Implementation of Two-pulse Phase-modulated (TPPM) 1H Decoupling in a Clinical MR Scanner for the Detection of the C1-glycogen Peak in 13C MRS

Hideto Kuribayashi 1,*, Toshiro Inubushi 2
PMCID: PMC12671239  PMID: 39828341

Abstract

Two-pulse phase-modulated (TPPM) 1H-decoupling pulse sequence repeats a pair of 180o RF pulses while changing the signs of the RF phase modulation angle and has been widely used for the 13C NMR of organic solids. TPPM was introduced into the 13C MRS pulse sequence on a clinical 3T MR scanner, and the 1H-decoupling performance was compared with conventional 1H-decoupling schemes using aqueous solutions containing glucose and oyster glycogen. The 13C C1-glucose peaks were 1H-decoupled using TPPM with B2 = 500 Hz, and the optimal RF phase modulation angle was up to 30o. Cycling sidebands were not observed when TPPM was used but were observed when WALTZ-16 was used. The 13C C1-glycogen peak was 1H-decoupled even with reducing TPPM duration to 8 ms, which reduced simulated specific absorption rate (SAR) to 39%. In conclusion, the TPPM 1H decoupling is applicable to clinical MR scanners, and the low-SAR sequence may be more valuable at 7T.

Keywords: 13C MRS, decoupling, muscle glycogen, specific absorption rate, two-pulse phase-modulated

Introduction

Glycogen has attracted a great deal of interest with relation to the energy metabolism for physical fatigue in skeletal muscle. The gold standard of measuring skeletal muscle glycogen levels is needle biopsy.1 However, as this method is invasive, it is not suitable for dynamic studies, such as the tracking of changes of glycogen levels along with skeletal muscle activity. For a non-invasive assessment, the 13C MRS method has been developed via detecting the C1-glycogen peak at 100.5 ppm that is clearly distinguishable from other signals in the in vivo spectrum,2 instead of the higher cluster peaks from glycogen C2-C5 in the 72–78 ppm range overlapped with the much higher C2-glycerol peak at 69.5 ppm.3 Despite the challenges in technical development for MR acquisition with a low-natural abundance ratio and low-gyromagnetic ratio, the 13C MRS has provided valuable insights into glycogen content and kinetics since it can acquire a series of spectra repeatedly from a single subject.4 For improving SNR and/or reducing scan time, high magnetic fields are advantageous,3,5–6 and the 13C peaks coupled with adjacent 1H spins are detected as singlets using 1H decoupling.2–6

The 1H decoupling is an essential technique for high-resolution 13C NMR in vitro and is achieved with RF irradiation at 1H resonance frequency during 13C FID acquisition. The technique has been developed to cover a wider range of 1H chemical shifts at high magnetic fields and is called broadband decoupling. An important advance in its development was the invention of the composite-pulse sequences, WALTZ.7 The WALTZ 1H decoupling is also often used to detect the C1-glycogen peak as a singlet from human skeletal muscle, but this still requires 5–15 mins at 3T.4 Although the doublet peak detected without 1H decoupling provides the heteronuclear J-coupling constant, which is another evidence for identifying the 13C C1-glycogen peaks, the peak heights are halved. Therefore, 1H decoupling is an essential technique given the need for reliable signal detection even when glycogen levels in muscles are low. The challenges of using WALTZ 1H decoupling for human 13C MRS are the limited decoupling field strength (B2) due to RF system specifications optimized for RF coils covering large sample volume and specific absorption rate (SAR) limitations, as well as the inhomogeneity of B2 when using surface coils. Although the WALTZ-16 decoupling scheme requires large RF flip angles of over 180o, de Graaf8 demonstrated the feasibility of the WALTZ-16 for human calf muscle 13C MRS at 4T with B2 = 500 Hz generated using a half-volume 1H-decoupling coil. Since the duration of a rectangle RF pulse for a 180o flip angle with the B2 is 1 ms, the duration for a single WALTZ-16 composite pulse is as long as 48 ms, which is too long to detect the 13C C1-glycogen peak with short-T2* in vivo (~4 ms).3 The RF safety of the 1H-decoupled 13C MRS for human subjects is often ensured by reducing the duty cycle of the 1H-decoupling period during 13C FID acquisition. However, if the shortened 1H-decoupling period is shorter than the duration of a single composite pulse, the decoupling compensation cycle is no longer effective.

