. 2024 Nov 30;16(11):e74878. doi: 10.7759/cureus.74878

Safety and Efficacy of Phacoemulsification With Trabecular Microbypass Stent W Implantation in Primary Angle-Closure Glaucoma

Kazuyuki Hirooka 1,, Hiromitsu Onoe 2, Hideaki Okumichi 1, Yoshiaki Kiuchi 1
Editors: Alexander Muacevic, John R Adler
PMCID: PMC11685551  PMID: 39741605

Abstract

Background

Investigation of the safety and efficacy of phacoemulsification with trabecular microbypass Stent W implantation in patients with primary angle-closure glaucoma (PACG).

Methods

Between August and December of 2023, this prospective study evaluated PACG patients who underwent phacoemulsification with iStent inject W implantation. All patients were 18 years and older and were monitored for 6 months after surgery. Changes in intraocular pressure (IOP) and the number of glaucoma medications at 6 months after the surgery were defined as the primary endpoint, while the safety of the iStent inject W implantation in these patients was the secondary endpoint.

Results

Of the six patients in the study, one was male and five were female, with patient ages ranging from 62 to 83 years. Pre- and postoperative IOP (at 6 months) ranged from 12 to 18 mmHg and 9 to 17 mmHg, respectively, with only one patient exhibiting an increase in the IOP from 12 to 17 mmHg. The number of glaucoma medications decreased in all but one case at six months postoperatively. Neither hyphema nor an IOP spike was observed in the present study. There was no evidence of iStent inject W occlusion or malposition at six months postoperatively in any of the cases.

Conclusion

In patients with PACG, phacoemulsification with trabecular microbypass Stent W implantation was demonstrated to be a safe and effective procedure.

Keywords: efficacy, istent inject w, phacoemulsification cataract surgery, primary angle-closure glaucoma, safety

Introduction

The primary cause of irreversible blindness globally is glaucoma, with the most common type being primary open-angle glaucoma (POAG). In contrast to POAG, primary-angle closure glaucoma (PACG) is associated with the closure of the anterior chamber angle by synechial or appositional approximation between the iris and the trabecular meshwork. This type of glaucoma is more aggressive and more likely to lead to blindness. It was reported that the global prevalence of PACG in 2013 was 0.5% [1]. The number of people worldwide with PACG was estimated to be 23.36 million in 2020, with the number projected to increase to 32.04 million by 2040 [1]. In 2010, bilateral blindness was estimated to affect 4.5 million people with POAG and 3.9 million people with PACG. By 2020, these numbers had risen to 5.9 and 5.3 million people, respectively [2].

It has been reported that the angle closure mechanism is associated with an increased lens thickness, with age additionally playing an important role [3]. Pupillary block is caused by a small eyeball with a thickened and forward-moving lens [3]. Therefore, to relieve the pupillary blockage, removal of the thickened lens (cataract surgery) is performed to deepen the anterior chamber. In PACG, although cataract surgery has an angle-widening effect, eyes with chronic iridocorneal apposition and elevated intraocular pressure (IOP) often have trabecular meshwork damage that makes it unlikely to be repaired by simply widening the angle [4]. In order to adequately and sustainably control the IOP in these eyes, glaucoma surgery may be necessary. Although substantial IOP reduction can be achieved by using traditional filtering surgeries, such as trabeculectomy and tube shunt, patients can also incur short- and long-term risks such as hypotony, diplopia, and infection [5]. A diverse group of relatively new procedures referred to as minimally invasive glaucoma surgery (MIGS) have been shown to play an important role in glaucoma treatment. Due to these procedures' ability to reduce both the IOP and IOP-lowering medications with limited or no disruption to the conjunctiva and sclera, they can help patients avoid the risks associated with filtration surgery [6,7].

The United States FDA has approved the iStent trabecular microbypass (Glaukos Co., San Clemente, CA, USA), and more recently, the second-generation iStent inject trabecular microbypass stent (Glaukos Co.). These procedures are indicated in patients with mild to moderate open-angle glaucoma, with these stents designed to create a direct route from the anterior chamber to Schlemm’s canal for the aqueous humor to bypass the damaged trabecular meshwork. For example, there is a wide flange at the base of the iStent inject W (Glaukos Co.) designed to optimize stent visualization and placement. However, in most countries, these devices are designated as “off-label” for PACG patients. In the present investigation, we examined patients with PACG and evaluated the safety and efficacy of phacoemulsification with trabecular microbypass stent implantation.

