Abstract
Current methods for single-molecule orientation localization microscopy (SMOLM) require optical setups and algorithms that can be prohibitively slow and complex, limiting widespread adoption for biological applications. We present POLCAM, a simplified SMOLM method based on polarized detection using a polarization camera, which can be easily implemented on any wide-field fluorescence microscope. To make polarization cameras compatible with single-molecule detection, we developed theory to minimize field-of-view errors, used simulations to optimize experimental design and developed a fast algorithm based on Stokes parameter estimation that can operate over 1,000-fold faster than the state of the art, enabling near-instant determination of molecular anisotropy. To aid in the adoption of POLCAM, we developed open-source image analysis software and a website detailing hardware installation and software use. To illustrate the potential of POLCAM in the life sciences, we applied our method to study α-synuclein fibrils, the actin cytoskeleton of mammalian cells, fibroblast-like cells and the plasma membrane of live human T cells.
Subject terms: Single-molecule biophysics, Microscopy, Fluorescence imaging
Combining localization and polarization microscopy can yield detailed insights into subcellular structures. POLCAM uses a polarization camera and wide-field microscopy for rapid measurement of super-resolution orientation imaging in live cells.
Main
Single-molecule localization microscopy (SMLM)1–4 is a super-resolution microscopy technique that is widely used in biology to study cellular structures below the diffraction limit5–7. Single-molecule orientation localization microscopy (SMOLM) is a multidimensional variant of SMLM in which, in addition to the precise spatial position, the orientation of individual fluorescent molecules is also measured. The ability to measure the orientation of single molecules provides information about how molecules organize, orient, rotate and wobble in their environment, which is of key relevance across biological systems8–13. The widespread use of SMOLM by the biological imaging community has thus far been limited by the need for complex experimental setups and often computationally expensive image analysis. Additionally, this lack of accessibility has also slowed down the necessary development of a wider range of labeling protocols that are appropriate for SMOLM: labeling methods in which the orientation of the fluorescent probe is relatively fixed and rotationally restricted with respect to its target14–17.
Fluorescent molecules are not isotropic point sources, that is, they do not emit light equally in all directions. Fundamentally, fluorescent molecules emit like oscillating electric dipoles: the intensity I of the emitted fluorescence depends on the relative observation direction and follows the relationship , where η is the angle between the observation direction and the orientation of the emission dipole moment of the molecule18. In conventional SMLM experiments, this anisotropic emission is typically not noticeable because, with common SMLM labeling protocols, fluorescent molecules are free to rapidly wobble and rotate with respect to their target (for example, due to long linker chains), resulting in an orientation-averaged image19. However, when a labeling method is used that restricts the rotational freedom of the fluorescent molecules with respect to their target, the anisotropy in the emitted fluorescence allows for the measurement of molecular orientation20. Different approaches have been used to achieve molecular orientation imaging. Some methods are based on active modulation of the polarization of the excitation light21–23, but the majority of methods are based on modification of the detection path of the microscope. The image of a single molecule can be fitted using a dipole-spread function (DSF) that includes the position of the molecule (x, y or x, y, z), the orientation of the emission dipole moment (ϕ, θ)24,25 and often a rotational mobility parameter26. As the intensity distribution of a standard DSF does not contain notable information about a molecule’s orientation, the DSF can be engineered to increase the orientation information content. A simple example is imaging slightly out of focus to exaggerate the DSF shape27–29. More advanced DSF engineering can be performed using a spatial light modulator19,30–33, a special optic34–36 or pupil splitting37. A drawback of DSF engineering is that the optical setups required are highly complex (with the exception of the vortex DSF34) and are sensitive to optical aberrations (with the exception of pupil splitting37), as the orientation estimation algorithms rely on simulated DSF models, which can necessitate performing spatially (in)variant phase retrieval to match the DSF model to the experimental DSF34. Additionally, fitting a five- or six-dimensional DSF model is computationally expensive, making data analysis prohibitively slow.
An alternative method is splitting the emission into multiple polarized channels that form separate images on the same camera10,38–44 or multiple detectors45 or using continuous image displacement using a rotating calcite crystal46. The advantage of polarized detection-based methods is that the orientation estimation can be performed using simple intensity measurements, and does not necessarily require the fitting of a complex DSF model. As a result, the data analysis is fast and easily compatible with high-throughput data collection. The simplest polarized detection setup splits the emission into two orthogonal polarized channels using a polarizing beam splitter38,39. This method suffers from some degeneracies because both an isotropic emitter (for example, a freely rotating molecule or fluorescent bead) and an immobilized molecule with an emission dipole moment oriented at 45° in between the transmission axis of the two channels or parallel to the optical axis all give rise to equal intensities measured in both channels8,47. Splitting the emission into three or more polarized channels or cameras breaks this degeneracy15,48,49 but substantially complicates the experimental setup39,41–44.
Here, we present a new experimentally simplified SMOLM method called POLCAM that uses a polarization camera for four-channel polarized detection. Polarization cameras have become popular in the field of computer vision as they provide single-shot multi-channel polarized measurements50–53. The pixels of the sensor of a polarization camera are covered by small linear polarizers with transmission axes typically oriented at 0°, 45°, 90° and −45° in a 2 × 2-pixel mosaic pattern that is repeated over the entire sensor (Fig. 1a). As polarizers are integrated into the camera chip, no additional polarization optics are required. Polarization cameras have reached quantum efficiencies and noise levels that in theory are compatible with single-molecule detection, but the ease of use of polarization cameras does come at a cost in the form of instantaneous field-of-view (IFOV) errors near object edges54–56. To make POLCAM robust to IFOV errors, we used vectorial diffraction simulations of single dipole emitters to optimize microscope design and developed a Stokes parameter estimation-based reconstruction algorithm and a DSF-fitting algorithm.
We validated and characterized our method using samples with known structures: single fluorophores immobilized on coverglass and in polymer and lipid bilayer-coated glass beads labeled with membrane dyes. We next performed SMOLM on amyloid fibrils in vitro and on the actin network of fixed mammalian cells. To demonstrate that a polarization camera can also be used for conventional polarized detection microscopy, we imaged the actin network of COS-7 cells and the membrane of live human T cells interacting with an antibody-coated coverglass.
Our approach can be easily implemented by changing the regular camera on any single-molecule fluorescence microscope to a polarization camera. The camera used in this work is supported by the popular image acquisition software µManager57, and we provide image analysis software in the form of MATLAB applications for single-molecule and diffraction-limited image analysis and real-time image processing and rendering during acquisition and a napari58 plugin for processing multidimensional diffraction-limited polarization camera image datasets. We envisage that the combination of ease of use, ease of implementation, low cost, improved speed and open-source software will make POLCAM an accessible and powerful tool for the study of molecular orientation across diverse biological applications.
Results
Measuring molecular orientation using polarized detection
When a fluorescent molecule has one dominant emission dipole moment μ, as is the case for many common fluorescent molecules, its emission resembles the far field emitted by an oscillating electric dipole18,20 (Fig. 1b,c and Supplementary Notes 1 and 3). When its emission dipole moment is oriented parallel to the sample plane, the electric field in the back focal plane of the objective will mainly be linearly polarized along the direction of the emission dipole moment (Fig. 1c, first and second row). If the emission dipole moment is oriented parallel to the optical axis, the electric field in the back focal plane of the objective will be radially polarized59,60 (Fig. 1c, third row). When a molecule is rapidly rotating, it can appear unpolarized (Fig. 1c, bottom row). As the tube lens in conventional wide-field fluorescence microscopy has a low numerical aperture (NA), the described polarization is mostly conserved in the image plane61. As a result, the angle of the axis of maximum polarization determines the in-plane orientation ϕ of the emission dipole moment, and the degree of net linear polarization is related to the out-of-plane orientation θ (Fig. 1d,e).