Glycogen is a globular macromolecule with a diameter of 20–30 nm and a molecular weight of 106–107, and therefore has some physical properties of a solid. In 13C NMR of organic solids, which requires much higher 1H-decoupling power, the two-pulse phase-modulated (TPPM) 1H decoupling has been proposed and developed to increase the sensitivity of solid-state NMR with decoupling the heteronuclear scalar coupling and reducing cycling sidebands as well as cancelling the effects from strong 1H–1H dipolar interactions in a solid-state spin environment.9 Currently, the TPPM has been utilized as a more efficient method which achieves a higher SNR with a lower 1H-decoupling power than the continuous wave (CW) decoupling.9–10 The TPPM pulse sequence is one of the simplest composite pulses: a pair of 180° rectangular RF pulses while changing the positive and negative signs of the RF phase modulation angle. Even when used in clinical scanners, the TPPM 1H-decoupling period can be reduced on a pulse-pair basis, since the pulse pairs are completed within a few milliseconds, comparable to the T2* of the 13C C1-glycogen peak in vivo. Therefore, it would be valuable if the TPPM 1H decoupling could be applied to detect the 13C C1-glycogen peak from human skeletal muscle. In this study, the TPPM 1H-decoupling scheme was introduced into the 13C MRS pulse sequence for a clinical scanner. To begin with, a glucose phantom was used to optimize the RF phase modulation angle of the TPPM. Then, the line shape of 1H-decoupled 13C peaks and 1H-decoupling bandwidth were investigated to compare the performance of the TPPM with CW and WALTZ-16. Finally, an oyster glycogen phantom was used as a human surrogate to investigate SAR and the height of 1H-decoupled 13C peak to compare the TPPM with WALTZ-16.

Materials and Methods

All MR experiments were carried out using a 3T clinical scanner (Skyra; Siemens Healthineers, Erlangen, Germany) equipped with broadband transmitters and receivers. A multi-channel RF coil was built for human calf muscle 13C MRS measurement with 1H imaging and decoupling capability based on Ref. 6. The linear-13C/quadrature-1HRF coils (13C: a 75-mm loop, 1H: a pair of 95-mm loops) were formed on the surface of semi-cylindrical acrylic resin tubes with radii of 71 mm and 77 mm, respectively, shown in Supplementary Fig. 1. The coils were connected to the MR scanner via an interface box (Takashima Seisakusho, Tokyo, Japan) containing low-pass/high-pass filters, preamplifiers, and transmit/receive switches. A band-pass filter for 13C resonance frequency (FSY microwave, Columbia, MD, USA) was inserted between the 13C loop coil and the interface box. The materials of the coil housing, such as the acrylic resin, rubber, and wood, have been confirmed to generate no background 13C signals around the chemical shift of the C1-glycogen peak. Two aqueous solution phantoms containing [1-13C]glucose (ISOTEC, Miamisburg, OH, USA) and oyster glycogen (Nacalai Tesque, Kyoto, Japan) were prepared and were scanned at room temperature. The molar concentration of the C1-carbons in the glycogen phantom was ~150 mM.