Materials and methods

Patient selection

The study protocol for this prospective study, with no control group, was approved by the Institutional Review Board of Hiroshima University (Approval No. CRB2022-0012) and registered as jRCTs062230007 in the Japan Registry of Clinical Trials (jRCT). We aimed to recruit as many PACG patients as possible from April 2023 to December 2023. During this period, we analyzed six eyes from six PACG subjects who underwent phacoemulsification cataract extraction combined with iStent inject W implantation. This study was conducted in accordance with the principles of the Declaration of Helsinki. All patients consented to be included in the manuscript.

To be included in the analyses for this study, the eyes had to meet several criteria: patients must be ≥18 years of age; the posterior trabecular meshwork must not be visible for ≥180 degrees on gonioscopy, with or without peripheral anterior synechiae (PAS); there must be glaucomatous fundus abnormalities such as thinning of the rim in the optic disc or a defect in the retinal nerve fiber layer (RNFL) corresponding to a visual field defect; a cataract and a need for IOP reduction must be present; and there must be follow-up data available for 6 months. We excluded patients with severe glaucoma (mean deviation worse than -12 dB), PAS in the nasal quadrant precluding the placement of iStent inject W, corneal endothelium cell density below 1500 cells/mm2, or an IOP ≥25 mmHg.

Surgical technique

A temporal corneal incision was made prior to phacoemulsification. The intraocular lens (IOL) was implanted in the capsular bag following cataract removal. To enhance visibility of Schlemm’s canal, sodium hyaluronate was added to the anterior chamber after IOL implantation. Subsequently, the goniolens was placed on the cornea after tilting the patient’s head and the microscope to observe the nasal angle. Two iStent inject W devices, spaced one to two clock hours apart, were inserted into the nasal quadrant trabecular meshwork after identifying Schlemm’s canal. Goniosynechialysis was not performed even if PAS was present. Postoperatively, all patients were started on eye drops three to four times a day, including 1.5% levofloxacin (Nipro Co., Osaka, Japan) and 0.1% betamethasone (Shionogi & Co., Ltd., Osaka, Japan), and 0.1% bromfenac (Senju Pharmaceutical Co., Ltd., Osaka, Japan) twice a day for 3-4 weeks. If patients had been using glaucoma eye drops prior to surgery, these were discontinued. Subsequently, based on the postoperative IOP, glaucoma eye drops were gradually reintroduced on an as-needed basis.

Other study measures

Prior to the surgery, all patients underwent a baseline examination. The preoperative data collected included age at the time of surgery, gender, best-corrected visual acuity (BCVA), and mean IOP. At every visit, the IOP was measured using a Goldmann applanation tonometer. Gonioscopy was performed under dark conditions using a narrow light beam and a 4-mirror goniolens (Volk Optical Inc., Mentor, OH, USA). Automated perimetry was conducted using the Humphrey Visual Field 24-2 Swedish Interactive Threshold Algorithm (SITA) standard program (Humphrey Field Analyzer II; Carl Zeiss Meditec AG, Jena, Germany). Optical coherence tomography (OCT) of the RNFL was performed with an RTVue-XR Avanti (Optovue Inc., Fremont, CA, USA). Corneal endothelial cell density was measured using a specular microscope and autofocus device (Topcon SP-3000; Topcon Co., Tokyo, Japan).

Patients were required to return to the clinical center at least five times: at 1 day, 1 week, and 1, 3, and 6 months postoperatively. The data collected at all postoperative consecutive visits included IOP measurements, the number of glaucoma medications, BCVA, and complications. A 4-mirror gonio lens was used to confirm the position of the iStent inject W at one day and six months postoperatively. At baseline and six months after the surgery, Humphrey perimetry and corneal endothelial cell density were reassessed.

Primary and secondary outcome measures

Differences in the IOP and the number of glaucoma medications at six months were defined as the primary outcome measures, while the incidences of intraoperative and postoperative complications were defined as the secondary outcome measures.

Results

Table 1 presents the clinical characteristics of the 6 eyes from the six patients enrolled in this study. The patients included one male and five females, ranging in age from 62 to 84 years, with baseline IOPs ranging from 12 to 18 mmHg.

