Abstract
Point-of-care ultrasound (POCUS) of the internal jugular vein (IJV) offers a non-invasive means of estimating right atrial pressure (RAP), especially in cases where the inferior vena cava is inaccessible or unreliable due to conditions such as liver disease or abdominal surgery. While many clinicians are familiar with visually assessing jugular venous pressure through the internal jugular vein, this method lacks sensitivity. The utilization of POCUS significantly enhances the visualization of the vein, leading to a more accurate identification. It has been demonstrated that combining IJV POCUS with physical examination enhances the specificity of RAP estimation. This review aims to provide a comprehensive summary of the various sonographic techniques available for estimating RAP from the internal jugular vein, drawing upon existing data.
Keywords: Point-of-care ultrasound, Bedside ultrasound, Internal jugular vein, Right atrial pressure, Central venous pressure
Core Tip: Point of care ultrasound of the internal jugular vein serves as a non-invasive tool to evaluate right atrial pressure. This is particularly useful when neck vein inspection is challenging, or inferior vena cava sonography lacks reliability. When combined with the physical examination and other sonographic parameters, it becomes a valuable component of bedside hemodynamic evaluation.
INTRODUCTION
An accurate assessment of hemodynamic status at the bedside is of utmost importance in guiding proper management through altering blood volume and cardiac output, especially in critically ill patients. Central venous pressure (CVP) is a physiologic parameter that is used to assess cardiac preload within this evaluation. CVP is identical to right atrial pressure (RAP) in the absence of vena caval obstruction, and these terms are often used interchangeably. Assessing CVP clinically by performing a neck veins examination can be challenging at times and yield low sensitivity, especially in the elderly, obese, and short-necked patients[1-4]. The gold standard for measuring CVP is via inserting a central venous catheter (CVC) into the superior vena cava, however, this method is invasive, time-consuming, and can cause serious complications[4-6]. With the help of point-of-care ultrasound (POCUS), RAP can be non-invasively measured through various methods including using compression ultrasound on the forearm vein[7], evaluation of the inferior vena cava (IVC), and evaluation of the internal jugular vein (IJV). The most common method is the evaluation of IVC diameter and collapsibility[8,9] (Figure 1). However, the association between IVC parameters and RAP is moderate and lacks reliability in mechanically ventilated individuals or those with abdominal compartment syndrome[10]. Moreover, imaging the IVC can be challenging in situations like morbidly obese individuals, those unable to recline, patients with cirrhosis, ascites, or those who have undergone recent abdominal surgery[11]. Owing to its location and convenient accessibility, IJV POCUS presents a viable alternative in such instances. Unlike visual inspection, where identifying the vein can be difficult in many cases, POCUS makes the IJV easily visible, saving time for physicians and improving the accuracy of the examination[12,13].
Figure 1.
Estimation of right atrial pressure by inferior vena cava ultrasound in spontaneously breathing patients, based on current guidelines[8]. IVC: Inferior vena cava; RA: Right atrium; HV: Hepatic vein; RAP: Rright atrial pressure.
SONOGRAPHIC METHODS OF IJV ASSESSMENT
Multiple techniques have been described to assess CVP by IJV ultrasound. Based on available data, these methods encompass measuring the vertical height of the blood column in the IJV[13-16], utilizing the IJV height-to-width ratio (aspect ratio)[17,18], IJV collapsibility index (IJV-CI)[13,17,19-22], IJV distensibility index[23], measuring the maximal anteroposterior diameter of the IJV (AP-IJV Dmax)[18,24], calculating the difference in the percentage of the IJV cross-sectional area (IJV-CSA) or diameter before and after the Valsalva maneuver[25,26], and combining IJV vertical height with right atrial depth measured with echocardiography[27,28].