Conventionally, polarizing beam splitters are used to split the detected fluorescence into multiple polarized image channels. The number of photons that are detected from a single molecule in the different channels will depend on the three-dimensional (3D) orientation and rotational mobility of the molecule. These measured intensities can be used to estimate the angles ϕ and θ using analytically derived equations. Equations for the case of four polarized detection channels (0°, 45°, 90° and −45°) were derived by John T. Fourkas60. Here, we rewrote these expressions in terms of Stokes parameters (Supplementary Notes 4 and 5):
1 |
2 |
where A, B and C are constants that are a function of the half-maximum collection angle of the objective α (Supplementary Notes 4 and 5) and the net degree of linear polarization (netDoLP) given by
3 |
where the brackets (…) refer to averaging over a small region of interest around a single molecule (Supplementary Note 7) and S0, S1 and S2 are the first three Stokes parameters:
4a |
4b |
4c |
where I0, I45, I90 and I−45 refer to the measured intensities in the four polarized channels. The full derivation of equations (1) and (2) can be found in Supplementary Notes 4 and 5, and example simulated images are in Supplementary Figs. 3–19.
We note that equation (1) is simply the expression for the angle of linear polarization (AoLP)62 and that equation (2) depends only on the netDoLP and the objective used. This is in line with our intuition (Fig. 1c). Equations (1) and (2) are plotted in Fig. 1d,e for a 1.4-NA oil-immersion objective. We note that estimation of ϕ using equation (1) is very robust but that estimation of θ using equation (2) on the other hand is only possible under strict conditions (perfect rotational immobilization, high signal-to-noise ratio and no large refractive index mismatch) and is more reliably performed with DSF fitting as will be discussed in more detail in a later section.
As a proxy for rotational mobility, we use the average degree of linear polarization (avgDoLP), which we define as a local average of the degree of linear polarization (DoLP):
5 |
where m is the number of pixels in the region of interest around the molecule. If a DSF-fitting algorithm is used instead (‘Improving accuracy by considering the DSF shape’), a rotational mobility parameter γ can be estimated that inversely relates to the size of a cone in which the molecule has rotational freedom26: γ is 1 for perfect immobilization and 0 for complete rotational freedom. The exact mathematical relation between avgDoLP and γ is complex, as avgDoLP is also influenced by the signal-to-noise ratio, the out-of-plane angle and the refractive index of the sample medium. The relationship between avgDoLP and rotational mobility is numerically explored in Supplementary Fig. 27, showing that it is monotonic under all conditions and can therefore be used qualitatively but with care. For details, we refer to the Methods and Supplementary Note 7.
Overcoming IFOV errors
In conventional polarization-sensitive fluorescence imaging, where polarizing elements are placed in the optical path, a traditional camera sensor captures the full-intensity distribution everywhere in the image plane. However, similar to the operation of conventional color image sensors63, the polarization camera measures the intensity of each polarization channel in a subset of the image (one in four pixels), and the full-intensity distributions have to be recovered through interpolation54–56. This recovery can be performed accurately provided that the pixel size is small enough. If this is not the case, any measurements taken from the recovered channels will exhibit what is known as IFOV errors54–56. In standard applications of polarization cameras (for example, quality control in the manufacturing industry, removal of reflections in images64), IFOV errors can be ignored or avoided, as the pixel size can be much smaller than variations in neighboring pixels and artifacts mostly appear near the edges of objects. However, when imaging single molecules, the opposite is true, as the image of a single emitter varies substantially over each pixel and the limited photon budget prevents the use of a small pixel size.
We determined the optimal pixel size for the estimation of molecular orientation using vectorial diffraction simulations (Supplementary Note 6). We define the optimal pixel size as the largest pixel size that still allows for accurate recovery of the four polarized channels from a single polarization camera image. To assess whether accurate recovery is possible, we used an approach described by Tyo et al.55 that checks for overlap between the contributions of different Stokes parameters in the Fourier transform of the unprocessed polarization camera image (Supplementary Note 6.1 and Supplementary Figs. 20–25). If the contributions do not overlap, the recovery is assumed to be accurate.
Using this method, we find that, for our setup (1.4-NA oil-immersion objective, wavelength of 650 nm, sample in an aqueous medium), optimal sampling is achieved at a pixel size of ~60 nm × 60 nm (Supplementary Fig. 20). Practically, a pixel size of 57.5 nm × 57.5 nm was achieved using a ×60 magnification objective and the standard polarization camera pixel size of 3.45 μm × 3.45 μm (Supplementary Fig. 30). The calculated ideal pixel size as a function of wavelength, objective NA and sample medium can be found in Supplementary Fig. 20.
Next, we compared the performance of different algorithms54–56 for Stokes parameter estimation and channel interpolation on simulated polarization camera images of immobilized single molecules. We found that a Fourier-based approach55 and cubic spline interpolation performed the best (Supplementary Note 6.2 and Supplementary Figs. 21–25). Fig. 2a shows an experimental polarization camera image of semi-immobilized SYTOX Orange molecules on a coverglass. Fig. 2b shows the results of the Fourier-based interpolation. Fig. 2c shows examples of the emission of single molecules taken from this dataset: two in-plane-oriented molecules (molecules 1 and 2), one molecule oriented out of the plane (molecule 3) and a rapidly rotating molecule (molecule 4). An example SYTOX Orange dataset is included in Supplementary Dataset 1.
Experimental accuracy and precision
To measure the experimental accuracy of orientation estimation, silica microspheres with a diameter of 5 μm were coated with a lipid bilayer (dipalmitoylphosphatidylcholine (DPPC) containing 40% cholesterol; Fig 2d) as described in refs. 33,37 and labeled using different membrane dyes. Polarization-resolved diffraction-limited images were acquired using bulk labeling concentrations of different membrane dyes (Di-8-ANEPPS, DiI and Nile red). As expected, the emission dipole moments of Di-8-ANEPPS and Nile red orient perpendicular to the membrane surface, and DiI orients parallel to the membrane surface (Fig. 2e and Supplementary Fig. 33). Nile red was used to collect a point accumulation for imaging in nanoscale topography (PAINT)4 dataset of a lipid-coated microsphere to generate an experimental accuracy curve for the in-plane angle ϕ (Fig. 2f,g).
Experimental orientation estimation and localization precision curves as a function of the number of detected photons were generated using single Alexa Fluor 647 (AF647) dyes immobilized in poly(vinyl alcohol) (PVA). The data points in Fig. 2i are the measured standard deviation on the repeated measurement of the orientation of an AF647 molecule. At 500 detected photons (the default lower threshold on photon number that is used in all datasets), we achieve an experimental in-plane angle precision σϕ of 7.5° (Fig. 2i), that is, the upper bound on σϕ. The precision converges to a lower bound of 1° at higher photon numbers (Fig. 2i). The experimental localization precision at 500 detected photons is 25–30 nm and converges to 5 nm at high photon numbers (Fig. 2j). The same curves generated using simulations with a realistic noise model largely agree with the experimental data (Supplementary Fig. 35). A complete characterization of the bias and precision on the position, orientation and rotational mobility estimates based on simulations can be found in Supplementary Figs. 36–41. DNA origami with a spacing of 80 nm between binding sites can also be easily resolved using POLCAM (Supplementary Figs. 31 and 32).
TAB-PAINT imaging of α-synuclein fibrils in vitro
Orientationally resolved imaging of α-synuclein fibrils labeled with the dye Nile red has previously been demonstrated using transient amyloid binding PAINT (TAB-PAINT)4,65. Nile red reversibly binds to hydrophobic regions of the fibrils in a defined orientation12,35. Therefore, this serves as an excellent test sample for orientation-resolved super-resolution imaging in biologically relevant samples. Fig. 3 shows the POLCAM reconstruction of the α-synuclein fibrils, which were immobilized on a coverglass using poly-l-lysine (PLL) coating and imaged in phosphate-buffered saline (PBS). POLCAM can super-resolve morphologically consistent α-synuclein fibrils with widths of ~50 nm (full-width at half-maximum (FWHM)) over large fields of views of ~50 μm × 50 μm (Fig. 3a,b and Supplementary Fig. 34). When color coding the reconstructions by the in-plane angle ϕ estimate, we show that the majority of the Nile red molecules orient parallel to the long axis of the fibril (Fig. 3c–e). The distributions of the measured in-plane angle over short fibril sections have standard deviations around 7° and 9° (Fig. 3f), which approaches the expected precision for this dataset (Fig. 2i; ~5° at 1,000 detected photons). This indicates that the width of this distribution cannot be explained by measurement precision alone and must partly be due to Nile red molecules binding in a range of orientations that are not exactly parallel to the fibril axis.