Pulse sequences with TPPM and WALTZ-16 decoupling schemes were prepared as research prototypes based on the product pulse sequence for non-localized 13C MRS with 1H-decouping capability. The same 1H-decoupling amplitude of B2 = 500 Hz (i.e., a 1-ms pulse duration for each 180° pulse element) was used in those 1H-decoupling schemes. For the glucose measurement, the RF offset for 1H decoupling was set at 4.93 ppm, which is at the midpoint between the H1-α- and H1-β-glucose peaks. The decoupling amplitude was kept constant during 13C FID acquisition. To optimize the TPPM, glucose 13C spectra were acquired using different RF phase modulation angles from 10° to 90° in 10° steps with 4 averages. To observe the 1H-decoupling bandwidth of those decoupling schemes, glucose 13C spectra were acquired using different 1H-decoupling RF offsets within ±1 kHz in 100-Hz steps. Other identical 13C MRS parameters were TR = 15s, the flip angle for 13C excitation = 90°, FID acquisition time = 85 ms and data points = 256, i.e., acquisition bandwidth = 3 kHz. The spectral acquisitions were repeated 3 times. For the glycogen measurement using the TPPM or the WALTZ-16, the duration of 13C FID acquisition was divided into 2 1H-decoupling phases as shown in Fig. 1a. The first phase (Phase I) consists of the cycles of the TPPM or WALTZ-16 scheme, i.e., the duration is an integer multiple of 1 composite cycle, with B2 = 500 Hz. The aims for having the second phase (Phase II) of a ramping down of the 1H-decoupling amplitude are to reduce SAR and to avoid detecting noise spikes when the decoupling power is turned off momentarily. Glycogen 13C spectra were acquired using different Phase I durations per the duration of 13C FID acquisition (85 ms) from 10% to 100% in 10% steps using the TPPM and those of 50% and 100% (= 1 and 2 composite pulse cycles, respectively) using the WALTZ-16. The RF offset for 13C transmit and receive was shifted 100 Hz from the peak frequency of C1-glycogen to distinguish the signal from artifacts that may be observed at the center of the acquisition bandwidth in a low sensitivity spectrum. The RF offset for 1H decoupling was set at 5.4 ppm on the peak of H1-glycogen with α-1,4-glycosidic linkages.11 Other identical 13C MRS parameters for the glycogen spectra were TR = 200 ms, the flip angle for 13C excitation = 90°, which had been optimized using the glycogen phantom, and 256 averages. RF irradiation at 1H resonance frequency from the end of the 13C FID acquisition to the next 13C RF excitation, so-called nuclear Overhauser effect (NOE) pulses, was not applied in this study. SARs were simulated on the scanner for an adult male with a height of 170 cm and a weight of 70 kg and were recorded at the end of the measurements.

Fig. 1.

Fig. 1

(a) TPPM 1H-decoupling scheme for the 13C MRS pulse sequence. For the glycogen measurements, the decoupling power was linearly decreased from the middle to the end of 13C FID acquisition (Phase II). For the glucose phantom, the decoupling power was kept constant during 13C FID acquisition, i.e., the Phase II was not applied. (b) The C1-glucose peaks in 13C spectra from the glucose phantom as a function of the RF phase modulation angle for TPPM 1H decoupling. The spectra are shown as the phase-sensitive display mode and the display width is 1.5 kHz (= 48.4 ppm). (c) The peak height as a function of the RF phase modulation angle in 10° increments. The peak heights were measured in the absolute-value display mode and averaged among 3 measurements. The corresponding standard deviations are smaller than the size of the points on the graph. TPPM, two-pulse phase-modulated.

For spectral analysis of the glucose spectra, the maximum peak height in the spectrum was measured in the absolute-value display mode on the operational software for the scanner, and the average of 3 measurements was calculated. The decoupling bandwidth was measured as the bandwidth where the average height of the 1H-decoupled signal exceeded 90% of the maximum value. For spectral analysis of the glycogen spectra, the peaks in the Fourier-transformed spectra with 20-Hz line broadening were fitted as Lorentzian curves using the Advanced Method for Accurate, Robust, and Efficient Spectral fitting (AMARES) algorithm12 in the jMRUI software.13 To compare the effect of 1H decoupling, the height of the fitted 13C C1-glycogen peak was evaluated rather than the amplitude (area under the curve) and was calculated by dividing the amplitude by the linewidth, both of which were estimated by the fitting. Relative SAR (relSAR) was calculated by dividing the simulated SAR with a specific Phase-I duration by the SAR with full 1H irradiation during 13C FID acquisition.

Results

The TPPM 1H decoupling was operated using the clinical 3T scanner as shown below. Figure 1b shows a series of the 13C C1-glucose peaks from the glucose phantom as a function of the RF phase modulation angle for the TPPM 1H decoupling. The 13C peaks are 1H-decoupled and elevated at RF phase modulation angles up to 30° (Fig. 1c). Based on the results, 30° was chosen for the RF phase modulation angle in the subsequent experiments when using the TPPM 1H decoupling. Figure 2 shows the 13C C1-glucose peaks from the glucose phantom using the TPPM, CW and WALTZ-16 1H-decoupling schemes. Cycling sidebands, which symmetrically flank the main decoupled peak, are only observed in the spectra using the WALTZ-16. The height of the largest cycling sideband is more than 8% of that of the main 1H-decoupled 13C C1-β-glucose peak. Figure 3a shows a series of the 13C C1-glucose peaks from the glucose phantom as a function of the 1H RF offset for the different 1H-decoupling schemes. The 1H-decoupling bandwidth of the TPPM is slightly wider than that of the CW (<200 Hz in both) and narrower than that of the WALTZ-16 (<900 Hz, Fig. 3b). Figure 4 shows the results using the glycogen phantom. The 13C C1-glycogen peaks are 1H-decoupled using the TPPM (Fig. 4a). Using the TPPM, shortening Phase I to 8 ms does not reduce the height of the 1H-decoupled 13C C1-glycogen peak (Figs. 4b and 4c). It reduced the simulated SAR to 39% compared to the SAR with full 1H irradiation during 13C FID acquisition. No cycling sidebands are detected in 1H-decoupled 13C glycogen spectra (Fig. 4c).