The first technique, pioneered by Lipton[14], uses POCUS to visualize the height of the blood column in the IJV above the sternal angle of Louis where a venous collapse is seen (the ‘wine bottle’ or ‘paint brush’ appearance), thereby being able to measure the height of the IJV more accurately than compared to only performing neck veins examination (Figure 2). The height is then added to the hypothesized right atrial depth below the sternum of 5 cm to calculate the RAP, abiding by the doctrine of Sir Thomas Lewis, who first proposed the use of neck veins examination in 1930. This method’s accuracy is questionable, owing to the presumed right atrial depth of 5 cm, which is proven to be inaccurate[27,29]. In a study by Deol et al[30] ultrasound-assisted assessment of the column height underestimated CVP by 4.7 cm H2O, whereas visual inspection underestimated it by 6.1 cm H2O. Moreover, this approach can be time consuming as it requires two tools to assess the height of the column from the sternal angle (e.g., two rulers, a pen and a ruler, or a tongue depressor and a ruler or tape, commonly employed by physicians at the bedside), further compromising measurement accuracy[31].
Figure 2.
Internal jugular vein collapse point compared to a wine bottle and paint brush. Figures adapted from NephroPOCUS.com with permission.
The second technique uses the height-to-width ratio (aspect ratio) to estimate the CVP as a binary variable (e.g., </> 10 mmHg), without precisely quantifying it. The transducer is positioned at the level of the cricoid cartilage in the transverse plane, then the height (anteroposterior diameter) and the width (transverse diameter) are measured to obtain the aspect ratio. A ratio of < 0.75 has shown to correlate with RAP of < 10 mmHg with a sensitivity and specificity of 62% and 67% respectively in one study[17]. This method yields mixed results, and its utility is currently not well-demonstrated[18].
The third technique uses the collapsibility index of IJV (IJV-CI) to estimate RAP. The IJV-CI is calculated by its respiratory variation, (IJVmax – IJVmin)/IJVmax expressed in percentage (Figure 3). The actual RAP cannot be quantified, but studies[13,21,22] have shown that an IJV-CI of approximately ≤ 30% signifies an elevated CVP of ≥ 8 mmHg, demonstrating good sensitivity and specificity. Worthy of particular mention, in a study by Leal-villarreal et al[21] in patients with cirrhosis, IJV-CI (anteroposterior diameter 2 cm above the sternoclavicular joint) ≤ 24.8% was better at predicting a CVP ≥ 8 mmHg with 100% sensitivity and 97.1% specificity outperforming IVC POCUS. In this cohort, IVC POCUS was unattainable in 18% of the cases correlating with our real-life experience in cirrhotic patients.
Figure 3.

Anteroposterior diameter of the internal jugular vein. M-mode tracing depicts respiratory variation in the diameter.
The fourth technique uses the IJV distensibility index to predict the fluid responsiveness in patients undergoing mechanical ventilation. IJV is imaged at the cricoid cartilage level in the transverse plane and the distensibility index is calculated by (IJVmax – IJVmin)/IJVmin, expressed in percentage. In one study, an IJV distensibility index of > 18% before volume challenge (7 mL/kg crystalloid) had an 80% sensitivity and 85% specificity to predict fluid responsiveness[23]. While this technique does not quantify the RAP, it could be a valuable adjunct to other hemodynamic parameters in assessing fluid status in mechanically ventilated patients.
The fifth technique uses the IJV maximal anteroposterior diameter (AP-IJV Dmax) to predict the RAP whether it is low (< 8 mmHg) or high. Similar to the second technique of acquiring the aspect ratio, the transducer is positioned transversely 2 cm above the clavicle, at end-expiration and the diameter is collected to correlate with RAP. It has been shown that AP-IJV Dmax has the best correlation with RAP, the best validity in predicting its values, and a very good inter-rater reliability[24] along with its accuracy in predicting low RAP of < 8 mmHg[18].