While a range of avgDoLP and photon values was extracted using POLCAM (Fig. 3g,h), we apply lower-limit filtering thresholds to ensure high-accuracy results: (1) at least 500 detected photons (applied to all single-molecule data presented in this work) and (2) avgDoLP > 0.4 when ϕ-color-coded data are shown, as localizations with avgDoLP < 0.4 are too rotationally free to estimate a meaningful orientation. An example TAB-PAINT dataset is included in Supplementary Dataset 2.
dSTORM imaging of actin in fixed HeLa cells
Next, to demonstrate the versatility of POLCAM, we performed SMOLM on eukaryotic cells using direct stochastic optical reconstruction microscopy (dSTORM). Fixed HeLa cells were labeled using phalloidin–Alexa Fluor 488 (AF488) and phalloidin–Alexa Fluor 647 (AF647). In previous literature, this labeling has been shown to result in rotational restriction for AF488 (where the dyes on average align with the axis of actin fibers) and rotational freedom for AF647 (refs. 39,44). Single AF488 and AF647 molecules can be easily detected with POLCAM in the cellular environment, and the difference in rotational mobility between the rotationally constrained AF488 and the randomly oriented AF647 is also directly evident from the images of single molecules (Fig. 4a,b). The difference in rotational mobility between the two labeling approaches is also visible in the resulting super-resolution images (Fig. 4c,d) using a color map in which more orientationally random areas appear white (avgDoLP < 0.4) and the more ordered areas appear colored. Analysis of the distributions of the avgDoLP revealed an ordered subset of localizations (Fig. 4e) (25% for AF488 and only 5% for AF648). By filtering these localizations using this empirically determined threshold (avgDoLP > 0.4), a refined POLCAM image of actin in cells can be generated that excludes localizations that are too rotationally free to generate an accurate ϕ estimate (Fig. 4g). The resolution of the super-resolved images was estimated using Fourier ring correlation (FRC66): 70 nm for the AF488 dataset and 55 nm for the AF647 dataset (Fig. 4f).
Improving accuracy by considering the DSF shape
From equation (2), it is clear that the estimation of the polar angle θ will become biased in the presence of rotational mobility, as netDoLP will decrease with increasing rotational mobility. Additionally, equation (2) becomes biased in the presence of noise and depends on the refractive index of the sample medium. As a result, unbiased estimation of θ is more reliably performed by additionally taking the shape of the DSF into account. We adapted the previously published DSF-fitting algorithm RoSE-O67 for use with a polarization camera. This algorithm fits the shape of the image of a single emitter in all four polarized channels to estimate the orientation and rotational mobility of the emitter. Using simulated images, we compared the performance of the intensity-only algorithm and the DSF-fitting algorithm. As expected, the DSF-fitting algorithm is able to more accurately estimate θ in the presence of rotational mobility (Fig. 5a,b). For in-plane-oriented emitters, both algorithms achieve the same localization precision across a wide range of signal-to-noise ratios, but the DSF-fitting algorithm outperforms the intensity-only algorithm up to twofold for emitters that are oriented out of plane (Fig. 5c). The precision on the estimation of ϕ is similar for in-plane-oriented emitters, but the DSF-fitting algorithm is able to more precisely estimate ϕ for more out-of-plane-oriented emitters, especially at high signal-to-noise ratios (Fig. 5d). A complete comparison of the algorithms can be found in Supplementary Figs. 36–41.
Fig. 5f shows super-resolution reconstructions generated by the two algorithms of a subset (200 × 200 pixels and 10,000 frames and ~15 localizations per frame) of the phalloidin–AF488 dSTORM dataset from Fig. 4g. Although both algorithms generate similar-looking reconstructions, zooming in on specific regions (Fig. 5g) seems to confirm that the DSF-fitting algorithm produces a slightly higher-resolution image. A spatial comparison of the ϕ estimates shows that the widths of the ϕ distributions generated by the DSF-fitting algorithm are slightly more narrow (Supplementary Fig. 41). On a typical workstation (see the Methods for system specifications), the intensity-only algorithm is >750–3,700 times faster (2 minutes for total processing using centroid localization and 10 minutes using least-squares rotated asymmetric Gaussian fitting) than the DSF-fitting algorithm. Therefore, the DSF-fitting algorithm is recommended if the complete 3D orientation (ϕ, θ) needs to be estimated. If knowledge of ϕ and the rotational mobility proxy avgDoLP is sufficient, the fast intensity-only algorithm is recommended.
Live diffraction-limited polarization microscopy
The combination of instrumental simplicity and fast computation makes POLCAM compatible with real-time image processing. We developed napari-polcam (a napari plugin for the open-source multidimensional image viewer napari58 (Fig. 6a)) and a standalone application for on-the-fly processing and rendering called POLCAM-Live (Fig. 6b). Both software take in unprocessed data and convert it into different formats in an easy-to-use interface. The ability to process data live is useful for fast decision making during experiments and alignment (Supplementary Note 8).
We use these software tools to illustrate diffraction-limited polarization imaging, demonstrating 3D time-lapse imaging of the plasma membrane in live human T cells using the new probe NR4A68. The 3D images are acquired by axially scanning the objective. The filopodia of the T cells appear more unpolarized (Fig. 6d–f), allowing their simple identification from cell bodies, which is the subject of intense research interest with regard to surface receptor organization69. The unpolarized appearance of filopodia is likely due to their small diameter. The 3D motion of the filopodia can be tracked over time (~1 minute per volume, limited by the speed of our z stage). Furthermore, POLCAM can be used in a simplified mode to discriminate ordered versus non-ordered actin structures based on the DoLP. Using a simple DoLP color map, we can distinguish highly ordered regions (yellow) from more disordered regions (blue) in the actin network in COS-7 cells labeled with silicon rhodamine-actin (SiR–actin) (Fig. 6c).
Discussion
In this work, we present POLCAM, a new method for molecular orientation-resolved fluorescence microscopy that makes use of a polarization camera to dramatically simplify the experimental setup. The method can be used in two modes, SMOLM or diffraction-limited polarization fluorescence microscopy, and can be implemented by simply replacing the conventional detector on a wide-field fluorescence microscope with a polarization camera. The polarization camera used in this work is supported by the popular image acquisition software µManager57. Furthermore, we provide software to improve widespread user adaption as well as a comprehensive installation guide and a software user manual.
We present two SMOLM analysis algorithms for POLCAM: (1) a fast algorithm based on Stokes parameter estimation and simple intensity measurements for robust in-plane angle estimation compatible with high-throughput data collection and (2) an unbiased 3D orientation estimation algorithm that fits a DSF model. Due to the computational cost of the DSF-fitting algorithm, it is more suited to the analysis of small datasets.
In conventional four-channel polarized detection, where beam splitters and polarization optics are used to separate the fluorescence into four channels, it is more likely that channel-dependent aberrations will occur. Due to the simplicity of the experimental setup of POLCAM, this is avoided, simplifying the use of a DSF-fitting algorithm. Moreover, there is no need for channel registration and localization grouping because single-molecule localization is performed on the estimated incident intensity (S0) on the micropolarizer array, meaning that all in-plane orientations are equally detectable and no localizations are therefore missed. A diffraction-limited polarization color map (HSVmap) can be generated and displayed in real-time during experiments, for which we provide the standalone software POLCAM-Live. In theory, the intensity-only SMOLM algorithm is also fast enough to be compatible with real-time processing, as a 200 × 200-pixel image (10 × 10 μm) with ~15 localizations takes about 10 ms to process, which is less than typical camera exposure times. This could enable molecular orientation event-triggered microscopy70,71, something that would be extremely challenging if DSF fitting is required.
Because a polarization camera uses polarizers to achieve polarized detection, on average, the system is 50% efficient, as half of the photons that are captured by the objective will be absorbed by polarizers. Nevertheless, we demonstrate in this work that this approach is compatible with a wide variety of fluorophores (for example, AF488, Nile red, SiR, Cy5, SYTOX Orange). The current generation of polarization cameras lacks the sophistication of modern scientific complementary metal–oxide–semiconductor (sCMOS) cameras in areas such as the quantum efficiency of the detector (70% versus 95%) and the presence of onboard denoising algorithms. We expect this to improve over time. Further improvements to the analysis approach can be made by performing a pixel-dependent characterization of any defects of the micropolarizer array or deviation from an ideal micropolarizer array72.
