Abstract
Purpose:
High-resolution fMRI at 7T is challenged by suboptimal alignment quality between functional data and structural scans. This study aims to develop a rapid acquisition method that provides distortion-matched, artifact-mitigated structural reference data.
Methods:
We introduce an efficient sequence protocol termed T1234, which offers adjustable distortions. This approach involves a T1-weighted 2-inversion 3D-EPI sequence with four spatial encoding directions optimized for high-resolution fMRI. A forward Bloch model was used for T1 quantification and protocol optimization. Twenty participants were scanned at 7T using both structural and functional protocols to evaluate the utility of T1234.
Results:
Results from two protocols are presented. A fast distortion-free protocol reliably produced whole-brain segmentations at 0.8mm isotropic resolution within 3:00–3:40 minutes. It demonstrates robustness across sessions, participants, and three different 7T SIEMENS scanners. For a protocol with geometric distortions that matched functional data, T1234 facilitates layer-specific fMRI signal analysis with enhanced laminar precision.
Conclusion:
This structural mapping approach enables precise registration with fMRI data. T1234 has been successfully implemented, validated, and tested, and is now available to users at our center and at over 50 centers worldwide.
Keywords: layer-fMRI, 7T acquisition, distortion, T1 mapping
1.). Introduction
High-resolution fMRI promises to reveal fine-scale neural information flow across layers and columns. However, common approaches of high-resolution fMRI with Cartesian EPI are limited by EPI phase artifacts dominantly caused from eddy-current induced trajectory imperfections and B0-related geometrical distortions. A survey by the ISMRM brain function study group found that the most limiting factor of the field of high-resolution fMRI is image registration1. This is despite the fact that researchers commonly invest significant time (≈10 min) in acquiring high-quality anatomical data with MP2RAGE2.
In this study, we build on previous works3–12 by developing, implementing, characterizing and validating a novel acquisition method: T1234, involving T1-weighted acquisition with 2 inversions using a 3-dimensional EPI readout in 4 directions. This sequence provides structural reference images with high anatomical contrast, adjustable geometrical distortions, and mitigated artifacts in 3–4 minutes.
1.1). Historical heritage of T1-EPI
Since the advent of EPI-based fMRI13–14, geometric distortions have been a recognized challenge, and have subsequently been addressed through dewarping techniques15. T1-weighted EPI protocols were introduced around the same time16–18. With 7T fMRI at higher resolution, Huber et al3 used T1-weighted EPI in first attempts to estimate voxels of upper and deeper layers in the distorted functional space of constraint fMRI slabs. High-resolution T1-EPI was then combined with more advanced strategies, including slice-wise inversion time cycling for broader 2D-EPI coverage5–6, simultaneous multi-slice readouts9, 3D-EPI, and segmented readouts11,8. Van der Zwaag et al. named their approach T123, referring to T1-imaging with two 3D-EPIs. Further developments have combined advanced 3D-EPI techniques with alternative MT-preparation methods4,10. These advancements, achieving sub-millimeter resolutions on 7T scanners, were driven by the goal of facilitating mesoscale layer-fMRI analyses. However, none of these approaches has yet gained widespread adoption. We suspect that this is largely due to the inherent noise in EPI, which typically demands long acquisition times and averaging. Moreover, some noise sources stem from consistent EPI trajectory imperfections, which cannot be mitigated through averaging alone.
Our work here builds on the method by Stirnberg et al8. We further optimize this technique by integrating multi-phase encoding directions10, interleaved dual-polarity readouts19–20, CNR optimizations by means of TI-specific flip angles that are larger than in conventional Look-Locker approaches, and we combined it with a forward Bloch model16 for T1 quantification. This sequence is tested for two protocols. A fast distortion-free whole brain protocol that can serve as a layer-fMRI optimized alternative to conventional MP2RAGE. And a distortion-matched protocol that allows layer segmentation in the functional data without spatial warping.
2.). Methods
We tested the T1234 sequence in 20 MRI sessions across three 7T scanners: two SIEMENS Terra scanners and one SIEMENS 7T Plus scanner. MRI data were acquired under the NIH-IRB (93-M-0170, ClinicalTrials.gov: NCT00001360).