The sixth technique uses the difference in the percentage of the IJV cross-sectional area (IJV-CSA) before and after the Valsalva maneuver to assess for RAP. The Valsalva maneuver increases the IJV-CSA by about 20%-30% in patients with normal RAP, and it is assumed that in volume-overloaded patients with a decrease in venous compliance, the increase in IJV-CSA will be blunted. This technique measures the patient’s IJV-CSA at the vertical column height where a venous collapse is seen and re-measures that same parameter while the patient performs the Valsalva maneuver, then calculates the difference in the percentage of the IJV-CSA (Figures 4 and 5). One study[25] found that a < 17% increase in IJV-CSA with Valsalva predicted elevated RAP (≥ 12 mmHg) with 90% sensitivity, 74% specificity, and a 94% negative predictive value. Alternatively, the ratio between maximum IJV diameter during Valsalva maneuver and diameter at rest (JVD ratio) < 4 has also shown to predict elevated RAP[26].
Figure 4.
Increase in the size of internal jugular vein with Valsalva maneuver by several folds in a spontaneously breathing person with normal right atrial pressure.
Figure 5.
Minimal increase in the size of internal jugular vein with Valsalva maneuver in a spontaneously breathing heart failure patient with elevated right atrial pressure.
Another recently described technique uses the IJV vertical height combined with right atrial depth measured with echocardiography to quantify the RAP. While similar to the first technique in principle, this approach enhances accuracy by directly measuring the right atrial depth instead of relying on the assumed 5 cm. It involves marking the highest point of the venous collapse (wine bottle sign) and then performing a focused cardiac ultrasound to visualize the heart in long axis (parasternal long axis view). Right atrial depth is measured from the surface to the location of the non-coronary cusp of the aortic valve attachment to the posterior left ventricular outflow tract (Figure 6). The two parameters, the column height, and the right atrial depth, are then added together for an estimation of the RAP in cm H2O, later converted to mmHg. This method predicted actual RAP within 3 mmHg 74% of the time when compared to cardiac catheterization-derived value[27].
Figure 6.

Measurement of right atrial depth using parasternal long axis view on focused cardiac ultrasound. RVOT: Right ventricular outflow tract; LV: Left ventricle; LVOT: LV outflow tract; LA: Left atrium; NCC: Non-coronary cusp of aortic valve.
In an alternative technique described by Xing et al[28], the initial step of measuring the height of collapse point is the same but the right atrial depth is calculated using an adjusted apical 4-chamber view with the right atrium in the center of the view, where distance between the transducer contact point and the center of the right atrium is recorded. A pencil and a ruler are then used in relation to the transducer to derive the surface projection of the center of the right atrium. The RAP is then estimated by adding the IJV column height to the right atrial depth. This technique demonstrates a substantial correlation and high accuracy, with a mean difference of 0.22 mmHg in RAP compared to CVC. Nevertheless, it lacks practicality in a clinical setting where swift assessment is essential, as it involves multiple operators and the use of various measuring instruments.
With respect to estimation of RAP using IJV POCUS in mechanically ventilated patients, there is limited data. One noteworthy study is by Hilbert et al[32] in a cohort of 47 patients, where they evaluated the ratio between IJV diameter in the 30° and 0° head position (30/0 ratio). A 30/0 ratio of < 0.45 indicated a low CVP, whereas a cutoff of > 0.65 predicted a CVP ≥ 10 mmHg with reasonable accuracy.
CONCLUSION
In essence, IJV POCUS serves as a quick, non-invasive bedside tool for assessing RAP. However, the diversity in techniques requires cautious interpretation, avoiding overestimation of accuracy. Interpretation of IJV POCUS in the appropriate clinical context is crucial. Additionally, CVP is not a surrogate for volume status but just one component in bedside hemodynamic assessment, to be considered alongside other variables. For instance, Figure 7 presents an algorithm for hemodynamic assessment in cirrhotic patients with acute kidney injury, utilizing IJV for fluid tolerance assessment[33]. Future research should compare various techniques in different clinical settings to establish a standardized, practical method for routine use.
Figure 7.