The success of a molecular orientation-resolved experiment stands and falls with the labeling approach. It would thus be meaningless to attempt to measure molecular orientation if the relative orientation of a target and the probe is random. Therefore, the use of antibodies with multiple dyes conjugated at random orientations (for example, random lysine labeling) or staining using secondary antibodies is likely not suitable. Currently, the number of labeling protocols that restrict or control the orientation and rotational mobility of a fluorophore with respect to their target is still limited, and there is a strong need for the development of more labeling approaches and the discovery of suitable probes. Examples are the use of bifunctional rhodamines16 and the genetically engineered rigid protein linker POLArIS73. It is likely that many common fluorescent probes are suitable for molecular orientation-resolved microscopy, but they have simply never been evaluated for this purpose. We anticipate that, because of the accessibility and compatibility with high-throughput data acquisition, POLCAM will accelerate this much-needed development and discovery of new probes and expand the current toolkit to cover more biological systems.
We note that polarization cameras can also be used for label-free microscopy74–77, leaving the possibility for multiplexing of fluorescence and label-free techniques. Overall, we envisage that the combination of POLCAM’s simple implementation and ease of use, computational speed and open-source software will lead to new biological insight across diverse systems.
Methods
Optical setups
Experiments were performed on three very similar wide-field fluorescence microscopes. SYTOX Orange, AF647 immobilized in polymer, TAB-PAINT of α-synuclein fibrils and diffraction-limited polarization imaging of the actin network in COS-7 cells were performed on ‘microscope 1’. The dSTORM experiments were performed on ‘microscope 2’. The lipid bilayer-coated silica microsphere and live T cell experiments were performed on ‘microscope 3’.
‘Microscope 1’ is a wide-field fluorescence microscope (Eclipse Ti-U, Nikon), with illumination entering the microscope body through the back illumination port. The beams from three free-space lasers (515 nm, 150 mW, Spectra-Physics; 532 nm, 120 mW, Odic Force Lasers; 638 nm, 350 mW, Odic Force Lasers) were expanded, spectrally and spatially filtered, combined with dichroics and focused to a spot in the back focal plane of an oil-immersion objective (Plan Apo, ×60 NA-1.40 oil, DIC H, ∞/0.17, WD 0.21, Nikon) using an achromatic doublet lens (AC254-300-A, Thorlabs). This lens and a periscope were mounted on a linear translation stage to allow manual adjustment of the beam emerging from the objective and switching between EPI, HILO and TIRF illumination. Approximately circular polarization at the sample plane was achieved using quarter waveplates (WPQ10M-514, WPQ10M-532 and WPQ10M-633, Thorlabs). For diffraction-limited imaging of the actin network in COS-7 cells, a different laser source was used with a multimode fiber (LDI-7 Laser Diode Illuminator, 89 North). Fluorescence was filtered by a dichroic beam splitter (Di03-R405/488/532/635-t1 for AF647 and Di03-R532-t1 for SYTOX Orange and Nile red; Semrock) and emission filters (BLP01-532R for SYTOX Orange, BLP01-635R for AF647, BLP01-532R and FF01-650/200 for Nile red; Semrock). The fluorescence was focused on a polarization camera (CS505MUP, Thorlabs) that was placed directly at the microscope body camera port. The pixel size of the camera is 3.45 μm × 3.45 μm, resulting in a virtual pixel size of 57.5 × 57.5 nm. The microscope PC was a Dell OptiPlex 7070 Mini Tower running on Windows 10 (64 bit) with an Intel Core i9-9900 processor and 32 GB of RAM.
‘Microscope 2’ is functionally similar to ‘microscope 1’, with a different microscope body (Eclipse Ti-E, Nikon) and laser source (Omicron LightHUB with 405-, 488-, 561- and 638-nm lasers, single-mode fiber, collimator RC08APC-P01, Thorlabs) and a ×60 1.42-NA oil-immersion objective (Olympus, PlanApo N). A quadband imaging dichroic (Di03-R405/488/532/635-t1, Semrock) and emission filters were used (BLP01-488R and FF01-582/64 for AF488, BLP01-635R for AF647; Semrock). The fluorescence was focused on a polarization camera (CS505MUP, Thorlabs) that was placed directly at the microscope body camera port, resulting in a virtual pixel size of 51.7 × 51.7 nm. Note that an Olympus objective (assuming use of an Olympus body with a 180.0-mm focal length tube lens) was used with a Nikon body (200.0-mm focal length tube lens).
‘Microscope 3’ is also functionally similar to ‘microscope 1’, other than the microscope body (Eclipse Ti-E, Nikon), lasers (Cobolt C-FLEX combiner with 405-, 488-, 515-, 561- and two 638-nm lasers, free space) coupled into a square-core multimode fiber (M97L02, Thorlabs) with a custom vibration motor-based mode scrambler and a ×100 1.49-NA oil-immersion objective. A 4f system consisting of two achromatic lenses (AC254-050-A-ML and AC254-100-A-ML, Thorlabs) was included in the emission path to demagnify the image 2×, resulting in a total system magnification of ×50 and thus a virtual pixel size of 69 × 69 nm. Imaging dichroics (Di03-R515-t1 for NR4A and Nile red and Di03-R405/488/532/635-t1 for SiR–actin; Semrock) and emission filters (BLP01-532R and FF01-650/200 for NR4A, BLP01-635R for SiR–actin; Semrock) were used. A multimode fiber was used for all diffraction-limited imaging experiments, as this allowed us to achieve highly randomized polarization at the sample plane (compared to using a quarter-wave plate), resulting in negligible photoselection.
Simulations
The emission of a fluorescent molecule was modeled as the far field of an oscillating electric dipole as previously described in ref. 20. Transmission of the emission through the micropolarizer array on the camera sensor was modeled for each camera pixel using a Jones matrix JLP for a linear polarizer with an axis of transmission at an angle η from the x axis78:
6 |
7 |
Unless otherwise specified, the following system parameters were used: emission wavelength of 630 nm, ×60 oil-immersion objective with an NA of 1.4, a tube lens with a focal length of 200.0 mm, physical camera pixel size of 3.45 μm. The molecules are placed on a glass–water refractive index interface (nglass = noil = 1.518, nwater = 1.33) and in focus. A CMOS camera noise model was used (Supplementary Note 1.7).
Image acquisition
On all instruments, image acquisition was performed using µManager57 (µManager version 2.0.0, http://micro-manager.org, RRID:SCR_000415) or ThorCam (ThorCam version 3.6.0, https://www.thorlabs.com/software_pages/ViewSoftwarePage.cfm?Code=ThorCam). Data were always recorded in unprocessed format. All experiments on all three microscopes were performed using epi-illumination (with the exception of the dSTORM experiment where a steep HILO angle was used) to easily control the polarization state at the sample plane. Perfectly random or circular polarization at the sample plane is very challenging to achieve experimentally; therefore, to some degree, there will always be a dominant axis of polarization. Transitioning from EPI to HILO and TIRF will change the amount of photoselection at the sample plane, as the dominant axis and degree of randomness or ellipticity of the polarization will change in a way that is challenging to quantify and reproduce. Before each experiment and/or change of inclination of the excitation beam, the polarization of the excitation beam at the sample is tuned by rotating a quarter-wave plate. Alignment is deemed optimal when the DoLP of the background is minimized (or the appearance of gridding distinctive of polarized background disappears). This step is not necessary if lasers are coupled into a multimode fiber. The imaging parameters for the data shown in all main figures and supplementary figures are summarized in Supplementary Table 3.
Image analysis
Single-molecule data analysis was performed using the MATLAB (MATLAB R2022a, MathWorks, http://www.mathworks.com/products/matlab/, RRID:SCR_001622) application POLCAM-SR, which includes tools for localization, filtering, drift correction and data visualization. The source code and installer are available on GitHub at https://github.com/ezrabru/POLCAM-SR. Diffraction-limited data analysis was performed using POLCAM-SR and a napari58 plugin called napari-polcam for processing and visualization of multidimensional polarization camera datasets. The source code and installation instructions for napari-polcam are available on GitHub at https://github.com/ezrabru/napari-polcam. Refer to Supplementary Note 7 for a detailed description of the image analysis pipeline.
