Abstract
Purpose
The temporal course of spontaneous mechanical activities of musculature (SMAMs) is investigated using a novel multiple‐point diffusion‐weighted stimulated echo imaging sequence (MP‐DW‐STE) with adapted spatial and temporal resolution. For this purpose, different sequence settings and measurement parameters are applied.
Methods
A single‐shot MP‐DW‐STE imaging sequence with multiple signal rephasing by small flip angle RF pulses was developed to acquire image series during spontaneous muscle contractions with duration of several hundred milliseconds. Measurements were conducted on an incoherent motion phantom and in the calf muscles of eight healthy volunteers. The number and cross‐sectional area of SMAM visualizations as well as the contractile behavior in terms of onset and duration of visible SMAMs was analyzed.
Results
Measurements on the incoherent motion phantom confirmed the ability of the proposed technique to characterize dynamic incoherent motion as signal voids in series of images with a high temporal resolution of approximately 40 ms. All human subjects showed SMAMs with a median duration of 120 ms, therefore, visible in several diffusion‐weighted images in a row. Contraction time of SMAMs was in the range of 80 to 120 ms for the soleus muscle. The MP‐DW‐STE sequence settings have shown to significantly influence the mean frequency and cross‐sectional area of SMAMs.
Conclusion
MP‐DW‐STE imaging allows for time‐resolved recording of spontaneous muscular contractions and provides new insights into their dynamic course. This new feature can be used for better characterization of physiological SMAMs in healthy subjects and pathological SMAMs in patients suffering from neuromuscular diseases.
Keywords: diffusion‐weighted imaging, motion‐sensitizing, multiple‐point acquisition, spontaneous mechanical activities in musculature, spontaneous muscular activities, twitch contraction
1. INTRODUCTION
Resting human skeletal muscle of healthy subjects, and especially muscle affected by some types of neuromuscular diseases, shows spontaneous mechanical activity. 1 Frequency of fiber twitches and affected muscular regions differ between benign and pathological conditions and can be an unequivocal sign of abnormality. 2 Increased spontaneous muscular fasciculations in specific muscle areas represents one of the most specific symptoms in the diagnosis of some neuromuscular diseases in specific stages. 2 Spontaneous generation of action potentials in motor neurons and following excitation of motor units (several muscle fibers connected to the axon and its terminal branches) result in muscular twitch contractions (i.e., fasciculations). In particular, in patients suffering from amyotrophic lateral sclerosis, fasciculation is an early pathophysiological feature associated with other signs of the disease, such as muscle weakness, 3 and is, therefore, of great interest in clinical and biomedical research.
Spontaneous muscular twitch contraction is generated by the shortening of several spatially distributed muscle fibers belonging to a motor unit. This results in microscopic displacement of adjacent excited and non‐excited fibers. Hence, a 3D incoherent motion is formed. Diffusion‐weighted MRI (DW‐MRI) is inherently sensitive to incoherent motion during the dephasing and rephasing of the motion‐sensitizing gradients of the MR acquisition. Based on this, it has been shown that single‐shot diffusion‐weighted spin‐echo (DW‐SE) and stimulated‐echo (DW‐STE) imaging is able to visualize spontaneous muscular contractions as signal voids in diffusion‐weighted images (DWI), termed spontaneous mechanical activities in musculature (SMAM). 4 These signal voids in DWI have been shown to correlate directly with small myoelectric activities detectable on the skin by surface electromyography 5 and have a pronounced extension along the muscle fiber direction. 6 An increase of visible SMAMs in DWI for patients suffering from amyotrophic lateral sclerosis was found for specific muscular regions. 7 , 8 Recording of the spontaneous activities using DW‐MRI is commonly realized by repeating the same single‐shot echo‐planar DWI sequence with a fixed repetition interval (of at least about 500 ms) resulting in asynchronous sampling of the twitch contraction. DW‐MRI in combination with electrical nerve stimulation has shown the potential to visualize the muscular twitch contraction by stimulating the motor unit with an electrical pulse and repetition of the MR acquisition with different time delays. 9 , 10 , 11 , 12 , 13 Both techniques, with and without electrical nerve stimulation, use standard DW‐SE sequences 9 , 10 , 11 , 12 , 13 or DW‐STE imaging 4 , 5 , 6 with relatively long time delays between consecutive measurements. Without repeated electrical nerve stimulation, these DW‐MRI sequences are only able to sample one time point during a spontaneous muscular contraction. Therefore, visualization of the course of contraction over time is not feasible and, therefore, it is impossible to gain further knowledge on the temporal course of SMAMs.
STE sequences have been modified in previous works by introducing series of small flip angle refocusing pulses (instead of one 90° refocusing pulse) for high‐speed imaging 14 , 15 , 16 or multislice imaging 17 , 18 , 19 . The use of small flip angle refocusing pulses allows the acquisition of multiple echoes (with inherently increasing diffusion weighting for longer mixing times TM). 20 Further technical improvement relates to the acquisition of the full k‐space after a small flip angle refocusing pulse resulting in a multiple‐echo stimulated echo image series. 21 Reported preliminary studies 22 , 23 have shown that it might be possible to visualize the muscular twitch contraction using this kind of STE adaptation, called multiple‐point in time diffusion‐weighted stimulated echo (MP‐DW‐STE) imaging. Further studies 24 , 25 have shown diffusion measurements using a single‐shot multi‐b‐value sequence using the same sequence concept.
The objective of this work is to explore the influence of different settings of the MP‐DW‐STE sequence for recording multiple time points of single spontaneous muscular contractions and to adapt those parameters for potential clinical applications. Furthermore, the typical temporal course of spontaneous contractions in the lower leg of healthy subjects was investigated. A newly built incoherent motion phantom was used for sequence testing.
2. METHODS
2.1. Design of the MP‐DW‐STE sequence
The sequence scheme of the single‐shot MP‐DW‐STE imaging sequence is depicted in Figure 1. The sequence was implemented with Pulseq 26 . In contrast to a conventional STE scheme, the third RF pulse after the mixing time TM, which deflects the stored magnetization in the longitudinal direction into the transverse plane, is replaced by a series of NRF variable flip angles to deflect only a fraction of the stored magnetization per RF refocusing pulse. Before the signal echo is sampled, the rephasing gradient lobe for diffusion weighting is inserted. For reduction of the time TRF between two subsequent α‐RF pulses, diffusion‐gradients were used as crusher gradients for selection of the stimulated echo pathway. Next, k‐space is read out by an EPI trajectory achieving the acquisition of NRF single‐shot images over all NRF imaging cycles. 22 , 24 Interference of the transverse magnetization on signal echoes in subsequent imaging cycles was suppressed by inserting spoiler gradient pulses along all three gradient directions. For signal echo readout, EPI with ramp sampling 27 and partial Fourier was chosen to shorten the acquisition time and enable short TRF. Under consideration of the duration of the α‐RF pulses, crusher and spoiler gradients as well as signal readout using EPI, a temporal resolution of 40 ms was achieved on our 3 T whole body system.
