(4):041707. doi: 10.1118/1.4868461 [DOI] [PubMed] [Google Scholar]
- 6. Low DA, Moran JM, Dempsey JF, Dong L, Oldham M. Dosimetry tools and techniques for IMRT. Med Phys. 2011;38(3):1313‐1338. [DOI] [PubMed] [Google Scholar]
- 7. Mynampati DK, Yaparpalv R, Hong L, Kuo HC, Mah D. Application of AAPM TG 119 to Volumetric Arc Therapy (VMAT). J Appl Clin Med Phys. 2012;13(5):108‐116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Dieterich S, Cavedon C, Chuang CF, et al. Report of AAPM TG 135: quality assurance for robotic radiosurgery. Med Phys. 2011;38(6):2914‐2936. doi: 10.1118/1.3579139 [DOI] [PubMed] [Google Scholar]
- 9. Benedict SH, Yenice KM, Followil D, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37(8):4078‐4101. [DOI] [PubMed] [Google Scholar]
- 10. Gerbi BJ, Higgins PD, Cho KH, Hall WA. Linac‐based stereotactic radiosurgery for treatment of trigeminal neuralgia. J Appl Clin Med Phys. 2004;5(3):80‐92. doi: 10.1120/jacmp.v5i3.1997 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Pokhrel D, Sood S, McClinton C, et al. Linac‐based stereotactic radiosurgery (SRS) in the treatment of refractory trigeminal neuralgia: detailed description of SRS procedure and reported clinical outcomes. J Appl Clin Med Phys. 2017;18(2):136‐143. doi: 10.1002/acm2.12057 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Tuleasca C, Régis J, Sahgal A, et al. Stereotactic radiosurgery for trigeminal neuralgia: a systematic review. J Neurosurg. 2019;130(3):733‐757. doi: 10.3171/2017.9.jns17545 [DOI] [PubMed] [Google Scholar]
- 13. Berti A, Ibars G, Wu X, et al. Evaluation of CyberKnife radiosurgery for recurrent trigeminal neuralgia. Cureus. 2018. doi: 10.7759/cureus.2598 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Stevens MTR, Lobb EC, Yenice KM. Validation of MLC‐based linac radiosurgery for trigeminal neuralgia. J Appl Clin Med Phys. 2018;19(4):214‐221. doi: 10.1002/acm2.12381 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Popple RA, Wu X, Brezovich IA, et al. The virtual cone: a novel technique to generate spherical dose distributions using a multileaf collimator and standardized control‐point sequence for small target radiation surgery. Adv Radiat Oncol. 2018;3(3):421‐430. doi: 10.1016/j.adro.2018.02.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Brown TAD, Ayers RG, Popple RA. Commissioning a multileaf collimator virtual cone for the stereotactic radiosurgery of trigeminal neuralgia. J Appl Clin Med Phys. 2022;23(5). doi: 10.1002/acm2.13562 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Hanley J, Dresser S, Simon W, et al. AAPM Task Group 198 Report: an implementation guide for TG 142 quality assurance of medical accelerators. Med Phys. 2021;48(10). doi: 10.1002/mp.14992 [DOI] [PubMed] [Google Scholar]
- 18. Xu Z, Yao R, Podgorsak MB, Wang IZ. Effects of collimator angle, couch angle, and starting phase on motion‐tracking dynamic conformal arc therapy (4D DCAT). J Appl Clin Med Phys. 2017;18(5):54‐63. doi: 10.1002/acm2.12132 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Losasso T, Chui C‐S, Ling CC. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys. 1998;25(10):1919‐1927. doi: 10.1118/1.598381 [DOI] [PubMed] [Google Scholar]
- 20. Hernandez V, Abella R, Calvo JF, Jurado‐Bruggemann D, Sancho I, Carrasco P. Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques. Med Phys. 2015;42(4):1911‐1916. doi: 10.1118/1.4915541 [DOI] [PubMed] [Google Scholar]
- 21. Miften M, Olch A, Mihailidis D, et al. Tolerance limits and methodologies for IMRT measurement‐based verification QA:recommendations of AAPM Task Group No. 218 . Med Phys. 2018;45(4):e53‐e83. doi: 10.1002/mp.12810 [DOI] [PubMed] [Google Scholar]
- 22. Brezovich IA, Wu X, Popple RA, et al. Stereotactic radiosurgery with MLC‐defined arcs: verification of dosimetry, spatial accuracy, and end‐to‐end tests. J Appl Clin Med Phys. 2019;20(5):84‐98. doi: 10.1002/acm2.12583 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Kumaraswamy LK, Schmitt JD, Bailey DW, Xu ZZ, Podgorsak MB. Spatial variation of dosimetric leaf gap and its impact on dose delivery. Med Phys. 2014;41(11):111711. doi: 10.1118/1.4897572 [DOI] [PubMed] [Google Scholar]
- 24. Sendani NG, Karimian A, Mahdavi SR, Jabbari I, Alaei P. Effect of beam configuration with inaccurate or incomplete small field output factors on the accuracy of treatment planning dose calculation. Med Phys. 2019;46(11):5273‐5283. doi: 10.1002/mp.13796 [DOI] [PubMed] [Google Scholar]
