. 2025 Dec 19;26(1):15. doi: 10.3390/s26010015

How the Level of Noise Affects Temporal Accuracy of a QRS Detector—Case Study

Wojciech Reklewski 1, Piotr Augustyniak 1,*
Editors: Georg Fischer1, Yvonne Tran1
PMCID: PMC12787906  PMID: 41516448

Abstract

Background: QRS complex detection is a key processing step of automated ECG analysis and determines its overall quality. The purpose of this paper is to study the detection performance of probably the most frequently implemented ready-to-use QRS detector in the presence of noise and with tightened temporal tolerance of detection points. Methods: We applied commonly used detection statistics (Detection Error Rate, Sensitivity, Positive Predictive Value, and F1 score), but re-defined true positive detection based on variable time jitter between detected and reference points. We also applied a controlled level of mixed noise to assess the detector’s performance in true-to-life conditions. Results: We found the following: (1) the detector under test showed a considerable drop in quality when reducing the jitter between 97.23 ms (DER = 8.08%) and 86.12 ms (DER = 67.22%), which means that the detection points’ time series are not accurate enough to be directly used for ECG time analysis; (2) with jitter allowed to 163.90 ms and an increasing noise level (SNR from 20 dB to −7.96 dB), the detection quality drops (DER from 0.98% to 57.13% respectively); however, an analysis of individual files revealed records, where the algorithm performs better in the presence of noise; (3) with a step-by-step code execution analysis of ECG strips where better performance was the most prominent, the imprecise definition of the local maximum was the cause of DER errors. Conclusions: Our research clearly indicates that selecting a QRS-detection algorithm based solely on DER, Se, and PPV detection statistics may be incorrect. Two equally important detection quality parameters are the change in the DER error rate with tightened requirements of jitter and robustness of the detection statistics DER, Se, and PPV to noise level variations (algorithm’s detection points immunity to noise).

Keywords: QRS detector, electrocardiogram, detection noise immunity, detection jitter

1. Introduction

QRS complex detection is a crucial first processing step in almost any automated ECG analysis [1]. The end result of ECG signal processing includes heart rate determination, heart rate variability determination, heartbeat classification, ECG signal fiducial points determination, on-line patient heart monitoring, and other applications. Hence, the accuracy of QRS detection is of primary importance as any errors from the QRS detection stage will be magnified or propagated to consecutive processing steps negatively influencing the end result of the whole processing chain.

The techniques for automatic QRS detection have been actively developed since the 1950s, and many approaches have been proposed [2,3,4], alongside the techniques for heartbeat classification and automated ECG analysis [5,6,7,8,9,10]. The main difficulties faced by QRS detectors are the wide variety of ECG waveforms, especially in cases of heart disease, and various types of interference present in the ECG signal. There are also technical limitations and considerations influencing the effectiveness of QRS detection: sampling rate, ADC resolution, power consumption considerations (especially important for mobile and wearable ECG acquisition systems) [3,4], hardware platform processing capabilities (especially important for real time operation) [3,4].

Despite the long history of research, only recently, the authors have pointed out that, in addition to the standard statistics of correctly detected QRS complexes, an important feature of the detector is the precision of indicating the temporal location of the detection point, and it is most desirable to maintain high accuracy regardless of the noise level. High temporal precision makes fiducial points directly usable for (1) comparing R-waves in heart beats classification process, (2) calculations of rhythm variability parameters such as HRV and HRT, (3) assessments of S-T segments for ischemia monitoring, and many others. No need for additional resynchronization procedure (such as [11]) is also beneficial to reduce the computational complexity, execution time and power requirements.

In [12], we presented an analysis of the effect of detector time tolerance (DTT) on the detection statistics of QRS complex detectors. In [13], we compared four algorithms, including one of our own, in terms of the sensitivity of their detection statistics for individual heart beat morphologies to gradually increase accuracy requirements in the presence of noise with different signal-to-noise ratios (SNRs). Overall, for all four analyzed detectors, the QRS detection accuracy decreased with increasing noise levels and tightening DTT values. It was also shown that the QRS detection accuracy depends on the heartbeat type and is specific to each detector, with some detectors performing better for a given heartbeat type.

In the current work, our objective was to focus on the specific case of the modern implementation of the Pan Tompkins algorithm [14] available on GitHub version 1.3.5 [15] and described in [13] as Alg-4. Following intuition and fragmentary results obtained in [13], we hypothesize that the temporal stability of detection points (i.e., the reproducibility of detection points at the same locations in QRS complexes of the same type) deteriorates linearly with an increasing noise level (i.e., decreasing SNR). To this end, we increased the density of added noise levels and the expected accuracy of true positive detection. In addition to the comprehensive analysis performed for all records in the ECG database within the DTT range where this influence was greatest, we analyzed individual records’ contributions to change in the total DER. Surprisingly, for some records, adding controlled noise to the ECG signal improves the detection statistics of the Alg-4 algorithm for certain DTT values. We therefore analyzed the step-by-step execution of the algorithm for specific strips of the database record with the greatest influence to pinpoint the computational procedure used in Alg-4 that results in the observed non-linear and non-monotonic relation.

Regarding the specific contribution of the present paper, (1) we stress that applying the temporal tolerance analysis may turn over QRS detectors’ rankings based solely on detection statistics, and (2) we suggest revisiting any research related to temporal ECG parameters (such as HRV) based on inaccurate QRS-detection algorithms.

The remainder of this paper is organized as follows: In Section 2 we will first define benchmarks for detection performance. Then, we will look at known schemes for noise removal and known techniques for correcting the location of the R Peak. We analyze how others rate the influence of noise. With Section 3, we propose a method of how to systematically add artificial noise and what its properties should be. We then, in Section 4, analyze whether there is a monotonic relation between an increase in the noise level and the degradation of detection performance. We will see that this is not always given. As this is a surprising result, we dig for the root cause of this strange behavior. Based on our findings, we propose adjusted criteria for characterizing the detection performance of a QRS detector.

2. Related Work

QRS detection is the first step in automatic ECG analysis; it is followed by additional blocks necessary to achieve the goal of signal processing. This can be HRV parameter analysis, fiducial point determination, heartbeat classification, etc. All blocks and components should fit with each other, and there are many aspects to achieve the goals of ECG signal processing; therefore, we start this section with an overview of classical and new metrics to assess the QRS detection accuracy (and specifically temporal detection accuracy), followed by the following: noise-removal techniques, corrections of R-peak detection points, HRV accuracy, and finally the influence of noise on detection accuracy.

2.1. QRS Detection Benchmarking

The effectiveness of QRS complex detection is typically assessed using a classic binary classification process [2,16]. Reference beat labels (database R peak annotations) are paired with algorithm-produced beat labels (R peaks identified by the algorithm). In order to pair the labels, the maximum time difference between reference the beat label and the algorithm beat label must not exceed a specific time window—the detection time tolerance. In [16], detection time tolerance is specified at 150 ms, whereas in the literature, there are various values: 27.8 ms [17] (ten times sampling period, for MIT-BIH AD [18,19] it is 10 × 2.78 = 27.8 ms) and from 60 ms to 160 ms [12].

Pairing the algorithm’s beat label with the corresponding reference beat label defines binary classification as follows: True Positive (TP): a paired reference beat label and algorithm beat label within the detection time tolerance, False Positive detection (FP): algorithm beat label without a paired reference beat label, False Negative detection (FN): reference beat label without a paired algorithm beat label. The TP, FP, and FN values are then used to calculate the binary classifier metrics: Detection Error Rate (DER), Sensitivity (Se), Positive Predictive Value (PPV), and F1-score (F1), as follows:

DER=(FP+FN)/(TP+FN) (1)
Se=TP/(TP+FN) (2)
PPV=TP/(TP+FP) (3)
F1=2·PPV·Se/(PPV+Se) (4)

where:

  • FP—False Positive,

  • FN—False Negative,

  • TP—True Positive.

Recently, Porr and Macfarlane [17] proposed a new QRS -detection accuracy benchmarking measure, which integrates the detection time tolerance with the F1-score. They recognized that due to the wide detection time tolerance (for example 150 ms) many QRS-detection algorithms proposed in the literature achieve F1-scores very close to 100%. With small differences in accuracy results, it is not possible to realistically compare and differentiate the proposed QRS-detection techniques. The authors proposed a new benchmarking measure called JF. The process to calculate JF consists of several steps. The first step is the determination of TPJF, FPJF, and FNJF. The pairing for calculating TPJF, FPJF, and FNJF is performed without a set width window of detection time tolerance. The basis for pairing is not the window of fixed width (for example 150 ms) but proximity; the nearest algorithm beat label and reference beat label are paired. Any unpaired algorithm beat label between two matched pairs is considered FPJF, and any unpaired reference beat label between two matched pairs is considered FNJF. All matched pairs are counted as TPJF. For all TPJF over the database record, the average time distance between reference and algorithm labels is calculated and is called the average temporal jitter ∆¯. As a next step the average temporal jitter is converted into a performance measure between 0 and 100%:

f(∆¯)=11+∆¯12 ms (5)

According to (6), the average temporal jitter of 12 ms will produce a performance measure of 0.5 or 50%. As a next step, the F1JF-score is calculated according to (4) with TPJF, FPJF, and FNJF. Finally, the JF is calculated:

JF=F1 JF·f∆¯·100% (6)

In our opinion, the proposed evaluation JF has a strong point of integrating temporal accuracy. It can be criticized for joining Se and PPV into one metric F1 neglecting that FP and FN have vastly different impacts on the downstream processing chain. It can be solved by defining the JSe and JPPV in an analogous way as F1 is converted into JF.

2.2. Noise Removal

Noise removal is often the first step in the QRS-detection algorithm. The authors in [20] describe the types of noise present in the ECG signal then review existing denoising and preprocessing techniques. Then, they review data transformation and detection techniques and list 15 various ECG signal test databases followed by noise parameters, detection accuracy evaluation parameters. The goal of removing the noise from the ECG signal is to improve the accuracy of QRS detection. As there are many types of noise, the study showed that there is no single technique that is effective for all noise conditions. Also, each denoising operation has some impact (distortion) on the underlying ECG signal. And then, review 38 techniques of QRS detection. In conclusion, the authors state that because modern health care requires wearable devices, only a few from reviewed QRS detection algorithms have been tested in a mobile scenario, a noise and power/complexity requirement.