FIGURE 1.

Scheme of the single‐shot multiple‐point diffusion‐weighted stimulated echo imaging sequence. The number of α‐RF pulses NRF, time between the α‐RF pulses (TRF) and mixing time (TM) are adjustable. Motion‐sensitizing gradient lobes are shaded in gray with the motion‐sensitizing time Δ n of the nth acquisition. Motion‐sensitizing gradients were also applied as crusher gradients for selection of the stimulated echo pathway.
Cyclic readout of the stored magnetization leads to an increase of the mixing time TM in the nth cycle: . Therefore, the signal intensity in each voxel changes over the imaging cycles. This can be (at least partly) compensated by using a variable flip angle approach 18 , 24 , 28 to obtain constant signal intensity for a given T1. With and given the condition that the entire remaining stored magnetization should be deflected in the last cycle NRF with , the flip angle series is iteratively determined by 18 , 29 :
| (1) |
Increasing leads to an increase of the motion‐sensitizing time, and therefore, to a change of the motion‐sensitivity of each imaging cycle. The b‐value of the nth image acquisition without consideration of imaging gradients is
| (2) |
with ; : initial motion‐sensitizing time of the first acquisition; rise time; duration of motion‐sensitizing gradient. Therefore, the nth b‐value in the acquisition series depends on the initial mixing time TM and the initial b‐value b1 following
| (3) |
with and being the duration of the RF pulse. The implemented MP‐DW‐STE imaging sequence will be made available for other research groups on request.
2.2. Incoherent motion phantom
To evaluate the DW‐MP‐STE sequence, measurements were also conducted on a custom‐built incoherent motion phantom. Because the signal voids visible in DWI result from intra‐voxel dephasing because of 3D incoherent motion, a basic implementation to reproduce this kind of incoherent motion is the induction of small local water disturbances by motion beads in a restricted area of an MR phantom filled with fluid. 6 To increase the viscosity of the pure water and locally control the induced water disturbances, small “intra‐slice” tubes were filled with water and cellulose (2%). An overview on the structure of the phantom is given in Figure S1. Furthermore, four “inter‐slice” tubes were filled with a solution of soy‐lecithin and water to adjust different T1‐values mimicking the relaxation times of different tissues (T1: 800 ms, 1000 ms, 1300 ms, and 1600 ms). 30 , 31 Motion beads inside the intra‐slice tubes can be shifted by thin filaments made of polyamide and controlled by a synchronized motor unit (Arduino Due connected to servomotor devices) from the outside of the critical magnetic field (i.e., the 5 Gauss line). To prevent water exchange between intra‐slice tubes and the surrounding area, an additional layer of rubber was used to seal the intra‐slice tube for the duration of the examination. The motion bead position is, therefore, fully controllable regarding position and velocity. The microcontroller is also able to synchronize the MR imaging process by triggering the MR system using a fiber optical connection. This allows synchronizing the motion bead cycle and MR image acquisition. For testing the implemented MP‐DW‐STE sequence, a linear motion pattern was applied with synchronization of the sequence (i.e., start of the incoherent motion in the intra‐slice tube after the first DWI).
2.3. Study participants
Six male and two female volunteers (age: 31 ± 13 years) participated in the study after giving written informed consent. All subjects were examined in supine position. In total, 32 DW‐MRI datasets (4 different DWI measurements per subject) with summarized 74 400 DWIs were recorded. The study protocol was approved by the local ethics review board of the medical faculty of the Eberhard Karls University and the University Hospital of Tübingen in 830/2020BO2.
2.4. MRI acquisition settings—Healthy volunteers
All DW‐MRI measurements were conducted on a 3 T MR scanner (MAGNETOM PrismaFit, Siemens Healthcare) with a 15‐channel Tx/Rx knee‐coil. In volunteer examinations, transverse slices were acquired at the position with maximum diameter of the right calf. Sequence parameters were: matrix size of 64 × 64, FOV of 192 × 192 mm2, slice‐thickness of 8 mm, 5/8 readout, receiver bandwidth (BW) = 2404 Hz/px, TE = 22 ms, (between consecutive STE excitations) TR = 1000 ms. A total scan‐time of 5 min per parameter set was chosen leading to a series of 300 repetitive measurements (each of them with multiple‐point imaging). Spectral fat suppression was applied. The number of RF repetitions was NRF = 10 and time between two subsequent α‐RF refocusing pulses was set to TRF = 40 ms. This results in a temporal resolution of 40 ms for all in vivo measurements. Flip angle series of α‐RF pulses was adjusted for skeletal muscle tissue (T1 = 1300 ms) following Eq. (1): α n = [15.9°, 17.0°, 18.4°, 20.1°, 22.1°, 24.8°, 28.5°, 34.0°, 44.1°, 90.0°]. For investigation of the influence of the ratio between TM and TRF on the visible SMAMs, three different TM were selected: (I) short TM (29 ms) with constant increase in motion‐sensitizing time; (II) medium TM (145 ms), which equals the “standard sequence settings” used in former studies 4 , 5 ; and (III) long TM (189 ms). For reference (IV), a “standard” stimulated echo sequence with NRF = 1, b‐value = 100 s/mm2 and TM = 145 ms was chosen according to previous works. 4 , 5 All sequence parameters that are changed for the different experimental settings are given in Table 1. For anatomical reference, T1‐weighted fast spin‐echo (T1w‐FSE) images were recorded using the same slice positions with matrix size of 256 × 256, FOV of 192 × 192 mm2, slice‐thickness of 6 mm, BW of 180 Hz/px, TE = 10 ms, TR = 650 ms, and echo train length of 5.
TABLE 1.
Sequence parameters for the four different in vivo experimental settings (I–IV) and for the phantom measurements.
| Experimental setting | |||||||
|---|---|---|---|---|---|---|---|
| In vivo | |||||||
| Sequence parameter | Variable | Unit | I | II | III | IV | Phantom |
| Mixing time | TM | ms | 29 | 145 | 189 | 145 | 145 |
| Temporal resolution | TRF | ms | 40 | 40 | 40 | 40 | 200 |
| RF repetitions | NRF | 10 | 10 | 10 | 1 | 5 | |
| Motion‐sensitizing times | Δ1–ΔNRF | ms | 40, 80,…, 400 | 156, 196,…, 516 | 200, 240,…, 560 | 156 | 156, 356,…, 956 |
| b‐Value | b1–b10 | s/mm2 | 50–520 | 50–167 | 50–141 | 100 | 50–309 |
Note: b‐Values were calculated following Equation (3).