SYTOX Orange on coverglass
Glass coverslips (VWR Collection, 631-0124) were cleaned with argon plasma for 30 min (Expanded Plasma Cleaner, PDC-002, Harrick Plasma). An imaging chamber was created on the coverslips using Frame-Seal slide chambers (9 × 9 mm, SLF0201, Bio-Rad). The glass in the chamber was coated with 70 μl PLL (0.01% (wt/vol), P4707, Sigma-Aldrich) for 15 min. After removing excess PLL and washing three times with filtered PBS (0.02-μm syringe filter, Whatman, 6809-1102), 50 μl of 1 nM SYTOX Orange (S11368, Invitrogen) was added gently. The sample was imaged immediately.
This protocol is available on https://www.protocols.io as ‘Imaging single SYTOX Orange molecules on a PLL-coated cover glass’79.
PAINT imaging of lipid bilayer-coated silica microspheres
To prepare lipid bilayer-coated silica microspheres, a slightly modified version of the protocol in ref. 33 was used. First, lipid vesicles of a certain composition were prepared as follows: DPPC (850355C, Avanti Polar Lipids) and cholesterol (C8667-5G, Sigma-Aldrich) were dissolved in chloroform (366927, Sigma-Aldrich) to (respectively) 25 mg ml−1 and 10 mg ml−1. A DPPC–40% cholesterol mixture was prepared by combining 23 μl DPPC and 20 μl cholesterol. The solvent was evaporated overnight under vacuum. The lipid–cholesterol mixture was rehydrated using 1 ml Tris-Ca2+ buffer (100 mM NaCl, 3 mM CaCl2, 10 mM Tris base, pH 7.4) and vortexed for 30 s. The solution was sonicated using a tip sonicator (cycles of 45 s on, 15 s off, 60% amplitude) for 40 min until the solution ran clear. The sonicated solution was centrifuged for 90 s at 14,000 rcf to remove titanium residue from the sonicator probe.
Next, silica microspheres with a diameter of 5 μm (44054-5ML-F, Sigma-Aldrich) were diluted to approximately 2.8 mg ml−1 and cleaned by centrifuging and replacing the stock buffer with Tris-Ca2+. The microspheres and lipid vesicle solutions were heated to 65 °C using a heated water bath and mixed together in a 1:1 ratio. After 30 min at 65 °C, the mixtures were slowly cooled down to room temperature (by turning the heating bath off). The buffer was gradually replaced with Tris (100 mM NaCl, 10 mM Tris base, pH 7.4) by centrifugation (5 min at 0.3 rcf) and replacement of two-thirds of the supernatant with Tris, repeated six times. The lipid-coated microspheres were stored at 4 °C and used within less than 2 weeks of preparation.
For imaging, the lipid-coated microspheres were added to an argon plasma-cleaned, PLL-coated coverglass (VWR Collection, 631-0124), and 1 nM NR4A or Nile red was added for PAINT imaging of the lipid bilayer. The Nile red derivative NR4A was provided by A.S. Klymchenko at the Université de Strasbourg. For diffraction-limited imaging of lipid-coated microspheres, 100 nM dye in PBS was used (for Nile red, NR4A and Di-8-ANEPPS). All buffers (PBS, Tris, Tris-Ca2+) were filtered before use (0.02-μm syringe filter, Whatman, 6809-1102).
This protocol is available on https://www.protocols.io as ‘Preparation and imaging of lipid bilayer-coated silica microspheres’80.
AF647 immobilized in PVA
A PVA solution (1%, 1 g in 100 ml) was prepared by slow addition of solid PVA into filtered (0.02-μm syringe filter, Whatman, 6809-1102) Milli-Q water with stirring. The solution was then heated to 90 °C, stirred for 30 min and then removed from heat with continued stirring for 12 h. The PVA solution was then filtered (0.02-μm syringe filter, Whatman, 6809-1102) and stored at 4 °C. AF647 (500 pM) was diluted in the 1% PVA solution, and 10 μl was then spin-cast (3,000 rpm, 45 s) onto a glass coverslip cleaned with Ar plasma (ODC-002, Harrick Plasma) and sealed before imaging.
This protocol is available on https://www.protocols.io as ‘Imaging single AF647 molecules immobilised in PVA on a cover glass’81.
TAB-PAINT of α-synuclein fibrils
To prepare α-synuclein fibrils, α-synuclein monomer was diluted to a concentration of 70 μM in PBS (with 0.01% NaN3) and incubated at 37 °C in a shaker (200 rpm) to aggregate for >24 h. To prepare fibrils for imaging, glass coverslips (VWR Collection, 631-0124) were plasma cleaned for 1 h (argon plasma cleaner, PDC-002, Harrick Plasma). An imaging chamber was created on the coverslips using Frame-Seal slide chambers (9 × 9 mm, SLF0201, Bio-Rad). The glass in the chamber was coated with 70 μl PLL (0.01% (wt/vol), P4707, Sigma-Aldrich) for 30 min. After removing excess PLL and washing three times with filtered PBS (20-nm pore filters), 50 μl TetraSpeck beads (0.1 μM stock, 10× diluted) were added for lateral drift correction. Samples were washed again 3× with filtered PBS and 50 μl α-synuclein fibrils (diluted to 35 mM monomer concentration from a 70 mM monomer concentration stock that was stored at 4 °C). Fibrils were stuck to PLL by pipetting up and down a couple of times in the four corners of the chamber. Before imaging, excess solution was removed, followed by a gentle wash with filtered PBS. Imaging buffer (50 μl, 1 nM Nile red in PBS, diluted from a 1 mM aliquot in DMSO, stored at −20 °C) was added, and the sample was imaged immediately.
This protocol is available on https://www.protocols.io as ‘TAB-PAINT imaging of alpha-synuclein fibrils using Nile Red’82.
dSTORM of actin in fixed HeLa cells
HeLa TDS cells (RRID:CVCL_0030) were cultured in DMEM medium (Gibco, Invitrogen) supplemented with 10% FBS (Life Technologies), 1% penicillin–streptomycin (Life Technologies) and 1% glutamine (Life Technologies) at 37 °C with 5% CO2. Cells were periodically tested for mycoplasma contamination and passaged three times per week. Cells were plated at low density on high-precision glass coverslips (MatTek, P35G-0.170-14-C) 1 d before fixation for dSTORM experiments.
Cells were simultaneously fixed and permeabilized in cytoskeleton buffer (10 mM MES, 138 mM KCl, 3 mM MgCl2, 2 mM EGTA, 4.5% sucrose (wt/vol), pH 7.4) with 4% paraformaldehyde and 0.2% Triton for 6 min at 37 °C and further fixed in cytoskeleton buffer with 4% paraformaldehyde for 14 min at 37 °C. After fixation, cells were washed three times with PBST (PBS supplemented with 0.1% Tween) and permeabilized a second time in PBS with 0.5% Triton for 5 min at room temperature. The samples were then washed three times with PBST and blocked for 30 min with 5% BSA. Cells were washed three times with PBST and then incubated with AF488–phalloidin (A12379, Invitrogen, 1:50 in PBS) for 1 h in the dark, followed by two washes with PBS. Before dSTORM imaging, PBS was replaced with dSTORM imaging buffer (base buffer consisting of 0.56 M glucose, 50 mM Tris (pH 8.5) and 10 mM NaCl supplemented with 5 U ml−1 pyranose oxidase (Sigma, P4234), 10 mM cysteamine (Sigma, 30070), 40 μg ml−1 catalase (Sigma, C100) and 2 mM cyclooctatetraene (Sigma, 138924).
This protocol is available on https://www.protocols.io as ‘dSTORM of actin in fixed HeLa cells’83.
Fixed COS-7 cells labeled with SiR–actin
A commercial slide (GATTA-Cells 4C, GATTAquant) was used, containing fixed COS-7 cells (RRID:CVCL_0224) labeled with SiR–actin (CY-SC001, Cytoskeleton).
Live cell imaging of the plasma membrane of Jurkat T cells
J8 LFA-1 cells were incubated overnight (~18 h) in complete RPMI (StableCell RPMI-1640 medium, Sigma) supplemented with 10% (vol/vol) fetal calf serum, 1% (vol/vol) HEPES buffer and 1% (vol/vol) penicillin–streptomycin antibiotics. Cells (1 ml) were collected by centrifugation and resuspended in phenol red-free RPMI supplemented with 1% HEPES.