2.1. Acquisition
Key protocol parameters included: 0.8mm isotropic resolution, Skipped-CAIPI 1×3z1, segmentation factor of 147–8, matrix size of 232×232×186, and a phase partial Fourier factor of 6/8. We utilized a relatively high segmentation factor to minimize distortions and T2*-blurring. For a complete list of parameters, see: https://github.com/layerfMRI/Sequence_Github/tree/master/T1234. The loop structure of this sequence8,21 is illustrated in Figure 1 for the proposed protocol.
A single volume of a bias-field corrected T1-weighted signal was reconstructed from pairs of inversion times (comparable to the UNI-image in MP2RAGE, Equation 3, of the article by Marques et al.2). In addition to the increased SNR achieved through averaging, the following corrections were implemented:
The use of two read directions helps to mitigate typical EPI low spatial frequency artifacts, known as “Fuzzy Ripples,” which result from imperfections in k-space trajectories and off-resonance effects19,12.
Two images acquired in reverse phase encoding directions have opposite distortions. The adjustable segmentation level and phase bandwidth in T1234 can be matched to the effective echo spacing of the functional data, thereby aligning distortions at the acquisition level. Alternatively, with distortions in opposite directions, researchers can estimate the distortion field and retrospectively synthesize any desired level of geometric warping, similar to the topup method22, here implemented in AFNI.
In five participants, in addition to the T1234 acquisition, we also tested the acquisition using a single phase encoding direction while keeping the effective echo spacing identical to that of a conventional fMRI protocol for 7T at standard resolutions23.
Additionally, we conducted four functional experiments to assess the usability of this tool for functional T1 mapping. These included three 12-minute runs of checkerboard stimulation with four inversion times (TIs) and IR-TRs (Figure 1A) of 3.8 seconds, and a 40-second TR with two TIs.
2.2. Reconstruction and Analysis
Complex-valued averaging was performed after parallel image reconstruction and coil combination using IcePat24. Nyquist ghost correction was carried out using the vendor’s standard two-parameter implementation25,-27. The same navigator and GRAPPA reference data were used for both EPI read polarities. For each phase encoding direction polarity, a different set of phase navigators and GRAPPA reference data was acquired and used.
Motion correction of complex-valued image space data was performed using ANTS28. Layers were extracted using the Layer-Toolbox framework (Barilari: https://github.com/marcobarilari/layerfMRI-toolbox) with LayNii v2.6.029, FreeSurfer30 v7.4.1, and PreSurfer 1f5ad28 as a wrapper for SPM12 tools31.
2.3. Analysis of T1 estimation
To predict the optimal distribution of variable flip angles across the inversion-recovery evolution, based on the employed loop structure (Figure 1), we implemented a forward Bloch simulation. This simulation allowed us to determine the excitation flip angles that provide the maximum contrast-to-noise ratio (CNR) between gray matter and white matter (Figure 2). The Bloch solver was also used to generate a lookup table (Figure 3A) that converts relative signal changes across inversion times to physical units of T1 in milliseconds. Given that the protocol used here employs relatively large flip angles (>10°), a simple Look-Locker analysis approach that assumes small flip angels is not justified. The spin-history of each readout alters the T1-relaxation curve depending on local B1 inhomogeneity-dependent excitation flip angles. As a first-order approximation of locally dependent Mz-saturation, we utilized a universal low-resolution template B1 map, previously obtained across the cortex. Simulation code and analysis scripts are available at: https://github.com/layerfMRI/repository/tree/master/T1234.
3.). Results
Figure 2 depicts representative results from one participant. The signal maps from the first inversion time show strong gray matter-white matter (GM-WM) contrast and good signal-to-noise ratio (SNR) with minimal graininess. However, these images also display conventional EPI artifacts, including geometric distortions and “fuzzy ripples” (Figure 2A). By combining data from multiple read and phase polarity directions, these artifacts can be mitigated, making the images suitable for standard tissue segmentation (Figure 2B).
Fig. 3A illustrates how T1234 data can be used to estimate quantitative T1-values. This is done by using a lookup table derived from a forward Bloch model to convert measured signal ratios of TI1 and TI2 into T1 values in milliseconds. The resulting T1 values are compared with those derived from MP2RAGE (a standard in the field) as maps (Figure 3B) and as histograms (Figure 3C). Slight deviations might arise due to unmodeled MT effects during the readout in either sequence, potentially incomplete inversion efficiency, imperfections in the MP2RAGE T1 model for relatively short TRs (4.5s), or inaccuracies in flip angle estimation in the T1234 Bloch model based on the universal B1 distribution approach used here.