2.5. MRI acquisition settings—Phantom measurements
For measurements using the incoherent motion phantom, acquisition settings were adapted to NRF = 5. To avoid signal voids in the entire intra‐slice motion tube, the motion bead is moved rather slowly (20 mm/s) and sequence parameters were adapted to TM = 145 ms and TRF = 200 ms (Table 1).
For visualization and verification of the motion‐bead position in test measurements on the incoherent motion phantom, images were acquired using a balanced steady state free precession (bSSFP) sequence with high temporal resolution (matrix size of 96 × 96, FOV of 192 × 192 mm2, slice‐thickness of 10 mm, BW of 1532 Hz/px, TE = 0.97 ms, and TR = 130.57 ms).
2.6. DW data processing and evaluation
All image reconstruction and evaluation processing steps were implemented in custom‐written code in MATLAB (The MathWorks). Intra‐series and inter‐modality spatial accordance between T1w‐FSE and DWI was achieved by an affine registration using elastix 32 , 33 with mutual information as similarity metric before further processing. For detection and segmentation of SMAMs in DWI, a neural network approach 8 , 34 was used. The used neural network architecture consists of two different sub‐networks. First a neural network based on a shallow encoder‐decoder structure inspired by U‐Net 35 , 36 to generate intermediate spatio‐temporal feature maps and, second, a neural network block for processing of the temporal relations between each temporal step using bidirectional convolutional long short‐term memories (CLSTMs) 37 , 38 , 39 . Afterward, segmentation results were revised to ensure classification results and proper subsequent evaluation. Large signal dropouts across multiple muscle regions or whole muscles were discarded since these motion patterns were assumed to be a result of intended muscular movements. Studies on the incoherent motion phantom were solely evaluated qualitatively to prove the function of the sequence. For in vivo studies, number and cross‐sectional area (CSA) of SMAMs were analyzed per repetition of the DWI measurements as well as per DWI time‐series image for each set of MR sequence parameters. SMAM activities for each NRF were visualized in terms of percentage event count maps (pECMs) (i.e., the summation of SMAMs in time‐direction normalized by the number of acquired repetitions). Furthermore, temporal behavior of SMAMs was analyzed by grouping all activities regarding their onset in the DWI time‐series (i.e., the first DWI with visible signal void to the last DWI with spatial overlapping visible signal void). To analyze the dynamic contraction of SMAMs, a contraction time SMAMTC was defined as the temporal period between the first occurrence in DWI to the maximum CSA of the SMAM in the DWI time‐series. For this reason, only SMAMs that were not yet visible at the first recording time in the series were evaluated, because otherwise the start time of the SMAMs would remain unclear.
2.7. Statistical analyses
Preliminary evaluations using a Shapiro–Wilk test have shown that the results are not normally distributed. Therefore, non‐parametric statistical tests were used for analysis. For pair‐wise statistical comparison of all three measurements settings, a non‐parametric two‐tailed Wilcoxon test was used in SPSS (IBM, version 28.0.1.1[14]). The analysis of statistical differences over all three measurement settings was performed with a non‐parametric Friedman test. In both statistical tests, a Bonferroni‐correction for detection of significant differences with a significance level of Padj = 0.05 was applied.
3. RESULTS
3.1. Phantom measurements
To validate the implemented MP‐DW‐STE sequence, movement pattern and signal intensity within the DWI time‐series was investigated. An exemplary scheme of a linear motion pattern with sequence triggers and time points for DW imaging is depicted in Figure 2A. The signal void resulting from the water turbulence from the moving motion bead grows according to the movement visible in the bSSFP images (Figure 2B). Visualization of the dynamic motion bead movements using the MP‐DW‐STE sequence was possible in all test series (with 100 repetitions each). A maximum variation of the signal intensity over the DWI time‐series of 1.27% was measured indicating a good compensation of T1‐related signal intensity effects using an adjusted flip angle series following Eq. (1). Therefore, no signal interference from previous imaging cycles was assumed.
FIGURE 2.

(A) Scheme of a linear motion bead displacement dMB inside the incoherent motion phantom. Motion bead position represented by red circles was assessed by the short repetition balanced steady state free precession (bSSFP) sequence. Areas with signal voids in images recorded by the multiple‐point diffusion‐weighted stimulated echo (MP‐DW‐STE) sequence are depicted in blue. (B) Images of the motion bead using bSSFP (top row) and MP‐DW‐STE (bottom row) sequences. Images in the bottom row show extended signal voids resulting from incoherent motion caused by the bead displacement. The signal void propagates from the center to the left (red: moving bead, yellow: stationary bead).
3.2. In vivo measurements
All measurements in healthy volunteers led to sufficient image quality for reliable analysis. However, one subject had to be excluded in further evaluation steps because of increased rate of extended signal voids indicating intended motion of the lower leg, which leads to difficulties in the evaluation of spontaneous muscular activities. An increased rate of activity in the extensor digitorum longus muscle was detected that was directly related to toe movements. This muscle did not reveal a high rate of spontaneous muscular activities in previous studies. 4 , 5 Variation of the rate of spontaneous activities was expected to be low during the acquisition of the four measurements, because preliminary studies have shown only a small intra‐day variability. 40
A summarizing overview on the in vivo results is given in Table 2. SNR is inherently reduced compared to standard STE imaging, however, a minimum mean SNR of 26 (Table 2) was achieved, which is comparable to former studies. 4 A mean signal reduction of minimum 49% ± 10% was found, which shows a sufficient image quality for SMAM analyses. Figure 3 shows an exemplary time‐series of DWI (TM = 145 ms) with a distinct SMAM in the soleus muscle. The CSA of the SMAM reaches its maximum at the sixth time point of the image recording series. In relation to this maximum extension, the area of the SMAM in the image time‐series was: 59.5% (5th), 71.5% (7th), and 33.6% (8th). The SMAM contraction time (SMAMTC, given as number of imaging periods with 40 ms each) between first visible signal void and maximum CSA was quite short, as the next image after the first occurrence already shows the maximum CSA value. For comparison, a SMAM with long (Figure 3B) and short (Figure 3C) duration is depicted. The CSA of the SMAM with long duration also first increases and then decreases.
TABLE 2.