Diagnostic algorithm in a case of cirrhosis and suspected hemodynamic acute kidney injury. Incorrect angle of insonation is a frequent source of error when assessing LVOT VTI (surrogate for stroke volume) and other Doppler measurements listed. Adapted from Ref. 33 with kind permission of the publisher (corresponding author’s prior open access publication). Blue boxes: Right heart; Red boxes: Left heart-related sonographic parameters; Green outlines: Volume tolerance phenotype; Orange outlines: Volume intolerance. POCUS: Point-of-care ultrasonography; VTI: Velocity time integral; E/e′: Ratio of the early diastolic waves of the mitral inflow Doppler and mitral annular tissue Doppler; LA: Left atrium; RV: Right ventricle; RVSP: Right ventricular systolic pressure; TAPSE: Tricuspid annular plane systolic excursion; S′: Tricuspid annular systolic velocity; SVC: Superior vena cava; ARDS: Acute respiratory distress syndrome; HV: Hepatic vein; PV: Portal vein.
Footnotes
Conflict-of-interest statement: All authors have no conflicts of interest to declare.
Provenance and peer review: Invited article; Externally peer reviewed.
Peer-review model: Single blind
Peer-review started: December 20, 2023
First decision: January 11, 2024
Article in press: February 1, 2024
Specialty type: Medicine, general and internal
Country/Territory of origin: United States
Peer-review report’s scientific quality classification
Grade A (Excellent): 0
Grade B (Very good): 0
Grade C (Good): C
Grade D (Fair): 0
Grade E (Poor): 0
P-Reviewer: Uz Zaman M, Pakistan S-Editor: Gong ZM L-Editor: A P-Editor: Zhao S
Contributor Information
Vichayut Chayapinun, Department of Medicine, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok 10400, Thailand.
Abhilash Koratala, Division of Nephrology, Medical College of Wisconsin, Wauwatosa, WI 53226, United States. koratala.abhi@gmail.com.
Taweevat Assavapokee, Department of Medicine, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok 10400, Thailand.
References
- 1.Assavapokee T, Thadanipon K. Examination of the Neck Veins. N Engl J Med. 2020;383:e132. doi: 10.1056/NEJMvcm1806474. [DOI] [PubMed] [Google Scholar]
- 2.Jang T, Aubin C, Naunheim R, Char D. Ultrasonography of the internal jugular vein in patients with dyspnea without jugular venous distention on physical examination. Ann Emerg Med. 2004;44:160–168. doi: 10.1016/j.annemergmed.2004.03.014. [DOI] [PubMed] [Google Scholar]
- 3.Donahue SP, Wood JP, Patel BM, Quinn JV. Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med. 2009;27:851–855. doi: 10.1016/j.ajem.2008.06.005. [DOI] [PubMed] [Google Scholar]
- 4.Baumann UA, Marquis C, Stoupis C, Willenberg TA, Takala J, Jakob SM. Estimation of central venous pressure by ultrasound. Resuscitation. 2005;64:193–199. doi: 10.1016/j.resuscitation.2004.08.015. [DOI] [PubMed] [Google Scholar]
- 5.Parienti JJ, Mongardon N, Mégarbane B, Mira JP, Kalfon P, Gros A, Marqué S, Thuong M, Pottier V, Ramakers M, Savary B, Seguin A, Valette X, Terzi N, Sauneuf B, Cattoir V, Mermel LA, du Cheyron D 3SITES Study Group. Intravascular Complications of Central Venous Catheterization by Insertion Site. N Engl J Med. 2015;373:1220–1229. doi: 10.1056/NEJMoa1500964. [DOI] [PubMed] [Google Scholar]
- 6.Kusminsky RE. Complications of central venous catheterization. J Am Coll Surg. 2007;204:681–696. doi: 10.1016/j.jamcollsurg.2007.01.039. [DOI] [PubMed] [Google Scholar]