Quantitative T1 mapping can be valuable for calibrating geometric cortical layers32 to myelo-architectonic layers33, using the shape of quantitative T1 profiles as anatomical landmarks. This is exemplified in Figure 3D with a modified slab-selective high-resolution T1234 protocol at 0.47mm isotropic.
We found that T1234 is highly robust across participants, sessions, scanners (Figure 4A), and runs (Figure 4B).
To demonstrate the value of the distortion-matched T1234 sequence for layer-fMRI research, we compared laminar fMRI signal pooling using T1234-derived tissue borders with signal pooling using conventionally derived MP2RAGE borders. Fig. 5 shows that laminar signal pooling from distortion-matched T1234-derived layer segmentation retains spatial signal modulation across cortical depth with higher spatial fidelity.
4). Discussion
4.1). Importance of high-precision tissue type segmentation
Current high-resolution fMRI studies are often constrained by the spatial accuracy of tissue segmentation. Previous research has reported that manual correction of automatic and semi-automatic segmentation approaches require extensive labor34,35. This bottleneck in high-resolution fMRI has been recognized by the ISMRM Brain Function Study Group as the most significant challenge in mesoscale fMRI1. Our T1234 method has the potential to overcome this challenge by achieving high precision without the residual constraints and artifacts associated with other EPI-based structural mapping protocols.
4.2). Future work of T1234
4.2.1. Optimizing this general T1234 protocol for special cases.
Our T1234 approach is optimized for applications in layer-fMRI and whole-brain coverage. While this protocol facilitates straightforward use of FreeSurfer pipelines, it may be less efficient for applications focused solely on individual brain areas. For such cases, optimizing the acquisition protocol for higher resolutions and smaller field-of-views (FOV) prescriptions, as demonstrated in Fig 3D, might be more efficient.
4.2.3. Limitations of universal B1 calibration
The protocol used here enables quantitative T1 estimation in physical units of milliseconds, assuming a universal B1+ pattern. This method has been effective for B1+ shimming in previous studies36, but it may not fully account for participant-specific local deviations from the representative mean. While this approach is sufficient for anatomical landmark identification, as demonstrated in Figure 3D, applications requiring more precise T1 quantification might benefit from acquiring a participant-specific B1 map. This can typically be done in under a minute.
4.2.3. Limitations of EPI
The high segmentation factor proposed in this protocol allows for relatively short echo train lengths of 14, with an image echo time of 8ms. Although this is significantly shorter than conventional EPI readouts, it remains longer than those used in conventional FLASH-based structural protocols. As a result, the T1234 approach is more susceptible to fast signal decay induced by susceptibility gradients, particularly in brain regions with significant B0 inhomogeneity near air cavities.
5). Summary and conclusion
EPI-based acquisition of structural reference data has the potential to address registration challenges in high-resolution fMRI studies. However, except for VASO studies, where functional scans double as structural scans, such approaches have not gained widespread adoption. This is largely due to the long acquisition times and EPI artifacts associated with functional scanning, which diminish the appeal of using these methods for structural reference. In this study, we build on previous EPI protocols for structural imaging and propose a multi-directional EPI readout approach that addresses two critical limitations: distortion and artifact levels. Our protocol offers a fast, whole-brain scan in just 3–4 minutes and has proven to be robust for tissue type segmentation across different participants and scanners. Additionally, it has been found to improve cortical depth estimation in layer-fMRI studies. This sequence is available for sharing on 7T SIEMENS scanners and is currently being utilized by over 50 sites worldwide.
8). Diversity statement.
Recent work in several fields of science has identified a bias in citation practices such that papers from women and other minorities are under-cited relative to the number of such papers in the field37. In the human layer-fMRI community the average of the gender citation bias is 84% male, 16% female (https://layerfmri.com/papers/). We obtained the gender of the first author of each reference. By this measure (and excluding self-citations to all authors of our current paper), our references contain 83% male first and 17% female first. This method is limited in that: (i) names, pronouns, and social media profiles used to construct the databases may not, in every case, be indicative of gender identity, and (ii) it cannot account for intersex, non-binary, or transgender people. We look forward to future work that could help us to better understand how to support equitable practices in science.
6). Acknowledgements
Funding:
This work was is supported by the NIH Intramural Program of NIMH/NINDS (#ZIC MH002884). MRI scanning was performed in the FMRIF core.