A summarizing overview on the in vivo results in terms of image quality, the relative number of DWI indicating SMAMs and the temporal characteristic of visible SMAMs for all four experimental settings.
| Experimental setting | ||||||
|---|---|---|---|---|---|---|
| I (TM = 29 ms) | II (TM = 145 ms) | III (TM = 189 ms) | IV (standard STE) | |||
| SNR (mean 1st image to mean 10th image) | μ | 46 to 26 | 37 to 32 | 33 to 31 | 156 | |
| Min. mean SMAM signal drop | [%, μ ± σ] | 49 ± 10 | 50 ± 12 | 52 ± 10 | 52 ± 10 | |
| Relative amount of DWI‐series repetitions with SMAM | Subject 1 | [%] | 20.0 | 20.0 | 14.0 | 12.3 |
| Subject 2 | [%] | 16.3 | 19.7 | 14.7 | 10.0 | |
| Subject 3 | [%] | 3.3 | 5.7 | 6.3 | 3.0 | |
| Subject 4 | [%] | 18.0 | 18.0 | 13.0 | 4.0 | |
| Subject 5 | [%] | 2.0 | 5.0 | 5.0 | 4.0 | |
| Subject 6 | [%] | 16.7 | 13.3 | 14.3 | 9.6 | |
| Subject 7 | [%] | 22.3 | 14.0 | 15.7 | 10.7 | |
| [%, μ ± σ] | 14.1 ± 8.1 | 13.7 ± 6.2 | 11.9 ± 4.3 | 7.7 ± 3.8 | ||
| Relative amount of SMAMs with: | ||||||
| • Start: 1st image, end: 10th image | [%] | 15.2 | 25.3 | 25.1 | – | |
| • Start: 1st image, end before 10th image | [%] | 5.1 | 17.6 | 20.5 | – | |
| • Start: after 1st image, end: 10th image | [%] | 73.1 | 55.9 | 53.7 | – | |
| • Start and end within time‐series | [%] | 22.1 | 26.9 | 25.8 | – | |
| • Only present for two imaging points | [%] | 4.2 | 7.5 | 4.2 | – | |
| Maximum SMAMTC | [images] | 8 | 6 | 4 | – | |
| Amount of smallest possible SMAMTC | [%] | 29.9 | 55.4 | 56.0 | – | |
| SMAMTC (soleus) | Median (95% CI) |
2 (2, 3) n = 87 |
1 (1, 2) n = 56 |
1 (1, 2) n = 50 |
– | |
| SMAMTC (gastrocnemius medialis) | Median (95% CI) |
1 (1, 2) n = 28 |
2 (1, 2) n = 14 |
2 (1, 2) n = 10 |
– | |
| Duration (soleus) | Median (95% CI) |
3 (3, 4) n = 50 |
3 (3, 4) n = 49 |
3 (3, 4) n = 47 |
– | |
| Duration (gastrocnemius medialis) | Median (95% CI) |
3.5 (2, 5) n = 12 |
3 (3, 5) n = 18 |
4 (3, 5) n = 11 |
– | |
Abbreviations: CI, confidence interval; DWI, diffusion‐weighted images; Min, minimum; SMAM; spontaneous mechanical activities of musculature.
FIGURE 3.

(A) Exemplary multiple‐point diffusion‐weighted stimulated echo imaging sequence (MP‐DW‐STE) image series (TM = 145 ms) showing a distinct spontaneous mechanical activity of musculature (SMAM) (annotated in red) in the soleus muscle over multiple images (intra‐series image number 5–8) with corresponding normalized cross‐sectional area (CSA). An enlarged view of the SMAM is given in the upper row (highlighted in green). It can be observed that the size of the SMAM first increases and subsequently decreases. (B) Exemplary SMAM with a long temporal duration. (C) Exemplary SMAM with short temporal duration. The contraction time SMAMTC is defined as temporal period between first visible signal void and maximum CSA.
A share of 14.1% ± 8.1% of all repetitions have shown a visible SMAM in at least one DWI (experimental setup I) compared to 11.9% ± 4.3% (III) shown in Table 2. Standard STE (IV) has shown a reduced rate of repetitions showing SMAMs (7.7% ± 3.8%). A minimum number of repetitions with visible SMAMs were found for subject 5 with 2.0% (I) in contrast to 22.3% (I) in subject 7. A significant difference between I versus IV (Padj = 0.04) and II versus IV (Padj = 0.03) was found by analyzing the four paired measurements with a non‐parametric Wilcoxon test.
Figure 4 indicates detection rates of SMAMs for all four MP‐DW‐STE sequence settings as pECMs for subject 1 (highest mean rate of SMAMs) and subject 5 (lowest mean rate of SMAMs). pECMs for all seven subjects are summarized in Figure S2. It is obvious that the overall patterns of active muscle regions are the same for all four settings. However, differences in the SMAM rate are noticeable. The highest rate of subject 1 in the soleus muscle is 1.94% for I, which shows a decrease to 0.81% for III. For subject 5 with a rather low amount of SMAMs, an increase is visible (0.39% [I], 0.84% [III]). pECMs for all 10 imaging points in time of DWI time‐series for subject 1 are shown in Figure 5. It can be seen that there is an increase of the activity rate from first to fifth pECM for I. III shows a distinct decrease for all 10 imaging time points. Median numbers of images per DWI time‐series time point, which indicates SMAMs are given in Table S1. An increase in the median number of images showing SMAMs from the first point in time to the maximum at seventh to ninth imaging point in time was found. The CSA of SMAMs in the unit of voxels is given for each imaging point in time. It can be seen that the spread in CSA increases for I for later time points in comparison to other experimental settings. The relative distribution of the durations of SMAMs (i.e., referring to the number of DWI time‐series images with visible SMAM at the same location) in relation to the onset of the SMAM in the DWI time‐series is given in Figure 6 for all three measurement settings. A total of 15.2% of all DWI repetitions indicating SMAMs have shown activities, which are present over the whole DWI time‐series (i.e., visible on the same spatial region from the first to the 10th DWI time‐series image) for I in contrast to 25.3% (II) and 25.1% (III). The relative shares of SMAMs with different onset and end are given in Table 2 with calculations in Figure S3A–E. The relative maximum CSA normalized by muscle area of SMAMs in DWI time‐series in relation to the SMAMTC period (i.e., temporal period in number of DWI time‐series images between first visible signal void and maximum CSA) is given in Figure 7 for the soleus muscle. Interestingly, SMAMs with a prolonged SMAMTC are only present for short TM (I). Highest SMAMTC for II was found to be 6 and 4 for III. A median SMAMTC of 2 (95% CI: [2, 3]) with duration of 3 (95% CI: [3, 4]) was found for experimental setting I (Table 2) with differences to the two other experimental settings. Small differences in SMAMTC and duration were revealed in the gastrocnemius medialis muscle. Other muscles were not evaluated regarding SMAMTC and SMAM duration because of the low rate of SMAMs.
FIGURE 4.

Overall percentage event count maps (pECMs) visualized for subject 1 (highest mean rate of SMAMs) and subject 5 (lowest mean rate of SMAMs) for all four MR sequence settings. It can be seen that all pECMs show the same active muscular regions. However, differences in the detected rate are clearly noticeable.
FIGURE 5.

Visualization of the intra‐series percentage event count maps (pECMs) for MR sequence settings I to III for subject 1 (highest mean rate) for all 10 measurement points in time.
FIGURE 6.