- 7.Thalhammer C, Aschwanden M, Odermatt A, Baumann UA, Imfeld S, Bilecen D, Marsch SC, Jaeger KA. Noninvasive central venous pressure measurement by controlled compression sonography at the forearm. J Am Coll Cardiol. 2007;50:1584–1589. doi: 10.1016/j.jacc.2007.07.022. [DOI] [PubMed] [Google Scholar]
- 8.Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, Solomon SD, Louie EK, Schiller NB. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23:685–713; quiz 786. doi: 10.1016/j.echo.2010.05.010. [DOI] [PubMed] [Google Scholar]
- 9.Dipti A, Soucy Z, Surana A, Chandra S. Role of inferior vena cava diameter in assessment of volume status: a meta-analysis. Am J Emerg Med. 2012;30:1414–1419.e1. doi: 10.1016/j.ajem.2011.10.017. [DOI] [PubMed] [Google Scholar]
- 10.Ciozda W, Kedan I, Kehl DW, Zimmer R, Khandwalla R, Kimchi A. The efficacy of sonographic measurement of inferior vena cava diameter as an estimate of central venous pressure. Cardiovasc Ultrasound. 2016;14:33. doi: 10.1186/s12947-016-0076-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Koratala A, Kazory A. Point of Care Ultrasonography for Objective Assessment of Heart Failure: Integration of Cardiac, Vascular, and Extravascular Determinants of Volume Status. Cardiorenal Med. 2021;11:5–17. doi: 10.1159/000510732. [DOI] [PubMed] [Google Scholar]
- 12.Davison R, Cannon R. Estimation of central venous pressure by examination of jugular veins. Am Heart J. 1974;87:279–282. doi: 10.1016/0002-8703(74)90064-7. [DOI] [PubMed] [Google Scholar]
- 13.Nik Muhamad NA, Safferi RS, Robertson CE. Internal Jugular Vein Height and Inferior Vena Cava Diameter Measurement using Ultrasound to Determine Central Venous Pressure: A Correlation Study. Med J Malaysia. 2015;70:63–66. [PubMed] [Google Scholar]
- 14.Lipton B. Estimation of central venous pressure by ultrasound of the internal jugular vein. Am J Emerg Med. 2000;18:432–434. doi: 10.1053/ajem.2000.7335. [DOI] [PubMed] [Google Scholar]
- 15.Siva B, Hunt A, Boudville N. The sensitivity and specificity of ultrasound estimation of central venous pressure using the internal jugular vein. J Crit Care. 2012;27:315.e7–315.11. doi: 10.1016/j.jcrc.2011.09.008. [DOI] [PubMed] [Google Scholar]
- 16.Jenkins S, Knowles P, Briffa N. Portable ultrasound assessment of jugular venous pressure is an accurate method for estimating volaemic status in patients with cardiac disease. J Ultrasound. 2022;25:939–947. doi: 10.1007/s40477-022-00654-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Prekker ME, Scott NL, Hart D, Sprenkle MD, Leatherman JW. Point-of-care ultrasound to estimate central venous pressure: a comparison of three techniques. Crit Care Med. 2013;41:833–841. doi: 10.1097/CCM.0b013e31827466b7. [DOI] [PubMed] [Google Scholar]
- 18.Parenti N, Bastiani L, Tripolino C, Bacchilega I. Ultrasound imaging and central venous pressure in spontaneously breathing patients: a comparison of ultrasound-based measures of internal jugular vein and inferior vena cava. Anaesthesiol Intensive Ther. 2022;54:150–155. doi: 10.5114/ait.2022.114469. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Unluer EE, Kara PH. Ultrasonography of jugular vein as a marker of hypovolemia in healthy volunteers. Am J Emerg Med. 2013;31:173–177. doi: 10.1016/j.ajem.2012.07.003. [DOI] [PubMed] [Google Scholar]
- 20.Wang MK, Piticaru J, Kappel C, Mikhaeil M, Mbuagbaw L, Rochwerg B. Internal jugular vein ultrasound for the diagnosis of hypovolemia and hypervolemia in acutely ill adults: a systematic review and meta-analysis. Intern Emerg Med. 2022;17:1521–1532. doi: 10.1007/s11739-022-03003-y. [DOI] [PubMed] [Google Scholar]