Support:
We thank Hoan Le and Vinai Roopchansingh for help with computing hardware used for the simulation work conducted here.
Footnotes
Conflict of interest: Omer Faruk Gulban is an employee of Brain Innovation (Maastricht, NL). The work presented here may be partly specific to industrial design choices of SIEMENS Healthineers’ UHF scanners. This vendor is used in 83% of all human layer-fMRI papers (source: www.layerfmri.com/papers).
Ethics: MRI data were acquired under the NIH-IRB (93-M-0170, ClinicalTrials.gov: NCT00001360). We thank Shruti Japee for guidance and support with respect to getting privileges for checking pregnancy tests and IRB.
7). Data Availability Statement
Data of all participants can be found on Zenodo as two parts: https://doi.org/10.5281/zenodo.13366784 and https://doi.org/10.5281/zenodo.13376563
Analysis scripts are available here: https://github.com/layerfMRI/repository/tree/master/T1234
Scan protocols are available here: https://github.com/layerfMRI/Sequence_Github/tree/master/T1234
The T1234 sequence with on-scanner reconstruction of coil-specific dual-polarity combination can be accessed via the Siemens C2P exchange platform: For European IP addresses https://webclient.eu.api.teamplay.siemens-healthineers.com/c2p and for US IP addresses https://webclient.us.api.teamplay.siemens-healthineers.com/c2p. Search for 3D-EPI VASO (by Stirnberg and Huber) provided by DZNE.
9) References
- 1.ISMRM SG CIBF L., 2018. ISMRM Study Group Current Issues in Brain Function Survey results: biggest challenges of high-resolution fMRI, in: Proc. Intl. Soc. Mag. Res. Med. 10.7490/f1000research.1115658.1 [DOI] [Google Scholar]
- 2.Marques J.P., Kober T., Krueger G., van der Zwaag W., Van de Moortele P.-F., Gruetter R., MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. NeuroImage 2010;49(2):1271–1281. 10.1016/j.neuroimage.2009.10.002 [DOI] [PubMed] [Google Scholar]
- 3.Huber L., Ivanov D., Krieger S.N., Streicher M.N., Mildner T., Poser B.A., Möller H.E., Turner R., Slab-Selective BOLD -corrected VASO at 7 Tesla provides measures of cerebral blood volume reactivity with high signal-to-noise ratio. Magnetic Resonance in Medicine 2014;72(1):137–148. 10.1002/mrm.24916 [DOI] [PubMed] [Google Scholar]
- 4.Chai Y., Li L., Wang Y., Huber L., Poser B.A., Duyn J., Bandettini P.A., Magnetization transfer weighted EPI facilitates cortical depth determination in native fMRI space. NeuroImage 2021;242:118455. 10.1016/j.neuroimage.2021.118455 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Kashyap S., Ivanov D., Havlicek M., Poser B.A., Uludağ K., Impact of acquisition and analysis strategies on cortical depth-dependent fMRI. NeuroImage 2018;168:332–344. 10.1016/j.neuroimage.2017.05.022 [DOI] [PubMed] [Google Scholar]
- 6.Renvall V., Witzel T., Wald L.L., Polimeni J.R., Automatic cortical surface reconstruction of high-resolution T1 echo planar imaging data. NeuroImage 2016;134:338–354. 10.1016/j.neuroimage.2016.04.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stirnberg R., Stöcker T., Segmented K-space blipped-controlled aliasing in parallel imaging for high spatiotemporal resolution EPI. Magnetic Resonance in Medicine 2021;85(3):1540–1551. 10.1002/mrm.28486 [DOI] [PubMed] [Google Scholar]
- 8.Stirnberg R., Dong Y., Bause J., Ehses P., Stöcker T., T1 Mapping at 7T Using a Novel Inversion-Recovery Look_Locker 3D-EPI Sequence. Proceedings of the International Society of Magnetic Resonance in Medicine 2018 [Google Scholar]
- 9.Sanchez Panchuelo R.M., Mougin O., Turner R., Francis S.T., Quantitative T1 mapping using multi-slice multi-shot inversion recovery EPI. NeuroImage 2021;234:117976. 10.1016/j.neuroimage.2021.117976 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Malekian V., Graedel N.N., Hickling A., Aghaeifar A., Dymerska B., Corbin N., Josephs O., Maguire E., Callaghan M., Mitigating susceptibility-induced distortions in high-resolution 3D EPI fMRI at 7T. NeuroImage 2023;279:120294. 10.1016/j.neuroimage.2023.120294 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Van der Zwaag W., Buur P.F., Fracasso A., van Doesum T., Uludağ K., Versluis M.J., Marques J.P., Distortion-Matched T1 maps and unbiased T1-weighted images as anatomical reference for high-resolution fMRI. NeuroImage 2018;176:41–55. 10.1016/j.neuroimage.2018.04.026 [DOI] [PubMed] [Google Scholar]