Relative distribution of the duration of visible spontaneous mechanical activities of musculature (SMAMs) (i.e., referring to the number of intra‐series images) in relation to the onset of SMAM (first imaging point in time of the diffusion‐weighted images [DWI] time‐series) for all three measurement settings. Distribution per measurement settings was normalized by the overall number of images indicating SMAMs.
FIGURE 7.

Relative maximum cross‐sectional area (CSA) of visible spontaneous mechanical activities of musculature (SMAMs) in relation to SMAMTC with the number of occurrences for the soleus muscle. Relative CSA was normalized on the area of the soleus muscle per subject. SMAMTC was defined as period between first visible signal void in diffusion‐weighted images (DWI) and maximum CSA.
4. DISCUSSION
The incoherent motion phantom could be well used in the process of MR sequence development for assessment of the temporal course of spontaneous muscular activities. It provided a solid basis for the investigations when traversing different movement trajectories with specified speed of the motion beads.
The contractile characteristics of the human skeletal muscular system can be divided into two temporal sections: a contraction phase and a relaxation phase. The contractile characteristics seem to be well represented by the course of the CSA of the SMAM in the successive images. Most of SMAMs that are visible in more than two subsequent DWI images have shown an increase of CSA peaking a maximum area and subsequent decrease. An example of muscular relaxation is demonstrated after the 6th image in the series shown in Figure 3. Therefore, it can be concluded that the MP‐DW‐STE sequence is able to image the contractile characteristics of SMAMs. For the different measurement settings I to III, the activation patterns were quite similar regarding active muscular regions, but the rate of activations showed significant differences. The b‐value of DW MRI sequences is mainly dependent on the gradient amplitude GD, the gradient duration δ, and the motion‐sensitizing time Δ. Because the b‐value was defined on the first image, the gradient amplitude is fixed for all other dephasing and rephasing motion‐sensitizing gradients within one measurement setting but changes between the measurement settings I to IV: GD,I = 27 mT/m, GD,II = 13.5 mT/m, GD,III = 11.9 mT/m, and GD,IV = 19.0 mT/m. For this, a relative motion difference of 87 μm (I), 173 μm (II), 198 μm (III), and 123 μm (IV) between the dephasing and rephasing motion‐sensitizing gradients results in a 180° phase shift. Therefore, the sensitivity of the MP‐DW‐STE sequence to visualize small muscular contractions is increased for I and decreased for II and III in relation to the standard STE sequence (IV). Based on this, the difference in the rate of activation between I (14.1% ± 8.1%) and III (11.9% ± 4.3%) in pECMs might be reflected by these changes of sensitivity. Furthermore, the increase in CSA for experimental setting I could also be a consequence. A dependence of the measured duration of SMAMs in the image series on the time of onset was shown for sequence settings I to III. The temporal sensitivity of a DW sequence, which is the probability to visualize SMAMs, is highly related to the timing of the sequence (i.e., mixing time interval TM). 41 Because the muscular contractions occur asynchronous to the sampling pattern of the DW sequence, the timing of the onset of the muscular contraction during the imaging sequence in combination with the duration leads to certain likelihood that this movement can be visualized. 5 , 41 For example, SMAMs that start and end during the mixing interval TM cannot be visualized by a DW sequence. The different types of possible configurations of muscular contraction that leads to different visible outcomes are depicted schematically in Figure 8 with corresponding class‐distribution scheme in Figure S4. Five different “types” of muscular contraction can be distinguished.
MI: muscular contraction starts before the dephasing gradient in the initial phase of the MP‐DW‐STE sequence and relaxes during the mixing time interval TM (i.e., relaxation before the first rephasing gradient) might lead to a phase shift for all imaging points in time and, therefore, a signal void in all following imaging cycles. The share of this type of SMAM is increased in II and III (25.3% and 25.1% in contrast to 15.2% [I]) because of the prolonged mixing time interval TM.
MII: for MII, the muscular contraction is active after the dephasing gradient but before the rephasing gradient of the first image and ends during the imaging cycles. For this, the contraction time SMAMTC cannot be determined, hence, the starting point is not known. An increase of this type of SMAM was found for II (17.6%) and III (20.5%) in contrast to I (5.1%). This difference is caused by the prolonged TM in II and III in which the muscular contraction can evolve and start relaxation.
MIII: the muscular contraction starts after recording the first image (i.e., no visible SMAM in the first DWI) and ends before the last image in the cycle is acquired. In this case, the contraction and relaxation phase is visualized and reflected by increasing and decreasing CSA. II and III showed an increased relative amount of these contractions mainly because of the differences in motion sensitivity.
MIV: muscular contraction shows an onset during the imaging cycles, but without finalized relaxation before recording the last image. Therefore, the contractile behavior cannot fully be determined.
MV: complete contraction and relaxation within the mixing time interval TM or after recording the last image does not lead to visualized SMAMs in DWI. 5
FIGURE 8.

Asynchronous onset of muscular motion in relation to the multiple‐point diffusion‐weighted stimulated echo imaging sequence (MP‐DW‐STE) imaging sequence leads to five different kind of spontaneous mechanical activities of musculature (SMAM) motion pattern (MI to MV) in the diffusion‐weighted images (DWI) time‐series. Investigation of dynamic characterization is only possible for the subgroup MIII where muscular motion starts after the first image and ends before the last image. Group MV is not visible in the DWI time‐series because motion starts and relaxes between motion dephasing and rephasing gradients.
Considering the time period between two consecutive RF excitations (i.e., TRF = 40 ms) and the temporal period between first onset and maximum CSA (i.e., SMAMTC) the median contraction time for the soleus muscle was calculated to be 80 ms (95% CI: [80, 120 ms]) in (I) and 40 ms (95% CI: [40, 80 ms]) for (II) and (III). In contrast, the gastrocnemius medialis muscle showed a median contraction time of 40 ms (95% CI: [40, 80 ms]) to 80 ms (95% CI: [40, 80 ms]). Muscle fiber types can be mainly subdivided in slow twitch (fiber type I) and fast twitch (fiber type IIa/b/d/m). 1 The soleus muscle is predominantly composed by slow twitch fibers (type I) 42 with reported slow contraction times of 127 ± 6 ms (range, 64–251 ms) 43 to 156.5 ± 14.7 ms 44 . In comparison, the contraction time of the gastrocnemius medialis muscle was reported to be 113.7 ± 19.6 ms 44 . A rather large deviation of those values given in the literature to the measured SMAMTC is evident. This might be caused by the coarse temporal resolution of the MP‐DW‐STE in conjunction with the sensitivity of the sequence to visualize muscular motion depending on the amplitude of motion‐sensitizing gradients. For example, a relative motion difference of 87 μm leads to a phase shift of 180° for measurement setting I. Therefore, temporal regions with only a small motion difference, especially in the beginning of the muscular contraction or the end of muscular relaxation, might not be visualized by the sequence and the muscular contractions appears shorter than expected.