Table 1. Demographic findings of each case.

y: Year; R: Right; L: Left; IOP: Intraocular pressure.

Patient Age (y) Gender Side (R/L) Baseline IOP (mmHg) No. glaucoma medication Corneal endothelial cells (/mm2)
1 83 Female L 15 3 2691
2 64 Female L 15 2 2954
3 62 Female R 16 0 2876
4 65 Female R 16 4 2673
5 74 Female R 12 2 1985
6 64 Male L 18 0 2365

Two trabecular stents were successfully implanted in all eyes. None of the patients experienced any intraoperative complications such as iris prolapse, iris trauma, or iridodialysis. Intraoperative hyphema occurred in all cases immediately after the insertion of the iStent inject W into Schlemm’s canal. The progression of IOP and the number of glaucoma medications are presented in Table 2. Preoperative and 6-month postoperative IOPs were 15.3 ± 2.0 mmHg and 14.2 ± 2.8 mmHg, respectively (P = 0.51; paired t-test). At 6 months postoperatively, except for one case (patient 5), the IOP was the same or lower than that observed preoperatively. At 6 months, patient 5 resumed glaucoma medication. In all but one case (patient 6), the number of glaucoma medications decreased at 6 months postoperatively. No significant difference was observed in the mean number of glaucoma medications compared to the baseline (P = 0.18; paired t-test).

Table 2. Individual data for IOP change.

Preoperative and six-month postoperative IOPs were 15.3 ± 2.0 mmHg and 14.2 ± 2.8 mmHg, respectively (P = 0.51; paired t-test).

IOP: Intraocular pressure; d: Day; w: Week; m: Month.

Patient IOP (no. of glaucoma medications)    
  Baseline 1d 1w 1m 3m 6m
1 15 (3) 15 (0) 15 (0) 13 (0) 14 (0) 14 (0)
2 15 (2) 11 (0) 12 (0) 12 (0) 14 (0) 15 (0)
3 16 (0) 12 (0) 12 (0) 12 (0) 10 (0) 9 (0)
4 16 (4) 15 (0) 14 (0) 15 (0) 16 (0) 16 (2)
5 12 (2) 15 (0) 13 (0) 14 (0) 14 (0) 17 (0)
6 18 (0) 24 (0) 18 (2) 18 (0) 14 (2) 14 (2)

Changes in corneal endothelial cell density are shown in Table 3. Preoperative and 6-month postoperative corneal endothelial cell counts were 2591 ± 360 /mm² and 2481 ± 326 /mm², respectively (P = 0.31; paired t-test). No significant decreases in corneal endothelial cell density or instances of hyphema or transient elevation of IOP were observed in any of the patients. No evidence of iStent inject W occlusion or malposition was observed at 6 months postoperatively in any of the cases. 

Table 3. Changes in corneal endothelial cells.

Preoperative and six-month postoperative corneal endothelial cell densities were 2591 ± 360 cells/mm² and 2481 ± 326 cells/mm², respectively (P = 0.31; paired t-test).

m: month.

Patient Baseline (cells/mm2) 6m (cells/mm2)
1 2691 2205
2 2954 2810
3 2876 2673
4 2673 2832
5 1985 2124
6 2365 2239

Discussion

This study evaluated patients with PACG and investigated the safety and efficacy of phacoemulsification when using iStent inject W implantation. The efficacy of phacoemulsification with iStent inject W in patients with open-angle glaucoma has been evaluated and reported in several other previous studies [8-10].

A significant decrease in the IOP after six months was reported by Deneri S et al., with an IOP of 13.88 ± 2.57 mmHg (P < 0.01) compared to the baseline IOP of 16.08 ± 3.27 mmHg [8]. The preoperative IOP was 16.2 ± 4.4 mmHg, while it was 12.5 ± 2.6 mmHg at 6 months postoperatively (P < 0.001) [9]. As recently reported by Morita S et al., there was a significant decrease from the preoperative mean IOP of 15.0 ± 2.8 mmHg to the 12-month postoperative value of 13.8 ± 3.3 mmHg (P < 0.01) [10]. Although two previous studies reported a significant decrease in the mean number of IOP-lowering medications from the preoperative values of 2.1 ± 1.4 [9] and 2.9 ± 1.4 [10] to the postoperative values of 0.5 ± 0.8 (P < 0.001) at 6 months and 1.7 ± 1.6 (P < 0.01) at 12 months, respectively, a third study reported no significant decrease in the mean number of IOP-lowering medications preoperatively (2.69 ± 1.03) and at six months postoperatively (2.06 ± 1.34) [8].