- 21.Leal-Villarreal MAJ, Aguirre-Villarreal D, Vidal-Mayo JJ, Argaiz ER, García-Juárez I. Correlation of Internal Jugular Vein Collapsibility With Central Venous Pressure in Patients With Liver Cirrhosis. Am J Gastroenterol. 2023;118:1684–1687. doi: 10.14309/ajg.0000000000002315. [DOI] [PubMed] [Google Scholar]
- 22.Vaidya GN, Kolodziej A, Stoner B, Galaviz JV, Cao X, Heier K, Thompson M, Birks E, Campbell K. Bedside ultrasound of the internal jugular vein to assess fluid status and right ventricular function: The POCUS-JVD study. Am J Emerg Med. 2023;70:151–156. doi: 10.1016/j.ajem.2023.05.042. [DOI] [PubMed] [Google Scholar]
- 23.Guarracino F, Ferro B, Forfori F, Bertini P, Magliacano L, Pinsky MR. Jugular vein distensibility predicts fluid responsiveness in septic patients. Crit Care. 2014;18:647. doi: 10.1186/s13054-014-0647-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Parenti N, Scalese M, Palazzi C, Agrusta F, Cahill J, Agnelli G. Role of Internal Jugular Vein Ultrasound Measurements in the Assessment of Central Venous Pressure in Spontaneously Breathing Patients: A Systematic Review. J Acute Med. 2019;9:39–48. doi: 10.6705/j.jacme.201906_9(2).0001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Simon MA, Kliner DE, Girod JP, Moguillansky D, Villanueva FS, Pacella JJ. Detection of elevated right atrial pressure using a simple bedside ultrasound measure. Am Heart J. 2010;159:421–427. doi: 10.1016/j.ahj.2010.01.004. [DOI] [PubMed] [Google Scholar]
- 26.Pellicori P, Shah P, Cuthbert J, Urbinati A, Zhang J, Kallvikbacka-Bennett A, Clark AL, Cleland JGF. Prevalence, pattern and clinical relevance of ultrasound indices of congestion in outpatients with heart failure. Eur J Heart Fail. 2019;21:904–916. doi: 10.1002/ejhf.1383. [DOI] [PubMed] [Google Scholar]
- 27.Istrail L, Kiernan J, Stepanova M. A Novel Method for Estimating Right Atrial Pressure With Point-of-Care Ultrasound. J Am Soc Echocardiogr. 2023;36:278–283. doi: 10.1016/j.echo.2022.12.008. [DOI] [PubMed] [Google Scholar]
- 28.Xing CY, Liu YL, Zhao ML, Yang RJ, Duan YY, Zhang LH, Sun XD, Yuan LJ, Cao TS. New method for noninvasive quantification of central venous pressure by ultrasound. Circ Cardiovasc Imaging. 2015;8:e003085. doi: 10.1161/CIRCIMAGING.114.003085. [DOI] [PubMed] [Google Scholar]
- 29.Kovacs G, Avian A, Olschewski A, Olschewski H. Zero reference level for right heart catheterisation. Eur Respir J. 2013;42:1586–1594. doi: 10.1183/09031936.00050713. [DOI] [PubMed] [Google Scholar]
- 30.Deol GR, Collett N, Ashby A, Schmidt GA. Ultrasound accurately reflects the jugular venous examination but underestimates central venous pressure. Chest. 2011;139:95–100. doi: 10.1378/chest.10-1301. [DOI] [PubMed] [Google Scholar]
- 31.Mondal H, Mondal S, Das D, Alam N, Saha K. Reliability of a Jugular Venous Pressure Scale in Cardiovascular Clinical Examination. J Clin Prev Cardiol. 2020;9:61–66. [Google Scholar]
- 32.Hilbert T, Ellerkmann RK, Klaschik S, Putensen C, Thudium M. The Use of Internal Jugular Vein Ultrasonography to Anticipate Low or High Central Venous Pressure During Mechanical Ventilation. J Emerg Med. 2016;50:581–587. doi: 10.1016/j.jemermed.2015.11.033. [DOI] [PubMed] [Google Scholar]
- 33.Koratala A, Reisinger N. Point of Care Ultrasound in Cirrhosis-Associated Acute Kidney Injury: Beyond Inferior Vena Cava. Kidney360. 2022;3:1965–1968. doi: 10.34067/KID.0005522022. [DOI] [PMC free article] [PubMed] [Google Scholar]