A median duration of SMAMs of 120 ms was found for soleus muscle and 120 to 160 ms for gastrocnemius muscle. From muscle ultrasound examinations, it was reported that fasciculations have shown a duration of 200 to 500 ms. 45 , 46 It can be derived from this study that MP‐DW‐STE imaging might underestimate the duration of fasciculations. This finding is likely to be related to the lower sensitivity of the MR sequence to very small displacements in musculature in comparison to ultrasound measurements. Another aspect could also be the number of images within the DWI time‐series because the measurements showed that 73.1% (I) of the SMAMs are not fully relaxed at the end of acquisition. For a more precise assessment of the SMAM duration, the RF time interval or the number of recordings during the DWI time‐series must be increased. Dynamic muscle ultrasound provides information about duration of the fasciculation 47 because of the rather high temporal resolution and motion sensitivity (resolving changes in contractile length as low as 5 μm 48 ); however, the pick‐up area is restricted to a few centimeters in this technique. 49 Here, the visualization of fasciculations with DW‐MRI seems advantageous because of the large FOV without limitations in the transverse plane. This allows the simultaneous visualization of a larger amount of motor units in comparison to muscle ultrasound and enables imaging of the contraction time of fasciculation in deep muscles. Furthermore, assessment of fasciculating motor units in all muscular regions of the body is feasible enabling whole‐body imaging of fasciculation with determination of the rate and contraction time.
One limitation of this study is the small number of subjects some of whom have shown rather low frequency of SMAMs. However, the large number of repetitions per subject leads to sufficient data for the characterization of dynamic muscular contraction by MP‐DW‐STE sequences. Furthermore, preliminary investigations using a newly designed incoherent motion phantom underline the viability of this study. The presented sequence implementation is limited compared to nerve stimulation techniques 11 , 13 because of the relatively long time period between two RF excitations leading to only a coarse temporal resolution of SMAMs. For this, the onset and maximum CSA of the SMAM cannot be optimally estimated. Additionally, very short spontaneous muscular contractions may not be adequately visualized because of the limited temporal resolution of the MR sequence. In addition, the amplitude of the motion‐sensitizing gradients has a large impact on the measurements in terms of number of SMAMs and duration. However, no additional equipment is required to carry out fully non‐invasive MR examinations (without external nerve stimulation). Measurements were only conducted with the incoherent motion phantom in one specific direction (i.e., within the axial plane) and with a slow‐moving motion bead. Further studies evaluating the sensitivity of the MR sequence regarding frequency and duration of induced movements, as well as, the dependency of the angle to the imaging plane might be of interest. Therefore, an incoherent motion phantom with a more precise control allowing very small and fast movements has to be built. Evaluation was restricted on the two muscles soleus und gastrocnemius medialis because of the expected low activities in other muscular regions. However, further work in this field of research should also consider other muscles, like tibialis anterior, because spontaneous activities might be relevant especially in patient groups. Therefore, a study with patient groups potentially showing a distinct change in the activation pattern of motor units and the mechanical response is considered a reasonable next step further. In comparison to clinically established needle electromyography for the detection of spontaneous activity with a high temporal resolution of the myoelectric signal, this MR technique enables to visualize the temporal course of mechanical activities in a larger FOV. However, needle electromyography is much less expensive and established in the clinical setting. Clinical translation of the proposed MR technique seems possible, but its value is still to be determined.
5. CONCLUSION
The systematic MR studies on a fully controllable incoherent motion phantom as well as in vivo measurements in healthy subjects demonstrate that the acquisition of several points in time of the human spontaneous muscular twitch contraction (SMAM) using a multiple‐point in time DW‐STE MRI sequence is feasible. In addition, no external myoelectric stimulation is required to record the temporal course of a single spontaneous muscle twitch contraction. For imaging of the contraction time, large amplitude motion‐sensitivity gradients might be optimal with short temporal period between RF excitations. Faster sampling of k‐space is required to shorten the interval between two RF excitations allowing more sampling points within the acquisition window. However, this goes along with reduced spatial resolution and overall image quality. Furthermore, triggering the imaging sequence by recordable myoelectric signals might be challenging, but would improve the understanding of spontaneous muscular contraction in the human skeletal muscular system visible in MRI.
CONFLICT OF INTEREST STATEMENT
Nothing to declare.
Supporting information
Table S1. (A) Median number of images indicating SMAMs per intra‐series imaging point in time for all three measurement settings. Differences between all three measurement settings (p = 0.008, Friedman test) as well as between the 10 imaging time points for each measurement (P adj ≤0.001, Wilcoxon test) were significant. Pair‐wise test: †‡ (P adj <0.05, Wilcoxon test). (B) Median cross‐sectional areas of SMAMs in the unit of voxels for all three measurement settings. A significant difference was revealed for testing all three measurements (p = 0.006, Friedman test). (α: P adj = 0.012, Wilcoxon test, β: P adj = 0.011, Wilcoxon test).
Figure S1. (A) Photography of the disassembled incoherent motion phantom. (B) Schematic axial view of the incoherent motion phantom with two intra‐slice and four inter‐slice tubes. Intra‐slice tubes filled with water and cellulose (2%) were utilized for inducing incoherent motion with a motion bead (red) connected to thin filaments. An additional layer of rubber was inserted at the end of the intra‐slice tubes to reduce the exchange between surrounding water. Four inter‐slice tubes were filled with a soy‐lecithin solution for adjusting T1‐values mimicking relaxation times of different tissues enabling investigations regarding T1‐dependent signal intensity variation.
Figure S2. Overall percentage Event Count Maps (pECMs) visualized for all seven subjects and all four MR settings. The spatial distribution of the activation patterns within subjects remains similar using different sequence settings.
Figure S3. Exemplary calculation of the relative proportion of the individual subgroups of SMAMs for class: (A), (B), (C), (D) and (E).
Figure S4. Four different classes of visible SMAMs categorized by onset and duration in the course of the MR sequence. SMAMs of the 5th (MV) class are not visible in DWI because contraction and relaxation is not during the motion‐sensitive period of the MR sequence and therefore the relative amount cannot be depicted.
ACKNOWLEDGMENTS
This work was supported and funded by the German Research Foundation under grant SCHI 498/11‐2, YA 28/16‐2. Open Access funding enabled and organized by Projekt DEAL.
Schwartz M., Martirosian P., Fritz V., Steidle G., Yang B., and Schick F., “The dynamic course of spontaneous muscular contractions assessed using multiple‐point acquisition in diffusion‐weighted stimulated echo imaging,” Magnetic Resonance in Medicine 95, no. 1 (2026): 517–530, 10.1002/mrm.30576.