- 25. Ning Wen HL, Song K, Chin‐Snyder K, et al. Characteristics of a novel treatment system for linear accelerator–based stereotactic radiosurgery. J Appl Clin Med Phys. 2015;16(4). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Tuomas Torsti LK, Petaja V. Using Varian Photon Beam Source Model for Dose Calculation of Small Fields. Varian Medical Systems; 2013. [Google Scholar]
- 27. Fogliata A, Lobefalo F, Reggiori G, et al. Evaluation of the dose calculation accuracy for small fields defined by jaw or MLC for AAA and Acuros XB algorithms. Med Phys. 2016;43(10):5685. doi: 10.1118/1.4963219 [DOI] [PubMed] [Google Scholar]
- 28. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359‐5373. doi: 10.1118/1.3238104 [DOI] [PubMed] [Google Scholar]
- 29. Rose MS, Tirpak L, Van Casteren K, et al. Multi‐institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance. Med Phys. 2020;47(7):3153‐3164. doi: 10.1002/mp.14153 [DOI] [PubMed] [Google Scholar]
- 30. Casar B, Gershkevitsh E, Mendez I, Jurković S, Saiful Huq M. Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations — perpendicular and parallel. Med Phys. 2020;47(1):242‐259. doi: 10.1002/mp.13894 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Casar B, Gershkevitsh E, Mendez I, Jurković S, Huq MS. A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams. Med Phys. 2019;46(2):944‐963. doi: 10.1002/mp.13318 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Chaikh A, Giraud J‐Y, Perrin E, Bresciani J‐P, Balosso J. The choice of statistical methods for comparisons of dosimetric data in radiotherapy. Radiat Oncol. 2014;9(1):205. doi: 10.1186/1748-717x-9-205 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Kawrakow I. The EGSnrc Code System, Monte Carlo Simulation of Electron and photon Transport. 2001. NRCC Report PIRS‐701. 2001. [Google Scholar]
- 34. Bergman AM, Gete E, Duzenli C, Teke T. Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans. J Appl Clin Med Phys. 2014;15(3):148‐163. doi: 10.1120/jacmp.v15i3.4686 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Cheng JY, Ning H, Arora BC, Zhuge Y, Miller RW. Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter‐free stereotactic radiosurgery system. J Appl Clin Med Phys. 2016;17(3):100‐110. doi: 10.1120/jacmp.v17i3.5956 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Borges C, Zarza‐Moreno M, Heath E, Teixeira N, Vaz P. Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam. Med Phys. 2012;39(1):415‐423. doi: 10.1118/1.3671935 [DOI] [PubMed] [Google Scholar]
- 37. Taindra N. Efficacy of the Virtual Cone Method Using Fixed Small Multi‐Leaf Collimator Field for Stereotactic Radiosurgery. Florida Atlantic University; 2022. [Google Scholar]
- 38. Neupane Taindra SC, Wazir M. Leventouri Theodora Viability of the virtual cone technique using a small fixed MLC field for stereotactic radiosurgery. Med Phys. 2022;49(6). doi: 10.1002/mp.15769 (MO‐G‐202‐4) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Neupane Taindra SC, Wazir M. Leventouri Theodora Evaluation of dosimetric uncertainties in small fields of the Varian EDGE radiosurgery system using Monte Carlo simulations. Med Phys. 2021;48(6):e117‐e635. doi: 10.1002/mp.15041 (PO‐GePV‐T‐398) [DOI] [PubMed] [Google Scholar]
- 40. Wijesooriya K, Aliotta E, Benedict S, Read P, Rich T, Larner J. RapidArc patient specific mechanical delivery accuracy under extreme mechanical limits using linac log files. Med Phys. 2012;39(4):1846‐1853. doi: 10.1118/1.3690464 [DOI] [PubMed] [Google Scholar]
- 41. Oliver M, Gagne I, Bush K, Zavgorodni S, Ansbacher W, Beckham W. Clinical significance of multi‐leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol. 2010;97(3):554‐560. doi: 10.1016/j.radonc.2010.06.013 [DOI] [PubMed] [Google Scholar]
- 42. Dufreneix S, Bellec J, Josset S, Vieillevigne L. Field output factors for small fields: a large multicentre study. Physica Med. 2021;81:191‐196. doi: 10.1016/j.ejmp.2021.01.001 [DOI] [PubMed] [Google Scholar]
- 43. Lárraga‐Gutiérrez JM, García‐Garduño OA, Herrera‐González JA, Galván De La Cruz OO. Evaluation of Acuros® XB accuracy for static small fields dose calculations based on the IAEA/AAPM TRS‐483 recommendation. Physica Med. 2021;89:140‐146. doi: 10.1016/j.ejmp.2021.06.021 [DOI] [PubMed] [Google Scholar]
- 44. Chetty IJ, Curran B, Cygler JE, et al. Report of the AAPM Task Group No. 105: issues associated with clinical implementation of Monte Carlo‐based photon and electron external beam treatment planning. Med Phys. 2007;34(12):4818‐4853. doi: 10.1118/1.2795842 [DOI] [PubMed] [Google Scholar]
- 45. Kawrakow I, Walters BRB. Efficient photon beam dose calculations using DOSXYZnrc with BEAMnrc. Med Phys. 2006;33(8):3046‐3056. doi: 10.1118/1.2219778 [DOI] [PubMed] [Google Scholar]