In [21], the authors define and review six main domains of noise-removal techniques: Empirical Mode Decomposition (EMD), Deep-learning autoencoder models (DAEs), wavelet-based models (WT), Sparsity-based models, Bayesian-filter-based models, Hybrid Models. Describe classification of noise into four types: base-line wander, power-line interference, muscle artefacts, channel noise. The types of noise are described and techniques of removal alongside their performance. In summary, the authors give their assessment of which type of filter is best for which type of noise.

In contrast to numerically complex methods proposed in [21,22], the authors present simple practical digital filter implementation to suppress power grid noise using integer coefficients filters. The proposed filter does not introduce the ECG shape deformations and can be used in power-restricted mobile ECG analysis devices.

2.3. Corrections of R-Peak Detection Points

The temporal accuracy of R-peak detection is of primary importance for HRV analysis. The authors in [23] observed that filtering in the preprocessing stage introduces time lag in the filtered signal, with delay equal to the group-delay of employed filters. Further, they observed that many algorithms set the time location of the R-peak at the maximum amplitude of the filtered signal, resulting in constant per beat slackness. Many QRS-detection algorithms do not provide compensation for this slackness, so the authors propose compensation as a first step to improve QRS-detection accuracy. As a second step to improve the QRS-detection temporal accuracy, the authors propose a process called the Slackness Reduction Algorithm. After locating the approximate location of the R-peak in the filtered signal, the algorithm searches in the original ECG signal for the more accurate location of the R-peak by searching for the highest absolute differential maximum in the vicinity of the maximum found in the filtered signal. The authors evaluated the results of this modification on three algorithms. They found that the average slackness of this algorithms for MIT-BIH AD was 9 ms for normal beats and 13 ms for abnormal beats. With the Slackness Reduction Algorithm, they reduced the average slackness to 4 ms and 7 ms, respectively.

The sampling interval influences the accuracy of the R-peak location necessary for precise HRV measurements. For example, when the ECG signal is sampled with an interval of 8 ms (fs = 125 Hz) without modification, that would be a maximum accuracy of HRV. However, it is still possible to maintain the accuracy of the R-peak location on the level of 1–2 ms for precise heart rhythm measurements with the process proposed in [11]. The author reported that by employing simple calculations (eight multiplications, nine additions, and one division on floating point data) for one R-peak, it is possible to achieve the precision level of 1 to 2 ms. The author employed approximation of the R-wave with the quadratic function with the length of approximation of 32 ms. This way, with little additional computation, it is possible to achieve higher temporal accuracy than the sampling interval.

2.4. Heart Rate Variability

Accurate detection of QRS complexes is essential for reliable HRV computation. In [24], the authors analyze how the QRS-detection accuracy of various QRS detectors impacts the errors in HRV metrics. They analyze the performance of eight QRS detectors on ECG data from the Glasgow University Database (GUDB) [25]. The GUDB contains ECG signals from 25 subjects recorded under five conditions: sitting, walking, performing a math test, using a handbike, and jogging. In addition, each ECG record contains two simultaneously recorded versions, one recorded via elastic chest strap and one with ECG electrodes. The ECG data was therefore recorded under vastly different levels of noise and artifacts. The 23 HRV metrics were computed based on the outputs from eight QRS detectors and compared with HRV metrics computed based on database annotations. The results show that some detectors perform better than others and none is the best for all tested conditions. Another result is that the relationship between the QRS-detection accuracy and HRV accuracy is metric-dependent and nonlinear. In summary, the authors call for further research to investigate the impact of signal processing and RR interval correction techniques to further improve the robustness of HRV calculations in real-life noisy settings.

In [26], the authors analyzed the performance of a Modular Analysis System (MEANS) to assess its accuracy in heartbeat classification and to study the effects of undetected non-sinus heart beats on HRV parameters. The ECG data used in the study consisted of 20 min ECG records from 1674 patients from the CARLA study (CARdiovascular disease, Living and Ageing in Halle). The heartbeat classification was performed, into three categories: normal sinus rhythm, supraventricular systoles, ventricular systoles. The MEANS system correctly classified 99% of all beats in the study. Compared to reference data, the MEANS system gave a higher SDNN on average (4.9%), and other HRV parameters were also overestimated (from 6.5% to 29%). On the other hand, the results from MEANS did not substantially change the association of HRV and CVD risk factors. The authors also observed that one or few misclassified N-type heartbeats mean large changes in HRV values; at the same time, many false N-type heartbeats lead to small effects on the HRV. In summary, the authors see the need for further improvements in detecting supraventricular systoles or alternatively an indication of the difficulty for algorithmic analysis stretches of the ECG that must have an additional visual reading.

The critical influence of the QRS-detection accuracy on HRV calculations is analyzed in [27]. The analysis was performed on all RR intervals (for simplification all RR intervals were taken instead of the standard approach with only NN intervals) using all MIT-BIH AD records with various added levels of artificial accuracy error. The time domain, frequency domain, and selected nonlinear HRV parameters were calculated (in total 12 parameters) for each added accuracy error. The results show that, in general, time domain parameters are resistant to accuracy error with HRV parameter error values from 1.3% to 0.02% for accuracy error σ = 8 ms and with the exception of pRR50 with an error of 15% for same accuracy error, whereas frequency parameters exhibited errors from 38% to 55% for σ = 8 ms. The selected nonlinear HRV parameters are from 0.83% to 7.5%. In summary, the authors express the opinion that the results of HRV depend strongly on many conditions: sampling frequency, type of QRS detector used, QRS classification procedure, ECG device model, type of heart disease (as low variability of time series and arrhythmic behavior will influence the total HRV error).

2.5. Analysis of Noise Influence on Detection Accuracy

In [28], the authors analyze the influence of noise on nine different QRS-detection algorithms, which they selected based on low complexity and high performance. Low complexity criteria were that the algorithm could be run on an 8-bit microprocessor, and with high performance criteria, the authors rejected algorithms with poor accuracy performance for input ECG signals with low noise levels. The algorithms were published from 1971 to 1983. The algorithms were as follows: (1) based on amplitude and first derivative: Moriet-Mahoudeaux, Fraden-Neuman, Gustafson, (2) based on first derivative only: Menard, Holsinger, (3) based on first and second derivative: Balda, Ahlstom-Tompkins, (4) based on digital filters: Engelsee-Zeelenberg, Okada. The ECG signal noise sources identified by the authors were as follows: electromyografic, 60 Hz powerline, baseline drift due to respiration, abrupt baseline drift, and composite noise constructed from the mentioned noise sources. The authors created ECG signals with added simulated noise added to the uncorrupted ECG signal at four levels: 25, 50, 75, and 100% of the maximum amplitude. They adopted 88 ms as temporal time tolerance for the TP/FP/FN determination. The authors also measured and reported the changes in detection delay with an increased level of input noise. In summary, the authors stated that noise deteriorated the accuracy of the algorithms and there was no one best algorithm that was superior for all types of noise considered in their study. For electromyografic noise, the best performance was exhibited by algorithms based on the slope and amplitude. For composite noise, the best performance was recorded for algorithms based on digital filters.

The influence of noise on the accuracy of three well-known algorithms was analyzed in [29]: Pan-Tompkins, WQRS, and Hamilton. Algorithm selection criteria were as follows: real time operation and robust performance with input ECG noise. The authors evaluated TP, FN, FP, Se, and PPV on records from MIT-BIH AD with different levels of added noise coming from baseline wander (BA), muscle artifact (MA), and electrode motion (EM). Noise data was from the MIT-BIH Noise Stress Test Database. The level of input ECG signal added noise was from −12 dB to 12 dB with a step of 3 dB. The three algorithms showed similar accuracy results expressed in Se and PPV with an input signal with no added noise. The accuracy deteriorated with added noise and for all 48 MIT-BIH records for SNR = −12dB, the results reported by the authors were respectively for Pan-Tompkins, WQRS, and Hamilton:

  • BW noise: Se: 99.42%, 94.58%, 98.13%; PPV: 97.21%, 62.45%, 95.74%

  • MA noise: Se: 85.94%, 84.71%, 81.74%; PPV: 59.68%, 36.95%, 61.74%

  • EM noise: Se: 68.85%, 65.44%, 84.10%; PPV: 42.54, 33.09%, 44.05%

Is summary, the authors observed that added composite noise at −12 dB significantly degraded all analyzed algorithms’ accuracy expressed in Se and PPV. In detail, the EM had the highest influence on the detection accuracy, followed by MA and BW. The detection temporal tolerance was not specified by the authors, which makes comparisons to other works difficult.

In [30], the authors proposed an R-peak-detection algorithm based on Hilbert transform and evaluated the algorithm accuracy results Se, PP, and DER achieved with ECG signals contaminated with noise. The ECG signals from MIT-BIH NSTD were prepared with five SNR values from 24 dB to 0 dB with a 6 dB step. The authors compared the results with four detectors known from the literature (Pangerec U. et al., Antink C.H. et al., De Cooman T. et al., Vollmer M.) achieving similar or better results. The detection temporal tolerance adopted by the authors was 150 ms. The results of the proposed QRS-detection method show that up to SNR = 6 dB, the influence of noise on the detection accuracy is minimal with Se = 98.13% and PPV = 96.91%, whereas for SNR = 0 dB, the accuracy deteriorates significantly to Se = 78.98% and PPV = 75.25%.

The influence of the noise, detection temporal tolerance (DTT), and QRS morphology on QRS-detection accuracy was analyzed in [13]. The authors analyzed four QRS detectors known from the literature (Gutiérrez-Rivas R. et al., Reklewski W. et al., Ravanshad, N. et al., Pan-Tompkins). To measure the changes in the QRS-detection accuracy, a scaled amount of MA noise (four levels) was added to MIT-BIH AD records, and the influences of six QRS morphologies and five values of DTT were analyzed. The detection accuracy, expressed as a ratio of True Positive/Total Beats (TP/TB), generally deteriorated with added levels of noise and decreasing values of detection temporal tolerance. For one detector (Pan-Tompkins), there was an unexpected improvement with added levels of noise for certain values of DTT. In terms of the influence of the QRS morphology on detection accuracy, the most difficult for accurate QRS detection was V-type (Ventricular beats) for three algorithms and L-type (Left bundle branch block beat) for one algorithm under study. The average (across all DTT and QRS morphologies) detection accuracy results expressed as the True Positive/True Negative ratio (TP/TB) for algorithms under analysis were respectively 83.72% to 82.17%, 90.68% to 89.18%, 77.12% to 71.74%, and 62.03% to 70.32% for levels of noise increasing from SNR = “no noise added” to SNR = 3 dB. As mentioned earlier, for the Pan-Tomkins algorithm, there was an improvement observed in TP/TB with added noise levels from 62.03% to 70.32%.