Currently, there are no published studies on the safety and efficacy of the iStent inject W implantation in PACG patients. However, two studies have investigated patients with PACG and examined the efficacy and safety of phacoemulsification with iStent inject implantation [11,12]. One of the studies found a significant decrease in the preoperative medicated IOP from 18.6 ± 4.7 mmHg to 14.9 ± 2.8 mmHg at 6 months after phacoemulsification with iStent inject implantation [11]. Furthermore, at 12 months postoperatively, the mean number of glaucoma medications was 0.25 ± 0.68 [11]. After 12 months of follow-up, a study by Salimi et al. reported a decrease in the IOP from 17.5 ± 2.7 mmHg to 13.0 ± 2.3 mmHg, and a decrease in the glaucoma medications from 2.2 ± 1.3 to 1.3 ± 0.8 [12]. Based on the results found in these two previous studies, the authors concluded that in PACG eyes, phacoemulsification with iStent inject implantation was more effective than phacoemulsification alone. In this study, we did not compare standard phacoemulsification with the efficacy of combined phacoemulsification with iStent inject W implantation in PACG patients. However, it is our assumption that phacoemulsification with iStent inject W implantation is indeed more effective than phacoemulsification alone.

In this study, hyphema was resolved by the first postoperative visit, although in all eyes there was blood reflux into the anterior chamber when the iStent inject W was well inserted in Schlemm’s canal. In our recent study on POAG eyes, hyphema was observed in 1 eye (1.8%) [9]. However, on postoperative day 1 in this study, there was no hyphema. In both POAG and PACG eyes, hyphema is a rare complication. A study of PACG eyes by Chen DZ et al. reported that iStent occlusion with the iris occurred in five eyes (31.3%) [11]. However, we did not observe iStent inject W occlusion with the iris in this study. In open-angle glaucoma eyes, the incidence of iStent inject occlusion has been reported to be between 0% and 4% [13-16]. Therefore, even in PACG eyes, we speculate that iStent inject W occlusion with the iris is not higher.

The present study had a few limitations. First, our study did not contain a phacoemulsification-only control group. Therefore, distinguishing the effect of iStent inject W implantation versus the cataract extraction was not possible. Second, the present study only evaluated 6 eyes. Therefore, to obtain more rigorous and definitive evidence, we will need to conduct larger-scale, multi-center trials.

Conclusions

In conclusion, the results of this study demonstrate both the efficacy and safety of combined phacoemulsification with iStent inject W implantation in eyes with PACG. We believe that this indication can be expanded, as cataract surgery can open the corneal angle even in cases with a narrow angle.

Funding Statement

This study was funded by Glaukos Japan. The funders had no role in the study design, data collection, decision to publish, or preparation of the manuscript.

Disclosures

Human subjects: Consent for treatment and open access publication was obtained or waived by all participants in this study. Institutional Review Board of the Hiroshima University issued approval CRB2022-0012.

Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.

Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:

Payment/services info: This study was funded by Glaukos Japan. The funders had no role in the study design, data collection, decision to publish, or preparation of the manuscript.

Financial relationships: Kazuyuki Hirooka declare(s) This study was funded by Glaukos Japan. from Glaukos Japan.

Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.

Author Contributions

Concept and design:  Kazuyuki Hirooka, Yoshiaki Kiuchi

Acquisition, analysis, or interpretation of data:  Kazuyuki Hirooka, Hiromitsu Onoe, Hideaki Okumichi

Drafting of the manuscript:  Kazuyuki Hirooka

Critical review of the manuscript for important intellectual content:  Hiromitsu Onoe, Hideaki Okumichi, Yoshiaki Kiuchi