- 12.Huber L., Stirnberg R., Feinberg D., Ma S.J., Ehses P., Gulban O.F., Polimeni J.R., Koiso K., Ma E., Beckett A.JS., Stöcker T., Bandettini P.A., Poser B.A., Low spatial-frequency ripple artifacts in layer-fMRI EPI: Identification, cause, and mitigation strategies with Dual-polarity readout. BioRxiv. 10.1101/2024.09.04.611294. [DOI] [Google Scholar]
- 13.Bandettini P.A., Wong E.C., Hinks R.S., Tikofsky R.S., Hyde J.S., 1992. Time course EPI of human brain function during task activation. Magnetic Resonance in Medicine 25, 390–397. 10.1002/mrm.1910250220 [DOI] [PubMed] [Google Scholar]
- 14.Kwong K.K., Belliveau J.W., Chesler D.A., Goldberg I.E., Weisskoff R.M., Poncelet B.P., Kennedy D.N., Hoppel B.E., Cohen M.S., Turner R., 1992. Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation. Proceedings of the National Academy of Sciences 89, 5675–5679. 10.1073/pnas.89.12.5675 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Jezzard P., 2012. Correction of geometric distortion in fMRI data. NeuroImage 62, 648–651. 10.1016/j.neuroimage.2011.09.010 [DOI] [PubMed] [Google Scholar]
- 16.Ordidge R.J., Gibbs P., Chapman B., Stehling M.K., Mansfield P., 1990. High‐speed multislice T 1 mapping using inversion‐recovery echo‐planar imaging. Magnetic Resonance in Med 16, 238–245. 10.1002/mrm.1910160205 [DOI] [PubMed] [Google Scholar]
- 17.Gowland P., Mansfield P., 1993. Accurate measurement of T 1 in vivo in less than 3 seconds using echo‐planar imaging. Magnetic Resonance in Med 30, 351–354. 10.1002/mrm.1910300312 [DOI] [PubMed] [Google Scholar]
- 18.Stehling M.K., Ordidge R.J., Coxon R., Mansfield P., 1990. Inversion‐recovery echo‐planar imaging (ir‐epi) at 0.5 T. Magnetic Resonance in Med 13, 514–517. 10.1002/mrm.1910130320 [DOI] [PubMed] [Google Scholar]
- 19.Stirnberg R., Deistung A., Reichenbach J.R., Breteler M.M.B., Stöcker T., 2024. Rapid submillimeter QSM and R 2 * mapping using interleaved multishot 3D‐EPI at 7 and 3 Tesla. Magnetic Resonance in Med mrm.30216. 10.1002/mrm.30216 [DOI] [PubMed] [Google Scholar]
- 20.Van der Zwaag W., Marques J.P., Lei H., Just N., Kober T., Gruetter R., 2009. Minimization of Nyquist ghosting for echo-planar imaging at ultra-high fields based on a “negative readout gradient” strategy. J. Magn. Reson. Imaging 30, 1171–1178. 10.1002/jmri.21951 [DOI] [PubMed] [Google Scholar]
- 21.Huber L., Finn E.S., Chai Y., Goebel R., Stirnberg R., Stöcker T., Marrett S., Uludag K., Kim S.G., Han S.H., Bandettini P.A., Poser B.A., 2021. Layer-dependent functional connectivity methods. Progress in Neurobiology 207. 10.1016/j.pneurobio.2020.101835 [DOI] [PubMed] [Google Scholar]
- 22.Andersson J.L.R., Skare S., Ashburner J., How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. NeuroImage, 20(2):870–888, 2003. [DOI] [PubMed] [Google Scholar]
- 23.Clarke W.T., Mougin O., Driver I.D., Rua C., Morgan A.T., Asghar M., Clare S., Francis S., Wise R.G., Rodgers C.T., Carpenter A., Muir K., Bowtell R., Multi-site harmonization of 7 tesla MRI neuroimaging protocols. NeuroImage 2020;206:116335. 10.1016/j.neuroimage.2019.116335 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Jellúš V. and Kannengießer S. A. R. (2014). Improved Coil Combination using the Prescan Normalize Adjustment. In SIEMENS Idea meeting 2014, page 4406 [Google Scholar]
- 25.Ahn C B, and Cho Z H. “A New Phase Correction Method in NMR Imaging Based on Autocorrelation and Histogram Analysis.” IEEE Transactions on Medical Imaging 6, no. 1 (1987): 32–36. 10.1109/TMI.1987.4307795. [DOI] [PubMed] [Google Scholar]
- 26.Feiweier T, inventor; Siemens Aktiengesellschaft, assignee. Magnetic resonance method and apparatus to determine phase correction parameters. US Patent 8,497,681. July 30, 2013.