3. Materials and Methods

In order to analyze the effects of improving and worsening the DER (Equation (1)) with increasing levels of noise and a reduction in the detection tolerance time (DTT) on the QRS detector accuracy we have tested algorithm Alg-4 from [13] (Porr algorithm). The tests are carried out for a range of DTT values and a range of controlled added muscular noise achieved by mixing muscular noise patterns with MIT-BIH AD records (Figure 1).

Figure 1.

Figure 1

Diagram summarizing the workflow of the tests.

First, we reproduced detection accuracy obtained in [13] and expressed it as TP/TB. Next we extended the analysis from [13] with more levels of added noise, expressed as SNR, and conducted the calculations for the same DTT range as in [13] but with larger number of DTT values in order to reduce result graininess. We switched from measuring the QRS-detection accuracy as the TP/TB ratio to DER as it is a more universally employed measure. The TP/TB and DER values here are totals for all 48 MIT-BIH AD records.

To study the influence of an added controlled level of noise to the ECG input signal, we applied a test data set consisting of modified ECG signals from MIT-BIH AD with an added noise signal from MIT-BIH NTSD, a muscle artifact record (MA). We have used all 48 original MIT-BIH AD records, which contain intrinsic noise, and thus, our analysis starts from SNR with an “original” level of noise and not a noise-free ECG record. The added level of SNR has been calculated based on the average power factor of the original record and added noise pattern. The calculation method is described below in the Noise source and mixing section.

The algorithm is tested on the MIT-BIH Arrhythmia Database (MIT-BIH AD) with added muscular noise from the MIT-BIH Noise Stress Test Database (MIT-BIH NSTD) [31]. The tests were conducted on a Dell Latitude E6400, Intel Core2Duo P8400, 2.26 GHz, and 4 GB RAM running with Debian 10.13. Implementation of the algorithms, test tools, and data processing was performed in Python 3.11.2. Plots were created in Jupyter Notebook (server v6.4.12 with Python 3.11.2 [GCC 12.2.0]).

3.1. QRS-Detection Algorithm

Algorithm Alg-4 [13] (Porr algorithm) is an implementation of the Pan Tompkins algorithm. It was selected as a study case because of being the most often cited reference algorithm originating from 1985. The original algorithm preprocessing chain consisting of band pass filtering, differentiation, squaring, and the window moving average is retained as the ‘feature function’. The feature function is further analyzed to find local extrema. When a given local extrema surpasses the threshold value, then the local extrema becomes the algorithm R-peak detection and the algorithm enters a 300 ms waiting state, as it is not possible physiologically for the next R-peak to occur within 200 ms; this implementation extends this waiting state to 300 ms. After the pause, the algorithm resumes the local extrema search and comparisons with the threshold value. The searchback mechanism, a adjustment of the threshold value based on regular and irregular heart rates, is implemented.

3.2. Detection Time Tolerance Range

The range of acceptable tolerance (DTT) tested in this work is presented in Table 1.

Table 1.

Range of DTT values tested.

Approach No. Detector Temporal Tolerance—DTT
[Samples] [ms]
1 3 8.33
2 7 19.44
3 11 30.56
4 15 41.67
5 19 52.78
6 23 63.89
7 27 75.00
8 31 86.11
9 35 97.22
10 39 108.33
11 43 119.45
12 47 130.56
13 51 141.67
14 55 152.78
15 59 163.90

3.3. Noise Source and Mixing

In order to prepare versions of test ECG records with various levels of noise, we have used the upper channel signal from the MIT-BIH AD and added a noise signal from the MIT-BIH NSTD [31] multiplied by adequate scaling factors. For wearable applications and due to the omnipresence of muscle artifacts, we decided to use a “muscle artifact” (MA) record from the MIT BIH NSTD. According to physionet.org, MIT-BIH NSTD is a database related to MIT-BIH AD and includes 12 half-hour ECG recordings and 3 half-hour recordings of noise typical in ambulatory ECG recordings [31]. The three noise records were assembled from the recordings by selecting intervals that contained predominantly baseline wander (in record ‘bw’), muscle (EMG) artifact (in record ‘ma’), and electrode motion artifact (in record ‘em’) [31].

The records from MIT-BIH AD and MIT-BIH NSTD were captured with the same sampling parameters and have the same length. We have used original records from the MIT-BIH AD, and the intrinsic noise is already present in the data and is out of our control. Consequently, mixing additional noise creates a record with additional noise added on top of already existing noise in the particular record. So this will be called the relative signal-to-noise ratio (SNR). The relative SNR can be calculated based on the root mean square (RMS) of the original record RMSs (MIT-BIH AD) and added noise pattern RMSn (MIT-BIH NTSD record MA).

SNR=20·logRMSsRMSn (7)
RMSs=1N·∑i=1N(xsi)2 (8)
RMSn=1N·∑i=1N(xni)2 (9)

where:

  • xsi—a record from MIT-BIH AD, signal values

  • xni—MA record from MIT-BIH NSTD, noise values

  • RMSs—root mean square amplitude of signal

  • RMSn—root mean square amplitude of noise

The levels of noise mixing (SNR) tested in this work are presented in Table 2.

Table 2.

Range of added noise levels tested.

ECG Input Signal Version No. RMSsRMSn 20·log10RMSsRMSn
[-] [db]
1 No noise added No noise added
2 10 20
3 9 19.08
4 8 18.06
5 7 16.90
6 6 15.56
7 5 13.98
8 4 12.04
9 3 9.54
10 2 6.02
11 1.8 5.11
12 1.6 4.08
13 1.4 2.92
14 1.2 1.58
15 1 0
16 0.8 −1.94
17 0.6 −4.44
18 0.4 −7.96

The following procedure has been employed to achieve target relative SNR levels as listed in Table 2. In the first step, the RMSs for one selected record from MIT-BIH AD and RMSn for the MA record from MIT-BIH NSTD are calculated according to Equations (8) and (9). Next, we can rewrite Equation (7) to express the required target RMS as tRMS as in Equation (10). In the next step, the scaling factor m is calculated. And the last step includes preparing the record with the required target SNR value according to Equation (12).

tSNR=20·logRMSsm·RMSn (10)
m=RMSsRMSn·10−tSNR20 (11)

where:

  • m—scaling factor

  • tSNR—target SNR

ECGtSNR=ECG+m·MA (12)

where:

  • ECG—original ECG record from MIT-BIH AD e.g., record 100

  • MA—MA record from MIT-BIH NTSD

The above presented noise-mixing procedure is repeated for all 48 records from the MIT-BIH AD database and for all 18 target noise levels of values specified in Table 2. This produced a total of 864 half-hour test files used to determine the Detection Error Rate that is the main goal of the experiment.

3.4. Individual ECG Record Analysis

Typically, the MIT-BIH AD is used as a collection of typical ECG recordings with specific characteristics representing the true-to-life occurrence of individual heartbeat types. In addition to the global, statistical approach, we also analyzed the impact of individual records on the overall DER value. This allowed us to identify cases particularly sensitive to noise and the tight temporal tolerance of detection points.

4. Results

This section presents the results of DER calculated for the whole collection of 48 MIT-BIH AD files in the full range of the noise level, initially for the full range of DTT and next narrowed to the range where the DTT curve is the steepest representing the fastest drop of detector performance (Section 4.1). Next, we limit the range of considered noise levels and seek a performance drop file-by-file to detect the MIT-BIH AD record where the detector performance drop is the fastest (Section 4.2). Surprisingly we also found files in which the presence of noise improves detection statistics, in particular file 123 (Section 4.3). Finally, we focused on selected ECG strips in this file to find the reason for this unexpected result.

4.1. General Performance of the Detector Under Test

For each DTT and SNR pair, the aggregate detection efficiency (expressed by TB, TP, FN, FP, and DER) is calculated for 48 records from the MIT BIH AD database (Figure 2 and Figure 3).

Figure 2.

Figure 2

Detection efficiency of the Porr algorithm expressed as TP/TB. The results reflect those obtained and described in [13]. The current calculations were performed for a larger number of SNR values (18 values) and DTT values (15 values). In [13], 4 SNR values and 5 DTT values were tested.

Figure 3.

Figure 3

Detection performance results of the Porr algorithm expressed as DER.

Figure 4 displays magnification of Figure 3 within the range of detection temporaltolerance (DTT) of 35 to 31 samples (corresponding to 97.23 ms and 86.12 ms), where the detector’s performance statistics drop dramatically.

Figure 4.

Figure 4

Enlargement of the 35 to 31 sample interval (97.23 ms and 86.12 ms, respectively). For further analysis, the “no added noise” SNR values (blue on the graph) and the noise-added version of the input ECG signal, for added noise up to SNR = 6.02 dB (light blue on the graph), were selected.

The DER results for individual MIT-BIH AD records for DTT = 59, 35, 31, and 3 samples for no noise added and added noise for SNR = 6.02 dB are listed in Table A1, Table A2, Table A3 and Table A4 in the Appendix A.

4.2. Identification of Most Contributing ECG Files

Observing the results in Section 4.1, summarized in Table 3, we asked the following question: which of the 48 MIT-BIH AD records contribute the most to the changes in the total DER shown in Figure 3 and Figure 4? Research for an answer led us to study individual record contributions. For this study, we limited the noise level to two values: the “no added noise” (signal number 1 in Table 2) and the noise-added version of input signal (signal number 10 in Table 2), for added noise up to SNR = 6.02 dB. Results of the study are presented in Figure 5.

Table 3.

Detection performance results of the Porr algorithm expressed as DER for no noise added and added noise for SNR = 6.02 dB. Range of DTT from 3 sample (8.33 ms) to 59 sample (163.90 ms) totals for all 48 MIT-BIH AD records. Total DER pooled over all records (TP, FN, and FP added first then total DER calculated).