Fig. 2.

Fig. 2

The C1-glucose peaks in 13C spectra from the glucose phantom using TPPM, CW and WALTZ-16 1H-decoupling schemes. The spectra are shown as the phase-sensitive display mode. The spectra in (b) are vertically magnified by a factor of 5. CW, continuous wave; TPPM, two-pulse phase-modulated.

Fig. 3.

Fig. 3

(a) The C1-glucose peaks in 13C spectra from the glucose phantom as a function of the 1H RF offset using TPPM, CW and WALTZ-16 1H-decoupling schemes. The spectra are shown as the phase-sensitive display mode and the display width is 1.5 kHz (= 48.4 ppm). (b) The peak height as a function of the 1H RF offset in 100-Hz increments. The heights of the maximum C1-glucose peaks were measured in the absolute-value display mode and were averaged among three measurements. The corresponding standard deviations are smaller than the size of the points on the graph. CW, continuous wave; TPPM, two-pulse phase-modulated.

Fig. 4.

Fig. 4

(a) 13C spectra from the oyster glycogen phantom with (top) and without (bottom) TPPM 1H decoupling. (b) The average height of the 13C C1-glycogen peaks using different 1H-decoupling schemes and Phase I durations (see Fig. 1a). T_n and W_n indicate that the 1H-decoupling scheme (T: TPPM, W: WALTZ-16) and Phase I = n%. Error bars indicate the standard deviations (n = 3). (c) Representative 13C spectra with different Phase I durations of TPPM (top) and WALTZ-16 (bottom) 1H decoupling. relSAR is calculated by dividing the simulated SAR by the SAR of Phase I = 100% with the same decoupling scheme. Line broadening = 20 Hz. TPPM, two-pulse phase-modulated

Discussion

This study demonstrated the implementation of the TPPM 1H decoupling in the clinical 3T MR scanner. Despite the large difference in the B2 amplitude required for solid-state NMR (>50 kHz)9–10 and clinical MRI (<1 kHz),3,5,8,14–17 the TPPM pulse sequence in the clinical MRI was able to decouple proton spins with the same characteristics as the solid-state NMR, where small RF phase modulation angles are effective for 1H decoupling.9–10 This study was conducted with B2 = 500 Hz, commonly chosen for human studies.3,8,17 The decoupling bandwidth for the WALTZ-16 relative to B2 showed similar values in the study by de Graaf (=1.64)8 and the present study (<1.8), both of which used half-volume 1H coils. Thus, our experimental results using the TPPM 1H decoupling could be reproduced on the MR system by de Graaf.8 The disadvantage of the TPPM 1H decoupling in clinical MR scanners is the narrow decoupling bandwidth due to low B2. Although the development of 1H-decoupling schemes has been concentrated in the direction of expanding the decoupling bandwidth, de Graaf8 discussed that the direction for in vivo MRS may differ from that for analytical chemistry. The 1H-decouping of the 2 13C C1-glucose peaks in solution state using TPPM was successful, where the chemical shift difference of the 1H H1-α- and H1-β-glucose peaks is 0.6 ppm, the 1H-decoupling bandwidth of the TPPM was estimated to be wider than 70 Hz in this study. Therefore, the TPPM 1H decoupling may be sufficient when only specific peaks are focused to be analyzed, such as the case of 13C C1-glycogen peak from the muscle.