Round coverslips were rinsed with IPA and Milli-Q, dried and cleaned with Ar plasma for 20 min. Grace Bio-Labs CultureWells were attached, and the slide was incubated with OKT3 antibody (provided by the Human Immunology Unit, WIMM) for 30 min. The slide was washed five times with phenol red-free RPMI supplemented with 1% HEPES, and a final wash was performed with phenol red-free RPMI supplemented with 1% HEPES and 200 nM NR4A (MemGlow NR4A Membrane Polarity Probe, MG06, Cytoskeleton) before imaging.
This protocol is available on https://www.protocols.io as ‘Live-cell imaging of the plasma membrane of Jurkat T cells’84.
Reporting summary
Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article.
Online content
Any methods, additional references, Nature Portfolio reporting summaries, source data, extended data, supplementary information, acknowledgements, peer review information; details of author contributions and competing interests; and statements of data and code availability are available at 10.1038/s41592-024-02382-8.
Supplementary information
Acknowledgements
This research was funded in part by Aligning Science Across Parkinson’s ASAP-000509 through the Michael J. Fox Foundation for Parkinson’s Research. For the purpose of open access, we have applied a CC BY public copyright license to all author accepted manuscripts arising from this submission. R.P. was funded by the European Research Council (ERC Consolidator Grant 820188-NanoMechShape to E.K.P.). We thank the Cambridge Advanced Imaging Centre (University of Cambridge) for the use of a microscope for the dSTORM experiments and the mechanical workshop (Y. Hamied, Department of Chemistry, University of Cambridge) for building custom parts of optical setups used in this work. We thank A. Ponjavic and J. Beckwith for critical reading of the manuscript. We thank L. Elfeky for testing the POLCAM-SR software.
Author contributions
S.F.L. conceived the project. S.F.L., K.O.’H. and M.D.L. supervised the research. E.B. built the optical setups. E.B. acquired data. E.B. programmed and performed simulations. E.B. and O.Z. analyzed data. E.B. programmed the POLCAM-SR, POLCAM-Live and napari-polcam software. O.Z. adapted the RoSE-O algorithm for polarization cameras. L.-M.N. prepared the α-synuclein fibrils and dye in polymer samples. R.P. prepared the dSTORM actin samples. M.K. prepared the live T cell samples. E.B., M.C. and S.D. prepared the lipid bilayer-coated microsphere samples. T.W. provided the protocol for preparing lipid bilayer-coated microspheres. A.S.K. provided the dye NR4A. S.J.D. provided the J8 LFA-1 cell line and OKT3 antibodies. E.B. and S.F.L. wrote the manuscript with input from all authors.
Peer review
Peer review information
Nature Methods thanks Jörg Enderlein, Peng Xi and the other, anonymous, reviewer(s) for their contribution to the peer review of this work. Peer reviewer reports are available. Primary Handling Editor: Rita Strack, in collaboration with the Nature Methods team.
Data availability
The datasets generated as part of this study were uploaded to Zenodo: pixel-dependent camera calibration results (10.5281/zenodo.10578307)85, single SYTOX Orange on a coverglass (10.5281/zenodo.10469322)86, PAINT data of single Nile red dye molecules binding to lipid bilayer-coated silica microspheres (10.5281/zenodo.10469444)87, TAB-PAINT data (10.5281/zenodo.10470795)88, dSTORM phalloidin–AF488 (10.5281/zenodo.10470982)89, dSTORM phalloidin–AF647 (10.5281/zenodo.10732697)90 and T cells (10.5281/zenodo.10471496)91.
Code availability
The installers and source code of all custom code and software used in this work are available on GitHub and Zenodo: POLCAM-SR (https://github.com/ezrabru/POLCAM-SR, 10.5281/zenodo.10732422 (ref. 92), RRID:SCR_025343), POLCAM-Live (https://github.com/ezrabru/POLCAM-Live, 10.5281/zenodo.10732437 (ref. 93), RRID:SCR_025342), napari-polcam (https://github.com/ezrabru/napari-polcam, 10.5281/zenodo.10732441 (ref. 94), RRID:SCR_025341), RoSE-O_POLCAM (https://github.com/Lew-Lab/RoSE-O_polCam, RRID:SCR_025340) and the camera calibration software (https://github.com/TheLeeLab/cameraCalibrationCMOS, 10.5281/zenodo.10732469 (ref. 95), RRID:SCR_025339).
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
The online version contains supplementary material available at 10.1038/s41592-024-02382-8.
References
- 1.Betzig, E. et al. Imaging intracellular fluorescent proteins at nanometer resolution. Science313, 1642–1645 (2006). [DOI] [PubMed] [Google Scholar]
- 2.Rust, M., Bates, M. & Zhuang, X. Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM). Nat. Methods3, 793–795 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Hess, S., Girirajan, T. & Mason, M. Ultra-high resolution imaging by fluorescence photoactivation localization microscopy. Biophys. J.91, 4258–4272 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sharonov, A. & Hochstrasser, R. Wide-field subdiffraction imaging by accumulated binding of diffusing probes. Proc. Natl Acad. Sci. USA103, 18911–18916 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Xu, K., Zhong, G. & Zhuang, X. Actin, spectrin, and associated proteins form a periodic cytoskeletal structure in axons. Science339, 452–456 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Szymborska, A. et al. Nuclear pore scaffold structure analyzed by super-resolution microscopy and particle averaging. Science341, 655–658 (2013). [DOI] [PubMed] [Google Scholar]
- 7.Mund, M. Systematic nanoscale analysis of endocytosis links efficient vesicle formation to patterned actin nucleation. Cell174, 884–896 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Shroder, D., Lippert, L. & Goldman, Y. Single molecule optical measurements of orientation and rotations of biological macromolecules. Methods Appl. Fluoresc.4, 042004 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Sosa, H., Peterman, E. J., Moerner, W. E. & Goldstein, L. S. ADP-induced rocking of the kinesin motor domain revealed by single-molecule fluorescence polarization microscopy. Nat. Struct. Biol.8, 540–544 (2001). [DOI] [PubMed] [Google Scholar]
- 10.Lippert, L. et al. Angular measurements of the dynein ring reveal a stepping mechanism dependent on a flexible stalk. Proc. Natl Acad. Sci. USA114, E4564–E4573 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Backer, A. S. et al. Single-molecule polarization microscopy of DNA intercalators sheds light on the structure of S-DNA. Sci. Adv.5, eaav1083 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ding, T., Wu, T., Mazidi, H., Zhang, O. & Lew, M. Single-molecule orientation localization microscopy for resolving structural heterogeneities within amyloid fibrils. Optica7, 602–607 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Lu, J., Mazidi, H., Ding, T., Zhang, O. & Lew, M. Single-molecule 3D orientation imaging reveals nanoscale compositional heterogeneity in lipid membranes. Angew. Chem. Int. Ed. Engl.59, 17572–17579 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Corrie, J. E. T. et al. Dynamic measurement of myosin light-chain-domain tilt and twist in muscle contraction. Nature400, 425–430 (1999). [DOI] [PubMed] [Google Scholar]
- 15.Forkey, J. N., Quinlan, M. E., Alexander Shaw, M., Corrie, J. E. T. & Goldman, Y. E. Three-dimensional structural dynamics of myosin V by single-molecule fluorescence polarization. Nature422, 399–404 (2003). [DOI] [PubMed] [Google Scholar]
- 16.Beausang, J. F., Sun, Y., Quinlan, M. E., Forkey, J. N. & Goldman, Y. E. Fluorescent labeling of calmodulin with bifunctional rhodamine. Cold Spring Harbor Protoc.10.1101/pdb.prot069351 (2012). [DOI] [PMC free article] [PubMed]
- 17.Chen, C. et al. Elongation factor G initiates translocation through a power stroke. Proc. Natl Acad. Sci. USA113, 7515–7520 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Novotny, L. & Hecht, B. Principles of Nano-Optics (Cambridge Univ., 2006).