DTT
[Samples]
DTT
[ms]
Added Noise [dB] All MIT-BIH
AD Records
TB TP FN FP Se [%] PPV [%] DER [%]
3 8.33 no noise added TOTAL 109494 14630 94864 94957 13.36 13.35 173.36
3 8.33 6.02 TOTAL 109494 21633 87861 93060 19.76 18.86 165.23
7 19.45 no noise added TOTAL 109494 15182 94312 94405 13.87 13.85 172.35
7 19.45 6.02 TOTAL 109494 23118 86376 91575 21.11 20.16 162.52
11 30.56 no noise added TOTAL 109494 15182 94312 94405 13.87 13.85 172.35
11 30.56 6.02 TOTAL 109494 27355 82139 87338 24.98 23.85 154.78
15 41.67 no noise added TOTAL 109494 44112 65382 65475 40.29 40.25 119.51
15 41.67 6.02 TOTAL 109494 58599 50895 56094 53.52 51.09 97.71
19 52.78 no noise added TOTAL 109494 57224 52270 52363 52.26 52.22 95.56
19 52.78 6.02 TOTAL 109494 73178 36316 41515 66.83 63.80 71.08
23 63.89 no noise added TOTAL 109494 60023 49471 49564 54.82 54.77 90.45
23 63.89 6.02 TOTAL 109494 76186 33308 38507 69.58 66.43 65.59
27 75.01 no noise added TOTAL 109494 63067 46427 46520 57.60 57.55 84.89
27 75.01 6.02 TOTAL 109494 79415 30079 35278 72.53 69.24 59.69
31 86.12 no noise added TOTAL 109494 72740 36754 36847 66.43 66.38 67.22
31 86.12 6.02 TOTAL 109494 88007 21487 26686 80.38 76.73 44.00
35 97.23 no noise added TOTAL 109494 105118 4376 4469 96.00 95.92 8.08
35 97.23 6.02 TOTAL 109494 105625 3869 9068 96.47 92.09 11.82
39 108.34 no noise added TOTAL 109494 107949 1545 1638 98.59 98.51 2.91
39 108.34 6.02 TOTAL 109494 107651 1843 7042 98.32 93.86 8.11
43 119.45 no noise added TOTAL 109494 108159 1335 1428 98.78 98.70 2.52
43 119.45 6.02 TOTAL 109494 108015 1479 6678 98.65 94.18 7.45
47 130.57 no noise added TOTAL 109494 108525 969 1062 99.12 99.03 1.85
47 130.57 6.02 TOTAL 109494 108481 1013 6212 99.07 94.58 6.60
51 141.68 no noise added TOTAL 109494 108751 743 836 99.32 99.24 1.44
51 141.68 6.02 TOTAL 109494 108672 822 6021 99.25 94.75 6.25
55 152.79 no noise added TOTAL 109494 108894 600 693 99.45 99.37 1.18
55 152.79 6.02 TOTAL 109494 108820 674 5873 99.38 94.88 5.98
59 163.90 no noise added TOTAL 109494 109002 492 585 99.55 99.47 0.98
59 163.90 6.02 TOTAL 109494 108974 520 5719 99.53 95.01 5.70

Figure 5.

Figure 5

DER broken down into individual MIT-BIH AD records with added noise. Total DER levels for the 48 MIT-BIH AD records are shown as horizontal lines.

Consequently, we indicated records for which the difference in results without adding noise and after adding noise to SNR = 6.02 dB for fixed DTT = 31 and then for fixed DTT = 35 was the most important (Figure 6). Moreover, we indicated records for which the change in DER after moving from DTT = 35 to DTT = 31 at a constant SNR is the most significant (Figure 7). In Figure 7, it can be seen that the biggest differences in DER are for 115, 123, and 109 for both studied SNR levels (no added noise and noise up to 6.02 dB).

Figure 6.

Figure 6

Difference in DER for the signal without noise and the signal after adding noise to SNR = 6.02 dB for DTT = 31 and DTT = 35 of individual MIT-BIH AD records.

Figure 7.

Figure 7

The difference in DER of individual MIT BIH AD records. The red color is the DER difference for DTT = 31 and DTT = 35, both results for the signal without added noise, and the orange color is the analogous difference for the signal with added noise up to the value of SNR = 6.02 dB.

In Figure 6, it can be seen that the input signal with added noise improves the results the most for record 123 (the difference in DER is approximately +85), followed by record 219 (the difference in DER is approximately +70), and then records 115, 109, and 102. At the same time, it can be observed that for two records, the signal with added noise causes a significant deterioration in the DER result: for 121 (the difference in DER is approximately −70) and 117 (approximately −60). A decrease in detector performance with a growing noise level was somewhat expected; however, the opposite case is worth further studies. To this point, record 123 will be investigated to find a plausible reason for the improvement in the DER after adding noise.

4.3. Fiducial Points Instability in File 123

Surprisingly, adding noise may improve the fiducial points stability (i.e., reduce DER). In file 123, it can be observed that for the input ECG signal without added noise, some of the algorithm’s detection points are shifted by approximately 30 samples. Annotations from the MIT BIH AD database for time values on the x-axis are equal to the following:

  • 70 samples (115 − 70 = 45 samples of shift),

  • 1498 samples (1531 − 1498 = 33 samples of shift),

  • 8892 samples (8925 − 8892 = 33 samples of shift),

  • 9326 samples (9359 − 9326 = 33 samples),

  • 9793 samples (9827 − 9793 = 34 samples),

  • 10,273 samples (10,285 − 10,273 = 12 samples),

Figure 8a–d show detailed fiducial point timing in these particular strips of the 123 record.

Figure 8.

Figure 8

Figure 8

(a). Beginning of record 123 with marked detection points of the algorithm for the input signal without added noise and for the added noise for SNR = 6.02 dB and annotations from the MIT BIH AD database. (b). Record 123 strip between samples 8000 and 10,000 with marked detection points of the algorithm for the input signal without added noise and for the added noise for SNR = 6.02 dB and annotations from the MIT BIH AD database. (c). Record 123 strip between samples 10,000 and 12,000 with marked detection points of the algorithm for the input signal without added noise and for the added noise for SNR = 6.02 dB and annotations from the MIT BIH AD database. (d) Record 123 strip between samples 20,000 and 24,000 with marked detection points of the algorithm for the input signal without added noise and for the added noise for SNR = 6.02 dB and annotations from the MIT BIH AD database.

4.4. Internal Operation Analysis

To shed light on the background of this surprising result (Section 4.3, see particularly Figure 8c with a relatively clean ECG), we analyzed, in greater detail, the Porr algorithm (Figure 9) internal operation for two detections related to database annotation at sample 10,273 and 10,713 for the ECG input signal with no added noise and added noise for SNR = 6.02 dB. For the reference beat label at 10,273, algorithm detection for the ECG input signal with no noise added at 10,285 is close to the annotation with a delay of 12 samples, whereas for the reference beat label at the 10,713 algorithm detection for the ECG input signal with no noise added at 10,725, it is close to the annotation with a delay of 12 samples (Figure 8c). These two beats represent the two opposing results and are thus the subject of further analysis. As can be seen in Figure 10, in this excerpt, there are three cases (detections 2, 3, 4) where the Porr algorithm beat labels for the input ECG signal with added noise are closer to the reference beat label than the beat labels for the input ECG signal with no noise added and one case where detection for no added noise is closer to the reference beat label.

Figure 9.

Figure 9

Porr algorithm (Alg-4) flow diagram (based on [13]).

Figure 10.

Figure 10

Detector Porr algorithm selected internal signals and locations of beat labels, time period from sample number 10,200 to 10,400.

Further analysis is presented in Figure 10 and Figure 11, below which an additional signal is presented, the output of the Porr algorithm preprocessing stage—Moving Window Average—mwa[n] signal.

Figure 11.

Figure 11

Detector Porr algorithm selected internal signals and locations of beat labels, time from sample 10,640 to 10,840, amplitude of mwa[n] signal magnified to highlight the shape.

When comparing Figure 10 and Figure 11 it can be noted that in both of these figures, the mwa[n] feature signal (final preprocessing signal before decision stage) looks very similar between the figures and between the two versions of the ECG input signal—the one without added noise (brown line) and the one with added noise (green line). At the same time, the order of algorithm beat labels versus the reference beat label in both figures is different. In Figure 10 the beat label for the input signal without noise is closer to the reference beat label, whereas in Figure 11, the beat label for the input signal with added noise is closer to the reference beat label.

To find out why it is a case, we have analyzed the decision stage in the Porr algorithm. There are two crucial moments:

  • finding the peak in the mwa[n] signal,

  • checking if the mwa[n] signal value at the found peak is greater than the algorithm detection threshold

The above operations are conducted in the below instructions: Instruction (2) is looking for local maxima in the mwa[n] signal; this is performed for all samples in the ECG record, Instruction (1). Next, for each local maxima found called a ‘peak’, the condition is checked if the signal mwa[n] at the peak is greater than threshold_I1 and if the previously found R peak was more than 108 samples (300 ms) earlier, the refractory condition. If that is true, then, the location found is assigned the algorithm beat label (Algorithm 1).

Algorithm 1. Porr algorithm snippet performing peak detection.
for i in range (1, lenght_of_ECG_record):                        (1)
   if mwa[i-1] < mwa[i] and mwa[i + 1]<mwa[i]: # if True -> peak is detected   (2)
        if mwa [i] > threshold_I1 and (peak-signal_peaks[−1]) > 0.3·fs:      (3)
        # second condition checks if previous
        # peak is more than 108 samples
        # earlier (300 ms refractory blanking).
        # if True -> peak is R peak (algorithm
        # label)
# if False -> peak is a noise peak

Analyzing, in detail, Table 4 in connection with Figure 10, it can be noted that the algorithm-detection delay is smaller for the ECG input signal with no added noise as compared to noise added. This situation can also be observed in the Figure 8b database reference at sample number = 8409, the algorithm beat label for no added noise at 8421, and added noise of 6.02 dB at 8442 and in the Figure 8c database reference at 10,273, algorithm beat label for no added noise at 10,285, and added noise of 6.02 dB at 10,307.

Table 4.

Detector Porr algorithm locations of True-True evaluation of Instruction (2) meaning a local peak in the mwa[n] signal, and True-True evaluation of Instruction (3) meaning a feature signal greater than the threshold and more than 300 ms from the last algorithm detection (beat label), with fill colors corresponding to Figure 10, MIT-BIH AD record 123.