The glycogen phantom study demonstrated that the 13C C1-glycogen peaks were able to be 1H-decoupled using the TPPM without compromising the quality of 1H decoupling, even when the duration of Phase-I was shortened to 8 ms in the pulse sequence used. In the measurement condition, the shortened duration of the TPPM 1H decoupling reduced the SAR to less than 40% compared to full 1H irradiation during 13C FID acquisition. The SAR reduction using the TPPM would be much more valuable at magnetic fields higher than 3T. As discussed by Stephenson et al.,18 detecting the C1-glycogen peak at 7T results in a 1.6-fold higher SNR compared to 3T and can reduce scan times by over 50%. However, including other applications using the WALTZ-16 at 7T,5,17 TRs were always set much longer (0.8–1.0s) than the estimated T1 of the 13C C1-glycogen peak in vivo (~0.2s)3 for the primary aim of reducing the SAR. If the narrow 1H-decoupling bandwidth is sufficient, the WALTZ-16 would be replaced with the TPPM. By further shortening the TR and averaging a large number of FIDs within a limited scan time, the noise level can be reduced and the SNR improved. If no noise spikes are detected in the MR system when the decoupling power is turned off momentarily, the SAR can be further reduced without using Phase-II in the pulse sequence. Another advantage of the TPPM at 7T may be the small maximum RF flip angle: 180° for the TPPM versus 360° for the WALTZ-16. That is extremely important at 7T, where uniform RF excitation is difficult to achieve at large RF flip angles. Clinically approved 7T MRI for humans is not widely available, and studies of muscle glycogen using 1H-decoupled 13C MRS have even fewer reports at 7T.3,5,17–18 13C MRS using the low-SAR TPPM 1H-decoupling pulse sequence would be highly suitable for studying glycogen metabolism in in vivo human muscle at 7T.

Unlike the WALTZ-16, the cycling sidebands were not observed in the 1H-decoupled 13C spectra from the glucose phantom using the TPPM as well as the CW. The reduction of the undesirable cycling sidebands was attempted with the development of 1H-decoupling schemes for clinical applications.14 The cycling sidebands are particularly amplified when an 1H-decoupling cycle based on composite pulses is longer than the inverse of the heteronuclear J-coupling constant,14 e.g., 6 ms for the C1-peak of glucose and glycogen, which suggests low efficiency of the 1H-decoupling compensation cycle. The TPPM 1H-decoupling cycle at 500-Hz B2 is 2 ms, which may be the reason why no cycling sidebands were observed using the TPPM in this study. The cycling sidebands are also amplified under B2 inhomogeneity even in the solid-state NMR.9 The small RF flip angle requirement of the TPPM may be another reason for no cycling sidebands having been observed. No cycling sidebands were detected in 13C glycogen spectra using TPPM or WALTZ-16 due to high noise levels, however the peak intensities tend to be lower using WALTZ-16, likely due to leakage of signal amplitude into the cycling sidebands.

The TPPM pulse sequence has been improved9–10 and is widely used for 1H decoupling in 13C NMR of organic solids. Since its improvements mainly focus on the dipolar interactions, which is not likely to be strong in liquid state or biological samples, and since the composite pulse duration of such evolved pulse sequences tends to be long, only the TPPM pulse sequence was implemented in this study. The evolved TPPM pulse sequences have been implemented for water suppression in 1H-detected multidimensional NMR for protein structure determination19 and for electron spectral saturation in dynamic nuclear polarization.20 The aim of implementing the TPPM in the latter study was to minimize heating of the sample, which is similar to our in vivo 13C MRS study. In addition to the 1H-decoupling application demonstrated in this study, the TPPM pulse sequence could also be used for other purposes in in vivo MR studies. Possible candidates would be water suppression for 1H MRS and narrow spectral saturation for CEST imaging, magnetization transfer 31P MRS and j-editing 1H MRS.

Conclusion

The TPPM pulse sequence is applicable to clinical MR scanners as a 1H-decoupling method to detect the C1-glycogen peak in 13C MRS. The performance of this sequence has been confirmed using the RF coils designed for human calf muscle, and therefore the sequence should be suitable for future in vivo applications. The low-SAR sequence may be more valuable at 7T.

Footnotes

Conflicts of Interest: Hideto Kuribayashi is an employee of Siemens Healthcare K.K.. Toshiro Inubushi has no conflicts of interest to declare.

Supplementary Material

Supplementary figures are available online.

Supplementary Figure 1.

Supplementary Figure 1

RF coils for 13C MR measurements of human calf muscle with 1H imaging and decoupling capabilities. The linear-13C/quadrature-1H RF coils (13C: a 75-mm loop, 1H: a pair of 95-mm loops) were formed on the surface of semi-cylindrical acrylic resin tubes with radii of 71 mm and 77 mm, respectively.