- 27.Heid Oliver. Method for the phase correction of nuclear magnetic resonance signals, Patent: US006043651A 2000.
- 28.Avants Brian B., Tustison Nicholas J., Stauffer Michael, Song Gang, Wu Baohua, and Gee James C.. 2014. “The Insight ToolKit Image Registration Framework.” Frontiers in Neuroinformatics 8. doi: 10.3389/fninf.2014.00044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Huber L., Poser B.A., Bandettini P.A., Arora K., Wagstyl K., Cho S., Goense J., Nothnagel N., Morgan A.T., van den Hurk J., Müller A.K., Reynolds R.C., Glen D.R., Goebel R., Gulban O.F., LayNii: A software suite for layer-fMRI. NeuroImage 2021;237:118091. 10.1016/j.neuroimage.2021.118091 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Fischl B., 2012. FreeSurfer. NeuroImage 62, 774–781. 10.1016/j.neuroimage.2012.01.021 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Ashburner J., 2012. SPM: A history. NeuroImage 62, 791–800. 10.1016/j.neuroimage.2011.10.025 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Geyer S., Turner R., 2011. Microstructural parcellation of the human cerebral cortex: From Brodmann’s post-mortem map to in vivo mapping with high-field magnetic resonance imaging. Springer; Berlin Heidelberg. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Gulban O.F., Huber R., 2024. Computing geometric layers and columns on continuously improving human (f)MRI data, in: Reference Module in Neuroscience and Biobehavioral Psychology. Elsevier, p. B9780128204801001881. 10.1016/B978-0-12-820480-1.00188-1 [DOI] [Google Scholar]
- 34.Koiso K., Müller A.K., Akamatsu K., Dresbach S., Gulban O.F., Goebel R., Miyawaki Y., Poser B.A., Huber L., 2023. Acquisition and processing methods of whole-brain layer-fMRI VASO and BOLD: The Kenshu dataset. Aperture Neuro 34. 10.1101/2022.08.19.504502. [DOI] [Google Scholar]
- 35.Gulban O.F., Schneider M., Marquardt I., Haast R.A.M., De Martino F., 2018. A scalable method to improve gray matter segmentation at ultra high field MRI. Plos One 13, e0198335. 10.1371/journal.pone.0198335 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Gras V., Vignaud A., Amadon A., Le Bihan D., Boulant N., 2017. Universal pulses: A new concept for calibration‐free parallel transmission. Magnetic Resonance in Med 77, 635–643. 10.1002/mrm.26148 [DOI] [PubMed] [Google Scholar]
- 37.Dworkin J.D., Linn K.A., Teich E.G., Zurn P., Shinohara R.T., Bassett D.S., 2020. The extent and drivers of gender imbalance in neuroscience reference lists. Nature Neuroscience 23, 918–926. 10.1038/s41593-020-0658-y [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data of all participants can be found on Zenodo as two parts: https://doi.org/10.5281/zenodo.13366784 and https://doi.org/10.5281/zenodo.13376563
Analysis scripts are available here: https://github.com/layerfMRI/repository/tree/master/T1234
Scan protocols are available here: https://github.com/layerfMRI/Sequence_Github/tree/master/T1234
The T1234 sequence with on-scanner reconstruction of coil-specific dual-polarity combination can be accessed via the Siemens C2P exchange platform: For European IP addresses https://webclient.eu.api.teamplay.siemens-healthineers.com/c2p and for US IP addresses https://webclient.us.api.teamplay.siemens-healthineers.com/c2p. Search for 3D-EPI VASO (by Stirnberg and Huber) provided by DZNE.