Input ECG Signal No Added Noise Input ECG Signal with Added Noise for SNR = 6.02 dB
sample
number i
mwa[i-1]
< mwa[i]


rising slope
mwa[i+1]
< mwa[i]


falling slope
mwa[i]
> threshold_
I1[i]


feature signal
> threshold
(peak-signal_
peaks[-1])
> 0.3·fs

108 samples from last peak (300 ms)
sample
number i
mwa[i-1]
< mwa[i]


rising slope
mwa[i+1]
< mwa[i]


falling
slope
mwa[i]
> threshold_
I1[i]


feature signal
> threshold
(peak-signal_
peaks[-1])
> 0.3·fs

108 samples from last peak (300 ms)
10200 False False False True 10200 False True False True
… …
10201 True True False True 10209 True True False True
10222 True True False True 10222 True True False True
… 10226 True True False True
10238 True True False True 10238 True True False True
10246 True True False True 10246 True True False True
10273 True False False True 10273 True False False True
10285 True True True True …
10307 True True True False 10307 True True True True
10354 True True False False 10353 True True False False
10364 True True False False 10364 True True False False
10371 True True False False 10370 True True False False
10380 True True False False 10385 True True False False
10382 True True False True …
10388 True True False True 10390 True True False False
10400 True False False True 10400 True False False False

Analyzing, in detail, Table 5 in connection with Figure 11, it can be noted that the algorithm-detection delay is smaller for the ECG input signal with added noise as compared to no noise added. This situation can also be observed in the Figure 8b database reference at sample number 8892 and Figure 8c database reference at sample number 10,713, 11,198, and 11,681.

Table 5.

Detector Porr algorithm locations of True-True evaluation of Instruction (2) meaning a local peak in the mwa[n] signal, and True-True evaluation of Instruction (3) meaning a feature signal greater than the threshold and more than 300 ms from the last algorithm detection (beat label), with fill colors corresponding to Figure 11, MIT-BIH AD record 123.

Input ECG Signal No Added Noise Input ECG Signal with Added Noise for SNR = 6.02 dB
sample
number i
mwa[i-1]
< mwa[i]


rising slope
mwa[i+1]
< mwa[i]


falling
slope
mwa[i]
> threshold_
I1[i]


feature signal
> threshold
(peak-signal_
peaks[-1])
> 0.3·fs

108 samples from last peak (300 ms)
sample
number i
mwa[i-1]
< mwa[i]


rising slope
mwa[i+1]
< mwa[i]


falling
slope
mwa[i]
> threshold_ I1[i]


feature signal
> threshold
(peak-signal_
peaks[-1])
> 0.3·fs

108 samples from last peak (300 ms)
10640 False True False True 10640 False False True True
10660 True True False True 10642 True True False True
… 10654 True True False True
… 10660 True True False True
10670 True True False True 10670 True True False True
10681 True True False True 10681 True True False True
… 10685 True True False True
… 10697 True True False True
10713 True False False True 10713 True False False True
10714 True True False True …
… 10725 True True True True
10746 True True True True 10744 True True True False
10758 True True True False 10758 True True True False
10769 True True False False 10769 True True False False
10793 True True False False 10789 True True False False
… 10797 True True False False
… 10807 True True False False
10815 True True False False 10816 True True False False
10822 True True False False 10823 True True False False
10825 True True False False …
10829 True True False False 10828 True True False False
10840 True False False False 10840 True False False False

5. Discussion

A regular QRS detector with large jitter of detection points needs additional procedures to center (i.e., synchronize) the QRS waves e.g., parabola fitting. The purpose is twofold: (1) provide a fiducial point to start time-based calculations of parameters (such as HRV) or amplitude measurement points (such as ST slope) and (2) provide a reference for a cross-correlation-based comparison of signals (for clustering and classification). Two approaches have been presented in the literature: either such centering was performed and precise R-points were calculated at the cost of additional computational load, or the centering was omitted saving machine cycles at the cost of parameters of only approximate quality. Our paper identifies another alternative: having a low complexity low jitter QRS detector robust to noise. In particular, the presence of noise together with low power requirements is typical for wearables; our findings may thus be of particular interest in pursuit for high quality ECG parameters from wearables.

The results obtained show that for almost all SNR values, when the DTT value is reduced from DTT = 97.23 ms to DTT = 86.12 ms (i.e., from DTT = 35 to DTT = 31 samples), there is a sharp drop in performance (i.e., rise of total DER) for all 48 MIT-BIH AD records. This is an interesting novelty in the methodology, because it would allow the detector to be examined as a black box, i.e., without knowledge of its structure. Regarding the detector under test, a delay (inaccuracy of the ‘R’ reading) of 80–100 ms is quite significant. The largest drop in the value of DER was observed for two SNR values: SNR = ‘no added noise’ and SNR = 6.02 dB (ratio of average amplitudes AsignalAnoise=2, i.e., the average amplitude of the useful signal is 2 times greater than the added noise). The amount of delay of this particular algorithm suggests that fiducial points, being its results, should rather be corrected before use in time-related ECG procedures. This statements challenges many papers in which the authors use the QRS-detection series directly for calculations of HRV, heart beat classifications, or estimations of the S-T segment.

In [32], we proposed a multiplierless QRS detector working fast (i.e., with minimum computational burden), precisely (i.e., with excellent detection statistics), and with far less delay (i.e., maintaining good results with less tolerant assumptions). We compared it to several other low-delay QRS detection methods developed recently.

In this paper, we examined the most popular implementation of the most popular QRS-detection algorithm, developed by Pan Tompkins [14], implemented by [15] and available on GitHub. Based on citation statistics, this implementation is considered the most frequently used as a ‘black box’ by the developers of ECG algorithms with a wide range of complexity. To shed more light on the implementation quality, we examined it, applying more true-to-life signal conditions and possibly more thoroughly than anyone previously, using the most accessible database of MIT-BIH AD records [19,20].

  • First, we generated regular statistics of detection performance (incl. Acc, PPV, DER etc.).

  • Next, we examined how detection-performance statistics decrease with increasing timing accuracy requirements by narrowing the range of temporal tolerance (DTT), i.e., allowed time between detection and reference points. We concluded that this algorithm’s results are not directly applicable for time series analyses of ECG, such as HRV, QT, S-T, etc.

  • Further, we examined how increased noise levels affect the stability of the detection point. We found that, as expected, in most cases, higher noise levels lead to greater inaccuracies. However, we also observed situations (files: 123, 219, 115, 109, and 102) where higher noise levels resulted in improved detection statistics.

  • We found several outlying cases and further seek the file in which this surprising situation was most pronounced. For selected ECG segments in file 123, we analyzed the algorithm’s execution step by step. We concluded that the source of the unexpected behavior is the ambiguous implementation of the local maximum definition applied to the feature function.

The observed phenomena of DER improvement based on the ECG input signal with added noise for SNR = 6.02 dB are caused by the local maximum detection mechanism used in the Decision Stage (Figure 9). The local maximum that is a candidate for the R break is searched in the mwa[n] signal. In idealized relationship [33] between the QRS complex and mwa[n] signal, the R peak, occurs approximately in the middle of the rising slope of mwa[n] (Figure 10 and Figure 11) (the time shift noticeable in Figure 10 and Figure 11 is due to group delay of the bandpass filter and a delay in differentiation and moving window average operations, Figure 9). In theory, the Porr algorithm is supposed to detect the first peak as soon as the rising slope ends (Figure 10 and Figure 11). In practice, due to the requirement for the local peak, the preceding and following samples have to have lower values, and considering that the mwa[n] signal is the time-averaged signal over 54 samples (150 ms), this condition for the local peak is often not met around the beginning of the plateau, and the detection of the peak often occurs later during or at the end of the plateau of the mwa[n] signal. This delay has been observed in this work to be about 33 samples (approx. 92 ms), which is a significant number and can be observed in Figure 12. Adding noise for SNR = 6.02 dB reduces the detection jitter by moving a significant number of detections to about 13 samples (approx. 36 ms); this can be observed in the histogram in Figure 12.

Figure 12.

Figure 12

Detection jitter histogram.

Adding noise increases variability of the mwa[n] signal, which often results in earlier detection of the local extrema, reducing the jitter, as has been observed in Figure 11 and Table 5, which in turn translates to improved DER results for a given DTT. Of course, for the tested algorithm, the DER is already poor at DTT = 31 samples (Table A3), but the fact still remains that the DER for the added noise of 6.02 dB is lower than the DER for no added noise for DTT = 31 and lower values of DTT.

Our research points out that Pan Tompkins as the QRS-detector reference has a limitation not described in the existing literature. Current detectors are compared with a flawed reference; then, the comparison is meaningless, and science is not advanced. Despite better results, we do not find a better commonly used reference in the literature. Moreover, in different variants, the Pan Tompkins algorithm is widely used as a front-end black-box procedure by researchers investigating further steps of the ECG-processing chain.

Results of this paper are potentially interesting as an example of detailed testing of the QRS-detector procedure, where most authors only provide selected detection statistics (DER, PPV, Se without DTT, and statistics noise resistance). Both [13] and [17] are consistent in showing that most detectors are converging to 100% accuracy if DTT is large enough (i.e., inaccurate pairing is allowed). Our results may also motivate one to revisit other papers focused on further steps of ECG processing, with a carelessly implemented flawed version of the reference QRS detector (Pan Tompkins) as a black-box ready-to-use procedure.

It is worth a remark that an imprecise definition of the local maximum was already noticed in an early improvement of the Pan Tompkins [14] algorithm, known as the Hamilton Tompkins method [33]. In the algorithm implementation selected in our study, the local peak was searched in the moving average signal (mwa[n]) with a series of if–then conditions and depending on the expected signal properties (sample to sample value changes), which are not met, thus omitting the maximum at its true position. The conclusion from our research is much more general and does not apply only to the Alg. 4 detector. Thorough testing of the detector (and virtually any kind of medical interpretive software) in any imaginable conditions helps in revealing software faults of which the effects are rarely revealed but medically important.

6. Conclusions

Our research clearly indicates that selecting a QRS-detection algorithm based solely on detection statistics may lead to incorrect results. A significant parameter is the change in statistics as the requirements for the temporal accuracy of the detection point change. Algorithms that maintain high true positive detection statistics, even within narrow time tolerances, enable direct application of their results to time-based ECG analyses such as HRV. Algorithms for which statistics deteriorate rapidly require the use of centering procedures to stabilize the QRS-detection point locations. Another important parameter is the robustness of the detection statistics and the temporal stability of the detection point to noise level variations. Long-term electrocardiogram recordings in normal human life are typically accompanied by variable muscle activity and the influence of external electromagnetic fields, so the QRS detector’s immunity to interference fluctuations is highly desirable.

Appendix A

Table A1.

DER Results for individual MIT-BIH AD records, and Total, for DTT = 59 samples (163.90 ms), no added noise, and added noise for SNR = 6.02 dB.