References

  • 1.Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Tham YC, Li X, Wong TY, Quigley HA, Aung T, Cheng CY. Ophthalmology. 2014;121:2081–2090. doi: 10.1016/j.ophtha.2014.05.013. [DOI] [PubMed] [Google Scholar]
  • 2.The number of people with glaucoma worldwide in 2010 and 2020. Quigley HA, Broman AT. Br J Ophthalmol. 2006;90:262–267. doi: 10.1136/bjo.2005.081224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Contemporary approach to narrow angle. Shamseldin SW, Reddy R, Razeghinejad R, Jay KL. J Ophthalmic Vis Res. 2024;19:88–108. doi: 10.18502/jovr.v19i1.15443. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Phacoemulsification versus combined phacotrabeculectomy in medically uncontrolled chronic angle closure glaucoma with cataracts. Tham CC, Kwong YY, Leung DY, et al. Ophthalmology. 2009;116:725–731. doi: 10.1016/j.ophtha.2008.12.054. [DOI] [PubMed] [Google Scholar]
  • 5.Postoperative complications in the Tube Versus Trabeculectomy (TVT) study during five years of follow-up. Gedde SJ, Herndon LW, Brandt JD, Budenz DL, Feuer WJ, Schiffman JC. Am J Ophthalmol. 2012;153:804–814. doi: 10.1016/j.ajo.2011.10.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.American Glaucoma Society position paper: microinvasible glaucoma surgery. Fellman RL, Mattox C, Singh K, et al. Ophthalmol Glaucoma. 2020;3:1–6. doi: 10.1016/j.ogla.2019.12.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Trends in glaucoma surgeries performed by glaucoma subspecialists versus nonsubspecialists on medicare beneficiaries from 2008 through 2016. Rathi S, Andrews CA, Greenfield DS, Stein JD. Ophthalmology. 2021;128:30–38. doi: 10.1016/j.ophtha.2020.06.051. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Comparison of the iStent Inject(®) versus the iStent Inject(®) W-Both in combination with cataract surgery-in open-angle glaucoma. Deneri S, Merté RL, Eter N, Brücher VC. J Clin Med. 2023;12:7259. doi: 10.3390/jcm12237259. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Comparison of surgical outcomes between iStent inject W implantation and microhook ab interno trabeculotomy in combination with phacoemulsification in primary open-angle glaucoma patients. Onoe H, Hirooka K, Namiguchi K, et al. Front Med (Lausanne) 2023;10:1266532. doi: 10.3389/fmed.2023.1266532. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Comparative evaluation of iStent versus iStent inject W combined with phacoemulsification in open angle glaucoma. Morita S, Sakanishi Y, Riyu I, Watanabe S, Ebihara N. PLoS One. 2024;19:0. doi: 10.1371/journal.pone.0297514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Phacoemulsification vs phacoemulsification with micro-bypass stent implantation in primary angle closure and primary angle closure glaucoma: a randomized single-masked clinical study. Chen DZ, Sng CC, Sangtam T, Thomas A, Shen L, Huang PK, Cheng J. Clin Exp Ophthalmol. 2020;48:450–461. doi: 10.1111/ceo.13721. [DOI] [PubMed] [Google Scholar]
  • 12.Matched cohort study of cataract surgery with and without trabecular microbypass stent implantation in primary angle-closure glaucoma. Salimi A, Abu-Nada M, Harasymowycz P. Am J Ophthalmol. 2021;224:310–320. doi: 10.1016/j.ajo.2020.12.032. [DOI] [PubMed] [Google Scholar]
  • 13.Second-generation trabecula micro-bypass (iStent inject) with cataract surgery in eyes with normal-tension glaucoma: one-year outcomes of a multi-center study. Salimi A, Clement C, Shiu M, Harasymowycz P. Ophthalmol Ther. 2020;9:585–596. doi: 10.1007/s40123-020-00266-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Prospective unmasked randomized evaluation of the iStent inject (®) versus two ocular hypotensive agents in patients with primary open-angle glaucoma. Fea AM, Belda JI, Rękas M, et al. Clin Ophthalmol. 2014;8:875–882. doi: 10.2147/OPTH.S59932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Prospective, randomized, controlled pivotal trial of an ab interno implanted trabecular micro-bypass in primary open-angle glaucoma and cataract: two-year results. Samuelson TW, Sarkisian SR Jr, Lubeck DM, et al. Ophthalmology. 2019;126:811–821. doi: 10.1016/j.ophtha.2019.03.006. [DOI] [PubMed] [Google Scholar]
  • 16.Prospective, unmasked evaluation of the iStent® inject system for open-angle glaucoma: synergy trial. Voskanyan L, García-Feijoó J, Belda JI, Fea A, Jünemann A, Baudouin C. Adv Ther. 2014;31:189–201. doi: 10.1007/s12325-014-0095-y. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.