References

  • 1.Greene J, Louis J, Korostynska O, Mason A. State-of-the-art methods for skeletal muscle glycogen analysis in athletes – The need for novel non-invasive techniques. Biosensors (Basel) 2017; 7:11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Jue T, Rothman DL, Tavitian BA, Shulman RG. Natural-abundance 13C NMR study of glycogen repletion in human liver and muscle. Proc Natl Acad Sci USA 1989; 86:1439–1442. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Serés Roig E. Toward structure and metabolism of glycogen C1-C6 in humans at 7T by localized 13C MRS using low-power bilevel broadband 1H decoupling. NMR Biomed 2024; 37:e5241. [DOI] [PubMed] [Google Scholar]
  • 4.Shiose K, Takahashi H, Yamada Y. Muscle glycogen assessment and relationship with body hydration status: A narrative review. Nutrients 2022; 15:155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Goluch S, Frass-Kriegl R, Meyerspeer M, et al. Proton-decoupled carbon magnetic resonance spectroscopy in human calf muscles at 7T using a multi-channel radiofrequency coil. Sci Rep 2018; 8:6211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Roig ES, Magill AW, Donati G, et al. A double-quadrature radiofrequency coil design for proton-decoupled carbon-13 magnetic resonance spectroscopy in humans at 7T. Magn Reson Med 2015; 73:894–900. [DOI] [PubMed] [Google Scholar]
  • 7.Shaka AJ, Keeler J, Freeman R. Evaluation of a new broadband decoupling sequence: WALTZ-16. J Magn Reson 1983; 53:313–340. [Google Scholar]
  • 8.de Graaf RA. Theoretical and experimental evaluation of broadband decoupling techniques for in vivo nuclear magnetic resonance spectroscopy. Magn Reson Med 2005; 53:1297–1306. [DOI] [PubMed] [Google Scholar]
  • 9.Bennett AE, Rienstra CM, Auger M, Lakshmi KV, Griffin RG. Heteronuclear decoupling in rotating solids. J Chem Phys 1995; 103:6951–6958. [Google Scholar]
  • 10.Ernst M. Heteronuclear spin decoupling in solid-state NMR under magic-angle sample spinning. J Magn Reson 2003; 162:1–34. [DOI] [PubMed] [Google Scholar]
  • 11.Zang L-H, Rothman DL, Shulman RG. 1H NMR visibility of mammalian glycogen in solution. Proc Natl Acad Sci USA 1990; 87:1678–1680. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Vanhamme L, van den Boogaart A, Van Huffel S. Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson 1997; 129:35–43. [DOI] [PubMed] [Google Scholar]
  • 13.Naressi A, Couturier C, Devos JM, et al. Java-based graphical user interface for the MRUI quantitation package. MAGMA 2001; 12:141–152. [DOI] [PubMed] [Google Scholar]
  • 14.Barker PB, Golay X, Artemov D, Ouwerkerk R, Smith MA, Shaka AJ. Broadband proton decoupling for in vivo brain spectroscopy in humans. Magn Reson Med 2001; 45:226–232. [DOI] [PubMed] [Google Scholar]
  • 15.Li S, Yang J, Shen J. Novel strategy for cerebral 13C MRS using very low RF power for proton decoupling. Magn Reson Med 2007; 57:265–271. [DOI] [PubMed] [Google Scholar]
  • 16.Li S, An L, Yu S, et al. 13C MRS of human brain at 7 tesla using [2-13C]glucose infusion and low power broadband stochastic proton decoupling. Magn Reson Med 2016; 75:954–961. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Heinicke K, Dimitrov IE, Romain N, et al. Reproducibility and absolute quantification of muscle glycogen in patients with glycogen storage disease by 13C NMR spectroscopy at 7 Tesla. PLoS One 2014; 9:e108706. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Stephenson MC, Gunner F, Napolitano A, et al. Applications of multi-nuclear magnetic resonance spectroscopy at 7T. World J Radiol 2011; 3:105–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Lamley JM, Iuga D, Öster C, et al. Solid-state NMR of a protein in a precipitated complex with a full-length antibody. J Am Chem Soc 2014; 136:16800–16806. [DOI] [PubMed] [Google Scholar]
  • 20.Redrouthu VS, Vinod-Kumar S, Mathies G. Dynamic nuclear polarization by two-pulse phase modulation. J Chem Phys 2023; 159:014201. [DOI] [PubMed] [Google Scholar]

Articles from Magnetic Resonance in Medical Sciences are provided here courtesy of Japanese Society for Magnetic Resonance in Medicine

RESOURCES