DTT = 59 Samples (163.90 ms) No Added Noise DTT = 59 Samples (163.90 ms) Added Noise for SNR = 6.02 dB
MIT-BIH AD
Record
TB TP FN FP Se [%] PPV [%] DER [%] TP FN FP Se [%] PPV [%] DER [%]
100 2273 2272 1 0 99.96 100 0.04 2271 2 3 99.91 99.87 0.22
101 1865 1863 2 5 99.89 99.73 0.38 1863 2 9 99.89 99.52 0.59
102 2187 2187 0 0 100 100 0 2182 5 69 99.77 96.93 3.38
103 2084 2080 4 1 99.81 99.95 0.24 2079 5 1 99.76 99.95 0.29
104 2229 2225 4 29 99.82 98.71 1.48 2223 6 73 99.73 96.82 3.54
105 2572 2564 8 38 99.69 98.54 1.79 2551 21 114 99.18 95.72 5.25
106 2027 2026 1 1 99.95 99.95 0.1 2022 5 32 99.75 98.44 1.83
107 2137 2133 4 0 99.81 100 0.19 2129 8 127 99.63 94.37 6.32
108 1763 1539 224 408 87.29 79.04 35.85 1698 65 508 96.31 76.97 32.5
109 2532 2529 3 0 99.88 100 0.12 2528 4 9 99.84 99.65 0.51
111 2124 2123 1 1 99.95 99.95 0.09 2117 7 50 99.67 97.69 2.68
112 2539 2539 0 0 100 100 0 2528 11 606 99.57 80.66 24.3
113 1795 1794 1 2 99.94 99.89 0.17 1794 1 1 99.94 99.94 0.11
114 1879 1878 1 6 99.95 99.68 0.37 1874 5 98 99.73 95.03 5.48
115 1953 1953 0 0 100 100 0 1953 0 3 100 99.85 0.15
116 2412 2388 24 2 99 99.92 1.08 2385 27 60 98.88 97.55 3.61
117 1535 1534 1 1 99.93 99.93 0.13 1516 19 905 98.76 62.62 60.2
118 2278 2278 0 0 100 100 0 2276 2 92 99.91 96.11 4.13
119 1987 1987 0 1 100 99.95 0.05 1977 10 324 99.5 85.92 16.81
121 1863 1863 0 0 100 100 0 1843 20 1086 98.93 62.92 59.37
122 2476 2476 0 0 100 100 0 2446 30 303 98.79 88.98 13.45
123 1518 1515 3 0 99.8 100 0.2 1511 7 56 99.54 96.43 4.15
124 1619 1619 0 1 100 99.94 0.06 1607 12 265 99.26 85.84 17.11
200 2601 2598 3 4 99.88 99.85 0.27 2598 3 10 99.88 99.62 0.5
201 1963 1913 50 0 97.45 100 2.55 1912 51 7 97.4 99.64 2.95
202 2136 2130 6 2 99.72 99.91 0.37 2131 5 5 99.77 99.77 0.47
203 2980 2919 61 23 97.95 99.22 2.82 2925 55 95 98.15 96.85 5.03
205 2656 2650 6 1 99.77 99.96 0.26 2648 8 12 99.7 99.55 0.75
207 1860 1857 3 8 99.84 99.57 0.59 1846 14 221 99.25 89.31 12.63
208 2955 2932 23 1 99.22 99.97 0.81 2930 25 10 99.15 99.66 1.18
209 3005 3004 1 1 99.97 99.97 0.07 3005 0 2 100 99.93 0.07
210 2650 2624 26 4 99.02 99.85 1.13 2621 29 9 98.91 99.66 1.43
212 2748 2748 0 0 100 100 0 2748 0 1 100 99.96 0.04
213 3251 3250 1 0 99.97 100 0.03 3250 1 1 99.97 99.97 0.06
214 2262 2258 4 0 99.82 100 0.18 2258 4 11 99.82 99.52 0.66
215 3363 3362 1 0 99.97 100 0.03 3362 1 0 99.97 100 0.03
217 2208 2205 3 3 99.86 99.86 0.27 2198 10 87 99.55 96.19 4.39
219 2154 2154 0 1 100 99.95 0.05 2151 3 64 99.86 97.11 3.11
220 2048 2047 1 1 99.95 99.95 0.1 2046 2 5 99.9 99.76 0.34
221 2427 2420 7 1 99.71 99.96 0.33 2421 6 11 99.75 99.55 0.7
222 2483 2482 1 5 99.96 99.8 0.24 2475 8 82 99.68 96.79 3.62
223 2605 2602 3 1 99.88 99.96 0.15 2601 4 34 99.85 98.71 1.46
228 2053 2049 4 24 99.81 98.84 1.36 2039 14 187 99.32 91.6 9.79
230 2256 2256 0 0 100 100 0 2256 0 0 100 100 0
231 1571 1571 0 0 100 100 0 1571 0 1 100 99.94 0.06
232 1780 1779 1 9 99.94 99.5 0.56 1779 1 67 99.94 96.37 3.82
233 3079 3074 5 0 99.84 100 0.16 3077 2 3 99.94 99.9 0.16
234 2753 2753 0 0 100 100 0 2753 0 0 100 100 0
TOTAL 109494 109002 492 585 99.55 99.47 0.98 108974 520 5719 99.53 95.01 5.70

Table A2.

DER Results for individual MIT-BIH AD records, and Total, for DTT = 35 samples (97.23 ms), no added noise, and added noise for SNR = 6.02 dB.

DTT = 35 Samples (97.23 ms) No Added Noise DTT = 35 Samples (97.23 ms) Added Noise for SNR = 6.02 dB
MIT-BIH AD
Record
TB TP FN FP Se[%] PPV[%] DER[%] TP FN FP Se[%] PPV[%] DER[%]
100 2273 2272 1 0 99.96 100 0.04 2269 4 5 99.82 99.78 0.4
101 1865 1833 32 35 98.28 98.13 3.59 1842 23 30 98.77 98.4 2.84
102 2187 1248 939 939 57.06 57.06 85.87 1688 499 563 77.18 74.99 48.56
103 2084 2080 4 1 99.81 99.95 0.24 2072 12 8 99.42 99.62 0.96
104 2229 1941 288 313 87.08 86.11 26.96 2124 105 172 95.29 92.51 12.43
105 2572 2548 24 54 99.07 97.92 3.03 2482 90 183 96.5 93.13 10.61
106 2027 1931 96 96 95.26 95.26 9.47 1943 84 111 95.86 94.6 9.62
107 2137 2117 20 16 99.06 99.25 1.68 2080 57 176 97.33 92.2 10.9
108 1763 972 791 975 55.13 49.92 100.17 1457 306 749 82.64 66.05 59.84
109 2532 2437 95 92 96.25 96.36 7.39 2413 119 124 95.3 95.11 9.6
111 2124 2116 8 8 99.62 99.62 0.75 2099 25 68 98.82 96.86 4.38
112 2539 2539 0 0 100 100 0 2386 153 748 93.97 76.13 35.49
113 1795 1750 45 46 97.49 97.44 5.07 1742 53 53 97.05 97.05 5.91
114 1879 1878 1 6 99.95 99.68 0.37 1844 35 128 98.14 93.51 8.67
115 1953 1952 1 1 99.95 99.95 0.1 1931 22 25 98.87 98.72 2.41
116 2412 2387 25 3 98.96 99.87 1.16 2362 50 83 97.93 96.61 5.51
117 1535 1441 94 94 93.88 93.88 12.25 1433 102 988 93.36 59.19 71.01
118 2278 2276 2 2 99.91 99.91 0.18 2221 57 147 97.5 93.79 8.96
119 1987 1984 3 4 99.85 99.8 0.35 1909 78 392 96.07 82.96 23.65
121 1863 1860 3 3 99.84 99.84 0.32 1677 186 1252 90.02 57.26 77.19
122 2476 2475 1 1 99.96 99.96 0.08 2358 118 391 95.23 85.78 20.56
123 1518 1514 4 1 99.74 99.93 0.33 1502 16 65 98.95 95.85 5.34
124 1619 1606 13 14 99.2 99.14 1.67 1562 57 310 96.48 83.44 22.67
200 2601 2586 15 16 99.42 99.39 1.19 2588 13 20 99.5 99.23 1.27
201 1963 1765 198 148 89.91 92.26 17.63 1811 152 108 92.26 94.37 13.25
202 2136 2093 43 39 97.99 98.17 3.84 2115 21 21 99.02 99.02 1.97
203 2980 2871 109 71 96.34 97.59 6.04 2879 101 141 96.61 95.33 8.12
205 2656 2639 17 12 99.36 99.55 1.09 2620 36 40 98.64 98.5 2.86
207 1860 1813 47 52 97.47 97.21 5.32 1771 89 296 95.22 85.68 20.7
208 2955 2729 226 204 92.35 93.04 14.55 2772 183 168 93.81 94.29 11.88
209 3005 2972 33 33 98.9 98.9 2.2 2970 35 37 98.84 98.77 2.4
210 2650 2620 30 8 98.87 99.7 1.43 2609 41 21 98.45 99.2 2.34
212 2748 2585 163 163 94.07 94.07 11.86 2627 121 122 95.6 95.56 8.84
213 3251 3219 32 31 99.02 99.05 1.94 3235 16 16 99.51 99.51 0.98
214 2262 2215 47 43 97.92 98.1 3.98 2214 48 55 97.88 97.58 4.55
215 3363 3339 24 23 99.29 99.32 1.4 3327 36 35 98.93 98.96 2.11
217 2208 2117 91 91 95.88 95.88 8.24 2120 88 165 96.01 92.78 11.46
219 2154 2154 0 1 100 99.95 0.05 2139 15 76 99.3 96.57 4.22
220 2048 2038 10 10 99.51 99.51 0.98 2036 12 15 99.41 99.27 1.32
221 2427 2418 9 3 99.63 99.88 0.49 2416 11 16 99.55 99.34 1.11
222 2483 2435 48 52 98.07 97.91 4.03 2427 56 130 97.74 94.92 7.49
223 2605 2596 9 7 99.65 99.73 0.61 2570 35 65 98.66 97.53 3.84
228 2053 2013 40 60 98.05 97.11 4.87 1968 85 258 95.86 88.41 16.71
230 2256 2255 1 1 99.96 99.96 0.09 2242 14 14 99.38 99.38 1.24
231 1571 1570 1 1 99.94 99.94 0.13 1560 11 12 99.3 99.24 1.46
232 1780 1152 628 636 64.72 64.43 71.01 1425 355 421 80.06 77.19 43.6
233 3079 3017 62 57 97.99 98.15 3.86 3039 40 41 98.7 98.67 2.63
234 2753 2750 3 3 99.89 99.89 0.22 2749 4 4 99.85 99.85 0.29
TOTAL 109494 105118 4376 4469 96 95.92 8.08 105625 3869 9068 96.47 92.09 11.82

Table A3.