- 19.Backer, A., Backlund, M., Lew, M. & Moerner, W. Single-molecule orientation measurements with a quadrated pupil. Opt. Lett.38, 1521–1523 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Backer, A. & Moerner, W. Extending single-molecule microscopy using optical Fourier processing. J. Phys. Chem. B118, 8313–8329 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Hafi, N. et al. Fluorescence nanoscopy by polarization modulation and polarization angle narrowing. Nat. Methods11, 579–584 (2014). [DOI] [PubMed] [Google Scholar]
- 22.Zhanghao, K. et al. Super-resolution dipole orientation mapping via polarization demodulation. Light Sci. Appl.5, e16166 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Guan, M. et al. Polarization modulation with optical lock-in detection reveals universal fluorescence anisotropy of subcellular structures in live cells. Light Sci. Appl.11, 4 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Aguet, F., Geissbühler, S., Märki, I., Lasser, T. & Unser, M. Super-resolution orientation estimation and localization of fluorescent dipoles using 3-D steerable filters. Opt. Express17, 6829–6848 (2009). [DOI] [PubMed] [Google Scholar]
- 25.Mortensen, K., Churchman, L., Spudich, J. & Flyvbjerg, H. Optimized localization analysis for single-molecule tracking and super-resolution microscopy. Nat. Methods7, 377–381 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Backer, A. S. & Moerner, W. E. Determining the rotational mobility of a single molecule from a single image: a numerical study. Opt. Express23, 4255–4276 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Böhmer, M. & Enderlein, J. Orientation imaging of single molecules by wide-field epifluorescence microscopy. J. Opt. Soc. Am. B20, 554–559 (2003). [Google Scholar]
- 28.Patra, D., Gregor, I. & Enderlein, J. Image analysis of defocused single-molecule images for three-dimensional molecule orientation studies. J. Phys. Chem. A108, 6836–6841 (2004). [Google Scholar]
- 29.Toprak, E. et al. Defocused orientation and position imaging (DOPI) of myosin V. Proc. Natl Acad. Sci. USA103, 6495–6499 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Backlund, M. et al. Simultaneous, accurate measurement of the 3D position and orientation of single molecules. Proc. Natl Acad. Sci. USA109, 19087–19092 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Backer, A., Backlund, M., von Diezmann, A., Sahl, S. & Moerner, W. A bisected pupil for studying single-molecule orientational dynamics and its application to three-dimensional super-resolution microscopy. Appl. Phys. Lett.104, 193701 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Zhang, O., Lu, J., Ding, T. & Lew, M. Imaging the three-dimensional orientation and rotational mobility of fluorescent emitters using the tri-spot point spread function. Appl. Phys. Lett.113, 031103 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Wu, T., Lu, J. & Lew, M. D. Dipole-spread-function engineering for simultaneously measuring the 3D orientations and 3D positions of fluorescent molecules. Optica9, 505–511 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Hulleman, C. N. et al. Simultaneous orientation and 3D localization microscopy with a Vortex point spread function. Nat. Commun.12, 5934 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Ding, T. & Lew, M. D. Single-molecule localization microscopy of 3D orientation and anisotropic wobble using a polarized vortex point spread function. J. Phys. Chem. B125, 12718–12729 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Curcio, V., Alemán-Castañeda, L. A., Brown, T. G., Brasselet, S. & Alonso, M. A. Birefringent Fourier filtering for single molecule coordinate and height super-resolution imaging with dithering and orientation. Nat. Commun.11, 5307 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Zhang, O. et al. Six-dimensional single-molecule imaging with isotropic resolution using a multi-view reflector microscope. Nat. Photonics17, 179–186 (2022). [DOI] [PMC free article] [PubMed]
- 38.Gould, T. J. et al. Nanoscale imaging of molecular positions and anisotropies. Nat. Methods5, 1027–1030 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Cruz, C. A. V. et al. Quantitative nanoscale imaging of orientational order in biological filaments by polarized superresolution microscopy. Proc. Natl Acad. Sci. USA7, E820–E828 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Lewis, J. & Lu, Z. Resolution of ångström-scale protein conformational changes by analyzing fluorescence anisotropy. Nat. Struct. Mol. Biol.26, 802–807 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Mehta, S. et al. Dissection of molecular assembly dynamics by tracking orientation and position of single molecules in live cells. Proc. Natl Acad. Sci. USA113, E6352–E6361 (2016). [DOI] [PMC free article] [PubMed]
- 42.Ohmachi, M. Fluorescence microscopy for simultaneous observation of 3D orientation and movement and its application to quantum rod-tagged myosin V. Proc. Natl Acad. Sci. USA109, 5294–5298 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Chung, I., Shimizu, K. & Bawendi, M. Room temperature measurements of the 3D orientation of single CdSe quantum dots using polarization microscopy. Proc. Natl Acad. Sci. USA100, 405–408 (2003). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Rimoli, C. V., Valades-Cruz, C. A., Curcio, V., Mavrakis, M. & Brasselet, S. 4polar-STORM polarized super-resolution imaging of actin filament organization in cells. Nat. Commun.13, 301 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Beckwith, J. S. & Yang, H. Sub-millisecond translational and orientational dynamics of a freely moving single nanoprobe. J. Phys. Chem. B125, 13436–13443 (2021). [DOI] [PubMed] [Google Scholar]
- 46.Monaghan, J. W. et al. Calcite-assisted localization and kinetics (CLocK) microscopy. J. Phys. Chem. Lett.13, 10527–10533 (2022). [DOI] [PubMed] [Google Scholar]
- 47.Foreman, M. R. & Török, P. Fundamental limits in single-molecule orientation measurements. New J. Phys.13, 093013 (2011). [Google Scholar]
- 48.Lewis, J. H., Beausang, J. F., Sweeney, H. L. & Goldman, Y. E. The azimuthal path of myosin V and its dependence on lever-arm length. J. Gen. Physiol.139, 101–120 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Beausang, J. F., Shroder, D. Y., Nelson, P. C. & Goldman, Y. E. Tilting and wobble of myosin V by high-speed single-molecule polarized fluorescence microscopy. Biophys. J.104, 1263–1273 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Chun, C. S. L., Fleming, D. L. & Torok, E. J. Polarization-sensitive thermal imaging. In Proc. of SPIE, Automatic Object Recognition IV2234, 275–286 (SPIE, 1994).
- 51.Gruev, V., Perkins, R. & York, T. CCD polarization imaging sensor with aluminum nanowire optical filters. Opt. Express18, 19087–19094 (2010). [DOI] [PubMed] [Google Scholar]
- 52.Gruev, V., Van der Spiegel, J. & Engheta, N. Dual-tier thin film polymer polarization imaging sensor. Opt. Express18, 19292–19303 (2010). [DOI] [PubMed] [Google Scholar]
- 53.Kulkarni, M. & Gruev, V. Integrated spectral-polarization imaging sensor with aluminum nanowire polarization filters. Opt. Express20, 22997–23012 (2012). [DOI] [PubMed] [Google Scholar]
- 54.Ratliff, B., LaCasse, C. & Tyo, J. Interpolation strategies for reducing IFOV artifacts in microgrid polarimeter imagery. Opt. Express17, 9112–9125 (2009). [DOI] [PubMed]
- 55.Tyo, J., LaCasse, C. & Ratliff, B. Total elimination of sampling errors in polarization imagery obtained with integrated microgrid polarimeters. Opt. Lett.34, 3187–3189 (2009). [DOI] [PubMed]
- 56.Gao, S. & Gruev, V. Bilinear and bicubic interpolation methods for division of focal plane polarimeters. Opt. Express19, 26161–26173 (2011). [DOI] [PubMed] [Google Scholar]
- 57.Edelstein, A., Amodaj, N., Hoover, K., Vale, R. & Stuurman, N. Computer control of microscopes using μManager. Curr. Protoc. Mol. Biol.92, 14.20.1–14.20.17 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Sofroniew, N. et al. napari: a multi-dimensional image viewer for Python. Zenodo10.5281/zenodo.7276432 (2022).
- 59.Axelrod, D. Carbocyanine dye orientation in red cell membrane studied by microscopic fluorescence polarization. Biophys. J.26, 557–574 (1979). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Fourkas, J. T. Rapid determination of the three-dimensional orientation of single molecules. Opt. Lett.26, 211–213 (2001). [DOI] [PubMed] [Google Scholar]
- 61.Ströhl, F., Bruggeman, E., Rowlands, C. J., Wolfson, D. L. & Ahluwalia, B. S. Quantification of the NA dependent change of shape in the image formation of a z-polarized fluorescent molecule using vectorial diffraction simulations. Microsc. Res. Tech.85, 2016–2022 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Born, M. & Wolf, E. Principles of Optics 7th edn (Cambridge Univ. Press, 2019).