REFERENCES
- 1. Engel AG, Franzini‐Armstrong C. Myology. 2nd ed. McGraw‐Hill; 1994. [Google Scholar]
- 2. Nojszewska M, Gawel M, Szmidt‐Salkowska E, et al. Abnormal spontaneous activity in primary myopathic disorders. Muscle Nerve. 2017;56:427‐432. [DOI] [PubMed] [Google Scholar]
- 3. Bashford JA, Wickham A, Iniesta R, et al. The rise and fall of fasciculations in amyotrophic lateral sclerosis. Brain Commun. 2020;2:fcaa018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Steidle G, Schick F. Addressing spontaneous signal voids in repetitive single‐shot DWI of musculature: spatial and temporal patterns in the calves of healthy volunteers and consideration of unintended muscle activities as underlying mechanism. NMR Biomed. 2015;28:801‐810. [DOI] [PubMed] [Google Scholar]
- 5. Schwartz M, Steidle G, Martirosian P, et al. Spontaneous mechanical and electrical activities of human calf musculature at rest assessed by repetitive single‐shot diffusion‐weighted MRI and simultaneous surface electromyography. Magn Reson Med. 2018;79:2784‐2794. [DOI] [PubMed] [Google Scholar]
- 6. Schwartz M, Martirosian P, Steidle G, et al. Volumetric assessment of spontaneous mechanical activities by simultaneous multi‐slice MRI techniques with correlation to muscle fiber orientation. NMR Biomed. 2018;31:e3959. [DOI] [PubMed] [Google Scholar]
- 7. Heskamp L, Birkbeck MG, Hall J, et al. Whole‐body fasciculation detection in amyotrophic lateral sclerosis using motor unit MRI. Clin Neurophysiol. 2024;161:246‐255. [DOI] [PubMed] [Google Scholar]
- 8. Schwartz M, Martirosian P, Steidle G, et al. Measuring spontaneous muscular activities in neuromuscular Disease: preliminary results. Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM). ISMRM; 2020. [Google Scholar]
- 9. Birkbeck MG, Blamire AM, Whittaker RG, Sayer AA, Dodds RM. The role of novel motor unit magnetic resonance imaging to investigate motor unit activity in ageing skeletal muscle. J Cachexia Sarcopenia Muscle. 2021;12:17‐29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Birkbeck MG, Heskamp L, Schofield IS, Blamire AM, Whittaker RG. Non‐invasive imaging of single human motor units. Clin Neurophysiol. 2020;131:1399‐1406. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Heskamp L, Birkbeck MG, Whittaker RG, Schofield IS, Blamire AM. The muscle twitch profile assessed with motor unit magnetic resonance imaging. NMR Biomed. 2021;34:e4466. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Heskamp L, Miller AR, Birkbeck MG, et al. In vivo 3D imaging of human motor units in upper and lower limb muscles. Clin Neurophysiol. 2022;141:91‐100. [DOI] [PubMed] [Google Scholar]
- 13. Whittaker RG, Porcari P, Braz L, Williams TL, Schofield IS, Blamire AM. Functional magnetic resonance imaging of human motor unit fasciculation in amyotrophic lateral sclerosis. Ann Neurol. 2019;85:455‐459. [DOI] [PubMed] [Google Scholar]
- 14. Frahm J, Hanicke W, Bruhn H, Gyngell ML, Merboldt KD. High‐speed STEAM MRI of the human heart. Magn Reson Med. 1991;22:133‐142. [DOI] [PubMed] [Google Scholar]
- 15. Merboldt KD, Hanicke W, Bruhn H, Gyngell ML, Frahm J. Diffusion imaging of the human brain in vivo using high‐speed STEAM MRI. Magn Reson Med. 1992;23:179‐192. [DOI] [PubMed] [Google Scholar]
- 16. Merboldt KD, Hanicke W, Frahm J. Diffusion imaging using stimulated echoes. Magn Reson Med. 1991;19:233‐239. [DOI] [PubMed] [Google Scholar]
- 17. Frahm J, Merboldt KD, Hänicke W, Haase A. Stimulated echo imaging. J Magn Reson (1969). 1985;64:81‐93. [Google Scholar]
- 18. Haase A, Frahm J. NMR imaging of spin‐lattice relaxation using stimulated echoes. J Magn Reson (1969). 1985;65:481‐490. [Google Scholar]
- 19. Matthaei D, Frahm J, Haase A, Merboldt KD, Hanicke W. Multipurpose NMR imaging using stimulated echoes. Magn Reson Med. 1986;3:554‐561. [DOI] [PubMed] [Google Scholar]
- 20. Franconi F, Lethimonnier F, Sonier CB, Pourcelot L, Akoka S. Diffusion imaging with a multi‐echo MISSTEC sequence. J Magn Reson Imaging. 1997;7:399‐404. [DOI] [PubMed] [Google Scholar]
- 21. Baete SH, Boada FE. Accelerated radial diffusion spectrum imaging using a multi‐echo stimulated echo diffusion sequence. Magn Reson Med. 2018;79:306‐316. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Schwartz M, Martirosian P, Steidle G, Yang B, Schick F. Single‐Shot Multi‐Point Stimulated Echo Imaging for Analysis of Spontaneous Muscular Activities. in ESMRMB 2020 Online, 37th Annual Scientific Meeting, September 30–October 2: Lightning Talks/Electronic Posters/Clinical Review Posters/Software Exhibits 2020.