DER Results for individual MIT-BIH AD records, and Total, for DTT = 31 samples (86.12 ms), no added noise, and added noise for SNR = 6.02 dB.

DTT = 31 Samples (86.12 ms) No Added Noise DTT = 31 Samples (86.12 ms) Added Noise for SNR = 6.02 dB
MIT-BIH AD
Record
TB TP FN FP Se [%] PPV [%] DER [%] TP FN FP Se [%] PPV [%] DER [%]
100 2273 2147 126 125 94.46 94.5 11.04 2160 113 114 95.03 94.99 9.99
101 1865 1408 457 460 75.5 75.37 49.17 1517 348 355 81.34 81.04 37.69
102 2187 766 1421 1421 35.03 35.03 129.95 1445 742 806 66.07 64.19 70.78
103 2084 1344 740 737 64.49 64.58 70.87 1593 491 487 76.44 76.59 46.93
104 2229 1639 590 615 73.53 72.72 54.06 1965 264 331 88.16 85.58 26.69
105 2572 2519 53 83 97.94 96.81 5.29 2418 154 247 94.01 90.73 15.59
106 2027 1810 217 217 89.29 89.29 21.41 1830 197 224 90.28 89.09 20.77
107 2137 1107 1030 1026 51.8 51.9 96.21 1707 430 549 79.88 75.66 45.81
108 1763 899 864 1048 50.99 46.17 108.45 1402 361 804 79.52 63.55 66.08
109 2532 613 1919 1916 24.21 24.24 151.46 1439 1093 1098 56.83 56.72 86.53
111 2124 2061 63 63 97.03 97.03 5.93 2054 70 113 96.7 94.79 8.62
112 2539 1510 1029 1029 59.47 59.47 81.06 2079 460 1055 81.88 66.34 59.67
113 1795 693 1102 1103 38.61 38.59 122.84 1170 625 625 65.18 65.18 69.64
114 1879 1836 43 48 97.71 97.45 4.84 1806 73 166 96.11 91.58 12.72
115 1953 397 1556 1556 20.33 20.33 159.34 1094 859 862 56.02 55.93 88.12
116 2412 1523 889 867 63.14 63.72 72.8 2007 405 438 83.21 82.09 34.95
117 1535 1441 94 94 93.88 93.88 12.25 1410 125 1011 91.86 58.24 74.01
118 2278 1068 1210 1210 46.88 46.88 106.23 1501 777 867 65.89 63.39 72.17
119 1987 1936 51 52 97.43 97.38 5.18 1854 133 447 93.31 80.57 29.19
121 1863 1619 244 244 86.9 86.9 26.19 1525 338 1404 81.86 52.07 93.51
122 2476 2471 5 5 99.8 99.8 0.4 2297 179 452 92.77 83.56 25.48
123 1518 317 1201 1198 20.88 20.92 158.04 987 531 580 65.02 62.99 73.19
124 1619 884 735 736 54.6 54.57 90.86 1228 391 644 75.85 65.6 63.93
200 2601 1885 716 717 72.47 72.44 55.09 2167 434 441 83.31 83.09 33.64
201 1963 1078 885 835 54.92 56.35 87.62 1400 563 519 71.32 72.95 55.12
202 2136 960 1176 1172 44.94 45.03 109.93 1420 716 716 66.48 66.48 67.04
203 2980 2047 933 895 68.69 69.58 61.34 2443 537 577 81.98 80.89 37.38
205 2656 2094 562 557 78.84 78.99 42.13 2325 331 335 87.54 87.41 25.08
207 1860 1123 737 742 60.38 60.21 79.52 1499 361 568 80.59 72.52 49.95
208 2955 2126 829 807 71.95 72.49 55.36 2369 586 571 80.17 80.58 39.15
209 3005 1731 1274 1274 57.6 57.6 84.79 2239 766 768 74.51 74.46 51.05
210 2650 1286 1364 1342 48.53 48.93 102.11 1881 769 749 70.98 71.52 57.28
212 2748 2309 439 439 84.02 84.02 31.95 2437 311 312 88.68 88.65 22.67
213 3251 2103 1148 1147 64.69 64.71 70.59 2774 477 477 85.33 85.33 29.34
214 2262 2139 123 119 94.56 94.73 10.7 2159 103 110 95.45 95.15 9.42
215 3363 1612 1751 1750 47.93 47.95 104.1 2283 1080 1079 67.89 67.91 64.2
217 2208 1431 777 777 64.81 64.81 70.38 1842 366 443 83.42 80.61 36.64
219 2154 787 1367 1368 36.54 36.52 126.97 1611 543 604 74.79 72.73 53.25
220 2048 1165 883 883 56.88 56.88 86.23 1593 455 458 77.78 77.67 44.58
221 2427 1125 1302 1296 46.35 46.47 107.05 1754 673 678 72.27 72.12 55.67
222 2483 2301 182 186 92.67 92.52 14.82 2349 134 208 94.6 91.87 13.77
223 2605 1956 649 647 75.09 75.14 49.75 2165 440 470 83.11 82.16 34.93
228 2053 1973 80 100 96.1 95.18 8.77 1931 122 295 94.06 86.75 20.31
230 2256 1390 866 866 61.61 61.61 76.77 1590 666 666 70.48 70.48 59.04
231 1571 1153 418 418 73.39 73.39 53.21 1229 342 343 78.23 78.18 43.6
232 1780 786 994 1002 44.16 43.96 112.13 1154 626 692 64.83 62.51 74.04
233 3079 2182 897 892 70.87 70.98 58.1 2574 505 506 83.6 83.57 32.84
234 2753 1990 763 763 72.28 72.28 55.43 2331 422 422 84.67 84.67 30.66
TOTAL 109494 72740 36754 36847 66.43 66.38 67.22 88007 21487 26686 80.38 76.73 44.00

Table A4.

DER Results for individual MIT-BIH AD records, and Total, for DTT = 3 samples (8.33 ms), no added noise, and added noise for SNR = 6.02 dB.

DTT = 3 Samples (8.33 ms) No Added Noise DTT = 3 Samples (8.33 ms) Added Noise for SNR = 6.02 dB
MIT-BIH AD
Record
TB TP FN FP Se [%] PPV [%] DER [%] TP FN FP Se [%] PPV [%] DER [%]
100 2273 376 1897 1896 16.54 16.55 166.87 562 1711 1712 24.73 24.71 150.59
101 1865 305 1560 1563 16.35 16.33 167.45 461 1404 1411 24.72 24.63 150.94
102 2187 134 2053 2053 6.13 6.13 187.75 429 1758 1822 19.62 19.06 163.69
103 2084 198 1886 1883 9.5 9.51 180.85 344 1740 1736 16.51 16.54 166.79
104 2229 142 2087 2112 6.37 6.3 188.38 367 1862 1929 16.46 15.98 170.08
105 2572 2188 384 414 85.07 84.09 31.03 1949 623 716 75.78 73.13 52.06
106 2027 401 1626 1626 19.78 19.78 160.43 459 1568 1595 22.64 22.35 156.04
107 2137 407 1730 1726 19.05 19.08 161.72 983 1154 1273 46 43.57 113.57
108 1763 99 1664 1848 5.62 5.08 199.21 223 1540 1983 12.65 10.11 199.83
109 2532 19 2513 2510 0.75 0.75 198.38 62 2470 2475 2.45 2.44 195.3
111 2124 123 2001 2001 5.79 5.79 188.42 289 1835 1878 13.61 13.34 174.81
112 2539 7 2532 2532 0.28 0.28 199.45 155 2384 2979 6.1 4.95 211.22
113 1795 199 1596 1597 11.09 11.08 177.88 318 1477 1477 17.72 17.72 164.57
114 1879 74 1805 1810 3.94 3.93 192.39 155 1724 1817 8.25 7.86 188.45
115 1953 4 1949 1949 0.2 0.2 199.59 113 1840 1843 5.79 5.78 188.58
116 2412 64 2348 2326 2.65 2.68 193.78 343 2069 2102 14.22 14.03 172.93
117 1535 142 1393 1393 9.25 9.25 181.5 292 1243 2129 19.02 12.06 219.67
118 2278 4 2274 2274 0.18 0.18 199.65 31 2247 2337 1.36 1.31 201.23
119 1987 793 1194 1195 39.91 39.89 120.23 864 1123 1437 43.48 37.55 128.84
121 1863 664 1199 1199 35.64 35.64 128.72 334 1529 2595 17.93 11.4 221.36
122 2476 2123 353 353 85.74 85.74 28.51 1603 873 1146 64.74 58.31 81.54
123 1518 7 1511 1508 0.46 0.46 198.88 100 1418 1467 6.59 6.38 190.05
124 1619 262 1357 1358 16.18 16.17 167.7 479 1140 1393 29.59 25.59 156.45
200 2601 173 2428 2429 6.65 6.65 186.74 320 2281 2288 12.3 12.27 175.66
201 1963 30 1933 1883 1.53 1.57 194.4 161 1802 1758 8.2 8.39 181.36
202 2136 12 2124 2120 0.56 0.56 198.69 72 2064 2064 3.37 3.37 193.26
203 2980 164 2816 2778 5.5 5.57 187.72 298 2682 2722 10 9.87 181.34
205 2656 574 2082 2077 21.61 21.65 156.59 791 1865 1869 29.78 29.74 140.59
207 1860 147 1713 1718 7.9 7.88 184.46 491 1369 1576 26.4 23.75 158.33
208 2955 342 2613 2591 11.57 11.66 176.11 489 2466 2451 16.55 16.63 166.4
209 3005 83 2922 2922 2.76 2.76 194.48 244 2761 2763 8.12 8.11 183.83
210 2650 64 2586 2564 2.42 2.44 194.34 367 2283 2263 13.85 13.95 171.55
212 2748 851 1897 1897 30.97 30.97 138.06 1128 1620 1621 41.05 41.03 117.94
213 3251 895 2356 2355 27.53 27.54 144.91 1786 1465 1465 54.94 54.94 90.13
214 2262 43 2219 2215 1.9 1.9 196.02 189 2073 2080 8.36 8.33 183.6
215 3363 18 3345 3344 0.54 0.54 198.9 51 3312 3311 1.52 1.52 196.94
217 2208 28 2180 2180 1.27 1.27 197.46 91 2117 2194 4.12 3.98 195.24
219 2154 76 2078 2079 3.53 3.53 192.99 496 1658 1719 23.03 22.39 156.78
220 2048 8 2040 2040 0.39 0.39 199.22 113 1935 1938 5.52 5.51 189.11
221 2427 45 2382 2376 1.85 1.86 196.04 251 2176 2181 10.34 10.32 179.52
222 2483 622 1861 1865 25.05 25.01 150.06 708 1775 1849 28.51 27.69 145.95
223 2605 16 2589 2587 0.61 0.61 198.69 103 2502 2532 3.95 3.91 193.24
228 2053 385 1668 1688 18.75 18.57 163.47 624 1429 1602 30.39 28.03 147.64
230 2256 0 2256 2256 0 0 200 3 2253 2253 0.13 0.13 199.73
231 1571 365 1206 1206 23.23 23.23 153.53 420 1151 1152 26.73 26.72 146.59
232 1780 12 1768 1776 0.67 0.67 199.1 51 1729 1795 2.87 2.76 197.98
233 3079 234 2845 2840 7.6 7.61 184.64 552 2527 2528 17.93 17.92 164.18
234 2753 708 2045 2045 25.72 25.72 148.57 919 1834 1834 33.38 33.38 133.24
TOTAL 109494 14630 94864 94957 13.36 13.35 173.36 21633 87861 93060 19.76 18.86 165.23

Author Contributions

Conceptualization and validation: W.R. and P.A.; analysis, software, investigation, and writing: W.R.; supervision: P.A. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.