- 63.Gunturk, B., Glotzbach, J., Altunbasak, Y., Schafer, R. & Mersereau, R. Demosaicking: color filter array interpolation. IEEE Signal Process. Mag.22, 44–54 (2005). [Google Scholar]
- 64.Li, N., Zhao, Y., Pan, Q. & Kong, S. G. Removal of reflections in LWIR image with polarization characteristics. Opt. Express26, 16488–16504 (2018). [DOI] [PubMed] [Google Scholar]
- 65.Spehar, K. et al. Super-resolution imaging of amyloid structures over extended times by using transient binding of single thioflavin T molecules. ChemBioChem19, 1944–1948 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Nieuwenhuizen, R. P. J. et al. Measuring image resolution in optical nanoscopy. Nat. Methods10, 557–562 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Mazidi, H., King, E. S., Zhang, O., Nehorai, A. & Lew, M. D. Dense super-resolution imaging of molecular orientation via joint sparse basis deconvolution and spatial pooling. In IEEE 16th International Symposium on Biomedical Imaging (ISBI 2019) 325–329 (IEEE, 2019).
- 68.Danylchuk, D. I., Moon, S., Xu, K. & Klymchenko, A. S. Switchable solvatochromic probes for live-cell super-resolution imaging of plasma membrane organization. Angew. Chem. Int. Ed. Engl.58, 14920–14924 (2019). [DOI] [PubMed] [Google Scholar]
- 69.Göhring, J., Schrangl, L., Schütz, G. J. & Huppa, J. B. Mechanosurveillance: tiptoeing T cells. Front. Immunol.13, 886328 (2022). [DOI] [PMC free article] [PubMed]
- 70.Mahecic, D. et al. Event-driven acquisition for content-enriched microscopy. Nat. Methods19, 1262–1267 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Alvelid, J., Damenti, M., Sgattoni, C. & Testa, I. Event-triggered STED imaging. Nat. Methods19, 1268–1275 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Liberatore, A. et al. PolarCam micropolarizer cameras characterization and usage. In International Conference on Space Optics — ICSO 2020 358–379 (SPIE, 2021).
- 73.Sugizaki, A. et al. POLArIS, a versatile probe for molecular orientation, revealed actin filaments associated with microtubule asters in early embryos. Proc. Natl Acad. Sci. USA118, e2019071118 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Shuhei Shibata, S., Takano, W., Hagen, N., Matsuda, M. & Otani, Y. Video-rate quantitative phase analysis by a DIC microscope using a polarization camera. Biomed. Opt. Express10, 1273–1281 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Yeh, L.-H. et al. uPTI: uniaxial permittivity tensor imaging of intrinsic density and anisotropy. In Biophotonics Congress 2021 (2021) NM3C.4 (Optica Publishing Group, 2021).
- 76.Kalita, R. et al. Single-shot phase contrast microscopy using polarisation-resolved differential phase contrast. J. Biophotonics14, e202100144 (2021). [DOI] [PubMed] [Google Scholar]
- 77.Song, S., Kim, J., Hur, S., Song, J. & Joo, C. Large-area, high-resolution birefringence imaging with polarization-sensitive Fourier ptychographic microscopy. ACS Photonics8, 158–165 (2021). [Google Scholar]
- 78.Hecht, E. Optics 4th edn (Addison Wesley, 2002).
- 79.Bruggeman, E. Imaging Single SYTOX Orange Molecules on a PLL-Coated Cover Glass. Protocols.iowww.protocols.io/view/imaging-single-sytox-orange-molecules-on-a-pll-coa-c672zhqe (2024).
- 80.Bruggeman, E. Preparation and Imaging of Lipid Bilayer-Coated Silica Microspheres. Protocols.iowww.protocols.io/view/preparation-and-imaging-of-lipid-bilayer-coated-si-ddca22se (2024).
- 81.Bruggeman, E. Imaging Single AF647 Molecules Immobilised in PVA on a Cover Glass. Protocols.iowww.protocols.io/view/imaging-single-af647-molecules-immobilised-in-pva-c68tzhwn (2024).
- 82.Bruggeman, E. TAB-PAINT Imaging of α-Synuclein Fibrils Using Nile Red. Protocols.iowww.protocols.io/view/tab-paint-imaging-of-alpha-synuclein-fibrils-using-c69bzh2n (2024).
- 83.Bruggeman, E. & Peters, R. dSTORM of Actin in Fixed HeLa Cells. Protocols.iowww.protocols.io/view/dstorm-of-actin-in-fixed-hela-cells-ddet23en (2024).
- 84.Bruggeman, E. & Körbel, M. Live-Cell Imaging of the Plasma Membrane of Jurkat T Cells. Protocols.iowww.protocols.io/view/live-cell-imaging-of-the-plasma-membrane-of-jurkat-ddeu23ew (2024).
- 85.Bruggeman, E. Pixel-dependent offset, gain, variance and readnoise maps for CS505MUP polarisation camera. Zenodo10.5281/zenodo.10578308 (2024).
- 86.Bruggeman, E. Polarisation camera movie of single SYTOX Orange molecules on a cover glass. Zenodo10.5281/zenodo.10469322 (2024).
- 87.Bruggeman, E. Polarisation camera PAINT movie of single Nile red dyes binding to lipid bilayer-coated silica microspheres. Zenodo10.5281/zenodo.10469445 (2024).
- 88.Bruggeman, E. Polarisation camera TAB-PAINT dataset of α-synuclein fibrils using Nile red. Zenodo10.5281/zenodo.10470795 (2024).
- 89.Bruggeman, E. Polarisation camera dSTORM datasets of actin in fixed HeLa cells labeled with phalloidin–Alexa Fluor 488. Zenodo10.5281/zenodo.10470982 (2024).
- 90.Bruggeman, E. Polarisation camera dSTORM datasets of actin in fixed HeLa cells labeled with phalloidin–Alexa Fluor 647. Zenodo10.5281/zenodo.10732697 (2024).
- 91.Bruggeman, E. Polarisation camera movies of T cells on a cover glass. Zenodo10.5281/zenodo.10471496 (2024).
- 92.Bruggeman, E. ezrabru/POLCAM-SR: version 1.4. Zenodo10.5281/zenodo.10732422 (2024).
- 93.Bruggeman, E. ezrabru/POLCAM-Live: v0.1. Zenodo10.5281/zenodo.10732437 (2024).
- 94.Bruggeman, E. ezrabru/napari-polcam: v0.1. Zenodo10.5281/zenodo.10732441 (2024).
- 95.Bruggeman, E. TheLeeLab/cameraCalibrationCMOS: v1.1. Zenodo10.5281/zenodo.10732469 (2024).
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets generated as part of this study were uploaded to Zenodo: pixel-dependent camera calibration results (10.5281/zenodo.10578307)85, single SYTOX Orange on a coverglass (10.5281/zenodo.10469322)86, PAINT data of single Nile red dye molecules binding to lipid bilayer-coated silica microspheres (10.5281/zenodo.10469444)87, TAB-PAINT data (10.5281/zenodo.10470795)88, dSTORM phalloidin–AF488 (10.5281/zenodo.10470982)89, dSTORM phalloidin–AF647 (10.5281/zenodo.10732697)90 and T cells (10.5281/zenodo.10471496)91.
The installers and source code of all custom code and software used in this work are available on GitHub and Zenodo: POLCAM-SR (https://github.com/ezrabru/POLCAM-SR, 10.5281/zenodo.10732422 (ref. 92), RRID:SCR_025343), POLCAM-Live (https://github.com/ezrabru/POLCAM-Live, 10.5281/zenodo.10732437 (ref. 93), RRID:SCR_025342), napari-polcam (https://github.com/ezrabru/napari-polcam, 10.5281/zenodo.10732441 (ref. 94), RRID:SCR_025341), RoSE-O_POLCAM (https://github.com/Lew-Lab/RoSE-O_polCam, RRID:SCR_025340) and the camera calibration software (https://github.com/TheLeeLab/cameraCalibrationCMOS, 10.5281/zenodo.10732469 (ref. 95), RRID:SCR_025339).