- 23. Schwartz M, Martirosian P, Steidle G, Yang B, Schick F. Imaging of the contraction pattern of spontaneous muscular activities using a multiple‐point diffusion‐weighted stimulated Echo sequence. Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM). ISMRM; 2023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Dan G, Sun K, Luo Q, Zhou XJ. Time‐dependent diffusion MRI using multiple stimulated echoes. Magn Reson Med. 2023;90:910‐921. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Dan G, Sun K, Luo Q, Zhou XJ. Single‐shot multi‐b‐value (SSMb) diffusion‐weighted MRI using spin echo and stimulated echoes with variable flip angles. NMR Biomed. 2024;37:e5261. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Layton KJ, Kroboth S, Jia F, et al. Pulseq: a rapid and hardware‐independent pulse sequence prototyping framework. Magn Reson Med. 2017;77:1544‐1552. [DOI] [PubMed] [Google Scholar]
- 27. Chen DQ, Marr RB, Lauterbur PC. Reconstruction from NMR data acquired with imaging gradients having arbitrary time dependence. IEEE Trans Med Imaging. 1986;5:162‐164. [DOI] [PubMed] [Google Scholar]
- 28. Hiepe P, Herrmann KH, Ros C, Reichenbach JR. Diffusion weighted inner volume imaging of lumbar disks based on turbo‐STEAM acquisition. Z Med Phys. 2011;21:216‐227. [DOI] [PubMed] [Google Scholar]
- 29. Jeong KE, Hsu EW, Jeong EK. Diffusion MRI using two‐dimensional single‐shot radial imaging (2D ss‐rDWI) with variable flip angle and random view ordering. Magn Reson Imaging. 2019;61:273‐284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Fritz V, Eisele S, Martirosian P, Machann J, Schick F. A straightforward procedure to build a non‐toxic relaxometry phantom with desired T1 and T2 times at 3T. MAGMA. 2024;37:899‐907. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Fritz V, Martirosian P, Machann J, Thorwarth D, Schick F. Soy lecithin: a beneficial substance for adjusting the ADC in aqueous solutions to the values of biological tissues. Magn Reson Med. 2023;89:1674‐1683. [DOI] [PubMed] [Google Scholar]
- 32. Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. Elastix: a toolbox for intensity‐based medical image registration. IEEE Trans Med Imaging. 2010;29:196‐205. [DOI] [PubMed] [Google Scholar]
- 33. Shamonin DP, Bron EE, Lelieveldt BP, et al. Fast parallel image registration on CPU and GPU for diagnostic classification of Alzheimer's disease. Front Neuroinform. 2013;7:50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Schwartz M, Martirosian P, Steidle G, Yang B, Schick F. Improved quantitative spatial analysis of spontaneous muscular activities using label uncertainty and feature analysis. Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM). ISMRM; 2024. [Google Scholar]
- 35. Milletari F, Navab N, Ahmadi SA. V‐net: fully convolutional neural networks for volumetric medical image segmentation. 2016 Fourth International Conference on 3D Vision (3DV). IEEE; 2016. [Google Scholar]
- 36. Ronneberger O, Fischer P, Brox T. U‐net: convolutional networks for biomedical image segmentation. Medical Image Computing and Computer‐Assisted Intervention – MICCAI 2015. Springer International Publishing; 2015. [Google Scholar]
- 37. Bai W, Suzuki H, Qin C, et al. Recurrent neural networks for aortic image sequence segmentation with sparse annotations. In: Frangi AF, Fichtinger G, Schnabel JA, Alberola‐López C, Davatzikos C, eds. 21st International Conference on Medical Image Computing and Computer Assisted Intervention, MICCAI 2018. Springer Verlag; 2018:586‐594. [Google Scholar]
- 38. Hochreiter S, Schmidhuber J. Long short‐term memory. Neural Comput. 1997;9:1735‐1780. [DOI] [PubMed] [Google Scholar]
- 39. Shi X, Chen Z, Wang H, Yeung D‐Y, Wong W‐k, Woo W‐c. Convolutional LSTM network: a machine learning approach for precipitation nowcasting. Proceedings of the 29th International Conference on Neural Information Processing Systems. Vol 1. MIT Press; 2015:802‐810. [Google Scholar]
- 40. Schwartz M, Steidle G, Martirosian P, et al. Longitudinal Analysis of Spontaneous Mechanical Activities in Resting Leg Musculature Assessed by Diffusion‐Weighted Imaging: Preliminary Results in Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM). Montreal. 2019.
- 41. Schwartz M, Martirosian P, Steidle G, Feiweier T, Yang B, Schick F. Capability of diffusion‐weighted stimulated‐Echo imaging to visualize spontaneous muscular activities in resting musculature. Annual Meeting of the International Society for Magnetic Resonance in Medicine (ISMRM). ISMRM; 2022. [Google Scholar]
- 42. Gollnick PD, Sjodin B, Karlsson J, Jansson E, Saltin B. Human soleus muscle: a comparison of fiber composition and enzyme activities with other leg muscles. Pflugers Arch. 1974;348:247‐255. [DOI] [PubMed] [Google Scholar]
- 43. Sanchez GN, Sinha S, Liske H, et al. In vivo imaging of human sarcomere twitch dynamics in individual motor units. Neuron. 2015;88:1109‐1120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Vandervoort AA, McComas AJ. A comparison of the contractile properties of the human gastrocnemius and soleus muscles. Eur J Appl Physiol Occup Physiol. 1983;51:435‐440. [DOI] [PubMed] [Google Scholar]
- 45. Walker FO, Cartwright MS, Wiesler ER, Caress J. Ultrasound of nerve and muscle. Clin Neurophysiol. 2004;115:495‐507. [DOI] [PubMed] [Google Scholar]
- 46. Walker FO, Donofrio PD, Harpold GJ, Ferrell WG. Sonographic imaging of muscle contraction and fasciculations: a correlation with electromyography. Muscle Nerve. 1990;13:33‐39. [DOI] [PubMed] [Google Scholar]
- 47. Reimers CD, Ziemann U, Scheel A, Rieckmann P, Kunkel M, Kurth C. Fasciculations: clinical, electromyographic, and ultrasonographic assessment. J Neurol. 1996;243:579‐584. [DOI] [PubMed] [Google Scholar]
- 48. Loram ID, Maganaris CN, Lakie M. Use of ultrasound to make noninvasive in vivo measurement of continuous changes in human muscle contractile length. J Appl Physiol (1985). 2006;100:1311‐1323. [DOI] [PubMed] [Google Scholar]
- 49. Wijntjes J, van Alfen N. Muscle ultrasound: present state and future opportunities. Muscle Nerve. 2021;63:455‐466. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1. (A) Median number of images indicating SMAMs per intra‐series imaging point in time for all three measurement settings. Differences between all three measurement settings (p = 0.008, Friedman test) as well as between the 10 imaging time points for each measurement (P adj ≤0.001, Wilcoxon test) were significant. Pair‐wise test: †‡ (P adj <0.05, Wilcoxon test). (B) Median cross‐sectional areas of SMAMs in the unit of voxels for all three measurement settings. A significant difference was revealed for testing all three measurements (p = 0.006, Friedman test). (α: P adj = 0.012, Wilcoxon test, β: P adj = 0.011, Wilcoxon test).
Figure S1. (A) Photography of the disassembled incoherent motion phantom. (B) Schematic axial view of the incoherent motion phantom with two intra‐slice and four inter‐slice tubes. Intra‐slice tubes filled with water and cellulose (2%) were utilized for inducing incoherent motion with a motion bead (red) connected to thin filaments. An additional layer of rubber was inserted at the end of the intra‐slice tubes to reduce the exchange between surrounding water. Four inter‐slice tubes were filled with a soy‐lecithin solution for adjusting T1‐values mimicking relaxation times of different tissues enabling investigations regarding T1‐dependent signal intensity variation.
Figure S2. Overall percentage Event Count Maps (pECMs) visualized for all seven subjects and all four MR settings. The spatial distribution of the activation patterns within subjects remains similar using different sequence settings.
Figure S3. Exemplary calculation of the relative proportion of the individual subgroups of SMAMs for class: (A), (B), (C), (D) and (E).
Figure S4. Four different classes of visible SMAMs categorized by onset and duration in the course of the MR sequence. SMAMs of the 5th (MV) class are not visible in DWI because contraction and relaxation is not during the motion‐sensitive period of the MR sequence and therefore the relative amount cannot be depicted.