Conflicts of Interest

The authors declare no conflicts of interest.

Funding Statement

This research was funded by AGH University of Krakow in 2025, as research project No. 16.16.120.773.

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.Sörnmo L., Laguna P. Bioelectrical Signal Processing in Cardiac and Neurological Applications. Academic Press; New York, NY, USA: 2005. [Google Scholar]
  • 2.Kohler B.-U., Hennig C., Orglmeister R. The principles of software QRS detection. IEEE Eng. Med. Biol. Mag. 2002;21:42–57. doi: 10.1109/51.993193. [DOI] [PubMed] [Google Scholar]
  • 3.Elgendi M., Eskofier B., Dokos S., Abbott D. Revisiting QRS detection methodologies for portable, wearable, battery-operated and wireless ECG systems. PLoS ONE. 2014;9:e84018. doi: 10.1371/journal.pone.0084018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Wang R., Veera S.C.M., Asan O., Liao T. A Systematic Review on the Use of Consumer-Based ECG Wearables on Cardiac Health Monitoring. IEEE J. Biomed. Health Inform. 2024;28:6525–6537. doi: 10.1109/JBHI.2024.3456028. [DOI] [PubMed] [Google Scholar]
  • 5.Berkaya S.K., Uysal A.K., Gunal E.S., Ergin S., Gunal S., Gulmezoglu M.B. A survey on ECG analysis. Biomed. Signal Process. Control. 2018;43:216–235. doi: 10.1016/j.bspc.2018.03.003. [DOI] [Google Scholar]
  • 6.Luz E.J.D.S., Schwartz W.R., Cámara-Chávez G., Menotti D. ECG-based heartbeat classification for arrhythmia detection: A survey. Comput. Methods Programs Biomed. 2016;127:144–164. doi: 10.1016/j.cmpb.2015.12.008. [DOI] [PubMed] [Google Scholar]
  • 7.Boulif A., Ananou B., Ouladsine M., Delliaux S. A literature review: ECG-based models for arrhythmia di-agnosis using artificial intelligence techniques. Bioinform. Biol. Insights. 2023;17:11779322221149600. doi: 10.1177/11779322221149600. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Gour A., Gupta M., Wadhvani R., Shukla S. Ecg based heart disease classification: Advancement and review of techniques. Procedia Comput. Sci. 2024;235:1634–1648. doi: 10.1016/j.procs.2024.04.155. [DOI] [Google Scholar]
  • 9.Clifford G.D., Azuaje F. In: Advanced Methods and Tools for ECG Data Analysis. McSharry P., editor. Volume 10 Artech House; Boston, MA, USA: 2006. [Google Scholar]
  • 10.Gupta R., Mitra M., Bera J. ECG Acquisition and Automated Remote Processing. Volume 20 Springer; New Delhi, India: 2014. [Google Scholar]
  • 11.Augustyniak P. Recovering the precise heart rate from sparsely sampled electrocardiograms. Comput. Med. 1999;59:23–25. [Google Scholar]
  • 12.Heryan K., Reklewski W., Szaflarski A., Ordowski M., Augustyniak P., Miśkowicz M. Sensitivity of QRS detection accuracy to detector temporal resolution; Proceedings of the 2021 Computing in Cardiology (CinC); Brno, Czech Republic. 13–15 September 2021; pp. 1–4. [Google Scholar]
  • 13.Reklewski W., Miśkowicz M., Augustyniak P. QRS Detector Performance Evaluation Aware of Temporal Accuracy and Presence of Noise. Sensors. 2024;24:1698. doi: 10.3390/s24051698. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Pan J., Tompkins W.J. A real-time QRS detection algorithm. IEEE Trans. Biomed. Eng. 1985;3:230–236. doi: 10.1109/TBME.1985.325532. [DOI] [PubMed] [Google Scholar]
  • 15.Howell L., Porr B. A Collection of 8 ECG Heartbeat Detection Algorithms Implemented in Python. Developed in Conjunction with a New ECG Database. 2019–2023. [(accessed on 20 August 2025)]. Available online: https://github.com/berndporr/py-ecg-detectors/blob/master/ecgdetectors.py.
  • 16.Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms. AAMI Recommended Practice/American National Standard; Washington, DC, USA: 1998. [(accessed on 20 August 2025)]. Available online: http://www.aami.org. [Google Scholar]
  • 17.Porr B., Macfarlane P.W. A new QRS detector stress test combining temporal jitter and F-score (JF) reveals significant performance differences amongst popular detectors. PLoS ONE. 2024;19:e0309739. doi: 10.1371/journal.pone.0309739. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Goldberger A.L., Amaral L.A., Glass L., Hausdorff J.M., Ivanov P.C., Mark R.G., Mietus J.E., Moody G.B., Peng C.K., Stanley H.E. PhysioBank, PhysioToolkit, and PhysioNet: Components of a new research resource for complex physiologic signals. Circulation. 2000;101:e215–e220. doi: 10.1161/01.CIR.101.23.e215. [DOI] [PubMed] [Google Scholar]
  • 19.Moody G.B., Mark R.G. The impact of the MIT-BIH Arrhythmia Data-base. [(accessed on 20 August 2025)];IEEE Eng. Med. Biol. Mag. 2001 20:45–50. doi: 10.1109/51.932724. Available online: https://physionet.org/content/mitdb/1.0.0/ [DOI] [PubMed] [Google Scholar]
  • 20.Tripathi P.M., Kumar A., Komaragiri R., Kumar M. A review on computational methods for denoising and detecting ECG signals to detect cardiovascular diseases. Arch. Comput. Methods Eng. 2022;29:1875–1914. doi: 10.1007/s11831-021-09642-2. [DOI] [Google Scholar]
  • 21.Chatterjee S., Thakur R.S., Yadav R.N., Gupta L., Raghuvanshi D.K. Review of noise removal techniques in ECG signals. IET Signal Process. 2020;14:569–590. doi: 10.1049/iet-spr.2020.0104. [DOI] [Google Scholar]
  • 22.Krajewski Z., Tkacz E. Power Noise Removal for Monitoring of Biomedical Signals in Daily Life Applications. J. Med. Imaging Health Inform. 2012;2:44–48. doi: 10.1166/jmihi.2012.1063. [DOI] [Google Scholar]
  • 23.Gradl S., Leutheuser H., Elgendi M., Lang N., Eskofier B.M. Temporal correction of detected R-peaks in ECG signals: A crucial step to improve QRS detection algorithms; Proceedings of the 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Milan, Italy. 25–29 August 2015; pp. 522–525. [DOI] [PubMed] [Google Scholar]
  • 24.Weinstein A., Rodiño J., Otero M. Effects of Electrocardiograms QRS Detection Algorithms in Heart Rate Variability Metrics. bioRxiv. 2025 doi: 10.1101/2025.04.23.650299. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Howell L., Porr B. High Precision ECG Database with Annotated R Peaks, Recorded and Filmed Under Realistic Conditions. University of Glasgow 2018. [(accessed on 20 August 2025)]. Available online: https://researchdata.gla.ac.uk/id/eprint/716.
  • 26.Sauerbier F., Haerting J., Sedding D., Mikolajczyk R., Werdan K., Nuding S., Kluttig A. Impact of QRS misclassifications on heart-rate-variability parameters (results from the CARLA cohort study) PLoS ONE. 2024;19:e0304893. doi: 10.1371/journal.pone.0304893. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Petelczyc M., Gierałtowski J.J., Żogała-Siudem B., Siudem G. Impact of observational error on heart rate variability analysis. Heliyon. 2020;6:e03984. doi: 10.1016/j.heliyon.2020.e03984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Friesen G.M., Jannett T.C., Jadallah M.A., Yates S.L., Quint S.R., Nagle H.T. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 1990;37:85–98. doi: 10.1109/10.43620. [DOI] [PubMed] [Google Scholar]
  • 29.Mohd Apandi Z.F., Ikeura R., Hayakawa S., Tsutsumi S. An analysis of the effects of noisy electrocardiogram signal on heartbeat detection performance. Bioengineering. 2020;7:53. doi: 10.3390/bioengineering7020053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.D’Aloia M., Longo A., Rizzi M. Noisy ECG signal analysis for automatic peak detection. Information. 2019;10:35. doi: 10.3390/info10020035. [DOI] [Google Scholar]
  • 31.Moody G., Muldrow W.E., Mark R. The MIT-BIH noise stress test database. Comput. Cardiol. 1984;11:381–384. [Google Scholar]
  • 32.Reklewski W., Heryan K., Augustyniak P., Miśkowicz M. Multiplierless QRS detection algorithm for mobile ECG monitoring based on approximate computing. Measurement. 2025;249:116969. doi: 10.1016/j.measurement.2025.116969. [DOI] [Google Scholar]
  • 33.Hamilton P.S., Tompkins W.J. Quantitative investigation of QRS detection rules using the MIT/BIH Arrhythmia database. IEEE Trans. Biomed. Eng. 1986;BME-33:1157–1165. doi: 10.1109/TBME.1986.325695. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.